GE Site Yearly Performance Evaluation GE Signa Excite HD - 3T 1-Sep-08. Table of Contents

Size: px
Start display at page:

Download "GE Site Yearly Performance Evaluation GE Signa Excite HD - 3T 1-Sep-08. Table of Contents"

Transcription

1 GE Site Yearly Performance Evaluation GE Signa Excite HD T Sep8 Table of Contents Summary and Signature Page 2 Specific Comments Site Information 4 Equipment Information 4 Table Position ccuracy 4 Magnetic Field Homogeneity 4 Slice ccuracy 4 Slice Crosstalk Soft Copy Displays 6 RF Coil Performance Evaluation Coil Inventory List 7 Body Integrated 8 Breast 8 Ch. 9 Cardiac 8 Ch. Head 8 Ch 7 Head Split Top 8 Knee 8 Ch. 2 Knee T/R 2 eurovascular rray 8 Ch. 2 Shoulder Ch. 26 Spine 8 Ch. 28 Torso Wrist ppendix : Magnet Homgeneity Map ppendix B: Slice / Profiles / RF Crosstalk ppendix C: CR Phantom nalysis ppendix D: Explanation of RF Coil Test Format 2 4 4

2 MRI Equipment Evaluation Summary & Signature Page Site ame: GE Site MRP # ddress: Survey Date: City, State, Zip Report Date: MRI Mfg: GE Model: Excite HD Field: 9 9//8 9//8 T MRI Scientist: Moriel essiver, Ph.D. Signature: Equipment Evaluation Tests Pass Fail * /. Magnetic field homogeneity: 2. Slice position accuracy:. Table positioning reproducibility: 4. Slice thickness accuracy:. RF coils' performance: a. Volume QD Coils b. Phase rray Coils c. Surface Coils 6. Interslice RF interference (Crosstalk): 7. Soft Copy Display Evaluation of Site's Technologist QC Program Pass Fail * /. Set up and positioning accuracy: (daily) 2. Center frequency: (daily). Transmitter attenuation or gain: (daily) 4. Geometric accuracy measurments: (daily). Spatial resolution measurements: (daily) 6. Low contrast detectability: (daily) 7. Head Coil (daily) 8. Body Coil (weekly) 9. Fast Spin Echo (F/T) ghosting levels: (daily). Film quality control: (weekly). Visual checklist: (weekly) *See comments page for description of any failures. GE Site GE Excite HD T 2

3 Specific Comments and Recommendations. The breast coil has an effectively dead channel. When using bilateral mode, 7 channels were good, one bad. When using EITHER the left side or the right side only, there were good channels and one bad There is no difference between ports & B (if the coil can use either.) The magnet homogeneity is good. The display console is good and there is good agreement between the screen and the film. Everything else looks fine O: Please be sure to read appendix D for an explanation of the format of this document. GE Site GE Excite HD T

4 Site ame: GE Site MRI Equipment Performance Evaluation Data Form Contact Title Phone Fax Equipment Information MRI Manufacturer: GE Model: Twinspeed S: 24 Software: 4._M4_629a Camera Manufacturer: Model: S: Software: PCS Manufacturer: Model: S: Software: CR Phantom umber used: J6. Table Positioning Reproducibility: Table motion out/in: IsoCenter Measured Phantom Center.4 Comment: Out/In.9 Out/In Out/In.4.4 Pass 2. Magnetic Field Homogeneity See appendix for field plots. PSS xial: Coronal: Sagittal: Last Year CF: 27,728,84 This Year CF: 27,727,86 CF Change: cm 2 cm 2 cm GRE :, : & Flip ngle: 4, : 4 mm skip mm, :.4KHz, x28, 2nex Comments: Shim is very good Slice ccuracy : 2mm Matrix: x (Slice # from CR Phantom) ll values in mm (CR) (Site T) (2/8) (2/8) Flip S Calc Target % Error.2%.2%.2% 2.2% F(6) T MEMP GRE % 8.% 4.% Comments: GE Site GE Excite HD T 4

5 4. Slice Crosstalk (RF interference) The following data were obtained using the CR phantom slice thickness wedges to measure the slice profile of three common T weighted sequences when the slice gap varies from 2% down to % (contiguous) s the slices get closer together it is expected that the edges of the slices will overlap causing a deterioration of the slice profile. The data shown below clearly demonstrates this effect. The T MEMP sequence is a spin echo sequence with a a 6 flip angle. Both this sequence and the standare sequence shows a drop in the profile once the gap drops below %. The interference is less with the smaller flip angle and longer (as expected) The GRE sequence is more complicated to evaluate because the images were not as sharp and the profile wedges were not well defined. ll of the slice profiles can be seen in ppendix B. There you can see how the profile is very square while the GREs are more triangular. Flip ngle (cm 2 ) Matrix S # of slices Full x T MEMP x GRE in Phs x 4 Skip Full T MEMP GRE In Phase T Weighted Slice es (mm) Full T MEMP GRE InPhase Slice (mm) GE Site GE Excite HD T

6 . Soft & Hard Copy Displays Luminance Meter Make/Model: Tektronix J6 Digital Photometer Monitor Description: Efilm workstation Cal Expires: 4/6/6 Which Monitor Luminance Measured: Ft. lamberts Center of Image Display Top Left Corner Top Right Corner Bottom Left Corner Bottom Right Corner MX MI Percent Delta Console % Y SMP OK? % delta =2% x (maxmin)/(max+center) (>% is action limit) imum Brightness must be > Ft. Lamberts The monitor is excellent and the agreemen between the screen and film is good. Density Ft Lamber Film Density LCD & Film Response Curve Log FtLambert.. Ideal Curve LCD Film % Density GE Site GE Excite HD T 6

7 Coil and Other Hardware Inventory List Site ame GE Site CR Magnet # ickname Signa ctive Coil Description Manufacturer Model Rev. Mfg. Date S Channels Body Integrated Breast 8 Ch GE 2444 Sep, Cardiac 8 Ch GE Oct, 2 8 Head 8 Ch GE 4742 ug, 26 U Head Split Top GE 79WH pr, 26 79WH Knee 8ch Invivo 44 Sep, 26 U297 8 Knee T/R Invivo U2229 eurovascular rray 8 Ch USI 2449 Sep, Shoulder Ch USI 244 May, 26 2 Spine 8 Ch USI 2442 Jul, Torso USI 244 Dec, Wrist HD Mayo Clinic BC HBR6 7

8 Coil: Body Integrated Mfg.: Mfg. Date: Coil ID: 97 Phantom: Body sphere Test Date: 9//28 Model: S: # of Channels 2 T 48 x y.6 S Body TX gain: 66 R: R2: 28 nalysis of Test Image ormalized,68, %,68,2, % The of the Body coil is 4 times higher than last year! G E S i t e Signa 8

9 Coil: Breast 8 Ch Mfg.: GE Mfg. Date: 9//26 Coil ID: 9 Phantom: HD Breast Phantom Test Date: 9//28 Model: 2444 S: 7 # of Channels 8 2 T 48 x y.6 S HD Breast Port TX gain: 6 R: R2: 29 nalysis of Composite Image,48,47,877,87,6, ormalized 72.% 72.4% nalysis of Uncombined Images Ch ,9,,7 999,4, 9, % 72% 7% 8% % 7% 24% 8% % 79% 8% 87% % 72% 27% 88% Channel #7 is one fourth of what it should be. Channel Channel 2 Channel Channel 4 Composites Channel Channel 6 Channel 7 Channel 8 G E S i t e Signa 9

10 Coil: Breast 8 Ch Mfg.: GE Mfg. Date: 9//26 Coil ID: 9 Phantom: HD Breast Phantom Test Date: 9//28 Model: 2444 S: 7 # of Channels 8 2 C 4 x y.6 S HD Breast Port,67,672,68,672,98,94,98,94,98,4,98, nalysis of Composite Image TX gain: 6 R: 2 R2: ormalized 82.8% 87.% 82.6% 87.% nalysis of Uncombined Images Ch , ,82,6,2,,4,28 997, % 66% 9% 8% 96% 4% 2% % % 7% 68% 8% % 6% 2% 88% Channel 7 is /4th of what it should be. Composites Channel Channel 2 Channel Channel 4 Composites Channel Channel 6 Channel 7 G E S i t e Signa

11 Coil: Breast 8 Ch Mfg.: GE Mfg. Date: 9//26 Coil ID: 9 Phantom: HD Breast Phantom Test Date: 9//28 Model: 2444 S: 7 # of Channels 8 2 T 48 x y.6 S HD Breast Port B TX gain: 6 R: R2: 29 nalysis of Composite Image,4,44,88,86,6, ormalized 7.7% 7.8% nalysis of Uncombined Images Ch ,4 97,44 986,64, 87, % 7% 7% 8% % 78% 8% 82% % 78% 8% 86% % 74% 2% 87% Channel 7 is /4th of what it should be. Channel Channel 2 Channel Channel 4 Composites Channel Channel 6 Channel 7 Channel 8 G E S i t e Signa

12 Coil: Breast 8 Ch Mfg.: GE Mfg. Date: 9//26 Coil ID: 9 Phantom: HD Breast Phantom Test Date: 9//28 Model: 2444 S: 7 # of Channels 8 2 C 4 x y.6 S HD Breast Port B,9,66,9,66,89,88,896,87,48,47,48,4 nalysis of Composite Image TX gain: 6 R: 2 R2: ormalized 8.2% 87.% 8.% 87.4% nalysis of Uncombined Images Ch , ,,4, 987,4,9 96 6, % 67% 6% 8% 97% 4% % % % 7% 67% 82% % 6% 4% 88% Channel 7 is /4th of what it should be. Composites Channel Channel 2 Channel Channel 4 Composites Channel Channel 6 Channel 7 G E S i t e Signa 2

13 Coil: Breast 8 Ch Mfg.: GE Mfg. Date: 9//26 Coil ID: 9 Phantom: HD Breast Phantom Test Date: 9//28 Model: 2444 S: 7 # of Channels 8 2 T 48 x y.6 S HD BreastLeft TX gain: 6 R: R2: 29 nalysis of Composite Image 76 76,, ormalized 7.% 6.9% nalysis of Uncombined Images Ch , , % 6% 2% 8% % 66% 26% 8% Channel is dead in both the Left and Right only modes. Channel Channel 2 Composites Channel Channel 4 G E S i t e Signa

14 Coil: Breast 8 Ch Mfg.: GE Mfg. Date: 9//26 Coil ID: 9 Phantom: HD Breast Phantom Test Date: 9//28 Model: 2444 S: 7 # of Channels 8 2 C 48 x y.6 S HD BreastRight TX gain: R: R2: nalysis of Composite Image ,4, ormalized 2.% 2.% nalysis of Uncombined Images Ch , 97 94, % 9% % % % 9% 22% % Channel is dead in both the Left and Right only modes. Channel Channel 2 Composites Channel Channel 4 G E S i t e Signa 4

15 Coil: Cardiac 8 Ch Mfg.: GE Mfg. Date: //2 Coil ID: 9 Phantom: Cardiac phantom Test Date: 9//28 Model: S: # of Channels 8 2 T 48 x y.6 S HD Cardiac Port TX gain: 69 R: R2: 29 nalysis of Composite Image ,2 2, ormalized 2.% 24.9% nalysis of Uncombined Images Ch ,76,46,8,,4,66,696, % 9% 9% 88% 84% % % 72% % 77% 69% 9% 72% 8% 82% 64% There is no significant difference between Ports & B Channel Channel 2 Channel Channel 4 Composites Channel Channel 6 Channel 7 Channel 8 G E S i t e Signa

16 Coil: Cardiac 8 Ch Mfg.: GE Mfg. Date: //2 Coil ID: 9 Phantom: Cardiac phantom Test Date: 9//28 Model: S: # of Channels 8 2 T 48 x y.6 S HD Cardiac Port B TX gain: 69 R: R2: 29 nalysis of Composite Image 84 84,97, ormalized 2.% 2.2% nalysis of Uncombined Images Ch ,64,424,2,49,2,8,96, % 94% 9% 88% 8% 99% % 72% % 78% 7% 92% 7% 8% 8% 6% There is no significant difference between Ports & B Channel Channel 2 Channel Channel 4 Composites Channel Channel 6 Channel 7 Channel 8 G E S i t e Signa 6

17 Coil: Head 8 Ch Mfg.: GE Mfg. Date: 8/6/26 Coil ID: 898 Phantom: Head TLT sphere Test Date: 9//28 Model: 4742 S: U2878 # of Channels 8 2 T 6 x y.6 S 8HRBrain TX gain: 6 R: R2: 29 nalysis of Composite Image,7,6,92,92,9, ormalized 77.% 77.% nalysis of Uncombined Images Ch ,26,24,72,224,9,84,9, % 99% 97% 8% 92% 94% % 96% % % 98% 86% 96% 99% 9% 92% There is a 7% drop in since last year. Channel Channel 2 Channel Channel 4 Composites Channel Channel 6 Channel 7 Channel 8 G E S i t e Signa 7

18 Coil: Head Split Top Mfg.: GE Mfg. Date: 4//26 Coil ID: 9 Phantom: CR Test Date: 9//28 Model: 79WH S: 79WH # of Channels 2 T 4 x y.6 S Head TX gain: R: R2: nalysis of Test Image ormalized,6 4,486 2, %,62 4,488 2, % G E S i t e Signa 8

19 Coil: Head Split Top Mfg.: GE Mfg. Date: 4//26 Coil ID: 9 Phantom: Oil filled sphere Test Date: 9//28 Model: 79WH S: 79WH # of Channels 2 T 6 x y.6 S Head TX gain: 2 R: R2: 29 nalysis of Test Image ormalized,6,2, %,64,9, % G E S i t e Signa 9

20 Coil: Knee 8ch Mfg.: Invivo Mfg. Date: 9//26 Coil ID: 92 Phantom: Knee phantom. Test Date: 9//28 Model: 44 S: U297 # of Channels 8 2 T x y.6 S HD Knee P TX gain: 68 R: R2: 29 nalysis of Composite Image 2,42 2,4 2,762 2,762,62, ormalized 66.% 66.% nalysis of Uncombined Images Ch ,978,878,89,472,6,9,79, % 87% 74% 6% 89% 72% 72% 99% % 86% 82% 74% 78% 76% 84% 7% Channel Channel 2 Channel Channel 4 Composites Channel Channel 6 Channel 7 Channel 8 G E S i t e Signa 2

21 Coil: Knee T/R Mfg.: Invivo Mfg. Date: Coil ID: 94 Phantom: 8ch Knee phantom and my wrist bottle in foot. Test Date: 9//28 Model: S: U2229 # of Channels 2 T 6 x y.6 S QUDKEE TX gain: 77 R: R2: 28 nalysis of Test Image ormalized 2,48,97, % 4,24 4,79, % 2,479,9, % 4,24 4,798, % Test Images G E S i t e Signa 2

22 Coil: Knee T/R Mfg.: Invivo Mfg. Date: Coil ID: 94 Phantom: 8ch Knee phantom and my wrist bottle in foot. Test Date: 9//28 Model: S: U2229 # of Channels 2 S 4 x y.6 S QUDKEE TX gain: 77 R: 9 R2: 29 nalysis of Test Image ormalized,89,44, % 4,92,274, %,898,446, % 4,9,27, % Test Images G E S i t e Signa 22

23 Coil: eurovascular rray 8 Ch Mfg.: USI Mfg. Date: 9//26 Coil ID: 92 Phantom: Twopiece V phantom set. Test Date: 9//28 Model: 2449 S: 82 # of Channels 8 2 T 6 x y.6 S a HDV Head TX gain: 6 R: R2: nalysis of Composite Image ormalized 8.8% 8.8% nalysis of Uncombined Images Ch % 78% 8% 68% 68% 7% % 97% % 82% 8% 72% 7% 7% % 98% Channel Channel 2 Channel Channel 4 Composites Channel Channel 6 Channel 7 Channel 8 G E S i t e Signa 2

24 Coil: eurovascular rray 8 Ch Mfg.: USI Mfg. Date: 9//26 Coil ID: 92 Phantom: Twopiece V phantom set. Test Date: 9//28 Model: 2449 S: 82 # of Channels 8 2 S 48 x y.6 S 2 HDV rray,22 96,2 9,6,879,6, nalysis of Composite Image TX gain: 77 R: R2: ormalized 8.2%.% 8.2%.% nalysis of Uncombined Images Ch , ,9, 897 7, % 66% 7% 79% % 8% 79% 98% % 6% 7% 79% % 92% 7% 99% Composites Channel Channel 2 Channel Channel 4 Composites Channel Channel 6 Channel 7 G E S i t e Signa 24

25 Coil: eurovascular rray 8 Ch Mfg.: USI Mfg. Date: 9//26 Coil ID: 92 Phantom: Twopiece V phantom set. Test Date: 9//28 Model: 2449 S: 82 # of Channels 8 2 C 48 x y.6 S. HDV rray ,,,2, nalysis of Composite Image TX gain: 77 R: R2: ormalized 8.4% 26.% 8.% 26.% nalysis of Uncombined Images Ch , % 87% 9% 8% 89% 9% % 8% % 74% 7% 27% 69% 82% % 6% Composites Channel Channel 2 Channel Channel 4 Composites Channel Channel 6 Channel 7 G E S i t e Signa 2

26 Coil: Shoulder Ch Mfg.: USI Mfg. Date: //26 Coil ID: 9 Phantom: Shoulder sphere Test Date: 9//28 Model: 244 S: 2 # of Channels 2 T 6 x y.6 S HD SHOULDER TX gain: 64 R: R2: 29 nalysis of Composite Image ,4, ormalized 4.% 4.% nalysis of Uncombined Images Ch , % 4% % % 7% % These axial images were acquired at the center of the phantom and OT at the + mark on the coil like last year. Channel Channel 2 Composites Channel G E S i t e Signa 26

27 Coil: Shoulder Ch Mfg.: USI Mfg. Date: //26 Coil ID: 9 Phantom: Shoulder sphere Test Date: 9//28 Model: 244 S: 2 # of Channels 2 C 6 x y.6 S HD SHOULDER TX gain: 64 R: R2: 29 nalysis of Composite Image 94 94,842, ormalized.8%.8% nalysis of Uncombined Images Ch ,787, % 48% % % 88% 78% Channel Channel 2 Composites Channel G E S i t e Signa 27

28 Coil: Spine 8 Ch Mfg.: USI Mfg. Date: 7//26 Coil ID: 899 Phantom: Dual T spine phantoms Test Date: 9//28 Model: 2442 S: 29 # of Channels 8 2 S 48 x y.6 S 2 HDCTL2 TX gain: 66 R: R2: 29 nalysis of Composite Image,4,4,9, ormalized.%.% nalysis of Uncombined Images Ch ,844,796,74,762,864, % 9% 7% 82% 2% 6% % 77% 7% 82% 74% 8% Composites Channel Channel 2 Channel Extra Pics Channel 4 Channel G E S i t e Signa 28

29 Coil: Spine 8 Ch Mfg.: USI Mfg. Date: 7//26 Coil ID: 899 Phantom: Dual T spine phantoms Test Date: 9//28 Model: 2442 S: 29 # of Channels 8 2 S 48 x y.6 S HDCTL4 PE P TX gain: 67 R: R2: 29 nalysis of Composite Image,4,4,26, ormalized 27.7% 27.6% nalysis of Uncombined Images Ch ,29,969 2,, % 9% 84% % % 96% 99% 94% Channel Channel 2 Composites Channel Channel 4 G E S i t e Signa 29

30 Coil: Spine 8 Ch Mfg.: USI Mfg. Date: 7//26 Coil ID: 899 Phantom: Dual T spine phantoms Test Date: 9//28 Model: 2442 S: 29 # of Channels 8 2 S 48 x y.6 S HDCTL6 PE P TX gain: 67 R: R2: 29 nalysis of Composite Image,6,6,62, ormalized 24.8% 24.7% nalysis of Uncombined Images Ch ,668 2,4 2,69 2, % 96% 89% % % 9% % 92% Channel Channel 2 Composites Channel Channel 4 G E S i t e Signa

31 Coil: Torso Mfg.: USI Mfg. Date: 2//26 Coil ID: 774 Phantom: Two large Torso phantoms... one of which is centered. Test Date: 9//28 Model: 244 S: 44 # of Channels 8 2 T 48 x y.6 S 8US TORSOP TX gain: 7 R: R2: 29 nalysis of Composite Image,69,68 2, 2, ormalized 4.% 4.% nalysis of Uncombined Images Ch ,28,2,29,74,72,64,2, % 82% 68% 98% 86% 8% 78% % % 4% 42% 88% 86% % 49% % Channel Channel 2 Channel Channel 4 Composites Channel Channel 6 Channel 7 Channel 8 G E S i t e Signa

32 Coil: Wrist HD Mfg.: Mayo Clinic Mfg. Date: Coil ID: 77 Phantom: Small wrist bottle. Test Date: 9//28 Model: BC S: HBR6 # of Channels 2 T x y.6 S HD WristCoil TX gain: R: R2: 29 nalysis of Test Image ormalized 8,2 8,872 7, % 8,4 8,879 7, % G E S i t e Signa 2

33 Coil: Wrist HD Mfg.: Mayo Clinic Mfg. Date: Coil ID: 77 Phantom: Small wrist bottle. Test Date: 9//28 Model: BC S: HBR6 # of Channels 2 C 2 x y.6 S HD WristCoil TX gain: 2 R: R2: 29 nalysis of Test Image ormalized 6,94 8,72, % 6,96 8,7, % G E S i t e Signa

34 Coil: Wrist HD Mfg.: Mayo Clinic Mfg. Date: Coil ID: 77 Phantom: Small wrist bottle. Test Date: 9//28 Model: BC S: HBR6 # of Channels 2 S 2 x y.6 S HD WristCoil TX gain: R: R2: 29 nalysis of Test Image ormalized 6,69 8,26, % 6,7 8,2, % G E S i t e Signa 4

35 ppendix : Magnet Homogeneity Field Maps GE Site Signa Excite HD.T central planes Measured September, 28 xial Right 2 nterior Left xial DIMER MI MX RGE PPM ME STDEV Posterior Coronal Right 2 Superior 4 2 Left Coronal DIMER MI MX RGE PPM ME STDEV Inferior Sagittal nteriror 2 2 Superior 2 2 Posterior Sagittal DIMER MI MX RGE PPM ME STDEV Inferior

36 ppendix : Magnet Homogeneity Field Maps GE Site Signa Excite HD.T Measured September, 28 Diameter cm cm 2cm 2cm 28cm cm GE Site Signa xial Field Plot 9//8..2. ppm cm from Isocenter (S/I) Diameter cm cm 2cm 2cm 28cm cm GE Site Signa Coronal Field Plot 9//8..2. ppm cm from Isocenter (P/) Diameter cm cm 2cm 2cm 28cm cm GE Site Signa Sagittal Field Plot 9//8..2. ppm cm from Isocenter (R/L)

37 xial Field Plots 2 2 H9 2 2 H8 2 2 H H H H H H H Isocenter F F F 2 2 F4 2 2 F 2 2 F6 2 F7 2 F8 2 F9 xial Field Plots

38 Coronal Field Plots Isocenter P P P P P P P7 2 2 P8 2 P9 Coronal Field Plots

39 Sagittal Field Plots R R R R R R R R R Isocenter L L L L L L L L L9 Sagittal Field Plots

40 ppendix B: RF Slice Profiles and Crosstalk Spin Echo / = 4/2 = 9.2 KHz nex = Scan time: :4 sk Upper=6.28 Lower=. sk Upper=6.4 Lower=2.84 sk 2. Upper=6.4 Lower= Slice =.46 Slice =.46 Slice =.4 sk 2 Upper=6.4 Lower=2.8 sk. Upper=.8 Lower=. sk. Upper=.9 Lower= Slice =.4 Slice =. Slice =.2 sk. Upper=2.29 Lower=48.4 sk.2 Upper=.4 Lower=48.6 sk. Upper=.94 Lower= Slice =. 6. Slice =. Slice =4.94 Slice thickness as a function of slice gap Slice Slice

41 ppendix B: RF Slice Profiles and Crosstalk Spin Echo T MEMP / = 6/ FLIP = 6 = 9.2 KHz nex = Scan time: 2:47 MEMP SK Upper=. Lower=2.6 MEMP SK Upper=.78 Lower=2. MEMP SK 2. Upper=.78 Lower= Slice =.4 Slice =.4 Slice =.44 MEMP SK 2 Upper=.9 Lower=2.6 MEMP SK. Upper=.66 Lower=2.4 MEMP SK. Upper=4.87 Lower= Slice =.4 Slice =.8 Slice =.2 MEMP SK. Upper=.6 Lower=. MEMP SK.2 Upper=. Lower=.2 MEMP SK. Upper=2.94 Lower= Slice =.9 6. Slice =.7 Slice =. Slice thickness as a function of slice gap Slice Slice

42 ppendix B: RF Slice Profiles and Crosstalk GRE In Phase / = 2/2. FLIP = 8 = 8.2 KHz nex = 4 Scan time: :2 InPhs sk Upper=4.9 Lower=2.84 InPhs sk Upper=4.4 Lower=.89 InPhs sk 2. Upper=4.4 Lower= Slice =4.8 Slice =4.9 Slice =4.82 InPhs sk 2 Upper=4.7 Lower=.24 InPhs sk. Upper=42.9 Lower=4.74 InPhs sk. Upper=4.47 Lower= Slice =4.76 Slice =4.8 Slice =4.89 InPhs sk. Upper=4.26 Lower=.44 InPhs sk.2 Upper=4.9 Lower=.7 InPhs sk. Upper=4.8 Lower= Slice = Slice =4.8 Slice =4.7 Slice thickness as a function of slice gap Slice Slice

43 GE Site Signa Coil Used: Head Split Top Sagittal Locator Length of phantom, end to end (mn 48± 2) 2 Slice Location # Resolution (.,.,.9 mm) Slice (fwhm in mm) Wedge (mm) Diameter (mm) (9±2) Slice Location # Diameter (mm) (9±2) Top Bottom Slice Location #7 4 6 Signal (mean only) Big ROI High Low 7 (>87.%) (mean ±std dev) Top Bottom Left Right 22 2 Ghosting Ratio (<2.%) (no spec) Calculated value.±.7 Low Con Detectability Slice Location #8 Slice Location #9 Slice Location # Slice Location # Total # of Spokes (>=9) Slice Location # Wedge (mm) Slice Position Error = + =.4% 2.%.6%.% = + = 48.4 The low uniformity is expected because I did not use any homogeneity correction. Test Date: 9//28 = calculated field ( /2) ( 2/2) ( 2/8) (Site T) (Site T2) CR T CR PD CR T2 Site T Site T % 76.8% 7.% 74.2% 72.9% 2.7 ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ±.4 2. ± ±..6%.%.%.%.2%

44 GE Site Signa parameters Test Date: 9//28 Coil Used: Head Split Top Test ID 9 Study Descrip tion Pulse (ETL) (ms) (ms) (cm) Phase Sample Ratio umber of Slices (mm) Slice S (ex) Freq Matrix Phase Matrix Band Width (khz) Scan Time (min:sec) CR T :9 CR PD Dual Echo :2 CR T2 Dual Echo :2 Site T :4 Site T2 F(2) :2 Magnet ID: 2 Coil ID: 9 TestID: 9

45 ppendix C: CR Phantom nalysis CR T

46 ppendix C: CR Phantom nalysis CR PD

47 ppendix C: CR Phantom nalysis CR T2

48 ppendix C: CR Phantom nalysis Site T

49 ppendix C: CR Phantom nalysis Site T2

50 ppendix D: Explanation of RF Coil Testing Report Introduction The primary goal of RF coil testing is to establish some sort of base line for tracking coil performance over time. The most common measure is the Signal to Ratio or. In addition, we can look at overall signal uniformity, ghosting level (or better lack of ghosting) and in the case of phased array coils we look at the of each and every channel and at symmetry between channels. Unfortunately, there is no single best method for measuring. Below I explain the different methods used and the rationale for each. One needs to measure the signal in the phantom (either mean or peak or both) and then divide that by the back noise. Measuring the signal is fairly straightforward, the noise can be more problematic. The simplest method is to measure the standard deviation () in the back air. However, MRI images are the magnitude of complex data. The noise in the underlying complex data is Gaussian but it follows a Rician distribution when the magnitude is used. The true noise can be estimated by multiplying the measured by.26. During the reconstruction process, most manufacturers perform various additional operations on the images, This could include geometric distortion correction, low pass filtering of the kspace data resulting in low signal at the edge of the images, RF coil intensity correction (PURE, CLER, SCIC, etc), and other processing during the combination of phased array data and parallel imaging techniques. ll of these methods distort the back noise making it impossible to obtain an accurate (and reproducible) estimate of the image noise in the air region. The alternative is to use a method which I shall refer to as the (ational Electrical Manufacturers ssociation) method. The signal in the phantom area is a sum of the proton signal and noise. Once the signal to noise ratio exceeds :, the noise in the magnitude image is effectively Gaussian. To eliminate the proton signal, you acquire an image twice and subtract them. The measured in the phantom region should now be the true times the square root of 2. When determining the using the method, calculate the mean signal of the average of the two source images then divide by.77 x the measured in the same area as the mean signal. Unfortunately, this doesn t always work. It is absolutely imperative that the RF channel scalings, both transmit and receive, be identical with both scans. ny ghosting in the system is not likely to repeat exactly for both scans and will cause a much higher. Finally, the phantom needs to be resting in place prior to the scan long enough for motion of the fluid to have died down. Depending on the size and shape of the phantom, this could take any where from to 2 minutes. One of the most common causes of ghosting is vibration from the helium coldhead. The best way to eliminate this artifact is to turn off the cold head, which will increase helium consumption. Because this vibration is periodic, the ghosting is usually of an over 2 (/2) nature. The affect inside the signal region of the phantom can be minimized by using a that is twice the diameter of the phantom (measured in the PE direction.) If the noise is to be measured in the air, then be sure to OT make measurements to either side of the phantom in the PE direction. Scan parameters also significantly affect measured. For most of the testing performed in this document I used a simple Spin Echo with a of, a of 2 and a slice thickness of mm and a receiver of.6 KHz. The was varied depending on the size of the coil and the phantom used. ll of the parameters used for each test can be found on each page immediately below the coil description. 2

51 Report Layout Each page of this report lists the data from a single test. The top third of the page describes the coil and phantom information, followed by the scan parameters used. The middle third contains the numbers measured and calculated results. This section will contain one table if the coil being tested is a single channel coil (i.e. quadrature or surface coils) and two tables if it is a multichannel phased array coil. The entries in the table will be described further below. The bottom section contains a few lines of comments (if necessary), a picture of the coil with the phantom as used for the testing and one or more of the images that were used for the measurements. There is usually one image for each composite image measurement and one image for each separate channel measurement. Each image shows the ROI (red line) where the mean signal was measured and two smaller ROIs (green lines) where the signal minimum and maximum was found. In the top left corner of each image is the mean signal in the large ROI. The bottom left corner contains the large ROI s area (in mm 2 ). The top right corner contains two numbers a mean and a standard deviation. If the method was used, then the top right corner will list the mean and of the large ROI (labeled ROI M and ROIsd) applied to the subtraction image. If the noise was measured in the back air the the numbers are labeled M and. Data Tables The meaning of most of the entries in the data table are should be self evident with a few exceptions. The first column in each table is labeled. In the composite analysis, this field may be empty or contain some sort of abbreviation to identify some aspect of the testing. Some possibilities are the letter for, for, L for Left, R for Right, C for CLER, oc for o CLER. In the Uncombined Image table, the label usually contains the channel number or similar descriptor. The column labeled will be either or SubSig which stands for Subtracted Signal, i.e. the method. Both tables contain a column for and which are the or signal divided by the of the noise scaled by either.26 () or.77 (). Composite Image Table: The final two columns in this table are ormalized and. It can be rather difficult to compare the performance of different coils particularly if different scan parameters are used. (Of course, it s even more difficult from one scanner to another.) I have standardized most of my testing to use a spin echo with a / of /2msec and a thickness of mm. The changes to depending on the size of the phantom used although I try to use a that is at least twice the diameter of the phantom as measured in the PE direction. For one reason or another, a change may be made in the scan parameters (either accidentally or intentionally such as turning on o Phase Wrap to eliminate aliasing, etc.). In order to make it easier to compare values I calculate a ormalized value. This value is theoretically what the would be if a of cm, x matrix, average, receiver of.6 KHz and slice thickness of mm had been used. Obviously, the final number is affected by the T/T2 values of the phantoms used as well as details of the coil and magnet field strength but it can be useful in certain situations. The value is defined by the CR as (maxmin)/(max+min). This is most important when looking at volume coils or for evaluating the effectiveness of surface coil intensity correction algorithms (such as PURE, CLER or SCIC). Uncombined Image Table: This table has two columns labeled and. When analyzing multichannel coils it is important to understand the relationship between the different channels, the inherent symmetry that usually exists between channels. In a 8 channel head or 4 channel torso phased array coil, all of the channels are usually have about the same. These two columns list how the (either or ) of each channel compares to the of the channel with the maximum value.

Philips Site Yearly Performance Evaluation Philips Intera 1.5T 2-Mar-08. Table of Contents

Philips Site Yearly Performance Evaluation Philips Intera 1.5T 2-Mar-08. Table of Contents Philips Site Yearly Performance Evaluation Philips Intera.5T 2Mar8 Table of Contents Summary and Signature Page 2 Specific Comments Site Information 4 Equipment Information 4 Table Position ccuracy 4 Magnetic

More information

Philips Site Yearly Performance Evaluation Philips Openview 16-Jan-08. Table of Contents

Philips Site Yearly Performance Evaluation Philips Openview 16-Jan-08. Table of Contents Philips Site Yearly Performance Evaluation Philips Openview 6-Jan-8 Table of Contents Summary and Signature Page 2 Specific Comments 3 Site Information 4 Equipment Information 4 Table Position Accuracy

More information

BACKGROUND. ** 78% of all MRI scanners have Image Quality problems. *** *** 25% of all Multi-Channel RF coils have at least one bad channel.

BACKGROUND. ** 78% of all MRI scanners have Image Quality problems. *** *** 25% of all Multi-Channel RF coils have at least one bad channel. Range of Results from over 534 ACR-mandated Annual MRI Performance Evaluations on over 204 Magnets from 8 Vendors Spanning a 10-year Period Moriel NessAiver, Ph.D. - Simply Physics - Baltimore, MD moriel@simplyphysics.com

More information

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 ACR MRI accreditation program MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics,

More information

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Advanced MSK MRI Protocols at 3.0T Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Outline Why High Field for MSK? SNR and Relaxation Times Technical Issues Example

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves Experiment 1 Goal: Examine the effect caused by changing flip angle on image contrast in a simple gradient echo sequence and derive T1-curves. Image

More information

1.5T HIGH FIELD SMALL ANIMAL MRI

1.5T HIGH FIELD SMALL ANIMAL MRI 1.5T HIGH FIELD SMALL ANIMAL MRI Designed Specifically for Veterinarians TECHNICAL GUIDE ADVANCING THE ART AND SCIENCE OF VETERINARY MRI The PetVet is the only high-field MRI system designed specifically

More information

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev. 1.0 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE DoseLab users may reference the following instructions to perform Tomotherapy Quality Assurance tests as recommended

More information

HETERONUCLEAR IMAGING. Topics to be Discussed:

HETERONUCLEAR IMAGING. Topics to be Discussed: HETERONUCLEAR IMAGING BioE-594 Advanced MRI By:- Rajitha Mullapudi 04/06/2006 Topics to be Discussed: What is heteronuclear imaging. Comparing the hardware of MRI and heteronuclear imaging. Clinical applications

More information

Pulse Sequence Design and Image Procedures

Pulse Sequence Design and Image Procedures Pulse Sequence Design and Image Procedures 1 Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant 2 A pulse sequence is a timing diagram designed with a series of RF pulses, gradients switching, and signal readout

More information

EE469B: Assignment 2 Solutions

EE469B: Assignment 2 Solutions EE469B Fall 26-7 RF Pulse Design for MRI EE469B: Assignment 2 s Due Thursday Oct 3 Introduction This assignment concerns the design of small-tip-angle 2D excitation pulses based on spiral k-space trajectories.

More information

NEMA Standards Publication MS (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging

NEMA Standards Publication MS (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging NEMA Standards Publication MS 1-2008 (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging Published by: National Electrical Manufacturers Association 1300 North

More information

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging JOURNAL OF MAGNETIC RESONANCE IMAGING 20:1046 1051 (2004) Technical Note The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging James W. Goldfarb, PhD* Purpose: To describe a known (but undocumented)

More information

Hardware. MRI System. MRI system Multicoil Microstrip. Part1

Hardware. MRI System. MRI system Multicoil Microstrip. Part1 Hardware MRI system Multicoil Microstrip MRI System Part1 1 The MRI system is made up of a variety of subsystems. the Operator Workspace Gradient Driver subsystem The Physiological Acquisition Controller

More information

2015 Spin echoes and projection imaging

2015 Spin echoes and projection imaging 1. Spin Echoes 1.1 Find f0, transmit amplitudes, and shim settings In order to acquire spin echoes, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week,

More information

Pulse Sequence Design Made Easier

Pulse Sequence Design Made Easier Pulse Sequence Design Made Easier Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant gurumri@gmail.com 1 2 Pulse Sequences generally have the following characteristics: An RF line characterizing RF Pulse applications

More information

MRI SYSTEM COMPONENTS Module One

MRI SYSTEM COMPONENTS Module One MRI SYSTEM COMPONENTS Module One 1 MAIN COMPONENTS Magnet Gradient Coils RF Coils Host Computer / Electronic Support System Operator Console and Display Systems 2 3 4 5 Magnet Components 6 The magnet The

More information

k y 2k y,max k x 2k x,max

k y 2k y,max k x 2k x,max EE225E/BIOE265 Spring 2012 Principles of MRI Miki Lustig Assignment 5 Due Feb 26, 2012 1. Finish reading Nishimura Ch. 5. 2. For the 16 turn spiral trajectory, plotted below, what is the a) Spatial resolution,

More information

Welcome at MR Achieva 3.0T TX. About the system. Index. 1.1 Introduction

Welcome at MR Achieva 3.0T TX. About the system. Index. 1.1 Introduction 1 Welcome at MR Achieva 3.0T TX Welcome to the Introduction of and e-learning module for the Achieva 3 Tesla TX-system. This CBT is setup for those engineers who will soon receive a Achieva 3Tesla TX system.

More information

2014 M.S. Cohen all rights reserved

2014 M.S. Cohen all rights reserved 2014 M.S. Cohen all rights reserved mscohen@g.ucla.edu IMAGE QUALITY / ARTIFACTS SYRINGOMYELIA Source http://gait.aidi.udel.edu/res695/homepage/pd_ortho/educate/clincase/syrsco.htm Surgery is usually recommended

More information

Background (~EE369B)

Background (~EE369B) Background (~EE369B) Magnetic Resonance Imaging D. Nishimura Overview of NMR Hardware Image formation and k-space Excitation k-space Signals and contrast Signal-to-Noise Ratio (SNR) Pulse Sequences 13

More information

6.S02 MRI Lab Acquire MR signals. 2.1 Free Induction decay (FID)

6.S02 MRI Lab Acquire MR signals. 2.1 Free Induction decay (FID) 6.S02 MRI Lab 1 2. Acquire MR signals Connecting to the scanner Connect to VMware on the Lab Macs. Download and extract the following zip file in the MRI Lab dropbox folder: https://www.dropbox.com/s/ga8ga4a0sxwe62e/mit_download.zip

More information

Page 1 of 9. Protocol: adult_other_adni3_study_human_ge_3t_25w_ _ _1. 3 Plane Localizer. 3 Plane Localizer PATIENT POSITION

Page 1 of 9. Protocol: adult_other_adni3_study_human_ge_3t_25w_ _ _1. 3 Plane Localizer. 3 Plane Localizer PATIENT POSITION 3 Localizer FOV 26.0 Slice Thickness 5.0 Slice Spacing 0.0 Freq 256 Phase 128 3-PLANE 3 Localizer Unswap Phase Correction Gradient Echo Imaging Options Seq, Fast Recon All Images Contrast Yes/ 3 Localizer

More information

MR Basics: Module 8 Image Quality

MR Basics: Module 8 Image Quality Module 8 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Huaiqun Guan,

More information

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Magnetic Resonance Imaging Spatial

More information

TimTX TrueShape. The parallel transmit architecture of the future. Answers for life.

TimTX TrueShape.  The parallel transmit architecture of the future. Answers for life. www.siemens.com/trueshape TimTX TrueShape The parallel transmit architecture of the future. The product/feature (mentioned herein) is not commercially available. Due to regulatory reasons its future availability

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

EE469B: Assignment 1 Solutions

EE469B: Assignment 1 Solutions EE469B Fall 26-7 RF Pulse Design for MRI EE469B: Assignment Solutions Due Thursday Oct 6 Introduction This assignment concerns typical Fourier transform designs of excitation pulses. This includes designing

More information

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

TITLE: Prostate Cancer Detection Using High-Spatial Resolution MRI at 7.0 Tesla: Correlation with Histopathologic Findings at Radical Prostatectomy

TITLE: Prostate Cancer Detection Using High-Spatial Resolution MRI at 7.0 Tesla: Correlation with Histopathologic Findings at Radical Prostatectomy Award Number: W81XWH-11-1-0253 TITLE: Prostate Cancer Detection Using High-Spatial Resolution MRI at 7.0 Tesla: Correlation with Histopathologic Findings at Radical Prostatectomy PRINCIPAL INVESTIGATOR:

More information

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements MR Data for Treatment Planning: Issues, Protocol Optimization, and s (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements TG-117 Use of MRI Data in Treatment Planning

More information

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010 (N)MR Imaging Lab Course Script FMP PhD Autumn School Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder Date: November 3rd, 2010 1 Purpose: Understanding the basic principles of MR imaging

More information

The ACR magnetic resonance accreditation phantom (ACR MRAP) has been designed to examine a broad range of instrument parameters.

The ACR magnetic resonance accreditation phantom (ACR MRAP) has been designed to examine a broad range of instrument parameters. OVERVIEW OF THE ACR MRI ACCREDITATION PHANTOM Geoffrey D. Clarke, Ph.D. University of Texas Southwestern Medical Center at Dallas email: GEOFFREY.CLARKE@MAIL.SWMED.EDU INTRODUCTION The ACR magnetic resonance

More information

MR Advance Techniques. Flow Phenomena. Class II

MR Advance Techniques. Flow Phenomena. Class II MR Advance Techniques Flow Phenomena Class II Flow Phenomena In this class we will explore different phenomenona produced from nuclei that move during the acquisition of data. Flowing nuclei exhibit different

More information

Weber State University Radiologic Technology 4603

Weber State University Radiologic Technology 4603 Weber State University Radiologic Technology 4603 MRI Physics and Instrumentation Instructor: Rex T. Christensen MHA R.T. (R) (MR) (CT) (ARRT) CIIP Contact Info: E-mail: rexchristensen@weber.edu Phone:

More information

ISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain

ISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain Available online at www.derpharmachemica.com ISSN 0975-413X CODEN (USA): PCHHAX Der Pharma Chemica, 2016, 8(17):15-20 (http://derpharmachemica.com/archive.html) The role of in increasing resolution in

More information

Image Quality/Artifacts Frequency (MHz)

Image Quality/Artifacts Frequency (MHz) The Larmor Relation 84 Image Quality/Artifacts (MHz) 42 ω = γ X B = 2πf 84 0.0 1.0 2.0 Magnetic Field (Tesla) 1 A 1D Image Magnetic Field Gradients Magnet Field Strength Field Strength / Gradient Coil

More information

SIEMENS MAGNETOM Skyra syngo MR D13

SIEMENS MAGNETOM Skyra syngo MR D13 Page 1 of 12 SIEMENS MAGNETOM Skyra syngo MR D13 \\USER\CIND\StudyProtocols\PTSA\*ep2d_M0Map_p2_TE15 TA:7.9 s PAT:2 Voxel size:2.5 2.5 3.0 mm Rel. SNR:1.00 :epfid Properties Routine Contrast Prio Recon

More information

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare GE Healthcare Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare There is excitement across the industry regarding the clinical potential of a hybrid

More information

Evaluation of MWT Materials Accusorb MRI Shield. Dr. E. Kanal, Department of Radiology, UPMC. Wednesday, April 21, 2010

Evaluation of MWT Materials Accusorb MRI Shield. Dr. E. Kanal, Department of Radiology, UPMC. Wednesday, April 21, 2010 Evaluation of MWT Materials Accusorb MRI Shield Dr. E. Kanal, Department of Radiology, UPMC Wednesday, April 21, 2010 Background: In almost all x-ray and ultrasonographic imaging examinations, in order

More information

Overview of Safety Code 35

Overview of Safety Code 35 Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)

More information

PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner

PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner September, 2017 Results submitted to Physics in Medicine & Biology Negar Omidvari 1, Jorge Cabello 1, Geoffrey Topping

More information

Methods. Experimental Stimuli: We selected 24 animals, 24 tools, and 24

Methods. Experimental Stimuli: We selected 24 animals, 24 tools, and 24 Methods Experimental Stimuli: We selected 24 animals, 24 tools, and 24 nonmanipulable object concepts following the criteria described in a previous study. For each item, a black and white grayscale photo

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Melissa Martin, M.S., FAAPM, FACR, FACMP AAPM Annual Meeting - Philadelphia July 19, 2010 MO-B-204C-1 Educational Objectives

More information

SIGNA Pioneer: Ultra High Efficiency Gradient System Advancing the gradient technology curve

SIGNA Pioneer: Ultra High Efficiency Gradient System Advancing the gradient technology curve GE Healthcare SIGNA Pioneer: Ultra High Efficiency Gradient System Advancing the gradient technology curve NEW TECHNOLOGY 40W Watts spec is irrelevant. 4W LED bulb delivers same brightness as 40W incandescent

More information

Using Frequency Diversity to Improve Measurement Speed Roger Dygert MI Technologies, 1125 Satellite Blvd., Suite 100 Suwanee, GA 30024

Using Frequency Diversity to Improve Measurement Speed Roger Dygert MI Technologies, 1125 Satellite Blvd., Suite 100 Suwanee, GA 30024 Using Frequency Diversity to Improve Measurement Speed Roger Dygert MI Technologies, 1125 Satellite Blvd., Suite 1 Suwanee, GA 324 ABSTRACT Conventional antenna measurement systems use a multiplexer or

More information

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter MR-guided RT in Heidelberg Funded by the German Research Foundation (DFG) Heidelberg consortium received

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey Implementation of the 2016 ACR Digital Mammography QC Manual Medical Physicist Mammography Equipment Evaluation and Annual Survey Eric A Berns, PhD, FACR Radiologist Responsibilities Radiologist Responsibilities

More information

Delivering Better Patient Care with SIGNA Architect

Delivering Better Patient Care with SIGNA Architect Delivering etter Patient Care with SIGNA Architect As a regional leader in outpatient-based diagnostic imaging, the radiologists and staff at Inova Fairfax MRI Center are focused on one thing: the patient.

More information

INSTRUCTIONS FOR THE CAROTID MRI COMPLETION FORM, CMR, Version B, 12/08/2005

INSTRUCTIONS FOR THE CAROTID MRI COMPLETION FORM, CMR, Version B, 12/08/2005 INSTRUCTIONS FOR THE CAROTID MRI COMPLETION FORM, CMR, Version B, 12/08/2005 I. General Instructions: This form is used to document the conduct and completeness of the MRI examination. The MRI examination

More information

The Script of ZST + Presentation. MIS Upstream Marketing Team [ 日期 ]

The Script of ZST + Presentation. MIS Upstream Marketing Team [ 日期 ] 1 The Script of ZST + Presentation MIS Upstream Marketing Team [ 日期 ] 1 The Script of ZST + Presentation Since Mindray was founded to develop ultrasound business, core technology has always been the engine

More information

COMPUTER PHANTOMS FOR SIMULATING ULTRASOUND B-MODE AND CFM IMAGES

COMPUTER PHANTOMS FOR SIMULATING ULTRASOUND B-MODE AND CFM IMAGES Paper presented at the 23rd Acoustical Imaging Symposium, Boston, Massachusetts, USA, April 13-16, 1997: COMPUTER PHANTOMS FOR SIMULATING ULTRASOUND B-MODE AND CFM IMAGES Jørgen Arendt Jensen and Peter

More information

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP MR Data for Treatment Planning and Stereotactic Procedures: Sources of Distortion, Protocol Optimization, and Assessment (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements

More information

Lab 8 6.S02 Spring 2013 MRI Projection Imaging

Lab 8 6.S02 Spring 2013 MRI Projection Imaging 1. Spin Echos 1.1 Find f0, TX amplitudes, and shim settings In order to acquire spin echos, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week, but these

More information

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

1. Patient size AEC. Large Patient High ma. Small Patient Low ma Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description

More information

EE225E/BIOE265 Spring 2011 Principles of MRI. Assignment 6 Solutions. (y 0 + vt) dt. 2 y 0T + 3 )

EE225E/BIOE265 Spring 2011 Principles of MRI. Assignment 6 Solutions. (y 0 + vt) dt. 2 y 0T + 3 ) EE225E/BIOE265 Spring 211 Principles of MRI Miki Lustig Handout Assignment 6 Solutions 1. Nishimura 6.7 (Thanks Galen!) a) After the 9 y pulse, the spin is in the ˆx direction (using left-handed rotations).

More information

EE225E/BIOE265 Spring 2014 Principles of MRI. Assignment 6. Due Friday March 7th, 2014, Self Grading Due Monday March 10th, 2014

EE225E/BIOE265 Spring 2014 Principles of MRI. Assignment 6. Due Friday March 7th, 2014, Self Grading Due Monday March 10th, 2014 EE225E/BIOE265 Spring 2014 Principles of MRI Miki Lustig 1. Read Nishimura Ch. 6 Assignment 6 Due Friday March 7th, 2014, Self Grading Due Monday March 10th, 2014 2. Nishimura assignment 6.5 3. Mimimum-Phase

More information

Medical Imaging (EL582/BE620/GA4426)

Medical Imaging (EL582/BE620/GA4426) Medical Imaging (EL582/BE620/GA4426) Jonathan Mamou, PhD Riverside Research Lizzi Center for Biomedical Engineering New York, NY jmamou@riversideresearch.org On behalf of Prof. Daniel Turnbull Outline

More information

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Dr John Ridgway Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Physics for clinicians: Part I Journal of Cardiovascular Magnetic Resonance 2010, 12:71 http://jcmr-online.com/content/12/1/71

More information

ACRIN 6686 / RTOG 0825

ACRIN 6686 / RTOG 0825 ACRIN 6686 (RTOG 0825) Advanced MRI Imaging Manual ACRIN 6686 / RTOG 0825 A phase III double blind placebo controlled trial of conventional chemoradiation and adjuvant temozolomide plus bevacizumab vs

More information

FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR

FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May 2010 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics, Inc. Physicist's Role in the

More information

Magnetic Resonance Imaging Principles, Methods, and Techniques

Magnetic Resonance Imaging Principles, Methods, and Techniques Magnetic Resonance Imaging Principles, Methods, and Techniques Perry Sprawls Jr., Emory University Publisher: Medical Physics Publishing Corporation Publication Place: Madison, Wisconsin Publication Date:

More information

Small, Low Power, High Performance Magnetometers

Small, Low Power, High Performance Magnetometers Small, Low Power, High Performance Magnetometers M. Prouty ( 1 ), R. Johnson ( 1 ) ( 1 ) Geometrics, Inc Summary Recent work by Geometrics, along with partners at the U.S. National Institute of Standards

More information

Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging

Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging Poster No.: C-0709 Congress: ECR 2014 Type: Scientific

More information

Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions.

Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions. Lesson 06: Pulse-echo Imaging and Display Modes These lessons contain 26 slides plus 15 multiple-choice questions. These lesson were derived from pages 26 through 32 in the textbook: ULTRASOUND IMAGING

More information

Accurate Utility Depth Measurements Using the Spar 300

Accurate Utility Depth Measurements Using the Spar 300 Accurate Utility Depth Measurements Using the Spar 3 This Application Note addresses how to obtain accurate subsurface utility depths using the model-based methods employed by the Spar 3. All electromagnetic

More information

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo Gradient Spoiling Average balanced SSFP magnetization Reduce sensitivity to off-resonance FFE, FISP, GRASS, GRE, FAST, Field Echo 1 Gradient-Spoiled Sequence (GRE, FFE, FISP, GRASS) RF TR G z G y G x Signal

More information

Magnetic Resonance Imaging

Magnetic Resonance Imaging Magnetic Resonance Imaging Principles, Methods, and Techniques Perry Sprawls, Ph.D., FACR, FAAPM, FIOMP Distinguished Emeritus Professor Department of Radiology Emory University Atlanta, Georgia Medical

More information

Improving high-field MRI using parallel excitation

Improving high-field MRI using parallel excitation review Improving high-field MRI using parallel excitation MRI at high magnetic field strengths promises to deliver clearer images of the body s structure and function. However, high-field MRI currently

More information

Coding & Signal Processing for Holographic Data Storage. Vijayakumar Bhagavatula

Coding & Signal Processing for Holographic Data Storage. Vijayakumar Bhagavatula Coding & Signal Processing for Holographic Data Storage Vijayakumar Bhagavatula Acknowledgements Venkatesh Vadde Mehmet Keskinoz Sheida Nabavi Lakshmi Ramamoorthy Kevin Curtis, Adrian Hill & Mark Ayres

More information

Physics 2310 Lab #5: Thin Lenses and Concave Mirrors Dr. Michael Pierce (Univ. of Wyoming)

Physics 2310 Lab #5: Thin Lenses and Concave Mirrors Dr. Michael Pierce (Univ. of Wyoming) Physics 2310 Lab #5: Thin Lenses and Concave Mirrors Dr. Michael Pierce (Univ. of Wyoming) Purpose: The purpose of this lab is to introduce students to some of the properties of thin lenses and mirrors.

More information

SIGNA Explorer Lift revives our MR

SIGNA Explorer Lift revives our MR Seiji Shiotani, MD, PhD Seirei Fuji Hospital in Fuji City, Shizuoka, Japan Masayoshi Sugimura Seirei Fuji Hospital in Fuji City, Shizuoka, Japan SIGN Explorer Lift revives our MR The clinical usefulness

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

Digital Imaging CT & MR

Digital Imaging CT & MR Digital Imaging CT & MR January 22, 2008 Digital Radiography, CT and MRI generate images in a digital format What is a Digital Image? A digital image is made up of picture elements, pixels row by column

More information

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook DIAGNOSTIC ACCREDITATION PROGRAM Radiology and CT Quality Control Procedures Workbook Quality Control Procedures Radiography/CR/DR Safety Code 35 Summary For more detail about each quality control (QC)

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

Stepping Motor. Applications. Structure and operation. Code names. Mobile equipment Digital cameras, Mobile equipments, PDA, etc.

Stepping Motor. Applications. Structure and operation. Code names. Mobile equipment Digital cameras, Mobile equipments, PDA, etc. Stepping Motor pplications Mobile equipment Digital cameras, Mobile equipments, PD, etc. Office automation equipment Printers, facsimiles, Typewriters, Photocopiers, FDD head drives, CD-ROM pickup drives,

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

Experience in implementing continuous arterial spin labeling on a commercial MR scanner

Experience in implementing continuous arterial spin labeling on a commercial MR scanner JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 1, WINTER 2005 Experience in implementing continuous arterial spin labeling on a commercial MR scanner Theodore R. Steger and Edward F. Jackson

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Magnetic Resonance Research Facility (MRRF) Resources

Magnetic Resonance Research Facility (MRRF) Resources Magnetic Resonance Research Facility (MRRF) Resources The Magnetic Resonance Research Facility (MRRF) has scanners located in both the hospital and research buildings on the campus of the University of

More information

American College of Radiology MR Accreditation Program. Testing Instructions

American College of Radiology MR Accreditation Program. Testing Instructions American College of Radiology MR Accreditation Program Testing Instructions (Revised May 28, 2017) This guide provides all of the instructions necessary for clinical tests, phantom tests and general submission

More information

Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution

Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Cynthia H. McCollough, PhD, DABR, FAAPM, FACR Director, CT Clinical Innovation Center Professor of Medical Physics and

More information

SmartExam helps to banish most repeat knee studies at Desert Medical Imaging

SmartExam helps to banish most repeat knee studies at Desert Medical Imaging I s s u e 3 2 - J u l y 2 0 0 7 F i e l d Strength Publication for the Philips MRI Community SmartExam helps to banish most repeat knee studies at Desert Medical Imaging Total scan automation boosts efficiency,

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

EE225E/BIOE265 Spring 2012 Principles of MRI. Assignment 7. Due March 16, 2012

EE225E/BIOE265 Spring 2012 Principles of MRI. Assignment 7. Due March 16, 2012 EE225E/BIOE265 Spring 2012 Principles of MRI Miki Lustig Assignment 7 Due March 16, 2012 1. From Midterm I 2010: You ve just programmed up your first 2DFT pulse sequence, and are trying it out on the scanner.

More information

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC? Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM

More information

CARING DESIGN. PRACTICAL TECHNOLOGY.

CARING DESIGN. PRACTICAL TECHNOLOGY. GE Healthcare CARING DESIGN. PRACTICAL TECHNOLOGY. Brivo * MR355 1.5T Inspire ITS ELEGANT DESIGN AND WARM LIGHT ARE INTRIGUING. Brivo MR355 Inspire product manager Every piece of equipment you own represents

More information

Ihor TROTS, Andrzej NOWICKI, Marcin LEWANDOWSKI

Ihor TROTS, Andrzej NOWICKI, Marcin LEWANDOWSKI ARCHIVES OF ACOUSTICS 33, 4, 573 580 (2008) LABORATORY SETUP FOR SYNTHETIC APERTURE ULTRASOUND IMAGING Ihor TROTS, Andrzej NOWICKI, Marcin LEWANDOWSKI Institute of Fundamental Technological Research Polish

More information

Field Simulation Software to Improve Magnetic Resonance Imaging

Field Simulation Software to Improve Magnetic Resonance Imaging Field Simulation Software to Improve Magnetic Resonance Imaging a joint project with the NRI in South Korea CST Usergroup Meeting 2010 Darmstadt Institute for Biometry and Medicine Informatics J. Mallow,

More information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information Ordering Information Please contact us if you have any questions or if you would like a quote or delivery schedule regarding the Catphan phantom. phone 800-525-1190, or 518-692-1190 fax 518-692-3329 mail

More information

Bar code Verifier Conformance Specifications. Using the INTEGRA-9000

Bar code Verifier Conformance Specifications. Using the INTEGRA-9000 Bar code Verifier Conformance Specifications Using the INTEGRA-9000 From: Label Vision Systems, Inc. (LVS) Document Created: 4-1998 Edit / Print Date: 2-2003 C:\My Documents\INTEGRA -9000 VERIFIER CONFORMANCE

More information

Tomophan TSP004 Manual

Tomophan TSP004 Manual T h e P h a n t o m L a b o r a t o r y 1 Tomophan TSP004 Manual Copyright 2016 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order and

More information

H 2 O and fat imaging

H 2 O and fat imaging H 2 O and fat imaging Xu Feng Outline Introduction benefit from the separation of water and fat imaging Chemical Shift definition of chemical shift origin of chemical shift equations of chemical shift

More information

MRI RF-Coils. Innovation with Integrity. Highest sensitivity for your preclinical MRI and MRS applications. Preclinical Imaging

MRI RF-Coils. Innovation with Integrity. Highest sensitivity for your preclinical MRI and MRS applications. Preclinical Imaging MRI RF-Coils Highest sensitivity for your preclinical MRI and MRS applications Innovation with Integrity Preclinical Imaging Molecular and Preclinical Imaging Preclinical magnetic resonance imaging of

More information

8800SX DMR Repeater Test Option 06

8800SX DMR Repeater Test Option 06 8800SX DMR Repeater Test Option 06 DMR Repeater Test Option The DMR Repeater test option allows testing of a DMR Repeater that is in conventional DMR Mode. Trunking or analog configurations are not supported.

More information