DOUBLE gated in vivo small animal cone beam micro

Size: px
Start display at page:

Download "DOUBLE gated in vivo small animal cone beam micro"

Transcription

1 Low Dose Phase Correlated Cone Beam Micro CT of Small Animals Stefan Sawall, Frank Bergner, Robert Lapp, Markus Mronz, Marek Karolczak, Andreas Hess, and Marc Kachelrieß, Member, IEEE I. INTRODUCTION DOUBLE gated in vivo small animal cone beam micro CT scans provide five dimensional information about the object: the three volume dimensions plus the temporal dimensions of the respiratory motion and the heart motion, respectively. Double gating is typically performed to separate respiratory from cardiac motion when imaging the animal s lung or heart [1], [2], [3], [4], [5]. We are aiming at significantly improving the image quality achievable in low dose single or double gated micro CT scans of small specimen. On the one hand we want to reduce streak artifacts that result from sparse angular sampling and on the other hand we want to reduce image noise resulting from the small amount of photons available in a given combination of respiratory and cardiac phase. A. Intrinsic Gating II. METHOD To correlate our reconstruction with the motion phases of the animal heart and lung we detect corresponding synchronization information directly from the rawdata (kymogram). In our case of a slowly rotating micro-ct scanner approaches that evaluate the center of mass of certain ROIs within each two dimensional projection image have been quite successful to detect the respiratory and cardiac motion [6], [7]. Fig. 1. Comparison between conventional phase correlated reconstruction (respiratory and cardiac gating) and our new iterative low dose phase correlated method of mouse 1. Values for image noise are shown in the sagittal slices. (C = 0 HU, W = 400 HU, Respiratory phase 0%, Cardiac phase 80%). B. Image Reconstruction Let f denote the object, X the x ray transform operator, and p be the projection data, such that p = Xf. Our standard image reconstruction is the Feldkamp algorithm which we denote with X and which results in the Std standard image f Std = X Stdp [8]. The standard Feldkamp algorithm is not phase correlated. We make use of it below to define the McKinnon Bates algorithm. To perform respiratory and cardiac correlated image reconstruction we use a phase correlated Feldkamp algorithm X PC that filters and backprojects only those projections that lie in the desired temporal window. The temporal window itself is defined by specifying the respiratory phase r and the cardiac phase c, both are values between 0 and 1 and count relative to one motion period, and by specifying the widths r and c of these temporal windows. The respiratory and cardiac phase correlated image is denoted as f PC = X PC p. Since only few projections fall into the desired temporal window, streak artifacts may occur unless a very large number of projections at very fine angular increments is acquired. The McKinnon Bates (MKB) algorithm can be used to address this issue [9], [10]. It works as follows. First, a standard reconstruction is performed to obtain a prior image. This prior image is blurry in those regions where motion is present, and it is of high image quality elsewhere. Then, a forward projection of the prior image is performed and subtracted from the measured rawdata. These subtracted data are then used for a phase correlated reconstruction which is added to the prior image. Mathematically: f MKB = f Std +X PC (p Xf Std). (1) To compensate for the loss in spatial resolution due to the additional forward and backprojection steps and to correctly compensate for the longitudinal truncation effects we rather

2 Fig. 2. The MKB algorithm is implemented by using complementary weights. A 20% wide phase weight is illustrated. Green weights would be applied to the measured data while red weights are applied to the forward projected data. The blending method is a weighted sum of measured rawdata and forward projected data followed by standard reconstruction. domain filters may be used as well. In this study, however, we restricted ourselves to considering Gaussian shaped filters. Since respiratory and cardiac gating yields five dimensional volumes f(x,y,z,r,c), with r denoting the respiratory and c denoting the cardiac phase, we can apply bilateral filtering in up to five dimensions. The corresponding domain filter parameters are denoted as σ x, σ y, σ z, σ r and σ c, respectively. We also apply the edge preserving filtration to the first prior volume (in x, y, and z), and to the second prior volume (in x, y, z, and r) using the same filter parameters as for the final five dimensional filtering. The final volume, obtained by bilateral filtering the MKB volume and the intermediate prior volumes, is the low dose phase correlated volume f LDPC = Bf MKB (6) which we want to compare against the conventional (and more simple) phase correlated volume f PC in the following. perform f MKB = X Std Xf Std +X PC (p Xf Std) (2) = (X Std X PC )Xf Std +X PC p (3) = X PC Xf Std +X PCp. (4) X The new operator PC denotes the complement of the phase correlated reconstruction operator such that the sum X PC +X PC is equal to the standard reconstruction X Std. Since our phase correlated reconstruction approach corresponds to nothing but a weighting of the projections prior to Feldkamp reconstruction the MKB algorithm can be realized by adequate weighted blending of the forward projected prior image Xf Std with the original rawdata p followed by another standard image reconstruction (figure 2). Since the respiratory motion dominates the cardiac motion we extend the MKB approach to a two step algorithm. In a first step we apply MKB only to respiratory gating and in the second step we use the respiratory gated MKB image as a prior for the additional cardiac gating. This procedure ensures the best possible prior for cardiac gating. To reduce noise we apply edge preserving denoising bilateral filters [11] in up to five dimensions. To define the bilateral filter let us restrict to one dimension, for convenience. Bilateral filtering of a function f(x) is then defined as dtd(x,t)r(x,t)f(t) Bf(x) = (5) dtd(x,t)r(x,t) with B denoting the bilateral filtering operator and ( x t ) 2 D(x,t) = e σ x R(x,t) = e ( f(x) f(t) ) 2 σ f being the domain and the range filter, respectively. The parameters σ x and σ f are the widths of the Gaussian domain and range filters, respectively. Note that other shapes of range and III. MEASUREMENTS We currently have several different data sets available for double gating that were measured in-house. Three of them shall be presented here. The first and second mouse was scanned with a dual source micro CT scanner in single source mode while the third mouse was scanned in a dual source micro CT scanner. Additionally data of recent publications were made available to us [4]. A. Mouse 1 and 2 The data of both mice were acquired with a dedicated in vivo cone beam micro CT scanner (TomoScope Synergy Twin, CT Imaging GmbH, Erlangen, Germany) in single source mode installed at the Institute of Medical Physics, Erlangen, Germany. The system consists of a micro focus x ray source mounted at a distance of 170 mm to the isocenter and a flat panel detector mounted at a distance of 39 mm to the isocenter. The size of the quadratic detector pixels was 50 µm. To increase the detector s readout rate to 25 frames per second a two by two binning of the detector pixels was performed. The scans were conducted at 65 kv tube voltage with a tube current of 0.3 ma projections were acquired per scan in a circular trajectory over an angular range of The scan time for these ten rotations was 288 s. The tube current time product was 87 mas, the absorbed dose was measured as 500 mgy in each case. The mice was anethesized using a combination of Ketamine and Xylazine. The contrast agents used were Binitio (Binitio Biomedical Inc., Canada) and Fenestra VC (ART Advanced Research Technologies, Canada), both injected via a tail vein. B. Mouse 3 The data were acquired with a dedicated in vivo dual source cone beam micro CT scanner (TomoScope Synergy Twin, CT Imaging GmbH, Erlangen, Germany) installed at the Weizmann Institute, Rehovot, Israel. Each source detector

3 Reference [2] Reference [3] Reference [4] Mouse 1 Mouse 2 Mouse 3 Type single source single source single source single source single source single source Gating 6 prospective retrospective retrospective retrospective retrospective retrospective R synchronization ventilation pressure chamber pneumatic cushion kymogram kymogram kymogram C synchronization ECG ECG ECG kymogram kymogram kymogram Source to isocenter 480 mm 540 mm 573 mm 170 mm 170 mm 170 mm Scan time 9 min 50 s 80 s 5 min 5 min 5 min Tube voltage 80 kv 80 kv 80 kv 65 kv 65 kv 65 kv Tube current 150 ma 50 ma 50 ma 0.3 ma 0.3 ma 0.3 ma Integration time 10 ms 12 ms 10 ms 40 ms 40 ms 40 ms Projections Current time product 2268 mas 2500 mas 4000 mas 86.4 mas 86.4 mas 86.4 mas Scan range View increment Spatial sampling 90 µm 150 µm 242 µm 80 µm 80 µm 80 µm Full cycle dose 1840 mgy 280 mgy 248 mgy 500 mgy 500 mgy 500 mgy Respiratory rate 90 rpm 51 rpm 30 rpm 150 rpm 120 rpm 120 rpm Heart rate 500 bpm 405 bpm 330 bpm 135 bpm 260 bpm 280 bpm Contrast agent Fenestra VC Fenestra VC Fenestra VC Fenestra VC Binitio NanoVista Enhancement 500 HU 360 HU 540 HU 110 HU 420 HU 230 HU Image noise 50 HU 36 HU 95 HU 175 HU / 30 HU 167 HU / 28 HU 130 HU / 24 HU Quality parameter 21% 21% 4% 15% / 87% 21% / 100% 13% / 61% TABLE I PARAMETERS FOR TWO DIFFERENT SCAN AND RECONSTRUCTION APPROACHES IN COMPARISON. NOTE THAT REFERENCE [2] DOES SAMPLE ONLY 50% OF THE CARDIAC CYCLE (SIX 10 MS WIDE SAMPLES WITHIN A 120 MS CARDIAC CYLCE). HENCE THE ACTUAL DOSE REPORTED AS 920 MGY IS ONLY HALF OF THE DOSE THAT WOULD BE REQUIRED TO SAMPLE THE COMPLETE CYCLE. RESPIRATORY AND HEART RATES ARE THE AVERAGES REPORTED IN EACH REFERENCE AND CORRESPOND TO MOUSE SCANS. THE SPATIAL SAMPLING CORRESPONDS TO THE DETECTOR PIXEL SIZE AFTER BINNING, IF APPLICABLE, AT THE ISOCENTER. thread consists of a micro focus x ray source mounted at a distance of 170 mm to the isocenter and a flat panel detector mounted at a distance of 62 mm to the isocenter. Both detectors were laterally shifted to increase the field of measurement to a diameter of 60 mm while leaving an 18 mm inner field of measurement [12]. Both threads were mounted under an angle of approximately 90. The size of the quadratic detector pixels was 50 µm. To increase the detector s readout rate to 25 frames per second a two by two binning of the detector pixels was performed. The scan was conducted at 40 kv tube voltage with a tube current of 1.0 ma per tube projections were acquired in a circular scan over an angular range of The scan time for these ten rotations was 288 s resulting in an overall tube current time product of 576 mas and an absorbed dose of 1077 mgy. The mouse was anethesized using a combination of Ketamine and Xylazine. The contrast agent used was Omnipaque Iohexol. [!t] IV. RESULTS Figure 1 compares the results of our new LDPC approach with a conventional PC approach of mouse 1. The PC reconstruction suffers from significantly increased image noise and from streak artifacts due to sparse view sampling. With the new algorithm we can reduce image noise from 175 HU to 30 HU and we can remove all streak artifacts while maintaining the full temporal resolution (as far as one can tell from the difference images). Furthermore we want to compare our results to those published in the literature. We have therefore prepared table I to see the most important differences at a glance. It should be noted that the data shown in this table only allow for a very rough comparison between the methods, mainly because important details of how certain parameters were measured or calculated are not disclosed in the publications. The last row of the table further shows a quality parameter Q given as [13] 1 Q σ (7) D 2 where σ is the image noise observed in unenhanced regions of the mouse tissue, D is the dose reported, and is the sampling distance of the rays scaled to the isocenter (which would typically be the diameter of the field of measurement divided by the number of detector rows or columns). The definition of the quality parameter reflects the fact that image noise variance is proportional to one over dose and to one over the fourth power of spatial resolution. The constant of proportionality is chosen to obtain a quality value of 100% for the LDPC reconstructions of mouse 2. The other quality value stated for our experiment corresponds to the conventional phase correlated reconstruction and therefore is far lower than 100%. It must be emphasized that the quality parameter is nothing but a rough estimate due to the following reasons. First of all, spatial resolution was assumed to be proportional to the sampling distance and the effect of the reconstruction process, that potentially lowers the spatial resolution, could not be taken into account. In addition the dose values reported are measured using different methods and different phantoms and therefore are not highly accurate. And finally the image noise quoted is neither measured in difference images, and therefore is likely to contain effects such as streak artifacts or non uniform background, nor was it measured in the same region of the mouse in all cases.

4 Fig. 3. Phase correlated reconstructions of mouse 2 centered at r = 60% and c = 0% with window widths r = 10% and c = 20%. The left panel shows the conventional phase correlated reconstructions while the right panel is the proposed low dose approach. Dose levels ranging from 60 mgy to 500 mgy were obtained by using only fractions of the data available. Values for image noise are shown in the axial slices. (400 HU / 800 HU) Fig. 4. Phase correlated reconstructions of the reference [4] mouse centered at r = 60% and c = 0% with window widths r = 10% and c = 20%. The left panel shows the conventional phase correlated reconstructions while the right panel is the proposed low dose approach. Dose levels ranging from 25% to 100% were obtained by using only fractions of the data available. Values for image noise are shown in the sagittal slices. (300 HU / 700 HU)

5 Nevertheless, there are two things to be learned regarding the evaluation of the quality parameter: Without advanced image reconstruction and signal processing techniques, our results lie well within the range of what has been published in the literature so far. And switching from standard phase correlated reconstruction to low dose phase correlated reconstruction has the potential to increase the image quality by a factor of about five. To further illustrate the dose reduction capabilities of the LDPC method a corresponding study has been performed based on the data of reference [4] that were made available to us and its results are shown in figure 4. A decrease in dose from 250 mgy to 63 mgy yields significant streak artifacts in the phase correlated images which are not visible in the low dose phase correlated images. V. CONCLUSION AND DISCUSSION LDPC is a new technique to significantly improve phase correlated imaging from highly undersampled data. We demonstrate that dose savings of about an order of magnitude are possible compared to standard phase correlated reconstruction approaches. With sophisticated reconstruction techniques such as the LDPC algorithm it is possible to perform 4D or 5D phase correlated imaging at about the same dose level as required for conventional 3D studies. Using LDPC appears to allow for longitudinal in vivo examinations of the rodent heart and enables long term studies with reduced metabolic inference. REFERENCES [1] C. Badea, L. Hedlund, and G. Johnson, Micro CT with respiratory and cardiac gating, Med. Phys., vol. 31, no. 12, pp , Dec [2] C. Badea, B. Fubara, L. Hedlund, and G. Johnson, 4D micro CT of the mouse heart, Molecular Imaging, vol. 4, no. 2, pp , Apr./Jun [3] M. Drangova, N. L. Ford, S. A. Detombe, A. R. Wheatley, and D. W. Holdsworth, Fast retrospectively gated quantitative four dimensional (4D) cardiac micro computed tomography imaging of free breathing mice, Investigative Radiology, vol. 42, no. 2, pp , Feb [4] S. H. Bartling, W. Stiller, M. Grasruck, B. Schmidt, P. Peschke, W. Semmler, and F. Kiessling, Retrospective motion gating in small animal CT of mice and rats, Investigative Radiology, vol. 42, no. 10, pp , oct [5] J. Dinkel, S. H. Bartling, J. Kuntz, M. Grasruck, A. Kopp-Schneider, M. Iwasaki, S. Dimmeler, R. Gupta, W. Semmler, H.-U. Kauczor, and F. Kiessling, Intrinsic gating for small animal computed tomography: A robust ECG less paradigm for deriving cardiac phase information and functional imaging, Circulation: Cardiovascular Imaging, vol. 1, no. 3, pp , [6] J.-J. Sonke, L. Zijp, P. Remeijer, and M. van Herk, Respiratory correlated cone beam CT, Med. Phys., vol. 32, no. 4, pp , Apr [7] S. H. Bartling, J. Dinkel, W. Stiller, M. Grasruck, I. Madisch, H.-U. Kauczor, W. Semmler, R. Gupta, and F. Kiessling, Intrinsic respiratory gating in small animal CT, European Radiology, vol. 18, no. 7, pp , Jul [8] L. Feldkamp, L. Davis, and J. Kress, Practical cone beam algorithm, Journal of the Optical Society of America, vol. 1, no. 6, pp , Jun [9] G. C. McKinnon and R. Bates, Towards imaging the beating heart usefully with a conventional CT scanner, IEEE Transactions on Biomedical Engineering, vol. BME-28, no. 2, pp , Feb [10] K. L. Garden and R. A. Robb, 3-D reconstruction of the heart from few projections: A practical implementation of the McKinnon Bates algorithm, IEEE Transactions on Medical Imaging, vol. MI-5, no. 4, pp , Dec [11] C. Tomasi and R. Manduchi, Bilateral filtering for gray and color images, Proc. 6th Int. Conf. Computer Vision, pp , [12] C. Maaß, M. Knaup, R. Lapp, M. Karolczak, W. A. Kalender, and M. Kachelrieß, A new weighting function to achieve high temporal resolution in circular cone beam CT with shifted detectors, Med. Phys., vol. 35, no. 12, pp , Dec [13] W. A. Kalender, Computed Tomography. Wiley & Sons, 2005, ISBN , 2nd Edition.

Preclinical X-Ray Computed Tomography

Preclinical X-Ray Computed Tomography Preclinical X-Ray Computed Tomography Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de/ct Is CT a Molecular Imaging Modality? Imaging Modality Molecular Sensitivity

More information

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

Software and Hardware in CCTA. Elly Castellano PhD

Software and Hardware in CCTA. Elly Castellano PhD Software and Hardware in CCTA Elly Castellano PhD Outline technical requirements for coronary CTA the modern cardiac CT scanner ECG-gating technology image reconstruction algorithms 2 Technical requirements

More information

Data. microcat +SPECT

Data. microcat +SPECT Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for

More information

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National

More information

Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2

Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2 Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2 1. Department of Medical Physics, University of Wisconsin-Madison 2. Department

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for

More information

Detector technology in simultaneous spectral imaging

Detector technology in simultaneous spectral imaging Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

More information

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT) Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in

More information

Computed tomography with single-shot dual-energy sandwich detectors

Computed tomography with single-shot dual-energy sandwich detectors Computed tomography with single-shot dual-energy sandwich detectors Seung Ho Kim, a Hanbean Youn, b,c Daecheon Kim, a Dong Woon Kim, a Hosang Jeon, b,c Ho Kyung Kim a,c a School of Mechanical Engineering,

More information

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT

More information

Efficacy of Beam-Hardening Correction for Vascular Morphometry Using 3D Conebeam Micro-CT

Efficacy of Beam-Hardening Correction for Vascular Morphometry Using 3D Conebeam Micro-CT Marquette University e-publications@marquette Biomedical Engineering Faculty Research and Publications Biomedical Engineering, Department of 1-1-1999 Efficacy of Beam-Hardening Correction for Vascular

More information

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Poster No.: C-2205 Congress: ECR 2012 Type: Scientific Paper

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

Geometric image distortion in flat-panel X-ray detectors and its influence on the accuracy of CT-based dimensional measurements

Geometric image distortion in flat-panel X-ray detectors and its influence on the accuracy of CT-based dimensional measurements Geometric image distortion in flat-panel X-ray detectors and its influence on the accuracy of CT-based dimensional measurements Daniel Weiß, Ronald Lonardoni, Andreas Deffner, Christoph Kuhn Carl Zeiss

More information

Electronic Noise in CT Detectors: Impact on Image Noise and Artifacts

Electronic Noise in CT Detectors: Impact on Image Noise and Artifacts Medical Physics and Informatics Original Research Duan et al. Electronic Noise in CT Detectors Medical Physics and Informatics Original Research Xinhui Duan 1 Jia Wang 1,2 Shuai Leng 1 ernhard Schmidt

More information

160-slice CT SCANNER / New Standard for the Future

160-slice CT SCANNER / New Standard for the Future TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

Suppression of metal artifacts using image-based monoenergetic DECT imaging

Suppression of metal artifacts using image-based monoenergetic DECT imaging Suppression of metal artifacts using image-based monoenergetic DECT imaging Poster No.: C-0519 Congress: ECR 2011 Type: Scientific Paper Authors: B. Krauss, B. Schmidt, M. Sedlmair, T. Flohr; Forchheim/DE

More information

Computerized Medical Imaging and Graphics

Computerized Medical Imaging and Graphics Computerized Medical Imaging and Graphics 36 (2012) 387 395 Contents lists available at SciVerse ScienceDirect Computerized Medical Imaging and Graphics jo ur n al homep age : www.elsevier.com/locate/compmedimag

More information

PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner

PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner September, 2017 Results submitted to Physics in Medicine & Biology Negar Omidvari 1, Jorge Cabello 1, Geoffrey Topping

More information

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc.

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. An Introduction to TG-142 Imaging QA Using Standard Imaging Products Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. Goals Understand the nature and intent of TG 142 imaging

More information

PD233: Design of Biomedical Devices and Systems

PD233: Design of Biomedical Devices and Systems PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/

More information

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

1. Patient size AEC. Large Patient High ma. Small Patient Low ma Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

Related topics Beam hardening, cupping effect, Beam hardening correction, metal artefacts, photon starvation

Related topics Beam hardening, cupping effect, Beam hardening correction, metal artefacts, photon starvation Beam hardening and metal artefacts TEP Related topics Beam hardening, cupping effect, Beam hardening correction, metal artefacts, photon starvation Principle X-ray sources produce a polychromatic spectrum

More information

Four-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development

Four-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development ORIGINAL ARTICLE ORIGINAL ARTICLE Radiation Medicine: Vol. 21 No. 1, 17 22 p.p., 2003 Four-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development Masahiro Endo,* Takanori Tsunoo,*

More information

Iterative Reconstruction in Image Space. Answers for life.

Iterative Reconstruction in Image Space. Answers for life. Iterative Reconstruction in Image Space Answers for life. Iterative Reconstruction in Image Space * (IRIS) * Please note: IRIS is used as an abbreviation for Iterative Reconstruction in Image Space throughout

More information

QUANTITATIVE COMPUTERIZED LAMINOGRAPHY. Suzanne Fox Buchele and Hunter Ellinger

QUANTITATIVE COMPUTERIZED LAMINOGRAPHY. Suzanne Fox Buchele and Hunter Ellinger QUANTITATIVE COMPUTERIZED LAMINOGRAPHY Suzanne Fox Buchele and Hunter Ellinger Scientific Measurement Systems, Inc. 2201 Donley Drive Austin, Texas 78758 INTRODUCTION Industrial computerized-tomography

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

diagnostic examination

diagnostic examination RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011

More information

Maximum Performance, Minimum Space

Maximum Performance, Minimum Space TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation

More information

Aquilion Precision Ultra-High Resolution CT: Quantifying diagnostic image quality

Aquilion Precision Ultra-High Resolution CT: Quantifying diagnostic image quality Aquilion Precision Ultra-High CT: Quantifying diagnostic image quality Kirsten Boedeker, PhD, DABR Senior Manager, Quantitative Image Quality Canon Medical Systems Corporation Introduction Over the last

More information

Philip Sperling. Sales Science and New Materials, YXLON International GmbH, Essener Bogen 15, Hamburg, Germany.

Philip Sperling. Sales Science and New Materials, YXLON International GmbH, Essener Bogen 15, Hamburg, Germany. A new generation of x-ray computed tomography devices for quality inspection and metrology inspection in the field of additive manufacturing and other sciences Philip Sperling Sales Science and New Materials,

More information

Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner

Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner Kisung Lee, Member, IEEE, Paul E. Kinahan, Senior Member, Robert S. Miyaoka, Member, IEEE, Jeffrey A. Fessler,

More information

Principles of CT scan

Principles of CT scan Related topics Detector calibration, saturation, CT acquisition, CT reconstruction Principle X-ray computed tomography consists of using X-rays that are converted to a digital signal by a detector and

More information

Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing. Koichi Hirokawa MEDIX VOL. 56 P.43 P.46

Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing. Koichi Hirokawa MEDIX VOL. 56 P.43 P.46 Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing Taiga Goto Koichi Hirokawa Hisashi Takahashi MEDIX VOL. 56 P.43 P.46 Advanced Noise Reduction Processing for X-ray CT System

More information

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life.

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life. Neuro Thoracic Abdominal Abdominal Cardiovascular Pediatric SAFIRE Sinogram Affirmed Iterative Reconstruction Answers for life. SAFIRE * (Sinogram Affirmed Iterative Reconstruction) * The information

More information

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation Image Quality Review I: Basics and Image Quality TH-A-16A-1 Thursday 7:30AM - 9:30AM Room: 16A J. Anthony

More information

Monte Carlo study on a new concept of a scanning photon beam system for IMRT

Monte Carlo study on a new concept of a scanning photon beam system for IMRT NUKLEONIKA 2011;56(4):291 297 ORIGINAL PAPER Monte Carlo study on a new concept of a scanning photon beam system for IMRT Anna M. Wysocka-Rabin, Günter H. Hartmann Abstract. Intensity-modulated radiation

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Clinical Importance on CT

Clinical Importance on CT 183 Truncated-View Clinical Importance on CT..,..,.... Artifacts:.... James L. Lehr1 A truncated-view artifact in CT is produced whenever any part of the patient or imaged object is present in some but

More information

Digital Image Processing

Digital Image Processing Digital Image Processing Part : Image Enhancement in the Spatial Domain AASS Learning Systems Lab, Dep. Teknik Room T9 (Fr, - o'clock) achim.lilienthal@oru.se Course Book Chapter 3-4- Contents. Image Enhancement

More information

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.125 pissn 2508-4445, eissn 2508-4453 Optimization of Energy Modulation Filter for Dual Energy CBCT

More information

Uncertainty in CT Metrology: Visualizations for Exploration and Analysis of Geometric Tolerances

Uncertainty in CT Metrology: Visualizations for Exploration and Analysis of Geometric Tolerances Uncertainty in CT Metrology: Visualizations for Exploration and Analysis of Geometric Tolerances Artem Amirkhanov 1, Bernhard Fröhler 1, Michael Reiter 1, Johann Kastner 1, M. Eduard Grӧller 2, Christoph

More information

The image reconstruction influence in relative measurement in SPECT / CT animal

The image reconstruction influence in relative measurement in SPECT / CT animal BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 0-01 (201) 01-09 The image reconstruction influence in relative measurement in SPECT / CT animal S.C.S. Soriano a ; S.A.L. Souza b ; T.Barboza b ; L.V. De Sá

More information

PET/CT Instrumentation Basics

PET/CT Instrumentation Basics / Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire

More information

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893 Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -

More information

X-RAY COMPUTED TOMOGRAPHY

X-RAY COMPUTED TOMOGRAPHY X-RAY COMPUTED TOMOGRAPHY Bc. Jan Kratochvíla Czech Technical University in Prague Faculty of Nuclear Sciences and Physical Engineering Abstract Computed tomography is a powerful tool for imaging the inner

More information

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE Takeyuki Hashimoto 1), Morio Onoe 2), Hiroshi Nakamura 3), Tamon Inouye 4), Hiromichi Jumonji 5), Iwao Takahashi 6); 1)Yokohama Soei

More information

CT Basics: Image Quality Module 6

CT Basics: Image Quality Module 6 Module 6 For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions. Any broadcast,

More information

Wide-Detector CT for TAVR Planning:

Wide-Detector CT for TAVR Planning: Wide-Detector CT for TAVR Planning: Impact on Iodine Dose, Radiation Dose, and Image Quality SCBTMR 2015 Annual Course Thursday, October 8 William P. Shuman MD FSCBTMR Department of Radiology University

More information

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) Hiroki Kadoya Yukiko Kitagawa MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria

More information

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations S. Baechler 1, P. Monnin 1, A. Aroua 1, J.F. Valley 1, M. Perrier, P. Trueb 3, F.R. Verdun 1 1 University

More information

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging JOURNAL OF MAGNETIC RESONANCE IMAGING 20:1046 1051 (2004) Technical Note The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging James W. Goldfarb, PhD* Purpose: To describe a known (but undocumented)

More information

SYLLABUS. 1. Identification of Subject:

SYLLABUS. 1. Identification of Subject: SYLLABUS Date/ Revision : 30 January 2017/1 Faculty : Life Sciences Approval : Dean, Faculty of Life Sciences SUBJECT : Biophysics 1. Identification of Subject: Name of Subject : Biophysics Code of Subject

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Non-contact Photoacoustic Tomography using holographic full field detection

Non-contact Photoacoustic Tomography using holographic full field detection Non-contact Photoacoustic Tomography using holographic full field detection Jens Horstmann* a, Ralf Brinkmann a,b a Medical Laser Center Lübeck, Peter-Monnik-Weg 4, 23562 Lübeck, Germany; b Institute of

More information

PET: New Technologies & Applications, Including Oncology

PET: New Technologies & Applications, Including Oncology PET: New Technologies & Applications, Including Oncology, PhD, FIEEE Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Disclosures Research Contract, GE Healthcare

More information

Metal Artifact Reduction for Orthopedic Implants (O-MAR)

Metal Artifact Reduction for Orthopedic Implants (O-MAR) Metal Artifact Reduction for Orthopedic Implants () Summary Since the inception of CT, numerous methods have been proposed to suppress metal artifacts with varying degrees of success. 1-4 (Metal Artifact

More information

COMPUTATIONAL IMAGING. Berthold K.P. Horn

COMPUTATIONAL IMAGING. Berthold K.P. Horn COMPUTATIONAL IMAGING Berthold K.P. Horn What is Computational Imaging? Computation inherent in image formation What is Computational Imaging? Computation inherent in image formation (1) Computing is getting

More information

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS Veterinary Science Preparatory Training for the Veterinary Assistant Floron C. Faries, Jr., DVM, MS Radiology Floron C. Faries, Jr., DVM, MS Objectives Determine the appropriate machine settings for making

More information

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Huaiqun Guan,

More information

Tomophan TSP004 Manual

Tomophan TSP004 Manual T h e P h a n t o m L a b o r a t o r y 1 Tomophan TSP004 Manual Copyright 2016 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order and

More information

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results NeuViz 16 Computed Tomography Elevating routine imaging for exceptional results Essence NeuViz 16 Raising the bar on clinical utility in routine imaging. Get more. More clinical information for patients.

More information

Conceptual Study of Brain Dedicated PET Improving Sensitivity

Conceptual Study of Brain Dedicated PET Improving Sensitivity Original Article PROGRESS in MEDICAL PHYSICS 27(4), Dec. 2016 https://doi.org/10.14316/pmp.2016.27.4.236 pissn 2508-4445, eissn 2508-4453 Conceptual Study of Brain Dedicated PET Improving Sensitivity Han-Back

More information

Fig.2: Scanner VistaScan for image plates

Fig.2: Scanner VistaScan for image plates RADIOGRAPHIC INSPECTION OF WELDINGS BY DIGITAL SENSORS H. Thiele, H.-J. Friemel RADIS GmbH, Johanniskirchen, Germany Abstract: The newly available digital sensors for radiographic inspection are suitable

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

Image Processing by Bilateral Filtering Method

Image Processing by Bilateral Filtering Method ABHIYANTRIKI An International Journal of Engineering & Technology (A Peer Reviewed & Indexed Journal) Vol. 3, No. 4 (April, 2016) http://www.aijet.in/ eissn: 2394-627X Image Processing by Bilateral Image

More information

Defining the Optimal Beam Hardening Correction Parameters for CT Dimensional Metrology Applications

Defining the Optimal Beam Hardening Correction Parameters for CT Dimensional Metrology Applications International Conference on Competitive Manufacturing Defining the Optimal Beam Hardening Correction Parameters for CT Dimensional Metrology Applications Y. Tan 1,2, K. Kiekens 1,2, F. Welkenhuyzen 2,

More information

Testing a wavelet based noise reduction method using computersimulated

Testing a wavelet based noise reduction method using computersimulated Testing a wavelet based noise reduction method using computersimulated mammograms Christoph Hoeschen 1, Oleg Tischenko 1, David R Dance 2, Roger A Hunt 2, Andrew DA Maidment 3, Predrag R Bakic 3 1 GSF-

More information

Synchrotron X-ray tomographic microscopy Theory vs. practice

Synchrotron X-ray tomographic microscopy Theory vs. practice Synchrotron X-ray tomographic microscopy Theory vs. practice Federica Marone Swiss Light Source, Paul Scherrer Institut, Villigen, Switzerland Theory Radon transform Rf x = Beer-Lambert law I E = I 0 (E)e

More information

Development of new dosimeter for measuring dose distribution in CT

Development of new dosimeter for measuring dose distribution in CT Development of new dosimeter for measuring dose distribution in CT Poster No.: C-2925 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology - Without Subtopic Authors: Y. Muramatsu, K.

More information

SPECT Reconstruction & Filtering

SPECT Reconstruction & Filtering SPECT Reconstruction & Filtering Goals Understand the basics of SPECT Reconstruction Filtered Backprojection Iterative Reconstruction Make informed choices on filter selection and settings Pre vs. Post

More information

Multimodal Co-registration Using the Quantum GX, G8 PET/CT and IVIS Spectrum Imaging Systems

Multimodal Co-registration Using the Quantum GX, G8 PET/CT and IVIS Spectrum Imaging Systems TECHNICAL NOTE Preclinical In Vivo Imaging Authors: Jen-Chieh Tseng, Ph.D. Jeffrey D. Peterson, Ph.D. PerkinElmer, Inc. Hopkinton, MA Multimodal Co-registration Using the Quantum GX, G8 PET/CT and IVIS

More information

Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT)

Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT) Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT) F Edward Boas, Roland Bammer, and Dominik Fleischmann Extended abstract for RSNA 2012 Purpose CT metal streak artifacts are

More information

Application Report 11/15/5471 Company: Kienle Spiess

Application Report 11/15/5471 Company: Kienle Spiess 11/15/5471 Company: Kienle Spiess Birger Niehaus, 12/07/2015 Radioscopic (DR) Inspection and Computer Tomography (CT) of Squirrel-Cage Motors Technology with Passion System description MU 2000-D CT X-ray

More information

Introduction of a Single Chip TLD System for Patient Dosimetry

Introduction of a Single Chip TLD System for Patient Dosimetry Introduction of a Single Chip TLD System for Patient Dosimetry C. Hranitzky a, M. Halda a, G. Müller a, B. Obryk b, H. Stadtmann a* a Austrian Research Centers GmbH ARC, 2444 Seibersdorf, Austria. b Institute

More information

Organ-Specific Context-Sensitive Single and Dual Energy CT (DECT) Image Reconstruction, Display and Analysis

Organ-Specific Context-Sensitive Single and Dual Energy CT (DECT) Image Reconstruction, Display and Analysis Organ-Specific Context-Sensitive Single and Dual Energy CT (DECT) Image Reconstruction, Display and Analysis Sabrina Dorn 1, Shuqing Chen², Francesco Pisana 1, Joscha Maier 1, Michael Knaup 1, Stefan Sawall

More information

Real Time Deconvolution of In-Vivo Ultrasound Images

Real Time Deconvolution of In-Vivo Ultrasound Images Paper presented at the IEEE International Ultrasonics Symposium, Prague, Czech Republic, 3: Real Time Deconvolution of In-Vivo Ultrasound Images Jørgen Arendt Jensen Center for Fast Ultrasound Imaging,

More information

Alignment of the camera

Alignment of the camera Related topics Detector Alignment, Rotation axis, tilt, Principle Alignment of the detector and the rotation stage is very important to get optimal quality images of a CT scan. In this experiment, the

More information

This document is published in:

This document is published in: Institutional Repository This document is published in: Bo Yu (ed.) (01). 01 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC): Anaheim, California, USA. October 9 - November 3, 01.

More information

OPTIMIZING SPATIAL RESOLUTION WITH THE MECHANICAL DESIGN OF AN X-RAY COMPU1ED TOMOGRAPHY SCANNER

OPTIMIZING SPATIAL RESOLUTION WITH THE MECHANICAL DESIGN OF AN X-RAY COMPU1ED TOMOGRAPHY SCANNER OPTIMIZING SPATIAL RESOLUTION WITH THE MECHANICAL DESIGN OF AN X-RAY COMPU1ED TOMOGRAPHY SCANNER Lowell D. Harris, RichardT. Bernardi, Simon H. C. Hughes, and Robert E. Slocum Bio-Imaging Research, Inc.

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET 2005 IEEE Nuclear Science Symposium Conference Record M11-126 Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET Jin Zhang, Member,

More information

Distributed source x-ray tube technology for tomosynthesis imaging

Distributed source x-ray tube technology for tomosynthesis imaging Distributed source x-ray tube technology for tomosynthesis imaging Authors: F. Sprenger a*, X. Calderon-Colon b, Y. Cheng a, K. Englestad a, J. Lu b, J. Maltz c, A. Paidi c, X. Qian b, D. Spronk a, S.

More information

The influence of bowtie filtration on cone-beam CT image quality

The influence of bowtie filtration on cone-beam CT image quality The influence of bowtie filtration on cone-beam CT image quality N. Mail Radiation Medicine Program, Princess Margaret Hospital, Toronto, Ontario M5G 2M9, Canada and Ontario Cancer Institute, University

More information

PIXSCAN CT scanner for Small Animal Imaging Based on hybrid pixel detectors

PIXSCAN CT scanner for Small Animal Imaging Based on hybrid pixel detectors PIXSCAN CT scanner for Small Animal Imaging Based on hybrid pixel detectors Centre de Physique des Particules de M arseille (CPPM -IN2P3), France S. B a s o lo, A. B o n is s e n t, P. B re u g n o n,

More information

Achim J. Lilienthal Mobile Robotics and Olfaction Lab, AASS, Örebro University

Achim J. Lilienthal Mobile Robotics and Olfaction Lab, AASS, Örebro University Achim J. Lilienthal Mobile Robotics and Olfaction Lab, Room T29, Mo, -2 o'clock AASS, Örebro University (please drop me an email in advance) achim.lilienthal@oru.se 4.!!!!!!!!! Pre-Class Reading!!!!!!!!!

More information

Alae Tracker: Tracking of the Nasal Walls in MR-Imaging

Alae Tracker: Tracking of the Nasal Walls in MR-Imaging Alae Tracker: Tracking of the Nasal Walls in MR-Imaging Katharina Breininger 1, Andreas K. Maier 1, Christoph Forman 1, Wilhelm Flatz 2, Catalina Meßmer 3, Maria Schuster 3 1 Pattern Recognition Lab, Friedrich-Alexander-Universität

More information

Minimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations

Minimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations uwe.ewert@bam.de Minimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations Uwe Ewert and Uwe Zscherpel BAM Federal Institute for

More information

NIH Public Access Author Manuscript Int J Cardiovasc Imaging. Author manuscript; available in PMC 2008 May 26.

NIH Public Access Author Manuscript Int J Cardiovasc Imaging. Author manuscript; available in PMC 2008 May 26. NIH Public Access Author Manuscript Published in final edited form as: Int J Cardiovasc Imaging. 2001 August ; 17(4): 287 296. A comparison of prospective and retrospective respiratory navigator gating

More information

Reconstruction Filtering in Industrial gamma-ray CT Application

Reconstruction Filtering in Industrial gamma-ray CT Application Reconstruction Filtering in Industrial gamma-ray CT Application Lakshminarayana Yenumula *, Rajesh V Acharya, Umesh Kumar, and Ashutosh Dash Industrial Tomography and Instrumentation Section, Isotope Production

More information