Characterization of Common Electron Multipliers in Harsh Environments

Size: px
Start display at page:

Download "Characterization of Common Electron Multipliers in Harsh Environments"

Transcription

1 ELECTRO-OPTICS Characterization of Common Electron Multipliers in Harsh Environments The Pittsburgh Conference 2005 Poster Paper Bruce Laprade and Raymond Cochran BURLE Electro-Optics INC

2 Introduction ELECTRO-OPTICS Mass Spectrometers are a valuable tool for use in demanding applications such as drug discovery, semiconductor manufacturing and food processing. The unique capability of a mass spectrometer to identify minute amounts of unknown materials sets this instrument apart from almost all others. The typical Mass Spectrometer has three key components: The ionization source, the mass filter and the detector.

3 Discussion ELECTRO-OPTICS Homeland security requirements are now driving the development of field portable instruments that can provide laboratory grade analysis. In order to reduce the size, weight and power consumption of a portable mass spectrometer, it is often necessary to deviate from the normal operating environment inside the vacuum system.

4 Objective The objective of this project was to characterize the performance of various electron multipliers under harsh test conditions.

5 ELECTRO-OPTICS There are 3 main types of Electron Multipliers: - Single Channel Electron Multipliers - Microchannel Plate based Detectors -Discrete Dynode Multipliers

6 Various Types of Electron Multipliers Microchannel Plates Single Channel Electron Multipliers Discrete Dynode Magnum 6 Channel Multipliers Microchannel Plate Detectors

7 Vacuum Conditions Typical MALDI Time-of-Flight Mass Spectrometers require an ion flight path length of 1 to 2 meters. Modern quadrupole based instruments typically operate at vacuum pressures in the high 10-6 Torr range because ions need to travel 25 cm. or more before they reach the detector without colliding with residual gas molecules. If it were possible to shrink the flight path length requirement to less than 5cm (approx... 2 ) then it should be possible to operate a system in the milli-torr range. Milli-Torr vacuum levels can be achieved with simple low cost vacuum pumps. Multiplier performance outside of the normal 10-6 Torr operating range has not been well characterized.

8 The Effects of Poor Operating Pressure on Detector Gain and Noise

9 Output Current (amps) 1.00E-06 Single Stage MCP Analog Operation in Argon at Various Gain Settings 1.00E-05 Gain E-07 Gain 1000 Gain E-08 Gain 100 Gain E E E E E E E E-01 System Pressure (torr)

10 Noise vs. Operating Pressure For Various Electron Multipliers 1000 Back-filled Air Detector Noise (cts./sec.) E E E E Operating Pressure (Torr) 5 um MCP Chevron 25 um MCP Chevron Single Channel Multiplier Discrete Dynode Multiplier Spiraltron

11 Miniature MCP Detector Noise as a Function of Vacuum Pressure ELECTRO-OPTICS Dark Current (amps) 1.00E E E E E E E E E E-02 Pressure (Torr)

12 The Effects of Poor Operating Pressure on Multiplier Lifetime

13 Lifetime at High Pressure: 100 MCP Chevron, 5 Micron Pore Assembly Back-filled Air Detector Gain (Millions) Extracted Charge (Coulombs) 1 X 10-3 Torr 5 X 10-3 Torr

14 Lifetime at High Pressure 100 MCP Chevron, 5 Micron Pore Assembly Back-filled Argon Detector Gain (Millions) Continuous Hours of Operation at 1uA Output Current 1 X 10-3 Torr Argon

15 Lifetime at High Pressure 100 MCP Chevron, 5 Micron Pore Assembly Back-filled Argon Detector Gain (Millions) Continuous Hours of Operation at 1uA Output Current 1 X 10-3 Torr Argon

16 Lifetime at High Pressure: 1000 Single MCP, 5 Micron Pore Assembly Operated at 1.2E-2 Torr In Argon 100 Gain Continuous Hours of Operation at 5E-8 amps Output Current

17 Lifetime at High Pressure: 100 MCP Chevron, 5 Micron Pore Assembly Back-filled Air Detector Gain (Millions) Continuous Hours of Operation at 1uA Output Current 1 X 10-3 Torr 5 X 10-3 Torr

18 Performance After 300 Hours: 100 Gain Measured at 10-3 Torr MCP Chevron, 5 Micron Pore Assembly Back-filled Air Detector Gain (Millions) Applied Voltage (kv)

19 Performance After 300 Hours: Detector Noise (cts./sec) Noise Measured at 10-3 Torr Model 3018MA 5 Micron Pore Assembly Back-filled Air Applied Voltage (kv)

20 Experimental Apparatus The experimental apparatus consisted of an Inficon XPR-III Residual Gas Analyzer (0-100 AMU) utilizing a microchannel plate/faraday detector mounted to a turbo pumped vacuum chamber. Chamber pressure was modulated using a Granville Philips precision leak valve connected to choice backfill gases. Sensor pressure was monitored using a combination of calibrated Veeco ionization tubes and Granville Philips convectrons.

21 Miniature Mass Spectrometer and Electron Multiplier Ionization Source and Mass Filter Faraday Cup and Electron Multiplier

22 Argon and Hydrogen at 5 milli-torr, Utilizing a Microchannel Plate Based Electron Multiplier Faraday Mode EM Mode

23 Lab Air Collected at 5.5 milli- Torr Faraday Mode EM Mode

24 Chemical Durability The performance of an electron multiplier will degrade as a result of everyday operation. Ions from the mass filter impinge on cone of the multiplier and produce a cascading of secondary electrons, ultimately resulting in the signal used to produce the mass spectrum. The constant bombardment of ions on the cone can result in the development of a surface coating which will reduce the efficiency of the secondary electron emission process. Equivalent Electron Multipliers from three manufacturers were subjected to an extended life test. The multipliers were loaded in a demountable test stand and bombarded with ions created from residual gas molecules of the species typically encountered in Residual Gas Analysis (RGA) monitored systems

25 Gain Stability of Various Multipliers for Bench top GCMS Applications 1.00E+08 BURLE MAGNUM 5900 K&M Model 7596M 1.00E+07 Detector Technology Model 2300 Gain at 1800 V 1.00E E E Extracted Charge (Coulombs)

26 Chemical Durability Multipliers from three manufacturers were tested for chemical durability in an Agilent 5971 GCMS Chemstation. The internal calibrant PFTBA was used to create ions ranging from mass 69 to mass 512. PFTBA is known to be an aggressive material which quickly degrades multiplier performance. The original instrument ion optics frame was used for this test with all multipliers. Each multiplier was subjected to repetitive test cycles consisting of a Maximum Sensitivity Auto tune, Gain and Linearity tests, then begin the cycle again. The instrument was not vented between cycles. The cycles were repeated until the instrument was no longer able to achieve the requirements of auto tune. Before each multiplier sequence began, the ion source was completely disassembled and cleaned.

27 Electron Multiplier Life Tests with PFTBA Autotune Multiplier Voltage Detector Technology Model 2300 K & M Model 7596M BURL MAGNUM 5900 Detector Technology Failed Autotune after 61 Cycles K & M Failed Autotune After 64 Cycles BURLE MAGNUM Continued to Operate beyond 119 Cycles Autotune Cycles with PFTBA

28 Conclusions Discrete dynode multipliers develop ion feedback at pressures above 10-5 Torr Single Channel Electron Multipliers Operate well at pressures into the mid 10-5 Torr Range. SPIRALTRON and MAGNUM Electron Multipliers operate well into the 10-4 Torr Range

29 A specially designed Microchannel Plate-based Electron multiplier has been successfully operated for over 1200 continuous hours at elevated pressures in excess of 10 milli-torr with good performance. A Mass Spectrometer (Inficon XPR-III) utilizing a very short (25mm) quadrupole mass filter and this microchannel plate electron multiplier has been commercialized for process monitoring in semiconductor manufacturing applications.

30 Multiplier lifetime testing, with low mass ions, for models manufactured by three companies highlighted the differences in gain performance. Multiplier lifetime testing in a bench top GCMS using the internal standard PFTBA to create ions indicated that the 6 channel MAGNUM Electron Multiplier utilizing SPIRALTRON Technology operated over 85% longer than the OEM multiplier.

The Development of Miniature Electron Multipliers for Use In Portable Mass Spectrometers

The Development of Miniature Electron Multipliers for Use In Portable Mass Spectrometers The Development of Miniature Electron Multipliers for Use In Portable Mass Spectrometers *Bruce N. Laprade, Lenny Erickson William G. Dunn and Reginald Farr BURLE Electro-Optics Sturbridge MA Paper 10400-2100

More information

The Development of a Novel Electron Multiplier with an Onboard Integral High Voltage Power Supply for use in Mass Spectrometers

The Development of a Novel Electron Multiplier with an Onboard Integral High Voltage Power Supply for use in Mass Spectrometers The Development of a Novel Electron Multiplier with an Onboard Integral High Voltage Power Supply for use in Mass Spectrometers Presented ASMS 2007 New Instrumentation Concepts Session, Poster 043 Bruce

More information

Performance Characterization Of A Simultaneous Positive and Negative Ion Detector For Mass Spectrometry Applications

Performance Characterization Of A Simultaneous Positive and Negative Ion Detector For Mass Spectrometry Applications Performance Characterization Of A Simultaneous Positive and Negative Ion Detector For Mass Spectrometry Applications Bruce Laprade and Raymond Cochran Introduction Microchannel Plates (Figures 1) are parallel

More information

Residual Gas Analyzers XT Series

Residual Gas Analyzers XT Series Residual Gas Analyzers XT Series Products from ExTorr Inc. - Pirani, Ion Gauge, Quadrupole - All Included The Extorr XT residual gas analyzer is a quadrupole mass spectrometer complete with a built-in

More information

835 Vacuum Quality Monitor (VQM TM ) 2012 Brooks Automation, Inc. Proprietary Information

835 Vacuum Quality Monitor (VQM TM ) 2012 Brooks Automation, Inc. Proprietary Information 835 Vacuum Quality Monitor (VQM TM ) 2012 Brooks Automation, Inc. Proprietary Information The Revolutionary New Vacuum Quality Monitor World s Fastest, Accurate at low mass Single Gas Calibration Small

More information

Transpector 2. Gas Analysis System. NEW-GENERATION RGAs WITH SUPERIOR SENSITIVITY AND PERFORMANCE

Transpector 2. Gas Analysis System. NEW-GENERATION RGAs WITH SUPERIOR SENSITIVITY AND PERFORMANCE Transpector 2 Gas Analysis System NEW-GENERATION RGAs WITH SUPERIOR SENSITIVITY AND PERFORMANCE We ve made the best RGAs better. Inficon, the leaders in RGA technology, have upgraded the Transpector gas

More information

Residual Gas Analyzers RGA100/200/ amu, 200 amu and 300 amu RGAs

Residual Gas Analyzers RGA100/200/ amu, 200 amu and 300 amu RGAs Residual Gas Analyzers RGA100/200/300 100 amu, 200 amu and 300 amu RGAs SRS Residual Gas Analyzers 100, 200 and 300 amu systems Better than 1 amu resolution 6 decades of dynamic range 5 10 14 Torr detection

More information

Residual Gas Analyzers

Residual Gas Analyzers Residual Gas Analyzers RGA100/200/300 100 amu, 200 amu and 300 amu systems SRS Residual Gas Analyzers 100, 200 and 300 amu systems Better than 1 amu resolution 6 decades of dynamic range 5 10 14 Torr detection

More information

Then the mass spectrometer should go to the field, in your hand, small as possible, light as possible, and operated with light battery.

Then the mass spectrometer should go to the field, in your hand, small as possible, light as possible, and operated with light battery. My name is Mo Yang from Korea Basic Science Institute. Everybody here knows that a mass spectrometer is the most powerful analytical tool in chemistry. However, the target samples you want to analyze are

More information

Autoresonant Trap Mass Spectrometry

Autoresonant Trap Mass Spectrometry 8 th Harsh Environment Mass Spectrometry Wokshop Autoresonant Trap Mass Spectrometry (ART MS) Sept 21, 2011 Gerardo A. Brucker and G. Jeffery Rathbone Granville-Phillips Vacuum Instrumentation Brooks Automation,

More information

Project by: Dr. Jorge A. Diaz. Physics School, University of Costa Rica, National Center for High Technology (CENAT)

Project by: Dr. Jorge A. Diaz. Physics School, University of Costa Rica, National Center for High Technology (CENAT) Test of a Miniature Double-Focusing Mass Spectrometer for the Variable Specific Impulse Magnetoplasma Rocket (VASIMR) at the Advanced Space Propulsion Laboratory Project by: Dr. Jorge A. Diaz Physics School,

More information

HPR-30 Vacuum Process Gas Analyser A differentially pumped RGA system for vacuum process monitoring

HPR-30 Vacuum Process Gas Analyser A differentially pumped RGA system for vacuum process monitoring HPR-30 Vacuum Process Gas Analyser A differentially pumped RGA system for vacuum process monitoring HPR-30 vacuum process gas analyser To analyse processes with high dynamic range operating at pressures

More information

830 Vacuum Quality Monitor

830 Vacuum Quality Monitor Granville-Phillips Vacuum Instrumentation 830 Vacuum Quality Monitor Featuring ART MS Technology February 2011 1 What is ART MS? ART MS Anharmonic Resonant Trap Mass Spectrometry or Auto Resonant Ion Trap

More information

HIQUAD. New high-end mass spectrometer! Fast, flexible and easy to operate.

HIQUAD. New high-end mass spectrometer! Fast, flexible and easy to operate. New high-end mass spectrometer! Fast, flexible and easy to operate. New high-end mass spectrometer! Fast, flexible and easy to operate. A modular solution for mass spectrometry With the new HiQuad mass

More information

PRISMAPLUS. The precise solution for mass spectrometry. Modular design. Powerful software. Wide range of applications.

PRISMAPLUS. The precise solution for mass spectrometry. Modular design. Powerful software. Wide range of applications. PRISMAPLUS The precise solution for mass spectrometry. Modular design. Powerful software. Wide range of applications. PRISMAPLUS The precise solution for mass spectrometry. The combination of high sensi

More information

PrismaPlus. The New Mass Spectrometer with the Added Plus! Modular Design. Powerful Software. Wide Range of Applications.

PrismaPlus. The New Mass Spectrometer with the Added Plus! Modular Design. Powerful Software. Wide Range of Applications. PrismaPlus The New Mass Spectrometer with the Added Plus! Modular Design. Powerful Software. Wide Range of Applications. PrismaPlus The PrismaPlus innovation The optimum combination of high sensitivity,

More information

Performance characteristics of a new wide range, fast settling electrometer design for a residual gas analysis mass spectrometer

Performance characteristics of a new wide range, fast settling electrometer design for a residual gas analysis mass spectrometer Performance characteristics of a new wide range, fast settling electrometer design for a residual gas analysis mass spectrometer MKS Spectra Products, January 2010 Design considerations for RGA components

More information

Improving Signal-to-Noise by Identifying Sources of Noise in Mass Spectrometer Systems

Improving Signal-to-Noise by Identifying Sources of Noise in Mass Spectrometer Systems GT-73A Improving Signal-to-Noise by Identifying Sources of Noise in Mass Spectrometer Systems Introduction: Kevin Kuchta Extrel CMS 575 Epsilon Drive, Pittsburgh, PA 538 (Poster presented at the 50th ASMS

More information

Electron Multipliers. Introduction to ETP Electron Multipliers Selection by Technique: LC-MS 242

Electron Multipliers. Introduction to ETP Electron Multipliers Selection by Technique: LC-MS 242 Introduction to ETP 240-241 Selection by Technique: GC-MS 242 LC-MS 242 ICP-MS 242 TOF-MS 243 Magnetic Sector 243 Selection by Instrument 243-244 ETP are individually tested to ensure the highest quality.

More information

Development of a Small Residual Gas Analyzer Utilizing the Quadrupole Array Structure Micropole System ~ QL Series ~

Development of a Small Residual Gas Analyzer Utilizing the Quadrupole Array Structure Micropole System ~ QL Series ~ F e a t u r e A r t i c l e Feature Article Development of a Small Residual Gas Analyzer Utilizing the Quadrupole Array Structure Micropole System ~ QL Series ~ Hirokazu Kitaura The Micropole System is

More information

Performance of Microchannel Plates Fabricated Using Atomic Layer Deposition

Performance of Microchannel Plates Fabricated Using Atomic Layer Deposition Performance of Microchannel Plates Fabricated Using Atomic Layer Deposition Andrey Elagin on behalf of the LAPPD collaboration Introduction Performance (timing) Conclusions Large Area Picosecond Photo

More information

Introducing the Agilent 7000A Triple Quadrupole GC/MS

Introducing the Agilent 7000A Triple Quadrupole GC/MS Introducing the Agilent 7000A Triple Quadrupole GC/MS Paul Zavitsanos, Joe Weitzel, Monty Benefiel and Terry Sheehan May 2008 Page 1 What s a Triple Quad? Page 2 GC/MS Triple Quad (QQQ) Collision Gas (N2

More information

Detectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014

Detectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014 Detectors for microscopy - CCDs, APDs and PMTs Antonia Göhler Nov 2014 Detectors/Sensors in general are devices that detect events or changes in quantities (intensities) and provide a corresponding output,

More information

The Extrel MAX Systems are UHV-compatible flange mounted Quadrupole Mass

The Extrel MAX Systems are UHV-compatible flange mounted Quadrupole Mass The Extrel MAX Systems are UHV-compatible flange mounted Quadrupole Mass Spectrometers. The Ionizer, Quadrupole Mass Filter, and Detector on a Mounting Flange are designed for inclusion in your experimental

More information

Ion optics innovations for increased sensitivity in hybrid MS systems

Ion optics innovations for increased sensitivity in hybrid MS systems Ion optics innovations for increased sensitivity in hybrid MS systems Abstract Sensitivity, mass resolution, scan speed, spectral fidelity, mass accuracy, and mass range are just a few of the parameters

More information

The 5975C Series MSDs: Normalized Instrument Tuning. Technical Overview. Introduction

The 5975C Series MSDs: Normalized Instrument Tuning. Technical Overview. Introduction The 5975C Series MSDs: Normalized Instrument Tuning Technical Overview Jeffrey T. Kernan and Harry Prest Introduction The automated optimization of ion detection in a mass spectrometer, commonly referred

More information

Sub-mm Linear Ion Trap Mass Spectrometer Made Using Lithographically Patterned Ceramic Plates

Sub-mm Linear Ion Trap Mass Spectrometer Made Using Lithographically Patterned Ceramic Plates Sub-mm Linear Ion Trap Mass Spectrometer Made Using Lithographically Patterned Ceramic Plates Ailin Li Brigham Young University, Provo, UT Coauthors: Qinghao Wu, Yuan Tian, Derek Andrews, Aaron Hawkins,

More information

Technical Specifications for the procurement of Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Technical Specifications for the procurement of Inductively Coupled Plasma Mass Spectrometer (ICP-MS) Technical Specifications for the procurement of Inductively Coupled Plasma Mass Spectrometer (ICP-MS) Preamble An Inductively Coupled Plasma Mass Spectrometer (Quadrupole ICP-MS) is proposed to be procured

More information

MCPs and MCP based detectors. Raquel Ortega

MCPs and MCP based detectors. Raquel Ortega MCPs and MCP based detectors Raquel Ortega r.ortega@photonisusa.com Outline Introduction MCP characteristics: Gain Detection efficiency Time resolution Spatial resolution Detecting different particles

More information

PRISMAPRO. The latest generation of compact mass spectrometers. Modular design. Powerful software. Low detection limit.

PRISMAPRO. The latest generation of compact mass spectrometers. Modular design. Powerful software. Low detection limit. The latest generation of compact mass spectrometers. Modular design. Powerful software. Low detection limit. The latest generation of compact mass spectrometers. The combination of high sensi tivity, maximum

More information

ICP-MS. plasma 3. Multi-Collector ICP-MS.

ICP-MS. plasma 3. Multi-Collector ICP-MS. ICP-MS plasma 3 Multi-Collector ICP-MS www.nu-ins.com plasma 3 is a third generation Multi Collector ICP Mass Spectrometer (MC-ICP-MS), designed to provide the best possible precision and accuracy for

More information

DN 16 ISO-KF. Dimensional drawings THERMOVAC TM 101 (mm)

DN 16 ISO-KF. Dimensional drawings THERMOVAC TM 101 (mm) Products Handheld Measuring Instruments Digital PIEZOVAC Sensor PV 101 Digital THERMOVAC Sensor TM 101 The digital compact PIEZOVAC PV 101 and THERMOVAC TM 101 combine high-quality sensor technology with

More information

CRL MASS SPECTROMETRY FACILITY INSTRUMENT USER MANUAL MALDI MICRO. Operating Instructions. Basement Spec Lab:

CRL MASS SPECTROMETRY FACILITY INSTRUMENT USER MANUAL MALDI MICRO. Operating Instructions. Basement Spec Lab: CRL MASS SPECTROMETRY FACILITY INSTRUMENT USER MANUAL MALDI MICRO Operating Instructions Basement Spec Lab: 00.097 crs Page of 2 This is a guide to using the MALDI Micro for those who have received training.

More information

Cylindrical Toroidal Ion Trap Mass Spectrometer. Daniel Austin and Nick Taylor Brigham Young University, Provo, Utah

Cylindrical Toroidal Ion Trap Mass Spectrometer. Daniel Austin and Nick Taylor Brigham Young University, Provo, Utah Cylindrical Toroidal Ion Trap Mass Spectrometer Daniel Austin and Nick Taylor Brigham Young University, Provo, Utah Why miniaturize ion traps Geometrically compact Higher tolerance to pressure Amenable

More information

Enable Highly-Stable Plasma Operations at High Pressures with the Right RPS Solution

Enable Highly-Stable Plasma Operations at High Pressures with the Right RPS Solution Enable Highly-Stable Plasma Operations at High Pressures with the Right RPS Solution Created by Advanced Energy Industries, Inc., Fort Collins, CO Abstract Conventional applications for remote plasma sources

More information

Evaluation of high power laser diodes for space applications: effects of the gaseous environment

Evaluation of high power laser diodes for space applications: effects of the gaseous environment Evaluation of high power laser diodes for space applications: effects of the gaseous environment Jorge Piris, E. M. Murphy, B. Sarti European Space Agency, Optoelectronics section, ESTEC. M. Levi, G. Klumel,

More information

Micro Analytical Instruments - A System Approach. Jörg Müller Institut für Mikrosystemtechnik

Micro Analytical Instruments - A System Approach. Jörg Müller Institut für Mikrosystemtechnik Micro Analytical Instruments - A System Approach Jörg Müller Institut für Mikrosystemtechnik Technische h Universität i Hamburg-Harburg H b Outline Motivation Political Political Issues Design Principles

More information

New Control Electronics for Helmer Gauges

New Control Electronics for Helmer Gauges New Control Electronics for Helmer Gauges Nikolaos Chatzigeorgiou TE-VSC-ICM Technology Department ICM Automation FORUM, 24th May 2016 Agenda The Helmer gauge Control electronics Supervision application

More information

A New Class of Robust Sub-nanosecond TOF Detectors with High Dynamic Range

A New Class of Robust Sub-nanosecond TOF Detectors with High Dynamic Range A New Class of Robust Sub-nanosecond TOF Detectors with High Dynamic Range Dick Stresau, Kevin Hunter, Wayne Shiels, Peter Raffin and Yair Benari ETP, Sydney, Australia Presented at the 54th ASMS Conference

More information

Mass spectrometers have evolved into standard

Mass spectrometers have evolved into standard FOCUS: FIELD-PORTABLE AND MINIATURE MS Trade-offs in Miniature Quadrupole Designs S. Boumsellek and R. J. Ferran Ferran Scientific Inc., San Diego, California, USA Pressing needs for miniature mass spectrometers

More information

The Perfect Solution for Mass Spectrometry. Modular Design. Powerful Software. Wide Range of Applications.

The Perfect Solution for Mass Spectrometry. Modular Design. Powerful Software. Wide Range of Applications. A P A S S I O N F O R P E R F E C T I O N PrismaPlus The Perfect Solution for Mass Spectrometry. Modular Design. Powerful Software. Wide Range of Applications. PrismaPlus The Perfect Solution for Mass

More information

Mass Spectrometry and the Modern Digitizer

Mass Spectrometry and the Modern Digitizer Mass Spectrometry and the Modern Digitizer The scientific field of Mass Spectrometry (MS) has been under constant research and development for over a hundred years, ever since scientists discovered that

More information

Agilent GC-MS Maintenance:

Agilent GC-MS Maintenance: Restek GC Accessories Agilent GC-MS Maintenance: Restek s Quick Reference Guide Replacement parts for Agilent ion sources now available. Optimize performance by upgrading ion source components and electron

More information

Collision cell: Fragmentation efficiency > 90 % Duty cycle: Variable between 1 and 50 % (user selectable trapping time/scan time ratio) Transmission:

Collision cell: Fragmentation efficiency > 90 % Duty cycle: Variable between 1 and 50 % (user selectable trapping time/scan time ratio) Transmission: Finnigan MAT 95 XP-API-GC-Trap tandem Mass spectrometer system Tandem Mass Spectrometer based on high resolution magnetic sector and Ion-Trap MS/MS system with API ( ESI, APCI ) and EI/CI ionization Mass

More information

Physics Requirements for the CXI 0.1 micron Sample Chamber

Physics Requirements for the CXI 0.1 micron Sample Chamber PHYSICS REQUIREMENT DOCUMENT (PRD) Doc. No. SP-391-000-20 R1 LUSI SUB-SYSTEM Coherent X-Ray Imaging Physics Requirements for the Sébastien Boutet CXI Scientist, Author Signature Date Paul Montanez CXI

More information

OPTI510R: Photonics. Khanh Kieu College of Optical Sciences, University of Arizona Meinel building R.626

OPTI510R: Photonics. Khanh Kieu College of Optical Sciences, University of Arizona Meinel building R.626 OPTI510R: Photonics Khanh Kieu College of Optical Sciences, University of Arizona kkieu@optics.arizona.edu Meinel building R.626 Photodetectors Introduction Most important characteristics Photodetector

More information

TRITON Plus Thermal Ionization MS. Another step ahead. in TIMS. Earth and Planetary Sciences Nuclear Sciences

TRITON Plus Thermal Ionization MS. Another step ahead. in TIMS. Earth and Planetary Sciences Nuclear Sciences TRITON Plus Thermal Ionization MS Another step ahead in TIMS Earth and Planetary Sciences Nuclear Sciences Based on more than 35 years of experience in variable multicollector instrumentation, we have

More information

Lecture 12 OPTICAL DETECTORS

Lecture 12 OPTICAL DETECTORS Lecture 12 OPTICL DETECTOS (eference: Optical Electronics in Modern Communications,. Yariv, Oxford, 1977, Ch. 11.) Photomultiplier Tube (PMT) Highly sensitive detector for light from near infrared ultraviolet

More information

The Absolute Detection Efficiency of Vacuum Electron Multipliers, to kev Protons and Ar + Ions.

The Absolute Detection Efficiency of Vacuum Electron Multipliers, to kev Protons and Ar + Ions. The Absolute Detection Efficiency of Vacuum Electron Multipliers, to kev Protons and Ar + Ions. S. Shchemelinin 1, S. Pszona 2, G. Garty 1,, A. Breskin 1, R. Chechik 1 1 Department of Particle Physics,

More information

A Sort-of Tissue Equivalent Proportional Counter (STEPC) for Space Radiation Dosimetry Applications

A Sort-of Tissue Equivalent Proportional Counter (STEPC) for Space Radiation Dosimetry Applications A Sort-of Tissue Equivalent Proportional Counter (STEPC) for Space Radiation Dosimetry Applications Eric Benton, Tyler Collums, and Art Lucas E. V. Benton Radiation Physics Laboratory Oklahoma State University

More information

Partial Replication of Storms/Scanlan Glow Discharge Radiation

Partial Replication of Storms/Scanlan Glow Discharge Radiation Partial Replication of Storms/Scanlan Glow Discharge Radiation Rick Cantwell and Matt McConnell Coolescence, LLC March 2008 Introduction The Storms/Scanlan paper 1 presented at the 8 th international workshop

More information

AccuTOF LC Training Course

AccuTOF LC Training Course 1 AccuTOF LC Training Course JEOL USA, Inc. 11 Dearborn Road Peabody, MA 01960 Area map Restaurants: Wendy s Bennigan s Bertucci s 2 AccuTOF LC Training Course Monday Basic principle and history of TOF

More information

Advantages of Multi-Turn-TOF Technology

Advantages of Multi-Turn-TOF Technology Advantages of Multi-Turn-TOF Technology Ordinary TOF vs. Multi-Turn TOF Ordinary TOF) TOF-MS is one of the most popular technologies for high resolution applications, particularly in ESI and MALDI instruments.

More information

LC/MS/MS. Page Header. triple quadrupole mass spectrometer.

LC/MS/MS. Page Header. triple quadrupole mass spectrometer. LC/MS/MS VARIAN, INC. 320-MS Page Header triple quadrupole mass spectrometer www.varianinc.com VARIAN, INC. 320-MS Unsurpassed commitment to innovation Varian, Inc. is an innovator and leader in mass spectrometry

More information

A Residual Gas Analyzer for Dry Etching Process

A Residual Gas Analyzer for Dry Etching Process FFeature Article Article Makoto MATSUHAMA Concerning the dry process of the semiconductor device manufacturing, the monitoring of etching chamber conditions (pressure, temperature, gas concentration,...)

More information

High Field Q-Slope in Superconducting RF Cavities

High Field Q-Slope in Superconducting RF Cavities High Field Q-Slope in Superconducting RF Cavities Jordan Webster Advisor: Matthias Liepe August 7, 2008 High Field Q-Slope in Superconducting RF Cavities A Tragic Experimental Tale Jordan Webster Advisor:

More information

Low-cost sensors and high-end instruments for mapping urban NO x pollution

Low-cost sensors and high-end instruments for mapping urban NO x pollution Low-cost sensors and high-end instruments for mapping urban NO x pollution Michael Müller, Morten Hundt, Lukas Emmenegger, Christoph Hüglin, Béla Tuzson, Christoph Zellweger and many other from the Empa,

More information

The Best Way to Get the Right Answers... is in a Flash

The Best Way to Get the Right Answers... is in a Flash FlashQuant Workstation The Best Way to Get the Right Answers... is in a Flash FlashQuant Workstation: Integrating MALDI Speed with Triple Quad Selectivity 5.03E+02 0.011 1.40E+03 2.680 5.26E+02 0.011 1.30E+03

More information

NEPTUNE Plus Multicollector ICPMS. Another step ahead. in MC-ICP-MS. Earth and Planetary Sciences Nuclear Sciences

NEPTUNE Plus Multicollector ICPMS. Another step ahead. in MC-ICP-MS. Earth and Planetary Sciences Nuclear Sciences NEPTNE Plus Multicollector ICPMS Another step ahead in MC-ICP-MS Earth and Planetary Sciences Nuclear Sciences Based on more than 35 years of experience in variable multicollector instrumentation, we have

More information

TABLE OF CONTENTS PAGE LIST OF FIGURES 4 PREFACE 5. CHAPTER 1 6 Introduction. CHAPTER 2 9 Fabrication and Operation. CHAPTER 3 19 Applications

TABLE OF CONTENTS PAGE LIST OF FIGURES 4 PREFACE 5. CHAPTER 1 6 Introduction. CHAPTER 2 9 Fabrication and Operation. CHAPTER 3 19 Applications CHANNELTRON ELECTRON MULTIPLIER HANDBOOK FOR MASS SPECTROMETRY APPLICATIONS TABLE OF CONTENTS PAGE LIST OF FIGURES 4 PREFACE 5 CHAPTER 1 6 Introduction CHAPTER 2 9 Fabrication and Operation CHAPTER 3 19

More information

Agilent 6400 Series Triple Quadrupole LC/MS System

Agilent 6400 Series Triple Quadrupole LC/MS System Agilent 6400 Series Triple Quadrupole LC/MS System Concepts Guide The Big Picture Agilent Technologies Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by

More information

Laser-Produced Sn-plasma for Highvolume Manufacturing EUV Lithography

Laser-Produced Sn-plasma for Highvolume Manufacturing EUV Lithography Panel discussion Laser-Produced Sn-plasma for Highvolume Manufacturing EUV Lithography Akira Endo * Extreme Ultraviolet Lithography System Development Association Gigaphoton Inc * 2008 EUVL Workshop 11

More information

Relative Quantum Efficiency Measurements of the ROSS Streak Camera Photocathode. Alex Grammar

Relative Quantum Efficiency Measurements of the ROSS Streak Camera Photocathode. Alex Grammar Relative Quantum Efficiency Measurements of the ROSS Streak Camera Photocathode Alex Grammar Relative Quantum Efficiency Measurements of the ROSS Streak Camera Photocathode Alex Grammar Advised by Dr.

More information

Dr. James McFadden, UC Berkeley

Dr. James McFadden, UC Berkeley Plasma Instrument Design: Outline Dr. James McFadden, UC Berkeley Outline: Particle detectors Analog Signal Processing Analyzers How to Design an instrument Calibration Issues Example Instruments James

More information

Our Company. We are major supplier to:

Our Company. We are major supplier to: VACUUM METROLOGY Our Company Hositrad Holland and Hositrad Deutschland combine more than 50 years of experience in vacuum and cryogenic technology. Customers profit from a broad range of capabilities,

More information

Herwig Schopper CERN 1211 Geneva 23, Switzerland. Introduction

Herwig Schopper CERN 1211 Geneva 23, Switzerland. Introduction THE LEP PROJECT - STATUS REPORT Herwig Schopper CERN 1211 Geneva 23, Switzerland Introduction LEP is an e + e - collider ring designed and optimized for 2 100 GeV. In an initial phase an energy of 2 55

More information

[NIGHT VISION TECHNOLOGY] SEMINAR REPORT

[NIGHT VISION TECHNOLOGY] SEMINAR REPORT 20 th JANUARY 2010 Night Vision Technology Introduction Night vision technology, by definition, literally allows one to see in the dark. Originally developed for military use. Federal and state agencies

More information

Historical. McPherson 15 Mount

Historical. McPherson 15 Mount McPherson 15 Mount Normal incidence designs include the McPherson 15 (classical 1.0 meter focal length) and modern NIM units. The latter features smaller included angles, longer focal lengths (e.g. 3,

More information

Quantum frequency standard Priority: Filing: Grant: Publication: Description

Quantum frequency standard Priority: Filing: Grant: Publication: Description C Quantum frequency standard Inventors: A.K.Dmitriev, M.G.Gurov, S.M.Kobtsev, A.V.Ivanenko. Priority: 2010-01-11 Filing: 2010-01-11 Grant: 2011-08-10 Publication: 2011-08-10 Description The present invention

More information

MuCool Test Area Experimental Program Summary

MuCool Test Area Experimental Program Summary MuCool Test Area Experimental Program Summary Alexey Kochemirovskiy The University of Chicago/Fermilab Alexey Kochemirovskiy NuFact'16 (Quy Nhon, August 21-27, 2016) Outline Introduction Motivation MTA

More information

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Takeshi Fujiwara 1, Yuki Mitsuya 2, Hiroyuki Takahashi 2, and Hiroyuki Toyokawa 2 1 National Institute of Advanced Industrial

More information

Progress towards Actinic Patterned Mask Inspection. Oleg Khodykin

Progress towards Actinic Patterned Mask Inspection. Oleg Khodykin Progress towards Actinic Patterned Mask Inspection Oleg Khodykin Outline Status (technical) of EUV Actinic Reticle Inspection program Xe based LPP source as bright and reliable solution Requirements Choice

More information

Ultraviolet Visible Infrared Instrumentation

Ultraviolet Visible Infrared Instrumentation Ultraviolet Visible Infrared Instrumentation Focus our attention on measurements in the UV-vis region of the EM spectrum Good instrumentation available Very widely used techniques Longstanding and proven

More information

PB T/R Two-Channel Portable Frequency Domain Terahertz Spectrometer

PB T/R Two-Channel Portable Frequency Domain Terahertz Spectrometer PB7220-2000-T/R Two-Channel Portable Frequency DATASHEET MA 2015 Compact, Portable Terahertz Spectroscopy System Bakman Technologies versatile PB7220-2000-T/R Spectroscopy Platform is designed for scanning

More information

A Low-Cost Approach to Teaching Transmission Line Fundamentals and Impedance Matching

A Low-Cost Approach to Teaching Transmission Line Fundamentals and Impedance Matching A Low-Cost Approach to Teaching Transmission Line Fundamentals and Impedance Matching David M. Hata Portland Community College Abstract: As part of a NSF-funded Project, Portland Community College has

More information

Simulations Guided Efforts to Understand MCP Performance

Simulations Guided Efforts to Understand MCP Performance University of Chicago Simulations Guided Efforts to Understand MCP Performance M. Wetstein, B. Adams, M. Chollet, A. Elagin, A. Vostrikov, R. Obaid, B. Hayhurst V. Ivanov, Z. Insepov, Q. Peng, A. Mane,

More information

Virtual Laboratory of Nuclear Fission Virtual practicum in the framework of the project Virtual Laboratory of Nuclear Fission

Virtual Laboratory of Nuclear Fission Virtual practicum in the framework of the project Virtual Laboratory of Nuclear Fission Virtual Laboratory of Nuclear Fission Virtual practicum in the framework of the project Virtual Laboratory of Nuclear Fission Khanyisa Sowazi, University of the Western Cape JINR SAR, September 2015 INDEX

More information

SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS

SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS Robert Edward Lee Electron Microscopy Center Department of Anatomy and Neurobiology Colorado State University P T R Prentice Hall, Englewood Cliffs,

More information

Seapoint Rhodamine Fluorometer

Seapoint Rhodamine Fluorometer Seapoint Rhodamine Fluorometer User Manual Standard Version 6/00 Dimensions Open Configuration (no pump) 2.5" (6.4 cm) 6.6" (16.8 cm) 5.1" (13.0 cm) Pumped Configuration Figure 1. 1/2 Scale Drawing. Page

More information

UV GAS LASERS PREPARED BY: STUDENT NO: COURSE NO: EEE 6503 COURSE TITLE: LASER THEORY

UV GAS LASERS PREPARED BY: STUDENT NO: COURSE NO: EEE 6503 COURSE TITLE: LASER THEORY UV GAS LASERS PREPARED BY: ISMAIL HOSSAIN FARHAD STUDENT NO: 0411062241 COURSE NO: EEE 6503 COURSE TITLE: LASER THEORY Introduction The most important ultraviolet lasers are the nitrogen laser and the

More information

Production of HPDs for the LHCb RICH Detectors

Production of HPDs for the LHCb RICH Detectors Production of HPDs for the LHCb RICH Detectors LHCb RICH Detectors Hybrid Photon Detector Production Photo Detector Test Facilities Test Results Conclusions IEEE Nuclear Science Symposium Wyndham, 24 th

More information

Status of the PRad Experiment (E )

Status of the PRad Experiment (E ) Status of the PRad Experiment (E12-11-106) NC A&T State University Outline Experimental apparatus, current status Installation plan Draft run plan Summary PRad Experimental Setup Main detectors and elements:

More information

Pulse Niru Company. General Catalogue.

Pulse Niru Company. General Catalogue. Pulse Niru Company General Catalogue www.pulseniru.com Pulse Niru Company initiated its activities since 2003 in manufacturing Pulsed Power equipment such as High Energy Pulse Discharge Capacitors for

More information

PB T/R Two-Channel Portable Frequency Domain Terahertz Spectrometer

PB T/R Two-Channel Portable Frequency Domain Terahertz Spectrometer Compact, Portable Terahertz Spectroscopy System Bakman Technologies versatile PB7220-2000-T/R Spectroscopy Platform is designed for scanning complex compounds to precise specifications with greater accuracy

More information

Miniature Spectrometer Technical specifications

Miniature Spectrometer Technical specifications Miniature Spectrometer Technical specifications Ref: MSP-ISI-TEC 001-02 Date: 2017-05-05 Contact Details Correspondence Address: Email: Phone: IS-Instruments Ltd. Pipers Business Centre 220 Vale Road Tonbridge

More information

DEVELOPMENT OF THE SPACE ACTIVE HYDROGEN MASER FOR THE ACES MISSION

DEVELOPMENT OF THE SPACE ACTIVE HYDROGEN MASER FOR THE ACES MISSION DEVELOPMENT OF THE SPACE ACTIVE HYDROGEN MASER FOR THE ACES MISSION D. Goujon (1), P. Rochat (1), P. Mosset (1), D. Boving (1), A. Perri (1), J. Rochat (1), N. Ramanan (1), D. Simonet (1), X. Vernez (1),

More information

Working Towards Large Area, Picosecond-Level Photodetectors

Working Towards Large Area, Picosecond-Level Photodetectors Working Towards Large Area, Picosecond-Level Photodetectors Matthew Wetstein - Enrico Fermi Institute, University of Chicago HEP Division, Argonne National Lab Introduction: What If? Large Water-Cherenkov

More information

PULSED BREAKDOWN CHARACTERISTICS OF HELIUM IN PARTIAL VACUUM IN KHZ RANGE

PULSED BREAKDOWN CHARACTERISTICS OF HELIUM IN PARTIAL VACUUM IN KHZ RANGE PULSED BREAKDOWN CHARACTERISTICS OF HELIUM IN PARTIAL VACUUM IN KHZ RANGE K. Koppisetty ξ, H. Kirkici Auburn University, Auburn, Auburn, AL, USA D. L. Schweickart Air Force Research Laboratory, Wright

More information

Characterizing Single Electron Radiation in an Intense Laser Field. Caleb Coburn

Characterizing Single Electron Radiation in an Intense Laser Field. Caleb Coburn Characterizing Single Electron Radiation in an Intense Laser Field Caleb Coburn A senior thesis submitted to the faculty of Brigham Young University in partial fulfillment of the requirements for the degree

More information

Compact High Intensity Light Source

Compact High Intensity Light Source Compact High Intensity Light Source General When a broadband light source in the ultraviolet-visible-near infrared portion of the spectrum is required, an arc lamp has no peer. The intensity of an arc

More information

Scanning electron microscope

Scanning electron microscope Scanning electron microscope 5 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Detectors

More information

SPRAY DROPLET SIZE MEASUREMENT

SPRAY DROPLET SIZE MEASUREMENT SPRAY DROPLET SIZE MEASUREMENT In this study, the PDA was used to characterize diesel and different blends of palm biofuel spray. The PDA is state of the art apparatus that needs no calibration. It is

More information

Applications of Cold Cathode PIG Ion Source in Lithography

Applications of Cold Cathode PIG Ion Source in Lithography Applications of Cold Cathode PIG Ion Source in Lithography N. I. Bassal Accelerator Department, Nuclear research Center, AE A, Cairo, Egypt. Email/nagatbassal@yahoo.com Received: 7/12/2011 Accepted: 17/10/2012

More information

Analytical Chemistry II

Analytical Chemistry II Analytical Chemistry II L3: Signal processing (selected slides) Semiconductor devices Apart from resistors and capacitors, electronic circuits often contain nonlinear devices: transistors and diodes. The

More information

Electronic Measurements & Instrumentation. 1. Draw the Maxwell s Bridge Circuit and derives the expression for the unknown element at balance?

Electronic Measurements & Instrumentation. 1. Draw the Maxwell s Bridge Circuit and derives the expression for the unknown element at balance? UNIT -6 1. Draw the Maxwell s Bridge Circuit and derives the expression for the unknown element at balance? Ans: Maxwell's bridge, shown in Fig. 1.1, measures an unknown inductance in of standard arm offers

More information

Study of Resistive Plate Chambers Scintillation detectors and Reconstruction of muon tracks in RPCs

Study of Resistive Plate Chambers Scintillation detectors and Reconstruction of muon tracks in RPCs Study of Resistive Plate Chambers Scintillation detectors and Reconstruction of muon tracks in RPCs Medha Soni VSRP Student 2012 Under the guidance of Prof. Sudeshna Banerjee What are we trying to do?

More information

A comparison of two high performance inverse photoemission bandpass detectors. Abstract

A comparison of two high performance inverse photoemission bandpass detectors. Abstract A comparison of two high performance inverse photoemission bandpass detectors I. G. Hill and A. B. McLean Department of Physics, Queen s University, Kingston, Ontario, Canada, K7L 3N6 (Dated: October 27,

More information

ENGINEERING CHANGE ORDER ECO No. COS-057 Center for Astrophysics & Space Astronomy Date 13 February 2001 University of Colorado, Boulder Sheet 1 of 6

ENGINEERING CHANGE ORDER ECO No. COS-057 Center for Astrophysics & Space Astronomy Date 13 February 2001 University of Colorado, Boulder Sheet 1 of 6 University of Colorado, Boulder Sheet 1 of 6 Description of Change: 1. Replace Table 5.3-2 in Section 5.3.2.1 with the following updated table, which includes a parameter called BFACTOR that is used in

More information

Thermal Conductivity Sensor for Leak or Pressure Detection MTCS2601. MTCS2601 silicon sensing die in SMD ceramic package

Thermal Conductivity Sensor for Leak or Pressure Detection MTCS2601. MTCS2601 silicon sensing die in SMD ceramic package Sensor Description Thermal Conductivity Sensor for Leak or Pressure Detection MTCS2601 MTCS2601 silicon sensing die in SMD ceramic package Thermal conductivity sensor for primary vacuum measurement Silicon

More information

A Thermal Compensation System for the gravitational wave detector Virgo

A Thermal Compensation System for the gravitational wave detector Virgo A Thermal Compensation System for the gravitational wave detector Virgo M. Di Paolo Emilio University of L Aquila and INFN Roma Tor Vergata On behalf of the Virgo Collaboration Index: 1) Thermal Lensing

More information