Electronic Brachytherapy Sources. Thomas W. Rusch

Size: px
Start display at page:

Download "Electronic Brachytherapy Sources. Thomas W. Rusch"

Transcription

1 Electronic Brachytherapy Sources Thomas W. Rusch

2 Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration Understand methods to characterize and confirm source spatial distribution

3 Source Construction & Operation

4 Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration Understand methods to characterize and confirm source spatial distribution

5 Axxent X-Ray Source 50 kv operating potential High vacuum x-ray tube technology Output: ~10 Gy/min 1cm into tissue Water cooled Fully disposable device Output described in TG-43 format HV connection Cooling connections X-Ray Tube HV Cable X-Ray Source Tip Detail

6 Axxent X-Ray Source Components Cable Anode Envelope Filament

7 Axxent Controller Features Handheld barcode scanner for data entry Display Screen/ Touch Screen Control Retractable Power Cord Wheel brakes for position stability during treatment

8 Axxent Controller Features High voltage power supply and connector Emergency stop button Well Chamber for source constancy checks Adjustable arm for source pullback (in storage position)

9 Axxent Applicators Balloon Applicators Vaginal Cylinders Skin-Surface Applicators 3-4 cm Spherical 2.0 cm diameter 1.0 cm diameter 4-5 cm Spherical 2.5 cm diameter 2.0 cm diameter 5-6 cm Spherical 3.0 cm diameter 3.5 cm diameter 5x7 cm Ellipsoidal 3.5 cm diameter 5.0 cm diameter 6x7 cm Ellipsoidal

10 Educational Objective #1 Understand key elements of ebx source construction & operation Diode x-ray source operates at 50 kv and 0.30 ma Water-cooled in a 5.6 mm diameter catheter Dose distribution is described in the TG-43 format Portable controller operates and positions the source FDA has cleared the source and controller for use anywhere in or on the body where radiation treatment is indicated Applicator choices: balloons for accelerated partial breast irradiation, vaginal cylinders, cones for skin or internal organ surfaces

11 Air Kerma Strength Characterization

12 Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration Understand methods to characterize and confirm source spatial distribution

13 2-D Brachytherapy Dosimetry Formalism, AAPM TG-43U1 (2004) D(r,θ) = S K Λ g L (r) G L (r,θ) F(r,θ) G L (r 0,θ 0 ) D(r,θ) S K Λ g L (r) G L (r,θ) F(r,θ) dose rate to water at point P(r,θ) air kerma strength dose rate constant radial dose function geometry function (line source approximation) 2-D anisotropy function

14 Air Kerma Standards National primary standards for air kerma are based on either free-air chambers or cavity ionization chambers Free-air chambers: < 300 kev photons e.g. primary x-ray standards, low-energy brachytherapy sources Cavity ionization chambers: > 300 kev photons e.g. 137 Cs (662 kev) and 60 Co (1250 kev) HDR 192 Ir calibrations are based on an ionization chamber calibrated using 250 kv x-rays and 137 Cs 3/24

15 Calibration Traceability - Most Low Energy Sources Primary standard (NIST Wide-Angle Free-Air Chamber) Secondary standard (ADCL well ionization chamber) Clinical well ionization chamber

16 Interim Calibration Traceability UW Attix free-air chamber Primary standard (NIST free-air chamber) Secondary standard (ADCL well ionization chamber) Clinical well ionization chamber

17 Comparison to 125 I Well chamber was also calibrated at UW using a I seed for interim NIST traceability Conversion coefficients between ebx source and 125 I well chamber calibrations are: Voltage Ratio to 125 I 50 kv /- 3.0% 45 kv /- 2.7% 40 kv /- 4.7% Reference: Application Note 06-01, University of Wisconsin Radiation Calibration Laboratory 21/24

18 HDR 1000 Plus Well Chamber Special aluminum insert designed by Standard Imaging Source strength is measured with the well chamber prior to treatment

19 Well Chamber Integrated into the Controller Source in well chamber for calibration Well chamber connected internally to electrometer

20 Educational Objective #2 Understand the rationale and methods for air kerma strength calibration An interim well chamber calibration is being used while a national standard is developed at NIST The interim method was established at the University of Wisconsin Medical Radiation Research Lab Attix free air chamber and a I seed An integral well chamber is used to check source strength constancy prior to each fraction

21 Spatial Distribution Characterization

22 Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration Understand methods to characterize and confirm source spatial distribution

23 Axxent Brachytherapy Dosimetry Formalism D(r,θ) = S K Λ g P (r) r 02 r 2 F(r,θ) Simplified geometry function due to L < 1 mm Published g P (r) and F(r,θ) (Rivard et al 2006) All sources are tested for agreement with accepted parameter set Uncertainties are added in quadrature

24 Motivation - it s got to be fast and it s got to be right! To assure an accurate delivery of radiation, must match TG-43 protocol parameters to within an overall error budget Depth dose distribution within 0.03 of TG-43 reference Polar distribution within 0.10 of TG-43 reference Azimuthal asymmetry 7% Dose rate constant within 7% of nominal Output level and stability Using a HDR well chamber 0 90 High throughput An apparatus was designed, built and validated to accomplish these goals

25 Manufacturing Test Fixture Water tank Shielded enclosure with leaded acrylic window Carousel assembly holding all detectors PTW ion chamber Photodiode devices Critical fixtures are Solid Water TM Shielded bridge and diode amplifiers on top plate Source to be tested is fixed, carousel assembly rotates Bridge Assembly Top Plate (raised) Carousel Assembly Water Tank Assembly Source

26 Sample printed output

27 In the Clinic: Shielded Test Fixture Provides a safe enclosure to evaluate source performance Attaches to controller pullback arm like an applicator Can accept a small pinpoint-style ion chamber Source stability Quantitative pullback comparisons Can accept 12.5 x 12.5 cm sheets of radiochromic film Source spatial distribution in one plane Pullback accuracy

28 Pretreatment QA: Ion Chamber and Film Completely shielded Ion chamber port and radio-opaque position markers

29 Educational Objective #3 Understand methods to characterize and confirm source spatial distribution Prior to certification for human use, each source is evaluated in manufacturing test fixtures and must perform within prescribed limits The shielded test fixture allows measurement of relative source spatial characteristics using radiochromic film or an ion chamber

30 Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration Understand methods to characterize and confirm source spatial distribution

31 Acknowledgements Stephen Davis and Prof. Larry DeWerd, University of Wisconsin Medical Radiation Research Lab Linda Kelley and Steve Axelrod, Xoft, Inc. 24/24

Cylindrical Ion Chambers Victoreen Model 550 Series

Cylindrical Ion Chambers Victoreen Model 550 Series Cylindrical Ion Chambers Victoreen Model 550 Series! Cylindrical Ion Chambers for use with Model 35040 and Model 530 electrometers! Wide range of applications in Diagnostic X-Ray and Radiation Oncology

More information

Cylindrical Ion Chambers

Cylindrical Ion Chambers Cylindrical Ion Chambers Radiation Oncology ON Victoreen Model 550T Series Cylindrical Ion Chambers for use with Model 35040 and Model 560 electrometers Wide range of applications in Diagnostic X-Ray and

More information

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Indra J. Das, PhD, FACR Department of Radiation Oncology Indiana University of School of Medicine & Midwest Proton Radiation

More information

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This

More information

Progress Report from the HDR Brachytherapy Working Party. Margaret Bidmead

Progress Report from the HDR Brachytherapy Working Party. Margaret Bidmead Progress Report from the HDR Brachytherapy Working Party Margaret Bidmead HDR Ir-192 calibration service provided by NPL since May 2004 Primary standard irradiated with a Nucletron classic HDR IR-192 source

More information

Calibration of KAP meters

Calibration of KAP meters Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area

More information

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film With improvements in targeting and localization, stereotactic treatments have become prevalent. To

More information

DETECTORS. Including Codes of Practice IONIZING RADIATION

DETECTORS. Including Codes of Practice IONIZING RADIATION DETECTORS Including Codes of Practice 2015 IONIZING RADIATION PTW History and General Remarks Looking back on a long history PTW-Freiburg is an internationally operating company, manufacturing and marketing

More information

DETECTORS UNCOMPROMISING QUALITY. The standard in dosimetry measurements for over 40 years. EXRADIN DETECTORS

DETECTORS UNCOMPROMISING QUALITY. The standard in dosimetry measurements for over 40 years. EXRADIN DETECTORS DETECTORS UNCOMPROMISING QUALITY The standard in dosimetry measurements for over 40 years. EXRADIN DETECTORS The Exradin Advantage Better Components Waterproof construction eliminates the need for sleeves

More information

Preliminary studies of a new monitor ionization chamber

Preliminary studies of a new monitor ionization chamber 1 Preliminary studies of a new monitor ionization chamber Maíra T. Yoshizumi, Vitor Vivolo and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN) Comissão Nacional de Energia

More information

Response characteristics of a tandem ionization chamber in standard X-ray beams

Response characteristics of a tandem ionization chamber in standard X-ray beams Applied Radiation and Isotopes 58 (2003) 495 500 Response characteristics of a tandem ionization chamber in standard X-ray beams Alessandro M. Costa*, Linda V.E. Caldas Instituto de Pesquisas Energ!eticas

More information

When small things matter. Small Field Dosimetry Application Guide

When small things matter. Small Field Dosimetry Application Guide R A D I AT I O N T H E R A P Y When small things matter. Small Field Dosimetry Application Guide Contents 1 Introduction 1 Introduction 2 2 The Physics of Small Fields 3 3 Detector Types 10 4 Detector

More information

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

3D Diode Array Commissioning: Building Confidence in 3D QA Technology 3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose

More information

IONIZING RADIATION 2018 / 2019 DETECTORS. Including Codes of Practice

IONIZING RADIATION 2018 / 2019 DETECTORS. Including Codes of Practice IONIZING RADIATION DETECTORS 2018 / 2019 Including Codes of Practice PTW History and General Remarks Looking back on a long history PTW-Freiburg is an internationally operating company, manufacturing and

More information

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array José A. Bencomo, * Geoffrey Ibbott, Seungsoo Lee, and Joao A. Borges Department of Radiation Physics.

More information

MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM *

MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM * MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM * C. Kessler and D.T. Burns December 2018 BUREAU INTERNATIONAL DES POIDS ET MESURES

More information

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage

More information

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion Leksell Gamma Knife Perfexion QA Considerations Paula L. Petti, Ph.D. Taylor McAdam Bell Neuroscience Institute Washington Hospital Healthcare System Fremont, CA 1 QA for LGK Perfexion : : Follow NRC licensing

More information

Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector

Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector ORIGINAL ARTICLES Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector Gopiraj ANNAMALAI 1, Ramasubramanian VELAYUDHAM 2 ABSTRACT Received: 7.07.2009 Accepted: 2.11.2009

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

Abstract. This investigation was supported by PHS grant CA10953 awarded by the NCI, DHHS. Introduction

Abstract. This investigation was supported by PHS grant CA10953 awarded by the NCI, DHHS. Introduction Practical Considerations for the Calibration of Low Energy/Low Activity Seeds Peter A. Balter, J. Francisco Aguirre, and William F. Hanson Department of Radiation Physics The University of Texas M. D.

More information

A Sort-of Tissue Equivalent Proportional Counter (STEPC) for Space Radiation Dosimetry Applications

A Sort-of Tissue Equivalent Proportional Counter (STEPC) for Space Radiation Dosimetry Applications A Sort-of Tissue Equivalent Proportional Counter (STEPC) for Space Radiation Dosimetry Applications Eric Benton, Tyler Collums, and Art Lucas E. V. Benton Radiation Physics Laboratory Oklahoma State University

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 David F. Lewis Advanced Materials Group Ashland Specialty Ingredients Spain, April 2014 What is Radiochromic Film? A film that instantly changes

More information

Peace of Mind. Automated.

Peace of Mind. Automated. 1 Peace of Mind. Automated. Automated and guided beam commissioning Why SMARTSCAN? Get thousands of beam scans done effortlessly. Save your valuable time and manual operations at the water phantom and

More information

SCINTILLATING FIBER DOSIMETER ARRAY

SCINTILLATING FIBER DOSIMETER ARRAY SCINTILLATING FIBER DOSIMETER ARRAY FIELD OF THE INVENTION [0001] This invention relates generally to the field of dosimetry and, more particularly, to rapid, high-resolution dosimeters for advanced treatment

More information

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film As clinics strive to treat with the smallest field possible, the demand for SRS/SBRT QA grows as well.

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

RADIATION THERAPY DETECTORS. For Small Field Dosimetry

RADIATION THERAPY DETECTORS. For Small Field Dosimetry RADIATION THERAPY DETECTORS For Small Field Dosimetry 0.125 cm 3 Semiflex Chamber Type 31010 Standard therapy chamber for scanning systems and for absolute dosimetry Waterproof, semiflexible design for

More information

PHYSICS QUESTIONNAIRE FORM

PHYSICS QUESTIONNAIRE FORM PHYSICS QUESTIONNAIRE FORM Institution Name: Date: Contact Information (name, address, phone, fax, email): Physicist: Radiation Oncologist: Dosimetrist (if applicable): Study Coordinator (if applicable):

More information

X-ray tube with needle-like anode

X-ray tube with needle-like anode NUKLEONIKA 2002;47(3):101 105 ORIGINAL PAPER X-ray tube with needle-like anode Mieczys aw S apa, W odzimierz StraÊ, Marek Traczyk, Jerzy Dora, Miros aw Snopek, Ryszard Gutowski, Wojciech Drabik Abstract

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 1, WINTER 2010 A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA Abdul Qadir Jangda,

More information

Introduction of a Single Chip TLD System for Patient Dosimetry

Introduction of a Single Chip TLD System for Patient Dosimetry Introduction of a Single Chip TLD System for Patient Dosimetry C. Hranitzky a, M. Halda a, G. Müller a, B. Obryk b, H. Stadtmann a* a Austrian Research Centers GmbH ARC, 2444 Seibersdorf, Austria. b Institute

More information

Experiment 6: Franck Hertz Experiment v1.3

Experiment 6: Franck Hertz Experiment v1.3 Experiment 6: Franck Hertz Experiment v1.3 Background This series of experiments demonstrates the energy quantization of atoms. The concept was first implemented by James Franck and Gustaf Ludwig Hertz

More information

LI-193 Spherical Quantum Sensor

LI-193 Spherical Quantum Sensor LI-193 Spherical Quantum Sensor The LI-193 Spherical Quantum Sensor measures PAR in air or underwater from all directions at depths up to 350 meters. This sensor is useful for studies of phytoplankton,

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

Rotational total skin electron irradiation with a linear accelerator

Rotational total skin electron irradiation with a linear accelerator JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 4, FALL 2008 Rotational total skin electron irradiation with a linear accelerator Eric P. Reynard, 1,a Michael D.C. Evans, 1 Slobodan Devic,

More information

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision Minnesota Rules, Chapter 4732 X-ray Revision DRAFT INDUSTRIAL X-RAY SYSTEMS DEFINTIONS, 1.0 4732.####. INDUSTRIAL X-RAY SYSTEMS DEFINITIONS. Subpart 1. Scope. For purposes of industrial x-ray systems under

More information

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017 The Current State of EPID-Based Linear Accelerator Quality Assurance Timothy Ritter, PhD, DABR, FAAPM 1 Disclosures Employed by the Veterans Health Administration Faculty appointment with the University

More information

LI-192 Underwater Quantum Sensor

LI-192 Underwater Quantum Sensor LI-192 Underwater Quantum Sensor The LI-192 Underwater Quantum Sensor measures PAR from all angles in one hemisphere. The LI-192 works in air or underwater at depths up to 560 meters. The measurements

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced

More information

PD233: Design of Biomedical Devices and Systems

PD233: Design of Biomedical Devices and Systems PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/

More information

SPECTROMETRIC DETECTION PROBE Model 310. Operator's manual

SPECTROMETRIC DETECTION PROBE Model 310. Operator's manual SPECTROMETRIC DETECTION PROBE Model 310 Operator's manual CONTENTS 1. INTRODUCTION... 3 2. SPECIFICATIONS... 4 3. DESIGN FEATURES... 6 4. INSTALLATION... 10 5. SAFETY AND PRECAUTIONS... 13 6. THEORY OF

More information

The on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO)

The on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO) The on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO) A. Ansarinejad1,2, A. Attili1, F. Bourhaleb2,R. Cirio1,2,M. Donetti1,3, M. A. Garella1, S. Giordanengo1,

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

The TSIS Spectral Irradiance Monitor: Prism Optical Degradation Studies

The TSIS Spectral Irradiance Monitor: Prism Optical Degradation Studies The TSIS Spectral Irradiance Monitor: Prism Optical Degradation Studies Lo Erik Richard, Dave Harber, Joel Rutkowski, Matt Triplett, Kasandra O Malia Laboratory for Atmospheric and Space Physics (LASP)

More information

OPERATION MANUAL. Model LET-SW5. April Far West Technology, Inc. 330 South Kellogg Ave, Suite D Goleta, CA 93117

OPERATION MANUAL. Model LET-SW5. April Far West Technology, Inc. 330 South Kellogg Ave, Suite D Goleta, CA 93117 OPERATION MANUAL Model LET-SW5 April 2010 Far West Technology, Inc. 330 South Kellogg Ave, Suite D Goleta, CA 93117 GENERAL INFORMATION This instrument is manufactured in the United States of America by:

More information

Total Ionizing Dose Test Report. Z-Series DC-DC Converter

Total Ionizing Dose Test Report. Z-Series DC-DC Converter Total Ionizing Dose Test Report Z-Series DC-DC Converter Revision A March, 2004 TOTAL DOSE TEST REPORT for Z - SERIES DC/DC CONVERTER Project Engineer: Engineering Director: Tom Hanson Peter Lee TABLE

More information

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System. Use of EPIDs for Non-Routine Linac QA Bin Cai PhD Disclosure Parts of this project received support from Varian Medical System. Learning Objectives Learn the recent development of EPID based Non-routine

More information

Summer Student project report

Summer Student project report Summer Student project report Mika Väänänen September 1, 2017 Abstract In this report I give a brief overview of my activities during the summer student project. I worked on the scintillating fibre (SciFi)

More information

200 MHz 350 MHz 750 MHz Linac2 RFQ2 202 MHz 0.5 MeV /m Weight : 1000 kg/m Ext. diameter : 45 cm

200 MHz 350 MHz 750 MHz Linac2 RFQ2 202 MHz 0.5 MeV /m Weight : 1000 kg/m Ext. diameter : 45 cm M. Vretenar, CERN for the HF-RFQ Working Group (V.A. Dimov, M. Garlasché, A. Grudiev, B. Koubek, A.M. Lombardi, S. Mathot, D. Mazur, E. Montesinos, M. Timmins, M. Vretenar) 1 1988-92 Linac2 RFQ2 202 MHz

More information

PHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002

PHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002 PHYSICS 334 - ADVANCED LABORATORY I COMPTON SCATTERING Spring 00 Purposes: Demonstrate the phenomena associated with Compton scattering and the Klein-Nishina formula. Determine the mass of the electron.

More information

Distributed source x-ray tube technology for tomosynthesis imaging

Distributed source x-ray tube technology for tomosynthesis imaging Distributed source x-ray tube technology for tomosynthesis imaging Authors: F. Sprenger a*, X. Calderon-Colon b, Y. Cheng a, K. Englestad a, J. Lu b, J. Maltz c, A. Paidi c, X. Qian b, D. Spronk a, S.

More information

ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES

ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES Romanian Reports in Physics, Vol. 65, No. 1, P. 168 177, 213 ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES F. SCARLAT 1, A. SCARISOREANU

More information

LBIR Fluid Bath Blackbody for Cryogenic Vacuum Calibrations

LBIR Fluid Bath Blackbody for Cryogenic Vacuum Calibrations LBIR Fluid Bath Blackbody for Cryogenic Vacuum Calibrations Timothy M. Jung*, Adriaan C. Carter*, Dale R. Sears*, Solomon I. Woods #, Dana R. Defibaugh #, Simon G. Kaplan #, Jinan Zeng * Jung Research

More information

DETERMINATION OF INHERENT AND ADDITIONAL FILTRATION IN ORDER TO ESTABLISH RADIATION QUALITIES ACCORDING TO IEC 61267

DETERMINATION OF INHERENT AND ADDITIONAL FILTRATION IN ORDER TO ESTABLISH RADIATION QUALITIES ACCORDING TO IEC 61267 2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 DETERMINATION

More information

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests Acknowledgements David Shepard, Ph.D. Daliang Cao, Ph.D. Muhammad K. N. Afghan, Ph.D. Jinsong Ye, M.S. Tony P. Wong, Ph.D. Fan Chen, Ph.D. Min Rao, Ph.D. Vivek Mehta, M.D. Igor Gomola, Ph.D. David Housley

More information

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications I. SCOPE The protocol applies to GafChromic EBT3P and EBT3+P films exposed in GafChromic QuiCk Phantom

More information

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure Accuracy and calibration of integrated radiation output indicators in diagnostic radiology: A report of the AAPM Imaging Physics Committee Task Group 190 Pei-Jan P. Lin a) Virginia Commonwealth University

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

- Water resistant. - Large size.

- Water resistant. - Large size. GAFCHROMIC EBT product brochure GAFCHROMIC EBT FEATURES GAFCHROMIC EBT dosimetry film has been developed specifically to address the needs of the medical physicist and dosimetrist working in the radiotherapy

More information

Technical Protocol. SIM Comparison of Calibration Coefficients at Radiotherapy Level for Orthovoltage X-ray Beams. Pilot laboratory: NIST

Technical Protocol. SIM Comparison of Calibration Coefficients at Radiotherapy Level for Orthovoltage X-ray Beams. Pilot laboratory: NIST SIM Comparison of Calibration Coefficients at Radiotherapy Level for Orthovoltage X-ray Beams Technical Protocol Pilot laboratory: NIST Contact person: Michelle O Brien National Institute of Standards

More information

A Scintillating Fiber Dosimeter for Radiology and Brachytherapy. with photodiode readout

A Scintillating Fiber Dosimeter for Radiology and Brachytherapy. with photodiode readout A Scintillating Fiber Dosimeter for Radiology and Brachytherapy with photodiode readout Florbela Rêgo a), and Maria da Conceição Abreu 5 Laboratório de Instrumentação e Física Experimental de Partículas,

More information

ADVANCED THERAPY DOSIMETER MODEL 35040

ADVANCED THERAPY DOSIMETER MODEL 35040 ADVANCED THERAPY DOSIMETER MODEL 35040 TECHNICAL SPECIFICATIONS Designed for ultra long-term stability error of approximately 0.1% per five years. Virtually removes effects of system leakage during measurement.

More information

Partial Replication of Storms/Scanlan Glow Discharge Radiation

Partial Replication of Storms/Scanlan Glow Discharge Radiation Partial Replication of Storms/Scanlan Glow Discharge Radiation Rick Cantwell and Matt McConnell Coolescence, LLC March 2008 Introduction The Storms/Scanlan paper 1 presented at the 8 th international workshop

More information

VERIFYING NUCLEAR WASTE TILE-HOLES USING GAMMA RADIATION SCANNING

VERIFYING NUCLEAR WASTE TILE-HOLES USING GAMMA RADIATION SCANNING FULL ARTICLE Nuclear waste management facilities at Chalk River Laboratories (CRL) use below-ground tile-holes to store solid waste from various activities such as medical isotope production. After long

More information

CT Scanner Dose Survey

CT Scanner Dose Survey CT Scanner Dose Survey Measurement Protocol Version 5.0 July 1997 Co-ordinated by ImPACT and The Medical Physics Department,, London SW17, UK. 0181-725-3366 CT Scanner Dose Survey: Measurement Protocol

More information

NOT FOR DISTRIBUTION JINST_083P_0914 v1

NOT FOR DISTRIBUTION JINST_083P_0914 v1 Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam F. Di Lillo a,b, D. Dreossi c, F. Emiro a,b, C. Fedon d,e, R. Longo d,e, G. Mettivier a,b,*, L. Rigon d,e, P. Russo

More information

GAFCHROMIC. Therapy Dosimetry Media Models to

GAFCHROMIC. Therapy Dosimetry Media Models to GAFCHROMIC Therapy Dosimetry Media Models 37-040 to 37-045! Superior uniformity and sensitivity! Dose rate and fractionation independent! Maps dose distribution! Provides quantitative measurements (via

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Detectors for small field dosimetry

Detectors for small field dosimetry Detectors or small ield dosimetry Hugo Palmans MedAustron, Wiener Neustadt, Austria and National Physical Laboratory, Teddington, UK 1 080915 Overview Ideal detector Water calorimeter Ionization chamber

More information

Scanning Electron Microscopy. EMSE-515 F. Ernst

Scanning Electron Microscopy. EMSE-515 F. Ernst Scanning Electron Microscopy EMSE-515 F. Ernst 1 2 Scanning Electron Microscopy Max Knoll Manfred von Ardenne Manfred von Ardenne Principle of Scanning Electron Microscopy 3 Principle of Scanning Electron

More information

Phantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging

Phantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging in partnership with Phantoms in Medical Physics (RT) U. Oelfke Division of Radiotherapy & Imaging uwe.oelfke@icr.ac.uk Making the discoveries that defeat cancer 1. Introduction What is a phantom? Wiki:

More information

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter MR-guided RT in Heidelberg Funded by the German Research Foundation (DFG) Heidelberg consortium received

More information

Measurements of Infrared Sources with the Missile Defense Transfer Radiometer

Measurements of Infrared Sources with the Missile Defense Transfer Radiometer Measurements of Infrared Sources with the Missile Defense Transfer Radiometer Simon G. Kaplan #, Solomon I. Woods #, Adriaan C. Carter, and Timothy M. Jung * # National Institute of Standards and Technology

More information

GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data

GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data Description GAFCHROMIC HD-810 dosimetry film is designed for the measurement of absorbed dose of high-energy

More information

Ludlum Medical Physics

Ludlum Medical Physics Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to

More information

Energy Monitoring Device for Electron Beam Facilities

Energy Monitoring Device for Electron Beam Facilities 1 SM/EB-18 Energy Monitoring Device for Electron Beam Facilities M. Lavalle 1, P.G. Fuochi 1, A. Martelli 1, U. Corda 1, A. Kovács 2, K. Mehta 3, F. Kuntz 4, S. Plumeri 4 1 CNR-ISOF, Via P. Gobetti 101,

More information

Radiation Lot Acceptance Testing (RLAT) of the RH1009MH 2.5V Voltage Reference for Linear Technology

Radiation Lot Acceptance Testing (RLAT) of the RH1009MH 2.5V Voltage Reference for Linear Technology Radiation Lot Acceptance Testing (RLAT) of the RH1009MH 2.5V Voltage Reference for Linear Technology Customer: Linear Technology, PO# 62118L RAD Job Number: 12-213 Part Type Tested: RH1009MH 2.5V Voltage

More information

ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung

ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung Treffen des Arbeitskreises IMRT der DGMP Würzburg, 26 + 27.03.2009 Salih Arican Sun Nuclear Corporation QA Challenge for Rotational Beams

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

Update on NIST X-Ray Air-Kerma Standards and Calibrations

Update on NIST X-Ray Air-Kerma Standards and Calibrations CCRI(I)/03-19 Update on NIST X-Ray Air-Kerma Standards and Calibrations Michelle O Brien and Stephen M. Seltzer Ionizing Radiation Division National Institute of Standards and Technology Gaithersburg,

More information

CyberKnife Iris Beam QA using Fluence Divergence

CyberKnife Iris Beam QA using Fluence Divergence CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery

More information

Products contained in this shipment may be subject to ITAR regulations.

Products contained in this shipment may be subject to ITAR regulations. Products contained in this shipment may be subject to ITAR regulations. Warning: The export of these commodity(ies), technology, or software are subject either to the U.S. Commerce Department Export Administration

More information

The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 BUSINESS UNIT OF ISP

The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 BUSINESS UNIT OF ISP The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 GAFCHROMIC EBT dosimetry film Designed and optimized for ALL RT procedures Can be cut into

More information

Technical aspects on DAP calibration and CT calibration

Technical aspects on DAP calibration and CT calibration EURADOS Report 2015-03 European Radiation Dosimetry Group e. V. Braunschweig, August 2015 Technical aspects on DAP calibration and CT calibration H. Järvinen, O. Turak, M. Ginjaume, J. Daures, C. Hourdakis,

More information

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS By Aravind Ravichandran arr192@mail.usask.ca University of Saskatchewan Address: 2424 Cumberland

More information

High Voltage Instrumentation Cables for the ITER Superconducting Magnet Systems

High Voltage Instrumentation Cables for the ITER Superconducting Magnet Systems High Voltage Instrumentation Cables for the ITER Superconducting Magnet Systems Summary for Call for Nominations 1. Background and scope ITER will be the world's largest experimental facility to demonstrate

More information

Dosimetry and Position Sensing Ionization Chamber for Ion Beam Tracking

Dosimetry and Position Sensing Ionization Chamber for Ion Beam Tracking Features Dosimetry and Position Sensing Ionization Chamber for Ion Beam Tracking 25 cm x 25 cm sensitive area Ionization chamber with integral plane readout for dosimetry and 128 by 128 strip readout for

More information

Status of Low-Background Infrared Calibration Facility at NIST

Status of Low-Background Infrared Calibration Facility at NIST Status of Low-Background Infrared Calibration Facility at NIST Simon G. Kaplan #, Solomon I. Woods #, Julia Scherschligt #, Joseph P. Rice #, Adriaan C. Carter *, and Timothy M. Jung * # National Institute

More information

Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs

Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs Jeffrey L. Guttman, John M. Fleischer, and Allen M. Cary Photon, Inc. 6860 Santa Teresa Blvd., San Jose,

More information

A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine

A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine Iwein Van de Vondel, 1 Koen Tournel, 1 Dirk

More information

Handheld Infrared Thermometers

Handheld Infrared Thermometers DX SERIES Handheld Infrared Thermometers User s Guide EXERGEN 400 Pleasant Street - Watertown, MA 02472 Phone: 617.923.9900 Fax: 617.923.9911 www.exergen.com e-mail: industrial@exergen.com Table of Contents

More information

Discover. Blue Box. the. Difference. High Resistance Metrology Products Guide

Discover. Blue Box. the. Difference. High Resistance Metrology Products Guide Discover the Blue Box Difference High Resistance Metrology Products Guide Metrology is our Science, Accuracy is Our Business Measurements International (MI) is the world s premier metrology company. MI

More information

Radiation Lot Acceptance Testing (RLAT) of the RH1014MW Quad Precision Operational Amplifier for Linear Technology

Radiation Lot Acceptance Testing (RLAT) of the RH1014MW Quad Precision Operational Amplifier for Linear Technology Radiation Lot Acceptance Testing (RLAT) of the RH1014MW Quad Precision Operational Amplifier for Linear Technology Customer: Linear Technology, PO# 61846L RAD Job Number: 12-085 Part Type Tested: RH1014MW

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

Radiation Lot Acceptance Testing (RLAT) of the RH1014MW Quad Precision Operational Amplifier for Linear Technology

Radiation Lot Acceptance Testing (RLAT) of the RH1014MW Quad Precision Operational Amplifier for Linear Technology Radiation Lot Acceptance Testing (RLAT) of the RH1014MW Quad Precision Operational Amplifier for Linear Technology Customer: Linear Technology, PO# 62118L RAD Job Number: 12-214 Part Type Tested: RH1014MW

More information

Products contained in this shipment may be subject to ITAR regulations.

Products contained in this shipment may be subject to ITAR regulations. Products contained in this shipment may be subject to ITAR regulations. Warning: The export of these commodity(ies), technology, or software are subject either to the U.S. Commerce Department Export Administration

More information

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R Building 3, No.9, Jiuxianqiaodong Road, Chaoyang District, Beijing 100015, P.R. China E-mail:

More information