Improved Image Quality in Dual- Energy Abdominal CT: Comparison of Iterative Reconstruction in Image Space and Filtered Back Projection Reconstruction

Size: px
Start display at page:

Download "Improved Image Quality in Dual- Energy Abdominal CT: Comparison of Iterative Reconstruction in Image Space and Filtered Back Projection Reconstruction"

Transcription

1 Medical Physics and Informatics Original Research Wang et al. Image Reconstruction Algorithms for Dual-Energy Abdominal CT Medical Physics and Informatics Original Research Rui Wang 1 Wei Yu 1 Runze Wu 2 Hua Yang 3 Dongxu Lu 1 Jiayi Liu 1 Zhaoqi Zhang 1 Chuanchen Zhang 1 Wang R, Yu W, Wu R, et al. Keywords: abdomen, dual-energy CT, image quality, iterative reconstruction DOI: /AJR Received April 27, 2011; accepted after revision December 21, R. Wu is a consultant at Siemens Healthcare. 1 Department of Radiology, Beijing Anzhen Hospital, Capital Medical University, 2 Anzhen Rd, Beijing, China. Address correspondence to Z. Zhang (zhaoqi5000@vip.sohu.com). 2 Siemens Healthcare China, Beijing, China. 3 Affiliated Hospital of Hebei United University, Tangshan, Hebei, China. AJR 2012; 199: X/12/ American Roentgen Ray Society Improved Image Quality in Dual- Energy Abdominal CT: Comparison of Iterative Reconstruction in Image Space and Filtered Back Projection Reconstruction OBJECTIVE. The purpose of this study is to investigate whether an iterative reconstruction in image space (IRIS) algorithm improves the image quality of dual-energy CT abdominal examinations performed during the hepatic arterial phase. MATERIALS AND METHODS. Seventy patients with suspected liver masses underwent contrast-enhanced multiphase abdominal examination and were enrolled in the study. A dual-energy CT protocol was performed in the hepatic arterial phase (parameters: tube A, 140 kv and 90 ma; tube B, 80 kv and 382 ma; automatic tube current modulation on; and collimation, mm). The reconstructions were performed with filtered back projection (FBP) and IRIS algorithms at a slice thickness of 3 mm and kernels of B30 and I30. The image noise was measured on the liver, aorta, and subcutaneous fat on the FBP and IRIS fusion images (m = 0.3) at the same position. The image noise and diagnostic acceptability of all images were scored by two radiologists. RESULTS. The image noise using the IRIS algorithm was lower than that using the standard FBP algorithm on the liver, aorta, and subcutaneous fat, respectively. The signal-tonoise ratio and contrast-to-noise ratio of images reconstructed with the IRIS algorithm also were significantly higher than for those reconstructed with the FBP algorithm. The diagnostic acceptability score using the IRIS algorithm was higher than that using the FBP algorithm at the same dose level (1.20 ± 0.40 vs 1.37 ± 0.57; p < 0.05). CONCLUSION. Compared with standard FBP reconstruction, an IRIS algorithm enables significant reduction of image noise and improvement of image quality and has the potential to decrease radiation exposure during contrast-enhanced dual-energy CT abdominal examination. D ual-energy CT provides the capability to differentiate between different materials at imaging [1]. This capability enables multiple dual-energy CT applications, including the triage of patients with nephrolithiasis on the basis of stone composition [2 4], evaluation of tissue perfusion blood volume [5, 6], the possibility of eliminating a separate unenhanced CT acquisition [7], and improved detection of pathologic hyperenhancement [8, 9]. Dual-energy CT acquisition includes images obtained at two different tube voltages: a high and a low peak kilovoltage. A fusion image of images obtained at two different peak kilovoltages is routinely generated to provide a 120-kVp-equivalent image for clinical use. However, the use of the low-peak-kilovoltage image frequently causes an increase in image noise and beam-hardening artifacts. This may result in a loss of image quality and problems in differential diagnosis. Recently, iterative reconstruction has been reintroduced to reduce radiation dose and improve image quality [10 12]. It has been shown that an iterative reconstruction in image space (IRIS) algorithm enables a significant reduction in image noise without a loss of diagnostic information for chest CT examinations [13]. However, to our knowledge, the IRIS algorithm has not been studied in dual-energy abdominal CT examinations. Therefore, the purpose of the current study was to determine the performance of the IRIS algorithm in improving the quality of fusion images of dual-energy CT abdominal examinations performed during the hepatic arterial phase. Materials and Methods Patient Population From June 2010 to July 2011, 70 patients (28 women and 42 men) with known or suspected liver masses referred for multiphase contrast-enhanced 402 AJR:199, August 2012

2 Image Reconstruction Algorithms for Dual-Energy Abdominal CT abdominal CT examination were retrospectively enrolled in this study. The mean (± SD) age of patients was 60 ± 10 years (range, years), the mean height was ± 7.47 cm (range, cm), the mean weight was ± kg (range, kg), and the mean body mass index was ± 2.90 kg/m 2 (range, kg/ m 2 ). Exclusion criteria were contraindications to iodinated contrast media, overt heart failure (New York Heart Association class III or IV), and renal insufficiency (serum creatine level > 1.4 mg/ dl). The study was approved by the institutional review board, and informed consent was obtained from all participating patients. Abdominal Contrast-Enhanced CT Scan Technique Multiphase abdominal CT was performed on a dual-source CT scanner (SOMATOM Definition, Siemens Healthcare). A dual-energy CT protocol was performed in the arterial phase. The parameters included helical scanning mode (tube A, 140 kv and 90 ma; tube B, 80 kv and 382 ma), automatic tube current modulation (CARE Dose4D, Siemens Healthcare), collimation of mm, pitch of 0.55, and gantry rotation time of 0.5 second. For triggering of the arterial phase acquisition, a bolus-tracking technique was used by placing a region of interest (ROI) at the supraceliac artery, and image acquisition was automatically triggered 12 seconds after the signal attenuation reached the predefined threshold of 100 HU [14]. The venous and equilibrium phases were at fixed time delays of 70 and 180 seconds, respectively. The venous and equilibrium phases were acquired during single-breath-hold helical scanning of the abdomen by using our routine protocol at a tube voltage of 120 kv, a tube current of 190 ma, and automatic tube current modulation. These sets were not included in the quantitative analyses because they were single-energy acquisition data. The volume CT dose index and doselength product were given by the CT scanner, and the effective dose was calculated by multiplying the dose-length product by the conversion coefficient (0.015 msv mgy 1 cm 1 ), leading to an effective dose for dual-energy CT of 3.39 ± 0.84 msv during the hepatic arterial phase. The contrast agent was injected with a dualhead power injector (Stellant D, Medrad) using an 18-gauge IV needle placed in the right antecubital vein. A routine injection protocol was used. According to patient weight, ml/kg of contrast agent (370 mg I/mL iopromide; Ultravist, Bayer Healthcare) was injected, followed by 30 ml of saline (0.9% sodium chloride) as a bolus chaser. The total volume of contrast agent was ml. The injection rate was ml/s. Dual-Energy CT Image Reconstruction The 140- and 80-kVp images and a fusion image (140- and 80-kVp images mixed with a weight factor of 0.3) were reconstructed by using a standard FBP algorithm with 3-mm slice thickness, 1.5-mm increment, and a medium-smooth kernel (D30f). The raw data were further transferred to an offline workstation for IRIS reconstruction. The IRIS images were reconstructed with the same parameters as the FBP algorithm and a Q30f kernel, which matches the resolution of a D30f kernel. In this study, images using five iterations were reconstructed, because image quality was better after five iterations, as described by Pontana et al. [13]. Image Quality Analysis The FBP and IRIS fusion images were loaded into a commercial workstation to evaluate objective and subjective image quality. The two fusion image sets obtained using FBP and IRIS for each patient were displayed side by side with a preset window (window width, 400 HU; window level, 60 HU). According to the European guidelines on quality criteria for CT for abdominal CT examinations [15], two radiologists independently assessed the subjective image quality of all 70 dual-energy CT datasets (70 image sets with FBP and 70 image sets with IRIS) on the workstation over a period of 4 weeks. Visually sharp abdominal structures, image noise, and diagnostic acceptability were assessed by two radiologists. According to the guidelines, each radiologist assessed critical reproduction of visually sharp abdominal structures, including the liver parenchyma and intrahepatic portal veins, liver veins, structures of liver hilum, common hepatic and common bile ducts, gallbladder wall, and aorta. Image noise was assessed on a 3-point scale, where grade 1 meant too little or less than usual noise, grade 2 meant optimum noise, and grade 3 was given for CT examinations with too much noise affecting the image interpretation. Two radiologists assessed independently the overall image diagnostic acceptability (1, excellent; 2, good; 3, acceptable; and 4, unacceptable), according to a procedure described elsewhere [11]. Interobserver disagreements were resolved by consensus in a joint session. Objective Analysis The image noise (SD of CT number) of the liver, aorta, and subcutaneous fat was measured by manually placing circular ROIs at the same image level. The SD of liver was recorded as the mean of measurement of three ROIs, not including any vessel and artifacts in the ROI area. When the SD of the aorta was measured, calcification and plaque on the aortic wall were avoided. The SD of subcutaneous fat was recorded as the mean measurement of three ROIs placed in the anterior, posterior, and lateral abdominal wall. For all measurements, the size, shape, and position of the ROIs were kept constant among the two protocols (Fig. 1). For each protocol, the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated with the following formulas: SNR = ROIm / SDm and CNR = (ROIm ROI fat ) / SD fat, where ROIm is the mean attenuation of liver, aorta, and fat tissue; ROI fat is the mean attenuation of subcutaneous fat; SDm is the mean of the image noise at the liver and aorta; and SD fat is the mean of the subcutaneous fat noise. Statistical Analysis Statistical analysis was performed using SPSS software (version 16.0, SPSS). Continuous variables were described as mean ± SD. The image noise (SD), SNR, and CNR of the FBP and IRIS groups were compared using paired Student t tests. Nonparametric tests (McNemar chi-square test) were used to compare subjective scores of image noise and diagnostic acceptability. A kappa test was used to check the agreement between the results of the two radiologists for subjective quality. The kappa statistics were weighted as follows: 0 0.2, low; , moderate; good, , substantial, and greater than 0.81, perfect agreement, and p values less than 0.05 were considered to be statistically significant. Results Subjective Image Quality There was moderate interobserver agreement between the two radiologists for subjective image quality assessed in all 70 dualenergy abdominal CT image sets (κ = 0.726; p < 0.05). Visually sharp abdominal structures, image noise, and diagnostic acceptability with FBP and IRIS are summarized in Table 1 (see also Figs. 1 and 2). There was a significant reduction in the level of subjective image noise overall for IRIS images compared with FBP images (p < 0.05); furthermore, the diagnostic acceptability score for IRIS images was higher than that for FBP images (1.20 ± 0.40 vs 1.37 ± 0.57; p < 0.05). Objective Image Quality Our results showed that, when the radiation dose was kept constant, the IRIS algorithm yielded significantly decreased image noise for the aorta, liver, and subcutaneous fat, respectively, and increased SNR and CNR for the aorta and liver and increased SNR for subcutaneous fat, respectively, compared with the standard FBP algorithm (all p < 0.001; Table AJR:199, August

3 Wang et al. 2 and Figs. 1 and 2). However, there was no statistically significant difference in CT values between IRIS and FBP images (all p > 0.05; Table 2). Discussion To our knowledge, this is the first study to evaluate the performance of an IRIS algorithm in abdominal dual-energy CT. Compared with FBP and IRIS imaging from TABLE 1: Comparison of Subjective Image Quality Between Filtered Back Projection (FBP) and Iterative Reconstruction in Image Space (IRIS) Algorithms Variable FBP IRIS Visual sharpness Positive Positive Image noise a Diagnostic acceptability b Note Data are no. of images assigned a particular grade for image noise and diagnostic acceptability. a Image noise was graded on a 3-point scale: grade 1, too little or less than usual noise; grade 2, optimum noise; and grade 3, too much noise affecting the image interpretation. b Overall image diagnostic acceptability was graded on a 4-point scale: 1, excellent; 2, good; 3, acceptable; and 4, unacceptable. Fig year-old woman. A and B, Axial contrast-enhanced dual-energy CT images obtained during hepatic arterial phase processed with iterative reconstruction in image space (A) and filtered back projection (B) algorithms show regions of interest manually drawn on aorta, liver, and subcutaneous fat on same level. Fig year-old woman with small hepatic cavernous hemangioma and two cysts. A and B, Contrast-enhanced dual-energy CT images obtained during hepatic arterial phase processed with iterative reconstruction in image space (IRIS) (A) and filtered back projection (FBP) (B) algorithms show significantly reduced image noise and improved image quality with IRIS compared with FBP. IRIS can enhance spatial resolution at edge of highcontrast tissues (arrowheads), maintaining boundary sharpness. Small vessel near cystic lesion (arrow) is shown clearly on IRIS compared with FBP. A A the same datasets, our results show that the IRIS algorithm significantly decreases image noise and improves image quality for abdominal dual-energy CT fusion images obtained during the hepatic arterial phase. A dual-energy CT scan generates three datasets by default: high peak kilovoltage, low peak kilovoltage, and a fusion image. The fusion image is a linear mix of the highand low-peak-kilovoltage images and is usually considered a 120-kVp-equivalent image. These images are routinely used for the purpose of diagnosis and the starting point for advanced dual-energy postprocessing. Therefore, maintaining sufficient image quality of the fusion image is desired. The IRIS algorithm introduces a loop of image correction in the image space, which iteratively reduces image noise while maintaining boundary sharpness [13]. It was found that IRIS can enhance spatial resolution at the edge of high-contrast tissues and reduce image noise on homogeneous areas at the same radiation dose [13]. In our study, 80- and 140-kVp image data, the fusion images, and virtual image data were reconstructed using the IRIS algorithm. Consistent with the results of previous studies [10 13, 16, 17], the image quality of these image sets was significantly improved by using IRIS instead of FBP. In the present study, IRIS reduced the image noise better than FBP did at the same doses level (p < 0.001). At the same time, IRIS yielded significantly higher SNR and CNR on the liver parenchyma and aorta than did FBP (p < 0.001). Clinically, it is vitally important that subjective assessments of image quality by radiologists are improved through various reconstruction algorithms. Our results show that image noise and diagnostic acceptability of images were significantly improved with the IRIS reconstruction algorithm over the FBP B B 404 AJR:199, August 2012

4 Image Reconstruction Algorithms for Dual-Energy Abdominal CT TABLE 2: Comparison of Objective Image Quality Between Filtered Back Projection (FBP) and Iterative Reconstruction in Image Space (IRIS) Algorithms 140-kV Image 80-kV Image Fusion Image a Virtual Unenhanced Image Variable FBP IRIS FBP IRIS FBP IRIS FBP IRIS p Liver CT value ± ± ± ± ± ± ± ± 5.11 > 0.05 Liver noise (HU) ± ± ± ± ± ± ± ± 1.07 < Liver SNR 3.29 ± ± ± ± ± ± ± ± 1.67 < Liver CNR ± ± ± ± ± ± ± ± 8.94 < Aorta CT value ± ± ± ± ± ± ± ± > 0.05 Aorta noise (HU) ± ± ± ± ± ± ± ± 1.76 < Aorta SNR ± ± ± ± ± ± ± ± 1.92 < Aorta CNR ± ± ± ± ± ± ± ± 8.59 < Fat noise (HU) ± ± ± ± ± ± v ± 2.59 < Fat SNR 6.17 ± ± ± ± ± ± ± ± 2.59 < Note Data are mean ± SD. CNR = contrast-to-noise ratio, HU = Hounsfield units, SNR = signal-to-noise ratio. a Fusion image is created by merging 140- and 80-kV datasets (m = 0.3) to yield a virtual 120-kV image. algorithm. A comparison of the two reconstruction algorithms showed better boundary sharpness of lesions on IRIS images, confirming that reduction of the objective image noise reinforces the diagnostic accuracy. Although the so-called plastic image impression in iterative reconstruction has been reported [18], the images reconstructed with IRIS are easily appreciated. However, the time needed for an iterative reconstruction was significantly longer than that needed for FBP reconstruction. The next-generation iterative algorithm with hardware acceleration will be desired for routine clinical usage. Our findings support that IRIS has important clinical applications. Compared with the standard FBP algorithm at the same level of radiation dose, the IRIS algorithm can reduce image noise and improve the image quality. Alternatively, the application of the IRIS technique could be another exercise for decreasing or optimizing radiation doses without losing the overall image quality [17] because the radiation dose of CT examinations and its associated risks are major concerns [19]. Thus, strategies aimed at lowering radiation dose have become particularly important. There were several limitations to this study. First, compared with standard FBP, IRIS significantly improved image noise and image quality, but these results reflect only a preliminary experience with a relatively small population. Second, we limited our assessment to the effect of IRIS I30 on fusion images of dual-energy CT. We did not assess whether alternative iteration would have resulted in substantially different results. In future work, we will evaluate a wide range of iterations on fusion image quality. Third, the assessment of the IRIS technique in this study, comparing it with standard FBP, was based on the same patients and the same levels of radiation dose. Future investigations of the effectiveness of new iterative reconstruction will be applied to reduced radiation doses, comparing the IRIS technique with standard FBP. In summary, this study shows that image quality is significantly improved in dual-energy CT abdominal examination using the IRIS algorithm compared with standard FBP. IRIS can be routinely applied on these studies, potentially reducing radiation exposure. Acknowledgments We thank the technical staff members of the CT department for invaluable contributions, particularly Hong Jiang, Weihua Zhang, Honghong Tie, Zhimin Li, Miao Guo, and Haixia Yang for technical assistance. References 1. Johnson TR, Krauss B, Sedlmair M, et al. Material differentiation by dual energy CT: initial experience. Eur Radiol 2007; 17: Takahashi N, Hartman RP, Vrtiska TJ, et al. Dualenergy CT iodine-subtraction virtual unenhanced technique to detect urinary stones in an iodinefilled collecting system: a phantom study. AJR 2008; 190: Graser A, Johnson TR, Bader M, et al. Dual energy CT characterization of urinary calculi: initial in vitro and clinical experience. Invest Radiol 2008; 43: Primak AN, Fletcher JG, Vrtiska TJ, et al. Noninvasive differentiation of uric acid versus nonuric acid kidney stones using dual-energy CT. Acad Radiol 2007; 14: Thieme SF, Becker CR, Hacker M, Nikolaou K, Reiser MF, Johnson TR. Dual energy CT for the assessment of lung perfusion: correlation to scintigraphy. Eur J Radiol 2008; 68: Ruzsics B, Lee H, Zwerner PL, Gebregziabher M, Costello P, Schoepf UJ. Dual-energy CT of the heart for diagnosing coronary artery stenosis and myocardial ischemia: initial experience. Eur Radiol 2008; 18: Stolzmann P, Frauenfelder T, Pfammatter T, et al. Endoleaks after endovascular abdominal aortic aneurysm repair: detection with dual-energy dual-source CT. Radiology 2008; 249: Schindera ST, Nelson RC, Mukundan S Jr, et al. Hypervascular liver tumors: low tube voltage, high tube current multi-detector row CT for enhanced detection phantom study. Radiology 2008; 246: Graser A, Johnson TR, Chandarana H, Macari M. Dual energy CT: preliminary observations and potential clinical applications in the abdomen. Eur Radiol 2009; 19: Marin D, Nelson RC, Schindera ST, et al. Lowtube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm initial clinical experience. Radiology 2010; 254: Prakash P, Kalra MK, Kambadakone AK, et al. Reducing abdominal CT radiation dose with adaptive statistical iterative reconstruction technique. Invest Radiol 2010; 45: Singh S, Kalra MK, Hsieh J, et al. Abdominal CT: comparison of adaptive statistical iterative and fil- AJR:199, August

5 Wang et al. tered back projection reconstruction techniques. Radiology 2010; 257: Pontana F, Pagniez J, Flohr T, et al. Chest computed tomography using iterative reconstruction vs filtered back projection. Part 1. Evaluation of image noise reduction in 32 patients. Eur Radiol 2011; 21: Sultana S, Awai K, Nakayama Y, et al. Hypervascular hepatocellular carcinomas: bolus tracking with a 40-detector CT scanner to time arterial phase imaging. Radiology 2007; 243: Menzel HG, Schibilla H, Teunen D, eds; European Commission s Radiation Protection Actions. Quality criteria for computed tomography. EUR Danish Society of Radiology Web site. www. drs.dk/guidelines/ct/quality/download/eur w51. Accessed December 29, Thibault JB, Sauer KD, Bouman CA, Hsieh J. A three-dimensional statistical approach to improved image quality for multislice helical CT. Med Phys 2007; 34: Pontana F, Duhamel A, Pagniez J, et al. Chest computed tomography using iterative reconstruction vs filtered back projection (Part 2): image quality of low-dose CT examinations in 80 patients. Eur Radiol 2011; 21: Leipsic J, Labounty TM, Heilbron B, et al. Adaptive statistical iterative reconstruction: assessment of image noise and image quality in coronary CT angiography. AJR 2010; 195: Brenner DJ, Hall EJ. Computed tomography: an increasing source of radiation exposure. N Engl J Med 2007; 357: FOR YOUR INFORMATION Malpractice Issues in Radiology, 3rd edition, by Leonard Berlin, is now available! For more information or to purchase a copy, see AJR:199, August 2012

Abdominal Radiology ISSN X. Abdom Radiol DOI /s

Abdominal Radiology ISSN X. Abdom Radiol DOI /s Assessment of 70-keV virtual monoenergetic spectral images in abdominal CT imaging: A comparison study to conventional polychromatic 120- kvp images Negin Rassouli, Hamid Chalian, Prabhakar Rajiah, Amar

More information

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National

More information

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

Iterative Reconstruction in Image Space. Answers for life.

Iterative Reconstruction in Image Space. Answers for life. Iterative Reconstruction in Image Space Answers for life. Iterative Reconstruction in Image Space * (IRIS) * Please note: IRIS is used as an abbreviation for Iterative Reconstruction in Image Space throughout

More information

Dual-Source Dual-Energy CT With Additional Tin Filtration: Dose and Image Quality Evaluation in Phantoms and In Vivo

Dual-Source Dual-Energy CT With Additional Tin Filtration: Dose and Image Quality Evaluation in Phantoms and In Vivo Medical Physics and Informatics Original Research Primak et al. Dose and Image Quality of Dual-Energy DSCT Medical Physics and Informatics Original Research Andrew N. Primak 1,2 Juan Carlos Ramirez Giraldo

More information

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

More information

160-slice CT SCANNER / New Standard for the Future

160-slice CT SCANNER / New Standard for the Future TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Poster No.: C-2205 Congress: ECR 2012 Type: Scientific Paper

More information

Electronic Noise in CT Detectors: Impact on Image Noise and Artifacts

Electronic Noise in CT Detectors: Impact on Image Noise and Artifacts Medical Physics and Informatics Original Research Duan et al. Electronic Noise in CT Detectors Medical Physics and Informatics Original Research Xinhui Duan 1 Jia Wang 1,2 Shuai Leng 1 ernhard Schmidt

More information

Wide-Detector CT for TAVR Planning:

Wide-Detector CT for TAVR Planning: Wide-Detector CT for TAVR Planning: Impact on Iodine Dose, Radiation Dose, and Image Quality SCBTMR 2015 Annual Course Thursday, October 8 William P. Shuman MD FSCBTMR Department of Radiology University

More information

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT

More information

Improvement of CT image quality with iterative reconstruction idose4

Improvement of CT image quality with iterative reconstruction idose4 Improvement of CT image quality with iterative reconstruction idose4 Poster No.: C-0387 Congress: ECR 2014 Type: Scientific Exhibit Authors: M.-L. Olsson, K. Norrgren, M. Söderberg; Malmö/SE Keywords:

More information

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) Hiroki Kadoya Yukiko Kitagawa MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria

More information

Iterative Reconstruction

Iterative Reconstruction RECENT ADVANCES IN CT RADIATION DOSE REDUCTION TECHNIQUES Iterative Reconstruction Kalpana Kanal, PhD, FSCBTMR, FACR, FAAPM Professor and Director, Diagnostic Physics Section University of Washington Seattle,

More information

Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise

Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise Jonas Rydberg, M.D. Professor of Radiology Indiana University School of Medicine Indianapolis, Indiana Medical

More information

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2

Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2 Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2 Medical Consult, SA; Establish a method to correlate image

More information

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life.

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life. Neuro Thoracic Abdominal Abdominal Cardiovascular Pediatric SAFIRE Sinogram Affirmed Iterative Reconstruction Answers for life. SAFIRE * (Sinogram Affirmed Iterative Reconstruction) * The information

More information

Maximum Performance, Minimum Space

Maximum Performance, Minimum Space TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

Dose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation

Dose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation Medical Physics and Informatics Original Research Christner et al. CT Dose Reduction Medical Physics and Informatics Original Research Downloaded from www.ajronline.org by 8.243.133.8 on 2/26/18 from IP

More information

Suppression of metal artifacts using image-based monoenergetic DECT imaging

Suppression of metal artifacts using image-based monoenergetic DECT imaging Suppression of metal artifacts using image-based monoenergetic DECT imaging Poster No.: C-0519 Congress: ECR 2011 Type: Scientific Paper Authors: B. Krauss, B. Schmidt, M. Sedlmair, T. Flohr; Forchheim/DE

More information

QC by the MPE in Belgium

QC by the MPE in Belgium Acceptance testing of state-of-the-art CT scanners using a new national protocol: first experience on a large number of scanners of different make and model the working group Radiology of the Belgian Hospital

More information

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

1. Patient size AEC. Large Patient High ma. Small Patient Low ma Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description

More information

Potential clinical usefulness and limitation of fast kv switch Dual energy CT

Potential clinical usefulness and limitation of fast kv switch Dual energy CT Potential clinical usefulness and limitation of fast kv switch Dual energy CT Poster No.: C-1859 Congress: ECR 2011 Type: Educational Exhibit Authors: M. Jinzaki, Y. Tanami, H. Sugiura, M. inoue, S. Koga,

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT) Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

CT Data Storage Reduction by Means of Compressing Projection Data Instead of Images: Feasibility Study 1

CT Data Storage Reduction by Means of Compressing Projection Data Instead of Images: Feasibility Study 1 Kyongtae T. Bae, MD, PhD Bruce R. Whiting, PhD Index terms: Computed tomography (CT), image display and recording Computed tomography (CT), image processing Computed tomography (CT), technology Data compression

More information

Reducing Radiation Exposure from Survey CT Scans

Reducing Radiation Exposure from Survey CT Scans Reducing Survey CT Scan Exposure Pediatric Imaging Original Research Jennifer C. O Daniel 1 Donna M. Stevens 2 Dianna D. Cody 2 O Daniel JC, Stevens DM, Cody DD Received July 28, 2004; accepted after revision

More information

Detector technology in simultaneous spectral imaging

Detector technology in simultaneous spectral imaging Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become

More information

Software and Hardware in CCTA. Elly Castellano PhD

Software and Hardware in CCTA. Elly Castellano PhD Software and Hardware in CCTA Elly Castellano PhD Outline technical requirements for coronary CTA the modern cardiac CT scanner ECG-gating technology image reconstruction algorithms 2 Technical requirements

More information

diagnostic examination

diagnostic examination RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

TISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY

TISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY TISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY By RYAN F. FISHER A THESIS PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT

More information

Aquilion Precision Ultra-High Resolution CT: Quantifying diagnostic image quality

Aquilion Precision Ultra-High Resolution CT: Quantifying diagnostic image quality Aquilion Precision Ultra-High CT: Quantifying diagnostic image quality Kirsten Boedeker, PhD, DABR Senior Manager, Quantitative Image Quality Canon Medical Systems Corporation Introduction Over the last

More information

Integrated Circuit Detector Technology in Abdominal CT: Added Value in Obese Patients

Integrated Circuit Detector Technology in Abdominal CT: Added Value in Obese Patients Medical Physics and Informatics Original Research Morsbach et al. CT in Obese Patients Medical Physics and Informatics Original Research Fabian Morsbach 1 Sebastian ickelhaupt 1 Susan Rätzer 1 ernhard

More information

Attikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece

Attikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece Radiation Protection Dosimetry (2005), Vol. 117, No. 1 3, pp. 291 297 doi:10.1093/rpd/nci742 Advance Access published on February 7, 2006 QUALITY ASSURANCE (QA) PROCEDURES FOR SOFTWARE: EVALUATION OF AN

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

Measurement of table feed speed in modern CT

Measurement of table feed speed in modern CT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Measurement of table feed speed in modern CT Atsushi Fukuda, 1,2a Pei-Jan P. Lin, 3 Kosuke Matsubara, 2 Tosiaki Miyati 2 Department

More information

Virtual monochromatic imaging in dual-source and dual-energy CT for visualization of acute ischemic stroke

Virtual monochromatic imaging in dual-source and dual-energy CT for visualization of acute ischemic stroke Virtual monochromatic imaging in dual-source and dual-energy CT for visualization of acute ischemic stroke Hidetake Hara and Hiroshi Muraishi School of Allied Health Sciences, Kitasato University, Japan

More information

Quantitation of clinical feedback on image quality differences between two CT scanner models

Quantitation of clinical feedback on image quality differences between two CT scanner models Received: 4 August 2016 Revised: 4 November 2016 Accepted: 12 December 2016 DOI: 10.1002/acm2.12050 MEDICAL IMAGING Quantitation of clinical feedback on image quality differences between two CT scanner

More information

Overview of Safety Code 35

Overview of Safety Code 35 Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2

Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2 Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2 1. Department of Medical Physics, University of Wisconsin-Madison 2. Department

More information

Organ-Specific Context-Sensitive Single and Dual Energy CT (DECT) Image Reconstruction, Display and Analysis

Organ-Specific Context-Sensitive Single and Dual Energy CT (DECT) Image Reconstruction, Display and Analysis Organ-Specific Context-Sensitive Single and Dual Energy CT (DECT) Image Reconstruction, Display and Analysis Sabrina Dorn 1, Shuqing Chen², Francesco Pisana 1, Joscha Maier 1, Michael Knaup 1, Stefan Sawall

More information

January 2012 February 1, Release Date: Expiration Date:

January 2012 February 1, Release Date: Expiration Date: CT for Technologists is a training program designed to meet the needs of radiologic technologists entering or working in the field of computed tomography (CT). This series is designed to augment classroom

More information

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology The Fundamentals of... Computed Tomography Computed Tomography (CT) systems use x-rays to produce images of slices through a patient s anatomy. Despite having lower spatial resolution than other x-ray

More information

Explain what is meant by a photon and state one of its main properties [2]

Explain what is meant by a photon and state one of its main properties [2] 1 (a) A patient has an X-ray scan taken in hospital. The high-energy X-ray photons interact with the atoms inside the body of the patient. Explain what is meant by a photon and state one of its main properties....

More information

Architecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare

Architecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare Architecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare 2 DOC1292532 Architecture of Quality Imaging Agile Acoustic Architecture E-Series and XDclear

More information

Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT)

Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT) Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT) F Edward Boas, Roland Bammer, and Dominik Fleischmann Extended abstract for RSNA 2012 Purpose CT metal streak artifacts are

More information

SOMATOM Esprit A Bundle of Energy

SOMATOM Esprit A Bundle of Energy SOMATOM Esprit A Bundle of Energy DATA SOMATOM Esprit An economical CT scanner designed for...... Excellent spiral image quality... A wide range of clinical applications... Value performance and reliabilty

More information

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard.

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard. Chiara Secco PET Performance measurements of the new LSO-Based Whole Body PET/CT Scanner biograph 16 HI-REZ using the NEMA NU 2-2001 Standard. INTRODUCTION Since its introduction, CT has become a fundamental

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

2D, 3D CT Intervention, and CT Fluoroscopy

2D, 3D CT Intervention, and CT Fluoroscopy 2D, 3D CT Intervention, and CT Fluoroscopy SOMATOM Definition, Definition AS, Definition Flash Answers for life. Siemens CT Vision Siemens CT Vision The justification for the existence of the entire medical

More information

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Wataru FUKUDA* Junya MORITA* and Masahiko YAMADA* Abstract Tomosynthesis is a three-dimensional imaging technology

More information

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them. In press 2004 1 2 Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Digital radiology An appropriate analogy that is easy for most people to understand

More information

Ludlum Medical Physics

Ludlum Medical Physics Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to

More information

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL TORNIER BLUEPRINT 3D Planning + PSI SCAN PROTOCOL Contents 3 Introduction 3 Patient preparation 3 Scanning instructions 4 Image instructions 5 Scanning parameters 6 Technical instructions 2 BLUEPRINT 3D

More information

PET/CT Instrumentation Basics

PET/CT Instrumentation Basics / Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire

More information

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results NeuViz 16 Computed Tomography Elevating routine imaging for exceptional results Essence NeuViz 16 Raising the bar on clinical utility in routine imaging. Get more. More clinical information for patients.

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Tomosynthesis and Motion

Tomosynthesis and Motion Tomosynthesis (3D) Motion Unsharpness Occurs at about the same frequency as conventional mammography (2D) Presents the same issues as 2D motion, EXCEPT that motion may go undetected Most common patient-related

More information

Metal Artifact Reduction for Orthopedic Implants (O-MAR)

Metal Artifact Reduction for Orthopedic Implants (O-MAR) Metal Artifact Reduction for Orthopedic Implants () Summary Since the inception of CT, numerous methods have been proposed to suppress metal artifacts with varying degrees of success. 1-4 (Metal Artifact

More information

Original Article Thoracic Imaging. Ki Baek Lee, RT, MS, Hyun Woo Goo, MD, PhD INTRODUCTION

Original Article Thoracic Imaging. Ki Baek Lee, RT, MS, Hyun Woo Goo, MD, PhD INTRODUCTION Original Article Thoracic Imaging https://doi.org/10.3348/kjr.2018.19.1.119 pissn 1229-6929 eissn 2005-8330 Korean J Radiol 2018;19(1):119-129 Quantitative Image Quality and Histogram-Based Evaluations

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

Radiographic Techniques, Contrast, and Noise in X-Ray Imaging

Radiographic Techniques, Contrast, and Noise in X-Ray Imaging Residents Section Physics Minimodule Huda and Abrahams Techniques, Contrast, and Noise in Radiography Residents Section Physics Minimodule Residents inradiology Walter Huda 1 R. Brad Abrahams 2 Huda W,

More information

Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution

Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Cynthia H. McCollough, PhD, DABR, FAAPM, FACR Director, CT Clinical Innovation Center Professor of Medical Physics and

More information

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC? Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM

More information

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.125 pissn 2508-4445, eissn 2508-4453 Optimization of Energy Modulation Filter for Dual Energy CBCT

More information

CT Basics: Data Acquisition Module 3

CT Basics: Data Acquisition Module 3 Module 3 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

COCIR SELF-REGULATORY INITIATIVE FOR MEDICAL IMAGING EQUIPMENT COMPUTED TOMOGRAPHY MEASUREMENT OF ENERGY CONSUMPTION

COCIR SELF-REGULATORY INITIATIVE FOR MEDICAL IMAGING EQUIPMENT COMPUTED TOMOGRAPHY MEASUREMENT OF ENERGY CONSUMPTION COCIR SELF-REGULATORY INITIATIVE FOR MEDICAL IMAGING EQUIPMENT COMPUTED TOMOGRAPHY MEASUREMENT OF ENERGY CONSUMPTION Revision: 1 Date: June 2015 Approved: June 2015 TABLE OF CONTENT 1. INTRODUCTION...

More information

Translating Protocols Between Scanner Manufacturer and Model

Translating Protocols Between Scanner Manufacturer and Model Translating Protocols Between Scanner Manufacturer and Model Robert J. Pizzutiello, MS, FAAPM, FACMP Sr. Vice-President, Global Physics Solutions President, Upstate Medical Physics Objectives Understand

More information

PET: New Technologies & Applications, Including Oncology

PET: New Technologies & Applications, Including Oncology PET: New Technologies & Applications, Including Oncology, PhD, FIEEE Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Disclosures Research Contract, GE Healthcare

More information

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging

More information

The Script of ZST + Presentation. MIS Upstream Marketing Team [ 日期 ]

The Script of ZST + Presentation. MIS Upstream Marketing Team [ 日期 ] 1 The Script of ZST + Presentation MIS Upstream Marketing Team [ 日期 ] 1 The Script of ZST + Presentation Since Mindray was founded to develop ultrasound business, core technology has always been the engine

More information

Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing. Koichi Hirokawa MEDIX VOL. 56 P.43 P.46

Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing. Koichi Hirokawa MEDIX VOL. 56 P.43 P.46 Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing Taiga Goto Koichi Hirokawa Hisashi Takahashi MEDIX VOL. 56 P.43 P.46 Advanced Noise Reduction Processing for X-ray CT System

More information

Brilliance in everything Philips CT products and services

Brilliance in everything Philips CT products and services Brilliance in everything Philips CT products and services Ready for anything No one does more than Philips to help you gain the productivity you need with a comprehensive approach to CT that marries significant

More information

T h e P h a n t o m L a b o r a t o r y

T h e P h a n t o m L a b o r a t o r y T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order

More information

Industry Breakthrough

Industry Breakthrough Industry Breakthrough Dynamic SPECT Acquisition Quantifying Myocardial Blood Flow Nuclear Cardiology in the 21st Century In the 21st century, most nuclear cameras are still relying on a technology invented

More information

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT A comparison of two methods for the determination of freein-air geometric efficiency in MDCT Theocharis Berris *1, Kostas Perisinakis 1,, Antonios E. Papadakis and John Damilakis 1, 1 Department of Medical

More information

GSI Xtream on RevolutionTM CT

GSI Xtream on RevolutionTM CT GE Healthcare Technology White Paper GSI Xtream on RevolutionTM CT Volume. Spectral. Simplified. Authors: Scott Slavic, M.S Priti Madhav, Ph.D. Mark Profio, M.S Dominic Crotty, Ph.D. Elizabeth Nett, Ph.D.

More information

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools

More information

Introduction. Parametric Imaging. The Ultrasound Research Interface: A New Tool for Biomedical Investigations

Introduction. Parametric Imaging. The Ultrasound Research Interface: A New Tool for Biomedical Investigations The Ultrasound Research Interface: A New Tool for Biomedical Investigations Shelby Brunke, Laurent Pelissier, Kris Dickie, Jim Zagzebski, Tim Hall, Thaddeus Wilson Siemens Medical Systems, Issaquah WA

More information

Gastrointestinal Imaging Original Research

Gastrointestinal Imaging Original Research Gastrointestinal Imaging Original Research Lee et al. Low-Dose CT in Cases of Suspected Appendicitis Gastrointestinal Imaging Original Research Yoon Jin Lee 1 Bohyoung Kim 1 Yousun Ko 1 Kyung Eun Cho 1

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

SOMATOM Sensation 4 Computed Tomography System for Multislice Spiral Scanning

SOMATOM Sensation 4 Computed Tomography System for Multislice Spiral Scanning SOMATOM Sensation 4 Computed Tomography System for Multislice Spiral Scanning SOMATOM Sensation 4 Computed Tomography System for Multislice Spiral Scanning Siemens has been a global player in CT for more

More information

Radiation Dose Modulation. the Multidetector CT Era: From Basics to Practice 1

Radiation Dose Modulation. the Multidetector CT Era: From Basics to Practice 1 Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. EDUCATION EXHIBIT

More information

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography Journal of Physics: Conference Series OPEN ACCESS The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography To cite this article: N A A Daud et al 2014 J. Phys.:

More information

Digital Breast Tomosynthesis

Digital Breast Tomosynthesis Digital Breast Tomosynthesis OLIVE PEART MS, RT(R) (M) HTTP://WWW.OPEART.COM 2D Mammography Not 100% effective Limited by tissue superimposition Overlapping tissue can mask tumors False negative Overlapping

More information

Computerized Medical Imaging and Graphics

Computerized Medical Imaging and Graphics Computerized Medical Imaging and Graphics 36 (2012) 387 395 Contents lists available at SciVerse ScienceDirect Computerized Medical Imaging and Graphics jo ur n al homep age : www.elsevier.com/locate/compmedimag

More information

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice nww.medical-devices.nhs.

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice   nww.medical-devices.nhs. abc September 2004 MHRA 04099 Philips Mx8000 IDT CT scanner technical evaluation Best choice best practice www.mhra.gov.uk nww.medical-devices.nhs.uk About MHRA evaluation reports. What you can expect.

More information

Appropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis

Appropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis Indian Journal of Science and Technology Vol 8(S8), 380-386, April 2015 ISSN (Print) : 0974-6846 ISSN (Online) : 0974-5645 DOI: 10.17485/ijst/2015/v8iS8/70528 Appropriate Inspection Distance of Digital

More information

Improved compressibility of multislice CT datasets using 3D JPEG2000 compression

Improved compressibility of multislice CT datasets using 3D JPEG2000 compression International Congress Series 1268 (2004) 57 62 www.ics-elsevier.com Improved compressibility of multislice CT datasets using 3D JPEG2000 compression Khan M. Siddiqui a,b, *, Eliot L. Siegel a,c, Bruce

More information

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System Manus

More information

2 nd generation TOMOSYNTHESIS

2 nd generation TOMOSYNTHESIS 2 nd generation TOMOSYNTHESIS 2 nd generation DBT true innovation in breast imaging synthesis graphy Combo mode Stereotactic Biopsy Works in progress: Advanced Technology, simplicity and ergonomics Raffaello

More information

Industry Breakthrough

Industry Breakthrough Industry Breakthrough Dynamic SPECT Acquisition Quantifying Myocardial Blood Flow D-S P EC T Cardiac Imaging System Nuclear Cardiology in the 21st Century In the 21st century, most nuclear cameras are

More information