A 3D Profile Parallel Detecting System Based on Differential Confocal Microscopy. Y.H. Wang, X.F. Yu and Y.T. Fei

Size: px
Start display at page:

Download "A 3D Profile Parallel Detecting System Based on Differential Confocal Microscopy. Y.H. Wang, X.F. Yu and Y.T. Fei"

Transcription

1 Key Engineering Materials Online: ISSN: , Vols , pp doi:10.408/ 005 Trans Tech Publications, Switzerland A 3D Profile Parallel Detecting System Based on Differential Confocal Microscopy Y.H. Wang, X.F. Yu and Y.T. Fei School of Instrumentation, Hefei University of Technology, Hefei, 30009, China Keywords: Differential, Confocal Microscopy, Parallel Detecting, 3D Profile, Non-Scanning Abstract: Based on the differential confocal microscopy principle, a non-scanning 3D profile detecting system is proposed. A D light source array produced by micro-optic components is used to form a parallel micro-confocal system. The whole field detecting of the measuring plane is realized. The reflected light beam is divided into two paths. Two CCD cameras work together to detect the facula array in a differential arrangement. A differential algorithm of facula intensity is used. The noise and the shift of light source can be avoided effectively. Using the linearity characteristic of the differential confocal system, the high axial resolution is achieved at a larger sampling interval. The measuring efficiency and accuracy can be increased. The construction and working principle of the 3D parallel detecting system and the 3D profile reconstruction method are investigated. Experiment results agree with the ones of theory analysis. It indicates that the differential method is applicable for non-scanning 3D profile detection. Introduction Due to its special advantages of high resolution, high accuracy and its easiness to realize the digitization of 3D image, the confocal measurement has been applied in many fields, such as biomedicine, semiconductor and 3D surface topography detecting [1,]. However, traditional confocal microscopy methods cannot meet the demand of high-speed on line measuring due to their complex construction, low speed and vibration of scanning system. Therefore, a non-scanning 3D profile detecting probe is developed in recent years [3,4]. In this method, multiple-beam light parallel detecting is used instead of the single-point collimates and scans. It can simultaneously realize the whole-field detecting of measured surface. As an alternative to a pinhole diaphragm, the pixels on the CCD sensor directly capture and register the light intensities of the confocal point image. In the traditional parallel confocal detecting system, when the object shifts along the axial direction, the distances between the corresponding point and the focus plane are different because of difference of the heights of each measuring point. This distance can figure out the height of the profile. Although this method is convenient, the noise and shift of the light source will make the references of the micro-light-path different, thus will affect the measuring accuracy. To increase the measuring accuracy, a non-scanning 3D profile parallel detecting system based on differential confocal microscopy method is proposed. Working Principle of the Differential Confocal Microscopy In conventional confocal system light source, measuring point and pinhole detector are conjugated. When the object is located at the focal plane, the reflected light is accurately focused on the detector unit. The light intensity captured by the pinhole detector is the highest. When the object offsets from the focal plane, the reflected light is focused on the front or back position of detector unit. Then, the light intensity captured by detector is reduced because of the pinhole obstructing. Therefore, the axial displacement information can be obtained through the change of the light intensity. The axial light intensity responding to confocal system can be written as [5] All rights reserved. No part of contents of this paper may be reproduced or transmitted in any form or by any means without the written permission of Trans Tech Publications, (ID: , Pennsylvania State University, University Park, USA-1/05/16,3:04:19)

2 50 Measurement Technology and Intelligent Instruments VI [ kz( 1 cosα )] ( 1 cosα ) sin I ( z) =, (1) kz where k is the wave number (k=π/λ, λ represents the wavelength of light wave), z is the offset from the focal plane, and sinα indicates the numerical aperture (NA) of the objective lens. However, the shift of the light source will cause the reference error of the light intensity. If used in the whole-field parallel measurement, the noise of light source will cause the light intensity variation of the different detecting point, thus bring the plane error. Focus plane 1 Detector 1 Detector Light source Beam-splitter Focus plane Object plane Beam-splitter 1 Objective lens Fig. 1 Schematic diagram of differential confocal microscopy The schematic diagram of differential confocal microscopy is shown in Fig. 1. Through the beam-splitter 1, the beam from the point light source is focused on the measured object by the object lens. The reflected beam from the object is divided into two ways by Beam-splitter, and detected by two detectors in a differential arrangement. When the object is located at the focal plane, the facula shape and light intensity captured by two detectors are the same, and the difference output between the signals of two detectors is zero. If the object offsets from the focal plane, the facula shape and light intensity will be different, and the differential signal of two detectors will not be zero. Then the axial shift information can be got through the change of the differential signal. When the offset of detector from the focal plane is ± Z, the axial response of the detector can be expressed as [6] Z sin k Z ± ( 1 cos α ) I ( Z ) ± = I. () 0 Z k Z ± ( 1 cos α ) When the detector offsets, the axial response curve has no change but an axial offset that produces a phase shift (shown in Fig. ). Obviously, the axial offset of the detector does not have much influence on axial response, which has been adopted in the differential focal method. The differential output of detector I D can be expressed as I = I( + Δ Z) I( Δ Z) D ΔZ ΔZ sin k Z + ( 1 cosα) sin k Z ( 1 cosα) = I ΔZ ΔZ k Z + ( 1 cosα) k Z ( 1 cosα) 0. (3)

3 Key Engineering Materials Vols Normalized light intensity I(ΔZ) I D I(-ΔZ) ΔZ 0 -ΔZ Axial offset Fig. Normalized output curve of the detector The output curve consists of several linear ranges (shown in Fig. ). It also shows that normalized light intensity is zero in focal plane. The linearity range and sensitivity of output curve are related to the parameters of the optical system. Since the differential algorithm of facula intensity is used, the noise and the shift of light source act on two detectors simultaneously, so the differential output can effectively reduce the influence of light source. Another advantage of differential confocal method is to improve the detecting speed. In the conventional confocal measurement, the axial sampling interval must be decreased to improve the axial resolution. It will certainly slow the measuring speed. However, since there is satisfactory linear range in the output characteristic curve the measuring speed can be increased by the differential confocal method. As long as the sampling interval is within the linear range, the sample point height can be got from the detector differential output curve. Thus, we can get higher axial resolution than sampling interval. Non-Scanning 3D Profile Parallel Detecting System Light source Focus plane 1 CCD1 Beam-splitter Micro-lens-array Focus plane CCD Object plane HP559A Collimating lens D light source array plane Beam-splitter 1 Objective lens Axial stage Fig. 3 Structure of non-scanning 3D profile detecting system Figure 3 shows the structure of the non-scanning 3D profile detecting system. The D array of point light source is produced by the micro-lens-array. The micro-lens-array is used to separates and then focus the light beam, in order to increases the illumination efficiency. Through the beam-splitter 1, the

4 504 Measurement Technology and Intelligent Instruments VI beam from the point light source is focused on the measured object by the objective lens. Beam-splitter divides the reflected beam from the object into two ways. Two monochrome CCD detectors are arranged in the differential position to detect the light intensity of the differential defocusing facula array. The pixels of CCD detecting surface are divided into many detecting units according to the area of each facula. Each detecting unit corresponds to a facula. Here, each CCD detecting unit is equivalent to a pinhole detector. The CCD image is collected into the computer by the image-sampling card, and used to reconstruct the 3D profile. The displacement device is made up of the displacement table and the displacement driving circuit, which drives the workpiece to make axial movement. 3D Profile Reconstruction While the object moves along the axial direction, the distance between the corresponding point and the focal plane is different because the height of each object point is different. The above distance can figure out the profile height. Therefore, the obtained 3D profile can be transformed to get the sampling position where the output T is zero passage. During the measurement, point-light-array forms the parallel section n times. Suppose the distances between each section are all equal to ΔZ, which is determined by the Z-axial step of micro-displacement driver. The sampling matrix of sampling point (i, j) is expressed as [ Z,I ( i, j) ], [ Z,I ( i, j) ],..., [ Z,I ( i, j) ] D1 D n Dn 1, (4) where I D represents the output of CCD unit and Z n is Z-axial scanning height. The comparison between the outputs T of each section can be made. It shows that, if I D (i,j) is equal to 0, the object point height is calculated by Z k = kδz, (5) where k is the section where I D (i, j) = 0. If none of the output is zero passage, the height can be alternatively obtained through the linearity range of differential output curve. The linearity range and sensitivity of differential output curve are related to the parameters of optical system. We can choose the appropriate parameters of optical system based on the conclusions and the demand of the axial resolution. Experimental Results and Discussion HP559A CCD1 Beam-splitter CCD Light source Objective lens Beam-splitter 1 Micro-lens array Collimating lens Fig. 4 Hardware setup of the 3D profile detecting experiment system

5 Key Engineering Materials Vols In order to validate the feasibility of differential confocal method in the parallel detecting, we have done some primary works for the experiment system (shown in Fig. 4). The surface of a standard gauge was chosen as the object to detect the axial response of the differential facula array. Micro-lens array is 4 by 4 with 100µm dimension. An achromatic doublet objective lens is used. The area array monochrome CCD is used to capture the reflected facula image. The image acquisition card adopts METEORII card made in Canada Matrox Electronic Systems Ltd. A two-frequency laser interferometer (HP559A) is used to calibrate axial shift. The images of facula array, captured by the monochrome CCD in different location of the focal plane, are shown in Fig. 5. These CCD images indicate that the luminous intensity reflected by the detecting system is in good accordance with the characteristics of confocal measurement. In other words, when the object is located at the focal plane, the reflected light is focused on the CCD detecting unit accurately, and the light intensity captured by CCD is the highest. While being out of focus, the reflected light is focused on the front or the back of CCD detecting unit, causing the vague facula and the decreased light intensity captured by CCD. Thus, we can get the axial displacement according to the change of the light intensity. a) negative defocused b) focused c)positive defocused Fig. 5 CCD images of facula array C D B Normalized light intensity Z/ Ì Z (µm) m Fig. 6 Axial response curve of the facula The experiment result of the facula is shown in Fig. 6. Curve C and curve D are the axial response curves of the facula captured by two CCD respectively, and curve B is differential output between C and D. From the curves, we know that the experiment results are in agreement with theory analysis with better linearity in limited range. It demonstrates that effectiveness and applicability of differential confocal method for parallel detecting, and the linearity range and sensitivity of curve are

6 506 Measurement Technology and Intelligent Instruments VI related to the parameters of optical system and CCD offset. However, in the experiment, there is still some non-linearity error between differential output curve and theoretical curve because of optical aberration, CCD noise and CCD gain inequality. The uncertainty and resolution of the experiment system is micro order. If appropriate parameters of optical system, high-quality optical components and CCD are chosen, the higher accuracy and resolution will be obtained. Moreover, the differences of focal length between each micro lens result in different axial resolution of different transverse points, and cause certain error. However, the error is system error, which could be eliminated by calibration and the processing of subsequent software. Conclusions Compared with the conventional confocal system, the differential algorithm of facula intensity can effectively restrain the noise and the shift of light source. By using the linear characteristic of differential confocal system, the high axial resolution is achieved at a larger sampling interval. Furthermore, the method is beneficial to increase the measuring efficiency and accuracy. The linearity range and sensitivity of differential output curve are related to the parameters of optical system. We can choose the appropriate parameters of optical system based on the conclusion and the demand of the axial resolution to meet the requirements of some specified measurement. In high-accuracy measurement, optical aberration and CCD noise have great impact on the accuracy of detecting system. Therefore, further study is need such as how to reduce the effects of optical aberration and noise of CCD in practical application. Acknowledgments This research is sponsored by the National Natural Science Foundation of China (No ), the Important Research Foundation of National Education Ministry of China (No ), and the Research Foundation of Hefei University of Technology (No F). References [1] T.M. Wilson: Confocal Microscopy (Academic Press, New York 1990). [] C.J.R. Sheppard and M. Gu: Journal of Microscopy, Vol. 165 (199), p [3] H.J. Tiziani and H.M. Uhde: Appl. Opt., Vol. 33 (1994), p [4] M. Ishihara and H. Sasaki: Opt. Eng., Vol. 38 (1994), p [5] T.R. Corle and C.H. Chou: Optics Letters, Vol. 11 (1986), p [6] T.M. Wilson and A.R.Carlini: Appl. Opt., Vol. 7 (1988), p

7 Measurement Technology and Intelligent Instruments VI / A 3D Profile Parallel Detecting System Based on Differential Confocal Microscopy / DOI References [5] T.R. Corle and C.H. Chou: Optics Letters, Vol. 11 (1986), p /OL

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI)

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI) Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI) Liang-Chia Chen 1#, Chao-Nan Chen 1 and Yi-Wei Chang 1 1. Institute of Automation Technology,

More information

Confocal Imaging Through Scattering Media with a Volume Holographic Filter

Confocal Imaging Through Scattering Media with a Volume Holographic Filter Confocal Imaging Through Scattering Media with a Volume Holographic Filter Michal Balberg +, George Barbastathis*, Sergio Fantini % and David J. Brady University of Illinois at Urbana-Champaign, Urbana,

More information

Three-dimensional quantitative phase measurement by Commonpath Digital Holographic Microscopy

Three-dimensional quantitative phase measurement by Commonpath Digital Holographic Microscopy Available online at www.sciencedirect.com Physics Procedia 19 (2011) 291 295 International Conference on Optics in Precision Engineering and Nanotechnology Three-dimensional quantitative phase measurement

More information

Innovative full-field chromatic confocal microscopy using multispectral sensors

Innovative full-field chromatic confocal microscopy using multispectral sensors Innovative full-field chromatic confocal microscopy using multispectral sensors Liang-Chia Chen 1, 2, a#, Pei-Ju Tan 2, b, Chih-Jer Lin 2,c, Duc Trung Nguyen 1,d, Yu-Shuan Chou 1,e, Nguyen Dinh Nguyen

More information

Automatic optical measurement of high density fiber connector

Automatic optical measurement of high density fiber connector Key Engineering Materials Online: 2014-08-11 ISSN: 1662-9795, Vol. 625, pp 305-309 doi:10.4028/www.scientific.net/kem.625.305 2015 Trans Tech Publications, Switzerland Automatic optical measurement of

More information

Parallel Mode Confocal System for Wafer Bump Inspection

Parallel Mode Confocal System for Wafer Bump Inspection Parallel Mode Confocal System for Wafer Bump Inspection ECEN5616 Class Project 1 Gao Wenliang wen-liang_gao@agilent.com 1. Introduction In this paper, A parallel-mode High-speed Line-scanning confocal

More information

Image Measurement of Roller Chain Board Based on CCD Qingmin Liu 1,a, Zhikui Liu 1,b, Qionghong Lei 2,c and Kui Zhang 1,d

Image Measurement of Roller Chain Board Based on CCD Qingmin Liu 1,a, Zhikui Liu 1,b, Qionghong Lei 2,c and Kui Zhang 1,d Applied Mechanics and Materials Online: 2010-11-11 ISSN: 1662-7482, Vols. 37-38, pp 513-516 doi:10.4028/www.scientific.net/amm.37-38.513 2010 Trans Tech Publications, Switzerland Image Measurement of Roller

More information

Supplementary Materials

Supplementary Materials Supplementary Materials In the supplementary materials of this paper we discuss some practical consideration for alignment of optical components to help unexperienced users to achieve a high performance

More information

Reflecting optical system to increase signal intensity. in confocal microscopy

Reflecting optical system to increase signal intensity. in confocal microscopy Reflecting optical system to increase signal intensity in confocal microscopy DongKyun Kang *, JungWoo Seo, DaeGab Gweon Nano Opto Mechatronics Laboratory, Dept. of Mechanical Engineering, Korea Advanced

More information

Dynamic beam shaping with programmable diffractive optics

Dynamic beam shaping with programmable diffractive optics Dynamic beam shaping with programmable diffractive optics Bosanta R. Boruah Dept. of Physics, GU Page 1 Outline of the talk Introduction Holography Programmable diffractive optics Laser scanning confocal

More information

Confocal principle for macro- and microscopic surface and defect analysis

Confocal principle for macro- and microscopic surface and defect analysis Confocal principle for macro- and microscopic surface and defect analysis Hans J. Tiziani, FELLOW SPIE Michael Wegner Daniela Steudle Institut für Technische Optik Pfaffenwaldring 9 70569 Stuttgart, Germany

More information

Quasi one-shot full-field surface profilometry using digital diffractive-confocal imaging correlation microscope

Quasi one-shot full-field surface profilometry using digital diffractive-confocal imaging correlation microscope Quasi one-shot full-field surface profilometry using digital diffractive-confocal imaging correlation microscope Duc Trung Nguyen 1, Liang-Chia Chen* 1,2, Nguyen Dinh Nguyen 1 1 Mechanical Engineering

More information

Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA

Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA Abstract: Speckle interferometry (SI) has become a complete technique over the past couple of years and is widely used in many branches of

More information

Use of Computer Generated Holograms for Testing Aspheric Optics

Use of Computer Generated Holograms for Testing Aspheric Optics Use of Computer Generated Holograms for Testing Aspheric Optics James H. Burge and James C. Wyant Optical Sciences Center, University of Arizona, Tucson, AZ 85721 http://www.optics.arizona.edu/jcwyant,

More information

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy Bi177 Lecture 5 Adding the Third Dimension Wide-field Imaging Point Spread Function Deconvolution Confocal Laser Scanning Microscopy Confocal Aperture Optical aberrations Alternative Scanning Microscopy

More information

Confocal Microscopy and Related Techniques

Confocal Microscopy and Related Techniques Confocal Microscopy and Related Techniques Chau-Hwang Lee Associate Research Fellow Research Center for Applied Sciences, Academia Sinica 128 Sec. 2, Academia Rd., Nankang, Taipei 11529, Taiwan E-mail:

More information

Single-shot three-dimensional imaging of dilute atomic clouds

Single-shot three-dimensional imaging of dilute atomic clouds Calhoun: The NPS Institutional Archive Faculty and Researcher Publications Funded by Naval Postgraduate School 2014 Single-shot three-dimensional imaging of dilute atomic clouds Sakmann, Kaspar http://hdl.handle.net/10945/52399

More information

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G APPLICATION NOTE M01 attocfm I for Surface Quality Inspection Confocal microscopes work by scanning a tiny light spot on a sample and by measuring the scattered light in the illuminated volume. First,

More information

Wavefront sensing by an aperiodic diffractive microlens array

Wavefront sensing by an aperiodic diffractive microlens array Wavefront sensing by an aperiodic diffractive microlens array Lars Seifert a, Thomas Ruppel, Tobias Haist, and Wolfgang Osten a Institut für Technische Optik, Universität Stuttgart, Pfaffenwaldring 9,

More information

An Indian Journal FULL PAPER. Trade Science Inc. Parameters design of optical system in transmitive star simulator ABSTRACT KEYWORDS

An Indian Journal FULL PAPER. Trade Science Inc. Parameters design of optical system in transmitive star simulator ABSTRACT KEYWORDS [Type text] [Type text] [Type text] ISSN : 0974-7435 Volume 10 Issue 23 BioTechnology 2014 An Indian Journal FULL PAPER BTAIJ, 10(23), 2014 [14257-14264] Parameters design of optical system in transmitive

More information

Bringing Answers to the Surface

Bringing Answers to the Surface 3D Bringing Answers to the Surface 1 Expanding the Boundaries of Laser Microscopy Measurements and images you can count on. Every time. LEXT OLS4100 Widely used in quality control, research, and development

More information

White-light interferometry, Hilbert transform, and noise

White-light interferometry, Hilbert transform, and noise White-light interferometry, Hilbert transform, and noise Pavel Pavlíček *a, Václav Michálek a a Institute of Physics of Academy of Science of the Czech Republic, Joint Laboratory of Optics, 17. listopadu

More information

PROCEEDINGS OF SPIE. Measurement of low-order aberrations with an autostigmatic microscope

PROCEEDINGS OF SPIE. Measurement of low-order aberrations with an autostigmatic microscope PROCEEDINGS OF SPIE SPIEDigitalLibrary.org/conference-proceedings-of-spie Measurement of low-order aberrations with an autostigmatic microscope William P. Kuhn Measurement of low-order aberrations with

More information

Point Autofocus Probe Surface Texture Measuring Instrument. PF-60 technical report

Point Autofocus Probe Surface Texture Measuring Instrument. PF-60 technical report Point Autofocus Probe Surface Texture Measuring Instrument PF-60 technical report ISO approved Mitaka measuring method for areal surface texture (ISO 25178-605) Document No, Title Published ISO 25178-6

More information

Single-shot depth-section imaging through chromatic slit-scan confocal microscopy

Single-shot depth-section imaging through chromatic slit-scan confocal microscopy Single-shot depth-section imaging through chromatic slit-scan confocal microscopy Paul C. Lin, Pang-Chen Sun, Lijun Zhu, and Yeshaiahu Fainman A chromatic confocal microscope constructed with a white-light

More information

EVALUATION OF A FOCUSED LASER SPOT DIAMETER FOR AN OPTICAL ANGLE SENSOR. Yuki Shimizu, Taiji Maruyama, Shota Nakagawa, Yuan-Liu Chen, Wei Gao

EVALUATION OF A FOCUSED LASER SPOT DIAMETER FOR AN OPTICAL ANGLE SENSOR. Yuki Shimizu, Taiji Maruyama, Shota Nakagawa, Yuan-Liu Chen, Wei Gao 59 th ILMENAU SCIENTIFIC COLLOQUIUM Technische Universität Ilmenau, 11 15 September 2017 URN: urn:nbn:de:gbv:ilm1-2017iwk-071:8 EVALUATION OF A FOCUSED LASER SPOT DIAMETER FOR AN OPTICAL ANGLE SENSOR Yuki

More information

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy Qiyuan Song (M2) and Aoi Nakamura (B4) Abstracts: We theoretically and experimentally

More information

Ron Liu OPTI521-Introductory Optomechanical Engineering December 7, 2009

Ron Liu OPTI521-Introductory Optomechanical Engineering December 7, 2009 Synopsis of METHOD AND APPARATUS FOR IMPROVING VISION AND THE RESOLUTION OF RETINAL IMAGES by David R. Williams and Junzhong Liang from the US Patent Number: 5,777,719 issued in July 7, 1998 Ron Liu OPTI521-Introductory

More information

Advanced 3D Optical Profiler using Grasshopper3 USB3 Vision camera

Advanced 3D Optical Profiler using Grasshopper3 USB3 Vision camera Advanced 3D Optical Profiler using Grasshopper3 USB3 Vision camera Figure 1. The Zeta-20 uses the Grasshopper3 and produces true color 3D optical images with multi mode optics technology 3D optical profiling

More information

In-line digital holographic interferometry

In-line digital holographic interferometry In-line digital holographic interferometry Giancarlo Pedrini, Philipp Fröning, Henrik Fessler, and Hans J. Tiziani An optical system based on in-line digital holography for the evaluation of deformations

More information

Nontranslational three-dimensional profilometry by chromatic confocal microscopy with dynamically configurable micromirror scanning

Nontranslational three-dimensional profilometry by chromatic confocal microscopy with dynamically configurable micromirror scanning Nontranslational three-dimensional profilometry by chromatic confocal microscopy with dynamically configurable micromirror scanning Sungdo Cha, Paul C. Lin, Lijun Zhu, Pang-Chen Sun, and Yeshaiahu Fainman

More information

Development of Laser Confocal Microscopy for Internal Defect Measurement

Development of Laser Confocal Microscopy for Internal Defect Measurement Development of Laser Confocal Microscopy for Internal Defect Measurement Chia-Liang Yeh*, Fu-Cheng Yang, Wei-Hsiung Tsai, and Keng-Li Lin Center for Measurement Standards, Industrial Technology Research

More information

AgilOptics mirrors increase coupling efficiency into a 4 µm diameter fiber by 750%.

AgilOptics mirrors increase coupling efficiency into a 4 µm diameter fiber by 750%. Application Note AN004: Fiber Coupling Improvement Introduction AgilOptics mirrors increase coupling efficiency into a 4 µm diameter fiber by 750%. Industrial lasers used for cutting, welding, drilling,

More information

The spectral colours of nanometers

The spectral colours of nanometers Reprint from the journal Mikroproduktion 3/2005 Berthold Michelt and Jochen Schulze The spectral colours of nanometers Precitec Optronik GmbH Raiffeisenstraße 5 D-63110 Rodgau Phone: +49 (0) 6106 8290-14

More information

Study of self-interference incoherent digital holography for the application of retinal imaging

Study of self-interference incoherent digital holography for the application of retinal imaging Study of self-interference incoherent digital holography for the application of retinal imaging Jisoo Hong and Myung K. Kim Department of Physics, University of South Florida, Tampa, FL, US 33620 ABSTRACT

More information

EE119 Introduction to Optical Engineering Spring 2002 Final Exam. Name:

EE119 Introduction to Optical Engineering Spring 2002 Final Exam. Name: EE119 Introduction to Optical Engineering Spring 2002 Final Exam Name: SID: CLOSED BOOK. FOUR 8 1/2 X 11 SHEETS OF NOTES, AND SCIENTIFIC POCKET CALCULATOR PERMITTED. TIME ALLOTTED: 180 MINUTES Fundamental

More information

Sensitive measurement of partial coherence using a pinhole array

Sensitive measurement of partial coherence using a pinhole array 1.3 Sensitive measurement of partial coherence using a pinhole array Paul Petruck 1, Rainer Riesenberg 1, Richard Kowarschik 2 1 Institute of Photonic Technology, Albert-Einstein-Strasse 9, 07747 Jena,

More information

Testing Aspheric Lenses: New Approaches

Testing Aspheric Lenses: New Approaches Nasrin Ghanbari OPTI 521 - Synopsis of a published Paper November 5, 2012 Testing Aspheric Lenses: New Approaches by W. Osten, B. D orband, E. Garbusi, Ch. Pruss, and L. Seifert Published in 2010 Introduction

More information

Experimental demonstration of polarization-assisted transverse and axial optical superresolution

Experimental demonstration of polarization-assisted transverse and axial optical superresolution Optics Communications 241 (2004) 315 319 www.elsevier.com/locate/optcom Experimental demonstration of polarization-assisted transverse and axial optical superresolution Jason B. Stewart a, *, Bahaa E.A.

More information

Comparison of resolution specifications for micro- and nanometer measurement techniques

Comparison of resolution specifications for micro- and nanometer measurement techniques P4.5 Comparison of resolution specifications for micro- and nanometer measurement techniques Weckenmann/Albert, Tan/Özgür, Shaw/Laura, Zschiegner/Nils Chair Quality Management and Manufacturing Metrology

More information

3D light microscopy techniques

3D light microscopy techniques 3D light microscopy techniques The image of a point is a 3D feature In-focus image Out-of-focus image The image of a point is not a point Point Spread Function (PSF) 1D imaging 1 1 2! NA = 0.5! NA 2D imaging

More information

Optical Coherence: Recreation of the Experiment of Thompson and Wolf

Optical Coherence: Recreation of the Experiment of Thompson and Wolf Optical Coherence: Recreation of the Experiment of Thompson and Wolf David Collins Senior project Department of Physics, California Polytechnic State University San Luis Obispo June 2010 Abstract The purpose

More information

Systematic Workflow via Intuitive GUI. Easy operation accomplishes your goals faster than ever.

Systematic Workflow via Intuitive GUI. Easy operation accomplishes your goals faster than ever. Systematic Workflow via Intuitive GUI Easy operation accomplishes your goals faster than ever. 16 With the LEXT OLS4100, observation or measurement begins immediately once the sample is placed on the stage.

More information

A broadband achromatic metalens for focusing and imaging in the visible

A broadband achromatic metalens for focusing and imaging in the visible SUPPLEMENTARY INFORMATION Articles https://doi.org/10.1038/s41565-017-0034-6 In the format provided by the authors and unedited. A broadband achromatic metalens for focusing and imaging in the visible

More information

Parallel Digital Holography Three-Dimensional Image Measurement Technique for Moving Cells

Parallel Digital Holography Three-Dimensional Image Measurement Technique for Moving Cells F e a t u r e A r t i c l e Feature Article Parallel Digital Holography Three-Dimensional Image Measurement Technique for Moving Cells Yasuhiro Awatsuji The author invented and developed a technique capable

More information

Open Access Structural Parameters Optimum Design of the New Type of Optical Aiming

Open Access Structural Parameters Optimum Design of the New Type of Optical Aiming Send Orders for Reprints to reprints@benthamscience.ae 208 The Open Electrical & Electronic Engineering Journal, 2014, 8, 208-212 Open Access Structural Parameters Optimum Design of the New Type of Optical

More information

Defocusing Effect Studies in MTF of CCD Cameras Based on PSF Measuring Technique

Defocusing Effect Studies in MTF of CCD Cameras Based on PSF Measuring Technique International Journal of Optics and Photonics (IJOP) Vol. 9, No. 2, Summer-Fall, 2015 Defocusing Effect Studies in MTF of CCD Cameras Based on PSF Measuring Technique Amir Hossein Shahbazi a, Khosro Madanipour

More information

Nanonics Systems are the Only SPMs that Allow for On-line Integration with Standard MicroRaman Geometries

Nanonics Systems are the Only SPMs that Allow for On-line Integration with Standard MicroRaman Geometries Nanonics Systems are the Only SPMs that Allow for On-line Integration with Standard MicroRaman Geometries 2002 Photonics Circle of Excellence Award PLC Ltd, England, a premier provider of Raman microspectral

More information

Investigation of an optical sensor for small angle detection

Investigation of an optical sensor for small angle detection Investigation of an optical sensor for small angle detection usuke Saito, oshikazu rai and Wei Gao Nano-Metrology and Control Lab epartment of Nanomechanics Graduate School of Engineering, Tohoku University

More information

Aberrations and adaptive optics for biomedical microscopes

Aberrations and adaptive optics for biomedical microscopes Aberrations and adaptive optics for biomedical microscopes Martin Booth Department of Engineering Science And Centre for Neural Circuits and Behaviour University of Oxford Outline Rays, wave fronts and

More information

University of Huddersfield Repository

University of Huddersfield Repository University of Huddersfield Repository Gao, F., Muhamedsalih, Hussam and Jiang, Xiang In process fast surface measurement using wavelength scanning interferometry Original Citation Gao, F., Muhamedsalih,

More information

SENSOR+TEST Conference SENSOR 2009 Proceedings II

SENSOR+TEST Conference SENSOR 2009 Proceedings II B8.4 Optical 3D Measurement of Micro Structures Ettemeyer, Andreas; Marxer, Michael; Keferstein, Claus NTB Interstaatliche Hochschule für Technik Buchs Werdenbergstr. 4, 8471 Buchs, Switzerland Introduction

More information

Measurement of the Modulation Transfer Function (MTF) of a camera lens. Laboratoire d Enseignement Expérimental (LEnsE)

Measurement of the Modulation Transfer Function (MTF) of a camera lens. Laboratoire d Enseignement Expérimental (LEnsE) Measurement of the Modulation Transfer Function (MTF) of a camera lens Aline Vernier, Baptiste Perrin, Thierry Avignon, Jean Augereau, Lionel Jacubowiez Institut d Optique Graduate School Laboratoire d

More information

Microscopic Structures

Microscopic Structures Microscopic Structures Image Analysis Metal, 3D Image (Red-Green) The microscopic methods range from dark field / bright field microscopy through polarisation- and inverse microscopy to techniques like

More information

REAL TIME THICKNESS MEASUREMENT OF A MOVING WIRE

REAL TIME THICKNESS MEASUREMENT OF A MOVING WIRE REAL TIME THICKNESS MEASUREMENT OF A MOVING WIRE Bini Babu 1, Dr. Ashok Kumar T 2 1 Optoelectronics and communication systems, 2 Associate Professor Model Engineering college, Thrikkakara, Ernakulam, (India)

More information

Digital Photographic Imaging Using MOEMS

Digital Photographic Imaging Using MOEMS Digital Photographic Imaging Using MOEMS Vasileios T. Nasis a, R. Andrew Hicks b and Timothy P. Kurzweg a a Department of Electrical and Computer Engineering, Drexel University, Philadelphia, USA b Department

More information

Displacement sensor by a common-path interferometer

Displacement sensor by a common-path interferometer Displacement sensor by a common-path interferometer Kazuhide KAMIYA *a, Takashi NOMURA *a, Shinta HIDAKA *a, Hatsuzo TASHIRO **b, Masayuki MINO +c, Seiichi OKUDA ++d a Facility of Engineering, Toyama Prefectural

More information

Shaping light in microscopy:

Shaping light in microscopy: Shaping light in microscopy: Adaptive optical methods and nonconventional beam shapes for enhanced imaging Martí Duocastella planet detector detector sample sample Aberrated wavefront Beamsplitter Adaptive

More information

Sensitivity Enhancement of Bimaterial MOEMS Thermal Imaging Sensor Array using 2-λ readout

Sensitivity Enhancement of Bimaterial MOEMS Thermal Imaging Sensor Array using 2-λ readout Sensitivity Enhancement of Bimaterial MOEMS Thermal Imaging Sensor Array using -λ readout O. Ferhanoğlu, H. Urey Koç University, Electrical Engineering, Istanbul-TURKEY ABSTRACT Diffraction gratings integrated

More information

Simple interferometric fringe stabilization by CCD-based feedback control

Simple interferometric fringe stabilization by CCD-based feedback control Simple interferometric fringe stabilization by CCD-based feedback control Preston P. Young and Purnomo S. Priambodo, Department of Electrical Engineering, University of Texas at Arlington, P.O. Box 19016,

More information

Numerical simulation of a gradient-index fibre probe and its properties of light propagation

Numerical simulation of a gradient-index fibre probe and its properties of light propagation Numerical simulation of a gradient-index fibre probe and its properties of light propagation Wang Chi( ) a), Mao You-Xin( ) b), Tang Zhi( ) a), Fang Chen( ) a), Yu Ying-Jie( ) a), and Qi Bo( ) c) a) Department

More information

CCD Automatic Gain Algorithm Design of Noncontact Measurement System Based on High-speed Circuit Breaker

CCD Automatic Gain Algorithm Design of Noncontact Measurement System Based on High-speed Circuit Breaker 2016 3 rd International Conference on Engineering Technology and Application (ICETA 2016) ISBN: 978-1-60595-383-0 CCD Automatic Gain Algorithm Design of Noncontact Measurement System Based on High-speed

More information

Macro Varon 4.5/85. Key features. Applications. Web and surface inspections

Macro Varon 4.5/85. Key features. Applications. Web and surface inspections The Macro Varon lens has been designed for high resolution 12k line scan cameras with 3.5 µm pixel pitch. They are optimized for an optical magnification range of.5x to 2.x. CAS-lens technology produces

More information

Fiber Optic Sensing Applications Based on Optical Propagation Mode Time Delay Measurement

Fiber Optic Sensing Applications Based on Optical Propagation Mode Time Delay Measurement R ESEARCH ARTICLE ScienceAsia 7 (1) : 35-4 Fiber Optic Sensing Applications Based on Optical Propagation Mode Time Delay Measurement PP Yupapin a * and S Piengbangyang b a Lightwave Technology Research

More information

WaveMaster IOL. Fast and accurate intraocular lens tester

WaveMaster IOL. Fast and accurate intraocular lens tester WaveMaster IOL Fast and accurate intraocular lens tester INTRAOCULAR LENS TESTER WaveMaster IOL Fast and accurate intraocular lens tester WaveMaster IOL is a new instrument providing real time analysis

More information

ADALAM Sensor based adaptive laser micromachining using ultrashort pulse lasers for zero-failure manufacturing

ADALAM Sensor based adaptive laser micromachining using ultrashort pulse lasers for zero-failure manufacturing 01/01/2015 Deliverable D2.3 Active alignment unit for beam coupling and sensor integration based on adaptive optics D2.3 Active alignment unit for beam coupling and sensor integration based on adaptive

More information

Conformal optical system design with a single fixed conic corrector

Conformal optical system design with a single fixed conic corrector Conformal optical system design with a single fixed conic corrector Song Da-Lin( ), Chang Jun( ), Wang Qing-Feng( ), He Wu-Bin( ), and Cao Jiao( ) School of Optoelectronics, Beijing Institute of Technology,

More information

WaveMaster IOL. Fast and Accurate Intraocular Lens Tester

WaveMaster IOL. Fast and Accurate Intraocular Lens Tester WaveMaster IOL Fast and Accurate Intraocular Lens Tester INTRAOCULAR LENS TESTER WaveMaster IOL Fast and accurate intraocular lens tester WaveMaster IOL is an instrument providing real time analysis of

More information

The extended-focus, auto-focus and surface-profiling techniques of confocal microscopy

The extended-focus, auto-focus and surface-profiling techniques of confocal microscopy JOURNAL OF MODERN OPTICS, 1988, voi,. 35, NO. 1, 145-154 The extended-focus, auto-focus and surface-profiling techniques of confocal microscopy C. J. R. SHEPPARD and H. J. MATTHEWS University of Oxford,

More information

University, Harbin, The 49th Research Institute of China Electronics Technology Group Corporation, Harbin,

University, Harbin, The 49th Research Institute of China Electronics Technology Group Corporation, Harbin, Key Engineering Materials Online: 2013-07-15 ISSN: 1662-9795, Vols. 562-565, pp 465-470 doi:10.4028/www.scientific.net/kem.562-565.465 2013 Trans Tech Publications, Switzerland Simulation research of the

More information

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION Revised November 15, 2017 INTRODUCTION The simplest and most commonly described examples of diffraction and interference from two-dimensional apertures

More information

Bias errors in PIV: the pixel locking effect revisited.

Bias errors in PIV: the pixel locking effect revisited. Bias errors in PIV: the pixel locking effect revisited. E.F.J. Overmars 1, N.G.W. Warncke, C. Poelma and J. Westerweel 1: Laboratory for Aero & Hydrodynamics, University of Technology, Delft, The Netherlands,

More information

Precision Machining by Optical Image Type Tool Measurement System

Precision Machining by Optical Image Type Tool Measurement System 10 Precision Machining by Optical Image Type Tool Measurement System YOSHIKATSU SATO *1 Due to the globalization of production bases and increasing demand for accuracy in recent years, machines and applications

More information

3D light microscopy techniques

3D light microscopy techniques 3D light microscopy techniques The image of a point is a 3D feature In-focus image Out-of-focus image The image of a point is not a point Point Spread Function (PSF) 1D imaging 2D imaging 3D imaging Resolution

More information

Development of a Low-order Adaptive Optics System at Udaipur Solar Observatory

Development of a Low-order Adaptive Optics System at Udaipur Solar Observatory J. Astrophys. Astr. (2008) 29, 353 357 Development of a Low-order Adaptive Optics System at Udaipur Solar Observatory A. R. Bayanna, B. Kumar, R. E. Louis, P. Venkatakrishnan & S. K. Mathew Udaipur Solar

More information

Aberrated Microlenses to Reduce Crosstalk in Free Space Optical Interconnects Systems

Aberrated Microlenses to Reduce Crosstalk in Free Space Optical Interconnects Systems Modern Applied Science; Vol., No. 5; 8 ISSN 93-844 E-ISSN 93-85 Published by Canadian Center of Science and Education Aberrated Microlenses to Reduce Crosstalk in Free Space Optical Interconnects Systems

More information

Rotation/ scale invariant hybrid digital/optical correlator system for automatic target recognition

Rotation/ scale invariant hybrid digital/optical correlator system for automatic target recognition Rotation/ scale invariant hybrid digital/optical correlator system for automatic target recognition V. K. Beri, Amit Aran, Shilpi Goyal, and A. K. Gupta * Photonics Division Instruments Research and Development

More information

Optimal Pupil Design for Confocal Microscopy

Optimal Pupil Design for Confocal Microscopy Optimal Pupil Design for Confocal Microscopy Yogesh G. Patel 1, Milind Rajadhyaksha 3, and Charles A. DiMarzio 1,2 1 Department of Electrical and Computer Engineering, 2 Department of Mechanical and Industrial

More information

(51) Int Cl.: G01B 9/02 ( ) G01B 11/24 ( ) G01N 21/47 ( )

(51) Int Cl.: G01B 9/02 ( ) G01B 11/24 ( ) G01N 21/47 ( ) (19) (12) EUROPEAN PATENT APPLICATION (11) EP 1 939 581 A1 (43) Date of publication: 02.07.2008 Bulletin 2008/27 (21) Application number: 07405346.3 (51) Int Cl.: G01B 9/02 (2006.01) G01B 11/24 (2006.01)

More information

1 Introduction. Research Article

1 Introduction. Research Article dv. Opt. Techn. 214; 3(4): 425 433 Research rticle Hiroki Yokozeki, Ryota Kudo, Satoru Takahashi* and Kiyoshi Takamasu Lateral resolution improvement of laser-scanning imaging for nano defects detection

More information

Development of innovative fringe locking strategies for vibration-resistant white light vertical scanning interferometry (VSI)

Development of innovative fringe locking strategies for vibration-resistant white light vertical scanning interferometry (VSI) Development of innovative fringe locking strategies for vibration-resistant white light vertical scanning interferometry (VSI) Liang-Chia Chen 1), Abraham Mario Tapilouw 1), Sheng-Lih Yeh 2), Shih-Tsong

More information

Comparison of an Optical-Digital Restoration Technique with Digital Methods for Microscopy Defocused Images

Comparison of an Optical-Digital Restoration Technique with Digital Methods for Microscopy Defocused Images Comparison of an Optical-Digital Restoration Technique with Digital Methods for Microscopy Defocused Images R. Ortiz-Sosa, L.R. Berriel-Valdos, J. F. Aguilar Instituto Nacional de Astrofísica Óptica y

More information

:... resolution is about 1.4 μm, assumed an excitation wavelength of 633 nm and a numerical aperture of 0.65 at 633 nm.

:... resolution is about 1.4 μm, assumed an excitation wavelength of 633 nm and a numerical aperture of 0.65 at 633 nm. PAGE 30 & 2008 2007 PRODUCT CATALOG Confocal Microscopy - CFM fundamentals :... Over the years, confocal microscopy has become the method of choice for obtaining clear, three-dimensional optical images

More information

EE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name:

EE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name: EE119 Introduction to Optical Engineering Fall 2009 Final Exam Name: SID: CLOSED BOOK. THREE 8 1/2 X 11 SHEETS OF NOTES, AND SCIENTIFIC POCKET CALCULATOR PERMITTED. TIME ALLOTTED: 180 MINUTES Fundamental

More information

Handbook of Optical Systems

Handbook of Optical Systems Handbook of Optical Systems Volume 5: Metrology of Optical Components and Systems von Herbert Gross, Bernd Dörband, Henriette Müller 1. Auflage Handbook of Optical Systems Gross / Dörband / Müller schnell

More information

Properties of Structured Light

Properties of Structured Light Properties of Structured Light Gaussian Beams Structured light sources using lasers as the illumination source are governed by theories of Gaussian beams. Unlike incoherent sources, coherent laser sources

More information

Instructions for the Experiment

Instructions for the Experiment Instructions for the Experiment Excitonic States in Atomically Thin Semiconductors 1. Introduction Alongside with electrical measurements, optical measurements are an indispensable tool for the study of

More information

Optical Components for Laser Applications. Günter Toesko - Laserseminar BLZ im Dezember

Optical Components for Laser Applications. Günter Toesko - Laserseminar BLZ im Dezember Günter Toesko - Laserseminar BLZ im Dezember 2009 1 Aberrations An optical aberration is a distortion in the image formed by an optical system compared to the original. It can arise for a number of reasons

More information

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the

More information

Fast Lateral Line-scan Model with Doublet-cylinder-lens in Intelligent OCT

Fast Lateral Line-scan Model with Doublet-cylinder-lens in Intelligent OCT International Conference on Advanced Computer Science and Electronics Information (ICACSEI 3) Fast Lateral Line-scan Model with Doublet-cylinder-lens in Intelligent OCT Shijun Xu, Xiaoling Ren, Chunmin

More information

Variable microinspection system. system125

Variable microinspection system. system125 Variable microinspection system system125 Variable micro-inspection system Characteristics Large fields, high NA The variable microinspection system mag.x system125 stands out from conventional LD inspection

More information

Very short introduction to light microscopy and digital imaging

Very short introduction to light microscopy and digital imaging Very short introduction to light microscopy and digital imaging Hernan G. Garcia August 1, 2005 1 Light Microscopy Basics In this section we will briefly describe the basic principles of operation and

More information

PROCEEDINGS OF SPIE. Measurement of the modulation transfer function (MTF) of a camera lens

PROCEEDINGS OF SPIE. Measurement of the modulation transfer function (MTF) of a camera lens PROCEEDINGS OF SPIE SPIEDigitalLibrary.org/conference-proceedings-of-spie Measurement of the modulation transfer function (MTF) of a camera lens Aline Vernier, Baptiste Perrin, Thierry Avignon, Jean Augereau,

More information

VC 14/15 TP2 Image Formation

VC 14/15 TP2 Image Formation VC 14/15 TP2 Image Formation Mestrado em Ciência de Computadores Mestrado Integrado em Engenharia de Redes e Sistemas Informáticos Miguel Tavares Coimbra Outline Computer Vision? The Human Visual System

More information

Fastest high definition Raman imaging. Fastest Laser Raman Microscope RAMAN

Fastest high definition Raman imaging. Fastest Laser Raman Microscope RAMAN Fastest high definition Raman imaging Fastest Laser Raman Microscope RAMAN - 11 www.nanophoton.jp Observation A New Generation in Raman Observation RAMAN-11 developed by Nanophoton was newly created by

More information

A novel solution for various monitoring applications at CERN

A novel solution for various monitoring applications at CERN A novel solution for various monitoring applications at CERN F. Lackner, P. H. Osanna 1, W. Riegler, H. Kopetz CERN, European Organisation for Nuclear Research, CH-1211 Geneva-23, Switzerland 1 Department

More information

Profile Measurement of Resist Surface Using Multi-Array-Probe System

Profile Measurement of Resist Surface Using Multi-Array-Probe System Sensors & Transducers 2014 by IFSA Publishing, S. L. http://www.sensorsportal.com Profile Measurement of Resist Surface Using Multi-Array-Probe System Shujie LIU, Yuanliang ZHANG and Zuolan YUAN School

More information

Adaptive optimisation of illumination beam profiles in fluorescence microscopy

Adaptive optimisation of illumination beam profiles in fluorescence microscopy Adaptive optimisation of illumination beam profiles in fluorescence microscopy T. J. Mitchell a, C. D. Saunter a, W. O Nions a, J. M. Girkin a, G. D. Love a a Centre for Advanced nstrumentation & Biophysical

More information

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

3. are adherent cells (ie. cells in suspension are too far away from the coverslip) Before you begin, make sure your sample... 1. is seeded on #1.5 coverglass (thickness = 0.17) 2. is an aqueous solution (ie. fixed samples mounted on a slide will not work - not enough difference in refractive

More information

Effective Cutoff Wavelength Measurement of Bend-insensitive Fiber by Longitudinal Misalignment Loss Method. Won-Taek Han

Effective Cutoff Wavelength Measurement of Bend-insensitive Fiber by Longitudinal Misalignment Loss Method. Won-Taek Han Advanced Materials Research Vols. 123-125 (2010) pp 419-422 Online available since 2010/Aug/11 at www.scientific.net (2010) Trans Tech Publications, Switzerland doi:10.4028/www.scientific.net/amr.123-125.419

More information