Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland

Size: px
Start display at page:

Download "Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland"

Transcription

1 Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland Center for proton therapy - PSI GANTRY DESIGN AND RECENT EXPERIENCE AT PSI 2nd Workshop on Hadron Beam Therapy of Cancer ERICE May 23 th, 2011 on behalf of the Gantry 2 group

2 Layout of the presentation Gantry 1 experience Facility expansion at PSI - new SC cyclotron The new Gantry 2 of PSI Layout Beam optics and beam size Energy variations with the beam Sweeper magnets calibration Advanced beam scanning techniques Possible future clinical use of Gantry 2 Page 2

3 Early 90's GANTRY 1 USED FOR PATIENT TREATMENTS SINCE 1996 GOAL in 1989 BASIC FEASIBILITY OF PENCIL BEAM SCANNING Page 3

4 The long term experience of using GANTRY 1 Designed in 1991 for protons On the basis of the scanning experience with pion therapy with inverse planning based on CT data Gantry 1 Page 4

5 System characteristics of Gantry 1 Upstream scanning applied only in the dispersion plane of the beam Magnetic scanning started before the last bending magnet Eccentric mounting of patient table on the gantry front wheel (with counter-rotation) Gantry radius reduce to 2 m Still the smallest proton gantry in the world α rotation β rotation φ rotation Page 5

6 If i could do it again Eccentric mounting as with Gantry 1 (R = 2 m) But with rotation only on one side (0 180 ) as with the new Gantry 2 Permanent floor underneath the patient table "Treatment cell" moving with the gantry 2 m compact eccentric gantry A new Gantry 3? Page 6

7 Pencil beam scanning Small pencil beam: 3 mm σ (7-8 mm FWHM) Cartesian scanning (infinite SSD) Discrete spot scanning step and shoot,method - on a 5 mm grid From 1996 until May 2008 the only scanning gantry in the world Elements of scanning: Dose Monitor+Kicker X Sweeper magnet 5 ms / step fast Y Range shifter 30 ms average Z Patient table 1 cm/s slow 100 us reaction time Page 7

8 Patient handling with Gantry 1 Use of a CT outside of the treatment area Position next patient while treating the previous one Transfer of positioned patients with automated transporter Very useful for treating children with anesthesia Patient positioning control with X-ray on Gantry 1 (optional) Page 8

9 Clinical use of GANTRY 1 In use since1996 Full fractionation ~30 fractions Treatments 8:00 16:00 Max 19 patients/day (2.5 per hour) 2.8 fields/fractions in average 1/3 of patients are children Under anesthesia 1/3 of treatments are IMPT Courtesy of B. Timmermann Weak points of Gantry 1 Table motion is part of scanning Not possible to use collimators Not possible to apply repainting We treat only non moving targets! Page 9

10 Advantages of using scanning (vs. scattering) Automated treatments all by software Minimal simple equipment The (pencil) beam "does it all" Same approach from small to large fields No individualized hardware (no fabrication of collimators compensators ) Apply dose fields in sequence without personnel entering the treatment room Simplify treatment - reduce treatment time and costs True 3-dimensional conformation Variable modulation of the range Minimal neutron background (for the patient) Lower risk of second cancer than scattering (depends on sophistication) Important for pediatric treatments Less activation of equipment in the nozzle (for the personnel) Less activation of the accelerator Better lateral fall-off as compared to scattering (?) No material in the beam path (except air and monitors) - Less MCS effects Page 10

11 Dose shaping within the target IMPT (intensity modulated PT) Delivery of IMPT and IGPT The term of comparison with IMRT (conventional therapy) Biological targeting (image guided proton therapy) Intentional non-homogeneous dose distributions Dose proportional to the tumor activity (biological signal) The topic for the future Courtesy of A. Lomax PSI Page 11

12 Major disadvantage: the organ motion sensitivity Interference of organ motion with the scanning sequence M.Phillips PSI 1992 Disturbance of the dose homogeneity within the target With Gantry 1 we treat only immobilized lesions attached to bony structures Tumors in the head, spinal chord and low pelvis We accept only movements < + -2 mm for treatments with full fractionation! Can we overcome this drawback? Possible remedies : Fast Scanning (Repainting) Gating Tracking? or a combination of those the possible points to be developed with Gantry 2 The point of scanning to be improved Page 12

13 2000 EXPANSION OF THE PROTON FACILITY AT PSI SUPERCONDUCTING CYCLOTRON FAST DEGRADER LAMINATED BEAM LINES GOALS: Stable DC beam Modulation of the beam intensity at the ion source Very fast energy changes Page 13

14 Layout of the newly expanded proton facility COMET - dedicated superconducting cyclotron [ACCEL - design H. Blosser] Beam for Gantry 1 all the year through patients treatments restarted in February 2007 no shut-downs since August 07 Horizontal beam line for OPTIS 2 transfer from OPTIS 1 last year Next generation scanning gantry : Gantry 2 1. patient planned for 2012 Medical cyclotron Exp. Area (PIF) OPTIS 2 Gantry 2 Disconnected from PSI ring cyclotron in 2006 Gantry 1 Page 14

15 Facility specifications mainly for the new Gantry 2 Super-conducting cyclotron Very stable beam at the ion source Aiming at 2-3% at the µs scale Deflector plate in the first orbit Dynamic control of the beam intensity µs time scale Fast degrader (moving carbon wedges) Continuous choice of the beam energy Beam line with laminated magnets Providing fast changes of beam energy Page 15

16 Excellent beam current stability an example Change of paradigm of testing the beam monitors Use the inherent stability of the beam to check the fine tuning of the electronics (capacitive matching) Result: improved linearity of the dose < 0.5% down to a 0.1 Gy dose Ideal beam for scanning 100% duty factor Dose measurements for diff. Plugin configurations 600 µs 33 nf 1.01 m e a s u re d D o s e /m e a s u re d D o s 1 G y applied Dose [cgy] ISO, 22pf ISO, 47 pf ISO, 33 pf ISO, 39 pf 22 nf Page 16

17 2005 A NEW PSI GANTRY - GANTRY 2 MAIN GOAL DEVELOP FURTHER SCANNING AS A UNIVERSAL BEAM DELIVERY TECHNIQUE Page 17

18 GANTRY 2 TOPIC 1: INNOVATIVE GANTRY LAYOUT READINESS FOR IMAGE GUIDED PROTON THERAPY EFFICIENT PATIENT HANDLING Page 18

19 Design started from the patient table Beam Line Support Bearing axle From -30 to +180 Patient table Room with fixed floor 0 to +180 would have been a better choice (cheaper) Page 19

20 Layout of the Gantry 2 room: patient table, compact nozzle Small compact nozzle Same patient table as at RPTC in Munich (Schär Engineering) Easy access to the patient table on fixed floor Fixed walls and ceiling for mounting commercial equipment (Vision RT?) A system open on both sides - lateral and front - Optimal for using an in-room sliding CT Page 20

21 In-room sliding CT - within reach with the patient table The beginning of IGPT? Patient positioning (tumor in soft tissues region) Setup for treating moving targets - 4D CT (relation external gate - internal motion) Page 21

22 BEV X-ray - synchronized with proton beam delivery BEV imaging - equivalent to portal imaging with photons Very large field-of-view (26 cm x 16 cm) not masked by equipment or collimators in the beam path QA control of gating and tracking (scanning + pulsed X-rays) Fluoroscopy mode Beam guidance? Proton beam Sweepers UPSTREAM SCANNING Bend and scan Scan and bend X rays tube Yoke hole Bending magnet Sweeper or Scatterer IGPT nozzle Patient Imager a) compact gantry b) long throw gantry Collimator Page 22

23 BEV - imager Check patient position at the isocenter Photograph of the retractable arm for holding the X-ray panel behind the patient on the side opposite to the nozzle (BEV X-ray). Page 23

24 GANTRY 2 TOPIC 2: PENCIL BEAM AND BEAM OPTICS GOALS: SMALL PENCIL BEAM FOR PRECISION THERAPY PARALLELISM OF SCANNING Page 24

25 GANTRY 2 beam optics M A 3 2 Q 6 W P T A 3 X Q 4 S Q 5 Q 3 Q 7 W U 3.2 m P M 1 Q C M2 S Q 1 Q 2 A 1 SPECIFICATIONS Rotational symmetric (large) phase space +- 3mm mrad % dp/p Complete achromatism Scanning-invariant beam focus Orthogonal (x-,y-) focal planes in T and U Focus to focus with 1:1 imaging from the coupling point of the gantry to the iso-center 2D- parallelism of scanning (in T and U) (on Gantry 1 only U) Page 25

26 GANTRY 2 beam optics (TRANSPORT and RAYTRACE) Q C A 1 A 2 W T W U H S Double parallel scanning A 3 Point to point focus Achromatic Q 1 Q 2 Q 3 Q 4 Q 5 Q 6 Q 7 Double parallel scanning TRANSPORT beam envelopes through the Gantry 2 Page 26

27 Advantages of coupling the gantry at a small beam focus At a small beam focus the spot image is insensitive to scattering Decouple vacuum (gantry from facility) Collimate the beam at the coupling point Forced beam centering on axis Measure the beam characteristics on-line before the gantry On-line independently of Gantry rotation and scanning Manipulate the beam shape imaged on the scanning beam Large Gaussian beam shape? Virtual dynamic collimation at coupling point? Disadvantage A focus coupling needs a longer beam line coupling with a parallel beam Page 27

28 Less skin dose Simplify treatment planning Advantage of using parallel scanning No inverse squared distance effects Simplify dosimetry and QA control Dose value and dose distribution in the patient and in the phantoms are the same Easier field patching (expansion of the scan range for treating long targets) Can be done without optimization within treatment planning Just exchange magnetic scan position with patient table offset Compensators (simulated scattering) No dose errors due to inverse R-square effects Collimators No tapered faces necessary Page 28

29 The difficult piece: the last 90 bending magnet A big beast - 34 tons 15 cm gap To accommodate a scan area of 20 x 12 cm Lamination leaks but in the end Vacuum problems well solved! Photograph of the mounting of the 90 bending magnet. Integrated vacuum chamber embracing the poles of the magnet 180 tons mechanical isocenter within mm Page 29

30 Compact optimized nozzle Vacuum up to the patient Sharp pencil beam - 3 mm sigma Two monitors and a strip monitor 2 mm strips (TERA collaboration) Removable pre-absorber IN and OUT of beam For ranges below 4 cm Telescopic motion of the nozzle To reduce air gap (keep patient at isocenter) Option to add collimator (compensator) To shield OAR on top of scanning To simulate passive scattering with a scanning beam Collision protection to treat patients remotely (multiple fields in one go) Page 30

31 Results: pencil beam size (on axis) Minimize material in the nozzle for keeping To have a sharp lateral fall-off the beam size between 3 and 4 mm sigma at all energies ( MeV) Region MeV 0.6 Adding piece by piece the materials in the nozzle Page 31

32 GANTRY 2 TOPIC 3 : ENERGY VARIATIONS WITH THE BEAM LINE COMPENSATED INTENSITY-ENERGY LOSSES VERY FAST ENERGY CHANGES Page 32

33 Compensation of beam losses in the range MeV na Constant ratio SC cyclotron Degrader Beam analysis Gantry 2 Coupling point na Rate Monitor Collimators for Intensity suppression khz Goal : reserve the use of the deflector plate for intensity modulated dose painting Energy [MeV] Page 33

34 Optimization for maximum transmission at 100 MeV Beam envelopes at 100 MeV. Horizontal (red) and vertical (blue) envelopes Green line = transmitted beam intensity Dots beam sizes calculated with TURTLE Page 34

35 Intentional beam defocusing at higher energies Beam envelopes at 200 MeV. yellow we four beam line sections. Beam defocusing at collimators (black bars) Same intensity losses as at 100 MeV Page 35

36 Taking into account hysteresis effects ( MeV) Energy loop Red: up-down Blue: down-up Uncorrected 1 mm Corrected Page 36

37 With very fast energy changes. 80 ms But - we observe Slow change of lateral beam position after a big energy change in the range MeV 1-3 mm with exponential decay with decay time of ~2 s < 0.3 mm position change after small energy changes within the SOBP We plan to correct these shift with sweeper offsets as a function of the time after the last energy change Strategy: Fixed targets - range precision of < 0.5 mm while working with full ramp MeV Moving targets - repainting - precision ~ 1mm repaint only the SOBP up and down Scintillator block the beam of Gantry 2 seen with a TV camera Page 37

38 GANTRY 2 TOPIC 4 : SWEEPER MAGNETS COMMISSIONING ISSUES NON LINEARITIES OF THE SWEEPERS Page 38

39 Need of a very precise mapping of the sweeper's action 210 MeV MeV 210 MEV 150 MEV T (cm) 0 T (cm) U (cm) 80 MeV 8 80 MEV T (cm) U (cm) U (cm) Measured (red) and calculated (blue) spot maps of the (linear input) action of the sweeper magnets on the scanned beam position. The non linearities are due to a curvature of the effective boundary of the magnetic field of the 90 bending magnet which is changing with energy (changing sign) Page 39

40 Invariant beam focusing needs some "tricks" First tests in MeV QMF1: 0% QMF1: -10% QMF1: -20% T U QMFC in series Anticipated by Raytrace calculations Tapering of the poles surface of the U sweeper Quadrupole corrector QMFC switched in series with the U-Sweeper Results: acceptable scanning-invariant focus Page 40

41 After a proper mapping of the sweepers 70 MEV 120 MEV 170 MEV 220 MEV Beam spots (2 cm steps) at the isocenter covering the scan region of 12 cm x 20 cm Page 41

42 GANTRY 2 Topic 5 : the main goal of Gantry 2 NEW ADVANCED BEAM DELIVERY TECHNIQUES TO PROMOTE SCANNING AS A UNIVERSAL BEAM DELIVERY TECHNIQUE Page 42

43 Aiming for highest scanning speed From spots For reducing organ motion errors Goal - Fast painting with volumetric repainting Painting lines < 10 ms per line (10cm + line change) Painting energy layers 200 ms per plane (20 lines x 5 mm) Change of energy (100 ms - 5mm range) Painting of volume 6 s per liter (20 energies at 5mm steps) Volumetric repainting capability (aim) repaintings / liter in 2 minutes To lines To contours Page 43

44 Conformation requires non-homogenous proton fluences U depth T 3d-Dose homogeneous Repainted for coping with organ motion Proton fluence of energy layer is non homogeneous Page 44

45 FPGA based delivery of tabulated energy layers Time driven beam delivery Beam path downloaded as tables of the sweepers U - T and beam intensity I Dose control with feed-back loop: Monitor 1 required dose -> vertical deflector plate A) If sweeper speed is the limit use variable intensity (IM) B) If beam dose rate is the limit use variable sweeper speed (SSM) C) or use both! Full range of dynamic dose control from zero IM to any dose SSM Delivered MUs Dose linearity of simple T-lines Required Dose 5ms - 10 cm lines - painted with IM 23 times Max T speed Variable intensity Page 45

46 GANTRY 2 POSSIBLE CLINICAL USE OF GANTRY 2 GOAL: PROMOTE SCANNING AS A UNIVERSAL BEAM DELIVERY TECHNIQUE to completely replace scattering Page 46

47 Often required Improve lateral fall-off for treating static targets Prostate sharp lateral fall-off at the boundary between tumor and sensitive structure Scanning alone better at high energy (edge enhancement with pencil beam) Scanning with added collimation better at low energy (beam size limitations) Scanning with varying beam energy superior to scattering? less material in the beam path Brain stem FWHM mm Idealized scattering (zero phase space) Realistic scanning Factor 1.4 gain Difference Gauss to error-function (1.7) Scanning with collimation better Scanning alone better Range mm Page 47

48 Very big tumors Combine the use of fast scanning with patient table displacements Needs remote control of the patient table (collision detection) Take advantage of the parallelism of the beam trivial patching - shift table - and apply intensity filter to spot pattern Medulloblastoma Dorsal irradiation Example: a paediatric case treated in 2004 with Gantry 1 Page 48

49 Moderately moving targets (0-5 mm) Discrete spot scanning applied with iso-layered repainting (max dose per spot visit) Treatments in the trunk (pancreas, cervix, colon), breast, lymph nodes, etc. Advantage of having fast energy changes -> Volumetric repainting FAST SCANNING IS MORE THAN JUST REPEATING COMPARE G1 spots, scaled G2 spots, scaled G2 spots, iso-layer G2 lines, scaled Pancreas Rectum Breast From Simulation of repainting strategies S. Zenklusen's thesis work PMB 55 (2010) Page 49

50 Largely moving targets Volume painting within a single breath hold? (repeated) Using Conformal line painting Speed of painting 0.2 Gy of a sphere of (17 layers) ½ liter in 7 secs If not working, Gating Lung liver 7s Page 50

51 Simulated scattering Dose shaping with compensators Uniform scanning Energy layer with homogeneous fluence Magnetic line scanning at max. speed Very high repainting number Collimator optional As a sub-mode of conformal scanning To simulate scattering on a scanning-gantry To replace completely scattering?? Retinoblastoma? Page 51

52 Feasibility demonstration of "simulated scattering" 8 cm diameter sphere Field shape = target projection minimal neutron background Variable modulation of the range Layer shrinking Parallel beam no compensator dose errors Highly repainted Distal layer: repainted 48 times in 30s From S. Zenklusen PhD - Medical Physics 2011 Page 52

53 Status of the Gantry 2 project Significant project delays due to the Priority given to OPTIS 2 in 2008 Difficulties with the vacuum of the 90 bending magnet PSI internal development of the software for the control of the patient table insufficient resources THANK YOU Page 53

Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland

Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland Center for proton therapy - PSI Gantry 2, the next generation gantry of PSI: a new system for promoting pencil beam scanning as a universal

More information

Gantry design and experience at PSI

Gantry design and experience at PSI Gantry design and experience at PSI Eros Pedroni for the R&D Technology Team Center for Proton Therapy Paul Scherrer Institute Villigen-PSI SWITZERLAND Workshop on Hadron Beam Therapy of Cancer Erice April

More information

The VARIAN 250 MeV Superconducting Compact Proton Cyclotron

The VARIAN 250 MeV Superconducting Compact Proton Cyclotron The VARIAN 250 MeV Superconducting Compact Proton Cyclotron VARIAN Medical Systems Particle Therapy GmbH Friedrich-Ebert-Str. 1 D-51429 BERGISCH GLADBACH GERMANY OUTLINE 1. Why having a Superconducting

More information

Beam Delivery Techniques: Passive Scattering Proton Beams. Zuofeng Li and Roelf Slopsema University of Florida Proton Therapy Institute

Beam Delivery Techniques: Passive Scattering Proton Beams. Zuofeng Li and Roelf Slopsema University of Florida Proton Therapy Institute Beam Delivery Techniques: Passive Scattering Proton Beams Zuofeng Li and Roelf Slopsema University of Florida Proton Therapy Institute 1 Disclaimer UFPTI uses IBA Proton Therapy System Proton Beam: production,

More information

HITACHI Proton Therapy System with Spot Scanning

HITACHI Proton Therapy System with Spot Scanning Workshop on Hadron Therapy of Cancer 27 th April, Erice, Sicily, Italy HITACHI Proton Therapy System with Spot Scanning Kazuo Hiramoto Energy & Environmental Systems Laboratory, Hitachi, Ltd. Contents

More information

Monte Carlo study on a new concept of a scanning photon beam system for IMRT

Monte Carlo study on a new concept of a scanning photon beam system for IMRT NUKLEONIKA 2011;56(4):291 297 ORIGINAL PAPER Monte Carlo study on a new concept of a scanning photon beam system for IMRT Anna M. Wysocka-Rabin, Günter H. Hartmann Abstract. Intensity-modulated radiation

More information

Evaluation of STPA in the Safety Analysis of the Gantry 2 Proton Radiation Therapy System Martin Rejzek, Paul Scherrer Institute, Switzerland

Evaluation of STPA in the Safety Analysis of the Gantry 2 Proton Radiation Therapy System Martin Rejzek, Paul Scherrer Institute, Switzerland Evaluation of STPA in the Safety Analysis of the Gantry 2 Proton Radiation Therapy System Martin Rejzek, Paul Scherrer Institute, Switzerland 11.04.2012 STAMP/STPA Workshop - Massachusetts Institute of

More information

Aperture Based Inverse Planning AAPM Summer School 2003

Aperture Based Inverse Planning AAPM Summer School 2003 Aperture Based Inverse Planning AAPM Summer School 003 D.M. Shepard, M.A. Earl, Y. Xiao, C.X. Yu Acknowledgements Ziping Jiang Allen Li Shahid Naqvi James Galvin Di Yan Prowess, Inc. University of Maryland

More information

Status of Proton Beam Commissioning at MedAustron Ion Beam Therapy Center

Status of Proton Beam Commissioning at MedAustron Ion Beam Therapy Center Status of Proton Beam Commissioning at MedAustron Ion Beam Therapy Center A. Garonna, A. Wastl, C. Kurfuerst, T. Kulenkampff, C. Schmitzer, L. Penescu, M. Pivi, M. Kronberger, F. Osmic, P. Urschuetz On

More information

New Tracking Gantry-Synchrotron Idea. G H Rees, ASTeC, RAL, U.K,

New Tracking Gantry-Synchrotron Idea. G H Rees, ASTeC, RAL, U.K, New Tracking Gantry-Synchrotron Idea G H Rees, ASTeC, RAL, U.K, Scheme makes use of the following: simple synchrotron and gantry magnet lattices series connection of magnets for 5 Hz tracking one main

More information

Initial setup and subsequent temporal position monitoring using implanted RF transponders

Initial setup and subsequent temporal position monitoring using implanted RF transponders Initial setup and subsequent temporal position monitoring using implanted RF transponders James Balter, Ph.D. University of Michigan Has financial interest in Calypso Medical Technologies Acknowledgements

More information

IMRT Delivery System QA. IMRT Dose Delivery. Acceptance testing. Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) MLC Alignment

IMRT Delivery System QA. IMRT Dose Delivery. Acceptance testing. Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) MLC Alignment 1 IMRT Delivery System Q Thomas LoSasso, PhD Memorial Sloan Kettering Cancer Center IMRT Dose Delivery cceptance testing Commissioning Quality assurance Verification Q Why: specific tests for IMRT? 2.

More information

CyberKnife Iris Beam QA using Fluence Divergence

CyberKnife Iris Beam QA using Fluence Divergence CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery

More information

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

3D Diode Array Commissioning: Building Confidence in 3D QA Technology 3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose

More information

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017 A Fast Monolithic System for Proton Imaging Fritz DeJongh ProtonVDA Inc October 2017 Disclosures I am a cofounder and co-owner of ProtonVDA Inc We hold intellectual property rights on our proton imaging

More information

SUPERCONDUCTING GANTRY AND OTHER DEVELOPMENTS AT HIMAC

SUPERCONDUCTING GANTRY AND OTHER DEVELOPMENTS AT HIMAC SUPERCONDUCTING GANTRY AND OTHER DEVELOPMENTS AT HIMAC Y. Iwata *, K. Noda, T. Shirai, T. Murakami, T. Fujita, T. Furukawa, K. Mizushima, Y. Hara, S. Suzuki, S. Sato, and K. Shouda, NIRS, 4-9-1 Anagawa,

More information

Distributed source x-ray tube technology for tomosynthesis imaging

Distributed source x-ray tube technology for tomosynthesis imaging Distributed source x-ray tube technology for tomosynthesis imaging Authors: F. Sprenger a*, X. Calderon-Colon b, Y. Cheng a, K. Englestad a, J. Lu b, J. Maltz c, A. Paidi c, X. Qian b, D. Spronk a, S.

More information

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017 The Current State of EPID-Based Linear Accelerator Quality Assurance Timothy Ritter, PhD, DABR, FAAPM 1 Disclosures Employed by the Veterans Health Administration Faculty appointment with the University

More information

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS 7:28-14.1 Scope (a) This subchapter covers therapeutic installations used in the healing arts. These therapeutic installations include x-ray, accelerator and teletherapy

More information

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images Aiping Ding, Bin Han, Lei Wang, Lei Xing Department of Radiation Oncology, Stanford University School of

More information

Volumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA

Volumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA Volumetric Modulated Arc Therapy David Shepard Swedish Cancer Institute Seattle, WA Disclaimer Our VMAT work has been sponsored in part by Elekta. Outline David Shepard VMAT Basics and VMAT Plan Quality

More information

Phantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging

Phantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging in partnership with Phantoms in Medical Physics (RT) U. Oelfke Division of Radiotherapy & Imaging uwe.oelfke@icr.ac.uk Making the discoveries that defeat cancer 1. Introduction What is a phantom? Wiki:

More information

Proton beam for UCN. UCN TAC-Meeting, May 12-13, 2005 Urs Rohrer, beam line physicist

Proton beam for UCN. UCN TAC-Meeting, May 12-13, 2005 Urs Rohrer, beam line physicist Proton beam for UCN UCN TAC-Meeting, May 12-13, 2005 Urs Rohrer, beam line physicist PSI Accelerator Division Department of Large Research Facilities Introduction Important parameters of the PSI proton

More information

Beam Production, Characteristics and Shaping

Beam Production, Characteristics and Shaping Beam Production, Characteristics and Shaping Dr. Manfred Sassowsky Outline X-ray production 60 Co units Linear Accelerators Beam characteristics Beam shaping Beam Production, Characteristics and Shaping

More information

8/3/2016. The EPID Strikes Back. Novel Applications for Current EPID Technology. Joerg Rottmann, PhD. Disclosures and acknowledgements

8/3/2016. The EPID Strikes Back. Novel Applications for Current EPID Technology. Joerg Rottmann, PhD. Disclosures and acknowledgements The EPID Strikes Back Joerg Rottmann Brigham and Women s Hospital / Dana-Farber Cancer Institute Harvard Medical School Disclosures and acknowledgements Disclosures Varian MRA grant Acknowledgements Boston

More information

NIRS accelerator present and future activities

NIRS accelerator present and future activities NIRS accelerator present and future activities Koji Noda Research Center for Charged Particle Therapy National Institute of Radiological Sciences 2 nd Workshop on Hadron Therapy of Cancer, Erice, 23rd

More information

ALIGNMENT METHODS APPLIED TO THE LEP MAGNET MEASUREMENTS. J. Billan, G. Brun, K. N. Henrichsen, P. Legrand, 0. Pagano, P. Rohmig and L. Walckiers.

ALIGNMENT METHODS APPLIED TO THE LEP MAGNET MEASUREMENTS. J. Billan, G. Brun, K. N. Henrichsen, P. Legrand, 0. Pagano, P. Rohmig and L. Walckiers. 295 ALIGNMENT METHODS APPLIED TO THE LEP MAGNET MEASUREMENTS J. Billan, G. Brun, K. N. Henrichsen, P. Legrand, 0. Pagano, P. Rohmig and L. Walckiers. CERN, CH-1211 Geneva 23, Switzerland Introduction Electromagnets

More information

200 MHz 350 MHz 750 MHz Linac2 RFQ2 202 MHz 0.5 MeV /m Weight : 1000 kg/m Ext. diameter : 45 cm

200 MHz 350 MHz 750 MHz Linac2 RFQ2 202 MHz 0.5 MeV /m Weight : 1000 kg/m Ext. diameter : 45 cm M. Vretenar, CERN for the HF-RFQ Working Group (V.A. Dimov, M. Garlasché, A. Grudiev, B. Koubek, A.M. Lombardi, S. Mathot, D. Mazur, E. Montesinos, M. Timmins, M. Vretenar) 1 1988-92 Linac2 RFQ2 202 MHz

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

The on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO)

The on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO) The on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO) A. Ansarinejad1,2, A. Attili1, F. Bourhaleb2,R. Cirio1,2,M. Donetti1,3, M. A. Garella1, S. Giordanengo1,

More information

Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System. Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX.

Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System. Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX. Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX. 1 Acknowledgements Collaborators: Jose Bencomo, Rafael. M. Landrove, Peter

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

OPTIMIZED MAGNET FOR A 250 MEV PROTON RADIOTHERAPY CYCLOTRON

OPTIMIZED MAGNET FOR A 250 MEV PROTON RADIOTHERAPY CYCLOTRON OPTIMIZED MAGNET FOR A 250 MEV PROTON RADIOTHERAPY CYCLOTRON J. Kim and H. Blosser 1. Introduction The design of a K250 superconducting cyclotron has been recently improved from the original design of

More information

SCINTILLATING FIBER DOSIMETER ARRAY

SCINTILLATING FIBER DOSIMETER ARRAY SCINTILLATING FIBER DOSIMETER ARRAY FIELD OF THE INVENTION [0001] This invention relates generally to the field of dosimetry and, more particularly, to rapid, high-resolution dosimeters for advanced treatment

More information

The Current Cyclotron Development Activities at CIAE. Current acyclotron

The Current Cyclotron Development Activities at CIAE. Current acyclotron Current Cyclotron Development Activities Shizhong An, Tianjue Zhang China Institute of Atomic Energy (CIAE) Beijing 2010-11.22 Greatful acknowledged is very fruitful and long lasting collaboration with

More information

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Indra J. Das, PhD, FACR Department of Radiation Oncology Indiana University of School of Medicine & Midwest Proton Radiation

More information

Superconducting Medical Accelerators at IBA

Superconducting Medical Accelerators at IBA Superconducting Medical Accelerators at IBA Wiel Kleeven and Eric Forton on behalf of IBA 1), AIMA 2) JINR 3) ASG 4) Sigmaphi 5) 1) Ion Beam Applications, Louvain-la-Neuve, Belgium 2) AIMA, Developpement,

More information

MXHF-1500RF is controlled by Digital key panel console that displays KV, ma and mas with APR menu programmed.

MXHF-1500RF is controlled by Digital key panel console that displays KV, ma and mas with APR menu programmed. R/F TV X-RAY SYSTEM DIAGNOSTIC RADIOGRAPHIC FLUOROSCOPIC TV SYSTEM MXHF-1500RF SYSTEM OUTLINE Product Data No. 041021-01 MXHF-1500RF is controlled by Digital key panel console that displays KV, ma and

More information

Thermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source

Thermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source Energy Level (ev) Multi-beam x-ray source array based on carbon nanotube field emission O. Zhou, JP Lu, X. Calderon-Colon, X. Qian, G. Yang, G. Cao, E. Gidcumb, A. Tucker, J. Shan University of North Carolina

More information

Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy

Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy Thomas LoSasso, a) Chen-Shou Chui, and C. Clifton Ling Department

More information

MIMS Workshop F. Hillion. MIMS Workshop

MIMS Workshop F. Hillion. MIMS Workshop MIMS Workshop 23 - F. Hillion MIMS Workshop 1/ Practical aspects of N5 Tuning Primary column : small probe, high current, influence of Z. Dynamic Transfer and scanning. Cy and P2/P3. LF4, Q and chromatic

More information

Study the Compact Photon Source Radiation Using FLUKA

Study the Compact Photon Source Radiation Using FLUKA Study the Compact Photon Source Radiation Using FLUKA Jixie Zhang, Donal Day, Rolf Ent Nov 30, 2017 This is a summary of radiation studies done for both the UVa target alone (for electron and photon beams)

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

X rays X-ray properties Denser material = more absorption = looks lighter on the x-ray photo X-rays CT Scans circle cross-sectional images Tumours

X rays X-ray properties Denser material = more absorption = looks lighter on the x-ray photo X-rays CT Scans circle cross-sectional images Tumours X rays X-ray properties X-rays are part of the electromagnetic spectrum. X-rays have a wavelength of the same order of magnitude as the diameter of an atom. X-rays are ionising. Different materials absorb

More information

1.8 MW Upgrade of the PSI Proton Accelerator Facility

1.8 MW Upgrade of the PSI Proton Accelerator Facility 1.8 MW Upgrade of the PSI Proton Accelerator Facility Pierre A. Schmelzbach for the PSI Accelerator Divisions This talk: analyzes the potential for improvements from the ion source to the spallation target

More information

Circumference 187 m (bending radius = 8.66 m)

Circumference 187 m (bending radius = 8.66 m) 4. Specifications of the Accelerators Table 1. General parameters of the PF storage ring. Energy 2.5 GeV (max 3.0 GeV) Initial stored current multi-bunch 450 ma (max 500 ma at 2.5GeV) single bunch 70 ma

More information

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images B R E A S T I M A G I N G S O L U T I O N S Setting the benchmark for mammography M-IV Series Innovations in breast imaging The Lorad M-IV Series exemplifies Hologic's commitment to developing advanced

More information

DEVELOPMENT OF CAPACITIVE LINEAR-CUT BEAM POSITION MONITOR FOR HEAVY-ION SYNCHROTRON OF KHIMA PROJECT

DEVELOPMENT OF CAPACITIVE LINEAR-CUT BEAM POSITION MONITOR FOR HEAVY-ION SYNCHROTRON OF KHIMA PROJECT DEVELOPMENT OF CAPACITIVE LINEAR-CUT BEAM POSITION MONITOR FOR HEAVY-ION SYNCHROTRON OF KHIMA PROJECT Ji-Gwang Hwang, Tae-Keun Yang, Seon Yeong Noh Korea Institute of Radiological and Medical Sciences,

More information

VIBRATING WIRE SENSORS FOR BEAM INSTRUMENTATION Suren Arutunian

VIBRATING WIRE SENSORS FOR BEAM INSTRUMENTATION Suren Arutunian VIBRATING WIRE SENSORS FOR BEAM INSTRUMENTATION Suren Arutunian Yerevan Physics Institute Yerevan Physics Institute S.Arutunian, VIBRATING WIRE SENSORS FOR BEAM INSTRUMENTATION BIW 2008, Lake Tahoe, USA

More information

BEAM HALO OBSERVATION BY CORONAGRAPH

BEAM HALO OBSERVATION BY CORONAGRAPH BEAM HALO OBSERVATION BY CORONAGRAPH T. Mitsuhashi, KEK, TSUKUBA, Japan Abstract We have developed a coronagraph for the observation of the beam halo surrounding a beam. An opaque disk is set in the beam

More information

Results of FE65-P2 Pixel Readout Test Chip for High Luminosity LHC Upgrades

Results of FE65-P2 Pixel Readout Test Chip for High Luminosity LHC Upgrades for High Luminosity LHC Upgrades R. Carney, K. Dunne, *, D. Gnani, T. Heim, V. Wallangen Lawrence Berkeley National Lab., Berkeley, USA e-mail: mgarcia-sciveres@lbl.gov A. Mekkaoui Fermilab, Batavia, USA

More information

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications I. SCOPE The protocol applies to GafChromic EBT3P and EBT3+P films exposed in GafChromic QuiCk Phantom

More information

LCLS-II SXR Undulator Line Photon Energy Scanning

LCLS-II SXR Undulator Line Photon Energy Scanning LCLS-TN-18-4 LCLS-II SXR Undulator Line Photon Energy Scanning Heinz-Dieter Nuhn a a SLAC National Accelerator Laboratory, Stanford University, CA 94309-0210, USA ABSTRACT Operation of the LCLS-II undulator

More information

ArcCHECKTM. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools. VMAT RapidArc TomoTherapy Pinnacle 3 SmartArc Conventional IMRT

ArcCHECKTM. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools. VMAT RapidArc TomoTherapy Pinnacle 3 SmartArc Conventional IMRT TM The Ultimate 4D QA Solution A 4D isotropical cylindrical detector array for arc delivery QA and Dosimetry U.S.Patent No. 8,044,359 What is? The world s first true 4D detector array The world s first

More information

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System. Use of EPIDs for Non-Routine Linac QA Bin Cai PhD Disclosure Parts of this project received support from Varian Medical System. Learning Objectives Learn the recent development of EPID based Non-routine

More information

Prompt-Gamma Based Range Verification in Particle Therapy: New prospects (also) for 4D?

Prompt-Gamma Based Range Verification in Particle Therapy: New prospects (also) for 4D? Prompt-Gamma Based Range Verification in Particle Therapy: New prospects (also) for 4D? Guntram Pausch 1-3 1 OncoRay National Center for Radiation Research in Oncology, Dresden 2 Technische Universität

More information

Nancy G. Leveson and Clark S. Turner, An Investigation of the Therac-25 Accidents. Computer 26(7), pp , Jul Presented by Dror Feitelson

Nancy G. Leveson and Clark S. Turner, An Investigation of the Therac-25 Accidents. Computer 26(7), pp , Jul Presented by Dror Feitelson Nancy G. Leveson and Clark S. Turner, An Investigation of the Therac-25 Accidents. Computer 26(7), pp. 18-41, Jul 1993. Presented by Dror Feitelson The Big Picture The Therac-25 was a computerized radiation

More information

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare GE Healthcare Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare There is excitement across the industry regarding the clinical potential of a hybrid

More information

Supplementary Figure 1

Supplementary Figure 1 Supplementary Figure 1 Technical overview drawing of the Roadrunner goniometer. The goniometer consists of three main components: an inline sample-viewing microscope, a high-precision scanning unit for

More information

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array José A. Bencomo, * Geoffrey Ibbott, Seungsoo Lee, and Joao A. Borges Department of Radiation Physics.

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Range Sensing strategies

Range Sensing strategies Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart and Nourbakhsh 4.1.6 Range Sensors (time of flight) (1) Large range distance measurement -> called

More information

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Huaiqun Guan,

More information

Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4

Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4 Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4 S.V. Roth, R. Döhrmann, M. Dommach, I. Kröger, T. Schubert, R. Gehrke Definition of the upgrade The wiggler beamline BW4 is dedicated to

More information

2010 Philips BrightView XCT SPECT/CT

2010 Philips BrightView XCT SPECT/CT 2010 Philips BrightView XCT SPECT/CT Unit was purchased from Philips training center in 2015. Installed but never been used by the current facility. (Scroll for pictures) BrightView XCT Camera with PinPoint

More information

Session 3 Summary: Orbit Feedback

Session 3 Summary: Orbit Feedback Session 3 Summary: Orbit Feedback Workshop on Ambient Ground Motion and Vibration Suppression for Low Emittance Storage Rings GM2017 12/13/2017 MMS System Design, Initial Results and Experiments with Orbit

More information

Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA

Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Arthur Olch, PhD, FAAPM AAPM Spring Clinical Meeting, March 21, 2017 Or.. What Dose are the Patients Really Getting???

More information

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Gena M.A.H 1, Ahmed L.El-Attar 2, Elbadry M. Zahran 3, Hany El-Gamal

More information

Radiation Test Report Paul Scherer Institute Proton Irradiation Facility

Radiation Test Report Paul Scherer Institute Proton Irradiation Facility the Large Hadron Collider project CERN CH-2 Geneva 23 Switzerland CERN Div./Group RadWG EDMS Document No. xxxxx Radiation Test Report Paul Scherer Institute Proton Irradiation Facility Responsibility Tested

More information

JEDI. Status of the commissioning of the waveguide RF Wien Filter

JEDI. Status of the commissioning of the waveguide RF Wien Filter COSY Beam Time Request For Lab. use Exp. No.: Session No. E 005.4 7 Collaboration: JEDI Status of the commissioning of the waveguide RF Wien Filter Spokespersons for the beam time: Ralf Gebel (Jülich)

More information

Radiological Safety Analysis Document for the CLAS12 Engineering and the first physics run of Run Group A

Radiological Safety Analysis Document for the CLAS12 Engineering and the first physics run of Run Group A Radiological Safety Analysis Document for the CLAS12 Engineering and the first physics run of Run Group A This Radiological Safety Analysis Document (RSAD) will identify the general conditions associated

More information

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM FILMQA STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE Supports all major radiotherapy technologies! FilmQA is optimized for use with Gafchromic film products, including EBT2 and RTQA2. FILMQA helps

More information

MAGNETIC LEVITATION SUSPENSION CONTROL SYSTEM FOR REACTION WHEEL

MAGNETIC LEVITATION SUSPENSION CONTROL SYSTEM FOR REACTION WHEEL IMPACT: International Journal of Research in Engineering & Technology (IMPACT: IJRET) ISSN 2321-8843 Vol. 1, Issue 4, Sep 2013, 1-6 Impact Journals MAGNETIC LEVITATION SUSPENSION CONTROL SYSTEM FOR REACTION

More information

FAST RF KICKER DESIGN

FAST RF KICKER DESIGN FAST RF KICKER DESIGN David Alesini LNF-INFN, Frascati, Rome, Italy ICFA Mini-Workshop on Deflecting/Crabbing Cavity Applications in Accelerators, Shanghai, April 23-25, 2008 FAST STRIPLINE INJECTION KICKERS

More information

Confirmation, refinement, and extension of a study in intrafraction motion interplay with sliding jaw motion

Confirmation, refinement, and extension of a study in intrafraction motion interplay with sliding jaw motion Confirmation, refinement, and extension of a study in intrafraction motion interplay with sliding jaw motion Michael W. Kissick, a Sarah A. Boswell, Robert Jeraj, and T. Rockwell Mackie Department of Medical

More information

SURVEY AND ALIGNMENT FOR THE SWISS LIGHT SOURCE

SURVEY AND ALIGNMENT FOR THE SWISS LIGHT SOURCE 1 SURVEY AND ALIGNMENT FOR THE SWISS LIGHT SOURCE F.Q. Wei, K. Dreyer, U. Fehlmann, J.L. Pochon and A. Wrulich SLS / Paul Scherrer Institute CH5232 Villigen PSI Switzerland ABSTRACT The Swiss Light Source

More information

MRI SYSTEM COMPONENTS Module One

MRI SYSTEM COMPONENTS Module One MRI SYSTEM COMPONENTS Module One 1 MAIN COMPONENTS Magnet Gradient Coils RF Coils Host Computer / Electronic Support System Operator Console and Display Systems 2 3 4 5 Magnet Components 6 The magnet The

More information

Development of toroidal bending magnets for Hadron Beam Therapy. L. Bromberg, P. Michael, J.V. Minervini MIT E. Pearson, E.

Development of toroidal bending magnets for Hadron Beam Therapy. L. Bromberg, P. Michael, J.V. Minervini MIT E. Pearson, E. Development of toroidal bending magnets for Hadron Beam Therapy L. Bromberg, P. Michael, J.V. Minervini MIT E. Pearson, E. Forton IBA IntroducEon Toroidal magnets Array of idenecal coils, revolved around

More information

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING 1 QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING QualiFormeD Phantoms A selection of test objects facilitating regulatory quality controls in radiation therapy and medical imaging Practical,

More information

Beam Transfer to Targets

Beam Transfer to Targets Volume III Update Report Chapter 3 Beam Transfer to Targets 3-1 Authors and Contributors Beam Transfer to Targets The executive summary was prepared by: R Maier 1 and KN Clausen 3 on behalf of the Beam

More information

COMMISSIONING OF THE ACCEL 250 MEV PROTON CYCLOTRON

COMMISSIONING OF THE ACCEL 250 MEV PROTON CYCLOTRON COMMISSIONING OF THE ACCEL 250 MEV PROTON CYCLOTRON A.E. Geisler, J. Hottenbacher, H.-U. Klein, D. Krischel, H. Röcken, M. Schillo, T. Stephani, J.H. Timmer, Accel Instruments GmbH, Bergisch Gladbach,

More information

Theoretical Study to calculate some parameters of Ion Optical System

Theoretical Study to calculate some parameters of Ion Optical System International Journal of ChemTech Research CODEN (USA): IJCRGG, ISSN: 97-9, ISSN(Online):55-9555 Vol.1 No.13, pp 1-18, 17 Theoretical Study to calculate some parameters of Ion Optical System *Bushra Joudah

More information

I. Introduction.

I. Introduction. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case

More information

Electron Linacs for Cargo Inspection and Other Industrial Applications

Electron Linacs for Cargo Inspection and Other Industrial Applications Electron Linacs for Cargo Inspection and Other Industrial Applications Chuanxiang Tang 1, Huaibi Chen 1,Yaohong Liu 2 Tang.xuh@tsinghua.edu.cn 1 Department of Engineering Physics, Tsinghua U., Beijing

More information

Tutorial: designing a converging-beam electron gun and focusing solenoid with Trak and PerMag

Tutorial: designing a converging-beam electron gun and focusing solenoid with Trak and PerMag Tutorial: designing a converging-beam electron gun and focusing solenoid with Trak and PerMag Stanley Humphries, Copyright 2012 Field Precision PO Box 13595, Albuquerque, NM 87192 U.S.A. Telephone: +1-505-220-3975

More information

1.6 Beam Wander vs. Image Jitter

1.6 Beam Wander vs. Image Jitter 8 Chapter 1 1.6 Beam Wander vs. Image Jitter It is common at this point to look at beam wander and image jitter and ask what differentiates them. Consider a cooperative optical communication system that

More information

REVIEW OF FAST BEAM CHOPPING F. Caspers CERN AB-RF-FB

REVIEW OF FAST BEAM CHOPPING F. Caspers CERN AB-RF-FB F. Caspers CERN AB-RF-FB Introduction Review of several fast chopping systems ESS-RAL LANL-SNS JAERI CERN-SPL Discussion Conclusion 1 Introduction Beam choppers are typically used for β = v/c values between

More information

Monica Kishore. Medical Physics Graduate Program Duke University. Approved: Jennifer O Daniel, Co-Supervisor. Fang-Fang Yin, Co-Supervisor

Monica Kishore. Medical Physics Graduate Program Duke University. Approved: Jennifer O Daniel, Co-Supervisor. Fang-Fang Yin, Co-Supervisor Accuracy of Planar Dosimetry for Volumetric Modulated Arc Therapy Quality Assurance by Monica Kishore Medical Physics Graduate Program Duke University Date: Approved: Jennifer O Daniel, Co-Supervisor Fang-Fang

More information

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity) Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular

More information

Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector

Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector ORIGINAL ARTICLES Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector Gopiraj ANNAMALAI 1, Ramasubramanian VELAYUDHAM 2 ABSTRACT Received: 7.07.2009 Accepted: 2.11.2009

More information

Beam Production, characteristics and shaping

Beam Production, characteristics and shaping Kantonsspital Luzern Beam Production, characteristics and shaping Dr. Manfred Sassowsky Cantonal Hospital Lucerne (KSL) Institute for Radio-Oncology 3.9.2007 X-ray production 60 Co units Linear Accelerators

More information

Status of the PRad Experiment (E )

Status of the PRad Experiment (E ) Status of the PRad Experiment (E12-11-106) NC A&T State University Outline Experimental apparatus, current status Installation plan Draft run plan Summary PRad Experimental Setup Main detectors and elements:

More information

The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 BUSINESS UNIT OF ISP

The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 BUSINESS UNIT OF ISP The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 GAFCHROMIC EBT dosimetry film Designed and optimized for ALL RT procedures Can be cut into

More information

Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator

Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator Aleksei E. Zhdanov 1 and Leonid G. Dorosinskiy 1 Ural Federal University named after the first President of Russia B. N.

More information

The CMS Silicon Strip Tracker and its Electronic Readout

The CMS Silicon Strip Tracker and its Electronic Readout The CMS Silicon Strip Tracker and its Electronic Readout Markus Friedl Dissertation May 2001 M. Friedl The CMS Silicon Strip Tracker and its Electronic Readout 2 Introduction LHC Large Hadron Collider:

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 David F. Lewis Advanced Materials Group Ashland Specialty Ingredients Spain, April 2014 What is Radiochromic Film? A film that instantly changes

More information

Hall C Polarimetry at 12 GeV Dave Gaskell Hall C Users Meeting January 14, 2012

Hall C Polarimetry at 12 GeV Dave Gaskell Hall C Users Meeting January 14, 2012 Hall C Polarimetry at 12 GeV Dave Gaskell Hall C Users Meeting January 14, 2012 1. Møller Polarimeter 2. Compton Polarimeter Hall C 12 GeV Polarimetry Møller Polarimeter 6 GeV operation: uses 2 quads to

More information

The European Spallation Source. Dave McGinnis Chief Engineer ESS\Accelerator Division IVEC 2013

The European Spallation Source. Dave McGinnis Chief Engineer ESS\Accelerator Division IVEC 2013 The European Spallation Source Dave McGinnis Chief Engineer ESS\Accelerator Division IVEC 2013 Overview The European Spallation Source (ESS) will house the most powerful proton linac ever built. The average

More information