Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11

Size: px
Start display at page:

Download "Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11"

Transcription

1 Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) Hiroki Kadoya Yukiko Kitagawa MEDIX VOL. 61 P.8 P.11

2 Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) Hiroki Kadoya Yukiko Kitagawa Kitagawa Hospital, Kouyuukai Medical Corporation (Hirado City, Nagasaki Prefecture) Clinical experience using the Supria 1 multislice (16 slice)ct system, manufactured by Hitachi Medical Corporation, introduced at our institution in March, 2014 is reported below. Supria enables smooth and efficient operations and settings to meet various scanning and patient positioning needs with its wide-open 750mm gantry bore and full data retention at the maximum FOV (500mmφ). Also, this CT system is capable of low exposure scanning, applying iterative reconstruction technology as a standard function. Furthermore, it realizes both the fast speed and wide range acquisition of thin slice volume data for short-time and high-resolution imaging. Key Words: Supria, FOV, Open Bore Gantry, Intelli IP 1. Introduction Our hospital, which opened in Hirado, Nagasaki, in 1954, currently has 52 beds. It has specialized departments in internal medicine, surgery, pediatrics, gastrointestinal medicine, cardiovascular medicine, proctology, radiology, and rehabilitation. As time had come to renew our CT system (Multislice CT ROBUSTO, Hitachi Medical Corporation), we decided to install a new system. In Hirado, Nagaski, the proportion of 65 years old or older is relatively high at 33.2% 1). Our hospital also has a geriatric health services facility in addition to a pediatrics department. Therefore, in selecting the device, we mainly considered the convenience of scan positioning, a reduced rate of rescanning for patients in a hunched position, and dose-reduction technology. Space and power requirements comparable to those of a conventional CT system additionally emerged as selection criteria, and therefore the decision was made in March 2014 to install the Supria 1 CT system, manufactured by Hitachi Medical Corporation (Fig. 1). Here, we report on our experiences using this CT system and present actual clinical images. Fig. 1: Supria

3 2. Outline of the installed system Supria has a minimum scan time of 0.75s/rot and is equipped with a 5MHU X-ray tube. It has a large gantry bore of 750mm. Regardless of the field of view (hereinafter referred to as FOV)which is set when planning the scan conditions, the system constantly retains all data at the maximum FOV (500mm). Given the frequent scanning of the elderly at our hospital, we judged that the system would contribute to reducing the rescanning rate. Furthermore, with regard to the scanning of children, we judged that use of the system would help to reduce their fear of examination. The system has a footswitch, which can control the movement of patient table based on the preset positions. This ensures safety by freeing up both hands of the operator while also enabling smooth positioning of the patient. The system is also equipped with a Breath guide-a useful function for elderly patients that allows breath-holding instructions to be displayed with characters. This function is implemented by miniature monitors, placed at 3 sites within the gantry. As for functions related to dose reduction, the system is equipped with IntelliEC 2, which forms an elliptical model of the human body for each patient to control the tube current, and Intelli IP 3, an image processing technique that applies iterative reconstruction which repeats the calculations to reduce the noise both in projection space and image space. We intend to reduce radiation exposure by utilizing these functions not only when scanning children but during all examinations. 3. Presentation of clinical images (1)Clinical case 1 Brain image of a child: 6-year old boy Scanning was performed with the protocol shown in Table 1. For scanning unsedated children, the scan time must be shortened. Although we considered performing a non-helical scan by increasing the rotation speed to 0.75s/ rot since Supria installed here could generate outputs of up to 400mA, we employed the scan conditions in Table 1 in consideration of the tradeoff with the view rate. The brain image of a child generally shows less contrast than that of an adult. However, the contrast obtained in this case (Fig. 2)is likely to be sufficient for making clinical diagnoses. In addition, we consider that by jointly using Intelli IP, it improves the image characteristics such as graininess, and this leads to obtaining the images which are easy to diagnose. For Intelli IP, the operator can select from among seven levels (Level 1 to Level 7). We confirmed how noise varied at each level by using a water phantom. Figure 3 shows the results for the rate of noise reduction measured using the water phantom. In the case of scanning conditions that assumed a pediatric brain scan, we confirmed that the noise-reduction effect increased by 7.7% on average for each increment in Intelli IP level. Table 1: Scanning conditions for Clinical case 1 Fig. 2: Brain image of a child: 6-year old boy SD value Normal Slice thickness Collimation Tilt Kernel F19 FBP Level 1 Level 2 Level 3 Level 4 Level 5 Level 6 Level 7 Intelli IP level 100mAs 200mAs 240mAs 250mAs 300mAs 350mAs 400mAs Fig. 3: Noise reduction effect of Intelli IP using a water phantom Figure 4 shows images in which Intelli IP was applied to pediatric brain scans. The scanning conditions are identical to those indicated in Table 1. We studied whether images with better visibility could be obtained owing to the noise-reduction effect by increasing the level of Intelli IP, even with actual clinical images. As a result of performing SD measurements for each case at the epiphysis level, it was confirmed that SD reduction effect is similar to that of the phantom experiment. However, when the level is increased excessively, our impression is that the images become unsuitable for diagnostic use. In the case

4 of the current pediatric brain scan conditions, we infer that Level 2 or Level 3 is approximately the appropriate level. Table 2. Because Supria constantly retains all data at the maximum FOV (500mm), it can restore the lost portion only by recalculation and can re-center the image (Fig. 5b). When scanning is performed with a fixed tube current, this reduces radiation exposure by approximately 69mGy per rotation compared to when rescanning is performed. FBP Level 1 Table 2: Scanning conditions for Clinical case Normal Slice thickness Collimation Tilt Kernel F12 Level 2 Level 3 a: Image loss example Level 4 Level 5 b: Re-calculation example Fig. 5: Brain image of patient with a hunched condition: 86-year old male Level 6 Level 7 Fig. 4: Example of Intelli IP applied to the brain of a child: 9-year old boy (2)Clinical case 2 Brain image of patient with a hunched condition: 86-year old male In scanning patients who are hunched, image loss may occur when the patient deviates from the FOV set at the time of scanning. Figure 5a shows an image in which a defect occurred under the scanning conditions shown in (3)Clinical case 3 Adult lung image (arms lowered): 92-year old female At our hospital, scanning with the arms lifted is difficult in some cases (e.g., with elderly patients). Figure 6 shows an image of the lungs (arms lowered)scanned under the conditions shown in Table 3. The image suggests that artifacts such as those that arise from the upper forelimb bone are reduced to a degree that is unproblematic for diagnosis. This is presumably due to the effect of the CORE technique for the newly installed 3D image reconstruction algorithm, in addition to the effects of ADNR (Automatic Data Noise Reduction), an artifact suppression software program that had also been installed in the previous system.

5 Table 3: Scanning conditions for Clinical case mA 0.75 Volume Slice thickness Collimation Beam pitch Kernel F22/F32 However, increasing the level of Intelli IP appears to change the visual quality. Specifically, image blurring tends to intensify in accordance with the level of Intelli IP, as shown in Fig. 4. This suggests that when considering the compatibility between dose reduction and image quality, the proper level of Intelli IP and scanning conditions should be continuously explored and established in accordance with conditions such as age, sex, and scanned area of the patient. Although here we primarily investigated dose reduction in children, it will be important to continue investigations on optimizing scanning conditions (including Intelli IP)so that radiation exposure can also be reduced during other CT examinations. Fig. 6: Adult lung image arms lowered : 92-year old female (4)Clinical case 4 Adult lung image (patient with difficulty assuming a dorsal position): 86-year old male When scanning elderly patients, we sometimes encounter those that have difficulty assuming the dorsal position, similar to cases of scanning with the arms lowered. Figure 7 shows an image obtained by scanning in the dorsal position. For Supria installed here, which has a large gantry bore of 750mm, the positioning of patients that need to be scanned in an irregular position is relatively easy. In addition, our impression is that artifacts occur infrequently even during scanning in an irregular position. 5. Summary Supria, manufactured by Hitachi Medical Corporation and installed in our hospital, has a high degree of freedom with regard to scan positioning during examination due to its large gantry bore and constant retention of data within the maximum FOV. In addition, because examinations with a reduced dose are possible, we confirmed that the system fits well with the classes of patient and regional characteristics of our hospital and has high clinical utility. High-resolution images can also be obtained even as the breath holding time is shortened. This allows high-resolution images and low-dose examinations to be achieved simultaneously-a feat that was difficult to achieve with the previous system. We suggest that this system can contribute to improving the quality of examinations (e.g., image quality)and providing further benefits for patients. 1 Supria, 2 IntelliEC, and 3 Intelli IP are registered trademarks of Hitachi Medical Corporation. Fig. 7: Adult lung image patient with difficulty assuming the dorsal position : 86-year old male 4. Future challenges At our hospital, we consider it necessary to implement further dose-reduction measures, particularly when scanning children. As has already been reported, when viewed in terms of image SD values, a maximum dose-reduction effect of approximately 80% can be expected by using Intelli IP 2). References 1) 2010 Population Census, Statistical Tables According to Prefecture, City, District, Town, and Village. (in Japanese) 2) K. Uchiyama et al.: Background Introduction and Usage Experience of MDCT Supria from Hitachi Medical Corporation. Radfan, Vol.11, No.15: 32-35, (in Japanese)

Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing. Koichi Hirokawa MEDIX VOL. 56 P.43 P.46

Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing. Koichi Hirokawa MEDIX VOL. 56 P.43 P.46 Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing Taiga Goto Koichi Hirokawa Hisashi Takahashi MEDIX VOL. 56 P.43 P.46 Advanced Noise Reduction Processing for X-ray CT System

More information

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT

More information

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

1. Patient size AEC. Large Patient High ma. Small Patient Low ma Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description

More information

diagnostic examination

diagnostic examination RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011

More information

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

160-slice CT SCANNER / New Standard for the Future

160-slice CT SCANNER / New Standard for the Future TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Poster No.: C-2205 Congress: ECR 2012 Type: Scientific Paper

More information

Iterative Reconstruction in Image Space. Answers for life.

Iterative Reconstruction in Image Space. Answers for life. Iterative Reconstruction in Image Space Answers for life. Iterative Reconstruction in Image Space * (IRIS) * Please note: IRIS is used as an abbreviation for Iterative Reconstruction in Image Space throughout

More information

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity) Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular

More information

Improvement of CT image quality with iterative reconstruction idose4

Improvement of CT image quality with iterative reconstruction idose4 Improvement of CT image quality with iterative reconstruction idose4 Poster No.: C-0387 Congress: ECR 2014 Type: Scientific Exhibit Authors: M.-L. Olsson, K. Norrgren, M. Söderberg; Malmö/SE Keywords:

More information

Maximum Performance, Minimum Space

Maximum Performance, Minimum Space TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

RADspeed Pro. EDGEpackage C501-E041C

RADspeed Pro. EDGEpackage C501-E041C RADspeed Pro EDGEpackage C501-E041C 2 Some of the FPDs may be not available in your country. Please contact us to check the availability in your country. 3 Tomosynthesis in the Standing Position Tomosynthesis

More information

Software and Hardware in CCTA. Elly Castellano PhD

Software and Hardware in CCTA. Elly Castellano PhD Software and Hardware in CCTA Elly Castellano PhD Outline technical requirements for coronary CTA the modern cardiac CT scanner ECG-gating technology image reconstruction algorithms 2 Technical requirements

More information

Detector technology in simultaneous spectral imaging

Detector technology in simultaneous spectral imaging Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become

More information

SIGNA Explorer Lift revives our MR

SIGNA Explorer Lift revives our MR Seiji Shiotani, MD, PhD Seirei Fuji Hospital in Fuji City, Shizuoka, Japan Masayoshi Sugimura Seirei Fuji Hospital in Fuji City, Shizuoka, Japan SIGN Explorer Lift revives our MR The clinical usefulness

More information

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Wataru FUKUDA* Junya MORITA* and Masahiko YAMADA* Abstract Tomosynthesis is a three-dimensional imaging technology

More information

QC by the MPE in Belgium

QC by the MPE in Belgium Acceptance testing of state-of-the-art CT scanners using a new national protocol: first experience on a large number of scanners of different make and model the working group Radiology of the Belgian Hospital

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

RAD. Experiences Using the RADspeed Pro EDGE Package. 1. Hospital Description. 2. Background of Adoption. Hirohito Tanaka

RAD. Experiences Using the RADspeed Pro EDGE Package. 1. Hospital Description. 2. Background of Adoption. Hirohito Tanaka RAD Experiences Using the RADspeed Pro EDGE Package Hirohito Tanaka, R.T. Department of Radiology, Chibune General Hospital Hirohito Tanaka 1. Hospital Description Our hospital was originally established

More information

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them. In press 2004 1 2 Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Digital radiology An appropriate analogy that is easy for most people to understand

More information

Translating Protocols Between Scanner Manufacturer and Model

Translating Protocols Between Scanner Manufacturer and Model Translating Protocols Between Scanner Manufacturer and Model Robert J. Pizzutiello, MS, FAAPM, FACMP Sr. Vice-President, Global Physics Solutions President, Upstate Medical Physics Objectives Understand

More information

Iterative Reconstruction

Iterative Reconstruction RECENT ADVANCES IN CT RADIATION DOSE REDUCTION TECHNIQUES Iterative Reconstruction Kalpana Kanal, PhD, FSCBTMR, FACR, FAAPM Professor and Director, Diagnostic Physics Section University of Washington Seattle,

More information

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT) Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in

More information

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life.

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life. Neuro Thoracic Abdominal Abdominal Cardiovascular Pediatric SAFIRE Sinogram Affirmed Iterative Reconstruction Answers for life. SAFIRE * (Sinogram Affirmed Iterative Reconstruction) * The information

More information

Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2

Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2 Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2 Medical Consult, SA; Establish a method to correlate image

More information

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL TORNIER BLUEPRINT 3D Planning + PSI SCAN PROTOCOL Contents 3 Introduction 3 Patient preparation 3 Scanning instructions 4 Image instructions 5 Scanning parameters 6 Technical instructions 2 BLUEPRINT 3D

More information

Redefining Ergonomics

Redefining Ergonomics Samsung Electronics Co., Ltd. inspires the world and shapes the future with transformative ideas and technologies, redefining the worlds of TVs, smartphones, wearable devices, tablets, cameras, digital

More information

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results NeuViz 16 Computed Tomography Elevating routine imaging for exceptional results Essence NeuViz 16 Raising the bar on clinical utility in routine imaging. Get more. More clinical information for patients.

More information

T h e P h a n t o m L a b o r a t o r y

T h e P h a n t o m L a b o r a t o r y T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order

More information

Multi-Access Biplane Lab

Multi-Access Biplane Lab Multi-Access Biplane Lab Advanced technolo gies deliver optimized biplane imaging Designed in concert with leading physicians, the Infinix VF-i/BP provides advanced, versatile patient access to meet the

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

Electronic Noise in CT Detectors: Impact on Image Noise and Artifacts

Electronic Noise in CT Detectors: Impact on Image Noise and Artifacts Medical Physics and Informatics Original Research Duan et al. Electronic Noise in CT Detectors Medical Physics and Informatics Original Research Xinhui Duan 1 Jia Wang 1,2 Shuai Leng 1 ernhard Schmidt

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

Advanced digital image processing for clinical excellence in fluoroscopy

Advanced digital image processing for clinical excellence in fluoroscopy Dynamic UNIQUE Digital fluoroscopy solutions Dynamic UNIQUE Advanced digital image processing for clinical excellence in fluoroscopy André Gooßen, PhD, Image Processing Specialist Dörte Hilcken, Clinical

More information

Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise

Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise Jonas Rydberg, M.D. Professor of Radiology Indiana University School of Medicine Indianapolis, Indiana Medical

More information

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice nww.medical-devices.nhs.

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice   nww.medical-devices.nhs. abc September 2004 MHRA 04099 Philips Mx8000 IDT CT scanner technical evaluation Best choice best practice www.mhra.gov.uk nww.medical-devices.nhs.uk About MHRA evaluation reports. What you can expect.

More information

SOMATOM Esprit A Bundle of Energy

SOMATOM Esprit A Bundle of Energy SOMATOM Esprit A Bundle of Energy DATA SOMATOM Esprit An economical CT scanner designed for...... Excellent spiral image quality... A wide range of clinical applications... Value performance and reliabilty

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

Inside Biograph mct.

Inside Biograph mct. Inside Biograph mct The technologies behind the world s first molecular CT. www.siemens.com/mi Large 78 cm bore helps reduce claustrophobia and provides more room for RTP positioning devices. 227 kg (500

More information

CT Basics: Data Acquisition Module 3

CT Basics: Data Acquisition Module 3 Module 3 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

Brilliance in everything Philips CT products and services

Brilliance in everything Philips CT products and services Brilliance in everything Philips CT products and services Ready for anything No one does more than Philips to help you gain the productivity you need with a comprehensive approach to CT that marries significant

More information

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology The Fundamentals of... Computed Tomography Computed Tomography (CT) systems use x-rays to produce images of slices through a patient s anatomy. Despite having lower spatial resolution than other x-ray

More information

2010 Philips BrightView XCT SPECT/CT

2010 Philips BrightView XCT SPECT/CT 2010 Philips BrightView XCT SPECT/CT Unit was purchased from Philips training center in 2015. Installed but never been used by the current facility. (Scroll for pictures) BrightView XCT Camera with PinPoint

More information

The shortest distance to diagnosis Philips BrightView SPECT

The shortest distance to diagnosis Philips BrightView SPECT The shortest distance to diagnosis Philips Closer is better Simplicity is seeing something better right from the start. And is a completely new vision of what patient care can be, in a system as compact

More information

2 nd generation TOMOSYNTHESIS

2 nd generation TOMOSYNTHESIS 2 nd generation TOMOSYNTHESIS 2 nd generation DBT true innovation in breast imaging synthesis graphy Combo mode Stereotactic Biopsy Works in progress: Advanced Technology, simplicity and ergonomics Raffaello

More information

R/F. Comparison of Long View Radiography Systems. 1. Introduction. 2. Methods of Long View Radiography

R/F. Comparison of Long View Radiography Systems. 1. Introduction. 2. Methods of Long View Radiography R/F Comparison of Long View Radiography Systems Department of Radiology, Tokyo Women's Medical University Medical Center East 1 Department of Central Radiology, Tokyo Women's Medical University Hospital

More information

CT Basics: Image Quality Module 6

CT Basics: Image Quality Module 6 Module 6 For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions. Any broadcast,

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC? Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM

More information

SONIALVISION G4 Multi-purpose Digital R/F System C506-E075

SONIALVISION G4 Multi-purpose Digital R/F System C506-E075 SONIALVISION G4 Multi-purpose Digital R/F System C506-E075 Selecting the best for every examination environment BEST in CLASS Multi-purpose Digital R/F System 2 With the Sonialvision G4, Shimadzu offers

More information

SONIALVISION G4 Multi-purpose Digital R/F System C506-E075

SONIALVISION G4 Multi-purpose Digital R/F System C506-E075 SONIALVISION G4 Multi-purpose Digital R/F System C506-E075 Selecting the best for every examination environment BEST in CLASS Multi-purpose Digital R/F System With the Sonialvision G4, Shimadzu offers

More information

Aquilion Precision Ultra-High Resolution CT: Quantifying diagnostic image quality

Aquilion Precision Ultra-High Resolution CT: Quantifying diagnostic image quality Aquilion Precision Ultra-High CT: Quantifying diagnostic image quality Kirsten Boedeker, PhD, DABR Senior Manager, Quantitative Image Quality Canon Medical Systems Corporation Introduction Over the last

More information

Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT)

Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT) Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT) F Edward Boas, Roland Bammer, and Dominik Fleischmann Extended abstract for RSNA 2012 Purpose CT metal streak artifacts are

More information

Digital radiography. bucky table and wall stand as a dual detector or wireless system. Amadeo DR Systems

Digital radiography. bucky table and wall stand as a dual detector or wireless system. Amadeo DR Systems Amadeo DR Systems with dicom PACS DX-R Software X-ray Systems for the Future Digital radiography with Amadeo R-DR including bucky table and wall stand as a dual detector or wireless system High contrast

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

2012 :15th SESSION of ESMP

2012 :15th SESSION of ESMP 2012 :15th SESSION of ESMP Lecture presented in Archamps (Salève Building) by : Elly CASTELLANO (London) Patient dosimetry in x-ray imaging and CT Elly Castellano Objectives measurable dose quantities

More information

ADVANCED MEDICAL SYSTEMS PTE LTD Singapore Malaysia India Australia

ADVANCED MEDICAL SYSTEMS PTE LTD Singapore Malaysia India Australia Innovative design is combined with cutting-edge technology to yield a definitive diagnosis and never before seen ergonomics GIOTTO CLASS is the result of 25 years of experience in the research and development

More information

MUSICA Nerve Center. Artificial Intelligence. Intelligent tools for your Digital Radiography workflow. Fluoroscopy. Workflow Optimization

MUSICA Nerve Center. Artificial Intelligence. Intelligent tools for your Digital Radiography workflow. Fluoroscopy. Workflow Optimization Image Quality Bariatric Abdomen Pediatric Imaging Diagnostic Confidence Fluoroscopy Neonatal Imaging Scatter Suppression Dental Full Leg Full Spine Exposure Control Index Artificial Intelligence General

More information

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist Topics Image Uniformity and Artifacts Image Quality - Detail

More information

Invisible sophistication. Visible simplicity. CS Welcome to the simplicity of compact panoramic imaging

Invisible sophistication. Visible simplicity. CS Welcome to the simplicity of compact panoramic imaging Invisible sophistication. Visible simplicity. CS 8100 Welcome to the simplicity of compact panoramic imaging Introducing the CS 8100 The Carestream Dental Factor Humanized technology We keep our technology

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s

P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s AN ALL AROUND PERFECT PICTURE. The perfect combination of image quality, efficiency and design. P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s Air Techniques experts in digital imaging ProVecta

More information

Virtual monochromatic imaging in dual-source and dual-energy CT for visualization of acute ischemic stroke

Virtual monochromatic imaging in dual-source and dual-energy CT for visualization of acute ischemic stroke Virtual monochromatic imaging in dual-source and dual-energy CT for visualization of acute ischemic stroke Hidetake Hara and Hiroshi Muraishi School of Allied Health Sciences, Kitasato University, Japan

More information

Ludlum Medical Physics

Ludlum Medical Physics Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to

More information

Introduction of a Single Chip TLD System for Patient Dosimetry

Introduction of a Single Chip TLD System for Patient Dosimetry Introduction of a Single Chip TLD System for Patient Dosimetry C. Hranitzky a, M. Halda a, G. Müller a, B. Obryk b, H. Stadtmann a* a Austrian Research Centers GmbH ARC, 2444 Seibersdorf, Austria. b Institute

More information

Philips EasyDiagnost Eleva

Philips EasyDiagnost Eleva Philips EasyDiagnost Eleva The Philips EasyDiagnost Eleva Recognized for its ease of use and superb image quality, the EasyDiagnost Eleva has for many years been entrusted with a variety of R/F applications

More information

Nuclear Associates , , CT Head and Body Dose Phantom

Nuclear Associates , , CT Head and Body Dose Phantom Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product

More information

INNOVATION BY DESIGN. Toshiba A History of Leadership REMOTE CONTROL R/F SYSTEM

INNOVATION BY DESIGN. Toshiba A History of Leadership REMOTE CONTROL R/F SYSTEM INNOVATION BY DESIGN For over 130 years, Toshiba has led the world in developing technology to improve the quality of life. This Made for Life TM commitment is reflected in our family of leading-edge imaging

More information

Wide-Detector CT for TAVR Planning:

Wide-Detector CT for TAVR Planning: Wide-Detector CT for TAVR Planning: Impact on Iodine Dose, Radiation Dose, and Image Quality SCBTMR 2015 Annual Course Thursday, October 8 William P. Shuman MD FSCBTMR Department of Radiology University

More information

Multislice HELICAL CT SCANNER

Multislice HELICAL CT SCANNER Multislice HELICAL CT SCANNER Product Data No. MPDCT0237EAE APPLICATION The Asteion TM is a multislice Helical CT scanner that supports whole-body scanning. The system generates 5.3 slices per second using

More information

Truly flexible to meet your clinical needs

Truly flexible to meet your clinical needs Truly flexible to meet your clinical needs 2 Adapting to meet your needs Flexible Fast and responsive Excellent image quality Designed with ergonomic efficiency Equipped with dose management tools 3 Three

More information

Q3D. Speak to a 3D Specialist. CBCT 3D / Panoramic Imaging GENERAL DIMENSIONS. Suni Imaging Product Lines GET.

Q3D. Speak to a 3D Specialist. CBCT 3D / Panoramic Imaging GENERAL DIMENSIONS. Suni Imaging Product Lines GET. GENERAL Q3D Q3D Ceph Exposure Time FOV Voxel Size Focal Spot Target Angle Tube Voltage Tube Current Line Voltage Warranty Panoramic CT 9 to 17 sec 9 to 17 sec 4 to 12 sec 7.7/14.5 sec 7.7/14.5 sec 4 x

More information

Dose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation

Dose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation Medical Physics and Informatics Original Research Christner et al. CT Dose Reduction Medical Physics and Informatics Original Research Downloaded from www.ajronline.org by 8.243.133.8 on 2/26/18 from IP

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

PET/CT Instrumentation Basics

PET/CT Instrumentation Basics / Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire

More information

Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis

Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis Poster No.: C-0585 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Minehiro,

More information

Luminos RF Classic. Where value meets performance.

Luminos RF Classic. Where value meets performance. Luminos RF Classic Where value meets performance www.siemens.com/healthcare What s good value in fluoroscopy? That s easy. Luminos RF Classic. 2 Whether for its handling convenience, outstanding image

More information

SONIALVISION G4 Multi-purpose Digital R/F System

SONIALVISION G4 Multi-purpose Digital R/F System C506-E075A SONIALVISION G4 Multi-purpose Digital R/F System Founded in 1875, Shimadzu Corporation, a leader in the development of advanced technologies, has a distinguished history of innovation built

More information

P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s

P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s AN ALL AROUND PERFECT PICTURE. The perfect combination of image quality, efficiency and design. P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s Air Techniques experts in digital imaging ProVecta

More information

The disclaimer on page 1 is an integral part of this document. Copyright November 30, 2017 by AAPM. All rights reserved.

The disclaimer on page 1 is an integral part of this document. Copyright November 30, 2017 by AAPM. All rights reserved. DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION

More information

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE Takeyuki Hashimoto 1), Morio Onoe 2), Hiroshi Nakamura 3), Tamon Inouye 4), Hiromichi Jumonji 5), Iwao Takahashi 6); 1)Yokohama Soei

More information

3016 HD 16 Slice CT Scanner

3016 HD 16 Slice CT Scanner Turnkey Laboratories Solutions 3016 HD 16 Slice CT Scanner EPC / PRODUCTS / APPLICATION / SOFTWARE / ACCESSORIES / CONSUMABLES / SERVICES Analytical Technologies Limited An ISO 9001 Certified Company www.analyticalgroup.net

More information

Four-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development

Four-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development ORIGINAL ARTICLE ORIGINAL ARTICLE Radiation Medicine: Vol. 21 No. 1, 17 22 p.p., 2003 Four-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development Masahiro Endo,* Takanori Tsunoo,*

More information

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

C501-E029C. RADspeed. GENERAL RADIOGRAPHIC SYSTEM Automatic

C501-E029C. RADspeed. GENERAL RADIOGRAPHIC SYSTEM Automatic C501-E029C RADspeed GENERAL RADIOGRAPHIC SYSTEM Automatic High-Performance General Radiographic System Improves Workflow and Achieves Dose Reduction GENERAL RADIOGRAPHIC SYSTEM Automatic This state-of-the-art

More information

WE MAKE RELATIONSHIP FOR LIFE.

WE MAKE RELATIONSHIP FOR LIFE. 2018 WE MAKE RELATIONSHIP FOR LIFE. www.sanrad.in Contents Corporate Profile 02 Company Profile Rich in heritage and synonymous with healthcare, the field of medicine is deeply ingrained in Sanrad s roots

More information

The future of nuclear imaging is clear

The future of nuclear imaging is clear Cardius X-ACT The future of nuclear imaging is clear Increased regulations, growing competition, and concerns about radiation exposure are just a sampling of the current challenges facing the nuclear medicine

More information

E5 Implementation Working Group Questions & Answers (R1) Current version dated June 2, 2006

E5 Implementation Working Group Questions & Answers (R1) Current version dated June 2, 2006 INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE E5 Implementation Working Group & (R1) Current version dated June 2, 2006 ICH Secretariat,

More information

make it easy, with Ray

make it easy, with Ray Lower dose - Quick scan times - Pulsed X-ray technology - Multiple scan modes 3 Dedicated detectors - Reliable performance - No damage - Long life span Easy upgrade - Ready to upgrade CBCT & Cephalometric

More information