IMRT Delivery System QA. IMRT Dose Delivery. Acceptance testing. Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) MLC Alignment
|
|
- Clifton Russell
- 5 years ago
- Views:
Transcription
1 1 IMRT Delivery System Q Thomas LoSasso, PhD Memorial Sloan Kettering Cancer Center IMRT Dose Delivery cceptance testing Commissioning Quality assurance Verification Q Why: specific tests for IMRT? 2. Gap error Dose error ccuracy of leaf positioning (gaps) Static IMRT Leaf affects borders only affects dose within target position 1-2 mm OK.2 mm OK % Dose error Range of gap width.5 Gap error (mm) Nominal gap (cm) MLC lignment cceptance testing MLC alignment Film Field Field MLC leaf / gap calibration Timing errors 1 1 eam characteristics at small MU Film 2 6 Field 9 1 Field
2 2 Doubly - exposed films MLCtable.txt corrections Film 1 Film 2 perfect alignment Correction (mm) MLCtable measurement "precise" Varian.5 mm misalignments Distance from central axis (cm) Gap measurement - mechanical DMLC - Symmetry and Flatness Gap at isocenter upper lower.95.9 open 5mm 5mm/open cm 12 cm DMLC - Symmetry and Flatness Gap correction - measurement open 5mm 5mm/open transmission gap only MV.2 y =.9945x R 2 = MV.2 y =.9954x R 2 = Nominal gap (mm)
3 3 Effective gap offset variation Mark 1 Mark 2 Millenium Phelps X 15X 6X 15X 6X 15X 6X 6X 15X Off-axis variation - "cupping" Millennium 12 Effective gap offset (mm) CX 5cm off 1cm off Series4 Series5 Series6 +/- 1% 15 MV 6 MV Eff gap offset iso) 2 4% Dose Off-axis variation - corrected Eff gap offset 1.2 S =.47 mm with cupping correction Eff gap offset 1.2 S =.27 mm ~1% dose with individual leaf corrections Frequency of deviations (%) # Global logged - prescribed leaf iso (mm) checks RMS Field Room ngle < LO Log file Leaf sequence file dose distribution LPO Sup P Post nt RPO RO cm Inf Overlay jaws Dose differences (log-dva) # checks = # moving leaves * OT / 55 msec
4 4 Difference (%) Timing error SMLC (Varian) MU/seg 16 MU/seg 25 MU/seg 4 MU/seg -5 Delivered with dose mode at 4MU/min 1 MU/seg (CL23, 6MV, 1 SSD, cm depth) MU/seg # Segment 16 MU/seg 25 MU/seg 4 mu/min Courtesy of Ping Xia, UCSF Timing error SMLC (Varian) Courtesy of Gary Ezzell and the Journal of pplied Clinical Medical Physics. Overlay 11 MU Dose differences (log-dva) 11 MU 55 MU 2X Timing / communication errors Distance from axis (mm) 28 MU 4X 14 MU 8X MU (%) 15 MU Timing / communication errors Distance from axis (mm) Distance from axis (mm) Comparison of prescribed and delivered doses Log file Leaf sequence file dose distribution 15 MU prescribed / 6 MU delivered F G F G MU (%) 15 MU MU (%) 6 MU D Overlay plan log C E D C Difference (log - plan) E
5 5 Relative dose (corrected to 15 MU) MU decrease dose errors increase Timing error Is there an analogous problem for DMLC? C D E F G Prescribed MU = MU delivered Trailing leaves (-side) compensate for the leading leaves (-side) Dose (MU) error = position error () position error leaf speed () leaf speed () However, some leaf sequencers actively utilize the beam hold-off to control leaf speed eam Stability: Flatness, Symmetry Stability of flatness and symmetry affects dose rate for small fields directed off the central axis. For an open 2x2cm 2 field, measure profiles for irradiations ranging from 1 to 1 MU. Sun Nuclear Profiler (46 diodes, 1 profiles/sec). Flatness is +/-3% if more than 5MU delivered. Symmetry +/-3% if more than 4MU delivered. Courtesy of M. Sharpe, William eaumont Hospital Transmission Transmission profiles across leaves Mark 1 (445) Millenium (444) average average Mark 2 (245) Mark 2 (442) average average Respiratory gating Parallel Splitting large fields Y Perpendicular Leaf motion X
6 6 Routine Q Quality assurance MLC Q MLC specific Mechanical stability ccuracy of leaf positioning (gaps) Verification Patient / field specific Tx plan is transferred and delivered correctly Motor fatigue MLC calibration acklash Grease Film pattern (semi-weekly) Dosimetry (monthly) Dosimetry (monthly) lcohol (semi-annually) Motors Motor replaced replacement history MLC motors replaced # Jan- 92 Jan- 93 MLC installed 1st MLC Tx Room Jan- Jan- Jan- Jan- Jan Date 1st dmlc Tx Jan- Jan- Jan Jan- Jan a Field Size Defined by MLC at 1 cm SD by Light Field: Leaf / carriage positioning vs off-axis distance Date: Tolerance:.5 mm Initials: Digital setting (cm) carriage carriage leaves leaves Note: If leaf positions exceed tolerance, reinitialize and recheck. If this does not correct the problem, verify that the correction parameters in files MLCTLE.TXT and MLCXCL.TXT have not changed. 4 cm
7 7 1 mm bands Leaf positioning errors introduced 17b DMLC Output (Gap width) Check 15 MV x-rays, File: DMLC_Q2.dva, Dose = 1 MU, Dose rate = 24 MU/min SD: 1cm, Depth = 3.1cm, ackup = 1cm, C.S. = 1x1cm2 Tolerance: (The ratios should be within 2% (<.2 mm gap) of the annual calibration.) -.5 mm -.2 mm Measure the output for the scanned field before and after reinitializing the MLC. Measure the output for a 1x1cm2 reference field size defined by the jaws (MLC retracted). Calculate the ratio of the scanned field to the reference field outputs. Reference 1x1 DMLC - before reinitialization DMLC - after reinitialization Reference 1x1 +.2 mm +.5 mm verage Ratio 17c eam Holdoff Functionality Date: Initials Same setup as 17b, Dose = 5 MU Record the output and the Tx time. Check if the holdoff is being applied. Reading Functioning nnual DMLC / 1x DMLC Output vs Time Jan-98 Jan-99 Jan- Jan-1 Jan-2 Jan-3 Jan-4 Date.2 mm ~1% 17d Output vs Gantry and Collimator ngles Date: Initials 15 MV x-rays, File: DMLC_Q2.dva, Dose = 1 MU, Dose rate = 24 MU/min SD: 1cm, Spokas in air w/buildup, C.S. = 1x1cm2 Tolerance: (The ratios should be within 3% (<.2 mm gap) of each other.) Position a cylindrical chamber in air at isocenter. For each gantry angle, measure the open field and the scanned field outputs. t gantry angles of 9 and 27, also measure output for collimator angles of 9. Calculate the ratio of the scanned field to the reference field outputs. Collimator angle Gantry Reference DMLC DMLC / Reference angle DMLC output vs gantry and collimator angle Mark 1 (445) Millennium (444) DMLC output vs gantry angle (diode array) 3 DMLC / 1x Mark 2 (245) Millennium (241) ±.2 mm (± 1%) mm Year
8 8 Summary 1. cceptance testing MLC capable of IMRT delivery Improved calculations needed for IMRT fields 2. Commissioning MLC modeled accurately in TPS 3. Routine MLC Q leaves and carriages - mechanics 4. Patient specific Q independent verification IMRT Dose ccuracy 2 cm Plan Film Overlay 2 cm Film - Plan Film - Plan Difference ~25% Dosimetry (film plan) T&G Source MLC scatter Lung P field Corrections (plan plan) x1 Tongue and Groove
9 9 Relative dose Tongue and groove Distance from CX (cm) -1 cm cm 1 cm 2 cm DMLC / 1x1 (norm to -) DMLC output vs gantry and collimator angle adjusted bearings replaced bearings adjusted bearings adjusted bearings Mark 1 (445) adjusted drive belts adjusted bearings
Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy
Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy Thomas LoSasso, a) Chen-Shou Chui, and C. Clifton Ling Department
More informationThe Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017
The Current State of EPID-Based Linear Accelerator Quality Assurance Timothy Ritter, PhD, DABR, FAAPM 1 Disclosures Employed by the Veterans Health Administration Faculty appointment with the University
More informationIQM Detector Characteristics: Signal reproducibility
The Integral Quality Monitor (IQM) System is a real-time beam verification system that monitors the accuracy of radiation delivery throughout each patient treatment without any user interaction. IQM continuously
More information8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.
Use of EPIDs for Non-Routine Linac QA Bin Cai PhD Disclosure Parts of this project received support from Varian Medical System. Learning Objectives Learn the recent development of EPID based Non-routine
More information7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators
Implementing a new digital medical accelerator John Wong Johns Hopkins University AAPM, Austin, 2014 Acknowledgements Yin Zhang, Ken Wang, Kai Ding (Commissioning - JHU) Esteban Velarde, Joe Moore (QA
More informationPHYSICS QUESTIONNAIRE FORM
PHYSICS QUESTIONNAIRE FORM Institution Name: Date: Contact Information (name, address, phone, fax, email): Physicist: Radiation Oncologist: Dosimetrist (if applicable): Study Coordinator (if applicable):
More informationCharacterization, Commissioning and Evaluation of Delta 4 IMRT QA System. Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX.
Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX. 1 Acknowledgements Collaborators: Jose Bencomo, Rafael. M. Landrove, Peter
More informationSUN NUCLEAR. EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm. corporation. Your Most Valuable QA and Dosimetry Tools
EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm SUN NUCLEAR corporation Your Most Valuable QA and Dosimetry Tools introduction Pre-treatment dose QA is an important process required
More informationCommissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests
Acknowledgements David Shepard, Ph.D. Daliang Cao, Ph.D. Muhammad K. N. Afghan, Ph.D. Jinsong Ye, M.S. Tony P. Wong, Ph.D. Fan Chen, Ph.D. Min Rao, Ph.D. Vivek Mehta, M.D. Igor Gomola, Ph.D. David Housley
More informationVolumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA
Volumetric Modulated Arc Therapy David Shepard Swedish Cancer Institute Seattle, WA Disclaimer Our VMAT work has been sponsored in part by Elekta. Outline David Shepard VMAT Basics and VMAT Plan Quality
More information3D Diode Array Commissioning: Building Confidence in 3D QA Technology
3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose
More informationEmerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA
Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Arthur Olch, PhD, FAAPM AAPM Spring Clinical Meeting, March 21, 2017 Or.. What Dose are the Patients Really Getting???
More informationEvaluation of a diode array for QA measurements on a helical tomotherapy unit
Evaluation of a diode array for QA measurements on a helical tomotherapy unit K. M. Langen, a S. L. Meeks, D. O. Poole, T. H. Wagner, T. R. Willoughby, O. A. Zeidan, and P. A. Kupelian Department of Radiation
More informationArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung
ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung Treffen des Arbeitskreises IMRT der DGMP Würzburg, 26 + 27.03.2009 Salih Arican Sun Nuclear Corporation QA Challenge for Rotational Beams
More informationSRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools
SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film With improvements in targeting and localization, stereotactic treatments have become prevalent. To
More informationDOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING
DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev. 1.0 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE DoseLab users may reference the following instructions to perform Tomotherapy Quality Assurance tests as recommended
More informationDevelopment of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA
Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Gena M.A.H 1, Ahmed L.El-Attar 2, Elbadry M. Zahran 3, Hany El-Gamal
More informationSRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools
SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film As clinics strive to treat with the smallest field possible, the demand for SRS/SBRT QA grows as well.
More informationInstallation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter
Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter MR-guided RT in Heidelberg Funded by the German Research Foundation (DFG) Heidelberg consortium received
More informationCommissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010
Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Indra J. Das, PhD, FACR Department of Radiation Oncology Indiana University of School of Medicine & Midwest Proton Radiation
More informationEffect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging
The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2008 Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging Dean
More informationComparison of peripheral dose measurements using Ionization chamber and MOSFET detector
ORIGINAL ARTICLES Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector Gopiraj ANNAMALAI 1, Ramasubramanian VELAYUDHAM 2 ABSTRACT Received: 7.07.2009 Accepted: 2.11.2009
More informationArcCHECKTM. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools. VMAT RapidArc TomoTherapy Pinnacle 3 SmartArc Conventional IMRT
TM The Ultimate 4D QA Solution A 4D isotropical cylindrical detector array for arc delivery QA and Dosimetry U.S.Patent No. 8,044,359 What is? The world s first true 4D detector array The world s first
More informationPerformance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 4, 2016 Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film Eric C. Lobb Department
More informationIsocenter and Field of View Accuracy Measurement Software for Linear Accelerator
Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator Aleksei E. Zhdanov 1 and Leonid G. Dorosinskiy 1 Ural Federal University named after the first President of Russia B. N.
More information8/3/2016. The EPID Strikes Back. Novel Applications for Current EPID Technology. Joerg Rottmann, PhD. Disclosures and acknowledgements
The EPID Strikes Back Joerg Rottmann Brigham and Women s Hospital / Dana-Farber Cancer Institute Harvard Medical School Disclosures and acknowledgements Disclosures Varian MRA grant Acknowledgements Boston
More informationA Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images
A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images Aiping Ding, Bin Han, Lei Wang, Lei Xing Department of Radiation Oncology, Stanford University School of
More informationArcCHECK. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools
ArcCHECK The Ultimate 4D QA Solution A 4D isotropical cylindrical detector array for arc delivery QA and Dosimetry U.S.Patent No. 8,044,359; 6,125,335 Compatible with: FFF Beams VMAT RapidArc TomoTherapy
More informationspecifications TrueBeam STx System
specifications TrueBeam STx System TrueBeam STx s The TrueBeam STx system specifications in this document are identified as belonging to two categories, performance specifications and descriptive specifications.
More informationMobius3D. Software based IMRT QA
Mobius3D Software based IMRT QA What is Mobius Medical Systems? Clinical Expertise Software Expertise Nathan Childress, Ph.D., Founder Eli Stevens, Chief Technical Officer Support Expertise Physicists
More informationISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM
FILMQA STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE Supports all major radiotherapy technologies! FilmQA is optimized for use with Gafchromic film products, including EBT2 and RTQA2. FILMQA helps
More informationA diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine Iwein Van de Vondel, 1 Koen Tournel, 1 Dirk
More informationA positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Huaiqun Guan,
More informationThe Ultimate 4D QA Solution A 4D isotropic cylindrical detector array for arc delivery QA and Dosimetry.
The Ultimate 4D QA Solution A 4D isotropic cylindrical detector array for arc delivery QA and Dosimetry. U.S.Patent No. 8,044,359; 6,125,335 Your Most Valuable QA and Dosimetry Tools 2 Y o u r M o s t
More informationMonica Kishore. Medical Physics Graduate Program Duke University. Approved: Jennifer O Daniel, Co-Supervisor. Fang-Fang Yin, Co-Supervisor
Accuracy of Planar Dosimetry for Volumetric Modulated Arc Therapy Quality Assurance by Monica Kishore Medical Physics Graduate Program Duke University Date: Approved: Jennifer O Daniel, Co-Supervisor Fang-Fang
More informationCOMPREHENSIVE TG-142 IMAGING AND MACHINE QA
QA SOFTWARE COMPREHENSIVE TG-142 IMAGING AND MACHINE QA Automate the analysis of over thirty TG-142 recommended QA tasks The rapid progress of Radiation Therapy has created the need for Quality Assurance
More informationQC Testing for Computed Tomography (CT) Scanner
QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including
More informationspecifications TrueBeam System
specifications TrueBeam System TrueBeam System s The TrueBeam system specifications in this document are identified as belonging to two categories, performance specifications and descriptive specifications.
More informationTable 1: Available X-ray Beam Energy Combinations (MV) 6 10/ Yes Yes 6 16/15 6 Yes No 6 23/18 6 Yes No 6 25/20 6 Yes No
SPECIFICATIONS Introduction This specification sheet provides information for the Trilogy linear accelerators. 1.0 Photon Beams 1.1 Energy: Three photon beams may be selected in accordance with the beam
More informationGafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications
GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications I. SCOPE The protocol applies to GafChromic EBT3P and EBT3+P films exposed in GafChromic QuiCk Phantom
More informationCharacterization of an in vivo diode dosimetry system for clinical use
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 2, SPRING 2003 Characterization of an in vivo diode dosimetry system for clinical use Kai Huang, 1, * William S. Bice, Jr., 2, and Oscar Hidalgo-Salvatierra
More informationSub-mm accuracy of accelerators: How manufacturers achieve it, how physicists verify it
Sub-mm accuracy of accelerators: How manufacturers achieve it, how physicists verify it Presented by Ivan A. Brezovich, PhD, at the 2015 Annual meeting of the SEAAPM, Raleigh, NC, April 24, 2015 Purpose
More informationCHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY
14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to
More informationAccuracy of rapid radiographic film calibration for intensity-modulated radiation therapy verification
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 2, SPRING 2006 Accuracy of rapid radiographic film calibration for intensity-modulated radiation therapy verification Ravi Kulasekere, a Jean
More informationEvaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array
Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array José A. Bencomo, * Geoffrey Ibbott, Seungsoo Lee, and Joao A. Borges Department of Radiation Physics.
More informationA proposed method for linear accelerator photon beam steering using EPID
Received: 13 January 2018 Revised: 11 May 2018 Accepted: 29 June 2018 DOI: 10.1002/acm2.12419 RADIATION ONCOLOGY PHYSICS A proposed method for linear accelerator photon beam steering using EPID Michael
More informationAnalysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose
J. Radiat. Res., 53, 301 305 (2012) Analysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose Katsumi SHIMA 1,2, Kunihiko TATEOKA 1 *, Yuichi SAITOH 1,2, Junji SUZUKI
More informationspecifications The Complete Radiation Oncology Solution
specifications The Complete Radiation Oncology Solution Specifications This specification sheet provides information for the UNIQUE radiation oncology solution and in particular the UNIQUE linear accelerator.
More informationAperture Based Inverse Planning AAPM Summer School 2003
Aperture Based Inverse Planning AAPM Summer School 003 D.M. Shepard, M.A. Earl, Y. Xiao, C.X. Yu Acknowledgements Ziping Jiang Allen Li Shahid Naqvi James Galvin Di Yan Prowess, Inc. University of Maryland
More informationImpact of energy variation on Cone Ratio, PDD10, TMR20 10 and IMRT doses for flattening filter free (FFF) beam of TomoTherapy Hi-Art TM machines
JBUON 2014; 19(4): 1105-1110 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com ORIGINAL ARTICLE Impact of energy variation on Cone Ratio, PDD10, TMR20 10 and IMRT
More informationComparison of measured Varian Clinac 21EX and TrueBeam accelerator electron field characteristics
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Comparison of measured Varian Clinac 21EX and TrueBeam accelerator electron field characteristics Samantha A.M. Lloyd, 1a Sergei Zavgorodni,
More informationA 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery
A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery Paul A. Jursinic a) Medical College of Wisconsin, Radiation Oncology Department, Milwaukee, Wisconsin
More informationMonte Carlo study on a new concept of a scanning photon beam system for IMRT
NUKLEONIKA 2011;56(4):291 297 ORIGINAL PAPER Monte Carlo study on a new concept of a scanning photon beam system for IMRT Anna M. Wysocka-Rabin, Günter H. Hartmann Abstract. Intensity-modulated radiation
More informationFast Real Time Calibration of a Varian 2100 (High Energy) series LINAC
Fast Real Time Calibration of a Varian 2100 (High Energy) series LINAC Kenneth Chu, Ph.D., D.A.B.R., P.Eng. Marquette General Hospital Marquette, MI And Shidong Tong, Ph.D., D.A.B.R. Inova Alexandria Hospital
More informationspecifications Novalis Tx image-guided radiosurgery linear accelerator
specifications Novalis Tx image-guided radiosurgery linear accelerator Specifications Introduction This specification sheet provides information for the Novalis Tx image-guided radiosurgery linear accelerator.
More informationThe evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response
Iran. J. Radiat. Res., 2005; 3 (1): 3-10 The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response M. Mohammadi 1,2,3* and E. Bezak 1,2
More informationStability of the Helical TomoTherapy Hi Art II detector for treatment beam irradiations
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 6, 2014 Stability of the Helical TomoTherapy Hi Art II detector for treatment beam irradiations Karin Schombourg, François Bochud, Raphaël
More informationUNIQUE. The Complete Radiation Oncology Solution SPECIFICATIONS
UNIQUE The Complete Radiation Oncology Solution SPECIFICATIONS UNIQUE SPECIFICATIONS I 2 UNIQUE SPECIFICATIONS I 3 Specifications This specification sheet provides information for the UNIQUE radiation
More informationSTEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS
STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%
More informationA feasibility study of using conventional jaws to deliver IMRT plans in the treatment of prostate cancer *
IOP PUBLISHING Phys. Med. Biol. 52 (7) 2147 2156 PHYSICS IN MEDICINE AND BIOLOGY doi:1.188/31-9155/52/8/7 A feasibility study of using conventional jaws to deliver IMRT plans in the treatment of prostate
More informationCommissioning an Elekta Versa HD linear accelerator
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 Commissioning an Elekta Versa HD linear accelerator Ganesh Narayanasamy, 1,2 Daniel Saenz, 1 Wilbert Cruz, 1,3 Chul S. Ha, 1 Niko
More informationClinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification
Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification L. N. McDermott, M. Wendling, B. van Asselen, J. Stroom, J.-J. Sonke, M. van Herk, and B. J. Mijnheer a Department
More informationSensitivity study of an automated system for daily patient QA using EPID exit dose images
Received: 27 June 2017 Revised: 8 December 2017 Accepted: 27 January 2018 DOI: 10.1002/acm2.12303 RADIATION ONCOLOGY PHYSICS Sensitivity study of an automated system for daily patient QA using EPID exit
More informationInitial setup and subsequent temporal position monitoring using implanted RF transponders
Initial setup and subsequent temporal position monitoring using implanted RF transponders James Balter, Ph.D. University of Michigan Has financial interest in Calypso Medical Technologies Acknowledgements
More informationInitial Experience with a Commercial System for Volumetric Analysis of Patient Specific QA. Katja Langen Mariana Guerrero Shifeng Chen Shh..
Initial Experience with a Commercial System for Volumetric Analysis of Patient Specific QA Katja Langen Mariana Guerrero Shifeng Chen Shh..Mobius3D ACKOWLEDGEMENTS Dr. Katja Langen Dr. Shifeng Chen Dr.
More informationAccurate two-dimensional IMRT verification using a back-projection EPID dosimetry method
Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method Markus Wendling, Robert J. W. Louwe, a Leah N. McDermott, Jan-Jakob Sonke, Marcel van Herk, and Ben J. Mijnheer
More informationFour-dimensional in vivo dosimetry by dose reconstruction using continuous EPID images and phase sorting method. JiHyung Yoon.
Four-dimensional in vivo dosimetry by dose reconstruction using continuous EPID images and phase sorting method Director of Dissertation: Dr. Jae Won Jung Major Department: Physics by JiHyung Yoon July,
More informationCharacterization and evaluation of an integrated quality monitoring system for online quality assurance of external beam radiation therapy
Received: 5 January 1 Accepted: 1 September 1 DOI: 1.1/acm.11 RADIATION ONCOLOGY PHYSICS Characterization and evaluation of an integrated quality monitoring system for online quality assurance of external
More informationDosimetric IMRT verification with a flat-panel EPID
Dosimetric IMRT verification with a flat-panel EPID B. Warkentin Department of Medical Physics, Cross Cancer Institute and Department of Physics, University of Alberta, 11 University Avenue, Edmonton,
More informationGantry angle determination during arc IMRT: evaluation of a simple EPID-based technique and two commercial inclinometers
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 6, 2012 Gantry angle determination during arc IMRT: evaluation of a simple EPID-based technique and two commercial inclinometers Pejman Rowshanfarzad,
More informationTOMOTHERAPY H SERIES TomoH, TomoHD and TomoHDA Systems Technical Specifications
TOMOTHERAPY H SERIES TomoH, TomoHD and TomoHDA Systems Technical Specifications The TomoTherapy H Series is designed to treat the entire spectrum of radiation therapy patients with enhanced speed, precise
More informationBeam Delivery Techniques: Passive Scattering Proton Beams. Zuofeng Li and Roelf Slopsema University of Florida Proton Therapy Institute
Beam Delivery Techniques: Passive Scattering Proton Beams Zuofeng Li and Roelf Slopsema University of Florida Proton Therapy Institute 1 Disclaimer UFPTI uses IBA Proton Therapy System Proton Beam: production,
More informationHITACHI Proton Therapy System with Spot Scanning
Workshop on Hadron Therapy of Cancer 27 th April, Erice, Sicily, Italy HITACHI Proton Therapy System with Spot Scanning Kazuo Hiramoto Energy & Environmental Systems Laboratory, Hitachi, Ltd. Contents
More informationUsing the frame averaging of as500 EPID for IMRT verification
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 Using the frame averaging of as500 EPID for IMRT verification J. Chang* and C. C. Ling Medical Physics Department, Memorial Sloan
More informationCyberKnife Iris Beam QA using Fluence Divergence
CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery
More informationBeam Production, Characteristics and Shaping
Beam Production, Characteristics and Shaping Dr. Manfred Sassowsky Outline X-ray production 60 Co units Linear Accelerators Beam characteristics Beam shaping Beam Production, Characteristics and Shaping
More informationPedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland
Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland Center for proton therapy - PSI GANTRY DESIGN AND RECENT EXPERIENCE AT PSI 2nd Workshop on Hadron Beam Therapy of Cancer ERICE May 23 th,
More informationDose-response characteristics of an amorphous silicon EPID
Dose-response characteristics of an amorphous silicon EPID Peter Winkler a Division of Medical Radiation Physics, Department of Radiotherapy and Radiobiology, Medical University of Vienna, Waehringer Guertel
More informationQA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion
Leksell Gamma Knife Perfexion QA Considerations Paula L. Petti, Ph.D. Taylor McAdam Bell Neuroscience Institute Washington Hospital Healthcare System Fremont, CA 1 QA for LGK Perfexion : : Follow NRC licensing
More informationNuclear Associates , , CT Head and Body Dose Phantom
Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product
More informationExposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?
Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago
More informationQUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING
1 QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING QualiFormeD Phantoms A selection of test objects facilitating regulatory quality controls in radiation therapy and medical imaging Practical,
More informationTotal body irradiation dose optimization based on radiological depth
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 3, 2012 Total body irradiation dose optimization based on radiological depth Amjad Hussain, 1,3a Peter Dunscombe, 1,2,3 J. Eduardo Villarreal-
More informationGantry design and experience at PSI
Gantry design and experience at PSI Eros Pedroni for the R&D Technology Team Center for Proton Therapy Paul Scherrer Institute Villigen-PSI SWITZERLAND Workshop on Hadron Beam Therapy of Cancer Erice April
More informationLudlum Medical Physics
Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to
More informationDesign and Testing of Indigenous Cost Effective Three Dimensional Radiation Field Analyser (3D RFA)
Technology in Cancer Research and Treatment ISSN 1533-0346 Volume 13, Number 3, June 2014 Adenine Press (2014) Design and Testing of Indigenous Cost Effective Three Dimensional Radiation Field Analyser
More informationSUBCHAPTER 14. THERAPEUTIC INSTALLATIONS
SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS 7:28-14.1 Scope (a) This subchapter covers therapeutic installations used in the healing arts. These therapeutic installations include x-ray, accelerator and teletherapy
More informationCharacterization of an in vivo diode dosimetry system for clinical use
Louisiana State University LSU Digital Commons LSU Master's Theses Graduate School 2002 Characterization of an in vivo diode dosimetry system for clinical use Kai Huang Louisiana State University and Agricultural
More informationA new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 1, WINTER 2010 A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA Abdul Qadir Jangda,
More informationDepartment of Physics, State University of New York at Buffalo, Buffalo NY, USA
124 research article A fully electronic intensity-modulated radiation therapy quality assurance (IMRT QA) process implemented in a network comprised of independent treatment planning, record and verify,
More informationof sufficient quality and quantity
of sufficient quality and quantity The patient s body attenuates the beam as it passes though the body More energy is deposited in organs located near the entry of the beam than near the exit of the beam
More informationConflict Disclosure. Rotational IMRT. Arc therapy. Dynamic Arc therapy. Intensity Modulated Arc Therapy Principles and Perspectives
Intensity Modulated Arc Therapy Principles and Perspectives Cedric Yu University of Maryland Conflict Disclosure Advisory Council on Advanced Treatment Delivery, Varian Medical Systems, Inc. Patent License:
More informationNathan Childress, Ph.D., DABR
Nathan Childress, Ph.D., DABR Introduction TG-142 is a comprehensive QA protocol Covers nearly every aspect of machine and safety QA Recommends quantitative results Recommends high testing frequencies
More informationX3D in Radiation Therapy Procedure Planning. Felix G. Hamza-Lup, Ph.D. Computer Science Armstrong Atlantic State University Savannah, Georgia USA
X3D in Radiation Therapy Procedure Planning Felix G. Hamza-Lup, Ph.D. Computer Science Armstrong Atlantic State University Savannah, Georgia USA Outline 1. What is radiation therapy? 2. Treatment planning
More informationAutomated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London
Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients
More informationDigital radiography (DR) post processing techniques for pediatric radiology
Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions
More informationUniversity of Wollongong. Research Online
University of Wollongong Research Online University of Wollongong Thesis Collection 1954-2016 University of Wollongong Thesis Collections 2013 Validation of a general model for Intensity Modulated Radiation
More informationState of the Art Film Dosimetry
State of the Art Film Dosimetry Micke A., Lewis D. Advanced Materials Ashland proprietary technology, patents pending Film Dosimetry Radiochromic Film EBT2/EBT3 One-Scan Protocol Multi-channel Film Dosimetry
More informationThe most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 BUSINESS UNIT OF ISP
The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 GAFCHROMIC EBT dosimetry film Designed and optimized for ALL RT procedures Can be cut into
More informationPhantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging
in partnership with Phantoms in Medical Physics (RT) U. Oelfke Division of Radiotherapy & Imaging uwe.oelfke@icr.ac.uk Making the discoveries that defeat cancer 1. Introduction What is a phantom? Wiki:
More information