Evaluation of a Chip LED Sensor Module at 770 nm for Fat Thickness Measurement of Optical Tissue Phantoms and Human Body Tissue

Size: px
Start display at page:

Download "Evaluation of a Chip LED Sensor Module at 770 nm for Fat Thickness Measurement of Optical Tissue Phantoms and Human Body Tissue"

Transcription

1 Journal of the Korean Physical Society, Vol. 51, No. 5, November 2007, pp Evaluation of a Chip LED Sensor Module at 770 nm for Fat Thickness Measurement of Optical Tissue Phantoms and Human Body Tissue Dong-Su Ho and Beop-Min Kim Department of Biomedical Engineering, Yonsei University, Wonju In Duk Hwang and Kunsoo Shin Bio & Health Laboratory, Samsung Advanced Institute of Technology, Suwon (Received 20 March 2007) We measured the fat thickness noninvasively by using a diffuse optical method for tissuesimulating phantoms and human tissues. A light source module composed of 770 nm low-power chip LEDs and a photo-detector were used in this study. The optical tissue phantoms were composed of a fat and a muscle layer made with gels with appropriate absorption/scattering coefficients. The fat thickness was varied from several to 30 mm. After a proper calibration procedure, we used this system to conduct human studies. Based on this preliminary study, noninvasive fat thickness measurements are possible with a simple curve-fitting procedure. PACS numbers: Be Keywords: Fat thickness, Diffuse optics, LED I. INTRODUCTION II. EXPERIMENTS There are various techniques used to study the lean body mass and subcutaneous fat distribution [1]. Imaging methods such as magnetic resonance imaging (MRI), computer tomography (CT), dual-energy X-ray absorptionmetry (DEXA), and ultrasound imaging are precise and accurate techniques. However, these methods are applicable in very limited cases due to the high cost and the risk of radiological burden (CT, DEXA). A noninvasive optical approach, such as diffuse optical spectroscopy, has many advantages over these technologies because it costs less and poses no radiation toxicity [2 4]. More importantly, it can be constructed in a small size and incorporated with mobile devices. If a portable, miniaturized optical sensor module is to be developed for human fat thickness measurements, a hardware layout, including an appropriate source-detector-distance (SD) and LED power, has to be configured. For this purpose, we analyzed the diffusely reflected light intensity at different SD s and LED powers by using tissue-simulating phantoms that had optical properties similar to those of human tissues at a wavelength of 770 nm. indhwang@samsung.com; Fax: The optical properties of the fat and the muscle of skin at 770 nm are shown in Table 1, which shows that the fat layer is highly scattering while the muscle is highly absorbing [2 4]. To accurately simulate the human skin, it is necessary to include the epidermal and the dermal layers, too. However, considering that those layers are very thin and they may contribute only to a linear attenuation of the diffusely reflected light for all SD s, we decided to construct a simple skin model with two layers (fat and muscle). The fat and the muscle phantoms were made separately from gelatin (Sigma Corp.) mixed with intralipid solution as a scatterer and neutral-red powder as an ab Phantom Preparation Table 1. Optical properties of human fat and muscle at 770 nm [2,3,6]. Tissue Layer µ a (1/cm) µ s (1/cm) Fat Phantom (Intralipid 20 %) Muscle Phantom 0.3 7

2 Journal of the Korean Physical Society, Vol. 51, No. 5, November 2007 Fig. 3. LED optical property acquisition module. 2. Experimental Setup Fig. 1. Photograph of fabricated fat (left) and muscle (right) phantoms. Fig. 2. Schematic of the two-layer tissue phantom experiment. sorber. The optical properties of the intralipid solution and the neutral-red powder were pre-determined by using an inverse adding-doubling (IAD) method from the transmittance and the reflectance measurements using an integrating sphere [5]. All measurements were made at 770 nm wavelength. Biological tissues are well known to be less opaque in the wavelength range, called the optical window of 600 1,100 nm. We picked the wavelength of 770 nm simply because less expensive and effective source module were available at that wavelength [7 14]. The final fat and muscle phantoms had optical properties similar to those of human tissues, as shown in Table 1. Once the phantoms had been made, we measured the optical properties of the phantoms separately to make sure that the phantoms had the same optical properties as designed. After each phantom had been made, a sliced fat layer with a specific thickness was placed over the absorber layer to obtain a two-layered skin model. The thickness of the fat layer were varied from 2 mm to 30 mm while the muscle layer was fixed at a thickness of 35 mm. The muscle thickness is not important because light cannot reach the bottom surface of the absorber phantom anyway. Figure 1 shows a photograph of the fabricated fat and muscle phantoms. The width and the length of the phantoms are 110 mm and 80 mm, respectively. Figure 2 shows the experimental setup for testing our sensor module with a two-layer phantom, which consists of a fat layer on top of a muscle layer. Two kinds of LEDs with a center wavelength of 770 nm were used in our experiment. One was a lamp LED (viewing angle 20 degrees), and the other was a miniaturized chip LED of the surface mount type (viewing angle 120 degrees). Figure 3 shows a picture of the chip LED module system. The frequency of the LED pulse operation was about 4.8 khz, and its duty was less than 10 %. In the case of the lamp LED, the source distances from the detector were 5, 10, 15, 20 and 25 mm, respectively. The applied current to each LED was 80 ma. In the case of the chip LED, the light sources were separated from the detector by 5 mm (S1), 10 mm (S2) and 20 mm (S3). As the SD increases, the detected intensity decreased exponentially. To compensate for this, multiple LEDs were used at longer distances from the detector. Figure 3 shows three LEDs were used at a distance of 10 mm and 5 LEDs were used at 20 mm. III. RESULTS 1. MonteCarlo Simulation In our Monte Carlo simulation, each layer of the model structure was assumed to be infinitely wide. The thickness of the fat layer was varied from 2 mm to 30.0 mm, in a step of 2.0 mm. The muscle layer was placed underneath the fat layer, and the muscle thickness was set large enough to guarantee that no light reached the bottom surface of the muscle layer. The refractive indices of the fat and the muscle layers were both set as 1.37, the diameter of the source beam was 2.5 mm, and the diffuse reflectance was measured up to a radial position 40 mm from the source. The number of photons was 105. Figure 4 show the simulation results on the detected diffuse reflectance as a function of the fat thickness at 770 nm. As expected, it shows an exponential decay of the diffuse reflectance as a function of the source-detector (SD) distance [6]. It also shows that the overall diffuse reflectance (integrated area under the curve) increases

3 Evaluation of a Chip LED Sensor Module at 770 nm for Fat Thickness Dong-Su Ho et al Fig. 4. Simulation results of the reflected intensity for different fat thicknesses as a function of the source-detector (SD) distances. Fig. 6. Detected intensity for various source-detector distances (SDs) SDs as a function of the fat thickness by using a llamp LED. sults are presented below. 2. Phantom Experiment Using Lamp LED s Fig. 5. Simulation results of the detected intensity for various source-detector distances (SDs) as a function of the fat thickness for thicknesses from 0.2 cm to 3 cm. as the thickness of the fat increases because the fat has much a lower absorption coefficient than muscle does. As the fat layer thickness increases, more light has a chance to travel through the fat and to be diffuse reflected on the surface. Figure 5 shows the simulation result for the relationship between the fat thickness and the diffuse reflectance for various source-detector (SD) separations. The diffuse reflectance increases gradually as the fat thickness increases for all SD s. The signal is ultimately saturated because most photons pass through the fat layer as the fat layer becomes thicker. For larger SD s, the photons pass though the deeper region before detection, and the signal is saturated for larger fat thickness. We verified our observation via a phantom experiment, and the re- We first investigated the sensor module, which consists of one detector and four lamp LED s (770 nm, divergence angle 20 degrees, 3 mm diameter) placed at various SD s. Starting 10 mm from the detector, the LED s are placed 5 mm apart from each other. The maximum distance between the detector and the source is 25 mm. The same current and amplifier gain were applied to all the LED s. Because we collected all photons backscattered with the same gain, the magnitudes at SD s of 20 mm and 25 mm were much lower than those of SD s of 10 mm and 5 mm. In Figure 6, the detected intensity at different SD s as a function of the fat thickness, shows trends similar to those of the simulated results (Figure 5). From these results, it is obvious that the fat thickness can be well-estimated only up to a thickness of 10 mm or so when a single LED is placed at a SD of 10 mm. It is possible, but difficult, to measure a fat thickness thicker than 10 mm. Especially, if the SD becomes too large, the signal magnitude becomes too small to be measured. It is important to notice that the SD has to be larger so that a thicker fat layer can be studied. Therefore, a large SD is needed, and more optical power is needed for that large SD. 3. Phantom Experiment Using Chip LED s Based on what we observed in the lamp LED experiment, we built and tested another sensor module, which consist of one detector and a series of chip LED s (770 nm). The light sources (770 nm, chip LED s, divergence

4 Journal of the Korean Physical Society, Vol. 51, No. 5, November 2007 Fig. 7. Reflected intensity variation for the three different separations between the source and the detector as a function of the fat thickness. as the fat became thicker. For a SD of 10 mm, the maximum measurable thickness was approximately 20 mm. However, since the optical power at large SD (20 mm) was increased, a fat thickness up to 30 mm could still be estimated by using this sensor module. In a separate experiment, to investigate the influence of the LED power, we measured the reflected intensity by using three LEDs and five LEDs for the same SD of 20 mm and for the same amplifier gain. In this test, all hardware conditions, except for the numbers of chip LED s, were the same. The results are shown in Figure 8. We observed that even if the signals showed similar increasing patterns, higher power (more chips) could produce a clearer signal. From the 3rd-order polynomial curve-fitting results of Figure 8, the data can be shown to be nicely fitted to a simple polynomial. Thus, we conclude that this sensor module with a larger SD and more optical power can measure a large range of fat thicknesses up to 30 mm. IV. CONCLUSION Fig. 8. Reflected intensity and the fitting results for two different numbers of LEDs as a function of the fat thickness at a fixed SD of 20 mm. angle of 120 degrees) were separated from a single Si detector by different distances of S1 (5 mm), S2 (10 mm) and S3 (20 mm). Since the diffuse reflectance decreases exponentially with radial distance, we used multiple LED s at larger distances (3 chips for S2 and 5 chips for S3). The cycle of the LED pulse operation was approximately 4.8 khz, and its duty was less than 10 %. We adjusted the gain of the amplifier so that the detected intensity (V) at fat thicknesses up to 30 mm could be detected for a SD of 20 mm with a maximum intensity signal of 3 V. The amplifier gain for the SD s of 5 mm, 10 mm and 20 mm were 1, 1.5 and 5, respectively. All data were collected at a LED driving current of 60 ma. Figure 7 shows the reflected intensity as a function of the fat thickness at three different separations between the source and the detector. The result is very similar to those of the lamp LED experiment. For a SD of 5 mm, the detected intensity rapidly increased up to 10 mm of fat thickness. After a thickness of 10 mm, it saturated We built an optical phantom by using gelatin mixed with intralipid solution as a scatterer and neutral-red powder as an absorber. The fat and muscle phantoms were made using the data in other literature. A Monte-Carlo simulation was performed, and the results were verified using the phantom experiments. We showed that a simple sensor module composed of a single detector and multiple LED s could measure fat thicknesses up to 30 mm. In further studies, the same module will be used for fat thickness measurements in humans. ACKNOWLEDGMENTS This study was supported by a grant of the Korea Health 21 R & D Project, Ministry of Health & Welfare, Republic of Korea (02-PJ3-PG6-EV ). This work was also supported by the Brain Research Center of the 21st Century Frontier Research Program funded by the Ministry of Science and Technology (M103KV K ). REFERENCES [1] S. J. Wallner, N. Luschnigg, W. J. Schnedl, T. Lahousen, K. Crailsheim, R. Möller, E. Tafeit and R. Horejsi, International J. Obesity 28, 1143 (2004). [2] Y. Yang, O. O. Soyemi, M. R. Landry and B. R. Soller, Opt. Express 13, 1570 (2005). [3] A. Kienle, L. Lilge, M. S. Patterson, B. C. Wilson, R. Hibst and R. Steiner, SPIE 2326, 212 (1995).

5 Evaluation of a Chip LED Sensor Module at 770 nm for Fat Thickness Dong-Su Ho et al [4] W. Cui and L. E. Ostrander, IEEE Trans Biomedical Eng. 39, 194 (1992). [5] B.-M. Kim, J. Korean Phys. Soc. 44, 427 (2004). [6] T. J. Farrell, M. S. Patterson and B. C. Wilson, Med. Phys. 19, 879 (1992). [7] J. T. Seo, U. Hommerich, J. D. Mackenzie, C. R. Abernathy and J. M. Zavada, J. Korean Phys. Soc. 36, 311 (2000). [8] H. M. Park, I. O. Chae, M. R. Lee, E. C. Jung, Y. J. Rhee and J. M. Lee, J. Korean Phys. Soc. 39, 886 (2001). [9] J. P. Culver, V. Ntziachristos, M. J. Holboke and A. G. Yodh, Opt. Lett. 26, 10 (2001). [10] M. A. Franceschini and D. A. Boas, Neurolimage 21, 372 (2004). [11] D. A. Boas, A. M. Dale and M. A. Franceschini, Neuroimage 23, S275 (2004). [12] J. Kim, C. Choi, H.-S. Soh, D.-S. Ho and B.-M. Kim, J. Korean Phys. Soc. 47, 375 (2005). [13] O.-Y. Jeon, M.-J. Jin, H.-H. Lim, B.-J. Kim and M. S. Cha, J. Korean Phys. Soc. 49, 589 (2006). [14] S. M. Kim, K. Kyhm, K. S. Hong and H.-S. Yang, J. Korean Phys. Soc. 49, 688 (2006).

Ultrasound-modulated optical tomography of absorbing objects buried in dense tissue-simulating turbid media

Ultrasound-modulated optical tomography of absorbing objects buried in dense tissue-simulating turbid media Ultrasound-modulated optical tomography of absorbing objects buried in dense tissue-simulating turbid media Lihong Wang and Xuemei Zhao Continuous-wave ultrasonic modulation of scattered laser light was

More information

Transmission- and side-detection configurations in ultrasound-modulated optical tomography of thick biological tissues

Transmission- and side-detection configurations in ultrasound-modulated optical tomography of thick biological tissues Transmission- and side-detection configurations in ultrasound-modulated optical tomography of thick biological tissues Jun Li, Sava Sakadžić, Geng Ku, and Lihong V. Wang Ultrasound-modulated optical tomography

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

Fig. 1

Fig. 1 PhysicsAndMathsTutor.com 1 1. Fig. 1 shows data for the intensity of a parallel beam of X-rays after penetration through varying thicknesses of a material. intensity / MW m 2 thickness / mm 0.91 0.40 0.69

More information

Dong-Eon Kim, a) Su-Mi Lee, and In-joon Jeon Department of Physics, Pohang University of Science and Technology, Pohang , Korea

Dong-Eon Kim, a) Su-Mi Lee, and In-joon Jeon Department of Physics, Pohang University of Science and Technology, Pohang , Korea Transmission characteristics of multilayer structure in the soft x-ray spectral region and its application to the design of quarter-wave plates at 13 and 4.4 nm Dong-Eon Kim, a) Su-Mi Lee, and In-joon

More information

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.125 pissn 2508-4445, eissn 2508-4453 Optimization of Energy Modulation Filter for Dual Energy CBCT

More information

Fabrication and Usage of a Multi-turn µ-coil and a PR Channel Combined with a Dual-type GMR-SV Device

Fabrication and Usage of a Multi-turn µ-coil and a PR Channel Combined with a Dual-type GMR-SV Device Journal of Magnetics 22(4), 649-653 (2017) ISSN (Print) 1226-1750 ISSN (Online) 2233-6656 https://doi.org/10.4283/jmag.2017.22.4.649 Fabrication and Usage of a Multi-turn µ-coil and a PR Channel Combined

More information

ASSESSMENT OF WALL-THINNING IN CARBON STEEL PIPE BY USING LASER-GENERATED GUIDED WAVE

ASSESSMENT OF WALL-THINNING IN CARBON STEEL PIPE BY USING LASER-GENERATED GUIDED WAVE ASSESSMENT OF WALL-THINNING IN CARBON STEEL PIPE BY USING LASER-GENERATED GUIDED WAVE DOYOUN KIM, YOUNHO CHO * and JOONHYUN LEE Graduate School of Mechanical Engineering, Pusan National University Jangjeon-dong,

More information

Confocal Imaging Through Scattering Media with a Volume Holographic Filter

Confocal Imaging Through Scattering Media with a Volume Holographic Filter Confocal Imaging Through Scattering Media with a Volume Holographic Filter Michal Balberg +, George Barbastathis*, Sergio Fantini % and David J. Brady University of Illinois at Urbana-Champaign, Urbana,

More information

On the dielectric properties of substrates with different surface conditions for submillimeter-wave and terahertz applications

On the dielectric properties of substrates with different surface conditions for submillimeter-wave and terahertz applications Invited Paper On the dielectric properties of substrates with different surface conditions for submillimeter-wave and terahertz applications Kung Bo Ng 1 and Chi Hou Chan 1*, 2 1 State Key Laboratory of

More information

3D near-infrared imaging based on a SPAD image sensor

3D near-infrared imaging based on a SPAD image sensor Zurich Open Repository and Archive University of Zurich Main Library Strickhofstrasse 39 CH-8057 Zurich www.zora.uzh.ch Year: 2011 3D near-infrared imaging based on a SPAD image sensor Mata Pavia, J; Niclass,

More information

Coaxial-type water load for measuring high voltage, high current and short pulse of a compact Marx system for a high power microwave source

Coaxial-type water load for measuring high voltage, high current and short pulse of a compact Marx system for a high power microwave source PHYSICAL REVIEW SPECIAL TOPICS - ACCELERATORS AND BEAMS 12, 113501 (2009) Coaxial-type water load for measuring high voltage, high current and short pulse of a compact Marx system for a high power microwave

More information

A miniature all-optical photoacoustic imaging probe

A miniature all-optical photoacoustic imaging probe A miniature all-optical photoacoustic imaging probe Edward Z. Zhang * and Paul C. Beard Department of Medical Physics and Bioengineering, University College London, Gower Street, London WC1E 6BT, UK http://www.medphys.ucl.ac.uk/research/mle/index.htm

More information

Development of Solid-State Detector for X-ray Computed Tomography

Development of Solid-State Detector for X-ray Computed Tomography Proceedings of the Korea Nuclear Society Autumn Meeting Seoul, Korea, October 2001 Development of Solid-State Detector for X-ray Computed Tomography S.W Kwak 1), H.K Kim 1), Y. S Kim 1), S.C Jeon 1), G.

More information

Supplementary Information. Highly conductive and flexible color filter electrode using multilayer film

Supplementary Information. Highly conductive and flexible color filter electrode using multilayer film Supplementary Information Highly conductive and flexible color filter electrode using multilayer film structure Jun Hee Han 1, Dong-Young Kim 1, Dohong Kim 1, and Kyung Cheol Choi 1,* 1 School of Electrical

More information

Author(s) Seki, D; Namita, T; Kato, Y; Shimiz.

Author(s) Seki, D; Namita, T; Kato, Y; Shimiz. Title A on human body Author(s) Seki, D; Namita, T; Kato, Y; Shimiz Citation 20th Symposium of the International Proceedings (2014): 29-31 Issue Date 2014-05 URL http://hdl.handle.net/2433/187846

More information

Design of the Diffuse Optical Tomography Device

Design of the Diffuse Optical Tomography Device Design of the Diffuse Optical Tomography Device A thesis submitted in partial fulfillment of the requirements for the degree of Bachelor of Science degree in Physics from the College of William and Mary

More information

Instruction manual and data sheet ipca h

Instruction manual and data sheet ipca h 1/15 instruction manual ipca-21-05-1000-800-h Instruction manual and data sheet ipca-21-05-1000-800-h Broad area interdigital photoconductive THz antenna with microlens array and hyperhemispherical silicon

More information

Photonic Crystal Slot Waveguide Spectrometer for Detection of Methane

Photonic Crystal Slot Waveguide Spectrometer for Detection of Methane Photonic Crystal Slot Waveguide Spectrometer for Detection of Methane Swapnajit Chakravarty 1, Wei-Cheng Lai 2, Xiaolong (Alan) Wang 1, Che-Yun Lin 2, Ray T. Chen 1,2 1 Omega Optics, 10306 Sausalito Drive,

More information

150 kj Compact Capacitive Pulsed Power System for an Electrothermal Chemical Gun

150 kj Compact Capacitive Pulsed Power System for an Electrothermal Chemical Gun J Electr Eng Technol Vol. 7, No. 6: 971-976, 2012 http://dx.doi.org/10.5370/jeet.2012.7.6.971 ISSN(Print) 1975-0102 ISSN(Online) 2093-7423 150 kj Compact Capacitive Pulsed Power System for an Electrothermal

More information

A STUDY ON THE VIBRATION CHARACTERISTICS OF CFRP COMPOSITE MATERIALS USING TIME- AVERAGE ESPI

A STUDY ON THE VIBRATION CHARACTERISTICS OF CFRP COMPOSITE MATERIALS USING TIME- AVERAGE ESPI A STUDY ON THE VIBRATION CHARACTERISTICS OF CFRP COMPOSITE MATERIALS USING TIME- AVERAGE ESPI Authors: K.-M. Hong, Y.-J. Kang, S.-J. Kim, A. Kim, I.-Y. Choi, J.-H. Park, C.-W. Cho DOI: 10.12684/alt.1.66

More information

A Real-time Photoacoustic Imaging System with High Density Integrated Circuit

A Real-time Photoacoustic Imaging System with High Density Integrated Circuit 2011 3 rd International Conference on Signal Processing Systems (ICSPS 2011) IPCSIT vol. 48 (2012) (2012) IACSIT Press, Singapore DOI: 10.7763/IPCSIT.2012.V48.12 A Real-time Photoacoustic Imaging System

More information

Sensors. CSE 666 Lecture Slides SUNY at Buffalo

Sensors. CSE 666 Lecture Slides SUNY at Buffalo Sensors CSE 666 Lecture Slides SUNY at Buffalo Overview Optical Fingerprint Imaging Ultrasound Fingerprint Imaging Multispectral Fingerprint Imaging Palm Vein Sensors References Fingerprint Sensors Various

More information

Pinhole collimator design for nuclear survey system

Pinhole collimator design for nuclear survey system Annals of Nuclear Energy 29 (2002) 2029 2040 www.elsevier.com/locate/anucene Pinhole collimator design for nuclear survey system Wanno Lee*, Gyuseong Cho Department of Nuclear Engineering, Korea Advanced

More information

THz Filter Using the Transverse-electric (TE 1 ) Mode of the Parallel-plate Waveguide

THz Filter Using the Transverse-electric (TE 1 ) Mode of the Parallel-plate Waveguide Journal of the Optical Society of Korea ol. 13 No. December 9 pp. 3-7 DOI: 1.387/JOSK.9.13..3 THz Filter Using the Transverse-electric (TE 1 ) Mode of the Parallel-plate Waveguide Eui Su Lee and Tae-In

More information

Development of Personal Dosimeter Using Electronic Dose Conversion Method

Development of Personal Dosimeter Using Electronic Dose Conversion Method Proceedings of the Korean Nuclear Spring Meeting Gyeong ju, Korea, May 2003 Development of Personal Dosimeter Using Electronic Dose Conversion Method Wanno Lee, Bong Jae Lee, and Chang Woo Lee Korea Atomic

More information

Dual-frequency Characterization of Bending Loss in Hollow Flexible Terahertz Waveguides

Dual-frequency Characterization of Bending Loss in Hollow Flexible Terahertz Waveguides Dual-frequency Characterization of Bending Loss in Hollow Flexible Terahertz Waveguides Pallavi Doradla a,b, and Robert H. Giles a,b a Submillimeter Wave Technology Laboratory, University of Massachusetts

More information

Microwave-induced acoustic imaging of biological tissues

Microwave-induced acoustic imaging of biological tissues REVIEW OF SCIENTIFIC INSTRUMENTS VOLUME 70, NUMBER 9 SEPTEMBER 1999 Microwave-induced acoustic imaging of biological tissues Lihong V. Wang, Xuemei Zhao, Haitao Sun, and Geng Ku Optical Imaging Laboratory,

More information

Microstrip even-mode half-wavelength SIR based I-band interdigital bandpass filter

Microstrip even-mode half-wavelength SIR based I-band interdigital bandpass filter Indian Journal of Engineering & Materials Sciences Vol. 9, October 0, pp. 99-303 Microstrip even-mode half-wavelength SIR based I-band interdigital bandpass filter Ram Krishna Maharjan* & Nam-Young Kim

More information

Interaction of Sound and. logarithms. Logarithms continued. Decibels (db) Decibels (db) continued. Interaction of Sound and Media continued

Interaction of Sound and. logarithms. Logarithms continued. Decibels (db) Decibels (db) continued. Interaction of Sound and Media continued Interaction of Sound and Media continued Interaction of Sound and Media Chapter 6 As sound travels through a media and interacts with normal anatomical structures its intensity weakens through what is

More information

Propagation of Single-Mode and Multi-Mode Terahertz Radiation Through a Parallel-Plate Waveguide

Propagation of Single-Mode and Multi-Mode Terahertz Radiation Through a Parallel-Plate Waveguide Journal of the Korean Physical Society, Vol. 53, No. 4, October 2008, pp. 18911896 Propagation of Single-Mode and Multi-Mode Terahertz Radiation Through a Parallel-Plate Waveguide Eui Su Lee, Jin Seok

More information

Comparative Study of Bio-implantable Acoustic Generator Architectures

Comparative Study of Bio-implantable Acoustic Generator Architectures Comparative Study of Bio-implantable Acoustic Generator Architectures D Christensen, S Roundy University of Utah, Mechanical Engineering, S. Central Campus Drive, Salt Lake City, UT, USA E-mail: dave.christensen@utah.edu

More information

Simultaneous Second Harmonic Generation of Multiple Wavelength Laser Outputs for Medical Sensing

Simultaneous Second Harmonic Generation of Multiple Wavelength Laser Outputs for Medical Sensing Sensors 2011, 11, 6125-6130; doi:10.3390/s110606125 OPEN ACCESS sensors ISSN 1424-8220 www.mdpi.com/journal/sensors Article Simultaneous Second Harmonic Generation of Multiple Wavelength Laser Outputs

More information

Jung Sik Kim, Seongchul Hong, Jae Uk Lee, Seung Min Lee, and Jinho Ahn*

Jung Sik Kim, Seongchul Hong, Jae Uk Lee, Seung Min Lee, and Jinho Ahn* Jung Sik Kim, Seongchul Hong, Jae Uk Lee, Seung Min Lee, and Jinho Ahn* Photon shot noise effect in EUVL Degrades stochastic imaging performance Suggestion of a thin attenuated PSM Comparing PSM with conventional

More information

Virtual monochromatic imaging in dual-source and dual-energy CT for visualization of acute ischemic stroke

Virtual monochromatic imaging in dual-source and dual-energy CT for visualization of acute ischemic stroke Virtual monochromatic imaging in dual-source and dual-energy CT for visualization of acute ischemic stroke Hidetake Hara and Hiroshi Muraishi School of Allied Health Sciences, Kitasato University, Japan

More information

Dual Core Differential Pulsed Eddy Current Probe to Detect the Wall Thickness Variation in an Insulated Stainless Steel Pipe

Dual Core Differential Pulsed Eddy Current Probe to Detect the Wall Thickness Variation in an Insulated Stainless Steel Pipe Journal of Magnetics 15(4), 204-208 (2010) DOI: 10.4283/JMAG.2010.15.4.204 Dual Core Differential Pulsed Eddy Current Probe to Detect the Wall Thickness Variation in an Insulated Stainless Steel Pipe C.

More information

Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based On CMOS

Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based On CMOS IOS Journal of Electrical and Electronics Engineering (IOS-JEEE) e-issn: 2278-1676,p-ISSN: 232-3331, Volume 12, Issue 1 Ver. IV (Jan. Feb. 217), PP 64-7 www.iosrjournals.org Design of the Pulse Oximetry

More information

NSERC Summer Project 1 Helping Improve Digital Camera Sensors With Prof. Glenn Chapman (ENSC)

NSERC Summer Project 1 Helping Improve Digital Camera Sensors With Prof. Glenn Chapman (ENSC) NSERC Summer 2016 Digital Camera Sensors & Micro-optic Fabrication ASB 8831, phone 778-782-319 or 778-782-3814, Fax 778-782-4951, email glennc@cs.sfu.ca http://www.ensc.sfu.ca/people/faculty/chapman/ Interested

More information

Capacitive Micromachined Ultrasonic Transducers (CMUTs) for Photoacoustic Imaging

Capacitive Micromachined Ultrasonic Transducers (CMUTs) for Photoacoustic Imaging Invited Paper Capacitive Micromachined Ultrasonic Transducers (CMUTs) for Photoacoustic Imaging Srikant Vaithilingam a,*, Ira O. Wygant a,paulinas.kuo a, Xuefeng Zhuang a, Ömer Oralkana, Peter D. Olcott

More information

Explain what is meant by a photon and state one of its main properties [2]

Explain what is meant by a photon and state one of its main properties [2] 1 (a) A patient has an X-ray scan taken in hospital. The high-energy X-ray photons interact with the atoms inside the body of the patient. Explain what is meant by a photon and state one of its main properties....

More information

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

Radiographic sensitivity improved by optimized high resolution X -ray detector design. DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Radiographic sensitivity improved by optimized high resolution X -ray detector

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

A Novel High Sensitive Optical Fiber Microphone Based on a Singlemode-Multimode-Singlemode Structure

A Novel High Sensitive Optical Fiber Microphone Based on a Singlemode-Multimode-Singlemode Structure Dublin Institute of Technology ARROW@DIT Articles School of Electrical and Electronic Engineering 2011-09-01 A Novel High Sensitive Optical Fiber Microphone Based on a Singlemode-Multimode-Singlemode Structure

More information

A Measurement of the Photon Detection Efficiency of Silicon Photomultipliers

A Measurement of the Photon Detection Efficiency of Silicon Photomultipliers A Measurement of the Photon Detection Efficiency of Silicon Photomultipliers A. N. Otte a,, J. Hose a,r.mirzoyan a, A. Romaszkiewicz a, M. Teshima a, A. Thea a,b a Max Planck Institute for Physics, Föhringer

More information

2. Pulsed Acoustic Microscopy and Picosecond Ultrasonics

2. Pulsed Acoustic Microscopy and Picosecond Ultrasonics 1st International Symposium on Laser Ultrasonics: Science, Technology and Applications July 16-18 2008, Montreal, Canada Picosecond Ultrasonic Microscopy of Semiconductor Nanostructures Thomas J GRIMSLEY

More information

Implementation of a Current-to-voltage Converter with a Wide Dynamic Range

Implementation of a Current-to-voltage Converter with a Wide Dynamic Range Journal of the Korean Physical Society, Vol. 56, No. 3, March 2010, pp. 863 867 Implementation of a Current-to-voltage Converter with a Wide Dynamic Range Jae-Hyoun Park and Hyung-Do Yoon Korea Electronics

More information

Characterization of Surface Structures using THz Radar Techniques with Spatial Beam Filtering and Out-of-Focus Detection

Characterization of Surface Structures using THz Radar Techniques with Spatial Beam Filtering and Out-of-Focus Detection ECNDT 2006 - Tu.2.8.3 Characterization of Surface Structures using THz Radar Techniques with Spatial Beam Filtering and Out-of-Focus Detection Torsten LÖFFLER, Bernd HILS, Hartmut G. ROSKOS, Phys. Inst.

More information

Simulation of All-Optical XOR, AND, OR gate in Single Format by Using Semiconductor Optical Amplifiers

Simulation of All-Optical XOR, AND, OR gate in Single Format by Using Semiconductor Optical Amplifiers Simulation of All-Optical XOR, AND, OR gate in Single Format by Using Semiconductor Optical Amplifiers Chang Wan Son* a,b, Sang Hun Kim a, Young Min Jhon a, Young Tae Byun a, Seok Lee a, Deok Ha Woo a,

More information

SMALL PROXIMITY COUPLED CERAMIC PATCH ANTENNA FOR UHF RFID TAG MOUNTABLE ON METALLIC OBJECTS

SMALL PROXIMITY COUPLED CERAMIC PATCH ANTENNA FOR UHF RFID TAG MOUNTABLE ON METALLIC OBJECTS Progress In Electromagnetics Research C, Vol. 4, 129 138, 2008 SMALL PROXIMITY COUPLED CERAMIC PATCH ANTENNA FOR UHF RFID TAG MOUNTABLE ON METALLIC OBJECTS J.-S. Kim, W.-K. Choi, and G.-Y. Choi RFID/USN

More information

Photoacoustic imaging using an 8-beam Fabry-Perot scanner

Photoacoustic imaging using an 8-beam Fabry-Perot scanner Photoacoustic imaging using an 8-beam Fabry-Perot scanner Nam Huynh, Olumide Ogunlade, Edward Zhang, Ben Cox, Paul Beard Department of Medical Physics and Biomedical Engineering, University College London,

More information

Abnormalities Detection in Superficial Layers of Skin using Optics

Abnormalities Detection in Superficial Layers of Skin using Optics Abnormalities Detection in Superficial Layers of Skin using Optics Saravana kumar T 1, Vignesh S M 2, Jeeva J B 3 1, 2, 3 Department of Biomedical Engineering, VIT University, India. 1 saravanakumar.t2013@vit.ac.in,

More information

Development of PET using 4 4 Array of Large Size Geiger-mode Avalanche Photodiode

Development of PET using 4 4 Array of Large Size Geiger-mode Avalanche Photodiode 2009 IEEE Nuclear Science Symposium Conference Record M09-8 Development of PET using 4 4 Array of Large Size Geiger-mode Avalanche Photodiode K. J. Hong, Y. Choi, J. H. Kang, W. Hu, J. H. Jung, B. J. Min,

More information

Fig. 1 Feeder pipes in the pressurized heavy water reactor.

Fig. 1 Feeder pipes in the pressurized heavy water reactor. DETECTION OF AXIAL CRACKS IN A BENT PIPE USING EMAT TORSIONAL GUIDED WAVES Yong-Moo Cheong 1, Sang-Soo Kim 1, Dong-Hoon Lee 1, Hyun-Kyu Jung 1, and Young H. Kim 2 1 Korea Atomic Energy Research Institute,

More information

Detection of a Surface-Breaking Crack Depth by Using the Surface Waves of Multiple Laser Beams

Detection of a Surface-Breaking Crack Depth by Using the Surface Waves of Multiple Laser Beams 17th World Conference on Nondestructive Testing, 25-28 Oct 2008, Shanghai, China Detection of a Surface-Breaking Crack Depth by Using the Surface Waves of Multiple Laser Beams Seung-Kyu PARK 1, Yong-Moo

More information

Appropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis

Appropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis Indian Journal of Science and Technology Vol 8(S8), 380-386, April 2015 ISSN (Print) : 0974-6846 ISSN (Online) : 0974-5645 DOI: 10.17485/ijst/2015/v8iS8/70528 Appropriate Inspection Distance of Digital

More information

Gallium Nitride PIN Avalanche Photodiode with Double-step Mesa Structure

Gallium Nitride PIN Avalanche Photodiode with Double-step Mesa Structure JOURNAL OF SEMICONDUCTOR TECHNOLOGY AND SCIENCE, VOL.18, NO.5, OCTOBER, 2018 ISSN(Print) 1598-1657 https://doi.org/10.5573/jsts.2018.18.5.645 ISSN(Online) 2233-4866 Gallium Nitride PIN Avalanche Photodiode

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

A Miniaturized Ultrasonic Power Delivery System Tzu-Chieh Chou, Ramkumar Subramanian, Jiwoong Park, and Patrick P. Mercier

A Miniaturized Ultrasonic Power Delivery System Tzu-Chieh Chou, Ramkumar Subramanian, Jiwoong Park, and Patrick P. Mercier A Miniaturized Ultrasonic Power Delivery System Tzu-Chieh Chou, Ramkumar Subramanian, Jiwoong Park, and Patrick P. Mercier 10/23/2014 Motivation: Powering Medical Implants Medical implants are fundamentally

More information

Real -time multi-spectral image processing for mapping pigmentation in human skin

Real -time multi-spectral image processing for mapping pigmentation in human skin Real -time multi-spectral image processing for mapping pigmentation in human skin Daisuke Nakao, Norimichi Tsumura, Yoichi Miyake Department of Information and Image Sciences, Chiba University, Japan Abstract

More information

Non-contact Photoacoustic Tomography using holographic full field detection

Non-contact Photoacoustic Tomography using holographic full field detection Non-contact Photoacoustic Tomography using holographic full field detection Jens Horstmann* a, Ralf Brinkmann a,b a Medical Laser Center Lübeck, Peter-Monnik-Weg 4, 23562 Lübeck, Germany; b Institute of

More information

Title detector with operating temperature.

Title detector with operating temperature. Title Radiation measurements by a detector with operating temperature cryogen Kanno, Ikuo; Yoshihara, Fumiki; Nou Author(s) Osamu; Murase, Yasuhiro; Nakamura, Masaki Citation REVIEW OF SCIENTIFIC INSTRUMENTS

More information

Integrated Blood Flowmeter Using Micromachining Technology

Integrated Blood Flowmeter Using Micromachining Technology Integrated Blood Flowmeter Using Micromachining Technology Yukie Omori 1, Renshi Sawada 2, Susumu Sakano 1 1 College of Engineering, Nihon University 1 Nakagawara-Tokusada, Tamura-cho, Koriyama, 963-8642,

More information

A STUDY ON NON-CONTACT ULTRASONIC TECHNIQUE FOR ON-LINE INSPECTION OF CFRP

A STUDY ON NON-CONTACT ULTRASONIC TECHNIQUE FOR ON-LINE INSPECTION OF CFRP 12 th A-PCNDT 6 Asia-Pacific Conference on NDT, 5 th 1 th Nov 6, Auckland, New Zealand A STUDY ON NON-CONTACT ULTRASONIC TECHNIQUE FOR ON-LINE INSPECTION OF CFRP Seung-Joon Lee 1, Won-Su Park 1, Joon-Hyun

More information

DEVELOPMENT OF CW AND Q-SWITCHED DIODE PUMPED ND: YVO 4 LASER

DEVELOPMENT OF CW AND Q-SWITCHED DIODE PUMPED ND: YVO 4 LASER DEVELOPMENT OF CW AND Q-SWITCHED DIODE PUMPED ND: YVO 4 LASER Gagan Thakkar 1, Vatsal Rustagi 2 1 Applied Physics, 2 Production and Industrial Engineering, Delhi Technological University, New Delhi (India)

More information

transducer. The result indicates that the system sensitivity limit is better than 10 nε dynamic range is around 80dB.

transducer. The result indicates that the system sensitivity limit is better than 10 nε dynamic range is around 80dB. International Conference on Information Science and Computer Applications (ISCA 2013 High-sensitivity ultrasound detection based on phase-shifted fiber Bragg grating Mingrui Xu1,a, Jingjing Guo1,b and

More information

FABRICATION AND SENSING CHARACTERISTICS OF THE CHEMICAL COMPOSITION GRATING SENSOR AT HIGH TEMPERATURES

FABRICATION AND SENSING CHARACTERISTICS OF THE CHEMICAL COMPOSITION GRATING SENSOR AT HIGH TEMPERATURES Figure 10 Measured peak gain of the proposed antenna REFERENCES 1. R.K. Mongia and P. Bhartia, Dielectric resonator antennas A review and general design relations for resonant frequency and bandwidth,

More information

Optical coherence tomography

Optical coherence tomography Optical coherence tomography Peter E. Andersen Optics and Plasma Research Department Risø National Laboratory E-mail peter.andersen@risoe.dk Outline Part I: Introduction to optical coherence tomography

More information

Quantitative diffuse optical tomography for small animals using an ultrafast gated image intensifier

Quantitative diffuse optical tomography for small animals using an ultrafast gated image intensifier Journal of Biomedical Optics 131, 011009 January/February 2008 Quantitative diffuse optical tomography for small animals using an ultrafast gated image intensifier Sachin V. Patwardhan Joseph P. Culver

More information

Tera-Hz Radiation Source by Deference Frequency Generation (DFG) and TPO with All Solid State Lasers

Tera-Hz Radiation Source by Deference Frequency Generation (DFG) and TPO with All Solid State Lasers Tera-Hz Radiation Source by Deference Frequency Generation (DFG) and TPO with All Solid State Lasers Jianquan Yao 1, Xu Degang 2, Sun Bo 3 and Liu Huan 4 1 Institute of Laser & Opto-electronics, 2 College

More information

Spatially Resolved Backscatter Ceilometer

Spatially Resolved Backscatter Ceilometer Spatially Resolved Backscatter Ceilometer Design Team Hiba Fareed, Nicholas Paradiso, Evan Perillo, Michael Tahan Design Advisor Prof. Gregory Kowalski Sponsor, Spectral Sciences Inc. Steve Richstmeier,

More information

X-rays in medical diagnostics

X-rays in medical diagnostics X-rays in medical diagnostics S.Dolanski Babić 2017/18. History W.C.Röntgen (1845-1923) discovered a new type of radiation Nature, Jan. 23. 1896.; Science, Feb.14. 1896. X- rays: Induced the ionization

More information

Light emitting diodes as an excitation source for biomedical photoacoustics

Light emitting diodes as an excitation source for biomedical photoacoustics Light emitting diodes as an excitation source for biomedical photoacoustics. J. llen and P.C. eard Department of Medical Physics and ioengineering, University College London, Malet Place Engineering uilding,

More information

D5.1 Report on the design of a fibre sensor based on NIRS

D5.1 Report on the design of a fibre sensor based on NIRS Optical Fibre Sensors Embedded into technical Textile for Healthcare Contract no.: FP6-027 869 Quality control Version : 2.0 Security: PU Nature: Prototype + Report (P, R) Workpackage: WP5 Start date of

More information

A New Imaging Technique Combining Diffusive Photon Density Waves

A New Imaging Technique Combining Diffusive Photon Density Waves A New Imaging Technique Combining Diffusive Photon Density Waves and Focussed Ultrasound by Charles A. DiMarzio Richard J. Gaudette Center for Electromagnetics Research Northeastern University Boston,

More information

Photoacoustic imaging with coherent light

Photoacoustic imaging with coherent light Photoacoustic imaging with coherent light Emmanuel Bossy Institut Langevin, ESPCI ParisTech CNRS UMR 7587, INSERM U979 Workshop Inverse Problems and Imaging Institut Henri Poincaré, 12 February 2014 Background:

More information

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE Takeyuki Hashimoto 1), Morio Onoe 2), Hiroshi Nakamura 3), Tamon Inouye 4), Hiromichi Jumonji 5), Iwao Takahashi 6); 1)Yokohama Soei

More information

Area Network Applications] Notice: This document has been prepared to assist the IEEE P It is

Area Network Applications] Notice: This document has been prepared to assist the IEEE P It is Project: IEEE P802.15 Working Group for Wireless Personal Area Networks N (WPANs) Submission Title: [RF Safety Considerations for Body Area Network Applications] Date Submitted: [] Source: [Kamya Yekeh

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

324 IEEE TRANSACTIONS ON PLASMA SCIENCE, VOL. 34, NO. 2, APRIL 2006

324 IEEE TRANSACTIONS ON PLASMA SCIENCE, VOL. 34, NO. 2, APRIL 2006 324 IEEE TRANSACTIONS ON PLASMA SCIENCE, VOL. 34, NO. 2, APRIL 2006 Experimental Observation of Temperature- Dependent Characteristics for Temporal Dark Boundary Image Sticking in 42-in AC-PDP Jin-Won

More information

Basic Guidelines for LED Lamp Package Design

Basic Guidelines for LED Lamp Package Design International Journal of Sustainable and Green Energy 2015; 4(5): 187-194 Published online September 11, 2015 (http://www.sciencepublishinggroup.com/j/ijsge) doi: 10.11648/j.ijrse.20150405.13 Basic Guidelines

More information

Co2Z hexaferrite T-DMB antenna for mobile phone applications. Journal: Transactions on Magnetics - Conferences

Co2Z hexaferrite T-DMB antenna for mobile phone applications. Journal: Transactions on Magnetics - Conferences CoZ hexaferrite T-DMB antenna for mobile phone applications Journal: Transactions on Magnetics - Conferences Manuscript ID: MAGCON--- Manuscript Type: Intermag Conference Date Submitted by the Author:

More information

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET 2005 IEEE Nuclear Science Symposium Conference Record M11-126 Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET Jin Zhang, Member,

More information

AN EXPERIMENT RESEARCH ON EXTEND THE RANGE OF FIBER BRAGG GRATING SENSOR FOR STRAIN MEASUREMENT BASED ON CWDM

AN EXPERIMENT RESEARCH ON EXTEND THE RANGE OF FIBER BRAGG GRATING SENSOR FOR STRAIN MEASUREMENT BASED ON CWDM Progress In Electromagnetics Research Letters, Vol. 6, 115 121, 2009 AN EXPERIMENT RESEARCH ON EXTEND THE RANGE OF FIBER BRAGG GRATING SENSOR FOR STRAIN MEASUREMENT BASED ON CWDM M. He, J. Jiang, J. Han,

More information

Measurement of Temperature, Soot Diameter and Soot Volume Fraction in a Gulder Burner

Measurement of Temperature, Soot Diameter and Soot Volume Fraction in a Gulder Burner Department of Engineering Science University of Oxford Measurement of Temperature, Soot Diameter and Soot Volume Fraction in a Gulder Burner Huayong Zhao, Ben William, Richard Stone Project Meeting in

More information

Gigabit Transmission in 60-GHz-Band Using Optical Frequency Up-Conversion by Semiconductor Optical Amplifier and Photodiode Configuration

Gigabit Transmission in 60-GHz-Band Using Optical Frequency Up-Conversion by Semiconductor Optical Amplifier and Photodiode Configuration 22 Gigabit Transmission in 60-GHz-Band Using Optical Frequency Up-Conversion by Semiconductor Optical Amplifier and Photodiode Configuration Jun-Hyuk Seo, and Woo-Young Choi Department of Electrical and

More information

Supporting Information

Supporting Information Electronic Supplementary Material (ESI) for Energy & Environmental Science. This journal is The Royal Society of Chemistry 2019 Supporting Information High Output Magneto-mechano-triboelectric Generator

More information

Optimized shield design for reduction of EMF from wireless power transfer systems

Optimized shield design for reduction of EMF from wireless power transfer systems This article has been accepted and published on J-STAGE in advance of copyediting. Content is final as presented. IEICE Electronics Express, Vol.*, No.*, 1 9 Optimized shield design for reduction of EMF

More information

Conceptual Design of a Table-top Terahertz Free-electron Laser

Conceptual Design of a Table-top Terahertz Free-electron Laser Journal of the Korean Physical Society, Vol. 59, No. 5, November 2011, pp. 3251 3255 Conceptual Design of a Table-top Terahertz Free-electron Laser Y. U. Jeong, S. H. Park, K. Lee, J. Mun, K. H. Jang,

More information

Lecture 19 Optical Characterization 1

Lecture 19 Optical Characterization 1 Lecture 19 Optical Characterization 1 1/60 Announcements Homework 5/6: Is online now. Due Wednesday May 30th at 10:00am. I will return it the following Wednesday (6 th June). Homework 6/6: Will be online

More information

EMBEDDED FBG SENSORS AND AWG-BASED WAVELENGTH INTERROGATOR FOR HEALTH MONITORING OF COMPOSITE MATERIALS

EMBEDDED FBG SENSORS AND AWG-BASED WAVELENGTH INTERROGATOR FOR HEALTH MONITORING OF COMPOSITE MATERIALS 16 TH INTERNATIONAL CONFERENCE ON COMPOSITE MATERIALS EMBEDDED FBG SENSORS AND AWG-BASED WAVELENGTH INTERROGATOR FOR HEALTH MONITORING OF COMPOSITE MATERIALS Shinji Komatsuzaki*, Seiji Kojima*, Akihito

More information

Femtosecond-stability delivery of synchronized RFsignals to the klystron gallery over 1-km optical fibers

Femtosecond-stability delivery of synchronized RFsignals to the klystron gallery over 1-km optical fibers FEL 2014 August 28, 2014 THB03 Femtosecond-stability delivery of synchronized RFsignals to the klystron gallery over 1-km optical fibers Kwangyun Jung 1, Jiseok Lim 1, Junho Shin 1, Heewon Yang 1, Heung-Sik

More information

Proceedings of Meetings on Acoustics

Proceedings of Meetings on Acoustics Proceedings of Meetings on Acoustics Volume 19, 2013 http://acousticalsociety.org/ ICA 2013 Montreal Montreal, Canada 2-7 June 2013 Signal Processing in Acoustics Session 1pSPc: Miscellaneous Topics in

More information

Imaging obscured subsurface inhomogeneity using laser speckle

Imaging obscured subsurface inhomogeneity using laser speckle Imaging obscured subsurface inhomogeneity using laser speckle Ralph Nothdurft, Gang Yao Department of Biological Engineering, University of Missouri-Columbia, Columbia, MO 65211 renothdurft@mizzou.edu,

More information

Monitoring the plant water status with terahertz waves

Monitoring the plant water status with terahertz waves Monitoring the plant water status with terahertz waves Dr. Gunter Urbasch Experimental Semiconductor Physics AG Martin Koch Fachbereich Physik Experimentelle Halbleiterphysik Arbeitsgruppe M. Koch Gunter

More information

INTERNATIONAL JOURNAL OF APPLIED ENGINEERING RESEARCH, DINDIGUL Volume 1, No 3, 2010

INTERNATIONAL JOURNAL OF APPLIED ENGINEERING RESEARCH, DINDIGUL Volume 1, No 3, 2010 All Optical Half Adder Design Using Equations Governing XGM and FWM Effect in Semiconductor Optical Amplifier V. K. Srivastava, V. Priye Indian School of Mines University, Dhanbad srivastavavikrant@hotmail.com

More information

Implementation of wireless ECG measurement system in ubiquitous health-care environment

Implementation of wireless ECG measurement system in ubiquitous health-care environment Implementation of wireless ECG measurement system in ubiquitous health-care environment M. C. KIM 1, J. Y. YOO 1, S. Y. YE 2, D. K. JUNG 3, J. H. RO 4, G. R. JEON 4 1 Department of Interdisciplinary Program

More information

THERMAL NOISE ANALYSIS OF THE RESISTIVE VEE DIPOLE

THERMAL NOISE ANALYSIS OF THE RESISTIVE VEE DIPOLE Progress In Electromagnetics Research Letters, Vol. 13, 21 28, 2010 THERMAL NOISE ANALYSIS OF THE RESISTIVE VEE DIPOLE S. Park DMC R&D Center Samsung Electronics Corporation Suwon, Republic of Korea K.

More information

Nanoimprint lithography with a focused laser beam for the fabrication of micro-/nano-hybrid patterns

Nanoimprint lithography with a focused laser beam for the fabrication of micro-/nano-hybrid patterns Supplementary Material (ESI) for Lab on a Chip This journal is The Royal Society of Chemistry 20XX Nanoimprint lithography with a focused laser beam for the fabrication of micro-/nano-hybrid patterns Hyungjun

More information

Overview of TEPC for high LET Radiation Monitoring in Complex Radiation Field

Overview of TEPC for high LET Radiation Monitoring in Complex Radiation Field Overview of TEPC for high LET Radiation Monitoring in Complex Radiation Field Uk-Won Nam, Jae Jin Lee, Jeonghyun Pyo, Bong-Kon Moon, Dae-Hee Lee, Youngsik Park Korea Astronomy and Space Science Institute

More information

THE THREE electrodes in an alternating current (ac) microdischarge

THE THREE electrodes in an alternating current (ac) microdischarge 488 IEEE TRANSACTIONS ON PLASMA SCIENCE, VOL. 32, NO. 3, JUNE 2004 Firing and Sustaining Discharge Characteristics in Alternating Current Microdischarge Cell With Three Electrodes Hyun Kim and Heung-Sik

More information