INSTRUCTIONS FOR THE CAROTID MRI COMPLETION FORM, CMR, Version B, 12/08/2005

Size: px
Start display at page:

Download "INSTRUCTIONS FOR THE CAROTID MRI COMPLETION FORM, CMR, Version B, 12/08/2005"

Transcription

1 INSTRUCTIONS FOR THE CAROTID MRI COMPLETION FORM, CMR, Version B, 12/08/2005 I. General Instructions: This form is used to document the conduct and completeness of the MRI examination. The MRI examination is a central component of the overall research protocol. The MRI scan should be conducted within two months of the Field Center visit and often will be done on the same day as this visit. Field Center Staff affix the participant ID label, name label and acrostic label (the last in item 0), fill out items 1 and 2 and pass the form to the MRI technician. The Field Center staff is also responsible for entering the form into the Data Entry System (DES). The MRI Technologist is to complete Items 3 12 at the time of the MRI exam either manually or using the MRI computer system. The completed form will be returned to the Field Center Staff within 24 hours (fax may be used). Field center staff will enter the data into DES within 24 hours upon receipt of the paper form. Whenever numerical responses are required, enter the number so that the last digit appears in the rightmost box. Enter leading zeroes where necessary to fill all boxes. If a number is entered incorrectly, mark through the incorrect entry with an X. Code the correct entry clearly above the incorrect entry. II. Specific Instructions: A. To Be Completed By Field Center Staff: 0. Apply the acrostic label to item Enter weight in pounds from the examination. Weight is recorded in Item 3 or Item 5 of the Anthropometry Form (ANT). 2. Circle the side selected by the Coordinating Center for detailed imaging (suspected side for plaque). This is the side shown on Item 17 of the Screening Form (CSC) which is based on the side with the thickest IMT that has not had carotid artery surgery. The AFU staff prints the screen from the Data Entry System for this Item and sends it to the Carotid MRI field center staff. B. To Be Completed by Radiology Technologist: 3a. Indicate whether it was safe for the participant to go through the MRI examination. If not, the participant became ineligible and MR imaging should not be performed for safety reasons. If not eligible, skip to Item 8 to document reasons that the participant was ineligible. 3b. Enter the date the MRI exam was performed. 4. Enter the series number in the left column and the number of images for that series in the right column. For example if series 1 has 4 images and there are 3 boxes in the left column followed MRI Completion Form QxQ (CMRB) Page 1 of 5

2 by 2 boxes on the right, it will be entered as series 001, # images 04. If the series wasn t done, it will be entered as 999 for the series number and 00 for the number of images in the series. Note: If an entire series needs to be repeated (e.g., the scanner malfunctioned, the subject continued to cough throughout the series, you noticed the wrong parameters were inadvertently entered for the scan, etc.), enter the series number and number of images only for the repeated series. 4.a. Record the series number and number of images for the scout series. For GE scanners, the Coronal and Axial scouts are separate series (i.e., items a and b, respectively). For the Siemens scanner, a 3-plane localizer is acquired and this series number is entered in a. 4.b.1. For GE scanners, an Asset Calibration must be acquired first and this series is recorded in item a. This image is used to optimize the parallel imaging capability of the coils. It looks like a cloud and isn t used for making measurements. 4.b.2. Carotid time of flight (TOF) MRA, enter the series number and number of images. 4.b.3. MIPs (TOF), enter the series number and number of images. 4.c.1. Oblique (black blood) DIR PD (Right), enter the series number and number of images (usually 3 images) 4.c.2. Oblique (black blood) DIR PD (Left), enter the series number and number of images (usually 3 images) 4.d.1. Axial (black blood) DIR T1 (Right CCA), enter the series number and number of images (usually 1 image) 4.d.2. For GE scanners write in the prescan values (R1, R2, TG, last 4 digits of AX, X, Y, and Z). These will be used for the postcontrast scan 4.m.1. For Siemens scanners leave this line blank. 4.e. Axial Multiecho (Right CCA), enter the series number and number of images (usually 4 images) 4.f.1. Axial (black blood) T1 DIR (Left CCA), enter the series number and number of images (usually 1 image) 4.f.2. For GE scanners write in the prescan values (R1, R2, TG, last 4 digits of AX, X, Y, and Z). These will be used for the postcontrast scan 4.m.2. For Siemens scanners leave this line blank. 4.g. Axial Multiecho (Left CCA), enter the series number and number of images (usually 4 images) 4.h.1. Circle the side selected to be imaged. This is usually the same side as indicated in item 2. Examine the TOF MRA and oblique black blood images (series 4.b-4.f) for the presence of plaque and carotid stenosis. Plaque usually forms along the outer wall of the internal carotid artery (ICA) opposite the flow divider but might occur higher in the ICA, or sometimes even just below the bifurcation. Be careful because flow artifact also occurs along the outer wall of the ICA but is usually less well-defined. Ignore plaque involving the external carotid artery. If no plaque is seen or if the stenosis is comparable on the two sides then image the side selected by the Coordinating Center. If a plaque is seen causing greater stenosis on the opposite side from MRI Completion Form QxQ (CMRB) Page 2 of 5

3 the side selected by the Coordinating Center (item 2) then change the selected side to the side with the greater stenosis. Please indicate your reason for the change. Examples of reasons for changing sides are: Left side selected by Coordinating Center (item 2) but: - Plaque causing stenosis is seen involving right ICA and none seen on the left side - Around 70% stenosis seen involving the right ICA but the stenosis on the left measures <50% 4.h.2. If no plaque was seen then enter a B in the box indicating that the bifurcation is imaged. If plaque was seen, enter a P in the box. 4.h.3 Enter the series number followed by the number of images for the PCMRA sequence. The side imaged is the side selected in 4.h.1. The images will be stored as two separate series for the Siemens scanner (i.e., phase and magnitude images), but as one series for the GE scanners. 4.h.4. Blood pressure recordings during the MRI are only obtained in centers that have equipment to do this (Wake Forest). The blood pressure should be timed during or close to the time of the PCMRA. All centers will measure blood pressure after the MRI. 4.i.1. Axial (black blood) DIR T1 (8 slices on bifurcation or plaque), enter the series number and number of images (usually 8 images) 4.i.2. For GE scanners enter the prescan values (R1, R2, TG, last 4 digits of AX, X, Y, and Z). These will be used for the postcontrast scan 4.m.3. For Siemens scanners leave this line blank. 4.j.1. Enter the image numbers for the two thickest contiguous slices selected from the 8 axial DIR T1 images. These should show the thickest cross-section of the plaque, or thickest carotid artery wall if no plaque is present. These 2 images will be reimaged in the subsequent multi-echo (ME) scan (4.j.2). 4.j.2-4.j.3. Enter the series number followed by the number of images for the two thickest slices scanned. For GE scanners, images from each slice will be stored as a separate series. For the Siemens scanner, both slices will be stored as one series. 4.k.1. Contrast enhanced MRA mask, enter the series number and number of images. 4.k.2. Record the exact time the contrast injection began. Circle AM or PM. 4.k.3. If the entire contrast dose was not successfully administered, circle N. Otherwise, circle Y and go to item 4.k.5. 4.k.4. Indicate the reason the entire contrast dose could not be administered. Examples: Nausea, injector malfunction, extravasation 4.k.5. Record the volume of contrast administered. If a complete dose was not given, estimate the volume successfully injected (e.g., based on remaining contrast in the syringe minus estimated extravasate if this occurred) and enter this number in the box. 4.k.6. Record the gadolinium lot number. MRI Completion Form QxQ (CMRB) Page 3 of 5

4 4.l.1. Coronal Gd-enhanced 3D MRA (10 sec after injection), enter the series number and number of images. 10 seconds is an approximation of the time it takes for the contrast to travel from the IV site to the carotids. Start the images 10 seconds after the injection begins. 4.l.2. Record the series number and number of images for the CEMRA MIPs. Do not construct these MIPs until after the 16 postcontrast axial T1 images are prescribed and/or scanning. 4.l.3. Wait exactly 5 minutes from the time the contrast administration began (recorded in 4.k.2) before starting the next sequence (4.m.1). 4.m.1. Repeat #4.d.1. Axial T1 DIR (Right CCA), enter the series number and number of images (usually 1 image). For GE scanners, use the pre-scan values entered in 4.d.2. 4.m.2. Repeat #4.f.1. Axial T1 DIR (Left CCA), enter the series number and number of images (usually 1 image). For GE scanners, use the pre-scan values entered in 4.f.2. 4.m.3. Repeat #4.i.1. Axial T1 DIR (16 slices on side selected), enter the series number and number of images (usually 16 images). For GE scanners, use the pre-scan values entered in 4.i.2. The 8 precontrast slices acquired in 4.i.1 will be repeated with 5 slices added above and 3 added below to make Enter the series numbers and number of images for the 4 Screen Saves. 5a. Oblique black blood image through right carotid used as a scout with slice position drawn for CCA 5b. Oblique black blood image through left carotid used as a scout with slice position drawn for CCA 5c. Obliqe black blood image through selected carotid used as a scout with slice positions drawn for precontrast T1 images through ICA/plaque 5d. Obliqe black blood image through selected carotid used as a scout with slice positions drawn for postcontrast T1 images through ICA/plaque 6. If one or more image needs to be repeated for some reason (e.g., motion artifact) but not the entire series, this section documents which images are repeated to help the Reading Center know which images are of the same location. In the left column, enter each series number that was repeated followed by the individual image numbers that were repeated. In the right column, enter the new series number followed by the number of images contained in this series. 7. Enter the BP (SBP/DBP) at the end of the study. The participant should be lying flat on his/her back for this measurement. In centers where automated blood pressure measurement is also done during the scan (Wake Forest) this recording should be taken using the automatic device approximately 1-5 minutes after the patient s head is out of the scanner. Note that blood pressure is not measured prior to the scan to allow the scan to start expediently and minimize the risk of a patient not being able to complete the scan. 8. If the MRI was completed, circle Y and go to item 9. Otherwise, circle N and continue with 8a. 8a. Indicate reason the MRI was not completed. 9. Enter the code for the MRI Technologist who performed this exam. MRI Completion Form QxQ (CMRB) Page 4 of 5

5 10. Enter any comments you have for the MRI Reading Center. This space can be used to elaborate on previous responses limited by boxes. 11. If the participant experienced a reaction to the contrast, please elaborate here. Examples: - Subject reported a mild headache which did not interfere with the scan - Subject reported difficulty breathing following administration of the contrast which required taking the subject out of the scanner and then evaluation was performed by Dr. X. 12. If you notice an abnormality on the scan which you think should be brought to a radiologist s attention such as a mass or a tight narrowing of the carotid artery, list the abnormality and notify the radiologist in your department. After the study is reviewed by the radiologist, enter his/her name. The Field Center should be notified immediately for urgent findings confirmed by the radiologist. C. Administratcion To Be Completed By Field Center Staff 13. Enter the date of data entry. 14. Enter the code number of staff entering the data into the Data Management System. MRI Completion Form QxQ (CMRB) Page 5 of 5

ACRIN 6686 / RTOG 0825

ACRIN 6686 / RTOG 0825 ACRIN 6686 (RTOG 0825) Advanced MRI Imaging Manual ACRIN 6686 / RTOG 0825 A phase III double blind placebo controlled trial of conventional chemoradiation and adjuvant temozolomide plus bevacizumab vs

More information

Delivering Better Patient Care with SIGNA Architect

Delivering Better Patient Care with SIGNA Architect Delivering etter Patient Care with SIGNA Architect As a regional leader in outpatient-based diagnostic imaging, the radiologists and staff at Inova Fairfax MRI Center are focused on one thing: the patient.

More information

MR Advance Techniques. Flow Phenomena. Class II

MR Advance Techniques. Flow Phenomena. Class II MR Advance Techniques Flow Phenomena Class II Flow Phenomena In this class we will explore different phenomenona produced from nuclei that move during the acquisition of data. Flowing nuclei exhibit different

More information

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Advanced MSK MRI Protocols at 3.0T Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Outline Why High Field for MSK? SNR and Relaxation Times Technical Issues Example

More information

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT

More information

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging JOURNAL OF MAGNETIC RESONANCE IMAGING 20:1046 1051 (2004) Technical Note The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging James W. Goldfarb, PhD* Purpose: To describe a known (but undocumented)

More information

Automatic Selection of Mask and Arterial Phase Images for Temporally-Resolved MR Digital Subtraction Angiography

Automatic Selection of Mask and Arterial Phase Images for Temporally-Resolved MR Digital Subtraction Angiography Automatic Selection of Mask and Arterial Phase Images for Temporally-Resolved MR Digital Subtraction Angiography 21 May 2002, ISMRM 2002 Junhwan Kim, Martin R. Prince, Ramin Zabih,, Jeff Bezanson, Richard

More information

IMAGE MANAGEMENT PLAN FOR ACRIN PA 4003 Evaluation of the Ability of a Novel [ 18 F] amyloid ligand ([ 18 F-AV-45]) to distinguish patients with a

IMAGE MANAGEMENT PLAN FOR ACRIN PA 4003 Evaluation of the Ability of a Novel [ 18 F] amyloid ligand ([ 18 F-AV-45]) to distinguish patients with a IMAGE MANAGEMENT PLAN FOR ACRIN PA 4003 Evaluation of the Ability of a Novel [ 18 F] amyloid ligand ([ 18 F-AV-45]) to distinguish patients with a clinical diagnosis of Alzheimer s disease from cognitively

More information

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 ACR MRI accreditation program MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics,

More information

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology The Fundamentals of... Computed Tomography Computed Tomography (CT) systems use x-rays to produce images of slices through a patient s anatomy. Despite having lower spatial resolution than other x-ray

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

IHE Radiology Technical Framework Supplement. Stereotactic Mammography Image (SMI) Trial Implementation

IHE Radiology Technical Framework Supplement. Stereotactic Mammography Image (SMI) Trial Implementation Integrating the Healthcare Enterprise 5 IHE Radiology Technical Framework Supplement 10 Stereotactic Mammography Image (SMI) 15 Trial Implementation 20 25 Date: June 11, 2013 Author: IHE Radiology Technical

More information

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC? Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM

More information

2D, 3D CT Intervention, and CT Fluoroscopy

2D, 3D CT Intervention, and CT Fluoroscopy 2D, 3D CT Intervention, and CT Fluoroscopy SOMATOM Definition, Definition AS, Definition Flash Answers for life. Siemens CT Vision Siemens CT Vision The justification for the existence of the entire medical

More information

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Dr John Ridgway Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Physics for clinicians: Part I Journal of Cardiovascular Magnetic Resonance 2010, 12:71 http://jcmr-online.com/content/12/1/71

More information

Weber State University Radiologic Technology 4603

Weber State University Radiologic Technology 4603 Weber State University Radiologic Technology 4603 MRI Physics and Instrumentation Instructor: Rex T. Christensen MHA R.T. (R) (MR) (CT) (ARRT) CIIP Contact Info: E-mail: rexchristensen@weber.edu Phone:

More information

GE Healthcare. Discovery MR T. Simply powerful. Powerfully simple.

GE Healthcare. Discovery MR T. Simply powerful. Powerfully simple. GE Healthcare Discovery MR750 3.0T Simply powerful. Powerfully simple. Break free. The breast images you need in only two sequences. A complete liver study in a 15-minute time slot. Routine fmri with shorter

More information

The Enlightened Choice for High-field MRI

The Enlightened Choice for High-field MRI The Enlightened Choice for High-field MRI ECHELON heralds the dawn of a new standard for 1.5T superconductive MRI. The ECHELON features a small footprint with economics that do not compromise diagnostic

More information

PET/CT Instrumentation Basics

PET/CT Instrumentation Basics / Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire

More information

CARING DESIGN. PRACTICAL TECHNOLOGY.

CARING DESIGN. PRACTICAL TECHNOLOGY. GE Healthcare CARING DESIGN. PRACTICAL TECHNOLOGY. Brivo * MR355 1.5T Inspire ITS ELEGANT DESIGN AND WARM LIGHT ARE INTRIGUING. Brivo MR355 Inspire product manager Every piece of equipment you own represents

More information

Image Quality, Artifacts and Hazards in Imaging. Laura Gruber, MBA, RT(R), RDMS, RVT Sr. Director Medical Imaging

Image Quality, Artifacts and Hazards in Imaging. Laura Gruber, MBA, RT(R), RDMS, RVT Sr. Director Medical Imaging Image Quality, Artifacts and Hazards in Imaging Laura Gruber, MBA, RT(R), RDMS, RVT Sr. Director Medical Imaging Case 1 2 Case 1 What is the image quality issue in this picture? A.) Improper exposure/technique

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

diagnostic examination

diagnostic examination RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011

More information

Evaluation of MWT Materials Accusorb MRI Shield. Dr. E. Kanal, Department of Radiology, UPMC. Wednesday, April 21, 2010

Evaluation of MWT Materials Accusorb MRI Shield. Dr. E. Kanal, Department of Radiology, UPMC. Wednesday, April 21, 2010 Evaluation of MWT Materials Accusorb MRI Shield Dr. E. Kanal, Department of Radiology, UPMC Wednesday, April 21, 2010 Background: In almost all x-ray and ultrasonographic imaging examinations, in order

More information

CHECKLIST FOR YOUR 1 ST APPOINTMENT

CHECKLIST FOR YOUR 1 ST APPOINTMENT CHECKLIST FOR YOUR 1 ST APPOINTMENT Please use this checklist to prepare for your first visit to the Neurosurgeon or Endocrinologist. 2 weeks before your appointment Make sure you have: 1) Brought the

More information

QC by the MPE in Belgium

QC by the MPE in Belgium Acceptance testing of state-of-the-art CT scanners using a new national protocol: first experience on a large number of scanners of different make and model the working group Radiology of the Belgian Hospital

More information

Patient-Assisted Compression Impact on Image Quality and Workflow

Patient-Assisted Compression Impact on Image Quality and Workflow Patient-Assisted Compression Impact on Image Quality and Workflow Senographe Pristina In 2017, GE Healthcare s Senographe Pristina ( Pristina ) was approved by the FDA using the standard technologist-controlled

More information

Works-in-Progress package Version 1.0. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. January 22, 2003

Works-in-Progress package Version 1.0. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. January 22, 2003 Works-in-Progress package Version 1.0 For the Installation and User s Guide NUMARIS/4VA21B January 22, 2003 Section of Medical Physics, University Hospital Freiburg, Germany Contact: Klaus Scheffler PhD,

More information

Experience in implementing continuous arterial spin labeling on a commercial MR scanner

Experience in implementing continuous arterial spin labeling on a commercial MR scanner JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 1, WINTER 2005 Experience in implementing continuous arterial spin labeling on a commercial MR scanner Theodore R. Steger and Edward F. Jackson

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

MR Basics: Module 8 Image Quality

MR Basics: Module 8 Image Quality Module 8 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

Page 1 of 9. Protocol: adult_other_adni3_study_human_ge_3t_25w_ _ _1. 3 Plane Localizer. 3 Plane Localizer PATIENT POSITION

Page 1 of 9. Protocol: adult_other_adni3_study_human_ge_3t_25w_ _ _1. 3 Plane Localizer. 3 Plane Localizer PATIENT POSITION 3 Localizer FOV 26.0 Slice Thickness 5.0 Slice Spacing 0.0 Freq 256 Phase 128 3-PLANE 3 Localizer Unswap Phase Correction Gradient Echo Imaging Options Seq, Fast Recon All Images Contrast Yes/ 3 Localizer

More information

SIEMENS MAGNETOM Skyra syngo MR D13

SIEMENS MAGNETOM Skyra syngo MR D13 Page 1 of 12 SIEMENS MAGNETOM Skyra syngo MR D13 \\USER\CIND\StudyProtocols\PTSA\*ep2d_M0Map_p2_TE15 TA:7.9 s PAT:2 Voxel size:2.5 2.5 3.0 mm Rel. SNR:1.00 :epfid Properties Routine Contrast Prio Recon

More information

HAVE A QUESTION? CALL ZEBRA CLIENT CARE

HAVE A QUESTION? CALL ZEBRA CLIENT CARE SCANNER FUNCTIONS SET TEAM MEMBER Using the Stylus, Tap the scanner screen twice to display the Key board. Use the keyboard to Key the Team Member and press ENT (Enter). NOTE: When entering the Team Member

More information

2 nd generation TOMOSYNTHESIS

2 nd generation TOMOSYNTHESIS 2 nd generation TOMOSYNTHESIS 2 nd generation DBT true innovation in breast imaging synthesis graphy Combo mode Stereotactic Biopsy Works in progress: Advanced Technology, simplicity and ergonomics Raffaello

More information

Ultrasound & Artifacts

Ultrasound & Artifacts ISSN 2005-7881 Journal of Neurosonology 3(Suppl. 2):1-17, 2011 Ultrasound & Artifacts Siryung Han The Catholic University of Korea Artifacts False image- echoes without anatomic correlate US image dose

More information

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National

More information

SmartExam helps to banish most repeat knee studies at Desert Medical Imaging

SmartExam helps to banish most repeat knee studies at Desert Medical Imaging I s s u e 3 2 - J u l y 2 0 0 7 F i e l d Strength Publication for the Philips MRI Community SmartExam helps to banish most repeat knee studies at Desert Medical Imaging Total scan automation boosts efficiency,

More information

KM-4800w. Copy/Scan Operation Manual

KM-4800w. Copy/Scan Operation Manual KM-4800w Copy/Scan Operation Manual NOTE: This Operation Manual contains information that corresponds to using both the metric and inch versions of these machines. The metric versions of these machines

More information

High Field MRI: Technology, Applications, Safety, and Limitations

High Field MRI: Technology, Applications, Safety, and Limitations High Field MRI: Technology, Applications, Safety, and Limitations R. Jason Stafford, Ph.D. The University of Texas M. D. Anderson Cancer Center, Houston, TX Introduction The amount of available signal

More information

Pulse Sequence Design and Image Procedures

Pulse Sequence Design and Image Procedures Pulse Sequence Design and Image Procedures 1 Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant 2 A pulse sequence is a timing diagram designed with a series of RF pulses, gradients switching, and signal readout

More information

BACKGROUND. ** 78% of all MRI scanners have Image Quality problems. *** *** 25% of all Multi-Channel RF coils have at least one bad channel.

BACKGROUND. ** 78% of all MRI scanners have Image Quality problems. *** *** 25% of all Multi-Channel RF coils have at least one bad channel. Range of Results from over 534 ACR-mandated Annual MRI Performance Evaluations on over 204 Magnets from 8 Vendors Spanning a 10-year Period Moriel NessAiver, Ph.D. - Simply Physics - Baltimore, MD moriel@simplyphysics.com

More information

American College of Radiology MR Accreditation Program. Testing Instructions

American College of Radiology MR Accreditation Program. Testing Instructions American College of Radiology MR Accreditation Program Testing Instructions (Revised May 28, 2017) This guide provides all of the instructions necessary for clinical tests, phantom tests and general submission

More information

Alzheimer s Disease Neuro Imaging III (ADNI3) Study

Alzheimer s Disease Neuro Imaging III (ADNI3) Study Alzheimer s Disease Neuro Imaging III (ADNI3) Study 3T MRI Technical Procedures Manual ADNI3 3T MRI Technical Procedures Manual Page 1 of 28 Contents 1. Contact Information For the ADNI3 Study... 4 2.

More information

Translating Protocols Between Scanner Manufacturer and Model

Translating Protocols Between Scanner Manufacturer and Model Translating Protocols Between Scanner Manufacturer and Model Robert J. Pizzutiello, MS, FAAPM, FACMP Sr. Vice-President, Global Physics Solutions President, Upstate Medical Physics Objectives Understand

More information

Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance

Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance FDA Workshop April 13, 2010 Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance Edward F. Jackson, PhD Professor and Chief, Section

More information

Forces on a 2D Plane

Forces on a 2D Plane C h a p t e r 3 Forces on a 2D Plane In this chapter, you will learn the following to World Class standards: 1. Measuring the Magnitude and Direction of a Force Vector 2. Computing the Resultant Force

More information

SIGNA Explorer Lift revives our MR

SIGNA Explorer Lift revives our MR Seiji Shiotani, MD, PhD Seirei Fuji Hospital in Fuji City, Shizuoka, Japan Masayoshi Sugimura Seirei Fuji Hospital in Fuji City, Shizuoka, Japan SIGN Explorer Lift revives our MR The clinical usefulness

More information

Vivid S5. Cardiovascular ultrasound system. GE Healthcare. Davis Medical

Vivid S5. Cardiovascular ultrasound system. GE Healthcare. Davis Medical GE Healthcare Vivid S5 Cardiovascular ultrasound system Versatility. It s a new design concept one that leverages our miniaturization expertise gained from the Vivid i and our performance expertise of

More information

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev. 1.0 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE DoseLab users may reference the following instructions to perform Tomotherapy Quality Assurance tests as recommended

More information

VX-2100/VX-2200 (UHF) Alignment

VX-2100/VX-2200 (UHF) Alignment VX-2100/VX-2200 (UHF) Introduction The VX-2100/2200 is carefully aligned at the factory for the specified performance across the frequency range specified for each version. Realignment should therefore

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

Wide-Detector CT for TAVR Planning:

Wide-Detector CT for TAVR Planning: Wide-Detector CT for TAVR Planning: Impact on Iodine Dose, Radiation Dose, and Image Quality SCBTMR 2015 Annual Course Thursday, October 8 William P. Shuman MD FSCBTMR Department of Radiology University

More information

CT Basics: Equipment and Instrumentation Module 2

CT Basics: Equipment and Instrumentation Module 2 Module 2 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational in-house or online educational course use only in educational and corporate institutions.

More information

Detector technology in simultaneous spectral imaging

Detector technology in simultaneous spectral imaging Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become

More information

CFMRI Mock Scanner User s Manual. (Last Updated 10/23/2014)

CFMRI Mock Scanner User s Manual. (Last Updated 10/23/2014) CFMRI Mock Scanner User s Manual (Last Updated 10/23/2014) Page 1 of 6 CFMRI 2014 UCSD Center for functional MRI (CFMRI) provides a mock Scanner that simulates the look and feel of a real MRI scanner.

More information

GETTING TO KNOW YOU. 1. What is the concern or reason for your visit today?

GETTING TO KNOW YOU. 1. What is the concern or reason for your visit today? GETTING TO KNOW YOU 1. What is the concern or reason for your visit today? 2. How did you learn about us? (circle one) Doctor Friend or Another Patient Community Agency Brochure or News Story Alzheimer

More information

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL TORNIER BLUEPRINT 3D Planning + PSI SCAN PROTOCOL Contents 3 Introduction 3 Patient preparation 3 Scanning instructions 4 Image instructions 5 Scanning parameters 6 Technical instructions 2 BLUEPRINT 3D

More information

Annual Ceiling Amount: $

Annual Ceiling Amount: $ High Field Magnetic Resonance Imaging Center Research MRI Suite, Basement Floor of Building 203 VAMC 4150 Clement Street, San Francisco, CA 94121 MR STUDY APPLICATION for Research Users Application date:

More information

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Val M. Runge, MD Institute for Diagnostic and Interventional Radiology Clinics for Neuroradiology and Nuclear Medicine University Hospital Zurich

More information

Publication for the Philips MRI Community

Publication for the Philips MRI Community FieldStrength Publication for the Philips MRI Community Issue 38 Summer 2009 Hospitals deploy Panorama HFO as an all purpose MRI scanner Panorama High Field Open scanner serves all MRI needs of both Community

More information

GE Healthcare. Vivid S5. Cardiovascular ultrasound system

GE Healthcare. Vivid S5. Cardiovascular ultrasound system GE Healthcare Vivid S5 Cardiovascular ultrasound system Versatility. It s a new design concept one that leverages our miniaturization expertise gained from the Vivid i and our performance expertise of

More information

Pulse Sequence Design Made Easier

Pulse Sequence Design Made Easier Pulse Sequence Design Made Easier Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant gurumri@gmail.com 1 2 Pulse Sequences generally have the following characteristics: An RF line characterizing RF Pulse applications

More information

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical 3T Unlimited ipat on MAGNETOM Allegra The Importance of ipat at 3T s medical ipat on MAGNETOM Allegra The Importance of ipat at 3T The rise of 3T MR imaging Ultra High Field MR (3T) has flourished during

More information

Diff Ganzkörper ep2d_diff_stir_b50_800_tra

Diff Ganzkörper ep2d_diff_stir_b50_800_tra Table of contents \\USER Ganzkörper clinical Diff Ganzkörper ep2d_diff_stir_b50_800_tra - 1 - \\USER\Ganzkörper\clinical\Diff Ganzkörper\ep2d_diff_stir_b50_800_tra TA: 2:23 PM: ISO Voxel size: 1.6 1.6

More information

MR Basics: Module 6 Pulse Sequences

MR Basics: Module 6 Pulse Sequences Module 6 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

Going Live with the Aquilion VISION Volume 4D CT

Going Live with the Aquilion VISION Volume 4D CT VISIONS 2-14 COMPUTED TOMOGRAPHY Going Live with the Aquilion VISION Volume 4D CT Interview with Dr. Rick Bhatia, Regional Clinical Chief Diagnostic Imaging Program, Eastern Health, Newfoundland What were

More information

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Magnetic Resonance Imaging Spatial

More information

Discover the new Prestige and experience 3D/4D imaging beyond your imagination.

Discover the new Prestige and experience 3D/4D imaging beyond your imagination. 3D/4D Beyond Imagination The Prestige ultrasound imaging system represents the pinnacle of more than a decade of technological advancement in 3D/4D ultrasound imaging at MEDISON. Inheriting a tradition

More information

Philips Site Yearly Performance Evaluation Philips Openview 16-Jan-08. Table of Contents

Philips Site Yearly Performance Evaluation Philips Openview 16-Jan-08. Table of Contents Philips Site Yearly Performance Evaluation Philips Openview 6-Jan-8 Table of Contents Summary and Signature Page 2 Specific Comments 3 Site Information 4 Equipment Information 4 Table Position Accuracy

More information

1 Introduction. 2 The basic principles of NMR

1 Introduction. 2 The basic principles of NMR 1 Introduction Since 1977 when the first clinical MRI scanner was patented nuclear magnetic resonance imaging is increasingly being used for medical diagnosis and in scientific research and application

More information

MRI with the VNS Therapy System October 2017

MRI with the VNS Therapy System October 2017 MRI with the VNS Therapy System October 2017 The information contained in this document is one part of the full labeling for the implanted portions of the VNS Therapy System. It is not intended to serve

More information

Chapter 6: TVA MR and Cardiac Function

Chapter 6: TVA MR and Cardiac Function Chapter 6 Cardiac MR Introduction Chapter 6: TVA MR and Cardiac Function The Time-Volume Analysis (TVA) optional module calculates time-dependent behavior of volumes in multi-phase studies from MR. An

More information

TradeSchool Start Guide

TradeSchool Start Guide TradeSchool Start Guide Welcome to TradeSchool! TradeSchool is the computer program that your Committee uses to track your progress through your apprentice program. As an apprentice, you also have access

More information

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey Implementation of the 2016 ACR Digital Mammography QC Manual Medical Physicist Mammography Equipment Evaluation and Annual Survey Eric A Berns, PhD, FACR Radiologist Responsibilities Radiologist Responsibilities

More information

Clinical Importance on CT

Clinical Importance on CT 183 Truncated-View Clinical Importance on CT..,..,.... Artifacts:.... James L. Lehr1 A truncated-view artifact in CT is produced whenever any part of the patient or imaged object is present in some but

More information

Compact yet Sophisticated

Compact yet Sophisticated Compact yet Sophisticated Hitachi has brought Open MRI one step further in its evolution, to better assist medical professionals who work at the forefront of healthcare. AIRIS Light MSK offers radiologists

More information

This is a limited subset of the types and difficulties of the questions from the 2013 State Competition.

This is a limited subset of the types and difficulties of the questions from the 2013 State Competition. This is a limited subset of the types and difficulties of the questions from the 2013 State Competition. It is NOT meant to be all inclusive or complete, or to limit the scope of the 2014 competition in

More information

Overview of Safety Code 35

Overview of Safety Code 35 Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)

More information

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

1. Patient size AEC. Large Patient High ma. Small Patient Low ma Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description

More information

Architecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare

Architecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare Architecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare 2 DOC1292532 Architecture of Quality Imaging Agile Acoustic Architecture E-Series and XDclear

More information

Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution

Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Cynthia H. McCollough, PhD, DABR, FAAPM, FACR Director, CT Clinical Innovation Center Professor of Medical Physics and

More information

IMAGE MANAGEMENT PLAN FOR ACRIN PA

IMAGE MANAGEMENT PLAN FOR ACRIN PA IMAGE MANAGEMENT PLAN FOR ACRIN PA 4003 Evaluation of the Ability of a Novel [ 18 F] amyloid ligand ([ 18 F- AV-45]) to distinguish patients with a clinical diagnosis of Alzheimer s disease from cognitively

More information

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

HETERONUCLEAR IMAGING. Topics to be Discussed:

HETERONUCLEAR IMAGING. Topics to be Discussed: HETERONUCLEAR IMAGING BioE-594 Advanced MRI By:- Rajitha Mullapudi 04/06/2006 Topics to be Discussed: What is heteronuclear imaging. Comparing the hardware of MRI and heteronuclear imaging. Clinical applications

More information

COSPAS-SARSAT ACCEPTANCE OF 406 MHz BEACON TYPE APPROVAL TEST FACILITIES

COSPAS-SARSAT ACCEPTANCE OF 406 MHz BEACON TYPE APPROVAL TEST FACILITIES COSPAS-SARSAT ACCEPTANCE OF 406 MHz BEACON TYPE APPROVAL TEST FACILITIES C/S T.008 Issue 2 T8NOV05 i C/S T.008 - Issue 2 COSPAS-SARSAT ACCEPTANCE OF 406 MHz BEACON TYPE APPROVAL TEST FACILITIES History

More information

23 CP Clarify Enhanced US Volume Image and Frame Type Values 3 and 4

23 CP Clarify Enhanced US Volume Image and Frame Type Values 3 and 4 23 CP-1463 - Clarify Enhanced US Volume Image and Frame Type Values 3 and 4 Page 1 1 Status Finale Text 2 Date of Last Update 2015/11/10 3 Person Assigned David Clunie 4 mailto:dclunie@dclunie.com 5 Submitter

More information

23 CP Clarify Enhanced US Volume Image and Frame Type Values 3 and 4

23 CP Clarify Enhanced US Volume Image and Frame Type Values 3 and 4 23 CP-1463 - Clarify Enhanced US Volume Image and Frame Type Values 3 and 4 Page 1 1 Status Letter Ballot 2 Date of Last Update 2015/09/16 3 Person Assigned David Clunie 4 mailto:dclunie@dclunie.com 5

More information

CT Scanner Dose Survey

CT Scanner Dose Survey CT Scanner Dose Survey Measurement Protocol Version 5.0 July 1997 Co-ordinated by ImPACT and The Medical Physics Department,, London SW17, UK. 0181-725-3366 CT Scanner Dose Survey: Measurement Protocol

More information

Software and Hardware in CCTA. Elly Castellano PhD

Software and Hardware in CCTA. Elly Castellano PhD Software and Hardware in CCTA Elly Castellano PhD Outline technical requirements for coronary CTA the modern cardiac CT scanner ECG-gating technology image reconstruction algorithms 2 Technical requirements

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

Siemens AG, Healthcare Sector. syngo MR D Operator Manual - Breast 0.0.

Siemens AG, Healthcare Sector. syngo MR D Operator Manual - Breast 0.0. Siemens AG, Healthcare Sector Cs2 syngo Breast Operator 2010-2012 MR-05019 630 02 English 06/2012 n.a. Informatik, Manual D13 Cape syngo MR D13 Operator Manual - Breast syngo MR D13 www.siemens.com/healthcare

More information

[Maestro Class] MAGNETOM Harmony The New Degree of Perfection

[Maestro Class] MAGNETOM Harmony The New Degree of Perfection [Maestro Class] MAGNETOM Harmony The New Degree of Perfection MAGNETOM Family The Perfection of Care The aim of Siemens MR is the perfection of care. To create products and services that improve the quality

More information

Publication for the Philips MRI Community Issue 33 December High demand leads to second Panorama HFO purchase at KNI

Publication for the Philips MRI Community Issue 33 December High demand leads to second Panorama HFO purchase at KNI FieldStrength Publication for the Philips MRI Community Issue 33 December 2007 High demand leads to second Panorama HFO purchase at KNI Utility and excellent image quality make decision easy This article

More information

PET: New Technologies & Applications, Including Oncology

PET: New Technologies & Applications, Including Oncology PET: New Technologies & Applications, Including Oncology, PhD, FIEEE Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Disclosures Research Contract, GE Healthcare

More information

CSE 123: Computer Networks

CSE 123: Computer Networks Total Points = 27 CSE 123: Computer Networks Homework 3 Solutions Out: 5/11, Due: 5/18 Problems 1. Distance Vector Routing [9 points] For the network shown below, give the global distance vector tables

More information

GE Site Yearly Performance Evaluation GE Signa Excite HD - 3T 1-Sep-08. Table of Contents

GE Site Yearly Performance Evaluation GE Signa Excite HD - 3T 1-Sep-08. Table of Contents GE Site Yearly Performance Evaluation GE Signa Excite HD T Sep8 Table of Contents Summary and Signature Page 2 Specific Comments Site Information 4 Equipment Information 4 Table Position ccuracy 4 Magnetic

More information

Downloaded from by on 02/07/18 from IP address Copyright ARRS. For personal use only; all rights reserved

Downloaded from  by on 02/07/18 from IP address Copyright ARRS. For personal use only; all rights reserved Downloaded from www.ajronline.org by 46.3.192.5 on 02/07/18 from IP address 46.3.192.5. Copyright RRS. For personal use only; all rights reserved C oil sensitivity encoding (SENSE) is a new technique that

More information

What language do you speak? English

What language do you speak? English A 1 What language do you speak? English Españo Português Tagalog ế Basically the hospital provides services in Japanese only however the multilingual contact center service is available if needed. (Fee:

More information

4.0 How to Turn On the Selenia Dimensions

4.0 How to Turn On the Selenia Dimensions Chapter 2 System Controls and Indicators How to Turn On the Selenia Dimensions 4.0 How to Turn On the Selenia Dimensions 4.1 Preparation 1. Reset all three Emergency Off switches. Emergency Off Switches

More information