First EURADOS intercomparison exercise of eye lens dosemeters for medical applications

Size: px
Start display at page:

Download "First EURADOS intercomparison exercise of eye lens dosemeters for medical applications"

Transcription

1 First EURADOS intercomparison exercise of eye lens dosemeters for medical applications I. Clairand 1, M. Ginjaume 2, F. Vanhavere 3, E. Carinou 4, J. Daures 5, M. Denoziere 5, E. Honorio da Silva 3, M. Roig 2, S. Principi 2, L. Van Rycheghem 1. 1 Institute for Radiological Protection and Nuclear Safety (IRSN), France 2 Universitat Politècnica de Catalunya (UPC), Spain 3 Belgian Nuclear Research Centre (SCK-CEN), Belgium 4 Greek Atomic Energy Commission (GAEC), Greece 5 CEA-LIST Laboratoire National Henri Becquerel (CEA-LNHB), France

2 Introduction Context: decrease of the eye lens dose limit for occupational exposure to 20 msv/year (European Directive 2013/59/EURATOM) EURADOS: organisation of an intercomparison exercise for eye lens dosemeters in the medical field Main objective: to assess the capabilities of the eye lens dosemeters currently in use in Europe. 2/18

3 Project phases and time schedule Phase Date Preparation Dec February 2014 Selection of participants March 2014 Execution April 2014 January 2015 Reception of dosemeters by coordinating laboratory June 2014 Irradiations July-August 2014 Irradiated dosemeters sent to the participants September 2014 Results reported by the participants October 2014 First draft individual result datasheets sent to participants January 2015 Interpretation of results and reporting February 2015 May 2015 FINAL individual result datasheets + certificates of attendance Under preparation 3/18

4 Eye lens dosemeter types included in this intercomparison A total of 20 European individual monitoring services from 15 different countries were registered. The dosemeters provided by the participants were all composed of thermoluminescent detectors, of various types and designs. SCK- CEN Belgium SURO Czech Republic GAEC Greece U.O.Fisica e Tecnologie Biomediche Italy Lithuanian Radiation Protection Center Lithuania IFJ Poland NIOM Poland SMU Slovakia RPI Ukraine 4/18

5 Eye lens dosemeter types included in this intercomparison IRSN, France AV-Controlatom, Belgium DOSILAB, Switzerland 5/18

6 Eye lens dosemeter types included in this intercomparison DOZIMED S.R.L., Rumania PHE, UK ENEA, Italy Seibersdorf labor GmbH, Austria 6/18

7 Eye lens dosemeter types included in this intercomparison UPC, Spain Tecnorad s.u.r.l., Italy Univ. Hospitals Birmingham, UK VINCA, Serbia 7/18

8 Radiation qualities and doses imparted Setup Radiation quality Ref. Mean energy Nb of irr. dosemeters Range H p (3) (msv) 1 S-Cs; 0 ISO kev S-Cs; 0 ISO kev S-Cs; 60 ISO kev N-40; 0 ISO kev N-60; 0 ISO kev N-80; 0 ISO kev RQR6; 0 IEC kev RQR6; 45 IEC kev RQR6; 75 IEC kev Realistic field (scattered field in int. rad.) CONRAD/ORAMED European projects (PCRD 7) 45 kev /18

9 Irradiations Irradiations were performed according to ISO 4037 standard in terms of personal dose equivalent H p (3) The head phantom (ORAMED design - 20*20) was used Irradiation laboratories UPC - Spain SCK-CEN - Belgium CEA - France IRSN - France For realistic configuration (scattered field in interventional radiology): head phantom was located in the scattered field, reference in H p (3), the scattered radiation produced by the patient was done with an ISO phantom. 9/18

10 Transit dose control/background/spare For each participant: 12 dosemeters were reserved for transit dose correction (3 dosemeters were dispatched to the 4 irradiation laboratories), this correction was made by the organisers. The background correction was made by participant according to the laboratory routine procedure. 6 dosemeters were reserved as spare in order to be used by the irradiation labs in case of damage or errors with the irradiations. Only a very few dosemeters had to be used for this purpose. 10/18

11 Results The numerical results in this intercomparison are reported as the dosemeter response R, where R is defined as R = H p (3) participant corrected for background and transit dose / H p (3) reference For the analysis of the global results, the performance limits according to ISO14146 («trumpet curves») were adopted: R is the response, the ratio between the participant measured value and the conventional true value F = 1.5 (ICRP 75) H C is the conventional true value, in this case, H p (3) reference H 0 : was chosen equal to msv for all participants, assuming a lower limit of the dose range for which the system has been approved of 1 msv in a year, and an issuing frequency of 12 per year, in order to be consistent with the EURADOS report EURADOS Intercomparison 2008 for Whole Body Dosemeters in Photon Fields EURADOS Report /18

12 Individual result datasheet First draft results sent in February Participants were asked: to confirm the results or if they considered that corrections should be made because of errors made by the organization group, to indicate what should be corrected and why. The organization group decided on the corrections on an individual basis. Final results + certificate of attendance: under preparation 12/18

13 Results Among the 20 participants: GENERAL COMMENTS 14 have 100% data within the trumpet curves for all setups, 17 have 90% data within the trumpet curves, 1 is out of the limits for setups with low energies, 3 are out of the limits for setups with large angles, 2 are out of the limits for both types of setup (low energies and large angles). 13/18

14 Calibration: general comments CALIBRATION ENERGY 9 participants use pure S-Cs, 3 participants use mixed S-Cs and X-ray, 8 use various X-ray spectra. CALIBRATION QUANTITY 16 participants use H p (3), 2 participants use H p (10), 2 participants use H p (0.07). PHANTOM USED FOR THE CALIBRATION 10 participants use a cylindrical phantom (20x20) (ORAMED design), 1 participant use a smaller cylindrical phantom (d=16 cm), 7 participants use ISO slab phantom, 1 participant uses the head of Rando phantom, 1 participant uses rod phantom. 14/18

15 Results R = H s /H c S-Cs, 0º, Low S-Cs, 0º, High S-Cs, 60º N-40, 0º N-60, 0º N-80, 0º RQR 6, 0º RQR 6, 45º RQR 6, 75º Realistic field H c (msv) 15/18

16 Results irradiation setups 2,5 Box plots showing the minimum, 1 st quartile, median, 3 rd quartile and maximum responses per irradiation setup 2 1,5 Response 1 0,5 0 S-Cs, 0, Low S-Cs, 0, High S-Cs,60 N-40, 0 N-60, 0 N-80, 0 RQR6, 0 RQR6, 45 RQR6, 75 Realistic field 16/18

17 Results - participants Box plots showing the minimum, 1 st quartile, median, 3 rd quartile, maximum values for each participant. 17/18

18 Conclusion A first European intercomparison exercise of eye lens dosemeters. This intercomparison gives an interesting overview of the different dosimetric systems available in Europe for eye lens dose monitoring. Results are satisfactory (some problems observed for large angles and low energies) Some participants could improve the performance of their system by improving their calibration. A larger eye lens dosemeter intercomparison should be organised by EURADOS in the future. 18/18

19 19/20

Barcelona, January 2011 Workshop on optimization of Radiation Protection of Medical Staff

Barcelona, January 2011 Workshop on optimization of Radiation Protection of Medical Staff USE OF ACTIVE PERSONAL DOSEMETERS IN INTERVENTIONAL RADIOLOGY/CARDIOLOGY: TESTS WITH CONTINUOUS AND PULSED FIELDS IN LABORATORY CONDITIONS ORAMED PROJECT I. Clairand 1, J-M. Bordy 2, E. Carinou 3, J. Daures

More information

ORAMED S MEASUREMENT AND SIMULATION CAMPAIGN FOR EXREMITY AND EYE LENS DOSES OF MEDICAL STAFF INVOLVED IN INTERVENTIONAL RADIOLOGY AND CARDIOLOGY

ORAMED S MEASUREMENT AND SIMULATION CAMPAIGN FOR EXREMITY AND EYE LENS DOSES OF MEDICAL STAFF INVOLVED IN INTERVENTIONAL RADIOLOGY AND CARDIOLOGY ORAMED S MEASUREMENT AND SIMULATION CAMPAIGN FOR EXREMITY AND EYE LENS DOSES OF MEDICAL STAFF INVOLVED IN INTERVENTIONAL RADIOLOGY AND CARDIOLOGY Carinou E., Brodecki M.,Domienik J., Donadille L., Ferrari

More information

Active Personal Dosemeters: EURADOS activities and application in interventional radiology. Filip Vanhavere

Active Personal Dosemeters: EURADOS activities and application in interventional radiology. Filip Vanhavere Active Personal Dosemeters: EURADOS activities and application in interventional radiology Filip Vanhavere Thessaloniki May 17 th, 2011 Introduction to EURADOS EURADOS working group on active personal

More information

Monitoring the eye lens

Monitoring the eye lens Funding TS7e.3 Rolf Behrens Physikalisch-Technische Bundesanstalt Introduction April 2011: Limit of H lens : 150 msv 20 msv per year Protecting and monitoring may be necessary! Main question: Which quantity

More information

Introduction of a Single Chip TLD System for Patient Dosimetry

Introduction of a Single Chip TLD System for Patient Dosimetry Introduction of a Single Chip TLD System for Patient Dosimetry C. Hranitzky a, M. Halda a, G. Müller a, B. Obryk b, H. Stadtmann a* a Austrian Research Centers GmbH ARC, 2444 Seibersdorf, Austria. b Institute

More information

Technical aspects on DAP calibration and CT calibration

Technical aspects on DAP calibration and CT calibration EURADOS Report 2015-03 European Radiation Dosimetry Group e. V. Braunschweig, August 2015 Technical aspects on DAP calibration and CT calibration H. Järvinen, O. Turak, M. Ginjaume, J. Daures, C. Hourdakis,

More information

nuclear science and technology

nuclear science and technology EUROPEAN COMMISSION nuclear science and technology European ALARA Network (EAN) Contract N o FIR1-CT-2001-20187 Final report (summary) Work performed as part of the European Atomic Energy Community's research

More information

Ideas for type tests of electronic dosemeters in pulsed fields

Ideas for type tests of electronic dosemeters in pulsed fields Ideas for type tests of electronic dosemeters in pulsed fields Oliver Hupe, Peter Ambrosi Physikalisch-Technische Bundesanstalt Fachbereich 6.3 Strahlenschutzdosimetrie Bundesallee 100, 38116 Braunschweig

More information

Calibration of KAP meters

Calibration of KAP meters Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area

More information

Education and Training in Nuclear Decommissioning

Education and Training in Nuclear Decommissioning Education and Training in Nuclear Decommissioning Presentation prepared by: Pierre Kockerols, Hans Günther Schneider European Commission, Joint Research Centre With the support of: Martin Freer University

More information

EURAMET project 1177 : Comparison protocol ( ) Comparison protocol for the calibration of KAP meters in terms of air kerma area product

EURAMET project 1177 : Comparison protocol ( ) Comparison protocol for the calibration of KAP meters in terms of air kerma area product Comparison protocol for the calibration of KAP meters in terms of air kerma area product 1. Introduction The air kerma area product, P KA, is the integral of the air kerma over the area of the X ray beam

More information

Half value layer and AEC receptor dose compliance survey in Estonia

Half value layer and AEC receptor dose compliance survey in Estonia Half value layer and AEC receptor dose compliance survey in Estonia K. Kepler, A. Vladimirov Training Centre of Medical Physics, University of Tartu Testing Centre of the University of Tartu, Estonia E-mail:

More information

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced

More information

Received 3 December 2016; revised 17 May 2017; editorial decision 21 June 2017; accepted 4 July 2017

Received 3 December 2016; revised 17 May 2017; editorial decision 21 June 2017; accepted 4 July 2017 Radiation Protection Dosimetry (2018), Vol. 178, No. 2, pp. 235 241 Advance Access publication 15 July 2017 doi:10.1093/rpd/ncx091 NOTE COMPARISON OF PERSONAL DOSE EQUIVALENT Hp(10) IN 137 CS RADIATION

More information

Determination of the tube voltage from clinic mammographic system using two types of detectors

Determination of the tube voltage from clinic mammographic system using two types of detectors BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 03-1A (2015) 01-06 Determination of the tube voltage from clinic mammographic system using two types of detectors J. S. Barreira a ; V. Vivolo a a Gerência

More information

Since 1996, the participation to the network has grown continuously, as well as the topics covered and the types of products and deliverables.

Since 1996, the participation to the network has grown continuously, as well as the topics covered and the types of products and deliverables. oral presentation IRPA Madrid 2004 THE EUROPEAN ALARA NETWORK (EAN) FROM THE EXPERTISE TO THE STAKEHOLDERS INVOLVEMENT ID 145 Dr C.Lefaure*; P.Crouail*; P. Shaw** *Centre d Etudes sur l Evaluation de la

More information

The European Platform on Preparedness for Nuclear and Radiological Emergency Response and Recovery: The NERIS Platform

The European Platform on Preparedness for Nuclear and Radiological Emergency Response and Recovery: The NERIS Platform The European Platform on Preparedness for Nuclear and Radiological Emergency Response and Recovery: The NERIS Platform Eduardo GALLEGO UPM (Based on commonly developed material by the NERIS management

More information

ILNAS-EN 14136: /2004

ILNAS-EN 14136: /2004 05/2004 National Foreword This European Standard EN 14136:2004 was adopted as Luxembourgish Standard in May 2004. Every interested party, which is member of an organization based in Luxembourg, can participate

More information

This document is a preview generated by EVS

This document is a preview generated by EVS INTERNATIONAL STANDARD ISO 14096-1 First edition 2005-06-15 Non-destructive testing Qualification of radiographic film digitisation systems Part 1: Definitions, quantitative measurements of image quality

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

ISO INTERNATIONAL STANDARD. Non-destructive testing Qualification of radiographic film digitisation systems Part 2: Minimum requirements

ISO INTERNATIONAL STANDARD. Non-destructive testing Qualification of radiographic film digitisation systems Part 2: Minimum requirements INTERNATIONAL STANDARD ISO 14096-2 First edition 2005-06-15 Non-destructive testing Qualification of radiographic film digitisation systems Part 2: Minimum requirements Essais non destructifs Qualification

More information

Heads of European Radiological

Heads of European Radiological Heads of European Radiological protection Competent Authorities Association O. Guzmán French Nuclear Safety Authority, ASN HERCA Secretariat Summary - Background - Terms of Reference (ToR): - Objectives

More information

Joint Convention on the Safety of Spent Fuel Management and on the Safety of Radioactive Waste Management

Joint Convention on the Safety of Spent Fuel Management and on the Safety of Radioactive Waste Management JC/RM3/02/Rev2 Joint Convention on the Safety of Spent Fuel Management and on the Safety of Radioactive Waste Management Third Review Meeting of the Contracting Parties 11 to 20 May 2009, Vienna, Austria

More information

Poland: Competitiveness Report 2015 Innovation and Poland s Performance in

Poland: Competitiveness Report 2015 Innovation and Poland s Performance in Poland: Competitiveness Report 2015 Innovation and Poland s Performance in 2007-2014 Marzenna Anna Weresa The World Economy Research Institute Collegium of the World Economy Key research questions How

More information

Economic and Social Council

Economic and Social Council United Nations Economic and Social Council ECE/CES/GE.41/2013/3 Distr.: General 15 August 2013 Original: English Economic Commission for Europe Conference of European Statisticians Group of Experts on

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 8442-5 First edition 2004-12-15 Materials and articles in contact with foodstuffs Cutlery and table holloware Part 5: Specification for sharpness and edge retention test of cutlery

More information

Dose conversion coefficients for electron exposure of the human eye lens

Dose conversion coefficients for electron exposure of the human eye lens Physics in Medicine & Biology CORRIGENDUM Dose conversion coefficients for electron exposure of the human eye lens To cite this article: R Behrens et al 2010 Phys. Med. Biol. 55 3937 View the article online

More information

H2020 Excellent science arie Skłodowska-Curie Actions. Your research career in Europe. 17 November 2015

H2020 Excellent science arie Skłodowska-Curie Actions. Your research career in Europe. 17 November 2015 H2020 Excellent science arie Skłodowska-Curie Actions Your research career in Europe 17 November 2015 As a researcher I want to undertake a project in Europe, in an academic or other (e.g. business) setting

More information

DETERMINATION OF INHERENT AND ADDITIONAL FILTRATION IN ORDER TO ESTABLISH RADIATION QUALITIES ACCORDING TO IEC 61267

DETERMINATION OF INHERENT AND ADDITIONAL FILTRATION IN ORDER TO ESTABLISH RADIATION QUALITIES ACCORDING TO IEC 61267 2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 DETERMINATION

More information

Development of Personal Dosimeter Using Electronic Dose Conversion Method

Development of Personal Dosimeter Using Electronic Dose Conversion Method Proceedings of the Korean Nuclear Spring Meeting Gyeong ju, Korea, May 2003 Development of Personal Dosimeter Using Electronic Dose Conversion Method Wanno Lee, Bong Jae Lee, and Chang Woo Lee Korea Atomic

More information

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance

More information

This document is a preview generated by EVS

This document is a preview generated by EVS CEN WORKSHOP AGREEMENT CWA 17327 September 2018 ICS 03.080.30; 03.200.01 English version Hotel General Manager - Knowledge, skills and competence requirements This CEN Workshop Agreement has been drafted

More information

FINAL DRAFT TECHNICAL REPORT CLC/FprTR RAPPORT TECHNIQUE TECHNISCHER BERICHT January English version

FINAL DRAFT TECHNICAL REPORT CLC/FprTR RAPPORT TECHNIQUE TECHNISCHER BERICHT January English version FINAL DRAFT TECHNICAL REPORT CLC/FprTR 50579 RAPPORT TECHNIQUE TECHNISCHER BERICHT January 2012 ICS English version Electricity metering equipment - Severity levels, immunity requirements and test methods

More information

English Version EUROPEAN COMMITTEE FOR STANDARDIZATION COMITÉ EUROPÉEN DE NORMALISATION EUROPÄISCHES KOMITEE FÜR NORMUNG

English Version EUROPEAN COMMITTEE FOR STANDARDIZATION COMITÉ EUROPÉEN DE NORMALISATION EUROPÄISCHES KOMITEE FÜR NORMUNG EUROPEAN STANDARD NORME EUROPÉENNE EUROPÄISCHE NORM EN 10226-2 August 2005 ICS 21.040.30 English Version Pipe threads where pressure tight joints are made on the threads - Part 2: Taper external threads

More information

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS By Aravind Ravichandran arr192@mail.usask.ca University of Saskatchewan Address: 2424 Cumberland

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

Group of Administrative Co-operation Under the R&TTE Directive. 5 th R&TTE Market Surveillance Campaign on WLAN 5 GHz

Group of Administrative Co-operation Under the R&TTE Directive. 5 th R&TTE Market Surveillance Campaign on WLAN 5 GHz Group of Administrative Co-operation Under the R&TTE Directive Ref. Ares(2015)1723904-23/04/2015 5 th R&TTE Market Surveillance Campaign on WLAN 5 GHz REPORT ON THE 5 TH JOINT CROSS-BORDER R&TTE MARKET

More information

OPERRA Stakeholder Management Strategy Version 1

OPERRA Stakeholder Management Strategy Version 1 OPERRA Stakeholder Management Strategy Version 1 Workpackage 4: Reaching out to new Member States, academic & professional partners, stakeholders & authorities Task 4.3 Reaching out to major stakeholders

More information

PEDIATRIC AND STAFF DOSE EVALUATION IN FLUOROSCOPY UPPER GASTROINTESTINAL SERIES

PEDIATRIC AND STAFF DOSE EVALUATION IN FLUOROSCOPY UPPER GASTROINTESTINAL SERIES Internationaljoint Conference RADIO 2014 Gramado, RS, Brazil, Augustl 26-29, 2014 SOCIEDADEBRASILEIRA DEPROTEÇÃO RADIOLÕGICA - SBPR PEDIATRIC AND STAFF DOSE EVALUATION IN FLUOROSCOPY UPPER GASTROINTESTINAL

More information

THE DIGITALISATION CHALLENGES IN LITHUANIAN ENGINEERING INDUSTRY. Darius Lasionis LINPRA Director November 30, 2018 Latvia

THE DIGITALISATION CHALLENGES IN LITHUANIAN ENGINEERING INDUSTRY. Darius Lasionis LINPRA Director November 30, 2018 Latvia THE DIGITALISATION CHALLENGES IN LITHUANIAN ENGINEERING INDUSTRY Darius Lasionis LINPRA Director November 30, 2018 Latvia THE ENGINEERING INDUSTRIES ASSOCIATION OF LITHUANIA (LINPRA) is an independent

More information

Radiation dose to physicians eye lens during interventional radiology

Radiation dose to physicians eye lens during interventional radiology Journal of Physics: Conference Series PAPER OPEN ACCESS Radiation dose to physicians eye lens during interventional radiology To cite this article: N A Bahruddin et al 2016 J. Phys.: Conf. Ser. 694 012035

More information

This document is a preview generated by EVS

This document is a preview generated by EVS INTERNATIONAL STANDARD ISO 17636-1 First edition 2013-01-15 Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film Contrôle non destructif des assemblages soudés

More information

Munkaanyag

Munkaanyag TECHNICAL SPECIFICATION SPÉCIFICATION TECHNIQUE TECHNISCHE SPEZIFIKATION CEN/TS 16555-6 December 2014 ICS 03.100.40; 03.100.50 English Version Innovation management - Part 6: Creativity management Management

More information

EC Guideline 109: Guidance on Diagnostic Reference Levels REFERENCE LEVELS AT EUROPEAN LEVEL FOR CARDIAC INTERVENTIONAL PROCEDURES

EC Guideline 109: Guidance on Diagnostic Reference Levels REFERENCE LEVELS AT EUROPEAN LEVEL FOR CARDIAC INTERVENTIONAL PROCEDURES SENTINEL WORKSHOP Delft, - April 7 EC Guideline : Guidance on Diagnostic Reference Levels REFERENCE LEVELS AT EUROPEAN LEVEL FOR CARDIAC INTERVENTIONAL PROCEDURES R. Padovani, E. Vano, A. Trianni, C. Bokou,

More information

EU Ecolabel EMAS Environmental Technology Verification (ETV) State-of-play and evaluations

EU Ecolabel EMAS Environmental Technology Verification (ETV) State-of-play and evaluations EU Ecolabel EMAS Environmental Technology Verification (ETV) State-of-play and evaluations Pierre Henry DG Environment B1 3 instruments of Circular Economy action plan Improving the efficiency and uptake

More information

Operating Manual. Differential Pressure Transmitter DPS 100

Operating Manual. Differential Pressure Transmitter DPS 100 Operating Manual Differential Pressure Transmitter DPS 100 Important notes: Please read this operating manual carefully before installing and starting up the pressure transmitter. This operating manual

More information

NERIS Platform An attempt to enhance European response to and recovery from radiological emergencies

NERIS Platform An attempt to enhance European response to and recovery from radiological emergencies Radioprotection 2013 Vol. 48, n o 5, pages S11 à S17 DOI: 10.1051/radiopro/20139902 Editorial NERIS Platform An attempt to enhance European response to and recovery from radiological emergencies R. Mustonen

More information

Preliminary studies of a new monitor ionization chamber

Preliminary studies of a new monitor ionization chamber 1 Preliminary studies of a new monitor ionization chamber Maíra T. Yoshizumi, Vitor Vivolo and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN) Comissão Nacional de Energia

More information

ISO INTERNATIONAL STANDARD. Geotechnical investigation and testing Field testing Part 3: Standard penetration test

ISO INTERNATIONAL STANDARD. Geotechnical investigation and testing Field testing Part 3: Standard penetration test INTERNATIONAL STANDARD ISO 22476-3 First edition 2005-01-15 Geotechnical investigation and testing Field testing Part 3: Standard penetration test Reconnaissance et essais géotechniques Essais en place

More information

Features and Weaknesses of Phantoms for CR/DR System Testing

Features and Weaknesses of Phantoms for CR/DR System Testing Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal

More information

Dosepix Detector as kvp-meter in Radiology and Mammography: First steps

Dosepix Detector as kvp-meter in Radiology and Mammography: First steps Dosepix Detector as kvp-meter in Radiology and Mammography: First steps F.Bisello, I.Ritter, F.Tennert, A.Zang MediPix Collaboration Meeting, 19th February 2014, CERN Protect, Enhance, and Save Lives -

More information

Munkaanyag

Munkaanyag TECHNICAL SPECIFICATION SPÉCIFICATION TECHNIQUE TECHNISCHE SPEZIFIKATION CEN/TS 16555-4 December 2014 ICS 03.100.40; 03.100.50; 03.140 English Version Innovation management - Part 4: Intellectual property

More information

English - Or. English NUCLEAR ENERGY AGENCY COMMITTEE ON THE SAFETY OF NUCLEAR INSTALLATIONS FINAL REPORT AND ANSWERS TO QUESTIONNAIRE

English - Or. English NUCLEAR ENERGY AGENCY COMMITTEE ON THE SAFETY OF NUCLEAR INSTALLATIONS FINAL REPORT AND ANSWERS TO QUESTIONNAIRE Unclassified NEA/CSNI/R(2003)3 NEA/CSNI/R(2003)3 Unclassified Organisation de Coopération et de Développement Economiques Organisation for Economic Co-operation and Development 04-Feb-2003 English - Or.

More information

AIR KERMA RATES MEASUREMENT IN AN INTERVENTIONAL CARDIOLOGY SUITE

AIR KERMA RATES MEASUREMENT IN AN INTERVENTIONAL CARDIOLOGY SUITE AIR KRMA RATS MASURMNT IN AN INTRVNTIONAL ARIOLOGY SUIT anevaro, L.V. 1 ; Gamarra Sanchez, M..; Pereira, L.S.; Maurício,.L.P. Instituto de Radioproteção e osimetria. omissão Nacional de nergia Nuclear.

More information

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Gena M.A.H 1, Ahmed L.El-Attar 2, Elbadry M. Zahran 3, Hany El-Gamal

More information

- Nuclear Radiation Sensor - Detects Beta and Gamma Radiation and X-Rays

- Nuclear Radiation Sensor - Detects Beta and Gamma Radiation and X-Rays - Nuclear Radiation Sensor - Detects Beta and Gamma Radiation and X-Rays Description The function of the RD2014 radiation sensor is based on an array of customized PIN diodes. The integrated pulse discriminator

More information

INTERNATIONAL CIVIL AVIATION ORGANIZATION

INTERNATIONAL CIVIL AVIATION ORGANIZATION EUR DOC 024 Attachment INTERNATIONAL CIVIL AVIATION ORGANIZATION EUROPEAN PRINCIPLES AND PROCEDURES FOR THE ALLOCATION OF SECONDARY SURVEILLANCE RADAR MODE S INTERROGATOR CODES (IC) 2011 ATTACHMENT MODE

More information

Improved Imaging Plate Dosimetry for X-rays in Interventional Radiology

Improved Imaging Plate Dosimetry for X-rays in Interventional Radiology Improved Imaging Plate Dosimetry for X-rays in Interventional Radiology H. Ohuchi 1, A. Yamadera 2, T.Satoh 3, Y,Eguchi 3 1 Graduate School of Pharmaceutical Sciences, Tohoku University, Aramaki,Aoba-ku,

More information

User Manual for 24 GHz Blind-Spot Radar Sensor

User Manual for 24 GHz Blind-Spot Radar Sensor User Manual for 24 GHz Blind-Spot Radar Sensor SRR2-A Department: 1 of 13 AUTHORS Name Organisation Section Frank Gruson Continental, Frequency Management WW Issue Document Date of Issue Document Owner

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

COPYRIGHT Danish Standards. NOT FOR COMMERCIAL USE OR REPRODUCTION. DS/EN ISO 9431:1999

COPYRIGHT Danish Standards. NOT FOR COMMERCIAL USE OR REPRODUCTION. DS/EN ISO 9431:1999 EUROPEAN STANDARD NORME EUROPÉENNE EUROPÄISCHE NORM EN ISO 9431 July 1999 ICS 01.100.00 English version Construction drawings - Spaces for drawing and for text, and title blocks on drawing sheets (ISO

More information

English version. Audio, video and similar electronic apparatus - Safety requirements

English version. Audio, video and similar electronic apparatus - Safety requirements EUROPEAN STANDARD EN 60065/A12 NORME EUROPÉENNE EUROPÄISCHE NORM February 2011 ICS 97.020 English version Audio, video and similar electronic apparatus - Safety requirements Appareils audio, vidéo et appareils

More information

A STUDY OF THE RESPONSE OF DEPLETED TYPE p-mosfets TO PHOTON AND ELECTRON DOSE

A STUDY OF THE RESPONSE OF DEPLETED TYPE p-mosfets TO PHOTON AND ELECTRON DOSE A STUDY OF THE RESPONSE OF DEPLETED TYPE p-mosfets TO PHOTON AND ELECTRON DOSE M. Fragopoulou 1, S. Stoulos 1, and M. Zamani 1, 1 School of Physics, Nuclear and Elementary Particle Section Aristotle University

More information

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Journal of Physics: Conference Series PAPER OPEN ACCESS Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Recent citations - Resolution Properties of a

More information

Regulatory status for using RFID in the UHF spectrum 3 May 2006

Regulatory status for using RFID in the UHF spectrum 3 May 2006 Regulatory status for using RFID in the UHF spectrum 3 May NOTE: The following countries were updated since the last publication of 3 March : Thailand, Romania. The table attached provides an overview

More information

NREP Conference 7 10 April 2014 Salt Lake City - UT, USA 08/10/2013 1

NREP Conference 7 10 April 2014 Salt Lake City - UT, USA 08/10/2013 1 Preliminary Results on How CATO Seeks to Address Public Reassurance During a Radiological Exposure Type Incident Carlos Rojas-Palma, Robby Nijs (SCK CEN), Belgium Clemens Woda (HMGU), Germany Michael Discher

More information

DOSE EVALUATION IN UROGRAPHY RADIOLOGICAL PROCEDURES

DOSE EVALUATION IN UROGRAPHY RADIOLOGICAL PROCEDURES 2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 DOSE EVALUATION IN

More information

Trade Barriers EU-Russia based in technical regulations

Trade Barriers EU-Russia based in technical regulations Trade Barriers EU-Russia based in technical regulations Introduction Russia is a large market that offers business opportunities for companies like yours. However, accessing this market can be somehow

More information

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations S. Baechler 1, P. Monnin 1, A. Aroua 1, J.F. Valley 1, M. Perrier, P. Trueb 3, F.R. Verdun 1 1 University

More information

National Census Geography Some lessons learned and future challenges in European countries

National Census Geography Some lessons learned and future challenges in European countries UNSD-AITRS Regional Workshop on the Integration of Statistical and Geospatial Information Amman, Jordan, 16-20 February, 2015 National Census Geography Some lessons learned and future challenges in European

More information

English version. Steel and steel products Location and preparation of samples and test pieces for mechanical testing (ISO 377:1997)

English version. Steel and steel products Location and preparation of samples and test pieces for mechanical testing (ISO 377:1997) EUROPEAN STANDARD NORME EUROPÉENNE EUROPÄISCHE NORM EN ISO 377 July 1997 ICS 77.040.10; 77.140.01 Descriptors: See ISO document English version Steel and steel products Location and preparation of samples

More information

LLNL Results from CALIBAN-PROSPERO Nuclear Accident Dosimetry Experiments in September 2014

LLNL Results from CALIBAN-PROSPERO Nuclear Accident Dosimetry Experiments in September 2014 LLNL-TR-671548 LLNL Results from CALIBAN-PROSPERO Nuclear Accident Dosimetry Experiments in September 2014 M. L. Lobaugh, D. P. Hickman, C. W. Wong, A. R Wysong, M. J. Merritt, D. P. Heinrichs, J. D. Topper

More information

Competition SyStem The championship will be played within 11 days (9 game days plus 2 rest days).

Competition SyStem The championship will be played within 11 days (9 game days plus 2 rest days). Horarios Competition SyStem The championship will be played within 11 days (9 game days plus 2 rest days). preliminary round The 16 teams are divided into 4 groups (A, B, C, D) of 4 team each. Each team

More information

Delicate tasks should be entrusted to safe hands: Medical gases from Messer.

Delicate tasks should be entrusted to safe hands: Medical gases from Messer. Delicate tasks should be entrusted to safe hands: Medical gases from Messer. Everything at hand: Messer as a medical partner. Stefan Messer Chief Executive Officer and owner of Messer Group When medical

More information

Centralised Services 7-2 Network Infrastructure Performance Monitoring and Analysis Service

Centralised Services 7-2 Network Infrastructure Performance Monitoring and Analysis Service EUROCONTROL Centralised Services 7-2 Network Infrastructure Performance Monitoring and Analysis Service Monitoring the performance of 1030/1090 MHz RF bands A COST-EFFICIENT SOLUTION To make best use of

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

Public Consultation: Science 2.0 : science in transition

Public Consultation: Science 2.0 : science in transition DIRECTORATES-GENERAL FOR RESEARCH AND INNOVATION (RTD) AND COMMUNICATIONS NETWORKS, CONTENT AND TECHNOLOGY (CONNECT) Public Consultation: Science 2.0 : science in transition QUESTIONNAIRE A. Information

More information

Radiological Protection: Old Questions Needing New Answers

Radiological Protection: Old Questions Needing New Answers Radiological Protection: Old Questions Needing New Answers William D. Magwood, IV Director-General Nuclear Energy Agency ICRP 2017 10 October 2017 2015 Organisation for Economic Co-operation and Development

More information

ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung

ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung Treffen des Arbeitskreises IMRT der DGMP Würzburg, 26 + 27.03.2009 Salih Arican Sun Nuclear Corporation QA Challenge for Rotational Beams

More information

Pre-Commercial Procurement (PCP) Actions

Pre-Commercial Procurement (PCP) Actions Pre-Commercial Procurement (PCP) Actions Open call in Objective 11.1 Targeted Calls in objectives 5.1(d), 11.2, 11.3, 8.2, 5.1(e)(1), 2.2(b) lieve.bos@ec.europa.eu EU Commission, DG INFSO Lisbon policy

More information

SVENSK STANDARD SS-EN 14136:2004

SVENSK STANDARD SS-EN 14136:2004 SVENSK STANDARD SS-EN 14136:2004 Fastställd 2004-07-16 Utgåva 1 In vitro-diagnostik Användning av externa kvalitetsbedömningsprogram för utvärdering av hur in vitro-diagnostiska undersökningar fungerar

More information

DETECTORS. Including Codes of Practice IONIZING RADIATION

DETECTORS. Including Codes of Practice IONIZING RADIATION DETECTORS Including Codes of Practice 2015 IONIZING RADIATION PTW History and General Remarks Looking back on a long history PTW-Freiburg is an internationally operating company, manufacturing and marketing

More information

The Composites Market

The Composites Market presents: The Composites Market - Applications, Developments & Trends - Hannover Messe, April 9th, 2014 page 1 Composites Germany - Members Composites Germany was founded by the four leading composites

More information

Communication systems for meters and remote reading of meters - Part 4: Wireless meter readout (Radio meter reading for operation in SRD bands)

Communication systems for meters and remote reading of meters - Part 4: Wireless meter readout (Radio meter reading for operation in SRD bands) Irish Standard Communication systems for meters and remote reading of meters - Part 4: Wireless meter readout (Radio meter reading for operation in SRD bands) CEN 2013 No copying without NSAI permission

More information

Computed Radiography

Computed Radiography BAM Berlin Computed Radiography --INDE 2007, Kalpakkam, India -- Uwe Zscherpel, Uwe Ewert BAM Berlin, Division VIII.3 Requests Requests and and information information to: to: Dr. Dr. U. U. Zscherpel Zscherpel

More information

A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering

A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering Sandra E. Burch Department of Radiology, Medical College of Georgia, Augusta, Georgia 30912 Kimberlee J. Kearfott

More information

30 A AC/DC Current Probe TCP0030A Datasheet

30 A AC/DC Current Probe TCP0030A Datasheet 30 A AC/DC Current Probe TCP0030A Datasheet Split-core construction allows easy circuit connection High accuracy with typically less than 1% DC gain error Low noise and DC drift 3rd party safety certification

More information

SECTEUR Ascertaining user needs

SECTEUR Ascertaining user needs Ascertaining user needs Marta Bruno Soares (Uni Leeds), Maria Noguer (IEA), Nigel Arnell (Uni Reading), Jorge Paz (Tecnalia) and Amanda Hall (Telespazio VEGA UK) What is? «The Sector Engagement for the

More information

Avery Dennison ICS Performance Guarantee Sign Cut and Wrapping Films Durability Bulletin 2.0

Avery Dennison ICS Performance Guarantee Sign Cut and Wrapping Films Durability Bulletin 2.0 Sign Cut and Wrapping Films Durability Bulletin 2.0 EU Revision number: 3 Date: 062016 ICS PERFORMANCE GUARANTEE The ICS Performance Guarantee combines our films with other qualified components to form

More information

In-circuit Measurements of Inductors and Transformers in Switch Mode Power Supplies APPLICATION NOTE

In-circuit Measurements of Inductors and Transformers in Switch Mode Power Supplies APPLICATION NOTE In-circuit Measurements of Inductors and Transformers in Switch Mode Power Supplies FIGURE 1. Inductors and transformers serve key roles in switch mode power supplies, including filters, step-up/step-down,

More information

IGD-TP 6 th Exchange Forum

IGD-TP 6 th Exchange Forum IGD-TP 6 th Exchange Forum IGD-TP Working Group on RD&D Dissemination Ray Kowe, RWM Gunnar Buckau JRC New Amba Hotel London, 3 rd November 2015 Ray Kowe 3 November, London, United Kingdom Page 1 SecIGD2

More information

This document is a preview generated by EVS

This document is a preview generated by EVS TECHNICAL REPORT CLC/TR 50579 RAPPORT TECHNIQUE TECHNISCHER BERICHT June 2012 ICS 91.140.50 English version Electricity metering equipment (a.c.) - Severity levels, immunity requirements and test methods

More information

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them. In press 2004 1 2 Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Digital radiology An appropriate analogy that is easy for most people to understand

More information

AC/DC Current Measurement Systems TCPA300, TCP312A, TCP305A, TCP303, TCPA400, TCP404XL Datasheet

AC/DC Current Measurement Systems TCPA300, TCP312A, TCP305A, TCP303, TCPA400, TCP404XL Datasheet AC/DC Current Measurement Systems TCPA300, TCP312A, TCP305A, TCP303, TCPA400, TCP404XL Datasheet Low DC drift and noise allows improved low-level current measurements 3rd party safety certification Applications

More information

The importance of radiation quality for optimisation in radiology

The importance of radiation quality for optimisation in radiology Available online at http://www.biij.org/2007/2/e38 doi: 10.2349/biij.3.2.e38 biij Biomedical Imaging and Intervention Journal COMMENTARY The importance of radiation quality for optimisation in radiology

More information

Passive Voltage Probes

Passive Voltage Probes Passive Voltage Probes TPP1000 TPP0500 TPP0502 Datasheet Connectivity Integrated Oscilloscope and Probe Measurement System provides Intelligent Communication that Automatically Scales and Adjusts Units

More information

Recommendations to harmonize European early warning dosimetry network systems

Recommendations to harmonize European early warning dosimetry network systems Journal of Instrumentation OPEN ACCESS Recommendations to harmonize European early warning dosimetry network systems To cite this article: H. Dombrowski et al View the article online for updates and enhancements.

More information

Tolerances. Alloy groups. Tolerances

Tolerances. Alloy groups. Tolerances Tolerances We apply tolerances for shape and dimension according to CEN the European standardisation organisation, CEN; European Standard EN 755-9. On the following pages we publish an excerpt from this

More information

News from CERN Ana Godinho Head of Education, Communications and Outreach

News from CERN Ana Godinho Head of Education, Communications and Outreach News from CERN Ana Godinho Head of Education, Communications and Outreach 23 rd EPPCN Meeting 23.04.2018 23 rd EPPCN Meeting 2 September Oct/Nov May/June October September http://hssip.web.cern.ch/ 23

More information