CHAPTER 6 QUALITY ASSURANCE OF VARIAN ON-BOARD IMAGER

Size: px
Start display at page:

Download "CHAPTER 6 QUALITY ASSURANCE OF VARIAN ON-BOARD IMAGER"

Transcription

1 127 CHAPTER 6 QUALITY ASSURANCE OF VARIAN ON-BOARD IMAGER 6.1 INTRODUCTION Accurate and repeatable setup of patients is a requisite in radiotherapy. In the treatment of head-and-neck tumors, accurate setup is particularly important because of the proximity of the treatment volume to critical structures such as the spinal cord, brainstem, and parotid glands. When intensity-modulated radiotherapy (IMRT) is used, proximity of this kind between the target and normal organs often leads to highly inhomogeneous fluence profiles with steep dose gradients. Thus, repeatable and accurate patient setup is crucial to prevent the potential hazard of encompassing critical structures for example, the spinal cord in the high-dose region. Indeed, studies have shown that improper setup in the head-and-neck region can have significant effects on tumor coverage and normal-tissue sparing (Manning et al 2001, Samuelsson et al 2003, Astreinidou et al 2005). Improvement of head-and-neck setup precision and reproducibility therefore continues to be an active field of study. One approach to evaluating setup errors is to classify them into two categories: interfractional and intrafractional. The former reflect setup differences from day to day, and the latter, changes in patient position during a treatment session. Setup verification studies more often measure the former (Hess et al 1995, Pissani et al 2000, Asselen et al 2004, Kim et al 2004, Hong et al 2005, Guckenberger et al 2006, Linthout et al 2006, Zhang et al 2006), but some measure the latter (Kim et al 2000, Hong et al 2005, Guckenberger

2 128 et al 2006, Linthout et al 2006). Measurement of interfractional and intrafractional setup error through the course of treatment therefore requires repeated imaging of the patient on the treatment table. This imaging may result in excessive dose to the patient if megavoltage (MV) imaging is used. Kilovoltage (kv) imaging results in a lower patient dose and it can therefore be used with sufficient frequency to acquire more setup information per patient than traditional MV imaging allows. The advent of onboard kv imaging devices has also introduced new and more effective means for measuring and correcting setup error at the treatment machines. These devices not only allow measurement of setup error at the treatment console, but also remote correction of the setup based on the measured error. The advantages of kv over MV imaging for setup correction have been shown to include not only lower dose, but also smaller interobserver variability and significantly better setup error reduction in the headand neck region (Pissani et al 2000). Because these imagers are of the projection variety, little information about soft tissue is available, and registration occurs by alignment of bony anatomy. But the imagers can also be used to generate cone-beam computed tomography (CT) scans if soft tissue registration is required. The On-Board Imager system (On-Board Imager, Varian Medical Systems, Inc., Palo Alto, CA) is designed to correct for motion and setup errors of patients undergoing radiation therapy. The OBI system provides three imaging modes: two-dimensional (2-D) radiographic acquisition, fluoroscopic image acquisition, and three-dimensional (3-D) cone-beam computed tomography (CBCT) acquisition. The fluoroscopic images are used to verify the gating thresholds of the respiratory gating system (RPM, Varian Medical Systems, Inc.) to account for intrafraction (i.e., respiratory) motion. The radiographic images manage interfractional motion and setup

3 129 errors. The CBCT images provide soft tissue and bony structure information in 3D and also manage interfractional motion and setup errors (Jaffray and Siewerdson 2000, Siewerdson and Jaffray 2001, Guan and Lu 2005). Using the 2D2D Match and 3D3D Match analysis tools a user can register the acquired kv or CBCT images with their associated reference image (e.g., digitally reconstructed radiograph (DRR) or planning CT). Couch corrections are then downloaded to the linear accelerator and the couch is moved remotely. The use of this new technology necessitates a comprehensive quality assurance (QA) program to maintain and monitor system performance characteristics, which have been established at the time of commissioning. Currently, there have been no published recommendations and guidelines for a QA program to verify the functionality, accuracy, stability, and image quality of the radiographic and CBCT modes of this device. Several early adopters of the OBI technology independently developed their own QA programs using similar but not identical methods. This paper combines the best of these methods to generate a comprehensive QA program for the OBI and CBCT system. The QA program has three components: safety, geometry, and image quality for both radiographic and tomographic images. The most critical tests are those that evaluate the geometric accuracy of the OBI system, since the OBI system is intended for repositioning of the patient before/during treatment. All the geometric tests described in this publication assume that the location of the MV isocenter is known and can be identified independently (Kutcher et al 1994). Therefore, the tools used to identity the location of the MV isocenter (e.g., wall lasers, field light) are assumed to be calibrated. Image quality needs to be checked frequently enough to endure consistent image quality. Tests of image quality do not need to be as frequent as the geometry tests, but a regular program to detect changes in the imaging chain is still important.

4 REVIEW OF LITERATURE Yoo et al (2006) developed a quality assurance (QA) program for the On-Board Imager (OBI) system and to summarize the results of these QA tests over extended periods from multiple institutions. Both the radiographic and cone-beam computed tomography (CBCT) mode of operation have been evaluated. The QA programs from four institutions have been combined to generate a series of tests for evaluating the performance of the On-Board Imager. The combined QA program consists of three parts: (1) safety and functionality, (2) geometry, and (3) image quality. Safety and functionality tests evaluate the functionality of safety features and the clinical operation of the entire system during the tube warm-up. Geometry QA verifies the geometric accuracy and stability of the OBI/CBCT hardware/software. Image quality QA monitors spatial resolution and contrast sensitivity of the radiographic images. Image quality QA for CBCT includes tests for Hounsfield Unit (HU) linearity, HU uniformity, spatial linearity, and scan slice geometry, in addition. All safety and functionality tests passed on a daily basis. The average accuracy of the OBI isocenter was better than 1.5 mm with a range of variation of less than 1 mm over 8 months. The average accuracy of arm positions in the mechanical geometry QA was better than 1 mm, with a range of variation of less than 1 mm over 8 months. Measurements of other geometry QA tests showed stable results within tolerance throughout the test periods. Radiographic contrast sensitivity ranged between 2.2% and 3.2% and spatial resolution ranged between 1.25 and 1.6 lp/mm. Over four months the CBCT images showed stable spatial linearity, scan slice geometry, contrast resolution (_1%; < 7 mm disk) and spatial resolution (6 lp/cm). The HU linearity was within ±40 HU for all measurements. By combining test methods from multiple institutions, we have developed a comprehensive, yet practical, set of QA tests for the OBI system. Use of the tests over extended periods show that the OBI system has reliable mechanical accuracy and stable

5 131 image quality. Nevertheless, the tests have been useful in detecting performance deficits in the OBI system that needed recalibration. It is important that all tests are performed on a regular basis. Mechalokos et al (2007) used a kilovoltage imaging device to measure interfractional and intrafractional setup deviations in patients with head-and-neck or brain cancers receiving intensity-modulated radiotherapy (IMRT) treatment. Before and after IMRT treatment, approximately 3 times weekly, 7 patients were imaged using the Varian On-Board Imager (OBI: Varian Medical Systems, Palo Alto, CA), a kilovoltage imaging device permanently mounted on the gantry of a Varian 21EX LINAC (Varian Medical Systems). Because of commissioning of the remote couch correction of the OBI during the study, online setup corrections were performed on 2 patients. For the other 5 patients, weekly corrections were made based on a sliding average of the measured data. From these data, we determined the interfractional setup deviation (defined as the shift from the original setup position suggested by the daily image), the residual error associated with the weekly correction protocol, and the intrafractional setup deviation, defined as the difference between the post-treatment and pretreatment images. We also used our own image registration software to determine interfractional and intrafractional rotational deviations from the images based on the templatematching method. In addition, we evaluated the influence of inter-observer variation on our results, and whether the use of various registration techniques introduced differences. Finally, translational data were compared with rotational data to search for correlations. Translational setup errors from all data were 0.0 ± 0.2 cm, 0.1 ± 0.3 cm, and 0.2 ± 0.3 cm in the right left (RL), anterior posterior (AP), and superior inferior (SI) directions respectively. Residual error for the 5 patients with a weekly correction protocol was 0.1 ± 0.2 cm (RL),

6 ± 0.3 cm (AP), and 0.0 ± 0.2 cm (SI). Intrafractional translation errors were small, amounting to 0.0 ± 0.1 cm, 0.1 ± 0.2 cm, and 0.0 ± 0.1 cm in the RL, AP, and SI directions respectively. In the sagittal and coronal views respectively, interfractional rotational errors were 1.1 ± 1.7 degrees and 0.5 ± 0.9 degrees, and intrafractional rotational errors were 0.3 ± 0.6 degrees and 0.2 ± 0.5 degrees. No significant correlation was seen between translational and rotational data. The OBI image data were used to study setup error in the head-and-neck patients. Nonzero systematic errors were seen in the interfractional translational and rotational data, but not in the intrafractional data, indicating that the mask is better at maintaining head position than at reproducing it. The performance of a flat panel detector was investigated by Ejere (2006) for linearity, uniformity and bad-pixels. Especially at lower and higher exposures, the flat-panel detector was non-linear. The same exposures that cause non-linearity of the flat-panel detector also caused non-uniformity. The detector was linearized piecewise and exposures that cause non-uniformity were sought and calibrated. Pixel-by-pixel standard deviation investigation was used in correcting non-uniformity. They have compromised between the number of calibration points and the costly calculations involved. Bad-pixels were also examined and replaced by the average of the neighboring pixels. The effect of X-ray scatter on image quality was examined. X-ray scatter has the most detrimental effect on cone-beam CT image quality. Streaks and cupping are the most prominent artifacts caused by x-ray scatter. On top of that, scatter reduces contrast and increases noise. Scatter was estimated and corrected using three methods: Estimating scatter as uniform given by scatter-to-primary ratio (SPR) of 0.33, beam stop array technique, and simulation. Scatter was estimated by choosing a constant SPR of This way of estimating and correcting scatter reduced the cupping artifact to

7 133 some extent but did not eliminate the artifact completely. To estimate scatter more accurately, scatter was sampled using beam stop array technique and interpolated to generate scatter-only image. Scatter-only image was then subtracted from the total image to give the scatter corrected primary image. The total image and the scatter corrected primary image were then compared in order to inspect the effect of scatter on image quality and evaluate the scatter correction methods. Effect of scatter on CT number accuracy, contrast and noise was investigated using an image quality phantom. The method of simulation considered the effect of X-ray source fluctuation, area occupied by the beam stoppers, transmission of the lead discs and transmission of the phantom itself. Inclusion of these factors in their beam stop array correction method produced considerable result. CT number inaccuracy or cupping artifact was corrected after the employment of the scatter correction methods. Contrast improved satisfactorily. However, all the scatter correction methods incurred noise. Knight (2006) used daily Computer Tomography (CT) imaging to comprehensively assess organ volumes, organ motion and their effect on dose. Daily CT images were obtained using a Siemens Primus Linear Accelerator equipped with an in-room Somatom CT unit in the accelerator suite, marketed as Primatom, to accurately position the patient prior to treatment delivery. The internal structures of interest were contoured on the planning workstation by the investigator. The daily volume and location of the organs were derived from the computer to assess and analyse internal organ motion. The planned dose distribution was then imported onto the treatment CT datasets and used to compare the planned dose to i) the actual isocentre, where the isocentre was actually placed for that fraction, ii) the uncorrected isocentre, by un-doing any on-line corrections performed by the treatment staff prior to treatment delivery, and iii) the future isocentre, by placing the isocentre relative to internal organ motion on a daily basis.

8 134 The results of this study did not confirm a statistically significant decrease in rectum volumes over time (hypothesis 1), however large fluctuations in bladder volume were confirmed (hypothesis 2). Internal organ motion for the rectum and bladder was demonstrated to be related to organ filling. Ideal planning volumes for these organs have been reported to minimise systematic and random uncertainty in the treatment volumes. An observed decrease in prostate volume over time, a systematic uncertainty in the location of the prostate at the time of the planning CT scan and a significant relationship between prostate centre of volume and rectum and bladder volumes has resulted in a recommendation that patients should be re-scanned during treatment to ensure appropriate clinical target volume coverage. A significant relationship between rectal and bladder volumes and the dose delivered to these organs was found (hypothesis 3). The dose delivered to the planning target volume was not related to the rectal or bladder volumes, although it was related to the motion of these organs. Despite these results only minimal effects on the dose delivered to any of the three isocentres occurred, indicating that the planned dose was accurately delivered using the methodology presented here (hypothesis 4). However the results do indicate that the patient preparation instructions need to be improved if margins are to be reduced in the future. Millos (2006) investigated to evaluate the performance of an On-Board Imager (OBI) mounted on a clinical Linear Accelerator. The measurements were divided into three parts, geometric accuracy, image registration and couch shift accuracy, and image quality. A cube phantom containing a radiation opaque marker was used to study the agreement with treatment isocenter for both kv-images and cone-beam CT(CBCT) images. The long term stability was investigated by acquiring frontal and lateral kv images twice a week over a 3-month period. Stability in vertical and

9 135 longitudinal robotic arm motion as well as the stability of the center of rotation was evaluated. Further, the agreement of kv-images and CBCT center with MV image center was examined. A marker seed phantom was used to evaluate and compare the three applications in image registration 2D/2D, 2D/3D and 3D/3D. Image registration using kv-kv images sets were compared with MV-MV and MV-kV image sets. Further, the accuracy in 2D/2D matches with images acquired at non-orthogonal gantry angles was evaluated. The image quality in CBCT images was evaluated using a Catphan phantom. Hounsfield unit (HU) uniformity and linearity was compared with planning CT. HU accuracy is crucial for dose verification using CBCT data. The geometric measurements showed good long term stability and accurate position reproducibility after robotic arm motions. A systematic error of about 1 mm in lateral direction of the kv-image center was detected. A small difference between kv and CBCT between kv and MV images center was > 2 mm at some gantry angles. Image registration with different match applications worked sufficiently. 2D/3D match was seen to correct more accurately than 2D/2D match for large translational and rotational shifts. CBCT images acquired with full-fan mode showed good HU uniformity but half-fan images were less uniform. In the soft tissue region the HU agreement with planning CT was reasonable while a larger disagreement was observed at higher densities. This work shows that the OBI is robust and stable in its performance. With regular QC and calibrations the geometric precision of the OBI can be maintained within I mm of treatment isocenter. Different aspects of the imaging system (called XVI) of Elekta Synergy machine has been evaluated by Renstrom (2005), such as image quality, flex of the gantry rotation, accuracy for patient setup correction, x-ray tube output and absorbed imaging doses. The stability in 2D and 3D image

10 136 quality as well as the mechanical stability during rotation (also called flex) was measured on a weekly basis during a three month period. The stability of the system was studied by imaging a dense sphere positioned at isocenter. The 3D imaging was evaluated with a Catphan 500 phantom and the 2D imaging with a Leeds phantom. Accuracy in patient setup corrections was evaluated by scanning a pelvis phantom in different locations. Readout from the software was compared to actual translational movements, which were defined by the room laser readouts on rulers mounted on the table-top in the x- y and z-direction. A gold marker matching study was performed using an Alderson/Rando phantom. Finally, both CT dose index (CTDI) and dose measurements with TLDs were performed. The TLDs were placed between the slices of the Alderson/Rando phantom using thin boards of polystyrene. The XVI showed minor long term variations in flex and stable image quality. The accuracy of detecting patient setup-errors based on bony structures was flawless, although matching with respect to gold markers failed. The received effective dose from a pelvis scan was about 11 msv and for a head-and-neck scan about 0.15 msv. Accurate and reproducible setup of patients is a important part in radiotherapy. In recent development the IGRT with On Board Imaging System helps to achieve this goal. A series of QA tests were performed with equipments and tools as recommended by Varian and as per their QA protocols OBI customer acceptance procedure and OBI CBCT customer acceptance procedure. The results of the OBI QA procedures were within the specified limits as per Varian protocol. This test should be carried at a specific regular frequency to ensure the quality use of OBI for IGRT.

11 RESULTS Accurate and reproducible setup of patients is a important part in radiotherapy. In recent development the IGRT with On Board Imaging System helps to achieve this goal. A series of QA tests were performed with equipments and tools as recommended by Varian and as per their QA protocols OBI customer acceptance procedure and OBI CBCT customer acceptance procedure. The results of mechanical test are given in Tables 6.1 to 6.9. Table 6.1 Mechanical Tests and Results Sr. No. Test Specification Observation Comments Visual and Sound Collision test Motion Enable Test Mechanical isocenter Door interlock Warning light Audio sound warning KVS Collimator Cover KVD Imager Cover KVD/KVS Arm Paddles Pendant Manual Motion Console Automatic Motion Collimator Rotation Gantry Rotation radius radius

12 138 Table 6.1 (Continued) Sr. No. Test Specification Observation Comments Source Vertical Position vs PRO Tube Warm-up Technique PRO Vertical Postion Actual vertical Position Spec cm (P1) ±2mm 80.0cm (P2) Vertical Travel Range Pulsed Fluoro Mode Full Travel Range 21.0 cm kv ma ms Beam-on Time 75 80mA 32ms seconds Tool Measurements on a 10 x 10 cm image Imager Panel Alignment Optical Isocenter Test Tool Width Height spec Measured Distance mm±2mm Measured Area mm±2mm Measured Angle 45º 45º±1.15º Pixel position Actul Specification KVS/KVD Isocenter 1024x764 Actual 0.3 mm 1024x768±10(±2.0mm) Specification

13 139 Table 6.2 Imager Positioning Vs. Pro and Travel Ranges PRO VERTICAL POSITION 0.0 cm (P5) cm(p4) cm (P2) cm (P1) cm (P2) Vertical Travel Range Vertical Position Longitudinal Travel Range Lateral Travel Range Actual (add No spec No Spec offset) 1mm Spec±2mm Actual (add Actual Range offset) No Spec Spec±2mm Spec -20.5cm to =24cm) Actual (add Actual Range Actual Ranger offset) Spec±2mm Spec -19cm to Spec -18cm to =24cm) =16cm) Actual (add Actual Range Actual Ranger offset) Spec±2mm Spec -19cm to Spec -18cm to =24cm) =16cm) Longitudinal Position at Longitudinal Position at cm cm PRO Actual PRO Actual Spec±2mm Spec±2mm Lateral position at Lateral position at cm cm PRO Actual PRO Actual Spec±2mm Full Travel Range 85.5 Spec±2mm Spec: 80.0 cm below isocenter)

14 140 Table 6.3 Digital Fluoroscopy kvp, ma and ms measurements Indicated kvp Indicated ma Pulsed fluoro Technique Indicated ms Actual kvp Actual ma Acutal ms Table 6.4 Digital Fluoroscopy kvp, ma and ms measurements Indicated kvp Sinlgle Standard-Single pulse half resolution technique Indicated ma Indicated ms Actual kvp Actual ma Acutal ms

15 141 Table 6.5 Half value layer (HVL) using digital Fluoroscopy Pulsed Mode Unfiltered Pc Filtered Pc mm of Aluminum 70 kvp kvp Table 6.6 Shows Density resolution (Hounsfield/CTNo.) Standard dose 150cm Bow Tie Mean CT- No. Tolerance Measured CT-No. Head Scan Air 1000 ± Acrylic 120 ± LDPE 100 ± Body Scan Air 1000 ± Acrylic 120 ± LDPE 100 ± Table 6.7 Shows the Spatial Linearity measurement Standard dose 150cm Bow Tie Distance Accuracy Catphan Performance Measurement Head Scan 50 mm ±1 % 49.8 Body Scan 50 mm ±1 % 49.7

16 142 Table 6.8 Shows the High resolution Standard dose 150cm Bow Tie Line Pair / cm Specification Actual Line Pair / cm row Head Scan Row 7 8 Body Scan Row 6 7 Table 6.9 Shows the Low Contrast Resolution Standard dose 150cm Bow Tie Specification Supra Slice % Specificatio n Target size Actual Supra Slice % Actual Target size Head Scan 1 % 15 mm 1% 6 mm Body Scan 1 % 15 mm 1% 6 mm 6.4 DISCUSSION Accurate and reproducible setup of patients is a important part in radiotherapy. In recent development the IGRT with On Board Imaging System helps to achieve this goal. A series of QA tests were performed with equipments and tools as recommended by Varian and as per their QA protocols OBI customer acceptance procedure and OBI CBCT customer acceptance procedure. The results of the OBI QA procedures were within the specified limits as per Varian protocol. This test should be carried at a specific regular frequency to ensure the quality use of OBI for IGRT.

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Huaiqun Guan,

More information

Imaging system QA of a medical accelerator, Novalis Tx, for IGRT per TG 142: our 1 year experience

Imaging system QA of a medical accelerator, Novalis Tx, for IGRT per TG 142: our 1 year experience JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 4, 2012 Imaging system QA of a medical accelerator, Novalis Tx, for IGRT per TG 142: our 1 year experience Zheng Chang, a James Bowsher, Jing

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

PHYSICS QUESTIONNAIRE FORM

PHYSICS QUESTIONNAIRE FORM PHYSICS QUESTIONNAIRE FORM Institution Name: Date: Contact Information (name, address, phone, fax, email): Physicist: Radiation Oncologist: Dosimetrist (if applicable): Study Coordinator (if applicable):

More information

Initial setup and subsequent temporal position monitoring using implanted RF transponders

Initial setup and subsequent temporal position monitoring using implanted RF transponders Initial setup and subsequent temporal position monitoring using implanted RF transponders James Balter, Ph.D. University of Michigan Has financial interest in Calypso Medical Technologies Acknowledgements

More information

ISO Cube Daily QA Package

ISO Cube Daily QA Package ISO Cube Daily QA Package Model 023-05 AFFORDABLE TURNKEY SOLUTION FOR DAILY MACHINE QA POWERED BY AQUILAB 2428 Almeda Avenue Suite 316 Norfolk, Virginia 23513 USA Tel: 757-855-2765 WWW.CIRSINC.COM CAPABILITIES

More information

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc.

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. An Introduction to TG-142 Imaging QA Using Standard Imaging Products Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. Goals Understand the nature and intent of TG 142 imaging

More information

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

3D Diode Array Commissioning: Building Confidence in 3D QA Technology 3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Overview of Safety Code 35

Overview of Safety Code 35 Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)

More information

Nathan Childress, Ph.D., DABR

Nathan Childress, Ph.D., DABR Nathan Childress, Ph.D., DABR Introduction TG-142 is a comprehensive QA protocol Covers nearly every aspect of machine and safety QA Recommends quantitative results Recommends high testing frequencies

More information

8/3/2016. The EPID Strikes Back. Novel Applications for Current EPID Technology. Joerg Rottmann, PhD. Disclosures and acknowledgements

8/3/2016. The EPID Strikes Back. Novel Applications for Current EPID Technology. Joerg Rottmann, PhD. Disclosures and acknowledgements The EPID Strikes Back Joerg Rottmann Brigham and Women s Hospital / Dana-Farber Cancer Institute Harvard Medical School Disclosures and acknowledgements Disclosures Varian MRA grant Acknowledgements Boston

More information

Image Quality and Dose of an Accelerator-Integrated kv CBCT Systems

Image Quality and Dose of an Accelerator-Integrated kv CBCT Systems Image Quality and Dose of an Accelerator-Integrated kv CBCT Systems by Amani Shaaer A thesis submitted in partial fulfillment of the requirements for the degree of Master of Science in Physics School of

More information

Phantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging

Phantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging in partnership with Phantoms in Medical Physics (RT) U. Oelfke Division of Radiotherapy & Imaging uwe.oelfke@icr.ac.uk Making the discoveries that defeat cancer 1. Introduction What is a phantom? Wiki:

More information

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING 1 QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING QualiFormeD Phantoms A selection of test objects facilitating regulatory quality controls in radiation therapy and medical imaging Practical,

More information

COMPREHENSIVE TG-142 IMAGING AND MACHINE QA

COMPREHENSIVE TG-142 IMAGING AND MACHINE QA QA SOFTWARE COMPREHENSIVE TG-142 IMAGING AND MACHINE QA Automate the analysis of over thirty TG-142 recommended QA tasks The rapid progress of Radiation Therapy has created the need for Quality Assurance

More information

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators Implementing a new digital medical accelerator John Wong Johns Hopkins University AAPM, Austin, 2014 Acknowledgements Yin Zhang, Ken Wang, Kai Ding (Commissioning - JHU) Esteban Velarde, Joe Moore (QA

More information

QC by the MPE in Belgium

QC by the MPE in Belgium Acceptance testing of state-of-the-art CT scanners using a new national protocol: first experience on a large number of scanners of different make and model the working group Radiology of the Belgian Hospital

More information

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity) Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev. 1.0 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE DoseLab users may reference the following instructions to perform Tomotherapy Quality Assurance tests as recommended

More information

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017 The Current State of EPID-Based Linear Accelerator Quality Assurance Timothy Ritter, PhD, DABR, FAAPM 1 Disclosures Employed by the Veterans Health Administration Faculty appointment with the University

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

IQM Detector Characteristics: Signal reproducibility

IQM Detector Characteristics: Signal reproducibility The Integral Quality Monitor (IQM) System is a real-time beam verification system that monitors the accuracy of radiation delivery throughout each patient treatment without any user interaction. IQM continuously

More information

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System. Use of EPIDs for Non-Routine Linac QA Bin Cai PhD Disclosure Parts of this project received support from Varian Medical System. Learning Objectives Learn the recent development of EPID based Non-routine

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications I. SCOPE The protocol applies to GafChromic EBT3P and EBT3+P films exposed in GafChromic QuiCk Phantom

More information

Distributed source x-ray tube technology for tomosynthesis imaging

Distributed source x-ray tube technology for tomosynthesis imaging Distributed source x-ray tube technology for tomosynthesis imaging Authors: F. Sprenger a*, X. Calderon-Colon b, Y. Cheng a, K. Englestad a, J. Lu b, J. Maltz c, A. Paidi c, X. Qian b, D. Spronk a, S.

More information

Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film

Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 4, 2016 Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film Eric C. Lobb Department

More information

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook DIAGNOSTIC ACCREDITATION PROGRAM Radiology and CT Quality Control Procedures Workbook Quality Control Procedures Radiography/CR/DR Safety Code 35 Summary For more detail about each quality control (QC)

More information

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

specifications TrueBeam System

specifications TrueBeam System specifications TrueBeam System TrueBeam System s The TrueBeam system specifications in this document are identified as belonging to two categories, performance specifications and descriptive specifications.

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation Image Quality Review I: Basics and Image Quality TH-A-16A-1 Thursday 7:30AM - 9:30AM Room: 16A J. Anthony

More information

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

1. Patient size AEC. Large Patient High ma. Small Patient Low ma Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description

More information

T h e P h a n t o m L a b o r a t o r y

T h e P h a n t o m L a b o r a t o r y T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order

More information

Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA

Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Arthur Olch, PhD, FAAPM AAPM Spring Clinical Meeting, March 21, 2017 Or.. What Dose are the Patients Really Getting???

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

Mobius3D. Software based IMRT QA

Mobius3D. Software based IMRT QA Mobius3D Software based IMRT QA What is Mobius Medical Systems? Clinical Expertise Software Expertise Nathan Childress, Ph.D., Founder Eli Stevens, Chief Technical Officer Support Expertise Physicists

More information

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Gena M.A.H 1, Ahmed L.El-Attar 2, Elbadry M. Zahran 3, Hany El-Gamal

More information

The influence of bowtie filtration on cone-beam CT image quality

The influence of bowtie filtration on cone-beam CT image quality The influence of bowtie filtration on cone-beam CT image quality N. Mail Radiation Medicine Program, Princess Margaret Hospital, Toronto, Ontario M5G 2M9, Canada and Ontario Cancer Institute, University

More information

CT Scanner Dose Survey

CT Scanner Dose Survey CT Scanner Dose Survey Measurement Protocol Version 5.0 July 1997 Co-ordinated by ImPACT and The Medical Physics Department,, London SW17, UK. 0181-725-3366 CT Scanner Dose Survey: Measurement Protocol

More information

specifications TrueBeam STx System

specifications TrueBeam STx System specifications TrueBeam STx System TrueBeam STx s The TrueBeam STx system specifications in this document are identified as belonging to two categories, performance specifications and descriptive specifications.

More information

Sub-mm accuracy of accelerators: How manufacturers achieve it, how physicists verify it

Sub-mm accuracy of accelerators: How manufacturers achieve it, how physicists verify it Sub-mm accuracy of accelerators: How manufacturers achieve it, how physicists verify it Presented by Ivan A. Brezovich, PhD, at the 2015 Annual meeting of the SEAAPM, Raleigh, NC, April 24, 2015 Purpose

More information

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter MR-guided RT in Heidelberg Funded by the German Research Foundation (DFG) Heidelberg consortium received

More information

LINEAR ACCELERATOR. Buyer's Guide. Version 1.1

LINEAR ACCELERATOR. Buyer's Guide. Version 1.1 PRE-OWNED LINEAR ACCELERATOR Buyer's Guide Version 1.1 Pre-Owned Linear Accelerator Buyer's Guide TABLE OF CONTENTS Considerations For Buying A Used Linear Accelerator... 3 Linear Accelerators Overview...

More information

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 1, WINTER 2010 A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA Abdul Qadir Jangda,

More information

Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging

Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2008 Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging Dean

More information

X-RAD SmART (Small Animal RadioTherapy) IMAGE GUIDED BIOLOGICAL IRRADIATOR Technical Specifications (updated March 2015)

X-RAD SmART (Small Animal RadioTherapy) IMAGE GUIDED BIOLOGICAL IRRADIATOR Technical Specifications (updated March 2015) X-RAD SmART (Small Animal RadioTherapy) IMAGE GUIDED BIOLOGICAL IRRADIATOR Technical Specifications (updated March 2015) The X-RAD SmART Small Animal IGRT research system provides a high accuracy cone

More information

Aperture Based Inverse Planning AAPM Summer School 2003

Aperture Based Inverse Planning AAPM Summer School 2003 Aperture Based Inverse Planning AAPM Summer School 003 D.M. Shepard, M.A. Earl, Y. Xiao, C.X. Yu Acknowledgements Ziping Jiang Allen Li Shahid Naqvi James Galvin Di Yan Prowess, Inc. University of Maryland

More information

Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator

Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator Aleksei E. Zhdanov 1 and Leonid G. Dorosinskiy 1 Ural Federal University named after the first President of Russia B. N.

More information

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM FILMQA STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE Supports all major radiotherapy technologies! FilmQA is optimized for use with Gafchromic film products, including EBT2 and RTQA2. FILMQA helps

More information

Evaluation of a diode array for QA measurements on a helical tomotherapy unit

Evaluation of a diode array for QA measurements on a helical tomotherapy unit Evaluation of a diode array for QA measurements on a helical tomotherapy unit K. M. Langen, a S. L. Meeks, D. O. Poole, T. H. Wagner, T. R. Willoughby, O. A. Zeidan, and P. A. Kupelian Department of Radiation

More information

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array José A. Bencomo, * Geoffrey Ibbott, Seungsoo Lee, and Joao A. Borges Department of Radiation Physics.

More information

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

More information

Accuracy of SRS dose delivery using the TomoTherapy Hi-Art System

Accuracy of SRS dose delivery using the TomoTherapy Hi-Art System Louisiana State University LSU Digital Commons LSU Master's Theses Graduate School 2010 Accuracy of SRS dose delivery using the TomoTherapy Hi-Art System Catherine Leigh Batte Louisiana State University

More information

AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF DIGITAL RADIOGRAPHY IMAGING SYSTEMS

AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF DIGITAL RADIOGRAPHY IMAGING SYSTEMS International Workshop SMART MATERIALS, STRUCTURES & NDT in AEROSPACE Conference NDT in Canada 2011 2-4 November 2011, Montreal, Quebec, Canada AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF

More information

QUALITY CONTROL SOFTWARE FOR RADIOTHERAPY AND MEDICAL IMAGING

QUALITY CONTROL SOFTWARE FOR RADIOTHERAPY AND MEDICAL IMAGING 1 QUALITY CONTROL SOFTWARE FOR RADIOTHERAPY AND MEDICAL IMAGING The QUALIMAGIQ platform and its modules An optimal software solution for automatic analyses and traceability of quality controls in radiation

More information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information Ordering Information Please contact us if you have any questions or if you would like a quote or delivery schedule regarding the Catphan phantom. phone 800-525-1190, or 518-692-1190 fax 518-692-3329 mail

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 76-700 Digital Subtraction Angiography Phantom Users Manual March 2005 Manual No. 76-700-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-649 CDRH Fluoroscopic Phantom Users Manual March 2005 Manual No. 07-649-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

SUN NUCLEAR. EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm. corporation. Your Most Valuable QA and Dosimetry Tools

SUN NUCLEAR. EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm. corporation. Your Most Valuable QA and Dosimetry Tools EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm SUN NUCLEAR corporation Your Most Valuable QA and Dosimetry Tools introduction Pre-treatment dose QA is an important process required

More information

MOVING LASER SYSTEMS IN RT

MOVING LASER SYSTEMS IN RT MOVING LASER SYSTEMS IN RT ESSENTIAL FOR PRECISE PATIENT ALIGNMENT NEXT GENERATION LAP IS YOUR IDEAL PARTNER FOR YOUR EXTERNAL LASER SYSTEMS OUR MOVING LASER SYSTEMS ARE RECOMMENDED BY LEADING CT SCANNER

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017 A Fast Monolithic System for Proton Imaging Fritz DeJongh ProtonVDA Inc October 2017 Disclosures I am a cofounder and co-owner of ProtonVDA Inc We hold intellectual property rights on our proton imaging

More information

Quality Control of Full Field Digital Mammography Units

Quality Control of Full Field Digital Mammography Units Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of

More information

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology The Fundamentals of... Computed Tomography Computed Tomography (CT) systems use x-rays to produce images of slices through a patient s anatomy. Despite having lower spatial resolution than other x-ray

More information

Chapter 3. Material and Methods

Chapter 3. Material and Methods Material and Methods 3 Material and Methods 3.1 Amorphous Silicon flat panel-type detector imager - Elekta iviewgt The Electronic Portal Imaging Device (EPID) used on this investigationn is of type Elekta

More information

Thermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source

Thermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source Energy Level (ev) Multi-beam x-ray source array based on carbon nanotube field emission O. Zhou, JP Lu, X. Calderon-Colon, X. Qian, G. Yang, G. Cao, E. Gidcumb, A. Tucker, J. Shan University of North Carolina

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

Maximum Performance, Minimum Space

Maximum Performance, Minimum Space TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

Detector technology in simultaneous spectral imaging

Detector technology in simultaneous spectral imaging Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become

More information

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film With improvements in targeting and localization, stereotactic treatments have become prevalent. To

More information

2D, 3D CT Intervention, and CT Fluoroscopy

2D, 3D CT Intervention, and CT Fluoroscopy 2D, 3D CT Intervention, and CT Fluoroscopy SOMATOM Definition, Definition AS, Definition Flash Answers for life. Siemens CT Vision Siemens CT Vision The justification for the existence of the entire medical

More information

IMRT Delivery System QA. IMRT Dose Delivery. Acceptance testing. Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) MLC Alignment

IMRT Delivery System QA. IMRT Dose Delivery. Acceptance testing. Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) MLC Alignment 1 IMRT Delivery System Q Thomas LoSasso, PhD Memorial Sloan Kettering Cancer Center IMRT Dose Delivery cceptance testing Commissioning Quality assurance Verification Q Why: specific tests for IMRT? 2.

More information

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results NeuViz 16 Computed Tomography Elevating routine imaging for exceptional results Essence NeuViz 16 Raising the bar on clinical utility in routine imaging. Get more. More clinical information for patients.

More information

High Energy Digital Radiography & 3D-CT for Industrial Systems

High Energy Digital Radiography & 3D-CT for Industrial Systems DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France High Energy Digital Radiography & 3D-CT for Industrial Systems Non-Destructive

More information

Four-dimensional in vivo dosimetry by dose reconstruction using continuous EPID images and phase sorting method. JiHyung Yoon.

Four-dimensional in vivo dosimetry by dose reconstruction using continuous EPID images and phase sorting method. JiHyung Yoon. Four-dimensional in vivo dosimetry by dose reconstruction using continuous EPID images and phase sorting method Director of Dissertation: Dr. Jae Won Jung Major Department: Physics by JiHyung Yoon July,

More information

Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy

Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy Thomas LoSasso, a) Chen-Shou Chui, and C. Clifton Ling Department

More information

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National

More information

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images Aiping Ding, Bin Han, Lei Wang, Lei Xing Department of Radiation Oncology, Stanford University School of

More information

Truly flexible to meet your clinical needs

Truly flexible to meet your clinical needs Truly flexible to meet your clinical needs 2 Adapting to meet your needs Flexible Fast and responsive Excellent image quality Designed with ergonomic efficiency Equipped with dose management tools 3 Three

More information

Confirmation, refinement, and extension of a study in intrafraction motion interplay with sliding jaw motion

Confirmation, refinement, and extension of a study in intrafraction motion interplay with sliding jaw motion Confirmation, refinement, and extension of a study in intrafraction motion interplay with sliding jaw motion Michael W. Kissick, a Sarah A. Boswell, Robert Jeraj, and T. Rockwell Mackie Department of Medical

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS Veterinary Science Preparatory Training for the Veterinary Assistant Floron C. Faries, Jr., DVM, MS Radiology Floron C. Faries, Jr., DVM, MS Objectives Determine the appropriate machine settings for making

More information

Wide beam CT dosimetry. Elly Castellano

Wide beam CT dosimetry. Elly Castellano Wide beam CT dosimetry Elly Castellano Outline revision: CT dose indices wide-beam CT: the end of the road for CTDI? the IEC rescue plan for CTDI 100 the american way AAPM report 111 better estimates of

More information

v tome x m microfocus CT

v tome x m microfocus CT GE Inspection Technologies v tome x m microfocus CT Uniting premium 3D metrology and inspection with quality and speed. gemeasurement.com/ct x plore precision CT line Inspect with precision, power, and

More information

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT

More information

University of Wollongong. Research Online

University of Wollongong. Research Online University of Wollongong Research Online University of Wollongong Thesis Collection 1954-2016 University of Wollongong Thesis Collections 2013 Validation of a general model for Intensity Modulated Radiation

More information

Sensitivity study of an automated system for daily patient QA using EPID exit dose images

Sensitivity study of an automated system for daily patient QA using EPID exit dose images Received: 27 June 2017 Revised: 8 December 2017 Accepted: 27 January 2018 DOI: 10.1002/acm2.12303 RADIATION ONCOLOGY PHYSICS Sensitivity study of an automated system for daily patient QA using EPID exit

More information

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology CODE: RADT 156 INSTITUTE: Health Science TITLE: Equipment Operation I DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of equipment operation and maintenance of radiographic

More information

Volumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA

Volumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA Volumetric Modulated Arc Therapy David Shepard Swedish Cancer Institute Seattle, WA Disclaimer Our VMAT work has been sponsored in part by Elekta. Outline David Shepard VMAT Basics and VMAT Plan Quality

More information

A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine

A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine Iwein Van de Vondel, 1 Koen Tournel, 1 Dirk

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information