PET: New Technologies & Applications, Including Oncology
|
|
- Philip Holland
- 6 years ago
- Views:
Transcription
1 PET: New Technologies & Applications, Including Oncology, PhD, FIEEE Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA
2 Disclosures Research Contract, GE Healthcare Advisory Board, Aposense Inc.
3 Objectives 1. Review history and uses of PET/CT 2. Understand recent technology developments and challenges 3. Learn the context of the potential future applications of PET/CT imaging in clinical practice Outline Respiratory motion compensation Time of flight imaging Advanced modeling of PET physics in image reconstruction Extended axial field of view New detector systems PET/MR scanners CT dose reduction methods Cost effective PET/CT scanners
4 PET Technology Innovations 1953 First coincidence positron imaging system 1975 PETT III 1989 whole body imaging D PET 1998 PET/CT respiratory gating 2008 time offlight PET/MR? 2,500,000 2,000,000 Procedures/yr PET/CT % of Sales 100% 80% Procedures/yr 1,500,000 1,000, ,000 60% 40% 20% % scanners 0 0%
5 Respiratory motion compensation
6 Impact of Respiratory Motion Breath hold (BH) PET SUVs are 26% greater than free breathing (FB) PET SUVs Truth is not known Kawano, JNM 2008
7 Impact of Respiratory Motion Static wholebody Single respiratory phase (1 of 7, so noisier) 1 cc lesion on CT The SUV of the lesion goes from 2 in the static image to 6 in one phase of the respiratory gated image sequence
8 A Hierarchy of Respiratory Motion Compensation Methods 1. Breath hold PET 2. Cine+Helical CTAC 3. Respiratory gated PET (4D PET) 4. Phase matched 4D PET and 4D CTAC 5. Quiescent Period Gating 6. Respiratory Aligned and Summed Phases (RASP) 7. Internal External Correlation (INTEX) 8. Dual respiratory / cardiac gating 9. Reconstruction based methods Increasing complexity
9 Helical+CINE CTAC Acquisition to Compensating For Patient Motion 1. Standard non contrast helical CT (diagnostic beam) for both CT imaging correlation and for CTbased attenuation correction (CTAC) 2. Cine CT acquired over the diaphragm region for respiratory motion (much like a PET transmission scan: based on method of Pan et al. JNM 2005) 3. Average of helical+cine CT acquired is used for CTAC of PET data Sum of all CT scans used for CTAC
10 Helical+CINE CTAC Protocol upper limit of diaphragm motion Cine CT range lower limit of diaphragm motion max inspiration max expiration
11 Helical+CINE CTAC Protocol area of impact reduced 'banana' artifacts new cine+helical CTAC standard helical CTAC
12 Respiratory Gated PET/CT Imaging reflective block CT PET IR LEDs and camera Gating inputs PC w/ frame grabber + trigger generator Respiratory trace information
13 Wholebody Respiratory Gated PET Notes: motion throughout body changes in moving lesion intensity apparent fixed diaphragm location QuickTime and a YUV420 codec decompressor are needed to see this picture.
14 Single Phase CT based Attenuation Correction for 4D PET 1 X = Respiratory phases 2 3 X X = = N X = CTAC Phase Image Volumes PET EM phase sinograms Attenuation corrected PET data Reconstructed PET images
15 Phase matched CT based Attenuation Correction 1 X = Respiratory phases 2 3 X X = = N X = CTAC Phase Image Volumes PET EM phase sinograms Attenuation corrected PET data Reconstructed PET images
16 Single phase CT attenuation correction (CTAC) image (e.g. helical CT) Phasematched CTAC QuickTime and a decompressor are needed to see this picture. QuickTime and a decompressor are needed to see this picture. We see the time average of these images, unless we use respiratory gating
17 Phase matched 4D PET and 4D CTAC CT PET/CT 4D CT 4D PET/CT
18 Time of flight imaging
19 Time of Flight (TOF) PET/CT best guess about location (d) c = 3x10 10 cm/s, so t = 600 ps ~ d = 10 cm in resolution Surti (JNM 2011) shows improved lesion detection in large patients Little information yet on effect on impact on quantitative accuracy timing resolution uncertainty d = c t/2 + + e - annihilation
20 Impact of TOF PET Detection QuickTime and a decompressor are needed to see this picture.
21 Impact on quantitation Typical coincident count levels 75 kg patient, 120 MBq, 3 min/bed (BG = ~1kBq/cc) VOI & EANM MAX & EANM VOI & PSF+TOF MAX & PSF+TOF High coincident count levels 75 kg patient, 120 MBq, 10 min/bed (BG = ~1kBq/cc) VOI & EANM VOI & PSF+TOF MAX & EANM MAX & PSF+TOF Recovery coefficient Recovery coefficient Sphere volume (ml) Sphere volume (ml) Courtesy Ronald Boellaard
22 Advanced modeling of PET physics in image reconstruction
23 Including improved physics modeling in image reconstruction In principle can remove detector blurring
24 Phantom measurements of ringing artifact QuickTime and a decompressor are needed to see this picture. real? QuickTime and a decompressor are needed to see this picture. QuickTime and a decompressor are needed to see this picture. Bai, 2010 IEEE MIC conf record
25 Extended axial field of view
26 Extended Axial Field of View PET scanner component has 3 rings of detector blocks. Siemens added one more ring (they quote "extra 33 %") to extend the axial field of view Increases scanner efficiency and/or allows for shorter scan times
27 NEMA NU performance measurements Noise Equivalent Count Rate Peak NECR (kcps) Biograph TP Peak NECR Biograph Biograph TP 34 kbq/ml 31 kbq/ml Biograph Courtesy David Townsend PhD: University of Tennessee at Knoxville
28 New detection systems Avalanche photodiodes Solid state silicon photomultipliers
29 Position sensitive avalanche photodiode (PSAPD) silicon based detectors Could ultimately be made cheaply in high volumes High optical quantum efficiency (up to 4x higher than PMTs) and wide spectral response Insensitivity to magnetic fields (used in PET/MR) Timing resolution of ~3 ns Lower SNR than PMTs Temperature sensitive gains Flood histogram from coupling a 8 x 8 mm 2 PSAPD to an LSO array (1 mm pixels, 20 mm tall) Shah et al. TNS 2004
30 The silicon photomultiplier (SiPM) Combine advantages of PMTs and APDs Still very expensive High gain (like PMTs) Insensitive to magnetic fields (e.g. for PET/MR) Excellent energy and timing resolution Temperature sensitive gain QuickTime and a decompressor are needed to see this picture. Simplified structure of a SiPM composed of G APD cells Roncali and Cherry, Ann Biomed Eng, 2011
31 PET/MR scanners
32 Clinical PET/MR scanner Siemens mmr first joint scanner (Philips and GE currently have 'tandem' systems) Uses APDs Expensive Attenuation correction for bone an open problem for all systems Clinical results from Drzezga et al, JNM June 2012 QuickTime and a decompressor are needed to see this picture. PET/CT PET/MR
33 CT dose reduction methods
34 Iterative CT image reconstruction Can be used in CT like it is in PET Computationally now tractable Several approaches under development for CT only scanners Gradually move to PET/CT scanners
35 AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation RSD Phantom: 120 kvp 100 mas 25 mas 12 mas FBP air water water+ contrast MBIR
36 Cost effective PET/CT scanners
37 'Value based' PET/CT Refurbished PET/CT scanners Strong market New systems GE Optima 560 Philips Gemini LXL Siemens Excel 20 mct Seem to be based on reduced capabilities compared to full featured scanners
Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard.
Chiara Secco PET Performance measurements of the new LSO-Based Whole Body PET/CT Scanner biograph 16 HI-REZ using the NEMA NU 2-2001 Standard. INTRODUCTION Since its introduction, CT has become a fundamental
More informationNew Technology in Nuclear Medicine
New Technology in Nuclear Medicine Reed G. Selwyn, PhD, DABR Vice Chair of Research & Imaging Sciences Associate Professor and Chief, Medical Physics Dept. of Radiology, University of New Mexico Objectives
More informationPET/CT Instrumentation Basics
/ Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire
More informationPerformance evaluation of a new highsensitivity time-of-flight clinical PET/CT system
Huo et al. EJNMMI Physics (2018) 5:29 https://doi.org/10.1186/s40658-018-0229-4 EJNMMI Physics ORIGINAL RESEARCH Open Access Performance evaluation of a new highsensitivity time-of-flight clinical PET/CT
More informationDesigning an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare
GE Healthcare Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare There is excitement across the industry regarding the clinical potential of a hybrid
More informationLSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option
LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option Y. Bercier, Member, IEEE, M. Casey, Member, IEEE, J. Young, Member, IEEE, T. Wheelock, Member, IEEE, T. Gremillion Abstract-- Factors which
More informationFocusing on high performance
Advanced Molecular Imaging Vereos PET/CT Focusing on high performance Michael A. Miller, PhD, Philips, Advanced Molecular Imaging Physics This white paper presents a description of the Vereos digital PET/CT
More informationAn innovative detector concept for hybrid 4D-PET/MRI Imaging
Piergiorgio Cerello (INFN - Torino) on behalf of the 4D-MPET* project *4 Dimensions Magnetic compatible module for Positron Emission Tomography INFN Perugia, Pisa, Torino; Polytechnic of Bari; University
More informationPitfalls and Remedies of MDCT Scanners as Quantitative Instruments
intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature
More informationHISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD
CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING
More informationCelesteion Time-of-Flight Technology
Celesteion Time-of-Flight Technology Bing Bai, PhD Clinical Sciences Manager, PET/CT Canon Medical Systems USA Introduction Improving the care for every patient while providing a high standard care to
More informationPositron Emission Tomography - PET
Positron Emission Tomography - PET Positron Emission Tomography Positron Emission Tomography (PET): Coincidence detection of annihilation radiation from positron-emitting isotopes followed by tomographic
More informationThe PennPET Explorer Scanner for Total Body Applications
The PennPET Explorer Scanner for Total Body Applications JS Karp, MJ Geagan, G Muehllehner, ME Werner, T McDermott, JP Schmall, V Viswanath, University of Pennsylvania, Philadelphia, PA AE Perkins, C-H
More informationCombined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8
Combined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8 UNIVERSITY OF CAMBRIDGE 1.2 Rob C. Hawkes 1, Tim D. Fryer 1, Alun J. Lucas 1,2, Stefan B. Siegel 3, Richard
More informationPET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002
PET Detectors William W. Moses Lawrence Berkeley National Laboratory March 26, 2002 Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes
More informationFuture directions in Nuclear Medicine Instrumentation
Future directions in Nuclear Medicine Instrumentation Where are we going - and why? First, the disclosure list My group at the University of Washington has research support from: NIH DOE General Electric
More informationFundamentals of Positron Emission Tomography (PET)
Fundamentals of Positron Emission Tomography (PET) NPRE 435, Principles of Imaging with Ionizing Radiation, Fall 2017 Content Fundamentals of PET Camera & Detector Design Real World Considerations Performance
More informationPET Performance Measurements for an LSO- Based Combined PET/CT Scanner Using the National Electrical Manufacturers Association NU Standard
PET Performance Measurements for an LSO- Based Combined PET/CT Scanner Using the National Electrical Manufacturers Association NU 2-2001 Standard Yusuf E. Erdi, DSc 1 ; Sadek A. Nehmeh, PhD 1 ; Tim Mulnix,
More informationWide-Detector CT for TAVR Planning:
Wide-Detector CT for TAVR Planning: Impact on Iodine Dose, Radiation Dose, and Image Quality SCBTMR 2015 Annual Course Thursday, October 8 William P. Shuman MD FSCBTMR Department of Radiology University
More informationLightburst Digital Detector
GE Healthcare Lightburst Digital Detector INTRODUCTION In clinical practice, PET/CT imaging helps clinicians visualize disease at an early stage, before it metastasizes and involves other organs, tissues
More informationPerformance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET
Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET A. Kuhn, S. Surti, Member, IEEE, J. S. Karp, Senior Member, IEEE, G. Muehllehner, Fellow, IEEE, F.M. Newcomer, R. VanBerg Abstract--
More informationIterative Reconstruction
RECENT ADVANCES IN CT RADIATION DOSE REDUCTION TECHNIQUES Iterative Reconstruction Kalpana Kanal, PhD, FSCBTMR, FACR, FAAPM Professor and Director, Diagnostic Physics Section University of Washington Seattle,
More informationTOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions
CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT
More informationPET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner
PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner September, 2017 Results submitted to Physics in Medicine & Biology Negar Omidvari 1, Jorge Cabello 1, Geoffrey Topping
More informationSoftware and Hardware in CCTA. Elly Castellano PhD
Software and Hardware in CCTA Elly Castellano PhD Outline technical requirements for coronary CTA the modern cardiac CT scanner ECG-gating technology image reconstruction algorithms 2 Technical requirements
More informationNoise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner
Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner Kisung Lee, Member, IEEE, Paul E. Kinahan, Senior Member, Robert S. Miyaoka, Member, IEEE, Jeffrey A. Fessler,
More informationSimultaneous Reconstruction of the Activity Image and Registration of the CT image in TOF-PET. Ahmadreza Rezaei, Johan Nuyts
Simultaneous Reconstruction of the Activity Image and Registration of the CT image in TOF-PET Ahmadreza Rezaei, Johan Nuyts Activity Reconstruction & Attenuation Registration Attenuation Correction, Background
More informationDevelopment of PET using 4 4 Array of Large Size Geiger-mode Avalanche Photodiode
2009 IEEE Nuclear Science Symposium Conference Record M09-8 Development of PET using 4 4 Array of Large Size Geiger-mode Avalanche Photodiode K. J. Hong, Y. Choi, J. H. Kang, W. Hu, J. H. Jung, B. J. Min,
More informationPhilips Astonish. Key advantages. including improved image quality and
Philips Key advantages including improved image quality and with AC offers improved image quality, interpretative certainty, diagnostic Alternatively, simplify patient care by exposing patients to reduced
More informationSlide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?
Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM
More informationCOMPUTED TOMOGRAPHY 1
COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the
More informationSPECT Reconstruction & Filtering
SPECT Reconstruction & Filtering Goals Understand the basics of SPECT Reconstruction Filtered Backprojection Iterative Reconstruction Make informed choices on filter selection and settings Pre vs. Post
More informationRecent Advances in Hybrid Molecular Imaging Systems
103 Recent Advances in Hybrid Molecular Imaging Systems Jae Sung Lee, PhD 1 Joong Hyun Kim, PhD 2 1 Departments of Nuclear Medicine and Biomedical Sciences, Seoul National University College of Medicine,
More informationSimulation and evaluation of a cost-effective high-performance brain PET scanner.
Research Article http://www.alliedacademies.org/biomedical-imaging-and-bioengineering/ Simulation and evaluation of a cost-effective high-performance brain PET scanner. Musa S Musa *, Dilber U Ozsahin,
More informationBreast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School
Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis
More informationDetector technology in simultaneous spectral imaging
Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become
More informationdiagnostic examination
RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011
More information1. Patient size AEC. Large Patient High ma. Small Patient Low ma
Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationUsing GATE to understand performance of a full-torso PET scanner
University of Pennsylvania Using GATE to understand performance of a full-torso PET scanner Varsha Viswanath 1 Margaret E. Daube-Witherspoon 1, Matthew E. Werner 1, Suleman Surti 1, Andreia Trindade 2,
More informationPD233: Design of Biomedical Devices and Systems
PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/
More informationWilliam Hallet - PRIMA IV 1
Quantitative and application specific imaging PET: the measurement process,reconstruction, calibration, quantification Dr William Hallett Centre for Imaging Sciences Imperial College Hammersmith Hospital
More informationThe Point Source. July 2011 Volume 2, Issue 1. Executive message. Rich Fabian. In this issue
The Point Source July 2011 Volume 2, Issue 1 In this issue Executive message from Rich Fabian...1 Handling anxiety...2 Product updates Introducing TruFlight Select PET/CT...3 BrightView XCT Version 2.5...4
More informationIterative Reconstruction in Image Space. Answers for life.
Iterative Reconstruction in Image Space Answers for life. Iterative Reconstruction in Image Space * (IRIS) * Please note: IRIS is used as an abbreviation for Iterative Reconstruction in Image Space throughout
More informationInside Biograph mct.
Inside Biograph mct The technologies behind the world s first molecular CT. www.siemens.com/mi Large 78 cm bore helps reduce claustrophobia and provides more room for RTP positioning devices. 227 kg (500
More informationInitial Certification
Initial Certification Nuclear Medical Physics (NMP) Study Guide Part 2 Content Guide and Sample Questions The content of all ABR exams is determined by a panel of experts who select the items based on
More informationPhotomultiplier Tube
Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or
More informationThe Influence of Crystal Configuration and PMT on PET Time-of-Flight Resolution
The Influence of Crystal Configuration and PMT on PET Time-of-Flight Resolution Christopher Thompson Montreal Neurological Institute and Scanwell Systems, Montreal, Canada Jason Hancock Cross Cancer Institute,
More informationPerformance evaluation of the Biograph mct Flow PET/CT system according to the NEMA NU standard
Rausch et al. EJNMMI Physics (2015) 2:26 DOI 10.1186/s40658-015-0132-1 ORIGINAL RESEARCH Open Access Performance evaluation of the Biograph mct Flow PET/CT system according to the NEMA NU2-2012 standard
More informationPerformance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET
2005 IEEE Nuclear Science Symposium Conference Record M11-126 Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET Jin Zhang, Member,
More informationAutomated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London
Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients
More informationData. microcat +SPECT
Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for
More information12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees
Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees
More information2594 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER /$ IEEE
2594 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER 2009 Investigation of Depth of Interaction Encoding for a Pixelated LSO Array With a Single Multi-Channel PMT Yongfeng Yang, Member, IEEE,
More informationSimulation of Algorithms for Pulse Timing in FPGAs
2007 IEEE Nuclear Science Symposium Conference Record M13-369 Simulation of Algorithms for Pulse Timing in FPGAs Michael D. Haselman, Member IEEE, Scott Hauck, Senior Member IEEE, Thomas K. Lewellen, Senior
More informationYinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2
Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2 1. Department of Medical Physics, University of Wisconsin-Madison 2. Department
More informationMonte Carlo Simulation Study of a Dual-Plate PET Camera Dedicated to Breast Cancer Imaging
IEEE Nuclear Science Symposium Conference Record M-9 Monte Carlo Simulation Study of a Dual-Plate PET Camera Dedicated to Breast Cancer Imaging Jin Zhang, Member, IEEE, Peter D. Olcott, Member, IEEE, Angela
More informationChanging the Shape of Nuclear Medicine
TRUTH IN IMAGING Changing the Shape of Nuclear Medicine Multi-Purpose SPECT Scanner Nothing Gets Closer Introducing 360 Body Contour Scanning With 360 degree detector coverage, and unique proximity sensors
More informationRadionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)
Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in
More informationPerformance Characteristics of a State of the Art Preclinical PET/SPECT/CT Scanner
Performance Characteristics of a State of the Art Preclinical PET/SPECT/CT Scanner Nya Mehnwolo Boayue 1 Samuel Kuttner 1 1 Center for Diagnostic Physics University Hospital of North-Norway Medfys, 2016
More informationResearch Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.
Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National
More information2/14/2019. Nuclear Medicine Artifacts. Symmetric energy windows
Nuclear Medicine Artifacts SCPMG Medical Imaging Technology & Informatics Medical Physics Group Brian Helbig, MS, DABR 1 2 Symmetric energy windows 3 1 Dynamic clinical study Energy peak shift Electrical
More informationAttenuation Correction in Hybrid MR-BrainPET Imaging
Mitglied der Helmholtz-Gemeinschaft Attenuation Correction in Hybrid MR-BrainPET Imaging Elena Rota Kops Institute of Neuroscience and Biophysics Medicine Brain Imaging Physics Interactions of 511 kev
More informationCHAPTER 8 GENERIC PERFORMANCE MEASURES
GENERIC PERFORMANCE MEASURES M.E. DAUBE-WITHERSPOON Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, United States of America 8.1. INTRINSIC AND EXTRINSIC MEASURES 8.1.1.
More informationConceptual Study of Brain Dedicated PET Improving Sensitivity
Original Article PROGRESS in MEDICAL PHYSICS 27(4), Dec. 2016 https://doi.org/10.14316/pmp.2016.27.4.236 pissn 2508-4445, eissn 2508-4453 Conceptual Study of Brain Dedicated PET Improving Sensitivity Han-Back
More informationPerformance Evaluation of SiPM Detectors for PET Imaging in the Presence of Magnetic Fields
2008 IEEE Nuclear Science Symposium Conference Record M02-4 Performance Evaluation of SiPM Detectors for PET Imaging in the Presence of Magnetic Fields Samuel España, Student Member, IEEE, Gustavo Tapias,
More informationA NOVEL CONCEPT FOR A POSITRON EMISSION TOMOGRAPHY SCANNER
A NOVEL CONCEPT FOR A POSITRON EMISSION TOMOGRAPHY SCANNER An Undergraduate Research Scholars Thesis by BRIAN KELLY, MATTHEW LEE ELLIOT LEVIN and JEENA KHATRI Submitted to Honors and Undergraduate Research
More informationImage Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose
Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation
More informationGE Healthcare. Discovery MI. with LightBurst Digital 4-Ring Detector
GE Healthcare Discovery MI with LightBurst Digital 4-Ring Detector Discovery TM MI is a PET/CT system conceptualized with lofty goals, equal only to your own. It was created to help you diagnose and stage
More informationComputed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology
The Fundamentals of... Computed Tomography Computed Tomography (CT) systems use x-rays to produce images of slices through a patient s anatomy. Despite having lower spatial resolution than other x-ray
More informationSIGNA Explorer Lift revives our MR
Seiji Shiotani, MD, PhD Seirei Fuji Hospital in Fuji City, Shizuoka, Japan Masayoshi Sugimura Seirei Fuji Hospital in Fuji City, Shizuoka, Japan SIGN Explorer Lift revives our MR The clinical usefulness
More informationPerformance measurements of a depth-encoding PET detector module based on positionsensitive
Home Search Collections Journals About Contact us My IOPscience Performance measurements of a depth-encoding PET detector module based on positionsensitive avalanche photodiode read-out This article has
More informationDevelopment of an innovative LSO-SiPM detector module for high-performance Positron Emission Tomography
Development of an innovative LSO-SiPM detector module for high-performance Positron Emission Tomography Maria Leonor Trigo Franco Frazão leonorfrazao@ist.utl.pt Instituto Superior Técnico, Lisboa, Portugal
More informationModel Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise
Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise Jonas Rydberg, M.D. Professor of Radiology Indiana University School of Medicine Indianapolis, Indiana Medical
More information7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline
Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation Image Quality Review I: Basics and Image Quality TH-A-16A-1 Thursday 7:30AM - 9:30AM Room: 16A J. Anthony
More informationLaBr 3 :Ce, the latest crystal for nuclear medicine
10th Topical Seminar on Innovative Particle and Radiation Detectors 1-5 October 2006 Siena, Italy LaBr 3 :Ce, the latest crystal for nuclear medicine Roberto Pani On behalf of SCINTIRAD Collaboration INFN
More informationInvestigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras
Egyptian J. Nucl. Med., Vol 2, No. 2, Dec. 2009 82 PHYSICS, Original Artical Investigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras Abdelsattar, M.B. Ph.D.; BuHumaid,
More information160-slice CT SCANNER / New Standard for the Future
TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.
More informationImprovement of CT image quality with iterative reconstruction idose4
Improvement of CT image quality with iterative reconstruction idose4 Poster No.: C-0387 Congress: ECR 2014 Type: Scientific Exhibit Authors: M.-L. Olsson, K. Norrgren, M. Söderberg; Malmö/SE Keywords:
More informationMTF and NPS of single-shot dual-energy sandwich detectors
MTF and NPS of single-shot dual-energy sandwich detectors Junwoo Kim, a Dong Woon Kim, a Hanbean Youn, b,c Ho Kyung Kim a,c a School of Mechanical Engineering, Pusan National University, Busan 609-735,
More informationIntroduction. Parametric Imaging. The Ultrasound Research Interface: A New Tool for Biomedical Investigations
The Ultrasound Research Interface: A New Tool for Biomedical Investigations Shelby Brunke, Laurent Pelissier, Kris Dickie, Jim Zagzebski, Tim Hall, Thaddeus Wilson Siemens Medical Systems, Issaquah WA
More informationThe Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017
The Current State of EPID-Based Linear Accelerator Quality Assurance Timothy Ritter, PhD, DABR, FAAPM 1 Disclosures Employed by the Veterans Health Administration Faculty appointment with the University
More informationHIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE
HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE Takeyuki Hashimoto 1), Morio Onoe 2), Hiroshi Nakamura 3), Tamon Inouye 4), Hiromichi Jumonji 5), Iwao Takahashi 6); 1)Yokohama Soei
More informationTutors Dominik Dannheim, Thibault Frisson (CERN, Geneva, Switzerland)
Danube School on Instrumentation in Elementary Particle & Nuclear Physics University of Novi Sad, Serbia, September 8 th 13 th, 2014 Lab Experiment: Characterization of Silicon Photomultipliers Dominik
More informationDevelopment of the LBNL Positron Emission Mammography Camera
Development of the LBNL Positron Emission Mammography Camera J.S. Huber, Member, IEEE, W.S. Choong, Member, IEEE, J. Wang, Member, IEEE, J.S. Maltz, Member, IEEE, J. Qi, Member, IEEE, E. Mandelli, Member,
More informationDose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp
Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division
More informationNuovi sensori di luce allo stato solido applicati alla medicina nucleare. Claudio Piemonte Chief Scientist, Fondazione Bruno Kessler
Nuovi sensori di luce allo stato solido applicati alla medicina nucleare Claudio Piemonte Chief Scientist, Fondazione Bruno Kessler piemonte@fbk.eu Indice La Fondazione Bruno Kessler - Chi siamo, cosa
More informationDigital radiography (DR) post processing techniques for pediatric radiology
Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions
More informationAttrius. product guide. a new perspective for cardiac imaging. contact. product guide. 530 oakmont lane westmont, illinois p
product guide Attrius a new perspective for cardiac imaging contact 530 oakmont lane westmont, illinois 60559 f 317.576.0358 sales@positron.com www.positron.com product guide p 317.576.0183 introducing
More informationCharacterisation of SiPM Index :
Characterisation of SiPM --------------------------------------------------------------------------------------------Index : 1. Basics of SiPM* 2. SiPM module 3. Working principle 4. Experimental setup
More informationRobert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center. Thursday September 8, 2011
Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center Thursday September 8, 2011 Quality Control Quality control should be performed because:
More informationHow Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram?
November 2014, Volume 1, Number 4 How Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram? Hojjat Mahani 1,2, Alireza Kamali-Asl 3, *, Mohammad Reza Ay 2, 4 1. Radiation Application
More informationX-Ray-Based Attenuation Correction for Positron Emission Tomography/Computed Tomography Scanners
X-Ray-Based Attenuation Correction for Positron Emission Tomography/Computed Tomography Scanners Paul E. Kinahan, Bruce H. Hasegawa, and Thomas Beyer A synergy of positron emission tomography (PET)/computed
More informationNeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results
NeuViz 16 Computed Tomography Elevating routine imaging for exceptional results Essence NeuViz 16 Raising the bar on clinical utility in routine imaging. Get more. More clinical information for patients.
More informationRadiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893
Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -
More informationTCSPC at Wavelengths from 900 nm to 1700 nm
TCSPC at Wavelengths from 900 nm to 1700 nm We describe picosecond time-resolved optical signal recording in the spectral range from 900 nm to 1700 nm. The system consists of an id Quantique id220 InGaAs
More informationAn Activity in Computed Tomography
Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).
More informationMaximum Performance, Minimum Space
TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.
More informationCharacterization of a 64 Channel PET Detector Using Photodiodes for Crystal Identification *
Characterization of a 64 Channel PET Detector Using Photodiodes for Crystal Identification * J. S. Huber, Member, IEEE, W.W. Moses, Senior Member, IEEE, S.E. Derenzo, Senior Member, IEEE, M.H. Ho, M.S.
More informationAssessment of Image Quality of a PET/CT scanner for a Standarized Image situation Using a NEMA Body Phantom
Assessment of Image Quality of a PET/CT scanner for a Standarized Image situation Using a NEMA Body Phantom The impact of Different Image Reconstruction Parameters on Image quality by QUAYE MICHAEL This
More information