Performance Characteristics of a State of the Art Preclinical PET/SPECT/CT Scanner
|
|
- Lester Norton
- 5 years ago
- Views:
Transcription
1 Performance Characteristics of a State of the Art Preclinical PET/SPECT/CT Scanner Nya Mehnwolo Boayue 1 Samuel Kuttner 1 1 Center for Diagnostic Physics University Hospital of North-Norway Medfys, 2016 Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, / 11
2 Motivation Motivation PET subsystem SPECT subsystem CT subsystem Summary Aim of study NEMA NU4 for PET No current standard for preclinical SPECT or CT Literature Study Lots of information in the literature but no current accepted standard NEMA NU 4 for PET but has limitations eg. Reconstruction Develop measurement protocols and analysis methods in matlab PET SPECT CT Image Reconstruction Signal to Noise Ratio Spatial Resolution Spatial Resolution Spatial Resolution Uniformity Recovery Coefficient Image Reconstruction Spill over Ratio Sensitivity Sensitivity Uniformity Quantification Noise Uniformity Energy Resolution Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
3 Motivation Motivation PET subsystem SPECT subsystem CT subsystem Summary Aim of study NEMA NU4 for PET No current standard for preclinical SPECT or CT Literature Study Lots of information in the literature but no current accepted standard NEMA NU 4 for PET but has limitations eg. Reconstruction Develop measurement protocols and analysis methods in matlab PET SPECT CT Image Reconstruction Signal to Noise Ratio Spatial Resolution Spatial Resolution Spatial Resolution Uniformity Recovery Coefficient Image Reconstruction Spill over Ratio Sensitivity Sensitivity Uniformity Quantification Noise Uniformity Energy Resolution Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
4 Motivation Motivation PET subsystem SPECT subsystem CT subsystem Summary Trifoil Imaging: Triumph II R small animal PET/SPECT/CT scanner Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
5 PET subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary PET Noise and Spatial Resolution 0.25 mm 22 Na point source 1.1MBq MLEM 3D Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
6 PET subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary PET Noise and Spatial Resolution 0.25 mm 22 Na point source 1.1MBq MLEM 3D Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
7 PET subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary PET Noise and Spatial Resolution 0.25 mm 22 Na point source 1.1MBq 7.4MBq 18 F in a cylindrical phantom MLEM 3D Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
8 PET subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary PET Noise and Spatial Resolution 0.25 mm 22 Na point source 1.1MBq 7.4MBq 18 F in a cylindrical phantom MLEM 3D Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
9 PET subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary PET Recovery Coefficient and Spill over Ratio NEMA NU4 Image Quality Phantom RC calculation SOR calculation RC decreases with rod diameter Spill over from hot surroundings to cold water or air compartment due to system PSF Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
10 SPECT subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary SPECT Spatial Resolution and Sensitivity.25mm Co-57 Point source 3.7MBq Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
11 SPECT subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary SPECT Spatial Resolution and Sensitivity.25mm Co-57 Point source 3.7MBq Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
12 SPECT subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary SPECT Spatial Resolution and Sensitivity.25mm Co-57 Point source 3.7MBq Cylindrical phantom Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
13 SPECT subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary SPECT Spatial Resolution and Sensitivity.25mm Co-57 Point source 3.7MBq Cylindrical phantom Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
14 SPECT subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary SPECT reconstruction optimization (OSEM algorithm) Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
15 SPECT subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary SPECT reconstruction optimization (OSEM algorithm) EM Equivalent iterations = iterations * subsets Brambilla et al, Ann Nucl Med Apr;19(2):75-82 Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
16 SPECT subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary SPECT reconstruction optimization (OSEM algorithm) EM Equivalent iterations = iterations * subsets Brambilla et al, Ann Nucl Med Apr;19(2):75-82 Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
17 SPECT subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary SPECT reconstruction optimization (OSEM algorithm) EM Equivalent iterations = iterations * subsets Brambilla et al, Ann Nucl Med Apr;19(2):75-82 Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
18 SPECT subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary SPECT reconstruction optimization (OSEM algorithm) 5 iterations 8 subsets Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
19 Nya Mehnwolo Boayue, Samuel Kuttner 5Preclinical iterations PET/SPECT/CT 8 subsets February 10, /11 SPECT subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary SPECT reconstruction optimization (OSEM algorithm) EM Equivalent iterations = iterations * subsets Brambilla et al, Ann Nucl Med Apr;19(2):75-82
20 Nya Mehnwolo Boayue, Samuel Kuttner 5Preclinical iterations PET/SPECT/CT 8 subsets February 10, /11 SPECT subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary SPECT reconstruction optimization (OSEM algorithm) EM Equivalent iterations = iterations * subsets Brambilla et al, Ann Nucl Med Apr;19(2): iterations 8 subsets
21 Nya Mehnwolo Boayue, Samuel Kuttner 5Preclinical iterations PET/SPECT/CT 8 subsets February 10, /11 SPECT subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary SPECT reconstruction optimization (OSEM algorithm) EM Equivalent iterations = iterations * subsets Brambilla et al, Ann Nucl Med Apr;19(2): iterations 8 subsets
22 CT subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary CT Spatial Resolution: Materials and Methods 50µm thick tungsten wire Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
23 CT subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary CT Spatial Resolution: Materials and Methods 50µm thick tungsten wire Kayugawa et al, Journal of Applied Clinical Medical Physics, Volume 14, Number 4, 2013 Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
24 CT subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary CT Spatial Resolution: Materials and Methods 50µm thick tungsten wire Kayugawa et al, Journal of Applied Clinical Medical Physics, Volume 14, Number 4, 2013 QRM SSP phantom: high-contrast platelet (diameter 1mm, thickness 10µm) Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
25 CT subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary CT Spatial Resolution: Materials and Methods 50µm thick tungsten wire Kayugawa et al, Journal of Applied Clinical Medical Physics, Volume 14, Number 4, 2013 QRM SSP phantom: high-contrast platelet (diameter 1mm, thickness 10µm) Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
26 CT subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary CT Spatial Resolution: Materials and Methods 50µm thick tungsten wire Kayugawa et al, Journal of Applied Clinical Medical Physics, Volume 14, Number 4, 2013 QRM SSP phantom: high-contrast platelet (diameter 1mm, thickness 10µm) Favazza et al, Phys. Med. Biol. 60 (2015) Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
27 CT subsystem Motivation PET subsystem SPECT subsystem CT subsystem Summary CT Spatial Resolution High res. settings Magnification 3.5 proj 1024 binning 1 1 voxel size µm 3 Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
28 Summary Summary Motivation PET subsystem SPECT subsystem CT subsystem Summary We have established protocols and analysis methods for preclinical PET/SPECT/CT in agreement with published literature PET: MLEM 3D SPECT: OSEM 5 iterations 8 subsets CT spatial resolution < 120µm manufacturer s tolerance for HR scans Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
29 Summary Summary Motivation PET subsystem SPECT subsystem CT subsystem Summary We have established protocols and analysis methods for preclinical PET/SPECT/CT in agreement with published literature PET: MLEM 3D SPECT: OSEM 5 iterations 8 subsets CT spatial resolution < 120µm manufacturer s tolerance for HR scans Outlook NEMA NU 4 NEMA NU 5? NEMA NU 6? 2030? Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
30 Motivation PET subsystem SPECT subsystem CT subsystem Summary Nya Mehnwolo Boayue, Samuel Kuttner Preclinical PET/SPECT/CT February 10, /11
Simulation and evaluation of a cost-effective high-performance brain PET scanner.
Research Article http://www.alliedacademies.org/biomedical-imaging-and-bioengineering/ Simulation and evaluation of a cost-effective high-performance brain PET scanner. Musa S Musa *, Dilber U Ozsahin,
More informationLSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option
LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option Y. Bercier, Member, IEEE, M. Casey, Member, IEEE, J. Young, Member, IEEE, T. Wheelock, Member, IEEE, T. Gremillion Abstract-- Factors which
More informationPET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner
PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner September, 2017 Results submitted to Physics in Medicine & Biology Negar Omidvari 1, Jorge Cabello 1, Geoffrey Topping
More informationNoise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner
Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner Kisung Lee, Member, IEEE, Paul E. Kinahan, Senior Member, Robert S. Miyaoka, Member, IEEE, Jeffrey A. Fessler,
More informationChiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard.
Chiara Secco PET Performance measurements of the new LSO-Based Whole Body PET/CT Scanner biograph 16 HI-REZ using the NEMA NU 2-2001 Standard. INTRODUCTION Since its introduction, CT has become a fundamental
More informationThe image reconstruction influence in relative measurement in SPECT / CT animal
BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 0-01 (201) 01-09 The image reconstruction influence in relative measurement in SPECT / CT animal S.C.S. Soriano a ; S.A.L. Souza b ; T.Barboza b ; L.V. De Sá
More informationPerformance evaluation of a new highsensitivity time-of-flight clinical PET/CT system
Huo et al. EJNMMI Physics (2018) 5:29 https://doi.org/10.1186/s40658-018-0229-4 EJNMMI Physics ORIGINAL RESEARCH Open Access Performance evaluation of a new highsensitivity time-of-flight clinical PET/CT
More informationSPECT Reconstruction & Filtering
SPECT Reconstruction & Filtering Goals Understand the basics of SPECT Reconstruction Filtered Backprojection Iterative Reconstruction Make informed choices on filter selection and settings Pre vs. Post
More informationImage Quality Assessment of Pixellated Systems
Image Quality Assessment of Pixellated Systems Andreas Goedicke, Herfried Wieczorek, Henrik Botterweck, Wolfgang Eckenbach, Ling Shao, Member, IEEE, Micheal Petrillo, Member, IEEE, Jinghan Ye, and John
More informationT h e P h a n t o m L a b o r a t o r y
T h e P h a n t o m L a b o r a t o r y 1 ECTphan Phantom SMR330 M a n u a l Copyright 2015 WARNING The use of this phantom requires radioactive fill solutions. Only people trained in the safe handling
More informationSimultaneous Reconstruction of the Activity Image and Registration of the CT image in TOF-PET. Ahmadreza Rezaei, Johan Nuyts
Simultaneous Reconstruction of the Activity Image and Registration of the CT image in TOF-PET Ahmadreza Rezaei, Johan Nuyts Activity Reconstruction & Attenuation Registration Attenuation Correction, Background
More informationCurrently, the spatial resolution of most dedicated smallanimal
A Prototype High-Resolution Small-Animal PET Scanner Dedicated to Mouse Brain Imaging Yongfeng Yang 1,2, Julien Bec 1, Jian Zhou 1, Mengxi Zhang 1, Martin S. Judenhofer 1, Xiaowei Bai 1, Kun Di 1, Yibao
More informationHow Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram?
November 2014, Volume 1, Number 4 How Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram? Hojjat Mahani 1,2, Alireza Kamali-Asl 3, *, Mohammad Reza Ay 2, 4 1. Radiation Application
More informationA Skew-Slit Collimator for Small-Animal SPECT
A Skew-Slit Collimator for Small-Animal SPECT Gengsheng L. Zeng Department of Radiology, Utah Center for Advanced Imaging Research (UCAIR), University of Utah, Salt Lake City, Utah The main objective of
More informationInvestigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras
Egyptian J. Nucl. Med., Vol 2, No. 2, Dec. 2009 82 PHYSICS, Original Artical Investigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras Abdelsattar, M.B. Ph.D.; BuHumaid,
More informationPET: New Technologies & Applications, Including Oncology
PET: New Technologies & Applications, Including Oncology, PhD, FIEEE Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Disclosures Research Contract, GE Healthcare
More informationRecovery and normalization of triple coincidences in PET
Universidad Carlos III de Madrid Repositorio institucional e-archivo Área de Imagen e Instrumentación (BiiG) http://e-archivo.uc3m.es DBIAB - BIIG - Journal Articles 2015-03 Recovery and normalization
More informationCombined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8
Combined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8 UNIVERSITY OF CAMBRIDGE 1.2 Rob C. Hawkes 1, Tim D. Fryer 1, Alun J. Lucas 1,2, Stefan B. Siegel 3, Richard
More informationACR Update in Nuclear Medicine Accreditation
Disclaimer ACR Update in Nuclear Medicine Accreditation Beth A. Harkness, MS, DABR, FACR Henry Ford Health System Detroit, MI ACR physics subcommittee for nuclear medicine accreditation. My facility is
More informationCelesteion Time-of-Flight Technology
Celesteion Time-of-Flight Technology Bing Bai, PhD Clinical Sciences Manager, PET/CT Canon Medical Systems USA Introduction Improving the care for every patient while providing a high standard care to
More informationData. microcat +SPECT
Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for
More informationAim. Images for this section: Page 2 of 13
Changes in CT number of high atomic number materials with field of view when using an extended CT number to electron density curve and a metal artifact reduction reconstruction algorithm Poster No.: R-0094
More informationPerformance evaluation of a multipinhole small animal SPECT system
University of Wollongong Research Online Faculty of Engineering - Papers (Archive) Faculty of Engineering and Information Sciences 23 Performance evaluation of a multipinhole small animal SPECT system
More informationThe Point Source. July 2011 Volume 2, Issue 1. Executive message. Rich Fabian. In this issue
The Point Source July 2011 Volume 2, Issue 1 In this issue Executive message from Rich Fabian...1 Handling anxiety...2 Product updates Introducing TruFlight Select PET/CT...3 BrightView XCT Version 2.5...4
More informationFactors Affecting the resolution of SPECT Imaging. h.
Factors Affecting the resolution of SPECT Imaging H. E. Mostafa *1, H. A. Ayoub 2 and Sh.Magraby 1 1 Kasr El-Ini Center for Oncology, Cairo University, 2 Faculty of Science, Suez Canal University hayamayoub@yahoo.com
More informationImplementing analytical geometric and penetration response correction for keel-edge pinhole SPECT image reconstruction
University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2008 Implementing analytical geometric and penetration
More informationChanging the Shape of Nuclear Medicine
TRUTH IN IMAGING Changing the Shape of Nuclear Medicine Multi-Purpose SPECT Scanner Nothing Gets Closer Introducing 360 Body Contour Scanning With 360 degree detector coverage, and unique proximity sensors
More informationComparing planar image quality of rotating slat and parallel hole collimation: influence of system modeling
Comparing planar image quality of rotating slat and parallel hole collimation: influence of system modeling Roel Van Holen, Stefaan Vandenberghe, Steven Staelens and Ignace Lemahieu ELIS Department, MEDISIP,
More informationIndustry Breakthrough
Industry Breakthrough Dynamic SPECT Acquisition Quantifying Myocardial Blood Flow Nuclear Cardiology in the 21st Century In the 21st century, most nuclear cameras are still relying on a technology invented
More informationCopyright American Association of Physicists in Medicine: Medical Physics
Document downloaded from: http://hdl.handle.net/10251/46182 This paper must be cited as: Sánchez Martínez, F.; Orero Palomares, A.; Soriano Asensi, A.; Correcher Salvador, C.; González Martínez, AJ.; Hernández
More informationAbsolute quantification in multi-pinhole micro-spect for different isotopes
Absolute quantification in multi-pinhole micro-spect for different isotopes B. Vandeghinste, Student Member, IEEE, C. Vanhove, J. De Beenhouwer, R. Van Holen, Member, IEEE, S. Vandenberghe, and S. Staelens
More informationFree running mouse brain PET imaging using point source motion tracking
Free running mouse brain PET imaging using point source motion tracking A. Miranda 1, J. Vleugels 2, G. De Bruyne 2, S. Stroobants 1, S. Staelens 1, J. Verhaeghe 1 1 Molecular Imaging Center Antwerp, University
More informationRadionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)
Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in
More informationConceptual Study of Brain Dedicated PET Improving Sensitivity
Original Article PROGRESS in MEDICAL PHYSICS 27(4), Dec. 2016 https://doi.org/10.14316/pmp.2016.27.4.236 pissn 2508-4445, eissn 2508-4453 Conceptual Study of Brain Dedicated PET Improving Sensitivity Han-Back
More informationAquilion Precision Ultra-High Resolution CT: Quantifying diagnostic image quality
Aquilion Precision Ultra-High CT: Quantifying diagnostic image quality Kirsten Boedeker, PhD, DABR Senior Manager, Quantitative Image Quality Canon Medical Systems Corporation Introduction Over the last
More informationC a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information
Ordering Information Please contact us if you have any questions or if you would like a quote or delivery schedule regarding the Catphan phantom. phone 800-525-1190, or 518-692-1190 fax 518-692-3329 mail
More informationQuality control in dual head γ- cameras: Comparison between methods and softwares used for image analysis
Quality control in dual head γ- cameras: Comparison between methods and softwares used for image analysis Abdalrhman Nayl Edam 1, *, Maria Rosa Fornasier 2, Mario de Denaro 2 Abdelmoneim Sulieman 3, Mohammed
More informationInveon. No Limits on Discovery.
Trademarks and service marks used in this material are property of Siemens Medical Solutions USA or Siemens AG. Inveon is a trademark of Siemens AG, its subsidiaries or affiliates. All other company, brand,
More informationarxiv: v1 [physics.med-ph] 29 Nov 2018
Expected performance of the TT-PET scanner E. Ripiccini, a,b,1 D. Hayakawa, a,b G. Iacobucci, a M. Nessi, a,c E. Nowak, c L. Paolozzi, a O. Ratib, b P. Valerio a and D. Vitturini a a University of Geneva,
More informationFocal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution
Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Cynthia H. McCollough, PhD, DABR, FAAPM, FACR Director, CT Clinical Innovation Center Professor of Medical Physics and
More informationTitle. CitationEuropean Journal of Nuclear Medicine and Molecular I. Issue Date Doc URL. Rights. Type. File Information
Title Performance characterization of the Inveon preclinic Author(s)Magota, Keiichi; Kubo, Naoki; Kuge, Yuji; Nishijima, CitationEuropean Journal of Nuclear Medicine and Molecular I Issue Date 2011-04
More informationBreast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School
Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis
More informationCHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY
14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to
More informationNEMA and clinical evaluation of a novel brain PET-CT scanner
Journal of Nuclear Medicine, published on December 23, 2015 as doi:10.2967/jnumed.115.159723 NEMA and clinical evaluation of a novel brain PET-CT scanner Kira S. Grogg 1, Terrence Toole 2, Jinsong Ouyang
More informationDesign of a Static Full-Ring Multi-Pinhole Collimator for Brain SPECT
Design of a Static Full-Ring Multi-Pinhole Collimator for Brain SPECT Karen Van Audenhaege, Student Member, IEEE, Roel Van Holen, Member, IEEE, Karel Deprez, Joel S. Karp, Senior Member, IEEE, Scott Metzler,
More informationof fifty-two camera heads using the PICKER thyroid phantom Université de Liège (ULg), Imagerie médicale expérimentale, Liège, Belgium
Hot and cold contrasts in high resolution Tc-99m planar scintigraphy: a survey of fifty-two camera heads using the PICKER thyroid phantom Alain Seret Université de Liège (ULg), Imagerie médicale expérimentale,
More informationStudy of a Prototype VP-PET Imaging System Based on highly. Pixelated CdZnTe Detectors
Study of a Prototype VP-PET Imaging System Based on highly Pixelated CdZnTe Detectors Zheng-Qian Ye 1, Ying-Guo Li 1, Tian-Quan Wang 1, Ya-Ming Fan 1, Yong-Zhi Yin 1,*, Xi-Meng Chen 1 Affiliations: 1 School
More information180 pinhole-spect with tilted detector and OS-EM. reconstruction: phantom studies and potential
180 pinhole-spect with tilted detector and OS-EM reconstruction: phantom studies and potential clinical applications Alain Seret 1, Michel Defrise 2 and Didier Blocklet 3. 1 Université de Liège (ULg),
More informationCHAPTER 15 DEVICES FOR EVALUATING IMAGING SYSTEMS
DEVICES FOR EVALUATING IMAGING SYSTEMS O. DEMIRKAYA, R. AL-MAZROU Department of Biomedical Physics, King Faisal Specialist Hospital and Research Centre, Riyadh, Saudi Arabia 15.1. DEVELOPING A QUALITY
More informationOptimization and Calibration of Slat Position for a SPECT With Slit-Slat Collimator and Pixelated Detector Crystals
2234 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 58, NO. 5, OCTOBER 2011 Optimization and Calibration of Slat Position for a SPECT With Slit-Slat Collimator and Pixelated Detector Crystals Xiao Deng, Tianyu
More informationPET Performance Measurements for an LSO- Based Combined PET/CT Scanner Using the National Electrical Manufacturers Association NU Standard
PET Performance Measurements for an LSO- Based Combined PET/CT Scanner Using the National Electrical Manufacturers Association NU 2-2001 Standard Yusuf E. Erdi, DSc 1 ; Sadek A. Nehmeh, PhD 1 ; Tim Mulnix,
More informationHIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE
HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE Takeyuki Hashimoto 1), Morio Onoe 2), Hiroshi Nakamura 3), Tamon Inouye 4), Hiromichi Jumonji 5), Iwao Takahashi 6); 1)Yokohama Soei
More informationDesign Studies of A High-Performance Onboard Positron Emission Tomography For Integrated Small Animal PET/CT/RT Radiation Research Systems
Proceedings of the International MultiConference of Engineers and Computer Scientists 2018 Vol II Design Studies of A High-Performance Onboard Positron Emission Tomography For Integrated Small Animal PET/CT/RT
More informationDesign Optimization of a Small-animal SPECT System Using LGSO Continuous Crystals and Micro Parallel-hole Collimators
Journal of the Korean Physical Society, Vol. 67, No. 1, July 2015, pp. 224 231 Design Optimization of a Small-animal SPECT System Using LGSO Continuous Crystals and Micro Parallel-hole Collimators Joong
More informationSources for Siemens Scanners/Cameras
Sources for Siemens Scanners/Cameras Siemens ECAT EXACT 47 - Annual Replacement SIE-ECAT-EXACT47 Ge-68 HEGL-0080-3M, Transmission Rod 3mCi Nominal 3 Siemens ECAT EXACT HR+ - Annual Replacement SIE-ECAT-EXACT-HR+
More informationComparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization
Comparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization Poster No.: C-1040 Congress: ECR 2016 Type: Scientific Exhibit
More informationParameters Affecting on Intrinsic Uniformity Test For MEDISO
ISPUB.COM The Internet Journal of Nuclear Medicine Volume 5 Number 2 Parameters Affecting on Intrinsic Uniformity Test For MEDISO S Zobly, A Osman Citation S Zobly, A Osman. Parameters Affecting on Intrinsic
More informationFirst Results From the High-Resolution mousespect Annular Scintillation Camera
First Results From the High-Resolution mousespect Annular Scintillation Camera Andrew L. Goertzen, Douglas W. Jones, Jurgen Seidel, King Li, and Michael V. Green Abstract High resolution SPECT imaging
More informationImaging with FDG PET is a valuable technique for tumor
Noise Reduction in Oncology FDG PET Images by Iterative Reconstruction: A Quantitative Assessment Cyril Riddell, Richard E. Carson, Jorge A. Carrasquillo, Steven K. Libutti, David N. Danforth, Millie Whatley,
More informationDesign Optimization of a Small-animal SPECT System Using LGSO Continuous. Crystal and a Micro Parallel-hole Collimator
1 Design Optimization of a Small-animal SPECT System Using LGSO Continuous Crystal and a Micro Parallel-hole Collimator 1 Joong Hyun Kim, 2 Mikiko Ito, 2 Soo Mee Kim, 3 Seong Jong Hong, 2,4 Jae Sung Lee,
More informationTechnical Aspects: Image Reconstruction
Annals of Nuclear Cardiology Vol. 2 No. 1 68-72 REVIEW ARTICLE Technical Aspects: Masahisa Onoguchi, RT, PhD 1), Takahiro Konishi, RT, MS 2), Takayuki Shibutani, RT, MS 1), Shinro Matsuo, MD, PhD 3) and
More informationPerformance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET
Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET A. Kuhn, S. Surti, Member, IEEE, J. S. Karp, Senior Member, IEEE, G. Muehllehner, Fellow, IEEE, F.M. Newcomer, R. VanBerg Abstract--
More informationPredicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities
Journal of Physics: Conference Series PAPER OPEN ACCESS Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Recent citations - Resolution Properties of a
More informationUCLA UCLA Previously Published Works
UCLA UCLA Previously Published Works Title Attenuation correction for small animal PET tomographs Permalink https://escholarship.org/uc/item/41n377p3 Journal Physics in Medicine and Biology, 5(8) ISSN
More informationMonte Carlo Simulation Study of a Dual-Plate PET Camera Dedicated to Breast Cancer Imaging
IEEE Nuclear Science Symposium Conference Record M-9 Monte Carlo Simulation Study of a Dual-Plate PET Camera Dedicated to Breast Cancer Imaging Jin Zhang, Member, IEEE, Peter D. Olcott, Member, IEEE, Angela
More informationDevelopment of a Small Animal SPECT System with. a High Purity Germanium Detector. Lindsay Craig Johnson. Dissertation
Development of a Small Animal SPECT System with a High Purity Germanium Detector By Lindsay Craig Johnson Dissertation Submitted to the Faculty of the Graduate School of Vanderbilt University in partial
More informationUsing GATE to understand performance of a full-torso PET scanner
University of Pennsylvania Using GATE to understand performance of a full-torso PET scanner Varsha Viswanath 1 Margaret E. Daube-Witherspoon 1, Matthew E. Werner 1, Suleman Surti 1, Andreia Trindade 2,
More informationIndustry Breakthrough
Industry Breakthrough Dynamic SPECT Acquisition Quantifying Myocardial Blood Flow D-S P EC T Cardiac Imaging System Nuclear Cardiology in the 21st Century In the 21st century, most nuclear cameras are
More informationdiagnostic examination
RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011
More informationPhilips Astonish. Key advantages. including improved image quality and
Philips Key advantages including improved image quality and with AC offers improved image quality, interpretative certainty, diagnostic Alternatively, simplify patient care by exposing patients to reduced
More informationTITLE: Commissioning and Characterization of a Dedicated High-Resolution Breast PET Camera
AWARD NUMBER: W81XWH 10 1 0393 TITLE: Commissioning and Characterization of a Dedicated High-Resolution Breast PET Camera PRINCIPAL INVESTIGATOR: Arne Vandenbroucke, Ph.D. CONTRACTING ORGANIZATION: Stanford
More informationEvaluation of Scatter Fraction and Count Rate Performance of Two Smallanimal PET scanners using dedicated phantoms
2011 IEEE Nuclear Science Symposium Conference Record MIC18.M-36 Evaluation of Scatter Fraction and Count Rate Performance of Two Smallanimal PET scanners using dedicated phantoms Rameshwar Prasad, Student
More informationGafChromic EBT2 and EBT3 Films for Ball Cube II Phantom
GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom Introduction: These EBT2/EBT3 films, shown in Figure 1a-c, are specially sized and formatted to uniquely fit the Accuray Ball Cube II Phantom. Each
More informationLightburst Digital Detector
GE Healthcare Lightburst Digital Detector INTRODUCTION In clinical practice, PET/CT imaging helps clinicians visualize disease at an early stage, before it metastasizes and involves other organs, tissues
More informationPrimer on molecular imaging technology
Primer on molecular imaging technology Craig S. Levin Division of Nuclear Medicine, Department of Radiology and Molecular Imaging Program at Stanford (MIPS), Stanford University School of Medicine, 300
More informationTomophan TSP004 Manual
T h e P h a n t o m L a b o r a t o r y 1 Tomophan TSP004 Manual Copyright 2016 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order and
More informationThis document is published in:
Institutional Repository This document is published in: Bo Yu (ed.) (01). 01 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC): Anaheim, California, USA. October 9 - November 3, 01.
More informationChapter 4 Semiconductor Detector-Based Scanners for Nuclear Medicine
Chapter 4 Semiconductor Detector-Based Scanners for Nuclear Medicine Wataru Takeuchi, Atsuro Suzuki, Yuichiro Ueno, Tohru Shiga, Kenji Hirata, Shozo Okamoto, Songji Zhao, Yuji Kuge, Naoki Kubo, Kentaro
More informationEffect of Post Reconstruction Gaussian Filtering on Image Quality and Myocardial Blood Flow Measurement with N 13 Ammonia PET
Effect of Post Reconstruction Gaussian Filtering on Image Quality and Myocardial Blood Flow Measurement with N 13 Ammonia PET Hyeon Sik Kim 1, Sang Geon Cho 2, Ju Han Kim 3, Seong Young Kwon 1, Byeongil
More informationA positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Huaiqun Guan,
More informationInitial Certification
Initial Certification Nuclear Medical Physics (NMP) Study Guide Part 2 Content Guide and Sample Questions The content of all ABR exams is determined by a panel of experts who select the items based on
More informationResearch Article Feasibility and Initial Performance of Simultaneous SPECT-CT Imaging Using a Commercial Multi-Modality Preclinical Imaging System
Hindawi Publishing Corporation International Journal of Molecular Imaging Volume 2015, Article ID 134768, 11 pages http://dx.doi.org/10.1155/2015/134768 Research Article Feasibility and Initial Performance
More informationImprovement of CT image quality with iterative reconstruction idose4
Improvement of CT image quality with iterative reconstruction idose4 Poster No.: C-0387 Congress: ECR 2014 Type: Scientific Exhibit Authors: M.-L. Olsson, K. Norrgren, M. Söderberg; Malmö/SE Keywords:
More informationPhotomultiplier Tube
Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or
More information2594 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER /$ IEEE
2594 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER 2009 Investigation of Depth of Interaction Encoding for a Pixelated LSO Array With a Single Multi-Channel PMT Yongfeng Yang, Member, IEEE,
More informationGUIDELINE FOR RADIATION PROTECTION AND PERFORMANCE EVALUATION OF PET-CT IMAGING FORUM FOR NUCLEAR COOPERATION IN ASIA (FNCA)
GUIDELINE FOR RADIATION PROTECTION AND PERFORMANCE EVALUATION OF PET-CT IMAGING FORUM FOR NUCLEAR COOPERATION IN ASIA (FNCA) 5 MARCH 2011 1 Contributors Dr. Noriah Jamal - Malaysia Dr. Muhd Noor Muhd Yunus
More information1396 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 5, OCTOBER 2005
1396 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 5, OCTOBER 2005 Performance Amelioration for Small Animal SPECT Using Optimized Pinhole Collimator and Image Correction Technique Tae Yong Song,
More informationAutomated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London
Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients
More informationWideband Focused Transducer Array for Optoacoustic Tomography
1st International Symposium on Laser Ultrasonics: Science, Technology and Applications July 16-18 2008, Montreal, Canada Wideband Focused Transducer Array for Optoacoustic Tomography Varvara A. SIMONOVA
More informationPerformance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET
2005 IEEE Nuclear Science Symposium Conference Record M11-126 Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET Jin Zhang, Member,
More informationHigh-Fidelity Modeling of Shift-Variant Focal-Spot Blur for High-Resolution CT
High-Fidelity Modeling of Shift-Variant Focal-Spot lur for High-Resolution CT Steven Tilley II, Wojciech Zbijewski, J. Webster Stayman bstract Dedicated application-specific CT systems are popular solutions
More informationAttenuation Correction in Hybrid MR-BrainPET Imaging
Mitglied der Helmholtz-Gemeinschaft Attenuation Correction in Hybrid MR-BrainPET Imaging Elena Rota Kops Institute of Neuroscience and Biophysics Medicine Brain Imaging Physics Interactions of 511 kev
More informationMC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE
MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for
More informationSuppression of metal artifacts using image-based monoenergetic DECT imaging
Suppression of metal artifacts using image-based monoenergetic DECT imaging Poster No.: C-0519 Congress: ECR 2011 Type: Scientific Paper Authors: B. Krauss, B. Schmidt, M. Sedlmair, T. Flohr; Forchheim/DE
More informationDose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp
Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division
More informationFeasibility Study of Compton Scattering Enchanced Multiple Pinhole Imager for Nuclear Medicine
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 5, OCTOBER 2003 1609 Feasibility Study of Compton Scattering Enchanced Multiple Pinhole Imager for Nuclear Medicine L. J. Meng, W. L. Rogers, N. H. Clinthorne,
More informationElectronic Noise in CT Detectors: Impact on Image Noise and Artifacts
Medical Physics and Informatics Original Research Duan et al. Electronic Noise in CT Detectors Medical Physics and Informatics Original Research Xinhui Duan 1 Jia Wang 1,2 Shuai Leng 1 ernhard Schmidt
More informationAn innovative detector concept for hybrid 4D-PET/MRI Imaging
Piergiorgio Cerello (INFN - Torino) on behalf of the 4D-MPET* project *4 Dimensions Magnetic compatible module for Positron Emission Tomography INFN Perugia, Pisa, Torino; Polytechnic of Bari; University
More informationImage Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose
Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation
More informationAn EM Reconstruction with Improved Signal-to. to-noise Ratio for Coded Aperture Imaging
An EM Reconstruction with Improved Signal-to to-noise Ratio for Coded Aperture Imaging Cynthia Tozian, PhD UMass Lowell Bristol-Myers Squibb Medical Imaging Young Investigators Symposium April 12, 2006
More information