Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter
|
|
- Cecilia Jennings
- 5 years ago
- Views:
Transcription
1 Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter
2
3 MR-guided RT in Heidelberg Funded by the German Research Foundation (DFG) Heidelberg consortium received one of two grants (project # ) for a comprehensive research program on MRgRT Timeline Mar. 2017: mid Aug. 2017: Dec. 2017: April 2018: start of vault remodeling start of Installation start of Commissioning first patient treatment
4 The MRIdian Linac Split superconducting magnet with 0.35T field strength Radiation gantry in 28cm gap between the two magnet halves, split gradient coils
5 The MRIdian Linac Buckets shield the Linac from the B 0 -Field as well as the MR from the Linac Linac: 6MV FFF, 600 MU/min, 90cm SAD Magnetron Circulator Puls Transformer Linac Gun Driver
6 RayZR MLC double stack, double focus MLC, 138 tungsten-alloy leaves max. Fieldsize 27.4 x 24.1 cm², min. Fieldsize 2 x 4.15 mm² each Leaf is 5.5cm high 8.3 mm Two different MLC configurations for a 4.15mm beam
7 The MR in MRIdian 0.35T magnetic field strength Split Gradients: slew 200mT/m/ms, 18mT/m peak pulse sequence: truefisp (bssfp) T2/T1-weighted contrast, highlighting fat and fluids cine-mr with 4 frames/s
8 MR Imaging: coils Gradient coils: Body coil:
9 MR Imaging: coils Surface coils: Torso Head and Neck Attenuation of treatment beam approx. 0.5%, Simulation coils for CT sim
10 Installation The Viewray MRIdian can be installed in a standard vault In Heidelberg: old Siemens MEVATRON machine in a two-storey vault
11 Installation The Viewray MRIdian can be installed in a standard vault In Heidelberg: old Siemens MEVATRON machine in a two-storey vault
12 Installation The Viewray MRIdian can be installed in a standard vault In Heidelberg: old Siemens MEVATRON machine in a two-storey vault
13 Installation: before: after:
14 Quench vent
15 Fringe fields 5 gauss line (usually the controlled zone ) is close to the machine 1 gauss line is (in our case) at bunker door comparison: the earth s magnetic field is around 0.5 gauss
16 RF shielding copper cage encompassing the whole room RF filters and waveguides for every cable or duct RF cage is customizable: additional RF filters and waveguides combined RT/RF door: 15.5 tons of lead and heavy concrete air gasket seals RF cage with compressed air
17 MR-Commissioning: B 0 -Homogeneity Measured with a 45cm FOV field camera: 16 ppm for our system within 24cm DSV: DSV = Diameter Spherical Volume, defines usable imaging volume at isocenter spherical phantom, filled with water + nickel sulfate hexahydrate spectral peak analysis: frequency spread measured for different gantry angles
18 MR-Commissioning: SNR and Uniformity SNR analysis as defined in NEMA MS : Noise-Image without excitation SNR = 0.66 ROI signal mean ROI Noise SD Uniformity Uniformity % =100 1 (ROI Signal max ROI Signal min ) (ROI Signal max +ROI Signal min ) SNR of single coil elements: 2x6 per Coil-pair
19 Spatial Integrity Viewray 2D Spatial Integrity Phantom, placed TRA/COR and SAG plus lateral shifts of ±7cm and ±12.5cm Specification: < 1mm over 20cm DSV < 2mm over 35cm DSV Analysis: software tool percentage of points passing
20 ACR phantom test: software for automated analysis
21 Commissioning: RT part Equipment needs to be MR-safe (not ferromagnetic / no artifacts) MR compatible ionization chambers detector arrays: Sun Nuclear: IC-Profiler MR, ArcCheck MR PTW: Octavius 1500 MR, StarCheck maxi MR Scandidos: prototype Delta4 MR manual water tanks: no motorized 3D-tanks commercially available automated rotation devices do not work Effects on dose response due to magnetic field??
22 B-field effects on measuring equipment Ionization chambers diodes ionization chamber arrays film OSLD TLD Alanine Water calorimeters Thijs Perik, Master-Thesis U of Heidelberg, 2017
23 Reference Dosimetry Absolute dose calibration Use real water and prevent air gaps Adequately sized ionization chamber, for example 0.125ccm, (k Q,B ±0.003) Always parallel to B 0 Reference condition for the MRIdian: 90cm SAD, 1.6cm depth Field size 10 x 9.96 cm² Measurement per calibration protocol (AAPM TG-51, IAEA TRS 398, DIN )
24 B-field effects on measuring equipment without B-field Example: Air cavity similar to an array of airfilled Ionization Chambers with B-field
25 Tests covered in FTP and ATP MR & RT Isocenter, Laser alignment Gantry rotation Isocentricity Output consistency with Gantry Angle Dose/MU linearity Uniformity of radiation field with Gantry Rotation MLC transmission (spec: av. < 0.375%, max. < 1%) Linac head leakage (spec: < 0.2% at 1m distance) MLC accuracy (spec: < 1mm) Couch attenuation Monte Carlo Beam Model validation IMRT delivery verification Gating tests...
26 Profile comparison Elekta Synergy 6MV FFF (blue) vs Viewray MRIdian Linac (red)
27 New Phantoms needed Viewray QA Phantom Filled with water for MR signal Can be used for several tests, holds film as well as different ionization chambers check for imaging isocenter, output consistency, couch (re)positioning, coil tests, MRI spot checks
28 Vielen Dank an: DFG Alle beteiligten Heidelberger Teams!
3D Diode Array Commissioning: Building Confidence in 3D QA Technology
3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose
More informationCommissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests
Acknowledgements David Shepard, Ph.D. Daliang Cao, Ph.D. Muhammad K. N. Afghan, Ph.D. Jinsong Ye, M.S. Tony P. Wong, Ph.D. Fan Chen, Ph.D. Min Rao, Ph.D. Vivek Mehta, M.D. Igor Gomola, Ph.D. David Housley
More informationThe Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017
The Current State of EPID-Based Linear Accelerator Quality Assurance Timothy Ritter, PhD, DABR, FAAPM 1 Disclosures Employed by the Veterans Health Administration Faculty appointment with the University
More informationWeber State University Radiologic Technology 4603
Weber State University Radiologic Technology 4603 MRI Physics and Instrumentation Instructor: Rex T. Christensen MHA R.T. (R) (MR) (CT) (ARRT) CIIP Contact Info: E-mail: rexchristensen@weber.edu Phone:
More informationSRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools
SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film With improvements in targeting and localization, stereotactic treatments have become prevalent. To
More informationA diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine Iwein Van de Vondel, 1 Koen Tournel, 1 Dirk
More informationIMRT Delivery System QA. IMRT Dose Delivery. Acceptance testing. Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) MLC Alignment
1 IMRT Delivery System Q Thomas LoSasso, PhD Memorial Sloan Kettering Cancer Center IMRT Dose Delivery cceptance testing Commissioning Quality assurance Verification Q Why: specific tests for IMRT? 2.
More informationCharacterization, Commissioning and Evaluation of Delta 4 IMRT QA System. Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX.
Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX. 1 Acknowledgements Collaborators: Jose Bencomo, Rafael. M. Landrove, Peter
More informationPHYSICS QUESTIONNAIRE FORM
PHYSICS QUESTIONNAIRE FORM Institution Name: Date: Contact Information (name, address, phone, fax, email): Physicist: Radiation Oncologist: Dosimetrist (if applicable): Study Coordinator (if applicable):
More informationArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung
ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung Treffen des Arbeitskreises IMRT der DGMP Würzburg, 26 + 27.03.2009 Salih Arican Sun Nuclear Corporation QA Challenge for Rotational Beams
More informationSRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools
SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film As clinics strive to treat with the smallest field possible, the demand for SRS/SBRT QA grows as well.
More information8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.
Use of EPIDs for Non-Routine Linac QA Bin Cai PhD Disclosure Parts of this project received support from Varian Medical System. Learning Objectives Learn the recent development of EPID based Non-routine
More informationMARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005
ACR MRI accreditation program MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics,
More informationIQM Detector Characteristics: Signal reproducibility
The Integral Quality Monitor (IQM) System is a real-time beam verification system that monitors the accuracy of radiation delivery throughout each patient treatment without any user interaction. IQM continuously
More informationMRI SYSTEM COMPONENTS Module One
MRI SYSTEM COMPONENTS Module One 1 MAIN COMPONENTS Magnet Gradient Coils RF Coils Host Computer / Electronic Support System Operator Console and Display Systems 2 3 4 5 Magnet Components 6 The magnet The
More informationArcCHECK. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools
ArcCHECK The Ultimate 4D QA Solution A 4D isotropical cylindrical detector array for arc delivery QA and Dosimetry U.S.Patent No. 8,044,359; 6,125,335 Compatible with: FFF Beams VMAT RapidArc TomoTherapy
More informationBACKGROUND. ** 78% of all MRI scanners have Image Quality problems. *** *** 25% of all Multi-Channel RF coils have at least one bad channel.
Range of Results from over 534 ACR-mandated Annual MRI Performance Evaluations on over 204 Magnets from 8 Vendors Spanning a 10-year Period Moriel NessAiver, Ph.D. - Simply Physics - Baltimore, MD moriel@simplyphysics.com
More information7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators
Implementing a new digital medical accelerator John Wong Johns Hopkins University AAPM, Austin, 2014 Acknowledgements Yin Zhang, Ken Wang, Kai Ding (Commissioning - JHU) Esteban Velarde, Joe Moore (QA
More informationGafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications
GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications I. SCOPE The protocol applies to GafChromic EBT3P and EBT3+P films exposed in GafChromic QuiCk Phantom
More informationNathan Childress, Ph.D., DABR
Nathan Childress, Ph.D., DABR Introduction TG-142 is a comprehensive QA protocol Covers nearly every aspect of machine and safety QA Recommends quantitative results Recommends high testing frequencies
More informationImpact of energy variation on Cone Ratio, PDD10, TMR20 10 and IMRT doses for flattening filter free (FFF) beam of TomoTherapy Hi-Art TM machines
JBUON 2014; 19(4): 1105-1110 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com ORIGINAL ARTICLE Impact of energy variation on Cone Ratio, PDD10, TMR20 10 and IMRT
More informationMR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements
MR Data for Treatment Planning: Issues, Protocol Optimization, and s (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements TG-117 Use of MRI Data in Treatment Planning
More informationCommissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010
Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Indra J. Das, PhD, FACR Department of Radiation Oncology Indiana University of School of Medicine & Midwest Proton Radiation
More informationThe Ultimate 4D QA Solution A 4D isotropic cylindrical detector array for arc delivery QA and Dosimetry.
The Ultimate 4D QA Solution A 4D isotropic cylindrical detector array for arc delivery QA and Dosimetry. U.S.Patent No. 8,044,359; 6,125,335 Your Most Valuable QA and Dosimetry Tools 2 Y o u r M o s t
More informationPhantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging
in partnership with Phantoms in Medical Physics (RT) U. Oelfke Division of Radiotherapy & Imaging uwe.oelfke@icr.ac.uk Making the discoveries that defeat cancer 1. Introduction What is a phantom? Wiki:
More informationCOMPREHENSIVE TG-142 IMAGING AND MACHINE QA
QA SOFTWARE COMPREHENSIVE TG-142 IMAGING AND MACHINE QA Automate the analysis of over thirty TG-142 recommended QA tasks The rapid progress of Radiation Therapy has created the need for Quality Assurance
More informationDevelopment of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA
Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Gena M.A.H 1, Ahmed L.El-Attar 2, Elbadry M. Zahran 3, Hany El-Gamal
More informationEmerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA
Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Arthur Olch, PhD, FAAPM AAPM Spring Clinical Meeting, March 21, 2017 Or.. What Dose are the Patients Really Getting???
More informationArcCHECKTM. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools. VMAT RapidArc TomoTherapy Pinnacle 3 SmartArc Conventional IMRT
TM The Ultimate 4D QA Solution A 4D isotropical cylindrical detector array for arc delivery QA and Dosimetry U.S.Patent No. 8,044,359 What is? The world s first true 4D detector array The world s first
More informationBeam Commissioning and Annual QA Phantoms, Detectors & Accessories. iba-dosimetry.com
Beam Commissioning and Annual QA Phantoms, Detectors & Accessories iba-dosimetry.com Benefit from efficient and smartly designed beam commissioning & annual QA tools that make your life EASIEST. For MORE
More informationThe most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 BUSINESS UNIT OF ISP
The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 GAFCHROMIC EBT dosimetry film Designed and optimized for ALL RT procedures Can be cut into
More informationCyberKnife Iris Beam QA using Fluence Divergence
CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery
More informationQUALITY CONTROL SOFTWARE FOR RADIOTHERAPY AND MEDICAL IMAGING
1 QUALITY CONTROL SOFTWARE FOR RADIOTHERAPY AND MEDICAL IMAGING The QUALIMAGIQ platform and its modules An optimal software solution for automatic analyses and traceability of quality controls in radiation
More informationI. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA
FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE
More informationSUN NUCLEAR. EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm. corporation. Your Most Valuable QA and Dosimetry Tools
EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm SUN NUCLEAR corporation Your Most Valuable QA and Dosimetry Tools introduction Pre-treatment dose QA is an important process required
More informationMonte Carlo study on a new concept of a scanning photon beam system for IMRT
NUKLEONIKA 2011;56(4):291 297 ORIGINAL PAPER Monte Carlo study on a new concept of a scanning photon beam system for IMRT Anna M. Wysocka-Rabin, Günter H. Hartmann Abstract. Intensity-modulated radiation
More informationA proposed method for linear accelerator photon beam steering using EPID
Received: 13 January 2018 Revised: 11 May 2018 Accepted: 29 June 2018 DOI: 10.1002/acm2.12419 RADIATION ONCOLOGY PHYSICS A proposed method for linear accelerator photon beam steering using EPID Michael
More informationInitial setup and subsequent temporal position monitoring using implanted RF transponders
Initial setup and subsequent temporal position monitoring using implanted RF transponders James Balter, Ph.D. University of Michigan Has financial interest in Calypso Medical Technologies Acknowledgements
More informationBeam Production, Characteristics and Shaping
Beam Production, Characteristics and Shaping Dr. Manfred Sassowsky Outline X-ray production 60 Co units Linear Accelerators Beam characteristics Beam shaping Beam Production, Characteristics and Shaping
More informationPeace of Mind. Automated.
1 Peace of Mind. Automated. Automated and guided beam commissioning Why SMARTSCAN? Get thousands of beam scans done effortlessly. Save your valuable time and manual operations at the water phantom and
More informationEvaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array
Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array José A. Bencomo, * Geoffrey Ibbott, Seungsoo Lee, and Joao A. Borges Department of Radiation Physics.
More informationSPECIFICATIONS FOR AN MRBR 7.0 TESLA / 210MM ACTIVELY SHIELDED MAGNET SYSTEM
SPECIFICATIONS FOR AN MRBR 7.0 TESLA / 210MM ACTIVELY SHIELDED MAGNET SYSTEM Prepared by:- Magnex Scientific Limited The Magnet Technology Centre 6 Mead Road Oxford Industrial Park Yarnton, Oxford OX5
More informationA Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images
A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images Aiping Ding, Bin Han, Lei Wang, Lei Xing Department of Radiation Oncology, Stanford University School of
More informationDOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING
DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev. 1.0 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE DoseLab users may reference the following instructions to perform Tomotherapy Quality Assurance tests as recommended
More information1.5T HIGH FIELD SMALL ANIMAL MRI
1.5T HIGH FIELD SMALL ANIMAL MRI Designed Specifically for Veterinarians TECHNICAL GUIDE ADVANCING THE ART AND SCIENCE OF VETERINARY MRI The PetVet is the only high-field MRI system designed specifically
More informationCHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY
14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to
More informationHardware. MRI System. MRI system Multicoil Microstrip. Part1
Hardware MRI system Multicoil Microstrip MRI System Part1 1 The MRI system is made up of a variety of subsystems. the Operator Workspace Gradient Driver subsystem The Physiological Acquisition Controller
More informationSPECIFICATION FOR A 7.0 TESLA/400MM ROOM TEMPERATURE BORE MAGNET SYSTEM
SPECIFICATION FOR A 7.0 TESLA/400MM ROOM TEMPERATURE BORE MAGNET SYSTEM Prepared by:- Magnex Scientific Limited The Magnet Technology Centre 6 Mead Road Oxford Industrial Park Yarnton, Oxford OX5 1QU,
More informationTOMOTHERAPY H SERIES TomoH, TomoHD and TomoHDA Systems Technical Specifications
TOMOTHERAPY H SERIES TomoH, TomoHD and TomoHDA Systems Technical Specifications The TomoTherapy H Series is designed to treat the entire spectrum of radiation therapy patients with enhanced speed, precise
More informationQC Testing for Computed Tomography (CT) Scanner
QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including
More informationMR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP
MR Data for Treatment Planning and Stereotactic Procedures: Sources of Distortion, Protocol Optimization, and Assessment (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements
More informationBackground (~EE369B)
Background (~EE369B) Magnetic Resonance Imaging D. Nishimura Overview of NMR Hardware Image formation and k-space Excitation k-space Signals and contrast Signal-to-Noise Ratio (SNR) Pulse Sequences 13
More informationStability of the Helical TomoTherapy Hi Art II detector for treatment beam irradiations
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 6, 2014 Stability of the Helical TomoTherapy Hi Art II detector for treatment beam irradiations Karin Schombourg, François Bochud, Raphaël
More informationISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM
FILMQA STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE Supports all major radiotherapy technologies! FilmQA is optimized for use with Gafchromic film products, including EBT2 and RTQA2. FILMQA helps
More informationQUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING
1 QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING QualiFormeD Phantoms A selection of test objects facilitating regulatory quality controls in radiation therapy and medical imaging Practical,
More informationPRELIMINARY SPECIFICATIONS MRBR 7.0 TESLA / 210MM ACTIVELY SHIELDED CRYO-COOLED MAGNET SYSTEM
PRELIMINARY SPECIFICATIONS MRBR 7.0 TESLA / 210MM ACTIVELY SHIELDED CRYO-COOLED MAGNET SYSTEM Prepared by:- Magnex Scientific Limited The Magnet Technology Centre 6 Mead Road Oxford Industrial Park Yarnton,
More informationSPECIFICATIONS FOR A 4.7 TESLA/310MM BORE ACTIVELY SHIELDED MAGNET SYSTEM
SPECIFICATIONS FOR A 4.7 TESLA/310MM BORE ACTIVELY SHIELDED MAGNET SYSTEM Prepared by:- Magnex Scientific Limited The Magnet Technology Centre 6 Mead Road Oxford Industrial Park Yarnton, Oxford OX5 1QU,
More informationPhilips Site Yearly Performance Evaluation Philips Openview 16-Jan-08. Table of Contents
Philips Site Yearly Performance Evaluation Philips Openview 6-Jan-8 Table of Contents Summary and Signature Page 2 Specific Comments 3 Site Information 4 Equipment Information 4 Table Position Accuracy
More informationEvaluation of a diode array for QA measurements on a helical tomotherapy unit
Evaluation of a diode array for QA measurements on a helical tomotherapy unit K. M. Langen, a S. L. Meeks, D. O. Poole, T. H. Wagner, T. R. Willoughby, O. A. Zeidan, and P. A. Kupelian Department of Radiation
More informationWhen small things matter. Small Field Dosimetry Application Guide
R A D I AT I O N T H E R A P Y When small things matter. Small Field Dosimetry Application Guide Contents 1 Introduction 1 Introduction 2 2 The Physics of Small Fields 3 3 Detector Types 10 4 Detector
More informationTOMOHD. Product Specifications
TOMOHD Treatment System Product Specifications 1. Complete Treatment System Dimensions Height Width Length Weight 8 ft., 3.25 in. (252.1 cm) 9 ft., 2.50 in. (280.6 cm) 15 ft., 2.25 in. (463.0 cm) 10,000
More informationSPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by
SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This
More informationMobius3D. Software based IMRT QA
Mobius3D Software based IMRT QA What is Mobius Medical Systems? Clinical Expertise Software Expertise Nathan Childress, Ph.D., Founder Eli Stevens, Chief Technical Officer Support Expertise Physicists
More informationComparison of peripheral dose measurements using Ionization chamber and MOSFET detector
ORIGINAL ARTICLES Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector Gopiraj ANNAMALAI 1, Ramasubramanian VELAYUDHAM 2 ABSTRACT Received: 7.07.2009 Accepted: 2.11.2009
More informationMonica Kishore. Medical Physics Graduate Program Duke University. Approved: Jennifer O Daniel, Co-Supervisor. Fang-Fang Yin, Co-Supervisor
Accuracy of Planar Dosimetry for Volumetric Modulated Arc Therapy Quality Assurance by Monica Kishore Medical Physics Graduate Program Duke University Date: Approved: Jennifer O Daniel, Co-Supervisor Fang-Fang
More informationQA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion
Leksell Gamma Knife Perfexion QA Considerations Paula L. Petti, Ph.D. Taylor McAdam Bell Neuroscience Institute Washington Hospital Healthcare System Fremont, CA 1 QA for LGK Perfexion : : Follow NRC licensing
More information8/3/2016. The EPID Strikes Back. Novel Applications for Current EPID Technology. Joerg Rottmann, PhD. Disclosures and acknowledgements
The EPID Strikes Back Joerg Rottmann Brigham and Women s Hospital / Dana-Farber Cancer Institute Harvard Medical School Disclosures and acknowledgements Disclosures Varian MRA grant Acknowledgements Boston
More informationFour-dimensional in vivo dosimetry by dose reconstruction using continuous EPID images and phase sorting method. JiHyung Yoon.
Four-dimensional in vivo dosimetry by dose reconstruction using continuous EPID images and phase sorting method Director of Dissertation: Dr. Jae Won Jung Major Department: Physics by JiHyung Yoon July,
More informationWide beam CT dosimetry. Elly Castellano
Wide beam CT dosimetry Elly Castellano Outline revision: CT dose indices wide-beam CT: the end of the road for CTDI? the IEC rescue plan for CTDI 100 the american way AAPM report 111 better estimates of
More informationQuality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT
Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance
More informationElectronic Brachytherapy Sources. Thomas W. Rusch
Electronic Brachytherapy Sources Thomas W. Rusch Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration
More informationIsocenter and Field of View Accuracy Measurement Software for Linear Accelerator
Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator Aleksei E. Zhdanov 1 and Leonid G. Dorosinskiy 1 Ural Federal University named after the first President of Russia B. N.
More informationAddressing Limitations of a Spatially Sensitive Large-Area Ion Chamber for Real-Time Verification of Intensity Modulated Radiation Therapy.
Addressing Limitations of a Spatially Sensitive Large-Area Ion Chamber for Real-Time Verification of Intensity Modulated Radiation Therapy. by Xun Lin A thesis submitted in conformity with the requirements
More informationA positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Huaiqun Guan,
More informationCombined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8
Combined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8 UNIVERSITY OF CAMBRIDGE 1.2 Rob C. Hawkes 1, Tim D. Fryer 1, Alun J. Lucas 1,2, Stefan B. Siegel 3, Richard
More informationAccurate two-dimensional IMRT verification using a back-projection EPID dosimetry method
Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method Markus Wendling, Robert J. W. Louwe, a Leah N. McDermott, Jan-Jakob Sonke, Marcel van Herk, and Ben J. Mijnheer
More informationVolumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA
Volumetric Modulated Arc Therapy David Shepard Swedish Cancer Institute Seattle, WA Disclaimer Our VMAT work has been sponsored in part by Elekta. Outline David Shepard VMAT Basics and VMAT Plan Quality
More informationTECHNICAL SPECIFICATIONS. FOR AN MRBR 7.0 TESLA / 160mm ACTIVELY SHIELDED ROOM TEMPERATURE BORE MAGNET SYSTEM
TECHNICAL SPECIFICATIONS FOR AN MRBR 7.0 TESLA / 160mm ACTIVELY SHIELDED ROOM TEMPERATURE BORE MAGNET SYSTEM Prepared by:- Magnex Scientific Limited The Magnet Technology Centre 6 Mead Road Oxford Industrial
More informationDETECTORS UNCOMPROMISING QUALITY. The standard in dosimetry measurements for over 40 years. EXRADIN DETECTORS
DETECTORS UNCOMPROMISING QUALITY The standard in dosimetry measurements for over 40 years. EXRADIN DETECTORS The Exradin Advantage Better Components Waterproof construction eliminates the need for sleeves
More informationNuclear Associates , , CT Head and Body Dose Phantom
Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product
More informationImpact of the high magnetic field and RF power in a superconducting cyclotron on the operation of a nearby MRI facility
ISSN 2469-5491 Impact of the high magnetic field and RF power in a superconducting cyclotron on the operation of a nearby MRI facility Chee-Wai Cheng1, Steven G Ferguson2, David Jordan3, Frederick Jesseph1,
More informationCommissioning an Elekta Versa HD linear accelerator
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 Commissioning an Elekta Versa HD linear accelerator Ganesh Narayanasamy, 1,2 Daniel Saenz, 1 Wilbert Cruz, 1,3 Chul S. Ha, 1 Niko
More information(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010
(N)MR Imaging Lab Course Script FMP PhD Autumn School Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder Date: November 3rd, 2010 1 Purpose: Understanding the basic principles of MR imaging
More informationCHAPTER 6 QUALITY ASSURANCE OF VARIAN ON-BOARD IMAGER
127 CHAPTER 6 QUALITY ASSURANCE OF VARIAN ON-BOARD IMAGER 6.1 INTRODUCTION Accurate and repeatable setup of patients is a requisite in radiotherapy. In the treatment of head-and-neck tumors, accurate setup
More informationMagnetic Resonance. Intera 1.5T Release 2.5 series Achieva 1.5T / 3.0T / XR Release 2.5 Panorama HFO Release 2.5 series. Technical Description
Magnetic Resonance 4522 132 68821 Intera 1.5T Release 2.5 series Achieva 1.5T / 3.0T / XR Release 2.5 Panorama HFO Release 2.5 series Technical Description Royal Philips Electronics N.V. 2007 All right
More informationLINEAR ACCELERATOR. Buyer's Guide. Version 1.1
PRE-OWNED LINEAR ACCELERATOR Buyer's Guide Version 1.1 Pre-Owned Linear Accelerator Buyer's Guide TABLE OF CONTENTS Considerations For Buying A Used Linear Accelerator... 3 Linear Accelerators Overview...
More informationBeam Production, characteristics and shaping
Kantonsspital Luzern Beam Production, characteristics and shaping Dr. Manfred Sassowsky Cantonal Hospital Lucerne (KSL) Institute for Radio-Oncology 3.9.2007 X-ray production 60 Co units Linear Accelerators
More informationUpdate on Fluoroscopy Physics AAPM MO-A-210A-1 Stephen Balter, Ph.D.
Update on Fluoroscopy Physics Stephen Balter, PhD Columbia University Draft for MO-A-210A-1 2009 AAPM Educational objectives Understand dosimetric concepts relating to interventional fluoroscopy Characterize
More informationThe evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response
Iran. J. Radiat. Res., 2005; 3 (1): 3-10 The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response M. Mohammadi 1,2,3* and E. Bezak 1,2
More informationC a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information
Ordering Information Please contact us if you have any questions or if you would like a quote or delivery schedule regarding the Catphan phantom. phone 800-525-1190, or 518-692-1190 fax 518-692-3329 mail
More information2 Hardware for Magnetic Resonance Imaging
Hardware for Magnetic Resonance Imaging 13 2 Hardware for Magnetic Resonance Imaging Kenneth W. Fishbein, Joseph C. McGowan, and Richard G. Spencer CONTENTS 2.1 Introduction 13 2.2 Magnets 13 2.2.1 Permanent
More informationAperture Based Inverse Planning AAPM Summer School 2003
Aperture Based Inverse Planning AAPM Summer School 003 D.M. Shepard, M.A. Earl, Y. Xiao, C.X. Yu Acknowledgements Ziping Jiang Allen Li Shahid Naqvi James Galvin Di Yan Prowess, Inc. University of Maryland
More informationBeam Diagnostics, Low Level RF and Feedback for Room Temperature FELs. Josef Frisch Pohang, March 14, 2011
Beam Diagnostics, Low Level RF and Feedback for Room Temperature FELs Josef Frisch Pohang, March 14, 2011 Room Temperature / Superconducting Very different pulse structures RT: single bunch or short bursts
More informationThe VARIAN 250 MeV Superconducting Compact Proton Cyclotron
The VARIAN 250 MeV Superconducting Compact Proton Cyclotron VARIAN Medical Systems Particle Therapy GmbH Friedrich-Ebert-Str. 1 D-51429 BERGISCH GLADBACH GERMANY OUTLINE 1. Why having a Superconducting
More informationThe promise of high-field MRI. High Field MRI Technology, Applications, Safety, and Limitations. High-field Scanners
High Field MRI Technology, Applications, Safety, and Limitations R. Jason Stafford, Ph.D. Department of Imaging Physics The University of Texas M. D. Anderson Cancer Center Houston, TX The promise of high-field
More informationMRI Metal Artifact Reduction
MRI Metal Artifact Reduction PD Dr. med. Reto Sutter University Hospital Balgrist Zurich University of Zurich OUTLINE Is this Patient suitable for MR Imaging? Metal artifact reduction Is this Patient suitable
More informationLudlum Medical Physics
Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to
More informationA Beam-Level Delivery Accuracy Study of the Robotic Image Guided Radiosurgery System Using a Scintillator/CCD Phantom
A Beam-Level Delivery Accuracy Study of the Robotic Image Guided Radiosurgery System Using a Scintillator/CCD Phantom Lei Wang 1, Shi Liu 1, Brett Nelson 2 1. Department of Radiation Oncology, Stanford
More informationGantry angle determination during arc IMRT: evaluation of a simple EPID-based technique and two commercial inclinometers
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 6, 2012 Gantry angle determination during arc IMRT: evaluation of a simple EPID-based technique and two commercial inclinometers Pejman Rowshanfarzad,
More informationSUPERCONDUCTING GANTRY AND OTHER DEVELOPMENTS AT HIMAC
SUPERCONDUCTING GANTRY AND OTHER DEVELOPMENTS AT HIMAC Y. Iwata *, K. Noda, T. Shirai, T. Murakami, T. Fujita, T. Furukawa, K. Mizushima, Y. Hara, S. Suzuki, S. Sato, and K. Shouda, NIRS, 4-9-1 Anagawa,
More information