Image Quality in Digital Mammography: Image Acquisition

Size: px
Start display at page:

Download "Image Quality in Digital Mammography: Image Acquisition"

Transcription

1 Image Quality in Digital Mammography: Image Acquisition Mark B. Williams, PhD a, Martin J. Yaffe, PhD b, Andrew D.A. Maidment, PhD c, Melissa C. Martin, MS d, J. Anthony Seibert, PhD e, Etta D. Pisano, MD f This paper on digital mammography image acquisition is 1 of 3 papers written as part of an intersociety effort to establish image quality standards for digital mammography. The information included in this paper is intended to support the development of an ACR guideline on image quality for digital mammography. The topics of the other 2 papers are digital mammography image display and digital mammography image storage, transmission, and retrieval. The societies represented in compiling this document were the Radiological Society of North America, the ACR, the American Association of Physicists in Medicine, and the Society for Computer Applications in Radiology. These papers describe in detail what is known to improve image quality for digital mammography and make recommendations about how digital mammography should be performed to optimize the visualization of breast cancers. Through the publication of these papers, the ACR is seeking input from industry, radiologists, and other interested parties on their contents so that the final ACR guideline for digital mammography will represent the consensus of the broader community interested in these topics. Key Words: Digital mammography, image acquisition, image quality J Am Coll Radiol 2006;3: Copyright 2006 American College of Radiology INTRODUCTION AND OVERVIEW OF MAMMOGRAPHIC IMAGE QUALITY Theoretical Quality of the Acquired Radiographic Image Task-Dependent Image Quality. The analysis of image quality has meaning only in the context of a particular imaging task [1]. For example, a mammographic imaging system must permit the detection and characterization of small microcalcifications (requiring very high spatial resolution), whereas a positron emission tomographic scanner must be able to identify subtle differences in radiopharmaceutical concentration (requiring high contrast transfer ability but at lower spatial resolution). Therefore, to be useful in assessing image quality, any figures of merit must take into account the properties of the specific imaging task. Accordingly, this paper on a University of Virginia, Charlottesville, Va. b Sunnybrook & Women s College Health Sciences Centre, University of Toronto, Toronto, Ont, Canada. c Hospital of the University of Pennsylvania, Philadelphia, Pa. d Therapy Physics Inc., Bellflower, Calif. e University of California, Davis, Medical Center, Sacramento, Calif. f University of North Carolina at Chapel Hill, Chapel Hill, NC. Corresponding author and reprints: Mark B. Williams, PhD, University of Virginia, Department of Radiology, PO Box , Charlottesville, VA ; mbwilliams@virginia.edu American College of Radiology /06/$32.00 DOI /j.jacr digital mammography image acquisition, together with its companion papers on digital mammography display and digital mammography storage, transmission, and retrieval, was developed with reference to the specific imaging tasks required by mammography, using the information available in the peer-reviewed medical literature. Ultimately, image quality is affected by both image acquisition and image display. Thus, particularly in digital systems, in which these functions are clearly separable, the analysis of image quality can be performed in 2 stages. In the first stage, the direct laboratory measurement of the physical properties of an image acquisition system itself (ie, its spatial resolution, noise, and sensitivity) can be used to predict system performance before an image is presented to an observer. The observer can be a human or a mathematical construct, one example of which is the ideal observer model. The ideal observer makes optimal use of both the signal and the noise contained in an acquired image, thus representing a quantitative measurement of the highest possible performance level. The measured physical properties are combined mathematically with a description of the particular imaging task to yield a figure of merit (such as the ideal observer signal-to-noise ratio [SNR I ]) that indicates how well positive (abnormal) and negative (normal) results can be differentiated [1]. The primary drawbacks to the ideal observer analysis are that (1) it is difficult to specify complex imaging tasks, including, for example, realistic 589

2 590 Journal of the American College of Radiology/ Vol. 3 No. 8 August 2006 background structure; (2) the results may not represent the performance of real human observers; and (3) accurate measurements of the physical properties of an image acquisition system are often difficult to make. In the second stage of analysis, images are presented to human observers in the context of a receiver operating characteristic (ROC) study, in which diagnostic sensitivity and specificity can be determined over a range of levels of observer aggressiveness (ie, how willing they are to accept false-positive results to avoid false-negative results) [2,3]. The major strengths of ROC studies are that they include all aspects of an imaging system, including the human observer, and they use a realistic and meaningful imaging task. The main drawbacks of ROC studies using human observers are that they are extremely resource and time consuming. A more efficient approach, therefore, is to first analyze an acquisition system using measured physical performance metrics and a mathematical observer model and to then use the results to identify promising sets of system configuration parameters to be tested in human observer performance studies. Mammography as an Example of the Decision- Making Paradigm. The presence or absence of a lesion in a mammogram is an example of a 2-class detection process. Simply put, a radiologist must make a decision as to whether a suspicious lesion is present (ie, the case is positive) or not (ie, the case is negative). According to statistical decision theory, each image can be considered to come from either a set of truly positive cases or a set of truly negative cases. The ability of an imaging system to differentiate between data taken from the 2 populations can be quantified by examining the results produced by a decision maker who, when given data from the system (ie, an image), is asked to decide from which of the 2 populations the data arose. The decision maker uses some type of decision variable based on 1 or more image features (eg, the presence of specific patterns in the image characteristic of masses, calcifications, or architectural distortion) to classify each image. If decision variable values are generated for each of a large number of images, a histogram of the resulting values forms 2 distributions, as shown in Figure 1, 1 corresponding to the normal (negative) population and 1 to the abnormal (positive) population. The horizontal axis represents the variable or variables used to make the decision. As shown in the figure, these distributions typically have some overlap. Case-to-case variations in lesion shape and size, noise introduced during the image acquisition process, and interobserver and intraobserver variability all serve to broaden the negative and positive distributions along the decision axis and result in overlap of the 2. Each observer must establish a decision threshold (shown as a vertical Fig 1. Schematic diagram showing the overlap between the decision variable histograms obtained by imaging cases drawn from both truly positive and truly negative populations. The fraction of truly negative cases with decision variable values greater than the decision threshold is the false-positive fraction (FPF). The fraction of truly positive cases with decision variable values greater than the decision threshold is the true-positive fraction (TPF). line in Figure 1), below which an image is considered negative and above which it is considered positive. The location of this threshold along the range of possible values of the decision variable(s), along with the widths and separations of the 2 populations, determines the false-positive fraction and true-positive fraction. As shown in the figure, the false-positive fraction and truepositive fraction are the fractions of truly negative and truly positive cases, respectively, determined to be positive. Measurement of the false-positive fraction and truepositive fraction as the decision threshold is systematically varied as the basis of the ROC methodology, which is described in more detail later in this paper. Given the distributions of the decision variable for the 2 hypotheses (Figure 1), it is possible to define an effective SNR as figure of merit describing the ability of the combined imaging system and decision maker to differentiate between negative and positive cases. If the signal is defined as the difference between the means of the distributions and the noise is defined as the average of their standard deviations, then the square of the SNR is SNR 2 L 2 L ( ), where L is the decision variable, denotes averaging, is the standard deviation of a distribution, and the subscripts 1 and 2 refer to normal and abnormal, respectively. This definition of the SNR is general and holds for all types of decision makers (observers) and all assumptions about the data (eg, how well the lesion shape and nature of the background are known). However, for some specific types of observers and specific assumptions

3 Williams et al/digital Mammography Image Acquisition 591 about the nature of the data and image noise, the SNR can be rewritten in terms of measurable quantities. One of the most useful observer types is the ideal Bayesian observer. The ideal observer SNR, SNR I, is a direct measure of the degree of overlap between the 2 probability distributions in the circumstance in which they have not been broadened by variability introduced during the decision-making process (ie, reader variability). The measurement of SNR I is thus the first stage of the quantitative assessment of image quality, namely, the assessment of the acquired data. The Ideal Observer Formalism for Assessment of the Quality of the Acquired Data. The ideal observer calculates the likelihood (probability), given the image data, that a patient is truly a member of the negative population and, similarly, the likelihood that she is truly a member of the positive population. The ratio of these 2 probabilities (ie, the likelihood ratio) is used as a decision variable. A given image is considered normal if the decision variable is less than a predetermined threshold value and abnormal if it is greater. The ideal aspect of this observer is that it decides in favor of the hypothesis that minimizes the probability of making an incorrect judgment [1]. The ideal observer is able to first identify spatial correlations in the image noise (eg, to recognize detector artifact noise in the digital mammogram) and to effectively remove it (in other words, to remove the spatial correlation or whiten the noise). In situations in which the signal shape (ie, lesion shape) is known, the ideal observer then uses the known shape as a filter to identify areas in the image that most closely match the filter. This strategy, therefore, is known as the prewhitened matched filter. It can be shown [1,4] that for linear, shift-invariant imaging systems, the square of the ideal observer SNR can be written as SNR I 2 K 2 f( ) 2 MFT 2 ( ) W n ( )d, where the quantities f( ), MTF( ), and W n ( ) are functions of the spatial frequency variable. For 2-D images, is actually 2-D and can be considered in terms of its components x and y. f( ) is the Fourier transform of the difference in signals under the 2 competing hypotheses (eg, lesion present vs lesion absent for the problem of the detection of disease or type 1 lesion present vs type 2 lesion present for the problem of characterization), MTF( ) is the modulation transfer function of the image acquisition system, and W n ( ) is the Wiener, or noise power, spectrum. The constant K is a scaling factor relating units at the output of the imaging system (eg, analogto-digital units or optical density) to units at its input (x-ray photons per unit area). The MTF describes the spatial resolution of the imaging system in terms of its ability to transfer signal from its input to output as a function of the spatial frequency of the signal. W n ( ) is the Fourier decomposition of the image variance and describes the spatial frequency dependence of the total noise, including system noise and x-ray photon noise. The quantities K, MTF( ), and W n ( ) can be determined from laboratory measurements made under a particular set of operating conditions (eg, particular x-ray beam quality and exposure settings) using simple phantoms [5,6]. They are often combined to form a quantity called the noise equivalent quanta (NEQ): NEQ( ) K 2 MTF 2 ( ) W n ( ). The NEQ is a direct measurement of the quality of the acquired image and has units of input quanta per unit area. It can be interpreted as the number of quanta per area at the input of a perfect imaging system that would be necessary to achieve the measured image quality. The quantity f( ) depends on the particular imaging task (eg, mass or microcalcification) and ensures that the derived SNR I is task specific. Thus, SNR I can also be written SNR I 2 f( ) 2 NEQ( )d. In digital mammography, the specific imaging task is accurate depiction of mass morphology, microcalcification morphology, architectural distortion, and left-right breast asymmetry, all in the presence of normal structure background [7]. So, for example, f( ) could be the Fourier transform of the spatial distribution of a mass (ie, the 2 hypotheses are that a mass is or is not present). The important physical performance characteristics for highquality mammographic image acquisition are high MTF, that the system itself not contribute appreciably to W n (ie, that the image noise is primarily x-ray quantum noise), a large dynamic range, and good contrast transfer. The latter property refers to the maximization of the subject contrast (the contrast in x-ray fluence incident at various points on the detector surface, produced by differences in attenuation along paths through the breast) by the use of optimal x-ray techniques, combined with high detector absorption efficiency (quantum detection efficiency). ROC Evaluation. The ROC curve is a graph describing, for a particular interpreter or averaged over multiple interpreters, their sensitivity in detecting breast cancer compared with 1 specificity (ie, their true-positive fraction vs their false-positive fraction). Thus, an important advantage of the ROC method is that the outcome variables are absolutely appropriate, providing an assessment of the accuracy of determining whether or not cancer is present. Two hypothetical ROC curves are illustrated in Figure 2. All ROC curves contain the points (0,0) (perfect specificity but zero sensitivity by calling all

4 592 Journal of the American College of Radiology/ Vol. 3 No. 8 August 2006 Fig 2. Hypothetical receiver operating characteristic curves. For any given sensitivity (true-positive fraction), the curve gives the false-positive fraction that would result for that system (averaged over many readers). Moving along a curve indicates reading with different levels of aggressiveness. Curve B represents a better system than curve A because at any falsepositive fraction, the sensitivity is higher than for curve A. examinations negative) and (1,1) (perfect sensitivity but zero specificity by calling all examinations positive). In between, the curve reflects the trade-off between sensitivity and specificity as an observer changes the level of aggressiveness or conservatism used in interpreting the image. Therefore, the path along a particular ROC curve indicates changing the approach taken in reading an image rather than the quality of the images or the readers. In Figure 2, curve B is higher than curve A, implying that at any chosen level of aggressiveness, a reader (or an imaging system) represented by this curve performs better than that represented by curve A. For a study to be meaningful, it would also be important that the images reflect the relevant target population in terms of breast characteristics and the prevalence of disease and that, through biopsy and follow-up, the truth be known regarding the disease status for each examination. In an ROC study, the entire diagnostic imaging chain is evaluated, including the acquisition components, display components, and human observer. Thus, such a study represents the ideal means of assessing mammographic image quality. The main drawbacks of observer performance studies are that they are extremely resource and time consuming. For example, the reading image set must contain an adequate number of cases to achieve an acceptable level of confidence in the result. If small differences between ROC curves are to be investigated, the number of images required to provide statistical significance can be very large. If a technology (rather than the performance of a particular reader) is being evaluated, the study will have to be performed with multiple readers and possibly with multiple readings per reader [8]. This imposes an enormous time and labor commitment on the readers, normally busy radiologists. The problem becomes especially difficult when it is desired to evaluate mammography in the context of screening. In screening mammography, the number of cancers may only be 3 to 5 per 1,000 examinations. For a meaningful evaluation, the reading set must then contain a relatively small fraction of cancer cases. Furthermore, because testing more than a single configuration of the imaging system or systems requires a separate study, performing such testing at regular intervals can cause the size of the image set to be increased to the point at which the task of reading is at best onerous and at worst impractical. Mammographic Image Quality Regardless of whether an image is obtained in analog or digital form, breast cancer is detected on the basis of 4 types of signs on the mammogram [7]: (1) the characteristic morphology of a tumor or mass, (2) the shape and spatial configuration of mineral deposits called microcalcifications, (3) distortion of the normal architecture of the breast tissue, and (4) asymmetry between images of the left and right breast. The primary goal of mammography is to accurately visualize these signs if they exist. At the same time, it is important that these signs not be falsely identified if they are not actually present in the breast. To achieve these goals, it is necessary that the images be of high quality. A high-quality image is one that provides high sensitivity in detecting breast cancer and, at the same time, high specificity to identify a normal breast. Two aspects of image quality can be distinguished: technical and clinical. It is relatively easy to make technical measurements describing the above attributes and reasonable to infer a connection between these technical measures and clinical image quality [9]. THE ACQUISITION PROCESS IN DIGITAL MAMMOGRAPHY The digital mammography acquisition process is defined as all steps in image formation up to the point at which a digital image is sent to a display device for viewing. This includes patient positioning; the generation and shaping of the x-ray beam (both spatially and from the standpoint of the energy spectrum); the interaction between the beam and the compression paddle, breast, and detector;

5 Williams et al/digital Mammography Image Acquisition 593 the digitization of the detector output; and acquisitionrelated digital image processing. Acquisition processing compensates for specific, consistent imperfections in the acquisition process through techniques such as removing offsets from detector dark current and correcting for spatial variations in detector sensitivity or in x-ray beam intensity. Some manufacturers may also perform compensation for a decrease in the MTF at this point. Subsequent image processing of the resulting for-processing (or raw) image from the acquisition device provides specialized image enhancement techniques to produce a for-presentation image for interpretation by a radiologist and for image archiving. A digital detector has a faithful response to the intensity of incident x-rays over a very wide range. It can be designed to efficiently absorb x-rays, produce an electronic signal, digitize the signal, and store the results in computer memory. The output image is saved as a 2-D matrix in which each element represents the x-ray transmission corresponding to a particular path through the breast. At present, there are several designs of digital mammography systems either being used clinically or being evaluated pending regulatory approval. Four designs are shown schematically in Figure 3. In all of these except photostimulated storage phosphor (PSP) systems, the detector is composed of discrete sensing elements. The geometry and other characteristics of these detector elements (dels; Figure 4) play a large role in defining the imaging performance of the digital mammography system. In PSP systems, the sensor is continuous, but when it is scanned by a laser beam for readout, the size of the beam and the distance it sweeps between sample measurements define an effective del. Tissue Coverage Tissue coverage refers to the need to project as much of the breast tissue as possible onto an image; otherwise, a breast cancer may not visualized. Tissue coverage depends on the chosen view (projection); the positioning of the breast; and the geometrical relationship of the x-ray source, collimation, compression device, patient, grid, and image receptor. The machine-dependent aspects of tissue coverage are illustrated in Figure 5. To maximize the inclusion of breast tissue in an image, the x-ray source must project a ray that is virtually tangential to the chest wall, intercepting the image receptor at the point closest to the chest wall. If the x-ray source is malpositioned (eg, after tube replacement), or if the collimation, compression plate, or grids are not properly aligned, tissue will be excluded from the image. Similarly, if the image receptor or its housing has an inactive region adjacent to the chest wall, tissue will be missed. Typically, film mammography systems miss between 4 and 8 mm of tissue at the chest wall because of the front wall of the breast support and the edge of the cassette. All current digital units are also within this range, and this is desirable from a clinical prospective and should remain a minimum standard. Tissue coverage can be clinically assessed by comparing the retromammary aspects of the breast between the craniocaudal and mediolateral oblique views. The distance between the nipple and the anterior edge of pectoralis muscle or the posterior edge of the image on the craniocaudal view should be not more than 1 cm less than the distance between the nipple and the pectoralis on the mediolateral oblique view. Also, the anterior edge of the mediolateral oblique image of the pectoralis muscle should be convex, and the muscle should be seen at least down to no less than 1 cm above the level of the nipple. To assess the alignment of components that affect tissue coverage, a simple tool shown in Figure 6 can be used. This gauge is fixed to a plastic block or plate on which there is a lip representing the chest wall of the patient, such that when the plate is placed on the breast support surface, the lip engages the edge of the breast support. When it is imaged, it is possible to read directly the number of millimeters of tissue that is missed at the chest-wall side of the image. For large breasts, the field of view of the mammographic detector may be insufficient to image the entire breast at once. In such situations, the breast must be imaged in sections, and the resulting subimages must be tiled together to form the complete mammogram. This procedure is identical to that used for film mammography. The larger the detector s field of view is, the lower the probability that such multiple-section imaging will be required. Obtaining an image in a single view is desirable not only because the total acquisition time is minimized but also because of the possibility that slight changes in the breast configuration between views due to differences in compression direction could make the image sections difficult to match at the boundaries. Note also that the process of imaging the breast in sections results in an increase in radiation dose to regions of the breast that are exposed to x-rays in more than 1 subimage. Spatial Resolution Also referred to as sharpness, spatial resolution describes the ability of the imaging system to allow 2 adjacent structures to be visualized as separate. Alternatively, it can be used to describe the distinctness of an edge in an image. The system spatial resolution can be assessed by imaging a bar pattern consisting of alternating radio-opaque bars and radiolucent spaces of equal width (Figure

6 594 Journal of the American College of Radiology/ Vol. 3 No. 8 August 2006 Fig 3. Current designs of digital mammography acquisition systems. (A) Large-area charge-coupled device (CCD; no longer commercially available): cesium iodide phosphor coupled to multiple CCD readouts through demagnifying fiber-optic tapers. (B) Photostimulated storage phosphor: photostimulable phosphor plate with dual-side laser scan readout. (C) Indirect flat-panel: cesium iodide phosphor on large-area amorphous silicon active-matrix photodiode array. (D) Scanned-slot CCD (no longer commercially available): cesium iodide coupled to multiple CCD modules through fiber optics. Acquisition through scanning x-ray beam and detector across the breast. (E) Direct flat-panel: amorphous selenium direct x-ray converter on large-area amorphous silicon active-matrix electrode readout. TFT thin-film transistor.

7 Williams et al/digital Mammography Image Acquisition 595 Fig 4. The detector element (del) determines, in part, the spatial resolution capability of the detector. The aperture, d, of the active area and the pitch, p, or spacing between elements are shown. 7). An adjacent bar and space are referred to as a line pair. In use, the pattern is imaged, and an observer determines the most closely spaced line pair for which 2 adjacent bars are seen as clearly separate. It is possible to perform the test so as to isolate and quantify individual blurring effects. For example, to measure the resolution of the image receptor, the bar pattern can be fixed firmly to the receptor so that there is negligible blurring due to motion or focal spot size. This is a subjective test, however, and is not very useful in the analysis of complex imaging systems. A preferable measurement is the MTF [10]. The MTF describes how well an entire imaging system or a component, such as an image receptor or a subcomponent such as a phosphor, transfers the amplitude (analogous to contrast) of sinusoidal patterns from the incident x-ray pattern to the output. In a system containing several components affecting spatial resolution, the overall MTF can easily be calculated if the MTFs of the individual components are known. The system MTF is determined by multiplying at each spatial frequency the MTFs of the individual components. For example, as illustrated schematically in Figure 8, the MTF of a radiographic system is the product of that due to the focal spot, the detector, and any motion of the patient during the exposure. Determining which part of the system is responsible for limiting performance is thus possible. Spatial resolution is limited by various sources of blurring. In digital mammography, these are primarily due to the size of the focal spot of the x-ray tube and the magnification factor of a given structure of interest; unsharpness due to diffusion of light in the phosphor screen of the image receptor; the del effective aperture and pitch; and the relative motion of the x-ray source, the breast, or the image receptor during the exposure. The size, shape, and intensity distribution of the x-ray tube focal spot in combination with focal spot-to-object and object-to-image receptor distances affect geometric blurring (Figure 9A). Each point in the focal spot casts a sharp shadow of structures within the breast. The size of the shadow increases with the degree of magnification between that structure and the plane of the image receptor. The entire focal spot can be thought of as a large number of adjacent point x-ray emitters. The overlap of the shadows from each causes blur. To minimize geometric blurring, the focal spot size and object-to-image receptor distance, d 2, should be minimized, whereas the focal spot-to-object distance, d 1, should be maximized. Most mammographic procedures are performed with a moving Bucky-type grid. With the grid in place, there is a gap of 1 to 2 cm in distance between the exit surface of the breast and the image receptor. In modern mammographic units, the nominal focal spot size for most procedures is 0.3 mm. The convention of the National Electrical Manufacturers Association for defining nominal focal spot size allows the actual distribution of radiation to be considerably larger (1.5 to 2 times) than the nominal value. Also, the effective size of the spot varies over the image plane, being largest near the chest wall. Therefore, geometric resolution also varies over the image. For

8 596 Journal of the American College of Radiology/ Vol. 3 No. 8 August 2006 Fig 5. The effect of system geometry on tissue coverage near the chest wall. (A) Proper alignment causes the central ray from the x-ray source to be projected tangent to the chest wall. (B) Improper position of the x-ray source causes the image of some tissue proximal to the chest wall to be projected inside the patient s thorax. (C) Incorrect setting of the collimator or (D) incorrect positioning of the compression paddle or grid will also cause tissue to be missed. geometric magnification, a second focal spot of nominal size (approximately 0.1 mm) should be used to avoid an unacceptable loss of resolution. Factors affecting unsharpness for digital mammography referable to the detector itself are signal diffusion between dels, the active area of each del (aperture size), the pitch or center-to-center spacing between dels, and the pixel size used to display the image. Most detectors are constructed as a set of discrete dels, as shown schematically in Figure 4. Each del has an active area with dimension d, and this may be surrounded by an area that is insensitive to the incident radiation. This causes the pitch, or spacing between dels, p, to be greater than d. For square dels, the relative area of sensitivity, d 2 /p 2 is called the fill factor, and this in part determines the geometric radiation efficiency of the detector. The del size also determines the basic spatial resolution associated with the del. Because information is smeared over d, the smaller d is, the less blurring results and therefore the higher the spatial resolution. As shown in Figure 10, the MTF associated with the del falls to zero at a spatial frequency of 1/d cycles/mm. A detector with 50- m dels passes spatial frequencies up to 20 cycles/ mm. Note that the MTF does rise again at frequencies beyond the first zero, but the information may not be reliably depicted beyond this point. For example, between the first and second zero points, there is a reversal of contrast, so that structures that should be dark appear light and vice versa. The pitch is also important in affecting image quality. The spacing between samples determines whether information is lost between measurements. If this occurs, a phenomenon called aliasing can result. To avoid aliasing, the highest spatial frequency of information in the image, f max, must be less than 1/(2p), which is referred to as the sampling frequency. Otherwise, aliasing causes information at spatial frequencies greater than 1/(2p) to be represented at lower spatial frequency, as illustrated in Figure 11.InFigure 11A are 2 sinusoidal patterns, 1 of low frequency and 1 of higher frequency, and the image in which both patterns are combined where there is no limitation due to sampling. Both patterns are seen clearly in the combined image. When the sampling pitch is too great, as in Figure 11B, we see that not only is the higher spatial frequency information poorly imaged, but the resulting erroneous (aliased) information that it produces at the lower spatial frequency also interferes with the proper representation of the lower frequency object. Therefore, for example, although 50- m dels will pass information up to 20 cycles/mm, they provide unaliased imaging only up to 10 cycles/mm. Any content in the image beyond 10 cycles/mm will cause aliasing to affect lower frequencies. Note that resolution specifications based on d and p

9 Williams et al/digital Mammography Image Acquisition 597 Fig 6. Device to measure the amount of tissue missed at the chest wall. On this system, 4.2 mm are cut off from the image. apply only to the directions parallel to the rows and columns of the image. For other directions, resolution will be lower. For example, at 45, the aperture and pitch will be about 40% greater than along the principal axes, and the resolution and aliasing will be affected accordingly. The MTF for a digital detector is determined partially by the effective del aperture and partly by physical signal spread between dels. The design philosophy for digital mammography balances many factors, including tradeoffs between spatial resolution, SNR, and radiation dose requirements, as well as manufacturing and economic considerations. The hypothesis of the design is that the improved dynamic range and SNR provided by digital mammography will outweigh a loss of limiting spatial resolution, and therefore, the del pitch has been chosen by different manufacturers to provide sampling frequencies between 5 and 10 cycles/mm. To isolate the amount Fig 7. Measurement of spatial resolution with a bar pattern test. From the top of the image, the first row illustrates the concept of line pairs (lp) per millimeter, the second row demonstrates a pattern containing sections spanning a range of line pairs per millimeter, and the third row illustrates the effects of an unsharp imaging system on the resolution of the bars. Fig 8. Modulation transfer function (MTF). This figure describes how well an imaging system conveys the modulation or contrast from the input to the output. The curve labeled S-F shows the MTF of a mammographic screen-film system, and the curve labeled FS shows the MTF due to the smaller focal spot and geometric magnification. The overall MTF, including both effects is shown, by the x values.

10 598 Journal of the American College of Radiology/ Vol. 3 No. 8 August 2006 Fig 9. Some of the sources of image blurring in mammography. (A) Geometric unsharpness due to the focal spot and magnification. (B) Unsharpness due to the lateral spread of signal (eg, light produced in a phosphor) in the image receptor. of intrinsic blurring caused by the detector from the effects of sampling, it is customary to measure the MTF before sampling or pixilation occurs and to specify the presampling MTF of the detector. This measurement can be done by imaging a sharp edge or a narrow slit that is tilted by a small angle with respect to the principal axes of the detector [11]. By combining measurements of this edge spread function or line spread function from several rows of the image, it is possible to simulate sampling at an interval much finer than the del pitch. The presampling MTFs for several digital mammography systems are shown in Figure 12. Another factor that increases image blur is long exposure time, which may result in patient motion. In digital mammography, most motion blurring is caused by movement of the breast during exposure. It can be minimized by using a short exposure time and by compressing the breast. Settings of kilovolt potential may be increased for thick, dense breasts to allow the reduction of exposure time. Magnification techniques generally require longer exposure times because small focal spots with lower tube current (milliamps) capacity are used. The amount of blurring depends on the speed of the motion in the patient and the duration of the exposure (Figure 13). For consideration of blur, the exposure time in the case of the large-area digital systems is the complete exposure time, whereas for a scanned-slot charge-coupled Fig 10. MTF associated with the del size, d. In this example, d 50 m, f spatial frequency; f N sampling frequency; f s cut off frequency due to del size, d; p pitch. MTF associated with the size, d. device (SSCCD) system, only part of the breast is exposed at any time. Even though the overall exposure time is generally longer for a scanning system, the time that x-rays expose a particular part of the breast (ie, the time that affects blurring) is only a small fraction of the total exposure time. For scanning systems, motion can, however, cause misregistration artifacts between the anatomy that is imaged before a motion occurs and that imaged after. For a given image receptor sensitivity (a specified amount of radiation required at the image plane), there is a trade-off between motion blur and geometric unsharpness. If one attempts to reduce motion blur, a greater x-ray output must be available. This is normally accomplished by increasing the tube current or reducing the distance from x-ray tube to image receptor. The former requires increasing the focal spot size. In either case, geometric unsharpness will become greater. Reducing distances also makes patient positioning for the examination more difficult. With digital mammography, an alternative approach may be to increase tube kilovoltage, because the loss in contrast may be compensated for by adjustment of the image display. In addition, it is useful to consider the mechanisms operating on spatial resolution in the different detector systems. For SSCCDs or indirect flat-panel (IFP) sys-

11 Williams et al/digital Mammography Image Acquisition 599 Fig 11. Aliasing. (A) Pattern containing sine waves at 2 spatial frequencies and their summation. (B) The effect of undersampling, which causes the higher spatial frequency pattern to be unresolved and also causes the higher frequency pattern to disturb the lower frequency pattern, which would otherwise be imaged correctly. IFP PSP DFP SSCCD Fig 12. Presampled modulation transfer functions (MTFs) for various current digital mammography detectors. tems, the first source of blur that occurs is spreading of emitted light within the cesium iodide crystals. This is determined both by the thickness of the phosphor (sharpness can be exchanged for quantum efficiency) and by the engineering of the phosphor (crystal structure, reflective and absorptive materials). The latter are closely guarded secrets of the manufacturers. In SSCCD systems, the quality of the fiberoptic coupling of light between the phosphor and the charge-coupled device (CCD), specifically, any deviation in the one-to-one correspondence of fibers between the entrance and exit surfaces of the coupler due to twisting, warping, and so on, will also affect resolution. In direct flat-panel detectors, the field across the direct conversion material must be adequate to ensure that there is negligible lateral spreading of charge created in the material before it is collected at the electrodes. The spatial resolution is affected by the aperture and pitch of the dels in the readout device, namely, the CCD cells or the photodiode or electrode elements in flat panel devices using thin-film transistor arrays.

12 600 Journal of the American College of Radiology/ Vol. 3 No. 8 August 2006 Fig 13. Motion unsharpness is due to relative motion between the x-ray source, structures within the breast, and the image receptor during image acquisition. It is determined by the velocity of motion and the exposure time per detector element. In an SSCCD system, the detector moves across the breast while the charge in the CCD elements is shifted down the columns of the CCD at the same speed as but in the opposite direction of the detector motion. The matching of the mechanical and electrical transfer speeds is critical, because a mismatch will cause blur. Any inefficiency of charge transfer in the CCD will cause additional blurring. There is also a minor effect on resolution because the detector moves continuously, whereas the charges are shifted in discrete jumps through the phases of each CCD element and from row to row down the columns. Although the readout structures of IFP and direct flatpanel devices are very similar, there is an important difference between the collection of light and charge, which can provide both advantages and disadvantages. For example, in IFP detectors, only the light produced in the phosphor above the active area of each del will be measured. In direct flat-panel detectors, charge travels along electric field lines, so that the charge produced in a region above an inactive part of the del can be redirected by the shape of the field to fall on the active electrode area. This makes the effective fill factor for the del higher than it would otherwise have been. On the other hand, a detector composed of adjacent dels will always suffer from undersampling, so that aliasing will occur unless some multiple sampling strategy is used. This is generally too time consuming and complex to be practical. The spreading of light in an IFP detector may provide adequate blurring of the high-spatial-frequency information to reduce the amount of aliasing compared with the sharper direct conversion detector of the same del dimensions. The mechanisms determining the resolution characteristics of PSP detectors are quite different from the others. Although this is a phosphor system, the spread of light produced by the phosphor is not of concern, because the emitted light is not used to provide spatial localization. Instead, the spatial information (ie, the effective del) is determined by the size of the scanned laser beam on the imaging plate used in the readout and the distance that the beam is moved (pitch) between successive lines of the laser-stimulated light. Light produced by the phosphor when the traps are emptied by the energy from the laser light must be collected as efficiently as possible. The effective size of the laser beam is determined by the actual beam size as well as the amount of scattering of the laser light that takes place within the phosphor. This determines the point spread function from which the MTF due to the laser can be calculated. The uniformity of motion of the scanning beam and the indexing of the phosphor plate also have an effect on the resolution and uniformity of the image. The effect of spatial resolution is most easily observed when considering the imaging of fine detail in the breast, such as spiculations radiating from a mass or calcifications. In addition to allowing the detection of calcifications, high spatial resolution is necessary to provide information on their shape and margin to assess whether their cause is more likely to result from a benign or malignant process. Calcifications are the most common presentation of the ductal carcinoma in situ, an early form of breast cancer confined to the duct itself. In Figure 14, a region of a digital mammogram containing calcifications is shown. In Figure 14A, the calcifications are depicted with a very high spatial resolution. In Figure 14B, they have been blurred in the manner that would occur if a system with a 50- m del were used to acquire the image, while Figures 14C and 14D represent 100- m and 200- m dels, respectively.

13 Williams et al/digital Mammography Image Acquisition 601 Fig 15. Calculated subject contrast for a small tumor and for calcifications compared with the x-ray energy used for imaging. Fig 14. Effect of del size on the spatial resolution of microcalcifications. Top left: Calcifications depicted with high spatial resolution. Top right: Blurring of image that would occur with a system that used 50 m dels. Bottom left: image with 100 m dels. Bottom right: Image with 200 m dels. Radiographic Contrast Radiographic contrast is the magnitude of the signal difference between the structure of interest and its surroundings in the displayed image. Radiographic contrast is influenced by 2 factors: subject contrast and display contrast. Contrast is typically considered for larger areas ( 1 cm 2 ) in an image. Subject contrast is measured in terms of the relative difference in x-ray exposure to the image receptor, transmitted through one part of the breast and through an adjacent part, whereas overall radiographic contrast, which depends on both the subject contrast and the display contrast, is expressed in terms of the optical density difference between 2 areas on the processed laser film or as the relative brightness difference between the corresponding areas in an image displayed on a monitor. Subject contrast is especially important in mammography because of the subtle differences in the soft-tissue density of normal and pathologic structures of the breast and because of the importance of detecting minute details such as calcifications and the marginal structural characteristics of soft-tissue masses. Subject contrast is caused by differences in the x-ray attenuation properties of the lesion and those of the surrounding tissue. There is less inherent subject contrast in breasts than elsewhere in the body, because there are no bony structures or gas. Cancers and fibroglandular tissue can show similar x-ray attenuation (defined here as mass per unit volume) because of the similarity in the atomic number of lesions relative to normal tissue. These differences and therefore the contrast also depend on the distribution of x-ray energies (spectrum) used for producing the mammogram. The x-ray spectrum is determined by the tube target material, kilovolt potential setting, and filtration (either inherent in the tube or added in its exit port). Figure 15 illustrates how the subject contrast of a tumor and calcification decrease as the energies of the x-rays increase. Maximizing subject contrast is particularly important. Most commonly, molybdenum target x-ray units are used. These provide high radiation output at the characteristic emission energies for molybdenum of 17.9 and 19.5 kev. When a molybdenum filter (typically to 0.03 mm) is used, the spectrum is strongly suppressed at lower photon energies and at energies greater than 20 kev because of the strong increase in x-ray absorption by molybdenum that occurs at its k-shell absorption edge. Therefore, the spectrum is rich in photons at and near the characteristic energies. These energies yield high subject contrast while avoiding the excessive radiation dose for breasts of average thickness, which would occur at lower energies. For thicker, denser breasts, few low-energy photons are able to pass through the breast and, because of the greater filtering action of a thick dense breast, absorption differences among structures become smaller in the resulting harder (or higher average energy) x-ray beam. Therefore, subject contrast is not as high as with averagesize or fatty breasts. In addition, as a dose-saving measure, higher energy incident x-ray beams are typically used to image these breasts. Although the effective energy

14 602 Journal of the American College of Radiology/ Vol. 3 No. 8 August 2006 can be adjusted by varying kilovoltage, the effect on the spectrum is somewhat limited because of the dominance of the fixed-energy characteristic x-rays from the target. This motivates the use of targets and filters of different materials to tune the spectral shape. Rhodium has a K- absorption edge at 23 kev, and a rhodium filter will pass energies between 20 and 23 kev. Used with a molybdenum target in combination with an increase in kilovoltage, this will give a more penetrating spectrum than that obtained with the molybdenum-molybdenum combination. This can be helpful in imaging thick breasts ( 5or 6 cm) of fatty or average composition. If an even more penetrating beam than available with molybdenum-rhodium is desired, it is possible to use a rhodium target in combination with a rhodium filter. Rhodium has characteristic emissions at 20 and 23 kev. The rhodium-rhodium combination is most effective with very dense, difficult-to-penetrate breasts, providing some dose reduction while preserving as much subject contrast as possible in these difficult-to-image breasts. It is also possible to provide a suitable spectrum for imaging dense breasts with a tungsten target tube and various metallic filters such as rhodium. Although this does not provide the quasi-monoenergetic x-rays available with the molybdenum and rhodium targets, careful choice of kilovoltage and filter material and thickness can yield an excellent result in terms of contrast and dose. For digital systems, the ability to adjust display contrast may make it advantageous to use slightly higher energy x-ray beams than are used with film mammography (for which 22 to 32 kvp is typically used, depending on the thickness and density of the breast). Provided that an adequate SNR is maintained, this may provide a dose reduction compared with screen-film mammography, especially for large or dense breasts. Digital Mammographic Imaging Screening Trial data suggest that this can be accomplished with improved diagnostic accuracy [12]. Another source of contrast degradation for digital mammography is the presence of scatter radiation. In soft tissue, even at the low energies used in mammography, scattering is an important mechanism by which x-rays interact with breast tissue. Scattered x-rays that escape the breast and are recorded by the image receptor reduce image contrast and the apparent sharpness. The amount of scattered radiation recorded compared with the useful, directly transmitted x-ray intensity, is characterized by the scatter-to-primary ratio. It is not unusual for the scatter-to-primary ratio to be greater than 1.0 [13,14]. For film systems, scattered x-rays recorded by the image receptor reduce the value of subject contrast, use up some of the available recording range or latitude of the film in recording essentially useless information, and add noise to the image, thereby reducing its SNR, which is a measure of the information content of the mammogram. In digital mammography, the same factors apply, but the effect of scattered radiation on the final radiographic contrast is somewhat different. Because x-rays may scatter multiple times within the breast, their spatial distribution is diffuse (ie, mainly affecting the low-spatialfrequency part of the MTF). For this reason, in digital systems, much of the contrast can be recovered by viewer adjustment of the computer image display. Similarly, the system can be designed such that the dynamic range of the image receptor is very large, so that recording of scattered radiation will not be a limiting factor. Under these conditions, only the third effect, the addition of random quantum noise, should be of any importance. The use of specifically designed grids for mammography reduces the amount of scattered radiation detected and improves subject contrast. This is particularly important when imaging thick, dense breasts [15,16]. Grids, consisting of lead strips separated by spacers of radiolucent material, are a standard feature of modern mammographic x-ray units. Most of these grids are 1-D and move during the x-ray exposure to allow for blurring of the image of the grid septa, rendering them invisible. Alternatively, one manufacturer provides a 2-D focused rhombic cellular structure grid in which the interspace material is air. These grids offer the potential of improved image contrast and transmission efficiency compared with conventional grids. For all digital systems except SSCCD systems, grids are typically used, because of the reduction in noise due to the scattered radiation and the consequent increase in the SNR. With a digital system, any exposure increase necessitated by the use of the grid is due only to its incomplete transmission of the primary radiation. An SSCCD system uses a narrow scanned beam of x-rays, which reduces the scatter-to-primary ratio. In addition, the long, narrow detector with collimation at its entrance surface rejects much of whatever scattered radiation is incident on the detector. Currently, a grid is not used with this system, and exposures can be lower. Subject contrast should also be optimized by good breast compression. Compression is an important factor in reducing scattered radiation. In a study using phantoms, Barnes and Brezovich [13] showed that reducing the thickness from 6 to 3 cm by compression reduced the scatter-to-primary ratio from 0.8 to 0.4. The use of a mammographic grid further reduced the scatter-to-primary ratio to In addition, compression can provide several other technical improvements in image quality, which can be achieved without compromising other image quality factors. These improvements include the immobilization of the breast, which reduces blurring caused by motion; the location of structures in the breast closer

15 Williams et al/digital Mammography Image Acquisition 603 Fig 16. The influence of the quantity of x-rays used to make a mammogram on its quality and the reliability of characterizing structures containing fine detail. Increasing milliamperage decreases the relative magnitude of noise and improves diagnostic detail. to the image receptor, which reduces geometric blurring; the production of a more uniformly thick breast, which in turn results in more even penetration by x-radiation and less difference in radiographic density in the area between the chest wall and the nipple; the reduction of radiation dose; and, finally, the spreading of breast tissue, enabling suspicious lesions to be more easily identified. Finally, for digital mammography, the signal stored in digital form is directly (or logarithmically) proportional to the amount of radiation transmitted through the breast. With a properly designed image acquisition system, the dynamic range should be adequate to measure the entire range of intensities from that of the unattenuated beam outside the breast to that through the densest, thickest part of the breast. For this reason, the stored image reflects the inherent subject contrast very faithfully. Noise In digital mammography, it is not very meaningful to discuss contrast without also considering noise. Radiographic noise or mottle is the unwanted variation in random optical density in a radiograph that has been given a uniform x-ray exposure [17-19]. Quantum mottle is caused by the random spatial variation of the x-ray quanta absorbed in the image receptor, and its effect is reduced as more x-rays are used to form the image. The effect of using fewer quanta on noise and on perception of subtle contrasts can also be seen in Figure 16, in which digital images have been produced of a phantom containing star-shaped objects mimicking calcifications. Noise can be quantified in terms of the standard deviation of the number of x-ray quanta recorded in a given area of the image receptor or the standard deviation in image signal (optical density or digital image value) over a given area (region of interest) of the image. This says nothing, however, about the spatial characteristics of the noise. This is better described by the noise power spectrum of the image [20,21]. More important than the absolute noise level is the consideration of the relative magnitudes of the noise and the image signal. This is described by the SNR or more commonly, the signal-to-noise transfer efficiency or detective quantum efficiency of the image [1,21-23]. Detective quantum efficiency describes the transfer of the SNR from the x-ray pattern incident on the imaging system to its output. It is generally plotted against spatial frequency, as shown in Figure 17. Figure 18 illustrates the effect of noise in imaging a dense breast. Because an insufficient amount of radiation is transmitted through the breast, the mammogram is quite noisy, making the reliable detection of small structures, such as calcifications, difficult. It is extremely important that an adequate x-ray exposure is used to ensure that detection and characterization of lesions will not be impaired by the presence of noise. An effective automatic exposure control system is valuable in helping achieve this goal. A structure in the breast must be visualized with respect to its surroundings, so it is useful to calculate the ratio of the signal difference (the difference in image value between 2 points of interest) to the noise in that difference. This is described by the signal difference-to-

16 604 Journal of the American College of Radiology/ Vol. 3 No. 8 August 2006 Fig 17. Detective quantum efficiency (DQE) versus spatial frequency for A, a modern mammographic screen-film image receptor, and B, a digital mammography detector (direct flat-panel [DFP]). Fig 18. The impact of excessive quantum noise due to underexposure on the quality of a digital mammogram. Noise interferes with detection and characterization of calcifications. Fig 19. Calculated signal difference-to-noise ratios (SDNRs) versus effective x-ray energy for different thicknesses of 50% fat and 50% fibroglandular tissue fora5mmdiameter tumor (A) and 200- m calcifications (B) in compressed breasts ranging in thickness from3to8cm. noise ratio (SDNR) of the image. To reliably detect a structure in the breast in the presence of noise, the SDNR must exceed some threshold value, typically 5. Calculated values of the SDNR are presented in Figures 19A and 19B for the case of detection of a small mass and calcifications, respectively, in breasts of different thicknesses. Note that for typical radiation exposure levels used in mammography, the SDNR is more than adequate for the mass but may be marginal for the calcifications. A simple method of measuring SDNR and an example of measurements are illustrated in Figure 20. Quantum noise should be the principal contributor to the signal fluctuation seen in a uniformly exposed radiograph. Factors affecting the perception of quantum mottle in mammography include x-ray interaction efficiency, efficiency of converting x-rays to light or electrons and collecting the signal, light diffusion in phosphors, and radiation quality. When sensitivity is increased because of increased x-ray absorption (higher quantum efficiency) for a given detector output, quantum mottle is not increased. When speed is increased because of other amplification mechanisms, fewer x-rays are used to form the image, and therefore quantum mottle is increased. Spreading of signal in the detector blurs the recording of quantum noise, so that it becomes less apparent but also causes a decrease in spatial resolution. Higher energy x-rays are absorbed with lower quantum efficiency, resulting in a noisier image for a given number of x-rays

17 Williams et al/digital Mammography Image Acquisition 605 determine the radiation dose used for a given examination. In digital mammography there may be spatial variations in sensitivity of the receptor, which would cause an image to have structure that is unrelated to the tissues in the breast. As long as the system design ensures that these variations are temporally stable, this fixed pattern noise can be eliminated by imaging a uniform field of x-rays and using the recorded image as a correction mask to make the image uniform (Figure 21). This procedure is often referred to as flat-field correction [24]. Fig 20. A simple test object for the measurement of the signal difference-to-noise ratio (SDNR). transmitted through the breast. More important, they produce more secondary quanta per x-ray, so that a given signal (pixel value) can be achieved with fewer x-ray photons (ie, higher noise). For mammographic imaging systems, quantum noise is a fundamental factor that can never be eliminated, only minimized. This is accomplished by attempting to maximize quantum efficiency and by using an adequate radiation dose. In digital mammography, the detector and electronics should be designed to have adequate dynamic range and number of bits of digitization to precisely record the entire range of x-ray intensities transmitted by the breast. If this is the case, the electronic image can be amplified as much as desired so that there is really no constraint on image brightness. If an inadequate number of quanta are used, however, the SDNR will be inadequate, and it is really the desired SDNR that should The Elimination of Artifacts and Other Image Quality Problems Acquisition processing (also known as image preprocessing) represents the manipulation of the direct output of a digital detector to correct for dead dels, nonfunctional columns or rows, stationary structured noise patterns, local gain variations, background (no signal) offsets, and global nonuniform response. Detector imperfections produce added noise in an output image that can overwhelm the signal variations inherent in the recorded x- ray image. Variations in the incident x-ray beam caused by the x-ray equipment itself (eg, the heel effect, added beam filtration, and grid transmission) can also detrimentally affect the output image quality. In some cases, corrections are possible for a fixed x-ray tube or detector geometry. Correction algorithms are specific to the type of digital detector. For all digital systems with discrete dels (the exception being cassette-based PSP detectors), initial corrections require the identification of the locations of dead or transient element responses, as well as row and column defects. A map is determined and the averaged response of adjacent dels is substituted into the bad del. More than 1 adjacent dead element, dead column, or dead row makes this correction potentially unsatisfactory, particularly if in the central area of the detector matrix, where diagnostic information is most crucial. The number of nonfunctional dels (both adjacent and nonadjacent) determines the acceptability of the detector. Over time, some dels might lose functionality, which will require a reanalysis of the bad-element map. A trigger for such an evaluation occurs when dropouts become visible in the processed digital images or flat-field images. Subsequent acquisition processing correction schemes are classified into 1-D and 2-D methods and typically require a linear (or modifiable linear) response over the useful dynamic range of the detector (eg, twice the incident radiation exposure results in twice the corresponding digital output value). Uncorrected digital detector response is determined by acquiring images with unattenuated or uniformly attenuated x-ray beam fluence of high incident exposure to reduce the quantum noise vari-

18 606 Journal of the American College of Radiology/ Vol. 3 No. 8 August 2006 Fig 21. The effect of flat-fielding correction. (A) The raw image. (B) The same image after flat-field correction. ability. For mammography, the x-ray beam is typically nonuniform because of heel-effect intensity variations, falling off significantly toward the anode (nipple side) of the field. Cassette-based PSP and SSCCD arrays use a 1-D based correction method. For PSP systems, gain variations are chiefly due to the pickup light guide, and the correction is applied in the scan direction only. For PSP, future 2-D correction methods could be used (above and beyond the 1-D corrections for the PSP reader response variations) by evaluating the response of each uniformly exposed imaging plate (after 1-D corrections are applied) and using methods described below for corrections of nonuniform response in 2 dimensions. This would require each PSP imaging plate in the inventory to have a 2-D correction map. In addition, because of the positional variability of the imaging plate within the mechanical translation device, the corrections would likely be applicable to the low spatial frequency spectrum and correct global variations but not local variations. Two-D correction techniques, often called flat-fielding, generally involve dividing each image by a mask image. If the mask image is not acquired with sufficient radiation, it will be noisy, and this noise will propagate into the corrected image. In addition, there can be noise associated with the electronic circuitry that amplifies and digitizes the detector signal. For a digital system to perform well, it must be designed to minimize these nonquantum noise sources such that the SNR is determined by the level of radiation used. As an example, the measured detective quantum efficiency for a direct amorphous selenium thin-film transistor array, based on the results of Yorker et al [25], is shown as curve B in Figure 17. The frequency of 1-D or 2-D calibration is dependent on the type of digital detector and recommendations for quality control (QC) by the manufacturer. For instance, daily calibration (or before patient use) is recommended for a digital mammography biopsy system using a fiberoptic taper-ccd sensor, with yearly (or as needed) for 1-D corrections for PSP systems and all other digital detector systems falling within that range. The basis for the calibration frequency is dependent on the amount of potential drift of the detector response at any time relative to the correction matrix acquisition time. Certainly, any drift or variation from the calibrated values will result in an incomplete correction and cause the partial injection of structured noise into the corrected image. For all digital detector systems, if variations caused by structured noise or other artifacts are noted at any time (during inspections, on patient images, etc), a flat-field calibration should immediately be performed before further patient imaging. Also, calibration should be implemented after any repairs of the system directly related to

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Mammography: Physics of Imaging

Mammography: Physics of Imaging Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness

More information

Acquisition, Processing and Display

Acquisition, Processing and Display Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

Amorphous Selenium Direct Radiography for Industrial Imaging

Amorphous Selenium Direct Radiography for Industrial Imaging DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial

More information

PRACTICE GUIDELINE FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY

PRACTICE GUIDELINE FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College

More information

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory

More information

CR Basics and FAQ. Overview. Historical Perspective

CR Basics and FAQ. Overview. Historical Perspective Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors

More information

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

10/26/2015. Study Harder

10/26/2015. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

10/3/2012. Study Harder

10/3/2012. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

LECTURE 1 The Radiographic Image

LECTURE 1 The Radiographic Image LECTURE 1 The Radiographic Image Prepared by:- KAMARUL AMIN ABDULLAH @ ABU BAKAR UiTM Faculty of Health Sciences Medical Imaging Department 11/23/2011 KAMARUL AMIN (C) 1 Lesson Objectives At the end of

More information

Breast Imaging Basics: Module 10 Digital Mammography

Breast Imaging Basics: Module 10 Digital Mammography Module 10 Transcript For educational and institutional use. This test bank is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893 Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -

More information

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring

More information

Beam-Restricting Devices

Beam-Restricting Devices Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the

More information

Do you have any other questions? Please call us at (Toll Free) or , or

Do you have any other questions? Please call us at (Toll Free) or , or INSTRUCTIONS Read the appropriate course/ textbook. This is an open book test. A score of 75% or higher is needed to receive CE credit. You will have a maximum of three attempts to pass this course. Please

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

Digital Detector Array Image Quality for Various GOS Scintillators

Digital Detector Array Image Quality for Various GOS Scintillators Digital Detector Array Image Quality for Various GOS Scintillators More info about this article: http://www.ndt.net/?id=22768 Brian S. White 1, Mark E. Shafer 2, William H. Russel 3, Eric Fallet 4, Jacques

More information

Collimation Assessment Using GAFCHROMIC XR-M2

Collimation Assessment Using GAFCHROMIC XR-M2 Collimation Assessment Using GAFCHROMIC XR-M2 I. Introduction A method of collimation assessment for GE Senographe full-field digital mammography (FFDM) systems is described that uses a self-developing

More information

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images B R E A S T I M A G I N G S O L U T I O N S Setting the benchmark for mammography M-IV Series Innovations in breast imaging The Lorad M-IV Series exemplifies Hologic s commitment to developing advanced

More information

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality Digital radiography: Digital radiography is set to become the most common form of processing radiographic images in the next 10 years. This is due to a number of practical and image quality issues. Practical

More information

radiography detector

radiography detector Clinical evaluation of a full field digital projection radiography detector Gary S. Shaber'1, Denny L. Leeb, Jeffrey Belib, Gregory Poweii1', Andrew D.A. Maidment'1 a Thomas Jefferson University Hospital,

More information

Features and Weaknesses of Phantoms for CR/DR System Testing

Features and Weaknesses of Phantoms for CR/DR System Testing Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal

More information

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org INTRODUCTION

More information

Digital Imaging Considerations Computed Radiography

Digital Imaging Considerations Computed Radiography Digital Imaging Considerations Digital Radiography Computed Radiography o Cassette based Direct or Indirect Digital Radiography o Cassetteless Computed Radiography 1 CR Image Acquisition Most like conventional

More information

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA RADT 2913 COMPREHENSIVE REVIEW 1 The CR cassette is backed by aluminum that: A. reflects x-rays B. absorbs x-rays C. captures the image D. transmits x-rays

More information

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY 12 th A-PCNDT 2006 Asia-Pacific Conference on NDT, 5 th 10 th Nov 2006, Auckland, New Zealand PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY Rajashekar

More information

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,

More information

Imaging Technique Optimization of Tungsten Anode FFDM System

Imaging Technique Optimization of Tungsten Anode FFDM System Imaging Technique Optimization of Tungsten Anode FFDM System Biao Chen a*, Andrew P. Smith b, Zhenxue Jing a, Elena Ingal a a Hologic, Inc. 600 Technology Drive, DE 1970 b Hologic, Inc. 35 Crosby Drive,

More information

Patient-Assisted Compression Impact on Image Quality and Workflow

Patient-Assisted Compression Impact on Image Quality and Workflow Patient-Assisted Compression Impact on Image Quality and Workflow Senographe Pristina In 2017, GE Healthcare s Senographe Pristina ( Pristina ) was approved by the FDA using the standard technologist-controlled

More information

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging The Physics of Radiation Oncology X-ray Imaging Charles E. Willis, Ph.D. DABR Associate Professor Department of Imaging Physics The University of Texas M.D. Anderson Cancer Center Houston, Texas Overview

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1 Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1 Gham Hur, M.D., Yoon Joon Hwang, M.D., Soon Joo Cha, M.D., Su Young Kim, M.D., Yong Hoon Kim, M.D.

More information

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology CODE: RADT 156 INSTITUTE: Health Science TITLE: Equipment Operation I DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of equipment operation and maintenance of radiographic

More information

2017 West Coast Educators Conference Orlando. Projection Geometry. 1. Review hierarchy of image qualities (amplified version):

2017 West Coast Educators Conference Orlando. Projection Geometry. 1. Review hierarchy of image qualities (amplified version): Spatial Resolution in the Digital Age: NOTES Quinn B. Carroll, MEd, RT 2017 West Coast Educators Conference Orlando Projection Geometry 1. Review hierarchy of image qualities (amplified version): a. Maximum

More information

Quality Control of Full Field Digital Mammography Units

Quality Control of Full Field Digital Mammography Units Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of

More information

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging

More information

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools

More information

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation Image Quality Review I: Basics and Image Quality TH-A-16A-1 Thursday 7:30AM - 9:30AM Room: 16A J. Anthony

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated

More information

Digital Images & Image Quality

Digital Images & Image Quality Introduction to Medical Engineering (Medical Imaging) Suetens 1 Digital Images & Image Quality Ho Kyung Kim Pusan National University Radiation imaging DR & CT: x-ray Nuclear medicine: gamma-ray Ultrasound

More information

Essentials of Digital Imaging

Essentials of Digital Imaging Essentials of Digital Imaging Module 1 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 1 Fundamentals 1. ASRT Animation 2. Welcome Welcome to Essentials of Digital Imaging

More information

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc Quality Control for Stereotactic Breast Biopsy Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc. 716-924-0350 Methods of Imaging Guided Breast Biopsy Ultrasound guided, hand-held needle

More information

Observer Performance of Reduced X-Ray Images on Liquid Crystal Displays

Observer Performance of Reduced X-Ray Images on Liquid Crystal Displays Original Paper Forma, 29, S45 S51, 2014 Observer Performance of Reduced X-Ray Images on Liquid Crystal Displays Akiko Ihori 1, Chihiro Kataoka 2, Daigo Yokoyama 2, Naotoshi Fujita 3, Naruomi Yasuda 4,

More information

RADIOGRAPHIC EXPOSURE

RADIOGRAPHIC EXPOSURE RADIOGRAPHIC EXPOSURE Receptor Exposure Receptor Exposure the that interacts with the receptor. Computed Radiography ( ) requires a. Direct Digital Radiography (DR) requires a. Exposure Indicators Exposure

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

Investigation of Effective DQE (edqe) parameters for a flat panel detector

Investigation of Effective DQE (edqe) parameters for a flat panel detector Investigation of Effective DQE (edqe) parameters for a flat panel detector Poster No.: C-1892 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit D. Bor 1, S. Cubukcu 1, A. Yalcin 1, O.

More information

Visibility of Detail

Visibility of Detail Visibility of Detail Radiographic Quality Quality radiographic images represents the, and information is for diagnosis. The of the anatomic structures and the accuracy of their ( ) determine the overall

More information

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) 2/ Overview Digital

More information

Testing a wavelet based noise reduction method using computersimulated

Testing a wavelet based noise reduction method using computersimulated Testing a wavelet based noise reduction method using computersimulated mammograms Christoph Hoeschen 1, Oleg Tischenko 1, David R Dance 2, Roger A Hunt 2, Andrew DA Maidment 3, Predrag R Bakic 3 1 GSF-

More information

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images B R E A S T I M A G I N G S O L U T I O N S Setting the benchmark for mammography M-IV Series Innovations in breast imaging The Lorad M-IV Series exemplifies Hologic's commitment to developing advanced

More information

X-ray light valve (XLV): a novel detectors technology for digital mammography*

X-ray light valve (XLV): a novel detectors technology for digital mammography* X-ray light valve (XLV): a novel detectors technology for digital mammography* Sorin Marcovici, Vlad Sukhovatkin, Peter Oakham XLV Diagnostics Inc., Thunder Bay, ON P7A 7T1, Canada ABSTRACT A novel method,

More information

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance

More information

Evaluating Commercial Scanners for Astronomical Images. The underlying technology of the scanners: Pixel sizes:

Evaluating Commercial Scanners for Astronomical Images. The underlying technology of the scanners: Pixel sizes: Evaluating Commercial Scanners for Astronomical Images Robert J. Simcoe Associate Harvard College Observatory rjsimcoe@cfa.harvard.edu Introduction: Many organizations have expressed interest in using

More information

X-Ray-Based Medical Imaging and Resolution

X-Ray-Based Medical Imaging and Resolution Residents Section Physics Minimodule Huda and Abrahams Resolution on Radiographs Residents Section Physics Minimodule Residents inradiology Walter Huda 1 R. Brad Abrahams 2 Huda W, Abrahams RB Keywords:

More information

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Unit thickness. Unit area. σ = NΔX = ΔI / I 0 Unit thickness I 0 ΔI I σ = ΔI I 0 NΔX = ΔI / I 0 NΔX Unit area Δx Average probability of reaction with atom for the incident photons at unit area with the thickness of Delta-X Atom number at unit area

More information

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 1 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT RADT 3463 COMPUTERIZED IMAGING Section I: Chapter

More information

Image Quality Artifacts in Digital Imaging

Image Quality Artifacts in Digital Imaging MAHIDOL UNIVERSITY Wisdom of the Land Image Quality Artifacts in Digital Imaging Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University, Bangkok,

More information

Published text: Institute of Cancer Research Repository Please direct all s to:

Published text: Institute of Cancer Research Repository   Please direct all  s to: This is an author produced version of an article that appears in: MEDICAL PHYSICS The internet address for this paper is: https://publications.icr.ac.uk/1316/ Copyright information: http://www.aip.org/pubservs/web_posting_guidelines.html

More information

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage

More information

Dental Radiography. One of the problems of dental radiography is having different dimensions than normal.

Dental Radiography. One of the problems of dental radiography is having different dimensions than normal. The prototype receptor (the recording medium) most commonly used in dental radiography is the radiographic film. However, there are many other new more efficient receptors than the formed one that can

More information

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY WHITE PAPER: IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY IBEX Innovations Ltd. Registered in England and Wales: 07208355 Address: Discovery 2, NETPark, William Armstrong Way, Sedgefield, UK Patents:

More information

PD233: Design of Biomedical Devices and Systems

PD233: Design of Biomedical Devices and Systems PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/

More information

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images.

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images. DIGITAL RADIOGRAPHY Digital radiography is a film-less technology used to record radiographic images. 1 The purpose of digital imaging is to generate images that can be used in the diagnosis and assessment

More information

RADIOGRAPHY TERMS TO KNOW SELF STUDY DENTALELLE TUTORING

RADIOGRAPHY TERMS TO KNOW SELF STUDY DENTALELLE TUTORING RADIOGRAPHY TERMS TO KNOW SELF STUDY DENTALELLE TUTORING PLEASE NOTE You DO NOT need to study these for the board exam if this is why you bought our Radiography course, however if you come across any terms

More information

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC R. Bellazzini a,b, G. Spandre a*, A. Brez a, M. Minuti a, M. Pinchera a and P. Mozzo b a INFN Pisa

More information

A Comprehensive Review of Image Production

A Comprehensive Review of Image Production A Comprehensive Review of Image Production Presented by: John Fleming, M.Ed., RT(R)(MR)(CT) St. Petersburg College Office: (727) 341-3758 E-mail: flemingj@spcollege.edu Lesson Objectives: ARRT Content

More information

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES The current multiplication mechanism offered by dynodes makes photomultiplier tubes ideal for low-light-level measurement. As explained earlier, there

More information

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. RA110 test 3 Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. An object 35 cm in width is radiographed at 100 cm SID and at a 50 cm SOD. What

More information

Determining MTF with a Slant Edge Target ABSTRACT AND INTRODUCTION

Determining MTF with a Slant Edge Target ABSTRACT AND INTRODUCTION Determining MTF with a Slant Edge Target Douglas A. Kerr Issue 2 October 13, 2010 ABSTRACT AND INTRODUCTION The modulation transfer function (MTF) of a photographic lens tells us how effectively the lens

More information

Seminar 8. Radiology S8 1

Seminar 8. Radiology S8 1 Seminar 8 Radiology Medical imaging. X-ray image formation. Energizing and controlling the X-ray tube. Image detectors. The acquisition of analog and digital images. Digital image processing. Selected

More information

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT) Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in

More information

A Study On Preprocessing A Mammogram Image Using Adaptive Median Filter

A Study On Preprocessing A Mammogram Image Using Adaptive Median Filter A Study On Preprocessing A Mammogram Image Using Adaptive Median Filter Dr.K.Meenakshi Sundaram 1, D.Sasikala 2, P.Aarthi Rani 3 Associate Professor, Department of Computer Science, Erode Arts and Science

More information

IBEX MATERIALS DETECTION TECHNOLOGY

IBEX MATERIALS DETECTION TECHNOLOGY WHITE PAPER: IBEX MATERIALS DETECTION TECHNOLOGY IBEX Innovations Ltd. Registered in England and Wales: 07208355 Address: Discovery 2, NETPark, William Armstrong Way, Sedgefield, TS21 3FH, UK Patents held

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

ADVANCED MEDICAL SYSTEMS PTE LTD Singapore Malaysia India Australia

ADVANCED MEDICAL SYSTEMS PTE LTD Singapore Malaysia India Australia Innovative design is combined with cutting-edge technology to yield a definitive diagnosis and never before seen ergonomics GIOTTO CLASS is the result of 25 years of experience in the research and development

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Rotation By: Michael Case and Roy Grayzel, Acton Research Corporation Introduction The majority of modern spectrographs and scanning

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 16371-1 First edition 2011-10-01 Non-destructive testing Industrial computed radiography with storage phosphor imaging plates Part 1: Classification of systems Essais non destructifs

More information

On spatial resolution

On spatial resolution On spatial resolution Introduction How is spatial resolution defined? There are two main approaches in defining local spatial resolution. One method follows distinction criteria of pointlike objects (i.e.

More information

Mammography Solution. AMULET Innovality. The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available.

Mammography Solution. AMULET Innovality. The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available. Mammography Solution AMULET Innovality The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available. FUJIFILM supports the Pink Ribbon Campaign for early detection of

More information

Maximizing clinical outcomes

Maximizing clinical outcomes Maximizing clinical outcomes Digital Tomosynthesis Dual Energy Subtraction Automated Long Length Imaging Improved image quality at a low dose Xray Xray Patented ISS capture technology promotes high sensitivity

More information

Film Replacement in Radiographic Weld Inspection The New ISO Standard

Film Replacement in Radiographic Weld Inspection The New ISO Standard BAM Berlin Film Replacement in Radiographic Weld Inspection The New ISO Standard 17636-2 Uwe Ewert, Uwe Zscherpel, Mirko Jechow Requests and information to: uwez@bam.de 1 Outline - The 3 essential parameters

More information

Teaching Digital Radiography and Fluoroscopic Radiation Protection

Teaching Digital Radiography and Fluoroscopic Radiation Protection Teaching Digital Radiography and Fluoroscopic Radiation Protection WCEC 20 th Student Educator Radiographer Conference Dennis Bowman, RT(R), CRT (R)(F) Community Hospital of the Monterey Peninsula (CHOMP)

More information

Abstract. ZEIGLER, GARY BOYCE, II. Direct Detection of Microcalcification Pairs in Simulated

Abstract. ZEIGLER, GARY BOYCE, II. Direct Detection of Microcalcification Pairs in Simulated Abstract ZEIGLER, GARY BOYCE, II. Direct Detection of Microcalcification Pairs in Simulated Digital Mammograms. (Under the Direction of Professor Kuruvilla Verghese.) Using the MCMIS (Monte Carlo for Mammography

More information

Essentials of Digital Imaging

Essentials of Digital Imaging Essentials of Digital Imaging Module 2 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 2 Processing 1. ASRT Animation 2. Welcome Welcome to Essentials of Digital Imaging

More information

Image Enhancement in Spatial Domain

Image Enhancement in Spatial Domain Image Enhancement in Spatial Domain 2 Image enhancement is a process, rather a preprocessing step, through which an original image is made suitable for a specific application. The application scenarios

More information

Image Display and Perception

Image Display and Perception Image Display and Perception J. Anthony Seibert, Ph.D. Department of Radiology UC Davis Medical Center Sacramento, California, USA Image acquisition, display, & interpretation X-rays kvp mas Tube filtration

More information

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING DIGITAL IMAGE PROCESSING IN X-RAY IMAGING Shalini Kumari 1, Bachan Prasad 2,Aliya Nasim 3 Department of Electronics And Communication Engineering R.V.S College of Engineering & Technology, Jamshedpur,

More information

Digital Mammography Quality Control for the Mammographic Physicist

Digital Mammography Quality Control for the Mammographic Physicist Ontario Breast Screening Program Digital Mammography Quality Control for the Mammographic Physicist Authors: G.E. Mawdsley, A.K. Bloomquist, M.J. Yaffe March 2014 Revision 3.2 Mammographic Physics Consulting

More information

of sufficient quality and quantity

of sufficient quality and quantity of sufficient quality and quantity The patient s body attenuates the beam as it passes though the body More energy is deposited in organs located near the entry of the beam than near the exit of the beam

More information

GE Healthcare. Performa. High-performance breast imaging

GE Healthcare. Performa. High-performance breast imaging GE Healthcare Performa High-performance breast imaging Moving mammography forward. And patients faster. GE Healthcare s unparalleled leadership across mammography begins with a deep understanding of breast

More information

Digital radiography (DR) post processing techniques for pediatric radiology

Digital radiography (DR) post processing techniques for pediatric radiology Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions

More information

Charged Coupled Device (CCD) S.Vidhya

Charged Coupled Device (CCD) S.Vidhya Charged Coupled Device (CCD) S.Vidhya 02.04.2016 Sensor Physical phenomenon Sensor Measurement Output A sensor is a device that measures a physical quantity and converts it into a signal which can be read

More information

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal Yashvinder Sabharwal, 1 James Joubert 2 and Deepak Sharma 2 1. Solexis Advisors LLC, Austin, TX, USA 2. Photometrics

More information