Suppression of metal artifacts using image-based monoenergetic DECT imaging
|
|
- Paul King
- 5 years ago
- Views:
Transcription
1 Suppression of metal artifacts using image-based monoenergetic DECT imaging Poster No.: C-0519 Congress: ECR 2011 Type: Scientific Paper Authors: B. Krauss, B. Schmidt, M. Sedlmair, T. Flohr; Forchheim/DE Keywords: Physics in radiology, CT, CT-Quantitative, Computer ApplicationsGeneral, Physics, Prostheses DOI: /ecr2011/C-0519 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. Page 1 of 15
2 Purpose Metal artifacts (figure below) are a common problem for interpreting CT-images of patients with metal prostheses. These artifacts can completely obscure delicate soft tissue and bone structures close to metal and are not removed by simply increasing the X-ray tube output. Dual Energy CT offers a way to reduce these artifacts and - as will be shown here commercially available image based processing can be used. Page 2 of 15
3 Fig.: Beam Hardening by hip prosthesis phantom scanned at 80kV. Methods and Materials Origin of metal artifacts It is known that metal artifacts are simultaneously caused by several mechanisms: a) beam hardening [1] (due to polychromatic X-ray spectrum) b) scattered radiation [2,3] (due to inefficiencies of the anti-scatter grid on the detector) c) photon starvation [1] (due to detector response for small signals) d) edge artifacts [1,4] (due to exponential absorption law and/or undersampling) e) cone beam artifacts [1] (due to data acquisition with X-ray cone beam) In the following we will mainly focus on effect a). In contrast to CT-images acquired with a monoenergetic photon beam at a synchrotron facility, X-ray spectra on commercially available CT-scanners are quite broad and range from about 30keV to the peak tube voltage times electron charge. The spectra of a commercially available CT-scanner (Somatom Definition Flash, Siemens AG, Healthcare Sector, Forchheim, Germany) have been simulated with DRASIM and are shown in the figure below. The typical Dual Energy Scan mode on this CT scanner operates at 100kV and 140kV with tin filter (Sn140kV). Page 3 of 15
4 Fig.: X-ray spectra of the SOMATOM Definition Flash (Siemens AG, Healthcare Sector, Forchheim, Germany) as simulated with DRASIM. As a consequence of the polychromatic spectrum, the logarithm of the measured signal attenuation for a certain X-ray beam is no longer proportional to the area density of the traversed material (which would be necessary to perform image reconstruction). However, for clinical CT-scanners the patient mainly consists of water and by assuming that it is again possible to calculate the effective area density of water or - more commonly 3 - the equivalent water thickness at standard density of 1g/cm. Unfortunately, this relation is material dependent: In the following figure the measured equivalent water thickness for 100kV and Sn140kV is plotted as a function of the material thickness of water and iodine for a true water thickness of up to 40cm and an iodine area 2 density of up to 400mg/cm (simulation with DRASIM). While the surface is almost flat for low amounts of iodine, there is an obvious curvature at higher iodine area densities. This curvature is the reason for beam hardening in the case of large amounts of iodine, bone or metal. Page 4 of 15
5 Fig.: Measured equivalent water thickness as a function of true water thickness and iodine area density for 100kV (left) and Sn140kV (right); simulation with DRASIM. Artifact Suppression To describe the observed curvature more quantitatively, an expansion into a power series can be done. This is shown in the next figure, where the measured equivalent water thickness dlow (dhigh) for the low (high) energy spectrum is plotted as a function of the iodine area density #I at a true water thickness of 20cm. A third order polynomial of the form 2 dlow=20cm+a1#i+a2#i +a3#i dhigh=20cm+b1#i+b2#i +b3#i Page 5 of 15
6 provides an almost perfect fit of the measured equivalent water thickness for both voltages. A fairly good approximation is obtained if terms up to second order are kept, while the first order approximation mainly works for iodine area densities below 75mg/ 2 cm. Fig.: Measured equivalent water thickness as a function of iodine area density at a true water thickness of 20cm; a third order polynomial fit (P3) is shown as well as its truncation to two orders (P2) or one order (P1). Simulation with DRASIM. Since image reconstruction is a linear process, the resulting images Ilow (Ihigh) for the low (high) energy spectrum can be similarly expressed after reconstruction: Ilow=Iwater+a1II+a2II2+... Ihigh=Iwater+b1II+b2II2+... Page 6 of 15
7 where the artifact-free iodine image II is related to #I, while the higher order images II2, II3,... are pure artifact images related to powers of #I. The image II is exactly the same for low and high voltage only if the measured X-ray trajectories are the same (e.g. for sequence scans on Dual Source scanners). Otherwise this relation is approximately true, if the object has little variation along the scan axis, which may be fulfilled for metal prostheses. It is now possible to eliminate higher order terms by performing a weighted average of the low and high kv image: Iw=w Ilow+(1-w) Ihigh. In order to eliminate the second order artifact contribution the following condition is required: w=b2/(b2-a2). As image based calculation of monoenergetic images also results in weighted average images, this weight can be transformed into a monoenergetic energy as shown by the blue curve in the following figure. The red line indicates that the optimum weight for 100/Sn140kV corresponds to a monoenergetic energy of 128keV. This statement is independent of the beamhardening material, as the X-ray absorption of all relevant materials can be described by an equivalent mixture of iodine and water [5]. Page 7 of 15
8 Fig.: Monoenergetic energy as a function of the image weight. The equivalent energies of the 100kV and Sn140kV spectrum are represented by green lines. The weight for vanishing artifacts (iodine contrast) is shown by a red (black) line. An estimated patient diameter of 20cm is assumed. Validation with Simulation In order to show that this prediction is self-consistent, the optimum image weight for all three voltage combinations was computed and DRASIM was used to simulate images of a water phantom with 20cm diameter and two Aluminum inserts of 2cm diameter, both at a distance of 2cm from the scan axis. By placing an ROI at the phantom center between the Aluminum rods, it was verified that the beam hardening artifact is indeed removed with high precision. Validation with Scanner Page 8 of 15
9 As the real X-ray spectra differ slightly from the simulated ones, phantom measurements were done in addition. The phantom had to fulfill the following criteria: - body/prosthesis-equivalent shape and materials (realistic beam-hardening artifacts) - small phantom diameter (avoid photon starvation) - preferably cylindrical metal rod facing along the scan axis (minimize cone beam and edge artifacts) Finally, a CTDI head phantom (lucite, 16cm diameter) was chosen, in which lucite rods were replaced with either aluminum (10mm diameter) or stainless steel (13mm diameter). As a more realistic case, a hip prosthesis phantom was used. Scans were performed on a SOMATOM Definition Flash Scanner with the following voltage combinations: 140kV/80kV, 80kV/Sn140kV, 100/Sn140kV. While the CTDI phantom was scanned in sequence mode (default scan protocol DE_AbdomenSeq: collimation 32x0.6mm; slice thickness 5mm, kernel D30f, rotation time 0.5s, CTDIvol=15.2mGy), the hip prosthesis phantom was scanned in spiral mode (default scan protocol DE_Abdomen_LiverVNC: collimation 32x0.6mm; slice thickness 0.75mm, kernel B30f, rotation time 0.5s, pitch 0.6, CTDIvol=18.2mGy). Images were analyzed using the commercially available software syngo Dual Energy (application class "Monoenergetic"). For each phantom and combination of spectra the monoenergetic energy was determined that gave the best visual image impression. The procedure is demonstrated in the attached movie on page. Results For the simulation the following results were obtained, where errorw is the remaining error in the weighted average image: spectra weight energy errorlow errorhigh errorw 100/ Sn140kV kev HU HU -0.3 HU 80/Sn140kV kev HU HU -0.4 HU 140/80kV (no equival.) HU HU -3.2 HU Page 9 of 15
10 Table 1: Beam hardening artifact at the center of the phantom for the original CTimages and weighted average images. In addition the optimum image weight and corresponding monoenergetic energy is listed. The maximum beam hardening error is observed for the 80kV spectrum, the error in the 140kV image without tin filter is 35% lower. Adding the tin filter reduces the artifact by an additional 65%, while the error in the weighted average is again lower by 98%. For the voltage combination 140/80kV it is mathematically impossible to calculate an energy corresponding to the optimum weight; this is in contradiction to the phantom measurement and hints at inaccuracies of the simulated spectra. For the CTDI phantom the following optimum energies were determined, irrespective of the metal insert (Al or steel): - 100/Sn140kV:130keV - 80/Sn140kV: 140keV - 140/80kV:140keV The results are shown in the following two figures, where the rows correspond to the voltage combinations and the columns correspond to the low tube voltage, the high tube voltage and the optimized image, respectively. Image windowing was kept constant at center=40hu and width=300hu, because soft tissue CT-values are almost independent of the spectrum. Page 10 of 15
11 Fig.: Metal artifacts by stainless steel rods for all voltage combinations (rows), the two original images (left, middle) and the weighted average image (right). Page 11 of 15
12 Fig.: Metal artifacts by aluminum rods for all voltage combinations (rows), the two original images (left, middle) and the weighted average image (right). For the hip prosthesis phantom the optimum energy was 130keV (100/Sn140kV) or 120keV (80/Sn140kV); results are shown in the figure below using the same image order as before. One should note that bone is even visible for very high monoenergetic energies as it has a higher effective atomic number than water but also a higher electron density. Page 12 of 15
13 Fig.: Metal artifacts by hip prosthesis phantom for all voltage combinations (rows), the two original images (left, middle) and the weighted average image (right). To quantify the results for the CTDI phantom, one ROI was placed in the area between the metal rods, while a second ROI was placed at the top of the CTDI phantom. The difference between the two measurements was taken as error. Results are shown in the following table: spectrum error for steel error for aluminum 80kV 678 HU 20 HU 100kV 627 HU 12 HU 140kV 243 HU 9 HU Sn140kV 160 HU 2 HU mono 100/Sn140kV 11 HU 1 HU mono 80/Sn140kV 28 HU 3 HU Page 13 of 15
14 mono 140/80kV 21 HU 1 HU Table 2: Beamhardening artifact at the middle between two rods for steel and aluminum inserts in the CTDI head phantom. The steel rods are actually big enough to make higher order artifact images (II3, II4,...) relevant. In addition photon starvation and scattered radiation should have some impact. It is remarkable that by choosing the optimum monoenergetic energy useful images are obtained from this mathematically simple approach (in comparison with published metal artifact reduction methods [6]). Conclusion Monoenergetic images derived from Dual Energy CT images have substantially lower beam hardening artifacts than the original images if the optimal photon energy is selected ( keV for the used phantoms). In clinical cases the optimum monoenergetic energy has been found to be within a larger range (100keV... >190keV). References 1) J.F. Barrett and N. Keat, "Artifacts in CT: Recognition and Avoidance", RadioGraphics 24: , ) Vetter JR, Holden JE, "Correction for scattered radiation and other background signals in dual-energy computed tomography material thickness measurements", Med Phys 15(5): , ) P.M. Joseph and R.D. Spital, "The effects of scatter in x-ray computed tomography", Med Phys 9(4): , ) P.M. Joseph and R.D. Spital, "The exponential edge-gradient effect in x-ray computed tomography", Phys Med Biol 26(3): , 1981 Page 14 of 15
15 5) R.E. Alvarez and A. Macovski, "Energy-selective Reconstruction in X-ray Computerized Tomography", Phys Med Biol 21(5): ,1976 6) E. Meyer et al., "Normalized metal artifact reduction (NMAR) in computed tomography", Med Phys 37(10): , 2010 Personal Information All authors are employees of Siemens AG, Healthcare Sector. Page 15 of 15
Aim. Images for this section: Page 2 of 13
Changes in CT number of high atomic number materials with field of view when using an extended CT number to electron density curve and a metal artifact reduction reconstruction algorithm Poster No.: R-0094
More informationEvaluation of no-grid radiography using the digital scattered x-ray removal processing
Evaluation of no-grid radiography using the digital scattered x-ray removal processing Poster No.: C-0416 Congress: ECR 2016 Type: Authors: Scientific Exhibit R. Suzuki 1, T. Goto 1, H. Ogawa 2, N. Amimoto
More informationInfluence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck
Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Poster No.: C-2205 Congress: ECR 2012 Type: Scientific Paper
More informationStudies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationStudies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationImprovement of CT image quality with iterative reconstruction idose4
Improvement of CT image quality with iterative reconstruction idose4 Poster No.: C-0387 Congress: ECR 2014 Type: Scientific Exhibit Authors: M.-L. Olsson, K. Norrgren, M. Söderberg; Malmö/SE Keywords:
More informationInvestigation of Effective DQE (edqe) parameters for a flat panel detector
Investigation of Effective DQE (edqe) parameters for a flat panel detector Poster No.: C-1892 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit D. Bor 1, S. Cubukcu 1, A. Yalcin 1, O.
More informationFundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging
Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging Poster No.: C-0709 Congress: ECR 2014 Type: Scientific
More informationA study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers
A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers Poster No.: C-3011 Congress: ECR 2010 Type: Topic: Authors: Scientific Exhibit
More informationComputer applications, PACS, Instrumentation, Equipment, Technology assessment /ecr2012/C-0655
The novel medical diagnostic color monitor automatically distinguishing monochrome/color images and displaying these with their own optimal tone curves Poster No.: C-0655 Congress: ECR 2012 Type: Authors:
More informationDose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis
Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis Poster No.: C-0585 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Minehiro,
More informationClear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE)
Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE) Poster No.: C-2459 Congress: ECR 2010 Type: Scientific Exhibit Topic: Neuro Authors: T. Yoneda,
More informationComparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization
Comparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization Poster No.: C-1040 Congress: ECR 2016 Type: Scientific Exhibit
More informationDevelopment of new dosimeter for measuring dose distribution in CT
Development of new dosimeter for measuring dose distribution in CT Poster No.: C-2925 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology - Without Subtopic Authors: Y. Muramatsu, K.
More informationMulti-slice computed tomography analysis of bullet trajectory in forensic investigation.
Multi-slice computed tomography analysis of bullet trajectory in forensic investigation. Poster No.: C-0510 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Usui, Y. Kawasumi, Y. Hosokai, A. Nakajima,
More informationVisualization of sources of scattered radiation from x-ray equipment used for interventional radiology
Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology Poster No.: C-1190 Congress: ECR 2011 Type: Scientific Exhibit Authors: K. Chida, T. Takahashi, D.
More informationCorrection of the local intensity nonuniformity artifact in high field MRI
Correction of the local intensity nonuniformity artifact in high field MRI Poster No.: C-0346 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Paper S. Kai, S. Kumazawa, H. Yabuuchi, F. Toyofuku;
More informationAnalysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging
Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging Award: Magna Cum Laude Poster No.: C-1988 Congress: ECR 2014 Type: Scientific Exhibit Authors: T. Hamaguchi,
More informationThe impact of increasing SID on patient dose in pa abdomen imaging
The impact of increasing SID on patient dose in pa abdomen imaging Poster No.: C-1074 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit T. Starc 1, N. Mekis 2 ; 1 Brezovica/SI, 2 Ljubljana/SI
More informationDetermining acceptance levels for automatic daily image quality control in magnetic resonance imaging
Determining acceptance levels for automatic daily image quality control in magnetic resonance imaging Poster No.: C-1125 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit J. I. Peltonen,
More informationDifference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing
Difference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing scanner Poster No.: C-0672 Congress: ECR 2014 Type: Scientific Exhibit Authors: F.
More informationPitfalls and Remedies of MDCT Scanners as Quantitative Instruments
intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature
More informationLumbar disc height assessment: a comparative study between EOS and digital radiography
Lumbar disc height assessment: a comparative study between EOS and digital radiography Poster No.: C-1178 Congress: ECR 2012 Type: Scientific Paper Authors: V. Freire, M. Benhamou, F. Rannou, S. Poiraudeau,
More informationA comparison between medical-grade liquid crystal display (LCD) and ipad color imaging
A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging Poster No.: C-1377 Congress: ECR 2014 Type: Scientific Exhibit Authors: K. Yoshimura 1, K. Shibata 2, T. Nihashi 1,
More informationPractical guidelines for color calibration and quality assurance of medical displays
Practical guidelines for color calibration and quality assurance of medical displays Poster No.: C-1140 Congress: ECR 2017 Type: Educational Exhibit Authors: T. Kimpe, J. Rostang, G. Van Hoey, A. Xthona
More informationAutomated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London
Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients
More informationElectronic Noise in CT Detectors: Impact on Image Noise and Artifacts
Medical Physics and Informatics Original Research Duan et al. Electronic Noise in CT Detectors Medical Physics and Informatics Original Research Xinhui Duan 1 Jia Wang 1,2 Shuai Leng 1 ernhard Schmidt
More informationWorldwide practice of breast MRI: insights from the MIPA study applications - the MIPA study group
Worldwide practice of breast MRI: insights from the MIPA study applications - the MIPA study group Poster No.: B-0757 Congress: ECR 2015 Type: Scientific Paper Authors: G. Di Leo, R. M. Trimboli, I. Ioan,
More informationNovel cassette-sized, flat-panel digital radiography (DR) system: Initial clinical and workflow results versus computed radiography (CR)
Novel cassette-sized, flat-panel digital radiography (DR) system: Initial clinical and workflow results versus computed radiography (CR) Poster No.: C-3027 Congress: ECR 2010 Type: Scientific Exhibit Topic:
More information1. Patient size AEC. Large Patient High ma. Small Patient Low ma
Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description
More informationThe role of CT in restoration of a very rare Christ wooden statue: a fundamental contribution.
The role of CT in restoration of a very rare Christ wooden statue: a fundamental contribution. Poster No.: C-1514 Congress: ECR 2015 Type: Scientific Exhibit Authors: M. Disaro', P. Sartori; Venice/IT
More informationMetal Artifact Reduction for Orthopedic Implants (O-MAR)
Metal Artifact Reduction for Orthopedic Implants () Summary Since the inception of CT, numerous methods have been proposed to suppress metal artifacts with varying degrees of success. 1-4 (Metal Artifact
More informationDetector technology in simultaneous spectral imaging
Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become
More informationResearch Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.
Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National
More informationThis document is published in:
Institutional Repository This document is published in: Bo Yu (ed.) (01). 01 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC): Anaheim, California, USA. October 9 - November 3, 01.
More informationOptimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation
Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.125 pissn 2508-4445, eissn 2508-4453 Optimization of Energy Modulation Filter for Dual Energy CBCT
More informationImage Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose
Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation
More informationHISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD
CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING
More informationThe feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE
The feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE Poster No.: C-0022 Congress: ECR 2010 Type: Scientific Exhibit Topic: Abdominal Viscera
More informationQC Testing for Computed Tomography (CT) Scanner
QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including
More informationMC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE
MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for
More informationQC by the MPE in Belgium
Acceptance testing of state-of-the-art CT scanners using a new national protocol: first experience on a large number of scanners of different make and model the working group Radiology of the Belgian Hospital
More informationdiagnostic examination
RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011
More informationPROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution
PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution Poster No.: C-1280 Congress: ECR 2011 Type: Educational Exhibit Authors: A. Monteiro, A.
More informationOptimal speed for dual-energy subtraction neck imaging with flat-panel detector radiography
Optimal speed for dual-energy subtraction neck imaging with flat-panel detector radiography Poster No.: C-2938 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology Authors: H. Machida
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationPROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution
PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution Poster No.: C-1280 Congress: ECR 2011 Type: Educational Exhibit Authors: A. Monteiro, A.
More informationPET/CT Instrumentation Basics
/ Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire
More informationTOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions
CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT
More informationNuclear Associates
Nuclear Associates 76-700 Digital Subtraction Angiography Phantom Users Manual March 2005 Manual No. 76-700-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names
More informationI. Introduction.
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case
More informationImaging Technique Optimization of Tungsten Anode FFDM System
Imaging Technique Optimization of Tungsten Anode FFDM System Biao Chen a*, Andrew P. Smith b, Zhenxue Jing a, Elena Ingal a a Hologic, Inc. 600 Technology Drive, DE 1970 b Hologic, Inc. 35 Crosby Drive,
More informationCOMPUTED TOMOGRAPHY 1
COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the
More informationMedical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging
Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition
More informationIterative Reconstruction in Image Space. Answers for life.
Iterative Reconstruction in Image Space Answers for life. Iterative Reconstruction in Image Space * (IRIS) * Please note: IRIS is used as an abbreviation for Iterative Reconstruction in Image Space throughout
More informationComputerized Medical Imaging and Graphics
Computerized Medical Imaging and Graphics 36 (2012) 387 395 Contents lists available at SciVerse ScienceDirect Computerized Medical Imaging and Graphics jo ur n al homep age : www.elsevier.com/locate/compmedimag
More informationLunar Technology Advantages
Lunar Technology Advantages DXA stands for Dual-Energy X-ray Absorptiometry. It is a measurement method that uses the differences in the absorption of high energy and low energy X-ray photons by different
More informationSECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS
RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT
More informationCR Basics and FAQ. Overview. Historical Perspective
Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors
More informationT h e P h a n t o m L a b o r a t o r y
T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order
More informationX-rays in medical diagnostics
X-rays in medical diagnostics S.Dolanski Babić 2017/18. History W.C.Röntgen (1845-1923) discovered a new type of radiation Nature, Jan. 23. 1896.; Science, Feb.14. 1896. X- rays: Induced the ionization
More informationAn Activity in Computed Tomography
Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).
More informationAn Activity in Computed Tomography
Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).
More informationIBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY
WHITE PAPER: IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY IBEX Innovations Ltd. Registered in England and Wales: 07208355 Address: Discovery 2, NETPark, William Armstrong Way, Sedgefield, UK Patents:
More informationBreast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School
Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis
More information12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees
Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees
More informationDose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation
Medical Physics and Informatics Original Research Christner et al. CT Dose Reduction Medical Physics and Informatics Original Research Downloaded from www.ajronline.org by 8.243.133.8 on 2/26/18 from IP
More informationCHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY
14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to
More informationRADspeed Pro. EDGEpackage C501-E041C
RADspeed Pro EDGEpackage C501-E041C 2 Some of the FPDs may be not available in your country. Please contact us to check the availability in your country. 3 Tomosynthesis in the Standing Position Tomosynthesis
More informationMammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over
Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on
More information2D, 3D CT Intervention, and CT Fluoroscopy
2D, 3D CT Intervention, and CT Fluoroscopy SOMATOM Definition, Definition AS, Definition Flash Answers for life. Siemens CT Vision Siemens CT Vision The justification for the existence of the entire medical
More informationIBEX MATERIALS DETECTION TECHNOLOGY
WHITE PAPER: IBEX MATERIALS DETECTION TECHNOLOGY IBEX Innovations Ltd. Registered in England and Wales: 07208355 Address: Discovery 2, NETPark, William Armstrong Way, Sedgefield, TS21 3FH, UK Patents held
More informationThe effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography
Journal of Physics: Conference Series OPEN ACCESS The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography To cite this article: N A A Daud et al 2014 J. Phys.:
More informationPD233: Design of Biomedical Devices and Systems
PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/
More informationClinical Importance on CT
183 Truncated-View Clinical Importance on CT..,..,.... Artifacts:.... James L. Lehr1 A truncated-view artifact in CT is produced whenever any part of the patient or imaged object is present in some but
More informationAn experimental method for ripple minimization in transmission data for industrial X-ray computed tomography imaging system
Sādhanā Vol. 27, Part 3, June 2002, pp. 393 404. Printed in India An experimental method for ripple minimization in transmission data for industrial X-ray computed tomography imaging system UMESH KUMAR,
More informationNuclear Associates , , CT Head and Body Dose Phantom
Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product
More informationRadionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)
Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in
More informationThe image reconstruction influence in relative measurement in SPECT / CT animal
BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 0-01 (201) 01-09 The image reconstruction influence in relative measurement in SPECT / CT animal S.C.S. Soriano a ; S.A.L. Souza b ; T.Barboza b ; L.V. De Sá
More informationReconstruction Filtering in Industrial gamma-ray CT Application
Reconstruction Filtering in Industrial gamma-ray CT Application Lakshminarayana Yenumula *, Rajesh V Acharya, Umesh Kumar, and Ashutosh Dash Industrial Tomography and Instrumentation Section, Isotope Production
More informationClinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11
Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) Hiroki Kadoya Yukiko Kitagawa MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria
More informationFocal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution
Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Cynthia H. McCollough, PhD, DABR, FAAPM, FACR Director, CT Clinical Innovation Center Professor of Medical Physics and
More informationSOMATOM Esprit A Bundle of Energy
SOMATOM Esprit A Bundle of Energy DATA SOMATOM Esprit An economical CT scanner designed for...... Excellent spiral image quality... A wide range of clinical applications... Value performance and reliabilty
More informationEffect of patient support pads on image quality and dose in fluoroscopy a)
Effect of patient support pads on image quality and dose in fluoroscopy a) William R. Geiser, Walter Huda, b) and Nikolaos A. Gkanatsios Department of Radiology, University of Florida, Gainesville, Florida
More informationDigital Imaging and Communications in Medicine (DICOM) Supplement 188: Multi-energy CT Images
Supplement 188: Multi-energy CT Images Page 1 2 4 6 Digital Imaging and Communications in Medicine (DICOM) 8 Supplement 188: Multi-energy CT Images 10 12 14 16 18 20 Prepared by: 22 DICOM Standards Committee,
More informationIMPROVEMENTS IN X-RAY CT
The First International Proficiency Testing Conference Sinaia, România 11 th 13 th October, 2007 IMPROVEMENTS IN X-RAY CT Emilia Dana Seleţchi Faculty of Physics, University of Bucharest, Măgurele, CP
More informationDigital Radiographic Inspection replacing traditional RT and 3D RT Development
Digital Radiographic Inspection replacing traditional RT and 3D RT Development Iploca Novel Construction Meeting 27&28 March 2014 Geneva By Jan van der Ent Technical Authority International Contents Introduction
More informationPredicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities
Journal of Physics: Conference Series PAPER OPEN ACCESS Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Recent citations - Resolution Properties of a
More information1. Define attenuation of radiation, half-value layer, and tenth value layer.
Objectives for Tutorial on Attenuation of Radiation 1. Define attenuation of radiation, half-value layer, and tenth value layer. 2. Explain what happens to intensity of a beam as one removes or adds HVLs
More informationTest Equipment for Radiology and CT Quality Control Contents
Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3
More informationEvaluation of magnetic resonance acoustic noise in 1.5 and 3 Tesla scanners
Evaluation of magnetic resonance acoustic noise in 1.5 and 3 Tesla scanners Poster No.: B-1014 Congress: ECR 2016 Type: Scientific Paper Authors: V. M. F. Silva, I. M. Ramos, J. Moreira, M. Marques; Porto/PT
More informationDose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp
Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division
More informationSAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life.
Neuro Thoracic Abdominal Abdominal Cardiovascular Pediatric SAFIRE Sinogram Affirmed Iterative Reconstruction Answers for life. SAFIRE * (Sinogram Affirmed Iterative Reconstruction) * The information
More informationLSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option
LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option Y. Bercier, Member, IEEE, M. Casey, Member, IEEE, J. Young, Member, IEEE, T. Wheelock, Member, IEEE, T. Gremillion Abstract-- Factors which
More informationX-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage
X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage
More informationVisibility of Detail
Visibility of Detail Radiographic Quality Quality radiographic images represents the, and information is for diagnosis. The of the anatomic structures and the accuracy of their ( ) determine the overall
More informationHigh Energy Digital Radiography & 3D-CT for Industrial Systems
DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France High Energy Digital Radiography & 3D-CT for Industrial Systems Non-Destructive
More informationAcceptance Testing of a Digital Breast Tomosynthesis Unit
Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures
More informationInvestigation of the line-pair pattern method for evaluating mammographic focal spot performance
Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann
More informationTISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY
TISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY By RYAN F. FISHER A THESIS PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT
More information