1 Application Note: ESCALAB 250: High Performance Imaging XPS Key Words Surface Analysis High Resolution High Sensitivity Multitechnique Parallel Imaging Introduction The Thermo Scientific ESCALAB 250, Figure 1, is a high-performance, imaging XPS system, capable of multitechnique analysis and which, optionally, has an extensive and versatile range of optional sample preparation facilities. X-ray Sources ESCALAB 250 is equipped with a monochromatic X-ray source. An optional twin anode source is also available. The twin-crystal, microfocusing monochromator has a 500 mm Rowland circle and uses an Al anode. The user is able to select any size of X-ray spot at the sample over a range of 120 µm to 650 µm. Some of the advantages of a microfocusing monochromator are: Small area XPS measurements can be made without reducing the sensitivity of the spectrometer. Consequently, the sensitivity is improved and the analysis time is reduced. Only that part of the sample being analyzed is exposed to X-rays. Sensitive samples do not become damaged in areas remote from the analysis position. In angle-resolved measurements, the whole of the X-ray spot is within the analysis area over the range of sample angles. With use, the aluminum coating on the monochromator anode will wear, affecting the intensity of the X-rays and therefore the sensitivity of the instrument. When this happens on instruments from other suppliers, the anode has to be replaced. On ESCALAB 250, the anode can be moved to expose a fresh area to the electron beam, increasing the lifetime of the anode significantly. The total anode movement of 25 mm can be accomplished without breaking the vacuum. Figure 1: ESCALAB 250 Description Its key features are: High sensitivity spectroscopy Small area XPS Image resolution <3 µm Depth profiling capability Angle resolved XPS Microfocusing monochromator Multi-technique analytical versatility Multiple sample preparation options Automated, unattended analysis Multiple sample analysis Avantage, Windows -based data system. Lens/Analyzer/Detectors It is the lens/analyzer/detector combination on ESCALAB 250 that makes the instrument unique for both imaging and small area XPS. The system consists of A set of input lenses, including a magnetic immersion lens A 180 hemispherical energy analyzer An array of 6 channel electron multipliers (which are the detectors for spectroscopy measurements, see Figure 2) An output lens A pair of channel plate detectors for imaging XPS
2 Figure 2: Schematic diagram of ESCALAB 250 operating in its spectroscopy mode The input lens system is equipped with two iris mechanisms, one controls the field of view and the other controls the angular acceptance of the lens. When using ESCALAB 250 for imaging, photoelectrons from the whole of the field of view are detected simultaneously (see Figure 3). Electrons of a given kinetic energy are focused on the channel plate detector to produce a direct image of the sample without scanning. This is known as parallel imaging. Figure 3: Schematic diagram of ESCALAB 250 operating in its imaging mode With some other instruments, an XPS map of the sample is produced pixel by pixel by scanning the X-ray beam, the acceptance area defined by the lens or the sample stage. Parallel imaging on ESCALAB 250 provides XPS images of the highest spatial resolution and is generally faster than other methods. There are many advantages of parallel imaging over scanned or serial acquisition methods but the following are the most important: Speed. Electrons are collected from the whole of the field of the imaged area and not point by point. This improves the signal to noise ratio and allows a given image to be acquired in a much shorter time (if a 256 x 256 pixel image is required then there are effectively 65,536 parallel collection channels). Interactivity. The operator can see the image forming from the whole field of view as it is being acquired. The time for the experiment can therefore be extended or reduced during the acquisition. Using serial acquisition, the experiment duration can only be multiples of complete frame times. This can further increase the time for an experiment. Resolution. Using the serial acquisition, the best spatial resolution is determined by the minimum field of view of the lens or the minimum spot size of the X-ray source. Therefore, field of view and spatial resolution are strongly coupled. Using parallel imaging, these are independently controllable. Thus, ESCALAB 250 can produce images whose resolution is in the region of 1 µm whereas the best that can be achieved from other instruments is ~10 µm. Physical Imaging. The parallel imaging facility allows real-time physical imaging. This is achieved by using a flood electron source and operating the spectrometer in its imaging mode. This produces a physical or topographic image of the sample. The detection of the scattered electrons uses the same optics and detector as the XPS analysis, and so the image formed by the scattered electrons is from precisely the same area as the XPS image and with the same spatial resolution. Thus, the area from which the XPS image is to be acquired is located rapidly and with great precision. Small Area XPS. Both the size and the position of a small area XPS analysis can be set rapidly and precisely using the physical image. The Magnetic Immersion Lens ESCALAB 250 is fitted with a magnetic immersion lens, which increases the sensitivity of the instrument by increasing its angular acceptance. Alternatively, for a given signal intensity, the spatial resolution is improved. An instrument equipped with this type of lens has a wide range of fields of view. Using only the electrostatic lens system, the field of view can be as large as 8 mm but, by using the combination of the electrostatic and magnetic lenses, spectra can be obtained with a spatial resolution below 20 µm. ESCALAB 250 can be operated with the magnetic lens switched off. This is important, especially when analyzing some magnetic samples or when using the instrument for Auger electron spectroscopy.
3 Small Area XPS In the range down to 120 µm, the area for analysis may be defined by the size of the X-ray spot (i.e. it is source-defined small area XPS). The relative merits of this approach are described in the application note AN31033 Below 120 µm the analysis area can be defined by the irises in the lens (lens-defined small area analysis). This extends the lateral resolution for spectroscopy down to 20 µm. The Flood Electron Source An electron source co-axial with the analyzer input lens, is provided with the instrument. This is used for charge compensation when analyzing non-conducting samples using the monochromatic X-ray source. The co-axial geometry of this source is ideal for XPS charge compensation because shadowing effects are eliminated (any feature visible to the transfer lens is also visible to the electron source). This geometry is essential when the magnetic lens is being used. Ion Gun In common with the other Thermo Scientific surface analysis instruments ESCALAB 250 is fitted with the EX05 ion gun. This is described in more detail in application note AN Digital Control All of the major analytical functions of ESCALAB 250 are controlled from the Windows-based Avantage data system. The Sample Sample Size ESCALAB 250 instruments have a fully motorized sample manipulator with 4 axes of movement. A fifth axis, rotation, is available as an option. The range of motion is 50 mm in X, 20 mm in Y and 12 mm in Z, each with 5 µm resolution. If fitted, the motorized rotation is continuous. Some of the sample holders available for ESCALAB 250 are shown in Figure 4. These include sample holders for heating and cooling the sample. Sample Alignment All axes of movement on the sample stage are controlled by the Avantage data system. As can be seen in Figure 1, there is a zoom microscope fitted to the instrument and aligned with the analysis position. The field of view from this optical system is in the range of 500 µm to 3 mm. The analysis position is accurately aligned with the cross hairs at the center of the graticule when the image is in focus. To align a feature for analysis it must be visible at the center of the graticule. This can be done by moving the sample using the tracker ball or pointing to a position on the optical image using the mouse pointer. Alternatively, the fast parallel imaging on ESCALAB 250 can be used to align the sample in real time, either using an XPS peak or using electrons elastically backscattered from an electron gun. Vacuum System The analysis chamber is constructed from 5 mm thick mumetal to maximize the efficiency of the magnetic shielding. The analysis chamber on all Thermo Scientific XPS instruments are constructed this way. The use of internal or external shields, the shielding method used by other XPS manufacturers, is not so effective. The chamber is pumped using both a turbomolecular pump and a titanium sublimation pump. This arrangement allows the analysis chamber to achieve a vacuum better than 5 x mbar. As well as ports for the energy analyzer and X-ray sources, the chamber is equipped with ports for Auger electron spectroscopy, UV lamp, ion guns, etc. This allows the system to be used as a true multi-technique facility. The entry lock is pumped using a turbo molecular pump backed by a rotary pump. This system is also used to provide differential pumping for the ion gun. The whole vacuum system can be baked in order to achieve the ultimate base pressure. Multi-technique Capability ESCALAB 250 is designed to accommodate other analytical techniques without compromise to the XPS performance. The power supplies for the lenses and analyzer are reversible and so, provided the standard ion gun is present, ISS (ion scattering spectroscopy) is always available. In addition, the following techniques are options on ESCALAB 250: XPS with non-monochromatic X-rays AES (Auger electron spectroscopy) UPS (Ultra-violet photoelectron spectroscopy) Figure 4: A selection of the sample holders available for use with ESCALAB 250 Preparation Chambers The base system is fitted with a simple air lock, which does not allow for any preparation facilities (see Figure 1). If preparation facilities are required then one of a range of PREPLOC chambers can be fitted. Figure 5, shows two examples of PREPLOC chambers.
4 Figure 7: A UHV Carousel Preparation chamber is available on ESCALAB 250. This chamber allows a number of radial preparation stages to be attached providing the basis for a very large range of in-vacuum preparation and analysis techniques Performance Figure 5: Two examples of PREPLOC chambers (combined preparation and entry lock chambers). The upper example is fitted with a parking stage and has ports for other accessories. The lower example is fitted with a heating stage, a gas cell and an ion gun. If a UHV preparation facility is required, an additional chamber can be supplied so that the preparation chamber is not exposed to the atmosphere when samples are introduced, Figure 6. Specifications The ESCALAB 250 specifications are defined under typical operating conditions. The microfocus monochromator is operated at the maximum power appropriate for each spot size. Count rate scaling is therefore not necessary. The sample is placed on the standard sample holder and at an angle such that its surface normal is parallel to the input lens axis. Sensitivity and spot size specifications are both achieved with the sample in this orientation. When comparing specifications from different instruments, it is important to ensure that the conditions under which the data are collected are the same. Spectroscopy The combination of an efficient transfer lens, a magnetic immersion lens, multi-channel detector and high intensity X-ray source ensures high sensitivity in all XPS applications. The detector is a multi-channeltron array providing maximum linear dynamic range Sensitivity Extremely high sensitivity and resolution are key features of ESCALAB 250. Figure 8 shows a survey spectrum of barium oxide, acquired in only 10 seconds. Note both the count rate and the extremely good signal-to-noise ratio. Figure 6: This ESCALAB 250 is equipped with both a PREPLOC and a UHV preparation chamber For more sophisticated sample preparation or analysis requirements, a UHV carousel preparation chamber is available, Figure 7. This consists of a large, disc-shaped chamber containing a series of radial ports. These ports can have further chambers added which can be designed to accommodate a wider variety of techniques. This chamber also has a fast entry air lock so that the preparation chamber is not exposed to atmospheric pressure during sample loading. Figure 8: XPS survey spectrum of barium oxide acquired in only 10 seconds using the monochromated X-ray source. Valence band and Fermi edge spectra may be acquired in just a few minutes.
5 Small Area XPS Small area XPS analysis, to < 20 µm, is fast and precise. Figure 9 shows the image from the zoom microscope of a set of aluminum bond pads with one pad centrally located in the analysis position. Figure 10 shows the Al 2p region of the XPS spectrum, acquired using the microfocus monochromator at a spot size of 120 µm. Figure 11: C 1s spectrum of polypropylene and flame treated polypropylene, the latter also showing the results of peak fitting Figure 9: Image of a set of aluminum bond pads on silicon, as viewed through the zoom microscope fitted to ESCALAB 250 A 60-point, 20 mm line scan was performed automatically under computer control across a defect area of an unmasked, polyester coated, paper sample. Spectra typical of the polyester coating are visible at most points, while spectra typical of the uncoated paper are seen from the defect area. This application illustrates the stability and effectiveness of the charge compensation system in unattended operation over a large analysis area. Figure 10: Al 2p spectrum from one of the bond pads in Figure 9. This was acquired using the monochromator set to 120 µm spot size. Note also, the excellent energy resolution in the elemental Al peaks which were fitted using asymmetric peak fitting. This spectrum was acquired using a low pass energy to obtain excellent energy resolution in the spectrum. The full width at half maximum (FWHM) of the elemental aluminum is < 0.3 ev. To obtain a good fit to the experimental data, asymmetric peak fitting was used. Asymmetric peak fitting is a feature of the Avantage data system. Insulating Samples Insulating samples are easily analyzed using the electron flood source for charge compensation. High-resolution polymer spectra enable C 1s envelopes to be fitted with confidence and precision, Figure 11. Figure 12: A C 1s spectrum line scan over a defect on a coated paper and a concentration profile of the two components present Energy Resolution For the Ag 3d 5/2 peak from clean silver, the guaranteed resolution is <0.45 ev, FWHM. The instrument can demonstrate better resolution on peaks whose natural line width is smaller. For example, Figure 13 shows a FWHM value of 0.29 ev (see application note AN31068). Figure 13: Part of the XPS spectrum from WSe 2 showing a FWHM of 0.29 ev
6 The Fermi edge data in Figure 14 was acquired on ESCALAB 250 using a microfocused monochromator with the sample at room temperature. The Fermi edge width is defined as the energy difference between 20% and 80% of the intensity measured below the Fermi edge. The relationship between the Fermi edge resolution and the intensity of XPS signals is discussed in detail in application note AN31007, which compares core level and valence band performance on silver. Figure 14: Fermi edge and Ag 4d region of a silver XPS spectrum measured at a pass energy of 3 ev. The measured Fermi edge width is <0.19 ev using the 20%-80% definition Parallel Imaging Parallel imaging produces rapid, high-resolution XPS images. ESCALAB 250 is the only instrument to use the same input lens and analyzer for both parallel imaging and spectroscopy. This eliminates the need to align two sets of electron optics and provides unambiguous location of features to be analyzed. The electron flood gun facilitates physical imaging of the sample, permitting rapid sample alignment and identification of the features to be analyzed. < 3 µm imaging resolution Variable field of view (120 µm - 8 mm) Imaging of large and small features Figure 15: Au 4f line scan over the image (above) showing a resolution of 2.8 µm Elemental Imaging In Figure 16 and Figure 17, two samples of metal brazing were analyzed. The overlays of the silver (green) and copper (red) show the differences between a good and poor quality braze. Image Resolution ESCALAB 250 produces an image resolution of < 3 µm. Figure 15 shows parallel images from a gold feature on glass which can be used to define small areas from which spectra can be acquired. The resolution of 2.8 µm is measured from the line scan. Figure 16: Image of a poor quality braze Figure 17: Image of a good quality braze
7 In the good quality braze, the silver more effectively wets the copper forming a strong joint. Where the wetting is less effective, the silver forms a single layer between the copper surfaces. Chemical State Imaging Figure 18 shows a set of images taken from the edge of a razor blade coated with a fluorocarbon lubricant. These C 1s images show the distribution of C-C and C-F on the blade edge and show that the lubricant distribution is not uniform and does not reach the edge of the blade. Indicating that the blade quality is poor. Azimuthal rotation of the sample during sputtering (optional on ESCALAB 250) minimizes the development of sputter induced topography. Figure 19 shows an example of a profile from ESCALAB 250. This is from a Ta/Si multi-layer on silicon. and shows excellent depth resolution (at a depth of 450 nm, the interface width is less than 6 nm). Figure 19: Depth profile from a Ta/Si multi-layer Figure 18: C 1s images from the edge of a razor blade. Top image is C-C, the middle image is C-F and the lower image is an overlay of the other two. Depth Profiling ESCALAB 250 provides concentration depth profiles from a small area with high sensitivity. Small sputtered areas ensure high etch rates and short acquisition times. Many features of ESCALAB 250 have been developed for optimum sputter profiling performance: The EX05 ion gun may be operated at high current for maximum profiling speed and at low energy (down to 100 ev) for optimum depth resolution Multi-sample sputter profile acquisition allows unattended operation for maximum sample throughput The data system includes Target Factor Analysis (TFA) and Linear and Non-Linear Least Squares Fitting (LLSF and NLLSF). These functions are fully integrated and ensure that the maximum possible chemical information is extracted from every profile. Angle Resolved XPS For very thin surface layers, ARXPS is the technique of choice. ESCALAB 250 features a variable aperture allowing the analyst to optimize the angular acceptance of the analyzer for each sample. Using the microfocused monochromator, photoelectrons from the whole X-ray spot are analyzed throughout the tilt range. This means: Peak resolution remains constant Peak position remains constant. Figure 20 shows the results of an ARXPS analysis of hafnium oxide on silicon dioxide on silicon. Two oxygen species are present, one at high concentration near to the surface, associated with Hf, and one in the bulk material, associated with silicon. Figure 20: Results of an ARXPS measurement on HfO 2 on SiO 2 on Si
8 Using the multi-overlayer thickness calculator, which is an integral part of the Avantage data system, the HfO 2 layer in this sample was found to be 1.6 nm thick and the SiO 2 was 0.58 nm. ISS All of the appropriate power supplies on ESCALAB 250 have reversible polarity. This allows the instrument to be used for ion scattering spectroscopy. The ISS technique is sensitive only to the top atomic layer of a sample. ISS measurements are made using the EX05 ion gun (fitted as standard) as the primary ion source. The spectrum, Figure 21, illustrates the high resolution that can be achieved using this technique on ESCALAB 250. Figure 22: Twin anode X-ray source (yellow) fitted to an ESCALAB 250. Figure 23: Comparison of copper XPS spectra acquired using an Al anode and a Mg anode Auger Electron Spectroscopy The optional Thermo Scientific FEG1000 field emission electron gun allows AES spectra and images to be acquired with a spatial resolution of <95 nm. The Avantage data system controls all aspects of the FEG 1000 operation. The gun is mounted in the optimum position on the chamber, such that its axis is orthogonal to the sample tilt axis, Figure 24. Figure 21: Part of the ISS spectrum from a phosphor bronze surface showing the separation of the two copper isotopes. This spectrum was acquired using argon ions. Multi-technique Capability In addition to ISS, which is present as standard on ESCALAB 250, the instrument was designed to accommodate a number of optional analytical facilities. Non-Monochromated XPS The sensitivity and resolution of the monochromator make it the first choice X-ray source for routine analysis. However, a non-monochromatic, twin anode source can be used when there are interferences with Auger transitions or when large areas need to be analyzed. This optional source is usually supplied with aluminum and magnesium anodes but other materials are available if required. Figure 22 shows the position of the X-ray source on the instrument while Figure 23 compares the XPS spectrum of copper acquired using the Al anode with that using the Mg anode. The shift of the Auger peaks on the binding energy scale is clear in this Figure. Figure 24: FEG1000 field emission electron gun (yellow) fitted to an ESCALAB 250 Figure 25 shows an SEM image of part of the surface of a copper nickel alloy. The SEM revealed the presence of surface corrosion pits and was used to define the points for the subsequent Auger analysis (P1 and P2 on Figure 25). Figure 25: SEM image of the surface of a copper-nickel alloy showing the presence of corrosion pits. The SEM was acquired using the FEG1000 field emission electron gun. Auger analysis, again using FEG1000, showed chlorine to be associated with these pits.
9 UPS An optional ultra-violet source fitted to ESCALAB 250 can be operated either as a source of He I or He II radiation. The position of the high-intensity UV lamp on the ESCALAB 250 chamber is shown in Figure 28. Figure 26: Spectra from the points labeled P1 and P2 on Figure 25 This was confirmed by the Auger maps shown in Figure 27. Figure 28: The high intensity UV lamp (yellow) fitted to an ESCALAB 250 The He I valence band spectrum shown below was acquired from a silver sample. Figure 27: Scanning Auger maps of Cl (red) and Cu (green) confirming the presence of chlorine in the corrosion pits Figure 29: He I spectrum of silver Figure 30: The silver Fermi edge spectrum acquired from silver at room temperature showing a resolution of 98 mev
10 Configuration ESCALAB 250 comprises a UHV mu-metal analysis chamber fitted with: Electron Analyzer Double-focusing full 180 spherical sector analyzer Magnetic and multi-element electrostatic input lenses for spectroscopy Multi-channel spectroscopic detector High resolution 2-D image detector Microfocused Monochromated XPS Facility 0.5 meter Rowland circle monochromator Digitally controlled microfocused electron gun and multi-position aluminum anode for extended life Two toroidal quartz crystals Flood Gun Charge compensation Electron imaging Digital control Ion Source Digital control Depth profiling Sample cleaning Secondary electron imaging for rapid and accurate alignment ISS Automated 4-axis Sample Manipulator Interfaced to the data system Avantage Data System Package Instrument control Data acquisition and processing Multi-sample, multi-point data acquisition Unattended data acquisition Options Twin Anode Non-monochromated XPS Facility Dual anode (MgKα /AlKα) X-ray source with linear motion drive Digital control Other anode coatings available UV Photoelectron Spectroscopy (UPS) Facility High intensity UV lamp with two-stage differential pumping Helium gas admission system with high precision leak valve Field Emission 95 nm Electron Gun for AES/SEM/SAM FEG1000 electron gun (Schottky type field emission source) Digital control Source ion pump SEM detector System vibration isolation Sample Rotation 5-axes sample manipulator including motorized azimuthal rotation for enhanced depth profiling resolution Full computer control of all 5 axes Sample Heating Sample holder with integral heater and thermocouple Sample Cooling Liquid nitrogen/chilled gas circuit PREPLOC Combined Fast Entry and Preparation Chamber Ports available for options including sample heating and cooling probe, sputter cleaning ion gun and sample scraper UHV Preparation Chamber with Separate Fast Entry Chamber Maintain preparation chamber at UHV during sample introduction In addition to these offices, Thermo Fisher Scientific maintains a network of representative organizations throughout the world. Africa Australia Austria Belgium Canada China Denmark Europe-Other France Germany India Italy Japan Latin America Middle East Netherlands South Africa Spain Sweden/Norway/ Finland Switzerland UK USA Thermo Fisher Scientific Inc. All rights reserved. Windows is a registered trademark of Microsoft Corporation. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. VG Systems Ltd. Trading as Thermo Fisher Scientific, East Grinstead, UK is ISO Certified. AN31063_E 04/08M Part of Thermo Fisher Scientific
Introduction of New Products
Field Emission Electron Microscope JEM-3100F For evaluation of materials in the fields of nanoscience and nanomaterials science, TEM is required to provide resolution and analytical capabilities that can
More informationAdd CLUE to your SEM. High-efficiency CL signal-collection. Designed for your SEM and application. Maintains original SEM functionality
Add CLUE to your SEM Designed for your SEM and application The CLUE family offers dedicated CL systems for imaging and spectroscopic analysis suitable for most SEMs. In addition, when combined with other
More informationNicolet Almega XR. High Performance Micro and Macro Dispersive Raman Analysis Systems
m o l e c u l a r s p e c t r o s c o p y Nicolet Almega XR High Performance Micro and Macro Dispersive Raman Analysis Systems Materials Research Identification/Verification Characterization Failure Analysis
More informationIntroduction... 3 Slits for AIR Operation... 4 Slits in Vacuum Vessels... 5 Slits for High Vacuum Operation... 6 Custom Slits... 7 Steel Slits...
Introduction... 3 Slits for AIR Operation... 4 Slits in Vacuum Vessels... 5 Slits for High Vacuum Operation... 6 Custom Slits... 7 Steel Slits... 10 Non-magnetic Options for Slits... 12 Slits with Passive
More informationThermo Scientific SPECTRONIC 200 Education
molecular spectroscopy Thermo Scientific SPECTRONIC 200 Education Part of Thermo Fisher Scientific Designed for the Teaching Laboratory Classroom Friendly Sample Compartment Whether you measure in 10 mm
More informationThermo Scientific SPECTRONIC 200
molecular spectroscopy Thermo Scientific SPECTRONIC 200 Part of Thermo Fisher Scientific The New Standard for Routine Measurements Robust, Multifunction Sample Compartment Whether you measure in 10 mm
More informationNanoSpective, Inc Progress Drive Suite 137 Orlando, Florida
TEM Techniques Summary The TEM is an analytical instrument in which a thin membrane (typically < 100nm) is placed in the path of an energetic and highly coherent beam of electrons. Typical operating voltages
More informationComponents for Surface Analysis PHOIBOS 150 WAL. Hemispherical Energy Analyzer with Wide Angle Lens
Customized Systems and Solutions Nanostructures and Thin Film Deposition Surface Analysis and Preparation Components Surface Science Applications Components for Surface Analysis PHOIBOS 150 WAL Hemispherical
More informationImproving the Collection Efficiency of Raman Scattering
PERFORMANCE Unparalleled signal-to-noise ratio with diffraction-limited spectral and imaging resolution Deep-cooled CCD with excelon sensor technology Aberration-free optical design for uniform high resolution
More informationSCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS
SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS Robert Edward Lee Electron Microscopy Center Department of Anatomy and Neurobiology Colorado State University P T R Prentice Hall, Englewood Cliffs,
More informationComponents for Surface Analysis PHOIBOS 150 NAP. Near Ambient Pressure Hemispherical Energy Analyzer
Customized Systems and Solutions Nanostructures and Thin Film Deposition Surface Analysis and Preparation Components Surface Science Applications Components for Surface Analysis PHOIBOS 150 NAP Near Ambient
More informationM4 TORNADO PLUS. Innovation with Integrity. Super Light Element Micro-XRF Spectrometer. Micro-XRF
M4 TORNADO PLUS Super Light Element Micro-XRF Spectrometer Innovation with Integrity Micro-XRF M4 TORNADO PLUS - A New Era in Micro-XRF M4 TORNADO PLUS is the world's first Micro-XRF spectrometer that
More informationThermo Scientific SPECTRONIC 200 Visible Spectrophotometer. The perfect. teaching instrument
Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer The perfect teaching instrument Designed for the Teaching Laboratory Thermo Scientific SPECTRONIC spectrophotometers have served as core analytical
More informationA Parallel Radial Mirror Energy Analyzer Attachment for the Scanning Electron Microscope
142 doi:10.1017/s1431927615013288 Microscopy Society of America 2015 A Parallel Radial Mirror Energy Analyzer Attachment for the Scanning Electron Microscope Kang Hao Cheong, Weiding Han, Anjam Khursheed
More informationThermo Scientific SPECTRONIC 200 Visible Spectrophotometer. The perfect tool. for routine measurements
Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer The perfect tool for routine measurements The Standard for Routine Measurements Thermo Scientific SPECTRONIC spectrophotometers have served as
More informationattocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G
APPLICATION NOTE M01 attocfm I for Surface Quality Inspection Confocal microscopes work by scanning a tiny light spot on a sample and by measuring the scattered light in the illuminated volume. First,
More informationPark NX-Hivac The world s most accurate and easy to use high vacuum AFM for failure analysis.
Park NX-Hivac The world s most accurate and easy to use high vacuum AFM for failure analysis www.parkafm.com Park NX-Hivac High vacuum scanning for failure analysis applications 4 x 07 / Cm3 Current (µa)
More informationThermo Scientific icap 7000 Plus Series ICP-OES: Innovative ICP-OES optical design
TECHNICAL NOTE 43333 Thermo Scientific icap 7000 Plus Series ICP-OES: Innovative ICP-OES optical design Keywords Optical design, Polychromator, Spectrometer Key Benefits The Thermo Scientific icap 7000
More informationHistorical. McPherson 15 Mount
McPherson 15 Mount Normal incidence designs include the McPherson 15 (classical 1.0 meter focal length) and modern NIM units. The latter features smaller included angles, longer focal lengths (e.g. 3,
More informationNOCOLOK Technical Brazing Center and Technical Service
NOCOLOK Technical Brazing Center and Technical Service The NOCOLOK Brazing Technical Center NOCOLOK flux brazing technology is the industry standard for brazing aluminum heat exchangers and other components.
More informationRound Robin: Quantitative Lateral Resolution of PHI XPS microprobes Quantum 2000 / Quantera SXM
Round Robin: Quantitative Lateral Resolution of PHI XPS microprobes Quantum 2000 / Quantera SXM U. SCHEITHAUER a, M. KOLB b, G. A.M. KIP c, E. NABURGH d, J.H.M. SNIJDERS d a 82008 Unterhaching, Germany
More informationMODULE I SCANNING ELECTRON MICROSCOPE (SEM)
MODULE I SCANNING ELECTRON MICROSCOPE (SEM) Scanning Electron Microscope (SEM) Initially, the plan of SEM was offered by H. Stintzing in 1927 (a German patent application). His suggested procedure was
More informationSystem Options. Magnetic Property Measurement System. AC Susceptibility. AC Susceptibility Specifications
System Options AC Susceptibility Magnetic Property Measurement System Many materials display dissipative mechanisms when exposed to an oscillating magnetic field, and their susceptibility is described
More informationSCIENTIFIC INSTRUMENT NEWS. Introduction. Design of the FlexSEM 1000
SCIENTIFIC INSTRUMENT NEWS 2017 Vol. 9 SEPTEMBER Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation The FlexSEM 1000: A Scanning Electron Microscope Specializing
More informationThe equipment used share any common features regardless of the! being measured. Electronic detection was not always available.
The equipment used share any common features regardless of the! being measured. Each will have a light source sample cell! selector We ll now look at various equipment types. Electronic detection was not
More informationObserving Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens. Compound Light Micros
PHARMACEUTICAL MICROBIOLOGY JIGAR SHAH INSTITUTE OF PHARMACY NIRMA UNIVERSITY Observing Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens.
More informationSchottky Emission VP FE-SEM
Schottky Emission VP FE-SEM Variable Pressure The Scanning Electron Microscope (SEM) has played an important role for many years for research and development of advanced materials in the leading edge of
More informationScanning Electron Microscopy Basics and Applications
Scanning Electron Microscopy Basics and Applications Dr. Julia Deuschle Stuttgart Center for Electron Microscopy MPI for Solid State Research Room: 1E15, phone: 0711/ 689-1193 email: j.deuschle@fkf.mpg.de
More informationDual-FL. World's Fastest Fluorometer. Measure absorbance spectra and fluorescence simultaneously FLUORESCENCE
Dual-FL World's Fastest Fluorometer Measure absorbance spectra and fluorescence simultaneously FLUORESCENCE 100 Times Faster Data Collection The only simultaneous absorbance and fluorescence system available
More informationApplications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region
Feature Article JY Division I nformation Optical Spectroscopy Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region Raymond Pini, Salvatore Atzeni Abstract Multichannel
More informationInstructions for the Experiment
Instructions for the Experiment Excitonic States in Atomically Thin Semiconductors 1. Introduction Alongside with electrical measurements, optical measurements are an indispensable tool for the study of
More informationThermo Scientific DXRxi Raman Imaging Microscope. Accelerate your work Visualize your answers
Thermo Scientific DXRxi Raman Imaging Microscope Accelerate your work Visualize your answers Raman imaging evolved Walk-up-and-run ease of use Visually driven image acquisition A microscopy-first approach
More informationSpotlight 150 and 200 FT-IR Microscopy Systems
S P E C I F I C A T I O N S Spotlight 150 and 200 FT-IR Microscopy Systems FT-IR Microscopy Spotlight 200 with Frontier FT-IR Spectrometer Introduction PerkinElmer Spotlight FT-IR Microscopy Systems are
More informationQuick and simple installation and no maintenance needed. 3 Times More affordable Than a normal SEM. Obtaining results in less than 4 minutes
INTRODUCTION We believe that every laboratory working in the field of nanotechnology needs an SEM, therefore we would like to introduce to you our IEM series of SEM. In short space of time, our device
More informationApproachable Raman Solutions The Shortest Path from Problem to Answer
Approachable Raman Solutions The Shortest Path from Problem to Answer Michael S. Bradley The world leader in serving science Thermo Scientific Raman Spectroscopy: Discover. Solve. Assure. Raman Spectroscopy
More informationJEM-F200. Multi-purpose Electron Microscope. Scientific / Metrology Instruments Multi-purpose Electron Microscope
Scientific / Metrology Instruments Multi-purpose Electron Microscope JEM-F200 Multi-purpose Electron Microscope JEM-F200/F2 is a multi-purpose electron microscope of the new generation to meet today's
More informationLow Voltage Electron Microscope
LVEM5 Low Voltage Electron Microscope Nanoscale from your benchtop LVEM5 Delong America DELONG INSTRUMENTS COMPACT BUT POWERFUL The LVEM5 is designed to excel across a broad range of applications in material
More informationScanning electron microscope
Scanning electron microscope 6 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Chamber and
More informationMS260i 1/4 M IMAGING SPECTROGRAPHS
MS260i 1/4 M IMAGING SPECTROGRAPHS ENTRANCE EXIT MS260i Spectrograph with 3 Track Fiber on input and InstaSpec IV CCD on output. Fig. 1 OPTICAL CONFIGURATION High resolution Up to three gratings, with
More informationTRITON Plus Thermal Ionization MS. Another step ahead. in TIMS. Earth and Planetary Sciences Nuclear Sciences
TRITON Plus Thermal Ionization MS Another step ahead in TIMS Earth and Planetary Sciences Nuclear Sciences Based on more than 35 years of experience in variable multicollector instrumentation, we have
More informationDual Magnetron Sputtering of Aluminum and Silicon Oxides for Low Temperature, High Rate Processing Abstract Background
Dual Magnetron Sputtering of Aluminum and Silicon Oxides for Low Temperature, High Rate Processing Christopher Merton and Scott Jones, 3M Corporate Research Lab, St. Paul, Minnesota, USA and Doug Pelleymounter,
More informationINTRODUCTION We believe that every laboratory working in the field of nanotechnology needs an SEM, therefore we would like to introduce to you our IEM
INTRODUCTION We believe that every laboratory working in the field of nanotechnology needs an SEM, therefore we would like to introduce to you our IEM series of SEM. In short space of time, our device
More informationICP-MS. plasma 3. Multi-Collector ICP-MS.
ICP-MS plasma 3 Multi-Collector ICP-MS www.nu-ins.com plasma 3 is a third generation Multi Collector ICP Mass Spectrometer (MC-ICP-MS), designed to provide the best possible precision and accuracy for
More informationCOMPONENTS FOR SURFACE ANALYSIS. The State of the Art Energy Analyzer Series
Surface Analysis Technology Vacuum Components Surface Analysis System Software Computer Technology COMPONENTS FOR SURFACE ANALYSIS The State of the Art Energy Analyzer Series PHOIBOS 100/150 Superior performance
More information2.Components of an electron microscope. a) vacuum systems, b) electron guns, c) electron optics, d) detectors. Marco Cantoni 021/
2.Components of an electron microscope a) vacuum systems, b) electron guns, c) electron optics, d) detectors, 021/693.48.16 Centre Interdisciplinaire de Microscopie Electronique CIME Summary Electron propagation
More informationMaterial analysis by infrared mapping: A case study using a multilayer
Material analysis by infrared mapping: A case study using a multilayer paint sample Application Note Author Dr. Jonah Kirkwood, Dr. John Wilson and Dr. Mustafa Kansiz Agilent Technologies, Inc. Introduction
More informationattosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G
APPLICATION NOTE M06 attosnom I: Topography and Force Images Scanning near-field optical microscopy is the outstanding technique to simultaneously measure the topography and the optical contrast of a sample.
More informationOriel MS260i TM 1/4 m Imaging Spectrograph
Oriel MS260i TM 1/4 m Imaging Spectrograph MS260i Spectrograph with 3 Track Fiber on input and InstaSpec CCD on output. The MS260i 1 4 m Imaging Spectrographs are economical, fully automated, multi-grating
More informationDevelopment of a Low Cost, Low Power, Miniature Sector Mass Spectrometer with IonCCD Detection
Leidos Proprietary Development of a Low Cost, Low Power, Miniature Sector Mass Spectrometer with IonCCD Detection Leidos; MIT MTL; Northeastern University; OI Analytical Development Team Noah Christian,
More informationMeasurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation
238 Hitachi Review Vol. 65 (2016), No. 7 Featured Articles Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation AFM5500M Scanning Probe Microscope Satoshi Hasumura
More informationThe SIRAD irradiation facility at the INFN - Legnaro National Laboratory
The SIRAD irradiation facility at the INFN - Legnaro National Laboratory I. Introduction 2 The INFN - Legnaro National Laboratory (LNL) SIRAD beamline http://www.lnl.infn.it 3 What is SIRAD? SIRAD is the
More informationLYNXEYE XE. Innovation with Integrity. High-Resolution Energy-Dispersive Detector for 0D, 1D, and 2D Diffraction XRD
High-Resolution Energy-Dispersive Detector for 0D, 1D, and 2D Diffraction The is the first energy dispersive 0D, 1D, and 2D detector operating at room temperature for ultra fast X-ray diffraction measurements.
More informationCompositional depth profile analysis of coatings on hard disks by X-ray photoelectron spectroscopy and imaging
Surface and Coatings Technology 176 (2003) 93 102 Compositional depth profile analysis of coatings on hard disks by X-ray photoelectron spectroscopy and imaging Jianxia Gao*, Erjia Liu, David Lee Butler,
More informationProcess Control Calibration Made Easy with Agilent U1401A
Process Control Calibration Made Easy with Agilent U1401A Application Note This application note explains how the Agilent U1401A with simultaneous source and measure functions eases technicians calibration
More informationDiamond X-ray Rocking Curve and Topograph Measurements at CHESS
Diamond X-ray Rocking Curve and Topograph Measurements at CHESS G. Yang 1, R.T. Jones 2, F. Klein 3 1 Department of Physics and Astronomy, University of Glasgow, Glasgow, UK G12 8QQ. 2 University of Connecticut
More informationMicroscopic Structures
Microscopic Structures Image Analysis Metal, 3D Image (Red-Green) The microscopic methods range from dark field / bright field microscopy through polarisation- and inverse microscopy to techniques like
More informationMEASUREMENT APPLICATION GUIDE OUTER/INNER
MEASUREMENT APPLICATION GUIDE OUTER/INNER DIAMETER Measurement I N D E X y Selection Guide P.2 y Measurement Principle P.3 y P.4 y X and Y Axes Synchronous Outer Diameter Measurement P.5 y of a Large Diameter
More informationIntroduction to Scanning Electron Microscopy
Introduction to Scanning Electron Microscopy By: Brandon Cheney Ant s Leg Integrated Circuit Nano-composite This document was created as part of a Senior Project in the Materials Engineering Department
More informationMicro Precision Apertures
National Aperture, Inc. Micro Precision Apertures Product Guide Precision apertures, targets and patterns, including round, slit, square, etc., for universal and high power applications National Aperture,
More informationScanning Electron Microscopy. EMSE-515 F. Ernst
Scanning Electron Microscopy EMSE-515 F. Ernst 1 2 Scanning Electron Microscopy Max Knoll Manfred von Ardenne Manfred von Ardenne Principle of Scanning Electron Microscopy 3 Principle of Scanning Electron
More informationA Penning Trap for Precision Spectroscopy of Highly Charged Ions at HITRAP. Jörg Krämer University of Mainz
A Penning Trap for Precision Spectroscopy of Highly Charged Ions at HITRAP University of Mainz Experimental Goal Precise measurement of the hyperfine splitting in highly charged ions (HCI) as a test of
More informationStarBright XLT Optical Coatings
StarBright XLT Optical Coatings StarBright XLT is Celestron s revolutionary optical coating system that outperforms any other coating in the commercial telescope market. Our most popular Schmidt-Cassegrain
More informationObservational Astronomy
Observational Astronomy Instruments The telescope- instruments combination forms a tightly coupled system: Telescope = collecting photons and forming an image Instruments = registering and analyzing the
More informationScanning electron microscope
Scanning electron microscope 5 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Detectors
More informationMajor Fabrication Steps in MOS Process Flow
Major Fabrication Steps in MOS Process Flow UV light Mask oxygen Silicon dioxide photoresist exposed photoresist oxide Silicon substrate Oxidation (Field oxide) Photoresist Coating Mask-Wafer Alignment
More informationA Laser-Based Thin-Film Growth Monitor
TECHNOLOGY by Charles Taylor, Darryl Barlett, Eric Chason, and Jerry Floro A Laser-Based Thin-Film Growth Monitor The Multi-beam Optical Sensor (MOS) was developed jointly by k-space Associates (Ann Arbor,
More informationAutomated Spectrophotometric Spatial Profiling of Coated Optical Wafers
Automated Spectrophotometric Spatial Profiling of Coated Optical Wafers Application note Materials testing and research Authors Travis Burt Fabian Zieschang Agilent Technologies, Inc. Parts of this work
More informationSupplementary Materials for
advances.sciencemag.org/cgi/content/full/3/10/eaao4204/dc1 Supplementary Materials for Enhanced mobility CsPbI3 quantum dot arrays for record-efficiency, high-voltage photovoltaic cells Erin M. Sanehira,
More informationSTEM Spectrum Imaging Tutorial
STEM Spectrum Imaging Tutorial Gatan, Inc. 5933 Coronado Lane, Pleasanton, CA 94588 Tel: (925) 463-0200 Fax: (925) 463-0204 April 2001 Contents 1 Introduction 1.1 What is Spectrum Imaging? 2 Hardware 3
More informationPhotoresist erosion studied in an inductively coupled plasma reactor employing CHF 3
Photoresist erosion studied in an inductively coupled plasma reactor employing CHF 3 M. F. Doemling, N. R. Rueger, and G. S. Oehrlein a) Department of Physics, University at Albany, State University of
More informationRoman Pots. Marco Oriunno SLAC, PPA. M.Oriunno, SLAC
Roman Pots Marco Oriunno SLAC, PPA The Roman Pot technique 1. The Roman Pot, an historically successful technique for near beam physics: ISR, SPS, TEVATRON, RICH, DESY 2. A CERN in-house technology: ISR,
More informationSection 2: Lithography. Jaeger Chapter 2. EE143 Ali Javey Slide 5-1
Section 2: Lithography Jaeger Chapter 2 EE143 Ali Javey Slide 5-1 The lithographic process EE143 Ali Javey Slide 5-2 Photolithographic Process (a) (b) (c) (d) (e) (f) (g) Substrate covered with silicon
More informationSection 2: Lithography. Jaeger Chapter 2 Litho Reader. The lithographic process
Section 2: Lithography Jaeger Chapter 2 Litho Reader The lithographic process Photolithographic Process (a) (b) (c) (d) (e) (f) (g) Substrate covered with silicon dioxide barrier layer Positive photoresist
More informationLYNXEYE XE-T. < 380 ev. Innovation with Integrity. Energy. Resolution. High-Resolution Position Sensitive Detector with Superb Energy Resolution XRD
Energy < 380 ev Resolution High-Resolution Position Sensitive Detector with Superb Energy Resolution The is the next generation "Compound Silicon Strip" detector with superb energy resolution for ultrafast
More informationPerformance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation
Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Rotation By: Michael Case and Roy Grayzel, Acton Research Corporation Introduction The majority of modern spectrographs and scanning
More informationSection 2: Lithography. Jaeger Chapter 2 Litho Reader. EE143 Ali Javey Slide 5-1
Section 2: Lithography Jaeger Chapter 2 Litho Reader EE143 Ali Javey Slide 5-1 The lithographic process EE143 Ali Javey Slide 5-2 Photolithographic Process (a) (b) (c) (d) (e) (f) (g) Substrate covered
More informationHollow Cathode Lamps
Hollow Cathode Lamps Competitive Comparison The hollow cathode lamp is a discharge lamp designed for use as a spectral line source with atomic absorption (AA) spectrometers. A single or multi-element hollow
More informationMicro-manipulated Cryogenic & Vacuum Probe Systems
Janis micro-manipulated probe stations are designed for non-destructive electrical testing using DC, RF, and fiber-optic probes. They are useful in a variety of fields including semiconductors, MEMS, superconductivity,
More informationFabrication of Probes for High Resolution Optical Microscopy
Fabrication of Probes for High Resolution Optical Microscopy Physics 564 Applied Optics Professor Andrès La Rosa David Logan May 27, 2010 Abstract Near Field Scanning Optical Microscopy (NSOM) is a technique
More informationCONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application
CONFIGURING Your Spectroscopy System For PEAK PERFORMANCE A guide to selecting the best Spectrometers, s, and s for your application Spectral Measurement System Spectral Measurement System Spectrograph
More informationChemical Imaging. Whiskbroom Imaging. Staring Imaging. Pushbroom Imaging. Whiskbroom. Staring. Pushbroom
Chemical Imaging Whiskbroom Chemical Imaging (CI) combines different technologies like optical microscopy, digital imaging and molecular spectroscopy in combination with multivariate data analysis methods.
More informationQuantum Efficiency Measurement System with Internal Quantum Efficiency Upgrade
Quantum Efficiency Measurement System with Internal Quantum Efficiency Upgrade QE / IPCE SYSTEM Upgraded with Advanced Features Includes IV Testing, Spectral Response, Quantum Efficiency System/ IPCE System
More informationUVISEL. Spectroscopic Phase Modulated Ellipsometer. The Ideal Tool for Thin Film and Material Characterization
UVISEL Spectroscopic Phase Modulated Ellipsometer The Ideal Tool for Thin Film and Material Characterization High Precision Research Spectroscopic Ellipsometer The UVISEL ellipsometer offers the best combination
More informationLVEM 25. Low Voltage Electron Microscope Fast Compact Powerful.... your way to electron microscopy
LVEM 25 Low Voltage Electron Microscope Fast Compact Powerful... your way to electron microscopy INTRODUCING THE LVEM 25 High Contrast & High Resolution Unmatched contrast of biologic and light material
More informationANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 325 ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER ABSTRACT William Chang, Jonathan Kerner, and Edward
More information190 nm (6.5 ev) Laser-ARPES. Laser ARPES using a Tunable ps UV Source
190 nm (6. ev) Laser-ARPES Laser ARPES using a Tunable ps UV Source 1 190 nm (6. ev) UV Source Laser ARPES/ARUPS with a High-Power UV Source Tunable deep-uv pulses from 190... 21 nm (6.....77 ev) are available
More informationFirst Results from a Retarding Field Angle-Resolved Analyzer
First Results from a Retarding Field Angle-Resolved Analyzer J. Antel Jr., G. R. Harp Department of Physics and Astronomy Ohio University Athens, OH 45701 USA A LEED-like display analyzer is discussed.
More information835 Vacuum Quality Monitor (VQM TM ) 2012 Brooks Automation, Inc. Proprietary Information
835 Vacuum Quality Monitor (VQM TM ) 2012 Brooks Automation, Inc. Proprietary Information The Revolutionary New Vacuum Quality Monitor World s Fastest, Accurate at low mass Single Gas Calibration Small
More informationSystem Options. Magnetic Property Measurement System. AC Susceptibility. AC Susceptibility Specifications
System Options AC Susceptibility Magnetic Property Measurement System Many materials display dissipative mechanisms when exposed to an oscillating magnetic field, and their susceptibility is described
More informationSäntis 300 Full wafer cathodoluminescence control up to 300 mm diameter
Säntis 300 Full wafer cathodoluminescence control up to 300 mm diameter Overview The Säntis 300 system has been designed for fully automated control of 150, 200 and 300 mm wafers. Attolight s Quantitative
More informationLow Voltage Electron Microscope
LVEM 25 Low Voltage Electron Microscope fast compact powerful Delong America FAST, COMPACT AND POWERFUL The LVEM 25 offers a high-contrast, high-throughput, and compact solution with nanometer resolutions.
More informationLVEM 25. Low Voltage Electron Mictoscope. fast compact powerful
LVEM 25 Low Voltage Electron Mictoscope fast compact powerful FAST, COMPACT AND POWERFUL The LVEM 25 offers a high-contrast, high-throughput, and compact solution with nanometer resolutions. All the benefits
More informationSpectroscopy Lab 2. Reading Your text books. Look under spectra, spectrometer, diffraction.
1 Spectroscopy Lab 2 Reading Your text books. Look under spectra, spectrometer, diffraction. Consult Sargent Welch Spectrum Charts on wall of lab. Note that only the most prominent wavelengths are displayed
More informationThe only simultaneous absorbance and f uorescence system for water quality analysis! Aqualog
The only simultaneous absorbance and fluorescence system for water quality analysis! Aqualog CDOM measurements made easy. The only simultaneous absorbance and fluorescence system for water quality analysis!
More informationAqualog. Water Quality Measurements Made Easy PARTICLE CHARACTERIZATION ELEMENTAL ANALYSIS FLUORESCENCE
Aqualog Water Quality Measurements Made Easy ELEMENTAL ANALYSIS FLUORESCENCE GRATINGS & OEM SPECTROMETERS OPTICAL COMPONENTS PARTICLE CHARACTERIZATION RAMAN SPECTROSCOPIC ELLIPSOMETRY SPR IMAGING Water
More informationPicoMaster 100. Unprecedented finesse in creating 3D micro structures. UV direct laser writer for maskless lithography
UV direct laser writer for maskless lithography Unprecedented finesse in creating 3D micro structures Highest resolution in the market utilizing a 405 nm diode laser Structures as small as 300 nm 375 nm
More informationNano Beam Position Monitor
Introduction Transparent X-ray beam monitoring and imaging is a new enabling technology that will become the gold standard tool for beam characterisation at synchrotron radiation facilities. It allows
More informationNEPTUNE Plus Multicollector ICPMS. Another step ahead. in MC-ICP-MS. Earth and Planetary Sciences Nuclear Sciences
NEPTNE Plus Multicollector ICPMS Another step ahead in MC-ICP-MS Earth and Planetary Sciences Nuclear Sciences Based on more than 35 years of experience in variable multicollector instrumentation, we have
More informationFull-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right
Quick Guide to Operating FEI Titan Themis G2 200 (S)TEM: TEM mode Susheng Tan Nanoscale Fabrication and Characterization Facility, University of Pittsburgh Office: M104/B01 Benedum Hall, 412-383-5978,
More information