1 Image Display and Perception J. Anthony Seibert, Ph.D. Department of Radiology UC Davis Medical Center Sacramento, California, USA Image acquisition, display, & interpretation X-rays kvp mas Tube filtration Collimation Patient Size Restraints Exam type ESE, dose Detector Efficiency Resolution Scatter grid DQE Technologist/engineering domain Computer Digitization Preprocessing Postprocessing Configuration Physicist & Radiologist domain PACS Data delivery Data display Data storage Workflow PACS admin domain Human Radiologist Physician Experience Condition 1
2 Outline of Presentation Human visual response Monitor pixel size Types of monitors Luminance, Illuminance and ambient lighting Pan and zoom DICOM GSDF and monitor calibration Viewing pitfalls Quality control Visual Accommodation Determines optimal field of view and monitor dimensions 2
3 Visual Accommodation Average resting point of accommodation 31 inches (75 cm), or arms-length viewing distance Field of view for a specific viewing distance Diagonal dimension should be ~80% of distance Normal viewing at 31 (75 cm) : about 21 (50 cm) From Mike Flynn Henry Ford Hospital Human visual response Frequency variations at fixed contrast Contrast variations at fixed spatial frequency 3
4 Contrast Sensitivity 10/26/2012 Spatial Acuity 60 cm viewing distance 100 cd/m 2 Luminance 21 mm 2 test pattern # pixels / degree and viewing distance determines pixel dimension and matrix size required mm mm % max Spatial Frequency, mm -1 Corresponds to a pixel dimension of ~200 m Monitor pixel size? 4
5 Retina Display Pixel density is such that the human eye is unable to notice pixelation at a normal viewing distance Portable imaging displays Calgary Scientific Resolution MD FDA approved for use iphone, ipad, web 5
6 Types of Monitors Medical Grade Higher luminance (>300 cd/m 2 ), good warranty > 3y Larger bit depth for display, bits Built-in calibration, typically dedicated video card High pixel density, smaller pixels $$$$ Professional / Pro-sumer Commodity/Consumer Grade Lower luminance, longevity/warranty limited Typical 8 bit performance Calibration external $$ Diagnostic Monitors 5 MP (2500 x 2000) 600 Cd/m 2 Monochrome 3 MP (2000 x 1500) 400 Cd/m 2 Monochrome or Color 6
7 Intermediate pro-sumer 76 cm diagonal 2560 x 1600 ~ 0.21 mm/pixel 4 MP = 2 x 2 MP >300 cd/m 2 <$1200/monitor Dual-link DVI video card ($100) Dual monitor setup Technologist QC monitors 2 MP and lower 1600x x x768 Often suboptimal --insufficient resolution --insufficient luminance 7
8 Technologist QC monitor Insist on high quality Mammography: 3 MP, >250 cd/m 2 Technologist now able to discern motion at point of care. Medical or Consumer Grade? Medical: Higher luminance (up to 1000 cd/m 2 ), higher pixel count, higher up-front costs, longer warranty, self-calibration Consumer: Lower luminance, much lower cost, reduced matrix size, external calibration. wide range of models Overall long-term cost? Perhaps similar 8
9 Considerations for Displaying Medical Images on LCD panels LCD glass varies in quality Types of LC switching TN (Twisted Nematic): low grade, severe angular brightness loss PVA (patterned vertical array): intermediate grade, much better viewing performance IPS (In-Plane-Switching): high grade, good angular response, good speed Unpolarized fluorescent light source Horizontal Polarizing Filter LC layer Vertical Polarizing Filter Off Dark Pixel Output Bright On Horizontally Polarized Light Vertically Polarized Light 9
10 Considerations for Displaying Medical Images The matrix size of a radiographic image is often much greater than display matrix size In order to view the whole projection image field of view (FOV), spatial down-sampling (interpolation) is required In order to view the image at its native resolution, a one-toone image-to-monitor pixel mapping is necessary For large images, a sub-image can be displayed at native resolution, and panning the image is necessary This takes more time for the radiologist s diagnosis Typical consumer-grade display monitor Typical radiograph Choices: 1. Display the full image 2. Display full resolution 3. Other (in-between or mag) 10
11 1760 Image 1200 x 1600, full resolution Zoom and Pan 2184 Display with full image resolution 1200 Display x 1600, partial (interpolated) resolution 2560 x 3328 image displayed in an 850 x 1100 display matrix 66% reduction Can see full image FOV 50% reduction in FOV provides better resolution Resolution test chart 11
12 25% reduction in FOV provides enhanced resolution True size (1 monitor pixel = 1 image pixel) 50% magnification (pixel replication) does not increase visible resolution, but does present expanded information for a better match to the human visual system response (you can view the monitor at a further distance) Magnification with bilinear interpolation 12
13 DICOM image quality GSDF (DICOM part 14): identifies a method to calibrate the luminance response of monitors according to just noticeable differences and perceptual linearization Concept: equal change in input pixel value (grayscale) results in equally perceived change in output luminance Barten s model fits, for a given luminance level, the change in luminance that an observer would just notice Each JND step: a single observable change in luminance DICOM monitor calibration Why not just calibrate to a linear response? Human contrast sensitivity varies with luminance Low levels very sensitive Perception difference of Cd/m 2 at ~0.05 Cd/m 2 High levels not as sensitive Perception difference of ~25 Cd/m 2 at ~4000 Cd/m 2 13
14 Luminance (Cd/m 2 ) 10/26/2012 DICOM GSDF (part 14) Perceptual Linearity Grayscale Standard Display Function Film CRT, LCD Displays Just Noticeable Difference (JND) Index Display Calibration Qualitative (simple, fast check) SMPTE pattern Grayscale Standard Display Function Quantitative adjustment of monitor Optimize human visual response to JND s AAPM TG-18 protocols 14
15 SMPTE pattern Pre Calibration Post Calibration (GSDF) AAPM TG 18: Monitor Calibration Defines methodology for acceptance testing and quality control of workstation monitors Provides test images (beyond the SMPTE image) with instructions for use Information is available on AAPM website: aapm.org/pubs/reports/or_03/pdf 15
16 General Purpose QC pattern 0% / 5% 100% / 95% QUALITY CONTROL 16
17 TG-18 Image test patterns.. and a lot more Perceptual Contrast Contrast-Luminance Bit-depth, continuous grayscale Pattern Descriptions / Uses 17
18 Methods / Analysis Detailed in executive summary of TG-18 Useful for initial comparison of monitors and acceptance test procedures.. 18
19 Open Source Calibration Software 19
20 Open Source: Henry Ford Hospital Measure GSDF for particular types of consumer-grade monitors Allow LUTS to be distributed to the enterprise (5000+ monitors..) Quantitative and qualitative verification tools are available Photometers.. $2000+ $300. Major difference: Speed Accuracy +/- 1% 20
21 Note: It s more than DICOM GSDF.. Max Luminance (L max ) to Min Luminance (L min ) Ratio: ~350 Many monitors have L max /L min > 800; add offset to L min Calibration software / photometer Drives the display and interfaces to a photometer for measurement Loads Look-Up-Table for video card to implement the Barten curve response Usable on most consumer-grade monitors and displays 21
22 GSDF calibration Pre-calibration Purchase photometer Download software from Internet Post-calibration GSDF conformance 22
23 Viewing and Environmental Considerations Display considerations Must assure uniform appearance at all calibrated display devices Uncalibrated QC monitors Images seen on PACS don t look right Tech or PACS get the blame for the poor monitor Viewing condition precautions Room illuminance, ambient lighting Angular luminance variations of LCD s Small, low quality Tech QC monitor Ambient lighting Maximum performance achieved at lux Higher illuminance allowed with higher luminance monitor 23
24 Incompatible lighting issues Mammography reading area Mammo Diagnostic Workstation PACS Workstation Breast MRI Workstation Film Viewer Ambient lighting.. 20 to 40 lux Internet Access Workstation Brenna, PC, McEntee M, Evanoff, M, et al. Ambient lighting: effect of illumination on soft-copy viewing of radiographs of the wrist. AJR 2007;188:W
25 Viewing Pitfalls Viewing angle dependence of brightness and contrast Displays are non-lambertian Molecular orientation symmetry within LC layer Some (expensive) LCD monitors correct this: Birefringent filter layers Multidomain pixels In-Plane Switching Combinations of the above Viewing pitfalls Angular luminance variation of LCD monitors Appearance of image 25
26 Image presentation might not be a calibration problem! After contrast adjustment Perception Reality on PACS workstation Compliments of Jeff Shepard, M.D. Anderson Cancer Center Quality Improvement LCD is not optimal as a QC monitor unless: The monitor is calibrated to DICOM PS3.14 (GSDF) and The angular dependence of brightness and contrast is adequately corrected (high quality monitor) Techs should be trained to view LCD s from a position that ensures a viewing angle that is normal to the plane of the display surface 26
27 Display Monitor Quality Control Who? IT staff, PACS administrator, mammo QC tech Vendor Physicist Periodic measurements? Primary: Semi-annually (weekly, mammo) Secondary: Annually Automated GSDF: Monthly What? AAPM TG-18 documentation Simple cleaning chores, weekly? At a minimum a qualitative check of calibration 27
28 Statement on Quality Control requirements Any facility using a digital image data management system must have documented policies and procedures for monitoring and evaluating the effective management, safety, and proper performance of acquisition, digitization, processing, compression, transmission, display, archiving, and retrieval functions of the system. The quality control program should be designed to maximize the quality and accessibility of diagnostic information. Summary Display monitors provide the human / image interface Careful consideration of monitor types, viewing conditions and continued quality control and consistency are crucial! Calibration is necessary for optimal information transfer The weak link in an electronic network is often the display 28
Technical Paper CONSISTENT PRESENTATION OF MEDICAL IMAGES
Technical Paper CONSISTENT PRESENTATION OF MEDICAL IMAGES A REPORT ON MEDICAL WORKSTATION CALIBRATION By Tom Schulte In years past, the quality of medical images was strictly controlled during the exposure,
More informationOutline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control
Understanding Digital Modalities: Image Quality and Dose S. Jeff Shepard, M.S. University of Texas M. D. Anderson Cancer Center Houston, Texas Special Acknowledgement: Stephen K. Thompson, M.S. William
More informationSUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES.
SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES. Tingberg, Anders Published in: Radiation Protection Dosimetry DOI: 10.1093/rpd/ncs302 Published: 2013-01-01 Link to publication Citation for published
More informationDigital radiography (DR) post processing techniques for pediatric radiology
Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions
More informationAcceptance Testing of a Digital Breast Tomosynthesis Unit
Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures
More informationPractical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities
Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Melissa Martin, M.S., FAAPM, FACR, FACMP AAPM Annual Meeting - Philadelphia July 19, 2010 MO-B-204C-1 Educational Objectives
More informationQuality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc
Quality Control for Stereotactic Breast Biopsy Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc. 716-924-0350 Methods of Imaging Guided Breast Biopsy Ultrasound guided, hand-held needle
More informationWhite Paper. The Advantages of Hybrid Gamma PXL. 1 Introduction Outline of Hybrid Gamma PXL Study Results...
White Paper The Advantages of Hybrid Gamma PXL CONTENTS 1 Introduction... 2 2 Outline of Hybrid Gamma PXL... 4 3 Study... 5 4 Results... 7 5 Conclusions... 10 No.17-001 Revision A November 2017 Product
More informationQuality Control of Full Field Digital Mammography Units
Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of
More informationAcquisition, Processing and Display
Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image
More informationThings you may want to know
Things you may want to know 1 / 9 Last updated 18-Nov-05 Why medical monitors are special? Unlike commercial displays, Medical displays are used in the hospital and they are displaying contain life critical
More informationTest Equipment for Radiology and CT Quality Control Contents
Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3
More informationA Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA
A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations Shirley Weddle, RT(R)(M), CIIP, BBA OBJECTIVES Define Computed Radiography (CR) Discuss CR vendor
More informationDigital Image Management: the Basics
Digital Image Management: the Basics Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University Outline From screen/film to digital radiography PACS/Tele
More informationA Novel 12-bit Grayscale Topology PACSmate Revolutionary GrayBoost Technology
A Novel 12-bit Grayscale Topology PACSmate Revolutionary GrayBoost Technology Author: Kenny Jan Date: August 3, 2009 Presented by PACSmate Technology Inc. Table of Contents 1. Introduction... 2 2. GrayBoost
More informationLearning Objectives. Outline. Getting Started with CR. Converting the Radiology Department from Film-Screen to Digital: Making the Transition
Converting the Radiology Department from Film-Screen to Digital: Making the Transition S. Jeff Shepard, MS, DABR University of Texas M. D. Anderson Cancer Center Houston, Texas jshepard@di.mdacc.tmc.edu
More informationFFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR
FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May 2010 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics, Inc. Physicist's Role in the
More informationEssentials of Digital Imaging
Essentials of Digital Imaging Module 7 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 7 Quality 1. ASRT Animation 2. Welcome Welcome to the Essentials of Digital Imaging:
More informationBarco medical display systems. Product catalog
Barco medical display systems Product catalog TECHNOLOGy THAT MAkES A DIFFERENCE Backlight Output Stabilization (BLOS) Backlight Output Stabilization (BLOS) is an image stabilization system driven by a
More informationBreast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School
Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis
More informationChin J Radiol 2003; 28:
Chin J Radiol 2003; 28: 175-180 175 modality console Image acquisition Quality assurance ISO Quality control PACS Picture Archiving and Communication System - Communication in Medicine [1] service WorkList
More informationVisual Requirements for High-Fidelity Display 1
Michael J Flynn, PhD Visual Requirements for High-Fidelity Display 1 The digital radiographic process involves (a) the attenuation of x rays along rays forming an orthographic projection, (b) the detection
More informationSmall Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL
Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL Section 1 Theory and Equipment 1 Introduction to Digital Imaging Small animal radiography has changed dramatically in the past
More informationSTEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS
STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%
More informationIncreasing the Number of Gray Shades in Medical Display Systems How Much is Enough?
Increasing the Number of Gray Shades in Medical Display Systems How Much is Enough? Tom Kimpe and Tom Tuytschaever Medical images produced by x-ray detectors, computed tomography (CT) scanners, and other
More informationGE Healthcare. Senographe 2000D Full-field digital mammography system
GE Healthcare Senographe 2000D Full-field digital mammography system Digital has arrived. The Senographe 2000D Full-Field Digital Mammography (FFDM) system gives you a unique competitive advantage. That
More informationGE Healthcare. Essential for life. Senographe Essential Full-Field Digital Mammography system
GE Healthcare Essential for life Senographe Essential Full-Field Digital Mammography system Excellence in FFDM is a process. An ongoing quest, fueled by our continuing breakthroughs in breast cancer detection
More informationof sufficient quality and quantity
of sufficient quality and quantity The patient s body attenuates the beam as it passes though the body More energy is deposited in organs located near the entry of the beam than near the exit of the beam
More informationTECHNICAL DATA. GIOTTO IMAGE SDL/W is pre-arranged for Full Field Digital Biopsy examination with the patient in prone position.
Ver. 01/06/07 TECHNICAL DATA GIOTTO IMAGE SDL/W LOW DOSE, FULL FIELD DIGITAL MAMMOGRAPHY UNIT USING AMORPHOUS SELENIUM (a-se) TECHNOLOGY DETECTOR (pre-arranged for stereotactic biopsy with the same digital
More informationClinical and management aspects of digital imaging and PACS
Clinical and management aspects of digital imaging and PACS พญ.จามร เช อเพชระโสภณ นายกร งส ว ทยาสมาคมแห งประเทศไทย chamareec@gmail.com www.radiologythailand.org Digital imaging Abbreviation and terminology
More informationBreast Imaging Basics: Module 10 Digital Mammography
Module 10 Transcript For educational and institutional use. This test bank is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.
More informationPerformance and care. all in one
Performance and care all in one INNOVATION IS WHAT DRIVES US THINKING ABOUT THE FUTURE Preventive diagnostics remains an essential weapon in defeating breast cancer. Metaltronica s forward-thinking design
More informationMedical Imaging Displays: Psychophysics and Quality Assurance: Psychophysics and the Human Visual System
Medical Imaging Displays: Psychophysics and Quality Assurance: Psychophysics and the Human Visual System Rationale Display options present significant challenge - which use, under what circumstances, how
More informationFeatures and Weaknesses of Phantoms for CR/DR System Testing
Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal
More informationSECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING
RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 1 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT RADT 3463 COMPUTERIZED IMAGING Section I: Chapter
More informationOpen. the Digitized world. Fuji Computed Radiography
Open the Digitized world Fuji Computed Radiography If just one of these applies to you... Managing developing fluid is hard and darkroom work is a hassle... Images are not stable... Isn t digitalization
More informationSYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology
CODE: RADT 156 INSTITUTE: Health Science TITLE: Equipment Operation I DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of equipment operation and maintenance of radiographic
More informationCR35-SP+ Computed Radiography (CR) System. Innovations in Digital Imaging
CR35-SP+ Computed Radiography (CR) System Innovations in Digital Imaging Solution for Quantum Medical Imaging CR35-SP+ Package An Agfa Healthcare Solution for Quantum Medical Imaging The CR35-SP+ Package
More informationSurveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation AAPM Annual Clinical Meeting Indianapolis, IN August 5, 2013 Learning Objectives Become familiar with the recommendations and
More informationCOST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS
COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS DRive is a digital imaging DR hardware & Software solution designed for General Radiography of anatomy. It intended to replace film/screen
More informationPRACTICE GUIDELINE FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY
The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College
More information8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey
Implementation of the 2016 ACR Digital Mammography QC Manual Medical Physicist Mammography Equipment Evaluation and Annual Survey Eric A Berns, PhD, FACR Radiologist Responsibilities Radiologist Responsibilities
More informationTESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist
TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist Topics Image Uniformity and Artifacts Image Quality - Detail
More informationIHE Radiology Technical Framework Supplement. Stereotactic Mammography Image (SMI) Trial Implementation
Integrating the Healthcare Enterprise 5 IHE Radiology Technical Framework Supplement 10 Stereotactic Mammography Image (SMI) 15 Trial Implementation 20 25 Date: June 11, 2013 Author: IHE Radiology Technical
More informationMEDICAL DISPLAYS. NEC 90 Series. NEC MD Series
MEDICAL DISPLAYS NEC MD Series Are you really sure your display is as accurate as your diagnosis? In medical environments such as radiology and emergency rooms, the right equipment and technology can help
More informationIntroduction to Grayscale Calibration and Related Aspects of Medical Imaging Grade Liquid Crystal Displays
Introduction to Grayscale Calibration and Related Aspects of Medical Imaging Grade Liquid Crystal Displays Kenneth A. Fetterly, 1 Hartwig R. Blume, 2 Michael J. Flynn, 3 and Ehsan Samei 4,5,6 Consistent
More informationChina Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW Digital RF System - HF51-5
China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW Digital RF System - HF51-5 #3, No.9, Jiuxianqiaodong Road, Chaoyang District, Beijing 100015, P.R. China E-mail: international@wandong.com.cn
More informationImage Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images
B R E A S T I M A G I N G S O L U T I O N S Setting the benchmark for mammography M-IV Series Innovations in breast imaging The Lorad M-IV Series exemplifies Hologic s commitment to developing advanced
More informationImage evaluation method for digital mammography with soft-copy reading using a digital phantom
Image evaluation method for digital mammography with soft-copy reading using a digital phantom Norimitsu Shinohara 1, Katsuhei Horita 2, Tokiko Endo 3 1 Department of Radiological Technology, Faculty of
More informationSECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS
RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT
More informationA comparison between medical-grade liquid crystal display (LCD) and ipad color imaging
A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging Poster No.: C-1377 Congress: ECR 2014 Type: Scientific Exhibit Authors: K. Yoshimura 1, K. Shibata 2, T. Nihashi 1,
More informationEnhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)
Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular
More informationNew spectral benefi ts, proven low dose
New spectral benefi ts, proven low dose Philips MicroDose mammography SI, technical data sheet Philips MicroDose SI with single-shot spectral imaging is a fullfi eld digital mammography solution that delivers
More informationACR AAPM SIIM Practice Guideline for Determinants of Image Quality in Digital Mammography
J Digit Imaging (2013) 26:10 25 DOI 10.1007/s10278-012-9521-3 EDITORIAL ACR AAPM SIIM Practice Guideline for Determinants of Image Quality in Digital Mammography Kalpana M. Kanal & Elizabeth Krupinski
More informationX-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto
X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring
More informationLearning Objectives: What s my motivation? (unknown screen actor) Workshop Overview
Practical Medical Physics Adapting Traditional Clinical Medical Physics to Digital Radiography Charles E. Willis, Ph.D., DABR Associate Professor Department of Imaging Physics The University of Texas M.D.
More informationMUSICA Nerve Center. Artificial Intelligence. Intelligent tools for your Digital Radiography workflow. Fluoroscopy. Workflow Optimization
Image Quality Bariatric Abdomen Pediatric Imaging Diagnostic Confidence Fluoroscopy Neonatal Imaging Scatter Suppression Dental Full Leg Full Spine Exposure Control Index Artificial Intelligence General
More information4/19/2016. Quality Control Activities for the RadiologicTechnologist. Objectives. 3D Tomosynthesis QC differences
Quality Control Activities for the RadiologicTechnologist Quality Control Tests 2D QC Tomosynthesis QC DICOM Printer Quality Control Weekly Detector Flat Field Calibration Weekl Artifact Evaluation Weekly
More informationARCO Rk.5 Digital Mobile C-Arm
Product Data ARCO Rk.5 Digital Mobile C-Arm Product Data 1 Easy to move INTRODUCTION Designed with the latest technology throughout, the ARCO Rk.5 set new standards for the excellence image quality ease
More informationDELWORKS DR MEDICAL. take the next step
DELWORKS DR MEDICAL take the next step DELWORKS MEDICAL DR If you are thinking of taking the next step to digital radiography, consider a DelWorks Medical DR Retrofit Package, the easy and affordable way
More informationRAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II
RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated
More informationSlide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?
Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM
More informationDISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter
DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools
More informationDR _ solutions. We understand that customers don t need just products, they want. solutions
DR _ solutions We understand that customers don t need just products, they want solutions index company profile 1974-2005 2006-2007 - 2008 ITALRAY Srl was founded in 1974 as the production branch of Marzocchi
More informationREQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS
REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS DEPARTMENT OF HEALTH DIRECTORATE: RADIATION CONTROL Implementation date: 31 March 2009 Contents
More informationMammography: Physics of Imaging
Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness
More informationDR Retrofit Kit Package Configuration Integrated PACS
DR Retrofit Kit Package Configuration Integrated PACS Description: The DR Retrofit Kit - Configuration Integrated PACS makes the use of CR Cassettes/Imaging Plates or the use of analogue x-ray films, the
More informationMulti-Access Biplane Lab
Multi-Access Biplane Lab Advanced technolo gies deliver optimized biplane imaging Designed in concert with leading physicians, the Infinix VF-i/BP provides advanced, versatile patient access to meet the
More informationVISION V FLOOR MOUNTED ТUBE SYSTEM
VISION V FLOOR MOUNTED ТUBE SYSTEM FLEXIBLE DESIGN for optimal results Vision V is a modular x-ray system with a standard configuration of floor mounted tube stand, vertical bucky stand and bucky table.
More informationExposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?
Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago
More informationFMT18 FLOOR MOUNTED SYSTEM
mas Time AEC 320 kvp 64 mas 320 ma 320 ma 320 DEN 0.0 mm Cu 17 in X 17 in 72.0 in FMT18 FLOOR MOUNTED SYSTEM with Synchronized Tracking System Overview Clinical Efficiency The FMT18 System was designed
More informationCR30-ORACLE CR. with Office Viewer. A Revelation in Computed Radiography. Innovations in Digital Imaging. Solution for Quantum Medical Imaging
CR30-ORACLE CR with Office Viewer A Revelation in Computed Radiography Innovations in Digital Imaging Solution for Quantum Medical Imaging CR30-ORACLE CR with Office Viewer The CR30-ORACLE CR System Package
More informationCR Basics and FAQ. Overview. Historical Perspective
Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors
More informationGet more from your images with Symphony Image Processing
DIRECT RADIOGRAPHY The user-friendly DelWorks image acquisition and processing software provides a wide range of tools for a variety of image enhancements. Its user interface simplifies every step of the
More informationTouch Screen Technology. Classic WAIV. with Touch Screen Technology. Legendary Reputation...Sensitivity to your budget
Touch Screen Technology Classic WAIV with Touch Screen Technology Legendary Reputation...Sensitivity to your budget Streamline Workflow For Higher Productivity And Patient Throughput. The CRescendo Classic
More informationCR Retrofit Kit Package Configuration Integrated PACS
CR Retrofit Kit Package Configuration Integrated PACS Description: The CR Retrofit Kit Integrated PACS makes the use of analogue x-ray films, the use of film cassettes and intensifying screens, the use
More informationQuantum Medical Imaging
Quantum Medical Imaging Pioneering technology from the premiere innovator in veterinary imaging Corporate Overview Quantum Medical Imaging is a highly innovative company which designs and manufactures
More informationPractical guidelines for color calibration and quality assurance of medical displays
Practical guidelines for color calibration and quality assurance of medical displays Poster No.: C-1140 Congress: ECR 2017 Type: Educational Exhibit Authors: T. Kimpe, J. Rostang, G. Van Hoey, A. Xthona
More informationDigital Imaging Considerations Computed Radiography
Digital Imaging Considerations Digital Radiography Computed Radiography o Cassette based Direct or Indirect Digital Radiography o Cassetteless Computed Radiography 1 CR Image Acquisition Most like conventional
More informationDigital Mammography Quality Control for the Mammographic Technologist
Ontario Breast Screening Program Digital Mammography Quality Control for the Mammographic Technologist Authors: G.E. Mawdsley, A.K. Bloomquist, M.J. Yaffe October 2011 Revision 3.1 Mammographic Physics
More informationWhile digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.
In press 2004 1 2 Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Digital radiology An appropriate analogy that is easy for most people to understand
More informationOutline ASRT Changes Impact on current curriculum Potential new courses WECM Changes Last update Resources and needs
Change nd Annual Blinn College 2 nd Educator s Workshop For Radiologic Sciences July 28, 2007 Christi Carter, MSRS, RT(R) Outline ASRT Changes Impact on current curriculum Potential new courses WECM Changes
More informationOTC18 OVERHEAD TUBE CRANE SYSTEM
OTC18 OVERHEAD TUBE CRANE SYSTEM System Overview Clinical Performance Versatile and intuitive, the OTC18M System delivers enhanced patient comfort and optimized workflow. Precisely designed to withstand
More informationTeaching Digital Radiography and Fluoroscopic Radiation Protection
Teaching Digital Radiography and Fluoroscopic Radiation Protection WCEC 20 th Student Educator Radiographer Conference Dennis Bowman, RT(R), CRT (R)(F) Community Hospital of the Monterey Peninsula (CHOMP)
More informationDigiMam Conformance Statement for DICOM V3.0
DigiMam Conformance Statement for DICOM V3.0 Copyright 2004 by I.M.S. s.r.l. DOCUMENT VERSIONS Version Date Author Changes 1.00 15-Feb-05 IMS s.r.l. First Version DOCUMENT VERSIONS Page 2 of 29 TABLE OF
More informationMammograph FFDM PRODUCT DATA
PRODUCT DATA Product data Rev.1 (May 2010) Page Description and Configurations 3 Technical Features 7 Mammography Unit 8 Acquisition Unit 12 Diagnostic Unit (Optional) 12 Size and Dimensions 13 Classifications
More informationOverview of Safety Code 35
Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)
More informationUltimate DR flexibility
Ultimate DR flexibility to fit your room, workflow, and budget KODAK DIRECTVIEW DR 7500 System KODAK DIRECTVIEW DR 7500 System YOU HAVE NEVER SEEN A DIGITAL RADIOGRAPHY SYSTEM LIKE THIS! Your radiography
More informationI. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA
FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE
More informationDICOM Implementations for Digital Radiography
The Medicine Behind the Image DICOM Implementations for Digital Radiography David A. Clunie Princeton Radiology Pharmaceutical Research Disclosure & Acknowledgements CTO RadPharm Proprietor of PixelMed
More informationCOMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED
More informationIntroduction. Digital Mammography QA: Comparing the Manufacturers Recommendations. What is QC and why is it important? Review & compare QC tests
Slide 1 Digital Mammography QA: Comparing the Manufacturers Recommendations Eric A. Berns, Ph.D. Slide 2 Introduction What is QC and why is it important? Review & compare QC tests Key take home points
More informationDR Retrofit Kit Package Configuration DICOM
DR Retrofit Kit Package Configuration DICOM Description: The DR Retrofit Kit - Configuration DICOM makes the use of CR Cassettes/Imaging Plates or the use of analogue x-ray films, the use of film cassettes
More informationImage Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images
B R E A S T I M A G I N G S O L U T I O N S Setting the benchmark for mammography M-IV Series Innovations in breast imaging The Lorad M-IV Series exemplifies Hologic's commitment to developing advanced
More informationDigital Radiography X-Ray System. X Twin with X Mobil Roesys GmbH [rshsmi] 1/11
Digital Radiography X-Ray System 2017 Roesys GmbH [rshsmi] 1/11 1 General specifications The system is intended to be installed with floor mounted components only. It enables spacesaving installation without
More informationPACS Fundamentals. By: Eng. Valentino T. Mvanga Ministry of Health and Social Welfare Tanzania
PACS Fundamentals By: Eng. Valentino T. Mvanga Ministry of Health and Social Welfare Tanzania 1 Learning Goals To Understand the importance of PACS To Understand PACS infrastructure requirement Introduction
More informationGet more from your images with Symphony Image Processing
DIRECT RADIOGRAPHY The user-friendly DelWorks image acquisition and processing software possesses a wide range of tools for a variety of image manipulations. Its user interface simplifies every step of
More informationMammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over
Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on
More informationThe Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System
The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System Manus
More informationDo you have any other questions? Please call us at (Toll Free) or , or
INSTRUCTIONS Read the appropriate course/ textbook. This is an open book test. A score of 75% or higher is needed to receive CE credit. You will have a maximum of three attempts to pass this course. Please
More information