Performance characteristics of a new wide range, fast settling electrometer design for a residual gas analysis mass spectrometer

Size: px
Start display at page:

Download "Performance characteristics of a new wide range, fast settling electrometer design for a residual gas analysis mass spectrometer"

Transcription

1 Performance characteristics of a new wide range, fast settling electrometer design for a residual gas analysis mass spectrometer MKS Spectra Products, January 2010

2 Design considerations for RGA components The maximum linear outputs from either a micro-channel plate or single channel electron multiplier are at the A level, typically requiring a moderate value resistor in the feedback circuit of the electrometer amplifier (in the MKS case 50M ) The noise across this 50M resistor would be ~1e -14 A RMS for a 1 second integration leading to ~e7 dynamic range of measurement Typically, signal measurements tend to be required at faster than 1Hz so it is desirable to have lower noise in the measurement. This can typically be achieved by using a higher resistor value in the electrometer amplifier feedback circuit A 20G resistor would have a 4e-16 A noise RMS for a 1 second integration, however the maximum measurable signal would then only be 0.5 na before the amplifier saturated The traditional approach to signal measurement has been to use gain switching either by switching between scans or within a scan on a channel by channel basis The setting times required to drain saturated circuits these very low current levels when switching is typically not published by the amplifier manufacturers because they are very long (hundreds of ms) For these very low currents a logarithmic amplifier circuit is not feasible due to the poor drift characteristics and the capacitive effects of typical devices used Traditional dual range electrometer, requires either: Two separate scans to cover the complete dynamic range Or long settling time to discharge circuit as relays switch the resistor values

3 MKS s approach to an electrometer design The signal from the detector is available to two amplifiers, one with a 50M feedback resistor and one with a 20G feedback resistor Each amplifier is sampled simultaneously by two ADCs A a fast acting diode circuit ensures that detector currents below a trigger level are available to the high feedback resistor amplifier Above the trigger level the diode proportions the signal between both the amplifiers where most of the signal is now measured by the low gain amplifier due to it s much lower resistor value The combination of the two ADC outputs is now the detector signal Due to the proportioning between the amplifiers, neither amplifier reaches saturation on the maximum signal from the detector Below the trigger threshold, the software is programmed to ignore the low gain amplifier (and therefore the noise generated from the 50M resistor) After the signal measurement a charge injection to the diode counters any residual charge resetting it before the next measurement

4 Microvision 2 proprietary technology to reduce settling time from large peaks 5.0 Graph showing detector settling time when jumping to a mass with no signal from different masses within the measurement range of the detector Starting signal 1E7 times higher than noise floor 4.0 Starting ignal 1E6 times higher than noise floor Starting signal 1E5 times higher than noise floor Part s per m illion of full range signal Starting signal 1E4 times higher than noise floor Statistical detection limit at specific measurement times Microvision 2 can jump from the highest peaks to the lowest without any danger of false positive measurements as the settled data for a no signal peak will always be below the theoretical detection limits Measurement starts after ms for smallest signals Settling and completed measurement time, milliseconds Signals close to the noise floor can still be reliably measured, even immediately after measuring very large signals

5 Application example 1: Transfer chamber RGA data quality Most transfer chamber slit valves are open a few seconds, during which time the partial pressure differences between chambers will balance The initial immediate change when the valve opens and the equalised partial pressures give key indicators of leaking mechanisms and pumping performances The detector must be calibrated so the highest partial pressure pulse is still less than the maximum signal so all events can be captured In the following data the maximum signal for Microvision Plus is 1.4E-5mbar but the maximum signal for Microvision 2 is >5e-4mbar Often more than one slit valve can be open at any one time or one valve can open within <1 second of another Adequate time resolution of data is required to correctly isolate a faulty component to save costly diagnostic time during which the tool is down Failures are often air and water leaks but process gas and wafer contaminant issues can also be seen A balance is required, more gases monitored = less time resolution Standard TOOLweb RGA conditions are to use a 1 to 70 mass barchart with a scan time of ~1.7 seconds

6 XFR chamber monitoring with 1 to 70 Barchart A B C Clipped signal on traditional RGA Model&Settings Mass 1 Mass 2 Mass 4 Mass 18 Mass 20 Mass 28 Mass 32 Mass 36 Mass 38 Mass 40 Mass 44 Mass 68 Combined mass MV+ (Acc3) Plateau Signal -4.43E E E E E E E E E E E E s %RSD % 1.28% 4.50% 1.65% 1.18% 2.14% 5.91% 2.32% 4.04% 0.72% 12.01% 28.89% (Trace A) BKG 5.557E E E E E E E E E E E E E-11 LOD 2.864E E E E E E E E E E E E E-10 MV2 (Acc3) Plateau Signal 5.245E E E E E E E E E E E E s %RSD 5.04% 1.98% 5.04% 1.15% 0.78% 1.52% 2.96% 1.17% 3.02% 0.47% 8.14% 16.16% (Trace B) BKG 4.028E E E E E E E E E E E E E-12 LOD 5.554E E E E E E E E E E E E E-11 MV2 (Acc1) Plateau Signal 5.463E E E E E E E E E E E E s %RSD 8.80% 3.54% 6.77% 2.77% 0.68% 3.48% 6.80% 3.34% 4.17% 0.46% 12.41% 36.09% (Trace C) BKG 3.993E E E E E E E E E E E E E-12 LOD 9.227E E E E E E E E E E E E E-10 Microvision 2 can run at the same speed as traditional RGA with x6 better detection and similar precision Microvision 2 can run 40% faster than traditional RGA with x3 better detection and similar precision

7 Often a fast event will occur while another slit valve is open Chart showing a pulse witin a transient event for Microvision Plus (1 to 70 amu barchart, 5ms delay, 7.5ms measurement) Chart showing a pulse witin a transient event for Microvision 2 (1 to 70 amu barchart, 10ms delay, 2ms measurement) Partial pressure, mbar 1.00E E E E E E-08 Scan time = 1.10secs DL = 4E-10mbar Mass 4 Mass 28 Mass 32 Mass 36 Mass 40 Partial pressure, mbar 1.00E E E E E E-08 Scan time = 1.01secs DL = 2E-10mbar Mass 4 Mass 28 Mass 32 Mass 36 Mass E E E E E-11 00: : : : : : :51.8 Time, mm:ss E-11 00: : : : : : :51.8 Time, mm:ss.0 By changing to a lower accuracy setting, faster data can be acquired The clipped maximum signal of traditional RGA is now more noticeable Detection limits at 2x worse for traditional RGA If the events are any faster then switching to peak jump mode will offer better time resolution

8 By choosing to measure only the gas peaks of interest Chart showing a pulse witin a transient event for Microvision Plus (10 mass peak jump, 5ms delay, 3.6ms measurement, default zero for each scan) Chart showing a pulse witin a transient event for Microvision 2 (10 mass peak jump, 10ms delay, 2ms measurement, default zero for each scan) 1.00E E E-05 Mass 4 Mass 28 Mass 32 Mass 36 Mass E E E-05 Mass 4 Mass 28 Mass 32 Mass 36 Mass 40 Partial pressure, mbar 1.00E E E-08 Scan time = 0.32secs DL = 1.4E-9mbar Partial pressure, mbar 1.00E E E-08 Scan time = 0.19secs DL = 2.7E-10mbar 1.00E E E E E-11 00: : : : : : :51.8 Time, mm:ss E-11 00: : : : : : :51.8 Time, mm:ss.0 At fast peak jumping acquisition speeds with full dynamic range The clipped maximum signal of traditional RGA is now more noticeable Traditional RGA can not resolve all the changes in partial pressure Microvision 2 is 1.5x faster with 5x better detection limits

9 Conclusion for XFR type data Microvision 2 can capture data for standard TOOLweb RGA applications with six times better detection limits Microvision 2 can capture up to five times more data points leading to better definition of transient events without any loss in detection limit or coverage of gas species Microvision 2 data acquisition parameters can be optimised for up to 10x faster scan times with 8x better detection limits when peak jumping for specific gas species only

10 Application example 2: Transient gas pulses - ALD Atomic layer deposition uses bursts of gas into a large chamber to deposit single molecule thick layers of reactant species onto a substrate During each inlet phase it is preferred that the reactant gas species should be introduced at exactly the correct level to prevent wasting expensive pre-cursor chemicals and extending the time between PM cycles The recipe can be changed to vary the ALD pulse width to introduce the correct level of reactant if the source of the reactant is degrading After each inlet pulse the decay of the reactant gas in the chamber and the monitoring of product gases from the surface reaction give indications of the quality of the ALD process

11 Data acquisition experiment to prove the effectiveness of Microvision 2 for ALD To simulate the effects of an ALD chamber with pulsed gases, a 10 litre vacuum chamber was turbo pumped to 1e-7 mbar using an open ion source Microvision 2 and an ALD valve fed with 0.5mbar of air The ALD valve was pulsed every 10 seconds and the pulse width varied from 80 to 300ms. The effect of the chamber was to broaden the gas pulse to less than 1 second (50% peak height) and reduce the pressure into a range where an open ion source RGA could be used In normal ALD conditions the chamber would be much larger, with much greater pumping capacity and a differentially pumped RGA inlet with a closed ion source would be used (such as a Vision 2000-C).

12 Data quality concerns For ALD there is interest in both the major gas species (the reactants) and trace gas species (reaction byproducts and chamber contaminants) so the ideal RGA is capable of measuring the widest dynamic range with each scan As the species of interest are transient and short lived, the highest quality of information requires a high rate of data collection Change masses quickly with no loss in stability Quickly stabilise on the new signal Quickly switch between gain ranges for the widest measurement range Have a high stability zero so it is possible to run without having to spend a long time on each scan re-zeroing

13 Data collections parameters Peak jump was used to maximise the time spent measuring species of interest Masses 28 (N2), 32 (O2) and 44 (CO2) Masses 36, 38 and 40 Argon isotopes Scan time for six masses was 70ms This is possible as the RF electronics are temperature controlled to allow accurate mass positioning in <1ms The dynamic range at this speed is possible as the detector electronics have a very fast diode based gain switching system for high and low signal For the first data set shown, one zero was taken at the start of the data collection and data collected for 2 hours Baseline changed by <0.5ppm of full scale This is due to having temperature controlled electrometer

14 Example traces for varying ALD pulses

15 Data collected during an 80ms pulse of an ALD valve 1.00E ms 300ms pulse pulse has has broader broader peaks peaks which which are are not not proportional proportional in in peak peak height height to to the the 80ms 80ms pulses pulses A A data data integration integration method method is required to give more is required to give more accurate accurate measurements measurements for for variable variable ALD ALD pulse pulse widths widths Almost Almost 66 decades decades of of signal measurement signal measurement range range Maximum Maximum signal signal 1E 1E-5-5 mbar mbar Noise Noise between between pulses pulses for for mass mass is is 2E 2E mbar mbar Mass E-08 Mass 32 Mass 36 Mass 38 Mass E-09 Mass E E E-12 28: : : : : : : : : : : Time (mm:ss.000) Data collected during a 300ms pulse of an ALD valve 1.00E E E-07 Partial pressure, mbar 1.00E-07 P artial pressure, mbar Example data for 80ms and 300ms ALD pulses 1.00E-06 Mass E-08 Mass 32 Mass 36 Mass 38 Mass E-09 Mass E E E-12 30: : : : : : : Time (mm:ss.000) 31: : : :09.696

16 Comparison of peak shapes for different ALD pulse times and schematic of the data integration method Comparision of O2 measured peak for different ALD valve pulse times in ms. Chart showing the integration technique used for ALD pulse measurement 1.00E E-06 Mass 32 (80ms) 1.40E-06 Mass 32 (100ms) Mass 32 (150ms) Mass 32 (200ms) Mass 32 (80ms) Mass 32 (300ms) Partial pressure, mbar Partial press ure, mbar 1.20E-06 Mass 32 area Background 1.00E E E E E E E E E E+00 00: E-10 00: : : : : Time (mm:ss.000) 00: : : : : : : : : : : : Time (mm.ss.000) Integrating Integrating over over the the peak peak area area and and subtracting subtracting the the baseline baseline offset offset under under the the peak peak will will lead lead to to aa linear linear relationship relationship between between the the measured measured partial partial pressures pressures and and the the ALD ALD pulse pulse times times Integrating Integrating under under the the peak peak is is only only possible possible with with fast fast enough enough data data acquisition, acquisition, aa peak peak must must have have 77 or or more more points points over over the the top top 90% 90% of of the the peak peak for for best best accuracy. accuracy. This This data data has has nine nine points. points.

17 Graph showing pulse peak areas for N2 and O2 with varying pulse widths Graph showing pulse peak areas of 40Ar and CO2 with varying pulse widths 1.60E E-05 Mass 28 Mass E E-05 Mass E E E E-06 Pulse peak area Pulse peak area Mass E E E E E E E E E E Comparison of peak area and peak height for ALD chemical delivery 1.20E E-05 Interated value for several pu lses different different ALD ALD pulse pulse widths widths Comparison Comparison of of peak peak area area vs. vs. peak peak height height for for accurately accurately checking checking gas gas species species Peak height gives non-linear data 200 ALD valve pulse width, ms ALD valve pulse width, ms Comparison Comparison of of peak peak areas areas for for 150 Mass 32 Mass 32 height 1.00E E E E E E E ALD valve pulse width, ms

18 Control of the ALD process Relies on knowing the relative levels of chemicals in each pulse It is critical that peak ratios do not change with pulse width so correct mixture changes can be made if required Chart showing ratios of peak areas for ALD pulses of different width 3.00E+02 Ratio 40/36 Ratio 32/44 Ratio 40/ E+02 Peak area ratios 2.00E E E E E ALD valve pulse width, ms

19 Conclusion for ALD type data Microvision 2 can capture data for multiple peaks on fast transient signals 6 gas partial pressures with no re-zero With 14Hz acquisition rate With almost 6 decades of dynamic range No baseline drift over several hours With linear correlation between ALD pulse width and peak area With equal gas peak ratios for all ALD pulse widths

HPR-30 Vacuum Process Gas Analyser A differentially pumped RGA system for vacuum process monitoring

HPR-30 Vacuum Process Gas Analyser A differentially pumped RGA system for vacuum process monitoring HPR-30 Vacuum Process Gas Analyser A differentially pumped RGA system for vacuum process monitoring HPR-30 vacuum process gas analyser To analyse processes with high dynamic range operating at pressures

More information

835 Vacuum Quality Monitor (VQM TM ) 2012 Brooks Automation, Inc. Proprietary Information

835 Vacuum Quality Monitor (VQM TM ) 2012 Brooks Automation, Inc. Proprietary Information 835 Vacuum Quality Monitor (VQM TM ) 2012 Brooks Automation, Inc. Proprietary Information The Revolutionary New Vacuum Quality Monitor World s Fastest, Accurate at low mass Single Gas Calibration Small

More information

Transpector 2. Gas Analysis System. NEW-GENERATION RGAs WITH SUPERIOR SENSITIVITY AND PERFORMANCE

Transpector 2. Gas Analysis System. NEW-GENERATION RGAs WITH SUPERIOR SENSITIVITY AND PERFORMANCE Transpector 2 Gas Analysis System NEW-GENERATION RGAs WITH SUPERIOR SENSITIVITY AND PERFORMANCE We ve made the best RGAs better. Inficon, the leaders in RGA technology, have upgraded the Transpector gas

More information

Residual Gas Analyzers RGA100/200/ amu, 200 amu and 300 amu RGAs

Residual Gas Analyzers RGA100/200/ amu, 200 amu and 300 amu RGAs Residual Gas Analyzers RGA100/200/300 100 amu, 200 amu and 300 amu RGAs SRS Residual Gas Analyzers 100, 200 and 300 amu systems Better than 1 amu resolution 6 decades of dynamic range 5 10 14 Torr detection

More information

Residual Gas Analyzers

Residual Gas Analyzers Residual Gas Analyzers RGA100/200/300 100 amu, 200 amu and 300 amu systems SRS Residual Gas Analyzers 100, 200 and 300 amu systems Better than 1 amu resolution 6 decades of dynamic range 5 10 14 Torr detection

More information

830 Vacuum Quality Monitor

830 Vacuum Quality Monitor Granville-Phillips Vacuum Instrumentation 830 Vacuum Quality Monitor Featuring ART MS Technology February 2011 1 What is ART MS? ART MS Anharmonic Resonant Trap Mass Spectrometry or Auto Resonant Ion Trap

More information

HIQUAD. New high-end mass spectrometer! Fast, flexible and easy to operate.

HIQUAD. New high-end mass spectrometer! Fast, flexible and easy to operate. New high-end mass spectrometer! Fast, flexible and easy to operate. New high-end mass spectrometer! Fast, flexible and easy to operate. A modular solution for mass spectrometry With the new HiQuad mass

More information

Characterization of Common Electron Multipliers in Harsh Environments

Characterization of Common Electron Multipliers in Harsh Environments ELECTRO-OPTICS Characterization of Common Electron Multipliers in Harsh Environments The Pittsburgh Conference 2005 Poster Paper 1340-20 Bruce Laprade and Raymond Cochran BURLE Electro-Optics INC Introduction

More information

Residual Gas Analyzers XT Series

Residual Gas Analyzers XT Series Residual Gas Analyzers XT Series Products from ExTorr Inc. - Pirani, Ion Gauge, Quadrupole - All Included The Extorr XT residual gas analyzer is a quadrupole mass spectrometer complete with a built-in

More information

The Development of a Novel Electron Multiplier with an Onboard Integral High Voltage Power Supply for use in Mass Spectrometers

The Development of a Novel Electron Multiplier with an Onboard Integral High Voltage Power Supply for use in Mass Spectrometers The Development of a Novel Electron Multiplier with an Onboard Integral High Voltage Power Supply for use in Mass Spectrometers Presented ASMS 2007 New Instrumentation Concepts Session, Poster 043 Bruce

More information

Development of a Small Residual Gas Analyzer Utilizing the Quadrupole Array Structure Micropole System ~ QL Series ~

Development of a Small Residual Gas Analyzer Utilizing the Quadrupole Array Structure Micropole System ~ QL Series ~ F e a t u r e A r t i c l e Feature Article Development of a Small Residual Gas Analyzer Utilizing the Quadrupole Array Structure Micropole System ~ QL Series ~ Hirokazu Kitaura The Micropole System is

More information

Notes on DYNALYZER HVU Usage

Notes on DYNALYZER HVU Usage December 19, 1986 Notes on DYNALYZER HVU Usage It has been shown that the Dynalyzer HVU gives accurate reproduction of tube waveforms for almost all applications. Possible cautions, a) kv frequency response

More information

Autoresonant Trap Mass Spectrometry

Autoresonant Trap Mass Spectrometry 8 th Harsh Environment Mass Spectrometry Wokshop Autoresonant Trap Mass Spectrometry (ART MS) Sept 21, 2011 Gerardo A. Brucker and G. Jeffery Rathbone Granville-Phillips Vacuum Instrumentation Brooks Automation,

More information

PRISMAPLUS. The precise solution for mass spectrometry. Modular design. Powerful software. Wide range of applications.

PRISMAPLUS. The precise solution for mass spectrometry. Modular design. Powerful software. Wide range of applications. PRISMAPLUS The precise solution for mass spectrometry. Modular design. Powerful software. Wide range of applications. PRISMAPLUS The precise solution for mass spectrometry. The combination of high sensi

More information

ICP-MS. plasma 3. Multi-Collector ICP-MS.

ICP-MS. plasma 3. Multi-Collector ICP-MS. ICP-MS plasma 3 Multi-Collector ICP-MS www.nu-ins.com plasma 3 is a third generation Multi Collector ICP Mass Spectrometer (MC-ICP-MS), designed to provide the best possible precision and accuracy for

More information

ToF-AMS DAQ. Joel Kimmel Univ. of Colorado at Boulder & Aerodyne Research, Inc AMS Users Meeting 2007

ToF-AMS DAQ. Joel Kimmel Univ. of Colorado at Boulder & Aerodyne Research, Inc AMS Users Meeting 2007 ToF-AMS DAQ Joel Kimmel Univ. of Colorado at Boulder & Aerodyne Research, Inc AMS Users Meeting 2007 1 Web Resources http://cires.colorado.edu/jimenez-group/tofamsresources/tofsoftware/ Downloads Release

More information

Project by: Dr. Jorge A. Diaz. Physics School, University of Costa Rica, National Center for High Technology (CENAT)

Project by: Dr. Jorge A. Diaz. Physics School, University of Costa Rica, National Center for High Technology (CENAT) Test of a Miniature Double-Focusing Mass Spectrometer for the Variable Specific Impulse Magnetoplasma Rocket (VASIMR) at the Advanced Space Propulsion Laboratory Project by: Dr. Jorge A. Diaz Physics School,

More information

Application Note (A13)

Application Note (A13) Application Note (A13) Fast NVIS Measurements Revision: A February 1997 Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1 407 422 3171 Fax: 1 407 648 5412 Email: sales@goochandhousego.com In

More information

A Simplified Test Set for Op Amp Characterization

A Simplified Test Set for Op Amp Characterization A Simplified Test Set for Op Amp Characterization INTRODUCTION The test set described in this paper allows complete quantitative characterization of all dc operational amplifier parameters quickly and

More information

Dust Measurements With The DIII-D Thomson system

Dust Measurements With The DIII-D Thomson system Dust Measurements With The DIII-D Thomson system The DIII-D Thomson scattering system, consisting of eight ND:YAG lasers and 44 polychromator detection boxes, has recently been used to observe the existence

More information

ALPHASENSE USER MANUAL. Toxic Sensor Evaluation Board Issue 4

ALPHASENSE USER MANUAL. Toxic Sensor Evaluation Board Issue 4 ALPHASENSE USER MANUAL Toxic Sensor Evaluation Board 072-0128 Issue 4 Introduction This Evaluation Board accepts Alphasense A, B and D Series toxic gas sensors. The purposes of this evaluation board is

More information

OBSOLETE. High Performance, BiFET Operational Amplifiers AD542/AD544/AD547 REV. B

OBSOLETE. High Performance, BiFET Operational Amplifiers AD542/AD544/AD547 REV. B a FEATURES Ultralow Drift: 1 V/ C (AD547L) Low Offset Voltage: 0.25 mv (AD547L) Low Input Bias Currents: 25 pa max Low Quiescent Current: 1.5 ma Low Noise: 2 V p-p High Open Loop Gain: 110 db High Slew

More information

PrismaPlus. The New Mass Spectrometer with the Added Plus! Modular Design. Powerful Software. Wide Range of Applications.

PrismaPlus. The New Mass Spectrometer with the Added Plus! Modular Design. Powerful Software. Wide Range of Applications. PrismaPlus The New Mass Spectrometer with the Added Plus! Modular Design. Powerful Software. Wide Range of Applications. PrismaPlus The PrismaPlus innovation The optimum combination of high sensitivity,

More information

LM231A/LM231/LM331A/LM331 Precision Voltage-to-Frequency Converters

LM231A/LM231/LM331A/LM331 Precision Voltage-to-Frequency Converters LM231A/LM231/LM331A/LM331 Precision Voltage-to-Frequency Converters General Description The LM231/LM331 family of voltage-to-frequency converters are ideally suited for use in simple low-cost circuits

More information

A NEW GENERATION PROGRAMMABLE PHASE/AMPLITUDE MEASUREMENT RECEIVER

A NEW GENERATION PROGRAMMABLE PHASE/AMPLITUDE MEASUREMENT RECEIVER GENERAL A NEW GENERATION PROGRAMMABLE PHASE/AMPLITUDE MEASUREMENT RECEIVER by Charles H. Currie Scientific-Atlanta, Inc. 3845 Pleasantdale Road Atlanta, Georgia 30340 A new generation programmable, phase-amplitude

More information

Special-Purpose Operational Amplifier Circuits

Special-Purpose Operational Amplifier Circuits Special-Purpose Operational Amplifier Circuits Instrumentation Amplifier An instrumentation amplifier (IA) is a differential voltagegain device that amplifies the difference between the voltages existing

More information

Gas Monitor Detector Electronics

Gas Monitor Detector Electronics Gas Monitor Detector Electronics from charged particles to digital value Fini Jastrow, TTF/VUV-FEL TTF/VUV-FEL meeting, meeting, 21. Mar 21. 06Mar 06 The Detector Ion Current Measurement Measured Data

More information

Response time reduction of the ZXCT1009 Current Monitor

Response time reduction of the ZXCT1009 Current Monitor Response time reduction of the ZXCT1009 Current Monitor Geoffrey Stokes, Systems Engineer, Diodes Incorporated Introduction and Summary The transient response of the ZXCT1009 and ZXCt1008 Current Monitors

More information

Operational Amplifiers

Operational Amplifiers Operational Amplifiers Table of contents 1. Design 1.1. The Differential Amplifier 1.2. Level Shifter 1.3. Power Amplifier 2. Characteristics 3. The Opamp without NFB 4. Linear Amplifiers 4.1. The Non-Inverting

More information

New Control Electronics for Helmer Gauges

New Control Electronics for Helmer Gauges New Control Electronics for Helmer Gauges Nikolaos Chatzigeorgiou TE-VSC-ICM Technology Department ICM Automation FORUM, 24th May 2016 Agenda The Helmer gauge Control electronics Supervision application

More information

PRISMAPRO. The latest generation of compact mass spectrometers. Modular design. Powerful software. Low detection limit.

PRISMAPRO. The latest generation of compact mass spectrometers. Modular design. Powerful software. Low detection limit. The latest generation of compact mass spectrometers. Modular design. Powerful software. Low detection limit. The latest generation of compact mass spectrometers. The combination of high sensi tivity, maximum

More information

Experiment 2: Transients and Oscillations in RLC Circuits

Experiment 2: Transients and Oscillations in RLC Circuits Experiment 2: Transients and Oscillations in RLC Circuits Will Chemelewski Partner: Brian Enders TA: Nielsen See laboratory book #1 pages 5-7, data taken September 1, 2009 September 7, 2009 Abstract Transient

More information

Multianode Photo Multiplier Tubes as Photo Detectors for Ring Imaging Cherenkov Detectors

Multianode Photo Multiplier Tubes as Photo Detectors for Ring Imaging Cherenkov Detectors Multianode Photo Multiplier Tubes as Photo Detectors for Ring Imaging Cherenkov Detectors F. Muheim a edin]department of Physics and Astronomy, University of Edinburgh Mayfield Road, Edinburgh EH9 3JZ,

More information

css Custom Silicon Solutions, Inc.

css Custom Silicon Solutions, Inc. css Custom Silicon Solutions, Inc. GENERAL PART DESCRIPTION The is a micropower version of the popular timer IC. It features an operating current under µa and a minimum supply voltage of., making it ideal

More information

Subject: Best Practices for Improving Tafel Plots of High Capacitance Cells with Low Series Resistance

Subject: Best Practices for Improving Tafel Plots of High Capacitance Cells with Low Series Resistance Technical Note Subject: Best Practices for Improving Tafel Plots of High Capacitance Cells with Low Series Resistance Date: April 2014 The PARSTAT4000 is designed with both function and versatility at

More information

XRF Instrumentation. Introduction to spectrometer

XRF Instrumentation. Introduction to spectrometer XRF Instrumentation Introduction to spectrometer AMPTEK, INC., Bedford, MA 01730 Ph: +1 781 275 2242 Fax: +1 781 275 3470 sales@amptek.com 1 Instrument Excitation source Sample X-ray tube or radioisotope

More information

Hello, and welcome to this presentation of the STM32 Digital Filter for Sigma-Delta modulators interface. The features of this interface, which

Hello, and welcome to this presentation of the STM32 Digital Filter for Sigma-Delta modulators interface. The features of this interface, which Hello, and welcome to this presentation of the STM32 Digital Filter for Sigma-Delta modulators interface. The features of this interface, which behaves like ADC with external analog part and configurable

More information

Chem466 Lecture Notes. Spring, 2004

Chem466 Lecture Notes. Spring, 2004 Chem466 Lecture Notes Spring, 004 Overview of the course: Many of you will use instruments for chemical analyses in lab. settings. Some of you will go into careers (medicine, pharmacology, forensic science,

More information

RCTrms Technical Notes

RCTrms Technical Notes RCTrms Technical Notes All measuring instruments are subject to limitations. The purpose of these technical notes is to explain some of those limitations and to help the engineer maximise the many advantages

More information

Analogue to Digital Conversion

Analogue to Digital Conversion Analogue to Digital Conversion Turns electrical input (voltage/current) into numeric value Parameters and requirements Resolution the granularity of the digital values Integral NonLinearity proportionality

More information

50 MHz Voltage-to-Frequency Converter

50 MHz Voltage-to-Frequency Converter Journal of Physics: Conference Series OPEN ACCESS 50 MHz Voltage-to-Frequency Converter To cite this article: T Madden and J Baldwin 2014 J. Phys.: Conf. Ser. 493 012008 View the article online for updates

More information

Hollow Cathode Lamps

Hollow Cathode Lamps Hollow Cathode Lamps Competitive Comparison The hollow cathode lamp is a discharge lamp designed for use as a spectral line source with atomic absorption (AA) spectrometers. A single or multi-element hollow

More information

AccuTOF LC Training Course

AccuTOF LC Training Course 1 AccuTOF LC Training Course JEOL USA, Inc. 11 Dearborn Road Peabody, MA 01960 Area map Restaurants: Wendy s Bennigan s Bertucci s 2 AccuTOF LC Training Course Monday Basic principle and history of TOF

More information

Front-End and Readout Electronics for Silicon Trackers at the ILC

Front-End and Readout Electronics for Silicon Trackers at the ILC 2005 International Linear Collider Workshop - Stanford, U.S.A. Front-End and Readout Electronics for Silicon Trackers at the ILC M. Dhellot, J-F. Genat, H. Lebbolo, T-H. Pham, and A. Savoy Navarro LPNHE

More information

HA Features. 12MHz, High Input Impedance, Operational Amplifier. Applications. Pinout. Part Number Information. Data Sheet May 2003 FN2893.

HA Features. 12MHz, High Input Impedance, Operational Amplifier. Applications. Pinout. Part Number Information. Data Sheet May 2003 FN2893. OBSOLETE PRODUCT POSSIBLE SUBSTITUTE PRODUCT HA-2525 HA-2515 Data Sheet May 23 FN2893.5 12MHz, High Input Impedance, Operational Amplifier HA-2515 is a high performance operational amplifier which sets

More information

PAiA 4780 Twelve Stage Analog Sequencer Design Analysis Originally published 1974

PAiA 4780 Twelve Stage Analog Sequencer Design Analysis Originally published 1974 PAiA 4780 Twelve Stage Analog Sequencer Design Analysis Originally published 1974 DESIGN ANALYSIS: CLOCK As is shown in the block diagram of the sequencer (fig. 1) and the schematic (fig. 2), the clock

More information

Technical Specifications for the procurement of Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Technical Specifications for the procurement of Inductively Coupled Plasma Mass Spectrometer (ICP-MS) Technical Specifications for the procurement of Inductively Coupled Plasma Mass Spectrometer (ICP-MS) Preamble An Inductively Coupled Plasma Mass Spectrometer (Quadrupole ICP-MS) is proposed to be procured

More information

Applications of the LM392 Comparator Op Amp IC

Applications of the LM392 Comparator Op Amp IC Applications of the LM392 Comparator Op Amp IC The LM339 quad comparator and the LM324 op amp are among the most widely used linear ICs today The combination of low cost single or dual supply operation

More information

Applications of the LM392 Comparator Op Amp IC

Applications of the LM392 Comparator Op Amp IC Applications of the LM392 Comparator Op Amp IC The LM339 quad comparator and the LM324 op amp are among the most widely used linear ICs today. The combination of low cost, single or dual supply operation

More information

PCS-150 / PCI-200 High Speed Boxcar Modules

PCS-150 / PCI-200 High Speed Boxcar Modules Becker & Hickl GmbH Kolonnenstr. 29 10829 Berlin Tel. 030 / 787 56 32 Fax. 030 / 787 57 34 email: info@becker-hickl.de http://www.becker-hickl.de PCSAPP.DOC PCS-150 / PCI-200 High Speed Boxcar Modules

More information

LINEAR IC APPLICATIONS

LINEAR IC APPLICATIONS 1 B.Tech III Year I Semester (R09) Regular & Supplementary Examinations December/January 2013/14 1 (a) Why is R e in an emitter-coupled differential amplifier replaced by a constant current source? (b)

More information

2302 Battery Simulator 2306, 2306-PJ Battery/Charger Simulators

2302 Battery Simulator 2306, 2306-PJ Battery/Charger Simulators Ultrafast response to transient load currents Choice of single- or dualchannel supplies Optimized for development and testing of battery-powered devices Variable output resistance for simulating battery

More information

LF155/LF156/LF355/LF356/LF357 JFET Input Operational Amplifiers

LF155/LF156/LF355/LF356/LF357 JFET Input Operational Amplifiers JFET Input Operational Amplifiers General Description These are the first monolithic JFET input operational amplifiers to incorporate well matched, high voltage JFETs on the same chip with standard bipolar

More information

Non-Destructive Ion Trap Mass Analysis at High Pressure. Supporting information

Non-Destructive Ion Trap Mass Analysis at High Pressure. Supporting information Non-Destructive Ion Trap Mass Analysis at High Pressure Supporting information Wei Xu 1, Jeffrey Maas 2, Frank Boudreau 3, William J. Chappell 2 and Zheng Ouyang 1,2,3* 1. Weldon School of Biomedical Engineering,

More information

MWPC Gas Gain with Argon-CO 2 80:20 Gas Mixture

MWPC Gas Gain with Argon-CO 2 80:20 Gas Mixture IMA Journal of Mathematical Control and Information Page 1 of 10 doi:10.1093/imamci/dri000 1. Principles of Operation MWPC Gas Gain with Argon-CO 2 80:20 Gas Mixture Michael Roberts A multi-wire proportional

More information

Biosensors and Instrumentation: Tutorial 3

Biosensors and Instrumentation: Tutorial 3 Biosensors and Instrumentation: Tutorial 3 1 1. A schematic cross section of an ion sensitive field effect transistor (ISFET) is shown in figure 1. Vref Solution eference Electrode Encapsulation SiO2 nsi

More information

hij Teacher Resource Bank GCE Electronics Exemplar Examination Questions ELEC2 Further Electronics

hij Teacher Resource Bank GCE Electronics Exemplar Examination Questions ELEC2 Further Electronics hij Teacher Resource Bank GCE Electronics Exemplar Examination Questions ELEC2 Further Electronics The Assessment and Qualifications Alliance (AQA) is a company limited by guarantee registered in England

More information

Optimized Process Performance Using the Paramount /Navigator Power- Delivery/Match Solution

Optimized Process Performance Using the Paramount /Navigator Power- Delivery/Match Solution Optimized Process Performance Using the Paramount /Navigator Power- Delivery/Match Solution Dan Carter, Advanced Energy Industries, Inc. Numerous challenges face designers and users of today s RF plasma

More information

Voltage-to-Frequency and Frequency-to-Voltage Converter ADVFC32

Voltage-to-Frequency and Frequency-to-Voltage Converter ADVFC32 a FEATURES High Linearity 0.01% max at 10 khz FS 0.05% max at 100 khz FS 0.2% max at 500 khz FS Output TTL/CMOS Compatible V/F or F/V Conversion 6 Decade Dynamic Range Voltage or Current Input Reliable

More information

Battery Simulator Battery/Charger Simulators

Battery Simulator Battery/Charger Simulators Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com 2302, 2302-PJ, Ultrafast response to transient load currents Choice of single- or dualchannel supplies

More information

HOM Based Diagnostics at the TTF

HOM Based Diagnostics at the TTF HOM Based Diagnostics at the TTF Nov 14, 2005 Josef Frisch, Nicoleta Baboi, Linda Hendrickson, Olaf Hensler, Douglas McCormick, Justin May, Olivier Napoly, Rita Paparella, Marc Ross, Claire Simon, Tonee

More information

A high-sensitivity drinkometer circuit with 60-Hz filtering

A high-sensitivity drinkometer circuit with 60-Hz filtering Behavior Research Methods 2007, 39 (1), 118-122 A high-sensitivity drinkometer circuit with 60-Hz filtering ROGER L. OVERTON Huntington Station, New York AND DONALD A. OVERTON Temple University, Philadelphia,

More information

DATASHEET HA Features. Applications. Pinout. Part Number Information. 12MHz, High Input Impedance, Operational Amplifier

DATASHEET HA Features. Applications. Pinout. Part Number Information. 12MHz, High Input Impedance, Operational Amplifier 12MHz, High Input Impedance, Operational Amplifier OBSOLETE PRODUCT POSSIBLE SUBSTITUTE PRODUCT HA-2525 DATASHEET FN289 Rev 6. HA-255 is an operational amplifier whose design is optimized to deliver excellent

More information

Control of Sputter Process for Improved Run-to-run Repeatability

Control of Sputter Process for Improved Run-to-run Repeatability Control of Sputter Process for Improved Run-to-run Repeatability S. Ghosal, R.L. Kosut, J.L. Ebert, L. Porter SC Solutions, Santa Clara, CA 95054 E-mail ghosal@scsolutions.com D. Brownell, D. Wang Nonvolatile

More information

Tel: Fax:

Tel: Fax: B Tel: 78.39.4700 Fax: 78.46.33 SPECIFICATIONS (T A = +5 C, V+ = +5 V, V = V or 5 V, all voltages measured with respect to digital common, unless otherwise noted) AD57J AD57K AD57S Model Min Typ Max Min

More information

Model 6600A Dual Source High Resistance Bridge

Model 6600A Dual Source High Resistance Bridge Dual Source High Resistance Bridge Based on proven NMI Design Range: 100 kω to 10 PΩ Voltages: 1 V to 1000 V (5000 V Optional) Automatic and Manual Operation Not affected by Temperature change 10 and 20

More information

AD8232 EVALUATION BOARD DOCUMENTATION

AD8232 EVALUATION BOARD DOCUMENTATION One Technology Way P.O. Box 9106 Norwood, MA 02062-9106 Tel: 781.329.4700 Fax: 781.461.3113 www.analog.com AD8232 EVALUATION BOARD DOCUMENTATION FEATURES Ready to use Heart Rate Monitor (HRM) Front end

More information

Summer 2015 Examination

Summer 2015 Examination Summer 2015 Examination Subject Code: 17445 Model Answer Important Instructions to examiners: 1) The answers should be examined by key words and not as word-to-word as given in the model answer scheme.

More information

Transient Current Measurement for Advance Materials & Devices

Transient Current Measurement for Advance Materials & Devices & Devices 8 May 2017 Brian YEO Application Engineer Keysight Technologies Agenda 2 High speed data acquisition basics Challenges & solutions for transient current measurement. Considerations when making

More information

XR-4151 Voltage-to-Frequency Converter

XR-4151 Voltage-to-Frequency Converter ...the analog plus company TM XR-45 Voltage-to-Frequency Converter FEATURES APPLICATIONS June 99- Single Supply Operation (+V to +V) Voltage-to-Frequency Conversion Pulse Output Compatible with All Logic

More information

potentiostat/galvanostat/impedance analyser

potentiostat/galvanostat/impedance analyser potentiostat/galvanostat/impedance analyser Rev. 9-2014 potentiostat/galvanostat/impedance PalmSens3 is a battery-powered, handheld instrument which allows the application of most of the relevant voltammetric,

More information

HA-2520, HA MHz, High Slew Rate, Uncompensated, High Input Impedance, Operational Amplifiers. Features. Applications. Ordering Information

HA-2520, HA MHz, High Slew Rate, Uncompensated, High Input Impedance, Operational Amplifiers. Features. Applications. Ordering Information HA-22, HA-22 Data Sheet August, 2 FN2894. 2MHz, High Slew Rate, Uncompensated, High Input Impedance, Operational Amplifiers HA-22/22 comprise a series of operational amplifiers delivering an unsurpassed

More information

HA-2520, HA-2522, HA-2525

HA-2520, HA-2522, HA-2525 HA-, HA-, HA- Data Sheet September 99 File Number 9. MHz, High Slew Rate, Uncompensated, High Input Impedance, Operational Amplifiers HA-// comprise a series of operational amplifiers delivering an unsurpassed

More information

Achieving accurate measurements of large DC currents

Achieving accurate measurements of large DC currents Achieving accurate measurements of large DC currents Victor Marten, Sendyne Corp. - April 15, 2014 While many instruments are available to accurately measure small DC currents (up to 3 A), few devices

More information

The Extrel MAX Systems are UHV-compatible flange mounted Quadrupole Mass

The Extrel MAX Systems are UHV-compatible flange mounted Quadrupole Mass The Extrel MAX Systems are UHV-compatible flange mounted Quadrupole Mass Spectrometers. The Ionizer, Quadrupole Mass Filter, and Detector on a Mounting Flange are designed for inclusion in your experimental

More information

High resolution measurements The differential approach

High resolution measurements The differential approach Electrical characterisation of nanoscale samples & biochemical interfaces: methods and electronic instrumentation High resolution measurements The differential approach Giorgio Ferrari Dipartimento di

More information

SUMMARY/DIALOGUE 2 PRESHAPE PIXEL OVERVIEW 3 BRIEF OPERATING INSTRUCTIONS 3 PRESHAPE PIXEL SIMULATION: EXAMPLE OPERATION 4 PRESHAPE PIXEL SIMULATION:

SUMMARY/DIALOGUE 2 PRESHAPE PIXEL OVERVIEW 3 BRIEF OPERATING INSTRUCTIONS 3 PRESHAPE PIXEL SIMULATION: EXAMPLE OPERATION 4 PRESHAPE PIXEL SIMULATION: SUMMARY/DIALOGUE 2 PRESHAPE PIXEL OVERVIEW 3 BRIEF OPERATING INSTRUCTIONS 3 PRESHAPE PIXEL SIMULATION: EXAMPLE OPERATION 4 PRESHAPE PIXEL SIMULATION: SMALL SIGNALS AROUND THRESHOLD 5 PRESHAPE PIXEL SIMULATION:

More information

Product Range Electronic Units

Product Range Electronic Units Pyramid Technical Consultants, Inc. 1050 Waltham Street Suite 200 Lexington, MA 02421 TEL: +1 781 402-1700 TEL (UK): +44 1273 492001 FAX: (781) 402-1750 EMAIL: SUPPORT@PTCUSA.COM Product Range Electronic

More information

Improved Pre-Sample pixel

Improved Pre-Sample pixel Improved Pre-Sample pixel SUMMARY/DIALOGUE 2 PRESAMPLE PIXEL OVERVIEW 3 PRESAMPLE PIXEL SIMULATION: EXAMPLE OPERATION 4 PRESAMPLE PIXEL SIMULATION: SMALL SIGNALS AROUND THRESHOLD 6 PRESAMPLE PIXEL SIMULATION:

More information

CHAPTER 3. Instrumentation Amplifier (IA) Background. 3.1 Introduction. 3.2 Instrumentation Amplifier Architecture and Configurations

CHAPTER 3. Instrumentation Amplifier (IA) Background. 3.1 Introduction. 3.2 Instrumentation Amplifier Architecture and Configurations CHAPTER 3 Instrumentation Amplifier (IA) Background 3.1 Introduction The IAs are key circuits in many sensor readout systems where, there is a need to amplify small differential signals in the presence

More information

Chapter 13: Introduction to Switched- Capacitor Circuits

Chapter 13: Introduction to Switched- Capacitor Circuits Chapter 13: Introduction to Switched- Capacitor Circuits 13.1 General Considerations 13.2 Sampling Switches 13.3 Switched-Capacitor Amplifiers 13.4 Switched-Capacitor Integrator 13.5 Switched-Capacitor

More information

BioInstrumentation Laboratory

BioInstrumentation Laboratory BioInstrumentation Laboratory Ian Hunter Vienna, May 22 2013 BioInstrumentation Lab, Mechanical Engineering, MIT - Robotic endoscopes - Needle-free drug delivery devices - Eye micro-surgery robots - High

More information

A New Conjoined RF Ion Guide for Enhanced Ion Transmission

A New Conjoined RF Ion Guide for Enhanced Ion Transmission A New Conjoined RF Ion Guide for Enhanced Ion Transmission Kevin Giles and David Gordon Waters Corporation, Manchester, UK APPLICATION BENEFITS We describe a novel conjoined ion guide that operates at

More information

Distributed by: www.jameco.com 1-800-831-4242 The content and copyrights of the attached material are the property of its owner. LM231A/LM231/LM331A/LM331 Precision Voltage-to-Frequency Converters General

More information

A COMPACT, AGILE, LOW-PHASE-NOISE FREQUENCY SOURCE WITH AM, FM AND PULSE MODULATION CAPABILITIES

A COMPACT, AGILE, LOW-PHASE-NOISE FREQUENCY SOURCE WITH AM, FM AND PULSE MODULATION CAPABILITIES A COMPACT, AGILE, LOW-PHASE-NOISE FREQUENCY SOURCE WITH AM, FM AND PULSE MODULATION CAPABILITIES Alexander Chenakin Phase Matrix, Inc. 109 Bonaventura Drive San Jose, CA 95134, USA achenakin@phasematrix.com

More information

Technical Information

Technical Information Technical Information Introduction to force sensors Driving long cable lengths Conversions, article reprints, glossary INTRODUCTION TO QUARTZ FORCE SENSORS Quartz Force Sensors are well suited for dynamic

More information

Considerations for Analog Input and Output

Considerations for Analog Input and Output Considerations for Analog Input and Output Useful information can be found in the text in Sections 6.7.1 (Data Rates), 6.7.5 (Analog Input Signals), 6.7.6 (Multiple Signal Sources: Data Loggers), 6.7.9

More information

Analogue to Digital Conversion

Analogue to Digital Conversion Analogue to Digital Conversion Turns electrical input (voltage/current) into numeric value Parameters and requirements Resolution the granularity of the digital values Integral NonLinearity proportionality

More information

HA MHz, High Slew Rate, High Output Current Buffer. Description. Features. Applications. Ordering Information. Pinouts.

HA MHz, High Slew Rate, High Output Current Buffer. Description. Features. Applications. Ordering Information. Pinouts. SEMICONDUCTOR HA-2 November 99 Features Voltage Gain...............................99 High Input Impedance.................... kω Low Output Impedance....................... Ω Very High Slew Rate....................

More information

Thermo Scientific icap 7000 Plus Series ICP-OES: Innovative ICP-OES optical design

Thermo Scientific icap 7000 Plus Series ICP-OES: Innovative ICP-OES optical design TECHNICAL NOTE 43333 Thermo Scientific icap 7000 Plus Series ICP-OES: Innovative ICP-OES optical design Keywords Optical design, Polychromator, Spectrometer Key Benefits The Thermo Scientific icap 7000

More information

OUTPUT UP TO 300mA C2 TOP VIEW FAULT- DETECT OUTPUT. Maxim Integrated Products 1

OUTPUT UP TO 300mA C2 TOP VIEW FAULT- DETECT OUTPUT. Maxim Integrated Products 1 19-1422; Rev 2; 1/1 Low-Dropout, 3mA General Description The MAX886 low-noise, low-dropout linear regulator operates from a 2.5 to 6.5 input and is guaranteed to deliver 3mA. Typical output noise for this

More information

Radiation Detection Instrumentation

Radiation Detection Instrumentation Radiation Detection Instrumentation Principles of Detection and Gas-filled Ionization Chambers Neutron Sensitive Ionization Chambers Detection of radiation is a consequence of radiation interaction with

More information

MSP430 Teaching Materials

MSP430 Teaching Materials MSP430 Teaching Materials Chapter 9 Data Acquisition A/D Conversion Introduction Texas Instruments t Incorporated University of Beira Interior (PT) Pedro Dinis Gaspar, António Espírito Santo, Bruno Ribeiro,

More information

Multi-function Gain-Phase Analyzer (Frequency Response Analyzer) Model 2505

Multi-function Gain-Phase Analyzer (Frequency Response Analyzer) Model 2505 OTHER PRODUCTS.. Multi-function Gain-Phase Analyzer ( Response Analyzer) Model 2505 Standard Configurations Gain phase analyzer response analyzer Phase Angle Voltmeter (PAV) Fast dual channel wide-band

More information

Method for digital particle spectrometry Khryachkov Vitaly

Method for digital particle spectrometry Khryachkov Vitaly Method for digital particle spectrometry Khryachkov Vitaly Institute for physics and power engineering (IPPE) Obninsk, Russia The goals of Analog Signal Processing Signal amplification Signal filtering

More information

OPTICAL BACKSCATTER REFLECTOMETER TM (Model OBR 5T-50)

OPTICAL BACKSCATTER REFLECTOMETER TM (Model OBR 5T-50) OPTICAL BACKSCATTER REFLECTOMETER TM (Model OBR 5T-50) The Luna OBR 5T-50 delivers fast, accurate return loss, insertion loss, and length measurements with 20 micron spatial resolution. PERFORMANCE HIGHLIGHTS

More information

TOP VIEW. OUTPUT PRESET 2.5V TO 5V 200mA SHDN 3 4 BP GND. Maxim Integrated Products 1

TOP VIEW. OUTPUT PRESET 2.5V TO 5V 200mA SHDN 3 4 BP GND. Maxim Integrated Products 1 19-2584; Rev ; 1/2 Low-Noise, Low-Dropout, 2mA General Description The low-noise, low-dropout linear regulator operates from a 2.5V to 6.5V input and delivers up to 2mA. Typical output noise is 3µV RMS,

More information

Zeiss Process Gas Analyzer (PGA) based on Ion Trap technology: Zeiss-PGA shows outstanding performances as in-line, real time gas analyzer

Zeiss Process Gas Analyzer (PGA) based on Ion Trap technology: Zeiss-PGA shows outstanding performances as in-line, real time gas analyzer Zeiss Process Gas Analyzer (PGA) based on Ion Trap technology: Zeiss-PGA shows outstanding performances as in-line, real time gas analyzer Dr.-Ing. M. Aliman Principal Scientist M. Aliman / ZEISS Agenda

More information

NEEP 427 PROPORTIONAL COUNTERS. Knoll, Chapters 6 & 14 Sect. I & II

NEEP 427 PROPORTIONAL COUNTERS. Knoll, Chapters 6 & 14 Sect. I & II NEEP 427 PROPORTIONAL COUNTERS References: Knoll, Chapters 6 & 14 Sect. I & II a proportional counter the height of the output pulse is proportional to the number of ion pairs produced in the counter gas.

More information

Transform. Isolate. Regulate

Transform. Isolate. Regulate 4707 DEY ROAD LIVERPOOL, NY 13088 PHONE: (315) 701-6751 FAX: (315) 701-6752 M.S. KENNEDY CORPORATION MSK Web Site: http://www.mskennedy.com/ DC - DC Converters MS Kennedy Corp.; Revised 9/19/2013 Application

More information