SPECT Reconstruction & Filtering

Size: px
Start display at page:

Download "SPECT Reconstruction & Filtering"

Transcription

1 SPECT Reconstruction & Filtering

2 Goals Understand the basics of SPECT Reconstruction Filtered Backprojection Iterative Reconstruction Make informed choices on filter selection and settings Pre vs. Post Filtering Filter Parameters Resolution Recovery 2 /

3 Acquisition Major components of a tomographic acquisition Matrix : 64 vs 128 Acquisition Zoom Number of projections Time per view, including half-time and half-dose What do these factors have to do with filters? Matrix & zoom determine pixel size. Cardiac studies need to be 6 8 mm/slice for most quantification packages Projections and time per view affect total counts 3 /

4 Acquisition With a 1.33 zoom: 128 matrix = 3.4mm/pixel 64 matrix = 6.8mm/pixel 4 /

5 Reconstruction - Simple Backprojection Mathematical method to reconstruct tomographic image Reconstruction of an Image of a point source Each projection sees the point source as a strip or ray-sum Simple backprojection plots the counts per pixel across each ray-sum 5 /

6 Reconstruction - Simple Backprojection Simple backprojection reconstructs an image by spreading each raysum evenly among the pixels along the projection line This simple method results in streaking or star artifacts 6 /

7 Reconstruction - Filtered Backprojection Filtered Backprojection modifies the original projection profiles with a filter function before projecting the ray sums into the image plane The streak and star artifacts are further reduced or cancelled when more projections are acquired. A pre-filter is used along with the RAMP filter to produce the final image 7 /

8 Image Components To choose the optimal filter, need to understand the three major components of any image: Background Generated from the backprojection RAMP filter eliminates most of this Image Data Composed of a series of frequency waves - high, medium, low Frequency depends on type of study Noise Unwanted high frequency contribution to an image due to statistical fluctuation Inversely proportional to the number of counts in the raw data Pre-filter used to remove noise 8 /

9 Frequency of Image Data High Frequency Large difference in pixel to pixel counts Images with sharp edges May need less filtering Low Frequency Less detail means lower frequency May need more filtering 9 /

10 High Frequency Noise Low Count Study 25 counts per pixel 20 percent noise High Count Study 100 counts per pixel 10 percent noise Lots High Frequency Noise Less Noise 10 /

11 Filtering to Optimize Image Data RAMP Filter Removes background Pre-Filter Removes Noise RAMP Filter Pre-Filter Remaining Image Data 11 /

12 RAMP Filter Only RAMP Filter Removes background Very noisy images 12 /

13 Filtering to Optimize Image Data Heavy Filter Light Filter High Count Data Lighter Filter Keeps higher frequency components More noise included Sharper image More image data included Low Count Data Heavier Filter More high frequency removed Less noise included Smoother image More image data removed 13 /

14 Filtering to Optimize Image Data Trade-Off Light Filter Smoothing versus Resolution Heavy Filter What constitutes a light or heavy filter depends on counts in image 14 /

15 Cutoff Frequency Defines the point at which the filter is defined to be zero Units may be cycles/pixel or cycles/cm Vendors may define filters differently Lower number = smoother image 15 /

16 Hann Filter Examples Cutoff Frequency: 0.5 cycles/cm Cutoff Frequency: 0.75 cycles/cm Cutoff Frequency: 1.00 cycles/cm 16 /

17 Butterworth Filter Critical Frequency Defines the point at which the filters starts to roll-off towards zero Units may vary by Vendor A lower number = smoother image (heavier filter) Order or Power Factor Determines the slope of the filter curve A higher number = steeper slope A higher number = smoother image More subtle of an effect than critical frequency 17 /

18 Butterworth Filter Examples Critical Frequency Critical Frequency: 0.25 cycles/cm Critical Frequency: 0.50 cycles/cm Critical Frequency: 0.75 cycles/cm 18 /

19 Butterworth Filter Examples Power Factor Critical Frequency: 0.40 cycles/cm Power Factor: 5 Critical Frequency: 0.40 cycles/cm Power Factor: 10 Critical Frequency: 0.40 cycles/cm Power Factor: /

20 Iterative Reconstruction An alternative to filtered backprojection Can accommodate corrections for attenuation, scatter, and noise and used for half-time/half-dose imaging Can produce higher image quality than FBP in terms of less noise, better resolution, fewer artifacts, and less distortion Computationally intense requires more time to complete FBP IR 20 /

21 Iterative Reconstruction vs FBP FBP IR 21 /

22 Iterative Reconstruction Initial estimate of transaxial distribution Creation of projection profiles Comparison of computed profiles to acquired patient profiles Error image used to update image estimate Repeat per iteration 22 /

23 Types of Iterative Reconstruction MLEM maximum likelihood expectationmaximization Original method used Computationally intense OSEM ordered subset expectation-maximization Projection data is grouped in ordered subsets o Generally groups of projections MLEM is then applied to each of the subsets Faster processing 23 /

24 Iterative Reconstruction Uses 3D Post-Filter to smooth images and reduce noise If enough counts, may not need post-filter Iterative Reconstruction 2 iterations, 10 subsets No Post-Filter Butterworth Filter Critical Frequency 0.50, Power Factor /

25 Iterative Reconstruction Number of Iterations More does not always mean better Profiles may diverge as iterations increase Subjective More iterations = increased noise 25 /

26 IR Number of Iterations (Same Post Filter) 2 iterations 10 subsets 6 iterations 10 subsets 4 iterations 10 subsets 8 iterations 10 subsets 26 /

27 Iterative Reconstruction Number of Subsets As with iterations, more does not always mean better PET lesion detection has been shown to decrease in some studies as subsets increase over More subsets = increased noise 27 /

28 IR Number of Subsets (Same Post Filter) 4 iterations 5 subsets 4 iterations 20 subsets 4 iterations 10 subsets 4 iterations 30 subsets 28 /

29 Beyond Iterative Reconstruction 29 /

30 Beyond Iterative Reconstruction Evolution Resolution Recovery Models the collimator-detector response in order to correct for noise and resolution degradation Corrects the blurring effect by including an accurate model of collimatordetector response function in an iterative SPECT reconstruction algorithm Vendor specific algorithm Utilizes intrinsic system response and collimator specific geometric response for each combination of acquisition system, radiopharmaceutical, and acquisition protocol 30 /

31 Evolution Half-Time Imaging 31 /

32 Evolution Comparison Filtered Back Projection OSEM Iterative Reconstruction Evolution Images courtesy of Baylor University Medical Center in Dallas 32 /

33 Filter Considerations Optimal Filter will depend upon: Study Type Organ studied will determine frequency of image data (e.g. Bone vs. In-111 Octreoscan) Radiopharmaceutical used will determine total counts and noise component (Tc-99m Sestamibi vs. I-123 mibg) Patient Body Habitus Patient size will effect counts and noise component Imaging parameters Time per projection Matrix size Equipment Single head vs. dual head Filter description (order/power and units for cutoff frequency are not uniform among vendors) Reconstruction Type Filtered backprojection, iterative reconstruction or resolution recovery Physician preference 33 /

34 Filter Conclusion There is no perfect filter!!! Filter is a preference Many quantitative programs (i.e. polar maps, gated SPECT ejection fraction) require a specific filter (and imaging protocol) for proper analysis 34 /

35 Resources Computers in Nuclear Medicine: A Practical Approach, 2 nd Edition, Kai Lee, PhD, Society of Nuclear Medicine, SPECT: A Primer, 3rd Edition, Robert English, CNMT, Society of Nuclear Medicine, Effective Use of Computers in Nuclear Medicine, Michael Gelfand, McGraw- Hill, Characterization of ordered-subsets expectation maximization with 3D postreconstruction Gauss filtering and comparison with filtered backprojection in 99mTc SPECT, Marco Brambilla et al, Annals of Nuclear Medicine Vol. 19, No. 2, 75 82, 2005 Effect of varying number of OSEM subsets on PET lesion detectability, Morey AM1, Kadrmas DJ., J Nucl Med Technol Dec;41(4): Accelerated Image Reconstruction using Ordered Subsets of Projection Data, H. Malcolm Hudson, Richard S. Larkin, IEEE Transactions on Medical Imaging, Vol. XX, No. Y, Month /

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT) Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in

More information

... In vivo imaging in Nuclear Medicine. 1957: Anger camera (X;Y) X Y

... In vivo imaging in Nuclear Medicine. 1957: Anger camera (X;Y) X Y József Varga, PhD EMISSION IMAGING BASICS OF QUANTIFICATION Imaging devices Aims of image processing Reconstruction University of Debrecen Department of Nuclear Medicine. In vivo imaging in Nuclear Medicine

More information

Reconstruction Filtering in Industrial gamma-ray CT Application

Reconstruction Filtering in Industrial gamma-ray CT Application Reconstruction Filtering in Industrial gamma-ray CT Application Lakshminarayana Yenumula *, Rajesh V Acharya, Umesh Kumar, and Ashutosh Dash Industrial Tomography and Instrumentation Section, Isotope Production

More information

The image reconstruction influence in relative measurement in SPECT / CT animal

The image reconstruction influence in relative measurement in SPECT / CT animal BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 0-01 (201) 01-09 The image reconstruction influence in relative measurement in SPECT / CT animal S.C.S. Soriano a ; S.A.L. Souza b ; T.Barboza b ; L.V. De Sá

More information

Continuing Education. Filtering in Frequency Space. THE FREQUENCY DOMAIN Frequency Space

Continuing Education. Filtering in Frequency Space. THE FREQUENCY DOMAIN Frequency Space Continuing Education Filtering in Frequency Space James R. Galt, H. Lee Hise, Ernest V. Garcia, and David J. Nowakt Emory University School of Medicine, Atlanta, Georgia; and tgeneral Electric Company

More information

Initial Certification

Initial Certification Initial Certification Nuclear Medical Physics (NMP) Study Guide Part 2 Content Guide and Sample Questions The content of all ABR exams is determined by a panel of experts who select the items based on

More information

Factors Affecting the resolution of SPECT Imaging. h.

Factors Affecting the resolution of SPECT Imaging. h. Factors Affecting the resolution of SPECT Imaging H. E. Mostafa *1, H. A. Ayoub 2 and Sh.Magraby 1 1 Kasr El-Ini Center for Oncology, Cairo University, 2 Faculty of Science, Suez Canal University hayamayoub@yahoo.com

More information

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance

More information

Industry Breakthrough

Industry Breakthrough Industry Breakthrough Dynamic SPECT Acquisition Quantifying Myocardial Blood Flow D-S P EC T Cardiac Imaging System Nuclear Cardiology in the 21st Century In the 21st century, most nuclear cameras are

More information

Industry Breakthrough

Industry Breakthrough Industry Breakthrough Dynamic SPECT Acquisition Quantifying Myocardial Blood Flow Nuclear Cardiology in the 21st Century In the 21st century, most nuclear cameras are still relying on a technology invented

More information

Hideo ONISHI * 1 * 3 Yuki MATSUTAKE * 2 Norikazu MATSUTOMO * 3 Hizuru AMIJIMA * 4. Abstract

Hideo ONISHI * 1 * 3 Yuki MATSUTAKE * 2 Norikazu MATSUTOMO * 3 Hizuru AMIJIMA * 4. Abstract Validation of optimal cut-off frequency using a Butterworth filter in single photon emission computed tomography reconstruction for the target organ: Spatial domain and frequency domain Hideo ONISHI *

More information

Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner

Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner Kisung Lee, Member, IEEE, Paul E. Kinahan, Senior Member, Robert S. Miyaoka, Member, IEEE, Jeffrey A. Fessler,

More information

Investigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras

Investigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras Egyptian J. Nucl. Med., Vol 2, No. 2, Dec. 2009 82 PHYSICS, Original Artical Investigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras Abdelsattar, M.B. Ph.D.; BuHumaid,

More information

J of Nuclear Medicine Technology, first published online July 27, 2011 as doi: /jnmt

J of Nuclear Medicine Technology, first published online July 27, 2011 as doi: /jnmt J of Nuclear Medicine Technology, first published online July 27, 2011 as doi:10.2967/jnmt.110.084814 Extrinsic Versus Intrinsic Uniformity Correction for g-cameras Randy Bolstad 1, Jody Brown, RT(N),

More information

Philips Astonish. Key advantages. including improved image quality and

Philips Astonish. Key advantages. including improved image quality and Philips Key advantages including improved image quality and with AC offers improved image quality, interpretative certainty, diagnostic Alternatively, simplify patient care by exposing patients to reduced

More information

An Investigation of Filter Choice for Filtered Back-Projection Reconstruction in PET

An Investigation of Filter Choice for Filtered Back-Projection Reconstruction in PET An nvestigation of Filter Choice for Filtered BackProjection Reconstruction in PET T. H. Farauhar, A. Chatziioannou, G. Chinn, M. Dahlbom, and E. J. Hoffman Division of Nuclear Medicine & Biophysics, Department

More information

NM Module Section 2 6 th Edition Christian, Ch. 3

NM Module Section 2 6 th Edition Christian, Ch. 3 NM 4303 Module Section 2 6 th Edition Christian, Ch. 3 Gas Filled Chamber Voltage Gas filled chamber uses Hand held detectors cutie pie Geiger counter Dose calibrators Cutie pie Chamber voltage in Ionization

More information

New Technology in Nuclear Medicine

New Technology in Nuclear Medicine New Technology in Nuclear Medicine Reed G. Selwyn, PhD, DABR Vice Chair of Research & Imaging Sciences Associate Professor and Chief, Medical Physics Dept. of Radiology, University of New Mexico Objectives

More information

CHAPTER 8 GENERIC PERFORMANCE MEASURES

CHAPTER 8 GENERIC PERFORMANCE MEASURES GENERIC PERFORMANCE MEASURES M.E. DAUBE-WITHERSPOON Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, United States of America 8.1. INTRINSIC AND EXTRINSIC MEASURES 8.1.1.

More information

Iterative Reconstruction

Iterative Reconstruction RECENT ADVANCES IN CT RADIATION DOSE REDUCTION TECHNIQUES Iterative Reconstruction Kalpana Kanal, PhD, FSCBTMR, FACR, FAAPM Professor and Director, Diagnostic Physics Section University of Washington Seattle,

More information

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard.

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard. Chiara Secco PET Performance measurements of the new LSO-Based Whole Body PET/CT Scanner biograph 16 HI-REZ using the NEMA NU 2-2001 Standard. INTRODUCTION Since its introduction, CT has become a fundamental

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

The Point Source. July 2011 Volume 2, Issue 1. Executive message. Rich Fabian. In this issue

The Point Source. July 2011 Volume 2, Issue 1. Executive message. Rich Fabian. In this issue The Point Source July 2011 Volume 2, Issue 1 In this issue Executive message from Rich Fabian...1 Handling anxiety...2 Product updates Introducing TruFlight Select PET/CT...3 BrightView XCT Version 2.5...4

More information

Changing the Shape of Nuclear Medicine

Changing the Shape of Nuclear Medicine TRUTH IN IMAGING Changing the Shape of Nuclear Medicine Multi-Purpose SPECT Scanner Nothing Gets Closer Introducing 360 Body Contour Scanning With 360 degree detector coverage, and unique proximity sensors

More information

Contrast Evaluation for a Dual Head SPECT System with Different Energy Peaking- Drift

Contrast Evaluation for a Dual Head SPECT System with Different Energy Peaking- Drift PHYSICS, ORIGINAL ARTICLE Contrast Evaluation for a Dual Head SPECT System with Different Energy Peaking- Drift Saad, I. *, Mattar, E. **, Ashour, A. *** * Department of Clinical oncology and Nuclear Medicine,

More information

Performance Characteristics of a State of the Art Preclinical PET/SPECT/CT Scanner

Performance Characteristics of a State of the Art Preclinical PET/SPECT/CT Scanner Performance Characteristics of a State of the Art Preclinical PET/SPECT/CT Scanner Nya Mehnwolo Boayue 1 Samuel Kuttner 1 1 Center for Diagnostic Physics University Hospital of North-Norway Medfys, 2016

More information

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

Imaging with FDG PET is a valuable technique for tumor

Imaging with FDG PET is a valuable technique for tumor Noise Reduction in Oncology FDG PET Images by Iterative Reconstruction: A Quantitative Assessment Cyril Riddell, Richard E. Carson, Jorge A. Carrasquillo, Steven K. Libutti, David N. Danforth, Millie Whatley,

More information

LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option

LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option Y. Bercier, Member, IEEE, M. Casey, Member, IEEE, J. Young, Member, IEEE, T. Wheelock, Member, IEEE, T. Gremillion Abstract-- Factors which

More information

A Skew-Slit Collimator for Small-Animal SPECT

A Skew-Slit Collimator for Small-Animal SPECT A Skew-Slit Collimator for Small-Animal SPECT Gengsheng L. Zeng Department of Radiology, Utah Center for Advanced Imaging Research (UCAIR), University of Utah, Salt Lake City, Utah The main objective of

More information

PET: New Technologies & Applications, Including Oncology

PET: New Technologies & Applications, Including Oncology PET: New Technologies & Applications, Including Oncology, PhD, FIEEE Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Disclosures Research Contract, GE Healthcare

More information

Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center. Thursday September 8, 2011

Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center. Thursday September 8, 2011 Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center Thursday September 8, 2011 Quality Control Quality control should be performed because:

More information

Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM

Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM Preamplifiers and amplifiers The current from PMT must be further amplified before it can be processed and counted (the number of electrons yielded

More information

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893 Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -

More information

The future of nuclear imaging is clear

The future of nuclear imaging is clear Cardius X-ACT The future of nuclear imaging is clear Increased regulations, growing competition, and concerns about radiation exposure are just a sampling of the current challenges facing the nuclear medicine

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

Synchrotron X-ray tomographic microscopy Theory vs. practice

Synchrotron X-ray tomographic microscopy Theory vs. practice Synchrotron X-ray tomographic microscopy Theory vs. practice Federica Marone Swiss Light Source, Paul Scherrer Institut, Villigen, Switzerland Theory Radon transform Rf x = Beer-Lambert law I E = I 0 (E)e

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

First Applications of the YAPPET Small Animal Scanner

First Applications of the YAPPET Small Animal Scanner First Applications of the YAPPET Small Animal Scanner Guido Zavattini Università di Ferrara CALOR2 Congress, Annecy - FRANCE YAP-PET scanner Scintillator: YAP:Ce Size: matrix of 2x2 match like crystals

More information

Iterative Reconstruction in Image Space. Answers for life.

Iterative Reconstruction in Image Space. Answers for life. Iterative Reconstruction in Image Space Answers for life. Iterative Reconstruction in Image Space * (IRIS) * Please note: IRIS is used as an abbreviation for Iterative Reconstruction in Image Space throughout

More information

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017 A Fast Monolithic System for Proton Imaging Fritz DeJongh ProtonVDA Inc October 2017 Disclosures I am a cofounder and co-owner of ProtonVDA Inc We hold intellectual property rights on our proton imaging

More information

Celesteion Time-of-Flight Technology

Celesteion Time-of-Flight Technology Celesteion Time-of-Flight Technology Bing Bai, PhD Clinical Sciences Manager, PET/CT Canon Medical Systems USA Introduction Improving the care for every patient while providing a high standard care to

More information

Image Quality Assessment of Pixellated Systems

Image Quality Assessment of Pixellated Systems Image Quality Assessment of Pixellated Systems Andreas Goedicke, Herfried Wieczorek, Henrik Botterweck, Wolfgang Eckenbach, Ling Shao, Member, IEEE, Micheal Petrillo, Member, IEEE, Jinghan Ye, and John

More information

ACR Update in Nuclear Medicine Accreditation

ACR Update in Nuclear Medicine Accreditation Disclaimer ACR Update in Nuclear Medicine Accreditation Beth A. Harkness, MS, DABR, FACR Henry Ford Health System Detroit, MI ACR physics subcommittee for nuclear medicine accreditation. My facility is

More information

How Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram?

How Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram? November 2014, Volume 1, Number 4 How Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram? Hojjat Mahani 1,2, Alireza Kamali-Asl 3, *, Mohammad Reza Ay 2, 4 1. Radiation Application

More information

NON-UNIFORM ATTENUATION CORRECTION USING SIMULTANEOUS TRANSMISSION AND EMISSION CONVERGING TOMOGRAPHY

NON-UNIFORM ATTENUATION CORRECTION USING SIMULTANEOUS TRANSMISSION AND EMISSION CONVERGING TOMOGRAPHY 1134 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 39, NO. 4,1992 NON-UNIFORM ATTENUATION CORRECTION USING SIMULTANEOUS TRANSMISSION AND EMISSION CONVERGING TOMOGRAPHY C-H Tung, G. T. Gullberg, G. L. Zeng,

More information

Simulation and evaluation of a cost-effective high-performance brain PET scanner.

Simulation and evaluation of a cost-effective high-performance brain PET scanner. Research Article http://www.alliedacademies.org/biomedical-imaging-and-bioengineering/ Simulation and evaluation of a cost-effective high-performance brain PET scanner. Musa S Musa *, Dilber U Ozsahin,

More information

Technical Aspects: Image Reconstruction

Technical Aspects: Image Reconstruction Annals of Nuclear Cardiology Vol. 2 No. 1 68-72 REVIEW ARTICLE Technical Aspects: Masahisa Onoguchi, RT, PhD 1), Takahiro Konishi, RT, MS 2), Takayuki Shibutani, RT, MS 1), Shinro Matsuo, MD, PhD 3) and

More information

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation

More information

Effect of Post Reconstruction Gaussian Filtering on Image Quality and Myocardial Blood Flow Measurement with N 13 Ammonia PET

Effect of Post Reconstruction Gaussian Filtering on Image Quality and Myocardial Blood Flow Measurement with N 13 Ammonia PET Effect of Post Reconstruction Gaussian Filtering on Image Quality and Myocardial Blood Flow Measurement with N 13 Ammonia PET Hyeon Sik Kim 1, Sang Geon Cho 2, Ju Han Kim 3, Seong Young Kwon 1, Byeongil

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

More information

Image Deblurring. This chapter describes how to deblur an image using the toolbox deblurring functions.

Image Deblurring. This chapter describes how to deblur an image using the toolbox deblurring functions. 12 Image Deblurring This chapter describes how to deblur an image using the toolbox deblurring functions. Understanding Deblurring (p. 12-2) Using the Deblurring Functions (p. 12-5) Avoiding Ringing in

More information

The Watershed Algorithm: A Method to Segment Noisy PET Transmission Images

The Watershed Algorithm: A Method to Segment Noisy PET Transmission Images The Watershed Algorithm: A Method to Segment Noisy PET Transmission Images C. Riddell, P. Brigger, R.E. Carson and S.L. Bacharach National Institutes of Health, Bldg. 10 Room 1C401, Bethesda, MD 20892

More information

Software and Hardware in CCTA. Elly Castellano PhD

Software and Hardware in CCTA. Elly Castellano PhD Software and Hardware in CCTA Elly Castellano PhD Outline technical requirements for coronary CTA the modern cardiac CT scanner ECG-gating technology image reconstruction algorithms 2 Technical requirements

More information

Feasibility Study of Compton Scattering Enchanced Multiple Pinhole Imager for Nuclear Medicine

Feasibility Study of Compton Scattering Enchanced Multiple Pinhole Imager for Nuclear Medicine IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 5, OCTOBER 2003 1609 Feasibility Study of Compton Scattering Enchanced Multiple Pinhole Imager for Nuclear Medicine L. J. Meng, W. L. Rogers, N. H. Clinthorne,

More information

Frequency Domain Enhancement

Frequency Domain Enhancement Tutorial Report Frequency Domain Enhancement Page 1 of 21 Frequency Domain Enhancement ESE 558 - DIGITAL IMAGE PROCESSING Tutorial Report Instructor: Murali Subbarao Written by: Tutorial Report Frequency

More information

Conceptual Study of Brain Dedicated PET Improving Sensitivity

Conceptual Study of Brain Dedicated PET Improving Sensitivity Original Article PROGRESS in MEDICAL PHYSICS 27(4), Dec. 2016 https://doi.org/10.14316/pmp.2016.27.4.236 pissn 2508-4445, eissn 2508-4453 Conceptual Study of Brain Dedicated PET Improving Sensitivity Han-Back

More information

Design of a Static Full-Ring Multi-Pinhole Collimator for Brain SPECT

Design of a Static Full-Ring Multi-Pinhole Collimator for Brain SPECT Design of a Static Full-Ring Multi-Pinhole Collimator for Brain SPECT Karen Van Audenhaege, Student Member, IEEE, Roel Van Holen, Member, IEEE, Karel Deprez, Joel S. Karp, Senior Member, IEEE, Scott Metzler,

More information

The shortest distance to diagnosis Philips BrightView SPECT

The shortest distance to diagnosis Philips BrightView SPECT The shortest distance to diagnosis Philips Closer is better Simplicity is seeing something better right from the start. And is a completely new vision of what patient care can be, in a system as compact

More information

1.Discuss the frequency domain techniques of image enhancement in detail.

1.Discuss the frequency domain techniques of image enhancement in detail. 1.Discuss the frequency domain techniques of image enhancement in detail. Enhancement In Frequency Domain: The frequency domain methods of image enhancement are based on convolution theorem. This is represented

More information

SAR AUTOFOCUS AND PHASE CORRECTION TECHNIQUES

SAR AUTOFOCUS AND PHASE CORRECTION TECHNIQUES SAR AUTOFOCUS AND PHASE CORRECTION TECHNIQUES Chris Oliver, CBE, NASoftware Ltd 28th January 2007 Introduction Both satellite and airborne SAR data is subject to a number of perturbations which stem from

More information

2/14/2019. Nuclear Medicine Artifacts. Symmetric energy windows

2/14/2019. Nuclear Medicine Artifacts. Symmetric energy windows Nuclear Medicine Artifacts SCPMG Medical Imaging Technology & Informatics Medical Physics Group Brian Helbig, MS, DABR 1 2 Symmetric energy windows 3 1 Dynamic clinical study Energy peak shift Electrical

More information

FRCR Nuclear Medicine

FRCR Nuclear Medicine FRCR Nuclear Medicine FRCR LECTURES Lecture I 20/09/2016: Nuclear Medicine and Image Formation Lecture II 22/09/2016: Positron Emission Tomography & QA Lecture III 27/09/2016: Radiation Detectors - Radiation

More information

Study on the usefulness of whole body SPECT coronal image, MIP image in 67 Ga scintigraphy

Study on the usefulness of whole body SPECT coronal image, MIP image in 67 Ga scintigraphy ORIGINAL ARTICLE Annals of Nuclear Medicine Vol. 16, No. 3, 221 226, 2002 Study on the usefulness of whole body SPECT coronal image, MIP image in 67 Ga scintigraphy Seiji KAWAMURA,* Masatoshi ISHIBASHI,**

More information

GS Introduction to Medical Physics IV Laboratory 5 Gamma Camera Characteristics

GS Introduction to Medical Physics IV Laboratory 5 Gamma Camera Characteristics GS02 0193 Introduction to Medical Physics IV Laboratory 5 Gamma Camera Characteristics Purpose: To introduce some of the basic characteristics of a gamma camera. This lab will introduce gamma camera QC

More information

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for

More information

Aquilion Precision Ultra-High Resolution CT: Quantifying diagnostic image quality

Aquilion Precision Ultra-High Resolution CT: Quantifying diagnostic image quality Aquilion Precision Ultra-High CT: Quantifying diagnostic image quality Kirsten Boedeker, PhD, DABR Senior Manager, Quantitative Image Quality Canon Medical Systems Corporation Introduction Over the last

More information

Parameters Affecting on Intrinsic Uniformity Test For MEDISO

Parameters Affecting on Intrinsic Uniformity Test For MEDISO ISPUB.COM The Internet Journal of Nuclear Medicine Volume 5 Number 2 Parameters Affecting on Intrinsic Uniformity Test For MEDISO S Zobly, A Osman Citation S Zobly, A Osman. Parameters Affecting on Intrinsic

More information

180 pinhole-spect with tilted detector and OS-EM. reconstruction: phantom studies and potential

180 pinhole-spect with tilted detector and OS-EM. reconstruction: phantom studies and potential 180 pinhole-spect with tilted detector and OS-EM reconstruction: phantom studies and potential clinical applications Alain Seret 1, Michel Defrise 2 and Didier Blocklet 3. 1 Université de Liège (ULg),

More information

Pinhole collimator design for nuclear survey system

Pinhole collimator design for nuclear survey system Annals of Nuclear Energy 29 (2002) 2029 2040 www.elsevier.com/locate/anucene Pinhole collimator design for nuclear survey system Wanno Lee*, Gyuseong Cho Department of Nuclear Engineering, Korea Advanced

More information

Image Display and Perception

Image Display and Perception Image Display and Perception J. Anthony Seibert, Ph.D. Department of Radiology UC Davis Medical Center Sacramento, California, USA Image acquisition, display, & interpretation X-rays kvp mas Tube filtration

More information

International Journal of Scientific & Engineering Research, Volume 4, Issue 9, September ISSN

International Journal of Scientific & Engineering Research, Volume 4, Issue 9, September ISSN International Journal of Scientific & Engineering Research, Volume 4, Issue 9, September-013 06 Evaluating the effect of acquisition parameters on image quality and acquisition time with SPECT using collimator

More information

Comparing planar image quality of rotating slat and parallel hole collimation: influence of system modeling

Comparing planar image quality of rotating slat and parallel hole collimation: influence of system modeling Comparing planar image quality of rotating slat and parallel hole collimation: influence of system modeling Roel Van Holen, Stefaan Vandenberghe, Steven Staelens and Ignace Lemahieu ELIS Department, MEDISIP,

More information

Digital radiography (DR) post processing techniques for pediatric radiology

Digital radiography (DR) post processing techniques for pediatric radiology Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions

More information

2010 Philips BrightView XCT SPECT/CT

2010 Philips BrightView XCT SPECT/CT 2010 Philips BrightView XCT SPECT/CT Unit was purchased from Philips training center in 2015. Installed but never been used by the current facility. (Scroll for pictures) BrightView XCT Camera with PinPoint

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

Compensating for Nonstationary Blurring by Further Blurring and Deconvolution

Compensating for Nonstationary Blurring by Further Blurring and Deconvolution Compensating for Nonstationary Blurring by Further Blurring and Deconvolution Gengsheng L. Zeng Department of Radiology, Utah Center for Advanced Imaging Research, University of Utah, Salt Lake City, UT

More information

Design of a Multi-Pinhole Collimator and Its Evaluation for Application to High-Resolution Pre-Clinical SPECT system for Small Animal Imaging

Design of a Multi-Pinhole Collimator and Its Evaluation for Application to High-Resolution Pre-Clinical SPECT system for Small Animal Imaging Design of a Multi-Pinhole Collimator and Its Evaluation for Application to High-Resolution Pre-Clinical SPECT system for Small Animal Imaging Hyun-Ju Ryu The Graduate School Yonsei University Department

More information

Assessment of Image Quality of a PET/CT scanner for a Standarized Image situation Using a NEMA Body Phantom

Assessment of Image Quality of a PET/CT scanner for a Standarized Image situation Using a NEMA Body Phantom Assessment of Image Quality of a PET/CT scanner for a Standarized Image situation Using a NEMA Body Phantom The impact of Different Image Reconstruction Parameters on Image quality by QUAYE MICHAEL This

More information

Primer on molecular imaging technology

Primer on molecular imaging technology Primer on molecular imaging technology Craig S. Levin Division of Nuclear Medicine, Department of Radiology and Molecular Imaging Program at Stanford (MIPS), Stanford University School of Medicine, 300

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life.

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life. Neuro Thoracic Abdominal Abdominal Cardiovascular Pediatric SAFIRE Sinogram Affirmed Iterative Reconstruction Answers for life. SAFIRE * (Sinogram Affirmed Iterative Reconstruction) * The information

More information

International Journal of Advancedd Research in Biology, Ecology, Science and Technology (IJARBEST)

International Journal of Advancedd Research in Biology, Ecology, Science and Technology (IJARBEST) Gaussian Blur Removal in Digital Images A.Elakkiya 1, S.V.Ramyaa 2 PG Scholars, M.E. VLSI Design, SSN College of Engineering, Rajiv Gandhi Salai, Kalavakkam 1,2 Abstract In many imaging systems, the observed

More information

X-RAY COMPUTED TOMOGRAPHY

X-RAY COMPUTED TOMOGRAPHY X-RAY COMPUTED TOMOGRAPHY Bc. Jan Kratochvíla Czech Technical University in Prague Faculty of Nuclear Sciences and Physical Engineering Abstract Computed tomography is a powerful tool for imaging the inner

More information

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY 12 th A-PCNDT 2006 Asia-Pacific Conference on NDT, 5 th 10 th Nov 2006, Auckland, New Zealand PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY Rajashekar

More information

Intrinsic and Tomographic Evaluation of Siemens e.cam SPECT System at the Korle-Bu Teaching Hospital (Ghana)

Intrinsic and Tomographic Evaluation of Siemens e.cam SPECT System at the Korle-Bu Teaching Hospital (Ghana) Research Journal of Applied Sciences, Engineering and Technology 3(10): 1152-1158, 2011 ISSN: 2040-7467 Maxwell Scientific Organization, 2011 Submitted: July 17, 2011 Accepted: September 05, 2011 Published:

More information

Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2

Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2 Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2 1. Department of Medical Physics, University of Wisconsin-Madison 2. Department

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

Results of the Measurement of the Collimator Hole Angulation for Different Collimators of SPECT with Adaptive Quality Control Phantom

Results of the Measurement of the Collimator Hole Angulation for Different Collimators of SPECT with Adaptive Quality Control Phantom Modern Instrumentation, 2012, 1, 4953 http://dx.doi.org/10.4236/mi.2012.14007 Published Online October 2012 (http://www.scirp.org/journal/mi) Results of the Measurement of the Collimator Hole Angulation

More information

SKYLight. Nuclear Camera Platform. Setting new standards. in molecular imaging

SKYLight. Nuclear Camera Platform. Setting new standards. in molecular imaging SKYLight Nuclear Camera Platform Setting new standards in molecular imaging The next generation of imaging platforms A VISION WITHOUT LIMITATIONS We started with a clean slate and no preconceived notions

More information

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology CODE: RADT 156 INSTITUTE: Health Science TITLE: Equipment Operation I DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of equipment operation and maintenance of radiographic

More information

PET Performance Measurements for an LSO- Based Combined PET/CT Scanner Using the National Electrical Manufacturers Association NU Standard

PET Performance Measurements for an LSO- Based Combined PET/CT Scanner Using the National Electrical Manufacturers Association NU Standard PET Performance Measurements for an LSO- Based Combined PET/CT Scanner Using the National Electrical Manufacturers Association NU 2-2001 Standard Yusuf E. Erdi, DSc 1 ; Sadek A. Nehmeh, PhD 1 ; Tim Mulnix,

More information

30 lesions. 30 lesions. false positive fraction

30 lesions. 30 lesions. false positive fraction Solutions to the exercises. 1.1 In a patient study for a new test for multiple sclerosis (MS), thirty-two of the one hundred patients studied actually have MS. For the data given below, complete the two-by-two

More information

Applications of Flash and No-Flash Image Pairs in Mobile Phone Photography

Applications of Flash and No-Flash Image Pairs in Mobile Phone Photography Applications of Flash and No-Flash Image Pairs in Mobile Phone Photography Xi Luo Stanford University 450 Serra Mall, Stanford, CA 94305 xluo2@stanford.edu Abstract The project explores various application

More information

Comparison of Reconstruction Algorithms for Images from Sparse-Aperture Systems

Comparison of Reconstruction Algorithms for Images from Sparse-Aperture Systems Published in Proc. SPIE 4792-01, Image Reconstruction from Incomplete Data II, Seattle, WA, July 2002. Comparison of Reconstruction Algorithms for Images from Sparse-Aperture Systems J.R. Fienup, a * D.

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

Appendix I: Artefacts and Trouble-shooting

Appendix I: Artefacts and Trouble-shooting Appendix I: Artefacts and Trouble-shooting Slide set of 101 slides based on the chapter authored by E. Busemann Sokole, N.J. Forwood of the publication (ISBN 978 92 0 143810 2): Nuclear Medicine Physics:

More information

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory

More information

Distortion Correction in LODOX StatScan X-Ray Images

Distortion Correction in LODOX StatScan X-Ray Images Distortion Correction in LODOX StatScan X-Ray Images Matthew Paul Beets November 27, 2007 X-Ray images produced by the LODOX StatScan machine contain a non-linear distortion in the direction of the beam

More information