Development of PET using 4 4 Array of Large Size Geiger-mode Avalanche Photodiode

Size: px
Start display at page:

Download "Development of PET using 4 4 Array of Large Size Geiger-mode Avalanche Photodiode"

Transcription

1 2009 IEEE Nuclear Science Symposium Conference Record M09-8 Development of PET using 4 4 Array of Large Size Geiger-mode Avalanche Photodiode K. J. Hong, Y. Choi, J. H. Kang, W. Hu, J. H. Jung, B. J. Min, H. K. Lim, S. H. Shin, Y. S. Huh, Y. H. Chung, P. Hughes and C. Jackson Abstract Geiger-mode avalanche photodiode (GAPD) has been demonstrated to be a high performance PET sensor because of high gain, fast response, low excess noise, low bias voltage operation and magnetic field insensitivity. The purpose of this study is to develop a PET for human brain imaging using 4 4 array of large size GAPD. PET detector modules were designed and built to develop a prototype PET. The PET consisted of 72 detector modules arranged in a ring with an inner diameter of 330 mm. The LYSO arrays consisted of 4 4 array of mm 3 pixels, which were 1-to-1 coupled to 4 4 arrays of 9 mm 2 GAPD pixels (SensL, Ireland). The GAPDs were tiled together using flip chip technology on glass and operated at a bias voltage of 32 V for a gain of The signals of the each module were amplified by a 16 channel preamplifier circuit with differential outputs and then sent to a position decoder circuit (PDC), which readout digital address and analog pulse of the one interacted channel from 64 signals of 4 preamplifier boards. The PDC output signals were fed into FPGA-embedded DAQ boards. The analog signal was sampled with 100 MHz, and arrival time and energy of the digitized signal were calculated and stored. The coincidence data were sorted and reconstructed by standard filtered back projection. The energy and time resolution of LYSO-GAPD block detector for 511-keV was 20.4% and 2.4 ns, respectively. The developed PDC could accurately provide the interacted PET signal and reduce the number of the readout channels of PET detector modules based on array type GAPD. The rods down to a diameter of 3.5 mm were resolved in hot-rod phantom image acquired with the brain PET which is similar to the image obtained by Monte Carlo simulation. Activity distribution pattern between white and gray matter in Hoffman brain phantom was well imaged. These results demonstrate that high performance PET could be developed using the GAPD-based PET detectors, analog and digital signal processing methods designed in this work. The prototype brain PET will be tested in a clinical 3T MRI to construct a combined PET-MRI. Manuscript received November 13, This study was supported by the Ministry of Knowledge Economy, by the Korea Science & Engineering Foundation, the Ministry of Education, Science & Technology through their Radiation Technology Development Programs and by a Grant of the Bio Industry Promotion Technology Development Program, the Ministry of Health, Welfare & Family affairs, Republic of Korea. K. J. Hong, Y. Choi, J. H. Kang, W. Hu, J. H. Jung, B. J. Min, H. K. Lim, S. H. Shin and Y. S. Huh are with the Department of Nuclear Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, , Republic of Korea ( ychoi@skku.edu). Y. H. Chung is with the Department of Radiological Science, Yonsei University, College of Health Science, 234 Meaji, Heungup Wonju, Kangwon-Do, , Republic of Korea. P. Hughes and C. Jackson are with SensL, Blackrock, Cork, Ireland. I. INTRODUCTION OSITRON Emission Tomography (PET) is a powerful P molecular imaging modality that uses positron-emitting radionuclides to provide exceptionally sensitive assays of a wide range of biological processes [1-3]. Several studies have been reported on the fabrication of PET system based on avalanche photodiode (APD) instead of the photomultiplier tube (PMT) for MRI compatible PET [4-6]. However, APD has relatively large excess noise, low gain (in the range of ,000) and poor timing capabilities [7, 8]. In recent years, Geiger-mode avalanche photodiode (GAPD) has attracted interests for its use as scintillator readout in PET applications [9]. GAPD consists of an array of avalanche photodiode microcells operated in Geiger-mode with the output summed from all the micro-cells, each with an integrated quenching resistor. It provides a number of advantages over APD detectors including; high gain (~10 6 ), low bias voltage and fast timing response [10]. II. PURPOSE Develop a PET system consisting of 72 detector modules, 18 position decoder circuits (PDC) and 3 FPGA based DAQ boards for human brain imaging using 4 Ý 4 array of large size GAPD Assess 80 LYSO crystal arrays to ensure the uniform performance of each crystal pixel Evaluate the performance of GAPD arrays for application as PET photosensor Evaluate the effect of gain correction by gain adjust circuit implemented in the PDC Estimate the performance of the designed PET using Monte Carlo simulation method Evaluate the performance of the prototype brain PET including imaging capability of hot-rod and Hoffman brain phantoms A. PET detector module III. MATERIALS AND METHODS Detector module components (Fig. 1) - Scintillator: Cerium-doped Lutetium Yttrium Orthosilicate (LYSO) /09/$ IEEE 3032

2 4 Ý 4 matrix of 3 Ý 3 Ý 20 mm3 crystals (Sinoceramics, China) 3.3 mm crystal pitch, polish treatment on all sides of crystal Optically isolated with a 0.3 mm white epoxy resin - Photosensor: Geiger-mode avalanche photodiode (GAPD) (SensL, Ireland) 3-side scalable structure 3.5 Ý 10 6 gain at 32 V low bias voltage 4 Ý 4 array of 2.85 Ý 2.85 mm 2 pixels, 3.3 mm pixel pitch 3640 microcells per pixel, 35 m microcell Preamplifier (SensL, Ireland) - Amplify 1,000 times the detector module signals - Supply voltage: 5.2 V - Output impedance: 47 Ω - Output signal characteristics (at 511 kev γ-ray) Rise/Fall time: 30/170 ns Amplitude of differential output signals: 250 mv Position decoder circuit (PDC) (Custom made) - Readout channel address and analog pulse of the channel interacted with coincidence event among 64 preamplified signals transmitted from 4 detector modules (Fig. 4) Fig matrix LYSO crystals and a 4 4 array GAPD used to construct PET detector module Components coupling and packing (Fig. 2) - Coupling: direct coupling LYSO to GAPD without optical-coupling material - Packing: each detector module independently encapsulated from light Fig. 4. Configuration of position decoder circuit Data acquisition module (DAQ) [13] - Hardware: VHS-ADC Virtex-4 boards (Lyrtech, Canada) - Sampling by 105 MHz ADC - Calculate energy and arrival time of the digitized signal by initial rise interpolation method implemented using FPGA [14] - Store the data in list mode format C. Performance Evaluation of PET Components 1) Assess LYSO array performance Fig. 2. LYSO-GAPD detector assembly used in this study B. Analog and digital signal processing Connection between PET detector module and preamplifier - Use flexible flat cable (FFC) of 300 cm long (Fig. 3) - Purpose: to separately locate PET detector module (inside MR bore) and preamplifier (outside MR bore) Experimental configuration - Purpose: to ensure the uniform performance of each crystal pixel of 80 4 Ý 4 LYSO arrays - 1 PET detector module + 1 preamplifier + 1 DAQ module (Fig.5) - Radiation source: point source of 220 kbq Na-22 - Duration of data acquisition: 5 min (per crystal block) Fig. 3. PET detector module and preamplifier connected using flexible flat cable of 300 cm long Fig. 5. Configuration of test set-up for LYSO array selection 3033

3 Measurement - Acquire energy spectra of 80 4 Ý 4 LYSO arrays to compare photopeak position and energy resolution - Compare crystals located at a same sensor pixel to rule out variability originated from sensor and to examine the variability only from LYSO crystals - Calculate means of energy resolution and photo peak position using the energy spectrum of 80 crystal pixels of the same position among 16 pixels of an array - 2 PET detector modules + 2 preamplifiers + 1 DAQ module - Radiation source: point source of 220 kbq Na-22 - Duration of data acquisition: 5 min Measurement - Acquire energy spectra of thirty-two outputs from the two detector modules Selection range [11] - The specification is as below list. If any crystal is out of the range, the crystal block was exchanged. TABLE I. SCINTILLATOR SELECTION RANGE 511 kev Photo peak position Energy resolution Specification Mean 23% of mean value Mean 31% of mean value Fig. 6. Schematic diagram of analog circuit capable of adjusting DC offset level and gain of the PET detector output signals D. Brain PET System 2) Evaluate the basic performance of GAPD array Experimental configuration - A pair of PET detector modules + 2 preamplifiers + 2 PDCs + 1 DAQ module - Radiation source: point source of 220 kbq Na-22 - Duration of data acquisition: 10 min Measurement [12] - Acquire energy spectrum, count rate, flood histogram and variation of 511-keV photopeak channel using one PET detector module - The resulting flood histogram was analyzed in terms of uniformity given by the ratio between minimum and maximum counts and variation of the 511 kev photopeak position for all 16 channels. PET scanner configuration (Fig. 7) - Detector module number: 72 unit - Ring diameter: 330 mm - Axial field-of-view: 12.9 mm PET electronics configuration - Preamplifier number: 72 unit - PDC: 18 unit - DAQ (VHS-ADC Virtex-4 board): 3 unit = = - Acquire coincidence timing spectrum using a pair of PET detector modules 3) Evaluate the effect of gain correction in the PDC Experimental configuration - Gain adjustment in PDC: analog circuit capable of adjusting DC offset level and gain of the PET detector output signals (Fig. 6) - Adjusting the value of a variable resistor connected in series to the output of offset level translation circuit. - Gain range: 1 to 2 Fig. 7. Brain PET system consisting of PET and signal processing parts E. Performance Estimation of the Brain PET Using Monte Carlo Simulation Simulation parameters (Fig. 8) - Simulation tool: GATE (Geant4 Application for Tomographic Emission) 3034

4 - Point source: Size: 1 mm diameter Activity: 3.7 MBq Source location: move from center to 100 mm offcenter, in 20 mm steps - Hot-rod phantom Rod diameter: range from 2.5 to 6.5 mm Activity: 13 MBq - 3D digital Hoffman brain phantom Activity: 96 MBq - Energy window: 250 to 650 kev - Image reconstruction: 2D FBP Measurement - Acquire Hot-rod and 3D Hoffman brain phantom images Fig. 9. Prototype brain PET consisting of 72 detector modules and connected to preamplifier using 300 cm FFC IV. RESULTS A. Performance Measurement of LYSO Crystal Fig. 8. Brain PET detector and 3D digital Hoffman brain phantom simulated to estimate the PET imaging performance Measurement - Draw radial profiles of reconstructed images of point source as a function of source location to evaluate the spatial resolution over the FOV - Simulate hot-rod and 3D Hoffman brain phantom images to evaluate the imaging performance of the designed PET Average photopeak position and energy resolution of one pixel at the same position in the 80 crystal blocks were and 21.7%, respectively. The values of the most crystals were within the selection range, however, some of pixels were out of selection range (Fig.10). These crystal blocks having abnormal pixel were exchanged to normal crystal. F. Performance Measurement of the Prototype PET Experimental configuration (Fig. 9) - Detector: 72 detector modules consisted of 4x4 array of LYSO-GAPD - Cable length between detector module and preamplifier: 300 cm - Hot-rod phantom Rod diameter: range between 2.5 and 6.5 mm Radiation source: 74 MBq F-18 Total acquisition counts: 1 million - Hoffman brain phantom Radiation source: 55 MBq F-18 Total acquisition counts: 6.8 million - Energy and time windows: 250 to 650 kev, 60 ns - Image reconstruction: 2D FBP - Image corrections: normalization, random (a) (b) Fig. 10. Photopeak position (a) and energy resolution (b) of one pixel at the same position in the 80 crystal blocks 3035

5 B. Performance Measurement of the GAPD Array TABLE II. PERFORMANCE MEASUREMENTS OF THE GAPD ARRAY Average energy resolution (n=16) Average count rate Flood histogram uniformity Variation of 511 kev photopeak channel Timing resolution 20.4±3% 530 cps/ Ci 1: % 2.4 ns D. Performance Estimation of the Brain PET Using Monte Carlo Simulation Radial resolution of reconstructed point source image was degraded from 3.3 mm to 6.1 mm, at a 100 mm off-center (Fig. 14). Fig. 14. Reconstructed images of the point source obtained by moving a source from the center to the radial offset of the scanner s FOV, at 20 mm intervals (left), and their radial resolutions (right) Fig. 11. Flood histogram of data acquired with 4 x 4 array of LYSO- GAPD C. Gain Correction The ratio of the highest and lowest photopeak positions was improved from 1.8 to 1.0 after gain correction by using the gain adjust circuit implemented in PDC (Fig 12,13). E. Performance Measurement of the Prototype PET PET images representing activity distribution patterns of hot-rod and Hoffman brain phantoms were successfully acquired using the brain PET developed in this study (Fig. 15). The rods down to a diameter of 3.5 mm were resolved in hot-rod phantom image acquired with the brain PET which is similar to the image obtained by Monte Carlo simulation. Activity distribution pattern between white and gray matter in Hoffman brain phantom was well imaged. Fig. 12. Relative photopeak position before and after the gain correction Fig. 15. Hot-rod (top row) and Hoffman brain phantom (bottom row) images acquired with simulation and with the brain PET developed in this study Fig channel energy spectra (Na-22) obtained after gain correction by using the gain adjust circuit V. SUMMARY AND CONCLUSION The results of this study demonstrate that the GAPD array can provide good energy resolution, count rate, timing resolution and flood histogram uniformity as a PET detector. Gain uniformity for all channels of the PET detector modules was improved 80% by using the gain correction circuit 3036

6 implemented in PDC. The developed PDC could accurately provide the interacted PET signal and reduce the number of the readout channels of PET detector modules based on array type GAPD. PET images were successfully acquired using the prototype brain PET consisting of 72 detector modules and these results demonstrate that high performance PET could be developed using the GAPD-based PET detectors, analog and digital signal processing methods designed in this work. The prototype brain PET will be tested in a clinical 3T MRI to construct a combined PET-MRI [15, 16]. [16] Y. Huh, Y. Choi, W. Hu, J. Kang, J. Jung, K. Hong, et al., A FPGA- Based PET Data Acquisition Method for Simultaneous PET/MRI Imaging, IEEE NSS-MIC, M13-258, REFERENCES [1] D. W. Townsend, J. P. Carney, J. T. Yap et al., PET/CT today and tomorrow, J Nucl Med, vol. 45 Suppl 1, pp. 4S-14S, Jan, [2] G. Antoch, N. Saoudi, H. Kuehl et al., Accuracy of whole-body dualmodality fluorine-18-2-fluoro-2-deoxy-d -glucose positron emission tomography and computed tomography (FDG-PET/CT) for tumor staging in solid tumors: comparison with CT and PET, J Clin Oncol, vol. 22, no. 21, pp , Nov 1, [3] T. Beyer, G. Antoch, S. Muller et al., Acquisition protocol considerations for combined PET/CT imaging, J Nucl Med, vol. 45 Suppl 1, pp. 25S-35S, Jan, [4] C. Catana, Y. Wu, M. S. Judenhofer et al., Simultaneous acquisition of multislice PET and MR images: initia results with a MR-compatible PET scanner, J Nucl Med, vol. 47, no. 12, pp , Dec, [5] B. J. Pichler, M. S. Judenhofer, C. Catana et al., Performance Test of an LSO-APD Detector in a 7-T MRI Scanner for Simultaneous PET/MRI, The Journal of nuclear medicine, vol. 47, no. 4, pp , [6] M. S. Judenhofer, H. F. Wehrl, D. F. Newport et al., Simultaneous PET- MRI: a new approach for functional and morphological imaging, Nat Med, vol. 14, no. 4, pp , Apr, [7] S. I. Ziegler, B. J. Pichler, G. Boening et al., A prototype highresolution animal positron tomograph with avalanche photodiode arrays and LSO crystals, European journal of nuclear medicine, vol. 28, no. 2, pp , [8] R. Grazioso, N. Zhanga, J. Corbeil et al., APD-based PET detector for simultaneous PET/MR imaging, Nucl. Instr. Meth., vol. A569, pp , [9] V. C. Spanoudaki, A. B. Mann, A. N. Otte et al., Use of single photon counting detector arrays in combined PET/MR: Characterization of LYSO-SiPM detector modules and comparison with a LSO-APD detector, Journal of instrumentation an IOP and SISSA journal, vol. 2, no. 12, pp. P12002-P12002, [10] Y. Shao, H. Li, and K. Gao, Initial experimental studies of using solidstate photomultiplier for PET applications, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, vol. 580, no. 2, pp , [11] J. Trummer, D. Aimard, E. Auffray et al., "Scintillation Properties of LuYAP and LYSO Crystals Measured with MiniACCOS, an Automatic Crystal Quality Control System," IEEE NSS-MIC, J03-4, [12] K. J. Hong, Y. Choi, J.H. Kang, W. Hu, J.H. Jung, B.J. Min, et al., "Performance Evaluation of 4 x 4 Array of Large Size Silicon Photomultiplier for MR-Compatible PET Photosensor," IEEE NSS-MIC, M06-61, [13] W. Hu, Y. Choi, J. Jung, K. Hong, J. Kang, B. Min, et al., "A FPGAbased high speed multi-channel simultaneous signal acquisition method for positron emission tomography," IEEE NSS-MIC, M05-334, [14] W. Hu, Y. Choi, J. Jung, K. Hong, J. Kang, B. Min, et al., "An improved simple digital timing method for positron emission tomography," IEEE NSS-MIC, M13-336, [15] J. H. Kang, Y. Choi, K. J. Hong, J. H. Jung, W. Hu, G. H. Im, et al., "Characterization of cross-compatibility of PET components and MRI," IEEE NSS-MIC, M13-261,

Design and development of compact readout electronics with silicon photomultiplier array for a compact imaging detector *

Design and development of compact readout electronics with silicon photomultiplier array for a compact imaging detector * CPC(HEP & NP), 2012, 36(10): 973 978 Chinese Physics C Vol. 36, No. 10, Oct., 2012 Design and development of compact readout electronics with silicon photomultiplier array for a compact imaging detector

More information

Design and development of compact readout electronics with silicon photomultiplier array for a compact imaging detector

Design and development of compact readout electronics with silicon photomultiplier array for a compact imaging detector University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2012 Design and development of compact readout

More information

Performance Evaluation of SiPM Detectors for PET Imaging in the Presence of Magnetic Fields

Performance Evaluation of SiPM Detectors for PET Imaging in the Presence of Magnetic Fields 2008 IEEE Nuclear Science Symposium Conference Record M02-4 Performance Evaluation of SiPM Detectors for PET Imaging in the Presence of Magnetic Fields Samuel España, Student Member, IEEE, Gustavo Tapias,

More information

Conceptual Study of Brain Dedicated PET Improving Sensitivity

Conceptual Study of Brain Dedicated PET Improving Sensitivity Original Article PROGRESS in MEDICAL PHYSICS 27(4), Dec. 2016 https://doi.org/10.14316/pmp.2016.27.4.236 pissn 2508-4445, eissn 2508-4453 Conceptual Study of Brain Dedicated PET Improving Sensitivity Han-Back

More information

PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner

PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner September, 2017 Results submitted to Physics in Medicine & Biology Negar Omidvari 1, Jorge Cabello 1, Geoffrey Topping

More information

Cross-Strip Multiplexed Electro-Optical Coupled Scintillation Detector for Integrated PET/MRI

Cross-Strip Multiplexed Electro-Optical Coupled Scintillation Detector for Integrated PET/MRI IEEE TRANSACTIONS ON NUCLEAR SCIENCE 1 Cross-Strip Multiplexed Electro-Optical Coupled Scintillation Detector for Integrated PET/MRI Peter D. Olcott, Member, IEEE, GaryGlover, Member, IEEE, and CraigS.Levin,

More information

Design Studies of A High-Performance Onboard Positron Emission Tomography For Integrated Small Animal PET/CT/RT Radiation Research Systems

Design Studies of A High-Performance Onboard Positron Emission Tomography For Integrated Small Animal PET/CT/RT Radiation Research Systems Proceedings of the International MultiConference of Engineers and Computer Scientists 2018 Vol II Design Studies of A High-Performance Onboard Positron Emission Tomography For Integrated Small Animal PET/CT/RT

More information

Investigation of Solid-State Photomultipliers for Positron Emission Tomography Scanners

Investigation of Solid-State Photomultipliers for Positron Emission Tomography Scanners Journal of the Korean Physical Society, Vol. 50, No. 5, May 2007, pp. 1332 1339 Investigation of Solid-State Photomultipliers for Positron Emission Tomography Scanners Jae Sung Lee Department of Nuclear

More information

An innovative detector concept for hybrid 4D-PET/MRI Imaging

An innovative detector concept for hybrid 4D-PET/MRI Imaging Piergiorgio Cerello (INFN - Torino) on behalf of the 4D-MPET* project *4 Dimensions Magnetic compatible module for Positron Emission Tomography INFN Perugia, Pisa, Torino; Polytechnic of Bari; University

More information

Journal of Radiology in press. Simultaneous PET/MR Images, acquired with a Compact MRI Compatible PET Detector in a 7 Tesla Magnet

Journal of Radiology in press. Simultaneous PET/MR Images, acquired with a Compact MRI Compatible PET Detector in a 7 Tesla Magnet Journal of Radiology in press Simultaneous PET/MR Images, acquired with a Compact MRI Compatible PET Detector in a 7 Tesla Magnet Martin S. Judenhofer BS 1, Ciprian Catana 2, Brian, K. Swann 3, Stefan

More information

2594 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER /$ IEEE

2594 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER /$ IEEE 2594 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER 2009 Investigation of Depth of Interaction Encoding for a Pixelated LSO Array With a Single Multi-Channel PMT Yongfeng Yang, Member, IEEE,

More information

A PET detector module using FPGA-only MVT digitizers

A PET detector module using FPGA-only MVT digitizers A PET detector module using FPGA-only MVT digitizers Daoming Xi, Student Member, IEEE, Chen Zeng, Wei Liu, Student Member, IEEE, Xiang Liu, Lu Wan, Student Member, IEEE, Heejong Kim, Member, IEEE, Luyao

More information

Performance measurements of a depth-encoding PET detector module based on positionsensitive

Performance measurements of a depth-encoding PET detector module based on positionsensitive Home Search Collections Journals About Contact us My IOPscience Performance measurements of a depth-encoding PET detector module based on positionsensitive avalanche photodiode read-out This article has

More information

PET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002

PET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002 PET Detectors William W. Moses Lawrence Berkeley National Laboratory March 26, 2002 Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes

More information

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET 2005 IEEE Nuclear Science Symposium Conference Record M11-126 Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET Jin Zhang, Member,

More information

APD Quantum Efficiency

APD Quantum Efficiency APD Quantum Efficiency Development of a 64-channel APD Detector Module with Individual Pixel Readout for Submillimeter Spatial Resolution in PET Philippe Bérard a, Mélanie Bergeron a, Catherine M. Pepin

More information

Simulation and evaluation of a cost-effective high-performance brain PET scanner.

Simulation and evaluation of a cost-effective high-performance brain PET scanner. Research Article http://www.alliedacademies.org/biomedical-imaging-and-bioengineering/ Simulation and evaluation of a cost-effective high-performance brain PET scanner. Musa S Musa *, Dilber U Ozsahin,

More information

Design of a Novel Front-End Readout ASIC for PET Imaging System *

Design of a Novel Front-End Readout ASIC for PET Imaging System * Journal of Signal and Information Processing, 2013, 4, 129-133 http://dx.doi.org/10.4236/jsip.2013.42018 Published Online May 2013 (http://www.scirp.org/journal/jsip) 129 Design of a Novel Front-End Readout

More information

Investigation of low noise, low cost readout electronics for high sensitivity PET systems based on Avalanche Photodiode arrays

Investigation of low noise, low cost readout electronics for high sensitivity PET systems based on Avalanche Photodiode arrays Investigation of low noise, low cost readout electronics for high sensitivity PET systems based on Avalanche Photodiode arrays Frezghi Habte, Member, IEEE and Craig S.Levin, Member, IEEE Abstract A compact,

More information

Index terms: PET, Silicon Photo-multiplier, Small Animal PET insert for MRI scanner. Size: 1.2x1.2 mm, Pitch: mm Thickness: 4 mm

Index terms: PET, Silicon Photo-multiplier, Small Animal PET insert for MRI scanner. Size: 1.2x1.2 mm, Pitch: mm Thickness: 4 mm Measurement of Energy and Timing Resolution of Very Highly Pixellated LYSO Crystal Blocks with Multiplexed SiPM Readout for Use in a Small Animal PET/MR Insert Christopher J. Thompson, Senior Member IEEE,

More information

Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET

Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET A. Kuhn, S. Surti, Member, IEEE, J. S. Karp, Senior Member, IEEE, G. Muehllehner, Fellow, IEEE, F.M. Newcomer, R. VanBerg Abstract--

More information

Development of an innovative LSO-SiPM detector module for high-performance Positron Emission Tomography

Development of an innovative LSO-SiPM detector module for high-performance Positron Emission Tomography Development of an innovative LSO-SiPM detector module for high-performance Positron Emission Tomography Maria Leonor Trigo Franco Frazão leonorfrazao@ist.utl.pt Instituto Superior Técnico, Lisboa, Portugal

More information

POSITRON emission tomography (PET) scanners, which

POSITRON emission tomography (PET) scanners, which 882 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 55, NO. 3, JUNE 2008 An Investigation Into the Use of Geiger-Mode Solid-State Photomultipliers for Simultaneous PET and MRI Acquisition Seong Jong Hong, In

More information

Effects of Dark Counts on Digital Silicon Photomultipliers Performance

Effects of Dark Counts on Digital Silicon Photomultipliers Performance Effects of Dark Counts on Digital Silicon Photomultipliers Performance Radosław Marcinkowski, Samuel España, Roel Van Holen, Stefaan Vandenberghe Abstract Digital Silicon Photomultipliers (dsipm) are novel

More information

A feasibility study of photosensor charge signal transmission to preamplifier using long cable for development of hybrid PET-MRI

A feasibility study of photosensor charge signal transmission to preamplifier using long cable for development of hybrid PET-MRI A feasibility study of photosensor charge signal transmission to preamplifier using long cable for development of hybrid PET-MRI Jihoon Kang, Yong Choi, a Key Jo Hong, Jin Ho Jung, Wei Hu, Yoon Suk Huh,

More information

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 1, FEBRUARY

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 1, FEBRUARY IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 1, FEBRUARY 2005 217 Optimization of Dual Layer Phoswich Detector Consisting of LSO and LuYAP for Small Animal PET Yong Hyun Chung, Yong Choi, Member,

More information

764 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 3, JUNE 2004

764 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 3, JUNE 2004 764 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 3, JUNE 2004 Study of Low Noise Multichannel Readout Electronics for High Sensitivity PET Systems Based on Avalanche Photodiode Arrays Frezghi Habte,

More information

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A Nuclear Instruments and Methods in Physics Research A 614 (2010) 308 312 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima

More information

Initial results on Sipm array based on a symmetric resistive voltage division readout

Initial results on Sipm array based on a symmetric resistive voltage division readout Initial results on Sipm array based on a symmetric resistive voltage division readout S. David, M. Georgiou, E. Fysikopoulos, N. Efthimiou, T. Paipais, L. Kefalidis and G. Loudos Abstract The aim of this

More information

Solid-State Photomultiplier in CMOS Technology for Gamma-Ray Detection and Imaging Applications

Solid-State Photomultiplier in CMOS Technology for Gamma-Ray Detection and Imaging Applications Solid-State Photomultiplier in CMOS Technology for Gamma-Ray Detection and Imaging Applications Christopher Stapels, Member, IEEE, William G. Lawrence, James Christian, Member, IEEE, Michael R. Squillante,

More information

Development of the LBNL Positron Emission Mammography Camera

Development of the LBNL Positron Emission Mammography Camera Development of the LBNL Positron Emission Mammography Camera J.S. Huber, Member, IEEE, W.S. Choong, Member, IEEE, J. Wang, Member, IEEE, J.S. Maltz, Member, IEEE, J. Qi, Member, IEEE, E. Mandelli, Member,

More information

Development of a simplified readout for a compact gamma camera based on 2 2 H8500 multi-anode PSPMT array

Development of a simplified readout for a compact gamma camera based on 2 2 H8500 multi-anode PSPMT array University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2010 Development of a simplified readout for a

More information

Study of a Prototype VP-PET Imaging System Based on highly. Pixelated CdZnTe Detectors

Study of a Prototype VP-PET Imaging System Based on highly. Pixelated CdZnTe Detectors Study of a Prototype VP-PET Imaging System Based on highly Pixelated CdZnTe Detectors Zheng-Qian Ye 1, Ying-Guo Li 1, Tian-Quan Wang 1, Ya-Ming Fan 1, Yong-Zhi Yin 1,*, Xi-Meng Chen 1 Affiliations: 1 School

More information

Development of FPGA-based Coincidence Units with Veto Function

Development of FPGA-based Coincidence Units with Veto Function Biomed Eng Lett (2011) 1:27-31 DOI 10.1007/s13534-011-0001-3 ORIGINAL ARTICLE Development of FPGA-based Coincidence Units with Veto Function Guen Bae Ko, Hyun Suk Yoon, Sun Il Kwon, Seong Jong Hong, Dong

More information

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare GE Healthcare Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare There is excitement across the industry regarding the clinical potential of a hybrid

More information

PoS(PhotoDet 2012)022

PoS(PhotoDet 2012)022 SensL New Fast Timing Silicon Photomultiplier Kevin O`Neill 1 SensL Technologies Limited 6800 Airport Business Park, Cork, Ireland E-mail: koneill@sensl.com Nikolai Pavlov SensL Technologies Limited 6800

More information

CHAPTER 8 GENERIC PERFORMANCE MEASURES

CHAPTER 8 GENERIC PERFORMANCE MEASURES GENERIC PERFORMANCE MEASURES M.E. DAUBE-WITHERSPOON Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, United States of America 8.1. INTRINSIC AND EXTRINSIC MEASURES 8.1.1.

More information

Celesteion Time-of-Flight Technology

Celesteion Time-of-Flight Technology Celesteion Time-of-Flight Technology Bing Bai, PhD Clinical Sciences Manager, PET/CT Canon Medical Systems USA Introduction Improving the care for every patient while providing a high standard care to

More information

Timing Resolution Performance Comparison for Fast and Standard Outputs of SensL SiPM

Timing Resolution Performance Comparison for Fast and Standard Outputs of SensL SiPM Timing Resolution Performance Comparison for Fast and Standard Outputs of SensL SiPM Sergei Dolinsky, Geng Fu, and Adrian Ivan Abstract A new silicon photomultiplier (SiPM) with a unique fast output signal

More information

Arrays of digital Silicon Photomultipliers Intrinsic performance and Application to Scintillator Readout

Arrays of digital Silicon Photomultipliers Intrinsic performance and Application to Scintillator Readout Arrays of digital Silicon Photomultipliers Intrinsic performance and Application to Scintillator Readout Carsten Degenhardt, Ben Zwaans, Thomas Frach, Rik de Gruyter Philips Digital Photon Counting NSS-MIC

More information

PUBLISHED BY IOP PUBLISHING FOR SISSA MEDIALAB. Imaging performance of silicon photomultipliers coupled to BGO and CsI:Na arrays

PUBLISHED BY IOP PUBLISHING FOR SISSA MEDIALAB. Imaging performance of silicon photomultipliers coupled to BGO and CsI:Na arrays PUBLISHED BY IOP PUBLISHING FOR SISSA MEDIALAB RECEIVED: August 2, 2013 REVISED: October 16, 2013 ACCEPTED: November 27, 2013 PUBLISHED: December 12, 2013 Imaging performance of silicon photomultipliers

More information

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud University of Groningen Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you

More information

PET: New Technologies & Applications, Including Oncology

PET: New Technologies & Applications, Including Oncology PET: New Technologies & Applications, Including Oncology, PhD, FIEEE Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Disclosures Research Contract, GE Healthcare

More information

First Applications of the YAPPET Small Animal Scanner

First Applications of the YAPPET Small Animal Scanner First Applications of the YAPPET Small Animal Scanner Guido Zavattini Università di Ferrara CALOR2 Congress, Annecy - FRANCE YAP-PET scanner Scintillator: YAP:Ce Size: matrix of 2x2 match like crystals

More information

Study of Silicon Photomultipliers for Positron Emission Tomography (PET) Application

Study of Silicon Photomultipliers for Positron Emission Tomography (PET) Application Study of Silicon Photomultipliers for Positron Emission Tomography (PET) Application Eric Oberla 5 June 29 Abstract A relatively new photodetector, the silicon photomultiplier (SiPM), is well suited for

More information

Future directions in Nuclear Medicine Instrumentation

Future directions in Nuclear Medicine Instrumentation Future directions in Nuclear Medicine Instrumentation Where are we going - and why? First, the disclosure list My group at the University of Washington has research support from: NIH DOE General Electric

More information

Characterization of Silicon Photomultiplier Readout Designs for Use in Positron. Emission Tomography Systems. Chen-Yi Liu. the University of Manitoba

Characterization of Silicon Photomultiplier Readout Designs for Use in Positron. Emission Tomography Systems. Chen-Yi Liu. the University of Manitoba Characterization of Silicon Photomultiplier Readout Designs for Use in Positron Emission Tomography Systems by Chen-Yi Liu A thesis submitted to the Faculty of Graduate Studies of the University of Manitoba

More information

P ILC A. Calcaterra (Resp.), L. Daniello (Tecn.), R. de Sangro, G. Finocchiaro, P. Patteri, M. Piccolo, M. Rama

P ILC A. Calcaterra (Resp.), L. Daniello (Tecn.), R. de Sangro, G. Finocchiaro, P. Patteri, M. Piccolo, M. Rama P ILC A. Calcaterra (Resp.), L. Daniello (Tecn.), R. de Sangro, G. Finocchiaro, P. Patteri, M. Piccolo, M. Rama Introduction and motivation for this study Silicon photomultipliers ), often called SiPM

More information

Currently, the spatial resolution of most dedicated smallanimal

Currently, the spatial resolution of most dedicated smallanimal A Prototype High-Resolution Small-Animal PET Scanner Dedicated to Mouse Brain Imaging Yongfeng Yang 1,2, Julien Bec 1, Jian Zhou 1, Mengxi Zhang 1, Martin S. Judenhofer 1, Xiaowei Bai 1, Kun Di 1, Yibao

More information

Thomas Frach, Member, IEEE, Walter Ruetten, Member, IEEE, Klaus Fiedler, Gunnar Maehlum, Member, IEEE, Torsten Solf, and Andreas Thon

Thomas Frach, Member, IEEE, Walter Ruetten, Member, IEEE, Klaus Fiedler, Gunnar Maehlum, Member, IEEE, Torsten Solf, and Andreas Thon Assessment of Photodiodes as a Light Detector for PET Scanners Thomas Frach, Member, IEEE, Walter Ruetten, Member, IEEE, Klaus Fiedler, Gunnar Maehlum, Member, IEEE, Torsten Solf, and Andreas Thon Abstract

More information

60 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 1, FEBRUARY /$ IEEE

60 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 1, FEBRUARY /$ IEEE 60 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 1, FEBRUARY 2007 Prototype Parallel Readout System for Position Sensitive PMT Based Gamma Ray Imaging Systems Frezghi Habte, Member, IEEE, Peter D.

More information

976 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 57, NO. 3, JUNE /$ IEEE

976 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 57, NO. 3, JUNE /$ IEEE 976 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 57, NO. 3, JUNE 2010 A Four-Layer DOI Detector With a Relative Offset for Use in an Animal PET System Mikiko Ito, Jae Sung Lee, Sun Il Kwon, Geon Song Lee,

More information

SOLID state photodiode and avalanche photodiode scintillation

SOLID state photodiode and avalanche photodiode scintillation 2007 IEEE Nuclear Science Symposium Conference Record M14-1 Data acquisition system design for a 1 mm 3 resolution PSAPD-based PET system Peter D. Olcott,,Student Member, IEEE, Frances W. Y. Lau, Student

More information

MPPC and Liquid Xenon technologies from particle physics to medical imaging

MPPC and Liquid Xenon technologies from particle physics to medical imaging CANADA S NATIONAL LABORATORY FOR PARTICLE AND NUCLEAR PHYSICS Owned and operated as a joint venture by a consortium of Canadian universities via a contribution through the National Research Council Canada

More information

LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option

LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option Y. Bercier, Member, IEEE, M. Casey, Member, IEEE, J. Young, Member, IEEE, T. Wheelock, Member, IEEE, T. Gremillion Abstract-- Factors which

More information

Characterisation of SiPM Index :

Characterisation of SiPM Index : Characterisation of SiPM --------------------------------------------------------------------------------------------Index : 1. Basics of SiPM* 2. SiPM module 3. Working principle 4. Experimental setup

More information

SPMMicro. SPMMicro. Low Cost High Gain APD. Low Cost High Gain APD. Page 1

SPMMicro. SPMMicro. Low Cost High Gain APD. Low Cost High Gain APD. Page 1 SPMMicro Page 1 Overview Silicon Photomultiplier (SPM) Technology SensL s SPMMicro series is a High Gain APD provided in a variety of miniature, easy to use, and low cost packages. The SPMMicro detector

More information

EndoTOFPET-US: an endoscopic Positron Emission Tomography detector for a novel multimodal medical imaging tool

EndoTOFPET-US: an endoscopic Positron Emission Tomography detector for a novel multimodal medical imaging tool : an endoscopic Positron Emission Tomography detector for a novel multimodal medical imaging tool DESY and University of Hamburg E-mail: daniele.cortinovis@desy.de The EndoTOFPET-US collaboration aims

More information

Characterization of a 64 Channel PET Detector Using Photodiodes for Crystal Identification *

Characterization of a 64 Channel PET Detector Using Photodiodes for Crystal Identification * Characterization of a 64 Channel PET Detector Using Photodiodes for Crystal Identification * J. S. Huber, Member, IEEE, W.W. Moses, Senior Member, IEEE, S.E. Derenzo, Senior Member, IEEE, M.H. Ho, M.S.

More information

Tutors Dominik Dannheim, Thibault Frisson (CERN, Geneva, Switzerland)

Tutors Dominik Dannheim, Thibault Frisson (CERN, Geneva, Switzerland) Danube School on Instrumentation in Elementary Particle & Nuclear Physics University of Novi Sad, Serbia, September 8 th 13 th, 2014 Lab Experiment: Characterization of Silicon Photomultipliers Dominik

More information

CAEN. Electronic Instrumentation. CAEN Silicon Photomultiplier Kit

CAEN. Electronic Instrumentation. CAEN Silicon Photomultiplier Kit CAEN Tools for Discovery Electronic Instrumentation CAEN Silicon Photomultiplier Kit CAEN realized a modular development kit dedicated to Silicon Photomultipliers, representing the state-of-the art in

More information

Simulation of Algorithms for Pulse Timing in FPGAs

Simulation of Algorithms for Pulse Timing in FPGAs 2007 IEEE Nuclear Science Symposium Conference Record M13-369 Simulation of Algorithms for Pulse Timing in FPGAs Michael D. Haselman, Member IEEE, Scott Hauck, Senior Member IEEE, Thomas K. Lewellen, Senior

More information

A Study of Silicon Photomultiplier Sensor Prototypes for Readout of a Scintillating Fiber / Lead Sheet Barrel Calorimeter

A Study of Silicon Photomultiplier Sensor Prototypes for Readout of a Scintillating Fiber / Lead Sheet Barrel Calorimeter 2007 IEEE Nuclear Science Symposium Conference Record N41-6 A Study of Silicon Photomultiplier Sensor Prototypes for Readout of a Scintillating Fiber / Lead Sheet Barrel Calorimeter Carl J. Zorn Abstract:

More information

Electronic Instrumentation for Radiation Detection Systems

Electronic Instrumentation for Radiation Detection Systems Electronic Instrumentation for Radiation Detection Systems January 23, 2018 Joshua W. Cates, Ph.D. and Craig S. Levin, Ph.D. Course Outline Lecture Overview Brief Review of Radiation Detectors Detector

More information

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard.

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard. Chiara Secco PET Performance measurements of the new LSO-Based Whole Body PET/CT Scanner biograph 16 HI-REZ using the NEMA NU 2-2001 Standard. INTRODUCTION Since its introduction, CT has become a fundamental

More information

A test of silicon photomultipliers as readout for PET

A test of silicon photomultipliers as readout for PET Nuclear Instruments and Methods in Physics Research A 545 (25) 75 715 www.elsevier.com/locate/nima A test of silicon photomultipliers as readout for PET A.N. Otte a,, J. Barral b, B. Dolgoshein c, J. Hose

More information

The digital Silicon Photomultiplier A novel Sensor for the Detection of Scintillation Light

The digital Silicon Photomultiplier A novel Sensor for the Detection of Scintillation Light The digital Silicon Photomultiplier A novel Sensor for the Detection of Scintillation Light Carsten Degenhardt, Gordian Prescher, Thomas Frach, Andreas Thon, Rik de Gruyter, Anja Schmitz, Rob Ballizany

More information

Review of Solidstate Photomultiplier. Developments by CPTA & Photonique SA

Review of Solidstate Photomultiplier. Developments by CPTA & Photonique SA Review of Solidstate Photomultiplier Developments by CPTA & Photonique SA Victor Golovin Center for Prospective Technologies & Apparatus (CPTA) & David McNally - Photonique SA 1 Overview CPTA & Photonique

More information

Positron Emission Tomography - PET

Positron Emission Tomography - PET Positron Emission Tomography - PET Positron Emission Tomography Positron Emission Tomography (PET): Coincidence detection of annihilation radiation from positron-emitting isotopes followed by tomographic

More information

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud University of Groningen Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you

More information

Journal of Radiation Protection and Research

Journal of Radiation Protection and Research 1) WOO JIN JO et al: CZT BASED PET SYSTEM IN KAERI Journal of Radiation Protection and Research pissn 2508-1888 eissn 2466-2461 http://dx.doi.org/10.14407/jrpr.2016.41.2.081 Paper Received July 17, 2015

More information

Highlights of Poster Session I: SiPMs

Highlights of Poster Session I: SiPMs Highlights of Poster Session I: SiPMs Yuri Musienko* FNAL(USA)/INR(Moscow) NDIP 2011, Lyon, 5.07.2011 Y. Musienko (Iouri.Musienko@cern.ch) 1 Poster Session I 21 contributions on SiPM characterization and

More information

A High-Resolution GSO-based Brain PET Camera

A High-Resolution GSO-based Brain PET Camera A High-Resolution GSO-based Brain PET Camera J.S. Karp', Senior Member IEEE, L.E. Adam', R.Freifelder', Member IEEE, G. Muehllehner3 Senior Member IEEE, F. Liu"', Student Member IEEE, S. Surti"', Student

More information

Characterization of Silicon Photomultipliers and their Application to Positron Emission Tomography. Zhiwei Yang. Abstract

Characterization of Silicon Photomultipliers and their Application to Positron Emission Tomography. Zhiwei Yang. Abstract DESY Summer Student Program 2009 Report No. Characterization of Silicon Photomultipliers and their Application to Positron Emission Tomography Zhiwei Yang V. N. Karazin Kharkiv National University E-mail:

More information

Small Prototype Gamma Spectrometer Using CsI(Tl) Scintillator Coupled to a Solid-State Photomultiplier

Small Prototype Gamma Spectrometer Using CsI(Tl) Scintillator Coupled to a Solid-State Photomultiplier 10P-58 1 Small Prototype Gamma Spectrometer Using CsI(Tl) Scintillator Coupled to a Solid-State Photomultiplier Eric M. Becker, Member IEEE, Abdollah T. Farsoni, Member, IEEE, Abdulsalam M. Alhawsawi,

More information

Initial evaluation of the Indiana small animal PET scanner

Initial evaluation of the Indiana small animal PET scanner Initial evaluation of the Indiana small animal PET scanner Ned C. Rouze, Member, IEEE, Victor C. Soon, John W. Young, Member, IEEE, Stefan Siegel, Member, IEEE, and Gary D. Hutchins, Member, IEEE Abstract

More information

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud University of Groningen Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you

More information

Overview 256 channel Silicon Photomultiplier large area using matrix readout system The SensL Matrix detector () is the largest area, highest channel

Overview 256 channel Silicon Photomultiplier large area using matrix readout system The SensL Matrix detector () is the largest area, highest channel 技股份有限公司 wwwrteo 公司 wwwrteo.com Page 1 Overview 256 channel Silicon Photomultiplier large area using matrix readout system The SensL Matrix detector () is the largest area, highest channel count, Silicon

More information

A NOVEL CONCEPT FOR A POSITRON EMISSION TOMOGRAPHY SCANNER

A NOVEL CONCEPT FOR A POSITRON EMISSION TOMOGRAPHY SCANNER A NOVEL CONCEPT FOR A POSITRON EMISSION TOMOGRAPHY SCANNER An Undergraduate Research Scholars Thesis by BRIAN KELLY, MATTHEW LEE ELLIOT LEVIN and JEENA KHATRI Submitted to Honors and Undergraduate Research

More information

Monte Carlo Simulation Study of a Dual-Plate PET Camera Dedicated to Breast Cancer Imaging

Monte Carlo Simulation Study of a Dual-Plate PET Camera Dedicated to Breast Cancer Imaging IEEE Nuclear Science Symposium Conference Record M-9 Monte Carlo Simulation Study of a Dual-Plate PET Camera Dedicated to Breast Cancer Imaging Jin Zhang, Member, IEEE, Peter D. Olcott, Member, IEEE, Angela

More information

Application of Silicon Photomultipliers to Positron Emission Tomography

Application of Silicon Photomultipliers to Positron Emission Tomography Annals of Biomedical Engineering, Vol. 39, No. 4, April 2011 (Ó 2011) pp. 1358 1377 DOI: 10.1007/s10439-011-0266-9 Application of Silicon Photomultipliers to Positron Emission Tomography EMILIE RONCALI

More information

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance

More information

A Measurement of the Photon Detection Efficiency of Silicon Photomultipliers

A Measurement of the Photon Detection Efficiency of Silicon Photomultipliers A Measurement of the Photon Detection Efficiency of Silicon Photomultipliers A. N. Otte a,, J. Hose a,r.mirzoyan a, A. Romaszkiewicz a, M. Teshima a, A. Thea a,b a Max Planck Institute for Physics, Föhringer

More information

Design Evaluation of A-PET: A High Sensitivity Animal PET Camera

Design Evaluation of A-PET: A High Sensitivity Animal PET Camera IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 5, OCTOBER 2003 1357 Design Evaluation of A-PET: A High Sensitivity Animal PET Camera S. Surti, Member, IEEE, J. S. Karp, Senior Member, IEEE, A. E. Perkins,

More information

Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM

Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM Preamplifiers and amplifiers The current from PMT must be further amplified before it can be processed and counted (the number of electrons yielded

More information

A new operative gamma camera for Sentinel Lymph Node procedure

A new operative gamma camera for Sentinel Lymph Node procedure A new operative gamma camera for Sentinel Lymph Node procedure A physicist device for physicians Samuel Salvador, Virgile Bekaert, Carole Mathelin and Jean-Louis Guyonnet 12/06/2007 e-mail: samuel.salvador@ires.in2p3.fr

More information

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC R. Bellazzini a,b, G. Spandre a*, A. Brez a, M. Minuti a, M. Pinchera a and P. Mozzo b a INFN Pisa

More information

SensL B-Series Silicon Photomultipliers for TOF- PET. NDIP2014 Kevin O Neill 4 th July, 2014

SensL B-Series Silicon Photomultipliers for TOF- PET. NDIP2014 Kevin O Neill 4 th July, 2014 SensL B-Series Silicon Photomultipliers for TOF- PET NDIP2014 Kevin O Neill 4 th July, 2014 1 Outline Performance-limiting physics of SiPM sensors Photon Detection Efficiency Dark count rate Crosstalk

More information

High granularity scintillating fiber trackers based on Silicon Photomultiplier

High granularity scintillating fiber trackers based on Silicon Photomultiplier High granularity scintillating fiber trackers based on Silicon Photomultiplier A. Papa Paul Scherrer Institut, Villigen, Switzerland E-mail: angela.papa@psi.ch Istituto Nazionale di Fisica Nucleare Sez.

More information

Combined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8

Combined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8 Combined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8 UNIVERSITY OF CAMBRIDGE 1.2 Rob C. Hawkes 1, Tim D. Fryer 1, Alun J. Lucas 1,2, Stefan B. Siegel 3, Richard

More information

Simulations of the J-PET detector response with the GATE package

Simulations of the J-PET detector response with the GATE package Simulations of the J-PET detector response with the GATE package Author: pawel.kowalski@ncbj.gov.pl 22nd to 24th September 2014 II Symposium on Positron Emission Tomography Outline 1. Introduction 2. Simulation

More information

Design of a High Resolution and High Sensitivity Scintillation Crystal Array with Nearly Perfect Light Collection

Design of a High Resolution and High Sensitivity Scintillation Crystal Array with Nearly Perfect Light Collection Design of a High Resolution and High Sensitivity Scintillation Crystal Array with Nearly Perfect Light Collection Craig S. Levin, Member, IEEE Abstract-- Spatial resolution improvements in Positron Emission

More information

FPGA-Based Data Acquisition System for a Positron Emission Tomography (PET) Scanner

FPGA-Based Data Acquisition System for a Positron Emission Tomography (PET) Scanner FPGA-Based Data Acquisition System for a Positron Emission Tomography (PET) Scanner Michael Haselman 1, Robert Miyaoka 2, Thomas K. Lewellen 2, Scott Hauck 1 1 Department of Electrical Engineering, 2 Department

More information

Over the past decade, many small-animal PET scanners

Over the past decade, many small-animal PET scanners A Prototype PET Scanner with DOI-Encoding Detectors Yongfeng Yang 1, Yibao Wu 1, Jinyi Qi 1, Sara St. James 1, Huini Du 1, Purushottam A. Dokhale 2, Kanai S. Shah 2, Richard Farrell 2, and Simon R. Cherry

More information

Concept and status of the LED calibration system

Concept and status of the LED calibration system Concept and status of the LED calibration system Mathias Götze, Julian Sauer, Sebastian Weber and Christian Zeitnitz 1 of 14 Short reminder on the analog HCAL Design is driven by particle flow requirements,

More information

LaBr 3 :Ce, the latest crystal for nuclear medicine

LaBr 3 :Ce, the latest crystal for nuclear medicine 10th Topical Seminar on Innovative Particle and Radiation Detectors 1-5 October 2006 Siena, Italy LaBr 3 :Ce, the latest crystal for nuclear medicine Roberto Pani On behalf of SCINTIRAD Collaboration INFN

More information

Design, Calibration, and Evaluation of Depth-of- Interaction-Capable PET Detector Modules

Design, Calibration, and Evaluation of Depth-of- Interaction-Capable PET Detector Modules Texas Medical Center Library DigitalCommons@TMC UT GSBS Dissertations and Theses (Open Access) Graduate School of Biomedical Sciences 12-2012 Design, Calibration, and Evaluation of Depth-of- Interaction-Capable

More information

Development of the Pixelated Photon Detector. Using Silicon on Insulator Technology. for TOF-PET

Development of the Pixelated Photon Detector. Using Silicon on Insulator Technology. for TOF-PET July 24, 2015 Development of the Pixelated Photon Detector Using Silicon on Insulator Technology for TOF-PET A.Koyama 1, K.Shimazoe 1, H.Takahashi 1, T. Orita 2, Y.Arai 3, I.Kurachi 3, T.Miyoshi 3, D.Nio

More information

Exprerimental Evaluation of a Dedicated Pinhole SPECT System for Small Animal Imaging and Scintimammography

Exprerimental Evaluation of a Dedicated Pinhole SPECT System for Small Animal Imaging and Scintimammography ETASR - Engineering, Technology & Applied Science Research Vol. 1, o. 1, 211, 17-22 17 Exprerimental Evaluation of a Dedicated Pinhole SPECT System for Small Animal Imaging and Scintimammography S. David

More information

A comparative study of the time performance between NINO and FlexToT ASICs

A comparative study of the time performance between NINO and FlexToT ASICs Journal of Instrumentation OPEN ACCESS A comparative study of the time performance between NINO and FlexToT ASICs To cite this article: I. Sarasola et al View the article online for updates and enhancements.

More information