TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL
|
|
- Terence McDowell
- 6 years ago
- Views:
Transcription
1 TORNIER BLUEPRINT 3D Planning + PSI SCAN PROTOCOL
2 Contents 3 Introduction 3 Patient preparation 3 Scanning instructions 4 Image instructions 5 Scanning parameters 6 Technical instructions 2 BLUEPRINT 3D Planning Software Scan Protocol
3 Scan Protocol Introduction This document describes the guidelines for a Shoulder CT-Scan to be processed by BLUEPRINT 3D Planning Software. Following this scanning protocol will optimize the likelihood of successful scan reconstruction in BLUEPRINT 3D Planning Software. Note: Please contact the helpline if you require further information. Figure 1: Artifacts generated by metallic implants Patient Preparation» Verify that metal is not present within the scanned portion of the operative side (Figure 1) Note: Pacemakers, dental and cervical hardware can cause artifact if too close to the GlenoHumeral joint. If artifact too intense, do not scan.» Do NOT use contrast material» Place the patient supine on the scanner table, Humerus along the trunk (Figure 2) Figure 2: Position of the patient in the machine Note: Position must be craniocaudal oriented (head first) and iso-centered, with the hand in a neutral position.» It is strongly recommended to place a sponge or a small pillow between humerus and trunk Scanning instructions» No gantry tilt» Limit the field of view to the desired/operative shoulder Figure 3: Example of incomplete projections» Avoid incomplete projections of soft-tissue wrap by having all shoulder softtissue within the field of view (causes pseudo-metal artifact) (Figure 3)» It is strongly recommended to perform a full scan of the scapula (Figure 4) Note: Scan must include several slices about the AC joint and several slices (around 1 cm) below the inferior angle.» All slices must have the same field of view and the same slice spacing» DO NOT use auto-ma setting Figure 4: Full scan of the scapula BLUEPRINT 3D Planning Software Scan Protocol 3
4 Images instructions» Use only axial images, no other reconstructions or reformations in different planes (sagittal or coronal)» The DICOM file format is defined in the Transfer Syntax (0002, 0010) Tag field, the supported formats are the following: Uncompressed Transfer Syntax UID Description Implicit VR Little Endian Explicit VR Little Endian Explicit VR Big Endian Lossless compressed Transfer Syntax UID Description RLE Lossless JPEG Lossless, Non-Hierarchical (Process 14) JPEG Lossless, Hierarchical, First-Order Prediction JPEG 2000 Image Compression (Lossless Only) Lossy compressed Transfer Syntax UID Description JPEG 2000 Image Compression 4 BLUEPRINT 3D Planning Software Scan Protocol
5 Scanning parameters» Recommended parameters Parameter Modality Kernel / Algorithm kvp ma Slice Increment Image Thickness Exposure time Recommended CT Bone or Bone+ 120 or 140 kvp Manual 240 ma DO NOT use auto-ma 1.2 mm - Detector Coverage should be maximum - Helical Thickness mm or 1.25 mm - Pitch 0.9 or less - Rotation time 1 sec or less 1000 ms Note: Using the manual-ma setting of 240 ma will provide compatible images. Not doing so will produce images that are incompatible with BLUEPRINT 3D Planning Software.» Display Field Of View (DFOV) Parameter DFOV (axial plane) (Figure 5) DFOV (frontal plane) (Figure 6) Recommended Please include the entire scapula 25 cm or no more 32 cm Please include the entire scapula More than 10 cm Matrix size 512 x 512 Figure 5: DFOV (axial plane) Figure 6: DFOV (frontal plane) BLUEPRINT 3D Planning Software Scan Protocol 5
6 Technical Instructions» GE HEALTHCARE In order to define the recommended ma, please follow the next steps: 1. Click on ma button 2. Set manual-ma to 240 ma and dose reduction 0% 3. Set the desired slice thickness for helical CT by clicking on Slice Thickness: - Set the detector coverage to maximum depending on each machine - Set the helical thickness to mm or no more than 1.25 mm - Set the pitch to 0.9 or less - Set the rotation time to 1.0 sec» SIEMENS This example works for Somatom Sensation 64 but the user interface is the same in most of Siemens systems. You have two main tabs to configure the scan protocol on Siemens machines: 1. On the Reconstruction tab: - Set the Kernel/Algorithm to Bone 70s, sharp or very sharp - Set the window on Bone - Set the reconstruction increment to 0.6 mm or less 2. On the Scan tab: - Set the quality ref mas to 300 mas - Set the kvp to 120 or 140 kvp - Set the rotation time to 1.0 sec - Set the slice thickness to 1.25 mm or less - Set the pitch to 0.9 or less» OTHER SCANNER BRANDS For any other brand of CT scan machine please refer to the standard recommendations in pages 4, 5 and 6. 6 BLUEPRINT 3D Planning Software Scan Protocol
7 Notes BLUEPRINT 3D Planning Software Scan Protocol 7
8 BLUEPRINT 3D PLANNING SOFTWARE HELPLINE Phone: Managed 8:00am 5:00pm CST Nesbitt Avenue South Bloomington, MN Rue Lavoisier Montbonnot Saint Martin France +33 (0) Proper surgical procedures and techniques are the responsibility of the medical professional. This material is furnished for information purposes only. Each surgeon must evaluate the appropriateness of the material based on his or her personal medical training and experience. Prior to use of any Tornier implant system, the surgeon should refer to the product package insert for complete warnings, precautions, indications, contraindications, and adverse effects. Package inserts are also available by contacting Wright. Contact information can be found in this document and the package insert. The BLUEPRINT Glenoid Guides are intended to be used as surgical instruments to assist in the intraoperative positioning of glenoid components used with total anatomic or reversed should arthroplasty procedures. and denote Trademarks and Registered Trademarks of Wright Medical Group N.V. or its affiliates Wright Medical Group N.V. or its affiliates. All Rights Reserved. CAW-7230 Rev C DCR Feb-2017
surgical technique addendum
s h o u l d e r Solutions by Tornier Aequalis reversed II shoulder system surgical technique addendum Aequalis reversed II threaded post baseplate surgical technique addendum s u r g i c a l t e c h n
More informationTORNIER MAXLOCK EXTREME. Clavicle Plating System SURGICAL TECHNIQUE
TORNIER MAXLOCK EXTREME Clavicle Plating System SURGICAL TECHNIQUE 2 Table of Contents: Key Design Features...4 Surgical Technique...5 Implants & Instruments...9 3 Key Design Features There are 7 anatomically
More informationTOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions
CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT
More informationDICOM Conformance Statement
DICOM Conformance Statement Application Annex: MR-CT Roadmap R1.1 On Interventional Workspot R1.4 Koninklijke Philips N.V. 2017 All rights are reserved. ICAP-T-030001.09b Corresponds to ICAP-W-030001.02
More informationORTHOLINK. 2 - Hole Wedge Osteotomy Plate SURGICAL TECHNIQUE
MAXLOCK EXTREME ORTHOLINK 2 - Hole Wedge Osteotomy Plate SURGICAL TECHNIQUE Contents Key Design Features 3 Surgical Technique 4 Implants and Instruments 7 Proper surgical procedures and techniques are
More informationDistal Fibula Plate SURGICAL TECHNIQUE
MAXLOCK EXTREME Distal Fibula Plate SURGICAL TECHNIQUE Contents Key Design Features 3 Surgical Technique 4 Implants and Instruments 8 Proper surgical procedures and techniques are the responsibility of
More informationISO Plate SURGICAL TECHNIQUE
MINI MAXLOCK EXTREME ISO Plate SURGICAL TECHNIQUE Contents Table of Contents Key Design Features 2 Surgical Technique 3 Implants and Instruments 8 Key Design Features The MINI MAXLOCK EXTREME ISO (Intraosseous
More informationSmall Plate and Screw System SURGICAL TECHNIQUE
MINI MAXLOCK EXTREME Small Plate and Screw System SURGICAL TECHNIQUE Contents Key Design Features 2 Surgical Technique 3 Implants and Instruments 9 Proper surgical procedures and techniques are the responsibility
More informationMaxTorque. surgical technique. Cannulated Screw System. Foot & Ankle. OrthoHelix Technology
MaxTorque Cannulated Screw System OrthoHelix Technology surgical technique Foot & Ankle 2 M A X T O R Q U E C A N N U L A T E D S C R E W S Y S T E M Table of Contents Advantages 3 Indications 4 Contraindications
More information1. Patient size AEC. Large Patient High ma. Small Patient Low ma
Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description
More informationConformance Statement for DICOM Viewer
MedDream Conformance Statement for DICOM Viewer (version 5.00) 2015, Softneta UAB, Kaunas All rights reserved in the event of granting of patents or registration as a utility patent. All names of companies
More informationAutomated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London
Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients
More informationMEDIALMAX System SURGICAL TECHNIQUE
MAXLOCK EXTREME MEDIALMAX System SURGICAL TECHNIQUE Contents The MEDIALMAX Advantage Design Feature Various flex options POCKETLOCK Technology Anatomic design Advantage Provides the surgeon with multiple
More informationPitfalls and Remedies of MDCT Scanners as Quantitative Instruments
intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature
More information2D, 3D CT Intervention, and CT Fluoroscopy
2D, 3D CT Intervention, and CT Fluoroscopy SOMATOM Definition, Definition AS, Definition Flash Answers for life. Siemens CT Vision Siemens CT Vision The justification for the existence of the entire medical
More informationDICOM Conformance Statement
DICOM Conformance Statement Application Annex: 3D Roadmap R1.1.5 Koninklijke Philips N.V. 2015 All rights are reserved. Document Number: ICAP-PF.0015381 Issued by: Philips Medical Systems Nederland BV,
More informationMTP Set SURGICAL TECHNIQUE
MINI MAXLOCK EXTREME MTP Set SURGICAL TECHNIQUE Contents Table of Contents Key Design Features 3 Surgical Technique Standard MTP Plate 4 MTP Plate with POCKETLOCK Technology 10 Implants and Instruments
More informationDICOM Conformance Statement
DICOM Conformance Statement Application Annex: Stentboost R4.2.5 Koninklijke Philips N.V. 2015 All rights are reserved. Document Number: ICAP-PF.0015387 Issued by: Philips Medical Systems Nederland BV,
More informationConformance Statement for DICOM Viewer
MedDream Conformance Statement for DICOM Viewer (version 5.1) 2015, Softneta UAB, Kaunas All rights reserved in the event of granting of patents or registration as a utility patent. All names of companies
More informationDICOM Conformance Statement
DICOM Conformance Statement Application Annex: US Applications on Philips IntelliSpace Portal V6.0 Koninklijke Philips Electronics N.V. 2013 All rights are reserved. Document Number: PIIOffc.0001323.01
More informationCOMPUTED TOMOGRAPHY 1
COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the
More informationSoftware and Hardware in CCTA. Elly Castellano PhD
Software and Hardware in CCTA Elly Castellano PhD Outline technical requirements for coronary CTA the modern cardiac CT scanner ECG-gating technology image reconstruction algorithms 2 Technical requirements
More informationTranslating Protocols Between Scanner Manufacturer and Model
Translating Protocols Between Scanner Manufacturer and Model Robert J. Pizzutiello, MS, FAAPM, FACMP Sr. Vice-President, Global Physics Solutions President, Upstate Medical Physics Objectives Understand
More informationResearch Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.
Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National
More informationMaximum Performance, Minimum Space
TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.
More informationDICOM Conformance Statement
DICOM Conformance Statement Application Annex: US Applications on Philips IntelliSpace Portal V7.0 Koninklijke Philips N.V. 2014 All rights are reserved. Document Number: ICAP-PF.0013672 Issued by: Philips
More informationMaximizing clinical outcomes
Maximizing clinical outcomes Digital Tomosynthesis Dual Energy Subtraction Automated Long Length Imaging Improved image quality at a low dose Xray Xray Patented ISS capture technology promotes high sensitivity
More information12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees
Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees
More informationSlide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?
Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM
More informationHISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD
CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING
More informationWide-Detector CT for TAVR Planning:
Wide-Detector CT for TAVR Planning: Impact on Iodine Dose, Radiation Dose, and Image Quality SCBTMR 2015 Annual Course Thursday, October 8 William P. Shuman MD FSCBTMR Department of Radiology University
More informationHIP Solutions by Tornier MEIJE DUO 12/14 TAPER SURGICAL TECHNIQUE. surgical technique
HIP Solutions by Tornier MEIJE DUO 12/14 TAPER SURGICAL TECHNIQUE surgical technique MEIJE DUO 12/14 TAPER SURGICAL TECHNIQUE Meije duotm H I P P R O S T H E S I S 1. Locating the pelvic horizontal line
More informationDICOM Conformance Statement
DICOM Conformance Statement Application Annex: CT Applications on Philips IntelliSpace Portal V5.0 Koninklijke Philips Electronics N.V. 2012 All rights are reserved. Document Number: PIIOffc.0000143.01
More informationUniversal Humeral Nail
990210009 INDEX Indications Preoperative Planning Patient Position Surgical Technique - Step 1 Open Humerus - Step 2 Calibrate The Nail - Step 3 Insert Nail - Step 4 Proximal Locking - Step 5 Assemble
More informationDICOM Conformance Statement
DICOM Conformance Statement Application Annex: MultiModality Applications on Philips IntelliSpace Portal V7.0 Koninklijke Philips N.V. 2014 All rights are reserved. Document Number: ICAPPF.0013673 Issued
More informationCHARLOTTE. Multi-Use Compression Screw
CHARLOTTE Multi-Use Compression Screw CHARLOTTE multi-use compression screw surgical technique SURGICAL ADVISORS ROBERT ANDERSON, MD BRUCE COHEN, MD W. HODGES DAVIS, MD Proper surgical procedures and techniques
More informationQC Testing for Computed Tomography (CT) Scanner
QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including
More informationTechnique Guide. Modular Sternal Cable System. Flexibility and strength in sternal closure and repair.
Technique Guide Modular Sternal Cable System. Flexibility and strength in sternal closure and repair. Table of Contents Introduction Overview 2 Indications and Contraindications 3 Surgical Technique A.
More informationRedefining Ergonomics
Samsung Electronics Co., Ltd. inspires the world and shapes the future with transformative ideas and technologies, redefining the worlds of TVs, smartphones, wearable devices, tablets, cameras, digital
More informationDICOM3.0 Conformance Statement
DICOM3.0 Conformance Statement INTRODUCTION 1 IMPLEMENTATION MODEL 2 AE Specifications 3 DIRECT DIGITIZER Communication Profiles 4 Extensions/Specializations/Privatizations 5 Configuration 6 0197 Support
More informationDigiMam Conformance Statement for DICOM V3.0
DigiMam Conformance Statement for DICOM V3.0 Copyright 2004 by I.M.S. s.r.l. DOCUMENT VERSIONS Version Date Author Changes 1.00 15-Feb-05 IMS s.r.l. First Version DOCUMENT VERSIONS Page 2 of 29 TABLE OF
More informationDICOM Conformance Statement
DICOM Conformance Statement Application Annex: NM Applications on Philips IntelliSpace Portal V5.0 Koninklijke Philips Electronics N.V. 2012 All rights are reserved. Document Number: PIIOffc.0000147.00
More informationTechnique Guide. 4.5 mm Cannulated Screws. Part of the Synthes Cannulated Screw System.
Technique Guide 4.5 mm Cannulated Screws. Part of the Synthes Cannulated Screw System. TableofContents Introduction 4.5 mm Cannulated Screws 2 AO Principles 3 Indications 4 Surgical Technique Surgical
More informationImage Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose
Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation
More informationNeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results
NeuViz 16 Computed Tomography Elevating routine imaging for exceptional results Essence NeuViz 16 Raising the bar on clinical utility in routine imaging. Get more. More clinical information for patients.
More informationSOMATOM Esprit A Bundle of Energy
SOMATOM Esprit A Bundle of Energy DATA SOMATOM Esprit An economical CT scanner designed for...... Excellent spiral image quality... A wide range of clinical applications... Value performance and reliabilty
More informationQ3D. Speak to a 3D Specialist. CBCT 3D / Panoramic Imaging GENERAL DIMENSIONS. Suni Imaging Product Lines GET.
GENERAL Q3D Q3D Ceph Exposure Time FOV Voxel Size Focal Spot Target Angle Tube Voltage Tube Current Line Voltage Warranty Panoramic CT 9 to 17 sec 9 to 17 sec 4 to 12 sec 7.7/14.5 sec 7.7/14.5 sec 4 x
More informationDART-FIRE. Small Screw System SURGICAL TECHNIQUE
DART-FIRE Small Screw System SURGICAL TECHNIQUE DART-FIRE Small Screw System SURGICAL TECHNIQUE Contents Chapter 1 4 Chapter 2 6 Appendix 1 9 Appendix 2 11 Introduction DART-FIRE Small Screw System Surgical
More informationDART-FIRE. Small Screw System SURGICAL TECHNIQUE
DART-FIRE Small Screw System SURGICAL TECHNIQUE DART-FIRE Small Screw System Surgical Technique Contents Chapter 1 4 Chapter 2 6 Chapter 3 7 Appendix 1 10 Appendix 2 12 Introduction Intended Use DART-FIRE
More information7.0 mm Cannulated Screws
Part of the DePuy Synthes Cannulated Screw System 7.0 mm Cannulated Screws Surgical Technique Table of Contents Introduction 7.0 mm Cannulated Screws 2 AO Principles 3 Indications 4 Surgical Technique
More informationPart of the DePuy Synthes Cannulated Screw System. 3.5 mm Cannulated Screws
Part of the DePuy Synthes Cannulated Screw System 3.5 mm Cannulated Screws Surgical Technique Table of Contents Introduction 3.5 mm Cannulated Screws 2 AO Principles 3 Indications 4 Surgical Technique
More informationClinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11
Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) Hiroki Kadoya Yukiko Kitagawa MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria
More informationSpeedTip CCS 5.0, 7.0
SURGICAL TECHNIQUE STEP BY STEP SpeedTip CCS 5.0, 7.0 Cannulated Compression Screws APTUS 2 SpeedTip CCS 5.0, 7.0 Cannulated Compression Screws Contents 3 Introduction Product Materials Indications Contraindications
More informationOPERATIVE TECHNIQUE RIVAL REDUCE FRACTURE PLATING SYSTEM. foot & ankle trauma procedures
OPERATIVE TECHNIQUE RIVAL REDUCE FRACTURE PLATING SYSTEM foot & ankle trauma procedures INTRODUCTION 3 SYSTEM DESCRIPTION 3 TECHNICAL DETAILS 4 SALES AND MARKETING CONFIGURATION 5 OPERATIVE TECHNIQUE 7
More informationDICOM Conformance Statement
GE Medical Systems Kretz Ultrasound DICOM Conformance Statement 105952 Revision 0 VOLUSON 730Expert/Pro V. 4.0.x 0123 Copyright 2000, 2001, 2002, 2003, 2004 by GE Medical Systems Kretztechnik GmbH & Co
More informationQC by the MPE in Belgium
Acceptance testing of state-of-the-art CT scanners using a new national protocol: first experience on a large number of scanners of different make and model the working group Radiology of the Belgian Hospital
More informationFocal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution
Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Cynthia H. McCollough, PhD, DABR, FAAPM, FACR Director, CT Clinical Innovation Center Professor of Medical Physics and
More informationTechnique Guide. 7.0 mm Cannulated Screws. Part of the Synthes Cannulated Screw System.
Technique Guide 7.0 mm Cannulated Screws. Part of the Synthes Cannulated Screw System. Table of Contents Introduction 7.0 mm Cannulated Screws 2 AO Principles 3 Indications 4 Surgical Technique Surgical
More informationBreast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School
Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis
More informationVirage OCT Spinal Fixation System
Virage OCT Spinal Fixation System Virage OCT Spinal Fixation System Change Your Perspective Become a Part of the Posterior Fixation Revolution The Virage System is an Occipital-Cervico-Thoracic (OCT) spinal
More informationREGIUS CONSOLE CS-3. DICOM 3.0 Conformance Statement CODE NO Manufacturer: 1 Sakura-machi, Hino-shi Tokyo , Japan
REGIS CONSOLE CS-3 DICOM 3.0 Conformance Statement CODE NO. 0862 Manufacturer: 1 Sakura-machi, Hino-shi Tokyo 191-8511, Japan Revision History Date Version Description 1 Contents 1 INTRODCTION... 4 1.1
More informationReflex Hybrid System Overview
Spine Reflex Hybrid System Overview Anterior Cervical Plating System Introduction The Reflex Hybrid ACP System offers a low-profile anterior cervical plate along with a selection of bone screw types to
More informationPET/CT Instrumentation Basics
/ Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire
More informationCT Basics: Data Acquisition Module 3
Module 3 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.
More informationdiagnostic examination
RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011
More informationOPERATIVE TECHNIQUE RIVAL BITE HEADED CANNULATED AND HEADLESS COMPRESSION SCREWS. foot & ankle applications
OPERATIVE TECHNIQUE RIVAL BITE HEADED CANNULATED AND HEADLESS COMPRESSION SCREWS foot & ankle applications INTRODUCTION 3 SYSTEM DESCRIPTION 3 TECHNICAL DETAILS 4 SALES AND MARKETING CONFIGURATION 6 OPERATIVE
More informationComputed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology
The Fundamentals of... Computed Tomography Computed Tomography (CT) systems use x-rays to produce images of slices through a patient s anatomy. Despite having lower spatial resolution than other x-ray
More information160-slice CT SCANNER / New Standard for the Future
TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.
More informationDetector technology in simultaneous spectral imaging
Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become
More informationT h e P h a n t o m L a b o r a t o r y
T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order
More informationCHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY
14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to
More informationRADspeed Pro. EDGEpackage C501-E041C
RADspeed Pro EDGEpackage C501-E041C 2 Some of the FPDs may be not available in your country. Please contact us to check the availability in your country. 3 Tomosynthesis in the Standing Position Tomosynthesis
More informationMetal Artifact Reduction for Orthopedic Implants (O-MAR)
Metal Artifact Reduction for Orthopedic Implants () Summary Since the inception of CT, numerous methods have been proposed to suppress metal artifacts with varying degrees of success. 1-4 (Metal Artifact
More informationMulti-Access Biplane Lab
Multi-Access Biplane Lab Advanced technolo gies deliver optimized biplane imaging Designed in concert with leading physicians, the Infinix VF-i/BP provides advanced, versatile patient access to meet the
More informationDART-FIRE. Small Screw System SURGIC A L T ECHNIQUE
DART-FIRE Small Screw System SURGIC A L T ECHNIQUE Contents Headline Headline PREFACE Chapter 1 4 Chapter 2 6 Chapter 3 7 Appendix A 10 Appendix B 12 Introduction Intended Use DART-FIRE Small Screw System
More informationCETRA ANTERIOR CERVICAL PLATE
CETRA TM ANTERIOR CERVICAL PLATE The CETRA Anterior Cervical Plate System offers a low profile titanium alloy plate with an intuitive locking mechanism, large graft windows, a high degree of screw angulation
More informationYinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2
Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2 1. Department of Medical Physics, University of Wisconsin-Madison 2. Department
More informationTechnique Guide. Modular Sternal Cable System. Flexibility and strength in sternal closure and repair.
Technique Guide Modular Sternal Cable System. Flexibility and strength in sternal closure and repair. Table of Contents Introduction Modular Sternal Cable System 2 Indications 4 Modular Sternal Closure
More informationGE Healthcare. Essential for life. Senographe Essential Full-Field Digital Mammography system
GE Healthcare Essential for life Senographe Essential Full-Field Digital Mammography system Excellence in FFDM is a process. An ongoing quest, fueled by our continuing breakthroughs in breast cancer detection
More information2.4 mm Cannulated Screw. An integral part of the Synthes Cannulated Screw System (CSS).
2.4 mm Cannulated Screw. An integral part of the Synthes Cannulated Screw System (CSS). Surgical Technique This publication is not intended f distribution in the USA. and implants approved by the AO Foundation.
More informationChiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard.
Chiara Secco PET Performance measurements of the new LSO-Based Whole Body PET/CT Scanner biograph 16 HI-REZ using the NEMA NU 2-2001 Standard. INTRODUCTION Since its introduction, CT has become a fundamental
More informationKretztechnik AG. Voluson 730 Ultrasound System
Kretztechnik Voluson 730 Ultrasound System DICOM Conformance Statement Rev 1.02 Voluson 730 DICOM Coformance Rev 1.02 2001-07-18 Kretztechnik AG Zipf/Austria KRETZTECHNIK AG TIEFENBACH 15 Telefon: +43
More informationLCP Pilon Plate 2.7/3.5
LCP Pilon Plate 2.7/3.5 Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation. Table of contents Indications 2 Implants
More informationEnhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)
Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular
More informationIterative Reconstruction in Image Space. Answers for life.
Iterative Reconstruction in Image Space Answers for life. Iterative Reconstruction in Image Space * (IRIS) * Please note: IRIS is used as an abbreviation for Iterative Reconstruction in Image Space throughout
More informationINNOVATION BY DESIGN. Toshiba A History of Leadership REMOTE CONTROL R/F SYSTEM
INNOVATION BY DESIGN For over 130 years, Toshiba has led the world in developing technology to improve the quality of life. This Made for Life TM commitment is reflected in our family of leading-edge imaging
More informationData. microcat +SPECT
Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More information3.5 mm Cannulated Screw Technique Guide
3.5 mm Cannulated Screw Technique Guide An Integral Part of the SYNTHES Cannulated Screw System Original Instruments and Implants of the Association for the Study of Internal Fixation AO ASIF The 3.5 mm
More informationThe disclaimer on page 1 is an integral part of this document. Copyright November 30, 2017 by AAPM. All rights reserved.
DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION
More informationPatient-Assisted Compression Impact on Image Quality and Workflow
Patient-Assisted Compression Impact on Image Quality and Workflow Senographe Pristina In 2017, GE Healthcare s Senographe Pristina ( Pristina ) was approved by the FDA using the standard technologist-controlled
More informationCT Basics: Image Quality Module 6
Module 6 For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions. Any broadcast,
More informationGE Healthcare. Senographe 2000D Full-field digital mammography system
GE Healthcare Senographe 2000D Full-field digital mammography system Digital has arrived. The Senographe 2000D Full-Field Digital Mammography (FFDM) system gives you a unique competitive advantage. That
More informationOEC Brivo 785 Essential
GE Healthcare OEC Brivo 785 Essential Your essential partner in surgery. The OEC Brivo 785 Essential is not approved for sale in the US Introducing the OEC Brivo 785 Essential For more than three decades,
More informationCorrelation of 2D Reconstructed High Resolution CT Data of the Temporal Bone and Adjacent Structures to 3D Images
Correlation of 2D Reconstructed High Resolution CT Data of the Temporal Bone and Adjacent Structures to 3D Images Rodt T 1, Ratiu P 1, Becker H 2, Schmidt AM 2, Bartling S 2, O'Donnell L 3, Weber BP 2,
More informationX-RAYS - NO UNAUTHORISED ENTRY
Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are
More informationSurgical Technique 1
Surgical Technique 1 D-RAD SMART PACK Single-Use Volar Distal Radius Plating System Surgical Technique Table of Contents Indications... 3 Contraindications... 3 D-RAD SMART PACK product overview... 4 Instrumentation...
More informationOrthopedic Bone Nail System Universal Humeral Nail
Orthopedic Bone Nail System Universal Humeral Nail Surgical Technique Manual Note: The surgical procedures should be performed under the guidance of qualified skilled orthopedic surgeons, and this surgical
More informationReducing Radiation Exposure from Survey CT Scans
Reducing Survey CT Scan Exposure Pediatric Imaging Original Research Jennifer C. O Daniel 1 Donna M. Stevens 2 Dianna D. Cody 2 O Daniel JC, Stevens DM, Cody DD Received July 28, 2004; accepted after revision
More informationChampion Instrumentation. Shoulder Repair Made Simpler
Champion Instrumentation Shoulder Repair Made Simpler Champion Instruments Simplicity & Versatility The Champion Shoulder instrumentation system was designed with simplicity in mind. Clever instruments
More information