Small-Animal X-ray Dose from Micro-CT

Size: px
Start display at page:

Download "Small-Animal X-ray Dose from Micro-CT"

Transcription

1 RESEARCH ARTICLE, pp Small-Animal X-ray Dose from Micro-CT John M. Boone, Orlando Velazquez, and Simon R. Cherry University of California Davis Abstract The use of micro-ct in small animals has increased in recent years. Although the radiation levels used for micro-ct are generally not lethal to the animal, they are high enough where changes in the immune response and other biological pathways may alter the experimental outcomes. Therefore, it is important to understand what the doses are for a specific imaging procedure. Monte Carlo simulation was used to evaluate the radiation dose to small animals (5 40 mm in diameter) as a result of X-ray exposure. Both monoenergetic (6 100 kev) and polyenergetic ( kvp) X-ray sources were simulated under typical mouse imaging geometries. X-ray spectral measurements were performed on a mouse imaging X-ray system using a commercially available X-ray spectrometer, and spectra from high-energy systems were used as well. For a typical X-ray system with 1.0 mm of added Al at 40 kvp, the dose coefficients (dose to mouse per air kerma at isocenter) were 0.80, 0.63, 0.52, and 0.44 mgy/mgy for mouse diameters of 10, 20, 30, and 40 mm, respectively. A number of tables and figures are provided for dose estimation over a range of mouse imaging geometries. Mol Imaging (2004) 3, Keywords: Radiation dose, X-ray, Monte Carlo, mouse imaging, computed tomography (CT), X-ray spectrum. Introduction Anatomical evaluation of the mouse, rat, and other small animals using micro-computed tomography (micro-ct) has proved useful in a number of research applications [1 5]. However, to achieve high signal-to-noise ratio (SNR) images with extremely small voxels (e.g., mm), relatively large radiation doses are required. Previous studies [1,6] have suggested that mouse imaging using micro-ct delivers between 100 and 300 mgy, about a factor of 10 larger than typical human CT examinations. Radiation doses for diagnostic imaging of the small animal in general are not lethal, but the dose levels may be sufficient to induce deterministic effects in the animal that could confound the biological outcome of the experiment [7 10]. In this study, we employed validated Monte Carlo procedures coupled with spectroscopic analysis of a micro-ct X-ray source to determine radiation dose levels as a function of the diameter of the animal scanned. We report here both dose coefficients (air kerma to tissue absorbed dose) as well as tissue doses for typical mouse dimensions and micro-ct examination parameters. These data should allow investigators to precisely calculate the radiation dose to small animals when scan parameters are known or to estimate doses when the exact scan parameters are not available. Materials and Methods Measurement of X-ray Spectra X-ray spectra were measured using a cadmium zinc telluride (CdZnT) spectrometer (Model XR-l00T-2T, Amptek, Bedford, MA) and an X-ray system typical of mouse CT (Oxford X-ray tube, tungsten anode, 0.04 mm beryllium window, Scotts Valley, CA, Spellman X-ray generator model SL50P300, Hauppauge, NY). X-ray spectra were measured from 15 to 50 kvp in 5-kVp intervals, with no added filtration to the X-ray system except for thin Be window of the X-ray tube. Each X-ray spectrum was acquired using approximately 10 6 photons. An analog-to-digital converter (Model DT 3010, Data Translation, Marlboro, MA) was used to acquire the signal output from the spectrometer. Custom software, utilizing the Data Translation Measurement Foundry software toolkit, was written for this application. X-ray spectral measurements were made at differing applied tube voltages (kvp s), at the same tube current setting (0.5 ma). The X-ray spectrometer was placed 40.3 cm from the X-ray source, and spectral acquisition times ranged from approximately 200 to 1000 sec. The energy scale of the spectrometer was calibrated using both the known kvp of the X-ray generator and the emission lines of several radioactive isotopes (e.g., Ba- 33 and Co-57). The L emission lines of the tungsten anode were also useful for energy calibration purposes. The tube output (air kerma per ma at each kvp) was measured using a recently calibrated ionization chamber Abbreviations: kev, kilo-electron volts, a unit of energy; kvp, peak kilovolts applied to the X-ray source; ma, milliamperes the X-ray tube current; mas, milliampere seconds, the product of the tube current and exposure time; SIC, source-to-isocenter distance; SID, source-to-image distance distance between X-ray source and detector. Corresponding author: John M. Boone, PhD, Department of Radiology, UC Davis Medical Center, 4701 X Street, X-ray Imaging Laboratory Sacramento, CA 95817, USA; jmboone@ucdavis.edu. Received in revised form 12 April 2004; Accepted 7 June D 2004 Massachusetts Institute of Technology.

2 150 Small-Animal X-Ray Dose from Micro-CT Boone et al. designed for mammographic applications (MDH Model 9010, Monrovia, CA). Kerma values were made at each kvp, and were corrected for different source to isocenter distances using the inverse square law. A previously reported spectral modeling technique based on polynomial interpolation [11,12] was used to generate X-ray spectra which were subsequently used for dose computations. By modeling the measured spectra, it was possible to compute spectra from 15 to 50 kvp in 1-kVp intervals, instead of the 5-kVp intervals that were measured. Raw (unfiltered) X-ray spectra were physically measured and modeled, and filtered X-ray spectra were calculated from the modeled spectra using mathematical calculations and tabulated mass attenuation coefficients [13]. The mouse X-ray system was capable of operation between 15 and 50 kvp. In some micro-ct scanners and imaging applications, X-ray sources with a higher kvp range are used. To provide dose coefficients for X-ray beams up to 100 kvp, a previous spectral model whose range was 40 to 140 kvp was used [12]. This model was based on an X-ray source from a cardiac catheterization laboratory, and thus the inherent filtration of the metal X-ray tube was much higher (about 2.0 mm Al) than the spectral model below 50 kvp (which used a Bewindowed tube). Several different thickness values (0.5, 1.0, and 2.0 mm) of added aluminum filtration were added mathematically to the lower kvp X-ray spectral measurements made on the Be-windowed X-ray system. Because the X-ray tube had only a thin Be window, the added Al filtration represents essentially the total (added + inherent) Al filtration of the spectrum. For some assessments, the modeled spectra from the Be-windowed X-ray system were mathematically filtered to match the inherent filtration of the cardiac tube (see Appendix). specimen with moving X-ray source and detector system. Indeed, the data generated are also valid for radiographic imaging of the small animal. Most small-animal micro-ct systems make use of cone-beam geometry [1], and z-axis collimation lengths of 1, 4, 8, 12, and 16 cm were studied here. For a 30-cm source-to-isocenter (SIC) distance, these z-axis fields of view (zfov) correspond to cone half-angles of 0.9, 3.8, 7.8, 11.3, and 14.9, respectively (see Figure 1 for the geometrical details). The methodology for computing dose (energy/mass) was similar to that used for human dose computations: The z-dimension of the field of view (at isocenter), combined with the animal diameter, defines the tissue volume and (given a density of 1.0) its mass. The X-ray energy imparted throughout the animal s tissues (except in the dermal layer) was tallied and included in the dose computation, including X-ray energy scattered outside the tissue volume. This geometry therefore assumes that the animal will be scanned over multiple contiguous volumes, since adjacent scans would contribute the same scattered energy to the volume under study as the modeled scan contributes to energy scattered outside this volume. This geometry has relatively small influence in this study, due to the low X-ray energies typically used in mouse imaging and due to the small scattering volumes involved (compared with humans). The SIERRA Monte Carlo code was written in C (C6.0 SP5, Microsoft, Redmond, WA). The Monte Carlo studies were performed on a 2.5-GHz Pentium-based personal computer. The tissue-equivalent cylinders ranged from Monte Carlo Simulation The SIERRA (Simple Investigational Environment for Radiology Research Applications) Monte Carlo code was used to compute the absorbed X-ray energy in unitdensity muscle, and the animal was modeled as a cylinder. Although this simple homogeneous solid-cylinder geometry does not accurately represent the more complex anatomy of the lungs or more attenuating bones of the animal, including such structures reduces the generality of the dose coefficients, probably has a very small effect, and is beyond the capability of the SIERRA code. The simulation included the 360 rotation of the X-ray source around a small animal as in micro-ct applications, and this simulation is valid for both stationary X-ray/ detectors with a moving specimen stage, or a stationary Figure 1. The geometry for the Monte Carlo computation of absorbed dose in the mouse is illustrated. The mouse was modeled as a cylinder of a specific diameter (Dia), located with the center of the cylinder a distance away (sourceto-isocenter, SIC) from the X-ray tube. The field of view along the z-dimension of the animal (zfov) is also illustrated, and this relates to the cone angle used for mouse radiography or cone-beam CT imaging procedures.

3 Small-Animal X-Ray Dose from Micro-CT Boone et al to 40 mm in diameter. A 1-mm-thick layer of dermis (and fur) was modeled, and X-ray energy deposition interior to the dermal/fur layer was tallied. For example, for a 20-mm mouse diameter, the dose was tallied to a concentric cylinder 18 mm in diameter. The fur and metabolically inactive dermis are relatively insensitive to the effects of ionization, and the dose of interest to the mouse is that deposited in the internal organs. This approach is similar to that used in mammography, except human skin layer is typically modeled as being 4.0 mm thick [14]. X ray photons were evaluated at each monoenergetic energy from 6 to 100 kev, in 1-keV intervals. For each X-ray photon energy and mouse phantom diameter, the energy deposition from 10 6 incident photons was tallied. The radiation dose to the mathematical phantom due to the monoenergetic X-ray exposure was normalized to the units of mgy per 10 6 photons/mm 2 incident on the animal at the plane of the isocenter (in air). Let this function be D m (d, E), where D m is the monoenergetic dose, d is the animal diameter, and E is the X-ray energy (kev). The unit (mgy/10 6 photons per mm 2 ) for D m (d, E) was used because it provides for relatively easy computation of polyenergetic dose deposition, when the X-ray spectrum is known. The function D m (d,e) was used to compute the polyenergetic dose coefficients, D p (d,kvp), using the modeled polyenergetic X-ray spectrum, F(E). D p ðd; kvpþ ¼k X E max FðEÞ D m ðd; EÞ ð1þ E¼E min where k is a normalization constant. In practice, the spectrum F(E) is normalized such that it corresponds to 1 mgy of air kerma, and then k is simply With this normalization, the polyenergetic dose coefficients D p (d,kvp) are in the units of absorbed tissue dose per air kerma measured at the plane of the isocenter (mgy/mgy). The D p (d,kvp) coefficients can then be used to convert measured air kerma values at isocenter of a micro-ct scanner to mean absorbed dose in the animal. It is useful to review the basic properties of an X-ray system for readers who may not use X-ray radiation routinely. A stable X-ray system when operating at a constant kvp will produce a relatively constant air kerma rate (formerly called exposure rate), which is a linear function of the ma (e.g., double the ma, and the air kerma rate will double as well). The kerma rate is also strongly dependent (r 2 ) upon distance r from the X-ray focal spot, obeying the inverse square law. The air kerma is a linear function of the product of the tube current (ma) and the exposure time in seconds (sec), typically called the mas. Thus, air kerma rate (mgy/sec) exposure time (sec) equates to air kerma. Precise calculation of the radiation dose to the mouse requires that the air kerma rate per ma, or the air kerma per mas, at the kvp of interest be determined at the isocenter of the scanner. Thus, to accurately measure the X-ray dose to the small animal, it is necessary to make air kerma measurements (in air) using an ionization chamber positioned at the isocenter of the micro-ct scanner. The imaging parameters (kvp, mas per projection image, number of projections, SIC distance, and mouse diameter) need to be recorded as well. Exposure measurements, X, made in mr, can be converted to air kerma, K, using K (mgy) = X (mr). Monte Carlo Comparisons with Measured Data The SIERRA code used for this investigation is a direct daughter-code of the previously validated [14 16] Monte Carlo routines. An extensive set of physical measurements were made and were compared against the SIERRA Monte Carlo (MC) derived coefficients for polymethyl methacrylate (PMMA) cylinders ranging from 10 to 20 cm in diameter [14]. Physical measurements were performed by placing a long thin pencil ionization chamber in a 10-mm diameter concentric hole drilled into each cylinder. Although the presence of the hole caused minimal perturbation of the PMMA attenuation for these large diameter cylinders, the influence of a 10-mm diameter concentric hole would have significant influence on dose measurements in phantoms in the 2- to 3-cm diameter range. Due to this limitation, it was not feasible to make physical measurements using the pencil chamber for comparisons in this study. Goertzen [17] measured the dose in a 30-mm PMMA phantom using thermoluminescent dosimeters (TLDs) placed at six different depths (radii) within the phantom. Dose measurements were performed for a 40-kVp W-anode spectrum filtered with 1.0 mm Al, and for a 50-kVp W-anode system filtered with 2.0 mm Al using 800 projection images. The mean absorbed dose (averaged over the whole phantom) was computed in the current study by weighting the dose values measured at each radii by the area of the annulus corresponding to each TLD position. The air kerma measured at the isocenter reported in Goertzen was combined with the appropriate Monte Carlo coefficient developed in this study, and these computed values were used for comparison with Goertzen s measured TLD dose data. Results and Discussion The raw (unfiltered except by the thin Be-window) X-ray spectra typical of an X-ray system used for mouse

4 152 Small-Animal X-Ray Dose from Micro-CT Boone et al. imaging are illustrated in Figure 2. In the absence of added metallic filtration to the X-ray tube, the raw X-ray spectra emitted from a tungsten anode tube demonstrate very high peaks corresponding to tungsten s characteristic L-shell X-rays. These peaks, visible at 8.4 and 9.7 kev, would lead to exceptionally high doses to the animal if unfiltered spectra were used for imaging. The low-energy X-rays contribute to dose, but play little role in image formation because very few are transmitted through the animal. For example, a 10-keV X-ray photon has a 1 in 45,900 chance of penetrating a 20-mmthick mouse. Therefore, metallic filters (usually aluminum) need to be placed over the X-ray source to attenuate the low-energy X-ray fluence corresponding to the L-emission lines. We have found that a minimum of 0.5 mm of added aluminum filtration is necessary to eliminate the L-fluorescence radiation from tungsten, when a beryllium windowed X-ray tube is used. The dashed lines in Figure 2 correspond to spectra with 0.5 mm of total Al filtration. The monoenergetic dose D m (d,e), as a function of X-ray energy is illustrated in Figure 3, for different imaging geometries and mouse diameters. The curves labeled A in Figure 3 correspond to a 20-mm mouse Figure 2. Two X-ray spectra (solid lines) measured on a typical mouse X-ray imaging system are illustrated. Corresponding spectra mathematically filtered with 0.5 mm of total filtration are illustrated as dashed lines. (A) The 30-kVp tungsten anode X-ray spectra, with and without added filtration, are illustrated. The spectra without aluminum filtration demonstrate the very significant L-emission lines of tungsten. (B) The 50-kVp X-ray spectra are illustrated. Figure 3. The results of the monoenergetic Monte Carlo dose evaluations are illustrated. The graphs labeled A illustrate the dose coefficients D m (d,e) for a 2-cm diameter mouse with an 8-cm zfov, for three different SIC distances (20, 30, 40 cm). The curves labeled B are for a 3-cm diameter mouse and SIC = 15 cm, and show four overlapping curves for different zfov values (4, 8, 12, and 16 cm). The arrow labeled C shows the maximum dependency on FOV, for a 4-cm diameter animal, SID = 15 cm, and zfovs from 1 to 16 cm. diameter and a field of view (zfov defined in Figure 1) of 8 cm. The results for three different SIC distances are shown (10, 20, and 30 cm) in curve A. The similarity in the profiles for the three different SICs demonstrate that dose has little dependence on the SIC, and this is because the ordinate parameter is normalized to the same incident fluence. The curve marked B in Figure 3 corresponds to a 30-mm diameter mouse imaged at SIC = 15 cm. The curves for four zfovs are shown (4, 8, 12, and 16 cm). These curves show small variations as well, and thus dose is not a strong function of the zfov. From geometrical considerations, there will be greater dependence on zfov for bigger animals (larger diameter) and smaller SICs, and the full extent of this is illustrated by the arrow marked C. The upper curve corresponds to a 40-mm diameter animal, with SIC = 15 cm and zfov = 16 cm. The lower curve corresponds to a 40-mm diameter animal, with SIC = 15 cm and zfov = 1 cm. Figure 4 illustrates the dose profiles for different animal diameters, ranging from 10 to 40 mm. Low energy X-rays (<10 kev) contribute little dose to the animal due to the 1-mm thick dermal layer assumed in these dose computations. The dermal layer acts to essentially protect the animal s deeper tissues from X-ray dose deposition for very low X-ray energies. Smaller diameter animals demonstrate higher relative (normalized to unit air kerma at isocenter) dose because they experience less self-shielding (self-attenuation) than larger diameter animals. This is similar to the dose coefficients for mammography, where the coefficients for thinner breasts are higher [18]. Although these coefficients are larger, the overall doses to smaller animals are actually lower, because substantially less radiation is

5 Small-Animal X-Ray Dose from Micro-CT Boone et al. 153 trends: (1) larger diameter animals experience significantly higher doses than smaller diameter animals when appropriate technique factors are chosen such that the signal to the detector is the same, and (2) higher X-ray tube voltages result in substantially lower mean doses to the animal than lower kvps, a well-known trend found in human imaging as well. While going to higher energies reduces dose, X-ray contrast is reduced as well. Different X-ray detector systems have different response functions Figure 4. The monoenergetic dose coefficients D m (d,e) are shown as a function of X-ray energy for eight mouse diameters ranging from 10 to 80 mm. The two dotted lines correspond to 15 and 25 mm mouse diameters. These data are for a zfov of 8 cm, and an SIC of 30 cm. needed to penetrate the animal and deliver adequate X-ray signal levels to the detector. The data illustrated in Figure 4 are the D m (d, E) functions for monoenergetic X-ray beams, and are the direct output of the Monte Carlo computations. The dotted lines emanating from the origin of the graph in Figure 4 correspond to the situation where all X-ray energy incident upon the animal is absorbed and contributes to dose these are marked total absorption curves. These data result from simple calculations where no dermal layer exists, and where no X-ray transmission or scatter escape from the mouse occurs. Figure 5 illustrates the normalized dose profiles D p (d,kvp) as a function of X-ray tube voltage for polyenergetic X-ray spectra. The curves illustrated in Figure 5A C correspond to an X-ray tube with 0.5, 1.0, and 2.0 mm of total (inherent + added) aluminum filtration, respectively. The normalized dose profiles are shown for animal diameters ranging from 10 to 40 mm. These coefficients, multiplied by the air kerma values measured or calculated at the isocenter of a specific CT system, allow investigators to compute the absorbed dose to the animal. The curves illustrated in Figure 6 show the mean dose to the animal as a function of X-ray tube voltage (kvp), for small animals ranging in diameter from 10 to 40 mm. As with Figure 5, Figure 6A C shows the curves for 0.5, 1.0, and 2.0 mm of total aluminum filtration, respectively. These data were computed assuming that the mean air kerma to the detector under the thickest part of the animal was 10 mgy (1.145 mr). An SIC distance of 30 cm and a source-to-detector distance of 35 cm was used for the computation of the curves in Figure 6. The dose profiles in Figure 6 illustrate two relatively strong Figure 5. The normalized dose coefficients D p (d,kvp) as a function of the X-ray tube voltage for polyenergetic X-ray spectra are shown. The data for mouse diameters ranging from 10 to 40 mm are illustrated, and the two dotted lines correspond to 15 and 25 mm mouse diameters. These data are shown for a tungsten anode X-ray system with a beryllium window. The normalized dose coefficients D p (d,kvp) are the mean absorbed dose to the entire mouse (mgy) per air kerma (mgy) measured at the isocenter of the micro-ct scanner. The data shown in these figures are for an 8-cm zfov and an SIC of 30 cm. The three figures are for (A) 0.5 mm total aluminum filtration, (B) 1.0 mm aluminum filtration, and (C) 2.0 mm of total aluminum filtration to the beam.

6 154 Small-Animal X-Ray Dose from Micro-CT Boone et al. Figure 6. The mean absorbed dose as a function of X-ray tube voltage is illustrated for mice ranging from 10 to 40 mm in diameter. These graphs assume that the air kerma to the detector under the center of the mouse was mgy (or 10 Gy). These values are for SIC = 30 cm, and the source-todetector distance was 35 cm. The curves are for (A) 0.5 mm added aluminum filtration, (B) 1.0 mm added Al, and (C) 2.0 mm of added Al filtration. The dose in (B) is referred to in the text (see Discussion). In (C), the doses were smoothed to reduce small differences imposed by the transition in spectral models at 40 kvp. to X-ray exposure, and the dose curves shown in Figure 6 are valid for an X-ray detector with a response such that 10 mgy (1.14 mr) to the detector delivers a useful imaging signal near the low end of the detectors characteristic curve (this corresponds to the signal under the thickest part of the mouse). For X-ray detectors which have different sensitivities, the mean dose levels will vary. High-resolution detector systems with small pixels may require exposures on the order of 20 to 40 mgy per frame to produce high-quality images. For example, the detector described in Goertzen et al. [1] required on the order of 30 mgy per image. Thus, the curves in Figure 6 will probably need to be adjusted upwards for most X-ray detector systems used in smallanimal imaging. For the purposes of approximation, if an X-ray system is 10 times more sensitive (requires 10 times less radiation to produce the same signal level) than that described for Figure 6, the mean dose levels will be reduced by a factor of 10. It is very important to note that the mean dose values illustrated in Figure 6 correspond to the acquisition of only one projection image on a micro-ct scanner. Micro-CT scanners typically acquire between 300 and 1000 individual projections, which are necessary to calculate the CT images. The dose from all of the individual projection images needs to be summed to evaluate the total dose from CT. The dose per projection discussed here assumes that the X-ray tube was shuttered or turned off between the acquisition of individual projection images, while the system or mouse is being rotated. If, however, X-ray exposure occurs between the acquisition of projection data, the integrated entrance kerma from the entire procedure (all projection acquisitions and total dead time in between acquisitions) should be included in the computation. Previously reported comparisons [14] between SIER- RA Monte Carlo dose measurements and physically measured results for the same CT geometry as that studied here but with larger cylinder diameters (10 to 20 cm) demonstrated excellent correlation (r =.9877) over the 60 data points compared, and the Monte Carlo and measured data were not significantly different from each other ( p =.743). Using a 40-kVp/1.0 mm Al spectrum and a 30-mm polycarbonate cylindrical phantom, the mean absorbed dose (averaged over six different depths in the phantom) reported by Goertzen [17] was 30.9 mgy. A dose of 26.3 mgy was computed using the air kerma measured at the phantom s isocenter and the appropriate coefficient in Table 3, a difference of about 15%. For a 50-kVp/2.0 mm spectrum, Goertzen s TLD measurements gave a dose of 22.2 mgy, which compared well to the dose of 23.5 mgy computed using the appropriate coefficient in Table 4, with only a 6% difference. These comparisons suggest that the coefficients reported in this study are capable of reasonably accurate dose assessment in the mouse. Discussion In this investigation, conversion factors (Figure 5) were computed so that researchers can calculate the mean dose to a small animal in an X-ray radiographic or CT

7 Small-Animal X-Ray Dose from Micro-CT Boone et al. 155 Table 1. Mean Dose (mgy) per 10 6 photons/mm 2 at Isocenter Mouse Diameter kev 5mm 10mm 15mm 20mm 25mm 30mm 40mm These data were computed for a 30-cm SIC and a 12-cm zfov. The monoenergetic coefficients D m (d, E ) are given in the units of mgy per 10 6 photons/mm 2 where the fluence is determined at the plane of the isocenter of the scanner. imaging procedure, and specific examples of doses based on a hypothetical detector system were provided (Figure 6) which allow dose approximation. Below, examples are given both for dose estimation and for precise calculation. Dose Estimation ( from Figure 6) When a rough estimate of the dose to the small animal is required, the dose profiles shown in Figure 6 can be used to allow investigators to estimate the radiation dose to their animals from a specific CT procedure. An example calculation is provided to illustrate this: Assume that a micro-ct system has a detector sensitivity where producing a signal near the low end of its dynamic range requires 100 mgy (10 times the values shown in Figure 6), and that a 20-mm diameter mouse is scanned at 40 kvp using 600 projection images. The X-ray system was filtered with about 1.0 mm Al. On Figure 6B, the mean dose for a 20-mm animal diameter Table 2. Mean Dose to Animal (mgy) Per Unit Kerma at Isocenter (mgy) Diameter of Small Animal kvp 5mm 10mm 15mm 20mm 25mm 30mm 40mm These data are for a 30-cm SID and 12-cm zfov. The units are mgy/mgy. The polyenergetic coefficients, D p (d,kvp) for a W-anode X-ray beam with 0.5 mm of added aluminum filtration.

8 156 Small-Animal X-Ray Dose from Micro-CT Boone et al. Table 3. Mean Dose to Animal (mgy) Per Unit kerma at Isocenter (mgy) Diameter of Small Animal kvp 5mm 10mm 15mm 20mm 25mm 30mm 40mm These data are for a 30-cm SID and 12-cm zfov. The units are mgy/mgy. The polyenergetic coefficients, D p (d,kvp) for a W-anode X-ray beam with 1.0 mm of added aluminum filtration are provided. at 40 kvp is mgy (this point is marked). This number is multiplied by 10 to account for the different sensitivity of the detector system, and another factor of 600 is included to account for the fact that 600 images are acquired to produced the CT image dataset. The mean dose to the mouse is thus: Mean mouse dose ðmgyþ ¼ 0:0375 mgy ¼ 226 mgy ð22:6 RadsÞ Dose Computation ( from Table 4) For more precise computations of dose, air kerma measurements and the technique factors specific to the CT scan are required. Air kerma measurements should be determined at the plane of the isocenter without backscatter, using free-in-air measurement geometry (structures which cause scatter should be removed before the kerma measurements are made). The following parameters are assumed in the calculation: beam energy = 80 kvp mouse diameter = 15 mm field of view (along z-axis) = 12 cm SIC distance = 30 cm inherent filtration = 2.0 mm Al air kerma at isocenter = 1.64 mgy/mas product of tube current and exposure time per projection image = 0.32 mas number of projection images per CT scan: 500 images Table 4. Mean Dose to Animal (mgy) Per Unit Kerma at Isocenter (mgy) Diameter of Small Animal kvp 5mm 10mm 15mm 20mm 25mm 30mm 40mm These data were computed for a 30-cm SIC and a 12-cm zfov. The units are mgy/mgy. The polyenergetic coefficients D p (d,kvp) are given for W X-ray spectra with 2.0 mm of added al.

9 Small-Animal X-Ray Dose from Micro-CT Boone et al. 157 Table 4 (where inherent filtration is 2.0 mm Al, and SIC = 30 cm) is consulted at the entry corresponding to a 15-mm mouse diameter and 80 kvp, and this value (bolded on the table) is mgy/mgy. The kerma for the CT acquisition is: Total kerma ¼ 500 images 0:32 mas=image 1:64 mgy=mas ¼ 262:4 mgy ð 30 RoentgensÞ The dose to the mouse is computed as: Mean absorbed dose ðmgyþ ¼ 262:4mGy ðkermaþ 0:9946 ðdose=kermaþ ¼ 261 mgy ðdoseþð26:1 RadsÞ Figure A1. This figure illustrates the half value layer (HVL) as a function of tube voltage for three different levels of added filtration, 0.5 mm, 1.0 mm, and 2.0 mm Al. This figure is provided for readers to compare their relative beam quality to the spectra used in this study. The 2.0-mm curve represents the combined results of two spectral models, one from 15 to 45 kvp, the other from 45 to 100 kvp. The dotted lines adjacent to the 2.0-mm curve (from 15 to 45 kvp) were filtered with 2.1 mm Al (top dotted line) and 1.9 mm (bottom dotted line). Summary Monte Carlo simulations combined with physical measurements of X-ray spectra typical of mouse imaging were used to compute mean absorbed dose coefficients [D p (d,kvp)], normalized to kerma at isocenter. These coefficients will allow investigators to compute the radiation dose to small-animals from X-ray exposure in radiography or micro-ct. Dose computations were performed over a range of realistic mouse imaging geometries. Table 1 gives the dose coefficients (mgy per 10 6 photons/mm 2 ) for monoenergetic X-ray beams. Tables 2, 3, and 4 are dose coefficients (mgy/mgy) for X-ray systems with differing inherent filtration. Coefficient tables for a more comprehensive range of parameters are available by request to the first author. The available tables include dose calculation coefficients [D m (d,e) for E = 6 to 100 kev, and for d = 5 to 40 mm]. The polyenergetic dose coefficients, D p (d,kvp), are available for kvps ranging from 15 to 100 kvp, and for d from 10 to 40 mm. The polyenergetic tables are also available for differing amounts of added aluminum filtration. Appendix Two X-ray spectral models were used in this investigation, a previously reported model which was tuned to emulate a cardiac catheterization system ( kvp, CATH spectra) [12] and the other was produced from the X-ray spectra measured (15 50 kvp) in this study (MOUSE spectra), as described in the Materials and Methods section. To adjust the unfiltered MOUSE spectra to the same level of inherent filtration as the CATH spectra, aluminum filtration was added mathematically to the MOUSE spectra and the resulting half value layer (HVL) was computed. It was found by iteration that 2.0 mm of added filtration to the MOUSE spectra provided HVLs that were consistent with the unfiltered CATH spectral model. Figure A1 illustrates the HVLs for the X-ray spectra discussed in this study. Acknowledgments We thank the reviewers for their comments. This research was supported in part by a grant from the California Breast Cancer Research Program (7EB-075), from the National Cancer Institute (R01 CA89260), and from the National Institute for Biomedical Imaging and Bioengineering (R01 EB and EB00230). The UC Davis Cancer Center also provided funding for this project. References [1] Goertzen AL, Meadors AK, Silverman RW, Cherry SR (2002). Simultaneous molecular and anatomical imaging of the mouse in vivo. Phys Med Biol. 47: [2] Kapadia RD, Stroup GB, Badger AM, Koller B, Levin JM, Coatney RW, Dodds RA, Liang X, Lark MW, Gowen M (1998). Applications of micro-ct and MR microscopy to study pre-clinical models of osteoporosis and osteoarthritis. Technol Health Care. 6: [3] Gleason SS, Sari-Sarraf H, Abidi MA, Karakashian O, Morandi F (2002). A new deformable model for analysis of X-ray CT images in preclinical studies of mice for polycystic kidney disease. IEEE Trans Med Imaging. 21: [4] Paulus MJ, Gleason SS, Easterly ME, Foltz CJ (2001). A review of high-resolution X-ray computed tomography and other imaging modalities for small animal research. Lab Anim (NY). 30: [5] Paulus MJ, Gleason SS, Kennel SJ, Hunsicker PR, Johnson DK (2000). High resolution X-ray computed tomography: An

10 158 Small-Animal X-Ray Dose from Micro-CT Boone et al. emerging tool for small animal cancer research. Neoplasia. 2: [6] Ford NL, Thornton MM, Holdsworth DW (November 2003). Fundamental image quality limits for microcomputed tomography in small animals. Med Phys. 30: [7] Mitchel RE, Jackson JS, Morrison DP, Carlisle SM (2003). Low doses of radiation increase the latency of spontaneous lymphomas and spinal osteosarcomas in cancer-prone, radiationsensitive Trp53 heterozygous mice. Radiat Res. 159: [8] Ohnishi T, Takahashi A, Ohnishi K, Yonezawa M (2001). Tumor suppressor p53 response is blunted by low-dose radiation. Phys Med. 1: [9] Wan H, Gong SL, Liu SZ (2001). Effects of low dose radiation on signal transduction of neurons in mouse hypothalamus. Biomed Environ Sci. 14: [10] Chen SL, Cai L, Meng QY, Xu S, Wan H, Liu SZ (2000). Low-dose whole-body irradiation (LD-WBI) changes protein expression of mouse thymocytes: Effect of a LD-WBI-enhanced protein RIP10 on cell proliferation and spontaneous or radiation-induced thymocyte apoptosis. Toxicol Sci. 55: [11] Boone JM, Fewell TR, Jennings RJ (1997). Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography. Med Phys. 24: [12] Boone JM, Seibert JA (1997). An accurate method for computergenerating tungsten anode x-ray spectra from 30 to 140 kv. Med Phys. 24: [13] Boone JM, Chavez AE (1996). Comparison of x-ray cross sections for diagnostic and therapeutic medical physics. Med Phys. 23: [14] Boone JM, Shah N, Nelson TR (2004). A comprehensive analysis of DgN(CT) coefficients for pendant-geometry cone-beam breast computed tomography. Med Phys. 31: [15] Boone JM, Buonocore MH, Cooper VN III (2000). Monte Carlo validation in diagnostic radiological imaging. Med Phys. 27: [16] Boone JM, Cooper VN III (2000). Scatter/primary in mammography: Monte Carlo validation. Med Phys. 27: [17] Goertzen AL (2003). Development of a combined micropet and microct system for mouse imaging. University of California Los Angeles, PhD Dissertation. [18] Boone JM (2002). Normalized glandular dose (DgN) coefficients for arbitrary X-ray spectra in mammography: Computer-fit values of Monte Carlo derived data. Med Phys. 29:

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Journal of Physics: Conference Series PAPER OPEN ACCESS Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Recent citations - Resolution Properties of a

More information

Imaging Technique Optimization of Tungsten Anode FFDM System

Imaging Technique Optimization of Tungsten Anode FFDM System Imaging Technique Optimization of Tungsten Anode FFDM System Biao Chen a*, Andrew P. Smith b, Zhenxue Jing a, Elena Ingal a a Hologic, Inc. 600 Technology Drive, DE 1970 b Hologic, Inc. 35 Crosby Drive,

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

I. Introduction.

I. Introduction. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case

More information

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.125 pissn 2508-4445, eissn 2508-4453 Optimization of Energy Modulation Filter for Dual Energy CBCT

More information

PD233: Design of Biomedical Devices and Systems

PD233: Design of Biomedical Devices and Systems PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

Evaluation of X-Ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem78 Models

Evaluation of X-Ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem78 Models Original Article Evaluation of X-Ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem78 Models Suk Chiang Chen, Wei Loong Jong, Ahmad Zaky Harun Submitted: 20 Aug 2011 Accepted:

More information

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage

More information

Mammography: Physics of Imaging

Mammography: Physics of Imaging Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This

More information

Effect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies

Effect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies ARTICLE IN PRESS Radiation Physics and Chemistry 68 (2003) 707 720 Effect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies M. Assiamah, D. Mavunda,

More information

2012 :15th SESSION of ESMP

2012 :15th SESSION of ESMP 2012 :15th SESSION of ESMP Lecture presented in Archamps (Salève Building) by : Elly CASTELLANO (London) Patient dosimetry in x-ray imaging and CT Elly Castellano Objectives measurable dose quantities

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

Introduction of a Single Chip TLD System for Patient Dosimetry

Introduction of a Single Chip TLD System for Patient Dosimetry Introduction of a Single Chip TLD System for Patient Dosimetry C. Hranitzky a, M. Halda a, G. Müller a, B. Obryk b, H. Stadtmann a* a Austrian Research Centers GmbH ARC, 2444 Seibersdorf, Austria. b Institute

More information

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS By Aravind Ravichandran arr192@mail.usask.ca University of Saskatchewan Address: 2424 Cumberland

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

Characterization of photon counting CZT detectors for medical x-ray imaging and spectroscopy

Characterization of photon counting CZT detectors for medical x-ray imaging and spectroscopy Louisiana State University LSU Digital Commons LSU Doctoral Dissertations Graduate School 2011 Characterization of photon counting CZT detectors for medical x-ray imaging and spectroscopy Shannon Fritz

More information

X-rays in medical diagnostics

X-rays in medical diagnostics X-rays in medical diagnostics S.Dolanski Babić 2017/18. History W.C.Röntgen (1845-1923) discovered a new type of radiation Nature, Jan. 23. 1896.; Science, Feb.14. 1896. X- rays: Induced the ionization

More information

The importance of radiation quality for optimisation in radiology

The importance of radiation quality for optimisation in radiology Available online at http://www.biij.org/2007/2/e38 doi: 10.2349/biij.3.2.e38 biij Biomedical Imaging and Intervention Journal COMMENTARY The importance of radiation quality for optimisation in radiology

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

Derivation of total filtration thickness for diagnostic x-ray source assembly

Derivation of total filtration thickness for diagnostic x-ray source assembly Physics in Medicine & Biology PAPER OPEN ACCESS Derivation of total filtration thickness for diagnostic x-ray source assembly To cite this article: Michiharu Sekimoto and Yoh Katoh 2016 Phys. Med. Biol.

More information

Update on NIST X-Ray Air-Kerma Standards and Calibrations

Update on NIST X-Ray Air-Kerma Standards and Calibrations CCRI(I)/03-19 Update on NIST X-Ray Air-Kerma Standards and Calibrations Michelle O Brien and Stephen M. Seltzer Ionizing Radiation Division National Institute of Standards and Technology Gaithersburg,

More information

Monte Carlo study on a new concept of a scanning photon beam system for IMRT

Monte Carlo study on a new concept of a scanning photon beam system for IMRT NUKLEONIKA 2011;56(4):291 297 ORIGINAL PAPER Monte Carlo study on a new concept of a scanning photon beam system for IMRT Anna M. Wysocka-Rabin, Günter H. Hartmann Abstract. Intensity-modulated radiation

More information

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging The Physics of Radiation Oncology X-ray Imaging Charles E. Willis, Ph.D. DABR Associate Professor Department of Imaging Physics The University of Texas M.D. Anderson Cancer Center Houston, Texas Overview

More information

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 325 ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER ABSTRACT William Chang, Jonathan Kerner, and Edward

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

X-ray tube with needle-like anode

X-ray tube with needle-like anode NUKLEONIKA 2002;47(3):101 105 ORIGINAL PAPER X-ray tube with needle-like anode Mieczys aw S apa, W odzimierz StraÊ, Marek Traczyk, Jerzy Dora, Miros aw Snopek, Ryszard Gutowski, Wojciech Drabik Abstract

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

NOT FOR DISTRIBUTION JINST_083P_0914 v1

NOT FOR DISTRIBUTION JINST_083P_0914 v1 Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam F. Di Lillo a,b, D. Dreossi c, F. Emiro a,b, C. Fedon d,e, R. Longo d,e, G. Mettivier a,b,*, L. Rigon d,e, P. Russo

More information

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure Accuracy and calibration of integrated radiation output indicators in diagnostic radiology: A report of the AAPM Imaging Physics Committee Task Group 190 Pei-Jan P. Lin a) Virginia Commonwealth University

More information

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013).

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013). CODE: RADT 151 INSTITUTE: Health Science TITLE: Radiographic Exposure DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of radiographic exposure selection and manipulation

More information

New spectral benefi ts, proven low dose

New spectral benefi ts, proven low dose New spectral benefi ts, proven low dose Philips MicroDose mammography SI, technical data sheet Philips MicroDose SI with single-shot spectral imaging is a fullfi eld digital mammography solution that delivers

More information

MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM *

MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM * MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM * C. Kessler and D.T. Burns December 2018 BUREAU INTERNATIONAL DES POIDS ET MESURES

More information

Development of new dosimeter for measuring dose distribution in CT

Development of new dosimeter for measuring dose distribution in CT Development of new dosimeter for measuring dose distribution in CT Poster No.: C-2925 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology - Without Subtopic Authors: Y. Muramatsu, K.

More information

Beam-Restricting Devices

Beam-Restricting Devices Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

Dosimetry for spectral molecular imaging of small animals with MARS-CT

Dosimetry for spectral molecular imaging of small animals with MARS-CT Dosimetry for spectral molecular imaging of small animals with MARS-CT Noémie Ganet a,d, Nigel Anderson b, Stephen Bell c, Anthony Butler a,b,c,d, Phil Butler a,b,c,d, Pierre Carbonez a, Nicholas Cook*

More information

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann

More information

Wide beam CT dosimetry. Elly Castellano

Wide beam CT dosimetry. Elly Castellano Wide beam CT dosimetry Elly Castellano Outline revision: CT dose indices wide-beam CT: the end of the road for CTDI? the IEC rescue plan for CTDI 100 the american way AAPM report 111 better estimates of

More information

Effect of patient support pads on image quality and dose in fluoroscopy a)

Effect of patient support pads on image quality and dose in fluoroscopy a) Effect of patient support pads on image quality and dose in fluoroscopy a) William R. Geiser, Walter Huda, b) and Nikolaos A. Gkanatsios Department of Radiology, University of Florida, Gainesville, Florida

More information

In-Vivo Imaging: IVIS Lumina XR. William R. Anderson IVIS Product Specialist

In-Vivo Imaging: IVIS Lumina XR. William R. Anderson IVIS Product Specialist In-Vivo Imaging: IVIS Lumina XR William R. Anderson IVIS Product Specialist 1 What will be covered? Introduction Principles of optical In Vivo Imaging Key IVIS Hardware components Overview of Living Image

More information

Pinhole collimator design for nuclear survey system

Pinhole collimator design for nuclear survey system Annals of Nuclear Energy 29 (2002) 2029 2040 www.elsevier.com/locate/anucene Pinhole collimator design for nuclear survey system Wanno Lee*, Gyuseong Cho Department of Nuclear Engineering, Korea Advanced

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY 12477 (800)VIP-XRAY (845)247-9028 Fax or 800-847-9729 local phone 845-246-9013 http://www.eccxray.com sales@eccxray.com INSTRUCTION MANUAL

More information

Calibration of KAP meters

Calibration of KAP meters Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area

More information

Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging

Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2008 Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging Dean

More information

Nuclear Associates , , CT Head and Body Dose Phantom

Nuclear Associates , , CT Head and Body Dose Phantom Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product

More information

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET 2005 IEEE Nuclear Science Symposium Conference Record M11-126 Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET Jin Zhang, Member,

More information

Clinical Experiences with a Patient Skin Dose Monitoring and Tracking Program

Clinical Experiences with a Patient Skin Dose Monitoring and Tracking Program Clinical Experiences with a Patient Skin Dose Monitoring and Tracking Program Allen R. Goode, MS, DABR Chief Diagnostic Medical Physicist Department of Radiology & Medical Imaging University of Virginia

More information

Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography

Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography Current Implementation Status of IEC Standard 62494-1: Exposure Index (EI) for Digital Radiography July 31, 2017 Ryan Fisher, PhD, DABR Katie Hulme, MS, DABR None Disclosures Outline Review of IEC Standard

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision Minnesota Rules, Chapter 4732 X-ray Revision DRAFT FLUOROSCOPIC X-RAY SYSTEMS, 1.0 Subpart 1. Applicability. Subpart 2. Limitation of the useful beam. Subpart 3. Measuring compliance; primary protective

More information

Radiographic Techniques, Contrast, and Noise in X-Ray Imaging

Radiographic Techniques, Contrast, and Noise in X-Ray Imaging Residents Section Physics Minimodule Huda and Abrahams Techniques, Contrast, and Noise in Radiography Residents Section Physics Minimodule Residents inradiology Walter Huda 1 R. Brad Abrahams 2 Huda W,

More information

MWPC Gas Gain with Argon-CO 2 80:20 Gas Mixture

MWPC Gas Gain with Argon-CO 2 80:20 Gas Mixture IMA Journal of Mathematical Control and Information Page 1 of 10 doi:10.1093/imamci/dri000 1. Principles of Operation MWPC Gas Gain with Argon-CO 2 80:20 Gas Mixture Michael Roberts A multi-wire proportional

More information

Ludlum Medical Physics

Ludlum Medical Physics Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam

Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam Home Search Collections Journals About Contact us My IOPscience Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam This content has been downloaded from IOPscience.

More information

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation Image Quality Review I: Basics and Image Quality TH-A-16A-1 Thursday 7:30AM - 9:30AM Room: 16A J. Anthony

More information

K-edge subtraction X-ray imaging with a pixellated spectroscopic detector

K-edge subtraction X-ray imaging with a pixellated spectroscopic detector K-edge subtraction X-ray imaging with a pixellated spectroscopic detector Silvia Pani Department of Physics, University of Surrey Summary Hyperspectral imaging K-edge subtraction X-ray imaging for mammography

More information

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

Effective energy measurement using radiochromic film: application of a mobile scanner

Effective energy measurement using radiochromic film: application of a mobile scanner Polish Journal of Medical Physics and Engineering 2016;22(4):85-92 December 2016 doi: 10.1515/pjmpe-2016-0015 ISSN 1898-0309 Scientific Paper Effective energy measurement using radiochromic film: application

More information

CR Basics and FAQ. Overview. Historical Perspective

CR Basics and FAQ. Overview. Historical Perspective Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors

More information

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images Aiping Ding, Bin Han, Lei Wang, Lei Xing Department of Radiation Oncology, Stanford University School of

More information

DALLA LUCE VISIBILE AI RAGGI X: NUOVI RIVELATORI DI IMMAGINI PER RAGGI X A DISCRIMINAZIONE IN ENERGIA ED APPLICAZIONI

DALLA LUCE VISIBILE AI RAGGI X: NUOVI RIVELATORI DI IMMAGINI PER RAGGI X A DISCRIMINAZIONE IN ENERGIA ED APPLICAZIONI DALLA LUCE VISIBILE AI RAGGI X: NUOVI RIVELATORI DI IMMAGINI PER RAGGI X A DISCRIMINAZIONE IN ENERGIA ED APPLICAZIONI D. Pacella ENEA - Frascati LIMS, Frascati 14-15 ottobre 2015 Come per la fotografia:

More information

THE USE OF CdTe DETECTORS FOR DENTAL X-RAY SPECTROMETRY

THE USE OF CdTe DETECTORS FOR DENTAL X-RAY SPECTROMETRY 2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 THE USE OF CdTe DETECTORS

More information

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring

More information

Study of increased radiation when an x-ray tube is placed in a strong magnetic field

Study of increased radiation when an x-ray tube is placed in a strong magnetic field Study of increased radiation when an x-ray tube is placed in a strong magnetic field Zhifei Wen Departments of Radiology and Physics, Stanford University, Stanford, California 94305 Norbert J. Pelc Departments

More information

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC R. Bellazzini a,b, G. Spandre a*, A. Brez a, M. Minuti a, M. Pinchera a and P. Mozzo b a INFN Pisa

More information

LECTURE 1 The Radiographic Image

LECTURE 1 The Radiographic Image LECTURE 1 The Radiographic Image Prepared by:- KAMARUL AMIN ABDULLAH @ ABU BAKAR UiTM Faculty of Health Sciences Medical Imaging Department 11/23/2011 KAMARUL AMIN (C) 1 Lesson Objectives At the end of

More information

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations S. Baechler 1, P. Monnin 1, A. Aroua 1, J.F. Valley 1, M. Perrier, P. Trueb 3, F.R. Verdun 1 1 University

More information

Dosepix Detector as kvp-meter in Radiology and Mammography: First steps

Dosepix Detector as kvp-meter in Radiology and Mammography: First steps Dosepix Detector as kvp-meter in Radiology and Mammography: First steps F.Bisello, I.Ritter, F.Tennert, A.Zang MediPix Collaboration Meeting, 19th February 2014, CERN Protect, Enhance, and Save Lives -

More information

Vanderbilt University

Vanderbilt University Vanderbilt University Department of Biomedical Engineering Using Fluorescence X-ray for Non-invasive Biopsy Team Members: Savannah Gill Kelvin Lin Mike McHugh Trey Reece Derric Williams Advisor: Frank

More information

Data. microcat +SPECT

Data. microcat +SPECT Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for

More information

Published text: Institute of Cancer Research Repository Please direct all s to:

Published text: Institute of Cancer Research Repository   Please direct all  s to: This is an author produced version of an article that appears in: MEDICAL PHYSICS The internet address for this paper is: https://publications.icr.ac.uk/1316/ Copyright information: http://www.aip.org/pubservs/web_posting_guidelines.html

More information

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

More information

Phase Imaging Using Focused Polycapillary Optics

Phase Imaging Using Focused Polycapillary Optics Phase Imaging Using Focused Polycapillary Optics Sajid Bashir, Sajjad Tahir, Jonathan C. Petruccelli, C.A. MacDonald Dept. of Physics, University at Albany, Albany, New York Abstract Contrast in conventional

More information

Preliminary studies of a new monitor ionization chamber

Preliminary studies of a new monitor ionization chamber 1 Preliminary studies of a new monitor ionization chamber Maíra T. Yoshizumi, Vitor Vivolo and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN) Comissão Nacional de Energia

More information

Determination of the tube voltage from clinic mammographic system using two types of detectors

Determination of the tube voltage from clinic mammographic system using two types of detectors BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 03-1A (2015) 01-06 Determination of the tube voltage from clinic mammographic system using two types of detectors J. S. Barreira a ; V. Vivolo a a Gerência

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

Master of Science Thesis. SIMIND Based Pinhole Imaging

Master of Science Thesis. SIMIND Based Pinhole Imaging Master of Science Thesis SIMIND Based Pinhole Imaging * Development and Validation Kurt Sundin Supervisor: Michael Ljungberg, PhD Medical Radiation Physics Clinical Sciences, Lund Lund University, 2006

More information

Investigation of Effective DQE (edqe) parameters for a flat panel detector

Investigation of Effective DQE (edqe) parameters for a flat panel detector Investigation of Effective DQE (edqe) parameters for a flat panel detector Poster No.: C-1892 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit D. Bor 1, S. Cubukcu 1, A. Yalcin 1, O.

More information

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS Veterinary Science Preparatory Training for the Veterinary Assistant Floron C. Faries, Jr., DVM, MS Radiology Floron C. Faries, Jr., DVM, MS Objectives Determine the appropriate machine settings for making

More information

Image Quality Artifacts in Digital Imaging

Image Quality Artifacts in Digital Imaging MAHIDOL UNIVERSITY Wisdom of the Land Image Quality Artifacts in Digital Imaging Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University, Bangkok,

More information

RULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER USE OF X-RAY APPARATUS

RULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER USE OF X-RAY APPARATUS RULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER 0400-20-06 USE OF X-RAY APPARATUS TABLE OF CONTENTS 0400-20-06-.01 Purpose 0400-20-06-.06 Veterinary

More information

IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film

IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film CYRIC Annual Report 2005 IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film Ohuchi H. 1, and Abe K. 2 1 Graduate School of Pharmaceutical Sciences,

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

PATIENT EFFECTIVE DOSES IN DIAGNOSTIC RADIOLOGY, NA

PATIENT EFFECTIVE DOSES IN DIAGNOSTIC RADIOLOGY, NA Title of Paper: Patient effective doses in diagnostic radiology Authors: N.A. Gkanatsios, and W. Huda * Corresponding Author: Department of Radiology, University of Florida, P.O. Box 100374, Gainesville,

More information

The effect of focal spot size on the spatial resolution of variable resolution X-ray CT scanner

The effect of focal spot size on the spatial resolution of variable resolution X-ray CT scanner The effect of focal spot size on the spatial resolution of variable resolution X-ray CT scanner H. Arabi, A.R. Kamali Asl *, S.M. Aghamiri Iran. J. Radiat. Res., 2010; 8 (1): 37-43 Radiation Medicine Department,

More information

CT Basics: Data Acquisition Module 3

CT Basics: Data Acquisition Module 3 Module 3 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-649 CDRH Fluoroscopic Phantom Users Manual March 2005 Manual No. 07-649-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information