1 Electronic Brachytherapy Sources Thomas W. Rusch
2 Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration Understand methods to characterize and confirm source spatial distribution
3 Source Construction & Operation
4 Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration Understand methods to characterize and confirm source spatial distribution
5 Axxent X-Ray Source 50 kv operating potential High vacuum x-ray tube technology Output: ~10 Gy/min 1cm into tissue Water cooled Fully disposable device Output described in TG-43 format HV connection Cooling connections X-Ray Tube HV Cable X-Ray Source Tip Detail
6 Axxent X-Ray Source Components Cable Anode Envelope Filament
7 Axxent Controller Features Handheld barcode scanner for data entry Display Screen/ Touch Screen Control Retractable Power Cord Wheel brakes for position stability during treatment
8 Axxent Controller Features High voltage power supply and connector Emergency stop button Well Chamber for source constancy checks Adjustable arm for source pullback (in storage position)
9 Axxent Applicators Balloon Applicators Vaginal Cylinders Skin-Surface Applicators 3-4 cm Spherical 2.0 cm diameter 1.0 cm diameter 4-5 cm Spherical 2.5 cm diameter 2.0 cm diameter 5-6 cm Spherical 3.0 cm diameter 3.5 cm diameter 5x7 cm Ellipsoidal 3.5 cm diameter 5.0 cm diameter 6x7 cm Ellipsoidal
10 Educational Objective #1 Understand key elements of ebx source construction & operation Diode x-ray source operates at 50 kv and 0.30 ma Water-cooled in a 5.6 mm diameter catheter Dose distribution is described in the TG-43 format Portable controller operates and positions the source FDA has cleared the source and controller for use anywhere in or on the body where radiation treatment is indicated Applicator choices: balloons for accelerated partial breast irradiation, vaginal cylinders, cones for skin or internal organ surfaces
11 Air Kerma Strength Characterization
12 Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration Understand methods to characterize and confirm source spatial distribution
13 2-D Brachytherapy Dosimetry Formalism, AAPM TG-43U1 (2004) D(r,θ) = S K Λ g L (r) G L (r,θ) F(r,θ) G L (r 0,θ 0 ) D(r,θ) S K Λ g L (r) G L (r,θ) F(r,θ) dose rate to water at point P(r,θ) air kerma strength dose rate constant radial dose function geometry function (line source approximation) 2-D anisotropy function
14 Air Kerma Standards National primary standards for air kerma are based on either free-air chambers or cavity ionization chambers Free-air chambers: < 300 kev photons e.g. primary x-ray standards, low-energy brachytherapy sources Cavity ionization chambers: > 300 kev photons e.g. 137 Cs (662 kev) and 60 Co (1250 kev) HDR 192 Ir calibrations are based on an ionization chamber calibrated using 250 kv x-rays and 137 Cs 3/24
15 Calibration Traceability - Most Low Energy Sources Primary standard (NIST Wide-Angle Free-Air Chamber) Secondary standard (ADCL well ionization chamber) Clinical well ionization chamber
16 Interim Calibration Traceability UW Attix free-air chamber Primary standard (NIST free-air chamber) Secondary standard (ADCL well ionization chamber) Clinical well ionization chamber
17 Comparison to 125 I Well chamber was also calibrated at UW using a I seed for interim NIST traceability Conversion coefficients between ebx source and 125 I well chamber calibrations are: Voltage Ratio to 125 I 50 kv /- 3.0% 45 kv /- 2.7% 40 kv /- 4.7% Reference: Application Note 06-01, University of Wisconsin Radiation Calibration Laboratory 21/24
18 HDR 1000 Plus Well Chamber Special aluminum insert designed by Standard Imaging Source strength is measured with the well chamber prior to treatment
19 Well Chamber Integrated into the Controller Source in well chamber for calibration Well chamber connected internally to electrometer
20 Educational Objective #2 Understand the rationale and methods for air kerma strength calibration An interim well chamber calibration is being used while a national standard is developed at NIST The interim method was established at the University of Wisconsin Medical Radiation Research Lab Attix free air chamber and a I seed An integral well chamber is used to check source strength constancy prior to each fraction
21 Spatial Distribution Characterization
22 Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration Understand methods to characterize and confirm source spatial distribution
23 Axxent Brachytherapy Dosimetry Formalism D(r,θ) = S K Λ g P (r) r 02 r 2 F(r,θ) Simplified geometry function due to L < 1 mm Published g P (r) and F(r,θ) (Rivard et al 2006) All sources are tested for agreement with accepted parameter set Uncertainties are added in quadrature
24 Motivation - it s got to be fast and it s got to be right! To assure an accurate delivery of radiation, must match TG-43 protocol parameters to within an overall error budget Depth dose distribution within 0.03 of TG-43 reference Polar distribution within 0.10 of TG-43 reference Azimuthal asymmetry 7% Dose rate constant within 7% of nominal Output level and stability Using a HDR well chamber 0 90 High throughput An apparatus was designed, built and validated to accomplish these goals
25 Manufacturing Test Fixture Water tank Shielded enclosure with leaded acrylic window Carousel assembly holding all detectors PTW ion chamber Photodiode devices Critical fixtures are Solid Water TM Shielded bridge and diode amplifiers on top plate Source to be tested is fixed, carousel assembly rotates Bridge Assembly Top Plate (raised) Carousel Assembly Water Tank Assembly Source
26 Sample printed output
27 In the Clinic: Shielded Test Fixture Provides a safe enclosure to evaluate source performance Attaches to controller pullback arm like an applicator Can accept a small pinpoint-style ion chamber Source stability Quantitative pullback comparisons Can accept 12.5 x 12.5 cm sheets of radiochromic film Source spatial distribution in one plane Pullback accuracy
28 Pretreatment QA: Ion Chamber and Film Completely shielded Ion chamber port and radio-opaque position markers
29 Educational Objective #3 Understand methods to characterize and confirm source spatial distribution Prior to certification for human use, each source is evaluated in manufacturing test fixtures and must perform within prescribed limits The shielded test fixture allows measurement of relative source spatial characteristics using radiochromic film or an ion chamber
30 Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration Understand methods to characterize and confirm source spatial distribution
31 Acknowledgements Stephen Davis and Prof. Larry DeWerd, University of Wisconsin Medical Radiation Research Lab Linda Kelley and Steve Axelrod, Xoft, Inc. 24/24
Cylindrical Ion Chambers Victoreen Model 550 Series
Cylindrical Ion Chambers Victoreen Model 550 Series! Cylindrical Ion Chambers for use with Model 35040 and Model 530 electrometers! Wide range of applications in Diagnostic X-Ray and Radiation Oncology
More informationCylindrical Ion Chambers
Cylindrical Ion Chambers Radiation Oncology ON Victoreen Model 550T Series Cylindrical Ion Chambers for use with Model 35040 and Model 560 electrometers Wide range of applications in Diagnostic X-Ray and
More informationCommissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010
Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Indra J. Das, PhD, FACR Department of Radiation Oncology Indiana University of School of Medicine & Midwest Proton Radiation
More informationSPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by
SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This
More informationProgress Report from the HDR Brachytherapy Working Party. Margaret Bidmead
Progress Report from the HDR Brachytherapy Working Party Margaret Bidmead HDR Ir-192 calibration service provided by NPL since May 2004 Primary standard irradiated with a Nucletron classic HDR IR-192 source
More informationCalibration of KAP meters
Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area
More informationSRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools
SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film With improvements in targeting and localization, stereotactic treatments have become prevalent. To
More informationDETECTORS. Including Codes of Practice IONIZING RADIATION
DETECTORS Including Codes of Practice 2015 IONIZING RADIATION PTW History and General Remarks Looking back on a long history PTW-Freiburg is an internationally operating company, manufacturing and marketing
More informationDETECTORS UNCOMPROMISING QUALITY. The standard in dosimetry measurements for over 40 years. EXRADIN DETECTORS
DETECTORS UNCOMPROMISING QUALITY The standard in dosimetry measurements for over 40 years. EXRADIN DETECTORS The Exradin Advantage Better Components Waterproof construction eliminates the need for sleeves
More informationPreliminary studies of a new monitor ionization chamber
1 Preliminary studies of a new monitor ionization chamber Maíra T. Yoshizumi, Vitor Vivolo and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN) Comissão Nacional de Energia
More informationResponse characteristics of a tandem ionization chamber in standard X-ray beams
Applied Radiation and Isotopes 58 (2003) 495 500 Response characteristics of a tandem ionization chamber in standard X-ray beams Alessandro M. Costa*, Linda V.E. Caldas Instituto de Pesquisas Energ!eticas
More informationWhen small things matter. Small Field Dosimetry Application Guide
R A D I AT I O N T H E R A P Y When small things matter. Small Field Dosimetry Application Guide Contents 1 Introduction 1 Introduction 2 2 The Physics of Small Fields 3 3 Detector Types 10 4 Detector
More information3D Diode Array Commissioning: Building Confidence in 3D QA Technology
3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose
More informationIONIZING RADIATION 2018 / 2019 DETECTORS. Including Codes of Practice
IONIZING RADIATION DETECTORS 2018 / 2019 Including Codes of Practice PTW History and General Remarks Looking back on a long history PTW-Freiburg is an internationally operating company, manufacturing and
More informationEvaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array
Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array José A. Bencomo, * Geoffrey Ibbott, Seungsoo Lee, and Joao A. Borges Department of Radiation Physics.
More informationMEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM *
MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM * C. Kessler and D.T. Burns December 2018 BUREAU INTERNATIONAL DES POIDS ET MESURES
More informationX-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage
X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage
More informationQA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion
Leksell Gamma Knife Perfexion QA Considerations Paula L. Petti, Ph.D. Taylor McAdam Bell Neuroscience Institute Washington Hospital Healthcare System Fremont, CA 1 QA for LGK Perfexion : : Follow NRC licensing
More informationComparison of peripheral dose measurements using Ionization chamber and MOSFET detector
ORIGINAL ARTICLES Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector Gopiraj ANNAMALAI 1, Ramasubramanian VELAYUDHAM 2 ABSTRACT Received: 7.07.2009 Accepted: 2.11.2009
More informationTest Equipment for Radiology and CT Quality Control Contents
Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3
More informationAbstract. This investigation was supported by PHS grant CA10953 awarded by the NCI, DHHS. Introduction
Practical Considerations for the Calibration of Low Energy/Low Activity Seeds Peter A. Balter, J. Francisco Aguirre, and William F. Hanson Department of Radiation Physics The University of Texas M. D.
More informationA Sort-of Tissue Equivalent Proportional Counter (STEPC) for Space Radiation Dosimetry Applications
A Sort-of Tissue Equivalent Proportional Counter (STEPC) for Space Radiation Dosimetry Applications Eric Benton, Tyler Collums, and Art Lucas E. V. Benton Radiation Physics Laboratory Oklahoma State University
More informationAn Activity in Computed Tomography
Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).
More informationA Guide to Radiochromic Film Dosimetry with EBT2 and EBT3
A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 David F. Lewis Advanced Materials Group Ashland Specialty Ingredients Spain, April 2014 What is Radiochromic Film? A film that instantly changes
More informationPeace of Mind. Automated.
1 Peace of Mind. Automated. Automated and guided beam commissioning Why SMARTSCAN? Get thousands of beam scans done effortlessly. Save your valuable time and manual operations at the water phantom and
More informationSCINTILLATING FIBER DOSIMETER ARRAY
SCINTILLATING FIBER DOSIMETER ARRAY FIELD OF THE INVENTION [0001] This invention relates generally to the field of dosimetry and, more particularly, to rapid, high-resolution dosimeters for advanced treatment
More informationSRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools
SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film As clinics strive to treat with the smallest field possible, the demand for SRS/SBRT QA grows as well.
More informationAn Activity in Computed Tomography
Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).
More informationRADIATION THERAPY DETECTORS. For Small Field Dosimetry
RADIATION THERAPY DETECTORS For Small Field Dosimetry 0.125 cm 3 Semiflex Chamber Type 31010 Standard therapy chamber for scanning systems and for absolute dosimetry Waterproof, semiflexible design for
More informationPHYSICS QUESTIONNAIRE FORM
PHYSICS QUESTIONNAIRE FORM Institution Name: Date: Contact Information (name, address, phone, fax, email): Physicist: Radiation Oncologist: Dosimetrist (if applicable): Study Coordinator (if applicable):
More informationX-ray tube with needle-like anode
NUKLEONIKA 2002;47(3):101 105 ORIGINAL PAPER X-ray tube with needle-like anode Mieczys aw S apa, W odzimierz StraÊ, Marek Traczyk, Jerzy Dora, Miros aw Snopek, Ryszard Gutowski, Wojciech Drabik Abstract
More informationDose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp
Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division
More informationA new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 1, WINTER 2010 A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA Abdul Qadir Jangda,
More informationIntroduction of a Single Chip TLD System for Patient Dosimetry
Introduction of a Single Chip TLD System for Patient Dosimetry C. Hranitzky a, M. Halda a, G. Müller a, B. Obryk b, H. Stadtmann a* a Austrian Research Centers GmbH ARC, 2444 Seibersdorf, Austria. b Institute
More informationExperiment 6: Franck Hertz Experiment v1.3
Experiment 6: Franck Hertz Experiment v1.3 Background This series of experiments demonstrates the energy quantization of atoms. The concept was first implemented by James Franck and Gustaf Ludwig Hertz
More informationLI-193 Spherical Quantum Sensor
LI-193 Spherical Quantum Sensor The LI-193 Spherical Quantum Sensor measures PAR in air or underwater from all directions at depths up to 350 meters. This sensor is useful for studies of phytoplankton,
More informationMammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over
Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on
More informationRotational total skin electron irradiation with a linear accelerator
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 4, FALL 2008 Rotational total skin electron irradiation with a linear accelerator Eric P. Reynard, 1,a Michael D.C. Evans, 1 Slobodan Devic,
More informationMinnesota Rules, Chapter 4732 X-ray Revision
Minnesota Rules, Chapter 4732 X-ray Revision DRAFT INDUSTRIAL X-RAY SYSTEMS DEFINTIONS, 1.0 4732.####. INDUSTRIAL X-RAY SYSTEMS DEFINITIONS. Subpart 1. Scope. For purposes of industrial x-ray systems under
More informationThe Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017
The Current State of EPID-Based Linear Accelerator Quality Assurance Timothy Ritter, PhD, DABR, FAAPM 1 Disclosures Employed by the Veterans Health Administration Faculty appointment with the University
More informationLI-192 Underwater Quantum Sensor
LI-192 Underwater Quantum Sensor The LI-192 Underwater Quantum Sensor measures PAR from all angles in one hemisphere. The LI-192 works in air or underwater at depths up to 560 meters. The measurements
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced
More informationPD233: Design of Biomedical Devices and Systems
PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/
More informationSPECTROMETRIC DETECTION PROBE Model 310. Operator's manual
SPECTROMETRIC DETECTION PROBE Model 310 Operator's manual CONTENTS 1. INTRODUCTION... 3 2. SPECIFICATIONS... 4 3. DESIGN FEATURES... 6 4. INSTALLATION... 10 5. SAFETY AND PRECAUTIONS... 13 6. THEORY OF
More informationThe on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO)
The on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO) A. Ansarinejad1,2, A. Attili1, F. Bourhaleb2,R. Cirio1,2,M. Donetti1,3, M. A. Garella1, S. Giordanengo1,
More informationCHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY
14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to
More informationThe TSIS Spectral Irradiance Monitor: Prism Optical Degradation Studies
The TSIS Spectral Irradiance Monitor: Prism Optical Degradation Studies Lo Erik Richard, Dave Harber, Joel Rutkowski, Matt Triplett, Kasandra O Malia Laboratory for Atmospheric and Space Physics (LASP)
More informationOPERATION MANUAL. Model LET-SW5. April Far West Technology, Inc. 330 South Kellogg Ave, Suite D Goleta, CA 93117
OPERATION MANUAL Model LET-SW5 April 2010 Far West Technology, Inc. 330 South Kellogg Ave, Suite D Goleta, CA 93117 GENERAL INFORMATION This instrument is manufactured in the United States of America by:
More informationTotal Ionizing Dose Test Report. Z-Series DC-DC Converter
Total Ionizing Dose Test Report Z-Series DC-DC Converter Revision A March, 2004 TOTAL DOSE TEST REPORT for Z - SERIES DC/DC CONVERTER Project Engineer: Engineering Director: Tom Hanson Peter Lee TABLE
More information8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.
Use of EPIDs for Non-Routine Linac QA Bin Cai PhD Disclosure Parts of this project received support from Varian Medical System. Learning Objectives Learn the recent development of EPID based Non-routine
More informationSummer Student project report
Summer Student project report Mika Väänänen September 1, 2017 Abstract In this report I give a brief overview of my activities during the summer student project. I worked on the scintillating fibre (SciFi)
More information200 MHz 350 MHz 750 MHz Linac2 RFQ2 202 MHz 0.5 MeV /m Weight : 1000 kg/m Ext. diameter : 45 cm
M. Vretenar, CERN for the HF-RFQ Working Group (V.A. Dimov, M. Garlasché, A. Grudiev, B. Koubek, A.M. Lombardi, S. Mathot, D. Mazur, E. Montesinos, M. Timmins, M. Vretenar) 1 1988-92 Linac2 RFQ2 202 MHz
More informationPHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002
PHYSICS 334 - ADVANCED LABORATORY I COMPTON SCATTERING Spring 00 Purposes: Demonstrate the phenomena associated with Compton scattering and the Klein-Nishina formula. Determine the mass of the electron.
More informationDistributed source x-ray tube technology for tomosynthesis imaging
Distributed source x-ray tube technology for tomosynthesis imaging Authors: F. Sprenger a*, X. Calderon-Colon b, Y. Cheng a, K. Englestad a, J. Lu b, J. Maltz c, A. Paidi c, X. Qian b, D. Spronk a, S.
More informationABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES
Romanian Reports in Physics, Vol. 65, No. 1, P. 168 177, 213 ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES F. SCARLAT 1, A. SCARISOREANU
More informationLBIR Fluid Bath Blackbody for Cryogenic Vacuum Calibrations
LBIR Fluid Bath Blackbody for Cryogenic Vacuum Calibrations Timothy M. Jung*, Adriaan C. Carter*, Dale R. Sears*, Solomon I. Woods #, Dana R. Defibaugh #, Simon G. Kaplan #, Jinan Zeng * Jung Research
More informationDETERMINATION OF INHERENT AND ADDITIONAL FILTRATION IN ORDER TO ESTABLISH RADIATION QUALITIES ACCORDING TO IEC 61267
2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 DETERMINATION
More informationCommissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests
Acknowledgements David Shepard, Ph.D. Daliang Cao, Ph.D. Muhammad K. N. Afghan, Ph.D. Jinsong Ye, M.S. Tony P. Wong, Ph.D. Fan Chen, Ph.D. Min Rao, Ph.D. Vivek Mehta, M.D. Igor Gomola, Ph.D. David Housley
More informationGafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications
GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications I. SCOPE The protocol applies to GafChromic EBT3P and EBT3+P films exposed in GafChromic QuiCk Phantom
More informationKey words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure
Accuracy and calibration of integrated radiation output indicators in diagnostic radiology: A report of the AAPM Imaging Physics Committee Task Group 190 Pei-Jan P. Lin a) Virginia Commonwealth University
More informationI. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA
FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE
More information- Water resistant. - Large size.
GAFCHROMIC EBT product brochure GAFCHROMIC EBT FEATURES GAFCHROMIC EBT dosimetry film has been developed specifically to address the needs of the medical physicist and dosimetrist working in the radiotherapy
More informationTechnical Protocol. SIM Comparison of Calibration Coefficients at Radiotherapy Level for Orthovoltage X-ray Beams. Pilot laboratory: NIST
SIM Comparison of Calibration Coefficients at Radiotherapy Level for Orthovoltage X-ray Beams Technical Protocol Pilot laboratory: NIST Contact person: Michelle O Brien National Institute of Standards
More informationA Scintillating Fiber Dosimeter for Radiology and Brachytherapy. with photodiode readout
A Scintillating Fiber Dosimeter for Radiology and Brachytherapy with photodiode readout Florbela Rêgo a), and Maria da Conceição Abreu 5 Laboratório de Instrumentação e Física Experimental de Partículas,
More informationADVANCED THERAPY DOSIMETER MODEL 35040
ADVANCED THERAPY DOSIMETER MODEL 35040 TECHNICAL SPECIFICATIONS Designed for ultra long-term stability error of approximately 0.1% per five years. Virtually removes effects of system leakage during measurement.
More informationPartial Replication of Storms/Scanlan Glow Discharge Radiation
Partial Replication of Storms/Scanlan Glow Discharge Radiation Rick Cantwell and Matt McConnell Coolescence, LLC March 2008 Introduction The Storms/Scanlan paper 1 presented at the 8 th international workshop
More informationVERIFYING NUCLEAR WASTE TILE-HOLES USING GAMMA RADIATION SCANNING
FULL ARTICLE Nuclear waste management facilities at Chalk River Laboratories (CRL) use below-ground tile-holes to store solid waste from various activities such as medical isotope production. After long
More informationCT Scanner Dose Survey
CT Scanner Dose Survey Measurement Protocol Version 5.0 July 1997 Co-ordinated by ImPACT and The Medical Physics Department,, London SW17, UK. 0181-725-3366 CT Scanner Dose Survey: Measurement Protocol
More informationNOT FOR DISTRIBUTION JINST_083P_0914 v1
Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam F. Di Lillo a,b, D. Dreossi c, F. Emiro a,b, C. Fedon d,e, R. Longo d,e, G. Mettivier a,b,*, L. Rigon d,e, P. Russo
More informationGAFCHROMIC. Therapy Dosimetry Media Models to
GAFCHROMIC Therapy Dosimetry Media Models 37-040 to 37-045! Superior uniformity and sensitivity! Dose rate and fractionation independent! Maps dose distribution! Provides quantitative measurements (via
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationDetectors for small field dosimetry
Detectors or small ield dosimetry Hugo Palmans MedAustron, Wiener Neustadt, Austria and National Physical Laboratory, Teddington, UK 1 080915 Overview Ideal detector Water calorimeter Ionization chamber
More informationScanning Electron Microscopy. EMSE-515 F. Ernst
Scanning Electron Microscopy EMSE-515 F. Ernst 1 2 Scanning Electron Microscopy Max Knoll Manfred von Ardenne Manfred von Ardenne Principle of Scanning Electron Microscopy 3 Principle of Scanning Electron
More informationPhantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging
in partnership with Phantoms in Medical Physics (RT) U. Oelfke Division of Radiotherapy & Imaging uwe.oelfke@icr.ac.uk Making the discoveries that defeat cancer 1. Introduction What is a phantom? Wiki:
More informationInstallation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter
Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter MR-guided RT in Heidelberg Funded by the German Research Foundation (DFG) Heidelberg consortium received
More informationMeasurements of Infrared Sources with the Missile Defense Transfer Radiometer
Measurements of Infrared Sources with the Missile Defense Transfer Radiometer Simon G. Kaplan #, Solomon I. Woods #, Adriaan C. Carter, and Timothy M. Jung * # National Institute of Standards and Technology
More informationGAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data
GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data Description GAFCHROMIC HD-810 dosimetry film is designed for the measurement of absorbed dose of high-energy
More informationLudlum Medical Physics
Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to
More informationEnergy Monitoring Device for Electron Beam Facilities
1 SM/EB-18 Energy Monitoring Device for Electron Beam Facilities M. Lavalle 1, P.G. Fuochi 1, A. Martelli 1, U. Corda 1, A. Kovács 2, K. Mehta 3, F. Kuntz 4, S. Plumeri 4 1 CNR-ISOF, Via P. Gobetti 101,
More informationRadiation Lot Acceptance Testing (RLAT) of the RH1009MH 2.5V Voltage Reference for Linear Technology
Radiation Lot Acceptance Testing (RLAT) of the RH1009MH 2.5V Voltage Reference for Linear Technology Customer: Linear Technology, PO# 62118L RAD Job Number: 12-213 Part Type Tested: RH1009MH 2.5V Voltage
More informationArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung
ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung Treffen des Arbeitskreises IMRT der DGMP Würzburg, 26 + 27.03.2009 Salih Arican Sun Nuclear Corporation QA Challenge for Rotational Beams
More informationX-RAYS - NO UNAUTHORISED ENTRY
Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are
More informationUpdate on NIST X-Ray Air-Kerma Standards and Calibrations
CCRI(I)/03-19 Update on NIST X-Ray Air-Kerma Standards and Calibrations Michelle O Brien and Stephen M. Seltzer Ionizing Radiation Division National Institute of Standards and Technology Gaithersburg,
More informationCyberKnife Iris Beam QA using Fluence Divergence
CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery
More informationProducts contained in this shipment may be subject to ITAR regulations.
Products contained in this shipment may be subject to ITAR regulations. Warning: The export of these commodity(ies), technology, or software are subject either to the U.S. Commerce Department Export Administration
More informationThe most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 BUSINESS UNIT OF ISP
The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 GAFCHROMIC EBT dosimetry film Designed and optimized for ALL RT procedures Can be cut into
More informationTechnical aspects on DAP calibration and CT calibration
EURADOS Report 2015-03 European Radiation Dosimetry Group e. V. Braunschweig, August 2015 Technical aspects on DAP calibration and CT calibration H. Järvinen, O. Turak, M. Ginjaume, J. Daures, C. Hourdakis,
More informationTitle: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS
Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS By Aravind Ravichandran arr192@mail.usask.ca University of Saskatchewan Address: 2424 Cumberland
More informationHigh Voltage Instrumentation Cables for the ITER Superconducting Magnet Systems
High Voltage Instrumentation Cables for the ITER Superconducting Magnet Systems Summary for Call for Nominations 1. Background and scope ITER will be the world's largest experimental facility to demonstrate
More informationDosimetry and Position Sensing Ionization Chamber for Ion Beam Tracking
Features Dosimetry and Position Sensing Ionization Chamber for Ion Beam Tracking 25 cm x 25 cm sensitive area Ionization chamber with integral plane readout for dosimetry and 128 by 128 strip readout for
More informationStatus of Low-Background Infrared Calibration Facility at NIST
Status of Low-Background Infrared Calibration Facility at NIST Simon G. Kaplan #, Solomon I. Woods #, Julia Scherschligt #, Joseph P. Rice #, Adriaan C. Carter *, and Timothy M. Jung * # National Institute
More informationReal-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs
Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs Jeffrey L. Guttman, John M. Fleischer, and Allen M. Cary Photon, Inc. 6860 Santa Teresa Blvd., San Jose,
More informationA diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine Iwein Van de Vondel, 1 Koen Tournel, 1 Dirk
More informationHandheld Infrared Thermometers
DX SERIES Handheld Infrared Thermometers User s Guide EXERGEN 400 Pleasant Street - Watertown, MA 02472 Phone: 617.923.9900 Fax: 617.923.9911 www.exergen.com e-mail: industrial@exergen.com Table of Contents
More informationDiscover. Blue Box. the. Difference. High Resistance Metrology Products Guide
Discover the Blue Box Difference High Resistance Metrology Products Guide Metrology is our Science, Accuracy is Our Business Measurements International (MI) is the world s premier metrology company. MI
More informationRadiation Lot Acceptance Testing (RLAT) of the RH1014MW Quad Precision Operational Amplifier for Linear Technology
Radiation Lot Acceptance Testing (RLAT) of the RH1014MW Quad Precision Operational Amplifier for Linear Technology Customer: Linear Technology, PO# 61846L RAD Job Number: 12-085 Part Type Tested: RH1014MW
More informationY11-DR Digital Radiography (DR) Image Quality
Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures
More informationRadiation Lot Acceptance Testing (RLAT) of the RH1014MW Quad Precision Operational Amplifier for Linear Technology
Radiation Lot Acceptance Testing (RLAT) of the RH1014MW Quad Precision Operational Amplifier for Linear Technology Customer: Linear Technology, PO# 62118L RAD Job Number: 12-214 Part Type Tested: RH1014MW
More informationProducts contained in this shipment may be subject to ITAR regulations.
Products contained in this shipment may be subject to ITAR regulations. Warning: The export of these commodity(ies), technology, or software are subject either to the U.S. Commerce Department Export Administration
More informationChina Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R
China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R Building 3, No.9, Jiuxianqiaodong Road, Chaoyang District, Beijing 100015, P.R. China E-mail:
More information