Correlation of 2D Reconstructed High Resolution CT Data of the Temporal Bone and Adjacent Structures to 3D Images

Size: px
Start display at page:

Download "Correlation of 2D Reconstructed High Resolution CT Data of the Temporal Bone and Adjacent Structures to 3D Images"

Transcription

1 Correlation of 2D Reconstructed High Resolution CT Data of the Temporal Bone and Adjacent Structures to 3D Images Rodt T 1, Ratiu P 1, Becker H 2, Schmidt AM 2, Bartling S 2, O'Donnell L 3, Weber BP 2, Jolesz FA 1, Kikinis R 1 ; Surgical Planning Laboratory, Brigham and Women's Hospital, Harvard Medical School 1, Boston, MA; Medical School Hannover 2, Hannover, Germany; Massachusetts Institute of Technology 3, Boston, MA Presented as a Poster at the 39 th Annual Meeting of the American Society of Neuroradiology (ASNR) 4/21/2001-4/27/2001, Boston MA Purpose The temporal bone anatomy is difficult to comprehend by looking at cross-sectional CT-images alone. 3D-visualization can help to understand the spatial relationship of the bony parts of the skull base, nerve structures and the middle and inner ear (1,2,3). 2D-images, however, often reflect the anatomy more accurately as some detail is lost by post-processing (3,4). Correlation of 2D-images to 3D-models might help to combine the advantages of these two modalities. Materials & Methods Multi-Slice CT scans covering both temporal bones were obtained in 17 patients with 1.25 mm collimation (140 kv, 40 ma, pitch 3). Two normal and 15 temporal bones with different pathologies including malformation, trauma, implants, tumor and inflammatory disease were investigated. A bone reconstruction algorithm was applied (0.3 mm reconstruction interval, 9.6 cm FOV) (5). Post-processing was performed using the in-house software "3D-Slicer". Segmentation was performed using thresholding and manual segmentation. The temporal bone was segmented as well as the ossicles, the tensor tympani muscle, the inner ear structures, the cranial nerves VII and VIII, and additional relevant structures in the case of pathology (6). 3Dmodels of the individual structures were generated in different colors using a surface-rendering algorithm. The 2D-slices were correlated to the 3D-models of the entire region by means of orienting the models along with the cross-sectional images into standardized views. The segmented areas for each individual structure were displayed in different colors in the axial, coronal and sagital images.

2 Results 3D-visualization of all the different anatomical structures displayed the correct anatomy and pathology in all the cases. The software "3D-Slicer" allowed various ways of interaction with the 3D-model, enabling a structured presentation of the anatomy and pathology. Post-processing took up to six hours due to time-consuming manual segmentation, merging of segmentationvolumes and generating the models. With more advanced post-processing algorithms this likely to be improved. In all the cases diagnosis would have been possible with the 2D-images alone. But correlation of the cross-sectional image to the 3D-model helped to understand the location and orientation of the 2D-slice in the 3D-anatomy. The complex spatial relationship of the individual structures of the temporal bone was easier to understand in the 2D-image when 3D- information was added. Conclusion Correlation of cross-sectional images to 3D-models is a more effective way of reaching a spatial understanding of the individual anatomy. Diagnosis can thus be made easier and the method can further be used for surgical planning and medical education. References 1)Howard JD, Elster AD, May JS. Temporal Bone: Three-dimensional CT. Part I. Normal Anatomy, Techniques, and Limitations. Radiology 1990; 177: )Reisser C, Schubert O, Forsting M, Sartor K. Anatomy of the temporal bone: detailed threedimensional display based on image data from high-resolution helical CT: a preliminary report. Am J Otol. 1996; 17: )Schubert O, Sartor K, Forsting M, Reisser C. Three-dimensional computed display of otosurgical operation sites by spiral CT. Neuroradiology 1996; 38: )Howard JD, Elster AD, May JS. Temporal Bone: Three-dimensional CT. Part II. Pathologic Alterations. Radiology 1990; 177: )Casselman JW. Temporal bone imaging. Neuroimaging Clin N Am. 1996; 6: )Janecka IP, Tiedemann K. Skull Base Surgery; Anatomy, Biology and Technology, Lippincott-Raven Publishers 1997

3

4 Fig.1: Generating a 3D-model and correlation to the 2D-image: 1A: reconstructed axial CT image. 1B: segmentation of the anatomical structures. 1C: 3D-model is generated based on segmentation and correlated to axial 2D-image. 1D: superior view of 3D-model of middle and inner ear structures correlated to the axial 2D-slice.

5 Fig.2: Tumor of the temporal bone 2A: reconstructed axial CT shows bone erosion, the overall amount of erosion is difficult to evaluate. 2B: lateral view of correlated 2D-image and 3D-model. The ossicular chain is still intact. 2C: lateral view of 3D-model shows amount of bone erosion but spatial relation to 2Dimage is difficult to comprehend. Fig.3: Total Ossicular Replacement: 3A: reconstructed axial CT shows tip of prosthesis posterior to oval window. 3B: anterior view of correlated 2D-image and 3D-model displays spatial relationship of inner ear structures and 2D-image. 3C: 3D-model shows discontinuity but spatial relation to 2D-image is difficult to comprehend. Fig.4: Fracture of the temporal bone and dislocation 4A: axial CT shows dislocation of the incudomallear joint and fracture fragment. 4B: inferior view of correlated 2D-image and 3D-model shows dislocation of the ossicular chain caused by impingement of the fracture fragment. The fracture course is shown. 4C: inferior view of 3D-

6 model shows pathology but spatial relation to 2D-image is difficult to comprehend.

Scopis Hybrid Navigation with Augmented Reality

Scopis Hybrid Navigation with Augmented Reality Scopis Hybrid Navigation with Augmented Reality Intelligent navigation systems for head surgery www.scopis.com Scopis Hybrid Navigation One System. Optical and electromagnetic measurement technology. As

More information

Image Interpretation System for Informed Consent to Patients by Use of a Skeletal Tracking

Image Interpretation System for Informed Consent to Patients by Use of a Skeletal Tracking Image Interpretation System for Informed Consent to Patients by Use of a Skeletal Tracking Naoki Kamiya 1, Hiroki Osaki 2, Jun Kondo 2, Huayue Chen 3, and Hiroshi Fujita 4 1 Department of Information and

More information

BodyViz fact sheet. BodyViz 2321 North Loop Drive, Suite 110 Ames, IA x555 www. bodyviz.com

BodyViz fact sheet. BodyViz 2321 North Loop Drive, Suite 110 Ames, IA x555 www. bodyviz.com BodyViz fact sheet BodyViz, the company, was established in 2007 at the Iowa State University Research Park in Ames, Iowa. It was created by ISU s Virtual Reality Applications Center Director James Oliver,

More information

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL TORNIER BLUEPRINT 3D Planning + PSI SCAN PROTOCOL Contents 3 Introduction 3 Patient preparation 3 Scanning instructions 4 Image instructions 5 Scanning parameters 6 Technical instructions 2 BLUEPRINT 3D

More information

Reflex TM Surgical Technique. Anterior Cervical Plate

Reflex TM Surgical Technique. Anterior Cervical Plate Reflex TM Surgical Technique Anterior Cervical Plate Surgical Technique Acknowledgement: Stryker Spine extends their thanks to the following surgeons for their participation in the development of the Reflex

More information

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT

More information

Gesture-controlled interactive three dimensional anatomy: a novel teaching tool in head and neck surgery

Gesture-controlled interactive three dimensional anatomy: a novel teaching tool in head and neck surgery Hochman et al. Journal of Otolaryngology - Head and Neck Surgery 2014, 43:38 ORIGINAL RESEARCH ARTICLE Open Access Gesture-controlled interactive three dimensional anatomy: a novel teaching tool in head

More information

Implants for surgery Orthopaedic joint prosthesis. Part 1: Procedure for producing parametric 3D bone models from CT data of the knee

Implants for surgery Orthopaedic joint prosthesis. Part 1: Procedure for producing parametric 3D bone models from CT data of the knee Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 19233-1 First edition 2017-05 Implants for surgery Orthopaedic joint prosthesis Part 1: Procedure for producing parametric 3D bone models from

More information

Technique Guide. Occipito-Cervical Fusion System. Implants and instruments designed to optimize fixation to the occiput.

Technique Guide. Occipito-Cervical Fusion System. Implants and instruments designed to optimize fixation to the occiput. Technique Guide Occipito-Cervical Fusion System. Implants and instruments designed to optimize fixation to the occiput. Table of Contents Introduction Overview 2 AO ASIF Principles 4 Indications and Contraindications

More information

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National

More information

METAL artifact reduction (MAR) is a major problem

METAL artifact reduction (MAR) is a major problem 1 Metal artifact reduction based on the combined prior image Yanbo Zhang, Xuanqin Mou arxiv:1408.5198v [physics.med-ph] 4 Sep 014 Abstract Metallic implants introduce severe artifacts in CT images, which

More information

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation

More information

Software and Hardware in CCTA. Elly Castellano PhD

Software and Hardware in CCTA. Elly Castellano PhD Software and Hardware in CCTA Elly Castellano PhD Outline technical requirements for coronary CTA the modern cardiac CT scanner ECG-gating technology image reconstruction algorithms 2 Technical requirements

More information

Multi-slice computed tomography analysis of bullet trajectory in forensic investigation.

Multi-slice computed tomography analysis of bullet trajectory in forensic investigation. Multi-slice computed tomography analysis of bullet trajectory in forensic investigation. Poster No.: C-0510 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Usui, Y. Kawasumi, Y. Hosokai, A. Nakajima,

More information

www.anatomage.com info@anatomage.com Why The Anatomage Table? Advanced Educational Tool Both the accuracy of the real human anatomy and the quantity of pathological examples are unique aspects of the Anatomage

More information

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) Hiroki Kadoya Yukiko Kitagawa MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria

More information

Cerclage Passer. For minimally invasive application of cerclage cables.

Cerclage Passer. For minimally invasive application of cerclage cables. Cerclage Passer. For minimally invasive application of cerclage cables. Handling Technique Cable application This publication is not intended for distribution in the USA. Instruments and implants approved

More information

H OLOGRAPUY is a method used mainly in technical

H OLOGRAPUY is a method used mainly in technical J Neurosurg 58:729-733, 1983 A study of human head vibrations using time-averaged holography HEINz-E. HOYER, D.S.c., AND Ji3nCEN D/)RHEIDE, ENC. Division of Functional and Applied Anatomy, Medical School

More information

Q3D. Speak to a 3D Specialist. CBCT 3D / Panoramic Imaging GENERAL DIMENSIONS. Suni Imaging Product Lines GET.

Q3D. Speak to a 3D Specialist. CBCT 3D / Panoramic Imaging GENERAL DIMENSIONS. Suni Imaging Product Lines GET. GENERAL Q3D Q3D Ceph Exposure Time FOV Voxel Size Focal Spot Target Angle Tube Voltage Tube Current Line Voltage Warranty Panoramic CT 9 to 17 sec 9 to 17 sec 4 to 12 sec 7.7/14.5 sec 7.7/14.5 sec 4 x

More information

8/3/2016. The EPID Strikes Back. Novel Applications for Current EPID Technology. Joerg Rottmann, PhD. Disclosures and acknowledgements

8/3/2016. The EPID Strikes Back. Novel Applications for Current EPID Technology. Joerg Rottmann, PhD. Disclosures and acknowledgements The EPID Strikes Back Joerg Rottmann Brigham and Women s Hospital / Dana-Farber Cancer Institute Harvard Medical School Disclosures and acknowledgements Disclosures Varian MRA grant Acknowledgements Boston

More information

Green Innovation for the Next Generation

Green Innovation for the Next Generation Green Innovation for the Next Generation Green Innovation for the Next Generation PaX-i3D Green The New Digital Environment Green CBCT - VATECH s innovative ultra low dose X-ray technology - Green CBCT

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

For dental students, acquisition of clinical competence. Stereoscopy in Dental Education: An Investigation. The Anaglyph Technique

For dental students, acquisition of clinical competence. Stereoscopy in Dental Education: An Investigation. The Anaglyph Technique Stereoscopy in Dental Education: An Investigation Shumei Murakami, PhD, DDS; Rinus G. Verdonschot, PhD; Sven Kreiborg, PhD, DDS; Naoya Kakimoto, PhD, DDS; Asuka Kawaguchi, DDS Abstract: The aim of this

More information

EYE ANATOMY. Multimedia Health Education. Disclaimer

EYE ANATOMY. Multimedia Health Education. Disclaimer Disclaimer This movie is an educational resource only and should not be used to manage your health. The information in this presentation has been intended to help consumers understand the structure and

More information

160-slice CT SCANNER / New Standard for the Future

160-slice CT SCANNER / New Standard for the Future TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Wataru FUKUDA* Junya MORITA* and Masahiko YAMADA* Abstract Tomosynthesis is a three-dimensional imaging technology

More information

High-fidelity haptic and visual rendering for patient-specific simulation of temporal bone surgery

High-fidelity haptic and visual rendering for patient-specific simulation of temporal bone surgery Computer Assisted Surgery ISSN: (Print) 2469-9322 (Online) Journal homepage: https://www.tandfonline.com/loi/icsu21 High-fidelity haptic and visual rendering for patient-specific simulation of temporal

More information

Maximum Performance, Minimum Space

Maximum Performance, Minimum Space TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

Vectra, Vectra-T and Vectra-One. Anterior cervical plating for spinal fusion.

Vectra, Vectra-T and Vectra-One. Anterior cervical plating for spinal fusion. Vectra, Vectra-T and Vectra-One. Anterior cervical plating for spinal fusion. Technique Guide Instruments and implants approved by the AO Foundation Table of Contents Introduction Vectra, Vectra-T and

More information

ACLP Anterior Cervical Locking Plate System TECHNIQUE GUIDE

ACLP Anterior Cervical Locking Plate System TECHNIQUE GUIDE ACLP Anterior Cervical Locking Plate System TECHNIQUE GUIDE Instruments and implants approved by the AO Foundation ACLP Anterior Cervical Locking Plate System The ACLP System is designed to reduce the

More information

Vitreous imaging system New method for medical diagnosis

Vitreous imaging system New method for medical diagnosis Vitreous imaging system New method for medical diagnosis Dr. Boucherit Taieb, Privet Laboratory, 12 impasse de Venise, Oran, Algeria Privet Laboratory, Dr. Boucherit Taieb, Oran, Algeria Abstract - The

More information

Why the Anatomage Table?

Why the Anatomage Table? Why the Anatomage Table? Compare Anatomage Table Cadaver Lab Chemicals No Formaldehyde, methanol, phenol, & other solvents Facility No special requirements Ventilation, freezer, storage, & disposal Restrictions

More information

ACCS Anterior Cervical Compression System TECHNIQUE GUIDE

ACCS Anterior Cervical Compression System TECHNIQUE GUIDE ACCS Anterior Cervical Compression System TECHNIQUE GUIDE Original Instruments and Implants of the Association for the Study of Internal Fixation AO ASIF ACCS Anterior Cervical Compression System The Anterior

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

Methods. Experimental Stimuli: We selected 24 animals, 24 tools, and 24

Methods. Experimental Stimuli: We selected 24 animals, 24 tools, and 24 Methods Experimental Stimuli: We selected 24 animals, 24 tools, and 24 nonmanipulable object concepts following the criteria described in a previous study. For each item, a black and white grayscale photo

More information

Picasso-Trio 3-D Dental Imaging

Picasso-Trio 3-D Dental Imaging Picasso-Trio 3-D Dental Imaging Digital Panoramic & Cephalometric, CBCT X-ray Imaging System 3 in 1 System Product Data Model Picasso-Trio - NP Panoramic only [CT upgradeable] Picasso-Trio - NC Panoramic

More information

Occipito-Cervical Fusion System. Implants and instruments designed to optimize fixation to the occiput.

Occipito-Cervical Fusion System. Implants and instruments designed to optimize fixation to the occiput. Occipito-Cervical Fusion System. Implants and instruments designed to optimize fixation to the occiput. Technique Guide This publication is not intended for distribution in the USA. Instruments and implants

More information

Universal Humeral Nail

Universal Humeral Nail 990210009 INDEX Indications Preoperative Planning Patient Position Surgical Technique - Step 1 Open Humerus - Step 2 Calibrate The Nail - Step 3 Insert Nail - Step 4 Proximal Locking - Step 5 Assemble

More information

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Huaiqun Guan,

More information

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

1. Patient size AEC. Large Patient High ma. Small Patient Low ma Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description

More information

2D, 3D CT Intervention, and CT Fluoroscopy

2D, 3D CT Intervention, and CT Fluoroscopy 2D, 3D CT Intervention, and CT Fluoroscopy SOMATOM Definition, Definition AS, Definition Flash Answers for life. Siemens CT Vision Siemens CT Vision The justification for the existence of the entire medical

More information

Data. microcat +SPECT

Data. microcat +SPECT Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for

More information

Robot assisted craniofacial surgery: first clinical evaluation

Robot assisted craniofacial surgery: first clinical evaluation Robot assisted craniofacial surgery: first clinical evaluation C. Burghart*, R. Krempien, T. Redlich+, A. Pernozzoli+, H. Grabowski*, J. Muenchenberg*, J. Albers#, S. Haßfeld+, C. Vahl#, U. Rembold*, H.

More information

5HDO 7LPH 6XUJLFDO 6LPXODWLRQ ZLWK +DSWLF 6HQVDWLRQ DV &ROODERUDWHG :RUNV EHWZHHQ -DSDQ DQG *HUPDQ\

5HDO 7LPH 6XUJLFDO 6LPXODWLRQ ZLWK +DSWLF 6HQVDWLRQ DV &ROODERUDWHG :RUNV EHWZHHQ -DSDQ DQG *HUPDQ\ nsuzuki@jikei.ac.jp 1016 N. Suzuki et al. 1). The system should provide a design for the user and determine surgical procedures based on 3D model reconstructed from the patient's data. 2). The system must

More information

Orthopedic Bone Nail System Universal Humeral Nail

Orthopedic Bone Nail System Universal Humeral Nail Orthopedic Bone Nail System Universal Humeral Nail Surgical Technique Manual Note: The surgical procedures should be performed under the guidance of qualified skilled orthopedic surgeons, and this surgical

More information

Technique Guide. 2.4/2.7 mm Locking Tarsal Plates. Talus Plate, Navicular Plate and Cuboid Plate.

Technique Guide. 2.4/2.7 mm Locking Tarsal Plates. Talus Plate, Navicular Plate and Cuboid Plate. Technique Guide 2.4/2.7 mm Locking Tarsal Plates. Talus Plate, Navicular Plate and Cuboid Plate. Table of Contents Introduction 2.4/2.7 mm Locking Tarsal Plates 2 AO Principles 4 Indications 5 Clinical

More information

Geometric modelling of the temporal bone for cochlea implant simulation

Geometric modelling of the temporal bone for cochlea implant simulation University of Wollongong Research Online Faculty of Engineering - Papers (Archive) Faculty of Engineering and Information Sciences 2004 Geometric modelling of the temporal bone for cochlea implant simulation

More information

Diagnostic imaging of the lateral skull base requires high

Diagnostic imaging of the lateral skull base requires high ORIGINAL RESEARCH O. Majdani K. Thews S. Bartling M. Leinung C. Dalchow R. Labadie T. Lenarz G. Heidrich Temporal Bone Imaging: Comparison of Flat Panel Volume CT and Multisection CT BACKGROUND AND PURPOSE:

More information

VBOSS. Surgical Technique. Vertebral Body Support System

VBOSS. Surgical Technique. Vertebral Body Support System VBOSS Surgical Technique Vertebral Body Support System 1. System Description 1.1 Implants...3 1.2 Instruments...4 2. Indications...8 3. Patient Position...8 4. Surgical Approach 4.1 Choice of adequate

More information

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

More information

Robots in the Field of Medicine

Robots in the Field of Medicine Robots in the Field of Medicine Austin Gillis and Peter Demirdjian Malden Catholic High School 1 Pioneers Robots in the Field of Medicine The use of robots in medicine is where it is today because of four

More information

Gibralt. Occipito-Cervico Thoracc System. Operative Technique

Gibralt. Occipito-Cervico Thoracc System. Operative Technique Gibralt Occipito-Cervico Thoracc System Operative Technique TABLE OF CONTENTS GIBRALT...1 OPERATIVE TECHNIQUE OVERVIEW...2 HOOK PLACEMENT...2 SCREW PLACEMENT...2 DETAILED OPERATIVE TECHNIQUE...4 PLACEMENT

More information

MR Basics: Module 8 Image Quality

MR Basics: Module 8 Image Quality Module 8 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

URS Degen. Top loading pedicle screw system for posterior stabilization.

URS Degen. Top loading pedicle screw system for posterior stabilization. URS Degen. Top loading pedicle screw system for posterior stabilization. Technique Guide This publication is not intended for distribution in the USA. Table of Contents Introduction URS Degen 2 AO Principles

More information

Stereoscopic Augmented Reality System for Computer Assisted Surgery

Stereoscopic Augmented Reality System for Computer Assisted Surgery Marc Liévin and Erwin Keeve Research center c a e s a r, Center of Advanced European Studies and Research, Surgical Simulation and Navigation Group, Friedensplatz 16, 53111 Bonn, Germany. A first architecture

More information

1571. Artifacts in computer tomography imaging: how it can really affect diagnostic image quality and confuse clinical diagnosis?

1571. Artifacts in computer tomography imaging: how it can really affect diagnostic image quality and confuse clinical diagnosis? 1571. Artifacts in computer tomography imaging: how it can really affect diagnostic image quality and confuse clinical diagnosis? Vincentas Veikutis 1, Tomas Budrys 2, Algidas Basevicius 3, Saulius Lukosevicius

More information

SOMATOM Esprit A Bundle of Energy

SOMATOM Esprit A Bundle of Energy SOMATOM Esprit A Bundle of Energy DATA SOMATOM Esprit An economical CT scanner designed for...... Excellent spiral image quality... A wide range of clinical applications... Value performance and reliabilty

More information

(12) Patent Application Publication (10) Pub. No.: US 2007/ A1

(12) Patent Application Publication (10) Pub. No.: US 2007/ A1 (19) United States US 200701 18100A1 (12) Patent Application Publication (10) Pub. No.: US 2007/0118100 A1 Mahesh et al. (43) Pub. Date: (54) SYSTEM AND METHOD FOR IMPROVED ABLATION OF TUMIORS (75) Inventors:

More information

PACS Fundamentals. By: Eng. Valentino T. Mvanga Ministry of Health and Social Welfare Tanzania

PACS Fundamentals. By: Eng. Valentino T. Mvanga Ministry of Health and Social Welfare Tanzania PACS Fundamentals By: Eng. Valentino T. Mvanga Ministry of Health and Social Welfare Tanzania 1 Learning Goals To Understand the importance of PACS To Understand PACS infrastructure requirement Introduction

More information

CTAND MRI DIAGNOSTIC IMAGING GUIDELINES

CTAND MRI DIAGNOSTIC IMAGING GUIDELINES CHAPTER 46 CTAND MRI DIAGNOSTIC IMAGING GUIDELINES FOR FOOT AND ANKLE PATFIOLOGY Craig A. Camasta, D.P.M. Diagnostic imaging of the lower extremity has become a popular and useful adjunct to the practice

More information

Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2

Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2 Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2 1. Department of Medical Physics, University of Wisconsin-Madison 2. Department

More information

Optima ZS Spinal Fixation System

Optima ZS Spinal Fixation System Surgical Technique Optima ZS Spinal Fixation System The low-profile, in-line, polyaxial pedicle screw system. Optima ZS Surgical Technique 1 Optima ZS Spinal Fixation System The Optima ZS Spinal Fixation

More information

THE USE OF OPEN REDUCtion

THE USE OF OPEN REDUCtion ORIGINAL ARTICLE Comparison of 3 Optical Navigation Systems for Computer-Aided Maxillofacial Surgery E. Bradley Strong, MD; Amir Rafii, MD; Bettina Holhweg-Majert, MD, DMD; Scott C. Fuller, MD; Marc Christian

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

Technique Guide. Quadrilateral Surface Plates 3.5. Part of the Low Profile Pelvic System 3.5.

Technique Guide. Quadrilateral Surface Plates 3.5. Part of the Low Profile Pelvic System 3.5. Technique Guide Quadrilateral Surface Plates 3.5. Part of the Low Profile Pelvic System 3.5. Table of Contents Introduction Quadrilateral Surface Plates 3.5 2 AO Principles 4 Indications 5 Surgical Technique

More information

SCIENTIFIC COMPUTING AND IMAGING INSTITUTE

SCIENTIFIC COMPUTING AND IMAGING INSTITUTE SCIENTIFIC COMPUTING AND IMAGING INSTITUTE University of Utah Scientific Computing and Imaging Institute The SCI research group was founded in 1994 by Drs. Chris Johnson and Rob MacLeod along with five

More information

Occipito-Cervical Fusion System

Occipito-Cervical Fusion System Implants and Instruments designed to enhance Fixation to the Occiput Occipito-Cervical Fusion System Surgical Technique Image intensifier control This description alone does not provide sufficient background

More information

MED-LIFE: A DIAGNOSTIC AID FOR MEDICAL IMAGERY

MED-LIFE: A DIAGNOSTIC AID FOR MEDICAL IMAGERY MED-LIFE: A DIAGNOSTIC AID FOR MEDICAL IMAGERY Joshua R New, Erion Hasanbelliu and Mario Aguilar Knowledge Systems Laboratory, MCIS Department Jacksonville State University, Jacksonville, AL ABSTRACT We

More information

T h e P h a n t o m L a b o r a t o r y

T h e P h a n t o m L a b o r a t o r y T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order

More information

Invisible sophistication. Visible simplicity. CS Welcome to the simplicity of compact panoramic imaging

Invisible sophistication. Visible simplicity. CS Welcome to the simplicity of compact panoramic imaging Invisible sophistication. Visible simplicity. CS 8100 Welcome to the simplicity of compact panoramic imaging Introducing the CS 8100 The Carestream Dental Factor Humanized technology We keep our technology

More information

Gamma3 Long Nail R1.5 and R2.0. Operative technique

Gamma3 Long Nail R1.5 and R2.0. Operative technique Gamma3 Long Nail R1.5 and R2.0 Gamma3 Long Nail R1.5 and R2 Gamma3 Long Nail R1.5 and R2.0 Contents 1. Design of the Gamma3 System...4 Lag screw and set screw function... 5 Distal locking screws... 5 2.

More information

A Training Simulator for the Angioplasty Intervention with a Web Portal for the Virtual Environment Searching

A Training Simulator for the Angioplasty Intervention with a Web Portal for the Virtual Environment Searching A Training Simulator for the Angioplasty Intervention with a Web Portal for the Virtual Environment Searching GIOVANNI ALOISIO, LUCIO T. DE PAOLIS, LUCIANA PROVENZANO Department of Innovation Engineering

More information

An immersive virtual reality environment for diagnostic imaging

An immersive virtual reality environment for diagnostic imaging An immersive virtual reality environment for diagnostic imaging Franklin King 1, 2, Jagadeesan Jayender 2, Steve Pieper 2, 3, Tina Kapur 2, Andras Lasso 1, and Gabor Fichtinger 1 1 Laboratory for Percutaneous

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Brilliance in everything Philips CT products and services

Brilliance in everything Philips CT products and services Brilliance in everything Philips CT products and services Ready for anything No one does more than Philips to help you gain the productivity you need with a comprehensive approach to CT that marries significant

More information

Wide-Detector CT for TAVR Planning:

Wide-Detector CT for TAVR Planning: Wide-Detector CT for TAVR Planning: Impact on Iodine Dose, Radiation Dose, and Image Quality SCBTMR 2015 Annual Course Thursday, October 8 William P. Shuman MD FSCBTMR Department of Radiology University

More information

D. Greg Anderson, MD Thomas Jefferson University Hospital Philadelphia, PA

D. Greg Anderson, MD Thomas Jefferson University Hospital Philadelphia, PA Surgical Technique D E S I G N I N G S U R G E O N S D. Greg Anderson, MD Thomas Jefferson University Hospital Philadelphia, PA Robert Heary, MD University of Medicine and Dentistry of New Jersey Newark,

More information

Incorporating novel image processing methods in a hospital-wide PACS

Incorporating novel image processing methods in a hospital-wide PACS International Congress Series 1281 (2005) 1016 1021 www.ics-elsevier.com Incorporating novel image processing methods in a hospital-wide PACS Erwin Bellon a, T, Michel Feron a, Paul Neyens a, Klaas Peeters

More information

RENDERING MEDICAL INTERVENTIONS VIRTUAL AND ROBOT

RENDERING MEDICAL INTERVENTIONS VIRTUAL AND ROBOT RENDERING MEDICAL INTERVENTIONS VIRTUAL AND ROBOT Lavinia Ioana Săbăilă Doina Mortoiu Theoharis Babanatsas Aurel Vlaicu Arad University, e-mail: lavyy_99@yahoo.com Aurel Vlaicu Arad University, e mail:

More information

A Study On Preprocessing A Mammogram Image Using Adaptive Median Filter

A Study On Preprocessing A Mammogram Image Using Adaptive Median Filter A Study On Preprocessing A Mammogram Image Using Adaptive Median Filter Dr.K.Meenakshi Sundaram 1, D.Sasikala 2, P.Aarthi Rani 3 Associate Professor, Department of Computer Science, Erode Arts and Science

More information

PET/CT Instrumentation Basics

PET/CT Instrumentation Basics / Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire

More information

CHARLOTTE. Multi-Use Compression Screw

CHARLOTTE. Multi-Use Compression Screw CHARLOTTE Multi-Use Compression Screw CHARLOTTE multi-use compression screw surgical technique SURGICAL ADVISORS ROBERT ANDERSON, MD BRUCE COHEN, MD W. HODGES DAVIS, MD Proper surgical procedures and techniques

More information

CT Basics: Equipment and Instrumentation Module 2

CT Basics: Equipment and Instrumentation Module 2 Module 2 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational in-house or online educational course use only in educational and corporate institutions.

More information

NIH Public Access Author Manuscript Otol Neurotol. Author manuscript; available in PMC 2012 January 1.

NIH Public Access Author Manuscript Otol Neurotol. Author manuscript; available in PMC 2012 January 1. NIH Public Access Author Manuscript Published in final edited form as: Otol Neurotol. 2011 January ; 32(1): 11 16. doi:10.1097/mao.0b013e3181fcee9e. Robotic Mastoidectomy Andrei Danilchenko, B.S., Graduate

More information

Medical Imaging and its Associated Analysis

Medical Imaging and its Associated Analysis Medical Imaging and its Associated Analysis Saurabh Singh 1, Anurag Singh 2,Pranay Surana3, Priyen Dang 4, Anand Ranka 5, Saurabh Burange 6 1 Department of Electronics and Communication Engineering 2,3,4,5,6

More information

Fretting and Corrosion Damage at the Head-Neck Taper is Reduced with Ceramic Femoral Heads: A Retrieval Study

Fretting and Corrosion Damage at the Head-Neck Taper is Reduced with Ceramic Femoral Heads: A Retrieval Study Fretting and Corrosion Damage at the Head-Neck Taper is Reduced with Ceramic Femoral Heads: A Retrieval Study Kasisin Klunklin, M.D. 1,2, Christina M. Schmidt 3,2, Marcel E. Roy 2, Leo A. Whiteside, M.D.

More information

Translating Protocols Between Scanner Manufacturer and Model

Translating Protocols Between Scanner Manufacturer and Model Translating Protocols Between Scanner Manufacturer and Model Robert J. Pizzutiello, MS, FAAPM, FACMP Sr. Vice-President, Global Physics Solutions President, Upstate Medical Physics Objectives Understand

More information

CMF Surgery. Angulus2. Angled Screwdriver. ref

CMF Surgery. Angulus2. Angled Screwdriver. ref CMF Surgery Angulus2 Angled Screwdriver ref. 90-126-02-04 It s the head that counts and the face. There is nothing with which we identify ourselves more than with the face. We are how we see ourselves.

More information

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice nww.medical-devices.nhs.

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice   nww.medical-devices.nhs. abc September 2004 MHRA 04099 Philips Mx8000 IDT CT scanner technical evaluation Best choice best practice www.mhra.gov.uk nww.medical-devices.nhs.uk About MHRA evaluation reports. What you can expect.

More information

Fourier Domain (Spectral) OCT OCT: HISTORY. Could OCT be a Game Maker OCT in Optometric Practice: A THE TECHNOLOGY BEHIND OCT

Fourier Domain (Spectral) OCT OCT: HISTORY. Could OCT be a Game Maker OCT in Optometric Practice: A THE TECHNOLOGY BEHIND OCT Could OCT be a Game Maker OCT in Optometric Practice: A Hands On Guide Murray Fingeret, OD Nick Rumney, MSCOptom Fourier Domain (Spectral) OCT New imaging method greatly improves resolution and speed of

More information

3016 HD 16 Slice CT Scanner

3016 HD 16 Slice CT Scanner Turnkey Laboratories Solutions 3016 HD 16 Slice CT Scanner EPC / PRODUCTS / APPLICATION / SOFTWARE / ACCESSORIES / CONSUMABLES / SERVICES Analytical Technologies Limited An ISO 9001 Certified Company www.analyticalgroup.net

More information

P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s

P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s AN ALL AROUND PERFECT PICTURE. The perfect combination of image quality, efficiency and design. P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s Air Techniques experts in digital imaging ProVecta

More information

High-Resolution Stereoscopic Surgical Display Using Parallel Integral Videography and Multi-projector

High-Resolution Stereoscopic Surgical Display Using Parallel Integral Videography and Multi-projector High-Resolution Stereoscopic Surgical Display Using Parallel Integral Videography and Multi-projector Hongen Liao 1, Nobuhiko Hata 2, Makoto Iwahara 2, Susumu Nakajima 3, Ichiro Sakuma 4, and Takeyoshi

More information

BLACKBIRD Spinal System

BLACKBIRD Spinal System BLACKBIRD Spinal System Cervical-Thoracic Spinal Fixation System The ChoiceSpine BLACKBIRD Cervical-Thoracic Spinal Fixation System is a comprehensive system for posterior fixation of the cervical and upper

More information

Technique Guide. Modular Sternal Cable System. Flexibility and strength in sternal closure and repair.

Technique Guide. Modular Sternal Cable System. Flexibility and strength in sternal closure and repair. Technique Guide Modular Sternal Cable System. Flexibility and strength in sternal closure and repair. Table of Contents Introduction Overview 2 Indications and Contraindications 3 Surgical Technique A.

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

SYLLABUS CHAPTER - 2 : INTENSITY TRANSFORMATIONS. Some Basic Intensity Transformation Functions, Histogram Processing.

SYLLABUS CHAPTER - 2 : INTENSITY TRANSFORMATIONS. Some Basic Intensity Transformation Functions, Histogram Processing. Contents i SYLLABUS UNIT - I CHAPTER - 1 : INTRODUCTION TO DIGITAL IMAGE PROCESSING Introduction, Origins of Digital Image Processing, Applications of Digital Image Processing, Fundamental Steps, Components,

More information

An Augmented Reality Application for the Enhancement of Surgical Decisions

An Augmented Reality Application for the Enhancement of Surgical Decisions An Augmented Reality Application for the Enhancement of Surgical Decisions Lucio T. De Paolis, Giovanni Aloisio Department of Innovation Engineering Salento University Lecce, Italy lucio.depaolis@unisalento.it

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information