Image Quality Assessment of Pixellated Systems

Size: px
Start display at page:

Download "Image Quality Assessment of Pixellated Systems"

Transcription

1 Image Quality Assessment of Pixellated Systems Andreas Goedicke, Herfried Wieczorek, Henrik Botterweck, Wolfgang Eckenbach, Ling Shao, Member, IEEE, Micheal Petrillo, Member, IEEE, Jinghan Ye, and John Vesel, Member, IEEE Abstract-- The recent development of pixellated solid-state detectors for SPECT provides new opportunities for better image quality when the impact of energy resolution, spatial response and reconstruction technique is combined. In this investigation we assess the image quality of cadmium zinc telluride (CZT) based solid-state detector SPECT systems by use of Monte Carlo simulation, implemented in GEANT 4. Images are evaluated assuming different spatial and energy resolution, reconstruction algorithms and filtering. Phantoms for NEMA resolution and efficiency measurements, test phantoms for noise properties, and the Jaszczak phantom are used for this study. Monte Carlo results have found to be consistent with measured data and with analytical results. Jaszczak phantom simulations show that the better spatial resolution of pixellated detectors enables better lesion detectability compared to NaI(Tl) detectors. A hypothetical value of E = 3% gives better contrast than the typical E = 9.5% of an Anger camera. On the other hand, there is hardly any difference when energy resolution is improved from E = 5% to E = 3% for a CZT detector. Reconstruction shows a strong influence on SPECT image quality. Appropriate filters in the back-projection, and statistical reconstruction give better small lesion detectability and higher contrast-to-noise ratio, though peak-to-valley ratios and spatial resolution are reduced. P I. INTRODUCTION ixellated detectors, based on semiconductor detector materials like cadmium zinc telluride (CZT), will provide new opportunities to enhance the image quality of SPECT systems [1]-[6]. Their performance, expressed in NEMA defined parameters like system planar efficiency or spatial resolution, depends strongly on collimator design, detector energy resolution, and image reconstruction method. Those NEMA parameters, as well as signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR), are described by an analytical model that we developed recently [7]. Additionally, attenuation and scatter in the object, penetration of collimator septa, and reconstruction effects have to be considered. For Manuscript received October 4, Andreas Goedicke, Herfried Wieczorek, Henrik Botterweck and Wolfgang Eckenbach are with Philips Research Laboratories, Aachen, Germany (telephone: , Ling Shao, Micheal Petrillo, and Jinghan Ye are with Philips Medical Systems, Milpitas, CA 95035, USA (telephone: , ling.shao@philips.com). John Vesel is with Philips Medical Systems, Cleveland, OH 44143, USA (telephone: , john.vesel@philips.com). that reason we have set up a Monte Carlo (MC) simulation, calibrated it against analytical model and measurement results, assessed image quality, mainly on the Jaszczak phantom, and evaluated the impact of reconstruction methods. II. METHODS 1) Simulation environment GEANT4 is a software toolkit developed for simulating the passage of particles through material objects. It was developed at CERN and intended for detector design in high-energy physics, but due to its extension to low energy particle physics it is widely used in many nuclear physics related areas like nuclear medicine and bio-physics. The problem with SPECT calculations is the low collimator efficiency that requires a large number of primary events to be simulated. We used various optimization techniques to speed up simulation: Pre-allocated dynamically allocated memory, no parameterized volumes, energy cut for low-energy particles, restricted angles of incidence, and scatter and penetration background parameterization for large phantoms. 2) Simulation setup Simulation has been done for Tc-99m, kev radiation. A 150 mm diameter, 10 mm high water disk inside an acrylic disk holder was employed for system planar efficiency, and a line source of 150 mm length, slightly tilted with respect to one detector axis, was used to determine system planar spatial resolution. Phantom-collimator distance was 100 mm. A Jaszczak Phantom with cold rods of mm diameter inside an acrylic cylinder, half-filled with 740 MBq (20 mci) activity of Tc-99m in water, was implemented to visualize SPECT image quality. Projection images were taken under 128 different SPECT angles with 200 mm rotational radius at the collimator surface and 15 min total imaging time (Fig. 1). Two types of detectors, a NaI camera with VXGP collimator and CZT detectors with parallel-hole (PH) collimator, 5 mm CZT thickness, 2.46 mm pixel pitch, 0.35 mm septa thickness, and 40 mm or 45 mm septa length were used. For each detector we employed different energy resolution and different reconstruction methods to separate the influence of the essential parameters: spatial resolution, energy resolution, and reconstruction method, including different filters.

2 III. SIMULATION RESULTS A. NEMA system parameters Table 1 shows a comparison of MC simulated values with measurement results for Anger cameras with VXGP collimator, and with analytical values for a CZT detector with parallel-hole collimator of 45 mm septa length. TABLE 1 NEMA PARAMETER SIMULATION RESULTS Efficiency (cps / MBq) Spatial resolution (mm) NaI-VXGP, MC (measured) ( 80) ( 7.04) CZT-PH, MC (analytical) (74) (6.80) We see that Monte Carlo results compare well with the best values measured on real collimators and with analytical model calculations. The slight difference in numbers for the CZT detector is explained by penetration in the collimator septa which had not been included in the analytical model. Effects like limited absorption and fluorescence loss in the NaI or CZT detector, and an assumed electrical efficiency of 89% for the CZT detector, had been included in the model. B. Jaszczak Phantom Simulation of the cold rod section in the Jaszczak phantom was done to evaluate the impact of detector and reconstruction parameters on the image quality of a strongly scattering phantom. 1) Impact of spatial resolution Fig. 2 shows Jaszczak images obtained for a NaI-VXGP camera and two different CZT-PH detectors. Energy resolution E = 5% was assumed for all detectors. For reconstruction we used filtered back-projection (FBP) with Ram-Lak filter. We took the average of 40 transaxial slices to account for the high noise level in the images, caused by the relatively low dose that we used in the simulation. Contrast and detectability of small lesions, especially in the fifth and sixth sector, is increased for the CZT-PH detector with 45 mm septa length (center image) in comparison to a NaI-VXGP camera (left image) though the difference in spatial resolution is only 6.85 mm vs. 7.0 mm. The dependence may be amplified by depth-of-interaction effects in the thicker NaI layer that were not accounted for in the analytical model. When lower spatial resolution is simulated, 7.45 mm for a CZT-PH detector with 40 mm septa length (right image), image quality is worse than for the Anger camera. The impact of spatial resolution on contrast and detectability of small structures agrees well with previously published results [8]. Fig. 1: Geometry of Jaszczak Phantom and VXGP collimator Fig. 2: Jaszczak phantom images of a NaI-VXGP camera and two CZT-PH detectors with 45 and 40 mm septa height, respectively (from left to right). An energy resolution of 5% was assumed for all detectors. Reconstruction was done by filtered back-projection with Ram-Lak filter; 40 slices were averaged.

3 2) Impact of energy resolution Fig. 3 shows the influence of detector energy resolution on image quality. A theoretical improvement in energy resolution from E = 9.5% to E = 3% is accompanied by a slight increase in lesion detectability in the upper three sectors and of contrast on the whole phantom area (left and center image). There is also some difference seen when the two images are compared with the one simulated with E = 5% (Fig. 2, left image). We find, however, no statistically significant increase in the peak-to-valley ratio for the first sector. This may be due to the relatively low detector dose that we used in the simulation. The slightly better spatial resolution of the CZT detector, compared to a NaI camera, is expressed again in the better lesion detectability in the upper three sectors (right image). A comparison between E = 5% (Fig. 2, center image) and E = 3% (Fig. 3, right image) for the CZT detector, however, shows hardly any difference. 3) Impact of reconstruction method Reconstruction has a major influence on image quality and lesion detectability. We tried FBP with Ram-Lak, Shepp- Logan, Cosine, Hamming or Hann filter, and statistical methods, Maximum Likelihood Expectation Maximization algorithm (MLEM) and Ordered Subset Expectation Maximization (OSEM). Fig. 4 shows three examples. Fine details are preserved but noise is high for FBP with Ram-Lak filter while application of high frequency filters smoothes all structures but increases lesion detectability (left and center image). For the innermost rods in the first sector we found a 2-3 higher contrast-tonoise ratio (CNR) accompanied by only 20% reduction in the peak-to-valley ratio. Detectability of small lesions is better increased by application of OSEM (right image). Spatial resolution is high, but care has to be taken since the peak-tovalley ratio is reduced by 40%, compared to FBP with Ram- Lak filter. Fig. 3: Jaszczak phantom images of NaI-VXGP cameras with E = 9.5% and E = 3%, and a CZT-PH detector, 45 mm septa length, with E = 3% (from left to right). Reconstruction was done by filtered back-projection with Ram-Lak filter; average of 40 slices. Fig. 4: Jaszczak phantom images of a CZT-PH detector, 45 mm septa length, E = 5%, reconstruction by filtered back-projection with Ram-Lak filter, Hann filter, and by 5-pass OSEM (from left to right); average of 40 slices.

4 C. Image quality test phantoms We used planar hot disk phantoms [9] to investigate noise propagation in the image reconstruction and compared the resulting image quality with SNR values derived from theory. Planar disks with 40 mm diameter were simulated, imaged by a rotating slat detector of 2.46 mm pixel pitch at 256 different 7 angles, with a total number of decays per voxel. The image was reconstructed from the resulting line integrals by FBP, MLEM, OSEM, and Algebraic Reconstruction Technique (ART). This reconstruction of planar images from line integrals is equivalent to the reconstruction of transaxial slices in SPECT, except for attenuation in three-dimensional phantoms [7]. 1) Impact of reconstruction on image noise Fig. 5 shows the impact of different filters, used in FBP, on noise in the reconstructed image. The Ram-Lak filter is known to preserve fine details as well as noise, while high frequency filters like Cosine, Hamming and Hann filters are seen to smooth shot noise in the image considerably. A similar noise reduction is seen when reconstruction is done by 16-pass MLEM (Fig. 6). The same result is achieved much faster by use of 1-pass OSEM with 16 subsets. Repeated application of OSEM, however, results in an increasingly noisy image. ART, a non-statistical iterative method, causes even higher noise. TABLE 2 SNR VALUES OF HOT DISK PHANTOM SNR lin SNR near SNR th ξ A(ξ ) Q Ram-Lak ξ 0.82 Shepp- Logan Cosine πξ ξ sin c 0.5 ξ N πξ ξ cos 0.16 ξ N 2πξ Hamming ξ cos ξ N 2πξ Hann ξ cos ξ N OSEM OSEM OSEM ART 9.4 Fig. 5: Hot disk phantoms, reconstruction by FBP with Ram-Lak, Cosine, Hamming and Hann filter (from upper left to lower right). Fig. 6: Reconstruction by 16-pass MLEM, 2-pass and and 8-pass OSEM with 16 subsets, and ART (from upper left to lower right). Table 2 shows SNR values determined from a 52-pixel region of interest in the center of the hot disk, calculated for different reconstruction methods. SNR lin refers to Fig. 5 and 6, using linear interpolation in FBP, while SNR near uses nearest neighbor values and SNR th gives theory values [10]. The FBP filter function in Fourier space is defined by the variable ξ, normalized by Nyquist frequency ξ N, and commonly known filter functions A (ξ ). We define the noise reduction factor Q: Q = π ξ A( ξ) dξ. (1) The squared signal-to-noise ratio is given by p SNR 2 = AvTE (2) 2rQ where Av is the activity per voxel, T the imaging time, E the detector efficiency and 2 r / p the disk diameter divided by the pixel pitch [7]. We see from Table 2 that this noise theory describes FBP with nearest neighbor interpolation very well, and for statistical reconstruction we get comparable SNR.

5 2) Impact of reconstruction on spatial resolution From the different spacing of grid lines in Fig. 5 and 6 we see that spatial resolution is reduced by filtering but higher for statistical reconstruction. We used line phantom simulations for further evaluation. Spatial resolution, calculated as the FWHM of the line spread function (LSF), is calculated from all lines in the image perpendicular to the line source (Fig. 7). Fig. 7: Line phantom image, reconstructed by FBP with Ram-Lak filter, and line spread function calculated from the image data. Fig. 8 shows the correlation of SNR values from Table 2 with the FWHM of LSF. Comparing different filters in FBP, the higher SNR achieved by filtering is accompanied by slightly lower spatial resolution. Repeated application of OSEM gives lower SNR values, shown for 1-pass up to 5-pass iteration, while spatial resolution is increased, even beyond that of a parallel-hole detector applied without reconstruction. S / N ratio OSEM parallel hole 5x 3x 2x FBP Shepp-Logan Ram-Lak 1x Cosine Hann Hamming 0 6,5 7,0 7,5 8,0 8,5 9,0 9,5 spatial resolution / mm Fig. 8: Signal-to-noise ratio, measured on disk phantoms, vs. spatial resolution, measured on line phantoms, for different reconstruction methods. The value for a parallel-hole detector is given for comparison We have seen, however, that convergence properties of the OSEM reconstruction depend strongly on the object simulated so that we cannot generalize the correlation shown in Fig. 8. IV. CONCLUSIONS We have set up a Monte Carlo simulator for the evaluation of new SPECT detector concepts. As a first test, NEMA system parameters were simulated and shown to be well comparable to analytical results and best values measured on SPECT systems. Different test phantoms for signal-to-noise ratio and for spatial resolution, and the Jaszczak phantom were simulated to evaluate the impact of detection concept, energy resolution, and reconstruction method on image quality. Contrast and the detectability of small lesions are enhanced by a small improvement in spatial resolution. There seems to be a further advantage of pixellated systems in comparison to Anger cameras that is not yet explained. An improvement in energy resolution from 9.5% to a theoretical value of 3% for Anger cameras causes an increase in lesion detectability, but for pixellated detectors we hardly see any difference for energy resolution values of 3% and 5%. Results are, however, statistically not significant. High frequency filters in the back-projection as well as the application of statistical reconstruction increase the signal-tonoise ratio in reconstructed images, which is well understood. Contrast and lesion detectability are enhanced, but the peak-tovalley ratio in the Jaszczak phantom images is reduced and spatial resolution is diminished. A further evaluation will use forced detection in the Monte Carlo code to enable simulation of a much higher phantom or patient dose. This is necessary to get statistically significant values of image quality parameters and evaluate the in depth impact of reconstruction methods on lesion detectability. V. REFERENCES [1] J.F. Butler, C.L. Lingren, S.J. Friesenhahn, W.L. Ashburn, R.L. Conwell, F.L. Augustine, et al, CdZnTe solid-state gamma camera, IEEE Trans. Nucl. Sci., vol. 45, pp , June [2] R. Amrami, G. Shani, Y. Hefetz, I. Blevis, A. Pansky, A Comparison Between The Performance Of a Pixellated CdZnTe Based Gamma Camera and Anger NaI(Tl) Scintillator Gamma Camera, Proceedings of the 22 nd Annual EMBS International Conference, July 23-28, 2000, Chicago, IL pp [3] Y. Eisen, I. Mardor, A. Shor, Z. Baum, D. Bar, G. Feldman, et al., NUCAM3 A Gamma Camera Based on Segmented Monolithic CdZnTe Detectors, IEEE TNS, vol. 49, No 4, pp , [4] D.J. Wagenaar, S. Chowdhury, J.C. Engdahl, D.D. Burckhardt, Planar image quality comparison between a CdZnTe prototype and a standard NaI(Tl) gamma camera, Nuclear Instruments and Methods in Physics Research, vol. A 505, pp , [5] B. Mueller, M. O Connor, I. Blevis, D. Rhodes, R. Smith, D. Collins, et al, Evaluation of a small CZT detector for scintimammography, JNM, Vol 44, No 4, pp , April 2003 [6] M. Petrillo, J. Ye, J. Vesel, L. Shao, H. Wieczorek, and A. Goedicke, "Imaging performance of tiled solid-state detectors", presented at IEEE MIC 2004, Rome, Oct [7] H. Wieczorek, A. Goedicke, L. Shao, M. Petrillo, J. Ye, and J. Vesel, "Analytical model for pixellated SPECT detector concepts", presented at IEEE MIC 2004, Rome, Oct [8] G. Muehllehner, "Effect of resolution improvement on required count density in ECT imaging: a computer simulation", Phys. Med. Biol., vol. 30, pp , 1985 [9] T. F. Budinger, S. E. Derenzo, G. T. Gullberg, W. L. Greenberg, and R. H. Huesmann, "Emission computer assisted tomography with singlephoton and positron annihilation photon emitters",, J. Comput. Assist. Tomogr., vol. 1, pp , 1977 [10] M. A. Lodge, D. M. Binnie, M. A. Flower, and S. Webb, "Experimental evaluation of a prototype rotating slat collimator for planar gamma camera imaging", Phys. Med. Biol., vol. 40, pp , 1995.

Simulation and evaluation of a cost-effective high-performance brain PET scanner.

Simulation and evaluation of a cost-effective high-performance brain PET scanner. Research Article http://www.alliedacademies.org/biomedical-imaging-and-bioengineering/ Simulation and evaluation of a cost-effective high-performance brain PET scanner. Musa S Musa *, Dilber U Ozsahin,

More information

Reconstruction Filtering in Industrial gamma-ray CT Application

Reconstruction Filtering in Industrial gamma-ray CT Application Reconstruction Filtering in Industrial gamma-ray CT Application Lakshminarayana Yenumula *, Rajesh V Acharya, Umesh Kumar, and Ashutosh Dash Industrial Tomography and Instrumentation Section, Isotope Production

More information

LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option

LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option Y. Bercier, Member, IEEE, M. Casey, Member, IEEE, J. Young, Member, IEEE, T. Wheelock, Member, IEEE, T. Gremillion Abstract-- Factors which

More information

Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET

Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET A. Kuhn, S. Surti, Member, IEEE, J. S. Karp, Senior Member, IEEE, G. Muehllehner, Fellow, IEEE, F.M. Newcomer, R. VanBerg Abstract--

More information

NM Module Section 2 6 th Edition Christian, Ch. 3

NM Module Section 2 6 th Edition Christian, Ch. 3 NM 4303 Module Section 2 6 th Edition Christian, Ch. 3 Gas Filled Chamber Voltage Gas filled chamber uses Hand held detectors cutie pie Geiger counter Dose calibrators Cutie pie Chamber voltage in Ionization

More information

Design of a Static Full-Ring Multi-Pinhole Collimator for Brain SPECT

Design of a Static Full-Ring Multi-Pinhole Collimator for Brain SPECT Design of a Static Full-Ring Multi-Pinhole Collimator for Brain SPECT Karen Van Audenhaege, Student Member, IEEE, Roel Van Holen, Member, IEEE, Karel Deprez, Joel S. Karp, Senior Member, IEEE, Scott Metzler,

More information

New Technology in Nuclear Medicine

New Technology in Nuclear Medicine New Technology in Nuclear Medicine Reed G. Selwyn, PhD, DABR Vice Chair of Research & Imaging Sciences Associate Professor and Chief, Medical Physics Dept. of Radiology, University of New Mexico Objectives

More information

... In vivo imaging in Nuclear Medicine. 1957: Anger camera (X;Y) X Y

... In vivo imaging in Nuclear Medicine. 1957: Anger camera (X;Y) X Y József Varga, PhD EMISSION IMAGING BASICS OF QUANTIFICATION Imaging devices Aims of image processing Reconstruction University of Debrecen Department of Nuclear Medicine. In vivo imaging in Nuclear Medicine

More information

How Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram?

How Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram? November 2014, Volume 1, Number 4 How Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram? Hojjat Mahani 1,2, Alireza Kamali-Asl 3, *, Mohammad Reza Ay 2, 4 1. Radiation Application

More information

Pinhole collimator design for nuclear survey system

Pinhole collimator design for nuclear survey system Annals of Nuclear Energy 29 (2002) 2029 2040 www.elsevier.com/locate/anucene Pinhole collimator design for nuclear survey system Wanno Lee*, Gyuseong Cho Department of Nuclear Engineering, Korea Advanced

More information

CHAPTER 8 GENERIC PERFORMANCE MEASURES

CHAPTER 8 GENERIC PERFORMANCE MEASURES GENERIC PERFORMANCE MEASURES M.E. DAUBE-WITHERSPOON Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, United States of America 8.1. INTRINSIC AND EXTRINSIC MEASURES 8.1.1.

More information

Investigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras

Investigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras Egyptian J. Nucl. Med., Vol 2, No. 2, Dec. 2009 82 PHYSICS, Original Artical Investigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras Abdelsattar, M.B. Ph.D.; BuHumaid,

More information

Comparing planar image quality of rotating slat and parallel hole collimation: influence of system modeling

Comparing planar image quality of rotating slat and parallel hole collimation: influence of system modeling Comparing planar image quality of rotating slat and parallel hole collimation: influence of system modeling Roel Van Holen, Stefaan Vandenberghe, Steven Staelens and Ignace Lemahieu ELIS Department, MEDISIP,

More information

SPECT Reconstruction & Filtering

SPECT Reconstruction & Filtering SPECT Reconstruction & Filtering Goals Understand the basics of SPECT Reconstruction Filtered Backprojection Iterative Reconstruction Make informed choices on filter selection and settings Pre vs. Post

More information

The quality of scintimammographic images obtained. Evaluation of a Small Cadmium Zinc Telluride Detector for Scintimammography

The quality of scintimammographic images obtained. Evaluation of a Small Cadmium Zinc Telluride Detector for Scintimammography Evaluation of a Small Cadmium Zinc Telluride Detector for Scintimammography Bryon Mueller, PhD 1 ; Michael K. O Connor, PhD 1 ; Ira Blevis, PhD 2 ; Deborah J. Rhodes, MD 3 ; Robin Smith, MD 3 ; Douglas

More information

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET 2005 IEEE Nuclear Science Symposium Conference Record M11-126 Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET Jin Zhang, Member,

More information

Study of a Prototype VP-PET Imaging System Based on highly. Pixelated CdZnTe Detectors

Study of a Prototype VP-PET Imaging System Based on highly. Pixelated CdZnTe Detectors Study of a Prototype VP-PET Imaging System Based on highly Pixelated CdZnTe Detectors Zheng-Qian Ye 1, Ying-Guo Li 1, Tian-Quan Wang 1, Ya-Ming Fan 1, Yong-Zhi Yin 1,*, Xi-Meng Chen 1 Affiliations: 1 School

More information

Design Optimization of a Small-animal SPECT System Using LGSO Continuous Crystals and Micro Parallel-hole Collimators

Design Optimization of a Small-animal SPECT System Using LGSO Continuous Crystals and Micro Parallel-hole Collimators Journal of the Korean Physical Society, Vol. 67, No. 1, July 2015, pp. 224 231 Design Optimization of a Small-animal SPECT System Using LGSO Continuous Crystals and Micro Parallel-hole Collimators Joong

More information

Celesteion Time-of-Flight Technology

Celesteion Time-of-Flight Technology Celesteion Time-of-Flight Technology Bing Bai, PhD Clinical Sciences Manager, PET/CT Canon Medical Systems USA Introduction Improving the care for every patient while providing a high standard care to

More information

Optimizing Pinhole and Parallel Hole Collimation for Scintimammography With Compact Pixellated Detectors

Optimizing Pinhole and Parallel Hole Collimation for Scintimammography With Compact Pixellated Detectors IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 3, JUNE 2003 321 Optimizing Pinhole and Parallel Hole Collimation for Scintimammography With Compact Pixellated Detectors Mark F. Smith, Member, IEEE,

More information

First Applications of the YAPPET Small Animal Scanner

First Applications of the YAPPET Small Animal Scanner First Applications of the YAPPET Small Animal Scanner Guido Zavattini Università di Ferrara CALOR2 Congress, Annecy - FRANCE YAP-PET scanner Scintillator: YAP:Ce Size: matrix of 2x2 match like crystals

More information

Design Optimization of a Small-animal SPECT System Using LGSO Continuous. Crystal and a Micro Parallel-hole Collimator

Design Optimization of a Small-animal SPECT System Using LGSO Continuous. Crystal and a Micro Parallel-hole Collimator 1 Design Optimization of a Small-animal SPECT System Using LGSO Continuous Crystal and a Micro Parallel-hole Collimator 1 Joong Hyun Kim, 2 Mikiko Ito, 2 Soo Mee Kim, 3 Seong Jong Hong, 2,4 Jae Sung Lee,

More information

LaBr 3 :Ce, the latest crystal for nuclear medicine

LaBr 3 :Ce, the latest crystal for nuclear medicine 10th Topical Seminar on Innovative Particle and Radiation Detectors 1-5 October 2006 Siena, Italy LaBr 3 :Ce, the latest crystal for nuclear medicine Roberto Pani On behalf of SCINTIRAD Collaboration INFN

More information

Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner

Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner Kisung Lee, Member, IEEE, Paul E. Kinahan, Senior Member, Robert S. Miyaoka, Member, IEEE, Jeffrey A. Fessler,

More information

Performance Characteristics of a State of the Art Preclinical PET/SPECT/CT Scanner

Performance Characteristics of a State of the Art Preclinical PET/SPECT/CT Scanner Performance Characteristics of a State of the Art Preclinical PET/SPECT/CT Scanner Nya Mehnwolo Boayue 1 Samuel Kuttner 1 1 Center for Diagnostic Physics University Hospital of North-Norway Medfys, 2016

More information

CZT Technology: Fundamentals and Applications

CZT Technology: Fundamentals and Applications GE Healthcare CZT Technology: Fundamentals and Applications White Paper Abstract Nuclear Medicine traces its technology roots to the 1950 s, and while it has continued to evolve since the invention of

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

Conceptual Study of Brain Dedicated PET Improving Sensitivity

Conceptual Study of Brain Dedicated PET Improving Sensitivity Original Article PROGRESS in MEDICAL PHYSICS 27(4), Dec. 2016 https://doi.org/10.14316/pmp.2016.27.4.236 pissn 2508-4445, eissn 2508-4453 Conceptual Study of Brain Dedicated PET Improving Sensitivity Han-Back

More information

Optimization and Calibration of Slat Position for a SPECT With Slit-Slat Collimator and Pixelated Detector Crystals

Optimization and Calibration of Slat Position for a SPECT With Slit-Slat Collimator and Pixelated Detector Crystals 2234 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 58, NO. 5, OCTOBER 2011 Optimization and Calibration of Slat Position for a SPECT With Slit-Slat Collimator and Pixelated Detector Crystals Xiao Deng, Tianyu

More information

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance

More information

Implementing analytical geometric and penetration response correction for keel-edge pinhole SPECT image reconstruction

Implementing analytical geometric and penetration response correction for keel-edge pinhole SPECT image reconstruction University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2008 Implementing analytical geometric and penetration

More information

Marten Bosma 1, Alex Fauler 2, Michael Fiederle 2 en Jan Visser Nikhef, Amsterdam, The Netherlands 2. FMF, Freiburg, Germany

Marten Bosma 1, Alex Fauler 2, Michael Fiederle 2 en Jan Visser Nikhef, Amsterdam, The Netherlands 2. FMF, Freiburg, Germany Marten Bosma 1, Alex Fauler 2, Michael Fiederle 2 en Jan Visser 1 1. Nikhef, Amsterdam, The Netherlands 2. FMF, Freiburg, Germany Digital Screen film Digital radiography advantages: Larger dynamic range

More information

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard.

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard. Chiara Secco PET Performance measurements of the new LSO-Based Whole Body PET/CT Scanner biograph 16 HI-REZ using the NEMA NU 2-2001 Standard. INTRODUCTION Since its introduction, CT has become a fundamental

More information

Factors Affecting the resolution of SPECT Imaging. h.

Factors Affecting the resolution of SPECT Imaging. h. Factors Affecting the resolution of SPECT Imaging H. E. Mostafa *1, H. A. Ayoub 2 and Sh.Magraby 1 1 Kasr El-Ini Center for Oncology, Cairo University, 2 Faculty of Science, Suez Canal University hayamayoub@yahoo.com

More information

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for

More information

Feasibility Study of Compton Scattering Enchanced Multiple Pinhole Imager for Nuclear Medicine

Feasibility Study of Compton Scattering Enchanced Multiple Pinhole Imager for Nuclear Medicine IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 5, OCTOBER 2003 1609 Feasibility Study of Compton Scattering Enchanced Multiple Pinhole Imager for Nuclear Medicine L. J. Meng, W. L. Rogers, N. H. Clinthorne,

More information

T h e P h a n t o m L a b o r a t o r y

T h e P h a n t o m L a b o r a t o r y T h e P h a n t o m L a b o r a t o r y 1 ECTphan Phantom SMR330 M a n u a l Copyright 2015 WARNING The use of this phantom requires radioactive fill solutions. Only people trained in the safe handling

More information

Primer on molecular imaging technology

Primer on molecular imaging technology Primer on molecular imaging technology Craig S. Levin Division of Nuclear Medicine, Department of Radiology and Molecular Imaging Program at Stanford (MIPS), Stanford University School of Medicine, 300

More information

A Skew-Slit Collimator for Small-Animal SPECT

A Skew-Slit Collimator for Small-Animal SPECT A Skew-Slit Collimator for Small-Animal SPECT Gengsheng L. Zeng Department of Radiology, Utah Center for Advanced Imaging Research (UCAIR), University of Utah, Salt Lake City, Utah The main objective of

More information

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT) Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in

More information

236 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 1, FEBRUARY 2012

236 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 1, FEBRUARY 2012 236 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 1, FEBRUARY 2012 Characterization of the H3D ASIC Readout System and 6.0 cm 3-D Position Sensitive CdZnTe Detectors Feng Zhang, Cedric Herman, Zhong

More information

Parameters Affecting on Intrinsic Uniformity Test For MEDISO

Parameters Affecting on Intrinsic Uniformity Test For MEDISO ISPUB.COM The Internet Journal of Nuclear Medicine Volume 5 Number 2 Parameters Affecting on Intrinsic Uniformity Test For MEDISO S Zobly, A Osman Citation S Zobly, A Osman. Parameters Affecting on Intrinsic

More information

LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging

LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging 8th International Workshop on Radiation Imaging Detectors Pisa 2-6 July 2006 LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging Roberto Pani On behalf of SCINTIRAD Collaboration

More information

Changing the Shape of Nuclear Medicine

Changing the Shape of Nuclear Medicine TRUTH IN IMAGING Changing the Shape of Nuclear Medicine Multi-Purpose SPECT Scanner Nothing Gets Closer Introducing 360 Body Contour Scanning With 360 degree detector coverage, and unique proximity sensors

More information

A Compact 16-Module Camera Using 64-Pixel CsI(Tl)/Si PIN Photodiode Imaging Modules

A Compact 16-Module Camera Using 64-Pixel CsI(Tl)/Si PIN Photodiode Imaging Modules 1 A Compact 16-Module Camera Using 64-Pixel CsI(Tl)/Si PIN Photodiode Imaging Modules W.-S. Choong, Member, IEEE, G. J. Gruber, W. W. Moses, Senior Member, IEEE, S. E. Derenzo, Fellow, IEEE, S. E. Holland,

More information

Fundamentals of Positron Emission Tomography (PET)

Fundamentals of Positron Emission Tomography (PET) Fundamentals of Positron Emission Tomography (PET) NPRE 435, Principles of Imaging with Ionizing Radiation, Fall 2017 Content Fundamentals of PET Camera & Detector Design Real World Considerations Performance

More information

Master of Science Thesis. SIMIND Based Pinhole Imaging

Master of Science Thesis. SIMIND Based Pinhole Imaging Master of Science Thesis SIMIND Based Pinhole Imaging * Development and Validation Kurt Sundin Supervisor: Michael Ljungberg, PhD Medical Radiation Physics Clinical Sciences, Lund Lund University, 2006

More information

NON-UNIFORM ATTENUATION CORRECTION USING SIMULTANEOUS TRANSMISSION AND EMISSION CONVERGING TOMOGRAPHY

NON-UNIFORM ATTENUATION CORRECTION USING SIMULTANEOUS TRANSMISSION AND EMISSION CONVERGING TOMOGRAPHY 1134 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 39, NO. 4,1992 NON-UNIFORM ATTENUATION CORRECTION USING SIMULTANEOUS TRANSMISSION AND EMISSION CONVERGING TOMOGRAPHY C-H Tung, G. T. Gullberg, G. L. Zeng,

More information

Monte Carlo study on a new concept of a scanning photon beam system for IMRT

Monte Carlo study on a new concept of a scanning photon beam system for IMRT NUKLEONIKA 2011;56(4):291 297 ORIGINAL PAPER Monte Carlo study on a new concept of a scanning photon beam system for IMRT Anna M. Wysocka-Rabin, Günter H. Hartmann Abstract. Intensity-modulated radiation

More information

Charge Sharing Effect on 600 µm Pitch Pixelated CZT Detector for Imaging Applications *

Charge Sharing Effect on 600 µm Pitch Pixelated CZT Detector for Imaging Applications * Charge Sharing Effect on 600 µm Pitch Pixelated CZT Detector for Imaging Applications * Yin Yong-Zhi( 尹永智 ), Liu Qi( 刘奇 ), Xu Da-Peng( 徐大鹏 ), Chen Xi-Meng( 陈熙萌 ) School of Nuclear Science and Technology,

More information

Monte Carlo Simulation Study of a Dual-Plate PET Camera Dedicated to Breast Cancer Imaging

Monte Carlo Simulation Study of a Dual-Plate PET Camera Dedicated to Breast Cancer Imaging IEEE Nuclear Science Symposium Conference Record M-9 Monte Carlo Simulation Study of a Dual-Plate PET Camera Dedicated to Breast Cancer Imaging Jin Zhang, Member, IEEE, Peter D. Olcott, Member, IEEE, Angela

More information

Development of the LBNL Positron Emission Mammography Camera

Development of the LBNL Positron Emission Mammography Camera Development of the LBNL Positron Emission Mammography Camera J.S. Huber, Member, IEEE, W.S. Choong, Member, IEEE, J. Wang, Member, IEEE, J.S. Maltz, Member, IEEE, J. Qi, Member, IEEE, E. Mandelli, Member,

More information

Initial Certification

Initial Certification Initial Certification Nuclear Medical Physics (NMP) Study Guide Part 2 Content Guide and Sample Questions The content of all ABR exams is determined by a panel of experts who select the items based on

More information

PET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002

PET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002 PET Detectors William W. Moses Lawrence Berkeley National Laboratory March 26, 2002 Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes

More information

PHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002

PHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002 PHYSICS 334 - ADVANCED LABORATORY I COMPTON SCATTERING Spring 00 Purposes: Demonstrate the phenomena associated with Compton scattering and the Klein-Nishina formula. Determine the mass of the electron.

More information

CdZnTe strip detector SPECT imaging with a slit collimator

CdZnTe strip detector SPECT imaging with a slit collimator Home Search Collections Journals About Contact us My IOPscience CdZnTe strip detector SPECT imaging with a slit collimator This content has been downloaded from IOPscience. Please scroll down to see the

More information

Technical Aspects: Image Reconstruction

Technical Aspects: Image Reconstruction Annals of Nuclear Cardiology Vol. 2 No. 1 68-72 REVIEW ARTICLE Technical Aspects: Masahisa Onoguchi, RT, PhD 1), Takahiro Konishi, RT, MS 2), Takayuki Shibutani, RT, MS 1), Shinro Matsuo, MD, PhD 3) and

More information

Detector technology challenges for nuclear medicine and PET

Detector technology challenges for nuclear medicine and PET Nuclear Instruments and Methods in Physics Research A 513 (2003) 1 7 Detector technology challenges for nuclear medicine and PET Paul K. Marsden Guy s and St. Thomas Clinical PET Centre, King s College

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center. Thursday September 8, 2011

Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center. Thursday September 8, 2011 Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center Thursday September 8, 2011 Quality Control Quality control should be performed because:

More information

A new operative gamma camera for Sentinel Lymph Node procedure

A new operative gamma camera for Sentinel Lymph Node procedure A new operative gamma camera for Sentinel Lymph Node procedure A physicist device for physicians Samuel Salvador, Virgile Bekaert, Carole Mathelin and Jean-Louis Guyonnet 12/06/2007 e-mail: samuel.salvador@ires.in2p3.fr

More information

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud University of Groningen Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you

More information

The image reconstruction influence in relative measurement in SPECT / CT animal

The image reconstruction influence in relative measurement in SPECT / CT animal BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 0-01 (201) 01-09 The image reconstruction influence in relative measurement in SPECT / CT animal S.C.S. Soriano a ; S.A.L. Souza b ; T.Barboza b ; L.V. De Sá

More information

DISCRETE crystal detector modules have traditionally been

DISCRETE crystal detector modules have traditionally been IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 5, OCTOBER 2006 2513 Performance Comparisons of Continuous Miniature Crystal Element (cmice) Detectors Tao Ling, Student Member, IEEE, Kisung Lee, and

More information

Future directions in Nuclear Medicine Instrumentation

Future directions in Nuclear Medicine Instrumentation Future directions in Nuclear Medicine Instrumentation Where are we going - and why? First, the disclosure list My group at the University of Washington has research support from: NIH DOE General Electric

More information

PERFORMANCE EVALUATION OF A SMALL FIELD-OF-VIEW, MOBILE PET/SPECT SYSTEM

PERFORMANCE EVALUATION OF A SMALL FIELD-OF-VIEW, MOBILE PET/SPECT SYSTEM PERFORMANCE EVALUATION OF A SMALL FIELD-OF-VIEW, MOBILE PET/SPECT SYSTEM By MATTHEW THOMAS STUDENSKI A THESIS PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT OF THE

More information

Gamma-ray spectral imaging using a single-shutter radiation camera

Gamma-ray spectral imaging using a single-shutter radiation camera Nuclear Instruments and Methods in Physics Research A299 (1990) 495-500 North-Holland 495 Gamma-ray spectral imaging using a single-shutter radiation camera T.A. DeVol, D.K. Wehe and G.F. Knoll The University

More information

Digital Images & Image Quality

Digital Images & Image Quality Introduction to Medical Engineering (Medical Imaging) Suetens 1 Digital Images & Image Quality Ho Kyung Kim Pusan National University Radiation imaging DR & CT: x-ray Nuclear medicine: gamma-ray Ultrasound

More information

J of Nuclear Medicine Technology, first published online July 27, 2011 as doi: /jnmt

J of Nuclear Medicine Technology, first published online July 27, 2011 as doi: /jnmt J of Nuclear Medicine Technology, first published online July 27, 2011 as doi:10.2967/jnmt.110.084814 Extrinsic Versus Intrinsic Uniformity Correction for g-cameras Randy Bolstad 1, Jody Brown, RT(N),

More information

PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner

PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner September, 2017 Results submitted to Physics in Medicine & Biology Negar Omidvari 1, Jorge Cabello 1, Geoffrey Topping

More information

HIGH SPEED, HIGH RESOLUTION AND LOW COST DIGITAL RADIOGRAPHY

HIGH SPEED, HIGH RESOLUTION AND LOW COST DIGITAL RADIOGRAPHY HIGH SPEED, HIGH RESOLUTION AND LOW COST DIGITAL RADIOGRAPHY AND COMPUTED TOMOGRAPHY SYSTEM Kasiviswanathan Rangarajan1,2 and T. Jensen 1 Department of Computer Engineering 2 Center for Nondestructive

More information

The role ofcompact PSPMTs for image quality enhancement in nuclear medicine

The role ofcompact PSPMTs for image quality enhancement in nuclear medicine Nuclear Instruments and Methods in Physics Research A 505 (2003) 599 603 The role ofcompact PSPMTs for image quality enhancement in nuclear medicine M.N. Cinti a, R. Pani b, *, R. Pellegrini b, F. Garibaldi

More information

Intrinsic and Tomographic Evaluation of Siemens e.cam SPECT System at the Korle-Bu Teaching Hospital (Ghana)

Intrinsic and Tomographic Evaluation of Siemens e.cam SPECT System at the Korle-Bu Teaching Hospital (Ghana) Research Journal of Applied Sciences, Engineering and Technology 3(10): 1152-1158, 2011 ISSN: 2040-7467 Maxwell Scientific Organization, 2011 Submitted: July 17, 2011 Accepted: September 05, 2011 Published:

More information

Industry Breakthrough

Industry Breakthrough Industry Breakthrough Dynamic SPECT Acquisition Quantifying Myocardial Blood Flow Nuclear Cardiology in the 21st Century In the 21st century, most nuclear cameras are still relying on a technology invented

More information

GS Introduction to Medical Physics IV Laboratory 5 Gamma Camera Characteristics

GS Introduction to Medical Physics IV Laboratory 5 Gamma Camera Characteristics GS02 0193 Introduction to Medical Physics IV Laboratory 5 Gamma Camera Characteristics Purpose: To introduce some of the basic characteristics of a gamma camera. This lab will introduce gamma camera QC

More information

Positron Emission Tomography - PET

Positron Emission Tomography - PET Positron Emission Tomography - PET Positron Emission Tomography Positron Emission Tomography (PET): Coincidence detection of annihilation radiation from positron-emitting isotopes followed by tomographic

More information

180 pinhole-spect with tilted detector and OS-EM. reconstruction: phantom studies and potential

180 pinhole-spect with tilted detector and OS-EM. reconstruction: phantom studies and potential 180 pinhole-spect with tilted detector and OS-EM reconstruction: phantom studies and potential clinical applications Alain Seret 1, Michel Defrise 2 and Didier Blocklet 3. 1 Université de Liège (ULg),

More information

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE Takeyuki Hashimoto 1), Morio Onoe 2), Hiroshi Nakamura 3), Tamon Inouye 4), Hiromichi Jumonji 5), Iwao Takahashi 6); 1)Yokohama Soei

More information

X-RAY COMPUTED TOMOGRAPHY

X-RAY COMPUTED TOMOGRAPHY X-RAY COMPUTED TOMOGRAPHY Bc. Jan Kratochvíla Czech Technical University in Prague Faculty of Nuclear Sciences and Physical Engineering Abstract Computed tomography is a powerful tool for imaging the inner

More information

arxiv: v1 [physics.med-ph] 29 Nov 2018

arxiv: v1 [physics.med-ph] 29 Nov 2018 Expected performance of the TT-PET scanner E. Ripiccini, a,b,1 D. Hayakawa, a,b G. Iacobucci, a M. Nessi, a,c E. Nowak, c L. Paolozzi, a O. Ratib, b P. Valerio a and D. Vitturini a a University of Geneva,

More information

First Results From the High-Resolution mousespect Annular Scintillation Camera

First Results From the High-Resolution mousespect Annular Scintillation Camera First Results From the High-Resolution mousespect Annular Scintillation Camera Andrew L. Goertzen, Douglas W. Jones, Jurgen Seidel, King Li, and Michael V. Green Abstract High resolution SPECT imaging

More information

Development of Solid-State Detector for X-ray Computed Tomography

Development of Solid-State Detector for X-ray Computed Tomography Proceedings of the Korea Nuclear Society Autumn Meeting Seoul, Korea, October 2001 Development of Solid-State Detector for X-ray Computed Tomography S.W Kwak 1), H.K Kim 1), Y. S Kim 1), S.C Jeon 1), G.

More information

A high energy gamma camera using a multiple hole collimator

A high energy gamma camera using a multiple hole collimator ELSEVIER Nuclear Instruments and Methods in Physics Research A 353 (1994) 328-333 A high energy gamma camera using a multiple hole collimator and PSPMT SV Guru *, Z He, JC Ferreria, DK Wehe, G F Knoll

More information

Feasibility study of a novel general purpose CZT-based digital SPECT camera: initial clinical results

Feasibility study of a novel general purpose CZT-based digital SPECT camera: initial clinical results Goshen et al. EJNMMI Physics (2018) 5:6 https://doi.org/10.1186/s40658-018-0205-z EJNMMI Physics ORIGINAL RESEARCH Open Access Feasibility study of a novel general purpose CZT-based digital SPECT camera:

More information

Charge Loss Between Contacts Of CdZnTe Pixel Detectors

Charge Loss Between Contacts Of CdZnTe Pixel Detectors Charge Loss Between Contacts Of CdZnTe Pixel Detectors A. E. Bolotnikov 1, W. R. Cook, F. A. Harrison, A.-S. Wong, S. M. Schindler, A. C. Eichelberger Space Radiation Laboratory, California Institute of

More information

The future of nuclear imaging is clear

The future of nuclear imaging is clear Cardius X-ACT The future of nuclear imaging is clear Increased regulations, growing competition, and concerns about radiation exposure are just a sampling of the current challenges facing the nuclear medicine

More information

Journal of Radiation Protection and Research

Journal of Radiation Protection and Research 1) WOO JIN JO et al: CZT BASED PET SYSTEM IN KAERI Journal of Radiation Protection and Research pissn 2508-1888 eissn 2466-2461 http://dx.doi.org/10.14407/jrpr.2016.41.2.081 Paper Received July 17, 2015

More information

CHAPTER 15 DEVICES FOR EVALUATING IMAGING SYSTEMS

CHAPTER 15 DEVICES FOR EVALUATING IMAGING SYSTEMS DEVICES FOR EVALUATING IMAGING SYSTEMS O. DEMIRKAYA, R. AL-MAZROU Department of Biomedical Physics, King Faisal Specialist Hospital and Research Centre, Riyadh, Saudi Arabia 15.1. DEVELOPING A QUALITY

More information

Industry Breakthrough

Industry Breakthrough Industry Breakthrough Dynamic SPECT Acquisition Quantifying Myocardial Blood Flow D-S P EC T Cardiac Imaging System Nuclear Cardiology in the 21st Century In the 21st century, most nuclear cameras are

More information

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A Nuclear Instruments and Methods in Physics Research A 614 (2010) 308 312 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima

More information

Evaluation of XRI-UNO CdTe detector for nuclear medical imaging

Evaluation of XRI-UNO CdTe detector for nuclear medical imaging Home Search Collections Journals About Contact us My IOPscience Evaluation of XRI-UNO CdTe detector for nuclear medical imaging This content has been downloaded from IOPscience. Please scroll down to see

More information

Performance evaluation of a multipinhole small animal SPECT system

Performance evaluation of a multipinhole small animal SPECT system University of Wollongong Research Online Faculty of Engineering - Papers (Archive) Faculty of Engineering and Information Sciences 23 Performance evaluation of a multipinhole small animal SPECT system

More information

Compensating for Nonstationary Blurring by Further Blurring and Deconvolution

Compensating for Nonstationary Blurring by Further Blurring and Deconvolution Compensating for Nonstationary Blurring by Further Blurring and Deconvolution Gengsheng L. Zeng Department of Radiology, Utah Center for Advanced Imaging Research, University of Utah, Salt Lake City, UT

More information

Results of the Measurement of the Collimator Hole Angulation for Different Collimators of SPECT with Adaptive Quality Control Phantom

Results of the Measurement of the Collimator Hole Angulation for Different Collimators of SPECT with Adaptive Quality Control Phantom Modern Instrumentation, 2012, 1, 4953 http://dx.doi.org/10.4236/mi.2012.14007 Published Online October 2012 (http://www.scirp.org/journal/mi) Results of the Measurement of the Collimator Hole Angulation

More information

Initial evaluation of the Indiana small animal PET scanner

Initial evaluation of the Indiana small animal PET scanner Initial evaluation of the Indiana small animal PET scanner Ned C. Rouze, Member, IEEE, Victor C. Soon, John W. Young, Member, IEEE, Stefan Siegel, Member, IEEE, and Gary D. Hutchins, Member, IEEE Abstract

More information

Development of Personal Dosimeter Using Electronic Dose Conversion Method

Development of Personal Dosimeter Using Electronic Dose Conversion Method Proceedings of the Korean Nuclear Spring Meeting Gyeong ju, Korea, May 2003 Development of Personal Dosimeter Using Electronic Dose Conversion Method Wanno Lee, Bong Jae Lee, and Chang Woo Lee Korea Atomic

More information

A High-Resolution GSO-based Brain PET Camera

A High-Resolution GSO-based Brain PET Camera A High-Resolution GSO-based Brain PET Camera J.S. Karp', Senior Member IEEE, L.E. Adam', R.Freifelder', Member IEEE, G. Muehllehner3 Senior Member IEEE, F. Liu"', Student Member IEEE, S. Surti"', Student

More information

Prompt-Gamma Based Range Verification in Particle Therapy: New prospects (also) for 4D?

Prompt-Gamma Based Range Verification in Particle Therapy: New prospects (also) for 4D? Prompt-Gamma Based Range Verification in Particle Therapy: New prospects (also) for 4D? Guntram Pausch 1-3 1 OncoRay National Center for Radiation Research in Oncology, Dresden 2 Technische Universität

More information

2010 Philips BrightView XCT SPECT/CT

2010 Philips BrightView XCT SPECT/CT 2010 Philips BrightView XCT SPECT/CT Unit was purchased from Philips training center in 2015. Installed but never been used by the current facility. (Scroll for pictures) BrightView XCT Camera with PinPoint

More information

International Journal of Scientific & Engineering Research, Volume 4, Issue 9, September ISSN

International Journal of Scientific & Engineering Research, Volume 4, Issue 9, September ISSN International Journal of Scientific & Engineering Research, Volume 4, Issue 9, September-013 06 Evaluating the effect of acquisition parameters on image quality and acquisition time with SPECT using collimator

More information