1 Leica TCS SP2 User Manual 1.1 Markus Friedrich and James Lee Department of Biological Sciences Wayne State University Confocal Manual 1

2 Helpful hints The confocal room is cold. Bring a sweater. Confocaling is like playing an instrument. It takes a while until you consistently generate beauty. Be patient while learning. Otherwise, it may get expensive. Confocal Manual 2

3 Important rules If you are having a problem don t start playing around. It may get expensive. Call James Lee or Markus Friedrich. Never exchange objectives. This would delete important microscope software settings. If you need to add on additional objectives, let James Lee know. Put your stuff on the counter in the back of the room and not on the microsocope table to prevent misalignment of the potentially vibration sensitive dichroics. Sample Preparation Use of coverslips with 170ul (+/-10um) thickness is recommended. Make sure nail polish sealing has dried for at least one day. Oil immersion: Use Leica oil only! Confocal Manual 3

4 Operating the Leica DM IRB This is an electronically operated scope. Take the time for making yourself familiar. Otherwise you risk damaging the scope or your samples. The part identification numbers used below refer to the DM IRBE description on page 11 of the Leica DM IRBE manual. Starting up the microscope It is important to start up the microscope before you start up the confocal system. The confocal system needs to communicate with the already running microscope software during startup. You may use the epifluorescence addition on the microscope to examine your samples before confocaling. In this case, you need start up the mercury bulb first (box underneath the microscope table). Note however that it is not necessary to have the mercury bulb running while confocaling. Each mercury bulb costs $ We therefore suggest that you either leave the mercury off or if you need to use it for prior on site evalutation of your specimens turn off the mercury bulb before you start confocaling. Leace the mercury bulb only on if there is a reason to double-check your samples during confocaling. If you turned off the mercury bulb you should wait a minimum of 20 minutes before reigniting it. Start up the scope using the main switch (7) on the left side. The LCD panel on the front of the scope body will first idnicate the initializiation process. One finished you will see a panel shown on the next page: Objective Focus position Focus step lower/upper focus limit Confocal Manual 4

5 Prizm ring Dry or Immersion Make sure the LCD panel shows the 5x objective in position. If this is not the case change to the 5x by operating the objective wheel which is done using the push buttons on the right side of the instrument. Pushing the lower button gets you to the next lower objective. Pushing the upper button gets you to the objective with the next higher mag. Note: Although stated in the manual the current version of the scope does not discriminate between Immersion and Dry objectives. Note: the objectives are parafocal. That means that each objective will automatically be set at the focus level of the previous objective. You may need to be careful when working with thick specimens and changing to high mag objectives with lower working distance. In this case parafocality may mean squishing your specimen. After positioning the 5x objective bend the upper microscope neck backwards. This opens up plenty of room to mount your sample on the slide holder. Confocal Manual 5

6 Place your slide with the coverslip facing down. Note again: Never mount slides sealed with finger nail polish where the nail polish is still fresh. Nail polish preparations need a least one day to dry! Left side: Set filter wheel (11) to position 4 Push in UV light block level (behind filter wheel) Right side: Make sure the switch knob on the right side of the transmitted light field diaphragm (27) is set to Vis Turn tube lens module wheel (22) to position 1x Push in path shifter changer, which is close to the tube lens module wheel. Turn the brightness adjustment wheel in your direction to start up the halogen illumination. You should see your sample now looking through the eyepieces. Focus your sample using the either the focussing drive (10) or the push buttons behind the focussing wheel on the left side of the instrument (up is up and down is down). Note: You can regulate the focusing speed of the push buttons by pushing the STEP key on the manual bord. The focussing stepsize is indicated on the LCD screen. S0 = 0.05um S1 = 0.1um S2 = 0.7um S3 = 1.5um Once you located your tissue, switch to 10x objective After switching to 10x mag you should now set an upper focus position. This will minimize the risk that you damage your sample later on by focussing to deep into the tissue. Confocal Manual 6

7 You can focus by either using the push buttons on the left side of the scope. Upper button moves objective up and thus the focal plane deeper into the tissue. Lower button moves objective down. Focus at the deepest level necessary to investigate you sample. Push the image key for longer than 1 sec. This deletes the previously stored threshold. Push again for 1 sec. This stores your new threshold. Confocal Manual 7

8 Koehler illumination For optimal illumation of your sample you need to do the Koehler adjustment. Move specimen out of the light path. Close transmitted light field diaphragm (27) by turning holder to the left. Look through the eyepieces and focus the borders of the transmitted light field diaphragm by changing the position of the condenser unit (7 and 8) with the large wheel on the right of it. Center the transmitted light field using the turning knobs on the front of the condenser unit (7 and 8). Open transmitted light field diaphragm (27) until the transmitted light field diaphragm margins disappear from the visual field. Confocal Manual 8

9 Oil immersion Note: Use Leica oil only! Works best for these objectives. Push the Learn key once to position your objective in 60degree position which makes it best accessible. Add a little aliquot of Leica oil to the lens with the plastic tip, which is linked to the lid of the oil bottle. Be careful not to touch the lens with the plastic tip. Don t use too much oil. It unnecessarily contaminates the objective periphery and increases the tendency of the oil to move away from the lens. Push the Learn key again once. This will bring the objective back to its position underneath the slide. Cleaning the objective after oil immersion use: Note that you do not need to remove the oil completely from the objective lens after you are done. Leica claims that it is not necessary (provided the environment is reasonably dust free). Simply clean off excess oil from the metal periphery of the objective using a clean sheet of lens paper. Water immersion Same as water immersion exept that you simply have to carefully drain off the water from the lens after use. Confocal Manual 9

10 DIC optics Push in Analyser (21) Chooser correct ICT prism (12) with wheel on the left side. Choice of the suitable objective prism is confirmed on LCD screen. If wrong filter is chosen this value keeps on flickering on the screen. Confocal Manual 10

11 Switching from DIC to epifluorescence Close transmitted light using brightness adjustment wheel (8). Pull out Analyser (21) Pull out UV light block level (behind filter wheel). You should see your sample after choosing the adequate filter with the fluorescence modul filter wheel. Confocal Manual 11

12 Getting the scope ready for confocaling Close transmitted light using brightness adjustment wheel (8). Left side: Set filter wheel to position 4 Pull out incident light path switch rod which is behind filter change wheel (11). Make sure Analyzer (21) is pulled out. Right side: Turn right filter wheel to UV confocal lens Pull out path shifter Switch upper left switch from Vis to Scan You are now read to receive signal in the confocal Confocal Manual 12

13 Starting up the confocal Start up the PC using red switch Log in. Every user has his own account. To start with your last name is username and password. You can change your password whenever you want. Here is an example: Username: Password: family family Push the Scanner switch Push the Laser switch and wait for 30 sec You can now start the lasers you need. Attention: You should only start those lasers, which you actually need for your dyes. If you are not sure about the laser required for your dye see emission spectra table in the appendix (xxx). Argon is for blue dyes (FITC) (458, 475, 488, 514nm and transmitted light channel) HeNe green is for green dyes (TRITC) (543nm) HeNe red is for red dyes (Cy5) (633nm) HeNe Lasers can be simply switched on Argon Laser needs more attention: Make sure that the Argon output knob is at lowest level (full turn counterclockwise). Put key in start position. It will automatically switch back to running position. You are now ready to start the Leica Confocal SoftwareYou are now ready to confocal. Confocal Manual 13

14 Confocaling This chapter covers the very basics for getting an image for the typical single or multilabelled specimens (GFP, FITC, TRITC, Cy5 or any combination thereof). Any step further may be explored using the Leica User Manual (LUM) or the online help function from within the Leica Confocal Software. Start up Leica Confocal Software by double clicking the xxx file. The welcome window appears. Log in as personal. It takes about 30sec until the program environment appears. At the end of startup we currently get the message DM microscope error in command ( 50013). Code! and the advice to initialize DM IRBE. Press initialize. You will get immediately the message DM IRBE sucessfully initialized. The left monitor shows windows with function menus to operate the confocal, while right monitor shows the Viewer Window, which shows your imaging and image manipulation results (LUM page 48-71). Focus on the left monitor. In the upper panel you will find experiment design functions In the lower panel confocal parameter setting functions The basic operating unit is called experiment Experiment files can contain any number of images from the same experiment, i.e. imaging the same specimens under constant conditions plus accessory files containing the experiment parameters. If you click on NEW a new experiment is started. A new image panel appaers on the right screen. If you close this panel you delete the experiment. The following will lead you through the steps to get a first transmitted light image of your specimen on the screen. All the required function buttons are in the lower half of your visual field. Beam button: A pretty much selfexplanatory Beam Path Setting panel appaers. Choose your wavelenth settings. Activate transmission on the right chooser panel of preconfigured excitation/emission wavelength settings. Our You can leave the panel open or close it. Confocal Manual 14

15 Unexperienced users are advised to begin with activating TRANSMISSION. You can use this setting to generate a transmission light image of your specimen on the screen to make sure you have your specimen in an adequate optical plane before continuing with activating fluorescent dyes which in the beginning often looks very different from expected. Note the field view is limited at 5, 10 and 20x magnification. From there move to the Signal button: Open this window and leave it open. Lets you modulate contrast (gain) and black value (offset) of the confocal image. This is the main tool to optimize your image contrast. If you activated TRANSMISSION in the Beam button field you will now find that only the PMT Trans is active. Move the Gain level to the bottom. Move on to the Continuous scan button: Triggers continuous scanning. Each new image replaces the previous one. Useful to look through different layers of a thick specimen and for adjusting beam path, pinhole size, electronic zoom and phase (LUM page 36). Sstart confocaling by pushing the continuous scan button. Dim the light now. This will help you to see the image on the screen better. If you don t see your specimen right away, don t worry. Increasing the Gain level for the transmitted light channel (PMT Trans) will help you to visualize your specimen. You will be a little out of focus from the specimen. Use the Z Pos Knob or the microscope focus drive to focus into your specimen. You may have to redefine your upper focussing level on the scope. One you can see your specimen in transmitted light mode you can swith to fluorescent dye excitation settings. Bit depth button Choose between 8 bit and 12 bit gray value. 12 bit is not recognized by some image processing programs. Scan Mode button: Let you choose spatial, temporal and wavelength parameters of your scan (LUM 31-32). Leave at xyz for a start. Confocal Manual 15

16 Scan Speed button: Fast speed lets you work fast and minimize bleaching but results in background pixels. Slower speed increases signal to background ratio (LUM page 32-33). 800 and 1000 Mh automatically changes to Zoom to 2 and 4 respectively. Start with 400 Mh. Scan Format button: Choose scanning format, which means the number of points that are scanned per length unit along each scanning axis. The lower the format the faster the scan but you might also loose information. Higher formats lead to higher resolution but take longer and may increase bleaching (LUM page 30-31). Start with 256x256. Pinhole button: Let you set the diameter of the detection pinhole. This lets you modulate the thickness of your optical section. Click the airy 1 button to set the detection pinhole diameter to the optimal diameter corresponding to the objective used. Larger pinhole increases signal but also background. Smaller pinhole gives thinner optical section at the cost of signal intensity (LMU page 29). Single Scan button: Triggers one scan. Recommended if imaging bleach sensitive specimens. Confocal Manual 16

17 Series Scan button: Series Scan button can create an image series, either record serial layers of three dimensional image or generate images of different wavelength one by one. Before recording an image series, configure all required scan parameters using the Continuous Scan function to ensure optimal image quality. To get spatial image stacks: Select xyz or xzy scan mode. Continuously scan and use z pos knob of the control panel to determine the start point. Store the begin point using the Begin button in the Series Scan dialog window. Use the same step to set and store the end point of your image. Select the number of spatial sections using the Sections button. When you choose others, you can define your own desired distance of each layer and the computer will calculate the number of sections automatically. The more sections you take within a specified z-range the more comprehensively you cover you specimen. However, with the number of images also increases the time needed for the experiment, storage space required and memory needed for subsequent manipulation of the images. You therefore will want avoid oversampling. The number of sections needed to continuously cover a z-range depends on many parameters including magnification, wave length and pinhole size. Below are approximate step size values for sampling each section area 2-3times: Magnification 100x 63x 40x 20x 10x step size nm 300nm 400nm 500nm 1um When you have finished configuring these settings, press the Series Scan button to get the serial images. To view them, you can either press the Gall button in the overview dialog window to show them on the same screen or touch the Play button to see the animated effect. To record a wavelength series: Select a scan mode of either xyλ or xzλ, thus the λscan window will be available now. Confocal Manual 17

18 Define the wavelength at which you want to begin and save to the scan Begin Point. Define the wavelength at which you want to end and save to the scan End Point. Select the desired number of recordings between the begin and end points of the wavelength series. Open the Sequential Scan order on Beam Path Setting panel. Add your serial wavelength one by one and if you will use them again, save them. Now you are set to get the wavelength series images. Defining user-specific instrument parameter sets: Predefined instrument parameters may not satisfy you. Don t be frustrated. You can have your own settings. On Beam Path Setting panel, you can find the predefined instrument parameters for typical applications. Choose one that is most close to what you want to create and modify the appropriate instrument parameters until you get the satisfied image. Click the Save button to save your new setting into User files. You still can modify your own setting. Just open it, modify it and save again. Other useful buttons: Objective button: Let you switch objectives (see LUM page 25-26) (xxx discrimiates dry vs immersion?) Electronic Zoom button: Let s you zoom into your specimen at increased pixel resolution. Increases bleaching rate. (see LUM page 28) Time button: For setting time lapse series (LUM 34-35). Unidirectional/Bidirectional Scan button: You can double the scanning speed by activating bidirectional scanning (LUM page 45). You may have to adjust phase shift using the phase button. Confocal Manual 18

19 Phase button: Adjust phase shift when bidirectional scanning (LUM page 46). Scan field rotation button: Rotate scan field. Apply button: Create new experiment taking over settings from previous experiment (LUM 47). Average button: Increase signal background ration using different averaging options (LUM page 48). Saving your data This is covered in the LUM on pages Confocal Manual 19

20 Computer crash Desktop Turn of scanner Wait 10 sec Turn on scanner Wait 30 seconds This may reinitiate your computer. If still unsuccessful, press: contl alt delete Task manager End task or reset button Start program Confocal Manual 20

21 Shutting down the confocal The most important thing here is the Argon Laser. This Laser must be turned off while the cooling fan is still running. Therefore, take care to proceed step by step. Shut down Leica confocal software Log off system software Wait for message: It is safe now to shut down the computer Switch off computer on knob panel Switch off scanner on knob panel Switch off HeNe Lasers Minimize Argon Laser output on level wheel Make sure the laser fan is still running Turn off Argon Laser using the key Wait 20 min Switch off Argon Laser fan Confocal Manual 21

22 Shutting down the microscope Clean objectives Set objective in Learn position Excess oil may be removed with cotton swaps but needs not completely removed due to the new Leice formula Severe contaminations: get James Lee or Markus Friedrich. Brush of contamination with Sparkle soaked cotton swap by stroking only once in one direction. Make sure the lowest mag objective is in position in lowest gear Switch of microscope Switch of mercury bulb if used Confocal Manual 22

23 Important phone numbers Leica Brad Calloway (Leica) x5926 James Lee beeper Home Markus Friedrich office Lab Home Confocal Manual 23

24 Controling leak through Symultanous vs dual scanning Simultaneous Optimizing AOTF, gain and emission spectrum Saving/calling up parameters dual Optimizing AOTF, gain and emission spectrum Scanning speed difference? Saving/calling up parameterr Effcient data acquisition Good dye combinations Saving files At present the microscope focus is not parafocal with confocal focus which is a little higher. Confocal Manual 24

Things to check before start-up.

Things to check before start-up. Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your

More information

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left

More information

Quick Start Guide. Leica SP5 X

Quick Start Guide. Leica SP5 X Quick Start Guide Leica SP5 X Please note: Some of the information in this guide was taken from Leica Microsystems Leica TCS SP5 LAS AF Guide for New Users. This work is licensed under the Creative Commons

More information

LEICA TCS SP5 AOBS TANDEM USER MANUAL

LEICA TCS SP5 AOBS TANDEM USER MANUAL LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE

More information

Guide to Confocal 5. Starting session

Guide to Confocal 5. Starting session Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope

More information

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire Nikon Ti-E Microscope Manual Rightmire Hall Ohio State University Director: Tony Brown Rightmire 060 292-1205 brown.2302@osu.edu Facility Manager: Paula Monsma Rightmire 062 293-0939 292-1367 monsma.1@osu.edu

More information

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5. Smith College August 2010 Operating Checklist for using the Laser Scanning Confocal Microscope Leica TCS SP5. CONTENT, page no. Startup, 1 Initial set-up, 1 Software, 2 Microscope Specimen observation

More information

LSM 510 Training Notes

LSM 510 Training Notes LSM 510 Training Notes Turning on the system Turn on the arc lamp, found on the bench top left of the microscope. This supplies light for epifluorescence for viewing your samples through the microscope.

More information

TRAINING MANUAL. Olympus FV1000

TRAINING MANUAL. Olympus FV1000 TRAINING MANUAL Olympus FV1000 September 2014 TABLE OF CONTENTS A. Start-Up Procedure... 1 B. Visual Observation under the Microscope... 1 C. Image Acquisition... 4 A brief Overview of the Settings...

More information

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope Improved Optics More Lasers 405 diode 440 diode 488 Argon 515 Argon 559

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power

More information

Operating Instructions for Zeiss LSM 510

Operating Instructions for Zeiss LSM 510 Operating Instructions for Zeiss LSM 510 Location: GNL 6.312q (BSL3) Questions? Contact: Maxim Ivannikov, maivanni@utmb.edu 1 Attend A Complementary Training Before Using The Microscope All future users

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front

More information

LSM 510 Meta Training Notes

LSM 510 Meta Training Notes LSM 510 Meta Training Notes Turning on the system Turn on X-Cite power supply. This supplies light for epifluorescence for viewing your samples through the microscope. Turn on the remote control switch.

More information

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: Confocal imaging on the Leica TCS SP8 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: 4) Do not touch the microscope while the software is initializing. Choose your options: Turn

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Leica SP8 TCS Users Manual Follow the procedure for start up and log on as posted in the lab. Please log on with your account only and do not share your password with anyone. We track and confirm usage

More information

Leica Sp5 II Confocal User Guide

Leica Sp5 II Confocal User Guide Leica Sp5 II Confocal User Guide Turning on the Confocal System (instructions are posted in the room) 1. Turn on Laser Power Button 2. Turn Key to On position 3. Turn on Scanner Power Button 4. Turn on

More information

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling Zeiss LSM 510 Confocor III Training Notes Center for Cell Analysis & Modeling Confocor 3 Start Up Go to System Module Turn on Main Switch, System/ PC, and Components Switches Do you need the arc lamp?

More information

ZEISS LSM510META confocal manual

ZEISS LSM510META confocal manual ZEISS LSM510META confocal manual Switching on the system 1) Switch on the Remote Control button located on the table to the right of the microscope. This is the main switch for the whole system including

More information

Leica SPEII confocal microscope. Short Manual

Leica SPEII confocal microscope. Short Manual Leica SPEII confocal microscope Short Manual Switching ON sequence: 1. Turn on the Workstation under the bench (top, far right). 2. Turn on the Supply Unit - Laser box (big green switch first and then

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

Leica SP8 Resonant Confocal. Quick-Start Guide

Leica SP8 Resonant Confocal. Quick-Start Guide Leica SP8 Resonant Confocal Quick-Start Guide Contents Start-up Preparing for Imaging Part 1 On the scope Part 2 Software interface Part 3 Heat & CO2 incubation Part 4 Other hardware options Shut-down

More information

Cell Biology and Bioimaging Core

Cell Biology and Bioimaging Core Cell Biology and Bioimaging Core Leica TCS SP5 Operating Instructions Starting up the instrument 1. First, log in the log book located on the confocal desk. Include your name, your lab s PI, an account

More information

Microscopy from Carl Zeiss

Microscopy from Carl Zeiss Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path

More information

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL START THE SYSTEM... 2 START ZEN SOFTWARE... 3 SET THE TEMPERATURE AND THE CO2 CONTROLLERS... OBSERVATION AT OCULARS... 5 STATIF PRESENTATION... 6 ACQUIRE ONE

More information

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2)

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2) Brief manual how to start and close the Leica sp2 Confocal (TCS SP2 AOBS system mounted on a DM IRE2) A. Switching on hardware B. Acquiring and saving images C. Switching off the microscope D. Good working

More information

Zeiss 880 Training Notes Zen 2.3

Zeiss 880 Training Notes Zen 2.3 Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start

More information

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017. Nikon AZ100 Laser Scanning Macro Confocal Microscope Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin May 2017 Contents 1 Introduction 2 2 Hardware - Startup 2 3 Software/Operation 4 3.1 Multidimensional

More information

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Training Guide for Leica SP8 Confocal/Multiphoton Microscope Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence

More information

b. Turn the power switch and key to on position for blue laser.

b. Turn the power switch and key to on position for blue laser. OLYMPUS FLUOVIEW 300 CONFOCAL MICOSCOPE OPERATION PROCEDURE 1. Turn ON microscope in this order: 1) Turn on mercury lamp (Note: once the mercury lamp is turned off, DO NOT turn it back on for at least

More information

Using the Nikon TE2000 Inverted Microscope

Using the Nikon TE2000 Inverted Microscope Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core Using the Nikon TE2000 Inverted Microscope Fluorescence image acquisition using Scanalytic s IPLab software and the B&W

More information

Leica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager

Leica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager Leica TCS SL Confocal Training Neuroscience Imaging Facility The Ohio State University Rightmire Hall 614-292-1367 Staff Core Director Anthony Brown, Ph. D. 060 Rightmire Hall 614-292-1205 brown.2302@osu.edu

More information

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue

More information

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s LSM 5 MP, LSM 510 and LSM 510 META M i c r o s c o p y f r o m C a r l Z e i s s Quick Guide Laser Scanning Microscopes LSM Software ZEN 2007 August 2007 We make it visible. Contents Page Contents... 1

More information

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 CIAN LSM1 or LSM2 short instructions, version 1.4, September 2014 page 1 of 6 SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 Before starting To work with LSM1

More information

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010 Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS Zeiss LSM 510 M eta Confocal M icroscope OPERATING INSTRUCTIONS Starting the System: 1. Turn the black knob on the laser box one-quarter turn from Off to On. You will hear the laser cooling mechanisms

More information

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL MIF ZEISS LSM510 CONFOCAL USER PROTOCOL START-UP Turn on the Mercury Bulb Power Supply (if needed). Power-on the Control Box. Turn on the computer. Open the LSM 510 software. Choose Scan New Images and

More information

Microscope Confocal LSM510 META

Microscope Confocal LSM510 META Microscope Confocal LSM510 META Welcome to the Zeiss LSM 510 Meta Confocal tutorial. Before using the LSM 510 META, Log off any other computer that is open with your personal login. You will need to put

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and

More information

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 Leica DMI AF6000LX Table of contents START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 ACQUIRE MODULE 6 SETTING THE LIGHTPATH 6

More information

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,

More information

Practical work no. 3: Confocal Live Cell Microscopy

Practical work no. 3: Confocal Live Cell Microscopy Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside

More information

Nikon E800 Operating Instructions.

Nikon E800 Operating Instructions. Nikon E800 Operating Instructions. You can request electronic copies of this manual by contacting lshats@jhsph.edu Copies are also available on the JHU MMI Department web site. Please send your comments

More information

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL START-UP On the Switchbox, turn both black switches to the ON position. Wait for the microscope to boot up completely (watch the screen on the side of the microscope).

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,

More information

Basics of confocal imaging (part I)

Basics of confocal imaging (part I) Basics of confocal imaging (part I) Swiss Institute of Technology (EPFL) Faculty of Life Sciences Head of BIOIMAGING AND OPTICS BIOP arne.seitz@epfl.ch Lateral resolution BioImaging &Optics Platform Light

More information

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging Bi/BE 227 Winter 2016 Assignment #3 Adding the third dimension: 3D Confocal Imaging Schedule: Jan 20: Assignment Jan 20-Feb 8: Work on assignment Feb 10: Student PowerPoint presentations. Goals for this

More information

CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2)

CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2) Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2) 1) STARTING THE SYSTEM Abridged INSTRUCTIONS Switch on the mercury bulb

More information

Leica Confocal - 2. Instructions for Leica SP2 Confocal Equipped with Visible Laser Lines

Leica Confocal - 2. Instructions for Leica SP2 Confocal Equipped with Visible Laser Lines Leica Confocal - 2 Instructions for Leica SP2 Confocal Equipped with Visible Laser Lines Imaging Technology Group Revised as of 08/2011 by Dianwen Zhang (Email: zhangdw@illinois.edu) Revision History Authors

More information

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide Nikon C1si Spectral Laser Scanning Confocal Microscope User Guide Contents: C1Si Turn-On/ShutDown Procedures... 2 Overview... 4 Setup for epi-illumination to view through the eyepieces:... 5 Setup for

More information

Instructions for Making On-Line Reservations for Microscopes in NB11-204

Instructions for Making On-Line Reservations for Microscopes in NB11-204 Instructions for Making On-Line Reservations for Microscopes in NB11-204 1. Log into Mail using Mail.swmed.edu 2. Log in using your university id and password. 3. Click the Calendar Tab at the top right

More information

CMI STANDARD OPERATING PROCEDURE. Fluoview 300 laser scanning confocal microscope

CMI STANDARD OPERATING PROCEDURE. Fluoview 300 laser scanning confocal microscope CMI STANDARD OPERATING PROCEDURE Fluoview 300 laser scanning confocal microscope CMI documentid:sop001 CONTACT INFORMATION: Peter Owens: 091 494036 (office) Peter.owens@nuigalway.ie Kerry Thompson: 091

More information

Microscope Confocal Sp2 Upright.

Microscope Confocal Sp2 Upright. Microscope Confocal Sp2 Upright. Welcome to the Leica Sp2 Confocal Upright tutorial. Before using the Sp2 Invert, You will need to put down your name on the reservation system = http://svintranet.epfl.ch/index.php?optio

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Nikon SIM-E & A1-R System

Nikon SIM-E & A1-R System Nikon SIM-E & A1-R System USER GUIDE LSU Health Sciences Center Shreveport Research Core Facility June 01 2017 Chaowei Shang 1 Table of Content 1. Start Up the System... Page 3 Hardware and microscope

More information

LSM 510 META in Chang Gung University

LSM 510 META in Chang Gung University Content LSM 510 META in Chang ung University LSM 510 META 路 理 The features and applications of LSM 510 META 01-09 Introduction of the hardware 10-12 Fluorescence observation in conventional microscope

More information

Internal Medicine Imaging Core Emory University Department of Medicine

Internal Medicine Imaging Core Emory University Department of Medicine Internal Medicine Imaging Core Emory University Department of Medicine 1 OPERATION OF THE ZEISS LSM 510 META YOU MUST SIGN UP TO USE THE MICROSCOPE OR COMPUTER EVERY TIME NO EXCEPTIONS Before attempting

More information

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION Olympus VS120-L100 Slide Scanner Standard Operating Procedure Startup 1) Red power bar switch (behind monitor) 2) Computer 3) Login: UserVS120 account (no password) 4) Double click: WAIT FOR INITIALIZATION

More information

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide Matyas Molnar, Biovis 2016 Starting the microscpe 1. Check the microscope if everything looks clean and normal. If not, report it in the logbook.

More information

Contents. Introduction

Contents. Introduction Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path and lasers... 12 Scanning

More information

Zeiss Axio Imager.A1 manual

Zeiss Axio Imager.A1 manual Zeiss Axio Imager.A1 manual Power-up protocol 1. Mercury lamp 2. Power strip on shelf 3. Computer The Mercury lamp should always be first-on and last-off. This prevents any electrical surges caused by

More information

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide Operation 58. UltraVIEW VoX Getting Started Guide Overview This chapter outlines the basic methods used to operate the UltraVIEW VoX system. About other software Volocity places great demands on the computer

More information

Nikon A1Rsi Confocal Start-Up Sequence

Nikon A1Rsi Confocal Start-Up Sequence 1. Turn the key on the Nikon LUN-V Laser Launch. Nikon A1Rsi Confocal Start-Up Sequence 2. Press the button the left side of the A1Rsi Controller unit. 3. Turn on the power strip underneath the microscope.

More information

Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal

Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal Todays Goal: Introduce some additional functionalities of the Leica SP8 confocal HyD vs. PMT detectors Dye Assistant Scanning By

More information

BASICS OF CONFOCAL IMAGING (PART I)

BASICS OF CONFOCAL IMAGING (PART I) BASICS OF CONFOCAL IMAGING (PART I) INTERNAL COURSE 2012 LIGHT MICROSCOPY Lateral resolution Transmission Fluorescence d min 1.22 NA obj NA cond 0 0 rairy 0.61 NAobj Ernst Abbe Lord Rayleigh Depth of field

More information

Zeiss LSM 510 Multiphoton Confocal Microscope

Zeiss LSM 510 Multiphoton Confocal Microscope Zeiss LSM 510 Multiphoton Confocal Microscope User Guide LSU Health Sciences Center-Shreveport Research Core Facility Table of Contents 1 Safety... Page 3 2 Turn On the System... Page 4 3 Start Up the

More information

Nikon E800 Operating Instructions.

Nikon E800 Operating Instructions. Nikon E800 Operating Instructions. You can request electronic copies of this manual by contacting imaging@fhcrc.org. Copies are also available on the Scientific Imaging web site. Please send your comments

More information

LSM 800 Confocal Microscope Standard Operation Protocol

LSM 800 Confocal Microscope Standard Operation Protocol LSM 800 Confocal Microscope Standard Operation Protocol Turning on the system 1. Switch on the Main switch (labeled 1 and 2 ) mounted on the wall. 2. Turn the Laser Key (labeled 3 ) 90 clockwise for power

More information

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions Starting up: Schematic of Olympus BX-61. 1. Turn on Olympus microscope power box (left of microscope) with toggle switch on

More information

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope Nikon Eclipse Ti A1-A Confocal Operating Manual Start-up 1. Turn on Excite Fluorescent light power supply- metal halide. a. Cool down as for mercury bulb b. Wheel closed liquid light guide 2. Turn on power

More information

Zeiss LSM 510 Multiphoton Confocal Microscope

Zeiss LSM 510 Multiphoton Confocal Microscope Zeiss LSM 510 Multiphoton Confocal Microscope Quick Start User Guide LSU Health Sciences Research Core Facility Table of Contents 1 Safety... Page 3 2 Turn On the System... Page 4 3 Start Up the ZEN Software.

More information

Manual for BMS E1 eplan series, compound microscope

Manual for BMS E1 eplan series, compound microscope Manual for BMS E1 eplan series, compound microscope The compound microscope allows it to study, at cell level, structures of textures of botanical and zoological nature. (e.g. slides of roots, leaves and

More information

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the

More information

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol Nikon Eclipse Ti-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol Please sign on the log sheet before switching on system. Turn on system Turn on A only if confocal mode or laser

More information

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope Training Guide for Carl Zeiss LSM 510 META Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON Components and System/PC switches

More information

Title: Leica SP5 Confocal User Manual

Title: Leica SP5 Confocal User Manual Title: Leica SP5 Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG07 or 05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk Website:

More information

Nikon E800 Microscope. Operating Instructions

Nikon E800 Microscope. Operating Instructions Nikon E800 Microscope Operating Instructions B Watson 12/2005 Table of contents: 1. The Nikon E800 Microscope 2. Turning the system ON and OFF 3. Selecting the light path 4. Operating in transmitted light

More information

Widefield 1. Switching on

Widefield 1. Switching on Widefield 1 Switching on 1. Ignite DG5 lamp - must be switched on first (if previous user has switched off, wait 30 min before igniting) 2. Wait 5s and then turn on the main DG5 controller switch. 3. DG5

More information

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005 Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev Why use confocal microscopy? Principles of the laser scanning confocal microscope. Image resolution. Manipulating the

More information

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

3. are adherent cells (ie. cells in suspension are too far away from the coverslip) Before you begin, make sure your sample... 1. is seeded on #1.5 coverglass (thickness = 0.17) 2. is an aqueous solution (ie. fixed samples mounted on a slide will not work - not enough difference in refractive

More information

Last updated: May 2014 Y.DeGraaf

Last updated: May 2014 Y.DeGraaf FLINDERS MICROSCOPY BIOMEDICAL SERVICES AVAILABLE MICROSCOPES AND SPECIFICATIONS & INFORMATION REGARDING TRAINING FOR NEW USERS Last updated: May 2014 Y.DeGraaf If you have new staff or students (Honours/Masters

More information

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report Introduction and Background Two-photon microscopy is a type of fluorescence microscopy using two-photon excitation. It

More information

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator Topics - How to calibrate the LSM scanner - How to measure the PSF - How to clean the microscope - How to adjust the pinhole alignment - How to adjust the Collimator How to calibrate the LSM scanner The

More information

1 Co Localization and Working flow with the lsm700

1 Co Localization and Working flow with the lsm700 1 Co Localization and Working flow with the lsm700 Samples -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ BrDU with alexa 488. -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ no BrDU (but with

More information

The Zeiss AiryScan System, Confocal Four.

The Zeiss AiryScan System, Confocal Four. The Zeiss AiryScan System, Confocal Four. Overview. The Zeiss AiryScan module is a segmented, radially stacked GaASP detector and collector system designed to subsample the airy disk of a point emission

More information

Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018

Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018 Zeiss LSM880 Operating Instructions UTMB Optical Microscopy Core Jan. 16, 2018 1 1. Power up the microscope Sing the LOGBOOK Steps below will provide power to the computer and all of the microscope components.

More information

User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope

User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope This version: 7.24.14. Introduction The IBIF confocal microscope is made available on a fee-for-use-hour basis to all users who have been trained.

More information

Renishaw InVia Raman microscope

Renishaw InVia Raman microscope Laser Spectroscopy Labs Renishaw InVia Raman microscope Operation instructions 1. Turn On the power switch, system power switch is located towards the back of the system on the right hand side. Wait ~10

More information

Laboratory Introduction

Laboratory Introduction Laboratory Introduction There are two basic categories of microscopes: light microscopes and electron microscopes. Light, or optical, microscopes require light waves to provide the illumination while electron

More information

Confocal Raman Microscopy (WITec Alpha 300R)

Confocal Raman Microscopy (WITec Alpha 300R) Confocal Raman Microscopy (WITec Alpha 300R) Please refer to Witec Alpha300R Confocal Raman Microscope User Manual for the details of the operating procedure. Sample preparation 1. Attach your sample on

More information

Swept-Field User Guide

Swept-Field User Guide Swept-Field User Guide Note: for more details see the Prairie user manual at http://www.prairietechnologies.com/resources/software/prairieview.html Please report any problems to Julie Last (jalast@wisc.edu)

More information

Supplemental Method Information Zeiss LSM710

Supplemental Method Information Zeiss LSM710 Supplemental Method Information Zeiss LSM710 1 Under the Light Path window set up the confocal for imaging a green dye (Alexa488-EGFP). For example, set up the light path as shown here using the 488 nm

More information

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each lens with a 1 Airy unit pinhole setting. Many water lenses

More information

Confocal Application Letter No. 13. Sequential Scan for Leica TCS NT/SP systems

Confocal Application Letter No. 13. Sequential Scan for Leica TCS NT/SP systems Confocal Application Letter No. 13 Sequential Scan for Leica TCS NT/SP systems Leica Microsystems Heidelberg GmbH Im Neuenheimer Feld 518 D-69120 Heidelberg Telephone +49 6221 4148 0 Fax +49 6221 414833

More information

Microscopy Suite. The Multi-Photon Confocal Microscope. Instructions for Leica SP2 Microscope #1 By Jon Ekman & Karl Garsha. Start-up.

Microscopy Suite. The Multi-Photon Confocal Microscope. Instructions for Leica SP2 Microscope #1 By Jon Ekman & Karl Garsha. Start-up. Microscopy Suite The Multi-Photon Confocal Microscope Instructions for Leica SP2 Microscope #1 Use of the Multiphoton Laser Scanning Confocal Microscope. Basic operation instructions as of 05/04/2007 Start-up

More information