Introduction. Digital Mammography QA: Comparing the Manufacturers Recommendations. What is QC and why is it important? Review & compare QC tests
|
|
- Clyde Little
- 5 years ago
- Views:
Transcription
1 Slide 1 Digital Mammography QA: Comparing the Manufacturers Recommendations Eric A. Berns, Ph.D. Slide 2 Introduction What is QC and why is it important? Review & compare QC tests Key take home points Northwestern University Medical School Lynn Sage Comprehensive Breast Center Chicago, IL Slide 3 Introduction Slide 4 MQSA Screen-Film Mammography Quality Standards Act ACR American College of Radiology
2 Slide 5 Slide 6 Screen-Film Screen-Film Slide 7 Slide 8 Screen-Film Screen-Film
3 Slide 9 Slide 10 Screen-Film FFDM In FFDM, the manufacturer designs and mandates their own QC program In FFDM, you must the manufacturers QC program Slide 11 Slide 12 Quality Control: Why? Reduce exposure to patients and personnel Consistent image quality RWS Detect and correct for potential problems, before they impact image quality
4 Slide 13 What is Quality Control? Slide 14 Before You Begin QC Determination of what is Normal Detection of what is Abnormal Understanding of how to return to Normal from Abnormal In particular, in FFDM, how do you know what you are seeing is what it is supposed to be? Golden Rules for Medical Physicist & Tech QC Must use manufacturer s s QC procedures Manufacturers QC may refer to Monitor Manufacturers QC Multimodality Workstations have own separate QC Printers may have their own QC Slide 15 Before You Begin ACR Accreditation - Physics forms GE Senographe 2000D, DS, and Essential Fischer Senoscan Lorad Selenia Siemens Novation Note: : if using unit for both screen-film & CR, you must accredit for both FDA Accreditation Other vendors when approved Slide 16 Before You Begin
5 Slide 17 Before You Begin Slide 18 Before You Begin Slide 19 Before You Begin Slide 20 Before You Begin
6 Slide 21 Before You Begin Slide 22 Note: Note: For new unit: : Must use most current version For renewal unit: : Can use older version (version used when tested previously) Vs. Slide 23 Mammo Unit Evaluation Slide 24 Collimation Screen-Film Mammo Ensure locks, detents, angulation indicators, and mechanical support devices operate properly FFDM Same Screen-Film Mammo Assure -ray field aligns with the light field Collimator allows full coverage of receptor Chest wall edge of compression paddle aligns with chest wall of receptor FFDM Same Multiple targe/filters and/or multiple paddle positions Chest Wall Missed Tissue Compression Plate Overlap on the Chest Wall Side
7 Slide 25 Collimation Slide 26 GE 2000D, DS, Essential Fischer Senoscan Lorad Selenia Siemens Novation Same as Screen-Film Multiple target/filter combo s Multiple paddle positions Chest Wall & Missed Tissue Compression Plate Overlap on Chest Wall Collimation Test for routinely used Collimator, bucky, compression paddle combo s target materials Slide 27 GE Collimation Mo/Mo Rh/Rh Different paddle positions Slide 28 Fischer Senoscan Collimation
8 Slide 29 Fischer Senoscan Slide 30 Lorad Selenia Collimation Collimation Assessment Use coin techniques as described in ACR Manual 24 x 30 cm 18 x 24 cm (left, center, right paddle shifts) Slide 31 Siemens Novation DR Slide 32 Siemens Novation DR Chest Wall and Missed Tissue Place coin tangent to chest wall edge of bucky Measure coin and subtract from full diameter Missed tissue must be < 5 mm Compression Plate Overlap on Chest Wall Mount paddle Measure overlapping distance Add the missed tissue value for total Must be < 6.5 mm Must evaluate for ALL compression paddles including mag
9 Slide 33 Limiting Spatial Resolution Slide 34 Spatial Resolution Screen-Film Mammo Line-pair test tool FFDM Focal spot Line-pair test tool Focal spot Detector & imaging chain Same as Screen-Film (Focal Spot Eval) Measure using LP phantom on top of acrylic GE 2000D, DS, Essential Fischer Senoscan Lorad Selenia Siemens Novation MTF Slide 35 Slide 36 Fischer Senoscan System Resolution Up to 10 lp/mm Test Pattern 4 cm acrylic Parallel & Perpendicular to tube with 3-53 degree angle Evaluation of System Resolution lp/mm Test Pattern Modulation in Normal 7 lp/mm must be > 10% Modulation in Hi Res mode at 11.1 lp/mm must be > 5% The system limiting spatial resolution must be: > 8 lp/mm + 2 lp/mm (6 to 10 lp/mm)
10 Slide 37 GE 2000D, DS, & Essential Focal Spot Evaluation Slide 38 GE 2000D MTF Measurement Monthly Raw Image Mo/Mo Rh/Rh Large & Small Spots (1.5 mag) MTF (%) = (Std. Dev.) x 222 / (Mean Dark ROI Mean Light ROI) MTF 2 lp/mm > 58% MTF 4 lp/mm > 25% Slide 39 GE 2000D 2000D Addendum: Sub-system MTF Optional replacement for MTF & Focal spot No need for film Use specific bar pattern & 4.5 cm acrylic block Use ROI tools to calculate sub-system system MTF Repeat at 1.8 mag Slide 40 GE DS & Essential MTF and CNR Measurement - Weekly Test consistency of CNR and ensure contrast is adequate IQST Test tool Action Limit: 2 lp/mm > 58%, 4 lp/mm > 25% Change in CNR must not exceed 0.2
11 Slide 41 Lorad & Siemens Slide 42 AEC Performance Evaluation of System Resolution lp/mm Test Pattern The system limiting spatial resolution must be: > 7 lp/mm Screen-Film Mammo Measure optical densities Different thicknesses using clinical modes Different density settings (-2,( -1, 0, +1, +2, etc.) FFDM Measure resultant techniques Measure signal, exposure, & SNR values Slide 43 AEC Slide 44 Same as Screen-Film GE 2000D, DS, Essential Fischer Senoscan Lorad Selenia Siemens Novation Density Control Function ACR Reproducibility Image Mode Tracking CNR SNR Density Control Function 4 cm acrylic, ACR phantom technique Repeat at -2, -1, 0, +1, +2 etc. density Record mas mas change should be 5 to 15% per step
12 Slide 45 Slide 46 Reproducibility & Image Mode Tracking 4 cm acrylic with clinical technique Position ion chamber in beam Record mas and Repeat 3 times Repeat in each mode (small, large, mag, no grid) Action Limits: Coefficient of variation for Exposure or mas must not exceed 0.05 No significant difference in exposure between small and large bucky when using similar grids CNR Per Object Thickness CNR using clinical technique for 2 cm Repeat for 4 and 6 cm Action Limits: CNR of 2 cm relative to 4 cm must be > 100% CNR of 6 cm relative to 4 cm must be > 75% Slide 47 GE 2000D Slide 48 GE DS & Essential AOP Mode and SNR Check Variable thicknesses of acrylic Std, Auto Evaluate: Acrylic Thickness Correct techniques? Adequate SNR? Target- Filter Selected kvp Selected mas 2.5 cm Mo-Mo 27 kvp cm Mo-Rh 28 kvp cm Rh-Rh 32 kvp cm Acrylic Each raw image must have a measured SNR of at least cm Acrylic 6.0 cm Acrylic DS specific QC tests AOP Mode and SNR Check Use built-in in software for image acquisition AOP: STD/Auto Acrylic Thickness (mm) Track/Filter 25 Mo/Mo 50 Rh/Rh 60 Rh/Rh SNR must be > 50 Exposure Parameters For AOP STD mode oly mas kv Mo/Mo Rh/Rh Rh/Rh
13 Slide 49 Lorad Selenia Slide 50 Siemens Novation Automatic Exposure Control AUTO-FILTER 2, 4, 6, 8 cm 4 cm (-5( 5 thru +5) Use ROI to measure mean pixel value inside virtual AEC detector Performance Criteria Pixel value should not vary by more than 10% of mean Exposure compensation steps shall meet requirements in pixel value table AEC Image Stability and Reproducibility and SNR ACR Phantom Mo/Mo, 28 kv, H,, sensor 1 Record mas, SNR, Mean, Entrance Exposure (5 times) Action Limits: Coef. of Var for mas and < 5% SNR and Mean shall not vary by + 15% of Mean Slide 51 Siemens Novation DR AEC Thickness Tracking 2, 4, and 6 cm Slide 52 Uniformity of Screen Speed & Detector Performance Screen-Film Mammo H mode Measure if OD s s are consistent Max deviation = (Max difference / mean value ) * 100 Check for artifacts FFDM Flat-field uniformity Detector calibration Pixel correction test CR Reader Scanner Performance
14 Slide 53 Detector Performance GE 2000D, DS, Fischer Senoscan Lorad Selenia Siemens Novation Essential Same as Screen-Film (Screen Uniformity) Flat-Field Field Uniformity Detector Calibration CR Reader Scanner Performance Dynamic Range Primary Erasure Inter-Plate Consistency Geometric Distortion Pixel Correction Ghosting Slide 54 Dynamic Range To confirm the dynamic range of reader and plate Use 2 & 4 cm acrylic Technique for 6 cm View image to determine if 3 regions are visible and discernable Slide 55 Slide 56 CR Reader Scanner Performance Establish that IP reader & optics do not exhibit scan or print jitter Non-grid exposure Position rulers in T formation Expose & process Action Limit: Image must have smooth borders free from jagged edges or defects Primary Erasure (Additive & Multiplicative) To assess the erasure performance of reader and plates Additive: Shoot phantom, process, wait 1 minute, reprocess, change S value to 10 original, inspect for visibility of phantom image Multiplicative: Shoot & process phantom, shoot 4 cm acrylic on same cassette, process and inspect for visibility of phantom image
15 Slide 57 Inter-Plate Consistency To confirm x-rayx absorption & SNR consistency Expose plates using clinical technique with 4cm acrylic Record mas (must be within +10%) Calculate SNR (must be within +15%) 15%) Slide 58 GE 2000D, DS, & Essential Flat-field uniformity Test to ensure detector performance acceptable Measures detector uniformity (signal & noise) Measures bad pixels System automatically calculates pass/fail QAP Slide 59 Fischer Slide 60 Fischer Senoscan Flat Field and Detector Calibration Run daily calibration using software Use 4 cm acrylic Exposure resulting in 1000 ADU s No Artifacts at WW > 800 Compute deviation between corner and center means Must be within + 20% of center ROI Geometric Distortion Use a 40 mesh Cu screen Acquire image at low technique Visually inspect for distortions or blurring
16 Slide 61 Lorad Selenia Slide 62 Lorad Selenia Detector Calibration Every two weeks 4 cm acrylic block no paddle Built in Calibration software Follow on-screen instructions Review image for artifacts Repeat until you have 8 images End calibration Calibration is performed automatically Detector Ghosting Assure ghosting does not interfere with image quality 4 cm acrylic covering half of detector Set AEC clinically Make exposure 4 cm acrylic covering entire detector Place 0.1 mm AL on top Make 2 nd exposure (AEC) as soon as possible (~ 1 minute after first) Calculate Ghost Image Factor = (Mean R3 Mean R2 / Mean R1 Mean R2 ) Must be within Slide 63 Siemens Novation DR Slide 64 Siemens Novation Detector Uniformity Make exposure: 28 kv, 50 mas, Mo/Mo Select ROI 128 x 128 pixels Record measurements in corners and center Pixel value inside each of 5 ROI s shall not differ by more than 10% of mean Detector Calibration Test Objective: to determine if system was correctly gain calibrated Must be done weekly (technologist) Follow on-screen instructions Acquire 8 images Images should be free of artifacts
17 Slide 65 Siemens Novation DR Slide 66 Siemens Novation DR Ghost Image Evaluation Setup with 2mm steel bar Exp 1: Mo/Mo, 23 kv, 2 mas record mean Exp 2: Mo/Mo, 28 kv, 200 mas - remove steel Wait 180 seconds, place 4 cm Acrylic on top Exp 3: Mo/Mo, 28 kv, 14 mas Calculate % Ghosting = (PV1 PV2) / PV1 +PV2 Action Limit: : % Ghosting < 3% Pixel Correction Test Calibration mode Create new pixel map Follow on-screen instructions Acquire image Evaluate image is free from collimator or debris artifacts System compiles new pixel map System display results must meet requirements Slide 67 Artifacts Slide 68 Artifacts Screen-Film Mammo Evaluate cassettes Same as Screen-Film GE 2000D, DS, Essential Fischer Senoscan Lorad Selenia Siemens Novation FFDM Expose Using Acrylic Evaluate detector CR: Imaging Plates
18 Slide 69 Artifacts Slide 70 Artifacts Artifact evaluation Mo/Mo Mo/Rh Rh/Rh Slide 71 Artifacts Slide 72 Artifacts
19 Slide 73 Artifacts Slide 74 Artifacts Possible cause: condensation on detector Possible Cause: Detector dropout or mis-registration Slide 75 Image Detection Calibration File Artifacts Processing Steps for Digital Images Image Correction Raw Image Processing Processed Image Display Slide 76 Image Quality Screen-Film Mammo ACR Phantom Scores Optical Density & Contrast FFDM ACR Phantom Scores Pass/fail requirements differ by vendor Signal-to to-noise Ratio (SNR) Contrast-to to-noise Ratio (CNR)
20 Slide 77 Image Quality Slide 78 Image Quality Same as Screen-Film GE 2000D, DS, Essential Fischer Senoscan Manual Techniques Lorad Selenia Siemens Novation Clinical Technique Partial CNR ACR Phantom Imaging GE & Fischer & Fuji Lorad & Siemens Fibers 4 5 Specks 3 4 Masses 3 4 Slide 79 GE 2000D Slide 80 GE 2000D ACR Phantom Imaging Manual technique (Mo/Mo, 26 kvp, 125 mas) Score the processed image Acquisition workstation Each monitor of the RWS Contrast-to to-noise Test (CNR) To examine consistency of CNR ratio measured over time Use the raw image + 20% of baseline Background ROI Mass ROI Laser imager CNR = (Mean background - Mean Mean mass )/SD background
21 Slide 81 GE DS Slide 82 Fischer Senoscan DS specific QC tests Phantom Image Quality Manual technique: Rh/Rh, 29 kvp, 56 mas MTF and CNR Measurement Phantom Image Acquisition Test and Image Quality Select Techniques to give 1000 ADU Compare background mean, StdDev, and ADU level difference to baseline values Score : 4 fibers, 3 speck groups, and 3 masses Use IQST test tool Use built-in in software for image acquisition Manual technique: Rh/Rh 30 kvp, 56 mas Results are automatically displayed (pass/fail) Same action limits as 2000D Slide 83 Lorad Selenia Slide 84 Siemens Novation DR Phantom Image Quality Select clinical exposure mode (i.e. AUTO-FILTER) Print film Measure background OD and density difference Plot on tech worksheets Background must be > 1.20 OD DD must be > Score on each Soft Copy Workstation 5 fibers 4 speck groups 4 masses Phantom Image Quality Position phantom 1 cm over chest wall edge Select: 28 kv, AEC-Auto, Auto, Mo/Mo Score on RWS or Film Fiber > 5 Specks > 4 Masses > 4
22 Slide 85 Lorad & Siemens Slide 86 SNR and CNR Measurements SNR at least > 40 CNR should stay within ±15% of baseline Contrast-to to-noise Test (CNR) To examine consistency of CNR ratio measured over time Use 4 cm acrylic & 0.2 mm Al Manual technique (Mo/Mo, 26 kvp, 125 mas) Calculate CNR using software + 20% of baseline Obtained during acceptance testing Slide 87 kvp Slide 88 HVL Screen-Film Mammo Assure that actual kvp is FFDM Accurate (+( 5% of indicated) Reproducible (Coeff. Var. < 2%) Same Screen-Film Mammo Assure HVL is adequate to minimize patient dose HVL shall meet FDA s s Performance Standards FFDM Same Note the lead sheet
23 Slide 89 Dose Slide 90 Radiation Output Screen-Film Mammo Dose for single CC view of ACR phantom shall not exceed 3.0 mgy per exposure FFDM Same Screen-Film Mammo System shall be capable of producing a minimum output of kvp Mo/Mo over 3.0 second period FFDM Same S value confirmation Exposure Linearity Slide 91 Radiation Output Slide 92 GE 2000D, DS, Essential Fischer Senoscan Lorad Selenia Siemens Novation Same as Screen-Film S Value Confirmation Exposure Linearity S Value Confirmation Make exposure at 25 kv Mo/Mo to give > 20 mr Exposure cassette at this technique (wait 10 min. & process) Record S value Calculate Corrected S value Action Limit: : Corrected S value should not exceed the range of % (96 < S < 144)
24 Slide 93 Fischer Senoscan Slide 94 Viewing & Viewing Conditions Exposure Linearity and Reproducibility Procedure: Measure exposure at specified techniques Calculate: Output = Average exposure reading/ma Linearity = (A-B)/(A+B) < 0.08 Reproducibility = StdDev/Avg < for each technique Screen-Film Mammo Luminance (>( 3000 cd/m 2 ) Room illuminance (<( 50 lux) Viewbox cleaning FFDM Same Slide 95 Assessing Tech QC Slide 96 Film Processing Screen-Film Mammo Assess site s s QC program Check that QC tests are properly performed Independent checks of tech QC tests FFDM Same Screen-Film Mammo Measure optical densities FFDM Density difference, mid-density, density, base+fog Measures consistency Manufacturers recommendations Some refer to printer manufacturers recommendations Typically identical to ACR SFM Manual
25 MD Step=12 Mid-Density Upper Control Limit = 2.06 Mid-Density Daily Value Mid-Density Daily Value Operating Level = 1.91 Mid-Density Lower Control Limit = Mid-Density Upper Control Limit = 1.48 Mid-Density Daily Value Mid-Density Daily Value Operating Level = 1.33 Mid-Density Lower Control Limit = HD Step=13 LD Step=10 Base + Fog Action Limit = 0.22 Base + Fog Daily Value Base + Fog - Operating Level = 0.19 Density Difference Upper Control Limit = 2.25 Density Difference Daily Value Density Difference Daily Value Operating Level = 2.1 Density Difference Lower Control Limit = Base + Fog Action Limit = 0.23 Base + Fog Daily Value Base + Fog - Operating Level = 0.2 Density Difference Upper Control Limit = 1.77 Density Difference Daily Value Density Difference Daily Value Operating Level = 1.62 Density Difference Lower Control Limit = 1.47 Slide 97 Film Processor QC Slide 98 Film Processing GE 2000D, DS, Essential Fischer Senoscan Follow Printer Manufacturers QC If Not, Use Theirs Lorad Selenia Siemens Novation Follow FFDMs QC Wet Processing Northwestern Memorial Hospital - Processor ID: 2 - MAR 2006 Base + Fog Base + Fog Laser Printer Northwestern Memorial Hospital - Processor ID: Kodak Oct 2006 Base + Fog 1-Mar-06 2-Mar-06 3-Mar-06 6-Mar-06 7-Mar-06 8-Mar-06 9-Mar Mar Mar Mar-06 service 3/14 15-Mar Mar-06 Date 17-Mar Mar Mar Mar Mar Mar Mar Mar Mar Mar Mar-06 October 2, 2006 October 3, 2006 October 4, 2006 October 5, 2006 October 6, 2006 October 9, 2006 October 10, 2006 October 11, 2006 October 12, 2006 October 13, 2006 October 16, 2006 October 17, 2006 Date October 18, 2006 October 19, 2006 October 20, 2006 October 23, 2006 October 24, 2006 October 25, 2006 October 26, 2006 October 27, 2006 October 30, 2006 Slide 99 Film Processing Slide 100 Film Processing Mid-Density Density Difference Wet Processing Laser Printer Wet Processing Laser Printer Northwestern Memorial Hospital - Processor ID: 2 - MAR 2006 Mid-Density Northwestern Memorial Hospital - Processor ID: Kodak Oct 2006 Mid-Density Northwestern Memorial Hospital - Processor ID: 2 - MAR 2006 Density Difference Northwestern Memorial Hospital - Processor ID: Kodak Oct 2006 Density Difference 1-Mar-06 2-Mar-06 3-Mar-06 6-Mar-06 7-Mar-06 8-Mar-06 9-Mar Mar Mar Mar-06 service 3/14 15-Mar Mar-06 Date 17-Mar Mar Mar Mar Mar Mar Mar Mar Mar Mar Mar-06 October 2, 2006 October 3, 2006 October 4, 2006 October 5, 2006 October 6, 2006 October 9, 2006 October 10, 2006 October 11, 2006 October 12, 2006 October 13, 2006 October 16, 2006 October 17, 2006 Date October 18, 2006 October 19, 2006 October 20, 2006 October 23, 2006 October 24, 2006 October 25, 2006 October 26, 2006 October 27, 2006 October 30, Mar-06 2-Mar-06 3-Mar-06 6-Mar-06 7-Mar-06 8-Mar-06 9-Mar Mar Mar Mar-06 service 3/14 15-Mar Mar-06 Date 17-Mar Mar Mar Mar Mar Mar Mar Mar Mar Mar Mar-06 October 2, 2006 October 3, 2006 October 4, 2006 October 5, 2006 October 6, 2006 October 9, 2006 October 10, 2006 October 11, 2006 October 12, 2006 October 13, 2006 October 16, 2006 October 17, 2006 Date October 18, 2006 October 19, 2006 October 20, 2006 October 23, 2006 October 24, 2006 October 25, 2006 October 26, 2006 October 27, 2006 October 30, 2006
26 Slide 101 Film Processing Slide 102 Lorad October 2, 2006 D max Northwestern Memorial Hospital - Processor ID: Kodak Oct 2006 Dmax October 3, 2006 October 4, 2006 October 5, 2006 October 6, 2006 October 9, 2006 October 10, 2006 Laser Printer October 11, 2006 October 12, 2006 October 13, 2006 October 16, 2006 October 17, 2006 October 18, 2006 Date October 19, 2006 Dmax Daily Value Dmax - Operating Level = 3.44 Dmax Action Limit = 3.19 October 20, 2006 October 23, 2006 October 24, 2006 October 25, 2006 October 26, 2006 October 27, 2006 October 30, 2006 October 31, % Laser Printer QC - Daily 10% 90% SMPTE pattern from AWS Measure OD 10%, 40% and 90% Plot mid-density density (40%), density difference (40-10%), lower density (90%) + 15 from baseline Slide 103 RWS QC Slide 104 Multimodality Workstations GE 2000D, DS, Fischer Senoscan Lorad Selenia Essential Follow monitor manufacturers QC If Not, Use FFDM s Siemens Novation FDA Approved Multimodality Workstations AGFA IMPA MA3000 Sectra IDS5/mx Kodak DirectView PACS System 5 McKesson s s PACS Mammo Station Use FFDM manufacturers s s QC icad Second Look 500M Cedera I-ReadMammoI Fuji Synapse GE Seno Advantage & Centricity Fischer Senoview Plus (Cedera) Hologic SecureView D Siemens MammoReportPlus
27 Slide 105 Slide 106 AGFA IMPA MA3000 Clean monitor surface Measure display white & black Measure quality level Measure uniformity Calibrate displays View SMPTE pattern Viewing conditions GE Seno Advantage Viewing conditions check and setting Monitor calibration Image quality SMPTE pattern Analysis of screen uniformity Siemens MammoReport Plus Geometric distortion Reflection Luminance Response Luminance Uniformity Resolution Noise Overall Evaluation Barco I-Guard Check Quality Level Display White Calibration Settings Check Eizo Pattern Check Luminance Check Grayscale Check Uniformity Check Planar Monitor Cleaning Viewing Conditions Check Monitor Calibration Check Image Quality SMPTE Visual Inspection for Display Defects Slide 107 Suggestions Slide 108 Troubleshooting QC Keep a good file system for naming & numbering the different QC tests Archive all digital QC images on a CD Keep QC test tools in a safe place Perform flat-field field imaging before phantom imaging tests Perform testing in the morning Have a system for monitoring the Auto Print/Auto Push functions Repeat test Clean plexiglass & move slightly Questions to ask yourself: Are you using the correct image? Raw vs. Processed? Are the ROI s s in the correct location? Re-boot system Call service Re-calibration solves most image quality problems
28 Slide 109 Key Take Home Points Slide 110 Key Take Home Points Obtain proper hands-on training Must perform manufacturer specific QC tests Artifacts most problems can be seen here Re-booting and/or re-calibrating fixes most problems Laser Printer D max at least 3.5 OD Mid-density density about 1.5 OD Review workstation monitors look at the clinical images! Do they match? Appropriate dark and light levels Be aware of separate QC for review workstations Know, and be involved with, your site QC program Slide 111 Slide 112 Thank You SAM Questions Digital Mammography Quality Assurance: Comparing the Manufacturers Recommendations
29 Slide % 62% 0% 18% In digital mammography, who mandates the pass/fail criteria for site QC? 1. The American College of Radiology 2. The FDA 3. The FFDM unit manufacturer 4. NEMA 5. MQSA 4% 1. Slide 114 Answer: 3 - The FFDM unit manufacturer References MQSA Regulations (e)(6) PHY/frmamcom2.html#s90012 Slide 115 Answer 3: The FFDM unit manufacturer Explanation the quality assurance program shall be substantially the same as the quality assurance program recommended by the image receptor manufacturer, except that the minimum allowable dose for screen- film systems in this section Slide 116 4% 1. 92% 4% 0% How do you accredit a Fuji CR Mammography system which uses both CR and screen-film on the same x-ray system? 1. As 1 mammography unit? 2. As 2 mammography units? 3. You cannot use CR and film on the same unit. 4. CR is exempt from accreditation
30 Slide 117 Answer: 2 As 2 mammography units References Slide 118 Answer: 2 - As 2 mammography units Explanation mammo_faq/mammo_faq_mamac.aspx#8.0.1 As of November 15, 2006 facilities using both screen-film and CR on the same mammography units must accredit these 2 systems as 2 separate units. Slide 119 4% 1. 11% 80% 0% 4% What are the minimum passing ACR phantom scores for the Lorad Selenia for weekly QC? 1. 5 Fibers, 4 Speck Groups, 3 Masses 2. 4 Fibers, 3 Speck Groups, 3 Masses 3. 5 Fibers, 4 Speck Groups, 4 Masses 4. 4 Fibers, 4 Speck Groups, 3 Masses 5. 4 Fibers, 4 Speck Groups, 4 Masses Slide 120 Answer 3: 3 5 Fibers, 4 Speck Groups, 4 Masses References Hologic. Lorad Selenia Quality Control Manual MAN-00093, Revision 003, Bedford (MA): Hologic, 2005
31 Slide 121 Answer 3: 3 5 Fibers, 4 Speck Groups, 4 Masses Explanation The Lorad Selenia QC manual states that the ACR Phantom minimum passing scores are as follows: 5 Fibers 4 Speck Groups 4 Masses
Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities
Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Melissa Martin, M.S., FAAPM, FACR, FACMP AAPM Annual Meeting - Philadelphia July 19, 2010 MO-B-204C-1 Educational Objectives
More informationQuality Control of Full Field Digital Mammography Units
Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of
More informationAcceptance Testing of a Digital Breast Tomosynthesis Unit
Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures
More informationFFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR
FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May 2010 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics, Inc. Physicist's Role in the
More informationSTEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS
STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%
More informationSurveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation AAPM Annual Clinical Meeting Indianapolis, IN August 5, 2013 Learning Objectives Become familiar with the recommendations and
More informationBreast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School
Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis
More informationAspire HD. Program Manual. 2nd Edition - October 2012
Quality Control 1 Aspire HD Quality Control Program Manual 2nd Edition - October 2012 Overview Installation of FDR Mammography QC Program Weekly Test 2 3 4 Quarterly Test 5 Semi-annual Test 6 Annual Test
More informationQuality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc
Quality Control for Stereotactic Breast Biopsy Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc. 716-924-0350 Methods of Imaging Guided Breast Biopsy Ultrasound guided, hand-held needle
More informationCollimation Assessment Using GAFCHROMIC XR-M2
Collimation Assessment Using GAFCHROMIC XR-M2 I. Introduction A method of collimation assessment for GE Senographe full-field digital mammography (FFDM) systems is described that uses a self-developing
More informationMammography: Physics of Imaging
Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness
More informationFFDM -FCRm QC Requirements- What You REALLY Need to Know
FFDM -FCRm QC Requirements- What You REALLY Need to Know Melissa C. Martin, M.S., FACR, FAAPM, FACMP AAPM Annual Meeting July 28, 2008 FDA Approval for Mammography Fuji FCRm System approved for Mammography
More information4/19/2016. Quality Control Activities for the RadiologicTechnologist. Objectives. 3D Tomosynthesis QC differences
Quality Control Activities for the RadiologicTechnologist Quality Control Tests 2D QC Tomosynthesis QC DICOM Printer Quality Control Weekly Detector Flat Field Calibration Weekl Artifact Evaluation Weekly
More information8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey
Implementation of the 2016 ACR Digital Mammography QC Manual Medical Physicist Mammography Equipment Evaluation and Annual Survey Eric A Berns, PhD, FACR Radiologist Responsibilities Radiologist Responsibilities
More informationFacility, Unit and Test Equipment Data
Facility, Unit and Test Equipment Data Medical hysicist's Tests - SAMLE FORMS Facility Information Facility Name Happy Valley Mammography Address Suite 1 Address 1 Oak Street City, State, Zip Anywhere
More informationMAMMOGRAPHY - HIGH LEVEL TROUBLESHOOTING
MAMMOGRAPHY - HIGH LEVEL TROUBLESHOOTING Maynard High New York Medical College SS2001-M.High 1 Objectives: Review MQSA and ACR annual QC tests as opportunities for troubleshooting before a significant
More information7/20/2014. Outline. Outline. Disclosures. Learning Objectives. SBB: Practical Aspects of ACR Accreditation, QC and ACR On Site Surveys
Outline SBB: Practical Aspects of ACR Accreditation, QC and ACR On Site Surveys Robert J. Pizzutiello, MS, FACR, FAAPM, FAC Residency Program Director, Upstate Medical Physics, PC Senior Vice President,
More informationOverview of Safety Code 35
Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)
More informationAnsur TNT Users Manual. Plug-In
Ansur TNT 12000 Plug-In Users Manual August 2009, Rev. 2, 12/09 2009 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their
More informationCOMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED
More informationREQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS
REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS DEPARTMENT OF HEALTH DIRECTORATE: RADIATION CONTROL Implementation date: 31 March 2009 Contents
More informationI. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA
FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE
More informationThe Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System
The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System Manus
More informationTHE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY
THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org INTRODUCTION
More informationNuclear Associates
Nuclear Associates 07-647 R/F QC Phantom Operators Manual March 2005 Manual No. 07-647-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. All product names are trademarks of their respective companies
More informationY11-DR Digital Radiography (DR) Image Quality
Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures
More informationInvestigation of the line-pair pattern method for evaluating mammographic focal spot performance
Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann
More informationImage Display and Perception
Image Display and Perception J. Anthony Seibert, Ph.D. Department of Radiology UC Davis Medical Center Sacramento, California, USA Image acquisition, display, & interpretation X-rays kvp mas Tube filtration
More informationQC Testing for Computed Tomography (CT) Scanner
QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including
More informationQuality control for digital mammography: Part II recommendations from the ACRIN DMIST trial
Quality control for digital mammography: Part II recommendations from the ACRIN DMIST trial Martin J. Yaffe, Aili K. Bloomquist, Gordon E. Mawdsley, Etta D. Pisano, R. Edward Hendrick, Laurie L. Fajardo,
More informationDigital Breast Tomosynthesis
Digital Breast Tomosynthesis OLIVE PEART MS, RT(R) (M) HTTP://WWW.OPEART.COM 2D Mammography Not 100% effective Limited by tissue superimposition Overlapping tissue can mask tumors False negative Overlapping
More informationDISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter
DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools
More informationACPSEM Position Paper RECOMMENDATIONS FOR A DIGITAL MAMMOGRAPHY QUALITY ASSURANCE PROGRAM V4.0
Heggie et al ACPSEM Position Paper: Digital Mammography V4.0 ACPSEM Position Paper RECOMMENDATIONS FOR A DIGITAL MAMMOGRAPHY QUALITY ASSURANCE PROGRAM V4.0 JCP Heggie 1, P Barnes 2, L Cartwright 3, J Diffey
More informationDIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook
DIAGNOSTIC ACCREDITATION PROGRAM Radiology and CT Quality Control Procedures Workbook Quality Control Procedures Radiography/CR/DR Safety Code 35 Summary For more detail about each quality control (QC)
More informationHistory of digital imaging
CR/QA RADCHEX History of digital imaging Early, crude digital detectors were developed in the 1970 s Image quality was problematic Processing time of digital images was untenable Viewing, transfer and
More informationNJDEP Medical Physicist s Radiographic QC Survey Registration Number:
Facility Name NJDEP ID # NJDEP Medical Physicist s Radiographic QC Survey PLEASE PRINT Facility Information Unit Information Manufacturer Model Console Model # Console serial # Tube serial # Location (room)
More informationA Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA
A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations Shirley Weddle, RT(R)(M), CIIP, BBA OBJECTIVES Define Computed Radiography (CR) Discuss CR vendor
More informationLearning Objectives: What s my motivation? (unknown screen actor) Workshop Overview
Practical Medical Physics Adapting Traditional Clinical Medical Physics to Digital Radiography Charles E. Willis, Ph.D., DABR Associate Professor Department of Imaging Physics The University of Texas M.D.
More informationMILADY. Product Data. Page 1 of 8
Page 1 of 8 The MILADY Mammographic Unit offers the best quality-to-price ratio to our customers worldwide. The unit advanced technology together with the application of industrial production standards,
More informationFeatures and Weaknesses of Phantoms for CR/DR System Testing
Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal
More informationTest Equipment for Radiology and CT Quality Control Contents
Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3
More informationExposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?
Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago
More informationBreast Imaging Basics: Module 10 Digital Mammography
Module 10 Transcript For educational and institutional use. This test bank is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.
More informationDigital radiography (DR) post processing techniques for pediatric radiology
Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions
More informationPatient-Assisted Compression Impact on Image Quality and Workflow
Patient-Assisted Compression Impact on Image Quality and Workflow Senographe Pristina In 2017, GE Healthcare s Senographe Pristina ( Pristina ) was approved by the FDA using the standard technologist-controlled
More informationRAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II
RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated
More informationMammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over
Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on
More informationof sufficient quality and quantity
of sufficient quality and quantity The patient s body attenuates the beam as it passes though the body More energy is deposited in organs located near the entry of the beam than near the exit of the beam
More informationEssentials of Digital Imaging
Essentials of Digital Imaging Module 7 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 7 Quality 1. ASRT Animation 2. Welcome Welcome to the Essentials of Digital Imaging:
More informationExposure System Selection
Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others
More informationComputed Radiography
BAM Berlin Computed Radiography --INDE 2007, Kalpakkam, India -- Uwe Zscherpel, Uwe Ewert BAM Berlin, Division VIII.3 Requests Requests and and information information to: to: Dr. Dr. U. U. Zscherpel Zscherpel
More informationNuclear Associates
Nuclear Associates 07-649 CDRH Fluoroscopic Phantom Users Manual March 2005 Manual No. 07-649-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More informationSlide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?
Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM
More informationEvaluation of a quality control phantom for digital chest radiography
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 2, SPRING 2001 Evaluation of a quality control phantom for digital chest radiography Eugene Mah* Department of Radiology, Medical University
More informationProtocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems
Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems Draft version 0.10 February 2013 European Reference Organisation for Quality Assured Breast
More informationTECHNICAL DATA. GIOTTO IMAGE SDL/W is pre-arranged for Full Field Digital Biopsy examination with the patient in prone position.
Ver. 01/06/07 TECHNICAL DATA GIOTTO IMAGE SDL/W LOW DOSE, FULL FIELD DIGITAL MAMMOGRAPHY UNIT USING AMORPHOUS SELENIUM (a-se) TECHNOLOGY DETECTOR (pre-arranged for stereotactic biopsy with the same digital
More informationX-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto
X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring
More informationMinimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations
uwe.ewert@bam.de Minimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations Uwe Ewert and Uwe Zscherpel BAM Federal Institute for
More informationImaging Technique Optimization of Tungsten Anode FFDM System
Imaging Technique Optimization of Tungsten Anode FFDM System Biao Chen a*, Andrew P. Smith b, Zhenxue Jing a, Elena Ingal a a Hologic, Inc. 600 Technology Drive, DE 1970 b Hologic, Inc. 35 Crosby Drive,
More informationCompliance Guidance for RADIOGRAPHIC QUALITY CONTROL (5 th Edition)
Compliance Guidance for RADIOGRAPHIC QUALITY CONTROL (5 th Edition) New Jersey Department of Environmental Protection Bureau of X-ray Compliance PO Box 420 Mail Code 25-01 Trenton, NJ 08625-0420 FAX 609-984-5811
More informationHalf value layer and AEC receptor dose compliance survey in Estonia
Half value layer and AEC receptor dose compliance survey in Estonia K. Kepler, A. Vladimirov Training Centre of Medical Physics, University of Tartu Testing Centre of the University of Tartu, Estonia E-mail:
More informationOutline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control
Understanding Digital Modalities: Image Quality and Dose S. Jeff Shepard, M.S. University of Texas M. D. Anderson Cancer Center Houston, Texas Special Acknowledgement: Stephen K. Thompson, M.S. William
More informationBeam-Restricting Devices
Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the
More informationAcquisition, Processing and Display
Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image
More informationDigital Mammography Quality Control for the Mammographic Technologist
Ontario Breast Screening Program Digital Mammography Quality Control for the Mammographic Technologist Authors: G.E. Mawdsley, A.K. Bloomquist, M.J. Yaffe October 2011 Revision 3.1 Mammographic Physics
More informationTESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist
TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist Topics Image Uniformity and Artifacts Image Quality - Detail
More informationDigital Imaging Considerations Computed Radiography
Digital Imaging Considerations Digital Radiography Computed Radiography o Cassette based Direct or Indirect Digital Radiography o Cassetteless Computed Radiography 1 CR Image Acquisition Most like conventional
More informationDigital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of
Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging
More informationDisclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography
Current Implementation Status of IEC Standard 62494-1: Exposure Index (EI) for Digital Radiography July 31, 2017 Ryan Fisher, PhD, DABR Katie Hulme, MS, DABR None Disclosures Outline Review of IEC Standard
More informationTeaching Digital Radiography and Fluoroscopic Radiation Protection
Teaching Digital Radiography and Fluoroscopic Radiation Protection WCEC 20 th Student Educator Radiographer Conference Dennis Bowman, RT(R), CRT (R)(F) Community Hospital of the Monterey Peninsula (CHOMP)
More informationComparison of computed radiography and filmõscreen combination using a contrast-detail phantom
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 1, WINTER 2003 Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom Z. F. Lu,* E. L. Nickoloff, J.
More information7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline
Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation Image Quality Review I: Basics and Image Quality TH-A-16A-1 Thursday 7:30AM - 9:30AM Room: 16A J. Anthony
More informationRadiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893
Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -
More informationGE Healthcare. Essential for life. Senographe Essential Full-Field Digital Mammography system
GE Healthcare Essential for life Senographe Essential Full-Field Digital Mammography system Excellence in FFDM is a process. An ongoing quest, fueled by our continuing breakthroughs in breast cancer detection
More informationSUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES.
SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES. Tingberg, Anders Published in: Radiation Protection Dosimetry DOI: 10.1093/rpd/ncs302 Published: 2013-01-01 Link to publication Citation for published
More informationProtocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems
Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems Draft version 0.15 January 2014 European Reference Organisation for Quality Assured Breast
More informationMammography Solution. AMULET Innovality. The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available.
Mammography Solution AMULET Innovality The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available. FUJIFILM supports the Pink Ribbon Campaign for early detection of
More information3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development
Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory
More informationLudlum Medical Physics
Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to
More informationPublished text: Institute of Cancer Research Repository Please direct all s to:
This is an author produced version of an article that appears in: MEDICAL PHYSICS The internet address for this paper is: https://publications.icr.ac.uk/1316/ Copyright information: http://www.aip.org/pubservs/web_posting_guidelines.html
More informationRaySafe Xi. Specifications
RaySafe Xi Specifications raysafe Xi General EMC tested Exposure needed Reset Temp. range According to EN 61000-6-1:2007 and en 61000-6-3:2007 One Automatic 15 35 C (59 95 F) DetECtor cable length 2 and
More informationSYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology
CODE: RADT 156 INSTITUTE: Health Science TITLE: Equipment Operation I DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of equipment operation and maintenance of radiographic
More informationDRAFT Technical evaluation of Philips Microdose SI digital mammography system
DRAFT Technical evaluation of Philips Microdose SI digital mammography system NHSBSP Equipment Report 1310 August 2013 About the NHS Cancer Screening Programmes The national office of the NHS Cancer Screening
More informationUnit thickness. Unit area. σ = NΔX = ΔI / I 0
Unit thickness I 0 ΔI I σ = ΔI I 0 NΔX = ΔI / I 0 NΔX Unit area Δx Average probability of reaction with atom for the incident photons at unit area with the thickness of Delta-X Atom number at unit area
More informationKODAK DIRECTVIEW CR Mammography Feature User s Guide
KODAK DIRECTVIEW CR Mammography Feature User s Guide 17 September 2010 9G3741 Version 1.0 Carestream Health, Inc. 150 Verona Street Rochester, NY 14608 CARESTREAM, DIRECTVIEW, and DRYVIEW are trademarks
More informationLearning Objectives. Outline. Getting Started with CR. Converting the Radiology Department from Film-Screen to Digital: Making the Transition
Converting the Radiology Department from Film-Screen to Digital: Making the Transition S. Jeff Shepard, MS, DABR University of Texas M. D. Anderson Cancer Center Houston, Texas jshepard@di.mdacc.tmc.edu
More informationNuclear Associates EZ CR-DIN Phantoms
Nuclear Associates 07-605-7777 EZ CR-DIN Phantoms Users Manual August 2006 Manual No. 07-605-7777-1 Rev. 4 2006 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More informationImage Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images
B R E A S T I M A G I N G S O L U T I O N S Setting the benchmark for mammography M-IV Series Innovations in breast imaging The Lorad M-IV Series exemplifies Hologic's commitment to developing advanced
More informationSafelight Fog does what to contrast and density on film?
Terri Jurkiewicz Safelight Fog does what to contrast and density on film? ANSWER INCREASES DENSITY DECREASES CONTRAST Explain how you determine if the focal spot size is within appropriate limits.
More informationImage Quality Artifacts in Digital Imaging
MAHIDOL UNIVERSITY Wisdom of the Land Image Quality Artifacts in Digital Imaging Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University, Bangkok,
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 16371-1 First edition 2011-10-01 Non-destructive testing Industrial computed radiography with storage phosphor imaging plates Part 1: Classification of systems Essais non destructifs
More informationGE Healthcare. Senographe 2000D Full-field digital mammography system
GE Healthcare Senographe 2000D Full-field digital mammography system Digital has arrived. The Senographe 2000D Full-Field Digital Mammography (FFDM) system gives you a unique competitive advantage. That
More informationNathan Childress, Ph.D., DABR
Nathan Childress, Ph.D., DABR Introduction TG-142 is a comprehensive QA protocol Covers nearly every aspect of machine and safety QA Recommends quantitative results Recommends high testing frequencies
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationMARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005
ACR MRI accreditation program MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics,
More informationNew spectral benefi ts, proven low dose
New spectral benefi ts, proven low dose Philips MicroDose mammography SI, technical data sheet Philips MicroDose SI with single-shot spectral imaging is a fullfi eld digital mammography solution that delivers
More informationDigital Radiography : Flat Panel
Digital Radiography : Flat Panel Flat panels performances & operation How does it work? - what is a sensor? - ideal sensor Flat panels limits and solutions - offset calibration - gain calibration - non
More informationAn Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc.
An Introduction to TG-142 Imaging QA Using Standard Imaging Products Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. Goals Understand the nature and intent of TG 142 imaging
More informationDigital Mammography Quality Control for the Mammographic Physicist
Ontario Breast Screening Program Digital Mammography Quality Control for the Mammographic Physicist Authors: G.E. Mawdsley, A.K. Bloomquist, M.J. Yaffe March 2014 Revision 3.2 Mammographic Physics Consulting
More informationFilm Replacement in Radiographic Weld Inspection The New ISO Standard
BAM Berlin Film Replacement in Radiographic Weld Inspection The New ISO Standard 17636-2 Uwe Ewert, Uwe Zscherpel, Mirko Jechow Requests and information to: uwez@bam.de 1 Outline - The 3 essential parameters
More information