1 Fluoroscopy - Chapter 9 Kalpana Kanal, Ph.D., DABR Lecturer, Diagnostic Physics Dept. of Radiology UW Medicine a copy of this lecture may be found at: 05.html
2 1. Real-Time Imaging Fluoroscopy is an imaging procedure that allows real-time x-ray x viewing of the patient with high temporal resolution Use TV technology, which provides 30 frames per second imaging Allows acquisition of a real-time digital sequence of images (digital video), that can be played back as a movie loop Cine cameras offer up to 120 frame per second acquisition rates using 35- mm cine film. Digital cine also available Kalpana M. Kanal, Ph.D., DABR 2
3 1. Fluoroscopic Imaging Chain Components c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p Kalpana M. Kanal, Ph.D., DABR 3
4 1. The Image Intensifier (II) There are 4 principal components of an II: (a) a vacuum bottle to keep the air out (b) an input layer that converts the x-ray signal to electrons (c) electronic lenses that focus the electrons, and (d) an output phosphor that converts the accelerated electrons into visible light c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p Kalpana M. Kanal, Ph.D., DABR 4
5 1. Input Screen The input screen of the II consists of 4 different layers: (a) vacuum window, a 1 mm aluminum window that is part of the vacuum bottle keeps the air out of the II, and its curvature is designed to withstand the force of the air pressing against it a vacuum is necessary in all devices in which electrons are accelerated across open space c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p Kalpana M. Kanal, Ph.D., DABR 5
6 1. Input Screen The input screen of the II consists of 4 different layers: (b) support layer, which is strong enough to support the input phosphor and photocathode layers, but thin enough to allow most x-rays x to pass through it 0.5 mm of aluminum, is the first component in the electronic lens system, and its curvature is designed for accurate electronic focusing c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p Kalpana M. Kanal, Ph.D., DABR 6
7 The input screen of the II consists of 4 different layers: 1. Input Screen (c) input phosphor, whose function is to absorb the x-rays x and convert their energy into visible light cesium iodide (CsI) is used long, needle-like like crystals which function as light pipes, channeling the visible light toward the photochathode with minimal lateral spreading 400 µm m tall, 5 µm m in diameter c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p Kalpana M. Kanal, Ph.D., DABR 7
8 1. Input Screen The input screen of the II consists of 4 different layers: (d) photocathode is a thin layer of antimony and alkali metals that emits electrons when struck by visible light 10 to 20% conversion efficiency 23 to 35 cm diameter input image (FOV) c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p Kalpana M. Kanal, Ph.D., DABR 8
9 1. Input Phosphor Energy Conversion Aluminum Support Photocathode CsI Needles Figure courtesy from Jonathan Tucker, Brooke Army Medical Center, SA, TX Kalpana M. Kanal, Ph.D., DABR 9
10 1. Input Phosphor Energy Conversion 60 kev X-Ray Aluminum Support Photocathode CsI Needles Figure courtesy from Jonathan Tucker, Brooke Army Medical Center, SA, TX Kalpana M. Kanal, Ph.D., DABR 10
11 1. Input Phosphor Energy Conversion Aluminum Support 3,000 light photons λ = ~ 420 nm Photocathode CsI Needles Figure courtesy from Jonathan Tucker, Brooke Army Medical Center, SA, TX Kalpana M. Kanal, Ph.D., DABR 11
12 1. Input Phosphor Energy Conversion Aluminum Support ~ 400 electrons Photocathode CsI Needles To Anode Figure courtesy from Jonathan Tucker, Brooke Army Medical Center, SA, TX Kalpana M. Kanal, Ph.D., DABR 12
13 1. Electron Optics Electrons are accelerated by an electric field Energy of each electron is substantially increased and this gives rise to electron gain Focusing is achieved using an electronic lens, which requires the input screen to be a curved surface, and this results in unavoidable pincushion distortion of the image c.f. Bushberg, et al. The Essential Physics of Medical Kalpana M. Kanal, Ph.D., DABR Imaging, 2 nd ed., p
14 1. Electron Optics The G1, G2, G3 electrodes along with the input screen and the anode near the output phosphor comprise the five-component electronic lens system of the II The electrons under the influence of the 25K to 35K V electric field, are accelerated and arrive at the anode with high velocity and considerable kinetic energy After penetrating the very thin anode, the energetic electrons strike the output phosphor c.f. Bushberg, et al. The Essential Physics of Medical Kalpana M. Kanal, Ph.D., DABR Imaging, 2 nd ed., p
15 1. The Output Phosphor The output phosphor is made of zinc cadmium sulfide Anode is very thin coating of aluminum on the vacuum side of the output phosphor, which is electrically conductive to carry away the electrons once they deposit their energy in the phosphor Each electron causes the emission of approximately 1000 light photons from the output phosphor 2.5 cm diameter output phosphor c.f. Bushberg, et al. The Essential Physics of Medical Kalpana M. Kanal, Ph.D., DABR Imaging, 2 nd ed., p
16 1. The Output Phosphor The reduction in image diameter leads to amplification (analogy: magnifying glass and sunlight) Minification gain of an II is simply the ratio of the area of the input phosphor to that of the output phosphor, e.g., 9 input phosphor, 1 output phosphor, area is square of the diameter ratio, minification gain is 81 c.f. Bushberg, et al. The Essential Physics of Medical Kalpana M. Kanal, Ph.D., DABR Imaging, 2 nd ed., p
17 1. The Output Phosphor The output phosphor is coated right onto the output window Some fraction of the light emitted by the output phosphor is reflected at the glass window Light bouncing around the output window is called veiling glare,, and can reduce image contrast c.f. Bushberg, et al. The Essential Physics of Medical Kalpana M. Kanal, Ph.D., DABR Imaging, 2 nd ed., p
18 1. Raphex 2001 Diagnostic Question D24-D28. D28. For the image intensifier shown, match the following: (answers may be used more than once) A. Light photons. B. X-ray X photons. C. Microwaves. D. Electrons. E. Infrared photons. D24. I represents D25. II represents D26. III represents D27. IV represents D28. V represents B B A D A Kalpana M. Kanal, Ph.D., DABR 18
19 1. Characteristics of Image Intensifier Performance Conversion Factor Conversion Factor = Light out of image intensifier (cd/m 2 ) Exposure rate into image intensifier (mr/sec) Defined as a measure of the gain of an image intensifier ratio of light output to exposure rate input 100 to 200 for new image intensifier Degrades over time, ultimately can lead to II replacement Kalpana M. Kanal, Ph.D., DABR 19
20 1. Characteristics of Image Intensifier Performance Brightness Gain BG = minification gain x electronic gain (flux gain) Minification gain = increase in image brightness that results from reduction in image size from the input phosphor to output phosphor or size (d i /d o ) 2, d i is input diameter which varies, d o is output diameter typically 2.5 cm For 30 cm (12 ) II, minification gain = 144 Kalpana M. Kanal, Ph.D., DABR 20
21 1. Characteristics of Image Intensifier Performance Brightness Gain BG = minification gain x electronic gain (flux gain) Electronic gain or flux gain is typically 50 The brightness gain therefore ranges from about 2,500 7,000 As the effective diameter of the input phosphor decreases (magnification increases), the brightness gain decreases Kalpana M. Kanal, Ph.D., DABR 21
22 1. Field of View/Magnification Modes FOV specifies the size of the input phosphor of the image intensifier ifier Different sizes: 23 cm (9 ), 30 cm (12 ), 35 cm (14 ), 40 cm (16 ) FOV Magnification is accomplished electronically using electronic focusing that projects part of the input layer onto the output phosphor Since brightness gain decreases in mag. mode, the x-ray x exposure rate is boosted. (12/9) 2 = 1.8, (12/7) 2 = 2.9 c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 ed., Kalpana M. Kanal, Ph.D., DABR p
23 1. Raphex 2000 Diagnostic Question D29. In fluoroscopy, switching from the 9 inch to the 6 inch field-of of-view (FOV) results in in the T.V. resolution and in the e patient entrance radiation exposure. A. An increase, a decrease B. A decrease, an increase C. An increase, an increase D. A decrease, a decrease E. An increase, no change Kalpana M. Kanal, Ph.D., DABR 23
24 1. Optical Coupling Parallel rays of light enter the optical chamber, are focused by lenses, and strike the video camera where an electronic image in produced A partially silvered mirror is used to shunt the light emitted by the image intensifier to an accessory port c.f. Bushberg, et al. The Essential Physics of Medical Kalpana M. Kanal, Ph.D., DABR Imaging, 2 nd ed., p
25 1. Video Cameras c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p Kalpana M. Kanal, Ph.D., DABR 25
26 1. Video Camera Analog video systems typically have 30 frames/sec operation, but they work in an interlaced fashion to reduce flicker, the perception of the e image flashing on and off The human eye-brain system can detect temporal fluctuations slower than about 47 images/sec, and therefore at 30 frames/sec flicker would d be perceptible With interlaced systems, each frame is composed of two fields and d each field is refreshed at a rate of 60 times per second, which is fast enough to avoid perception of flicker Kalpana M. Kanal, Ph.D., DABR 26
27 1. Lag Lag means that each new TV image actually contains residual image information from the last several frames Lag is good and bad Lag acts to smooth the quantum noise in the image, but can also cause motion blurring Kalpana M. Kanal, Ph.D., DABR 27
28 1. Video Resolution Spatial resolution of a video in the vertical direction (top to bottom) of the TV image is governed by the number of scan lines By convention, 525 lines are used in N. America for TV 490 lines usable In the early days of TV, a man named Kell determined that about 70% of theoretical video resolution is appreciated visually, and this psychophysical effect is now called the Kell factor 490 x 0.7 = 343 lines or 172 line pairs useful for resolution For 9 field, resolution = 172 lp/229 mm = 0.75 lp/mm 17 cm or 7 field, resolution is 1.0 lp/mm 12 cm or 5 field, resolution is 1.4 lp/mm Kalpana M. Kanal, Ph.D., DABR 28
29 1. Video Resolution The horizontal resolution is determined by how fast the video electronics ectronics can respond to changes in light intensity This is influenced by the camera, the cable, the monitor but the horizontal resolution is governed by the bandwidth of the system The time necessary to scan each video line (525 lines at 30 frame/sec) is 63 µsec 11 µsec required for horizontal retrace, 52 µsec available To achieve 172 cycles in 52 msec, the bandwidth required is 172 cycles/52 x 10-6 sec = 3.3 x 10 6 cycles/sec = 3.3 MHz Higher bandwidths are required for high-line video systems Kalpana M. Kanal, Ph.D., DABR 29
30 1. Raphex 2001 Diagnostic Question D29. The maximum resolution of fluoroscopy images displayed on television in the 6" field of view with a 1024 line system with a Kell factor of 0.7 is about lp/mm. A. 0.7 B. 1.2 C. 1.8 D. 2.4 E x 0.7 = 717 lines or 358 line pairs useful for resolution For 6 field, resolution = 358 lp/6 *25.4 mm = 2.35 lp/mm Kalpana M. Kanal, Ph.D., DABR 30
31 1. Take Home Points Fluoroscopy is a live imaging procedure Image Intensifier main component and consists of the input phosphor, hor, electronic lens system and output phosphor Input phosphor Cesium Iodide, converts x-rays x to light Photocathode converts light into electrons Output phosphor Zinc cadmium sulphide, converts electrons into light Artifacts pincushion distortion, veiling glare, lag Brightness gain = minification gain x electronic (flux) gain Several magnification modes available, typically exposure rate increases i with magnification Video camera produces the electronic image which we see on the TV T monitor Use interlaced scanning to avoid flicker Horizontal (determined by bandwidth) and vertical (determined by the number of scan lines) video resolution Kalpana M. Kanal, Ph.D., DABR 31
32 Flat Panel Digital Fluoroscopy Flat panel devices are thin film transistor (TFT) arrays that are rectangular in format and are used as x-ray x detectors CsI, a scintillator is used to convert the incident x-ray x beam into light TFT systems have a photodiode at each detector element which converts light energy to an electronic signal Flat panel detectors would replace the image intensifier, video camera, and other peripheral devices c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p Kalpana M. Kanal, Ph.D., DABR 32
33 Peripheral Equipment Photo-spot camera used to generate images on photographic film, 100-mm cut film or 105- mm roll film full resolution of the II system, hardly seen nowadays Digital photo-spot high resolution, slow-scan scan TV cameras in which the TV signal is digitized and stored in computer memory Or CCD cameras with or pixel formats near-instantaneous viewing of the image on a video monitor allows the fluoroscopist to put together a number of images to demonstrate the anatomy important to the diagnosis digital images can be printed on a laser imager Kalpana M. Kanal, Ph.D., DABR 33
34 Peripheral Equipment Spot-film devices attaches to the front of the II, and produces conventional radiographic screen-film images better resolution than images produced by II Cine-radiography cameras attaches to a port and can record a very rapid sequence of images s on 35- mm film used in cardiac studies, 30 frames/sec to 120 frames/sec or higher uses very short radiographic pulses digital cine are typically CCD-based cameras that produce a rapid sequence of digital images instead of film sequence Kalpana M. Kanal, Ph.D., DABR 34
35 Fluoroscopy Modes of Operation Continuous fluoroscopy continuously on x-ray x beam, ma or higher display at 30 frames/sec, 33 msec/frame acquisition time blurring present due to patient motion, acceptable 10 R/min is the maximum legal limit High dose rate fluoroscopy specially activated fluoroscopy 20 R/min is the maximum legal limit audible signal required to sound used for obese patients Kalpana M. Kanal, Ph.D., DABR 35
36 Fluoroscopy Modes of Operation Pulsed fluoro: series of short x-ray x pulses, 30 pulses at ~10 msec per pulse exposure time is shorter, reduces blurring from patient motion Can be used where object motion is high, e.g., positioning catheters ters in highly pulsatile vessels 15 frames/sec, 7.5 frames/sec also available Variable frame pulsed fluoroscopy is instrumental in reducing dose Ex., initially guiding the catheter up from the femoral artery to the aortic arch does not require high temporal resolution and 7.5 frames/sec c could potentially be used instead of 30 frames/sec 7.5 frames/sec instead of 30 frames/sec, dose savings of (7.5/30) ) 25% Kalpana M. Kanal, Ph.D., DABR 36
37 Frame Averaging Fluoroscopy systems provide excellent temporal resolution However, fluoroscopy images are relatively noisy, and in some applications it is beneficial to compromise temporal resolution for lower noise images This can be achieved by averaging a series of images or frames Real-time averaging in the computer memory for display Can cause noticeable image lag but noise in image is reduced as well Could also reduce dose in some circumstances c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p Kalpana M. Kanal, Ph.D., DABR 37
38 Last-Frame Hold Last-frame hold when the fluoroscopist takes his or her foot off the fluoroscopy pedal, rather than seeing a blank monitor, last-frame frame-hold enables the last live image to be shown continuously useful at training institutions no unnecessary radiation used on patient Kalpana M. Kanal, Ph.D., DABR 38
39 Road Mapping Road Mapping software-enhanced enhanced variant of the last-frame frame-hold feature side-by by-side video monitors, one shows captured image, the other live image In angiography, subtracted image can be overlayed over live image e to give the angiographer a vascular road map right on the fluoroscopy image is useful for advancing catheters through tortuous vessels Kalpana M. Kanal, Ph.D., DABR 39
40 Automatic Brightness Control The purpose of the automatic brightness control (ABC) is to keep the brightness of the image constant at monitor It does this by regulating the x-ray x exposure rate (control kvp, ma or both) Automatic brightness control triggers with changing patient size and field modes Kalpana M. Kanal, Ph.D., DABR 40
41 Automatic Brightness Control The top curve increases ma more rapidly than kv as a function of patient thickness, and preserves subject contrast at the expense of higher dose The bottom curve increases kv more rapidly than ma with increasing patient thickness, and results in lower dose, but lower contrast as well c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p Kalpana M. Kanal, Ph.D., DABR 41
42 Raphex 2003 Diagnostic Question D36. A 9-in. 9 multi-mode mode image intensifier (II) is switched to the 6-in. 6 mode. As a result, the image will be, and the automatic brightness control system (ABC) will the exposure to the II and the t patient. A. magnified, decrease B. magnified, increase C. minified, increase D. magnified, not change E. minified, decrease Kalpana M. Kanal, Ph.D., DABR 42
43 Raphex 2000 Diagnostic Question D30. Pulsed fluoroscopy at 15 fps is utilized primarily to.. A. Reduce motion blur. B. Increase kw rating. C. Increase kvp stabilization. D. Reduce focal spot sizes. E. Reduce patient dose. Kalpana M. Kanal, Ph.D., DABR 43
44 Image Quality Spatial Resolution A 2D image really has 3 dimensions: height, width, and gray scale Height and width are spatial and have units such as millimeters The classic notion of spatial resolution is the ability of an image system to distinctly depict two objects as they become smaller and closer together The closer together they are, with the image still showing them as separate objects, the better the spatial resolution At some point, the two objects become so close that they appear as one, and at this point, spatial resolution is lost Kalpana M. Kanal, Ph.D., DABR 44
45 Image Quality Spatial Resolution The spatial domain simply refers to the two spatial dimensions of an image, width (x-dimension) and length (y-dimension) Another useful way to express the resolution of an imaging system m is to make use of the spatial frequency domain F (line pairs/mm or cycles/mm) =1/2,, where is the size of the object (mm) Smaller objects (small ) ) correspond to higher spatial frequencies and larger objects (large ) ) correspond to lower spatial frequencies So, objects that are 0.36 mm correspond to 1.4 lp/mm 0.19 mm corresponds to 2.7 lp/mm 1 mm correspond to 0.5 lp/mm Kalpana M. Kanal, Ph.D., DABR 45
46 Image Quality Spatial Resolution Spatial frequency is just another way of thinking of object size A device used to measure the spatial resolution is the bar pattern c.f. Bushberg, et al. The Essential Physics of Kalpana M. Kanal, Ph.D., DABR Medical Imaging, 2 nd ed., p
47 Image Quality Spatial Resolution The modulation transfer function, MTF of an image system is a very complete description of the resolution properties of an imaging system The MTF illustrates the fraction (or percentage) of an object s contrast that is recorded by the imaging system, as a function of the size (i.e., spatial frequency) of the object The limiting spatial resolution is the size of the smallest object that an imaging system can resolve The limiting resolution of modern image intensifiers is between 4 and 5 cycles/mm c.f. Bushberg, et al. The Essential Physics of Medical cycles/mm Imaging, 2 nd ed., p Kalpana M. Kanal, Ph.D., DABR 47
48 Image Quality Contrast Resolution The ability to detect a low-contrast object on an image is highly related to how much noise (quantum noise and otherwise) there is in the image The ability to visualize low-contrast objects is the essence of contrast resolution. Better contrast resolution implies that more subtle objects can be routinely seen on the image The contrast resolution of fluoroscopy is low by comparison to radiography, r because the low exposure levels produce images with relatively low l signal- to-noise ratio (SNR) Kalpana M. Kanal, Ph.D., DABR 48
49 Image Quality Contrast Resolution Contrast resolution is increased when higher exposure rates are used, but the disadvantage is more radiation dose to the patient Fluoroscopic systems with different dose settings allow the user flexibility from patient to patient to adjust the compromise between contrast resolution and patient exposure Kalpana M. Kanal, Ph.D., DABR 49
50 Image Quality Temporal Resolution Fluoroscopy has excellent temporal resolution, that is over time Blurring in the time domain is typically called image lag Lag implies that a fraction of the image data from one frame carries ries over into the next frame Video cameras such as the vidicon demonstrate a fair amount of lagl Kalpana M. Kanal, Ph.D., DABR 50
51 Image Quality Temporal Resolution Lag in general is undesirable, beneficial for DSA Frame averaging improves contrast resolution at the expense of temporal t resolution With DSA and digital cine, cameras with low-lag lag performance (plumbicons or CCD cameras) are used to maintain temporal resolution Kalpana M. Kanal, Ph.D., DABR 51
52 Raphex 2003 Diagnostic Question D37. For fluoroscopy performed in the 6-in. 6 II mode and displayed on a 525 line TV monitor, spatial resolution is most limited by the.. A. Scatter from the patient B. Grid C. II tube D. Optical system E. TV system Kalpana M. Kanal, Ph.D., DABR 52
53 Fluoroscopy Suites Gastrointestinal Suites R and F room, large table that can be rotated from horizontal to vertical to put the patient in a head-down down or head-up position II above or under the table, spot film device usually there Remote Fluoroscopy Rooms Designed for remote operation by the radiologist Tube above table, II under table Reduce dose to the physician and no lead apron needed c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p Kalpana M. Kanal, Ph.D., DABR 53
54 Fluoroscopy Suites Peripheral Angiography Suites Table floats, allows patient to be moved from side to side and head to toe C-arm or U-arm U configuration 30 to 40 cm image intensifier used Power injectors are normally ceiling- or table-mounted Cardiology Catheterization Suite Similar to angiography suite, 23 cm II used to permit more tilt in cranial caudal direction Cine cameras used, biplane rooms common c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p Kalpana M. Kanal, Ph.D., DABR 54
55 Fluoroscopy Suites Biplane Angiographic Systems Two complete x-ray x tube/ii systems used, PA and Lateral Simultaneous acquisition of 2 views allows a reduction of the volume of contrast media injected in patient Portable Fluoroscopy- C Arms C-Arm devices with an x-ray x tube placed opposite from the II 18-cm (7-inch) and 23-cm (9-inch) and several other field sizes available Operating rooms and ICUs c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p Kalpana M. Kanal, Ph.D., DABR 55
56 Radiation Dose Patient Dose The maximum exposure rate permitted in the US is governed by the Code of Federal Regulations (CFR), and is overseen by the Center for Devices and Radiological Health (CDRH), a branch of the Food and Drug Administration (FDA) The maximum legal entrance exposure rate for normal fluoroscopy to the patient is 10 R/min For specially activated fluoroscopy, the maximum exposure rate allowable a is 20 R/min Kalpana M. Kanal, Ph.D., DABR 56
57 Radiation Dose Patient Dose Typical entrance exposure rates for fluoroscopic imaging are About 1 to 2 R/min for thin (10- cm) body parts 3 to 5 R/min for the average patient 8 to 10 R/min for the heavy patient Maximum dose at 120 kvp for most vendors c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2 nd ed., p Kalpana M. Kanal, Ph.D., DABR 57
58 Dose to Personnel Rule of Thumb: standing 1 m from the patient, the fluoroscopist receives from scattered radiation (on the outside of apron) approximately 1/1,000 of the exposure incident upon the patient The scatter field incident upon the radiologist while performing a fluoroscopic procedure is shown A radiologist of average height, 178 cm (5 10 ) is shown overlaid on the graph and key anatomic levels are indicated c.f. Bushberg, et al. The Essential Physics of Medical Kalpana M. Kanal, Ph.D., DABR Imaging, 2 nd ed., p
59 Dose to Personnel The dose rate as a function of height above the floor in the room is shown for 6 different distances D, representing the distance between the edge of the patient and the radiologist 80 kvp beam and 20 cm patient thickness assumed for calculation Roentgen-area product (RAP) or dose- area product (DAP) meters can be used to provide real-time estimate of the amount of radiation the patient has received c.f. Bushberg, et al. The Essential Physics of Medical Kalpana M. Kanal, Ph.D., DABR Imaging, 2 nd ed., p
60 Raphex 2002 Diagnostic Question D36. The fluoroscopic operating factors displayed on a monitor are kvp and 10 ma. Which of the following is true? A. The skin entrance dose is unusually low. B. The five-minute timer is broken. C. The skin entrance dose is extremely high. D. The display must be wrong. E. The anti-scatter grid is not in the beam. Kalpana M. Kanal, Ph.D., DABR 60
BASICS OF FLUOROSCOPY
Medical Physics Residents Training Program BASICS OF FLUOROSCOPY Dr. Khalid Alyousef, PhD Department of Medical Imaging King Abdulaziz Medical City- Riyadh Edison examining the hand of Clarence Dally with
More information10/15/2012 SECTION III - CHAPTER 6 DIGITAL FLUOROSCOPY RADT 3463 COMPUTERIZED IMAGING
RADT 3463 - COMPUTERIZED IMAGING Section III: Chapter 6 RADT 3463 Computerized Imaging 1 SECTION III - CHAPTER 6 DIGITAL FLUOROSCOPY RADT 3463 COMPUTERIZED IMAGING Section III: Chapter 6 RADT 3463 Computerized
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationI. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA
FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE
More informationX-RAY FLUOROSCOPY IMAGING SYSTEMS. Dr Slavik Tabakov. Luminescence: Dept. Medical Eng. & Physics King s College London
X-RAY FLUOROSCOPY IMAGING SYSTEMS Dr Slavik Tabakov OBJECTIVES - Image Intensifier construction - Input window - Accelerating and focusing electrodes - Output window - Conversion factor - II characteristics
More informationDo you have any other questions? Please call us at (Toll Free) or , or
INSTRUCTIONS Read the appropriate course/ textbook. This is an open book test. A score of 75% or higher is needed to receive CE credit. You will have a maximum of three attempts to pass this course. Please
More information10/3/2012. Study Harder
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen
More information10/26/2015. Study Harder
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen
More informationX-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto
X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced
More informationSeminar 8. Radiology S8 1
Seminar 8 Radiology Medical imaging. X-ray image formation. Energizing and controlling the X-ray tube. Image detectors. The acquisition of analog and digital images. Digital image processing. Selected
More informationX-ray Tube and Generator Basic principles and construction
X-ray Tube and Generator Basic principles and construction Dr Slavik Tabakov - Production of X-rays and Patient Dose OBJECTIVES - X-ray tube construction - Anode - types, efficiency - Classical X-ray generator
More informationDIGITAL IMAGE PROCESSING IN X-RAY IMAGING
DIGITAL IMAGE PROCESSING IN X-RAY IMAGING Shalini Kumari 1, Bachan Prasad 2,Aliya Nasim 3 Department of Electronics And Communication Engineering R.V.S College of Engineering & Technology, Jamshedpur,
More informationTESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist
TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist Topics Image Uniformity and Artifacts Image Quality - Detail
More informationAmorphous Selenium Direct Radiography for Industrial Imaging
DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial
More informationAN ABSTRACT OF THE THESIS OF. W. Scott Helms for the degree of Master of Science in Radiation Health Physics
AN ABSTRACT OF THE THESIS OF W. Scott Helms for the degree of Master of Science in Radiation Health Physics presented on November 24, 2014 Title: A Quantitative Comparison of Cardiovascular Imaging Systems
More informationOverview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging
The Physics of Radiation Oncology X-ray Imaging Charles E. Willis, Ph.D. DABR Associate Professor Department of Imaging Physics The University of Texas M.D. Anderson Cancer Center Houston, Texas Overview
More informationPhotomultiplier Tube
Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or
More informationRadiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893
Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -
More informationUnit thickness. Unit area. σ = NΔX = ΔI / I 0
Unit thickness I 0 ΔI I σ = ΔI I 0 NΔX = ΔI / I 0 NΔX Unit area Δx Average probability of reaction with atom for the incident photons at unit area with the thickness of Delta-X Atom number at unit area
More informationX-RAYS - NO UNAUTHORISED ENTRY
Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are
More informationINTRODUCTION TO FLEXIBLE BRONCHOSCOPY. Fluoroscopy Synopsis HENRI G COLT MD SECOND EDITION THE BRONCHOSCOPY EDUCATION PROJECT SERIES
SECOND EDITION INTRODUCTION TO FLEXIBLE BRONCHOSCOPY Fluoroscopy Synopsis HENRI G COLT MD With contributions from Dr. S. Murgu THE BRONCHOSCOPY EDUCATION PROJECT SERIES FLUOROSCOPY SYNOPSIS The purpose
More informationSECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING
RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 1 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT RADT 3463 COMPUTERIZED IMAGING Section I: Chapter
More informationBeam-Restricting Devices
Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the
More informationX-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah
X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,
More informationDigital Detector Array Image Quality for Various GOS Scintillators
Digital Detector Array Image Quality for Various GOS Scintillators More info about this article: http://www.ndt.net/?id=22768 Brian S. White 1, Mark E. Shafer 2, William H. Russel 3, Eric Fallet 4, Jacques
More informationTruly flexible to meet your clinical needs
Truly flexible to meet your clinical needs 2 Adapting to meet your needs Flexible Fast and responsive Excellent image quality Designed with ergonomic efficiency Equipped with dose management tools 3 Three
More informationCOMPUTED TOMOGRAPHY 1
COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the
More informationDose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp
Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division
More informationCurrent technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016
Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) 2/ Overview Digital
More informationMedical Imaging: A Look inside. Medical Imaging. Medical Imaging. Visible Human Project
Medical Imaging: A Look inside Medical Imaging Allows physicians to see what had previously been unseeable: internal organs and tissues, bones, a beating heart, etc. Allows physicians to: detect brain
More informationLECTURE 1 The Radiographic Image
LECTURE 1 The Radiographic Image Prepared by:- KAMARUL AMIN ABDULLAH @ ABU BAKAR UiTM Faculty of Health Sciences Medical Imaging Department 11/23/2011 KAMARUL AMIN (C) 1 Lesson Objectives At the end of
More informationMultiple Choice Identify the letter of the choice that best completes the statement or answers the question.
RA110 test 3 Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. An object 35 cm in width is radiographed at 100 cm SID and at a 50 cm SOD. What
More informationAdvanced digital image processing for clinical excellence in fluoroscopy
Dynamic UNIQUE Digital fluoroscopy solutions Dynamic UNIQUE Advanced digital image processing for clinical excellence in fluoroscopy André Gooßen, PhD, Image Processing Specialist Dörte Hilcken, Clinical
More informationPLD5600A High Frequency Digital Gastrointestinal &DR System(630mA)
PLD5600A High Frequency Digital Gastrointestinal &DR System(630mA) Application: Full support perspective, gastrointestinal spot film, GI (barium meal, barium enema), orthopedic photography, pediatrics
More informationBCA 9SK. 60 Hz on request Absorbed current on Stationary Vac and Vac in fluoro mode
ELECTRICAL CHARACTERISTICS Voltage 230 Vac ±10% monophase standard 105 / 115 / 125 / 220 / 240 Vac ±10% monophase on request Frequency 50 Hz standard 60 Hz on request Absorbed current on Stationary 4.5
More informationBCA 9RK. 60 Hz on request Absorbed current on Stationary Vac and Vac in fluoro mode
ELECTRICAL CHARACTERISTICS RADIOLOGICAL DATA MONOBLOC Voltage 230 Vac ±10% monophase standard 105 / 115 / 125 / 220 / 240 Vac ±10% monophase on request Frequency 50 Hz standard 60 Hz on request Absorbed
More informationARCO Rk.5 Digital Mobile C-Arm
Product Data ARCO Rk.5 Digital Mobile C-Arm Product Data 1 Easy to move INTRODUCTION Designed with the latest technology throughout, the ARCO Rk.5 set new standards for the excellence image quality ease
More informationNuclear Associates
Nuclear Associates 76-700 Digital Subtraction Angiography Phantom Users Manual March 2005 Manual No. 76-700-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names
More informationIMAGE SENSOR SOLUTIONS. KAC-96-1/5" Lens Kit. KODAK KAC-96-1/5" Lens Kit. for use with the KODAK CMOS Image Sensors. November 2004 Revision 2
KODAK for use with the KODAK CMOS Image Sensors November 2004 Revision 2 1.1 Introduction Choosing the right lens is a critical aspect of designing an imaging system. Typically the trade off between image
More informationDigital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality
Digital radiography: Digital radiography is set to become the most common form of processing radiographic images in the next 10 years. This is due to a number of practical and image quality issues. Practical
More informationSYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology
CODE: RADT 156 INSTITUTE: Health Science TITLE: Equipment Operation I DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of equipment operation and maintenance of radiographic
More informationCHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC
CHAPTER 6 QC Test For Fluoroscopic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done. Identify QC
More informationMulti-Access Biplane Lab
Multi-Access Biplane Lab Advanced technolo gies deliver optimized biplane imaging Designed in concert with leading physicians, the Infinix VF-i/BP provides advanced, versatile patient access to meet the
More informationX-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope
X-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope Kenichi Ikeda 1, Hideyuki Kotaki 1 ' 2 and Kazuhisa Nakajima 1 ' 2 ' 3 1 Graduate University for Advanced
More informationMXHF-1500RF is controlled by Digital key panel console that displays KV, ma and mas with APR menu programmed.
R/F TV X-RAY SYSTEM DIAGNOSTIC RADIOGRAPHIC FLUOROSCOPIC TV SYSTEM MXHF-1500RF SYSTEM OUTLINE Product Data No. 041021-01 MXHF-1500RF is controlled by Digital key panel console that displays KV, ma and
More informationAcquisition, Processing and Display
Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image
More informationThe X-ray circuit: part II
The X-ray circuit: part II By Dr. Mohsen Dashti 357 Radiologic Processing & Procedure Lecture notes #2 Key issues Types of x-ray equipment. Power for x-ray generator. A basic x-ray circuit. Generators.
More informationFluoroscopy & Tomography. Rhodes
Fluoroscopy & Tomography Rhodes 10-526-194 1 Historical Development Dynamic examination Active Diagnosis Domain of radiologist Fluoroscope Invented by Edison in 1896 2 Fluoroscopic Imaging Chain Components
More informationPD233: Design of Biomedical Devices and Systems
PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/
More informationSECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS
RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT
More informationRadiographic sensitivity improved by optimized high resolution X -ray detector design.
DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Radiographic sensitivity improved by optimized high resolution X -ray detector
More informationInterventional Radiological Equipment selection and installation
Interventional Radiological Equipment selection and installation Renato Padovani ICTP Learning objectives To understand the main components of an interventional radiology equipment To understand the relevance
More informationDigital Radiography : Flat Panel
Digital Radiography : Flat Panel Flat panels performances & operation How does it work? - what is a sensor? - ideal sensor Flat panels limits and solutions - offset calibration - gain calibration - non
More informationR&F X-ray systsem. Savings With Every Exposure
R&F X-ray systsem Savings With Every Exposure Savings with every exposure Versatility meets value Dependable performance User-friendly interface Expandable imaging capabilities Excellent image quality
More informationEnhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)
Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular
More informationChromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC
Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC R. Bellazzini a,b, G. Spandre a*, A. Brez a, M. Minuti a, M. Pinchera a and P. Mozzo b a INFN Pisa
More informationDistributed by. Ecotron Anyview Series. The Intelligent C-Arm System
Distributed by Ecotron Anyview Series The Intelligent C-Arm System The Ecotron Anyview Series is a new standard for C-arm imaging Excellent radiographic imaging at an economical price Advanced X-ray Generator
More informationCHAPTER 10. DIGITAL SUBTRACTION ANGIOGRAPHY (DSA) 10.1 Introduction
PHYSICS OF MEDICAL X-RAY IMAGING (1)Chapter 10 CHAPTER 10. DIGITAL SUBTRACTION ANGIOGRAPHY (DSA) 10.1 Introduction In this chapter we will discuss a technique that is known as digital subtraction angiography
More informationDigital Radiography X-Ray System. X Twin with X Mobil Roesys GmbH [rshsmi] 1/11
Digital Radiography X-Ray System 2017 Roesys GmbH [rshsmi] 1/11 1 General specifications The system is intended to be installed with floor mounted components only. It enables spacesaving installation without
More informationDevelopment of a fast EUV movie camera for Caltech spheromak jet experiments
P1.029 Development of a fast EUV movie camera for Caltech spheromak jet experiments K. B. Chai and P. M. Bellan ` California Institute of Technology kbchai@caltech.edu Caltech Spheromak gun 2 Target: study
More informationPERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY
12 th A-PCNDT 2006 Asia-Pacific Conference on NDT, 5 th 10 th Nov 2006, Auckland, New Zealand PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY Rajashekar
More informationSONIALVISION G4 Multi-purpose Digital R/F System C506-E075
SONIALVISION G4 Multi-purpose Digital R/F System C506-E075 Selecting the best for every examination environment BEST in CLASS Multi-purpose Digital R/F System 2 With the Sonialvision G4, Shimadzu offers
More informationSONIALVISION G4 Multi-purpose Digital R/F System C506-E075
SONIALVISION G4 Multi-purpose Digital R/F System C506-E075 Selecting the best for every examination environment BEST in CLASS Multi-purpose Digital R/F System With the Sonialvision G4, Shimadzu offers
More informationComponents of Optical Instruments
Components of Optical Instruments General Design of Optical Instruments Sources of Radiation Wavelength Selectors (Filters, Monochromators, Interferometers) Sample Containers Radiation Transducers (Detectors)
More informationEssentials of Digital Imaging
Essentials of Digital Imaging Module 1 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 1 Fundamentals 1. ASRT Animation 2. Welcome Welcome to Essentials of Digital Imaging
More informationVisualization of sources of scattered radiation from x-ray equipment used for interventional radiology
Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology Poster No.: C-1190 Congress: ECR 2011 Type: Scientific Exhibit Authors: K. Chida, T. Takahashi, D.
More informationPhilips XPER FD10C R7.0.4
Philips XPER FD10C R7.0.4 Reconditioned 2005 System- Upgraded to R7 in Oct 2010 The Allura Xper FD10 (Ceiling) single-plane cardiovascular system is comprised of a ceiling mounted C-arm stand and digital
More informationVeterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS
Veterinary Science Preparatory Training for the Veterinary Assistant Floron C. Faries, Jr., DVM, MS Radiology Floron C. Faries, Jr., DVM, MS Objectives Determine the appropriate machine settings for making
More informationSONIALVISION G4 Multi-purpose Digital R/F System
C506-E075A SONIALVISION G4 Multi-purpose Digital R/F System Founded in 1875, Shimadzu Corporation, a leader in the development of advanced technologies, has a distinguished history of innovation built
More informationFlat-panel detectors: how much better are they?
Pediatr Radiol (2006) 36 (Suppl 2): 173 181 DOI 10.1007/s00247-006-0208-0 ALARA J. Anthony Seibert Flat-panel detectors: how much better are they? Published online: 22 July 2006 # Springer-Verlag 2006
More information3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development
Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory
More informationV SALAI SELVAM, AP & HOD, ECE, Sriram Engg. College, Perumalpattu 1 MEDICAL ELECTRONICS UNIT IV
V SALAI SELVAM, AP & HOD, ECE, Sriram Engg. College, Perumalpattu 1 MEDICAL ELECTRONICS UNIT IV Ionizing and non-ionizing radiations: The radiation that ionizes the gases through which it travels is known
More informationBasis of Computed Radiography & PACS
Basis of Computed Radiography & PACS Slavik Tabakov Computed Radiography (CR) refers to new types of X-ray detectors (i.e. replaces the X-ray Film) The CR output media is a digital image, which can be
More information1-1. GENERAL 1-2. DISCOVERY OF X-RAYS
1-1. GENERAL Radiography is a highly technical field, indispensable to the modern dental practice, but presenting many potential hazards. The dental radiographic specialist must be thoroughly familiar
More informationC-ARM FLUOROSCOPIC AND SPOT-FILM SYSTEMS
PART X C-ARM FLUOROSCOPIC AND SPOT-FILM SYSTEMS FORM FD 3260 REPRINTED APRIL 2000 ROUTINE COMPLIANCE TESTING C-ARM FLUOROSCOPES (Test Procedure CFA - Use Form FDA 3260) 1.0 GENERAL GUIDANCE 1.1 This procedure
More informationPROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, Edited by
X - R A Y M I C R O S C O P Y A N D M I C R O R A D I O G R A P H Y PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, 1956 Edited by V. E. COSSLETT Cavendish Laboratory, University
More informationRadionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM
Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM Preamplifiers and amplifiers The current from PMT must be further amplified before it can be processed and counted (the number of electrons yielded
More informationOptical basics for machine vision systems. Lars Fermum Chief instructor STEMMER IMAGING GmbH
Optical basics for machine vision systems Lars Fermum Chief instructor STEMMER IMAGING GmbH www.stemmer-imaging.de AN INTERNATIONAL CONCEPT STEMMER IMAGING customers in UK Germany France Switzerland Sweden
More informationMammography: Physics of Imaging
Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness
More informationDetectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014
Detectors for microscopy - CCDs, APDs and PMTs Antonia Göhler Nov 2014 Detectors/Sensors in general are devices that detect events or changes in quantities (intensities) and provide a corresponding output,
More informationFeatures and Weaknesses of Phantoms for CR/DR System Testing
Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal
More informationPerformance of Image Intensifiers in Radiographic Systems
DOE/NV/11718--396 LA-UR-00-211 Performance of Image Intensifiers in Radiographic Systems Stuart A. Baker* a, Nicholas S. P. King b, Wilfred Lewis a, Stephen S. Lutz c, Dane V. Morgan a, Tim Schaefer a,
More informationDigital Imaging Considerations Computed Radiography
Digital Imaging Considerations Digital Radiography Computed Radiography o Cassette based Direct or Indirect Digital Radiography o Cassetteless Computed Radiography 1 CR Image Acquisition Most like conventional
More informationCOPYRIGHT 2002 by Srinivasan Vedantham ALL RIGHTS RESERVED. Use or inclusion of any portion of this document in another work intended for
COPYRIGHT 00 by Srinivasan Vedantham ALL RIGHTS RESERVED Use or inclusion of any portion of this document in another work intended for commercial use will require permission from the copyright owner. ii
More informationProduct data Remote Controlled Table
Product data Remote Controlled Table Product data Remote Controlled Table 2 BRT-20 Remote controlled table has been designed as basic equipment, compact and extremely simplified. MAIN APPLICATION Gastroenterology
More informationDigital radiography. bucky table and wall stand as a dual detector or wireless system. Amadeo DR Systems
Amadeo DR Systems with dicom PACS DX-R Software X-ray Systems for the Future Digital radiography with Amadeo R-DR including bucky table and wall stand as a dual detector or wireless system High contrast
More informationChina Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW Digital RF System - HF51-5
China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW Digital RF System - HF51-5 #3, No.9, Jiuxianqiaodong Road, Chaoyang District, Beijing 100015, P.R. China E-mail: international@wandong.com.cn
More informationSarah Hughes, MS, DABR Radiation Safety Officer
Sarah Hughes, MS, DABR Radiation Safety Officer 502-852-6146 sarah.hughes@louisville.edu Mo my back is burnin!!! I got it MAG the cine! Sumthin s not right. Where s his heart? Fluoroscopy http://dccwww.bumc.bu.edu/fluoroscopy/def
More informationCPSC 4040/6040 Computer Graphics Images. Joshua Levine
CPSC 4040/6040 Computer Graphics Images Joshua Levine levinej@clemson.edu Lecture 04 Displays and Optics Sept. 1, 2015 Slide Credits: Kenny A. Hunt Don House Torsten Möller Hanspeter Pfister Agenda Open
More informationNuclear Associates
Nuclear Associates 07-706 Patient Phantom/Penetrometer System Users Manual March 2005 Manual No. 07-706-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are
More informationDigital Radiography System
www.drgem.co.kr GXR-SD SERIES GXR-SD SERIES Digital Radiography System GXR-SD Series www.drgem.co.kr High Performance and Reliability The Optimal System for Digital Solution The GXR-SD Series is a diagnostic
More informationSYLLABUS. 1. Identification of Subject:
SYLLABUS Date/ Revision : 30 January 2017/1 Faculty : Life Sciences Approval : Dean, Faculty of Life Sciences SUBJECT : Biophysics 1. Identification of Subject: Name of Subject : Biophysics Code of Subject
More informationSTUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA
STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA RADT 2913 COMPREHENSIVE REVIEW 1 The CR cassette is backed by aluminum that: A. reflects x-rays B. absorbs x-rays C. captures the image D. transmits x-rays
More informationX-rays in medical diagnostics
X-rays in medical diagnostics S.Dolanski Babić 2017/18. History W.C.Röntgen (1845-1923) discovered a new type of radiation Nature, Jan. 23. 1896.; Science, Feb.14. 1896. X- rays: Induced the ionization
More informationC506-E064. Full digital system. Printed in Japan A-NS
C506-E064 Full digital system Printed in Japan 6295-08807-30A-NS Full digital system Highest Image Quality in Its Class Comprehensive Full-Digital System FLEXAVISION is a full-digital R/F system equipped
More informationX-ray light valve (XLV): a novel detectors technology for digital mammography*
X-ray light valve (XLV): a novel detectors technology for digital mammography* Sorin Marcovici, Vlad Sukhovatkin, Peter Oakham XLV Diagnostics Inc., Thunder Bay, ON P7A 7T1, Canada ABSTRACT A novel method,
More informationX-ray detectors in healthcare and their applications
X-ray detectors in healthcare and their applications Pixel 2012, Inawashiro September 4th, 2012 Martin Spahn, PhD Clinical applications of X-ray imaging Current X-ray detector technology (case study radiography
More informationSAMSUNG SMART DIGITAL RADIOGRAPHY. XGEO GU60A series REDEFINING ERGONOMICS
SAMSUNG SMART DIGITAL RADIOGRAPHY XGEO GU60A series REDEFINING ERGONOMICS U-arm Positioner (Fully Automated) The XGEO GU60A/60A-65 is a universal, fully motorized system. Its unique U-arm rotates +120
More informationsiemens.com/luminos-fusion Luminos Fusion The 2-in-1 system that fits your needs and fits your budget
siemens.com/luminos-fusion Luminos Fusion The 2-in-1 system that fits your needs and fits your budget Luminos Fusion The 2-in-1 system that fits your needs and fits your budget Luminos Fusion provides
More information