Computed tomography with single-shot dual-energy sandwich detectors

Size: px
Start display at page:

Download "Computed tomography with single-shot dual-energy sandwich detectors"

Transcription

1 Computed tomography with single-shot dual-energy sandwich detectors Seung Ho Kim, a Hanbean Youn, b,c Daecheon Kim, a Dong Woon Kim, a Hosang Jeon, b,c Ho Kyung Kim a,c a School of Mechanical Engineering, Pusan National University, Busan 46241, South Korea b Department of Radiation Oncology, Pusan National University Yangsan Hospital, Yangsan, Gyeongsangnam-do 50612, South Korea c Center for Advanced Medical Engineering, Pusan National University, Busan 46241, South Korea ABSTRACT Single-shot dual-energy sandwich detector can produce sharp images because of subtraction of images from two sub-detector layers, which have different thick x-ray converters, of the sandwich detector. Inspired by this observation, the authors have developed a microtomography system with the sandwich detector in pursuit of highresolution bone-enhanced small-animal imaging. The preliminary results show that the bone-enhanced images reconstructed with the subtracted projection data are better in visibility of bone details than the conventionally reconstructed images. In addition, the bone-enhanced images obtained from the sandwich detector are relatively immune to the artifacts caused by photon starvation. The microtomography with the single-shot dual-energy sandwich detector will be useful for the high-resolution bone imaging. Keywords: Computed tomography, microtomography, micro-ct, sandwich detector, dual-energy imaging, single-shot dual-energy imaging, mouse imaging 1. INTRODUCTION We previously described the multilayer ( sandwich ) detector by stacking two flat-panel detectors (FPDs) and demonstrated its prospect for motion-artifact-free single-shot dual-energy imaging (DEI) by obtaining bone and soft-tissue images of a postmortem mouse. 1, 2 While the front FPD measures relatively low energy, as shown in Fig. 1, the rear one measures relatively high energy because of x-ray beam hardening through the front FPD. Onto the same CMOS photodiode platform, thus, we placed a thicker scintillator in the rear FPD than the front one to achieve high quantum efficiency with the relatively higher-energy x-ray spectrum. An intermediate copper (Cu) filter can be used to further increase spectral separation between the two FPD measurements, which may provides a better contrast-to-noise performance in the subtracted images. As shown in Fig. 2, the conventional dual-shot image obtained by switching the applied tube voltages (40/70 kvp) showed superior signal-to-noise ratio (SNR) performance but the single-shot image obtained from the sandwich detector was almost as good and had the advantage of being less susceptible to motion artifacts. We also observed that the single-shot method showed better SNR at higher spatial frequencies (e.g. edge regions and bone details) than the double-shot method. The reason can be explained by the inherent unsharp masking effect of the sandwich detector; 2 the rear FPD with a thicker scintillator provides a blurrier image than the front FPD, hence subtraction of the two images enhances edges in the resultant image. Since the physical characteristics of bone are affected by numerous factors, such as age, hormones, arthritis, and exercise, 3 a preclinical study with small animals with respect to those various factors would be helpful for theoretical and computational models of bone. 4 Moreover, longitudinal, histological studies without sacrificing animals are essential. 5, 6 In these regards, the x-ray microtomography (or micro-ct) has been popular for smallanimal bone studies Inspired by the aforementioned observation, we have applied the sandwich-detector concept to the microtomography (or micro-ct) for small-animal bone imaging. hokyung@pusan.ac.kr; phone ; fax

2 Figure 1. Conceptual illustration of the sandwich detector operation. Figure 2. Comparison of dual-energy postmortem mouse images obtained from (a) dual-shot (40/70 kvp switching) and (b) single-shot (using the sandwich detector at 70 kvp) methods. 2. MATERIALS AND METHODS As shown in Fig. 3, we have developed a bench-top micro-ct system with the sandwich detector. During continuous x-ray irradiation, the object rotates on its axis by an amount of prescribed step angle and then the rotation stays until the sandwich detector produces two projection images. These motion and image readout were computer-controlled and lasted till a single rotation completed. The distances from the x-ray focal spot to the detector (dsd ) and to the axis of rotation (dsa ) were computer-controlled variables. The traveling ranges of the object jig were 600, 300, and 75 mm along the x, y, and z directions, respectively. The x-ray source (Series 5000 XTF5011, Oxford Instruments, Inc., US) employed a tungsten anode and could operate up to the maximum power of 50 Watts. The tungsten x-ray spectra were further tailored by an additional 1-mm thick aluminum filter. According to the manufacturer, the nominal focal-spot size was mm. Each FPD layer constituting the sandwich detector employed a combination of a Gd2 O2 S:Tb phosphor screen for conversion of x-ray into optical quanta and a photodiode array for detection of them. The thicknesses of the front and rear phosphors were 34 and 67 mg cm 2, respectively. The same photodiode arrays (RadEye1TM, Teledyne Rad-icon Imaging Corp., Sunnyvale, US) were used for the front and rear FPDs. The pixel pitch of the photodiode array was mm, and the magnification of the pixel pitch of the rear detector was negligible. The active area of the sandwich detector was mm2. For a small-animal imaging, we prepared a postmortem mouse phantom ( 40 g) by replacing blood by paraformaldehyde as shown in Fig. 3. Bone-enhanced tomographic images may be obtained by reconstructing bone-enhanced projection data fbone (r) = FDK {WPF PR }, (1) where Pj denotes the projection data in a matrix form obtained from the jth FPD layer and W is a diagonal matrix consisting of weighting factors determined at each projection angle w(θ). The operator FDK { } implies the approximate filtered backprojection operation11 with the Hann filter. The w(θ) was determined by minimizing

3 Figure 3. Picture describing the micro-ct system with the sandwich detector. The enlarged views show the sandwich detector and a postmortem mouse phantom. contrast between the soft tissue to be subtracted and background. 2 If the weighting factor is independent upon the rotation angle, W becomes a scalar w. Head part of a mouse was scanned using two different designs of sandwich detector; one design used no intermediate filter and the other used a Cu filter with a thickness of 0.3 mm. Irradiation x-ray spectrum was from a tungsten target at 50 kvp/1 mm aluminum filter. 360 projection views were obtained for a single circular scan and they were used for reconstruction. All the reconstructed images were calibrated into the Hounsfield units (HU) using separate scans of water phantom. 3. RESULTS AND DISCUSSION Figure 4 compares projection images obtained from the each FPD layer of the two designs of sandwich detectors (i.e. one design included a 0.3 mm-thick Cu filter and the other did not) and their resultant DE images for the postmortem mouse phantom. The images were displayed with the level of their mean value (µ) and a window of two times their standard deviation (σ) (other images below were displayed with the same level and window). It was observed that the projections from the front FPD were sharper than those from the rear FPD as the font FPD employed a thinner phosphor than the rear FPD. Comparing Figs. 4(b) with (e), use of the intermediate filter resulted in higher noise in the rear FPD image, and the reason could be explained by the reduction in the number of x-ray photons reaching the rear FPD due to the attenuation through the filter layer. Weighted logarithmic subtraction successfully provided bone-enhanced images as shown in Figs. 4(c) and (f). Tomographic images reconstructed using each projection dataset, as exemplary shown in Fig. 4, are summarized in Fig. 5. The characteristics observed from the projection data were well reflected into the tomographic images. Comparing Figs. 5(c) with (f), use of the intermediate filter gave rise to a more reduction of soft tissues. It was also observed that the DE tomographic images showed less streak artifacts due to photon starvation compared to the images obtained using the front FPD. As shown in Fig. 6, the regions indicated as boxes in Fig. 5(d) have been investigated in detail. The rear FPD with a thick phosphor provided a blur image as shown in Fig. 6(b), and it became noisier, as shown in Fig. 6(e), when the Cu filter was additionally used. The DE tomographic image obtained form the sandwich detector without any filter layers showed the best visual performance for bone details. Figure 7 compares profiles extracted along the line A A (as possible as we can), as designated in Fig. 5(a), for each reconstruction image. Bone signal in the reconstructed image with projections from the front FPD was the largest, the rear FPD the second, and then the DE results. Although the DE tomographic image signals were noisy, they consisted mostly of bone signals. As observed from Figs. 4(e) and 5(e), the images obtained from the rear FPD of the sandwich detector with the Cu filter were noisy.

4 Figure 4. Projection images obtained from the each FPD layer of the sandwich detectors without and with a Cu filter layer and their resultant DE images for the postmortem mouse phantom. (L/W = µ/ ± 2σ).

5 Figure 5. Comparison of tomographic images reconstructed from the corresponding projection dataset. Figure 6. Enlarged images indicated by the boxes Fig. 5(d) for more detailed displays of tomographic images.

6 Figure 7. Profiles extracted along the line A A, as designated in Fig. 5(a), for each corresponding reconstruction image. In the present work, the bone-enhanced tomographic image was obtained by reconstructing weighted-subtraction sinogram. Alternatively, the bone-enhanced images may also be obtained by subtraction of two tomographic images reconstructed with projections obtained from the front and rear FPDs f bone (r) = wfdk {P F } FDK {P R }. (2) This approach is under progress and comparison with the present method will be made. Furthermore, the boneenhanced DE tomographic images may be combined with the images reconstructed with projections obtained from the front FPD as the conventional unsharp masking digital image processing. The authors anticipate the resultant images will be conventional tomographic images which include more pronounced bone details. These further studies will be a separate future study. 4. CONCLUSION Bone-enhanced tomographic images have been obtained using dual-energy sandwich detectors for a postmortem mouse phantom, and they outperformed the tomographic images obtained from the conventional detectors (i.e. the front and rear flat-panel detectors constituting the sandwich detectors) for bone details. Although use of an intermediate filter, which was placed between the front and rear flat-panel detectors, resulted in less residual soft tissues in the reconstructed bone-enhanced images, it degraded the visual image quality of bone details because of increased noise. Optimal filter design in terms of material and thickness is required for a more tissue separability and less noise performance in images. ACKNOWLEDGMENTS This work was supported by the National Research Foundation (NRF) grants funded by the Korea government (MSIP) (No. 2013M2A2A and No. 2014R1A2A2A ). REFERENCES 1. S. Yun, J. C. Han, D. W. Kim, H. Youn, H. K. Kim, J. Tanguay, and I. A. Cunningham, Feasibility of active sandwich detectors for single-shot dual-energy imaging, Proc. SPIE 9033, pp T 90335T 8, 2014.

7 2. J. C. Han, H. K. Kim, D. W. Kim, S. Yun, H. Youn, S. Kam, J. Tanguay, and I. A. Cunningham, Singleshot dual-energy x-ray imaging with a flat-panel sandwich detector for preclinical imaging, Cur. Appl. Phys. 14(12), pp , J.-T. Hsu, Y.-J. Chen, J.-T. Ho, H.-L. Huang, S.-P. Wang, F.-C. Cheng, J. Wu, and M.-T. Tsai, A comparison of micro-ct and dental CT in assessing cortical bone morphology and trabecular bone microarchitecture, PLoS ONE 9, p. e107545, N. A. Hanson and C. M. Bagi, Alternative approach to assessment of bone quality using micro-computed tomography, Bone 35(1), pp , E. L. Ritman, Molecular imaging in small animals - Roles for micro-ct, J. Cell. Biochem. 87(S39), pp , S. C. Lee, H. K. Kim, I. K. Chun, M. H. Cho, S. Y. Lee, and M. H. Cho, A flat-panel detector based micro- CT system: Performance evaluation for small-animal imaging, Phy. Med. Biol. 48(24), pp , M. J. Paulus, S. S. Gleason, S. J. Kennel, P. R. Hunsicker, and D. K. Johnson, High resolution x-ray computed tomography: An emerging tool for small animal cancer research, Neoplasia 2(1-2), pp , D. W. Holdsworth and M. M. Thornton, Micro-CT in small animal and specimen imaging, Trends Biotechnol. 20(8), pp. S34 S39, E. L. Ritman, Micro-computed tomography - Current status and developments, Annu. Rev. Biomed. Eng. 6(1), pp , D. Clark and C. Badea, Micro-CT of rodents: State-of-the-art and future perspectives, Phys. Medica 30(6), pp , L. A. Feldkamp, L. C. Davis, and J. W. Kress, Practical cone-beam algorithm, J. Opt. Soc. Am. A 1, pp , Jun 1984.

MTF and NPS of single-shot dual-energy sandwich detectors

MTF and NPS of single-shot dual-energy sandwich detectors MTF and NPS of single-shot dual-energy sandwich detectors Junwoo Kim, a Dong Woon Kim, a Hanbean Youn, b,c Ho Kyung Kim a,c a School of Mechanical Engineering, Pusan National University, Busan 609-735,

More information

VOLUMETRIC computed tomography (CT) based on flatpanel

VOLUMETRIC computed tomography (CT) based on flatpanel IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE 2009 1179 Radiation Effects on the Performance of CMOS Photodiode Array Detectors and the Role of Gain-Offset Corrections Ho Kyung Kim, Min Kook

More information

Performance Evaluation of a Flat-panel Detector-based Microtomography System for Small-animal Imaging

Performance Evaluation of a Flat-panel Detector-based Microtomography System for Small-animal Imaging Performance Evaluation of a Flat-panel Detector-based Microtomography System for Small-animal Imaging Ho Kyung Kim, Sang Chul Lee, In Kon Chun, Myung Hye Cho, Min Hyoung Cho, Member, IEEE, Soo Yeol Lee

More information

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.125 pissn 2508-4445, eissn 2508-4453 Optimization of Energy Modulation Filter for Dual Energy CBCT

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Digital Detector Array Image Quality for Various GOS Scintillators

Digital Detector Array Image Quality for Various GOS Scintillators Digital Detector Array Image Quality for Various GOS Scintillators More info about this article: http://www.ndt.net/?id=22768 Brian S. White 1, Mark E. Shafer 2, William H. Russel 3, Eric Fallet 4, Jacques

More information

The effect of focal spot size on the spatial resolution of variable resolution X-ray CT scanner

The effect of focal spot size on the spatial resolution of variable resolution X-ray CT scanner The effect of focal spot size on the spatial resolution of variable resolution X-ray CT scanner H. Arabi, A.R. Kamali Asl *, S.M. Aghamiri Iran. J. Radiat. Res., 2010; 8 (1): 37-43 Radiation Medicine Department,

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

An aging study of the signal and noise characteristics in large-area CMOS detectors

An aging study of the signal and noise characteristics in large-area CMOS detectors An aging study of the signal and noise characteristics in large-area CMOS detectors Jong Chul Han a, Seungman Yun a, Chang Hwy Lim a, Tae Gyun Youm a, Sung Kyn Heo b, Tae Woo Kim b, Ian Cunningham c, Ho

More information

Geometric image distortion in flat-panel X-ray detectors and its influence on the accuracy of CT-based dimensional measurements

Geometric image distortion in flat-panel X-ray detectors and its influence on the accuracy of CT-based dimensional measurements Geometric image distortion in flat-panel X-ray detectors and its influence on the accuracy of CT-based dimensional measurements Daniel Weiß, Ronald Lonardoni, Andreas Deffner, Christoph Kuhn Carl Zeiss

More information

Four-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development

Four-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development ORIGINAL ARTICLE ORIGINAL ARTICLE Radiation Medicine: Vol. 21 No. 1, 17 22 p.p., 2003 Four-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development Masahiro Endo,* Takanori Tsunoo,*

More information

Data. microcat +SPECT

Data. microcat +SPECT Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for

More information

Reconstruction Filtering in Industrial gamma-ray CT Application

Reconstruction Filtering in Industrial gamma-ray CT Application Reconstruction Filtering in Industrial gamma-ray CT Application Lakshminarayana Yenumula *, Rajesh V Acharya, Umesh Kumar, and Ashutosh Dash Industrial Tomography and Instrumentation Section, Isotope Production

More information

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for

More information

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

Scatter Correction by Modulation of Primary Radiation in Industrial X-ray CT: Beam-hardening Effects and their Correction

Scatter Correction by Modulation of Primary Radiation in Industrial X-ray CT: Beam-hardening Effects and their Correction International Symposium on Digital Industrial Radiology and Computed Tomography - Mo.3.2 Scatter Correction by Modulation of Primary Radiation in Industrial X-ray CT: Beam-hardening Effects and their Correction

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

X-RAY COMPUTED TOMOGRAPHY

X-RAY COMPUTED TOMOGRAPHY X-RAY COMPUTED TOMOGRAPHY Bc. Jan Kratochvíla Czech Technical University in Prague Faculty of Nuclear Sciences and Physical Engineering Abstract Computed tomography is a powerful tool for imaging the inner

More information

Amorphous Selenium Direct Radiography for Industrial Imaging

Amorphous Selenium Direct Radiography for Industrial Imaging DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial

More information

DOUBLE gated in vivo small animal cone beam micro

DOUBLE gated in vivo small animal cone beam micro Low Dose Phase Correlated Cone Beam Micro CT of Small Animals Stefan Sawall, Frank Bergner, Robert Lapp, Markus Mronz, Marek Karolczak, Andreas Hess, and Marc Kachelrieß, Member, IEEE I. INTRODUCTION DOUBLE

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC R. Bellazzini a,b, G. Spandre a*, A. Brez a, M. Minuti a, M. Pinchera a and P. Mozzo b a INFN Pisa

More information

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Takeshi Fujiwara 1, Yuki Mitsuya 2, Hiroyuki Takahashi 2, and Hiroyuki Toyokawa 2 1 National Institute of Advanced Industrial

More information

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices Moving from film to digital: A study of digital x-ray benefits, challenges and best practices H.U. Pöhler 1 and N. D Ademo 2 DÜRR NDT GmbH & Co. KG, Höpfigheimer Straße 22, Bietigheim-Bissingen, 74321,

More information

Detector technology in simultaneous spectral imaging

Detector technology in simultaneous spectral imaging Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become

More information

diagnostic examination

diagnostic examination RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011

More information

Thermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source

Thermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source Energy Level (ev) Multi-beam x-ray source array based on carbon nanotube field emission O. Zhou, JP Lu, X. Calderon-Colon, X. Qian, G. Yang, G. Cao, E. Gidcumb, A. Tucker, J. Shan University of North Carolina

More information

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) 2/ Overview Digital

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

Niklas Norrby 12/17/2010

Niklas Norrby 12/17/2010 LINKÖPINGS UNIVERSITET Nanotomography Synchrotron radiation course project Niklas Norrby 12/17/2010 Introduction Tomography is a method to image three-dimensional objects by illumination from different

More information

DALLA LUCE VISIBILE AI RAGGI X: NUOVI RIVELATORI DI IMMAGINI PER RAGGI X A DISCRIMINAZIONE IN ENERGIA ED APPLICAZIONI

DALLA LUCE VISIBILE AI RAGGI X: NUOVI RIVELATORI DI IMMAGINI PER RAGGI X A DISCRIMINAZIONE IN ENERGIA ED APPLICAZIONI DALLA LUCE VISIBILE AI RAGGI X: NUOVI RIVELATORI DI IMMAGINI PER RAGGI X A DISCRIMINAZIONE IN ENERGIA ED APPLICAZIONI D. Pacella ENEA - Frascati LIMS, Frascati 14-15 ottobre 2015 Come per la fotografia:

More information

Photon Counting and Energy Discriminating X-Ray Detectors - Benefits and Applications

Photon Counting and Energy Discriminating X-Ray Detectors - Benefits and Applications 19 th World Conference on Non-Destructive Testing 2016 Photon Counting and Energy Discriminating X-Ray Detectors - Benefits and Applications David WALTER 1, Uwe ZSCHERPEL 1, Uwe EWERT 1 1 BAM Bundesanstalt

More information

Suppression of metal artifacts using image-based monoenergetic DECT imaging

Suppression of metal artifacts using image-based monoenergetic DECT imaging Suppression of metal artifacts using image-based monoenergetic DECT imaging Poster No.: C-0519 Congress: ECR 2011 Type: Scientific Paper Authors: B. Krauss, B. Schmidt, M. Sedlmair, T. Flohr; Forchheim/DE

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

Tomosynthesis. Energy Subtraction. Long View Imaging

Tomosynthesis. Energy Subtraction. Long View Imaging Freedom and Flexibility in Imaging Experience a wide range of applications targeted to improve diagnostic capabilities combined with a precise design that facilitates imaging. With unique image processing

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

Distributed source x-ray tube technology for tomosynthesis imaging

Distributed source x-ray tube technology for tomosynthesis imaging Distributed source x-ray tube technology for tomosynthesis imaging Authors: F. Sprenger a*, X. Calderon-Colon b, Y. Cheng a, K. Englestad a, J. Lu b, J. Maltz c, A. Paidi c, X. Qian b, D. Spronk a, S.

More information

Philip Sperling. Sales Science and New Materials, YXLON International GmbH, Essener Bogen 15, Hamburg, Germany.

Philip Sperling. Sales Science and New Materials, YXLON International GmbH, Essener Bogen 15, Hamburg, Germany. A new generation of x-ray computed tomography devices for quality inspection and metrology inspection in the field of additive manufacturing and other sciences Philip Sperling Sales Science and New Materials,

More information

K-edge subtraction X-ray imaging with a pixellated spectroscopic detector

K-edge subtraction X-ray imaging with a pixellated spectroscopic detector K-edge subtraction X-ray imaging with a pixellated spectroscopic detector Silvia Pani Department of Physics, University of Surrey Summary Hyperspectral imaging K-edge subtraction X-ray imaging for mammography

More information

Metal Artifact Reduction for Orthopedic Implants (O-MAR)

Metal Artifact Reduction for Orthopedic Implants (O-MAR) Metal Artifact Reduction for Orthopedic Implants () Summary Since the inception of CT, numerous methods have been proposed to suppress metal artifacts with varying degrees of success. 1-4 (Metal Artifact

More information

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

More information

CT Basics: Data Acquisition Module 3

CT Basics: Data Acquisition Module 3 Module 3 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

Maximizing clinical outcomes

Maximizing clinical outcomes Maximizing clinical outcomes Digital Tomosynthesis Dual Energy Subtraction Automated Long Length Imaging Improved image quality at a low dose Xray Xray Patented ISS capture technology promotes high sensitivity

More information

Development of a large-area CMOS-based detector for real-time x-ray imaging

Development of a large-area CMOS-based detector for real-time x-ray imaging Development of a large-area CMOS-based detector for real-time x-ray imaging Sung Kyn Heo a, Sung Kyu Park a, Sung Ha Hwang a, Dong Ak Im a, Jari Kosonen a, Tae Woo Kim a, Seungman Yun b, Ho Kyung Kim b*

More information

Uncertainty in CT Metrology: Visualizations for Exploration and Analysis of Geometric Tolerances

Uncertainty in CT Metrology: Visualizations for Exploration and Analysis of Geometric Tolerances Uncertainty in CT Metrology: Visualizations for Exploration and Analysis of Geometric Tolerances Artem Amirkhanov 1, Bernhard Fröhler 1, Michael Reiter 1, Johann Kastner 1, M. Eduard Grӧller 2, Christoph

More information

Synchrotron X-ray tomographic microscopy Theory vs. practice

Synchrotron X-ray tomographic microscopy Theory vs. practice Synchrotron X-ray tomographic microscopy Theory vs. practice Federica Marone Swiss Light Source, Paul Scherrer Institut, Villigen, Switzerland Theory Radon transform Rf x = Beer-Lambert law I E = I 0 (E)e

More information

Applications of the Cracow X-Ray Microprobe in Tomography

Applications of the Cracow X-Ray Microprobe in Tomography Vol. 115 (2009) ACTA PHYSICA POLONICA A No. 2 Proceedings of the XLII Zakopane School of Physics, Zakopane 2008 Applications of the Cracow X-Ray Microprobe in Tomography J. Bielecki a, S. Bożek a,b, J.

More information

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc.

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. An Introduction to TG-142 Imaging QA Using Standard Imaging Products Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. Goals Understand the nature and intent of TG 142 imaging

More information

Mammography: Physics of Imaging

Mammography: Physics of Imaging Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness

More information

Efficacy of Beam-Hardening Correction for Vascular Morphometry Using 3D Conebeam Micro-CT

Efficacy of Beam-Hardening Correction for Vascular Morphometry Using 3D Conebeam Micro-CT Marquette University e-publications@marquette Biomedical Engineering Faculty Research and Publications Biomedical Engineering, Department of 1-1-1999 Efficacy of Beam-Hardening Correction for Vascular

More information

Digital Radiography : Flat Panel

Digital Radiography : Flat Panel Digital Radiography : Flat Panel Flat panels performances & operation How does it work? - what is a sensor? - ideal sensor Flat panels limits and solutions - offset calibration - gain calibration - non

More information

An EM Reconstruction with Improved Signal-to. to-noise Ratio for Coded Aperture Imaging

An EM Reconstruction with Improved Signal-to. to-noise Ratio for Coded Aperture Imaging An EM Reconstruction with Improved Signal-to to-noise Ratio for Coded Aperture Imaging Cynthia Tozian, PhD UMass Lowell Bristol-Myers Squibb Medical Imaging Young Investigators Symposium April 12, 2006

More information

Seminar 8. Radiology S8 1

Seminar 8. Radiology S8 1 Seminar 8 Radiology Medical imaging. X-ray image formation. Energizing and controlling the X-ray tube. Image detectors. The acquisition of analog and digital images. Digital image processing. Selected

More information

TDI Imaging: An Efficient AOI and AXI Tool

TDI Imaging: An Efficient AOI and AXI Tool TDI Imaging: An Efficient AOI and AXI Tool Yakov Bulayev Hamamatsu Corporation Bridgewater, New Jersey Abstract As a result of heightened requirements for quality, integrity and reliability of electronic

More information

Migrating from traditional to Digital Radiography in Aerospace

Migrating from traditional to Digital Radiography in Aerospace Migrating from traditional to Digital Radiography in Aerospace More info about this article: http://www.ndt.net/?id=22663 Abstract Lennart Schulenburg VisiConsult X-ray System & Solutions GmbH Brandenbrooker

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Observational Astronomy

Observational Astronomy Observational Astronomy Instruments The telescope- instruments combination forms a tightly coupled system: Telescope = collecting photons and forming an image Instruments = registering and analyzing the

More information

PD233: Design of Biomedical Devices and Systems

PD233: Design of Biomedical Devices and Systems PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/

More information

Development of Solid-State Detector for X-ray Computed Tomography

Development of Solid-State Detector for X-ray Computed Tomography Proceedings of the Korea Nuclear Society Autumn Meeting Seoul, Korea, October 2001 Development of Solid-State Detector for X-ray Computed Tomography S.W Kwak 1), H.K Kim 1), Y. S Kim 1), S.C Jeon 1), G.

More information

Evaluation of a Chip LED Sensor Module at 770 nm for Fat Thickness Measurement of Optical Tissue Phantoms and Human Body Tissue

Evaluation of a Chip LED Sensor Module at 770 nm for Fat Thickness Measurement of Optical Tissue Phantoms and Human Body Tissue Journal of the Korean Physical Society, Vol. 51, No. 5, November 2007, pp. 1663 1667 Evaluation of a Chip LED Sensor Module at 770 nm for Fat Thickness Measurement of Optical Tissue Phantoms and Human

More information

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity) Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular

More information

LECTURE 1 The Radiographic Image

LECTURE 1 The Radiographic Image LECTURE 1 The Radiographic Image Prepared by:- KAMARUL AMIN ABDULLAH @ ABU BAKAR UiTM Faculty of Health Sciences Medical Imaging Department 11/23/2011 KAMARUL AMIN (C) 1 Lesson Objectives At the end of

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

Modeling detector performance in digital mammography using the linear cascaded systems approach

Modeling detector performance in digital mammography using the linear cascaded systems approach 51 Modeling detector performance in digital mammography using the linear cascaded systems approach V. Spyropoulou a,b, N. Kalivas a,b, A. Gaitanis a,b, C. Michail a,b, G. Panayiotakis b, I. Kandarakis

More information

Digital Images & Image Quality

Digital Images & Image Quality Introduction to Medical Engineering (Medical Imaging) Suetens 1 Digital Images & Image Quality Ho Kyung Kim Pusan National University Radiation imaging DR & CT: x-ray Nuclear medicine: gamma-ray Ultrasound

More information

CT Basics: Image Quality Module 6

CT Basics: Image Quality Module 6 Module 6 For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions. Any broadcast,

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893 Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage

More information

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE Takeyuki Hashimoto 1), Morio Onoe 2), Hiroshi Nakamura 3), Tamon Inouye 4), Hiromichi Jumonji 5), Iwao Takahashi 6); 1)Yokohama Soei

More information

Electronic Noise in CT Detectors: Impact on Image Noise and Artifacts

Electronic Noise in CT Detectors: Impact on Image Noise and Artifacts Medical Physics and Informatics Original Research Duan et al. Electronic Noise in CT Detectors Medical Physics and Informatics Original Research Xinhui Duan 1 Jia Wang 1,2 Shuai Leng 1 ernhard Schmidt

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

High Energy Digital Radiography & 3D-CT for Industrial Systems

High Energy Digital Radiography & 3D-CT for Industrial Systems DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France High Energy Digital Radiography & 3D-CT for Industrial Systems Non-Destructive

More information

Investigation of Effective DQE (edqe) parameters for a flat panel detector

Investigation of Effective DQE (edqe) parameters for a flat panel detector Investigation of Effective DQE (edqe) parameters for a flat panel detector Poster No.: C-1892 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit D. Bor 1, S. Cubukcu 1, A. Yalcin 1, O.

More information

Related topics Beam hardening, cupping effect, Beam hardening correction, metal artefacts, photon starvation

Related topics Beam hardening, cupping effect, Beam hardening correction, metal artefacts, photon starvation Beam hardening and metal artefacts TEP Related topics Beam hardening, cupping effect, Beam hardening correction, metal artefacts, photon starvation Principle X-ray sources produce a polychromatic spectrum

More information

Phase Imaging Using Focused Polycapillary Optics

Phase Imaging Using Focused Polycapillary Optics Phase Imaging Using Focused Polycapillary Optics Sajid Bashir, Sajjad Tahir, Jonathan C. Petruccelli, C.A. MacDonald Dept. of Physics, University at Albany, Albany, New York Abstract Contrast in conventional

More information

Simulated validation and quantitative analysis of the blur of an integral image related to the pickup sampling effects

Simulated validation and quantitative analysis of the blur of an integral image related to the pickup sampling effects J. Europ. Opt. Soc. Rap. Public. 9, 14037 (2014) www.jeos.org Simulated validation and quantitative analysis of the blur of an integral image related to the pickup sampling effects Y. Chen School of Physics

More information

Lecture 25. Wind Lidar (3) Direct Detection Doppler Lidar

Lecture 25. Wind Lidar (3) Direct Detection Doppler Lidar Lecture 25. Wind Lidar (3) Direct Detection Doppler Lidar Overview of Direct Detection Doppler Lidar (DDL) Fringe imaging DDL Scanning FPI DDL FPI edge-filter DDL Iodine absorption-line edge-filter DDL

More information

Performance of Image Intensifiers in Radiographic Systems

Performance of Image Intensifiers in Radiographic Systems DOE/NV/11718--396 LA-UR-00-211 Performance of Image Intensifiers in Radiographic Systems Stuart A. Baker* a, Nicholas S. P. King b, Wilfred Lewis a, Stephen S. Lutz c, Dane V. Morgan a, Tim Schaefer a,

More information

Objective Evaluation of Radiographic Contrast- Enhancement Masks

Objective Evaluation of Radiographic Contrast- Enhancement Masks Chapter 8 Objective Evaluation of Radiographic Contrast- Enhancement Masks The development and application of radiographic contrast-enhancement masks (RCMs) in digital radiography (DR) were discussed in

More information

SYLLABUS. 1. Identification of Subject:

SYLLABUS. 1. Identification of Subject: SYLLABUS Date/ Revision : 30 January 2017/1 Faculty : Life Sciences Approval : Dean, Faculty of Life Sciences SUBJECT : Biophysics 1. Identification of Subject: Name of Subject : Biophysics Code of Subject

More information

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Poster No.: C-2205 Congress: ECR 2012 Type: Scientific Paper

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

X-rays in medical diagnostics

X-rays in medical diagnostics X-rays in medical diagnostics S.Dolanski Babić 2017/18. History W.C.Röntgen (1845-1923) discovered a new type of radiation Nature, Jan. 23. 1896.; Science, Feb.14. 1896. X- rays: Induced the ionization

More information

Development of a Virtual Simulation Environment for Radiation Treatment Planning

Development of a Virtual Simulation Environment for Radiation Treatment Planning Journal of Medical and Biological Engineering, 25(2): 61-66 61 Development of a Virtual Simulation Environment for Radiation Treatment Planning Tai-Sin Su De- Kai Chen Wen-Hsu Sung Ching-Fen Jiang * Shuh-Ping

More information

X-ray detectors in healthcare and their applications

X-ray detectors in healthcare and their applications X-ray detectors in healthcare and their applications Pixel 2012, Inawashiro September 4th, 2012 Martin Spahn, PhD Clinical applications of X-ray imaging Current X-ray detector technology (case study radiography

More information

Rad-icon Imaging Corp A Division of DALSA Corporation

Rad-icon Imaging Corp A Division of DALSA Corporation Rad-icon Imaging Corp A Division of DALSA Corporation el: 408-486-0886 Fax: 408-486-0882 www.rad-icon.com PRELIMINARY DAA SHEE SkiaGraph 8 Very Large Area X-Ray Camera Key Features: Active area of 20 cm

More information

REAL-TIME X-RAY IMAGE PROCESSING; TECHNIQUES FOR SENSITIVITY

REAL-TIME X-RAY IMAGE PROCESSING; TECHNIQUES FOR SENSITIVITY REAL-TIME X-RAY IMAGE PROCESSING; TECHNIQUES FOR SENSITIVITY IMPROVEMENT USING LOW-COST EQUIPMENT R.M. Wallingford and J.N. Gray Center for Aviation Systems Reliability Iowa State University Ames,IA 50011

More information

X-ray phase-contrast imaging

X-ray phase-contrast imaging ...early-stage tumors and associated vascularization can be visualized via this imaging scheme Introduction As the selection of high-sensitivity scientific detectors, custom phosphor screens, and advanced

More information

Clinical Importance on CT

Clinical Importance on CT 183 Truncated-View Clinical Importance on CT..,..,.... Artifacts:.... James L. Lehr1 A truncated-view artifact in CT is produced whenever any part of the patient or imaged object is present in some but

More information

Manufacturing Metrology Team

Manufacturing Metrology Team The Team has a range of state-of-the-art equipment for the measurement of surface texture and form. We are happy to discuss potential measurement issues and collaborative research Manufacturing Metrology

More information

Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography

Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography ECNDT 26 - Poster 17 Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography Khairul Anuar MOHD SALLEH, Ab. Razak HAMZAH and Mohd Ashhar

More information

CHAPTER 10. DIGITAL SUBTRACTION ANGIOGRAPHY (DSA) 10.1 Introduction

CHAPTER 10. DIGITAL SUBTRACTION ANGIOGRAPHY (DSA) 10.1 Introduction PHYSICS OF MEDICAL X-RAY IMAGING (1)Chapter 10 CHAPTER 10. DIGITAL SUBTRACTION ANGIOGRAPHY (DSA) 10.1 Introduction In this chapter we will discuss a technique that is known as digital subtraction angiography

More information

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING DIGITAL IMAGE PROCESSING IN X-RAY IMAGING Shalini Kumari 1, Bachan Prasad 2,Aliya Nasim 3 Department of Electronics And Communication Engineering R.V.S College of Engineering & Technology, Jamshedpur,

More information

Diffraction-enhanced X-ray Imaging (DEXI) Medical Solutions. More information using less radiation

Diffraction-enhanced X-ray Imaging (DEXI) Medical Solutions. More information using less radiation Diffraction-enhanced X-ray Imaging (DEXI) Medical Solutions More information using less radiation Medical Small Animal Security NDE/NDT Diffraction-Enhanced X-ray Imaging Medical Solutions Safe non-invasive

More information

This document is published in:

This document is published in: Institutional Repository This document is published in: Bo Yu (ed.) (01). 01 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC): Anaheim, California, USA. October 9 - November 3, 01.

More information

Principles of CT scan

Principles of CT scan Related topics Detector calibration, saturation, CT acquisition, CT reconstruction Principle X-ray computed tomography consists of using X-rays that are converted to a digital signal by a detector and

More information