1 Image Quality, Artifacts and Hazards in Imaging Laura Gruber, MBA, RT(R), RDMS, RVT Sr. Director Medical Imaging

2 Case 1 2

3 Case 1 What is the image quality issue in this picture? A.) Improper exposure/technique B.) Detector burn-in or ghosting artifact C.) Electromagnetic interference artifact (EMI) D.) Artifact from a thermoregulating blanket Case Courtesy of Judy Le, R.T. (CT) (R) 3

4 Case 1 What is the image quality issue in this picture? A.) Improper exposure/technique B.) Detector burn-in or ghosting artifact C.) Electromagnetic interference artifact (EMI) D.) Artifact from a thermoregulating blanket Case Courtesy of Judy Le, R.T. (CT) (R) 4

5 Case 1 Key Concepts: 1.) Thermo-regulating blankets can cause honey-comb like artifacts if they are left in place during xray. These artifacts do not obscure visibility of lines, however, they can be distracting when evaluating bowel gas and potential free air. 3.) When evaluating lines and lungs, thermoregulating blankets are acceptable to leave in if patient condition prohibits removal. If evaluating pathology in the abdomen, artifacts caused by these blankets may affect diagnostic interpretation. Case Courtesy of Judy Le, R.T. (CT) (R) 5

6 Case 2 6

7 Case 2 What is the image quality issue in this picture? A.) Baby should be prone B.) Immobilization board will create significant artifact C.) Lead shield in the fluoroscopic field of view D.) Using an image intensifier instead of flat panel fluoroscope 7

8 Case 2 What is the image quality issue in this picture? A.) Baby should be prone B.) Immobilization board will create significant artifact C.) Lead shield in the fluoroscopic field of view D.) Using an image intensifier instead of flat panel fluoroscope 8

9 Case 2 Key Concepts: 1.) All modern fluoroscopes uses automatic brightness control (ABC) which means the machine adjusts the xray tube output based on how much attenuation it detects. 2.) A lead gonad shield in the fluoroscopy field of view will cause the fluoroscopy system to substantially increase radiation to the patient AND compromise image quality. 3.) You should also not use lead shields in the field of view for other modalities that use automatic exposure control (ex. xray=phototiming, CT=dose modulation) 9

10 Case 3 10

11 Case 3 What is the image quality issue in this picture? A.) Improper exposure/technique B.) Detector burn-in or ghosting artifact C.) Electromagnetic interference artifact (EMI) D.) Artifact from a positioning device 11

12 Case 3 What is the image quality issue in this picture? A.) Improper exposure/technique B.) Detector burn-in or ghosting artifact C.) Electromagnetic interference artifact (EMI) D.) Artifact from a positioning device 12

13 Case 3 Key Concepts: 1.) Scoliosis exams are often performed with a series of 2 or 3 images exposing the same detector in rapid succession. 2.) Though rare, the burn-in artifact can be seen in high mas scoli exams, it is most often noted as a repeated image of the xray marker. 13

14 Case 4 14

15 Case 4 What is the image quality issue in this picture? A.) Use of Lap Shield B.) Room is not suitable for portable xray imaging C.) Source-to-image distance is too short D.) Portable xray tube should be under the table. 15

16 Case 4 What is the image quality issue in this picture? A.) Use of Lap Shield B.) Room is not suitable for portable xray imaging C.) Source-to-image distance is too short D.) Portable xray tube should be under the table. 16

17 Case 4 There is a ruler provided to help you determine the SID 17

18 Case 4 Key Concepts: 1.) Portable xray protocols have defined SID s, make sure you try to approximate the correct distance so the dose is appropriate and the image quality is consistent. 2.) Use the ruler provided by the system and pick the appropriate protocol (27, 40, or 72 ) 18

19 Case 5 19

20 Case 5 What is the image quality issue in this picture? A.) Improper exposure/technique B.) Detector burn-in or ghosting artifact C.) Electromagnetic interference artifact (EMI) D.) Artifact from a positioning device Case Courtesy of Judy Le, R.T. (CT) (R) 20

21 Case 5 What is the image quality issue in this picture? A.) Improper exposure/technique B.) Detector burn-in or ghosting artifact C.) Electromagnetic interference artifact (EMI) D.) Artifact from a positioning device Case Courtesy of Judy Le, R.T. (CT) (R) 21

22 Case 5 Key Concepts: 1.) Gel pillows are filled with a non-toxic water based gel that can provide the support needs of micro-preemie, preemie, and full-term infants 2.) Removal of gel pillows is highly recommended for any indication. Artifacts caused by gel pillows can mimic pathology, such as free air, which can lead to misinterpretation by Radiologists. 3.) These gel positioning devices can lead to additional x-rays to confirm artifact vs. pathology, resulting in increased radiation dose to the patient. Case Courtesy of Judy Le, R.T. (CT) (R) 22

23 Case 6 23

24 Case 6 What is the image quality issue in this picture? A.) Chest chair is too large B.) Baby should have arms down C.) Holding is never acceptable in pediatric radiology D.) Poor collimation 24

25 Case 6 What is the image quality issue in this picture? A.) Chest chair is too large B.) Baby should have arms down C.) Holding is never acceptable in pediatric radiology D.) Poor collimation 25

26 Case 6 Key Concepts: 1.) Poor collimation is a source of poor image quality in digitally processed images. 2.) Differences in collimation introduces variability in the source images and inconsistent results on processed images. 2.) Collimation is also very effective at reducing our patient s exposure (and it also reduces the exposure of anyone who is holding the child). Always collimate as much as reasonable! 26

27 Case 7 27

28 Case 7 What is the image quality issue in this picture? A.) Patient should be feet first first B.) Patient is off iso-center C.) Incorrect pediatric table insert in place D.) Incorrect position on the table in the head-to-foot direction 28

29 Case 7 What is the image quality issue in this picture? A.) Patient should be feet first first B.) Patient is off iso-center C.) Incorrect pediatric table insert in place D.) Incorrect position on the table in the head-to-foot direction 29

30 Case 7 Automatic dose control requires good isocentering. 30

31 Case 7 Magnification = SID/SOD 31

32 Z-axis dose modulation Bushberg JT, et al. The Essential Physics of Medical Imaging, Wolters Kluwer, Philadelphia, 3 rd Edition,

33 Z-axis dose modulation Bushberg JT, et al. The Essential Physics of Medical Imaging, Wolters Kluwer, Philadelphia, 3 rd Edition,

34 Case 7 Key Concepts: 1.) Use your lasers to ensure very good centering of the patient in the AP and RL directions. 2.) If you do notice you are off-isocenter, do not move your planning box to accommodate. Go back in the room and physically reposition your table and acquire another topogram. 34

35 Case 8 35

36 Case 8 What is the image quality issue in this picture? A.) Cosmetics containing glitter may lead to tissue heating and discomfort during MRI B.) Cosmetics containing glitter may lead to imaging artifacts in MRI C.) Glitter tattoos can lead to bright spots in the MRI image D.) Both A and B 36

37 Case 8 What is the image quality issue in this picture? A.) Cosmetics containing glitter may lead to tissue heating and discomfort during MRI B.) Cosmetics containing glitter may lead to imaging artifacts in MRI C.) Glitter tattoos can lead to bright spots in the MRI image D.) Both A and B 37

38 Case 8 38

39 Case 8 39

40 Case 8 40

41 41

42 Case 9 42

43 Case 9 What is the image quality issue in this picture? A.) Artifact due to patient motion B.) Iterative reconstruction artifact C.) Artifact due to contrast extravasation D.) Ring artifact from detector failure 43

44 Case 9 What is the image quality issue in this picture? A.) Artifact due to patient motion B.) Iterative reconstruction artifact C.) Artifact due to contrast extravasation D.) Ring artifact from detector failure 44

45 Case 9 Key Concepts: 1.) Contrast extravasation means that the scan quality will be compromised and possibly non-diagnostic. 2.) Always check with an attending before continuing a scan where extravasation is suspected. 45

46 Case 10 46

47 Case 10 What is the image quality issue in this picture? A.) Post-processing error B.) Improper exposure/technique C.) Electromagnetic interference artifact (EMI) D.) Artifact from a positioning device TG Sandridge, M Karastanovic, Radiologic Technology, January/February 2015, Volume 86, Number 3 47

48 Case 10 What is the image quality issue in this picture? A.) Post-processing error B.) Improper exposure/technique C.) Electromagnetic interference artifact (EMI) D.) Artifact from a positioning device TG Sandridge, M Karastanovic, Radiologic Technology, January/February 2015, Volume 86, Number 3 48

49 Case 10 Key Concepts: 1.) Some devices such as this thermoregulation unit can interfere electromagnetically with digital detectors. 2.) Electromagnetic interference artifacts often manifest as parallel lines and can be eliminated if the source of the interference can be removed or turned off. TG Sandridge, M Karastanovic, Radiologic Technology, January/February 2015, Volume 86, Number 3 49

50 Case 10 Thermoregulation Unit ON Thermoregulation Unit OFF TG Sandridge, M Karastanovic, Radiologic Technology, January/February 2015, Volume 86, Number 3 50

51 Case 11 51

52 Case 11 What is the hazard or compliance issue in this picture? 52

53 Case 11 Answer: Objects depressing the fluoroscopy foot-pedal! 53

54 Case 11 Key Concepts: 1.) Always remember to practice safety with foot-pedal operated devices. Keep them clear of equipment which can fall on to the pedal. Inhibit xrays if possible when the device is not in use. 2.) Become familiar with the beam-on alerts for your device, both visual and audio. 3.) Always wear your radiation badge in case of unexpected exposures. 54

55 Case 12 55

56 Case 12 What is the hazard or compliance issue in this picture? 56

57 Case 12 Answer: Someone ripped off the spacer cone! 57

58 Case 12 Answer: The spacer cone is there to protect the patient s skin. 58

59 Case 12 Key Concepts: 1.) Never remove the spacer cone. This is in place to ensure the entrance skin dose to the patient is not too high and is regulated by the State and Joint Commission. 2.) Remember that the xray radiation is highest directly at the tube output and decreases by 1/distance^2. Entrance skin dose is highest if anatomy is closer to tube output, so the spacer cone is there to reduce the likelihood of skin burns. 59

60 Case 13 60

61 Case 13 What is the hazard or compliance issue in this picture? 61

62 Case 13 Answer: This physician is entering the MRI room with his pager and stethoscope! 62

63 Case 13 Key Concepts: 1.) Most metal object MRI accidents occur because of untrained or absent-minded people entering with common items. 2.)As the most trained staff in the hospital, it s our job to always be vigilant with screening (even if it is a doctor or supervisor who is entering the MRI room). 63

64 Case 14 64

65 Case 14 What is the hazard or compliance issue in this picture? 65

66 Case 14 Answer: A magnet quench may cause vaporized helium (a white smoky substance) to fill the MRI scanner room. 66

67 Case 14 MRI scanners are built with a exhaust for vaporized helium. 67

68 Case 14 Example of vaporized helium exhaust from a quench. Failure of the exhaust system can lead to very high gas pressures in the scanner room. 68

69 Case 14 Key Concepts: 1.) MRI system use superconducting technology that must be kept at cryogenic temperatures. Any disturbance to this cryo system can lead to a quench. 2.) Quenches happen more often than you might think. Generally there is a audible loud bang, an immediate loss of field strength, and the room will fill with helium gas. 2.)Immediately evacuate all personnel from the area to avoid breathing in helium vapor and contact HTM. 69

70 Case 15 70

71 Case 15 What is the hazard or compliance issue in this picture? 71

72 Case 15 Answer: Drinking cups in the Nuclear Medicine Hotlab! 72

73 Case 15 Key Concepts: 1.) Eating, drinking, applying cosmetics or creams can introduce radio-isotopes to your body and lead to internal exposure. 2.) Refrain from these activities in all radioactive material work areas and avoid touching your eyes, nose, and lips. 73

74 Case 16 74

75 Case 16 What is the hazard or compliance issue in this picture? 75

76 Case 16 Answer: This staff member is standing in the highest radiation scatter zone! 76

77 Case 16 Scatter profile for CT scanners is forward along the table. 77

78 Case 16 The side of the CT scanner has lower scatter and is a preferable place to stand than next to the table. 78

79 Case 16 Key Concepts: 1.) The CT scanner tend to throw radiation forward and backward along the length of the patient bed. 2.) If possible, it s better for staff and family member to sneak around to the side of the scanner during the acquisition to reduce their own exposure. 79

80 Case 17 80

81 Case 17 What is the hazard or compliance issue in this picture? 81

82 Case 17 Answer: Staff member is using frontal apron but has back turned to source of radiation! 82

83 Frontal Drop Lead Apron Type 83

84 Skirt Vest Type Apron 84

85 Case 17 Key Concepts: 1.) Be careful if you are wearing a front apron that you are facing the source of radiation and do not get your body rotated. 2.) Wear a skirt/vest style apron in circumstances when you cannot always face the procedure (ie. if you are circulating). 3.) If you have no other option, wear two front lead aprons, one front-ward, the other back-ward to protect your organs from all sides. 85

86 Acknowledgements Jackie Choragwicki Merima Karastanovic Judy Le THANK YOU! Christina L. Sammet, Ph.D., DABR Medical Physicist Radiation and Laser Safety Officer 86

Teaching Digital Radiography and Fluoroscopic Radiation Protection

Teaching Digital Radiography and Fluoroscopic Radiation Protection Teaching Digital Radiography and Fluoroscopic Radiation Protection WCEC 20 th Student Educator Radiographer Conference Dennis Bowman, RT(R), CRT (R)(F) Community Hospital of the Monterey Peninsula (CHOMP)

More information

Multi-Access Biplane Lab

Multi-Access Biplane Lab Multi-Access Biplane Lab Advanced technolo gies deliver optimized biplane imaging Designed in concert with leading physicians, the Infinix VF-i/BP provides advanced, versatile patient access to meet the

More information

INNOVATION BY DESIGN. Toshiba A History of Leadership REMOTE CONTROL R/F SYSTEM

INNOVATION BY DESIGN. Toshiba A History of Leadership REMOTE CONTROL R/F SYSTEM INNOVATION BY DESIGN For over 130 years, Toshiba has led the world in developing technology to improve the quality of life. This Made for Life TM commitment is reflected in our family of leading-edge imaging

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

of sufficient quality and quantity

of sufficient quality and quantity of sufficient quality and quantity The patient s body attenuates the beam as it passes though the body More energy is deposited in organs located near the entry of the beam than near the exit of the beam

More information

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS Veterinary Science Preparatory Training for the Veterinary Assistant Floron C. Faries, Jr., DVM, MS Radiology Floron C. Faries, Jr., DVM, MS Objectives Determine the appropriate machine settings for making

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced

More information

Truly flexible to meet your clinical needs

Truly flexible to meet your clinical needs Truly flexible to meet your clinical needs 2 Adapting to meet your needs Flexible Fast and responsive Excellent image quality Designed with ergonomic efficiency Equipped with dose management tools 3 Three

More information

INTRODUCTION TO FLEXIBLE BRONCHOSCOPY. Fluoroscopy Synopsis HENRI G COLT MD SECOND EDITION THE BRONCHOSCOPY EDUCATION PROJECT SERIES

INTRODUCTION TO FLEXIBLE BRONCHOSCOPY. Fluoroscopy Synopsis HENRI G COLT MD SECOND EDITION THE BRONCHOSCOPY EDUCATION PROJECT SERIES SECOND EDITION INTRODUCTION TO FLEXIBLE BRONCHOSCOPY Fluoroscopy Synopsis HENRI G COLT MD With contributions from Dr. S. Murgu THE BRONCHOSCOPY EDUCATION PROJECT SERIES FLUOROSCOPY SYNOPSIS The purpose

More information

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Sarah Hughes, MS, DABR Radiation Safety Officer

Sarah Hughes, MS, DABR Radiation Safety Officer Sarah Hughes, MS, DABR Radiation Safety Officer 502-852-6146 sarah.hughes@louisville.edu Mo my back is burnin!!! I got it MAG the cine! Sumthin s not right. Where s his heart? Fluoroscopy http://dccwww.bumc.bu.edu/fluoroscopy/def

More information

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory

More information

Ludlum Medical Physics

Ludlum Medical Physics Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

Digital radiography (DR) post processing techniques for pediatric radiology

Digital radiography (DR) post processing techniques for pediatric radiology Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions

More information

SYLLABUS. 1. Identification of Subject:

SYLLABUS. 1. Identification of Subject: SYLLABUS Date/ Revision : 30 January 2017/1 Faculty : Life Sciences Approval : Dean, Faculty of Life Sciences SUBJECT : Biophysics 1. Identification of Subject: Name of Subject : Biophysics Code of Subject

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

PD233: Design of Biomedical Devices and Systems

PD233: Design of Biomedical Devices and Systems PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/

More information

The Architecture of Medical Imaging

The Architecture of Medical Imaging University of Kansas Architecture 731 Systems and Components of Healthcare Facilities F. Zilm The Architecture of Medical Imaging Designing Healthcare Facilities for Advanced Radiologic Diagnostic and

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

R/F. Comparison of Long View Radiography Systems. 1. Introduction. 2. Methods of Long View Radiography

R/F. Comparison of Long View Radiography Systems. 1. Introduction. 2. Methods of Long View Radiography R/F Comparison of Long View Radiography Systems Department of Radiology, Tokyo Women's Medical University Medical Center East 1 Department of Central Radiology, Tokyo Women's Medical University Hospital

More information

Image Quality Artifacts in Digital Imaging

Image Quality Artifacts in Digital Imaging MAHIDOL UNIVERSITY Wisdom of the Land Image Quality Artifacts in Digital Imaging Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University, Bangkok,

More information

Advanced digital image processing for clinical excellence in fluoroscopy

Advanced digital image processing for clinical excellence in fluoroscopy Dynamic UNIQUE Digital fluoroscopy solutions Dynamic UNIQUE Advanced digital image processing for clinical excellence in fluoroscopy André Gooßen, PhD, Image Processing Specialist Dörte Hilcken, Clinical

More information

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology The Fundamentals of... Computed Tomography Computed Tomography (CT) systems use x-rays to produce images of slices through a patient s anatomy. Despite having lower spatial resolution than other x-ray

More information

diagnostic examination

diagnostic examination RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011

More information

FOUR CATEGORIES OF SAFETY

FOUR CATEGORIES OF SAFETY OCTOBER 2013 FOUR CATEGORIES OF SAFETY DOSIMETRY PERSONAL SAFETY EQUIPMENT EQUIPMENT KNOWLEDGE PHYSICAL SAFETY DOSIMETRY THERMAL LUMINISCENT DEVICES AND FILM BADGES CNSC PERMISSIBLE DOSES WHOLE BODY DOSE

More information

CR Basics and FAQ. Overview. Historical Perspective

CR Basics and FAQ. Overview. Historical Perspective Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors

More information

Overview of Safety Code 35

Overview of Safety Code 35 Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)

More information

Maximum Performance, Minimum Space

Maximum Performance, Minimum Space TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-706 Patient Phantom/Penetrometer System Users Manual March 2005 Manual No. 07-706-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are

More information

Maximizing clinical outcomes

Maximizing clinical outcomes Maximizing clinical outcomes Digital Tomosynthesis Dual Energy Subtraction Automated Long Length Imaging Improved image quality at a low dose Xray Xray Patented ISS capture technology promotes high sensitivity

More information

1-1. GENERAL 1-2. DISCOVERY OF X-RAYS

1-1. GENERAL 1-2. DISCOVERY OF X-RAYS 1-1. GENERAL Radiography is a highly technical field, indispensable to the modern dental practice, but presenting many potential hazards. The dental radiographic specialist must be thoroughly familiar

More information

GE Healthcare. Senographe 2000D Full-field digital mammography system

GE Healthcare. Senographe 2000D Full-field digital mammography system GE Healthcare Senographe 2000D Full-field digital mammography system Digital has arrived. The Senographe 2000D Full-Field Digital Mammography (FFDM) system gives you a unique competitive advantage. That

More information

CHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

CHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC CHAPTER 6 QC Test For Fluoroscopic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done. Identify QC

More information

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. RA110 test 3 Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. An object 35 cm in width is radiographed at 100 cm SID and at a 50 cm SOD. What

More information

Digital Radiographic Inspection replacing traditional RT and 3D RT Development

Digital Radiographic Inspection replacing traditional RT and 3D RT Development Digital Radiographic Inspection replacing traditional RT and 3D RT Development Iploca Novel Construction Meeting 27&28 March 2014 Geneva By Jan van der Ent Technical Authority International Contents Introduction

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-647 R/F QC Phantom Operators Manual March 2005 Manual No. 07-647-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. All product names are trademarks of their respective companies

More information

Fluoroscopy - Chapter 9

Fluoroscopy - Chapter 9 Fluoroscopy - Chapter 9 Kalpana Kanal, Ph.D., DABR Lecturer, Diagnostic Physics Dept. of Radiology UW Medicine a copy of this lecture may be found at: http://courses.washington.edu/radxphys/physicscourse04-05.html

More information

JEFFERSON COLLEGE COURSE SYLLABUS BET220 DIAGNOSTIC IMAGING. 3 Credit Hours. Prepared by: Scott Sebaugh Date: 2/20/2012

JEFFERSON COLLEGE COURSE SYLLABUS BET220 DIAGNOSTIC IMAGING. 3 Credit Hours. Prepared by: Scott Sebaugh Date: 2/20/2012 JEFFERSON COLLEGE COURSE SYLLABUS BET220 DIAGNOSTIC IMAGING 3 Credit Hours Prepared by: Scott Sebaugh Date: 2/20/2012 Mary Beth Ottinger, Division Chair Elizabeth Check, Dean, Career & Technical Education

More information

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) Hiroki Kadoya Yukiko Kitagawa MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria

More information

NJDEP Medical Physicist s Radiographic QC Survey Registration Number:

NJDEP Medical Physicist s Radiographic QC Survey Registration Number: Facility Name NJDEP ID # NJDEP Medical Physicist s Radiographic QC Survey PLEASE PRINT Facility Information Unit Information Manufacturer Model Console Model # Console serial # Tube serial # Location (room)

More information

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

2 nd generation TOMOSYNTHESIS

2 nd generation TOMOSYNTHESIS 2 nd generation TOMOSYNTHESIS 2 nd generation DBT true innovation in breast imaging synthesis graphy Combo mode Stereotactic Biopsy Works in progress: Advanced Technology, simplicity and ergonomics Raffaello

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

Nuclear Associates , , , , , ,

Nuclear Associates , , , , , , Nuclear Associates 57-411, 57-412, 57-413 57-426, 57-431, 57-432 57-433, 57-435, 57-436 CLEAR-Pb Transparent X-Ray Compensation Filters Users Manual March 2005 Manual No. 57-XXX-1 Rev. 2 2003, 2005 Fluke

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

QC by the MPE in Belgium

QC by the MPE in Belgium Acceptance testing of state-of-the-art CT scanners using a new national protocol: first experience on a large number of scanners of different make and model the working group Radiology of the Belgian Hospital

More information

RAD. Experiences Using the RADspeed Pro EDGE Package. 1. Hospital Description. 2. Background of Adoption. Hirohito Tanaka

RAD. Experiences Using the RADspeed Pro EDGE Package. 1. Hospital Description. 2. Background of Adoption. Hirohito Tanaka RAD Experiences Using the RADspeed Pro EDGE Package Hirohito Tanaka, R.T. Department of Radiology, Chibune General Hospital Hirohito Tanaka 1. Hospital Description Our hospital was originally established

More information

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R Building 3, No.9, Jiuxianqiaodong Road, Chaoyang District, Beijing 100015, P.R. China E-mail:

More information

Essentials of Digital Imaging

Essentials of Digital Imaging Essentials of Digital Imaging Module 6 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 6 Dose Reduction and Patient Safety 1. ASRT Animation 2. Welcome Welcome to Essentials

More information

Features and Weaknesses of Phantoms for CR/DR System Testing

Features and Weaknesses of Phantoms for CR/DR System Testing Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org INTRODUCTION

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

Artefacts found in computed radiography

Artefacts found in computed radiography The British Journal of Radiology, 74 (2001), 195 202 E 2001 The British Institute of Radiology Pictorial review Artefacts found in computed radiography L J CESAR, RT(R)(QM), B A SCHUELER, PhD, F E ZINK,

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

RADspeed Pro. EDGEpackage C501-E041C

RADspeed Pro. EDGEpackage C501-E041C RADspeed Pro EDGEpackage C501-E041C 2 Some of the FPDs may be not available in your country. Please contact us to check the availability in your country. 3 Tomosynthesis in the Standing Position Tomosynthesis

More information

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT) Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in

More information

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them. In press 2004 1 2 Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Digital radiology An appropriate analogy that is easy for most people to understand

More information

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision Minnesota Rules, Chapter 4732 X-ray Revision DRAFT FLUOROSCOPIC X-RAY SYSTEMS, 1.0 Subpart 1. Applicability. Subpart 2. Limitation of the useful beam. Subpart 3. Measuring compliance; primary protective

More information

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure Accuracy and calibration of integrated radiation output indicators in diagnostic radiology: A report of the AAPM Imaging Physics Committee Task Group 190 Pei-Jan P. Lin a) Virginia Commonwealth University

More information

Tomosynthesis and Motion

Tomosynthesis and Motion Tomosynthesis (3D) Motion Unsharpness Occurs at about the same frequency as conventional mammography (2D) Presents the same issues as 2D motion, EXCEPT that motion may go undetected Most common patient-related

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

Diagnostic X-Ray Shielding

Diagnostic X-Ray Shielding Diagnostic X-Ray Shielding Multi-Slice CT Scanners Using NCRP 147 Methodology Melissa C. Martin, M.S., FAAPM, FACR Therapy Physics Inc., Bellflower, CA AAPM Annual Meeting, Orlando, FL FL Refresher Course

More information

DIGITAL RADIOGRAPHY ARTIFACTS

DIGITAL RADIOGRAPHY ARTIFACTS IMAGING LAB MPHY 487 DIGITAL RADIOGRAPHY ARTIFACTS Mohammad Esmael Alsulimane B.Sc, M.Sc Medical Physics Lecturer - Physics Department All Rights Reserved: Some information and figures in this presentation

More information

CS 9300 Family User Guide Including

CS 9300 Family User Guide Including CS 9300 Family User Guide Including CS 9300 CS 9300 Select Notice Congratulations on purchasing this unit of the CS 9300 Family. Thank you for your confidence in our products and we will do all in our

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated

More information

Ch. 223 VETERINARY MEDICINE CHAPTER 223. VETERINARY MEDICINE GENERAL PROVISIONS X-RAYS RADIOACTIVE MATERIAL. Authority

Ch. 223 VETERINARY MEDICINE CHAPTER 223. VETERINARY MEDICINE GENERAL PROVISIONS X-RAYS RADIOACTIVE MATERIAL. Authority Ch. 223 VETERINARY MEDICINE 25 223.1 CHAPTER 223. VETERINARY MEDICINE Sec. 223.1. Purpose and scope. 223.2. [Reserved]. 223.2a. Definitions. 223.3 223.6. [Reserved]. 223.7. Structural shielding. 223.8.

More information

2217 US Highway 70 East Garner, NC Main: Fax:

2217 US Highway 70 East Garner, NC Main: Fax: Viztek is committed to providing the highest image quality possible in our CR & DR product lines. There are several factors that directly affect the overall quality of CR & DR based images. The eposure

More information

Some operation methods show in the catalog reguire optional eguipment

Some operation methods show in the catalog reguire optional eguipment Some operation methods show in the catalog reguire optional eguipment I'm interested in real-time imaging with a larger field of view. I wish to acquire high-definition images while reducing exposure dose.

More information

C501-E029C. RADspeed. GENERAL RADIOGRAPHIC SYSTEM Automatic

C501-E029C. RADspeed. GENERAL RADIOGRAPHIC SYSTEM Automatic C501-E029C RADspeed GENERAL RADIOGRAPHIC SYSTEM Automatic High-Performance General Radiographic System Improves Workflow and Achieves Dose Reduction GENERAL RADIOGRAPHIC SYSTEM Automatic This state-of-the-art

More information

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity) Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular

More information

Ultimate DR flexibility

Ultimate DR flexibility Ultimate DR flexibility to fit your room, workflow, and budget KODAK DIRECTVIEW DR 7500 System KODAK DIRECTVIEW DR 7500 System YOU HAVE NEVER SEEN A DIGITAL RADIOGRAPHY SYSTEM LIKE THIS! Your radiography

More information

10/3/2012. Study Harder

10/3/2012. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

R&F X-ray systsem. Savings With Every Exposure

R&F X-ray systsem. Savings With Every Exposure R&F X-ray systsem Savings With Every Exposure Savings with every exposure Versatility meets value Dependable performance User-friendly interface Expandable imaging capabilities Excellent image quality

More information

160-slice CT SCANNER / New Standard for the Future

160-slice CT SCANNER / New Standard for the Future TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

T h e P h a n t o m L a b o r a t o r y

T h e P h a n t o m L a b o r a t o r y T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order

More information

Wide-Detector CT for TAVR Planning:

Wide-Detector CT for TAVR Planning: Wide-Detector CT for TAVR Planning: Impact on Iodine Dose, Radiation Dose, and Image Quality SCBTMR 2015 Annual Course Thursday, October 8 William P. Shuman MD FSCBTMR Department of Radiology University

More information

Data. microcat +SPECT

Data. microcat +SPECT Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for

More information

Essentials of Digital Imaging

Essentials of Digital Imaging Essentials of Digital Imaging Module 7 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 7 Quality 1. ASRT Animation 2. Welcome Welcome to the Essentials of Digital Imaging:

More information

MaxRay Handheld X-ray Systems Operator Training Exam

MaxRay Handheld X-ray Systems Operator Training Exam MaxRay Handheld X-ray Systems Operator Training Exam Employee: Instructor: ate: Score: Instructions Read each question carefully and choose the best answer. 1) LR is 2) 3) 4) a. a safety principle meant

More information

2017 West Coast Educators Conference Orlando. Projection Geometry. 1. Review hierarchy of image qualities (amplified version):

2017 West Coast Educators Conference Orlando. Projection Geometry. 1. Review hierarchy of image qualities (amplified version): Spatial Resolution in the Digital Age: NOTES Quinn B. Carroll, MEd, RT 2017 West Coast Educators Conference Orlando Projection Geometry 1. Review hierarchy of image qualities (amplified version): a. Maximum

More information

MUSICA Nerve Center. Artificial Intelligence. Intelligent tools for your Digital Radiography workflow. Fluoroscopy. Workflow Optimization

MUSICA Nerve Center. Artificial Intelligence. Intelligent tools for your Digital Radiography workflow. Fluoroscopy. Workflow Optimization Image Quality Bariatric Abdomen Pediatric Imaging Diagnostic Confidence Fluoroscopy Neonatal Imaging Scatter Suppression Dental Full Leg Full Spine Exposure Control Index Artificial Intelligence General

More information

PET/CT Instrumentation Basics

PET/CT Instrumentation Basics / Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire

More information

Digital Imaging CT & MR

Digital Imaging CT & MR Digital Imaging CT & MR January 22, 2008 Digital Radiography, CT and MRI generate images in a digital format What is a Digital Image? A digital image is made up of picture elements, pixels row by column

More information

2D, 3D CT Intervention, and CT Fluoroscopy

2D, 3D CT Intervention, and CT Fluoroscopy 2D, 3D CT Intervention, and CT Fluoroscopy SOMATOM Definition, Definition AS, Definition Flash Answers for life. Siemens CT Vision Siemens CT Vision The justification for the existence of the entire medical

More information

Fabrício Sampaio Péres Kury Federal University of Rio de Janeiro Medical School

Fabrício Sampaio Péres Kury Federal University of Rio de Janeiro Medical School Fabrício Sampaio Péres Kury Federal University of Rio de Janeiro Medical School Harvard Medical School Exchange Clerkship Program Primary Care Radiology Clerkship Gillian Lieberman, M. D. Monday, September

More information

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring

More information

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview Practical Medical Physics Adapting Traditional Clinical Medical Physics to Digital Radiography Charles E. Willis, Ph.D., DABR Associate Professor Department of Imaging Physics The University of Texas M.D.

More information

Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology

Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology Poster No.: C-1190 Congress: ECR 2011 Type: Scientific Exhibit Authors: K. Chida, T. Takahashi, D.

More information