Philips Site Yearly Performance Evaluation Philips Intera 1.5T 2-Mar-08. Table of Contents

Size: px
Start display at page:

Download "Philips Site Yearly Performance Evaluation Philips Intera 1.5T 2-Mar-08. Table of Contents"

Transcription

1 Philips Site Yearly Performance Evaluation Philips Intera.5T 2Mar8 Table of Contents Summary and Signature Page 2 Specific Comments Site Information 4 Equipment Information 4 Table Position ccuracy 4 Magnetic Field Homogeneity 4 Slice ccuracy 4 Slice Crosstalk 5 Soft Copy Displays 6 RF Coil Performance Evaluation Coil Inventory List 7 Body Integrated 8 Body Sense Coil 9 Breast Coil Head and eck Head Sense 2 Knee/Foot Quadrature 4 Shoulder rray Large 6 Shoulder rray Small 8 Spine Synergy 2 Surface C 2 Surface C 24 Surface C4 25 Surface E 26 Synergy FlexM 27 Wrist Coil High Res. ppendix : Magnet Homgeneity Map ppendix B: Slice / Profiles / RF Crosstalk ppendix C: CR Phantom nalysis ppendix D: Explanation of RF Coil Test Format

2 MRI Equipment Evaluation Summary & Signature Page Site ame: Philips Site MRP # ddress: Survey Date: City, State, Zip Report Date: MRI Mfg: Philips Model: Intera Field: /8/8 /7/8.5T MRI Scientist: Moriel essiver, Ph.D. Signature: Equipment Evaluation Tests Pass Fail * /. Magnetic field homogeneity: 2. Slice position accuracy:. Table positioning reproducibility: 4. Slice thickness accuracy: 5. RF coils' performance: a. Volume QD Coils b. Phase rray Coils c. Surface Coils 6. Interslice RF interference (Crosstalk): 7. Soft Copy Display 8. Hard Copy Display Evaluation of Site's Technologist QC Program Pass Fail * /. Set up and positioning accuracy: (daily) 2. Center frequency: (daily). Transmitter attenuation or gain: (daily) 4. Geometric accuracy measurments: (daily) 5. Spatial resolution measurements: (daily) 6. Low contrast detectability: (daily) 7. Head Coil (daily) 8. Body Coil (weekly) 9. Fast Spin Echo (F/T) ghosting levels: (daily). Film quality control: (weekly). Visual checklist: (weekly) *See comments page for description of any failures. Philips Site Philips Intera.5T 2

3 Specific Comments and Recommendations. The following warning was often seen: "Batteries of Magnet Emergency Rundown Unit are Low Contact Philips Service" Edwin says this is a known issue and the batteries are actually fine. How are we sure? 2.. The wrist coil has one totally dead channel and one weak channel. The film response is very good. The monitor's response curve is good in the middle but it dims a little too fast at the high end. The 5% is also rather difficult to see at the low end. 4. In general, the shim is very good. s can be seen in ppendix, there is a small region in the Inferior Right region that has slightly less homogeneity (still within spec). 5. Overall image quality looks good. O: Please be sure to read appendix D for an explanation of the new format of this document. Philips Site Philips Intera.5T

4 Site ame: Philips Site MRI Equipment Performance Evaluation Data Form Contact Title Owner Chief Tech. Phone Equipment Information MRI Manufacturer: Philips Model: Intera S: 54 Software:..4. Camera Manufacturer: Model: S: Software: PCS Manufacturer: Model: S: Software: CR Phantom umber used: Table Positioning Reproducibility: Table motion out/in: IsoCenter Measured Phantom Center.2 Comment: Out/In.8 Out/In Out/In.5.4 Pass 2. Magnetic Field Homogeneity See appendix for field plots. PSS xial: Coronal: Sagittal: Last Year CF: / This Year CF: 6929 CF Change: 5 cm 2 cm 25 cm GRE : 5, : & 5 Flip ngle: 45, : 4 mm skip mm, :.4KHz, x28, 2nex Comments: In general, the shim is very good. s can be seen in ppendix,.2.2. there is a small region in the Inferior Right region that has slightly less homogeneity (still within spec). Slice ccuracy : 25mm Matrix: x (Slice # from CR Phantom) ll values in mm (CR) (2/8) (2/8) Flip Calc Target % Error.6%.8% 2.8%.% T(5) % % Comments: Philips Site Philips Intera.5T 4

5 4. Slice Crosstalk (RF interference) The following data were obtained using the CR phantom slice thickness wedges to measure the slice profile of two T weighted sequences, a Spin Echo with either 69 pr 9 flip angle, when the slice gap varies from 2% down to % (contiguous) s the slices get closer together it is expected that the edges of the slices will overlap causing a deterioration of the slice profile. The data shown below clearly demonstrates this effect. Once the slice gap reaches % () of the slice thickness, the measured slice profile begins to drop. The main difference between the two sequences is the lower flip angle has a slightly smaller thickness as would be expected. ll of the slice profiles can be seen in ppendix B. (cm 2 ) Matrix # of slices Slice Measured x x Skip T Weighted Slice es (mm) Slice (mm) Philips Site Philips Intera.5T 5

6 5. Soft & Hard Copy Displays Luminance Meter Make/Model: Tektronix J6 Digital Photometer Monitor Description: LCD Luminance Measured: Ft. lamberts Cal Expires: 4/6/6 SMP Which Monitor Center of Image Display Top Left Corner Top Right Corner Bottom Left Corner Bottom Right Corner MX MI Percent Delta OK? Console Fair % delta =2% x (maxmin)/(max+center) (>% is action limit) imum Brightness must be > Ft. Lamberts The film response is very good. The monitor's response curve is good in the middle but it dims a little too fast at the high end. The 5% is also rather difficult to see at the low end. Density Ft Lamber ts Film from scanner Film from Camera. LCD & Film Response Curve Log FtLambert. Ideal Curve LCD Film Film % Density Philips Site Philips Intera.5T 6

7 Coil and Other Hardware Inventory List Site ame CR Magnet # Philips Site ickname chieva ctive Coil Description Manufacturer Model Rev. Mfg. Date S Channels Body Integrated QD Philips Jan, 24 Body Sense Coil Philips pr, Breast Coil Philips Mar, Head and eck Philips /86 Jun, /62 Head Quadrature Philips Mar, Head Sense Invivo Mar, Knee/Foot Quadrature Med. dvances Jan, Shoulder rray Large MRI Devices 227 Oct, 24 U Shoulder rray Small MRI Devices 227 Oct, 24 U Spine Synergy Mar, /S 5 Surface C Philips Feb, Surface C Philips Feb, Surface C4 Philips ov, 2 8 Surface E Philips May, /29 Syn. Multi Connect IGC Med. dvances Px Oct, /9 Syn. MultiConnect Philips Px Mar, 22 25/ Synergy FlexM Philips Mar, Synergy MultiConnect IGC Med. dvances Px Jan, /9 Wrist Coil High Res. MRI Devices 5 ug, 25 U

8 Coil: Body Integrated QD Mfg.: Philips Mfg. Date: //24 Coil ID: 562 Phantom: 2 cm GE sphere (used on pillows without table.) Test Date: /2/28 Model: S: # of Channels 2 T Coil Mode: BodyQD nalysis of Test Image % % Test Images Philips Site Intera 8

9 Coil: Body Sense Coil Mfg.: Philips Mfg. Date: 4//2 Coil ID: 5 Phantom: Philips Body disk Test Date: /2/28 Model: S: 974 # of Channels 4 2 T Coil Mode: Body Synergy nalysis of Composite Image clr, ,45,47, % 5.% 49.7% nalysis of Uncombined Images Ch ,27,45,9, % 68% % 89% % 69% 95% % The first composite image used CLER hence a greater signal uniformity. Composites Channels Channel Channel 2 Channel Channel 4 Philips Site Intera 9

10 Coil: Breast Coil Mfg.: Philips Mfg. Date: //24 Coil ID: 58 Phantom: 2 liter bottles with pad between them Test Date: /2/28 Model: S: 255 # of Channels 2 C Coil Mode: Breast (linear) nalysis of Test Image,228,24, %,26,, % Test Images Philips Site Intera

11 Coil: Head and eck Mfg.: Philips Mfg. Date: 6//22 Coil ID: 529 Phantom: CR Phantom in Head, liter bottle in neck Test Date: /2/28 Model: /86 S: /62 # of Channels 2 T Coil Mode: head head neck neck Head/eck HP ,8,8,62,64,64, nalysis of Composite Image % 79.9% 82.8% 82.7% nalysis of Uncombined Images Ch H P, ,,74, % 66% 46% % % 29% 46% % Composites Channels Channel Channel 2 Channel Channel 4 Philips Site Intera

12 Coil: Head Sense Mfg.: Invivo Mfg. Date: //2 Coil ID: 528 Phantom: CR Phantom Test Date: /2/28 Model: S: 8 # of Channels 6 2 T Coil Mode: Head Port # nalysis of Composite Image CL, ,57,9,4, % 68.8% 68.7% nalysis of Uncombined Images Ch ,87,8,5,77,28, % % 62% 67% 97% 6% % % 66% 7% 98% 66% This is normally an 8 channel coil, not a 6. channels 2 and 5 are actually 2 coils combined each. The results from Ports # & 2 are virtually identical. The first composite image used CLER hence a greater signal uniformity. Channel Channel 2 Channel Composites Channel 4 Channel 5 Channel 6 Philips Site Intera 2

13 Coil: Head Sense Mfg.: Invivo Mfg. Date: //2 Coil ID: 528 Phantom: CR Phantom Test Date: /2/28 Model: S: 8 # of Channels 6 2 T Coil Mode: Head Port #2 nalysis of Composite Image CL, ,5,26,26, % 69.2% 69.% nalysis of Uncombined Images Ch ,95,5,64,75,22, % 98% 6% 69% % 62% % 96% 67% 72% % 66% This is normally an 8 channel coil, not a 6. channels 2 and 5 are actually 2 coils combined each. The results from Ports # & 2 are virtually identical. Channel Channel 2 Channel Composites Channel 4 Channel 5 Channel 6 Philips Site Intera

14 Coil: Knee/Foot Quadrature Mfg.: Med. dvances Mfg. Date: //24 Coil ID: 526 Phantom: liter bottle in Knee, wrist phantom in foot Test Date: /2/28 Model: S: 878 # of Channels 2 T Coil Mode: Knee/Foot nalysis of Test Image,94, % 994, %,42,47, %,4,47, % Test Images Philips Site Intera 4

15 Coil: Knee/Foot Quadrature Mfg.: Med. dvances Mfg. Date: //24 Coil ID: 526 Phantom: liter bottle in Knee, wrist phantom in foot Test Date: /2/28 Model: S: 878 # of Channels 2 S Coil Mode: Knee/Foot PW nalysis of Test Image % %,224,8, %,224,4, % o phase wrap is used to eliminate wrap around of ghost into phantom. Test Images Philips Site Intera 5

16 Coil: Shoulder rray Large Mfg.: MRI Devices Mfg. Date: //24 Coil ID: 59 Phantom: Breast coil phantom Test Date: /2/28 Model: 227 S: U7964 # of Channels 4 2 T Coil Mode: SMCMRID 245 (Shoulder) nalysis of Composite Image ,7, % 52.% nalysis of Uncombined Images Ch ,5, % 6% % 92% % 44% 88% % Channel Channel 2 Composites Channel Channel 4 Philips Site Intera 6

17 Coil: Shoulder rray Large Mfg.: MRI Devices Mfg. Date: //24 Coil ID: 59 Phantom: Breast coil phantom Test Date: /2/28 Model: 227 S: U7964 # of Channels 4 2 S Coil Mode: SMCMRID 245 (Shoulder) nalysis of Composite Image 54 54,, % 29.% nalysis of Uncombined Images Ch ,94,2,29, % 75% 86% % % 58% 45% % Channel Channel 2 Composites Channel Channel 4 Philips Site Intera 7

18 Coil: Shoulder rray Small Mfg.: MRI Devices Mfg. Date: //24 Coil ID: 56 Phantom: Breast coil phantom Test Date: /2/28 Model: 227 S: U7999 # of Channels 4 2 T Coil Mode: SMCMRID 245 (Shoulder) nalysis of Composite Image % 46.2% nalysis of Uncombined Images Ch ,,9 87, % 42% % 68% % 8% 92% 6% Channel Channel 2 Composites Channel Channel 4 Philips Site Intera 8

19 Coil: Shoulder rray Small Mfg.: MRI Devices Mfg. Date: //24 Coil ID: 56 Phantom: Breast coil phantom Test Date: /2/28 Model: 227 S: U7999 # of Channels 4 2 S Coil Mode: SMCMRID 245 (Shoulder) nalysis of Composite Image % 7.8% nalysis of Uncombined Images Ch ,2 25 9,47, % 74% % 84% % 6% 55% % Channel #2 is a little lower than normal for this coil... but.. it could be related to phantom position. I don t think it is significant enough of a difference to merit action at this time. Channel Channel 2 Composites Channel Channel 4 Philips Site Intera 9

20 Coil: Spine Synergy Mfg.: Mfg. Date: //998 Coil ID: 525 Phantom: Philips Body Disk Test Date: /2/28 Model: S: 2667/S # of Channels 5 2 S Coil Mode: Spine Coil &2 nalysis of Composite Image % 27.8% nalysis of Uncombined Images Ch % 9% % 94% Composites Channel Channel 2 Philips Site Intera 2

21 Coil: Spine Synergy Mfg.: Mfg. Date: //998 Coil ID: 525 Phantom: Philips Body Disk Test Date: /2/28 Model: S: 2667/S # of Channels 5 2 S Coil Mode: Spine Coil 2& nalysis of Composite Image ,62, % 2.9% nalysis of Uncombined Images Ch 2 2 4,7, % % % % Composites Channel Channel 2 Philips Site Intera 2

22 Coil: Spine Synergy Mfg.: Mfg. Date: //998 Coil ID: 525 Phantom: Philips Body Disk Test Date: /2/28 Model: S: 2667/S # of Channels 5 2 S Coil Mode: Spine Coil 4&5 nalysis of Composite Image % 6.4% nalysis of Uncombined Images Ch , % % % % Composites Channel Channel 2 Philips Site Intera 22

23 Coil: Surface C Mfg.: Philips Mfg. Date: 2//24 Coil ID: 5 Phantom: PIQT Test Date: /2/28 Model: S: 756 # of Channels 2 C Coil Mode: C Ports & 2 nalysis of Test Image P,8, % P,8, % P2,2, % P2,7, % Test Images Philips Site Intera 2

24 Coil: Surface C Mfg.: Philips Mfg. Date: 2//24 Coil ID: 5 Phantom: PIQT Test Date: /2/28 Model: S: 62 # of Channels 2 C Coil Mode: C nalysis of Test Image,4, % 995, % Test Images Philips Site Intera 24

25 Coil: Surface C4 Mfg.: Philips Mfg. Date: //2 Coil ID: 52 Phantom: PIQT Test Date: /2/28 Model: S: 8 # of Channels 2 T Coil Mode: C4 nalysis of Test Image 845, % 88, % Test Images Philips Site Intera 25

26 Coil: Surface E Mfg.: Philips Mfg. Date: 5//2 Coil ID: 54 Phantom: liter bottle Test Date: /2/28 Model: S: 4574/29 # of Channels 2 T Coil Mode: E nalysis of Test Image 874, % 868, % Test Images Philips Site Intera 26

27 Coil: Synergy FlexM Mfg.: Philips Mfg. Date: //24 Coil ID: 55 Phantom: 5 liter bottle Test Date: /2/28 Model: S: 57 # of Channels 2 2 T Coil Mode: Synergy Flex M nalysis of Composite Image % 76.6% nalysis of Uncombined Images Ch % 97% % 88% Composites Channel Channel 2 Philips Site Intera 27

28 Coil: Synergy FlexM Mfg.: Philips Mfg. Date: //24 Coil ID: 55 Phantom: 5 liter bottle Test Date: /2/28 Model: S: 57 # of Channels 2 2 S Coil Mode: Synergy Flex M nalysis of Composite Image % 54.% nalysis of Uncombined Images Ch 2 4 5, % 95% % % Composites Channel Channel 2 Philips Site Intera 28

29 Coil: Wrist Coil High Res. Mfg.: MRI Devices Mfg. Date: 8//25 Coil ID: 54 Phantom: Wrist Phantom Test Date: /2/28 Model: 5 S: U2426 # of Channels 4 2 T Coil Mode: Wrist nalysis of Composite Image,294,294,98,97,68, % 9.2% nalysis of Uncombined Images Ch % % 5% 22% % % 52% 24% Channel 2 is dead, channel 4 is very sick. Channel Channel 2 Composites Channel Channel 4 Philips Site Intera 29

30 Coil: Wrist Coil High Res. Mfg.: MRI Devices Mfg. Date: 8//25 Coil ID: 54 Phantom: Wrist Phantom Test Date: /2/28 Model: 5 S: U2426 # of Channels 4 2 C Coil Mode: Wrist nalysis of Composite Image 92 92,6, % 4.7% nalysis of Uncombined Images Ch ,497, % % 8% 2% % % % 8% Channel 2 is dead, channel 4 is very sick. Channel Channel 2 Composites Channel Channel 4 Philips Site Intera

31 Coil: Wrist Coil High Res. Mfg.: MRI Devices Mfg. Date: 8//25 Coil ID: 54 Phantom: Wrist Phantom Test Date: /2/28 Model: 5 S: U2426 # of Channels 4 2 S Coil Mode: Wrist nalysis of Composite Image ,4, % 7.5% nalysis of Uncombined Images Ch ,, % % 9% 2% % % % 8% Channel 2 is dead, channel 4 is very sick. Channel Channel 2 Composites Channel Channel 4 Philips Site Intera

32 ppendix : Magnet Homogeneity Field Maps Philips Intera.5T central planes Measured March 2, 28 xial Right nterior 2 Left xial DIMER MI MX RGE PPM ME STDEV Posterior Coronal Right Superior 2 Left Coronal DIMER MI MX RGE PPM ME STDEV Inferior Sagittal nteriror Superior Posterior Sagittal DIMER MI MX RGE PPM ME STDEV Inferior

33 ppendix : Magnet Homogeneity Field Maps Philips Intera.5T Measured March 2, 28 Diameter cm 5cm 2cm 25cm 28cm cm Philips Site Intera xial Field Plot /2/ ppm cm from Isocenter (S/I) Diameter cm 5cm 2cm 25cm 28cm cm Philips Site Intera Coronal Field Plot /2/ ppm cm from Isocenter (P/) Diameter cm 5cm 2cm 25cm 28cm cm Philips Site Intera Coronal Field Plot /2/ ppm cm from Isocenter (R/L)

34 H9 2 H8 2 H7 2 2 H6 2 2 H5 2 2 H4 2 2 H 2 2 H H Isocenter F F2 2 2 F 2 2 F4 2 F5 F6 F7 F8 F9 xial Field Plots

35 Isocenter P 2 P2 2 P 2 2 P4 2 P5 2 P6 2 P7 2 P8 2 P9 Coronal Field Plots

36 2 2 L L L L L L4 2 2 L 2 2 L2 2 L 2 Isocenter 2 R 2 2 R2 2 2 R 2 2 R4 2 2 R5 2 2 R6 2 2 R7 2 R8 2 2 R9 Sagittal Field Plots

37 ppendix B: RF Slice Profiles and Crosstalk Spin Echo : Flip = 69 / = 4/5 = 4.4 KHz nex = 2 Scan time: :25 RF Xtalk 69 sk Upper= Lower=48.6 RF Xtalk 69 sk 5. Upper=48.67 Lower= RF Xtalk 69 sk 2.5 Upper=48.6 Lower= Slice = RF Xtalk 69 sk 2. Upper=48.72 Lower=48. Slice = RF Xtalk 69 sk.5 Upper=48.7 Lower=48.44 Slice = RF Xtalk 69 sk. Upper=47.9 Lower= Slice = RF Xtalk 69 sk.5 Upper=46.9 Lower=45.82 Slice = RF Xtalk 69 sk.2 Upper=44.66 Lower=44.2 Slice = RF Xtalk 69 sk. Upper=4.7 Lower= Slice = Slice Slice = Slice =4.4 Slice thickness as a function of slice gap Slice

38 ppendix B: RF Slice Profiles and Crosstalk Spin Echo : Flip = 69 / = 4/5 = 4.4 KHz nex = 2 Scan time: :25 RF Xtalk 9 sk Upper=5.7 8 Lower=48.94 RF Xtalk 9 sk 5. Upper=49.59 Lower= RF Xtalk 9 sk 2.5 Upper=49.78 Lower= Slice = RF Xtalk 9 sk 2. Upper=49. Lower= Slice = RF Xtalk 9 sk.5 Upper=49.4 Lower= Slice = RF Xtalk 9 sk. Upper=47.88 Lower= Slice = RF Xtalk 9 sk.5 Upper=46.6 Lower= Slice = RF Xtalk 9 sk.2 Upper=44.8 Lower= Slice = RF Xtalk 9 sk. Upper=44.8 Lower= Slice = Slice = Slice = Slice Slice thickness as a function of slice gap Slice

39 Philips Site Coil Used: Head Sense Sagittal Locator Length of phantom, end to end (mn 48± 2) 2 Slice Location # Resolution (.,.,.9 mm) Slice (fwhm in mm) Wedge (mm) Diameter (mm) (9±2) Slice Location #5 Diameter (mm) (9±2) Top Bottom Slice Location # Signal (mean only) Big ROI High Low 7 (>87.5%) (mean ±std dev) Top Bottom Left Right 22 2 Ghosting Ratio (<2.5%) (no spec) Calculated value 5.±.7 Low Con Detectability Slice Location #8 Slice Location #9 Slice Location # Slice Location # Total # of Spokes (>=9) Slice Location # Wedge (mm) Slice Position Error = + =.4% 2.5%.6% 5.% = + = 46.6 chieva Test Date: /2/28 = calculated field ( 5/2) ( 2/2) ( 2/8) (Site T) (Site T2) CR T CR PD CR T2 Site T Site T % 9.2% 9.% 9.2% 92.5% 4.7 ± ± ± ± 5.8. ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± 5.2.%.2%.%.2%.%

40 Philips Site chieva parameters Test Date: /2/28 Coil Used: Head Sense Test ID Study Descrip tion Pulse (ETL) (ms) (ms) (cm) Phase Sample Ratio umber of Slices (mm) Slice (ex) Freq Matrix Phase Matrix Band Width (khz) Scan Time (min:sec) CR T :9 CR PD Dual Echo :2 CR T2 Dual Echo :2 Site T :5 Site T2 T(5) : Magnet ID: 95 Coil ID: 528 TestID:

41 ppendix C: CR Phantom nalysis CR T Sagital Length High Contrast Resolution Slice Slice Position Inferior Upper=5.98 Lower=49.7 Slice = Diff.=.49 xial Diameters # xial Diameters #5 & Ghosting #7 Slice Position Superior : :5 : PIU: 92.4% S.D S.D S.D S.D Diff.=.2 /2/8 : 499 : 2. Low Contrast #8 Low Contrast #9 Low Contrast # Low Contrast #

42 ppendix C: CR Phantom nalysis CR PD High Contrast Resolution Slice Slice Position Inferior Upper=5.7 Lower=49.47 Slice =5.4 Diff.=.56 xial Diameters # xial Diameters #5 & Ghosting #7 Slice Position Superior : :529 :4 PIU: 9.2% S.D S.D S.D S.D. 4.4 Diff.=.6 /2/8 :2 : 2. Low Contrast #8 Low Contrast #9 Low Contrast # Low Contrast #

43 ppendix C: CR Phantom nalysis CR T2 High Contrast Resolution Slice Slice Position Inferior Upper=5. Lower=47.2 Slice =4.86 Diff.=.52 xial Diameters # xial Diameters #5 & Ghosting #7 Slice Position Superior : :9 :794 PIU: 9.% S.D S.D S.D S.D Diff.=.4 /2/8 :2 : 8. Low Contrast #8 Low Contrast #9 Low Contrast # Low Contrast #

44 ppendix C: CR Phantom nalysis Site T High Contrast Resolution Slice Slice Position Inferior Upper=5.4 Lower=49.5 Slice =5. Diff.=.48 xial Diameters # xial Diameters #5 & Ghosting #7 Slice Position Superior : :547 :5 PIU: 9.2% S.D S.D S.D S.D Diff.=.2 /2/8 : 45 : 5. Low Contrast #8 Low Contrast #9 Low Contrast # Low Contrast #

45 ppendix C: CR Phantom nalysis Site T2 High Contrast Resolution Slice Slice Position Inferior Upper=54.97 Lower=52.9 Slice =5.9 Diff.=.64 xial Diameters # xial Diameters #5 & Ghosting #7 Slice Position Superior : :89 :769 PIU: 92.5% S.D S.D S.D S.D. 2.6 Diff.=.4 /2/8 :4 :. Low Contrast #8 Low Contrast #9 Low Contrast # Low Contrast #

46 ppendix D: Explanation of RF Coil Testing Report Introduction The primary goal of RF coil testing is to establish some sort of base line for tracking coil performance over time. The most common measure is the Signal to Ratio or. In addition, we can look at overall signal uniformity, ghosting level (or better lack of ghosting) and in the case of phased array coils we look at the of each and every channel and at symmetry between channels. Unfortunately, there is no single best method for measuring. Below I explain the different methods used and the rationale for each. One needs to measure the signal in the phantom (either mean or peak or both) and then divide that by the back noise. Measuring the signal is fairly straightforward, the noise can be more problematic. The simplest method is to measure the standard deviation () in the back air. However, MRI images are the magnitude of complex data. The noise in the underlying complex data is Gaussian but it follows a Rician distribution when the magnitude is used. The true noise can be estimated by multiplying the measured by.526. During the reconstruction process, most manufacturers perform various additional operations on the images, This could include geometric distortion correction, low pass filtering of the kspace data resulting in low signal at the edge of the images, RF coil intensity correction (PURE, CLER, SCIC, etc), and other processing during the combination of phased array data and parallel imaging techniques. ll of these methods distort the back noise making it impossible to obtain an accurate (and reproducible) estimate of the image noise in the air region. The alternative is to use a method which I shall refer to as the (ational Electrical Manufacturers ssociation) method. The signal in the phantom area is a sum of the proton signal and noise. Once the signal to noise ratio exceeds 5:, the noise in the magnitude image is effectively Gaussian. To eliminate the proton signal, you acquire an image twice and subtract them. The measured in the phantom region should now be the true times the square root of 2. When determining the using the method, calculate the mean signal of the average of the two source images then divide by.77 x the measured in the same area as the mean signal. Unfortunately, this doesn t always work. It is absolutely imperative that the RF channel scalings, both transmit and receive, be identical with both scans. ny ghosting in the system is not likely to repeat exactly for both scans and will cause a much higher. Finally, the phantom needs to be resting in place prior to the scan long enough for motion of the fluid to have died down. Depending on the size and shape of the phantom, this could take any where from 5 to 2 minutes. One of the most common causes of ghosting is vibration from the helium coldhead. The best way to eliminate this artifact is to turn off the cold head, which will increase helium consumption. Because this vibration is periodic, the ghosting is usually of an over 2 (/2) nature. The affect inside the signal region of the phantom can be minimized by using a that is twice the diameter of the phantom (measured in the PE direction.) If the noise is to be measured in the air, then be sure to OT make measurements to either side of the phantom in the PE direction. Scan parameters also significantly affect measured. For most of the testing performed in this document I used a simple Spin Echo with a of, a of 2 and a slice thickness of mm and a receiver of 28. KHz (a pixel fat/water chemical shift). The was varied depending on the size of the coil and the phantom used. ll of the parameters used for each test can be found on each page immediately below the coil description.

47 Report Layout Each page of this report lists the data from a single test. The top third of the page describes the coil and phantom information, followed by the scan parameters used. The middle third contains the numbers measured and calculated results. This section will contain one table if the coil being tested is a single channel coil (i.e. quadrature or surface coils) and two tables if it is a multichannel phased array coil. The entries in the table will be described further below. The bottom section contains a few lines of comments (if necessary), a picture of the coil with the phantom as used for the testing and one or more of the images that were used for the measurements. There is usually one image for each composite image measurement and one image for each separate channel measurement. Each image shows the ROI (red line) where the mean signal was measured and two smaller ROIs (green lines) where the signal minimum and maximum was found. In the top left corner of each image is the mean signal in the large ROI. The bottom left corner contains the large ROI s area (in mm 2 ). The top right corner contains two numbers a mean and a standard deviation. If the method was used, then the top right corner will list the mean and of the large ROI (labeled ROI M and ROIsd) applied to the subtraction image. If the noise was measured in the back air the the numbers are labeled M and. Data Tables The meaning of most of the entries in the data table are should be self evident with a few exceptions. The first column in each table is labeled. In the composite analysis, this field may be empty or contain some sort of abbreviation to identify some aspect of the testing. Some possibilities are the letter for, for, L for Left, R for Right, C for CLER, oc for o CLER. In the Uncombined Image table, the label usually contains the channel number or similar descriptor. The column labeled will be either or SubSig which stands for Subtracted Signal, i.e. the method. Both tables contain a column for and which are the or signal divided by the of the noise scaled by either.526 () or.77 (). Composite Image Table: The final two columns in this table are and. It can be rather difficult to compare the performance of different coils particularly if different scan parameters are used. (Of course, it s even more difficult from one scanner to another.) I have standardized most of my testing to use a spin echo with a / of /2msec and a thickness of mm. The changes to depending on the size of the phantom used although I try to use a that is at least twice the diameter of the phantom as measured in the PE direction. For one reason or another, a change may be made in the scan parameters (either accidentally or intentionally such as turning on o Phase Wrap to eliminate aliasing, etc.). In order to make it easier to compare values I calculate a value. This value is theoretically what the would be if a of cm, x matrix, average, receiver of 5.6 KHz and slice thickness of mm had been used. Obviously, the final number is affected by the T/T2 values of the phantoms used as well as details of the coil and magnet field strength but it can be useful in certain situations. The value is defined by the CR as (maxmin)/(max+min). This is most important when looking at volume coils or for evaluating the effectiveness of surface coil intensity correction algorithms (such as PURE, CLER or SCIC). Uncombined Image Table: This table has two columns labeled and. When analyzing multichannel coils it is important to understand the relationship between the different channels, the inherent symmetry that usually exists between channels. In a 8 channel head or 4 channel torso phased array coil, all of the channels are usually have about the same. These two columns list how the (either or ) of each channel compares to the of the channel with the maximum value.

GE Site Yearly Performance Evaluation GE Signa Excite HD - 3T 1-Sep-08. Table of Contents

GE Site Yearly Performance Evaluation GE Signa Excite HD - 3T 1-Sep-08. Table of Contents GE Site Yearly Performance Evaluation GE Signa Excite HD T Sep8 Table of Contents Summary and Signature Page 2 Specific Comments Site Information 4 Equipment Information 4 Table Position ccuracy 4 Magnetic

More information

Philips Site Yearly Performance Evaluation Philips Openview 16-Jan-08. Table of Contents

Philips Site Yearly Performance Evaluation Philips Openview 16-Jan-08. Table of Contents Philips Site Yearly Performance Evaluation Philips Openview 6-Jan-8 Table of Contents Summary and Signature Page 2 Specific Comments 3 Site Information 4 Equipment Information 4 Table Position Accuracy

More information

BACKGROUND. ** 78% of all MRI scanners have Image Quality problems. *** *** 25% of all Multi-Channel RF coils have at least one bad channel.

BACKGROUND. ** 78% of all MRI scanners have Image Quality problems. *** *** 25% of all Multi-Channel RF coils have at least one bad channel. Range of Results from over 534 ACR-mandated Annual MRI Performance Evaluations on over 204 Magnets from 8 Vendors Spanning a 10-year Period Moriel NessAiver, Ph.D. - Simply Physics - Baltimore, MD moriel@simplyphysics.com

More information

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 ACR MRI accreditation program MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics,

More information

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Advanced MSK MRI Protocols at 3.0T Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Outline Why High Field for MSK? SNR and Relaxation Times Technical Issues Example

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

EE469B: Assignment 2 Solutions

EE469B: Assignment 2 Solutions EE469B Fall 26-7 RF Pulse Design for MRI EE469B: Assignment 2 s Due Thursday Oct 3 Introduction This assignment concerns the design of small-tip-angle 2D excitation pulses based on spiral k-space trajectories.

More information

1.5T HIGH FIELD SMALL ANIMAL MRI

1.5T HIGH FIELD SMALL ANIMAL MRI 1.5T HIGH FIELD SMALL ANIMAL MRI Designed Specifically for Veterinarians TECHNICAL GUIDE ADVANCING THE ART AND SCIENCE OF VETERINARY MRI The PetVet is the only high-field MRI system designed specifically

More information

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging JOURNAL OF MAGNETIC RESONANCE IMAGING 20:1046 1051 (2004) Technical Note The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging James W. Goldfarb, PhD* Purpose: To describe a known (but undocumented)

More information

MRI SYSTEM COMPONENTS Module One

MRI SYSTEM COMPONENTS Module One MRI SYSTEM COMPONENTS Module One 1 MAIN COMPONENTS Magnet Gradient Coils RF Coils Host Computer / Electronic Support System Operator Console and Display Systems 2 3 4 5 Magnet Components 6 The magnet The

More information

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev. 1.0 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE DoseLab users may reference the following instructions to perform Tomotherapy Quality Assurance tests as recommended

More information

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves Experiment 1 Goal: Examine the effect caused by changing flip angle on image contrast in a simple gradient echo sequence and derive T1-curves. Image

More information

2014 M.S. Cohen all rights reserved

2014 M.S. Cohen all rights reserved 2014 M.S. Cohen all rights reserved mscohen@g.ucla.edu IMAGE QUALITY / ARTIFACTS SYRINGOMYELIA Source http://gait.aidi.udel.edu/res695/homepage/pd_ortho/educate/clincase/syrsco.htm Surgery is usually recommended

More information

k y 2k y,max k x 2k x,max

k y 2k y,max k x 2k x,max EE225E/BIOE265 Spring 2012 Principles of MRI Miki Lustig Assignment 5 Due Feb 26, 2012 1. Finish reading Nishimura Ch. 5. 2. For the 16 turn spiral trajectory, plotted below, what is the a) Spatial resolution,

More information

HETERONUCLEAR IMAGING. Topics to be Discussed:

HETERONUCLEAR IMAGING. Topics to be Discussed: HETERONUCLEAR IMAGING BioE-594 Advanced MRI By:- Rajitha Mullapudi 04/06/2006 Topics to be Discussed: What is heteronuclear imaging. Comparing the hardware of MRI and heteronuclear imaging. Clinical applications

More information

2015 Spin echoes and projection imaging

2015 Spin echoes and projection imaging 1. Spin Echoes 1.1 Find f0, transmit amplitudes, and shim settings In order to acquire spin echoes, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week,

More information

Hardware. MRI System. MRI system Multicoil Microstrip. Part1

Hardware. MRI System. MRI system Multicoil Microstrip. Part1 Hardware MRI system Multicoil Microstrip MRI System Part1 1 The MRI system is made up of a variety of subsystems. the Operator Workspace Gradient Driver subsystem The Physiological Acquisition Controller

More information

NEMA Standards Publication MS (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging

NEMA Standards Publication MS (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging NEMA Standards Publication MS 1-2008 (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging Published by: National Electrical Manufacturers Association 1300 North

More information

Page 1 of 9. Protocol: adult_other_adni3_study_human_ge_3t_25w_ _ _1. 3 Plane Localizer. 3 Plane Localizer PATIENT POSITION

Page 1 of 9. Protocol: adult_other_adni3_study_human_ge_3t_25w_ _ _1. 3 Plane Localizer. 3 Plane Localizer PATIENT POSITION 3 Localizer FOV 26.0 Slice Thickness 5.0 Slice Spacing 0.0 Freq 256 Phase 128 3-PLANE 3 Localizer Unswap Phase Correction Gradient Echo Imaging Options Seq, Fast Recon All Images Contrast Yes/ 3 Localizer

More information

Image Quality/Artifacts Frequency (MHz)

Image Quality/Artifacts Frequency (MHz) The Larmor Relation 84 Image Quality/Artifacts (MHz) 42 ω = γ X B = 2πf 84 0.0 1.0 2.0 Magnetic Field (Tesla) 1 A 1D Image Magnetic Field Gradients Magnet Field Strength Field Strength / Gradient Coil

More information

Pulse Sequence Design Made Easier

Pulse Sequence Design Made Easier Pulse Sequence Design Made Easier Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant gurumri@gmail.com 1 2 Pulse Sequences generally have the following characteristics: An RF line characterizing RF Pulse applications

More information

Pulse Sequence Design and Image Procedures

Pulse Sequence Design and Image Procedures Pulse Sequence Design and Image Procedures 1 Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant 2 A pulse sequence is a timing diagram designed with a series of RF pulses, gradients switching, and signal readout

More information

Background (~EE369B)

Background (~EE369B) Background (~EE369B) Magnetic Resonance Imaging D. Nishimura Overview of NMR Hardware Image formation and k-space Excitation k-space Signals and contrast Signal-to-Noise Ratio (SNR) Pulse Sequences 13

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

The ACR magnetic resonance accreditation phantom (ACR MRAP) has been designed to examine a broad range of instrument parameters.

The ACR magnetic resonance accreditation phantom (ACR MRAP) has been designed to examine a broad range of instrument parameters. OVERVIEW OF THE ACR MRI ACCREDITATION PHANTOM Geoffrey D. Clarke, Ph.D. University of Texas Southwestern Medical Center at Dallas email: GEOFFREY.CLARKE@MAIL.SWMED.EDU INTRODUCTION The ACR magnetic resonance

More information

Welcome at MR Achieva 3.0T TX. About the system. Index. 1.1 Introduction

Welcome at MR Achieva 3.0T TX. About the system. Index. 1.1 Introduction 1 Welcome at MR Achieva 3.0T TX Welcome to the Introduction of and e-learning module for the Achieva 3 Tesla TX-system. This CBT is setup for those engineers who will soon receive a Achieva 3Tesla TX system.

More information

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Huaiqun Guan,

More information

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010 (N)MR Imaging Lab Course Script FMP PhD Autumn School Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder Date: November 3rd, 2010 1 Purpose: Understanding the basic principles of MR imaging

More information

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements MR Data for Treatment Planning: Issues, Protocol Optimization, and s (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements TG-117 Use of MRI Data in Treatment Planning

More information

MR Basics: Module 8 Image Quality

MR Basics: Module 8 Image Quality Module 8 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

6.S02 MRI Lab Acquire MR signals. 2.1 Free Induction decay (FID)

6.S02 MRI Lab Acquire MR signals. 2.1 Free Induction decay (FID) 6.S02 MRI Lab 1 2. Acquire MR signals Connecting to the scanner Connect to VMware on the Lab Macs. Download and extract the following zip file in the MRI Lab dropbox folder: https://www.dropbox.com/s/ga8ga4a0sxwe62e/mit_download.zip

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

SIGNA Explorer Lift revives our MR

SIGNA Explorer Lift revives our MR Seiji Shiotani, MD, PhD Seirei Fuji Hospital in Fuji City, Shizuoka, Japan Masayoshi Sugimura Seirei Fuji Hospital in Fuji City, Shizuoka, Japan SIGN Explorer Lift revives our MR The clinical usefulness

More information

Digital Imaging CT & MR

Digital Imaging CT & MR Digital Imaging CT & MR January 22, 2008 Digital Radiography, CT and MRI generate images in a digital format What is a Digital Image? A digital image is made up of picture elements, pixels row by column

More information

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Magnetic Resonance Imaging Spatial

More information

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter MR-guided RT in Heidelberg Funded by the German Research Foundation (DFG) Heidelberg consortium received

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

Weber State University Radiologic Technology 4603

Weber State University Radiologic Technology 4603 Weber State University Radiologic Technology 4603 MRI Physics and Instrumentation Instructor: Rex T. Christensen MHA R.T. (R) (MR) (CT) (ARRT) CIIP Contact Info: E-mail: rexchristensen@weber.edu Phone:

More information

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

1. Patient size AEC. Large Patient High ma. Small Patient Low ma Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description

More information

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Melissa Martin, M.S., FAAPM, FACR, FACMP AAPM Annual Meeting - Philadelphia July 19, 2010 MO-B-204C-1 Educational Objectives

More information

Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging

Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging Poster No.: C-0709 Congress: ECR 2014 Type: Scientific

More information

EE469B: Assignment 1 Solutions

EE469B: Assignment 1 Solutions EE469B Fall 26-7 RF Pulse Design for MRI EE469B: Assignment Solutions Due Thursday Oct 6 Introduction This assignment concerns typical Fourier transform designs of excitation pulses. This includes designing

More information

FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR

FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May 2010 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics, Inc. Physicist's Role in the

More information

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare GE Healthcare Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare There is excitement across the industry regarding the clinical potential of a hybrid

More information

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

Evaluation of MWT Materials Accusorb MRI Shield. Dr. E. Kanal, Department of Radiology, UPMC. Wednesday, April 21, 2010

Evaluation of MWT Materials Accusorb MRI Shield. Dr. E. Kanal, Department of Radiology, UPMC. Wednesday, April 21, 2010 Evaluation of MWT Materials Accusorb MRI Shield Dr. E. Kanal, Department of Radiology, UPMC Wednesday, April 21, 2010 Background: In almost all x-ray and ultrasonographic imaging examinations, in order

More information

SIEMENS MAGNETOM Skyra syngo MR D13

SIEMENS MAGNETOM Skyra syngo MR D13 Page 1 of 12 SIEMENS MAGNETOM Skyra syngo MR D13 \\USER\CIND\StudyProtocols\PTSA\*ep2d_M0Map_p2_TE15 TA:7.9 s PAT:2 Voxel size:2.5 2.5 3.0 mm Rel. SNR:1.00 :epfid Properties Routine Contrast Prio Recon

More information

TimTX TrueShape. The parallel transmit architecture of the future. Answers for life.

TimTX TrueShape.  The parallel transmit architecture of the future. Answers for life. www.siemens.com/trueshape TimTX TrueShape The parallel transmit architecture of the future. The product/feature (mentioned herein) is not commercially available. Due to regulatory reasons its future availability

More information

Bar code Verifier Conformance Specifications. Using the INTEGRA-9000

Bar code Verifier Conformance Specifications. Using the INTEGRA-9000 Bar code Verifier Conformance Specifications Using the INTEGRA-9000 From: Label Vision Systems, Inc. (LVS) Document Created: 4-1998 Edit / Print Date: 2-2003 C:\My Documents\INTEGRA -9000 VERIFIER CONFORMANCE

More information

Magnetic Resonance Research Facility (MRRF) Resources

Magnetic Resonance Research Facility (MRRF) Resources Magnetic Resonance Research Facility (MRRF) Resources The Magnetic Resonance Research Facility (MRRF) has scanners located in both the hospital and research buildings on the campus of the University of

More information

H 2 O and fat imaging

H 2 O and fat imaging H 2 O and fat imaging Xu Feng Outline Introduction benefit from the separation of water and fat imaging Chemical Shift definition of chemical shift origin of chemical shift equations of chemical shift

More information

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP MR Data for Treatment Planning and Stereotactic Procedures: Sources of Distortion, Protocol Optimization, and Assessment (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements

More information

Tomophan TSP004 Manual

Tomophan TSP004 Manual T h e P h a n t o m L a b o r a t o r y 1 Tomophan TSP004 Manual Copyright 2016 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order and

More information

Methods. Experimental Stimuli: We selected 24 animals, 24 tools, and 24

Methods. Experimental Stimuli: We selected 24 animals, 24 tools, and 24 Methods Experimental Stimuli: We selected 24 animals, 24 tools, and 24 nonmanipulable object concepts following the criteria described in a previous study. For each item, a black and white grayscale photo

More information

Accurate Utility Depth Measurements Using the Spar 300

Accurate Utility Depth Measurements Using the Spar 300 Accurate Utility Depth Measurements Using the Spar 3 This Application Note addresses how to obtain accurate subsurface utility depths using the model-based methods employed by the Spar 3. All electromagnetic

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Experience in implementing continuous arterial spin labeling on a commercial MR scanner

Experience in implementing continuous arterial spin labeling on a commercial MR scanner JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 1, WINTER 2005 Experience in implementing continuous arterial spin labeling on a commercial MR scanner Theodore R. Steger and Edward F. Jackson

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

Encoding of inductively measured k-space trajectories in MR raw data

Encoding of inductively measured k-space trajectories in MR raw data Downloaded from orbit.dtu.dk on: Apr 10, 2018 Encoding of inductively measured k-space trajectories in MR raw data Pedersen, Jan Ole; Hanson, Christian G.; Xue, Rong; Hanson, Lars G. Publication date:

More information

Using Frequency Diversity to Improve Measurement Speed Roger Dygert MI Technologies, 1125 Satellite Blvd., Suite 100 Suwanee, GA 30024

Using Frequency Diversity to Improve Measurement Speed Roger Dygert MI Technologies, 1125 Satellite Blvd., Suite 100 Suwanee, GA 30024 Using Frequency Diversity to Improve Measurement Speed Roger Dygert MI Technologies, 1125 Satellite Blvd., Suite 1 Suwanee, GA 324 ABSTRACT Conventional antenna measurement systems use a multiplexer or

More information

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey Implementation of the 2016 ACR Digital Mammography QC Manual Medical Physicist Mammography Equipment Evaluation and Annual Survey Eric A Berns, PhD, FACR Radiologist Responsibilities Radiologist Responsibilities

More information

PREPARED BY: I. Miller DATE: 2004 May 23 CO-OWNERS REVISED DATE OF ISSUE/CHANGED PAGES

PREPARED BY: I. Miller DATE: 2004 May 23 CO-OWNERS REVISED DATE OF ISSUE/CHANGED PAGES Page 1 of 30 LIGHTMACHINERY TEST REPORT LQT 30.11-2 TITLE: HMI Michelson Interferometer Test Report Serial Number 2 - Narrowband FSR INSTRUCTION OWNER HMI Project Manager PREPARED BY: I. Miller DATE: 2004

More information

Overview of Safety Code 35

Overview of Safety Code 35 Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)

More information

SIGNA Pioneer: Ultra High Efficiency Gradient System Advancing the gradient technology curve

SIGNA Pioneer: Ultra High Efficiency Gradient System Advancing the gradient technology curve GE Healthcare SIGNA Pioneer: Ultra High Efficiency Gradient System Advancing the gradient technology curve NEW TECHNOLOGY 40W Watts spec is irrelevant. 4W LED bulb delivers same brightness as 40W incandescent

More information

Supplementary Material

Supplementary Material Supplementary Material Orthogonal representation of sound dimensions in the primate midbrain Simon Baumann, Timothy D. Griffiths, Li Sun, Christopher I. Petkov, Alex Thiele & Adrian Rees Methods: Animals

More information

Nuclear Associates EZ CR-DIN Phantoms

Nuclear Associates EZ CR-DIN Phantoms Nuclear Associates 07-605-7777 EZ CR-DIN Phantoms Users Manual August 2006 Manual No. 07-605-7777-1 Rev. 4 2006 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

INSTRUCTIONS FOR THE CAROTID MRI COMPLETION FORM, CMR, Version B, 12/08/2005

INSTRUCTIONS FOR THE CAROTID MRI COMPLETION FORM, CMR, Version B, 12/08/2005 INSTRUCTIONS FOR THE CAROTID MRI COMPLETION FORM, CMR, Version B, 12/08/2005 I. General Instructions: This form is used to document the conduct and completeness of the MRI examination. The MRI examination

More information

On spatial resolution

On spatial resolution On spatial resolution Introduction How is spatial resolution defined? There are two main approaches in defining local spatial resolution. One method follows distinction criteria of pointlike objects (i.e.

More information

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC? Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM

More information

TITLE: Prostate Cancer Detection Using High-Spatial Resolution MRI at 7.0 Tesla: Correlation with Histopathologic Findings at Radical Prostatectomy

TITLE: Prostate Cancer Detection Using High-Spatial Resolution MRI at 7.0 Tesla: Correlation with Histopathologic Findings at Radical Prostatectomy Award Number: W81XWH-11-1-0253 TITLE: Prostate Cancer Detection Using High-Spatial Resolution MRI at 7.0 Tesla: Correlation with Histopathologic Findings at Radical Prostatectomy PRINCIPAL INVESTIGATOR:

More information

Image Display and Perception

Image Display and Perception Image Display and Perception J. Anthony Seibert, Ph.D. Department of Radiology UC Davis Medical Center Sacramento, California, USA Image acquisition, display, & interpretation X-rays kvp mas Tube filtration

More information

PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner

PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner September, 2017 Results submitted to Physics in Medicine & Biology Negar Omidvari 1, Jorge Cabello 1, Geoffrey Topping

More information

CARING DESIGN. PRACTICAL TECHNOLOGY.

CARING DESIGN. PRACTICAL TECHNOLOGY. GE Healthcare CARING DESIGN. PRACTICAL TECHNOLOGY. Brivo * MR355 1.5T Inspire ITS ELEGANT DESIGN AND WARM LIGHT ARE INTRIGUING. Brivo MR355 Inspire product manager Every piece of equipment you own represents

More information

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc.

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. An Introduction to TG-142 Imaging QA Using Standard Imaging Products Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. Goals Understand the nature and intent of TG 142 imaging

More information

Magnetic Resonance Imaging Principles, Methods, and Techniques

Magnetic Resonance Imaging Principles, Methods, and Techniques Magnetic Resonance Imaging Principles, Methods, and Techniques Perry Sprawls Jr., Emory University Publisher: Medical Physics Publishing Corporation Publication Place: Madison, Wisconsin Publication Date:

More information

NJDEP Medical Physicist s Radiographic QC Survey Registration Number:

NJDEP Medical Physicist s Radiographic QC Survey Registration Number: Facility Name NJDEP ID # NJDEP Medical Physicist s Radiographic QC Survey PLEASE PRINT Facility Information Unit Information Manufacturer Model Console Model # Console serial # Tube serial # Location (room)

More information

Range Sensing strategies

Range Sensing strategies Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart and Nourbakhsh 4.1.6 Range Sensors (time of flight) (1) Large range distance measurement -> called

More information

Physics 2310 Lab #5: Thin Lenses and Concave Mirrors Dr. Michael Pierce (Univ. of Wyoming)

Physics 2310 Lab #5: Thin Lenses and Concave Mirrors Dr. Michael Pierce (Univ. of Wyoming) Physics 2310 Lab #5: Thin Lenses and Concave Mirrors Dr. Michael Pierce (Univ. of Wyoming) Purpose: The purpose of this lab is to introduce students to some of the properties of thin lenses and mirrors.

More information

Gentec-EO USA. T-RAD-USB Users Manual. T-Rad-USB Operating Instructions /15/2010 Page 1 of 24

Gentec-EO USA. T-RAD-USB Users Manual. T-Rad-USB Operating Instructions /15/2010 Page 1 of 24 Gentec-EO USA T-RAD-USB Users Manual Gentec-EO USA 5825 Jean Road Center Lake Oswego, Oregon, 97035 503-697-1870 voice 503-697-0633 fax 121-201795 11/15/2010 Page 1 of 24 System Overview Welcome to the

More information

Holography. Casey Soileau Physics 173 Professor David Kleinfeld UCSD Spring 2011 June 9 th, 2011

Holography. Casey Soileau Physics 173 Professor David Kleinfeld UCSD Spring 2011 June 9 th, 2011 Holography Casey Soileau Physics 173 Professor David Kleinfeld UCSD Spring 2011 June 9 th, 2011 I. Introduction Holography is the technique to produce a 3dimentional image of a recording, hologram. In

More information

Digitizing Color. Place Value in a Decimal Number. Place Value in a Binary Number. Chapter 11: Light, Sound, Magic: Representing Multimedia Digitally

Digitizing Color. Place Value in a Decimal Number. Place Value in a Binary Number. Chapter 11: Light, Sound, Magic: Representing Multimedia Digitally Chapter 11: Light, Sound, Magic: Representing Multimedia Digitally Fluency with Information Technology Third Edition by Lawrence Snyder Digitizing Color RGB Colors: Binary Representation Giving the intensities

More information

Reference Targets Complete Test and Recalibration Kit Type CTS

Reference Targets Complete Test and Recalibration Kit Type CTS Delta-T SCAN Reference Targets Complete Test and Recalibration Kit Type CTS WARNING DO NOT LET THESE FILMS GET WET OR THEY MAY SWELL AND LOSE THEIR ACCURACY PROTECT FROM HUMIDITY, DIRT AND SCRATCHES. Delta-T

More information

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review MRES 7005 - Fast Imaging Techniques Module 2 Artefacts and Imaging Optimisation for single shot methods Content: Introduction Phase error Phase bandwidth Chemical shift review Chemical shift in pixels

More information

EE225E/BIOE265 Spring 2011 Principles of MRI. Assignment 6 Solutions. (y 0 + vt) dt. 2 y 0T + 3 )

EE225E/BIOE265 Spring 2011 Principles of MRI. Assignment 6 Solutions. (y 0 + vt) dt. 2 y 0T + 3 ) EE225E/BIOE265 Spring 211 Principles of MRI Miki Lustig Handout Assignment 6 Solutions 1. Nishimura 6.7 (Thanks Galen!) a) After the 9 y pulse, the spin is in the ˆx direction (using left-handed rotations).

More information

FCC ID: A3LSLS-BD106Q. Report No.: HCT-RF-1801-FC003. Plot Data for Output Port 2_QPSK 9 khz ~ 150 khz Middle channel 150 khz ~ 30 MHz Low channel

FCC ID: A3LSLS-BD106Q. Report No.: HCT-RF-1801-FC003. Plot Data for Output Port 2_QPSK 9 khz ~ 150 khz Middle channel 150 khz ~ 30 MHz Low channel Plot Data for Output Port 2_QPSK 9 khz ~ 150 khz Middle channel 150 khz ~ 30 MHz Low channel 30 MHz ~ 1 GHz Middle channel 1 GHz ~ 2.491 GHz Low channel 2.695 GHz ~ 12.75 GHz High channel 12.75 GHz ~ 26.5

More information

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Dr John Ridgway Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Physics for clinicians: Part I Journal of Cardiovascular Magnetic Resonance 2010, 12:71 http://jcmr-online.com/content/12/1/71

More information

Small, Low Power, High Performance Magnetometers

Small, Low Power, High Performance Magnetometers Small, Low Power, High Performance Magnetometers M. Prouty ( 1 ), R. Johnson ( 1 ) ( 1 ) Geometrics, Inc Summary Recent work by Geometrics, along with partners at the U.S. National Institute of Standards

More information

Magnetic Resonance Imaging

Magnetic Resonance Imaging Magnetic Resonance Imaging Principles, Methods, and Techniques Perry Sprawls, Ph.D., FACR, FAAPM, FIOMP Distinguished Emeritus Professor Department of Radiology Emory University Atlanta, Georgia Medical

More information

Delivering Better Patient Care with SIGNA Architect

Delivering Better Patient Care with SIGNA Architect Delivering etter Patient Care with SIGNA Architect As a regional leader in outpatient-based diagnostic imaging, the radiologists and staff at Inova Fairfax MRI Center are focused on one thing: the patient.

More information

Annual Ceiling Amount: $

Annual Ceiling Amount: $ High Field Magnetic Resonance Imaging Center Research MRI Suite, Basement Floor of Building 203 VAMC 4150 Clement Street, San Francisco, CA 94121 MR STUDY APPLICATION for Research Users Application date:

More information

High Dynamic Range Imaging

High Dynamic Range Imaging High Dynamic Range Imaging 1 2 Lecture Topic Discuss the limits of the dynamic range in current imaging and display technology Solutions 1. High Dynamic Range (HDR) Imaging Able to image a larger dynamic

More information

Improving high-field MRI using parallel excitation

Improving high-field MRI using parallel excitation review Improving high-field MRI using parallel excitation MRI at high magnetic field strengths promises to deliver clearer images of the body s structure and function. However, high-field MRI currently

More information

Ihor TROTS, Andrzej NOWICKI, Marcin LEWANDOWSKI

Ihor TROTS, Andrzej NOWICKI, Marcin LEWANDOWSKI ARCHIVES OF ACOUSTICS 33, 4, 573 580 (2008) LABORATORY SETUP FOR SYNTHETIC APERTURE ULTRASOUND IMAGING Ihor TROTS, Andrzej NOWICKI, Marcin LEWANDOWSKI Institute of Fundamental Technological Research Polish

More information

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT

More information

MR Advance Techniques. Flow Phenomena. Class II

MR Advance Techniques. Flow Phenomena. Class II MR Advance Techniques Flow Phenomena Class II Flow Phenomena In this class we will explore different phenomenona produced from nuclei that move during the acquisition of data. Flowing nuclei exhibit different

More information

5/17/2009. Digitizing Color. Place Value in a Binary Number. Place Value in a Decimal Number. Place Value in a Binary Number

5/17/2009. Digitizing Color. Place Value in a Binary Number. Place Value in a Decimal Number. Place Value in a Binary Number Chapter 11: Light, Sound, Magic: Representing Multimedia Digitally Digitizing Color Fluency with Information Technology Third Edition by Lawrence Snyder RGB Colors: Binary Representation Giving the intensities

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

SmartExam helps to banish most repeat knee studies at Desert Medical Imaging

SmartExam helps to banish most repeat knee studies at Desert Medical Imaging I s s u e 3 2 - J u l y 2 0 0 7 F i e l d Strength Publication for the Philips MRI Community SmartExam helps to banish most repeat knee studies at Desert Medical Imaging Total scan automation boosts efficiency,

More information

Image Processing for feature extraction

Image Processing for feature extraction Image Processing for feature extraction 1 Outline Rationale for image pre-processing Gray-scale transformations Geometric transformations Local preprocessing Reading: Sonka et al 5.1, 5.2, 5.3 2 Image

More information