Iterative Reconstruction
|
|
- Natalie Wood
- 5 years ago
- Views:
Transcription
1 RECENT ADVANCES IN CT RADIATION DOSE REDUCTION TECHNIQUES Iterative Reconstruction Kalpana Kanal, PhD, FSCBTMR, FACR, FAAPM Professor and Director, Diagnostic Physics Section University of Washington Seattle, WA SCBT-MR
2 DISCLOSURE NONE 2
3 TALKING POINTS Iterative Reconstruction (IR) How it works? Different types of IR Examples from Literature IR and Full/Half Dose IR and Spatial Resolution IR and Lesion Detection IR and Radiation Dose Clinical Implementation of IR Advantages and Disadvantages of IR 3
4 INTRODUCTION IR reconstruction is not a new concept It was first used in the 1960s for SPECT Also used in first CT in 1972 Limited by computational capabilities Re-emergence in last few years 4
5 HOW DOES ITERATIVE RECONSTRUCTION (IR) WORK? Ref: Beister et al, Physica Medica, 2012, 28:
6 DIFFERENT TYPES OF IR Pure IR (no modeling) Statistical IR (modeling of photon counting statistics) Raw data domain (sinogram domain) Volumetric images (image domain) During IR process Model based IR (go beyond statistical IR) Try to model acquisition process as accurately as possible Ref: Beister et al, Physica Medica, 2012, 28:
7 GEOMETRIC MODELING Ref: Beister et al, Physica Medica, 2012, 28:
8 VARIOUS AVAILABLE IR ALGORITHMS IN THE USA Ref: Raman et al, JACR 2013, 10:37-41 GE ASIR-V Siemens ADMIRE Ref: Beister et al, Physica Medica, 2012, 28:
9 GENERATIONS OF ITERATIVE RECONSTRUCTION (GE) ASiR ASiR-V Veo Physics Noise Object Noise Object Optics Real-time performance (up to 35 ips) Reduce dose up to 82% Profound image quality under 1 msv Full model-based IR Longer reconstruction times Slide courtesy: GE Medical Systems Ref: Beister et al, Physica Medica, 2012, 28:
10 ASIR - V Slide courtesy: GE Medical Systems 10
11 SIEMENS ADMIRE Five image noise and sharpness levels Strength 1 Strength 4 Strength 5 There is an incremental reduction of noise in each of the images at a natural image impression. Slide courtesy: GE Medical Systems 11
12 FULL DOSE VS. HALF DOSE Images acquired with half dose have 1.4 times SD values without ASIR (28.57 vs 20.39) Reconstructing images with 30% ASIR for half-dose acquisitions produce images with noise nearly equivalent to that of full dose images without ASIR Degree of dose reduction is greater for patients with a lower BMI 57 yr old woman, BMI = 18 Ref: Hara et al, AJR, Sep kv, 3.75 mm, CTDI = 8 mgy, no ASIR 120 kv, 3.75 mm, CTDI = 8 mgy, ASIR 140 kv, 3 mm, CTDI = 22 mgy 12
13 SPATIAL RESOLUTION Spatial resolution has been shown to degrade with IR. Slightly decreased sharpness of cyst edges and a mildly irregular or jagged margin of solid organs 120 kv, 3.75 mm, CTDI = 11 mgy, no ASIR 120 kv, 3.75 mm, CTDI = 11 mgy, ASIR 140 kv, 3 mm, CTDI = 20 mgy 13
14 IR AND LESION DETECTION Ref: Jensen et al, Eur Radiol, July
15 IR and Rad Dose Ref: Padole et al, AJR,
16 CLINICAL IMPLEMENTATION OF IR How does one go about implementing IR? Section buy-in is important, have all radiologists on board Start at vendor recommendations based on clinical application, then fine-tune? Important to reduce dose upfront by at least 20-25% before using IR Be prepared for resistance to new technology/process Continuous iterative improvement process 16
17 ADVANTAGES OF IR Reduce image noise and artifacts depending on degree of modeling By modeling cause of artifacts during reconstruction, IR methods represent more intuitive and natural way of image reconstruction Potential for significant patient dose reduction Using model based IR has better lesion conspicuity than FBP Reading off thinner slices as low noise, improves lesion detection Using statistical based IR does not show significant lesion conspicuity 17
18 DISADVANTAGES OF IR Artificial appearance of image Increased computational effort Evaluation of image noise and resolution is more complex Statistical modeling for low dose CT electronic noise more dominant Difficult to adopt by radiologists as noise patterns and artifacts look different Are we missing diagnostically important information? Spatial resolution degrades slightly at high IR levels 18
19 ARE WE MISSING DIAGNOSTICALLY IMPORTANT INFORMATION? Hypodense seam along the stent indicating proliferation of the vessel intima which may ultimately lead to stent stenosis Hypodense seam not visible in IR image Hypodense seam confirmed by interventional angiography Ref: Beister et al, Physica Medica, 2012, 28:
20 CONCLUSIONS Iterative reconstruction methods are available on modern scanners Adoption and implementation varies depending on facility Faster computational speeds will make model-based IR more practical to use Radiologist adoption due to different image quality (less noise, more artificial look) and confidence level (are we missing something) varies 20
21 THANKS! TUESDAY 11:20 12:05: Workshop (flipped classroom) Recent Advances in CT Radiation Dose Reduction Techniques William Shuman, MD, FSCBTMR Kalpana Kanal, PhD, FSCBTMR Reduction-Techniques 21
22 22
23 ASiR-V Slide courtesy: GE Medical Systems 23
24 SIEMENS ADMIRE (Advanced Modeled Iterative Reconstruction) Potential to lower radiation Superb details, delineation and sharpness of organ borders Positive impact on the image quality of e.g. streak artifacts in the shoulder region Slide courtesy: Siemens Medical Systems 24
Software and Hardware in CCTA. Elly Castellano PhD
Software and Hardware in CCTA Elly Castellano PhD Outline technical requirements for coronary CTA the modern cardiac CT scanner ECG-gating technology image reconstruction algorithms 2 Technical requirements
More informationTOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions
CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT
More informationModel Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise
Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise Jonas Rydberg, M.D. Professor of Radiology Indiana University School of Medicine Indianapolis, Indiana Medical
More informationImage Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose
Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation
More informationIterative Reconstruction in Image Space. Answers for life.
Iterative Reconstruction in Image Space Answers for life. Iterative Reconstruction in Image Space * (IRIS) * Please note: IRIS is used as an abbreviation for Iterative Reconstruction in Image Space throughout
More informationInfluence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck
Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Poster No.: C-2205 Congress: ECR 2012 Type: Scientific Paper
More informationWide-Detector CT for TAVR Planning:
Wide-Detector CT for TAVR Planning: Impact on Iodine Dose, Radiation Dose, and Image Quality SCBTMR 2015 Annual Course Thursday, October 8 William P. Shuman MD FSCBTMR Department of Radiology University
More informationAutomated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London
Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients
More informationPET: New Technologies & Applications, Including Oncology
PET: New Technologies & Applications, Including Oncology, PhD, FIEEE Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Disclosures Research Contract, GE Healthcare
More informationYinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2
Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2 1. Department of Medical Physics, University of Wisconsin-Madison 2. Department
More informationSPECT Reconstruction & Filtering
SPECT Reconstruction & Filtering Goals Understand the basics of SPECT Reconstruction Filtered Backprojection Iterative Reconstruction Make informed choices on filter selection and settings Pre vs. Post
More informationSAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life.
Neuro Thoracic Abdominal Abdominal Cardiovascular Pediatric SAFIRE Sinogram Affirmed Iterative Reconstruction Answers for life. SAFIRE * (Sinogram Affirmed Iterative Reconstruction) * The information
More informationPitfalls and Remedies of MDCT Scanners as Quantitative Instruments
intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature
More informationImprovement of CT image quality with iterative reconstruction idose4
Improvement of CT image quality with iterative reconstruction idose4 Poster No.: C-0387 Congress: ECR 2014 Type: Scientific Exhibit Authors: M.-L. Olsson, K. Norrgren, M. Söderberg; Malmö/SE Keywords:
More information12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees
Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees
More informationClinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11
Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) Hiroki Kadoya Yukiko Kitagawa MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria
More informationEnhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)
Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular
More informationHISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD
CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING
More informationFocal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution
Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Cynthia H. McCollough, PhD, DABR, FAAPM, FACR Director, CT Clinical Innovation Center Professor of Medical Physics and
More informationOptimized CT metal artifact reduction using the Metal Deletion Technique (MDT)
Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT) F Edward Boas, Roland Bammer, and Dominik Fleischmann Extended abstract for RSNA 2012 Purpose CT metal streak artifacts are
More informationRadionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)
Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in
More informationAdvanced Noise Reduction Processing for X-ray CT System with Iterative Processing. Koichi Hirokawa MEDIX VOL. 56 P.43 P.46
Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing Taiga Goto Koichi Hirokawa Hisashi Takahashi MEDIX VOL. 56 P.43 P.46 Advanced Noise Reduction Processing for X-ray CT System
More informationdiagnostic examination
RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011
More informationSurveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation AAPM Annual Clinical Meeting Indianapolis, IN August 5, 2013 Learning Objectives Become familiar with the recommendations and
More informationDiagnostic X-Ray Shielding
Diagnostic X-Ray Shielding Multi-Slice CT Scanners Using NCRP 147 Methodology Melissa C. Martin, M.S., FAAPM, FACR Therapy Physics Inc., Bellflower, CA AAPM Annual Meeting, Orlando, FL FL Refresher Course
More informationAquilion Precision Ultra-High Resolution CT: Quantifying diagnostic image quality
Aquilion Precision Ultra-High CT: Quantifying diagnostic image quality Kirsten Boedeker, PhD, DABR Senior Manager, Quantitative Image Quality Canon Medical Systems Corporation Introduction Over the last
More informationElectronic Noise in CT Detectors: Impact on Image Noise and Artifacts
Medical Physics and Informatics Original Research Duan et al. Electronic Noise in CT Detectors Medical Physics and Informatics Original Research Xinhui Duan 1 Jia Wang 1,2 Shuai Leng 1 ernhard Schmidt
More information1. Patient size AEC. Large Patient High ma. Small Patient Low ma
Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description
More informationTISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY
TISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY By RYAN F. FISHER A THESIS PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT
More informationTranslating Protocols Between Scanner Manufacturer and Model
Translating Protocols Between Scanner Manufacturer and Model Robert J. Pizzutiello, MS, FAAPM, FACMP Sr. Vice-President, Global Physics Solutions President, Upstate Medical Physics Objectives Understand
More informationMetal Artifact Reduction for Orthopedic Implants (O-MAR)
Metal Artifact Reduction for Orthopedic Implants () Summary Since the inception of CT, numerous methods have been proposed to suppress metal artifacts with varying degrees of success. 1-4 (Metal Artifact
More informationSlide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?
Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM
More informationBreast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School
Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis
More information2D, 3D CT Intervention, and CT Fluoroscopy
2D, 3D CT Intervention, and CT Fluoroscopy SOMATOM Definition, Definition AS, Definition Flash Answers for life. Siemens CT Vision Siemens CT Vision The justification for the existence of the entire medical
More informationMaximum Performance, Minimum Space
TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.
More informationOverview of Safety Code 35
Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)
More informationClinical Experiences with a Patient Skin Dose Monitoring and Tracking Program
Clinical Experiences with a Patient Skin Dose Monitoring and Tracking Program Allen R. Goode, MS, DABR Chief Diagnostic Medical Physicist Department of Radiology & Medical Imaging University of Virginia
More informationResearch Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.
Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National
More informationThe disclaimer on page 1 is an integral part of this document. Copyright November 30, 2017 by AAPM. All rights reserved.
DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION
More informationDose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation
Medical Physics and Informatics Original Research Christner et al. CT Dose Reduction Medical Physics and Informatics Original Research Downloaded from www.ajronline.org by 8.243.133.8 on 2/26/18 from IP
More informationQC Testing for Computed Tomography (CT) Scanner
QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including
More informationWhile digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.
In press 2004 1 2 Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Digital radiology An appropriate analogy that is easy for most people to understand
More informationQuantitation of clinical feedback on image quality differences between two CT scanner models
Received: 4 August 2016 Revised: 4 November 2016 Accepted: 12 December 2016 DOI: 10.1002/acm2.12050 MEDICAL IMAGING Quantitation of clinical feedback on image quality differences between two CT scanner
More informationChanging the Shape of Nuclear Medicine
TRUTH IN IMAGING Changing the Shape of Nuclear Medicine Multi-Purpose SPECT Scanner Nothing Gets Closer Introducing 360 Body Contour Scanning With 360 degree detector coverage, and unique proximity sensors
More information160-slice CT SCANNER / New Standard for the Future
TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.
More informationTomosynthesis and Motion
Tomosynthesis (3D) Motion Unsharpness Occurs at about the same frequency as conventional mammography (2D) Presents the same issues as 2D motion, EXCEPT that motion may go undetected Most common patient-related
More informationRadiation Dose Modulation. the Multidetector CT Era: From Basics to Practice 1
Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. EDUCATION EXHIBIT
More informationExposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations
Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations S. Baechler 1, P. Monnin 1, A. Aroua 1, J.F. Valley 1, M. Perrier, P. Trueb 3, F.R. Verdun 1 1 University
More informationSIGNA Explorer Lift revives our MR
Seiji Shiotani, MD, PhD Seirei Fuji Hospital in Fuji City, Shizuoka, Japan Masayoshi Sugimura Seirei Fuji Hospital in Fuji City, Shizuoka, Japan SIGN Explorer Lift revives our MR The clinical usefulness
More informationThe image reconstruction influence in relative measurement in SPECT / CT animal
BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 0-01 (201) 01-09 The image reconstruction influence in relative measurement in SPECT / CT animal S.C.S. Soriano a ; S.A.L. Souza b ; T.Barboza b ; L.V. De Sá
More informationX-ray detectors in healthcare and their applications
X-ray detectors in healthcare and their applications Pixel 2012, Inawashiro September 4th, 2012 Martin Spahn, PhD Clinical applications of X-ray imaging Current X-ray detector technology (case study radiography
More informationQuality Control of Full Field Digital Mammography Units
Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of
More informationImproved compressibility of multislice CT datasets using 3D JPEG2000 compression
International Congress Series 1268 (2004) 57 62 www.ics-elsevier.com Improved compressibility of multislice CT datasets using 3D JPEG2000 compression Khan M. Siddiqui a,b, *, Eliot L. Siegel a,c, Bruce
More informationCHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY
14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to
More informationDetector technology in simultaneous spectral imaging
Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become
More informationAdvanced digital image processing for clinical excellence in fluoroscopy
Dynamic UNIQUE Digital fluoroscopy solutions Dynamic UNIQUE Advanced digital image processing for clinical excellence in fluoroscopy André Gooßen, PhD, Image Processing Specialist Dörte Hilcken, Clinical
More informationSimultaneous Multi-Slice (Slice Accelerated) Diffusion EPI
Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Val M. Runge, MD Institute for Diagnostic and Interventional Radiology Clinics for Neuroradiology and Nuclear Medicine University Hospital Zurich
More informationRobert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center. Thursday September 8, 2011
Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center Thursday September 8, 2011 Quality Control Quality control should be performed because:
More informationabc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice nww.medical-devices.nhs.
abc September 2004 MHRA 04099 Philips Mx8000 IDT CT scanner technical evaluation Best choice best practice www.mhra.gov.uk nww.medical-devices.nhs.uk About MHRA evaluation reports. What you can expect.
More informationImproved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques
Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Wataru FUKUDA* Junya MORITA* and Masahiko YAMADA* Abstract Tomosynthesis is a three-dimensional imaging technology
More informationWide beam CT dosimetry. Elly Castellano
Wide beam CT dosimetry Elly Castellano Outline revision: CT dose indices wide-beam CT: the end of the road for CTDI? the IEC rescue plan for CTDI 100 the american way AAPM report 111 better estimates of
More informationImproved Image Quality in Dual- Energy Abdominal CT: Comparison of Iterative Reconstruction in Image Space and Filtered Back Projection Reconstruction
Medical Physics and Informatics Original Research Wang et al. Image Reconstruction Algorithms for Dual-Energy Abdominal CT Medical Physics and Informatics Original Research Rui Wang 1 Wei Yu 1 Runze Wu
More informationCalculati tions - BIR meth thod and NCRP 147
Calculations l - BIR method and NCRP 147 David Sutton PhD 1. Cath Lab Indicative of all C-arm applications Only secondary radiation Primary Beam is stopped by the image intensifier Geometry BIR 9.5 m Preparation
More informationAuthors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2
Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2 Medical Consult, SA; Establish a method to correlate image
More informationAbdominal Radiology ISSN X. Abdom Radiol DOI /s
Assessment of 70-keV virtual monoenergetic spectral images in abdominal CT imaging: A comparison study to conventional polychromatic 120- kvp images Negin Rassouli, Hamid Chalian, Prabhakar Rajiah, Amar
More informationQC by the MPE in Belgium
Acceptance testing of state-of-the-art CT scanners using a new national protocol: first experience on a large number of scanners of different make and model the working group Radiology of the Belgian Hospital
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationCT Data Storage Reduction by Means of Compressing Projection Data Instead of Images: Feasibility Study 1
Kyongtae T. Bae, MD, PhD Bruce R. Whiting, PhD Index terms: Computed tomography (CT), image display and recording Computed tomography (CT), image processing Computed tomography (CT), technology Data compression
More informationPatient-Assisted Compression Impact on Image Quality and Workflow
Patient-Assisted Compression Impact on Image Quality and Workflow Senographe Pristina In 2017, GE Healthcare s Senographe Pristina ( Pristina ) was approved by the FDA using the standard technologist-controlled
More informationOriginal Article Thoracic Imaging. Ki Baek Lee, RT, MS, Hyun Woo Goo, MD, PhD INTRODUCTION
Original Article Thoracic Imaging https://doi.org/10.3348/kjr.2018.19.1.119 pissn 1229-6929 eissn 2005-8330 Korean J Radiol 2018;19(1):119-129 Quantitative Image Quality and Histogram-Based Evaluations
More information2 nd generation TOMOSYNTHESIS
2 nd generation TOMOSYNTHESIS 2 nd generation DBT true innovation in breast imaging synthesis graphy Combo mode Stereotactic Biopsy Works in progress: Advanced Technology, simplicity and ergonomics Raffaello
More informationGE Healthcare. Essential for life. Senographe Essential Full-Field Digital Mammography system
GE Healthcare Essential for life Senographe Essential Full-Field Digital Mammography system Excellence in FFDM is a process. An ongoing quest, fueled by our continuing breakthroughs in breast cancer detection
More informationSOMATOM Esprit A Bundle of Energy
SOMATOM Esprit A Bundle of Energy DATA SOMATOM Esprit An economical CT scanner designed for...... Excellent spiral image quality... A wide range of clinical applications... Value performance and reliabilty
More informationSynchrotron X-ray tomographic microscopy Theory vs. practice
Synchrotron X-ray tomographic microscopy Theory vs. practice Federica Marone Swiss Light Source, Paul Scherrer Institut, Villigen, Switzerland Theory Radon transform Rf x = Beer-Lambert law I E = I 0 (E)e
More informationDose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp
Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division
More informationA Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017
A Fast Monolithic System for Proton Imaging Fritz DeJongh ProtonVDA Inc October 2017 Disclosures I am a cofounder and co-owner of ProtonVDA Inc We hold intellectual property rights on our proton imaging
More informationMammography: Physics of Imaging
Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness
More informationIntroduction. Sam R. Kottamasu Lawrence R. Kuhns
Pediatr Radiol (1997) 27: 119 123 Springer-Verlag 1997 Sam R. Kottamasu Lawrence R. Kuhns Musculoskeletal computed radiography in children: scatter reduction and improvement in bony trabecular sharpness
More informationInterventional X-ray quality measure based on a psychovisual detectability model
Interventional X-ray quality measure based on a psychovisual detectability model Asli Kumcu, Benhur Ortiz-Jaramillo, Ljiljana Platisa, Bart Goossens, Wilfried Philips iminds-telin-ipi, Ghent University,
More informationMedical Images Analysis and Processing
Medical Images Analysis and Processing - 25642 Emad Course Introduction Course Information: Type: Graduated Credits: 3 Prerequisites: Digital Image Processing Course Introduction Reference(s): Insight
More information7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline
Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation Image Quality Review I: Basics and Image Quality TH-A-16A-1 Thursday 7:30AM - 9:30AM Room: 16A J. Anthony
More informationTORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL
TORNIER BLUEPRINT 3D Planning + PSI SCAN PROTOCOL Contents 3 Introduction 3 Patient preparation 3 Scanning instructions 4 Image instructions 5 Scanning parameters 6 Technical instructions 2 BLUEPRINT 3D
More informationMETAL artifact reduction (MAR) is a major problem
1 Metal artifact reduction based on the combined prior image Yanbo Zhang, Xuanqin Mou arxiv:1408.5198v [physics.med-ph] 4 Sep 014 Abstract Metallic implants introduce severe artifacts in CT images, which
More informationDifferentiation of Malignant and Benign Masses on Mammograms Using Radial Local Ternary Pattern
Differentiation of Malignant and Benign Masses on Mammograms Using Radial Local Ternary Pattern Chisako Muramatsu 1, Min Zhang 1, Takeshi Hara 1, Tokiko Endo 2,3, and Hiroshi Fujita 1 1 Department of Intelligent
More informationAn Activity in Computed Tomography
Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).
More informationThe Trend of Medical Image Work Station
The Trend of Medical Image Work Station Abstract Image Work Station has rapidly improved its efficiency and its quality along the development of biomedical engineering. The quality improvement of image
More informationNeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results
NeuViz 16 Computed Tomography Elevating routine imaging for exceptional results Essence NeuViz 16 Raising the bar on clinical utility in routine imaging. Get more. More clinical information for patients.
More informationGoing Live with the Aquilion VISION Volume 4D CT
VISIONS 2-14 COMPUTED TOMOGRAPHY Going Live with the Aquilion VISION Volume 4D CT Interview with Dr. Rick Bhatia, Regional Clinical Chief Diagnostic Imaging Program, Eastern Health, Newfoundland What were
More informationNew Technology in Nuclear Medicine
New Technology in Nuclear Medicine Reed G. Selwyn, PhD, DABR Vice Chair of Research & Imaging Sciences Associate Professor and Chief, Medical Physics Dept. of Radiology, University of New Mexico Objectives
More informationA Guide to Radiochromic Film Dosimetry with EBT2 and EBT3
A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 David F. Lewis Advanced Materials Group Ashland Specialty Ingredients Spain, April 2014 What is Radiochromic Film? A film that instantly changes
More informationCOMPUTED TOMOGRAPHY 1
COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the
More informationSTEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS
STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%
More informationRadiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893
Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -
More informationInvestigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras
Egyptian J. Nucl. Med., Vol 2, No. 2, Dec. 2009 82 PHYSICS, Original Artical Investigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras Abdelsattar, M.B. Ph.D.; BuHumaid,
More informationDURING the past 15 years the use of digitized
DIGITAL IMAGING BASICS Properties of Digital Images in Radiology DURING the past 15 years the use of digitized images in radiology has proliferated. It is reasonable to expect that within a few years virtually
More informationAttikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece
Radiation Protection Dosimetry (2005), Vol. 117, No. 1 3, pp. 291 297 doi:10.1093/rpd/nci742 Advance Access published on February 7, 2006 QUALITY ASSURANCE (QA) PROCEDURES FOR SOFTWARE: EVALUATION OF AN
More informationThe future of nuclear imaging is clear
Cardius X-ACT The future of nuclear imaging is clear Increased regulations, growing competition, and concerns about radiation exposure are just a sampling of the current challenges facing the nuclear medicine
More informationReconstruction Filtering in Industrial gamma-ray CT Application
Reconstruction Filtering in Industrial gamma-ray CT Application Lakshminarayana Yenumula *, Rajesh V Acharya, Umesh Kumar, and Ashutosh Dash Industrial Tomography and Instrumentation Section, Isotope Production
More information1571. Artifacts in computer tomography imaging: how it can really affect diagnostic image quality and confuse clinical diagnosis?
1571. Artifacts in computer tomography imaging: how it can really affect diagnostic image quality and confuse clinical diagnosis? Vincentas Veikutis 1, Tomas Budrys 2, Algidas Basevicius 3, Saulius Lukosevicius
More informationSECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS
RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT
More information