A novel and fast method for proton range verification using a step wedge and 2D scintillator

Size: px
Start display at page:

Download "A novel and fast method for proton range verification using a step wedge and 2D scintillator"

Transcription

1 A novel and fast method for proton range verification using a step wedge and 2D scintillator Jiajian Shen, a) Bryce C. Allred, Daniel G. Robertson, Wei Liu, and Terence T. Sio Department of Radiation Oncology, Mayo Clinic Arizona, Phoenix, AZ 85054, USA Nicholas B. Remmes Department of Radiation Oncology, Mayo Clinic Rochester, Rochester, MN 55905, USA Sameer R. Keole and Martin Bues Department of Radiation Oncology, Mayo Clinic Arizona, Phoenix, AZ 85054, USA (Received 5 May 2017; revised 11 June 2017; accepted for publication 22 June 2017; published 31 July 2017) Purpose: To implement and evaluate a novel and fast method for proton range verification by using a planar scintillator and step wedge. Methods: A homogenous proton pencil beam plan with 35 energies was designed and delivered to a 2D flat scintillator with a step wedge. The measurement was repeated 15 times (3 different days, 5 times per day). The scintillator image was smoothed, the Bragg peak and distal fall off regions were fitted by an analytical equation, and the proton range was calculated using simple trigonometry. The accuracy of this method was verified by comparing the measured ranges to those obtained using an ionization chamber and a scanning water tank, the gold standard. The reproducibility was evaluated by comparing the ranges over 15 repeated measurements. The sensitivity was evaluated by delivering to same beam to the system with a film inserted under the wedge. Results: The range accuracy of all 35 proton energies measured over 3 days was within 0.2 mm. The reproducibility in 15 repeated measurements for all 35 proton ranges was mm. The sensitivity to range variation is 0.1 mm for the worst case. This efficient procedure permits measurement of 35 proton ranges in less than 3 min. The automated data processing produces results immediately. The setup of this system took less than 5 min. The time saving by this new method is about two orders of magnitude when compared with the time for water tank range measurements. Conclusions: A novel method using a scintillator with a step wedge to measure the proton range was implemented and evaluated. This novel method is fast and sensitive, and the proton range measured by this method was accurate and highly reproducible American Association of Physicists in Medicine [ proton pencil beam scanning, proton range measurements, quality assurance, scintilla- Key words: tor, wedge 1. INTRODUCTION The interest in using protons for radiation therapy is rapidly increasing due to beneficial integral dose characteristics. 1 As compared to conventional x-ray radiation therapy, the sharp distal dose fall-off of protons enables significantly lower integral dose and the capability of improved sparing of healthy tissue, while still maintaining a conformal dose distribution to the target. 2 On the other hand, the finite range and sharp distal dose fall-off, if not accurately quantified and verified, can lead to significant under-dosing of the target and over-dosing of critical healthy tissue or organs at risk. 3,4 Therefore, proton range quality assurance (QA) is essential to ensure high-quality proton radiation therapy. The gold standard to quantify proton range is to measure the depth dose with an ionization chamber in a water tank. However, a water tank is typically used only during initial machine commissioning and annual QA. It is an inconvenient tool for frequent QA testing such as daily, weekly and monthly measurements. Alternative methods to quantify proton range exist but have other disadvantages. For example, Zebra (IBA Dosimetry, Schwarzenbruck, Germany), a commercial multiple layer ionization chamber device, has been used to measure proton ranges. However, its maximum measurement uncertainty was reported to be more than 1 mm. 5 Gafchromic EBT Film has been used for proton range measurements during commissioning, and a 0.5-mm accuracy was reported. 6 However, film is not efficient for a frequent constancy check of ranges for multiple proton energies because of postprocessing requirements and the number of films required. Volumetric scintillator detectors have also been used for fast range verification in proton therapy, 7 and a few recent studies reported that sub-millimeter accuracy was achieved, 8,9 but this is not currently a commercially available technology. In this letter, we propose a novel analytic method using a large 2D scintillator of 30 cm 9 40 cm combined with a customized wedge device to efficiently measure the proton range to 0.2 mm precision. Notably, the large 2D scintillator 4409 Med. Phys. 44 (9), September /2017/44(9)/4409/ American Association of Physicists in Medicine 4409

2 4410 Shen et al.: Fast proton range check by 2D scintillator 4410 device also allows for simultaneous quality assurance for spot size and spot position. 10,11 Other technologies discussed above do not currently offer such efficient and convenient allin-one functionality for range, spot size, and spot position QA. 2. METHODS AND MATERIALS 2.A. The 2D scintillator and the step wedge The 2D scintillator XRV4000 (Logos Systems Int l, Scotts Valley, CA, USA) has an effective measurement area of 30 cm 9 40 cm, and spatial resolution of mm/ pixel. XRV4000 has 4.9-mm of acrylic buildup which includes the top acrylic cover, acrylic under the cover for mechanical support of the scintillator material, and the scintillator itself. A sketch of the wedge LCW-200 Plus (Logos Systems Int l, Scotts Valley, CA, USA) is shown in Fig. 1. The wedge was made of acrylic with six graduating steps (named A to F from the highest to the lowest, respectively). On the top of each step is a pair of 45 wedges. The highest step (i.e., step A) is 20 cm tall in the center and 15 cm on both sides. The height decreases by 2 cm for each subsequent step. As a result, step F is 10 cm tall in the center and 5 cm on the sides. The footprint of the wedge is 10 cm (front dimension) 9 15 cm (side dimension). The wedge is attached to a 1.2 cm acrylic base. The combined thickness of the wedge base and the native device buildup is 1.69 cm, as shown in Fig B. The beam design and the range detection A scanning proton beam treatment plan with 35 energy layers was designed with energies from 121 to 189 MeV and beam ranges in water from 10.6 to cm. Each layer was composed of 625 spots, evenly distributed at 6-mm spot spacing. This gave an intended field size of 15 cm 9 15 cm which covered the entire footprint of the step wedge. The spots were equally weighted in order to deliver a laterally homogeneous fluence to the step wedge. As proton spots from a given energy layer are delivered to the step wedge, the range is determined by the water equivalent thickness (WET) of the step wedge through which it traveled. An example of the images obtained from the scintillator by this method for E = MeV (range = 174 mm) is shown in Fig. 2(a). The dark pixels at steps A and B are due to low energy beams unable to reach the scintillator. Two bright regions occur in both steps C and D, where the WET of the wedges is equivalent to the proton range. Hence, the proton range can be derived from the position of the brightest pixels on the scintillator image, by simple trigonometric principles. For steps E and F, the protons were energetic enough to pass completely through the wedges and the scintillator. After raw images were captured by scintillator and the wedge system, an in-house data analysis to derive the proton range was developed. The longer curve in the Fig. 2(b) was obtained by averaging the light intensity profiles in the top rectangular ROI as highlighted in the left of Fig. 2(a). The purpose of the ROI average was to get a smooth Bragg peak curve FIG. 1. The front and side-view sketch of the step wedge, with the 16.9 mm acrylic base and buildup of the scintillator also included. The peak heights of wedges A to F are 200 mm, 180 mm, 160 mm, 140 mm, 120 mm and 100 mm, respectively. The side heights are 50 mm shorter than the corresponding peak heights in a uniform fashion. The footprint is 100 mm mm. The wedge angle is 45.

3 4411 Shen et al.: Fast proton range check by 2D scintillator 4411 FIG. 2. (a): an image captured by the scintillator with the step wedge for protons at E = MeV. The bright pixels in wedges C and D were caused by the Bragg peak dose deposition. The rectangles identify the regions of interest (ROI) for data analysis. (b): the longer curve is the averaged light intensities of the top rectangular ROI as shown in left image. The shorter curves represent the fit to the longer curve. The x labels mark the locations of 80% of the peak intensities, which were used to calculate the half- distance between Bragg peaks (L). [Color figure can be viewed at wileyonlinelibrary.com] by removing the effects of camera readout noise. Equation 1 was used to fit the light intensities (y) of the smoothed Bragg peaks, 2 C2 ðc3 xþ y ¼ C1 exp C4 1 pffiffi 2 1 þ erf C3 x 2 C2 C5 pffiffi (1) 2 C5 C4 where C1 to C5 are the fit parameters, and x is its corresponding position on the scintillator. Equation 1 was drawn from the analytical formula for the proton integral depth dose, 12,13 but it was greatly simplified as we only needed to fit the critical region near the Bragg peak and distal fall-off. After curve fitting was performed, the points with 80% of the peak intensity on the distal fall-off side were identified [see cross labels in Fig. 2(b)]. The half-distance between Bragg peaks [L, as shown in Fig. 2(b)] is defined as half of the distance between the 80% points. 2.C. The range quantification method The proton range is the water equivalent thickness (WET) that the beam traveled between the position where proton pencil beam entered the wedge surface and the position where the same protons generated Bragg peak brightness in the scintillator. Since the proton beam has divergence in both X and Y directions, the distance that the beam travels can be calculated using the three projections in the X, Y, and Z directions, respectively. In Fig. 3, the beam path d was projected to the X plane and Y plane, respectively. Using trigonometry and the Pythagorean Theorem, we can derive, qffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi R wedge ¼ C WET d ¼ C WET d 2 x þ d 2 2 y þ d z (2) d z ¼ h a ¼ h SAD x h SAD x L L d x ¼ L a ¼ L SAD x h SAD x L L (2.a) (2.b) d y ¼ b d z (2.c) SAD y d z where C WET is the conversion factor from acrylic thickness to WET, h is the peak height of the wedges which includes the base and the buildup, L is the half peak distance measured by the scintillator as defined above, b is the distance from the isocenter to the center of the wedge that the protons pass through, and SAD x (135 cm) and SAD y (192.5 cm) are the virtual SAD of the beam in the X and Y directions, respectively. The WET of the wedge was measured using the scanning water tank, with a resulting value of C WET ¼ 1: D. Accuracy and reproducibility test The scintillator device (XRV4000) was placed on the couch and the center of the scintillator device was aligned

4 4412 Shen et al.: Fast proton range check by 2D scintillator 4412 FIG. 3. The projections of the beam passing through the wedge into the X and Y planes are represented. The d x,d y, and d z are the projections of the beam trajectory d in the X, Y, and Z directions, respectively. h is the peak height of the wedges which includes the base and the buildup, L is the half peak distance as defined above, b is the distance from the isocenter to the center of the wedge that protons pass through, and SAD x (135 cm) and SAD y (192.5 cm) are the virtual SAD of the beam in the X and Y directions, respectively. [Color figure can be viewed at wileyonlinelibrary.com] to isocenter using the laser. The step wedge was placed on the top of the scintillator device and the center of the wedge was aligned to the laser. The proton beam described in the above section was delivered 5 times a day for 3 days. The data analysis process was performed, and the proton range R wedge was derived. For each day, R wedge from five repeated measurements was averaged, and these average ranges were compared with the ranges measured by a scanning water tank (R 80, 80% of the peak at the distal falloff): ΔR =R wedge R 80. The reproducibility of this novel technique was then evaluated by comparing the R wedge results from all 15 measurements. In order to test the sensitivity to the change of ranges, one EB3 Gafchromic film (Ashland, Bridgewater, NJ, USA) was placed under the wedge and five repeated beams were delivered. The WET of 10 stacked films was measured using the scanning water tank, giving a measured single-film WET of 0.36 mm. R wedge was derived from this data set without accounting for the film, so that the use of the film can be treated as generating a new set of 35 energies whose corresponding ranges were smaller than the original set by 0.36 mm (film WET). R wedge from five repeated measurements with film was averaged and compared with the ranges by scanning water tank. 3. RESULTS 3.A. Range measurement accuracy The accuracy of ranges measured with the step wedge is shown in Fig. 4. Please note that the range measured by step wedge was systematically lower than the scanning water tank measurements (Δ = 0.51 mm). The cross, plus and circle symbols show the range accuracy of all 35 energies measured on days 1 through 3, after the application of a systematic shift of Δ = 0.51 mm. As a result, the accuracy of the range as measured by step wedge was within 0.2 mm after the systematic shift was applied. The standard deviation of the day-to-day variation was very small at only 0.04 mm. The star symbols show the range accuracy measured with the step wedge with a film inserted under the wedge. The star symbols are on average 0.35 mm (r = 0.04 mm) less than green symbols for all 35 energies, and with minimum and maximum deviations at 0.26 mm and 0.42 mm, respectively. 3.B. Range measurement reproducibility The reproducibility of proton beam ranges as measured with the step wedge for all 35 energies in 15 measurements is shown in Fig. 5. The reproducibility is within 0.3 mm, with a standard deviation of mm. The dashed lines mark one and two standard deviations (r), respectively. 4. DISCUSSION The underlying principle of this novel method was to project the depth dose into the lateral direction using a wedge, so that the depth dose could be measured with a flat scintillator placed perpendicular to the beam and imaged by a CCD camera. As shown in Fig. 4, the accuracy of the proton range measured with the scintillator and wedge was 0.2 mm after a systematic shift of 0.51 mm was applied, which is

5 4413 Shen et al.: Fast proton range check by 2D scintillator 4413 FIG. 4. The accuracy of the range measured with the step wedge and scintillator. The cross, plus and circle symbols are the range accuracy measured from day 1 through day 3, after a constant 0.51 mm shift was applied. The star symbols show the range accuracy measured with a 0.36 mm WET film inserted under the wedge. On average, the ranges of star symbols were 0.35 mm (solid line) shorter than those of the cross, plus and circle symbols. [Color figure can be viewed at wileyonlinelibrary.com] FIG. 5. The reproducibility of the range measurements with the scintillator and step wedge, including the entire set of 15 repeated measurements for all 35 energies. The dashed lines mark one and two standard deviations (r = mm), respectively. [Color figure can be viewed at wileyonlinelibrary.com] comparable to the precision of water tank measurements. As the star symbols demonstrate in Fig. 4, the test easily detects a 0.36 mm range change with the average and the worst sensitives at 0.01 mm and 0.1 mm, respectively. While the system precision is comparable to that of water tank measurements, our novel method is much faster, requiring only 5 min to set up the system and 3 min of beam delivery time to measure 35 beam ranges. If the water tank was used instead, it

6 4414 Shen et al.: Fast proton range check by 2D scintillator 4414 would take about 1 h for setup and several hours to measure all of these proton ranges. The setup is simple, fast, and reproducible, requiring only the alignment of the wedge with the setup lasers. The highly reproducible results as shown in Fig. 5 indicate that this method is robust. This method can be translated to the other scintillator, such as Lynx (Ion Beam Applications, SA, Louvain-la-Neuve, Belgium). The 2D scintillator is used in our institute and other institutes for the quality assurance of the spot sizes and spot positions. 10,11 These spot size and position measurements are not included in this paper because they are beyond the scope of this paper. However, another distinct advantage with this novel method is that it allows the proton range, along with the spot size and spot positions, to be measured using a single device. This saves cost by reducing the number of measurement devices required for QA, and it saves time by removing the need to set up multiple detectors for different measurements. A 0.51 mm system shift has to be applied in order to get 0.2 mm accuracy as shown in Fig. 4. The source of the submillimeter systematic shift will be investigated in future experiments. A number of uncertainties may cause this submillimeter deviation, including uncertainties in the acrylic WET conversion factor, the physical thickness of the system, and/or the pixel resolution. The linear energy transfer dependence of the scintillator response, also known as the quenching effect (the scintillator material not having a linear response to radiation), may also play a role in this systematic shift. We selected the distal 80% of the peak intensity to define R wedge, which follows our institution s proton range definition in water (i.e., R 80 ). A previous study has also found the distal 80% to be appropriate for scintillator-based range measurements. 8 However, the quenching effect may change the shape of the distal Bragg peak, leading to differences in the distal 80% location between the scintillation light curve and the actual dose distribution in water. In order to have a more meaningful comparison of the ranges measured by the wedge and the water tank, quenching correction should be applied in the future. 14 The current wedge design can only be used to measure ranges from about 10 to 24 cm. Although it is possible to add buildup to extend this range to higher energies, the current design does not allow measurement of lower energy beams. A future work is to modify the current step wedge, so that ranges from 4 cm to 32.4 cm (as available in our proton scanning machine) can be measured. 5. CONCLUSION A novel method to measure proton range using a 2D scintillator detector and a step wedge was implemented and evaluated. This novel method is fast, robust, sensitive and highly reproducible. The range accuracy is within 0.2 mm, which is comparable to that of a scanning water tank, the gold standard. ACKNOWLEDGMENTS We thank Logos Systems engineer Mr. Brett Nelson for providing specific information on XRV4000. CONFLICT OF INTEREST The authors have no relevant conflicts of interest to disclose. a) Author to whom correspondence should be addressed. Electronic mail: shen.jiajian@mayo.edu. REFERENCES 1. Particle Therapy Co-Operative Group. Particle therapy facilities in operation: Information about technical equipment and patient statistics; pp. Retrieved from 2. Smith AR. Vision 20/20: proton therapy. Med Phys. 2009;36: Lomax AJ. Intensity modulated proton therapy and its sensitivity to treatment uncertainties 1: the potential effects of calculational uncertainties. Phys Med Biol. 2008;53: Paganetti H. Range uncertainties in proton therapy and the role of Monte Carlo simulations. Phys Med Biol. 2012;57:R99 R Dhanesar S, Sahoo N, Kerr M, et al. Quality assurance of proton beams using a multilayer ionization chamber system. Med Phys. 2013;40: Zhao L, Das IJ. Gafchromic EBT film dosimetry in proton beams. Phys Med Biol. 2010;55:N291 N Fukushima Y, Hamada M, Nishio T, Maruyama K. Development of an easy-to-handle range measurement tool using a plastic scintillator for proton beam therapy. Phys Med Biol. 2006;51: Archambault L, Poenisch F, Sahoo N, et al. Verification of proton range, position, and intensity in IMPT with a 3D liquid scintillator detector system. Med Phys. 2012;39: Hui C, Robertson D, Alsanea F, Beddar S. Fast range measurement of spot scanning proton beams using a volumetric liquid scintillator detector. Biomed Phys Eng Express. 2015;1: Lin LY, Ainsley CG, Mertens T, De Wilde O, Talla PT, McDonough JE. A novel technique for measuring the low-dose envelope of pencil-beam scanning spot profiles. Phys Med Biol. 2013;58:N171 N Lin L, Huang S, Kang M, Solberg TD, McDonough JE, Ainsley CG. Technical note: validation of halo modeling for proton pencil beam spot scanning using a quality assurance test pattern. Med Phys. 2015;42: Ulmer W, Schaffner B. Foundation of an analytical proton beamlet model for inclusion in a general proton dose calculation system. Radiat Phys Chem. 2011;80: Shen J, Liu W, Stoker J, et al. An efficient method to determine double Gaussian fluence parameters in the eclipse proton pencil beam model. Med Phys. 2016;43: Robertson D, Mirkovic D, Sahoo N, Beddar S. Quenching correction for volumetric scintillation dosimetry of proton beams. Phys Med Biol. 2013;58:

CyberKnife Iris Beam QA using Fluence Divergence

CyberKnife Iris Beam QA using Fluence Divergence CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery

More information

A Beam-Level Delivery Accuracy Study of the Robotic Image Guided Radiosurgery System Using a Scintillator/CCD Phantom

A Beam-Level Delivery Accuracy Study of the Robotic Image Guided Radiosurgery System Using a Scintillator/CCD Phantom A Beam-Level Delivery Accuracy Study of the Robotic Image Guided Radiosurgery System Using a Scintillator/CCD Phantom Lei Wang 1, Shi Liu 1, Brett Nelson 2 1. Department of Radiation Oncology, Stanford

More information

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Gena M.A.H 1, Ahmed L.El-Attar 2, Elbadry M. Zahran 3, Hany El-Gamal

More information

Evaluation of a diode array for QA measurements on a helical tomotherapy unit

Evaluation of a diode array for QA measurements on a helical tomotherapy unit Evaluation of a diode array for QA measurements on a helical tomotherapy unit K. M. Langen, a S. L. Meeks, D. O. Poole, T. H. Wagner, T. R. Willoughby, O. A. Zeidan, and P. A. Kupelian Department of Radiation

More information

Beam Delivery Techniques: Passive Scattering Proton Beams. Zuofeng Li and Roelf Slopsema University of Florida Proton Therapy Institute

Beam Delivery Techniques: Passive Scattering Proton Beams. Zuofeng Li and Roelf Slopsema University of Florida Proton Therapy Institute Beam Delivery Techniques: Passive Scattering Proton Beams Zuofeng Li and Roelf Slopsema University of Florida Proton Therapy Institute 1 Disclaimer UFPTI uses IBA Proton Therapy System Proton Beam: production,

More information

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Huaiqun Guan,

More information

SCINTILLATING FIBER DOSIMETER ARRAY

SCINTILLATING FIBER DOSIMETER ARRAY SCINTILLATING FIBER DOSIMETER ARRAY FIELD OF THE INVENTION [0001] This invention relates generally to the field of dosimetry and, more particularly, to rapid, high-resolution dosimeters for advanced treatment

More information

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 1, WINTER 2010 A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA Abdul Qadir Jangda,

More information

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array José A. Bencomo, * Geoffrey Ibbott, Seungsoo Lee, and Joao A. Borges Department of Radiation Physics.

More information

NOT FOR DISTRIBUTION JINST_083P_0914 v1

NOT FOR DISTRIBUTION JINST_083P_0914 v1 Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam F. Di Lillo a,b, D. Dreossi c, F. Emiro a,b, C. Fedon d,e, R. Longo d,e, G. Mettivier a,b,*, L. Rigon d,e, P. Russo

More information

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System. Use of EPIDs for Non-Routine Linac QA Bin Cai PhD Disclosure Parts of this project received support from Varian Medical System. Learning Objectives Learn the recent development of EPID based Non-routine

More information

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Indra J. Das, PhD, FACR Department of Radiation Oncology Indiana University of School of Medicine & Midwest Proton Radiation

More information

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images Aiping Ding, Bin Han, Lei Wang, Lei Xing Department of Radiation Oncology, Stanford University School of

More information

GEM-based Dose Imaging Detectors for Proton Therapy Quality Assurance and Proton Radiography

GEM-based Dose Imaging Detectors for Proton Therapy Quality Assurance and Proton Radiography GEM-based Dose Imaging Detectors for Proton Therapy Quality Assurance and Proton Radiography Authors: A.V. Klyachko Affiliation: Phenix Medical LLC, Bloomington, IN, USA. Correspondense address: Email:

More information

Gantry design and experience at PSI

Gantry design and experience at PSI Gantry design and experience at PSI Eros Pedroni for the R&D Technology Team Center for Proton Therapy Paul Scherrer Institute Villigen-PSI SWITZERLAND Workshop on Hadron Beam Therapy of Cancer Erice April

More information

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017 The Current State of EPID-Based Linear Accelerator Quality Assurance Timothy Ritter, PhD, DABR, FAAPM 1 Disclosures Employed by the Veterans Health Administration Faculty appointment with the University

More information

Radiographic film dosimetry of proton beams for depth-dose constancy check and beam profile measurement

Radiographic film dosimetry of proton beams for depth-dose constancy check and beam profile measurement JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Radiographic film dosimetry of proton beams for depth-dose constancy check and beam profile measurement Inhwan J. Yeo, a Anthony Teran,

More information

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion Leksell Gamma Knife Perfexion QA Considerations Paula L. Petti, Ph.D. Taylor McAdam Bell Neuroscience Institute Washington Hospital Healthcare System Fremont, CA 1 QA for LGK Perfexion : : Follow NRC licensing

More information

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM FILMQA STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE Supports all major radiotherapy technologies! FilmQA is optimized for use with Gafchromic film products, including EBT2 and RTQA2. FILMQA helps

More information

- Water resistant. - Large size.

- Water resistant. - Large size. GAFCHROMIC EBT product brochure GAFCHROMIC EBT FEATURES GAFCHROMIC EBT dosimetry film has been developed specifically to address the needs of the medical physicist and dosimetrist working in the radiotherapy

More information

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017 A Fast Monolithic System for Proton Imaging Fritz DeJongh ProtonVDA Inc October 2017 Disclosures I am a cofounder and co-owner of ProtonVDA Inc We hold intellectual property rights on our proton imaging

More information

IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film

IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film CYRIC Annual Report 2005 IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film Ohuchi H. 1, and Abe K. 2 1 Graduate School of Pharmaceutical Sciences,

More information

Prompt-Gamma Based Range Verification in Particle Therapy: New prospects (also) for 4D?

Prompt-Gamma Based Range Verification in Particle Therapy: New prospects (also) for 4D? Prompt-Gamma Based Range Verification in Particle Therapy: New prospects (also) for 4D? Guntram Pausch 1-3 1 OncoRay National Center for Radiation Research in Oncology, Dresden 2 Technische Universität

More information

Test of a Multilayer Dose-Verification Gaseous Detector with Raster Scan Mode Proton Beams

Test of a Multilayer Dose-Verification Gaseous Detector with Raster Scan Mode Proton Beams Test of a Multilayer Dose-Verification Gaseous Detector with Raster Scan Mode Proton Beams Kyong Sei Lee*, Byungsik Hong, Minho Kang, Seunkyung Lee, and Sung Keun Park Korea University, Seoul 136-713 Sung

More information

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

3D Diode Array Commissioning: Building Confidence in 3D QA Technology 3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose

More information

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators Implementing a new digital medical accelerator John Wong Johns Hopkins University AAPM, Austin, 2014 Acknowledgements Yin Zhang, Ken Wang, Kai Ding (Commissioning - JHU) Esteban Velarde, Joe Moore (QA

More information

GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom

GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom Introduction: These EBT2/EBT3 films, shown in Figure 1a-c, are specially sized and formatted to uniquely fit the Accuray Ball Cube II Phantom. Each

More information

Monte Carlo study on a new concept of a scanning photon beam system for IMRT

Monte Carlo study on a new concept of a scanning photon beam system for IMRT NUKLEONIKA 2011;56(4):291 297 ORIGINAL PAPER Monte Carlo study on a new concept of a scanning photon beam system for IMRT Anna M. Wysocka-Rabin, Günter H. Hartmann Abstract. Intensity-modulated radiation

More information

Quality assurance in proton beam therapy using a plastic scintillator and a commercially available digital camera

Quality assurance in proton beam therapy using a plastic scintillator and a commercially available digital camera Received: 12 January 2017 Revised: 14 May 2017 Accepted: 14 June 2017 DOI: 10.1002/acm2.12143 RADIATION ONCOLOGY PHYSICS Quality assurance in proton beam therapy using a plastic scintillator and a commercially

More information

The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response

The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response Iran. J. Radiat. Res., 2005; 3 (1): 3-10 The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response M. Mohammadi 1,2,3* and E. Bezak 1,2

More information

Beam Commissioning and Annual QA Phantoms, Detectors & Accessories. iba-dosimetry.com

Beam Commissioning and Annual QA Phantoms, Detectors & Accessories. iba-dosimetry.com Beam Commissioning and Annual QA Phantoms, Detectors & Accessories iba-dosimetry.com Benefit from efficient and smartly designed beam commissioning & annual QA tools that make your life EASIEST. For MORE

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests Acknowledgements David Shepard, Ph.D. Daliang Cao, Ph.D. Muhammad K. N. Afghan, Ph.D. Jinsong Ye, M.S. Tony P. Wong, Ph.D. Fan Chen, Ph.D. Min Rao, Ph.D. Vivek Mehta, M.D. Igor Gomola, Ph.D. David Housley

More information

SUN NUCLEAR. EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm. corporation. Your Most Valuable QA and Dosimetry Tools

SUN NUCLEAR. EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm. corporation. Your Most Valuable QA and Dosimetry Tools EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm SUN NUCLEAR corporation Your Most Valuable QA and Dosimetry Tools introduction Pre-treatment dose QA is an important process required

More information

Attenuation length in strip scintillators. Jonathan Button, William McGrew, Y.-W. Lui, D. H. Youngblood

Attenuation length in strip scintillators. Jonathan Button, William McGrew, Y.-W. Lui, D. H. Youngblood Attenuation length in strip scintillators Jonathan Button, William McGrew, Y.-W. Lui, D. H. Youngblood I. Introduction The ΔE-ΔE-E decay detector as described in [1] is composed of thin strip scintillators,

More information

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING 1 QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING QualiFormeD Phantoms A selection of test objects facilitating regulatory quality controls in radiation therapy and medical imaging Practical,

More information

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc.

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. An Introduction to TG-142 Imaging QA Using Standard Imaging Products Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. Goals Understand the nature and intent of TG 142 imaging

More information

Comparison of measured Varian Clinac 21EX and TrueBeam accelerator electron field characteristics

Comparison of measured Varian Clinac 21EX and TrueBeam accelerator electron field characteristics JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Comparison of measured Varian Clinac 21EX and TrueBeam accelerator electron field characteristics Samantha A.M. Lloyd, 1a Sergei Zavgorodni,

More information

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 David F. Lewis Advanced Materials Group Ashland Specialty Ingredients Spain, April 2014 What is Radiochromic Film? A film that instantly changes

More information

Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam

Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam Home Search Collections Journals About Contact us My IOPscience Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam This content has been downloaded from IOPscience.

More information

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications I. SCOPE The protocol applies to GafChromic EBT3P and EBT3+P films exposed in GafChromic QuiCk Phantom

More information

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev. 1.0 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE DoseLab users may reference the following instructions to perform Tomotherapy Quality Assurance tests as recommended

More information

I. Introduction.

I. Introduction. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case

More information

Detectors for small field dosimetry

Detectors for small field dosimetry Detectors or small ield dosimetry Hugo Palmans MedAustron, Wiener Neustadt, Austria and National Physical Laboratory, Teddington, UK 1 080915 Overview Ideal detector Water calorimeter Ionization chamber

More information

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter MR-guided RT in Heidelberg Funded by the German Research Foundation (DFG) Heidelberg consortium received

More information

A proposed method for linear accelerator photon beam steering using EPID

A proposed method for linear accelerator photon beam steering using EPID Received: 13 January 2018 Revised: 11 May 2018 Accepted: 29 June 2018 DOI: 10.1002/acm2.12419 RADIATION ONCOLOGY PHYSICS A proposed method for linear accelerator photon beam steering using EPID Michael

More information

Beam Profiling. Introduction. What is Beam Profiling? by Michael Scaggs. Haas Laser Technologies, Inc.

Beam Profiling. Introduction. What is Beam Profiling? by Michael Scaggs. Haas Laser Technologies, Inc. Beam Profiling by Michael Scaggs Haas Laser Technologies, Inc. Introduction Lasers are ubiquitous in industry today. Carbon Dioxide, Nd:YAG, Excimer and Fiber lasers are used in many industries and a myriad

More information

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Takeshi Fujiwara 1, Yuki Mitsuya 2, Hiroyuki Takahashi 2, and Hiroyuki Toyokawa 2 1 National Institute of Advanced Industrial

More information

Imaging Cerenkov emission as a quality assurance tool in electron radiotherapy

Imaging Cerenkov emission as a quality assurance tool in electron radiotherapy TB, VB, UK, PMB/484054, 11/03/2014 Phys. Med. Biol. 59 (2014) 1 16 Institute of Physics and Engineering in Medicine Physics in Medicine and Biology UNCORRECTED PROOF Imaging Cerenkov emission as a quality

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film

Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 4, 2016 Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film Eric C. Lobb Department

More information

Peace of Mind. Automated.

Peace of Mind. Automated. 1 Peace of Mind. Automated. Automated and guided beam commissioning Why SMARTSCAN? Get thousands of beam scans done effortlessly. Save your valuable time and manual operations at the water phantom and

More information

Calibration of KAP meters

Calibration of KAP meters Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area

More information

Nathan Childress, Ph.D., DABR

Nathan Childress, Ph.D., DABR Nathan Childress, Ph.D., DABR Introduction TG-142 is a comprehensive QA protocol Covers nearly every aspect of machine and safety QA Recommends quantitative results Recommends high testing frequencies

More information

Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy

Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy Thomas LoSasso, a) Chen-Shou Chui, and C. Clifton Ling Department

More information

Monica Kishore. Medical Physics Graduate Program Duke University. Approved: Jennifer O Daniel, Co-Supervisor. Fang-Fang Yin, Co-Supervisor

Monica Kishore. Medical Physics Graduate Program Duke University. Approved: Jennifer O Daniel, Co-Supervisor. Fang-Fang Yin, Co-Supervisor Accuracy of Planar Dosimetry for Volumetric Modulated Arc Therapy Quality Assurance by Monica Kishore Medical Physics Graduate Program Duke University Date: Approved: Jennifer O Daniel, Co-Supervisor Fang-Fang

More information

Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland

Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland Pedroni Eros Paul Scherrer Institute - Villigen PSI, Switzerland Center for proton therapy - PSI GANTRY DESIGN AND RECENT EXPERIENCE AT PSI 2nd Workshop on Hadron Beam Therapy of Cancer ERICE May 23 th,

More information

A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering

A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering Sandra E. Burch Department of Radiology, Medical College of Georgia, Augusta, Georgia 30912 Kimberlee J. Kearfott

More information

Analysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose

Analysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose J. Radiat. Res., 53, 301 305 (2012) Analysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose Katsumi SHIMA 1,2, Kunihiko TATEOKA 1 *, Yuichi SAITOH 1,2, Junji SUZUKI

More information

Laser Beam Analysis Using Image Processing

Laser Beam Analysis Using Image Processing Journal of Computer Science 2 (): 09-3, 2006 ISSN 549-3636 Science Publications, 2006 Laser Beam Analysis Using Image Processing Yas A. Alsultanny Computer Science Department, Amman Arab University for

More information

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS 7:28-14.1 Scope (a) This subchapter covers therapeutic installations used in the healing arts. These therapeutic installations include x-ray, accelerator and teletherapy

More information

Diamond X-ray Rocking Curve and Topograph Measurements at CHESS

Diamond X-ray Rocking Curve and Topograph Measurements at CHESS Diamond X-ray Rocking Curve and Topograph Measurements at CHESS G. Yang 1, R.T. Jones 2, F. Klein 3 1 Department of Physics and Astronomy, University of Glasgow, Glasgow, UK G12 8QQ. 2 University of Connecticut

More information

Product Range Electronic Units

Product Range Electronic Units Pyramid Technical Consultants, Inc. 1050 Waltham Street Suite 200 Lexington, MA 02421 TEL: +1 781 402-1700 TEL (UK): +44 1273 492001 FAX: (781) 402-1750 EMAIL: SUPPORT@PTCUSA.COM Product Range Electronic

More information

Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA

Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Arthur Olch, PhD, FAAPM AAPM Spring Clinical Meeting, March 21, 2017 Or.. What Dose are the Patients Really Getting???

More information

COMPREHENSIVE TG-142 IMAGING AND MACHINE QA

COMPREHENSIVE TG-142 IMAGING AND MACHINE QA QA SOFTWARE COMPREHENSIVE TG-142 IMAGING AND MACHINE QA Automate the analysis of over thirty TG-142 recommended QA tasks The rapid progress of Radiation Therapy has created the need for Quality Assurance

More information

Initial Results from a Cryogenic Proton Irradiation of a p-channel CCD

Initial Results from a Cryogenic Proton Irradiation of a p-channel CCD Centre for Electronic Imaging Initial Results from a Cryogenic Proton Irradiation of a p-channel CCD Jason Gow Daniel Wood, David Hall, Ben Dryer, Simeon Barber, Andrew Holland and Neil Murray Jason P.

More information

A Laser-Based Thin-Film Growth Monitor

A Laser-Based Thin-Film Growth Monitor TECHNOLOGY by Charles Taylor, Darryl Barlett, Eric Chason, and Jerry Floro A Laser-Based Thin-Film Growth Monitor The Multi-beam Optical Sensor (MOS) was developed jointly by k-space Associates (Ann Arbor,

More information

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance

More information

State of the Art Film Dosimetry

State of the Art Film Dosimetry State of the Art Film Dosimetry Micke A., Lewis D. Advanced Materials Ashland proprietary technology, patents pending Film Dosimetry Radiochromic Film EBT2/EBT3 One-Scan Protocol Multi-channel Film Dosimetry

More information

On spatial resolution

On spatial resolution On spatial resolution Introduction How is spatial resolution defined? There are two main approaches in defining local spatial resolution. One method follows distinction criteria of pointlike objects (i.e.

More information

A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine

A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine Iwein Van de Vondel, 1 Koen Tournel, 1 Dirk

More information

Application of the HICAM Camera for Imaging of Prompt Gamma Rays in Measurements of Proton Beam Range

Application of the HICAM Camera for Imaging of Prompt Gamma Rays in Measurements of Proton Beam Range Application of the HICAM Camera for Imaging of Prompt Gamma Rays in Measurements of Proton Beam Range R. Peloso 1, P. Busca 1, A.Celani 2, C. Fiorini 1, I.Perali 1, M. Basilavecchia 2, T. Frizzi 2, J.

More information

GAFCHROMIC. Therapy Dosimetry Media Models to

GAFCHROMIC. Therapy Dosimetry Media Models to GAFCHROMIC Therapy Dosimetry Media Models 37-040 to 37-045! Superior uniformity and sensitivity! Dose rate and fractionation independent! Maps dose distribution! Provides quantitative measurements (via

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Post-irradiation colouration of Gafchromic EBT radiochromic film

Post-irradiation colouration of Gafchromic EBT radiochromic film INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 5 (25) N281 N285 PHYSICS IN MEDICINE AND BIOLOGY doi:1.188/31-9155/5/2/n4 NOTE Post-irradiation colouration of Gafchromic EBT radiochromic film Tsang Cheung

More information

AN INITIAL investigation into the effects of proton irradiation

AN INITIAL investigation into the effects of proton irradiation IEEE TRANSACTIONS ON ELECTRON DEVICES, VOL. 53, NO. 2, FEBRUARY 2006 205 Proton Irradiation of EMCCDs David R. Smith, Richard Ingley, and Andrew D. Holland Abstract This paper describes the irradiation

More information

Initial setup and subsequent temporal position monitoring using implanted RF transponders

Initial setup and subsequent temporal position monitoring using implanted RF transponders Initial setup and subsequent temporal position monitoring using implanted RF transponders James Balter, Ph.D. University of Michigan Has financial interest in Calypso Medical Technologies Acknowledgements

More information

Homogeneity of GAFCHROMIC EBT2 film among different lot numbers

Homogeneity of GAFCHROMIC EBT2 film among different lot numbers JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 4, 2012 Homogeneity of GAFCHROMIC EBT2 film among different lot numbers Hirokazu Mizuno, 1,2a Yutaka Takahashi, 3 Atsushi Tanaka, 1 Takamitsu

More information

ECEN 4606, UNDERGRADUATE OPTICS LAB

ECEN 4606, UNDERGRADUATE OPTICS LAB ECEN 4606, UNDERGRADUATE OPTICS LAB Lab 2: Imaging 1 the Telescope Original Version: Prof. McLeod SUMMARY: In this lab you will become familiar with the use of one or more lenses to create images of distant

More information

ArcCHECKTM. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools. VMAT RapidArc TomoTherapy Pinnacle 3 SmartArc Conventional IMRT

ArcCHECKTM. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools. VMAT RapidArc TomoTherapy Pinnacle 3 SmartArc Conventional IMRT TM The Ultimate 4D QA Solution A 4D isotropical cylindrical detector array for arc delivery QA and Dosimetry U.S.Patent No. 8,044,359 What is? The world s first true 4D detector array The world s first

More information

Bringing Answers to the Surface

Bringing Answers to the Surface 3D Bringing Answers to the Surface 1 Expanding the Boundaries of Laser Microscopy Measurements and images you can count on. Every time. LEXT OLS4100 Widely used in quality control, research, and development

More information

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

CALICE AHCAL overview

CALICE AHCAL overview International Workshop on the High Energy Circular Electron-Positron Collider in 2018 CALICE AHCAL overview Yong Liu (IHEP), on behalf of the CALICE collaboration Nov. 13, 2018 CALICE-AHCAL Progress, CEPC

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Accuracy of rapid radiographic film calibration for intensity-modulated radiation therapy verification

Accuracy of rapid radiographic film calibration for intensity-modulated radiation therapy verification JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 2, SPRING 2006 Accuracy of rapid radiographic film calibration for intensity-modulated radiation therapy verification Ravi Kulasekere, a Jean

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System. Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX.

Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System. Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX. Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX. 1 Acknowledgements Collaborators: Jose Bencomo, Rafael. M. Landrove, Peter

More information

Phase Imaging Using Focused Polycapillary Optics

Phase Imaging Using Focused Polycapillary Optics Phase Imaging Using Focused Polycapillary Optics Sajid Bashir, Sajjad Tahir, Jonathan C. Petruccelli, C.A. MacDonald Dept. of Physics, University at Albany, Albany, New York Abstract Contrast in conventional

More information

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools

More information

Stability of the Helical TomoTherapy Hi Art II detector for treatment beam irradiations

Stability of the Helical TomoTherapy Hi Art II detector for treatment beam irradiations JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 6, 2014 Stability of the Helical TomoTherapy Hi Art II detector for treatment beam irradiations Karin Schombourg, François Bochud, Raphaël

More information

PHYSICS QUESTIONNAIRE FORM

PHYSICS QUESTIONNAIRE FORM PHYSICS QUESTIONNAIRE FORM Institution Name: Date: Contact Information (name, address, phone, fax, email): Physicist: Radiation Oncologist: Dosimetrist (if applicable): Study Coordinator (if applicable):

More information

A Multichannel Dosimeter Based on Scintillating Fibers for Medical Applications

A Multichannel Dosimeter Based on Scintillating Fibers for Medical Applications A Multichannel Dosimeter Based on Scintillating Fibers for Medical Applications K.-H. Becks 1, J. Drees 1, K. Goldmann 1, I.M. Gregor 1 2, M. Heintz, Arndt Roeser* 1 Fachbereich Physik, Bergische Universität-Gesamthochschule

More information

Aim. Images for this section: Page 2 of 13

Aim. Images for this section: Page 2 of 13 Changes in CT number of high atomic number materials with field of view when using an extended CT number to electron density curve and a metal artifact reduction reconstruction algorithm Poster No.: R-0094

More information

Determining MTF with a Slant Edge Target ABSTRACT AND INTRODUCTION

Determining MTF with a Slant Edge Target ABSTRACT AND INTRODUCTION Determining MTF with a Slant Edge Target Douglas A. Kerr Issue 2 October 13, 2010 ABSTRACT AND INTRODUCTION The modulation transfer function (MTF) of a photographic lens tells us how effectively the lens

More information

Comparative performance evaluation of a new a-si EPID that exceeds quad high-definition resolution

Comparative performance evaluation of a new a-si EPID that exceeds quad high-definition resolution JBUON 2018; 23(2): 507-513 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com ORIGINAL ARTICLE Comparative performance evaluation of a new a-si EPID that exceeds quad

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 5, 2015

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 5, 2015 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 5, 15 Evaluation of monitor unit calculation based on measurement and calculation with a simplified Monte Carlo method for passive beam delivery

More information

BRANDON RICE UNIVERSITY OF FLORIDA

BRANDON RICE UNIVERSITY OF FLORIDA METHODS FOR PRODUCING OFF-AXIS RATIO TABLES FROM MINI-MULTILEAF COLLIMATOR SHAPED CIRCULAR FIELDS FOR INPUT INTO A STEREOTACTIC RADIOSURGERY TREATMENT PLANNING SYSTEM By BRANDON RICE A THESIS PRESENTED

More information

E LECTROOPTICAL(EO)modulatorsarekeydevicesinoptical

E LECTROOPTICAL(EO)modulatorsarekeydevicesinoptical 286 JOURNAL OF LIGHTWAVE TECHNOLOGY, VOL. 26, NO. 2, JANUARY 15, 2008 Design and Fabrication of Sidewalls-Extended Electrode Configuration for Ridged Lithium Niobate Electrooptical Modulator Yi-Kuei Wu,

More information

arxiv:physics/ v1 [physics.optics] 12 May 2006

arxiv:physics/ v1 [physics.optics] 12 May 2006 Quantitative and Qualitative Study of Gaussian Beam Visualization Techniques J. Magnes, D. Odera, J. Hartke, M. Fountain, L. Florence, and V. Davis Department of Physics, U.S. Military Academy, West Point,

More information