Evaluation of STPA in the Safety Analysis of the Gantry 2 Proton Radiation Therapy System Martin Rejzek, Paul Scherrer Institute, Switzerland

Size: px
Start display at page:

Download "Evaluation of STPA in the Safety Analysis of the Gantry 2 Proton Radiation Therapy System Martin Rejzek, Paul Scherrer Institute, Switzerland"

Transcription

1 Evaluation of STPA in the Safety Analysis of the Gantry 2 Proton Radiation Therapy System Martin Rejzek, Paul Scherrer Institute, Switzerland STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 1

2 Agenda Proton Therapy at the Paul Scherrer Institute, Switzerland Evaluation of STPA for the Advanced Scanning Technique Scope of the Project Examples Conclusions STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 2

3 Proton Therapy at the Paul Scherrer Institute, Switzerland STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 3

4 PSI Center for Proton Therapy 250 MeV Proton accelerator (superconducting cyclotron) Beamlines to 4 user areas OPTIS Gantry 1 Gantry 2 Experimental area STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 4

5 Gantry 1 Sweeper magnet (1 dimension) Beam enters rotating Gantry Dose monitoring Range shifter STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 5

6 Gantry 1 Spot Scanning Technique Elements of spot scanning: Beam on/off 50 µs Sweeper magnet 5 ms/step Range shifter 30 ms Patient table 1 cm/s spots to treat 1 liter volume STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 6

7 Gantry 2 Beam enters rotating Gantry Sweeper magnets (2 dimensions) Dose monitoring STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 7

8 Gantry 2 Advanced Scanning Technique Advanced scanning: Increased speed Increased flexibility New treatment modalities STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 8

9 Evaluation of STPA for the Advanced Scanning Technique STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 9

10 Scope of this Project Safety analysis done with classical methods Evaluation of STPA as supplementary method for advanced scanning explored different approaches considered different parts of whole facility Few examples: 1) STPA during workshops 2) How to model controllers that can insert veto 3) What is the reference for inadequate timing and the Thomas process Treatment Facility Irradiation Patient Treatment Operator Nurse STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 10

11 Evaluation of STPA for the Advanced Scanning Technique 1) Performing STPA during workshops STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 11

12 Recipe for a STPA Workshop Step 1 Performing STPA analysis during workshop with engineers Preparation: Hierarchical control structure / high-level hazards Table with guidewords STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 12

13 Recipe for a STPA Workshop Step 2 Performing STPA analysis during workshop with engineers Preparation: Hierarchical control structure / high-level hazards Table with guidewords Reduced process loop Influence1 Influence2... Controller (1) Control Algorithm (relevant part) Var1 Var2... Param1 Param2... (3) Sensor 1 (2) Controller / Actuator Param1 Param2... Var1 Var2... Var1 Var2... Param1 Param2... (3) Sensor STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 13

14 Recipe for a STPA Workshop Classification Work in progress! UCA = Unsafe Control Action PV = Process Variable STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 14

15 Evaluation of STPA for the Advanced Scanning Technique 2) How to model controllers which can insert veto STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 15

16 Controllers that can insert Veto Typical Situation: One Treatment Delivery System (TDS) Treatment Delivery System In Proton Therapy: One source of beam for all treatment areas Intensity Controller Dose Controller various actuators and sensors Patient Treatment STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 16

17 Controllers that can insert Veto First approach: Second approach: Beam Allocator Other Areas Treatment Delivery System Req. Master Set Intensity Set Intensity Other Areas Treatment Delivery System Beam Allocator Dose Controller Intensity Controller Intensity Controller Dose Controller various actuators and sensors various actuators and sensors Patient Treatment Patient Treatment STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 17

18 Evaluation of STPA for the Advanced Scanning Technique 3) What is the reference for inadequate timing STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 18

19 Reference for Inadequate Timing Dose Beam Patient Personnel Start null off Personnel brings patient and installs him null off Personnel leaves room null off Turn beam on Pres. dose reached? accumulating on yes Turn beam off prescribed dose reached prescribed off Personnel enters room prescribed off Personnel takes patient out prescribed off End STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 19

20 Reference for Inadequate Timing too early Dose Beam Patient Personnel Start null off Personnel brings patient and installs him Personnel leaves room Turn beam on null off accumulating on X accumulating on Turn beam on Pres. dose reached? accumulating on yes Turn beam off prescribed dose reached prescribed off Personnel enters room prescribed off Personnel takes patient out prescribed off End too late STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 20

21 Conclusions STPA is a very useful method Results achievable in straitforward way and rather short time Feasible to perform with non safety experts Discussion points raised for components not yet developed Next steps Finish this project Compare Results with those of existing safety analysis hopefully continue with STPA method STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 21

22 Acknowledgments Prof. Dr. Nancy Leveson Blandine Antoine Massachusetts Institute of Technology, US Prof. Dr. Christian Hilbes Zurich University of Applied Sciences, CH Dr. Martin Grossmann Dr. David Meer Center for Proton Therapy Paul Scherrer Institute, CH Contacts: Dipl. el. Ing. FH Martin Rejzek Prof. Dr. Nancy Leveson Blandine Antoine Dr. Christian Hilbes Dr. Martin Grossmann Dr. David Meer STAMP/STPA Workshop - Massachusetts Institute of Technology - April 17-19, (c) copyright by PSI, 2012 Page 22

Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland

Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland Center for proton therapy - PSI Gantry 2, the next generation gantry of PSI: a new system for promoting pencil beam scanning as a universal

More information

The VARIAN 250 MeV Superconducting Compact Proton Cyclotron

The VARIAN 250 MeV Superconducting Compact Proton Cyclotron The VARIAN 250 MeV Superconducting Compact Proton Cyclotron VARIAN Medical Systems Particle Therapy GmbH Friedrich-Ebert-Str. 1 D-51429 BERGISCH GLADBACH GERMANY OUTLINE 1. Why having a Superconducting

More information

Gantry design and experience at PSI

Gantry design and experience at PSI Gantry design and experience at PSI Eros Pedroni for the R&D Technology Team Center for Proton Therapy Paul Scherrer Institute Villigen-PSI SWITZERLAND Workshop on Hadron Beam Therapy of Cancer Erice April

More information

The Project Objectives

The Project Objectives STPA Software Module A Eurostars Funded Project 5 th European STAMP/STPA Workshop and Conference 13-15 September 2017 - Reykjavík, Iceland Christopher Brown and Jianfei Zheng The Project Objectives Provide

More information

Nancy G. Leveson and Clark S. Turner, An Investigation of the Therac-25 Accidents. Computer 26(7), pp , Jul Presented by Dror Feitelson

Nancy G. Leveson and Clark S. Turner, An Investigation of the Therac-25 Accidents. Computer 26(7), pp , Jul Presented by Dror Feitelson Nancy G. Leveson and Clark S. Turner, An Investigation of the Therac-25 Accidents. Computer 26(7), pp. 18-41, Jul 1993. Presented by Dror Feitelson The Big Picture The Therac-25 was a computerized radiation

More information

Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland

Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland Center for proton therapy - PSI GANTRY DESIGN AND RECENT EXPERIENCE AT PSI 2nd Workshop on Hadron Beam Therapy of Cancer ERICE May 23 th,

More information

STPA FOR LINAC4 AVAILABILITY REQUIREMENTS. A. Apollonio, R. Schmidt 4 th European STAMP Workshop, Zurich, 2016

STPA FOR LINAC4 AVAILABILITY REQUIREMENTS. A. Apollonio, R. Schmidt 4 th European STAMP Workshop, Zurich, 2016 STPA FOR LINAC4 AVAILABILITY REQUIREMENTS A. Apollonio, R. Schmidt 4 th European STAMP Workshop, Zurich, 2016 LHC colliding particle beams at very high energy 26.8 km Circumference LHC Accelerator (100

More information

HITACHI Proton Therapy System with Spot Scanning

HITACHI Proton Therapy System with Spot Scanning Workshop on Hadron Therapy of Cancer 27 th April, Erice, Sicily, Italy HITACHI Proton Therapy System with Spot Scanning Kazuo Hiramoto Energy & Environmental Systems Laboratory, Hitachi, Ltd. Contents

More information

A New Systems-Theoretic Approach to Safety. Dr. John Thomas

A New Systems-Theoretic Approach to Safety. Dr. John Thomas A New Systems-Theoretic Approach to Safety Dr. John Thomas Outline Goals for a systemic approach Foundations New systems approaches to safety Systems-Theoretic Accident Model and Processes STPA (hazard

More information

4 th European STAMP Workshop 2016

4 th European STAMP Workshop 2016 4 th European STAMP Workshop 2016 STPA Tutorial - Part 1 Introduction Objectives and Content Overview 2 Objectives and Organization The goal of this tutorial is to give you an overview of STPA. Targeted

More information

Application of STPA in Radiation Therapy: a Preliminary Study

Application of STPA in Radiation Therapy: a Preliminary Study Application of STPA in Radiation Therapy: a Preliminary Study Natalia Silvis-Cividjian Wilko Verbakel Marjan Admiraal MIT STAMP Workshop 2018 VU medical center Vrije Universiteit (VU) campus Amsterdam,

More information

Beam Delivery Techniques: Passive Scattering Proton Beams. Zuofeng Li and Roelf Slopsema University of Florida Proton Therapy Institute

Beam Delivery Techniques: Passive Scattering Proton Beams. Zuofeng Li and Roelf Slopsema University of Florida Proton Therapy Institute Beam Delivery Techniques: Passive Scattering Proton Beams Zuofeng Li and Roelf Slopsema University of Florida Proton Therapy Institute 1 Disclaimer UFPTI uses IBA Proton Therapy System Proton Beam: production,

More information

IQM Detector Characteristics: Signal reproducibility

IQM Detector Characteristics: Signal reproducibility The Integral Quality Monitor (IQM) System is a real-time beam verification system that monitors the accuracy of radiation delivery throughout each patient treatment without any user interaction. IQM continuously

More information

Engineering a Safer World. Prof. Nancy Leveson Massachusetts Institute of Technology

Engineering a Safer World. Prof. Nancy Leveson Massachusetts Institute of Technology Engineering a Safer World Prof. Nancy Leveson Massachusetts Institute of Technology Why Our Efforts are Often Not Cost-Effective Efforts superficial, isolated, or misdirected Too much effort on assuring

More information

Radiation Test Report Paul Scherer Institute Proton Irradiation Facility

Radiation Test Report Paul Scherer Institute Proton Irradiation Facility the Large Hadron Collider project CERN CH-2 Geneva 23 Switzerland CERN Div./Group RadWG EDMS Document No. xxxxx Radiation Test Report Paul Scherer Institute Proton Irradiation Facility Responsibility Tested

More information

The CMS Silicon Strip Tracker and its Electronic Readout

The CMS Silicon Strip Tracker and its Electronic Readout The CMS Silicon Strip Tracker and its Electronic Readout Markus Friedl Dissertation May 2001 M. Friedl The CMS Silicon Strip Tracker and its Electronic Readout 2 Introduction LHC Large Hadron Collider:

More information

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017 A Fast Monolithic System for Proton Imaging Fritz DeJongh ProtonVDA Inc October 2017 Disclosures I am a cofounder and co-owner of ProtonVDA Inc We hold intellectual property rights on our proton imaging

More information

CMS Beam Condition Monitoring Wim de Boer, Hannes Bol, Alexander Furgeri, Steffen Muller

CMS Beam Condition Monitoring Wim de Boer, Hannes Bol, Alexander Furgeri, Steffen Muller CMS Beam Condition Monitoring Wim de Boer, Hannes Bol, Alexander Furgeri, Steffen Muller BCM2 8diamonds BCM1 8diamonds each BCM2 8diamonds Beam Condition Monitoring at LHC BCM at LHC is done by about 3700

More information

The ALPHA facility at Indiana University New Capabilities for Dose Rate Testing

The ALPHA facility at Indiana University New Capabilities for Dose Rate Testing The ALPHA facility at Indiana University New Capabilities for Dose Rate Testing P.E. Sokol, S.Y.Lee, T. Rinckel, C. Romel, R. Ellis, G. East, Y. Jing, P. McChesney, X. Pang Indiana University S. Clark

More information

Andrés Sánchez González. FTEC 2nd Workshop. CERN and CIEMAT. CERN Graphic Charter: use of the outline version of the CERN logo.

Andrés Sánchez González. FTEC 2nd Workshop. CERN and CIEMAT. CERN Graphic Charter: use of the outline version of the CERN logo. Design, production and test of readout electronics for the assessment of the radiation hardness of pixel sensor based detector components for the ALICE ITS Upgrade project Andrés Sánchez González CERN

More information

Product Range Electronic Units

Product Range Electronic Units Pyramid Technical Consultants, Inc. 1050 Waltham Street Suite 200 Lexington, MA 02421 TEL: +1 781 402-1700 TEL (UK): +44 1273 492001 FAX: (781) 402-1750 EMAIL: SUPPORT@PTCUSA.COM Product Range Electronic

More information

Results of FE65-P2 Pixel Readout Test Chip for High Luminosity LHC Upgrades

Results of FE65-P2 Pixel Readout Test Chip for High Luminosity LHC Upgrades for High Luminosity LHC Upgrades R. Carney, K. Dunne, *, D. Gnani, T. Heim, V. Wallangen Lawrence Berkeley National Lab., Berkeley, USA e-mail: mgarcia-sciveres@lbl.gov A. Mekkaoui Fermilab, Batavia, USA

More information

Intro to Systems Theory and STAMP John Thomas and Nancy Leveson. All rights reserved.

Intro to Systems Theory and STAMP John Thomas and Nancy Leveson. All rights reserved. Intro to Systems Theory and STAMP 1 Why do we need something different? Fast pace of technological change Reduced ability to learn from experience Changing nature of accidents New types of hazards Increasing

More information

SUPERCONDUCTING GANTRY AND OTHER DEVELOPMENTS AT HIMAC

SUPERCONDUCTING GANTRY AND OTHER DEVELOPMENTS AT HIMAC SUPERCONDUCTING GANTRY AND OTHER DEVELOPMENTS AT HIMAC Y. Iwata *, K. Noda, T. Shirai, T. Murakami, T. Fujita, T. Furukawa, K. Mizushima, Y. Hara, S. Suzuki, S. Sato, and K. Shouda, NIRS, 4-9-1 Anagawa,

More information

Engineering a Safer and More Secure World

Engineering a Safer and More Secure World Engineering a Safer and More Secure World Nancy Leveson MIT Topics What is the problem? Why do we need something new? Applying systems theory to system safety engineering STAMP: a new model of accident

More information

DICOM Correction Item

DICOM Correction Item DICOM Correction Item Correction Number CP- 617 Log Summary: Type of Modification Addition Name of Standard PS 3.3 2006 Rationale for Correction: The motion of modern patient support devices is no longer

More information

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

3D Diode Array Commissioning: Building Confidence in 3D QA Technology 3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose

More information

SURVEY AND ALIGNMENT FOR THE SWISS LIGHT SOURCE

SURVEY AND ALIGNMENT FOR THE SWISS LIGHT SOURCE 1 SURVEY AND ALIGNMENT FOR THE SWISS LIGHT SOURCE F.Q. Wei, K. Dreyer, U. Fehlmann, J.L. Pochon and A. Wrulich SLS / Paul Scherrer Institute CH5232 Villigen PSI Switzerland ABSTRACT The Swiss Light Source

More information

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film As clinics strive to treat with the smallest field possible, the demand for SRS/SBRT QA grows as well.

More information

REDI. M. Wind (SL), P. Beck (SL), M. Latocha (SL), S. Metzger (INT), M. Poizat(ESA), M. Steffens (INT)

REDI. M. Wind (SL), P. Beck (SL), M. Latocha (SL), S. Metzger (INT), M. Poizat(ESA), M. Steffens (INT) REDI Radiation evaluation of digital isolators currently available, suitable for space missions in terms of radiation tolerance (TID and SEE) including the JUICE mission M. Wind (SL), P. Beck (SL), M.

More information

Initial setup and subsequent temporal position monitoring using implanted RF transponders

Initial setup and subsequent temporal position monitoring using implanted RF transponders Initial setup and subsequent temporal position monitoring using implanted RF transponders James Balter, Ph.D. University of Michigan Has financial interest in Calypso Medical Technologies Acknowledgements

More information

Application of STPA in Radiation Therapy: a Preliminary Study

Application of STPA in Radiation Therapy: a Preliminary Study Application of STPA in Radiation Therapy: a Preliminary Study Natalia Silvis-Cividjian Wilko Verbakel Marjan Admiraal MIT STAMP Workshop 2018 VU medical center Vrije Universiteit (VU) campus Amsterdam,

More information

Interfaces with MPS/PSS

Interfaces with MPS/PSS Interfaces with / European Spallation Source Accelerator Division TB, 16 November 2016, Lund, Sweden / interfaces 1/21 Outline 1 Introduction 2 3 4 Conclusions / interfaces 2/21 Outline 1 Introduction

More information

Herwig Schopper CERN 1211 Geneva 23, Switzerland. Introduction

Herwig Schopper CERN 1211 Geneva 23, Switzerland. Introduction THE LEP PROJECT - STATUS REPORT Herwig Schopper CERN 1211 Geneva 23, Switzerland Introduction LEP is an e + e - collider ring designed and optimized for 2 100 GeV. In an initial phase an energy of 2 55

More information

COCIR SELF-REGULATORY INITIATIVE FOR MEDICAL IMAGING EQUIPMENT COMPUTED TOMOGRAPHY MEASUREMENT OF ENERGY CONSUMPTION

COCIR SELF-REGULATORY INITIATIVE FOR MEDICAL IMAGING EQUIPMENT COMPUTED TOMOGRAPHY MEASUREMENT OF ENERGY CONSUMPTION COCIR SELF-REGULATORY INITIATIVE FOR MEDICAL IMAGING EQUIPMENT COMPUTED TOMOGRAPHY MEASUREMENT OF ENERGY CONSUMPTION Revision: 1 Date: June 2015 Approved: June 2015 TABLE OF CONTENT 1. INTRODUCTION...

More information

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Indra J. Das, PhD, FACR Department of Radiation Oncology Indiana University of School of Medicine & Midwest Proton Radiation

More information

Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator

Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator Aleksei E. Zhdanov 1 and Leonid G. Dorosinskiy 1 Ural Federal University named after the first President of Russia B. N.

More information

(Geant4) Monte Carlo benchmarking. Harald Paganetti

(Geant4) Monte Carlo benchmarking. Harald Paganetti (Geant4) Monte Carlo benchmarking Harald Paganetti Why benchmarking? 1. We are in Medical Physics! 2. Monte Carlo is considered to be the gold standard 3. There are various competing codes/algorithms 4.

More information

Applying systems thinking to safety assurance of Nuclear Power Plants

Applying systems thinking to safety assurance of Nuclear Power Plants Applying systems thinking to safety assurance of Nuclear Power Plants Francisco Luiz de Lemos Instituto de Pesquisas Energeticas/ Comissao Nacional de Energia Nuclear IPEN/CNEN _ Brazil IMPRO Dialog Forum

More information

MuCool Test Area Experimental Program Summary

MuCool Test Area Experimental Program Summary MuCool Test Area Experimental Program Summary Alexey Kochemirovskiy The University of Chicago/Fermilab Alexey Kochemirovskiy NuFact'16 (Quy Nhon, August 21-27, 2016) Outline Introduction Motivation MTA

More information

Cosmic Rays in MoNA. Eric Johnson 8/08/03

Cosmic Rays in MoNA. Eric Johnson 8/08/03 Cosmic Rays in MoNA Eric Johnson 8/08/03 National Superconducting Cyclotron Laboratory Department of Physics and Astronomy Michigan State University Advisors: Michael Thoennessen and Thomas Baumann Abstract:

More information

Test (Irradiate) Delivered Parts

Test (Irradiate) Delivered Parts Radiation Hardness Evaluation of the Analog Devices AD9042 ADC for use in the CMS Electromagnetic Calorimeter P. Denes, B. Lev, R. Wixted Physics Department, Princeton University, Princeton NJ 08544, USA

More information

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array José A. Bencomo, * Geoffrey Ibbott, Seungsoo Lee, and Joao A. Borges Department of Radiation Physics.

More information

FLASH at DESY. FLASH. Free-Electron Laser in Hamburg. The first soft X-ray FEL operating two undulator beamlines simultaneously

FLASH at DESY. FLASH. Free-Electron Laser in Hamburg. The first soft X-ray FEL operating two undulator beamlines simultaneously FLASH at DESY The first soft X-ray FEL operating two undulator beamlines simultaneously Katja Honkavaara, DESY for the FLASH team FEL Conference 2014, Basel 25-29 August, 2014 First Lasing FLASH2 > First

More information

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film With improvements in targeting and localization, stereotactic treatments have become prevalent. To

More information

Superconducting Medical Accelerators at IBA

Superconducting Medical Accelerators at IBA Superconducting Medical Accelerators at IBA Wiel Kleeven and Eric Forton on behalf of IBA 1), AIMA 2) JINR 3) ASG 4) Sigmaphi 5) 1) Ion Beam Applications, Louvain-la-Neuve, Belgium 2) AIMA, Developpement,

More information

The on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO)

The on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO) The on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO) A. Ansarinejad1,2, A. Attili1, F. Bourhaleb2,R. Cirio1,2,M. Donetti1,3, M. A. Garella1, S. Giordanengo1,

More information

1.8 MW Upgrade of the PSI Proton Accelerator Facility

1.8 MW Upgrade of the PSI Proton Accelerator Facility 1.8 MW Upgrade of the PSI Proton Accelerator Facility Pierre A. Schmelzbach for the PSI Accelerator Divisions This talk: analyzes the potential for improvements from the ion source to the spallation target

More information

Irradiation Measurements of the Hitachi H8S/2357 MCU.

Irradiation Measurements of the Hitachi H8S/2357 MCU. Irradiation Measurements of the Hitachi H8S/2357 MCU. A. Ferrando 1, C.F. Figueroa 2, J.M. Luque 1, A. Molinero 1, J.J. Navarrete 1, J.C. Oller 1 1 CIEMAT, Avda Complutense 22, 28040 Madrid, Spain 2 IFCA,

More information

Maurizio Vretenar Linac4 Project Leader EuCARD-2 Coordinator

Maurizio Vretenar Linac4 Project Leader EuCARD-2 Coordinator Maurizio Vretenar Linac4 Project Leader EuCARD-2 Coordinator Every accelerator needs a linac as injector to pass the region where the velocity of the particles increases with energy. At high energies (relativity)

More information

rones-vulnerable-to-terrorist-hijackingresearchers-say/

rones-vulnerable-to-terrorist-hijackingresearchers-say/ http://www.youtube.com/v/jkbabvnunw0 http://www.foxnews.com/tech/2012/06/25/d rones-vulnerable-to-terrorist-hijackingresearchers-say/ 1 The Next Step: A Fully Integrated Global Multi-Modal Security and

More information

Monte Carlo Simulation of the PRad Experiment at JLab 1

Monte Carlo Simulation of the PRad Experiment at JLab 1 Monte Carlo Simulation of the PRad Experiment at JLab 1 Li Ye Mississippi State University for the PRad collaboration 1.This work is supported in part by NSF MRI award PHY-1229153, the U.S. Department

More information

The SIRAD irradiation facility at the INFN - Legnaro National Laboratory

The SIRAD irradiation facility at the INFN - Legnaro National Laboratory The SIRAD irradiation facility at the INFN - Legnaro National Laboratory I. Introduction 2 The INFN - Legnaro National Laboratory (LNL) SIRAD beamline http://www.lnl.infn.it 3 What is SIRAD? SIRAD is the

More information

Electron Linacs for Cargo Inspection and Other Industrial Applications

Electron Linacs for Cargo Inspection and Other Industrial Applications Electron Linacs for Cargo Inspection and Other Industrial Applications Chuanxiang Tang 1, Huaibi Chen 1,Yaohong Liu 2 Tang.xuh@tsinghua.edu.cn 1 Department of Engineering Physics, Tsinghua U., Beijing

More information

IN THE UNITED STATES DISTRICT COURT FOR THE DISTRICT OF DELAWARE ) ) ) ) ) ) ) ) ) ) ) COMPLAINT FOR PATENT INFRINGEMENT

IN THE UNITED STATES DISTRICT COURT FOR THE DISTRICT OF DELAWARE ) ) ) ) ) ) ) ) ) ) ) COMPLAINT FOR PATENT INFRINGEMENT IN THE UNITED STATES DISTRICT COURT FOR THE DISTRICT OF DELAWARE BEST MEDICAL INTERNATIONAL, INC., v. Plaintiff, VARIAN MEDICAL SYSTEMS, INC., AND VARIAN MEDICAL SYSTEMS INTERNATIONAL AG, Defendants. )

More information

200 MHz 350 MHz 750 MHz Linac2 RFQ2 202 MHz 0.5 MeV /m Weight : 1000 kg/m Ext. diameter : 45 cm

200 MHz 350 MHz 750 MHz Linac2 RFQ2 202 MHz 0.5 MeV /m Weight : 1000 kg/m Ext. diameter : 45 cm M. Vretenar, CERN for the HF-RFQ Working Group (V.A. Dimov, M. Garlasché, A. Grudiev, B. Koubek, A.M. Lombardi, S. Mathot, D. Mazur, E. Montesinos, M. Timmins, M. Vretenar) 1 1988-92 Linac2 RFQ2 202 MHz

More information

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS 7:28-14.1 Scope (a) This subchapter covers therapeutic installations used in the healing arts. These therapeutic installations include x-ray, accelerator and teletherapy

More information

New Tracking Gantry-Synchrotron Idea. G H Rees, ASTeC, RAL, U.K,

New Tracking Gantry-Synchrotron Idea. G H Rees, ASTeC, RAL, U.K, New Tracking Gantry-Synchrotron Idea G H Rees, ASTeC, RAL, U.K, Scheme makes use of the following: simple synchrotron and gantry magnet lattices series connection of magnets for 5 Hz tracking one main

More information

Phantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging

Phantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging in partnership with Phantoms in Medical Physics (RT) U. Oelfke Division of Radiotherapy & Imaging uwe.oelfke@icr.ac.uk Making the discoveries that defeat cancer 1. Introduction What is a phantom? Wiki:

More information

Initial Results from a Cryogenic Proton Irradiation of a p-channel CCD

Initial Results from a Cryogenic Proton Irradiation of a p-channel CCD Centre for Electronic Imaging Initial Results from a Cryogenic Proton Irradiation of a p-channel CCD Jason Gow Daniel Wood, David Hall, Ben Dryer, Simeon Barber, Andrew Holland and Neil Murray Jason P.

More information

Five years of operational experience with digitally controlled Power Supplies for beam control at the Paul Scherrer Institut (PSI)

Five years of operational experience with digitally controlled Power Supplies for beam control at the Paul Scherrer Institut (PSI) Five years of operational experience with digitally controlled Power Supplies for beam control at the Paul Scherrer Institut (PSI) Keywords F. Jenni, R. Künzi, A. Lüdeke 1, L. Tanner 2 PSI, Paul Scherrer

More information

MR3T LASER SYSTEMS IN RT ESSENTIAL FOR PRECISE PATIENT ALIGNMENT

MR3T LASER SYSTEMS IN RT ESSENTIAL FOR PRECISE PATIENT ALIGNMENT MR3T LASER SYSTEMS IN RT ESSENTIAL FOR PRECISE PATIENT ALIGNMENT LAP IS YOUR IDEAL PARTNER FOR YOUR EXTERNAL LASER SYSTEMS OUR CUSTOMIZED MR3T LASER BRIDGE SYSTEMS ARE RECOMMENDED BY LEADING MRI SCANNER

More information

X3D in Radiation Therapy Procedure Planning. Felix G. Hamza-Lup, Ph.D. Computer Science Armstrong Atlantic State University Savannah, Georgia USA

X3D in Radiation Therapy Procedure Planning. Felix G. Hamza-Lup, Ph.D. Computer Science Armstrong Atlantic State University Savannah, Georgia USA X3D in Radiation Therapy Procedure Planning Felix G. Hamza-Lup, Ph.D. Computer Science Armstrong Atlantic State University Savannah, Georgia USA Outline 1. What is radiation therapy? 2. Treatment planning

More information

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images Aiping Ding, Bin Han, Lei Wang, Lei Xing Department of Radiation Oncology, Stanford University School of

More information

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests Acknowledgements David Shepard, Ph.D. Daliang Cao, Ph.D. Muhammad K. N. Afghan, Ph.D. Jinsong Ye, M.S. Tony P. Wong, Ph.D. Fan Chen, Ph.D. Min Rao, Ph.D. Vivek Mehta, M.D. Igor Gomola, Ph.D. David Housley

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 62220-1 First edition 2003-10 Medical electrical equipment Characteristics of digital X-ray imaging devices Part 1: Determination of the detective quantum efficiency Appareils

More information

Proton beam for UCN. UCN TAC-Meeting, May 12-13, 2005 Urs Rohrer, beam line physicist

Proton beam for UCN. UCN TAC-Meeting, May 12-13, 2005 Urs Rohrer, beam line physicist Proton beam for UCN UCN TAC-Meeting, May 12-13, 2005 Urs Rohrer, beam line physicist PSI Accelerator Division Department of Large Research Facilities Introduction Important parameters of the PSI proton

More information

Energy Monitoring Device for Electron Beam Facilities

Energy Monitoring Device for Electron Beam Facilities 1 SM/EB-18 Energy Monitoring Device for Electron Beam Facilities M. Lavalle 1, P.G. Fuochi 1, A. Martelli 1, U. Corda 1, A. Kovács 2, K. Mehta 3, F. Kuntz 4, S. Plumeri 4 1 CNR-ISOF, Via P. Gobetti 101,

More information

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL TORNIER BLUEPRINT 3D Planning + PSI SCAN PROTOCOL Contents 3 Introduction 3 Patient preparation 3 Scanning instructions 4 Image instructions 5 Scanning parameters 6 Technical instructions 2 BLUEPRINT 3D

More information

Think About Control Fundamentals Training. Terminology Control. Eko Harsono Control Fundamental

Think About Control Fundamentals Training. Terminology Control. Eko Harsono Control Fundamental Think About Control Fundamentals Training Terminology Control Eko Harsono eko.harsononus@gmail.com; 1 Contents Topics: Slide No: Process Control Terminology 3-10 Control Principles 11-18 Basic Control

More information

SCINTILLATION FIBER DETECTOR FOR DOSE VERIFICATION IN PROTON THERAPY

SCINTILLATION FIBER DETECTOR FOR DOSE VERIFICATION IN PROTON THERAPY SCINTILLATION FIBER DETECTOR FOR DOSE VERIFICATION IN PROTON THERAPY Suhyun Lee Nuclear Physics Lab. Korea University CONTENTS Introduction Detector - Scintillation-fiber array - Electronics Beam Test

More information

Dose Delivery Instrumentation

Dose Delivery Instrumentation Dose Delivery Instrumentation Simona Giordanengo Istituto Nazionale di Fisica Nucleare Torino 28 Maggio 2015 CERN Accelerator School Vienna - Dose Delivery Instrumentation - Simona Giordanengo INFN Torino

More information

NIRS accelerator present and future activities

NIRS accelerator present and future activities NIRS accelerator present and future activities Koji Noda Research Center for Charged Particle Therapy National Institute of Radiological Sciences 2 nd Workshop on Hadron Therapy of Cancer, Erice, 23rd

More information

An Integrated Approach to Requirements Development and Hazard Analysis

An Integrated Approach to Requirements Development and Hazard Analysis An Integrated Approach to Requirements Development and Hazard Analysis John Thomas, John Sgueglia, Dajiang Suo, and Nancy Leveson Massachusetts Institute of Technology 2015-01-0274 Published 04/14/2015

More information

Degradation of electrical properties of coaxial cables exposed to ionizing radiation

Degradation of electrical properties of coaxial cables exposed to ionizing radiation 3 elspec GmbH Degradation of electrical properties of coaxial cables exposed to ionizing radiation Dipl. Ing. (FH) Stefan Burger DeltaGamma RF-Expert Dr. Michael Kuntzsch, Helmholtz-Zentrum Dresden-Rossendorf

More information

PRELIMINARY. The following table outlines the specifications of our standard tunable 2D-mirror MR Custom mirror coatings are possible.

PRELIMINARY. The following table outlines the specifications of our standard tunable 2D-mirror MR Custom mirror coatings are possible. Datasheet: MR-1-3 Copyright 212 Optotune Dual axis mirror with position feedback MR-1-3 Optotune s dual axis mirror series MR-1-3 is the ideal choice for applications that require large deflections in

More information

High Power ISOL Radioactive Target Remote Handling at TRIUMF. Abstract

High Power ISOL Radioactive Target Remote Handling at TRIUMF. Abstract High Power ISOL Radioactive Target Remote Handling at TRIUMF Grant Minor 1, Isaac Earle 1, Alexander Gottberg 1 1 TRIUMF, 4004 Wesbrook Mall, Vancouver, BC, V6T2A3, Canada Abstract TRIUMF, Canada's national

More information

5th International Symposium on NDT in Aerospace, 13-15th November 2013, Singapore

5th International Symposium on NDT in Aerospace, 13-15th November 2013, Singapore 5th International Symposium on NDT in Aerospace, 13-15th November 2013, Singapore Accurate Area and Width Measurements from Neutron Beam Computed Tomography (NBCT) Images of Cooling Holes of a Used Tornado

More information

12 GeV Upgrade Project DESIGN SOLUTIONS DOCUMENT. Upgrade Hall A

12 GeV Upgrade Project DESIGN SOLUTIONS DOCUMENT. Upgrade Hall A 12 GeV Upgrade Project DESIGN SOLUTIONS DOCUMENT Upgrade Hall A Version 1.2 July 28, 2010 DESIGN SOLUTIONS DOCUMENT Upgrade Hall A APPROVALS Approved by: 12 GeV Upgrade Control Account Manager, Hall A

More information

Single Event Effects Testing of the ISL7124SRH Quad Operational Amplifier June 2002

Single Event Effects Testing of the ISL7124SRH Quad Operational Amplifier June 2002 Single Event Effects Testing of the ISL7124SRH Quad Operational Amplifier June 2002 Purpose - This report describes the results of single event effects testing of the ISL7124SRH quad operational amplifier

More information

CyberKnife Iris Beam QA using Fluence Divergence

CyberKnife Iris Beam QA using Fluence Divergence CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery

More information

IFE/HR/E-2017/002. Human factors in the design of control rooms for ESS

IFE/HR/E-2017/002. Human factors in the design of control rooms for ESS IFE/HR/E-2017/002 Human factors in the design of control rooms for ESS Report number ISSN Revision number Date IFE/HR/E-2017/002 0333-2039 2017-05-11 Client/ Client reference: ISBN Number of issues Number

More information

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators Implementing a new digital medical accelerator John Wong Johns Hopkins University AAPM, Austin, 2014 Acknowledgements Yin Zhang, Ken Wang, Kai Ding (Commissioning - JHU) Esteban Velarde, Joe Moore (QA

More information

The APOLLO Project: LV Power Supplies For The Next High Energy Physics Experiments

The APOLLO Project: LV Power Supplies For The Next High Energy Physics Experiments The A roject: V ower Supplies For The Next High Energy hysics Experiments Milano, ctober 19, 2011 Agostino anza, on behalf of the A Collaboration M. Alderighi (1,6), M. Citterio (1), M. Riva (1,8),. Cova

More information

Large-Area CdTe Photon-Counting Pixel Detectors

Large-Area CdTe Photon-Counting Pixel Detectors Large-Area CdTe Photon-Counting Pixel Detectors Tilman Donath, Application Scientist 22.6.2015, DIR2015, Ghent DECTRIS Ltd. 5400 Baden Switzerland www.dectris.com Agenda 1. Introduction Hybrid Photon Counting

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

OPTIMIZED MAGNET FOR A 250 MEV PROTON RADIOTHERAPY CYCLOTRON

OPTIMIZED MAGNET FOR A 250 MEV PROTON RADIOTHERAPY CYCLOTRON OPTIMIZED MAGNET FOR A 250 MEV PROTON RADIOTHERAPY CYCLOTRON J. Kim and H. Blosser 1. Introduction The design of a K250 superconducting cyclotron has been recently improved from the original design of

More information

Heavy Ion Test Report for the MSK5063RH Switching Regulator with the RH3845 and RH411 Die

Heavy Ion Test Report for the MSK5063RH Switching Regulator with the RH3845 and RH411 Die Heavy Ion Test Report for the MSK5063RH Switching Regulator with the RH3845 and RH411 Die Shirley Hart 1, Paul Musil 2, David Beebe 3, and Bryan Horton 2 Report prepared by: Dakai Chen 3 1. Previously

More information

Test results on 60 MeV proton beam at CYCLONE - UCL Performed on CAEN HV prototype module A June 2001 Introduction

Test results on 60 MeV proton beam at CYCLONE - UCL Performed on CAEN HV prototype module A June 2001 Introduction Test results on 60 MeV proton beam at CYCLONE - UCL Performed on CAEN HV prototype module A877 27-28 June 2001 (M. De Giorgi, M. Verlato INFN Padova, G. Passuello CAEN spa) Introduction The test performed

More information

Module Integration Sensor Requirements

Module Integration Sensor Requirements Module Integration Sensor Requirements Phil Allport Module Integration Working Group Sensor Geometry and Bond Pads Module Programme Issues Numbers of Sensors Required Nobu s Sensor Size Summary n.b. 98.99

More information

A systems approach to risk analysis of maritime operations

A systems approach to risk analysis of maritime operations A systems approach to risk analysis of maritime operations Børge Rokseth 1*, Ingrid Bouwer Utne 1, Jan Erik Vinnem 1 1 Norwegian University of Science and Technology (NTNU), Department of Marine Technology

More information

Radiological Safety Analysis Document for the CLAS12 Engineering and the first physics run of Run Group A

Radiological Safety Analysis Document for the CLAS12 Engineering and the first physics run of Run Group A Radiological Safety Analysis Document for the CLAS12 Engineering and the first physics run of Run Group A This Radiological Safety Analysis Document (RSAD) will identify the general conditions associated

More information

Overview of SINAP Timing System

Overview of SINAP Timing System Overview of SINAP Timing System Ming Liu Electronics Group, Division of Beam Diagnostics & Control Technology Shanghai Synchrotron Radiation Facility Outline System design Hardware modules Performance

More information

Michael Budde Head of Physics Design group Danfysik Taastrup, Denmark

Michael Budde Head of Physics Design group Danfysik Taastrup, Denmark Michael Budde Head of Physics Design group Danfysik Taastrup, Denmark Accelerating Technology Business We provide high performance particle accelerators and related equipment for research, healthcare and

More information

US 7,688,938 B2 Mar.30,2010

US 7,688,938 B2 Mar.30,2010 I 1111111111111111 11111 lllll lllll 111111111111111 1111111111111111 IIII IIII IIII US007688938B2 c12) United States Patent Paliwal et al. (IO) Patent No.: (45) Date of Patent: US 7,688,938 B2 Mar.30,2010

More information

FRAUNHOFER INSTITUTE FOR INTEGRATED CIRCUITS IIS. MANUAL PANORAMIC MICROSCOPY WITH istix

FRAUNHOFER INSTITUTE FOR INTEGRATED CIRCUITS IIS. MANUAL PANORAMIC MICROSCOPY WITH istix FRAUNHOFER INSTITUTE FOR INTEGRATED CIRCUITS IIS MANUAL PANORAMIC MICROSCOPY WITH istix CLINICAL DIAGNOSTICS AND MATERIAL SCIENCES IMPROVED BY DIGITAL MICROSCOPY B A C K G R O U N D Due to a high grade

More information

Modelling and Hazard Analysis for Contaminated Sediments Using STAMP Model

Modelling and Hazard Analysis for Contaminated Sediments Using STAMP Model Publications 5-2011 Modelling and Hazard Analysis for Contaminated Sediments Using STAMP Model Karim Hardy Mines Paris Tech, hardyk1@erau.edu Franck Guarnieri Mines ParisTech Follow this and additional

More information

Test of a Multilayer Dose-Verification Gaseous Detector with Raster Scan Mode Proton Beams

Test of a Multilayer Dose-Verification Gaseous Detector with Raster Scan Mode Proton Beams Test of a Multilayer Dose-Verification Gaseous Detector with Raster Scan Mode Proton Beams Kyong Sei Lee*, Byungsik Hong, Minho Kang, Seunkyung Lee, and Sung Keun Park Korea University, Seoul 136-713 Sung

More information

Overview over CERN SPS test beams

Overview over CERN SPS test beams Overview over CERN SPS test beams A. Gerbershagen On behalf of CERN Experimental Areas Group 17/1/2018 A. Gerbershagen - Overview over CERN SPS secondary beams 1 CERN Accelerator Complex SPS : protons/ions

More information

Volumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA

Volumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA Volumetric Modulated Arc Therapy David Shepard Swedish Cancer Institute Seattle, WA Disclaimer Our VMAT work has been sponsored in part by Elekta. Outline David Shepard VMAT Basics and VMAT Plan Quality

More information