1 CCRI(I)/03-19 Update on NIST X-Ray Air-Kerma Standards and Calibrations Michelle O Brien and Stephen M. Seltzer Ionizing Radiation Division National Institute of Standards and Technology Gaithersburg, MD USA March 31, 2003 Beginning in November of 2001 the 100 kv x-ray tube was replaced, and all the NIST and ISO reference radiation qualities have been reestablished. The beam uniformity remains the same, and all x-ray beam half-value layers were reproduced. In May of 2002 the 320 kv x-ray tube was replaced. The new 320 kv tube is a different design than the previous tube, featuring less inherent filtration and a slightly larger focal spot. All NIST reference radiation qualities have been redeveloped, without changing the additional filtration appreciably or the operating tube potential. Most of the ISO reference radiation qualities have been reestablished, with the others to be reestablished in the near future. The four BIPM reference beam qualities produced at 100, 135, 180 and 250 kv have now been developed at NIST. These are being used for an indirect comparison with the BIPM, which should conclude in April of Since November of 2001, two voltage dividers have been acquired for use on all three x-ray ranges. Both voltage dividers are in the process of being calibrated by NIST Electronics and Electrical Engineering Laboratory and will be used to verify the operating voltage of the x-ray generators. The recent acquisition of the second divider will allow the measurement of voltage on both the cathode and anode of the 320 kv system. These calibrated dividers will allow a more accurate determination of voltage and possibly permit the calibration of non-invasive kvp meters. The design process for a replacement of the Wyckoff-Attix free-air ionization chamber, which is the 300 kv primary standard, has begun. Monte Carlo calculations of correction factors for free-air chambers, by David Burns of the BIPM, were reported at the last two meetings of the CCRI(I) [1,2]. He obtained these data as a function of incident photon energy from 10 kev to 250 kev. Those pertaining to the three NIST standards, the Lamperti (up to 50 kev), Ritz (up to 100 kev) and Wyckoff-Attix (up to 250 kev) free-air chambers, were given to NIST and later augmented by additional calculations between 250 kev and 300 kev for the Wyckoff- Attix chamber. These data include results for the photon-scatter correction k sc, the electron-loss correction k e, and the new fluorescence correction k fl introduced at the CCRI(I) 2001 [2]. Using a library of photon spectra derived by others [3] from measurements, the corrections for our standard free-air chambers were calculated for the beam qualities maintained by NIST. In this work, we included also the calculation for the attenuation correction k att. As described in references [4, 5], previously at NIST k e and k sc were derived from the data in references [6-9], and k att was derived from 1
2 measurements. The final determinations of theses factors depended to varying degrees on skillful curve fitting and interpolation. NIST is now in the process of adopting the Monte Carlo database provided by the BIPM and available x-ray beam spectra for the determination of k sc, k e, and k fl, along with a blend of measured and calculated values for k att. The current changes in NIST standards are indicated in Tables 1 and 2. Although most of these data are not expected to change significantly, note that the work is ongoing and some of these results could be revised. References [1] D.T. Burns, Consistent set of calculated values for electron-loss and photonscatter corrections for parallel-plate free-air chambers, CCRI(I)/99-4 (1999). [2] D.T. Burns, The re-absorption of fluorescence photons in free-air chambers, CCRI(I)/01-32 (2001). [3] See, e.g., W.W.Seelentag, W. Panzer, G. Drexler, L. Platz and F. Santner, A Catologue of Spectra for the Calibration of Dosimeters, GSF Report 560 (1979). Electronic files provided by C.G. Soares (private communication). [4] P.J. Lamperti, T.P. Loftus and R. Loevinger, Calibration of X-Ray and Gamma- Ray Measuring Instruments, NBS Special Publication (1988). [5] P.J. Lamperti and M. O Brien, Calibration of X-Ray and Gamma-Ray Measuring Instruments, NIST Special Publication (2001). [6] V.H. Ritz, Standard free-air chamber for the measurement of Low-energy x rays ( kilovolts-constant-potential). J. Res. NBS 64C, (1960). [7] H.O. Wyckoff and F.H. Attix, Design of free-air ionization chambers, NBS Handbook 64 (1957). [8] A. Allisy and A.M. Roux, Contribution a la mesure des rayons roentgen dans le domaines de 5 a 50 kv, Acta Radiologica 55, (1961). [9] V.H. Ritz, Design of free-air ionization chambers for the soft x-ray region ( kv), Radiology 73, (1959). 2
3 Table 1. NIST Calibration Conditions for X-Ray Measuring Instruments, Additional Filtration a Half-value layer b Homogeneity Total Effective Beam (HVL) coefficient (HC) percent energy code Al Cu Sn Pb Al Cu change in Al Cu (kev) standard c L d L d L L L L L L M M M M M M e M M e M M M M H d H d H H H H H H H H H H S S a The additional filtration value does not include the inherent filtration. The inherent filtration is approximately 1.0 mm Be for beam codes L10-L100, M20-M50, H10-H40 and S75; and 3.0 mm Be for beam codes M60-M300, H50-H300 and S60. b The HVL values were measured directly using the two new x-ray tubes installed in November of 2001 and May of c Subject to modification as work progresses. d Total change includes only that due to k e k sc k fl, i.e., does not include k att, which is determined at the time of calibration for these low-energy beams. e New beam quality. 3
4 Table 2. ISO X-Ray Beam Quality Parameters Offered at NIST Additional Filtration a First HVL b Second HVL b Beam Code Al Cu Sn Pb Al Cu Al Cu Total percent change in standard c HK HK HK HK HK HK HK HK HK WS WS WS WS WS WS WS NS NS NS NS NS NS NS NS NS NS NS NS NS NS LK LK LK LK LK LK70 d LK100 d LK LK LK LK see next page for table notes 4
5 a The additional filtration does not include the inherent filtration. The inherent filtration is a combination of the filtration due to the monitor chamber plus 1 mm Be for beam codes LK10-LK30, NS10-NS30, HK10- HK30; for all other techniques the inherent filtration is adjusted to 4 mm Al. b The HVL values were measured directly using the two new x-ray tubes installed in November of 2001 and May of c Subject to modification as work progresses. d Might be redeveloped using the Wyckoff-Attic rather than the Ritz FAC. 5
MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM *
MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM * C. Kessler and D.T. Burns December 2018 BUREAU INTERNATIONAL DES POIDS ET MESURES
More informationSPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by
SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This
More informationEffect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies
ARTICLE IN PRESS Radiation Physics and Chemistry 68 (2003) 707 720 Effect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies M. Assiamah, D. Mavunda,
More informationCalibration of KAP meters
Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area
More informationTechnical Protocol. SIM Comparison of Calibration Coefficients at Radiotherapy Level for Orthovoltage X-ray Beams. Pilot laboratory: NIST
SIM Comparison of Calibration Coefficients at Radiotherapy Level for Orthovoltage X-ray Beams Technical Protocol Pilot laboratory: NIST Contact person: Michelle O Brien National Institute of Standards
More informationI. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA
FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE
More informationOptimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation
Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.125 pissn 2508-4445, eissn 2508-4453 Optimization of Energy Modulation Filter for Dual Energy CBCT
More informationDerivation of total filtration thickness for diagnostic x-ray source assembly
Physics in Medicine & Biology PAPER OPEN ACCESS Derivation of total filtration thickness for diagnostic x-ray source assembly To cite this article: Michiharu Sekimoto and Yoh Katoh 2016 Phys. Med. Biol.
More informationEvaluation of X-Ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem78 Models
Original Article Evaluation of X-Ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem78 Models Suk Chiang Chen, Wei Loong Jong, Ahmad Zaky Harun Submitted: 20 Aug 2011 Accepted:
More informationSpekCalc: a program to calculate photon spectra from tungsten anode x-ray tubes
NOTE SpekCalc: a program to calculate photon spectra from tungsten anode x-ray tubes 5 10 G Poludniowski 1, G Landry 2, F DeBlois 2,3, P M Evans 1 and F Verhaegen 2,4 1 Joint Department of Physics, Institute
More informationI. Introduction.
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case
More informationX-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage
X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage
More informationGE AMX 4+ Portable X-Ray
GE AMX 4+ Portable X-Ray Typical Manufacturer s Picture GE Healthcare s AMX-4+ analog X-ray system provides high-performance in a compact, easy-to-maneuver package. The rotating arm and tube simplify positioning
More informationNew Detectors for X-Ray Metal Thickness Measuring
ECNDT 2006 - Poster 132 New Detectors for X-Ray Metal Thickness Measuring Boris V. ARTEMIEV, Alexander I. MASLOV, Association SPEKTR- GROUP, Moscow, Russia Abstract. X-ray thickness measuring instruments
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced
More informationTitle: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS
Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS By Aravind Ravichandran arr192@mail.usask.ca University of Saskatchewan Address: 2424 Cumberland
More informationMammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over
Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on
More informationANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 325 ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER ABSTRACT William Chang, Jonathan Kerner, and Edward
More informationX-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah
X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,
More information1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013).
CODE: RADT 151 INSTITUTE: Health Science TITLE: Radiographic Exposure DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of radiographic exposure selection and manipulation
More informationReceived 3 December 2016; revised 17 May 2017; editorial decision 21 June 2017; accepted 4 July 2017
Radiation Protection Dosimetry (2018), Vol. 178, No. 2, pp. 235 241 Advance Access publication 15 July 2017 doi:10.1093/rpd/ncx091 NOTE COMPARISON OF PERSONAL DOSE EQUIVALENT Hp(10) IN 137 CS RADIATION
More informationDISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter
DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools
More informationThis document is a preview generated by EVS
INTERNATIONAL STANDARD ISO 17636-1 First edition 2013-01-15 Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film Contrôle non destructif des assemblages soudés
More informationTest Equipment for Radiology and CT Quality Control Contents
Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3
More informationPreliminary studies of a new monitor ionization chamber
1 Preliminary studies of a new monitor ionization chamber Maíra T. Yoshizumi, Vitor Vivolo and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN) Comissão Nacional de Energia
More information1-1. GENERAL 1-2. DISCOVERY OF X-RAYS
1-1. GENERAL Radiography is a highly technical field, indispensable to the modern dental practice, but presenting many potential hazards. The dental radiographic specialist must be thoroughly familiar
More informationEffective energy measurement using radiochromic film: application of a mobile scanner
Polish Journal of Medical Physics and Engineering 2016;22(4):85-92 December 2016 doi: 10.1515/pjmpe-2016-0015 ISSN 1898-0309 Scientific Paper Effective energy measurement using radiochromic film: application
More informationISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film
INTERNATIONAL STANDARD ISO 17636-1 First edition 2013-01-15 Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film Contrôle non destructif des assemblages soudés
More informationSUBCHAPTER 14. THERAPEUTIC INSTALLATIONS
SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS 7:28-14.1 Scope (a) This subchapter covers therapeutic installations used in the healing arts. These therapeutic installations include x-ray, accelerator and teletherapy
More informationDosepix Detector as kvp-meter in Radiology and Mammography: First steps
Dosepix Detector as kvp-meter in Radiology and Mammography: First steps F.Bisello, I.Ritter, F.Tennert, A.Zang MediPix Collaboration Meeting, 19th February 2014, CERN Protect, Enhance, and Save Lives -
More informationDETERMINATION OF INHERENT AND ADDITIONAL FILTRATION IN ORDER TO ESTABLISH RADIATION QUALITIES ACCORDING TO IEC 61267
2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 DETERMINATION
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationDose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp
Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division
More informationAPPLICATION NOTE. RTI Electronics AB, Sweden Revision A, July kvp and time measurements with the Barracuda or Piranha on a typical dental unit
APPLICATION NOTE RTI Electronics AB, Sweden Revision A, July 2010 kvp and time measurements with the Barracuda or Piranha on a typical dental unit Measurements of kvp and time on dental units with the
More information2012 :15th SESSION of ESMP
2012 :15th SESSION of ESMP Lecture presented in Archamps (Salève Building) by : Elly CASTELLANO (London) Patient dosimetry in x-ray imaging and CT Elly Castellano Objectives measurable dose quantities
More informationStudy of increased radiation when an x-ray tube is placed in a strong magnetic field
Study of increased radiation when an x-ray tube is placed in a strong magnetic field Zhifei Wen Departments of Radiology and Physics, Stanford University, Stanford, California 94305 Norbert J. Pelc Departments
More informationAssessment of Beam Alignment, Collimation and Half Value Layer of Some Selected X-Ray Machines in Plateau State, Nigeria
International Journal of Innovative Scientific & Engineering Technologies Research 5(4):-5, Oct.-Dec., 07 SEAHI PUBLICATIONS, 07 www.seahipaj.org ISSN: 60-896X Assessment of Beam Alignment, Collimation
More informationExposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?
Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago
More informationKey words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure
Accuracy and calibration of integrated radiation output indicators in diagnostic radiology: A report of the AAPM Imaging Physics Committee Task Group 190 Pei-Jan P. Lin a) Virginia Commonwealth University
More informationMH<RIA TECHNICAL REPORT. Laboratory x-ray irradiator for cellular radiobiology research studies: Dosimetry report DEFENSE NUCLEAR AGENCY
MH
More informationMC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE
MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for
More informationPATIENT EFFECTIVE DOSES IN DIAGNOSTIC RADIOLOGY, NA
Title of Paper: Patient effective doses in diagnostic radiology Authors: N.A. Gkanatsios, and W. Huda * Corresponding Author: Department of Radiology, University of Florida, P.O. Box 100374, Gainesville,
More informationRULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER USE OF X-RAY APPARATUS
RULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER 0400-20-06 USE OF X-RAY APPARATUS TABLE OF CONTENTS 0400-20-06-.01 Purpose 0400-20-06-.06 Veterinary
More informationElectronic Brachytherapy Sources. Thomas W. Rusch
Electronic Brachytherapy Sources Thomas W. Rusch Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration
More informationCh. 223 VETERINARY MEDICINE CHAPTER 223. VETERINARY MEDICINE GENERAL PROVISIONS X-RAYS RADIOACTIVE MATERIAL. Authority
Ch. 223 VETERINARY MEDICINE 25 223.1 CHAPTER 223. VETERINARY MEDICINE Sec. 223.1. Purpose and scope. 223.2. [Reserved]. 223.2a. Definitions. 223.3 223.6. [Reserved]. 223.7. Structural shielding. 223.8.
More informationPredicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities
Journal of Physics: Conference Series PAPER OPEN ACCESS Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Recent citations - Resolution Properties of a
More informationY11-DR Digital Radiography (DR) Image Quality
Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures
More informationSTEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS
STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%
More informationMWPC Gas Gain with Argon-CO 2 80:20 Gas Mixture
IMA Journal of Mathematical Control and Information Page 1 of 10 doi:10.1093/imamci/dri000 1. Principles of Operation MWPC Gas Gain with Argon-CO 2 80:20 Gas Mixture Michael Roberts A multi-wire proportional
More informationK-edge subtraction X-ray imaging with a pixellated spectroscopic detector
K-edge subtraction X-ray imaging with a pixellated spectroscopic detector Silvia Pani Department of Physics, University of Surrey Summary Hyperspectral imaging K-edge subtraction X-ray imaging for mammography
More informationResponse characteristics of a tandem ionization chamber in standard X-ray beams
Applied Radiation and Isotopes 58 (2003) 495 500 Response characteristics of a tandem ionization chamber in standard X-ray beams Alessandro M. Costa*, Linda V.E. Caldas Instituto de Pesquisas Energ!eticas
More informationImaging Technique Optimization of Tungsten Anode FFDM System
Imaging Technique Optimization of Tungsten Anode FFDM System Biao Chen a*, Andrew P. Smith b, Zhenxue Jing a, Elena Ingal a a Hologic, Inc. 600 Technology Drive, DE 1970 b Hologic, Inc. 35 Crosby Drive,
More informationSmall-Animal X-ray Dose from Micro-CT
RESEARCH ARTICLE, pp. 149 158 149 Small-Animal X-ray Dose from Micro-CT John M. Boone, Orlando Velazquez, and Simon R. Cherry University of California Davis Abstract The use of micro-ct in small animals
More informationORAMED S MEASUREMENT AND SIMULATION CAMPAIGN FOR EXREMITY AND EYE LENS DOSES OF MEDICAL STAFF INVOLVED IN INTERVENTIONAL RADIOLOGY AND CARDIOLOGY
ORAMED S MEASUREMENT AND SIMULATION CAMPAIGN FOR EXREMITY AND EYE LENS DOSES OF MEDICAL STAFF INVOLVED IN INTERVENTIONAL RADIOLOGY AND CARDIOLOGY Carinou E., Brodecki M.,Domienik J., Donadille L., Ferrari
More informationPhotons interaction with matter
ب س م هللا الر ح من الر حیم Photons interaction with matter Ionization Ionization is the process of removing an electron from an electrically neutral atom to produce an ion pair. An ion is an atom or subatomic
More informationQC in Diagnostic Radiology. Main steps for a QC survey in Diagnostic Radiology
EVALUATING X-RAY TUBE AND GENERATOR PERFORMANCE : DEMO for PRACTICAL QUALITY CONTROL (QC) Dr Slavik Tabakov Dept. Medical Eng. & Physics, King's College London slavik.tabakov@kcl.ac.uk QC in Diagnostic
More informationVeterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS
Veterinary Science Preparatory Training for the Veterinary Assistant Floron C. Faries, Jr., DVM, MS Radiology Floron C. Faries, Jr., DVM, MS Objectives Determine the appropriate machine settings for making
More informationExplain what is meant by a photon and state one of its main properties [2]
1 (a) A patient has an X-ray scan taken in hospital. The high-energy X-ray photons interact with the atoms inside the body of the patient. Explain what is meant by a photon and state one of its main properties....
More informationPRACTICAL CONSIDERATIONS AND EFFECTS OF METALLIC SCREEN FLUORESCENCE AND BACKSCATTER CONTROL IN GAMMA COMPUTED RADIOGRAPHY
19 th World Conference on Non-Destructive Testing 2016 PRACTICAL CONSIDERATIONS AND EFFECTS OF METALLIC SCREEN FLUORESCENCE AND BACKSCATTER CONTROL IN GAMMA COMPUTED RADIOGRAPHY Steven MANGO 1 1 Carestream
More informationDigital Radiography : Flat Panel
Digital Radiography : Flat Panel Flat panels performances & operation How does it work? - what is a sensor? - ideal sensor Flat panels limits and solutions - offset calibration - gain calibration - non
More informationRADIOGRAPHIC EXPOSURE
RADIOGRAPHIC EXPOSURE Receptor Exposure Receptor Exposure the that interacts with the receptor. Computed Radiography ( ) requires a. Direct Digital Radiography (DR) requires a. Exposure Indicators Exposure
More informationISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 2: X- and gamma-ray techniques with digital detectors
INTERNATIONAL STANDARD ISO 17636-2 First edition 2013-01-15 Non-destructive testing of welds Radiographic testing Part 2: X- and gamma-ray techniques with digital detectors Contrôle non destructif des
More informationX-RAYS - NO UNAUTHORISED ENTRY
Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are
More informationAmorphous Selenium Direct Radiography for Industrial Imaging
DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial
More informationDetermination of the tube voltage from clinic mammographic system using two types of detectors
BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 03-1A (2015) 01-06 Determination of the tube voltage from clinic mammographic system using two types of detectors J. S. Barreira a ; V. Vivolo a a Gerência
More informationTechnical data CAMARGUE CS-VH50/300. VARIABLE Height Bucky Table With Ceiling Suspension
Technical data VARIABLE Height Bucky Table With Ceiling Suspension Model Variations CAMARGUE FH (Fixed Height) CAMARGUE FH Tomo CAMARGUE FH Ceiling suspension CAMARGUE VH (Variable Height) CAMARGUE VH
More informationNuclear Associates
Nuclear Associates 07-706 Patient Phantom/Penetrometer System Users Manual March 2005 Manual No. 07-706-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are
More informationA COMPARATIVE STUDY ON THE PERFORMANCE OF DIGITAL DETECTOR SYSTEMS FOR HIGH ENERGY APPLICATIONS
11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic More Info at Open Access Database www.ndt.net/?id=16394 A COMPARATIVE STUDY ON THE PERFORMANCE
More informationAdvancing EDS Analysis in the SEM Quantitative XRF. International Microscopy Congress, September 5 th, Outline
Advancing EDS Analysis in the SEM with in-situ Quantitative XRF Brian J. Cross (1) & Kenny C. Witherspoon (2) 1) CrossRoads Scientific, El Granada, CA 94018, USA 2) ixrf Systems, Inc., Houston, TX 77059,
More informationX-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto
X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring
More informationDisclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography
Current Implementation Status of IEC Standard 62494-1: Exposure Index (EI) for Digital Radiography July 31, 2017 Ryan Fisher, PhD, DABR Katie Hulme, MS, DABR None Disclosures Outline Review of IEC Standard
More informationELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone
ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY 12477 (800)VIP-XRAY (845)247-9028 Fax or 800-847-9729 local phone 845-246-9013 http://www.eccxray.com sales@eccxray.com INSTRUCTION MANUAL
More informationBREMSSTRAHLUNG DIODE PERFORMANCE ON MERCURY MIVA
BREMSSTRAHLUNG DIODE PERFORMANCE ON MERCURY MIVA J.W. Schumer ξ, R.J. Allen, R.J. Commisso, G. Cooperstein, D.D. Hinshelwood, D.P. Murphy, S.J. Stephanakis t, S.B. Swanekamp t, F.C. Young t Plasma Physics
More informationEffect of patient support pads on image quality and dose in fluoroscopy a)
Effect of patient support pads on image quality and dose in fluoroscopy a) William R. Geiser, Walter Huda, b) and Nikolaos A. Gkanatsios Department of Radiology, University of Florida, Gainesville, Florida
More information35080M kvp Divider. Operators Manual
35080M kvp Divider Operators Manual July 2006 Manual No. 1210001000 Rev. 3 2004, 2005, 2006 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective
More informationX-ray Tube and Generator Basic principles and construction
X-ray Tube and Generator Basic principles and construction Dr Slavik Tabakov - Production of X-rays OBJECTIVES - X-ray tube construction - Anode - types, efficiency - X-ray tube working characteristics
More informationX-ray tube with needle-like anode
NUKLEONIKA 2002;47(3):101 105 ORIGINAL PAPER X-ray tube with needle-like anode Mieczys aw S apa, W odzimierz StraÊ, Marek Traczyk, Jerzy Dora, Miros aw Snopek, Ryszard Gutowski, Wojciech Drabik Abstract
More informationPD233: Design of Biomedical Devices and Systems
PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/
More informationHISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD
CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING
More informationHow to Avoid Thermal Sensor Damage & Out of Tolerance Conditions
About Ophir-Spiricon With over 30 years of experience, the Ophir Photonics Group provides a complete line of instrumentation including power and energy meters, beam profilers, spectrum analyzers, and goniometric
More informationDevelopment of Personal Dosimeter Using Electronic Dose Conversion Method
Proceedings of the Korean Nuclear Spring Meeting Gyeong ju, Korea, May 2003 Development of Personal Dosimeter Using Electronic Dose Conversion Method Wanno Lee, Bong Jae Lee, and Chang Woo Lee Korea Atomic
More informationX-rays in medical diagnostics
X-rays in medical diagnostics S.Dolanski Babić 2017/18. History W.C.Röntgen (1845-1923) discovered a new type of radiation Nature, Jan. 23. 1896.; Science, Feb.14. 1896. X- rays: Induced the ionization
More informationLearning Objectives: What s my motivation? (unknown screen actor) Workshop Overview
Practical Medical Physics Adapting Traditional Clinical Medical Physics to Digital Radiography Charles E. Willis, Ph.D., DABR Associate Professor Department of Imaging Physics The University of Texas M.D.
More informationDevelopment of new dosimeter for measuring dose distribution in CT
Development of new dosimeter for measuring dose distribution in CT Poster No.: C-2925 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology - Without Subtopic Authors: Y. Muramatsu, K.
More informationSeminar 8. Radiology S8 1
Seminar 8 Radiology Medical imaging. X-ray image formation. Energizing and controlling the X-ray tube. Image detectors. The acquisition of analog and digital images. Digital image processing. Selected
More informationCylindrical Ion Chambers Victoreen Model 550 Series
Cylindrical Ion Chambers Victoreen Model 550 Series! Cylindrical Ion Chambers for use with Model 35040 and Model 530 electrometers! Wide range of applications in Diagnostic X-Ray and Radiation Oncology
More informationStudies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationStudies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationEvaluation of large area polycrystalline CdTe detector for diagnostic x-ray imaging
The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2011 Evaluation of large area polycrystalline CdTe detector for diagnostic x-ray imaging Xiance Jin The University
More informationScienceDirect. Evaluation of Pipe Wall Thickness Based on Contrast Measurement using Computed Radiography (CR)
Available online at www.sciencedirect.com ScienceDirect Procedia Engineering 69 ( 2014 ) 1216 1224 24th DAAAM International Symposium on Intelligent Manufacturing and Automation, 2013 Evaluation of Pipe
More informationRaySafe X2. Effortless measurements of X-ray
RaySafe X2 Effortless measurements of X-ray At your fingertips We ve grown accustomed to intuitive interactions with our devices. After all, it s not the device that s most important, but what you can
More informationHASL-151 DETERMINATION OF ENERGY DISTRIBUTION OF X-RAY BEAMS FROM A 250 KV CONSTANT POTENTIAL MACHINE
HASL-151 DETERMINATION OF ENERGY DISTRIBUTION OF X-RAY BEAMS FROM A 25 KV CONSTANT POTENTIAL MACHINE By Arthur Shambon November 1964 Health and Safety Laboratory, AEC New York Operations Office New York,
More informationTHE PANTAK THERAPAX SXT 150
THE PANTAK THERAPAX SXT 150 combines the safety and ease-of-use of photon radiation therapy with state-of-the-art technology, including microprocessor control and solid-state electronics. Using Therapax,
More informationInvestigation of the line-pair pattern method for evaluating mammographic focal spot performance
Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann
More informationBarcelona, January 2011 Workshop on optimization of Radiation Protection of Medical Staff
USE OF ACTIVE PERSONAL DOSEMETERS IN INTERVENTIONAL RADIOLOGY/CARDIOLOGY: TESTS WITH CONTINUOUS AND PULSED FIELDS IN LABORATORY CONDITIONS ORAMED PROJECT I. Clairand 1, J-M. Bordy 2, E. Carinou 3, J. Daures
More informationNEMA XR X-ray Equipment for Interventional Procedures User Quality Control Mode
NEMA XR 27-2012 X-ray Equipment for Interventional Procedures User Quality Control Mode Published by: National Electrical Manufacturers Association 1300 North 17th Street, Suite 1752 Rosslyn, Virginia
More informationBeam-Restricting Devices
Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the
More informationFire testing: Calibration of smoke opacity measuring systems
Subject of Agreement EGOLF AGREEMENT 002-2016 Calibration of smoke opacity measuring systems Related test standard EN 13823 Date of issue Reference original query SM3:1996 Previous publication number (if
More informationUnfors EDD-30 Radiation Protection in Fluoroscopy
Unfors EDD-30 Radiation Protection in Fluoroscopy Immediate Warning Decrease Your Dose Interventional radiology procedures are considered to be essential to medical diagnosis and treatment. It is recognized,
More informationDate: July 31, 2017 Title: Design of an Aperture Assembly for X-Ray Diffraction Student: Christina Schmidt Mentor: Dr. Peter Ko
Date: July 31, 2017 Title: Design of an Aperture Assembly for X-Ray Diffraction Student: Christina Schmidt Mentor: Dr. Peter Ko Abstract: Dr. Ko and I designed and built a new X-Ray diffraction aperture
More information