Radiation Dose Modulation. the Multidetector CT Era: From Basics to Practice 1

Size: px
Start display at page:

Download "Radiation Dose Modulation. the Multidetector CT Era: From Basics to Practice 1"

Transcription

1 Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at EDUCATION EXHIBIT Radiation Dose Modulation Techniques in the Multidetector CT Era: From Basics to Practice TEACHING POINTS See last page Chang Hyun Lee, MD Jin Mo Goo, MD Hyun Ju Lee, MD Sung-Joon Ye, PhD Chang Min Park, MD Eun Ju Chun, MD Jung-Gi Im, MD Radiation exposure to the patient has become a concern for the radiologist in the multidetector computed tomography (CT) era. With the introduction of faster multidetector CT scanners, various techniques have been developed to reduce the radiation dose to the patient; one method is automatic exposure control (AEC). AEC systems make use of different types of control, including patient-size AEC, z-axis AEC, rotational or angular AEC, or a combination of two or more of these types. AEC systems operate on the basis of several methods: standard deviation, noise index, reference milliamperage, and reference image. A clear understanding of how to use different AEC systems on different multidetector CT scanners will allow users to modulate radiation dose, reduce photon starvation artifacts, and maintain image quality throughout the body. Further development of AEC systems and their successful introduction into clinical practice will require user education and good communication between users and manufacturers. RSNA, 2008 radiographics.rsnajnls.org Abbreviations: ACS = automatic current selection, AEC = automatic exposure control, DLP = dose-length product RadioGraphics 2008; 28: Published online /rg Content Codes: 1 From the Department of Radiology, Seoul National University College of Medicine and Institute of Radiation Medicine, 28 Yeongeon-dong, Jongno-gu, Seoul , South Korea (C.H.L., J.M.G., H.J.L., C.M.P., E.J.C., J.-G.I.); Gangnam Healthcare Center, Seoul National University Hospital, Seoul, South Korea (C.H.L.); and Department of Radiation Oncology, Seoul National University Medical Research Center, Seoul, South Korea (S.-J.Y.). Presented as an education exhibit at the 2006 RSNA Annual Meeting. Received April 17, 2007; revision requested July 23 and received September 4; accepted November 7. All authors have no financial relationships to disclose. Address correspondence to J.M.G. ( jmgoo@plaza.snu.ac.kr). RSNA, 2008

2 1452 September-October 2008 RG Volume 28 Number 5 Introduction Multidetector computed tomography (CT) scanners have led to greatly expanded clinical use of CT. With the simultaneous scanning of up to 64 sections, technologic advances have further improved the volume coverage, z resolution, and scanning speed of CT scanners. The result has been a considerable increase in the number of examinations and in the average scanned volume obtained per examination. In the past 25 years, the number of nuclear medicine studies has nearly tripled, but the number of CT studies has increased more than twentyfold (1). Unfortunately, these improvements in imaging have also inevitably led to an increase in radiation exposure to the patient. In the United States, citizens are being exposed to nearly six times more radiation from medical devices compared with 1980 levels, and CT accounted for 45% of the U.S. population s collective medical radiation exposure, even though it made up only 12% of all medical radiation procedures performed in the United States (2). This radiation exposure from CT is especially critical in pediatric patients and young adult females (3 5). Therefore, the increase in radiation exposure from CT examinations has been of concern to radiologists, medical physicists, government regulators, and CT equipment manufacturers. To cope with this increase in radiation exposure, scanner manufacturers have implemented their own dose modulation techniques to appropriately manage or reduce radiation doses. Apart from the various scanner geometry and noise filtering techniques, tube current modulation is one of the more effective methods of controlling radiation dose (6). However, with no set standard, CT equipment manufacturers currently provide different methods of applying automatic exposure control (AEC) (7). Thus, if these AEC systems are to be used properly, radiologists must be aware of the characteristics of dose modulation techniques with different multidetector CT scanners and learn how to apply these techniques to the patient. In this article, we discuss and illustrate the basic radiation dose parameters on multidetector CT scanners, types of AEC, control of AEC systems on different scanners, determination of protocols with AEC systems, and clinical applications of AEC systems. Reading Radiation Dose Parameters on CT Consoles Radiation Exposure The radiation exposure (expressed in coulombs per kilogram) is defined as the total charge produced in dry air when all electrons liberated by photons in a unit mass of air are completely stopped in air. The quantity must always be defined with respect to the specific material in which the interactions are taking place (eg, air kerma [kinetic energy released per unit mass], water kerma). From radiation exposure, one can calculate the skin entrance dose, which is important for deterministic effects such as skin erythema. These effects can be of potential concern in CT fluoroscopy or repeated multidetector CT and angiography, although they are not usually encountered at routine CT (8,9). Absorbed Dose For diagnostic x-rays, air kerma is the same as the absorbed dose delivered to a given volume of air in the absence of scatter. Absorbed dose represents the energy (in grays) absorbed per unit mass within an object. Absorbed dose does not take into account the differences in radiation sensitivity between organs and cannot provide an estimate of whole-body radiation risk. Equivalent dose (in sieverts) in tissue that incorporates tissue weighting factors has the same numeric value as absorbed dose because the radiation weighting factor is 1 for x-rays (5). CT Dose Index The CT dose index (10) is the most commonly used dose indicator. Display of this information on the user interface of modern CT scanners is required by European Union regulations, but outside the European Union some manufacturers provide display of the CT dose index only when required by the user (Fig 1). This allows direct comparison of the radiation dose at different scanning parameter settings, even between scanners made by different manufacturers (10). However, the CT dose index does not indicate the precise dose for any individual patient, but is rather an index of the dose as measured and calculated in a polymethylmethacrylate phantom. Although the CT dose index is a valuable tool for protocol comparison, it does not take into

3 RG Volume 28 Number 5 Lee et al 1453 Figure 1. Chart shows the basic dose parameters (as determined with an AEC system) displayed on the console of a Siemens multidetector CT scanner. Note that the reference milliamperage has been set at 100. From the volume CT dose index and dose-length product (DLP), the effective dose can be roughly calculated as DLP conversion factor. csl = section collimation (not section width), mas = average applied milliamperage, ref. = quality reference milliamperage, TI = rotation time, Total DLP = DLP value of the entire examination, Total mas = actual value of the entire examination in milliamperes. weighted CT dose index has to be corrected by the pitch factor (dose index divided by pitch) and is then termed the volume CT dose index. Figure 2. Photograph shows instruments used to determine the CT dose index. Plexiglas body and head phantoms are placed on a CT table. The ionization chamber probe is inserted into the center of the body phantom to measure the central CTDI 100 (CT dose index; subscript denotes that the measurement was made with a 100-mm ionization chamber). account patient-associated parameters such as size, shape, and inhomogeneous composition. For any given scanning technique, patient dose depends on the size and attenuation of the patient (ie, the greater the patient attenuation, the smaller the patient dose). Therefore, the displayed CT dose index is smaller than the actual dose delivered to young children and infants (11). The CT dose index is now commonly measured from one axial rotation of the scanner with use of a 100-mm pencil ionization chamber (Fig 2). An average or weighted CT dose index is calculated by adding one-third of the central value and two-thirds of the peripheral values. For scanning with a pitch that does not equal 1, the Dose-Length Product DLP is an indicator of the integrated radiation dose of an entire CT examination (5). The DLP incorporates the number of scans and the scan width (DLP = volume CT dose index total scan length). For conventional (nonspiral) scanning, the scan length is the sum of all section collimations for example, 25 mm (25 1 mm) for high-resolution CT. However, spiral and multidetector CT oversample data at the beginning and end of the scanning range because these data are needed for raw data interpolation of the first and last sections. There are differences between the various manufacturers, but approximately one half-rotation at the beginning and another halfrotation at the end have to be added to calculate the radiation exposure to the patient. Thus, the scan length, as provided by the scanner, should be increased by at least one table feed. Effective Dose The radiation risk to the patient can be estimated from the effective dose (in sieverts), which can be calculated by summing the absorbed doses to individual organs weighted for their radiation sensitivity (5). However, because we cannot obtain accurate measurements of all pertinent organ doses and the risk coefficients specific to age, gender, and organ being irradiated, the estimated dose is calculated for an idealized 70-kg, 30-year-old Teaching Point

4 1454 September-October 2008 RG Volume 28 Number 5 Teaching Point patient (5,12). In spite of these limitations, effective dose is the most widely used quantity for comparison between radiologic procedures. There are various computer programs that can calculate dose for individual organs using the volume CT dose index and organ weighting factors from International Commission on Radiological Protection Publication 60 (5). From effective dose, the risk estimates for stochastic effects can be determined with linear extrapolation of radiation exposure data from Japanese atomic bomb survivors (5,13). Calculation of the effective dose is usually made with a mathematic anthropomorphic phantom or computer-simulated irradiation Monte Carlo techniques (statistical calculations of photon interactions) (12). However, a reasonable approximation of the effective dose can be obtained with a conversion factor k (msv mgy -1 cm -1, where msv = millisieverts and mgy = milligrays) that varies depending on the body region being imaged (Fig 3, Table 1) (5,12). Figure 3. Diagram illustrates a simple algorithm for estimating radiation risk from CT examinations. Data in box are provided by the CT scanner. Teaching Point Automatic Exposure Control AEC systems for multidetector CT scanners are now available from all major scanner manufacturers under different names (3,14). These dose modulation systems operate in a variety of ways, but their main purpose is to adjust radiation dose according to the patient s attenuation and ultimately to reduce the radiation dose to the patient while sustaining diagnostic image quality (6). AEC systems have a number of potential advantages, including better control of patient radiation dose, avoidance of photon starvation artifacts, reduced load on the x-ray tube, and the maintenance of image quality in spite of different attenuation values on CT scans (15,16). With these benefits of AEC systems in mind, users should learn how to use and apply the systems properly. However, concerns about the routine use of AEC still remain. Although AEC systems generally reduce radiation dose, image noise inevitably increases, particularly in the region adjacent to contrast material and prosthesis-related artifacts. Table 1 Conversion Factors for Calculating Effective Dose Anatomic Region Conversion Factor (msv mgy -1 cm -1 ) Head Neck Chest Abdomen Pelvis Source. Reference 18. Note. mgy = milligrays, msv = millisieverts. control of tube current (milliamperage) (7). This tube current modulation can be performed using one or more of three basic methods: patientsize AEC, z-axis AEC, and rotational AEC (Fig 4). Most recent multidetector CT software and technology makes use of a combination of these methods. Teaching Point Types of AEC Reduction of radiation dose and variations in image quality in patients of different sizes and attenuations are generally achieved through the Patient-Size AEC and Z-axis AEC. Scan projection radiographs (topographic views) are applied in patient-size and z-axis AEC and are used mainly for the assessment of the size and attenuation of the patient. In reality, AEC systems tend to operate on the basis of a single anteropos-

5 RG Volume 28 Number 5 Lee et al 1455 Figure 4. Drawings illustrate types of AEC systems. With patient-size AEC (a), tube current is modulated according to patient size; with z-axis AEC (b), tube current is modulated according to patient attenuation along the z-axis; with rotational or angular AEC (c), tube current is modulated according to the asymmetry at each z-axis position; with combination AEC (d), tube current is modulated according to the combined effects of two or three of these types of AEC. terior topographic view to achieve compatibility with existing clinical practice (7). In patient-size AEC, the tube current is adjusted based on the overall size of the patient to reduce the variation in image quality between small patients and large patients. For a given patient size, the appropriate milliamperage is selected and is used for the entire examination or scan series. The AEC system (DoseRight) from Philips Medical Systems (Best, the Netherlands) has automatic current selection (ACS), which provides patient-based AEC. This system automatically suggests a milliamperage to be used throughout the examination for each patient depending on his or her size as seen on the topogram. The milliamperage is set with use of a reference image, whose noise value the system attempts to duplicate. The reference image is prestored at the factory and is based on a patient size of 33 cm. The suggested milliamperage will produce CT scans with image noise similar to that of the reference image. The system displays the volume CT dose index with ACS prior to scanning. Thus, the user can select the suggested milliamperage or individual settings at his or her discretion, with the dose difference between scans obtained with and without ACS technique. AEC systems from GE Medical Systems (Waukesha, Wis) (AutomA), Siemens Medical Solutions (Forchheim, Germany) (CARE Dose 4D), and Toshiba Medical (Tokyo, Japan) (SureExposure) provide both patient-size and z-axis AEC. The aim of z-axis AEC is to reduce the variation in the quality of images from the same series. AutomA (GE) adjusts the tube current to maintain a user-specified noise level in the image data. Z-axis AEC determines the tube current

6 1456 September-October 2008 RG Volume 28 Number 5 on the basis of projection data obtained from a topogram and a set of empirically determined noise prediction coefficients with use of the reference technique (17). These projection data can be used to determine the attenuation, size, and shape of the patient. AutomA provides a noise index to allow users to select the level of quantum noise that will be present in the reconstructed images, and the system modulates the tube currents to maintain the noise level in all images irrespective of patient size and attenuation. SureExposure (Toshiba) is also based on the projection data obtained from a topogram. For z-axis modulation, the water equivalent diameter at each level of the patient based on a topogram is calculated relative to the maximum attenuation (17). The appropriate tube current is applied at the maximum water equivalent diameter to achieve the selected standard deviation (noise level). Tube current is then modulated to maintain the standard deviation throughout the examination. Consequently, the milliamperage would be low through the thorax and high through the abdomen, which would be helpful when scanning both anatomic areas because, unlike the thorax, the abdomen contains solid organs. Rotational or Angular AEC. Rotational or angular dose modulation involves varying the tube current to equalize the photon flux to the detector as the x-ray tube rotates about the patient (eg, from anteroposterior to lateral). Angular dose modulation was first introduced for a single detector row helical CT scanner (SmartScan, GE) in This technique makes use of the information obtained from two topograms (anteroposterior and lateral views). With this information, sinusoidal modulation of tube current is achieved during 360 rotation for equalization of x-ray absorption. In a noncircular cross-sectional geometry, attenuation varies in different projection angles. At angular projections with a small patient diameter or body region, the tube current can be reduced without substantially increasing the image noise. Generally, lateral projections are more attenuating than anteroposterior projections, particularly in asymmetric regions of the body such as the shoulder or pelvis (compared with the head). With the advances in AEC techniques, an online modulation technique has been developed that does not require the information provided by topograms. This technique calculates the modulation function data (an objective image quality parameter) from the online patient attenuation. These data are sent to the generator control for dose modulation with a delay of 180 from the x-ray generation angle. Thus, the system makes use of attenuation data from the previous rotation and modulates tube current to accommodate patient attenuation on the fly (15,16). In angular dose modulation, more dose modulations occur in asymmetric regions and the variation in image noise throughout the examination can be minimized. This rotational AEC is also helpful in reducing photon starvation artifacts, especially in the shoulder (15). CT equipment manufacturers usually provide AEC systems with a combination of two or three of these types of AEC (3). Most 64-row scanners now provide these AEC systems, which include near real-time modulation techniques requiring x-ray tubes and generators to vary their output rapidly for subsecond rotation times (Table 2) (14). Operation of AEC Systems on Different Multidetector CT Scanners Standard Deviation based AEC. With Sure- Exposure 3D (Toshiba), users have to provide a standard deviation of pixel value for specifying image quality. A high standard deviation value will produce noisy images, whereas a low standard deviation value will produce less noisy images. SureExposure has a reference position selector for a given region, and the mean attenuation within this scanning region is used to calculate the tube current. The system makes use of a topogram to obtain the attenuation information used to calculate the tube current for each rotation. Next, on an image-by-image basis, the scanner sets the tube current that is required to achieve the requested standard deviation value. This method allows constant and consistent image quality for all patients and patient shapes. Therefore, the milliamperage is controlled by standard deviation and is dependent on the kernel for primary image reconstruction. Sure- Exposure aims to match the image noise to the target standard deviation value; however, with use of a standard kernel, the images are noisier than requested for higher target standard deviation values (7).

7 RG Volume 28 Number 5 Lee et al 1457 Table 2 Features of AEC Systems in a 64-section CT Scanner Feature Automatic ma control software Method for operator control of AEC Method for system control of ma ma adjustment for patient size ma adjustment along the z axis ma modulation during rotation GE LightSpeed VCT Philips Brilliance CT 64 CT Scanner Siemens Sensation 64 Toshiba Aquilion 64 Single or dual scout image, online control Source. Modified from reference 14. Note. ma = milliamperage. Single or dual scout image Single scout image, online control Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes SmartmA Doseright ACS, Doseright DOM (D DOM, Z DOM) CARE Dose 4D SureExposure Noise index Reference image Reference ma Standard deviation Single or dual scanogram Noise Index based AEC. SmartmA (GE) allows the user to set diagnostic image quality by entering Noise Index and a range of acceptable tube current settings (minimum and maximum milliamperage) (7). The current AEC system from GE has two elements: AutomA and SmartmA. AutomA provides the patient-size and z-axis AEC elements, and SmartmA provides the rotational AEC element. Users need to understand that image noise is inversely proportional to the square of the tube current. The noise index is used to set the estimated tube current, producing a noise level in reconstructed images that is based on anticipated patient attenuation from topograms. This system also aims to maintain a constant image noise level in each section. In this system, the standard deviation of images reflects the noise index most closely when GE s Standard reconstruction kernel is used. Images acquired with sharper or smoother kernels will result in images with standard deviations that are higher or lower than the noise index (7). Reference Milliamperage based AEC. CARE Dose 4D (Siemens) makes use of the effective milliamperage to compensate for the helical pitch for a given tube milliamperage. The effective milliamperage is defined as (tube current [in milliamperes] gantry rotation time [in sec- onds])/pitch factor. After selecting peak voltage, table feed, and detector configuration, the user must enter a reference effective milliamperage for the examination. This is the value that would be used with an average-sized patient. CARE Dose 4D assesses the size of the patient cross section being scanned and adjusts tube current relative to the reference effective milliamperage. It aims to provide adequate image noise, which varies depending on the size of the patient. This system operates on the principle that different-sized patients require different levels of noise to maintain adequate image quality. The degree to which tube current is adjusted for patient size can be selected using weak, average, or strong compensation settings. Reference effective milliamperage is designed for an average-sized patient (70 80 kg for adults, 20 kg [or 5 years of age] for pediatric patients). The scanner could be set up to provide average reductions in the reference effective milliamperage for slim patients and strong increases in the reference effective milliamperage for obese patients. These compensations result in less tube current adjustment than would be necessary to keep image noise constant for all patient sizes. With this approach, smaller or larger patients can be dealt

8 1458 September-October 2008 RG Volume 28 Number 5 with before the minimum or maximum tube current is reached. CARE Dose 4D adjusts the tube current over the patient s z-axis (one dimension) and in the x- and y-axis (three dimensions) based on the topogram. Real-time four-dimensional modulation is available during scanning and is controlled by using feedback from the previous rotation to set the tube current according to the attenuation measured at each tube angle. Reference Image based AEC. DoseRight by Philips consists of two types of dose modulations: DoseRight ACS, which provides patient-based AEC; and DoseRight DOM (dose modulation), which provides z-axis (Z DOM) and rotational (D DOM) AEC. DoseRight ACS makes use of a topogram to assess the patient s attenuation to set the tube current, which is used at all z-axis positions. Within a particular protocol, the user can select DoseRight ACS and DoseRight DOM either singly or together. On the basis of the system s prior attenuation measurements in crosssectional anatomy, ACS automatically suggests the lowest milliamperage settings to maintain consistent image quality at a low radiation dose throughout the examination. Tube current is set so that 90% of the images will have noise equal to or lower than that of the reference image, with the remaining 10% of images having noise equal to or higher than that of the reference image. In this system, the required image quality is expressed as an existing clinical image rather than an abstract standard deviation value or noise index. Therefore, it is difficult to compare scanning protocols, since there is no value associated with the image quality in the reference image. There is also the possibility of obtaining a better image with lower noise than one that has been deemed good enough for evaluation because of the temptation to pick a nice image. This could potentially lead to the system attempting to match images with higher quality than is needed and using higher doses than are necessary. Determining Protocols with AEC Systems Optimization of scanning protocol involves many parameters, including tube voltage, tube current, section thickness, collimation, and pitch. The Table 3 European Guidelines for the Diagnostic Reference Levels in Various CT Examinations Examination Diagnostic Reference Level CTDIw (mgy) DLP (mgy cm) Routine head Face and sinuses Vertebral trauma Routine chest High-resolution CT (lung) Routine abdomen Liver and spleen Routine pelvis Osseous pelvis Source. Reference 18. Note. CTDIw = weighted CT dose index, mgy = milligrays. image reconstruction kernel is also important. Changing even one of these parameters can affect different AEC systems depending on the CT scanner being used. For example, changing the tube voltage and section thickness will not affect the tube current on Siemens CARE Dose systems, which permit variable levels of image noise for different-sized patients, but will do so on other AEC systems (7). Changing the reconstruction kernel will alter the tube current used by the Toshiba SureExposure system, which tries to maintain the image noise in response to variations in scanning and reconstruction parameters, but will not alter the tube current used by other systems (7). Therefore, users should know about each of their system s characteristics and the effect that changing scanning and reconstruction parameters will have on the system. Nevertheless, in the use of AEC systems, selecting an appropriate noise index, standard deviation, reference milliamperage, or reference image is vital. This process is not straightforward, however. There are two ways to determine appropriate values for the use of AEC systems. The European Guidelines on Quality Criteria for Computed Tomography are a good standard for optimizing scanning protocol with reasonable radiation dose (Table 3) (18). These guideline values are provided to the extent available in relation to technique for a standard-sized patient for each type of CT examination considered

9 RG Volume 28 Number 5 Lee et al 1459 Teaching Point (Table 3). Another way to optimize protocol is to use simulation software before scanning, although this technique is not currently available to most CT practitioners. This software can simulate the effect of increasing image noise, and the resultant simulated data can be reconstructed. Thereafter, users will be able to evaluate image quality with radiation dose modulation. Clinical Application of AEC Systems AEC systems are now installed in all CT scanners from major manufacturers to reduce the radiation dose to the patient. Although the use of these systems generally leads to a decrease in radiation exposure, there remains the possibility of higher exposure than would occur without the application of an AEC system (7). Therefore, it is paramount that radiologists know how to use the AEC systems in their own scanners. AEC systems do not reduce radiation dose per se; rather, they control radiation exposure relative to the required image quality (7). Conclusions Reductions in radiation dose inevitably result in a corresponding reduction in image quality. However, an optimal diagnostic image can still be acquired with the aid of AEC systems. Unfortunately, with no set standard, CT equipment manufacturers have created their own ways of modulating radiation dose, thereby creating the need to understand how to use different AEC systems and what kinds of values or information the user must predetermine. Therefore, users must possess a good understanding of the concepts of noise index, standard deviation, reference images, and reference milliamperage as they relate to each AEC system. For further development of AEC systems and their successful introduction into clinical practice, user education and good communication between users and manufacturers are needed. A consensus regarding good default AEC systems may also help users more easily achieve diagnostic image quality with reduced radiation dose. References 1. Amis ES Jr, Butler PF, Applegate KE, et al. American College of Radiology white paper on radiation dose in medicine. J Am Coll Radiol 2007;4: Mettler FA Jr. Magnitude of radiation uses and doses in the United States: NCRP scientific committee 6-2 analysis of medical exposures. Fortythird NCRP annual meeting program, 2007; Brenner D, Elliston C, Hall E, Berdon W. Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR Am J Roentgenol 2001;176: Khursheed A, Hillier MC, Shrimpton PC, Wall BF. Influence of patient age on normalized effective doses calculated for CT examinations. Br J Radiol 2002;75: Recommendations of the International Commission on Radiological Protection. Ann ICRP 1991;21(1-3): Mulkens TH, Bellinck P, Baeyaert M, et al. Use of an automatic exposure control mechanism for dose optimization in multi detector row CT examinations: clinical evaluation. Radiology 2005;237: Keat N. Report CT scanner automatic exposure control systems. London, England: Im- PACT, Available at: org/reports/report Accessed June 16, Golding SJ, Shrimpton PC. Commentary. Radiation dose in CT: are we meeting the challenge? Br J Radiol 2002;75: Imanishi Y, Fukui A, Niimi H, et al. Radiationinduced temporary hair loss as a radiation damage only occurring in patients who had the combination of MDCT and DSA. Eur Radiol 2005;15: Shope TB, Gagne RM, Johnson GC. A method for describing the doses delivered by transmission x-ray computed tomography. Med Phys 1981;8: Fearon T. CT dose parameters and their limitations. Pediatr Radiol 2002;32: McCollough CH, Schueler BA. Calculation of effective dose. Med Phys 2000;27: National Council on Radiation Protection and Measurements. Risk estimates for radiation protection. NCRP report no Bethesda, Md: National Council on Radiation Protection and Measurements, Lewis M, Keat N, Edyvean S. Report 06013: 32 to 64 slice CT scanner comparison report version 14. London, England: ImPACT, Available at: impactscan.org/reports/report Accessed June 16, Kalender WA, Wolf H, Suess C. Dose reduction in CT by anatomically adapted tube current modulation: phantom measurements. Med Phys 1999;26: Gies M, Kalender WA, Wolf H, Suess C. Dose reduction in CT by anatomically adapted tube current modulation: simulation studies. Med Phys 1999;26: Kalra MK, Maher MM, Toth TL, et al. Techniques and applications of automatic tube current modulation for CT. Radiology 2004;233: Jessen K, Panzer W, Shrimpton P, et al. EUR 16262: European guidelines on quality criteria for computed tomography. Luxembourg: Office for Official Publications of the European Communities, Also available at: /guidelines/ct/quality/. Accessed June 16, 2008.

10 RG Volume 28 Volume 5 September-October 2008 Lee et al Radiation Dose Modulation Techniques in the Multidetector CT Era: From Basics to Practice Chang Hyun Lee, MD, et al RadioGraphics 2008; 28: Published online /rg Content Codes: Page 1453 DLP is an indicator of the integrated radiation dose of an entire CT examination. Page 1454 A reasonable approximation of the effective dose can be obtained with a conversion factor k (msv mgy [minus]1 cm [minus]1, where msv = millisieverts and mgy = milligrays) that varies depending on the body region being imaged. Page 1454 AEC systems have a number of potential advantages, including better control of patient radiation dose, avoidance of photon starvation artifacts, reduced load on the x-ray tube, and the maintenance of image quality in spite of different attenuation values on CT scans. Page 1454 Tube current modulation can be performed using one or more of three basic methods: patient-size AEC, z-axis AEC, and rotational AEC. Page 1459 Unfortunately, with no set standard, CT equipment manufacturers have created their own ways of modulating radiation dose, thereby creating the need to understand how to use different AEC systems and what kinds of values or information the user must predetermine. Therefore, users must possess a good understanding of the concepts of noise index, standard deviation, reference images, and reference milliamperage as they relate to each AEC system.

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

1. Patient size AEC. Large Patient High ma. Small Patient Low ma Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

TISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY

TISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY TISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY By RYAN F. FISHER A THESIS PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

diagnostic examination

diagnostic examination RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011

More information

Reducing Radiation Exposure from Survey CT Scans

Reducing Radiation Exposure from Survey CT Scans Reducing Survey CT Scan Exposure Pediatric Imaging Original Research Jennifer C. O Daniel 1 Donna M. Stevens 2 Dianna D. Cody 2 O Daniel JC, Stevens DM, Cody DD Received July 28, 2004; accepted after revision

More information

Electronic Noise in CT Detectors: Impact on Image Noise and Artifacts

Electronic Noise in CT Detectors: Impact on Image Noise and Artifacts Medical Physics and Informatics Original Research Duan et al. Electronic Noise in CT Detectors Medical Physics and Informatics Original Research Xinhui Duan 1 Jia Wang 1,2 Shuai Leng 1 ernhard Schmidt

More information

Diagnostic X-Ray Shielding

Diagnostic X-Ray Shielding Diagnostic X-Ray Shielding Multi-Slice CT Scanners Using NCRP 147 Methodology Melissa C. Martin, M.S., FAAPM, FACR Therapy Physics Inc., Bellflower, CA AAPM Annual Meeting, Orlando, FL FL Refresher Course

More information

Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2

Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2 Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2 Medical Consult, SA; Establish a method to correlate image

More information

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

Dose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation

Dose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation Medical Physics and Informatics Original Research Christner et al. CT Dose Reduction Medical Physics and Informatics Original Research Downloaded from www.ajronline.org by 8.243.133.8 on 2/26/18 from IP

More information

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

QC by the MPE in Belgium

QC by the MPE in Belgium Acceptance testing of state-of-the-art CT scanners using a new national protocol: first experience on a large number of scanners of different make and model the working group Radiology of the Belgian Hospital

More information

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation

More information

Improvement of CT image quality with iterative reconstruction idose4

Improvement of CT image quality with iterative reconstruction idose4 Improvement of CT image quality with iterative reconstruction idose4 Poster No.: C-0387 Congress: ECR 2014 Type: Scientific Exhibit Authors: M.-L. Olsson, K. Norrgren, M. Söderberg; Malmö/SE Keywords:

More information

Detector technology in simultaneous spectral imaging

Detector technology in simultaneous spectral imaging Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become

More information

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them. In press 2004 1 2 Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Digital radiology An appropriate analogy that is easy for most people to understand

More information

Quantitation of clinical feedback on image quality differences between two CT scanner models

Quantitation of clinical feedback on image quality differences between two CT scanner models Received: 4 August 2016 Revised: 4 November 2016 Accepted: 12 December 2016 DOI: 10.1002/acm2.12050 MEDICAL IMAGING Quantitation of clinical feedback on image quality differences between two CT scanner

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.125 pissn 2508-4445, eissn 2508-4453 Optimization of Energy Modulation Filter for Dual Energy CBCT

More information

Wide beam CT dosimetry. Elly Castellano

Wide beam CT dosimetry. Elly Castellano Wide beam CT dosimetry Elly Castellano Outline revision: CT dose indices wide-beam CT: the end of the road for CTDI? the IEC rescue plan for CTDI 100 the american way AAPM report 111 better estimates of

More information

FOREWORD. Acknowledgements

FOREWORD. Acknowledgements ΠΑΝΕΠΙΣΗΜΙΟ ΠΑΣΡΩΝ Διαημημαηικό Πρόγραμμα Μεηαπηστιακών ποσδών ζηην Ιαηρική Φσζική Διπλωμαηική εργαζία «ΔΟΙΜΕΣΡΙΑ ΑΘΕΝΩΝ Ε ΕΞΕΣΑΕΙ ΤΠΟΛΟΓΙΣΙΚΗ ΣΟΜΟΓΡΑΦΙΑ ΠΟΛΛΑΠΛΩΝ ΣΟΜΩΝ» ηέλλα Γ. Θαλαζζινού Α.Μ : 1575

More information

Suppression of metal artifacts using image-based monoenergetic DECT imaging

Suppression of metal artifacts using image-based monoenergetic DECT imaging Suppression of metal artifacts using image-based monoenergetic DECT imaging Poster No.: C-0519 Congress: ECR 2011 Type: Scientific Paper Authors: B. Krauss, B. Schmidt, M. Sedlmair, T. Flohr; Forchheim/DE

More information

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Poster No.: C-2205 Congress: ECR 2012 Type: Scientific Paper

More information

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

Iterative Reconstruction in Image Space. Answers for life.

Iterative Reconstruction in Image Space. Answers for life. Iterative Reconstruction in Image Space Answers for life. Iterative Reconstruction in Image Space * (IRIS) * Please note: IRIS is used as an abbreviation for Iterative Reconstruction in Image Space throughout

More information

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations S. Baechler 1, P. Monnin 1, A. Aroua 1, J.F. Valley 1, M. Perrier, P. Trueb 3, F.R. Verdun 1 1 University

More information

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology The Fundamentals of... Computed Tomography Computed Tomography (CT) systems use x-rays to produce images of slices through a patient s anatomy. Despite having lower spatial resolution than other x-ray

More information

Translating Protocols Between Scanner Manufacturer and Model

Translating Protocols Between Scanner Manufacturer and Model Translating Protocols Between Scanner Manufacturer and Model Robert J. Pizzutiello, MS, FAAPM, FACMP Sr. Vice-President, Global Physics Solutions President, Upstate Medical Physics Objectives Understand

More information

I. Introduction.

I. Introduction. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case

More information

Digital radiography (DR) post processing techniques for pediatric radiology

Digital radiography (DR) post processing techniques for pediatric radiology Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions

More information

160-slice CT SCANNER / New Standard for the Future

160-slice CT SCANNER / New Standard for the Future TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) Hiroki Kadoya Yukiko Kitagawa MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria

More information

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT A comparison of two methods for the determination of freein-air geometric efficiency in MDCT Theocharis Berris *1, Kostas Perisinakis 1,, Antonios E. Papadakis and John Damilakis 1, 1 Department of Medical

More information

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT) Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in

More information

The disclaimer on page 1 is an integral part of this document. Copyright November 30, 2017 by AAPM. All rights reserved.

The disclaimer on page 1 is an integral part of this document. Copyright November 30, 2017 by AAPM. All rights reserved. DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION

More information

Maximum Performance, Minimum Space

Maximum Performance, Minimum Space TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

Radiographic Techniques, Contrast, and Noise in X-Ray Imaging

Radiographic Techniques, Contrast, and Noise in X-Ray Imaging Residents Section Physics Minimodule Huda and Abrahams Techniques, Contrast, and Noise in Radiography Residents Section Physics Minimodule Residents inradiology Walter Huda 1 R. Brad Abrahams 2 Huda W,

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

Software and Hardware in CCTA. Elly Castellano PhD

Software and Hardware in CCTA. Elly Castellano PhD Software and Hardware in CCTA Elly Castellano PhD Outline technical requirements for coronary CTA the modern cardiac CT scanner ECG-gating technology image reconstruction algorithms 2 Technical requirements

More information

Measurement of table feed speed in modern CT

Measurement of table feed speed in modern CT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Measurement of table feed speed in modern CT Atsushi Fukuda, 1,2a Pei-Jan P. Lin, 3 Kosuke Matsubara, 2 Tosiaki Miyati 2 Department

More information

Development of new dosimeter for measuring dose distribution in CT

Development of new dosimeter for measuring dose distribution in CT Development of new dosimeter for measuring dose distribution in CT Poster No.: C-2925 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology - Without Subtopic Authors: Y. Muramatsu, K.

More information

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC? Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM

More information

Introduction of a Single Chip TLD System for Patient Dosimetry

Introduction of a Single Chip TLD System for Patient Dosimetry Introduction of a Single Chip TLD System for Patient Dosimetry C. Hranitzky a, M. Halda a, G. Müller a, B. Obryk b, H. Stadtmann a* a Austrian Research Centers GmbH ARC, 2444 Seibersdorf, Austria. b Institute

More information

1. Queries are issued to the image archive for information about computed tomographic (CT)

1. Queries are issued to the image archive for information about computed tomographic (CT) Appendix E1 Exposure Extraction Method examinations. 1. Queries are issued to the image archive for information about computed tomographic (CT) 2. Potential dose report screen captures (hereafter, dose

More information

CT Basics: Data Acquisition Module 3

CT Basics: Data Acquisition Module 3 Module 3 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

Kerma-Area Product in Diagnostic Radiology

Kerma-Area Product in Diagnostic Radiology Residents Section Physics Minimodule Huda Kerma-Area Product in Diagnostic Radiology Residents Section Physics Minimodule Walter Huda 1 Huda W Keywords: air kerma, fluoroscopy, interventional radiology,

More information

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life.

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life. Neuro Thoracic Abdominal Abdominal Cardiovascular Pediatric SAFIRE Sinogram Affirmed Iterative Reconstruction Answers for life. SAFIRE * (Sinogram Affirmed Iterative Reconstruction) * The information

More information

Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals

Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals DOI 10.1007/s00247-011-2290-1 ORIGINAL ARTICLE Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals Mervyn D. Cohen & Richard Markowitz

More information

2217 US Highway 70 East Garner, NC Main: Fax:

2217 US Highway 70 East Garner, NC Main: Fax: Viztek is committed to providing the highest image quality possible in our CR & DR product lines. There are several factors that directly affect the overall quality of CR & DR based images. The eposure

More information

Iterative Reconstruction

Iterative Reconstruction RECENT ADVANCES IN CT RADIATION DOSE REDUCTION TECHNIQUES Iterative Reconstruction Kalpana Kanal, PhD, FSCBTMR, FACR, FAAPM Professor and Director, Diagnostic Physics Section University of Washington Seattle,

More information

PET/CT Instrumentation Basics

PET/CT Instrumentation Basics / Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire

More information

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,

More information

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice nww.medical-devices.nhs.

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice   nww.medical-devices.nhs. abc September 2004 MHRA 04099 Philips Mx8000 IDT CT scanner technical evaluation Best choice best practice www.mhra.gov.uk nww.medical-devices.nhs.uk About MHRA evaluation reports. What you can expect.

More information

Teaching Digital Radiography and Fluoroscopic Radiation Protection

Teaching Digital Radiography and Fluoroscopic Radiation Protection Teaching Digital Radiography and Fluoroscopic Radiation Protection WCEC 20 th Student Educator Radiographer Conference Dennis Bowman, RT(R), CRT (R)(F) Community Hospital of the Monterey Peninsula (CHOMP)

More information

A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers

A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers Poster No.: C-3011 Congress: ECR 2010 Type: Topic: Authors: Scientific Exhibit

More information

Wide-Detector CT for TAVR Planning:

Wide-Detector CT for TAVR Planning: Wide-Detector CT for TAVR Planning: Impact on Iodine Dose, Radiation Dose, and Image Quality SCBTMR 2015 Annual Course Thursday, October 8 William P. Shuman MD FSCBTMR Department of Radiology University

More information

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography Journal of Physics: Conference Series OPEN ACCESS The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography To cite this article: N A A Daud et al 2014 J. Phys.:

More information

Essentials of Digital Imaging

Essentials of Digital Imaging Essentials of Digital Imaging Module 6 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 6 Dose Reduction and Patient Safety 1. ASRT Animation 2. Welcome Welcome to Essentials

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-706 Patient Phantom/Penetrometer System Users Manual March 2005 Manual No. 07-706-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are

More information

Overview of Safety Code 35

Overview of Safety Code 35 Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

Sarah Hughes, MS, DABR Radiation Safety Officer

Sarah Hughes, MS, DABR Radiation Safety Officer Sarah Hughes, MS, DABR Radiation Safety Officer 502-852-6146 sarah.hughes@louisville.edu Mo my back is burnin!!! I got it MAG the cine! Sumthin s not right. Where s his heart? Fluoroscopy http://dccwww.bumc.bu.edu/fluoroscopy/def

More information

Nuclear Associates , , CT Head and Body Dose Phantom

Nuclear Associates , , CT Head and Body Dose Phantom Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product

More information

CT Basics: Equipment and Instrumentation Module 2

CT Basics: Equipment and Instrumentation Module 2 Module 2 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational in-house or online educational course use only in educational and corporate institutions.

More information

CT Data Storage Reduction by Means of Compressing Projection Data Instead of Images: Feasibility Study 1

CT Data Storage Reduction by Means of Compressing Projection Data Instead of Images: Feasibility Study 1 Kyongtae T. Bae, MD, PhD Bruce R. Whiting, PhD Index terms: Computed tomography (CT), image display and recording Computed tomography (CT), image processing Computed tomography (CT), technology Data compression

More information

New Exposure Indicators for Digital Radiography Simplified for Radiologists and Technologists

New Exposure Indicators for Digital Radiography Simplified for Radiologists and Technologists Medical Physics and Informatics Technical Innovation Don et al. New Simplified Exposure Indicators Medical Physics and Informatics Technical Innovation Steven Don 1 ruce R. Whiting 2 Lois Jo Rutz 3 ruce

More information

The importance of radiation quality for optimisation in radiology

The importance of radiation quality for optimisation in radiology Available online at http://www.biij.org/2007/2/e38 doi: 10.2349/biij.3.2.e38 biij Biomedical Imaging and Intervention Journal COMMENTARY The importance of radiation quality for optimisation in radiology

More information

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging

More information

CR Basics and FAQ. Overview. Historical Perspective

CR Basics and FAQ. Overview. Historical Perspective Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors

More information

2D, 3D CT Intervention, and CT Fluoroscopy

2D, 3D CT Intervention, and CT Fluoroscopy 2D, 3D CT Intervention, and CT Fluoroscopy SOMATOM Definition, Definition AS, Definition Flash Answers for life. Siemens CT Vision Siemens CT Vision The justification for the existence of the entire medical

More information

Multi-Access Biplane Lab

Multi-Access Biplane Lab Multi-Access Biplane Lab Advanced technolo gies deliver optimized biplane imaging Designed in concert with leading physicians, the Infinix VF-i/BP provides advanced, versatile patient access to meet the

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

T h e P h a n t o m L a b o r a t o r y

T h e P h a n t o m L a b o r a t o r y T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced

More information

2012 :15th SESSION of ESMP

2012 :15th SESSION of ESMP 2012 :15th SESSION of ESMP Lecture presented in Archamps (Salève Building) by : Elly CASTELLANO (London) Patient dosimetry in x-ray imaging and CT Elly Castellano Objectives measurable dose quantities

More information

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL TORNIER BLUEPRINT 3D Planning + PSI SCAN PROTOCOL Contents 3 Introduction 3 Patient preparation 3 Scanning instructions 4 Image instructions 5 Scanning parameters 6 Technical instructions 2 BLUEPRINT 3D

More information

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging The Physics of Radiation Oncology X-ray Imaging Charles E. Willis, Ph.D. DABR Associate Professor Department of Imaging Physics The University of Texas M.D. Anderson Cancer Center Houston, Texas Overview

More information

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S. Practical Medical Physics Session: TG-151 Dose Monitoring August 5, 2013 Katie Hulme, M.S. Digital Imaging and Dose Creep Images courtesy of Agfa Healthcare Under-Exposed Over-Exposed Freedman et al.,

More information

CT Basics: Image Quality Module 6

CT Basics: Image Quality Module 6 Module 6 For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions. Any broadcast,

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

CT radiation profile width measurement using CR imaging plate raw data

CT radiation profile width measurement using CR imaging plate raw data JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 CT radiation profile width measurement using CR imaging plate raw data Thorarin A Bjarnason, 1,2,3a Chang-Ying Joseph Yang 3,4 Diagnostic

More information

Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing. Koichi Hirokawa MEDIX VOL. 56 P.43 P.46

Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing. Koichi Hirokawa MEDIX VOL. 56 P.43 P.46 Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing Taiga Goto Koichi Hirokawa Hisashi Takahashi MEDIX VOL. 56 P.43 P.46 Advanced Noise Reduction Processing for X-ray CT System

More information

Lunar Technology Advantages

Lunar Technology Advantages Lunar Technology Advantages DXA stands for Dual-Energy X-ray Absorptiometry. It is a measurement method that uses the differences in the absorption of high energy and low energy X-ray photons by different

More information

Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise

Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise Jonas Rydberg, M.D. Professor of Radiology Indiana University School of Medicine Indianapolis, Indiana Medical

More information

SOMATOM Esprit A Bundle of Energy

SOMATOM Esprit A Bundle of Energy SOMATOM Esprit A Bundle of Energy DATA SOMATOM Esprit An economical CT scanner designed for...... Excellent spiral image quality... A wide range of clinical applications... Value performance and reliabilty

More information

Features and Weaknesses of Phantoms for CR/DR System Testing

Features and Weaknesses of Phantoms for CR/DR System Testing Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal

More information

PD233: Design of Biomedical Devices and Systems

PD233: Design of Biomedical Devices and Systems PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/

More information

of sufficient quality and quantity

of sufficient quality and quantity of sufficient quality and quantity The patient s body attenuates the beam as it passes though the body More energy is deposited in organs located near the entry of the beam than near the exit of the beam

More information

Redefining Ergonomics

Redefining Ergonomics Samsung Electronics Co., Ltd. inspires the world and shapes the future with transformative ideas and technologies, redefining the worlds of TVs, smartphones, wearable devices, tablets, cameras, digital

More information

IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film

IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film CYRIC Annual Report 2005 IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film Ohuchi H. 1, and Abe K. 2 1 Graduate School of Pharmaceutical Sciences,

More information

SYLLABUS. 1. Identification of Subject:

SYLLABUS. 1. Identification of Subject: SYLLABUS Date/ Revision : 30 January 2017/1 Faculty : Life Sciences Approval : Dean, Faculty of Life Sciences SUBJECT : Biophysics 1. Identification of Subject: Name of Subject : Biophysics Code of Subject

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information