ABSTRACT 1. INTRODUCTION

Size: px
Start display at page:

Download "ABSTRACT 1. INTRODUCTION"

Transcription

1 High-resolulion, multipleoptical mode confocal microscope: I. System design, image acquisition and 3D visualization Carol J. Cogswell, Kiezan G. Larkin, John W. OByrne, and Matthew R. Amison Department of Physical Optics, School of Physics, University of Sydney, NSW 2006, Australia ABSTRACT We have designed and constructed an experimental confocal spccimen-miing microscope which has the capability of producing high resolution 3D images in a variety of optical modes, many of which are not currently available on commercial confocal microscopes. The transmission Nomarski differential interference contrast (DIC) mode is particularly interesting because it can be utilized to image small changes in refrative index within complex biological specimens which are transparent in standard brightfield. The three-color reflection configuration can produce a color 3D image, which means that stained or pigmented objects will be similar in appearance to images obtained from conventional white light micmscopes which makes them more recognizable. 1. INTRODUCTION Confocal microscopes have been under development for over a decade and commercial instruments are now commonly utilized in a variety of applications in the biological and materials sciences as well as for industrial inspection. However, commercial instruments have largely been developed for specific applications and disciplines, which have resulted in the instru.. ments having the capability of high-resolution imaging in only a few of the optical modes common to conventional full-field microscopes. This is a great disadvantage for many users of the instruments, because the confocal images xoduced are often so different in appearance to the conventional microscope images, that the biologist or materials scientist finds it extremely difficult to identify and interpret what is meaningful information even though the specimen can be viewed as a full 3D reconstruction. Perhaps the most widely used confocal configuration is fluorescence which has a very large nwnber of applications in the biological sciences. This mode produces images that are the most similar in appearance to those of the equivalent conventional microscope (ie. conventional fluorescence) and hence kes not present many problems for the typical user. This fxt alone may prove to be the major reason for its widespread popularity. However, when moving to the other most common commercial confocal optical mode, reflection brightfield, the problems of locating the areas of interest within the specimen volume and the interpretation of the resulting images are greatly increased. For example, a typical briglnfield reflection 2D confocal image will have bright pixels from the in-focus, reflective regions while the remaining areas of the image will be dark. In a true confocal set-up (ie. one where the detector pinhole is of a size to sample only the central 25percent of the image Airy disc) the resulting image will often have only a small proportion of bright pixels against a black bkground.t Hence, eh of the 2D images or "optical sections" will have a very different appearance to the conventional microscope view, especially (as in the field of biological microscopy) when the user is secustomed to observing the specimen in a transmission mode where the background is generally light and regions of the specimen appear as various shades of gray. In addition to the identification and interpretation problems encountered, a more fundamental limitation is that commercial confocal microscopes cannot, in general, image phase information in 3D. This is in contrast to conventional biological microscopes in which modes such as Zernike phase contrast and Nomarski differential interference contrast (DIC) are fundamental for the study of living or unstained materials. Furthermore, most confocal systems cannot give an accurate representation of the spectral response of the specimen (ie. relative absorption and reflection of various wavelengths throughout the visible spectrum) which again is in variance to the common use of a white light source and biological stains to provide full-color viewing in a conventional microscope. A principal reason why commercial confocal systems do not have the range of optical modes typical of a high quality conventional system is that these additional configurations are much more difficult to implement and utilize than the fluorescence mode. A true confocal design uses a small aperture pisced at the image plane in front of a photodetector. The correct positioning of this aperture or "pinhole" is crucial if high resolution images are to be obtained. It has been shown in our recent studies of multiple wavelength (ie. color) confocal brightfield or fluorescence imaging1 that optical effects such as chromatic and spherical aberration can so dominate the performance as to make the images worthless. In addition, in all transmission modes, we have shown that refraction alone can cause the focussed image spot to "wander" both axially and transversely from the position of the fixed detector pinlle which means that almost continual alignment correction is necessary to produce useable images. t This is true even if one ignores the problems of coherent speckle which are frequently also present when laser ifiumination is used and which further compound the problems of identification and interpretation of object features. 48 /SPIE Vol Three-Dimensional Microscopy (1994) O /94/$6.OO

2 2. A HIGH RESOLUTION, MULTIPLE OPTICAL MODE CONFOCAL MICROSCOPE In order to produce a much more versatile confocal microscope for 3D visualization of specimens, we have developed a system that has the capability of imaging in both a reflection and a transmission configuration and in addition has the capability of utilizing a range of optical modes such as 3-color (RGB) brightfleld, Nomarski DIC and fluorescence. Table 1 provides a summary of all of the optical modes currently undez development for high resolution imaging on our confocal microscope and indicates whether they can be used in reflection or transmission Design features Optical Mode Reflection Transmission Brightfield (single color) Brightfield (3-color ROB) Nomarski DIC I I Ruorescerxe Polarizing I I Table 1: Optical modes of the high resolution, specimen-scanning confocal microscope. The multiple optical mode capability of our experimental microscope has been realized primarily through one fundamental design feature: the illuminating and detected light remains stationary on the optical axis rather thanbeing transversely scanned as it is in most commercial confocal designs. This single feature greatly simplifies the overall optical design and limits all aberrations to an on-axis condition, which means they can be more easily corrected. To produce a 2D image, the specimen (rather than the beam) is mechanally scanned in a raster pattern. The microscope uses very high NA objective lenses (1.4 or 1.3) in order to obtain images having the best possible resolution. Magnification is changed simply by altering the distance over which the specimen is scanned Three-color reflection configuration The ability to image in color is a powerful tool for a great many applications in biology, geology and materials science and, because of this, white-light ifiumination has been used in conventional microscopes for centuries. In the twentieth century, due to the design and development of precision multi-element objective lenses ("apochromats"), it has become possible to obtain images that are fully corrected for lateral chromatic aberration throughout the entire visible spectrum. ll in addition, the apechromatic objective is corrected for curvature of field ("plan") then its axial resolution for visible wavelengths will also be good. However, in order to configure a confocal microscope for high resolution color imaging, even when using the most highly corrected planapochromat objectives, it is usually necessary to further correct for chromatic aberration, primarily because the demands for precise axial alignment are greater in confocal microscopes than in the conventional light microscopes for which the objectives are designed. Because of the constraints imposed by objectives, our experiments with color imaging in a confocal microscope3 indicate that axial chromatic aberration causes severe image degradation if a single multi-wavelength light source and a single detector is used in a confocal system. This is the case even when the most highly corrected objectives are used and despite the on-axis nature of our confocal design which virtually eliminates the lateral chromatic aberration usually present in beam-scanning microscopes. In order to overcome the axial chromatic aberration problem, we have chosen a compromise configuration, one which utilizes three separate ifiuminating wavelengths (red, green and blue), esch with its own individually-aligned ifiumination and detection path (Fig. 1). This allows three morochromatic images to be scquired simultaneously while the specimen is scanned. Finally, a color image is produced by writing the images to the red, green and blue channels of a display monitor. The three laser wavelengths (HeNe 633nm, frequcucy-deubled NdYAG 532nm, Argon ion 458mn) provide a reasonable color representation when viewing objects that have fairly broad spectral reflectivity across regions of the Visible spectrum, which is the case for many naturally occurring pigments or artificial dyes commonly used in briglnfleld microscopy. In acklition, preliminary results have indicated that our confocal color reflection mode can assist in the location and identification of immunogold particles, 5-10 nm in diameter. These are commonly used probes for labelling biological materials. However, in practice, the gold particles usually require additional enhancement with silver in order to be made large enough for viewing with a conventional light microscope. Our initial studies using the high-resolution color confocal system indicate that sub-wavelength gold particles can be identified by nature of their increased scattering of the green wavelength thus alleviating the need for the silver enhancement process.3 SPIE VoL Three-Dimensional Microscopy (1 994) I 49

3 Pir*oIe Macirtosh Quadra Compiler with A/D Converter Bosrd z x&y Specimen Stage with 2 pairs c Loudspeakers for x, y Scanning -3 Fig. 1: Schematic layout of the multiple opticai mode confocal scanning laser microscope for transmission and reflection imaging. From Eironka (WP = Wdlaston pdsm for Nornarskl DIC Imaging mode) SO ISPIE Vol Three-Dimensional Microscopy (1994)

4 2.3. Transmission brightfield and Nomarski DIC configurations Figure 1 also shows the layout of the transmission path of our confocal microscope. In order to obtain high resolution transmission images, several factors must be considered with regard to the overall microscope design and the transmission image formation process. For example, a second high NA objective must be inserted and ecisely aligned beneath the object plane and the specimen itself must be mounted between two covergiasses. Preliminary investigations2 also show that refraction by the specimen will cause the focussed image spot to wander both Iatezally and axially from the center of the detector pinhole. This factor tends to make high resolution transmission imaging virtually impossible for all but the least complex and thinnest of specimens if a standard fixed pinhole detector arrangement is used. Several mechanisms for tracking the wandering spot and correcting for this problem are currently wxler investigation in our gmup.'1 One technique uses a multiple-element array detector which can detect the position of the image spot and then send the information to a tilting mirror system that repositions the spot onto the center of the detector pinhole. Another method utilizes a large area detector such as a high speed CCD array to sample the image Airy disc and explores whether sampling the maximum intensity or the average intensity value at the centroid of the pattern will give the equivalent to a normal confocal detector with a pinhole aperture. In addition to investigating the xoblem of the wandering image spot, our studies have addressed a more fundamental issue: that of whether or not confocal transmission imaging modes actually give an improvement in axial resolution over conventional transmission microscopes. In order to conduct comparative studies, our microscope has been configured with two photomultiplier detectors (PMTs) in the transmission path. The first detector has a pinhole aperture at the image plane (confocal case) and the second utilizes the entire large area of the PMT detector (no pinhole) which simulates a conventional microscope. Using a beam spliuer, the light from the transmission objective is split into two image paths in ordez that a pair of images, one confocal and one conventional, can be acquired simultaneously for comparison jirposes. 3. EXPERIMENTAL RESULTS: CONFOCAL TRANSMISSION USING BRIGHTFIELD AND NOMARSKI DIC In a recent study of thin biological preparations (plant chromosomes) we acquired high resolution confocal images in both transmission brightfield and transmission Nomarski DIC in order to compare the potential of eh configuration for imaging lightly stained specimens. For each of these optical mocks, two simultaneous through-focus series of images were acquired one using a fixed pinhole in front of a photomultiplier tube which gave a confocal configuration and another using no pinhole in front of the PMT which, in effect, produced a conventional microscope image. For these initial experiments, simple manual methods for maintaining optical alignment appropriate to confocal transmission were utilized which were based on qualitative assessment of the scanned image and of the Airy disc in the plane of the pinhole. This required realigning the pinhole for eh step change in focus which gave a more consistent alignment for all the images in a through-focus stack but was still a compromise with respect to accurate alignment of every pixel in each 2D image. In both the brightfield and Nomarski DIC experiments, the confocal images showed much improved contrast as well as some subtle improvement in axial resolution over the conventional case. In addition, in this particular chromosome study, the Nomarski DIC images appeared to have much more high spatial frequency information (ie. more fine structure was discernible) than the brightfield images. This is largely due to the well-known capaity of Nomarski DIC to image phase in addition to the amplitude information standardly imaged by a brightfield mode.5 Thus, slight changes in refractive index in the unstained regions of the preparation are able to be visualized by DIC. Figures 2(a-b) and 3(a-b) show a comparison of confocal versus conventional microscope images of chromosomes using transmission Nomarski DIC, where Fig 3(a-b) represents a plane of focus 1.4 microns below Fig 2(a-b). The tendency of the confocal image to go out of focus more rapidly than the conventional microscope image is apparent This means that the confocal optical path rejects out of focus information more readily than the conventional case and hence gives an improvement in axial resolution. A quantitative assessment of the resolution of the confocal and conventional transmission images was obtained using Fourier analysis. By taking the Fourier transform of esch 2D image slice in a through-focus series, we were able to compare the degree of high spatial frequency information retained in the confocal versus the conventional case. Figures 2(c-d) and 3(c-d) are 2D Fourier transforms (FFs) which represern the spatial frequency content of each image. The center of each VF corresponds to the lowest spatial frequency with progressively higher spatial frequencies radiating outward vertically and horizontally. The lateral asymmetiy of the FFs is due to the optical characteristics of DIC which produces a shading gradient ("ts relief") effect in one of the transverse directions. SPIE Vol Three-Dimensional Microscopy (1994) / 51

5 Fig. 2. Focussed transmission DIC images of metaphase orchid root tip chromosomes acquired simultaneously (a) with a 10 micron detector pinhole (confocal) and (b) with no detector pinhole (conventional microscope). The Fourier transform for each image is shown in (c) and (d) respectively. The arrows indicate a feature which has gone out ot focus in the pinhole case, butis still visible in the no pinhole case. 52 ISPIE Vol Three-Dimensional Microscopy (1994)

6 Fig. 3. Images of the same chromosomes as Fig. 2, but &a plane of focus 1.4 microns below Fig. 2. (a) detector with pinhole and (b) detector with no pinhole. The respective 2D Fourier transforms are shown in (c) and (d). The pinhole case (a) shows more defocus of features than the conventional (no pinhole) case which demonstrates the ability of the confocal detector to reject out of focus information more readily. This is also clearly visible as a reduction in the percentage of high spatial frequencies in the confocal Fourier transform (c). SPIE Vol Three-Dimensional Microscopy (1994)153

7 4. CONCLUSION Use of confocal scanning laser microscopy (CSLM) in transmission presents new problems in optical alignment. In reflection or fluorescence CSLM the beam deflection introduced by refractive index boundaries within the specimen is cancelled out on the rewrn path, so that the focussed spot falls consistently upon the fixed pinhole detector. This does not occur in transmission, causing the spot to "wander" across the pinhole as it passes through inhomogenous media in the specimen. In a recent study of thin specimens (plant chromosomes) we have acquired high resolution confocal images in both transmission brightfield and transmission Nomarski differential interference contrast (DIC) using a fixed pinhole. In the process, methods for maintaining optical alignment approriate to confocal transmission have been found using qualitative assessment of the scanned image and of the Airy disc at the plane of the pinhole. Compaiisons have been made between the confocal (detector with pinhole) and conventional (detector without pinhole) microscope configurations by recording two images simultaneously. The confocal case showed much improved contrast, aixi reliminary results indicate improved axial resolution for the confocal case. Analysis of the Fourier transforms of the data also allowed a quantitative assessment of transmission imaging properties. Transmission CSLM also presents new problems for the 3-D visualization of confocal data. As predicted by theory, confocal transmission brightfield images acquired on our microscope do not exhibit total removal of out of focus regions from the image, in marked contrast to the reflection brightfield and fltxwescere modes of confocal microscopy. In rder to render such data in 3-D, image processing must be performed to remove the out of focus regions of the image. In addition, transmission Nomarski DIC produces 2-D sections with a has reliefappearance, which also requires image processing in order to present this information in 3-D. In both cases, the image processing techniques need to take account of the optical characteristics of the respective imaging mo&s of the confocal microscope. Using the multiple optical mode confocal system, other optical defects such as defocus and spherical aberration can also be readily monitored and we are furthering our investigations into methods for interactively detecting and correcting these types of aberrations. The ultimate goal of the project will be to produce a user friendly confocal microscope which is capable of 3-D imaging in all of the optical modes commonly found in conventional microscopes, and will include the capacity to produce simultaneous images which are a combination of reflection and transmission information. 5. ACKNOWLEDGMENTS The Physical Optics Department is supported by funds from the University of Sydney, the Science Foundation for Physics within the University of Sydney, and the Australian Research Council. One of us (CJC) is also pleased to acknowledge the support of her position as the School of Physics S1at Research Fellow by the Nell and Hermon Slade Trust and the G. Hermon Slade Foundation. 6. REFERENCES 1. Ci. Cogswell and CJ.R. Sheppard, "Confocal brightfiek! imaging techniques using an on-axis scanning optical microscope," in Confocal Microscopy, T. Wilson, ed., Academic Press, pp , CJ. Cogswell and J.W. OByrne, "A high resolution confocal transmission microscope: I. System design," Proceedings SPIE, 1660, , CJ. Cogswell, D.K. Hamilton and CJ.R. Sheppard, "Colour confocal reflection microscopy using red, green and blue lasers," J. Microsc. 165, , J.W. OByme and CJ. Cogswell, "A high resolution confocal transmission microscope: II. Determining image position and correcting aberrations," Proceedings SPIE, 1660, , CJ. Cogswell and C.J.R. Sheppard, "Confocal differential interference contrast (DIC) microscopy: including a theoretical analysis of conventional and confocal DIC imaging," I. Microsc. 165, , ISPIE Vol Three-Dimensional Microscopy (1994)

Reflecting optical system to increase signal intensity. in confocal microscopy

Reflecting optical system to increase signal intensity. in confocal microscopy Reflecting optical system to increase signal intensity in confocal microscopy DongKyun Kang *, JungWoo Seo, DaeGab Gweon Nano Opto Mechatronics Laboratory, Dept. of Mechanical Engineering, Korea Advanced

More information

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy Bi177 Lecture 5 Adding the Third Dimension Wide-field Imaging Point Spread Function Deconvolution Confocal Laser Scanning Microscopy Confocal Aperture Optical aberrations Alternative Scanning Microscopy

More information

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G APPLICATION NOTE M01 attocfm I for Surface Quality Inspection Confocal microscopes work by scanning a tiny light spot on a sample and by measuring the scattered light in the illuminated volume. First,

More information

Basics of confocal imaging (part I)

Basics of confocal imaging (part I) Basics of confocal imaging (part I) Swiss Institute of Technology (EPFL) Faculty of Life Sciences Head of BIOIMAGING AND OPTICS BIOP arne.seitz@epfl.ch Lateral resolution BioImaging &Optics Platform Light

More information

Light Microscopy. Upon completion of this lecture, the student should be able to:

Light Microscopy. Upon completion of this lecture, the student should be able to: Light Light microscopy is based on the interaction of light and tissue components and can be used to study tissue features. Upon completion of this lecture, the student should be able to: 1- Explain the

More information

Very short introduction to light microscopy and digital imaging

Very short introduction to light microscopy and digital imaging Very short introduction to light microscopy and digital imaging Hernan G. Garcia August 1, 2005 1 Light Microscopy Basics In this section we will briefly describe the basic principles of operation and

More information

Microscopy: Fundamental Principles and Practical Approaches

Microscopy: Fundamental Principles and Practical Approaches Microscopy: Fundamental Principles and Practical Approaches Simon Atkinson Online Resource: http://micro.magnet.fsu.edu/primer/index.html Book: Murphy, D.B. Fundamentals of Light Microscopy and Electronic

More information

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI)

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI) Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI) Liang-Chia Chen 1#, Chao-Nan Chen 1 and Yi-Wei Chang 1 1. Institute of Automation Technology,

More information

Confocal Imaging Through Scattering Media with a Volume Holographic Filter

Confocal Imaging Through Scattering Media with a Volume Holographic Filter Confocal Imaging Through Scattering Media with a Volume Holographic Filter Michal Balberg +, George Barbastathis*, Sergio Fantini % and David J. Brady University of Illinois at Urbana-Champaign, Urbana,

More information

Observing Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens. Compound Light Micros

Observing Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens. Compound Light Micros PHARMACEUTICAL MICROBIOLOGY JIGAR SHAH INSTITUTE OF PHARMACY NIRMA UNIVERSITY Observing Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens.

More information

Resolution. Diffraction from apertures limits resolution. Rayleigh criterion θ Rayleigh = 1.22 λ/d 1 peak at 2 nd minimum. θ f D

Resolution. Diffraction from apertures limits resolution. Rayleigh criterion θ Rayleigh = 1.22 λ/d 1 peak at 2 nd minimum. θ f D Microscopy Outline 1. Resolution and Simple Optical Microscope 2. Contrast enhancement: Dark field, Fluorescence (Chelsea & Peter), Phase Contrast, DIC 3. Newer Methods: Scanning Tunneling microscopy (STM),

More information

Imaging Introduction. September 24, 2010

Imaging Introduction. September 24, 2010 Imaging Introduction September 24, 2010 What is a microscope? Merriam-Webster: an optical instrument consisting of a lens or combination of lenses for making enlarged images of minute objects; especially:

More information

BASICS OF CONFOCAL IMAGING (PART I)

BASICS OF CONFOCAL IMAGING (PART I) BASICS OF CONFOCAL IMAGING (PART I) INTERNAL COURSE 2012 LIGHT MICROSCOPY Lateral resolution Transmission Fluorescence d min 1.22 NA obj NA cond 0 0 rairy 0.61 NAobj Ernst Abbe Lord Rayleigh Depth of field

More information

CCAM Microscope Objectives

CCAM Microscope Objectives CCAM Microscope Objectives Things to consider when selecting an objective Magnification Numerical Aperture (NA) resolving power and light intensity of the objective Working Distance distance between the

More information

CCAM s Selection of. Zeiss Microscope Objectives

CCAM s Selection of. Zeiss Microscope Objectives CCAM s Selection of Zeiss Microscope Objectives 1. Magnification Image scale 2. Resolution The minimum separation distance between two points that are clearly resolved. The resolution of an objective is

More information

ANSWER KEY Lab 2 (IGB): Bright Field and Fluorescence Optical Microscopy and Sectioning

ANSWER KEY Lab 2 (IGB): Bright Field and Fluorescence Optical Microscopy and Sectioning Phys598BP Spring 2016 University of Illinois at Urbana-Champaign ANSWER KEY Lab 2 (IGB): Bright Field and Fluorescence Optical Microscopy and Sectioning Location: IGB Core Microscopy Facility Microscope:

More information

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES Shortly after the experimental confirmation of the wave properties of the electron, it was suggested that the electron could be used to examine objects

More information

OPTICAL PRINCIPLES OF MICROSCOPY. Interuniversity Course 28 December 2003 Aryeh M. Weiss Bar Ilan University

OPTICAL PRINCIPLES OF MICROSCOPY. Interuniversity Course 28 December 2003 Aryeh M. Weiss Bar Ilan University OPTICAL PRINCIPLES OF MICROSCOPY Interuniversity Course 28 December 2003 Aryeh M. Weiss Bar Ilan University FOREWORD This slide set was originally presented at the ISM Workshop on Theoretical and Experimental

More information

NIH Public Access Author Manuscript Opt Lett. Author manuscript; available in PMC 2010 August 9.

NIH Public Access Author Manuscript Opt Lett. Author manuscript; available in PMC 2010 August 9. NIH Public Access Author Manuscript Published in final edited form as: Opt Lett. 2010 January 1; 35(1): 67 69. Autoconfocal transmission microscopy based on two-photon induced photocurrent of Si photodiodes

More information

Optical Microscopy and Imaging ( Part 2 )

Optical Microscopy and Imaging ( Part 2 ) 1 Optical Microscopy and Imaging ( Part 2 ) Chapter 7.1 : Semiconductor Science by Tudor E. Jenkins Saroj Kumar Patra, Department of Electronics and Telecommunication, Norwegian University of Science and

More information

Microscopy Techniques that make it easy to see things this small.

Microscopy Techniques that make it easy to see things this small. Microscopy Techniques that make it easy to see things this small. What is a Microscope? An instrument for viewing objects that are too small to be seen easily by the naked eye. Dutch spectacle-makers Hans

More information

INTRODUCTION THIN LENSES. Introduction. given by the paraxial refraction equation derived last lecture: Thin lenses (19.1) = 1. Double-lens systems

INTRODUCTION THIN LENSES. Introduction. given by the paraxial refraction equation derived last lecture: Thin lenses (19.1) = 1. Double-lens systems Chapter 9 OPTICAL INSTRUMENTS Introduction Thin lenses Double-lens systems Aberrations Camera Human eye Compound microscope Summary INTRODUCTION Knowledge of geometrical optics, diffraction and interference,

More information

Microscopy Training & Overview

Microscopy Training & Overview Microscopy Training & Overview Product Marketing October 2011 Stephan Briggs - PLE OVERVIEW AND PRESENTATION FLOW Glossary and Important Terms Introduction Timeline Innovation and Advancement Primary Components

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and

More information

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging Bi/BE 227 Winter 2016 Assignment #3 Adding the third dimension: 3D Confocal Imaging Schedule: Jan 20: Assignment Jan 20-Feb 8: Work on assignment Feb 10: Student PowerPoint presentations. Goals for this

More information

OCT Spectrometer Design Understanding roll-off to achieve the clearest images

OCT Spectrometer Design Understanding roll-off to achieve the clearest images OCT Spectrometer Design Understanding roll-off to achieve the clearest images Building a high-performance spectrometer for OCT imaging requires a deep understanding of the finer points of both OCT theory

More information

Microscopy. Matti Hotokka Department of Physical Chemistry Åbo Akademi University

Microscopy. Matti Hotokka Department of Physical Chemistry Åbo Akademi University Microscopy Matti Hotokka Department of Physical Chemistry Åbo Akademi University What s coming Anatomy of a microscope Modes of illumination Practicalities Special applications Basic microscope Ocular

More information

Aberrations and adaptive optics for biomedical microscopes

Aberrations and adaptive optics for biomedical microscopes Aberrations and adaptive optics for biomedical microscopes Martin Booth Department of Engineering Science And Centre for Neural Circuits and Behaviour University of Oxford Outline Rays, wave fronts and

More information

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each lens with a 1 Airy unit pinhole setting. Many water lenses

More information

2/4/15. Brightfield Microscopy! It s all about Magnification..! or is it?!

2/4/15. Brightfield Microscopy! It s all about Magnification..! or is it?! Brightfield Microscopy It s all about Magnification.. or is it? 1 What actually does go into chosing a microscope Choice depends on what you need the microscope to do. Do you want to magnify stained specimens?

More information

Practical work no. 3: Confocal Live Cell Microscopy

Practical work no. 3: Confocal Live Cell Microscopy Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside

More information

Heisenberg) relation applied to space and transverse wavevector

Heisenberg) relation applied to space and transverse wavevector 2. Optical Microscopy 2.1 Principles A microscope is in principle nothing else than a simple lens system for magnifying small objects. The first lens, called the objective, has a short focal length (a

More information

Basics of Light Microscopy and Metallography

Basics of Light Microscopy and Metallography ENGR45: Introduction to Materials Spring 2012 Laboratory 8 Basics of Light Microscopy and Metallography In this exercise you will: gain familiarity with the proper use of a research-grade light microscope

More information

1 Co Localization and Working flow with the lsm700

1 Co Localization and Working flow with the lsm700 1 Co Localization and Working flow with the lsm700 Samples -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ BrDU with alexa 488. -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ no BrDU (but with

More information

Microscopic Structures

Microscopic Structures Microscopic Structures Image Analysis Metal, 3D Image (Red-Green) The microscopic methods range from dark field / bright field microscopy through polarisation- and inverse microscopy to techniques like

More information

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005 Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev Why use confocal microscopy? Principles of the laser scanning confocal microscope. Image resolution. Manipulating the

More information

Application Note. The New 2D Superresolution Mode for ZEISS Airyscan 120 nm Lateral Resolution without Acquiring a Z-stack

Application Note. The New 2D Superresolution Mode for ZEISS Airyscan 120 nm Lateral Resolution without Acquiring a Z-stack The New 2D Superresolution Mode for ZEISS Airyscan 120 nm Lateral Resolution without Acquiring a Z-stack The New 2D Superresolution Mode for ZEISS Airyscan 120 nm Lateral Resolution without Acquiring a

More information

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton

More information

FLUORESCENCE MICROSCOPY. Matyas Molnar and Dirk Pacholsky

FLUORESCENCE MICROSCOPY. Matyas Molnar and Dirk Pacholsky FLUORESCENCE MICROSCOPY Matyas Molnar and Dirk Pacholsky 1 The human eye perceives app. 400-700 nm; best at around 500 nm (green) Has a general resolution down to150-300 μm (human hair: 40-250 μm) We need

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front

More information

Exp No.(8) Fourier optics Optical filtering

Exp No.(8) Fourier optics Optical filtering Exp No.(8) Fourier optics Optical filtering Fig. 1a: Experimental set-up for Fourier optics (4f set-up). Related topics: Fourier transforms, lenses, Fraunhofer diffraction, index of refraction, Huygens

More information

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report Introduction and Background Two-photon microscopy is a type of fluorescence microscopy using two-photon excitation. It

More information

Confocal Microscopy. Kristin Jensen

Confocal Microscopy. Kristin Jensen Confocal Microscopy Kristin Jensen 17.11.05 References Cell Biological Applications of Confocal Microscopy, Brian Matsumoto, chapter 1 Studying protein dynamics in living cells,, Jennifer Lippincott-Schwartz

More information

ADVANCED OPTICS LAB -ECEN Basic Skills Lab

ADVANCED OPTICS LAB -ECEN Basic Skills Lab ADVANCED OPTICS LAB -ECEN 5606 Basic Skills Lab Dr. Steve Cundiff and Edward McKenna, 1/15/04 Revised KW 1/15/06, 1/8/10 Revised CC and RZ 01/17/14 The goal of this lab is to provide you with practice

More information

Zeiss 880 Training Notes Zen 2.3

Zeiss 880 Training Notes Zen 2.3 Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start

More information

Confocal Laser Scanning Microscopy

Confocal Laser Scanning Microscopy Name of the Core Facility: Confocal Laser Scanning Microscopy CORE Forschungszentrum Immunologie Mainz Welcome to the CSLM Core Facility: The CLSM Core Facility enables working groups to incorporate high

More information

Dynamic Phase-Shifting Microscopy Tracks Living Cells

Dynamic Phase-Shifting Microscopy Tracks Living Cells from photonics.com: 04/01/2012 http://www.photonics.com/article.aspx?aid=50654 Dynamic Phase-Shifting Microscopy Tracks Living Cells Dr. Katherine Creath, Goldie Goldstein and Mike Zecchino, 4D Technology

More information

Lecture 23 MNS 102: Techniques for Materials and Nano Sciences

Lecture 23 MNS 102: Techniques for Materials and Nano Sciences Lecture 23 MNS 102: Techniques for Materials and Nano Sciences Reference: #1 C. R. Brundle, C. A. Evans, S. Wilson, "Encyclopedia of Materials Characterization", Butterworth-Heinemann, Toronto (1992),

More information

PROCEEDINGS OF SPIE. Three-dimensional transfer functions

PROCEEDINGS OF SPIE. Three-dimensional transfer functions PROCEEDINGS OF SPIE SPIEDigitalLibrary.org/conference-proceedings-of-spie Three-dimensional transfer functions Colin J. R. Sheppard Colin J. R. Sheppard, "Three-dimensional transfer functions," Proc. SPIE

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power

More information

Chapter 2 The Study of Microbial Structure: Microscopy and Specimen Preparation

Chapter 2 The Study of Microbial Structure: Microscopy and Specimen Preparation Chapter 2 The Study of Microbial Structure: Microscopy and Specimen Preparation 1 Lenses and the Bending of Light light is refracted (bent) when passing from one medium to another refractive index a measure

More information

Microscope anatomy, image formation and resolution

Microscope anatomy, image formation and resolution Microscope anatomy, image formation and resolution Ian Dobbie Buy this book for your lab: D.B. Murphy, "Fundamentals of light microscopy and electronic imaging", ISBN 0-471-25391-X Visit these websites:

More information

Modulation Transfer Function

Modulation Transfer Function Modulation Transfer Function The resolution and performance of an optical microscope can be characterized by a quantity known as the modulation transfer function (MTF), which is a measurement of the microscope's

More information

Transmission Electron Microscopy 9. The Instrument. Outline

Transmission Electron Microscopy 9. The Instrument. Outline Transmission Electron Microscopy 9. The Instrument EMA 6518 Spring 2009 02/25/09 Outline The Illumination System The Objective Lens and Stage Forming Diffraction Patterns and Images Alignment and Stigmation

More information

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy Qiyuan Song (M2) and Aoi Nakamura (B4) Abstracts: We theoretically and experimentally

More information

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal Yashvinder Sabharwal, 1 James Joubert 2 and Deepak Sharma 2 1. Solexis Advisors LLC, Austin, TX, USA 2. Photometrics

More information

Optical design of a high resolution vision lens

Optical design of a high resolution vision lens Optical design of a high resolution vision lens Paul Claassen, optical designer, paul.claassen@sioux.eu Marnix Tas, optical specialist, marnix.tas@sioux.eu Prof L.Beckmann, l.beckmann@hccnet.nl Summary:

More information

Instructions for the Experiment

Instructions for the Experiment Instructions for the Experiment Excitonic States in Atomically Thin Semiconductors 1. Introduction Alongside with electrical measurements, optical measurements are an indispensable tool for the study of

More information

Nontranslational three-dimensional profilometry by chromatic confocal microscopy with dynamically configurable micromirror scanning

Nontranslational three-dimensional profilometry by chromatic confocal microscopy with dynamically configurable micromirror scanning Nontranslational three-dimensional profilometry by chromatic confocal microscopy with dynamically configurable micromirror scanning Sungdo Cha, Paul C. Lin, Lijun Zhu, Pang-Chen Sun, and Yeshaiahu Fainman

More information

Parallel Mode Confocal System for Wafer Bump Inspection

Parallel Mode Confocal System for Wafer Bump Inspection Parallel Mode Confocal System for Wafer Bump Inspection ECEN5616 Class Project 1 Gao Wenliang wen-liang_gao@agilent.com 1. Introduction In this paper, A parallel-mode High-speed Line-scanning confocal

More information

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE Prior to lab: 1) Read these instructions (p 1-6) 2) Go through the online tutorial, the microscopy pre-lab

More information

Microscopy from Carl Zeiss

Microscopy from Carl Zeiss Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path

More information

Light Microscopy for Biomedical Research

Light Microscopy for Biomedical Research Light Microscopy for Biomedical Research Tuesday 4:30 PM Quantification & Digital Images Michael Hooker Microscopy Facility Michael Chua microscopy@unc.edu 843-3268 6007 Thurston Bowles http://microscopy.unc.edu/lmbr

More information

Compare and Contrast. Contrast Methods in Industrial Inspection Microscopy. Application Note. We explain how to

Compare and Contrast. Contrast Methods in Industrial Inspection Microscopy. Application Note. We explain how to Application Note Compare and Contrast Contrast Methods in Industrial Inspection Microscopy We explain how to E nhance materials inspection microscopy workflows Reveal surface and sub-surface imperfections

More information

ADAPTIVE CORRECTION FOR ACOUSTIC IMAGING IN DIFFICULT MATERIALS

ADAPTIVE CORRECTION FOR ACOUSTIC IMAGING IN DIFFICULT MATERIALS ADAPTIVE CORRECTION FOR ACOUSTIC IMAGING IN DIFFICULT MATERIALS I. J. Collison, S. D. Sharples, M. Clark and M. G. Somekh Applied Optics, Electrical and Electronic Engineering, University of Nottingham,

More information

Some of the important topics needed to be addressed in a successful lens design project (R.R. Shannon: The Art and Science of Optical Design)

Some of the important topics needed to be addressed in a successful lens design project (R.R. Shannon: The Art and Science of Optical Design) Lens design Some of the important topics needed to be addressed in a successful lens design project (R.R. Shannon: The Art and Science of Optical Design) Focal length (f) Field angle or field size F/number

More information

Technology Note ZEISS LSM 880 with Airyscan

Technology Note ZEISS LSM 880 with Airyscan Technology Note ZEISS LSM 880 with Airyscan Introducing the Fast Acquisition Mode ZEISS LSM 880 with Airyscan Introducing the Fast Acquisition Mode Author: Dr. Annette Bergter Carl Zeiss Microscopy GmbH,

More information

Improving the Collection Efficiency of Raman Scattering

Improving the Collection Efficiency of Raman Scattering PERFORMANCE Unparalleled signal-to-noise ratio with diffraction-limited spectral and imaging resolution Deep-cooled CCD with excelon sensor technology Aberration-free optical design for uniform high resolution

More information

2013 LMIC Imaging Workshop. Sidney L. Shaw Technical Director. - Light and the Image - Detectors - Signal and Noise

2013 LMIC Imaging Workshop. Sidney L. Shaw Technical Director. - Light and the Image - Detectors - Signal and Noise 2013 LMIC Imaging Workshop Sidney L. Shaw Technical Director - Light and the Image - Detectors - Signal and Noise The Anatomy of a Digital Image Representative Intensities Specimen: (molecular distribution)

More information

Sensitive measurement of partial coherence using a pinhole array

Sensitive measurement of partial coherence using a pinhole array 1.3 Sensitive measurement of partial coherence using a pinhole array Paul Petruck 1, Rainer Riesenberg 1, Richard Kowarschik 2 1 Institute of Photonic Technology, Albert-Einstein-Strasse 9, 07747 Jena,

More information

Opterra II Multipoint Scanning Confocal Microscope. Innovation with Integrity

Opterra II Multipoint Scanning Confocal Microscope. Innovation with Integrity Opterra II Multipoint Scanning Confocal Microscope Enabling 4D Live-Cell Fluorescence Imaging through Speed, Sensitivity, Viability and Simplicity Innovation with Integrity Fluorescence Microscopy The

More information

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,

More information

Chapter 18 Optical Elements

Chapter 18 Optical Elements Chapter 18 Optical Elements GOALS When you have mastered the content of this chapter, you will be able to achieve the following goals: Definitions Define each of the following terms and use it in an operational

More information

Studying of Reflected Light Optical Laser Microscope Images Using Image Processing Algorithm

Studying of Reflected Light Optical Laser Microscope Images Using Image Processing Algorithm IRAQI JOURNAL OF APPLIED PHYSICS Fatema H. Rajab Al-Nahrain University, College of Engineering, Department of Laser and Optoelectronic Engineering Studying of Reflected Light Optical Laser Microscope Images

More information

3.0 Alignment Equipment and Diagnostic Tools:

3.0 Alignment Equipment and Diagnostic Tools: 3.0 Alignment Equipment and Diagnostic Tools: Alignment equipment The alignment telescope and its use The laser autostigmatic cube (LACI) interferometer A pin -- and how to find the center of curvature

More information

Basics of INTERFEROMETRY

Basics of INTERFEROMETRY Basics of INTERFEROMETRY P Hariharan CSIRO Division of Applied Sydney, Australia Physics ACADEMIC PRESS, INC. Harcourt Brace Jovanovich, Publishers Boston San Diego New York London Sydney Tokyo Toronto

More information

Practical Flatness Tech Note

Practical Flatness Tech Note Practical Flatness Tech Note Understanding Laser Dichroic Performance BrightLine laser dichroic beamsplitters set a new standard for super-resolution microscopy with λ/10 flatness per inch, P-V. We ll

More information

Multifluorescence The Crosstalk Problem and Its Solution

Multifluorescence The Crosstalk Problem and Its Solution Multifluorescence The Crosstalk Problem and Its Solution If a specimen is labeled with more than one fluorochrome, each image channel should only show the emission signal of one of them. If, in a specimen

More information

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010 Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION Optically reconfigurable metasurfaces and photonic devices based on phase change materials S1: Schematic diagram of the experimental setup. A Ti-Sapphire femtosecond laser (Coherent Chameleon Vision S)

More information

PROCEEDINGS OF SPIE. Measurement of low-order aberrations with an autostigmatic microscope

PROCEEDINGS OF SPIE. Measurement of low-order aberrations with an autostigmatic microscope PROCEEDINGS OF SPIE SPIEDigitalLibrary.org/conference-proceedings-of-spie Measurement of low-order aberrations with an autostigmatic microscope William P. Kuhn Measurement of low-order aberrations with

More information

Ron Liu OPTI521-Introductory Optomechanical Engineering December 7, 2009

Ron Liu OPTI521-Introductory Optomechanical Engineering December 7, 2009 Synopsis of METHOD AND APPARATUS FOR IMPROVING VISION AND THE RESOLUTION OF RETINAL IMAGES by David R. Williams and Junzhong Liang from the US Patent Number: 5,777,719 issued in July 7, 1998 Ron Liu OPTI521-Introductory

More information

Bringing Answers to the Surface

Bringing Answers to the Surface 3D Bringing Answers to the Surface 1 Expanding the Boundaries of Laser Microscopy Measurements and images you can count on. Every time. LEXT OLS4100 Widely used in quality control, research, and development

More information

Quick Start Guide. Leica SP5 X

Quick Start Guide. Leica SP5 X Quick Start Guide Leica SP5 X Please note: Some of the information in this guide was taken from Leica Microsystems Leica TCS SP5 LAS AF Guide for New Users. This work is licensed under the Creative Commons

More information

STEM Spectrum Imaging Tutorial

STEM Spectrum Imaging Tutorial STEM Spectrum Imaging Tutorial Gatan, Inc. 5933 Coronado Lane, Pleasanton, CA 94588 Tel: (925) 463-0200 Fax: (925) 463-0204 April 2001 Contents 1 Introduction 1.1 What is Spectrum Imaging? 2 Hardware 3

More information

Invitation for a walk through microscopy. Sebastian Schuchmann Jörg Rösner

Invitation for a walk through microscopy. Sebastian Schuchmann Jörg Rösner Invitation for a walk through microscopy Sebastian Schuchmann Jörg Rösner joerg.roesner@charite.de Techniques in microscopy Conventional (light) microscopy bright & dark field, phase & interference contrast

More information

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION Revised November 15, 2017 INTRODUCTION The simplest and most commonly described examples of diffraction and interference from two-dimensional apertures

More information

J. C. Wyant Fall, 2012 Optics Optical Testing and Testing Instrumentation

J. C. Wyant Fall, 2012 Optics Optical Testing and Testing Instrumentation J. C. Wyant Fall, 2012 Optics 513 - Optical Testing and Testing Instrumentation Introduction 1. Measurement of Paraxial Properties of Optical Systems 1.1 Thin Lenses 1.1.1 Measurements Based on Image Equation

More information

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement CONFOCAL MICROSCOPY BioVis Uppsala, 2017 Jeremy Adler Matyas Molnar Dirk Pacholsky Widefield & Confocal Microscopy

More information

Confocal Microscopy and Related Techniques

Confocal Microscopy and Related Techniques Confocal Microscopy and Related Techniques Chau-Hwang Lee Associate Research Fellow Research Center for Applied Sciences, Academia Sinica 128 Sec. 2, Academia Rd., Nankang, Taipei 11529, Taiwan E-mail:

More information

Bio 407. Applied microscopy. Introduction into light microscopy. José María Mateos. Center for Microscopy and Image Analysis

Bio 407. Applied microscopy. Introduction into light microscopy. José María Mateos. Center for Microscopy and Image Analysis Center for Microscopy and Image Analysis Bio 407 Applied Introduction into light José María Mateos Fundamentals of light Compound microscope Microscope composed of an objective and an additional lens (eyepiece,

More information

Shreyash Tandon M.S. III Year

Shreyash Tandon M.S. III Year Shreyash Tandon M.S. III Year 20091015 Confocal microscopy is a powerful tool for generating high-resolution images and 3-D reconstructions of a specimen by using point illumination and a spatial pinhole

More information

Bias errors in PIV: the pixel locking effect revisited.

Bias errors in PIV: the pixel locking effect revisited. Bias errors in PIV: the pixel locking effect revisited. E.F.J. Overmars 1, N.G.W. Warncke, C. Poelma and J. Westerweel 1: Laboratory for Aero & Hydrodynamics, University of Technology, Delft, The Netherlands,

More information

Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA

Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA Abstract: Speckle interferometry (SI) has become a complete technique over the past couple of years and is widely used in many branches of

More information

High resolution extended depth of field microscopy using wavefront coding

High resolution extended depth of field microscopy using wavefront coding High resolution extended depth of field microscopy using wavefront coding Matthew R. Arnison *, Peter Török #, Colin J. R. Sheppard *, W. T. Cathey +, Edward R. Dowski, Jr. +, Carol J. Cogswell *+ * Physical

More information

No part of this material may be reproduced without explicit written permission.

No part of this material may be reproduced without explicit written permission. This material is provided for educational use only. The information in these slides including all data, images and related materials are the property of : Robert M. Glaeser Department of Molecular & Cell

More information

Single-shot depth-section imaging through chromatic slit-scan confocal microscopy

Single-shot depth-section imaging through chromatic slit-scan confocal microscopy Single-shot depth-section imaging through chromatic slit-scan confocal microscopy Paul C. Lin, Pang-Chen Sun, Lijun Zhu, and Yeshaiahu Fainman A chromatic confocal microscope constructed with a white-light

More information

Surface Finish Measurement Methods and Instrumentation

Surface Finish Measurement Methods and Instrumentation 125 years of innovation Surface Finish Measurement Methods and Instrumentation Contents Visual Inspection Surface Finish Comparison Plates Contact Gauges Inductive / Variable Reluctance (INTRA) Piezo Electric

More information

Test procedures Page: 1 of 5

Test procedures Page: 1 of 5 Test procedures Page: 1 of 5 1 Scope This part of document establishes uniform requirements for measuring the numerical aperture of optical fibre, thereby assisting in the inspection of fibres and cables

More information

Flatness of Dichroic Beamsplitters Affects Focus and Image Quality

Flatness of Dichroic Beamsplitters Affects Focus and Image Quality Flatness of Dichroic Beamsplitters Affects Focus and Image Quality Flatness of Dichroic Beamsplitters Affects Focus and Image Quality 1. Introduction Even though fluorescence microscopy has become a routine

More information