IMRT Delivery System QA. IMRT Dose Delivery. Acceptance testing. Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) MLC Alignment

Size: px
Start display at page:

Download "IMRT Delivery System QA. IMRT Dose Delivery. Acceptance testing. Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) MLC Alignment"

Transcription

1 1 IMRT Delivery System Q Thomas LoSasso, PhD Memorial Sloan Kettering Cancer Center IMRT Dose Delivery cceptance testing Commissioning Quality assurance Verification Q Why: specific tests for IMRT? 2. Gap error Dose error ccuracy of leaf positioning (gaps) Static IMRT Leaf affects borders only affects dose within target position 1-2 mm OK.2 mm OK % Dose error Range of gap width.5 Gap error (mm) Nominal gap (cm) MLC lignment cceptance testing MLC alignment Film Field Field MLC leaf / gap calibration Timing errors 1 1 eam characteristics at small MU Film 2 6 Field 9 1 Field

2 2 Doubly - exposed films MLCtable.txt corrections Film 1 Film 2 perfect alignment Correction (mm) MLCtable measurement "precise" Varian.5 mm misalignments Distance from central axis (cm) Gap measurement - mechanical DMLC - Symmetry and Flatness Gap at isocenter upper lower.95.9 open 5mm 5mm/open cm 12 cm DMLC - Symmetry and Flatness Gap correction - measurement open 5mm 5mm/open transmission gap only MV.2 y =.9945x R 2 = MV.2 y =.9954x R 2 = Nominal gap (mm)

3 3 Effective gap offset variation Mark 1 Mark 2 Millenium Phelps X 15X 6X 15X 6X 15X 6X 6X 15X Off-axis variation - "cupping" Millennium 12 Effective gap offset (mm) CX 5cm off 1cm off Series4 Series5 Series6 +/- 1% 15 MV 6 MV Eff gap offset iso) 2 4% Dose Off-axis variation - corrected Eff gap offset 1.2 S =.47 mm with cupping correction Eff gap offset 1.2 S =.27 mm ~1% dose with individual leaf corrections Frequency of deviations (%) # Global logged - prescribed leaf iso (mm) checks RMS Field Room ngle < LO Log file Leaf sequence file dose distribution LPO Sup P Post nt RPO RO cm Inf Overlay jaws Dose differences (log-dva) # checks = # moving leaves * OT / 55 msec

4 4 Difference (%) Timing error SMLC (Varian) MU/seg 16 MU/seg 25 MU/seg 4 MU/seg -5 Delivered with dose mode at 4MU/min 1 MU/seg (CL23, 6MV, 1 SSD, cm depth) MU/seg # Segment 16 MU/seg 25 MU/seg 4 mu/min Courtesy of Ping Xia, UCSF Timing error SMLC (Varian) Courtesy of Gary Ezzell and the Journal of pplied Clinical Medical Physics. Overlay 11 MU Dose differences (log-dva) 11 MU 55 MU 2X Timing / communication errors Distance from axis (mm) 28 MU 4X 14 MU 8X MU (%) 15 MU Timing / communication errors Distance from axis (mm) Distance from axis (mm) Comparison of prescribed and delivered doses Log file Leaf sequence file dose distribution 15 MU prescribed / 6 MU delivered F G F G MU (%) 15 MU MU (%) 6 MU D Overlay plan log C E D C Difference (log - plan) E

5 5 Relative dose (corrected to 15 MU) MU decrease dose errors increase Timing error Is there an analogous problem for DMLC? C D E F G Prescribed MU = MU delivered Trailing leaves (-side) compensate for the leading leaves (-side) Dose (MU) error = position error () position error leaf speed () leaf speed () However, some leaf sequencers actively utilize the beam hold-off to control leaf speed eam Stability: Flatness, Symmetry Stability of flatness and symmetry affects dose rate for small fields directed off the central axis. For an open 2x2cm 2 field, measure profiles for irradiations ranging from 1 to 1 MU. Sun Nuclear Profiler (46 diodes, 1 profiles/sec). Flatness is +/-3% if more than 5MU delivered. Symmetry +/-3% if more than 4MU delivered. Courtesy of M. Sharpe, William eaumont Hospital Transmission Transmission profiles across leaves Mark 1 (445) Millenium (444) average average Mark 2 (245) Mark 2 (442) average average Respiratory gating Parallel Splitting large fields Y Perpendicular Leaf motion X

6 6 Routine Q Quality assurance MLC Q MLC specific Mechanical stability ccuracy of leaf positioning (gaps) Verification Patient / field specific Tx plan is transferred and delivered correctly Motor fatigue MLC calibration acklash Grease Film pattern (semi-weekly) Dosimetry (monthly) Dosimetry (monthly) lcohol (semi-annually) Motors Motor replaced replacement history MLC motors replaced # Jan- 92 Jan- 93 MLC installed 1st MLC Tx Room Jan- Jan- Jan- Jan- Jan Date 1st dmlc Tx Jan- Jan- Jan Jan- Jan a Field Size Defined by MLC at 1 cm SD by Light Field: Leaf / carriage positioning vs off-axis distance Date: Tolerance:.5 mm Initials: Digital setting (cm) carriage carriage leaves leaves Note: If leaf positions exceed tolerance, reinitialize and recheck. If this does not correct the problem, verify that the correction parameters in files MLCTLE.TXT and MLCXCL.TXT have not changed. 4 cm

7 7 1 mm bands Leaf positioning errors introduced 17b DMLC Output (Gap width) Check 15 MV x-rays, File: DMLC_Q2.dva, Dose = 1 MU, Dose rate = 24 MU/min SD: 1cm, Depth = 3.1cm, ackup = 1cm, C.S. = 1x1cm2 Tolerance: (The ratios should be within 2% (<.2 mm gap) of the annual calibration.) -.5 mm -.2 mm Measure the output for the scanned field before and after reinitializing the MLC. Measure the output for a 1x1cm2 reference field size defined by the jaws (MLC retracted). Calculate the ratio of the scanned field to the reference field outputs. Reference 1x1 DMLC - before reinitialization DMLC - after reinitialization Reference 1x1 +.2 mm +.5 mm verage Ratio 17c eam Holdoff Functionality Date: Initials Same setup as 17b, Dose = 5 MU Record the output and the Tx time. Check if the holdoff is being applied. Reading Functioning nnual DMLC / 1x DMLC Output vs Time Jan-98 Jan-99 Jan- Jan-1 Jan-2 Jan-3 Jan-4 Date.2 mm ~1% 17d Output vs Gantry and Collimator ngles Date: Initials 15 MV x-rays, File: DMLC_Q2.dva, Dose = 1 MU, Dose rate = 24 MU/min SD: 1cm, Spokas in air w/buildup, C.S. = 1x1cm2 Tolerance: (The ratios should be within 3% (<.2 mm gap) of each other.) Position a cylindrical chamber in air at isocenter. For each gantry angle, measure the open field and the scanned field outputs. t gantry angles of 9 and 27, also measure output for collimator angles of 9. Calculate the ratio of the scanned field to the reference field outputs. Collimator angle Gantry Reference DMLC DMLC / Reference angle DMLC output vs gantry and collimator angle Mark 1 (445) Millennium (444) DMLC output vs gantry angle (diode array) 3 DMLC / 1x Mark 2 (245) Millennium (241) ±.2 mm (± 1%) mm Year

8 8 Summary 1. cceptance testing MLC capable of IMRT delivery Improved calculations needed for IMRT fields 2. Commissioning MLC modeled accurately in TPS 3. Routine MLC Q leaves and carriages - mechanics 4. Patient specific Q independent verification IMRT Dose ccuracy 2 cm Plan Film Overlay 2 cm Film - Plan Film - Plan Difference ~25% Dosimetry (film plan) T&G Source MLC scatter Lung P field Corrections (plan plan) x1 Tongue and Groove

9 9 Relative dose Tongue and groove Distance from CX (cm) -1 cm cm 1 cm 2 cm DMLC / 1x1 (norm to -) DMLC output vs gantry and collimator angle adjusted bearings replaced bearings adjusted bearings adjusted bearings Mark 1 (445) adjusted drive belts adjusted bearings

Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy

Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy Thomas LoSasso, a) Chen-Shou Chui, and C. Clifton Ling Department

More information

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017 The Current State of EPID-Based Linear Accelerator Quality Assurance Timothy Ritter, PhD, DABR, FAAPM 1 Disclosures Employed by the Veterans Health Administration Faculty appointment with the University

More information

IQM Detector Characteristics: Signal reproducibility

IQM Detector Characteristics: Signal reproducibility The Integral Quality Monitor (IQM) System is a real-time beam verification system that monitors the accuracy of radiation delivery throughout each patient treatment without any user interaction. IQM continuously

More information

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System. Use of EPIDs for Non-Routine Linac QA Bin Cai PhD Disclosure Parts of this project received support from Varian Medical System. Learning Objectives Learn the recent development of EPID based Non-routine

More information

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators Implementing a new digital medical accelerator John Wong Johns Hopkins University AAPM, Austin, 2014 Acknowledgements Yin Zhang, Ken Wang, Kai Ding (Commissioning - JHU) Esteban Velarde, Joe Moore (QA

More information

PHYSICS QUESTIONNAIRE FORM

PHYSICS QUESTIONNAIRE FORM PHYSICS QUESTIONNAIRE FORM Institution Name: Date: Contact Information (name, address, phone, fax, email): Physicist: Radiation Oncologist: Dosimetrist (if applicable): Study Coordinator (if applicable):

More information

Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System. Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX.

Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System. Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX. Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX. 1 Acknowledgements Collaborators: Jose Bencomo, Rafael. M. Landrove, Peter

More information

SUN NUCLEAR. EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm. corporation. Your Most Valuable QA and Dosimetry Tools

SUN NUCLEAR. EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm. corporation. Your Most Valuable QA and Dosimetry Tools EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm SUN NUCLEAR corporation Your Most Valuable QA and Dosimetry Tools introduction Pre-treatment dose QA is an important process required

More information

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests Acknowledgements David Shepard, Ph.D. Daliang Cao, Ph.D. Muhammad K. N. Afghan, Ph.D. Jinsong Ye, M.S. Tony P. Wong, Ph.D. Fan Chen, Ph.D. Min Rao, Ph.D. Vivek Mehta, M.D. Igor Gomola, Ph.D. David Housley

More information

Volumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA

Volumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA Volumetric Modulated Arc Therapy David Shepard Swedish Cancer Institute Seattle, WA Disclaimer Our VMAT work has been sponsored in part by Elekta. Outline David Shepard VMAT Basics and VMAT Plan Quality

More information

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

3D Diode Array Commissioning: Building Confidence in 3D QA Technology 3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose

More information

Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA

Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Arthur Olch, PhD, FAAPM AAPM Spring Clinical Meeting, March 21, 2017 Or.. What Dose are the Patients Really Getting???

More information

Evaluation of a diode array for QA measurements on a helical tomotherapy unit

Evaluation of a diode array for QA measurements on a helical tomotherapy unit Evaluation of a diode array for QA measurements on a helical tomotherapy unit K. M. Langen, a S. L. Meeks, D. O. Poole, T. H. Wagner, T. R. Willoughby, O. A. Zeidan, and P. A. Kupelian Department of Radiation

More information

ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung

ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung Treffen des Arbeitskreises IMRT der DGMP Würzburg, 26 + 27.03.2009 Salih Arican Sun Nuclear Corporation QA Challenge for Rotational Beams

More information

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film With improvements in targeting and localization, stereotactic treatments have become prevalent. To

More information

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev. 1.0 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE DoseLab users may reference the following instructions to perform Tomotherapy Quality Assurance tests as recommended

More information

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Gena M.A.H 1, Ahmed L.El-Attar 2, Elbadry M. Zahran 3, Hany El-Gamal

More information

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film As clinics strive to treat with the smallest field possible, the demand for SRS/SBRT QA grows as well.

More information

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter MR-guided RT in Heidelberg Funded by the German Research Foundation (DFG) Heidelberg consortium received

More information

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Indra J. Das, PhD, FACR Department of Radiation Oncology Indiana University of School of Medicine & Midwest Proton Radiation

More information

Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging

Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2008 Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging Dean

More information

Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector

Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector ORIGINAL ARTICLES Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector Gopiraj ANNAMALAI 1, Ramasubramanian VELAYUDHAM 2 ABSTRACT Received: 7.07.2009 Accepted: 2.11.2009

More information

ArcCHECKTM. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools. VMAT RapidArc TomoTherapy Pinnacle 3 SmartArc Conventional IMRT

ArcCHECKTM. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools. VMAT RapidArc TomoTherapy Pinnacle 3 SmartArc Conventional IMRT TM The Ultimate 4D QA Solution A 4D isotropical cylindrical detector array for arc delivery QA and Dosimetry U.S.Patent No. 8,044,359 What is? The world s first true 4D detector array The world s first

More information

Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film

Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 4, 2016 Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film Eric C. Lobb Department

More information

Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator

Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator Aleksei E. Zhdanov 1 and Leonid G. Dorosinskiy 1 Ural Federal University named after the first President of Russia B. N.

More information

8/3/2016. The EPID Strikes Back. Novel Applications for Current EPID Technology. Joerg Rottmann, PhD. Disclosures and acknowledgements

8/3/2016. The EPID Strikes Back. Novel Applications for Current EPID Technology. Joerg Rottmann, PhD. Disclosures and acknowledgements The EPID Strikes Back Joerg Rottmann Brigham and Women s Hospital / Dana-Farber Cancer Institute Harvard Medical School Disclosures and acknowledgements Disclosures Varian MRA grant Acknowledgements Boston

More information

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images Aiping Ding, Bin Han, Lei Wang, Lei Xing Department of Radiation Oncology, Stanford University School of

More information

ArcCHECK. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools

ArcCHECK. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools ArcCHECK The Ultimate 4D QA Solution A 4D isotropical cylindrical detector array for arc delivery QA and Dosimetry U.S.Patent No. 8,044,359; 6,125,335 Compatible with: FFF Beams VMAT RapidArc TomoTherapy

More information

specifications TrueBeam STx System

specifications TrueBeam STx System specifications TrueBeam STx System TrueBeam STx s The TrueBeam STx system specifications in this document are identified as belonging to two categories, performance specifications and descriptive specifications.

More information

Mobius3D. Software based IMRT QA

Mobius3D. Software based IMRT QA Mobius3D Software based IMRT QA What is Mobius Medical Systems? Clinical Expertise Software Expertise Nathan Childress, Ph.D., Founder Eli Stevens, Chief Technical Officer Support Expertise Physicists

More information

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM FILMQA STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE Supports all major radiotherapy technologies! FilmQA is optimized for use with Gafchromic film products, including EBT2 and RTQA2. FILMQA helps

More information

A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine

A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine Iwein Van de Vondel, 1 Koen Tournel, 1 Dirk

More information

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Huaiqun Guan,

More information

The Ultimate 4D QA Solution A 4D isotropic cylindrical detector array for arc delivery QA and Dosimetry.

The Ultimate 4D QA Solution A 4D isotropic cylindrical detector array for arc delivery QA and Dosimetry. The Ultimate 4D QA Solution A 4D isotropic cylindrical detector array for arc delivery QA and Dosimetry. U.S.Patent No. 8,044,359; 6,125,335 Your Most Valuable QA and Dosimetry Tools 2 Y o u r M o s t

More information

Monica Kishore. Medical Physics Graduate Program Duke University. Approved: Jennifer O Daniel, Co-Supervisor. Fang-Fang Yin, Co-Supervisor

Monica Kishore. Medical Physics Graduate Program Duke University. Approved: Jennifer O Daniel, Co-Supervisor. Fang-Fang Yin, Co-Supervisor Accuracy of Planar Dosimetry for Volumetric Modulated Arc Therapy Quality Assurance by Monica Kishore Medical Physics Graduate Program Duke University Date: Approved: Jennifer O Daniel, Co-Supervisor Fang-Fang

More information

COMPREHENSIVE TG-142 IMAGING AND MACHINE QA

COMPREHENSIVE TG-142 IMAGING AND MACHINE QA QA SOFTWARE COMPREHENSIVE TG-142 IMAGING AND MACHINE QA Automate the analysis of over thirty TG-142 recommended QA tasks The rapid progress of Radiation Therapy has created the need for Quality Assurance

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

specifications TrueBeam System

specifications TrueBeam System specifications TrueBeam System TrueBeam System s The TrueBeam system specifications in this document are identified as belonging to two categories, performance specifications and descriptive specifications.

More information

Table 1: Available X-ray Beam Energy Combinations (MV) 6 10/ Yes Yes 6 16/15 6 Yes No 6 23/18 6 Yes No 6 25/20 6 Yes No

Table 1: Available X-ray Beam Energy Combinations (MV) 6 10/ Yes Yes 6 16/15 6 Yes No 6 23/18 6 Yes No 6 25/20 6 Yes No SPECIFICATIONS Introduction This specification sheet provides information for the Trilogy linear accelerators. 1.0 Photon Beams 1.1 Energy: Three photon beams may be selected in accordance with the beam

More information

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications I. SCOPE The protocol applies to GafChromic EBT3P and EBT3+P films exposed in GafChromic QuiCk Phantom

More information

Characterization of an in vivo diode dosimetry system for clinical use

Characterization of an in vivo diode dosimetry system for clinical use JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 2, SPRING 2003 Characterization of an in vivo diode dosimetry system for clinical use Kai Huang, 1, * William S. Bice, Jr., 2, and Oscar Hidalgo-Salvatierra

More information

Sub-mm accuracy of accelerators: How manufacturers achieve it, how physicists verify it

Sub-mm accuracy of accelerators: How manufacturers achieve it, how physicists verify it Sub-mm accuracy of accelerators: How manufacturers achieve it, how physicists verify it Presented by Ivan A. Brezovich, PhD, at the 2015 Annual meeting of the SEAAPM, Raleigh, NC, April 24, 2015 Purpose

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

Accuracy of rapid radiographic film calibration for intensity-modulated radiation therapy verification

Accuracy of rapid radiographic film calibration for intensity-modulated radiation therapy verification JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 2, SPRING 2006 Accuracy of rapid radiographic film calibration for intensity-modulated radiation therapy verification Ravi Kulasekere, a Jean

More information

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array José A. Bencomo, * Geoffrey Ibbott, Seungsoo Lee, and Joao A. Borges Department of Radiation Physics.

More information

A proposed method for linear accelerator photon beam steering using EPID

A proposed method for linear accelerator photon beam steering using EPID Received: 13 January 2018 Revised: 11 May 2018 Accepted: 29 June 2018 DOI: 10.1002/acm2.12419 RADIATION ONCOLOGY PHYSICS A proposed method for linear accelerator photon beam steering using EPID Michael

More information

Analysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose

Analysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose J. Radiat. Res., 53, 301 305 (2012) Analysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose Katsumi SHIMA 1,2, Kunihiko TATEOKA 1 *, Yuichi SAITOH 1,2, Junji SUZUKI

More information

specifications The Complete Radiation Oncology Solution

specifications The Complete Radiation Oncology Solution specifications The Complete Radiation Oncology Solution Specifications This specification sheet provides information for the UNIQUE radiation oncology solution and in particular the UNIQUE linear accelerator.

More information

Aperture Based Inverse Planning AAPM Summer School 2003

Aperture Based Inverse Planning AAPM Summer School 2003 Aperture Based Inverse Planning AAPM Summer School 003 D.M. Shepard, M.A. Earl, Y. Xiao, C.X. Yu Acknowledgements Ziping Jiang Allen Li Shahid Naqvi James Galvin Di Yan Prowess, Inc. University of Maryland

More information

Impact of energy variation on Cone Ratio, PDD10, TMR20 10 and IMRT doses for flattening filter free (FFF) beam of TomoTherapy Hi-Art TM machines

Impact of energy variation on Cone Ratio, PDD10, TMR20 10 and IMRT doses for flattening filter free (FFF) beam of TomoTherapy Hi-Art TM machines JBUON 2014; 19(4): 1105-1110 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com ORIGINAL ARTICLE Impact of energy variation on Cone Ratio, PDD10, TMR20 10 and IMRT

More information

Comparison of measured Varian Clinac 21EX and TrueBeam accelerator electron field characteristics

Comparison of measured Varian Clinac 21EX and TrueBeam accelerator electron field characteristics JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Comparison of measured Varian Clinac 21EX and TrueBeam accelerator electron field characteristics Samantha A.M. Lloyd, 1a Sergei Zavgorodni,

More information

A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery

A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery Paul A. Jursinic a) Medical College of Wisconsin, Radiation Oncology Department, Milwaukee, Wisconsin

More information

Monte Carlo study on a new concept of a scanning photon beam system for IMRT

Monte Carlo study on a new concept of a scanning photon beam system for IMRT NUKLEONIKA 2011;56(4):291 297 ORIGINAL PAPER Monte Carlo study on a new concept of a scanning photon beam system for IMRT Anna M. Wysocka-Rabin, Günter H. Hartmann Abstract. Intensity-modulated radiation

More information

Fast Real Time Calibration of a Varian 2100 (High Energy) series LINAC

Fast Real Time Calibration of a Varian 2100 (High Energy) series LINAC Fast Real Time Calibration of a Varian 2100 (High Energy) series LINAC Kenneth Chu, Ph.D., D.A.B.R., P.Eng. Marquette General Hospital Marquette, MI And Shidong Tong, Ph.D., D.A.B.R. Inova Alexandria Hospital

More information

specifications Novalis Tx image-guided radiosurgery linear accelerator

specifications Novalis Tx image-guided radiosurgery linear accelerator specifications Novalis Tx image-guided radiosurgery linear accelerator Specifications Introduction This specification sheet provides information for the Novalis Tx image-guided radiosurgery linear accelerator.

More information

The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response

The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response Iran. J. Radiat. Res., 2005; 3 (1): 3-10 The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response M. Mohammadi 1,2,3* and E. Bezak 1,2

More information

Stability of the Helical TomoTherapy Hi Art II detector for treatment beam irradiations

Stability of the Helical TomoTherapy Hi Art II detector for treatment beam irradiations JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 6, 2014 Stability of the Helical TomoTherapy Hi Art II detector for treatment beam irradiations Karin Schombourg, François Bochud, Raphaël

More information

UNIQUE. The Complete Radiation Oncology Solution SPECIFICATIONS

UNIQUE. The Complete Radiation Oncology Solution SPECIFICATIONS UNIQUE The Complete Radiation Oncology Solution SPECIFICATIONS UNIQUE SPECIFICATIONS I 2 UNIQUE SPECIFICATIONS I 3 Specifications This specification sheet provides information for the UNIQUE radiation

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

A feasibility study of using conventional jaws to deliver IMRT plans in the treatment of prostate cancer *

A feasibility study of using conventional jaws to deliver IMRT plans in the treatment of prostate cancer * IOP PUBLISHING Phys. Med. Biol. 52 (7) 2147 2156 PHYSICS IN MEDICINE AND BIOLOGY doi:1.188/31-9155/52/8/7 A feasibility study of using conventional jaws to deliver IMRT plans in the treatment of prostate

More information

Commissioning an Elekta Versa HD linear accelerator

Commissioning an Elekta Versa HD linear accelerator JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 Commissioning an Elekta Versa HD linear accelerator Ganesh Narayanasamy, 1,2 Daniel Saenz, 1 Wilbert Cruz, 1,3 Chul S. Ha, 1 Niko

More information

Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification

Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification L. N. McDermott, M. Wendling, B. van Asselen, J. Stroom, J.-J. Sonke, M. van Herk, and B. J. Mijnheer a Department

More information

Sensitivity study of an automated system for daily patient QA using EPID exit dose images

Sensitivity study of an automated system for daily patient QA using EPID exit dose images Received: 27 June 2017 Revised: 8 December 2017 Accepted: 27 January 2018 DOI: 10.1002/acm2.12303 RADIATION ONCOLOGY PHYSICS Sensitivity study of an automated system for daily patient QA using EPID exit

More information

Initial setup and subsequent temporal position monitoring using implanted RF transponders

Initial setup and subsequent temporal position monitoring using implanted RF transponders Initial setup and subsequent temporal position monitoring using implanted RF transponders James Balter, Ph.D. University of Michigan Has financial interest in Calypso Medical Technologies Acknowledgements

More information

Initial Experience with a Commercial System for Volumetric Analysis of Patient Specific QA. Katja Langen Mariana Guerrero Shifeng Chen Shh..

Initial Experience with a Commercial System for Volumetric Analysis of Patient Specific QA. Katja Langen Mariana Guerrero Shifeng Chen Shh.. Initial Experience with a Commercial System for Volumetric Analysis of Patient Specific QA Katja Langen Mariana Guerrero Shifeng Chen Shh..Mobius3D ACKOWLEDGEMENTS Dr. Katja Langen Dr. Shifeng Chen Dr.

More information

Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method

Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method Markus Wendling, Robert J. W. Louwe, a Leah N. McDermott, Jan-Jakob Sonke, Marcel van Herk, and Ben J. Mijnheer

More information

Four-dimensional in vivo dosimetry by dose reconstruction using continuous EPID images and phase sorting method. JiHyung Yoon.

Four-dimensional in vivo dosimetry by dose reconstruction using continuous EPID images and phase sorting method. JiHyung Yoon. Four-dimensional in vivo dosimetry by dose reconstruction using continuous EPID images and phase sorting method Director of Dissertation: Dr. Jae Won Jung Major Department: Physics by JiHyung Yoon July,

More information

Characterization and evaluation of an integrated quality monitoring system for online quality assurance of external beam radiation therapy

Characterization and evaluation of an integrated quality monitoring system for online quality assurance of external beam radiation therapy Received: 5 January 1 Accepted: 1 September 1 DOI: 1.1/acm.11 RADIATION ONCOLOGY PHYSICS Characterization and evaluation of an integrated quality monitoring system for online quality assurance of external

More information

Dosimetric IMRT verification with a flat-panel EPID

Dosimetric IMRT verification with a flat-panel EPID Dosimetric IMRT verification with a flat-panel EPID B. Warkentin Department of Medical Physics, Cross Cancer Institute and Department of Physics, University of Alberta, 11 University Avenue, Edmonton,

More information

Gantry angle determination during arc IMRT: evaluation of a simple EPID-based technique and two commercial inclinometers

Gantry angle determination during arc IMRT: evaluation of a simple EPID-based technique and two commercial inclinometers JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 6, 2012 Gantry angle determination during arc IMRT: evaluation of a simple EPID-based technique and two commercial inclinometers Pejman Rowshanfarzad,

More information

TOMOTHERAPY H SERIES TomoH, TomoHD and TomoHDA Systems Technical Specifications

TOMOTHERAPY H SERIES TomoH, TomoHD and TomoHDA Systems Technical Specifications TOMOTHERAPY H SERIES TomoH, TomoHD and TomoHDA Systems Technical Specifications The TomoTherapy H Series is designed to treat the entire spectrum of radiation therapy patients with enhanced speed, precise

More information

Beam Delivery Techniques: Passive Scattering Proton Beams. Zuofeng Li and Roelf Slopsema University of Florida Proton Therapy Institute

Beam Delivery Techniques: Passive Scattering Proton Beams. Zuofeng Li and Roelf Slopsema University of Florida Proton Therapy Institute Beam Delivery Techniques: Passive Scattering Proton Beams Zuofeng Li and Roelf Slopsema University of Florida Proton Therapy Institute 1 Disclaimer UFPTI uses IBA Proton Therapy System Proton Beam: production,

More information

HITACHI Proton Therapy System with Spot Scanning

HITACHI Proton Therapy System with Spot Scanning Workshop on Hadron Therapy of Cancer 27 th April, Erice, Sicily, Italy HITACHI Proton Therapy System with Spot Scanning Kazuo Hiramoto Energy & Environmental Systems Laboratory, Hitachi, Ltd. Contents

More information

Using the frame averaging of as500 EPID for IMRT verification

Using the frame averaging of as500 EPID for IMRT verification JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 Using the frame averaging of as500 EPID for IMRT verification J. Chang* and C. C. Ling Medical Physics Department, Memorial Sloan

More information

CyberKnife Iris Beam QA using Fluence Divergence

CyberKnife Iris Beam QA using Fluence Divergence CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery

More information

Beam Production, Characteristics and Shaping

Beam Production, Characteristics and Shaping Beam Production, Characteristics and Shaping Dr. Manfred Sassowsky Outline X-ray production 60 Co units Linear Accelerators Beam characteristics Beam shaping Beam Production, Characteristics and Shaping

More information

Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland

Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland Center for proton therapy - PSI GANTRY DESIGN AND RECENT EXPERIENCE AT PSI 2nd Workshop on Hadron Beam Therapy of Cancer ERICE May 23 th,

More information

Dose-response characteristics of an amorphous silicon EPID

Dose-response characteristics of an amorphous silicon EPID Dose-response characteristics of an amorphous silicon EPID Peter Winkler a Division of Medical Radiation Physics, Department of Radiotherapy and Radiobiology, Medical University of Vienna, Waehringer Guertel

More information

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion Leksell Gamma Knife Perfexion QA Considerations Paula L. Petti, Ph.D. Taylor McAdam Bell Neuroscience Institute Washington Hospital Healthcare System Fremont, CA 1 QA for LGK Perfexion : : Follow NRC licensing

More information

Nuclear Associates , , CT Head and Body Dose Phantom

Nuclear Associates , , CT Head and Body Dose Phantom Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING 1 QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING QualiFormeD Phantoms A selection of test objects facilitating regulatory quality controls in radiation therapy and medical imaging Practical,

More information

Total body irradiation dose optimization based on radiological depth

Total body irradiation dose optimization based on radiological depth JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 3, 2012 Total body irradiation dose optimization based on radiological depth Amjad Hussain, 1,3a Peter Dunscombe, 1,2,3 J. Eduardo Villarreal-

More information

Gantry design and experience at PSI

Gantry design and experience at PSI Gantry design and experience at PSI Eros Pedroni for the R&D Technology Team Center for Proton Therapy Paul Scherrer Institute Villigen-PSI SWITZERLAND Workshop on Hadron Beam Therapy of Cancer Erice April

More information

Ludlum Medical Physics

Ludlum Medical Physics Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to

More information

Design and Testing of Indigenous Cost Effective Three Dimensional Radiation Field Analyser (3D RFA)

Design and Testing of Indigenous Cost Effective Three Dimensional Radiation Field Analyser (3D RFA) Technology in Cancer Research and Treatment ISSN 1533-0346 Volume 13, Number 3, June 2014 Adenine Press (2014) Design and Testing of Indigenous Cost Effective Three Dimensional Radiation Field Analyser

More information

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS 7:28-14.1 Scope (a) This subchapter covers therapeutic installations used in the healing arts. These therapeutic installations include x-ray, accelerator and teletherapy

More information

Characterization of an in vivo diode dosimetry system for clinical use

Characterization of an in vivo diode dosimetry system for clinical use Louisiana State University LSU Digital Commons LSU Master's Theses Graduate School 2002 Characterization of an in vivo diode dosimetry system for clinical use Kai Huang Louisiana State University and Agricultural

More information

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 1, WINTER 2010 A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA Abdul Qadir Jangda,

More information

Department of Physics, State University of New York at Buffalo, Buffalo NY, USA

Department of Physics, State University of New York at Buffalo, Buffalo NY, USA 124 research article A fully electronic intensity-modulated radiation therapy quality assurance (IMRT QA) process implemented in a network comprised of independent treatment planning, record and verify,

More information

of sufficient quality and quantity

of sufficient quality and quantity of sufficient quality and quantity The patient s body attenuates the beam as it passes though the body More energy is deposited in organs located near the entry of the beam than near the exit of the beam

More information

Conflict Disclosure. Rotational IMRT. Arc therapy. Dynamic Arc therapy. Intensity Modulated Arc Therapy Principles and Perspectives

Conflict Disclosure. Rotational IMRT. Arc therapy. Dynamic Arc therapy. Intensity Modulated Arc Therapy Principles and Perspectives Intensity Modulated Arc Therapy Principles and Perspectives Cedric Yu University of Maryland Conflict Disclosure Advisory Council on Advanced Treatment Delivery, Varian Medical Systems, Inc. Patent License:

More information

Nathan Childress, Ph.D., DABR

Nathan Childress, Ph.D., DABR Nathan Childress, Ph.D., DABR Introduction TG-142 is a comprehensive QA protocol Covers nearly every aspect of machine and safety QA Recommends quantitative results Recommends high testing frequencies

More information

X3D in Radiation Therapy Procedure Planning. Felix G. Hamza-Lup, Ph.D. Computer Science Armstrong Atlantic State University Savannah, Georgia USA

X3D in Radiation Therapy Procedure Planning. Felix G. Hamza-Lup, Ph.D. Computer Science Armstrong Atlantic State University Savannah, Georgia USA X3D in Radiation Therapy Procedure Planning Felix G. Hamza-Lup, Ph.D. Computer Science Armstrong Atlantic State University Savannah, Georgia USA Outline 1. What is radiation therapy? 2. Treatment planning

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

Digital radiography (DR) post processing techniques for pediatric radiology

Digital radiography (DR) post processing techniques for pediatric radiology Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions

More information

University of Wollongong. Research Online

University of Wollongong. Research Online University of Wollongong Research Online University of Wollongong Thesis Collection 1954-2016 University of Wollongong Thesis Collections 2013 Validation of a general model for Intensity Modulated Radiation

More information

State of the Art Film Dosimetry

State of the Art Film Dosimetry State of the Art Film Dosimetry Micke A., Lewis D. Advanced Materials Ashland proprietary technology, patents pending Film Dosimetry Radiochromic Film EBT2/EBT3 One-Scan Protocol Multi-channel Film Dosimetry

More information

The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 BUSINESS UNIT OF ISP

The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 BUSINESS UNIT OF ISP The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 GAFCHROMIC EBT dosimetry film Designed and optimized for ALL RT procedures Can be cut into

More information

Phantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging

Phantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging in partnership with Phantoms in Medical Physics (RT) U. Oelfke Division of Radiotherapy & Imaging uwe.oelfke@icr.ac.uk Making the discoveries that defeat cancer 1. Introduction What is a phantom? Wiki:

More information