Scanned light sheet microscopy with confocal slit detection

Size: px
Start display at page:

Download "Scanned light sheet microscopy with confocal slit detection"

Transcription

1 Scanned light sheet microscopy with confocal slit detection Eugen Baumgart * and Ulrich Kubitscheck Institute of Physical and Theoretical Chemistry, Rheinische Friedrich-Wilhelms-University Bonn, Wegelerstraße 12, D Bonn, Germany * baumgart@pc.uni-bonn.de Abstract: In light sheet fluorescence microscopy optical sectioning is achieved by illuminating the sample orthogonally to the detection pathway with a thin, focused sheet of light. However, light scattering within the sample often deteriorates the optical sectioning effect. Here, we demonstrate that contrast and degree of confocality can greatly be increased by combining scanned light sheet fluorescence excitation and confocal slit detection. A high frame rate was achieved by using the rolling shutter of a scientific CMOS camera as a slit detector. Synchronizing the rolling shutter with the scanned illumination beam results in confocal line detection. Acquiring image data with selective plane illumination minimizes photo-damage while simultaneously enhancing contrast, optical sectioning and signal-to-noise ratio. Thus the imaging principle presented here merges the benefits of scanned light sheet microscopy and linescanning confocal imaging Optical Society of America OCIS codes: ( ) Confocal microscopy; ( ) Fluorescence microscopy; ( ) Scanning microscopy; ( ) Three-dimensional microscopy; ( ) Medical and biological imaging. References and links 1. A. H. Voie, D. H. Burns, and F. A. Spelman, Orthogonal-plane fluorescence optical sectioning: threedimensional imaging of macroscopic biological specimens, J. Microsc. 170(3), (1993). 2. J. Huisken, J. Swoger, F. Del Bene, J. Wittbrodt, and E. H. K. Stelzer, Optical sectioning deep inside live embryos by selective plane illumination microscopy, Science 305(5686), (2004). 3. J. Huisken and D. Y. R. Stainier, Even fluorescence excitation by multidirectional selective plane illumination microscopy (mspim), Opt. Lett. 32(17), (2007). 4. H.-U. Dodt, U. Leischner, A. Schierloh, N. Jährling, C. P. Mauch, K. Deininger, J. M. Deussing, M. Eder, W. Zieglgänsberger, and K. Becker, Ultramicroscopy: three-dimensional visualization of neuronal networks in the whole mouse brain, Nat. Methods 4(4), (2007). 5. P. J. Keller and E. H. K. Stelzer, Quantitative in vivo imaging of entire embryos with digital scanned laser light sheet fluorescence microscopy, Curr. Opin. Neurobiol. 18(6), (2008). 6. P. J. Keller, A. D. Schmidt, J. Wittbrodt, and E. H. K. Stelzer, Reconstruction of zebrafish early embryonic development by scanned light sheet microscopy, Science 322(5904), (2008). 7. J. Mertz and J. Kim, Scanning light-sheet microscopy in the whole mouse brain with HiLo background rejection, J. Biomed. Opt. 15(1), (2010). 8. F. O. Fahrbach and A. Rohrbach, Propagation stability of self-reconstructing Bessel beams enables contrastenhanced imaging in thick media, Nat Commun 3(632), 632 (2012). 9. H. K. A. Spiecker, Method and arrangement for microscopy, PCT Patent 2011/ (2011). 10. J. G. Ritter, R. Veith, A. Veenendaal, J. P. Siebrasse, and U. Kubitscheck, Light sheet microscopy for single molecule tracking in living tissue, PLoS ONE 5(7), e11639 (2010). 11. T. Wilson and C. Sheppard, Theory and Practice of Scanning Optical Microscopy (Academic Press, London, 1984), Chap J. T. Bushberg, J. A. Seibert, E. M. Leidholdt, Jr., and J. M. Boone, The Essential Physics of Medical Imaging (Lippincott Williams & Wilkins, Philadelphia, 2002). 13. A. A. Michelson, Studies in Optics (University of Chicago Press 1927). 14. P. J. Keller, A. D. Schmidt, A. Santella, K. Khairy, Z. Bao, J. Wittbrodt, and E. H. Stelzer, Fast, high-contrast imaging of animal development with scanned light sheet-based structured-illumination microscopy, Nat. Methods 7(8), (2010). (C) 2012 OSA 10 September 2012 / Vol. 20, No. 19/ OPTICS EXPRESS 21805

2 1. Introduction Conventional epi-illumination fluorescence microscopes illuminate the focal plane of the detection objective as well as a large fraction of the sample outside the focal plane. This leads to background intensity due to out-of-focus fluorescence and reduces contrast and resolution. In confocal laser scanning microscopy a pinhole in a conjugate plane can be used to reject out-of-focus fluorescence since the sample is scanned point-wise but at the cost of low image acquisition rates. This type of sectioned imaging is advantageous for three-dimensional (3D) data acquisition but leads to high light exposure of the entire sample. Light Sheet Fluorescence Microscopy (LSFM) circumvents this disadvantage by illuminating the sample perpendicular to the detection direction. Typically, a laser beam with an elliptical cross section is formed by cylindrical lenses and focused into the sample by an illumination objective. Thus only fluorophores in a thin sheet near the focal plane are excited [1 4]. The result is an intrinsic optical sectioning effect, low photobleaching and reduced photo-toxicity. LSFM permits studies of bulk samples like whole embryonic mouse brains or the real time investigation of embryonic cell development in vivo, while keeping the total light exposure of the sample at a minimum. An alternative method to generate a light sheet is realized in digitally scanned light sheet microscopy [5,6]. Here, a circular Gaussian beam is scanned across the focal plane by a mirror, which is positioned in the back focal plane of the illumination objective. The beam is focused into the sample with its main axis perpendicular to the focal plane of the illumination objective. When the beam is swept across the focal plane at least once during an image acquisition cycle a planar illumination equivalent to a sheet but with improved illumination efficiency and a uniform intensity distribution is achieved. The fluorescence detection in LSFM does not differ from conventional wide-field microscopy. The image sensor detects ballistic photons as well as scattered photons which have lost their image information. In addition, excitation light is scattered in the sample and excites off-focal plane fluorophores. Especially in thick samples these effects can be noticed as image blur and reduction of contrast. Previously published methods to reduce the amount of scattered light either rely on structured illumination patterns [7] or the application of confocal slit masks to images taken sequentially with line illumination [8]. Though image contrast is enhanced these methods suffer from long data acquisition times since the final contrast-enhanced image is numerically reconstructed from multiple raw images. An elegant solution to speed up the data acquisition has recently been proposed [9]. The scanning of the illumination beam can be synchronized to the rolling shutter of a scientific CMOS (scmos) camera. An scmos is an active-pixel sensor with separate photodetector and amplification units for each pixel. In contrast to CCDs the pixels can be addressed and read out independently from each other. Currently available scmos can be operated in two data acquisition modes: global shutter and rolling shutter. When working with global shutter, all image pixels are simultaneously active, which is similar to image acquisition by CCD cameras. In rolling shutter mode, however, the active detection region is comprised of only a few adjacent pixel lines and thus equals a band of simultaneously exposed rows that moves across the image sensor. Choosing a sufficiently small exposure time results in a line-shaped detection region that represents a moving confocal slit detector, if synchronized to the illumination beam. In this manner fast line-scanning confocal light detection with a minimum number of mechanical parts can be realized. We developed a microscope that uses a circular Gaussian beam scanned synchronously to the rolling shutter detection. It offers true confocal imaging at high frame rates. We showed that scattered fluorescence is rejected during the acquisition of an image. Further, we characterized the improvements in contrast, signal-to-noise ratio (SNR) and optical (C) 2012 OSA 10 September 2012 / Vol. 20, No. 19/ OPTICS EXPRESS 21806

3 sectioning. The instrument presented here will be especially beneficial for 3D imaging of large specimen. 2. Methods 2.1 Illumination beam path Our scanned light sheet microscope setup was designed to augment an inverted microscope. Custom built sample chambers with a coverslip bottom (0.17 mm) allowed us to use conventional objective lenses for observation. The sample was held by a motorized three axis translation stage [10]. Fig. 1. Illumination beam path. (A) Top view. The excitation beam exits a single mode fiber and is guided by mirrors onto a scanning mirror. The scanning position is imaged by a relay objective into the back focal plane of the illumination objective lens, which is located inside the objective housing. The relay lens assembly is composed of two lens systems with overlapping focal points. It also serves to expand the illumination beam. (B) Side view of the illumination optics. A laser beam with a wavelength of 633 nm was guided through a single mode optical fiber onto a two-axis scanning mirror, Fig. 1. The scanning mirror was placed in the front focal plane of a relay lens system that imaged it into the back focal plane of a 10X achromat (Plan Apo Infinity-Corrected Long Working Distance, Mitutoyo Corp., Kawasaki, Japan) located inside the objective housing. The illuminated aperture radius in the back focal plane was 1.6 mm, which resulted in an effective illumination NA of The relay system comprised a lens doublet and a single lens in a 4f configuration. It was designed to compensate spherical aberration that originates from the specimen chamber glass walls and the sample medium by 2 ray-tracing using Zemax (Radiant Zemax, LLC, Redmond, WA, USA). The1/e beam waist radius w and the Rayleigh length z 0 r were w = 4.5 ±0.2 μ m and z 0 r =116 ±5 μ m for 633 nm excitation wavelength as determined by imaging the beam in a fluorescent dye solution. A Nikon CFI Plan Fluor 10X objective lens with a NA of 0.3 was used for fluorescence detection. Its axial FWHM was estimated to be 11.6 µm for a wavelength of 650 nm [11]. 2.2 Data acquisition We used a scmos camera (1920 x 1440 pixels, pixel size 3.63 µm, Orca Flash 2.8, Hamamatsu Photonics K.K., Hamamatsu City, Japan) for data acquisition. It was operated either in a global or a rolling shutter mode. In global shutter mode all pixels simultaneously detect photons for a predefined integration time after reset and begin of exposure. In rolling shutter mode the detector is activated, and the exposure starts sequentially with a time delay of 15 μs between neighboring rows beginning with the topmost pixel row. The activation process continues across the complete chip until the last row. The readout starts in a similar (C) 2012 OSA 10 September 2012 / Vol. 20, No. 19/ OPTICS EXPRESS 21807

4 manner at the first pixel row after a predefined integration time. It is continued row by row again with a delay of 15 μs between subsequent rows. In this manner the time required to activate all 1440 rows of the sensor is 22 ms. We defined the integration time per line to be below this value. Therefore, the readout of the topmost rows began before the activation of the last row. In this manner a rolling shutter was realized by effectively scanning an active pixel region across the scmos chip. The principle of rolling shutter data acquisition is illustrated in Fig. 2. Fig. 2. Rolling shutter mode. An activated row was marked in red, and a read-out row was marked in yellow. The band of pixel rows in between is exposed to light simultaneously. This band moves from the top to the bottom of the sensor. Its width is defined by the single row exposure time, and can be adjusted from a minimum of one line up to the whole chip. 2.3 Synchronization of illumination and detection The scanning mirror in the illumination beam path was driven by a saw-tooth voltage signal generated by a NI USB-6221 data acquisition and generation device (DAQ) from National Instruments (National Instruments Corp., Austin, TX, USA). It also generated step function trigger signals with amplitude of 5V to control the data acquisition of the camera and the acousto-optical tunable filter which defined the laser intensity. It was crucial to synchronize the output of the analog scanning curve and the two further trigger signals with utmost precision. This was achieved by using internal signal routes of the DAQ. In short, the procedure was as follows. At first, the saw-tooth and the trigger waveforms were loaded into the buffer of the device, and the channels were initialized. This was done via the USB interface and was not time-critical. Then, the device generated the time-dependent analog output voltage driving the scanning mirror. Simultaneously the two digital signals for activating data acquisition and laser illumination were issued. Pixel-line reset and readout were controlled by the internal camera clock. The data acquisition of the camera started on the rising edge of the camera trigger signal and the laser was switched on while the laser trigger output was at 5V. The scanning mirror voltage curve and the two trigger signals shared the same sample clock of 65,376 Hz. For optimal imaging performance the illumination beam position in the focal plane had to match the conjugate rolling shutter position on the scmos exactly during the entire image acquisition process. To achieve this we imaged the beam position in a homogeneous fluorescent solution for a set of different mirror angles. The position of the beam main axis was determined by column-wise fitting of a Gaussian function to the image data. A linear fit to this data supplied the required amplitude, offset and step size for the voltage curve driving the mirrors. When the voltage curve was # $15.00 USD (C) 2012 OSA Received 20 Jul 2012; revised 13 Aug 2012; accepted 5 Sep 2012; published 7 Sep September 2012 / Vol. 20, No. 19/ OPTICS EXPRESS 21808

5 subsequently applied to the scanning mirror, the maximum beam deflection corresponded to the first and last pixel rows, and the intermediate beam positions increased linearly with scanning mirror angle. 3. Results 3.1 Contrast improvement in scattering media A focused Gaussian beam provides illumination in scanned LSFM. This pattern could directly be imaged when it illuminated a weakly scattering homogeneously fluorescent solution of Alexa Fluor 647 (Life Technologies Corp., Carlsbad, CA, USA) in phosphate buffered saline (PBS), Fig. 3(A). However, when the same beam profile illuminated a strongly scattering sample such as fluorescent polystyrene beads (diameter 200 nm, Molecular Probes, Eugene, OR, USA) embedded in an agarose gel containing additional fluorescent molecules, an immense deterioration of the profile became visible, Fig. 3(B). Fig. 3. Image of the illumination beam entering from the left into a (A) homogeneously fluorescent solution of Alexa Fluor 647 in PBS buffer and (B) sample of fluorescent polysterene beads embedded in agarose containing Alexa Fluor 647. A significant amount of light was scattered leading to fluorescence excitation outside the directly illuminated probe region. Imaging conditions and image contrast settings were kept identical. Scale bar 100 μm. A large amount of light was scattered into the sample along the Gaussian illumination beam path. The scattered light excited in-focus and out-of-focus fluorescence at large distances off the illumination beam. This greatly reduced the image quality, when images were acquired in a global shutter mode during beam scanning. Fluorescence photons that were scattered on their way through the sample to the detection device added to a uniform background. These effects were shown in the image in Fig. 4(A). This situation could be greatly improved, when the image was acquired in a rolling shutter mode during beam scanning as shown in Figs. 4(B) and 4(C), with rolling shutter widths of 128 px. and 32 px., respectively. Figure 4(D) shows the same data taken with a rolling shutter size of 2 px. Here the rejection of light leads to a reduced signal to noise ratio. (C) 2012 OSA 10 September 2012 / Vol. 20, No. 19/ OPTICS EXPRESS 21809

6 Fig. 4. Images of beads in an agarose gel demonstrating the contrast difference between global shutter and rolling shutter image acquisition. (A) Image taken in global shutter mode, (B) in rolling shutter mode with a slit width of 128 rows, (C) in rolling shutter mode with a slit width of 32 rows and (D) in rolling shutter mode with a slit width of 2 rows. The image acquisition time was 22 ms in each image. Scale bar, 50 μm. The qualitative difference was most striking when performing 3D reconstructions. Again we used the sample with fluorescent beads, and imaged a 50 μm thick volume in steps of 1 μm. Figure 5 shows a comparison between 3D reconstructions of the sample imaged in global shutter and in rolling shutter mode, respectively. The global shutter image in Fig. 5(A) shows a strong background, and substantial post-processing would be required to remove it. In the confocal image in Fig. 5(B) the background is almost absent. The confocal data shown Fig. 5 was taken with a rolling shutter width of 2 pixels (8% of the width of the illumination beam) corresponding to an effective exposure time of 30 µs. A large amount of ballistic photons was excluded from detection, which reduced the overall intensity. In global shutter mode the effective exposure time approximated the time the beam needed to sweep over a certain point in the sample. This was approximately 375 μs, according to a line activation time of approximately 15 μs and a beam diameter equal to 25 pixels. Additionally, in-focus emitters were not only excited by the scanned Gaussian beam but also by light scattered from the excitation beam at other positions in the sample. Thus scattered light also caused illumination of fluorophores, in addition to the Gaussian beam profile. To achieve a comparable SNR in both data sets, and to emphasize the attainable sectioning improvement, the confocal data was averaged 40-fold. # $15.00 USD (C) 2012 OSA Received 20 Jul 2012; revised 13 Aug 2012; accepted 5 Sep 2012; published 7 Sep September 2012 / Vol. 20, No. 19/ OPTICS EXPRESS 21810

7 Fig. 5. 3D reconstruction of a 120x120x50 μm volume of beads fixed in agarose. The sample volume was imaged in (A) global shutter mode and (B) with a rolling shutter of two lines. The reduction of background in the reconstructed rolling shutter data is evident. The rolling shutter data were averaged 40x to yield an SNR comparable to the global shutter images. 3 The strongly scattering microbead-sample was used to characterize the SNR and contrast for global shutter and rolling shutter detection, respectively. We systematically varied the width of the rolling shutter from 2 to 256 rows and compared the imaging result with the global shutter mode. For the acquisition of an image the beam was only swept once across the sample resulting in an acquisition time of 22 ms for all data points. Fig. 6. SNR and contrast. (A) SNR for different rolling shutter sizes in units of the illumination beam diameter, and for global shutter. (B) Contrast as a function of rolling shutter size, and for global shutter. The rolling shutter sizes corresponding to each point are 2, 4, 8, 16, 32, 64, 128 and 256 pixels, starting from the left. The SNR was measured by fitting two-dimensional Gaussian functions to determine the signal amplitude IS of single fluorescent beads. Regions with no signal were used for determination of the background standard deviation σ bg. The SNR was then calculated as SNR= IS /σ bg [12]. Image contrast C was measured by determining the highest and lowest intensity values in each row (I,I ) and applying the Michelson formula C=(I - I )/(I + I ) [13]. Finally we determined the mean of all rows. The results are given in Fig. 6. The width of the rolling shutter was expressed in units normalized to the illumination beam diameter. Unity max # $15.00 USD (C) 2012 OSA min max min max min Received 20 Jul 2012; revised 13 Aug 2012; accepted 5 Sep 2012; published 7 Sep September 2012 / Vol. 20, No. 19/ OPTICS EXPRESS 21811

8 corresponded to the 1/e -diameter of the illumination beam on the chip (25 rows). As expected, the detection slit width determined the SNR, which showed a distinct maximum around a slit size of unity. The image contrast C for rolling shutter was about twice as large as that for global shutter detection Optical sectioning We measured the microscope point spread function (PSF) by fitting a 3D Gaussian function to images of single diffraction limited particles of the data shown in Fig. 5 and averaged the results of 100 particles. The full width at half maximum (FWHM) of the PSF in axial direction was w 0 = 5.7±1.0 μm and w 0 = 4.5±1.0 μm for global shutter and rolling shutter (2 pixels) detection, respectively. The latter is comparable to the FWHM of the illumination beam of 4.4±0.2 μm. 3.3 Confocal imaging of biological samples In order to demonstrate the superior imaging performance of the line scanning light sheet microscope with rolling shutter detection we imaged a 3D extended, scattering biological sample. To this end we used salivary gland cells of Chironomus tentans that were fixed in 4% formaldehyde solution. The transcription sites along the polytene chromosomes were fluorescently labeled using monoclonal mouse antibodies ARNA-3 CBL221 (Millipore, Billerica, MA, USA) against RNA polymerase II as primary, and Alexa Fluor 647 labeled goat anti-mouse IgG1 (Life Technologies Corp., Carlsbad, CA, USA) as secondary antibody. Fig. 7. Images of polytene chromosomes in C. tentans salivary gland cell nucleus. (A) Image taken in global shutter mode. The beam was swept once across the sample. (B) Nucleus imaged with a rolling shutter width corresponding to the illumination beam diameter. The bands are crisper and the background in dark areas of the sample is lower as compared to (A). Scale bar, 30 μm. (C) Normalized intensity along the blue and red line in (A) and (B), respectively. Figure 7 shows the transcription sites inside the nucleus of a gland cell. The sample was imaged in global shutter mode, Fig. 7(A), and rolling shutter mode, Fig. 7(B), with a width corresponding to the illumination beam width of approximately 9 µm ( 1/e -diameter). Hence, with an effective pixel size of μm, the illumination beam diameter corresponded to about 25 rows. The acquisition time for each frame was 22 ms for global and rolling shutter mode. In both images the prominent Balbiani ring 2 transcription site and further transcription bands were visible. In comparison to the global shutter image the rolling shutter image was sharper and had a higher contrast. Thus, it was easier to distinguish transcription sites from the background. Furthermore, the total background signal in the dark regions of the image was lower in the confocal mode as compared to the global shutter mode. Figure 7(C) shows an intensity plot normalized according to I = I/(I + I ) along a vertical line in the images as 2 norm # $15.00 USD (C) 2012 OSA max min Received 20 Jul 2012; revised 13 Aug 2012; accepted 5 Sep 2012; published 7 Sep September 2012 / Vol. 20, No. 19/ OPTICS EXPRESS 21812

9 marked in Figs. 7(A) and 7(B). The plot illustrates that the chromosome bands showed a better contrast when imaged in rolling shutter detection. 4. Discussion We developed a confocal line scanning light microscope combining the benefits of selective plane illumination and confocal line detection. A digitally scanned Gaussian laser beam, which was focused into the sample perpendicular to the detection axis, was used for fluorescence excitation. Rolling shutter data acquisition using a scientific CMOS camera provided synchronized confocal line detection. Similar results can be achieved by taking multiple images that were illuminated row by row, cutting out the illuminated rows and stitching them back together [8]. Obviously, this is a time and computationally intensive process. Alternatively, structured illumination created with a scanned laser beam can be used to enhance image contrast. Here, the illumination beam is switched on and off repeatedly during scanning to produce a sinusoidal illumination pattern [14]. The sample has to be illuminated three times with the pattern phase shifted by 60 degrees between each acquisition. The final contrast-enhanced image is then calculated from the structured illumination data. A related technique uses a uniformly illuminated image and a structured illuminated image to numerically remove contrast-degrading background [7]. However, contrast-enhancing techniques based on illuminating the sample with a structured pattern ultimately depend on the contrast between bright and dark parts of the pattern and thus may suffer from low SNR or high background signals. Compared to these existing approaches to confocal or contrast improved light sheet microscopy, our approach has several advantages. Only a single, intrinsically confocal image has to be acquired. This reduces the risk of artifacts due to sample drift between subsequent image acquisitions. No post-processing or numerical background elimination from multiple images is required. Notably, no special optical and mechanical elements for de-scanning like in conventional confocal microscopes are required, since the rolling shutter represents a slit detector that moves synchronously to the illumination beam in a conjugate image plane. This simplifies the setup, improves mechanical stability and most importantly improves fluorescence intensity. However it should be mentioned that the Hamamatsu Orca Flash 2.8 is an early prototype of a scientific CMOS camera where the rolling shutter moves continuously from the top to the bottom of the image sensor. Its successor, the Hamamatsu Orca Flash 4.0 as well as current models from other manufacturers, have a considerably larger image sensor. The trade-off is that the chip consists of two parts, where each half has its own rolling shutter. These are moving synchronously outwards from the center. Thus, in order to be able to take advantage of the whole image sensor, the sample would have to be illuminated from both sides with counter-directed moving beams, which might lead to artifacts at their overlap. Of course, the most elegant solution would be to utilize a camera with rolling shutters moving in the same direction and two parallel beams. Furthermore, the time needed for the rolling shutter to run from the top to the bottom of the chip is set to 22 ms by the camera electronics. In order to achieve a rolling shutter width of one line, the exposure time had to be set to 22 ms divided by the total number of pixel rows, which is approximately 15 μs and equals the line cycle time. Thus the only way to adjust the rolling shutter size is by setting the exposure time to the desired width multiplied by 15 μs. A straight forward solution would be to make the cycle time between activation of rows adjustable by the user. Despite the fact that optical confocality is enhanced only perpendicular to the direction of rolling shutter movement, we could significantly improve the image quality. We measured an increase in optical sectioning capability for small rolling shutter sizes and could demonstrate the improvement in contrast and SNR without the need of repetitive data acquisition like in structured illumination techniques. (C) 2012 OSA 10 September 2012 / Vol. 20, No. 19/ OPTICS EXPRESS 21813

10 Acknowledgments Helpful discussions with Jan-Hendrik Spille are gratefully acknowledged. Tim P. Kaminski kindly provided the fluorescently labeled C. tentans salivary gland cell samples. We thank Hamamatsu Photonics Deutschland GmbH for generously making the Orca Flash 2.8 available. (C) 2012 OSA 10 September 2012 / Vol. 20, No. 19/ OPTICS EXPRESS 21814

Supplementary Information. Stochastic Optical Reconstruction Microscopy Imaging of Microtubule Arrays in Intact Arabidopsis thaliana Seedling Roots

Supplementary Information. Stochastic Optical Reconstruction Microscopy Imaging of Microtubule Arrays in Intact Arabidopsis thaliana Seedling Roots Supplementary Information Stochastic Optical Reconstruction Microscopy Imaging of Microtubule Arrays in Intact Arabidopsis thaliana Seedling Roots Bin Dong 1,, Xiaochen Yang 2,, Shaobin Zhu 1, Diane C.

More information

Adaptive optimisation of illumination beam profiles in fluorescence microscopy

Adaptive optimisation of illumination beam profiles in fluorescence microscopy Adaptive optimisation of illumination beam profiles in fluorescence microscopy T. J. Mitchell a, C. D. Saunter a, W. O Nions a, J. M. Girkin a, G. D. Love a a Centre for Advanced nstrumentation & Biophysical

More information

Confocal Imaging Through Scattering Media with a Volume Holographic Filter

Confocal Imaging Through Scattering Media with a Volume Holographic Filter Confocal Imaging Through Scattering Media with a Volume Holographic Filter Michal Balberg +, George Barbastathis*, Sergio Fantini % and David J. Brady University of Illinois at Urbana-Champaign, Urbana,

More information

Multicolor 4D Fluorescence Microscopy using Ultrathin Bessel Light sheets

Multicolor 4D Fluorescence Microscopy using Ultrathin Bessel Light sheets SUPPLEMENTARY MATERIAL Multicolor 4D Fluorescence Microscopy using Ultrathin Bessel Light sheets Teng Zhao, Sze Cheung Lau, Ying Wang, Yumian Su, Hao Wang, Aifang Cheng, Karl Herrup, Nancy Y. Ip, Shengwang

More information

Nature Methods: doi: /nmeth Supplementary Figure 1. Schematic of 2P-ISIM AO optical setup.

Nature Methods: doi: /nmeth Supplementary Figure 1. Schematic of 2P-ISIM AO optical setup. Supplementary Figure 1 Schematic of 2P-ISIM AO optical setup. Excitation from a femtosecond laser is passed through intensity control and shuttering optics (1/2 λ wave plate, polarizing beam splitting

More information

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy Bi177 Lecture 5 Adding the Third Dimension Wide-field Imaging Point Spread Function Deconvolution Confocal Laser Scanning Microscopy Confocal Aperture Optical aberrations Alternative Scanning Microscopy

More information

3D light microscopy techniques

3D light microscopy techniques 3D light microscopy techniques The image of a point is a 3D feature In-focus image Out-of-focus image The image of a point is not a point Point Spread Function (PSF) 1D imaging 1 1 2! NA = 0.5! NA 2D imaging

More information

Fast, high-contrast imaging of animal development with scanned light sheet based structured-illumination microscopy

Fast, high-contrast imaging of animal development with scanned light sheet based structured-illumination microscopy nature methods Fast, high-contrast imaging of animal development with scanned light sheet based structured-illumination microscopy Philipp J Keller, Annette D Schmidt, Anthony Santella, Khaled Khairy,

More information

Nikon Instruments Europe

Nikon Instruments Europe Nikon Instruments Europe Recommendations for N-SIM sample preparation and image reconstruction Dear customer, We hope you find the following guidelines useful in order to get the best performance out of

More information

Nature Neuroscience: doi: /nn Supplementary Figure 1. Optimized Bessel foci for in vivo volume imaging.

Nature Neuroscience: doi: /nn Supplementary Figure 1. Optimized Bessel foci for in vivo volume imaging. Supplementary Figure 1 Optimized Bessel foci for in vivo volume imaging. (a) Images taken by scanning Bessel foci of various NAs, lateral and axial FWHMs: (Left panels) in vivo volume images of YFP + neurites

More information

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report Introduction and Background Two-photon microscopy is a type of fluorescence microscopy using two-photon excitation. It

More information

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy, KTH Applied Physics Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy, 2009-06-05, 8-13, FB51 Allowed aids: Compendium Imaging Physics (handed out) Compendium Light Microscopy

More information

3D light microscopy techniques

3D light microscopy techniques 3D light microscopy techniques The image of a point is a 3D feature In-focus image Out-of-focus image The image of a point is not a point Point Spread Function (PSF) 1D imaging 2D imaging 3D imaging Resolution

More information

Shreyash Tandon M.S. III Year

Shreyash Tandon M.S. III Year Shreyash Tandon M.S. III Year 20091015 Confocal microscopy is a powerful tool for generating high-resolution images and 3-D reconstructions of a specimen by using point illumination and a spatial pinhole

More information

Optical Sectioning Deep Inside Live Embryos by Selective Plane Illumination Microscopy

Optical Sectioning Deep Inside Live Embryos by Selective Plane Illumination Microscopy Supporting Online Material Optical Sectioning Deep Inside Live Embryos by Selective Plane Illumination Microscopy 1 Material and methods 1.1 Setup Jan Huisken, Jim Swoger, Filippo Del Bene, Joachim Wittbrodt,

More information

Multi-channel imaging cytometry with a single detector

Multi-channel imaging cytometry with a single detector Multi-channel imaging cytometry with a single detector Sarah Locknar 1, John Barton 1, Mark Entwistle 2, Gary Carver 1 and Robert Johnson 1 1 Omega Optical, Brattleboro, VT 05301 2 Philadelphia Lightwave,

More information

Applications of Optics

Applications of Optics Nicholas J. Giordano www.cengage.com/physics/giordano Chapter 26 Applications of Optics Marilyn Akins, PhD Broome Community College Applications of Optics Many devices are based on the principles of optics

More information

STORM/ PALM ANSWER KEY

STORM/ PALM ANSWER KEY STORM/ PALM ANSWER KEY Phys598BP Spring 2016 University of Illinois at Urbana-Champaign Questions for Lab Report 1. How do you define a resolution in STORM imaging? If you are given a STORM setup, how

More information

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G APPLICATION NOTE M01 attocfm I for Surface Quality Inspection Confocal microscopes work by scanning a tiny light spot on a sample and by measuring the scattered light in the illuminated volume. First,

More information

Maria Smedh, Centre for Cellular Imaging. Maria Smedh, Centre for Cellular Imaging

Maria Smedh, Centre for Cellular Imaging. Maria Smedh, Centre for Cellular Imaging Nonlinear microscopy I: Two-photon fluorescence microscopy Multiphoton Microscopy What is multiphoton imaging? Applications Different imaging modes Advantages/disadvantages Scattering of light in thick

More information

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name:

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name: EE119 Introduction to Optical Engineering Spring 2003 Final Exam Name: SID: CLOSED BOOK. THREE 8 1/2 X 11 SHEETS OF NOTES, AND SCIENTIFIC POCKET CALCULATOR PERMITTED. TIME ALLOTTED: 180 MINUTES Fundamental

More information

Light Microscopy. Upon completion of this lecture, the student should be able to:

Light Microscopy. Upon completion of this lecture, the student should be able to: Light Light microscopy is based on the interaction of light and tissue components and can be used to study tissue features. Upon completion of this lecture, the student should be able to: 1- Explain the

More information

Reflecting optical system to increase signal intensity. in confocal microscopy

Reflecting optical system to increase signal intensity. in confocal microscopy Reflecting optical system to increase signal intensity in confocal microscopy DongKyun Kang *, JungWoo Seo, DaeGab Gweon Nano Opto Mechatronics Laboratory, Dept. of Mechanical Engineering, Korea Advanced

More information

Bias errors in PIV: the pixel locking effect revisited.

Bias errors in PIV: the pixel locking effect revisited. Bias errors in PIV: the pixel locking effect revisited. E.F.J. Overmars 1, N.G.W. Warncke, C. Poelma and J. Westerweel 1: Laboratory for Aero & Hydrodynamics, University of Technology, Delft, The Netherlands,

More information

III III 0 IIOI DID IIO 1101 I II 0II II 100 III IID II DI II

III III 0 IIOI DID IIO 1101 I II 0II II 100 III IID II DI II (19) United States III III 0 IIOI DID IIO 1101 I0 1101 0II 0II II 100 III IID II DI II US 200902 19549A1 (12) Patent Application Publication (10) Pub. No.: US 2009/0219549 Al Nishizaka et al. (43) Pub.

More information

Shaping light in microscopy:

Shaping light in microscopy: Shaping light in microscopy: Adaptive optical methods and nonconventional beam shapes for enhanced imaging Martí Duocastella planet detector detector sample sample Aberrated wavefront Beamsplitter Adaptive

More information

Aberrations and adaptive optics for biomedical microscopes

Aberrations and adaptive optics for biomedical microscopes Aberrations and adaptive optics for biomedical microscopes Martin Booth Department of Engineering Science And Centre for Neural Circuits and Behaviour University of Oxford Outline Rays, wave fronts and

More information

Nature Methods: doi: /nmeth Supplementary Figure 1

Nature Methods: doi: /nmeth Supplementary Figure 1 . Supplementary Figure 1 Schematics and characterization of our AO two-photon fluorescence microscope. (a) Essential components of our AO two-photon fluorescence microscope: Ti:Sapphire laser; optional

More information

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal Yashvinder Sabharwal, 1 James Joubert 2 and Deepak Sharma 2 1. Solexis Advisors LLC, Austin, TX, USA 2. Photometrics

More information

Physics 431 Final Exam Examples (3:00-5:00 pm 12/16/2009) TIME ALLOTTED: 120 MINUTES Name: Signature:

Physics 431 Final Exam Examples (3:00-5:00 pm 12/16/2009) TIME ALLOTTED: 120 MINUTES Name: Signature: Physics 431 Final Exam Examples (3:00-5:00 pm 12/16/2009) TIME ALLOTTED: 120 MINUTES Name: PID: Signature: CLOSED BOOK. TWO 8 1/2 X 11 SHEET OF NOTES (double sided is allowed), AND SCIENTIFIC POCKET CALCULATOR

More information

Be aware that there is no universal notation for the various quantities.

Be aware that there is no universal notation for the various quantities. Fourier Optics v2.4 Ray tracing is limited in its ability to describe optics because it ignores the wave properties of light. Diffraction is needed to explain image spatial resolution and contrast and

More information

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON N-SIM guide NIKON IMAGING CENTRE @ KING S COLLEGE LONDON Starting-up / Shut-down The NSIM hardware is calibrated after system warm-up occurs. It is recommended that you turn-on the system for at least

More information

Practical Flatness Tech Note

Practical Flatness Tech Note Practical Flatness Tech Note Understanding Laser Dichroic Performance BrightLine laser dichroic beamsplitters set a new standard for super-resolution microscopy with λ/10 flatness per inch, P-V. We ll

More information

Very short introduction to light microscopy and digital imaging

Very short introduction to light microscopy and digital imaging Very short introduction to light microscopy and digital imaging Hernan G. Garcia August 1, 2005 1 Light Microscopy Basics In this section we will briefly describe the basic principles of operation and

More information

High resolution extended depth of field microscopy using wavefront coding

High resolution extended depth of field microscopy using wavefront coding High resolution extended depth of field microscopy using wavefront coding Matthew R. Arnison *, Peter Török #, Colin J. R. Sheppard *, W. T. Cathey +, Edward R. Dowski, Jr. +, Carol J. Cogswell *+ * Physical

More information

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005 Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev Why use confocal microscopy? Principles of the laser scanning confocal microscope. Image resolution. Manipulating the

More information

Zero Focal Shift in High Numerical Aperture Focusing of a Gaussian Laser Beam through Multiple Dielectric Interfaces. Ali Mahmoudi

Zero Focal Shift in High Numerical Aperture Focusing of a Gaussian Laser Beam through Multiple Dielectric Interfaces. Ali Mahmoudi 1 Zero Focal Shift in High Numerical Aperture Focusing of a Gaussian Laser Beam through Multiple Dielectric Interfaces Ali Mahmoudi a.mahmoudi@qom.ac.ir & amahmodi@yahoo.com Laboratory of Optical Microscopy,

More information

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION Revised November 15, 2017 INTRODUCTION The simplest and most commonly described examples of diffraction and interference from two-dimensional apertures

More information

On spatial resolution

On spatial resolution On spatial resolution Introduction How is spatial resolution defined? There are two main approaches in defining local spatial resolution. One method follows distinction criteria of pointlike objects (i.e.

More information

Εισαγωγική στην Οπτική Απεικόνιση

Εισαγωγική στην Οπτική Απεικόνιση Εισαγωγική στην Οπτική Απεικόνιση Δημήτριος Τζεράνης, Ph.D. Εμβιομηχανική και Βιοϊατρική Τεχνολογία Τμήμα Μηχανολόγων Μηχανικών Ε.Μ.Π. Χειμερινό Εξάμηνο 2015 Light: A type of EM Radiation EM radiation:

More information

Flatness of Dichroic Beamsplitters Affects Focus and Image Quality

Flatness of Dichroic Beamsplitters Affects Focus and Image Quality Flatness of Dichroic Beamsplitters Affects Focus and Image Quality Flatness of Dichroic Beamsplitters Affects Focus and Image Quality 1. Introduction Even though fluorescence microscopy has become a routine

More information

Introduction to Light Microscopy. (Image: T. Wittman, Scripps)

Introduction to Light Microscopy. (Image: T. Wittman, Scripps) Introduction to Light Microscopy (Image: T. Wittman, Scripps) The Light Microscope Four centuries of history Vibrant current development One of the most widely used research tools A. Khodjakov et al. Major

More information

Confocal Microscopy. Kristin Jensen

Confocal Microscopy. Kristin Jensen Confocal Microscopy Kristin Jensen 17.11.05 References Cell Biological Applications of Confocal Microscopy, Brian Matsumoto, chapter 1 Studying protein dynamics in living cells,, Jennifer Lippincott-Schwartz

More information

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

3. are adherent cells (ie. cells in suspension are too far away from the coverslip) Before you begin, make sure your sample... 1. is seeded on #1.5 coverglass (thickness = 0.17) 2. is an aqueous solution (ie. fixed samples mounted on a slide will not work - not enough difference in refractive

More information

Optically sectioned imaging by oblique plane microscopy

Optically sectioned imaging by oblique plane microscopy Optically sectioned imaging by oblique plane microscopy C. Dunsby Photonics Group, Department of Physics, Imperial College London, South Kensington Campus, London, SW7 2AZ, UK Corresponding author: christopher.dunsby@imperial.ac.uk

More information

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010 Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped

More information

INTRODUCTION THIN LENSES. Introduction. given by the paraxial refraction equation derived last lecture: Thin lenses (19.1) = 1. Double-lens systems

INTRODUCTION THIN LENSES. Introduction. given by the paraxial refraction equation derived last lecture: Thin lenses (19.1) = 1. Double-lens systems Chapter 9 OPTICAL INSTRUMENTS Introduction Thin lenses Double-lens systems Aberrations Camera Human eye Compound microscope Summary INTRODUCTION Knowledge of geometrical optics, diffraction and interference,

More information

product overview pco.edge family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology

product overview pco.edge family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology product overview family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology scmos knowledge base scmos General Information PCO scmos cameras are a breakthrough

More information

Nontranslational three-dimensional profilometry by chromatic confocal microscopy with dynamically configurable micromirror scanning

Nontranslational three-dimensional profilometry by chromatic confocal microscopy with dynamically configurable micromirror scanning Nontranslational three-dimensional profilometry by chromatic confocal microscopy with dynamically configurable micromirror scanning Sungdo Cha, Paul C. Lin, Lijun Zhu, Pang-Chen Sun, and Yeshaiahu Fainman

More information

More fancy SPIM, Even fancier SPIM

More fancy SPIM, Even fancier SPIM More fancy SPIM, Even fancier SPIM Last class Light sheet microscopy Fancy SPIM (ispim, dspim, etc ) This class Multi camera SPIM SIM SPIM Bessels d x,y = λ em 2 NA d z = 2 NA λ ex + n(1 cosθ λ em 1 IsoView

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

Multifluorescence The Crosstalk Problem and Its Solution

Multifluorescence The Crosstalk Problem and Its Solution Multifluorescence The Crosstalk Problem and Its Solution If a specimen is labeled with more than one fluorochrome, each image channel should only show the emission signal of one of them. If, in a specimen

More information

OCT Spectrometer Design Understanding roll-off to achieve the clearest images

OCT Spectrometer Design Understanding roll-off to achieve the clearest images OCT Spectrometer Design Understanding roll-off to achieve the clearest images Building a high-performance spectrometer for OCT imaging requires a deep understanding of the finer points of both OCT theory

More information

Opterra II Multipoint Scanning Confocal Microscope. Innovation with Integrity

Opterra II Multipoint Scanning Confocal Microscope. Innovation with Integrity Opterra II Multipoint Scanning Confocal Microscope Enabling 4D Live-Cell Fluorescence Imaging through Speed, Sensitivity, Viability and Simplicity Innovation with Integrity Fluorescence Microscopy The

More information

Digital confocal microscope

Digital confocal microscope Digital confocal microscope Alexandre S. Goy * and Demetri Psaltis Optics Laboratory, École Polytechnique Fédérale de Lausanne, Station 17, Lausanne, 1015, Switzerland * alexandre.goy@epfl.ch Abstract:

More information

Supplementary information, Figure S1A-S1H The thickness and the uniformity of the light sheet at different DOFs. By

Supplementary information, Figure S1A-S1H The thickness and the uniformity of the light sheet at different DOFs. By Supplementary information, Figure S1A-S1H The thickness and the uniformity of the light sheet at different DOFs. By imaging FITC-containing solution, the thickness of the light sheet generated by the P3A-DSLM

More information

a) How big will that physical image of the cells be your camera sensor?

a) How big will that physical image of the cells be your camera sensor? 1. Consider a regular wide-field microscope set up with a 60x, NA = 1.4 objective and a monochromatic digital camera with 8 um pixels, properly positioned in the primary image plane. This microscope is

More information

Heisenberg) relation applied to space and transverse wavevector

Heisenberg) relation applied to space and transverse wavevector 2. Optical Microscopy 2.1 Principles A microscope is in principle nothing else than a simple lens system for magnifying small objects. The first lens, called the objective, has a short focal length (a

More information

INTRODUCTION TO MICROSCOPY. Urs Ziegler THE PROBLEM

INTRODUCTION TO MICROSCOPY. Urs Ziegler THE PROBLEM INTRODUCTION TO MICROSCOPY Urs Ziegler ziegler@zmb.uzh.ch THE PROBLEM 1 ORGANISMS ARE LARGE LIGHT AND ELECTRONS: ELECTROMAGNETIC WAVES v = Wavelength ( ) Speed (v) Frequency ( ) Amplitude (A) Propagation

More information

pco.edge 4.2 LT 0.8 electrons 2048 x 2048 pixel 40 fps up to :1 up to 82 % pco. low noise high resolution high speed high dynamic range

pco.edge 4.2 LT 0.8 electrons 2048 x 2048 pixel 40 fps up to :1 up to 82 % pco. low noise high resolution high speed high dynamic range edge 4.2 LT scientific CMOS camera high resolution 2048 x 2048 pixel low noise 0.8 electrons USB 3.0 small form factor high dynamic range up to 37 500:1 high speed 40 fps high quantum efficiency up to

More information

Better Imaging with a Schmidt-Czerny-Turner Spectrograph

Better Imaging with a Schmidt-Czerny-Turner Spectrograph Better Imaging with a Schmidt-Czerny-Turner Spectrograph Abstract For years, images have been measured using Czerny-Turner (CT) design dispersive spectrographs. Optical aberrations inherent in the CT design

More information

Microscope anatomy, image formation and resolution

Microscope anatomy, image formation and resolution Microscope anatomy, image formation and resolution Ian Dobbie Buy this book for your lab: D.B. Murphy, "Fundamentals of light microscopy and electronic imaging", ISBN 0-471-25391-X Visit these websites:

More information

Nature Methods: doi: /nmeth Supplementary Figure 1. sospim principle and representation of the sospim beam-steering unit.

Nature Methods: doi: /nmeth Supplementary Figure 1. sospim principle and representation of the sospim beam-steering unit. Supplementary Figure 1 sospim principle and representation of the sospim beam-steering unit. Schematic representation of the sospim principle showing a sample holder comprising 45 micromirrored cavities

More information

NIH Public Access Author Manuscript Opt Lett. Author manuscript; available in PMC 2010 August 9.

NIH Public Access Author Manuscript Opt Lett. Author manuscript; available in PMC 2010 August 9. NIH Public Access Author Manuscript Published in final edited form as: Opt Lett. 2010 January 1; 35(1): 67 69. Autoconfocal transmission microscopy based on two-photon induced photocurrent of Si photodiodes

More information

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI)

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI) Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI) Liang-Chia Chen 1#, Chao-Nan Chen 1 and Yi-Wei Chang 1 1. Institute of Automation Technology,

More information

Life Science Instrumentation. New Generation. Light Sheet Fluorescence Microscope. Alph

Life Science Instrumentation. New Generation. Light Sheet Fluorescence Microscope. Alph Life Science Instrumentation Light Sheet Fluorescence Microscope New Generation Alph Modular Light Sheet Microscope Alpha 3 is a new generation of light sheet fluorescence microscope addressing the needs

More information

Nature Structural & Molecular Biology: doi: /nsmb Supplementary Figure 1

Nature Structural & Molecular Biology: doi: /nsmb Supplementary Figure 1 Supplementary Figure 1 Supplemental correlative nanomanipulation-fluorescence traces probing nascent RNA and fluorescent Mfd during TCR initiation. Supplemental correlative nanomanipulation-fluorescence

More information

Instant super-resolution imaging in live cells and embryos via analog image processing

Instant super-resolution imaging in live cells and embryos via analog image processing Nature Methods Instant super-resolution imaging in live cells and embryos via analog image processing Andrew G. York, Panagiotis Chandris, Damian Dalle Nogare, Jeffrey Head, Peter Wawrzusin, Robert S.

More information

Laser Speckle Reducer LSR-3000 Series

Laser Speckle Reducer LSR-3000 Series Datasheet: LSR-3000 Series Update: 06.08.2012 Copyright 2012 Optotune Laser Speckle Reducer LSR-3000 Series Speckle noise from a laser-based system is reduced by dynamically diffusing the laser beam. A

More information

Confocal Microscopy and Related Techniques

Confocal Microscopy and Related Techniques Confocal Microscopy and Related Techniques Chau-Hwang Lee Associate Research Fellow Research Center for Applied Sciences, Academia Sinica 128 Sec. 2, Academia Rd., Nankang, Taipei 11529, Taiwan E-mail:

More information

Experiment 1: Fraunhofer Diffraction of Light by a Single Slit

Experiment 1: Fraunhofer Diffraction of Light by a Single Slit Experiment 1: Fraunhofer Diffraction of Light by a Single Slit Purpose 1. To understand the theory of Fraunhofer diffraction of light at a single slit and at a circular aperture; 2. To learn how to measure

More information

Nature Protocols: doi: /nprot Supplementary Figure 1. Schematic diagram of Kőhler illumination.

Nature Protocols: doi: /nprot Supplementary Figure 1. Schematic diagram of Kőhler illumination. Supplementary Figure 1 Schematic diagram of Kőhler illumination. The green beam path represents the excitation path and the red represents the emission path. Supplementary Figure 2 Microscope base components

More information

1 Co Localization and Working flow with the lsm700

1 Co Localization and Working flow with the lsm700 1 Co Localization and Working flow with the lsm700 Samples -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ BrDU with alexa 488. -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ no BrDU (but with

More information

Microscopy Live Animal Imaging

Microscopy Live Animal Imaging Microscopy Live Animal Imaging A collaborative environment that provides the knowledge, instruments, and expertise needed to visualize life at scales ranging from single molecules to entire animals. Project

More information

Laser Telemetric System (Metrology)

Laser Telemetric System (Metrology) Laser Telemetric System (Metrology) Laser telemetric system is a non-contact gauge that measures with a collimated laser beam (Refer Fig. 10.26). It measure at the rate of 150 scans per second. It basically

More information

Rapid three-dimensional isotropic imaging of living cells using Bessel beam plane illumination

Rapid three-dimensional isotropic imaging of living cells using Bessel beam plane illumination Nature Methods Rapid three-dimensional isotropic imaging of living cells using beam plane illumination Thomas A Planchon, Liang Gao, Daniel E Milkie, Michael W Davidson, James A Galbraith, Catherine G

More information

Microscopic Structures

Microscopic Structures Microscopic Structures Image Analysis Metal, 3D Image (Red-Green) The microscopic methods range from dark field / bright field microscopy through polarisation- and inverse microscopy to techniques like

More information

Development of a High-speed Super-resolution Confocal Scanner

Development of a High-speed Super-resolution Confocal Scanner Development of a High-speed Super-resolution Confocal Scanner Takuya Azuma *1 Takayuki Kei *1 Super-resolution microscopy techniques that overcome the spatial resolution limit of conventional light microscopy

More information

Radial Polarization Converter With LC Driver USER MANUAL

Radial Polarization Converter With LC Driver USER MANUAL ARCoptix Radial Polarization Converter With LC Driver USER MANUAL Arcoptix S.A Ch. Trois-portes 18 2000 Neuchâtel Switzerland Mail: info@arcoptix.com Tel: ++41 32 731 04 66 Principle of the radial polarization

More information

Practical work no. 3: Confocal Live Cell Microscopy

Practical work no. 3: Confocal Live Cell Microscopy Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside

More information

Enhancement of the lateral resolution and the image quality in a line-scanning tomographic optical microscope

Enhancement of the lateral resolution and the image quality in a line-scanning tomographic optical microscope Summary of the PhD thesis Enhancement of the lateral resolution and the image quality in a line-scanning tomographic optical microscope Author: Dudás, László Supervisors: Prof. Dr. Szabó, Gábor and Dr.

More information

Observing Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens. Compound Light Micros

Observing Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens. Compound Light Micros PHARMACEUTICAL MICROBIOLOGY JIGAR SHAH INSTITUTE OF PHARMACY NIRMA UNIVERSITY Observing Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens.

More information

High-resolution, low light-dose lightsheet microscope LATTICE LIGHTSHEET

High-resolution, low light-dose lightsheet microscope LATTICE LIGHTSHEET LATTICE LIGHTSHEET High-resolution, low light-dose lightsheet microscope First developed by Nobel Laureate Dr. Eric Betzig, the 3i Lattice LightSheet microscope is capable of imaging biological systems

More information

Camera Test Protocol. Introduction TABLE OF CONTENTS. Camera Test Protocol Technical Note Technical Note

Camera Test Protocol. Introduction TABLE OF CONTENTS. Camera Test Protocol Technical Note Technical Note Technical Note CMOS, EMCCD AND CCD CAMERAS FOR LIFE SCIENCES Camera Test Protocol Introduction The detector is one of the most important components of any microscope system. Accurate detector readings

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

LENSES. INEL 6088 Computer Vision

LENSES. INEL 6088 Computer Vision LENSES INEL 6088 Computer Vision Digital camera A digital camera replaces film with a sensor array Each cell in the array is a Charge Coupled Device light-sensitive diode that converts photons to electrons

More information

Optical Performance of Nikon F-Mount Lenses. Landon Carter May 11, Measurement and Instrumentation

Optical Performance of Nikon F-Mount Lenses. Landon Carter May 11, Measurement and Instrumentation Optical Performance of Nikon F-Mount Lenses Landon Carter May 11, 2016 2.671 Measurement and Instrumentation Abstract In photographic systems, lenses are one of the most important pieces of the system

More information

Astigmatism Particle Tracking Velocimetry for Macroscopic Flows

Astigmatism Particle Tracking Velocimetry for Macroscopic Flows 1TH INTERNATIONAL SMPOSIUM ON PARTICLE IMAGE VELOCIMETR - PIV13 Delft, The Netherlands, July 1-3, 213 Astigmatism Particle Tracking Velocimetry for Macroscopic Flows Thomas Fuchs, Rainer Hain and Christian

More information

Introduction to light microscopy

Introduction to light microscopy Center for Microscopy and Image Anaylsis Introduction to light microscopy (an overview) Microscopy with light Components of a light microscope 1. Light source 2. Objective 3. Sample or specimen holder

More information

Dynamic Phase-Shifting Microscopy Tracks Living Cells

Dynamic Phase-Shifting Microscopy Tracks Living Cells from photonics.com: 04/01/2012 http://www.photonics.com/article.aspx?aid=50654 Dynamic Phase-Shifting Microscopy Tracks Living Cells Dr. Katherine Creath, Goldie Goldstein and Mike Zecchino, 4D Technology

More information

A 3D Profile Parallel Detecting System Based on Differential Confocal Microscopy. Y.H. Wang, X.F. Yu and Y.T. Fei

A 3D Profile Parallel Detecting System Based on Differential Confocal Microscopy. Y.H. Wang, X.F. Yu and Y.T. Fei Key Engineering Materials Online: 005-10-15 ISSN: 166-9795, Vols. 95-96, pp 501-506 doi:10.408/www.scientific.net/kem.95-96.501 005 Trans Tech Publications, Switzerland A 3D Profile Parallel Detecting

More information

UltraGraph Optics Design

UltraGraph Optics Design UltraGraph Optics Design 5/10/99 Jim Hagerman Introduction This paper presents the current design status of the UltraGraph optics. Compromises in performance were made to reach certain product goals. Cost,

More information

Bio 407. Applied microscopy. Introduction into light microscopy. José María Mateos. Center for Microscopy and Image Analysis

Bio 407. Applied microscopy. Introduction into light microscopy. José María Mateos. Center for Microscopy and Image Analysis Center for Microscopy and Image Analysis Bio 407 Applied Introduction into light José María Mateos Fundamentals of light Compound microscope Microscope composed of an objective and an additional lens (eyepiece,

More information

Technology Note ZEISS LSM 880 with Airyscan

Technology Note ZEISS LSM 880 with Airyscan Technology Note ZEISS LSM 880 with Airyscan Introducing the Fast Acquisition Mode ZEISS LSM 880 with Airyscan Introducing the Fast Acquisition Mode Author: Dr. Annette Bergter Carl Zeiss Microscopy GmbH,

More information

contents TABLE OF The SECOM platform Applications - sections Applications - whole cells Features Integrated workflow Automated overlay

contents TABLE OF The SECOM platform Applications - sections Applications - whole cells Features Integrated workflow Automated overlay S E C O M TABLE OF contents The SECOM platform 4 Applications - sections 5 Applications - whole cells 8 Features 9 Integrated workflow 12 Automated overlay ODEMIS - integrated software Specifications 13

More information

Application Note. The New 2D Superresolution Mode for ZEISS Airyscan 120 nm Lateral Resolution without Acquiring a Z-stack

Application Note. The New 2D Superresolution Mode for ZEISS Airyscan 120 nm Lateral Resolution without Acquiring a Z-stack The New 2D Superresolution Mode for ZEISS Airyscan 120 nm Lateral Resolution without Acquiring a Z-stack The New 2D Superresolution Mode for ZEISS Airyscan 120 nm Lateral Resolution without Acquiring a

More information

Systems Biology. Optical Train, Köhler Illumination

Systems Biology. Optical Train, Köhler Illumination McGill University Life Sciences Complex Imaging Facility Systems Biology Microscopy Workshop Tuesday December 7 th, 2010 Simple Lenses, Transmitted Light Optical Train, Köhler Illumination What Does a

More information

BEAM HALO OBSERVATION BY CORONAGRAPH

BEAM HALO OBSERVATION BY CORONAGRAPH BEAM HALO OBSERVATION BY CORONAGRAPH T. Mitsuhashi, KEK, TSUKUBA, Japan Abstract We have developed a coronagraph for the observation of the beam halo surrounding a beam. An opaque disk is set in the beam

More information

Introduction to light microscopy

Introduction to light microscopy Center for Microscopy and Image Anaylsis Introduction to light Basic concepts of imaging with light Urs Ziegler ziegler@zmb.uzh.ch Microscopy with light 1 Light interacting with matter Absorbtion Refraction

More information

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy Qiyuan Song (M2) and Aoi Nakamura (B4) Abstracts: We theoretically and experimentally

More information