A PET detector module using FPGA-only MVT digitizers

Size: px
Start display at page:

Download "A PET detector module using FPGA-only MVT digitizers"

Transcription

1 A PET detector module using FPGA-only MVT digitizers Daoming Xi, Student Member, IEEE, Chen Zeng, Wei Liu, Student Member, IEEE, Xiang Liu, Lu Wan, Student Member, IEEE, Heejong Kim, Member, IEEE, Luyao Wang, Chien-Min Kao, Senior Member, IEEE, and Qingguo Xie, Member, IEEE Abstract Multi-voltage threshold (MVT) is an amplitudebased sampling method. It takes timing samples when the event pulse crosses the user-defined thresholds. Only a few comparators and TDCs are required when implementing such digitizer. Previously, we have demonstrated an FPGA-only MVT digitizer based on this method. The FPGA-only MVT digitizer employs the differential I/Os in an FPGA as the required comparators and FPGA based TDCs. The implementation of this digitizer is entirely based on the FPGA. We have demonstrated that it is possible to implement a significant number of MVT digitizers by using a single FPGA. It is also flexible, as it allows us to readily modify, or add functions to, the implementation without requiring costly hardware changes. Currently, we are developing a PET detector module using the FPGA-only MVT digitizer. In this paper we describe the design and implementation of the detector module and report its performance properties. The detector module has a total detection sensitive area of 50mm 50mm, an overall energy resolution of 15.1% FWHM at 511keV, and a module-level coincidence timing resolution of 684ps FWHM. In addition, our preliminary imaging with such detector module successfully resolves 1.6mm-diameter rods separated by 3.2mm. Index Terms FPGA, LVDS, MVT, Digitizer, all digital data acquisition (DAQ) system, block detector, PET. I. INTRODUCTION RECENTLY, there is a strong interest in sampling the PET event waveform at the earliest possible stage of the signal detection chain and several methods to achieve this have emerged [1] [4]. By sampling signals early, information loss due to, for example, filtering/shapping and transmission This work was supported in part by the Natural Science Foundation of China (NSFC) Grant #U , # , # , in part by the National Key Technology R&D Program of China Grant #2012BAI13B06, in part by the Key Grant Project of Chinese Ministry of Education #313023, in part by the Ministry of Science and Technology of China (MOST) grant #2012DFG31970, the Research and Development Programme of Hubei Province grant #2011BFA005, the Wuhan Programs for Science and Technology Development grant # It was also supported in part by USA Clinical and Translational Science Awards (CTSA) grant CTSA UL1 TR and National Institutes of Health (NIH) grant R01 EB A1. Qingguo Xie is with the Wuhan National Laboratory for Optoelectronics, Wuhan, Hubei, China and also with the Department of Biomedical Engineering, Huazhong University of Science and Technology, Wuhan, Hubei, China (Tel: , qgxie@mail.hust.edu.cn). Daoming Xi and Luyao Wang are with the Department of Biomedical Engineering, Huazhong University of Science and Technology, Wuhan, Hubei, China and also with the Raycan Technology Co., Ltd (Suzhou), Suzhou, Jiang Su, China. Wei Liu, Chen Zeng, Xiang Liu and Lu Wan are with the Department of Biomedical Engineering, Huazhong University of Science and Technology, Wuhan, Hubei, China. Chien-Min Kao and Heejong Kim are with the Department of Radiology, the University of Chicago, Chicago, Illinois, USA. can be minimized. By converting the event signals into digital numbers, it also becomes possible to tap the power of modern digital electronics to apply nontrivial signal processing algorithms to achieve many benefits, including enhanced detection performance and easier to perform upgrade and adjust for varying detector properties due to, for example, aging and radiation exposure. Also, the same data acquisition (DAQ) electronics may be used for a variety of detector designs by changing the signal processing algorithms. By using such kind of universal DAQ, one can shorten the PET system development cycle and reduce the development cost. We have previously proposed one such sampling method, which we refer to as the multi-voltage threshold (MVT) method. This method provides amplitude-based sampling in the sense that the signal waveform is sampled with respect to a user-defined set of reference amplitudes [5] [7]. In contrast, the conventional sampling method is time-based in the sense that the signal waveform is sampled at a (approximately) regular interval in time. Modern PET detectors use fast scintillators; therefore, a critical challenge in using the time-based sampling method is the need to have a high sampling rate (above 1 Gsps). Several research groups have demonstrated the Domino Ring Sampler (DRS) can be used to provide such a high sampling rate but the count-rate performance is a concern that remains to be addressed [1], [8]. On the other hand, we have shown that it is sufficient to use only a small number of reference amplitudes with the MVT method [5]. We have also recently demonstrated that it is possible to implement a significant number of MVT digitizers by using a single FPGA [9]. We believe that this sampling technology can provide low-cost and versatile DAQ for developing PET detectors. In this work, we developed a PET detector module using the FPGA-only MVT digitizers and LYSO/SiPM detector block. In Sec. 2, we show the design of the detector module. In Sec. 3 we measured the performance of the detector module and also obtain a imaging by using of a stable dual-head detection system which is consisted of such 2 detector modules. Conclusions and discussion is followed in Sec. 4. II. DETECTOR DESIGN Previously, we have implemented a FPGA-only MVT digitizer by using of the differential I/Os in a FPGA which is configured to work in the low-voltage-differential-signaling (LVDS) receiver mode (refer as LVDS comparators in the

2 Fig. 1. A 4-level MVT digitizing channel in the FPGA-only implementation. Fig. 2. The digital PET detector module. following paragraph). Such LVDS receiver in the FPGA has a positive input and a negative input and its output is determined by the relative voltages appearing on these inputs. It sends a logic 1 when the voltage at its positive input is higher than that at the negative input; otherwise, it sends out a logic 0.. The schematics of the digitizer is shown in Fig. 1. It has four LVDS comparators for providing four reference voltages. As shown, the PET event pulse is split into four signals and each of which is fed to the positive node of an LVDS receiver. The negative node of the LVDS receiver is connected to a DAC to provide a programmable voltage threshold. The logic output of the LVDS comparator is then connected to two TDCs for determining the digital times of its positive and negative transitions. Thus, the four-level channel generates eight samples for a pulse that has a sufficient amplitude. The TDCs are also implemented inside the FPGA by using the Song s method [10]. In [9], we have reported a single channel implementation on the Altera Cyclone II EP2C70F896C7 FPGA and described methods for calibrating the implementation. Using this FPGA-only 4-channel MVT digitizer for an mm 3 LSO coupled to a SensL SiPM, we obtained an energy resolution of 16%@511keV and a coincidence timing resolution of 500 ps. By collaborating with the Raycan Technology Co., Ltd (Suzhou), we have recently developed a FPGA-only MVT digitizer board that has 48 4-level MVT digitizer channels. In this work, we setup a detector by connecting 3 such digitizer boards, which provide a total of 144 digitizing channels, to a 2x2 array of LYSO/SiPM detector blocks, each of which has 36 output signals (hence a total of 144 output signals). As shown in Fig.2, the detector module consists of two physical units, including the detector unit (DU) and the digital acquiring unit (DAU). The scintillator pulses generated from the DU are digitized by the DAU. The resulting samples are packaged into event frames and fed to a computer via an Ethernet interface. The event information is then extracted by digitally analyzing the samples on the computer. The energy calibration, crystal identification and coincidence detection are also performed on the computer. An important feature of our detector design is that the DAU can be readily used for other DUs that may use different scintillation materials, crystal sizes, or photodetectors, through Fig. 3. The LYSO/SiPM detector block. Left: A 6 6 SiPM array. Middle: A 6 6 LYSO matrix. Right: A LYSO/SiPM detector block the use of adequate adaptors. This is possible because, as discussed above, different processing algorithms can be readily developed and implemented to handle the different characteristics of the signal pulse once it is converted into digital form by the DAU. A. The Detector Unit The DU consists of a 2 2 array of LYSO/SiPM detector blocks. As shown in Fig. 3, each LYSO/SiPM detector consists of a 6 6 SiPM array and a 6 6 LYSO matrix using one-to-one coupling. The SiPM is SensL FM In the array, each SiPM pixel has an mm 2 active area with a mm 2 pitch. The LYSO matrix is consisted of mm 3 crystals with a 0.3 mm gap. Currently, the LYSO matrix is directly coupled to the SiPM array using epoxy optical adhesive (using GHJ-01(Z) from Chenguang research institute of chemical industry). Since the crystal size is larger than the SiPM s active area, we are considering using taped crystal bars to provide a better optical coupling between the LYSO and SiPM in future. B. The digital acquiring unit The DAU is consisted of 3 digitizer boards. We show one of the digitizer boards in Fig. 4. It contains 4 Altera Cyclone II EP2C50U484C6 FPGAs and each of them provides 12 4-level MVT digitizing channels described above, yielding a total of 48 digitizing channels. The waveform samples generated by the EP2C50U484C6 FPGAs are readout by another Altera Cyclone II EP2C50F484C7 FPGA and output via a Gigabit

3 Fig. 4. The 48 channel 4-level MVT digitizer board. Fig. 6. Count distribution obtained by placing a point-like source in front of the detector module. Fig. 5. The experiment setup for measuring the performance of the detector modules and obtaining the initial imaging result. Each module has a detection sensitive area of 50mm 50mm Ethernet communicator. This FPGA also controls the DACs (using DAC7678 from TI) to provide programmable reference voltages. C. The samples analyzing The event samples acquired by the detector module are first stored on a computer. The resulted samples are fitted, as described in [9], by a bi-exponential function for estimating the event energy and time. The bi-exponential function is given by ( y (t) =a exp t t ) [ ( 0 1 exp t t )] 0, (1) b d where a is determined by the pulse amplitude, b and d by the rise and decay time of the pulse, and t 0 by the occurrence time of the pulse. The event energy and timing is obtained with the re-generated pulse from the fitted function. III. PERFORMANCE MEASUREMENTS AND INITIAL IMAGING As shown in Fig. 5, we set up a dual head detection system by using a pair of detector modules. The distance between the two heads is 26mm. The resulted samples from each digitizer board is collected and analyzed. A clock distributor is used to synchronize all the six digitizer boards. The results reported below are all measured by using this detection system. A. Crystal Identification Since the LYSO crystals are coupled to the SiPMs one by one and the SiPMs output are individually sampled, we can directly get the events position by identifying the digitizer channel number. Fig. 6 shows the counts distribution in each crystal. As a point-like source is placed at the front of the detector, the count in the center crystal is higher than those in the border crystals. This result indicates that all the channels in the detector modules work properly. B. Energy Resolution The top graph in Fig.7 shows the pulse height spectra for individual crystals in a detector module. A Gaussian function fitting is applied to each spectra to obtain the photo-peak position and energy resolution for individual crystals. Then, based on the obtained photo-peak position the pulse height measurement is calibrated to event energy measurement. The energy resolutions of the crystals vary from 11.2% to 24.6% FWHM at 511keV for a detector module. The bottom plot of Fig. 7 shows the summed energy spectra of one detector module, after applying energy calibration. The overall energy resolution is 15.1% FWHM at 511keV. C. Timing Resolution Figure 8 shows the measured coincidence timing histogram (using the events between 450keV and 650keV ) of a pair of detector modules using the dual-head detection system shown in Fig.5. This result is obtained by filling a single 1.6 mmdiameter channel of the phantom shown in Fig. 9 with FDG and placing the phantom at the center of the setup, by using an energy window of 450keV to 650keV. By fitting a Gaussian function to this histogram, a module-level coincidence timing resolution of 684ps FWHM is obtained.

4 Fig. 9. A homemade micro-derenzo phantom consisting of six groups of hollow channels having diameters of 2.4mm, 2.0mm, 1.6mm, 1.2mm, 1.0mm, 0.7mm. Also shown is the transverse slices of the reconstructed image of the phantom generated by an maximum likelihood expectation maximization algorithm. The 1.6mm-diameter rods can be resolved. Fig. 7. Top: The pulse height spectra obtained for the crystals in a detector module. The energy resolutions of individual crystals range from 11.2% to 24.6%. Bottom: The average energy spectra which indicates an over-all energy resolution of 15.1% FWHM at 511keV. D. Phantom Imaging Figure 9 shows the the reconstructed images, and a photo, of the homemade micro-derenzo phantom that contains six groups of hollow channels having diameters ranging from 0.7mm to 2.4mm. The center-to-center spacing between adjacent channels in the same group is twice the channel diameter. All the hollow channels in the phantom is filled with FDG, and placed at the center of the dual-head detection system with its axis vertical to the detector module. When acquiring the data, both the phantom and the detection system are kept stationary. The coincidence event are then obtained by using a 10ns coincidence timing window and a 450keV -650keV energy window. A model-based image reconstruction approach is used to obtain the tomographic image of the phantom [11]. As shown in Fig.9, the 1.6mm-diameter rods can be resolved in the reconstructed image. In this initial study, no normalization, attenuation correction or random correction is employed. Fig. 8. The coincidence timing histogram obtained for a pair of detector module. By fitting this histogram with a Gaussian function (red curve), we obtain a module-level coincidence timing resolution of 684ps FWHM. IV. CONCLUSIONS AND DISCUSSION In this work, we develop a digital PET detector module. It employs FPGA-only MVT digitizers to generate samples from which event energy and timing are estimated by digital signal processing. The digital signal processing algorithms are implemented in the computer. The experimental results show that our detector module provides an energy resolution in the range of 11.2% to 24.6% FWHM at 511keV (with an overall energy resolution of 15.1%) and a module-level coincidence timing resolution of 684ps FWHM. These performance properties are adequate for human and small-animal imaging. Our

5 initial imaging result also shows that a PET scanner based on these modules has the capability to resolve 1.6mm structures. V. ACKNOWLEDGE The authors thank Dr. Peng Xiao and Mr. Xiaoqing Cao for discussion. We also thank the PET center of the Union Hospital, Tongji Medical College of Huazhong University of Science and Technology for their generous support with radioisotopes and other resources. REFERENCES [1] H. Kim, C.-M. Kao, S. Kim, and C.-T. Chen, A development of waveform sampling readout board for PET using DRS4, in Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE, 2011, pp [2] W. J. Ashmanskas, B. C. LeGeyt, F. M. Newcomer, R. Van Berg, R. Wiener, and J. Karp, Waveform-sampling electronics for timeof-flight PET scanner, in Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE, 2011, pp [3] R. Fontaine, F. Belanger, J. Cadorette, J. D. Leroux, J. P. Martin, J. B. Michaud, J.-F. Pratte, S. Robert, and R. Lecomte, Architecture of a dual-modality, high-resolution, fully digital positron emission tomography/computed tomography (PET/CT) scanner for small animal imaging, Nuclear Science, IEEE Transactions on, vol. 52, no. 3, pp , [4] Q. Xie, C.-M. Kao, Z. Hsiau, and C.-T. Chen, A new approach for pulse processing in positron emission tomography, Nuclear Science, IEEE Transactions on, vol. 52, no. 4, pp , [5] Q. Xie, Y. Chen, J. Zhu, J. Liu, X. Wang, W. Liu, X. Chen, M. Niu, Z. Wu, D. Xi, L. Wang, P. Xiao, C.-T. Chen, and C.-M. Kao, Implementation of LYSO/PSPMT block detector with all digital DAQ system, Nuclear Science, IEEE Transactions on, vol. 60, no. 3, pp , [6] Q. Xie, C.-M. Kao, X. Wang, N. Guo, C. Zhu, F. H., M. W. W., and C.- T. Chen, Potentials of digitally sampling scintillation pulses in timing determination in PET, Nuclear Science, IEEE Transactions on, vol. 56, no. 5, pp , [7] H. Kim, C.-M. Kao, Q. Xie, C.-T. Chen, Z. L., F. Tang, H. Frisch, M. W. W., and W.-S. Choong, A multi-threshold sampling method for TOF-PET signal processing, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, vol. 602, no. 2, pp , [8] H. Kim, C.-T. Chen, H. Frisch, F. Tang, and C.-M. Kao, A prototype tof pet detector module using a micro-channel plate photomultiplier tube with waveform sampling, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, vol. 662, no. 1, pp , [9] D. Xi, C.-M. Kao, W. Liu, C. Zeng, X. Liu, and Q. Xie, FPGA-only MVT digitizer for TOF PET, Nuclear Science, IEEE Transactions on, vol. 60, no. 5, pp , [10] J. Song, Q. An, and S. Liu, A high-resolution time-to-digital converter implemented in field-programmable-gate-arrays, IEEE Trans. Nucl. Sci., vol. 53, no. 1, pp , Feb [11] C.-M. Kao, Q. Xie, Y. Dong, L. Wan, and C.-T. Chen, A high-sensitivity small-animal pet scanner: Development and initial performance measurements, Nuclear Science, IEEE Transactions on, vol. 56, no. 5, pp , 2009.

PROGRESS in TOF PET timing resolution continues to

PROGRESS in TOF PET timing resolution continues to Combined Analog/Digital Approach to Performance Optimization for the LAPET Whole-Body TOF PET Scanner W. J. Ashmanskas, Member, IEEE, Z. S. Davidson, B. C. LeGeyt, F. M. Newcomer, Member, IEEE, J. V. Panetta,

More information

Simulation of Algorithms for Pulse Timing in FPGAs

Simulation of Algorithms for Pulse Timing in FPGAs 2007 IEEE Nuclear Science Symposium Conference Record M13-369 Simulation of Algorithms for Pulse Timing in FPGAs Michael D. Haselman, Member IEEE, Scott Hauck, Senior Member IEEE, Thomas K. Lewellen, Senior

More information

Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET

Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET A. Kuhn, S. Surti, Member, IEEE, J. S. Karp, Senior Member, IEEE, G. Muehllehner, Fellow, IEEE, F.M. Newcomer, R. VanBerg Abstract--

More information

Potentials of Digitally Sampling Scintillation Pulses in Timing Determination in PET

Potentials of Digitally Sampling Scintillation Pulses in Timing Determination in PET Potentials of Digitally Sampling Scintillation Pulses in Timing Determination in PET Journal: IEEE Transactions on Nuclear Science Manuscript ID: TNS--.R Manuscript Type: Imaging and Instrumentation for

More information

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud University of Groningen Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you

More information

Development of the LBNL Positron Emission Mammography Camera

Development of the LBNL Positron Emission Mammography Camera Development of the LBNL Positron Emission Mammography Camera J.S. Huber, Member, IEEE, W.S. Choong, Member, IEEE, J. Wang, Member, IEEE, J.S. Maltz, Member, IEEE, J. Qi, Member, IEEE, E. Mandelli, Member,

More information

An innovative detector concept for hybrid 4D-PET/MRI Imaging

An innovative detector concept for hybrid 4D-PET/MRI Imaging Piergiorgio Cerello (INFN - Torino) on behalf of the 4D-MPET* project *4 Dimensions Magnetic compatible module for Positron Emission Tomography INFN Perugia, Pisa, Torino; Polytechnic of Bari; University

More information

Study of Silicon Photomultipliers for Positron Emission Tomography (PET) Application

Study of Silicon Photomultipliers for Positron Emission Tomography (PET) Application Study of Silicon Photomultipliers for Positron Emission Tomography (PET) Application Eric Oberla 5 June 29 Abstract A relatively new photodetector, the silicon photomultiplier (SiPM), is well suited for

More information

Development of PET using 4 4 Array of Large Size Geiger-mode Avalanche Photodiode

Development of PET using 4 4 Array of Large Size Geiger-mode Avalanche Photodiode 2009 IEEE Nuclear Science Symposium Conference Record M09-8 Development of PET using 4 4 Array of Large Size Geiger-mode Avalanche Photodiode K. J. Hong, Y. Choi, J. H. Kang, W. Hu, J. H. Jung, B. J. Min,

More information

Design Studies of A High-Performance Onboard Positron Emission Tomography For Integrated Small Animal PET/CT/RT Radiation Research Systems

Design Studies of A High-Performance Onboard Positron Emission Tomography For Integrated Small Animal PET/CT/RT Radiation Research Systems Proceedings of the International MultiConference of Engineers and Computer Scientists 2018 Vol II Design Studies of A High-Performance Onboard Positron Emission Tomography For Integrated Small Animal PET/CT/RT

More information

A Simple All-digital PET System

A Simple All-digital PET System A Simple All-digital PET System Qingguo Xie 12,Chien-MinKao 2, Rongsheng Xia 3,XiWang 3,NaLi 3,XinJiang 3,LiZhi 2, Zhi Zhang 2, Zhonghua Deng 3, Chin-Tu Chen 1 1 Department of Radiology, The University

More information

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET 2005 IEEE Nuclear Science Symposium Conference Record M11-126 Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET Jin Zhang, Member,

More information

Development of a High-Resolution and Depth-of- Interaction Capable Detector for Time-of-Flight PET

Development of a High-Resolution and Depth-of- Interaction Capable Detector for Time-of-Flight PET Development of a High-Resolution and Depth-of- Interaction Capable Detector for Time-of-Flight PET Srilalan Krishnamoorthy, Member, IEEE, Rony I. Wiener, Madhuri Kaul, Joseph Panetta, Joel S. Karp, Senior

More information

PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner

PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner September, 2017 Results submitted to Physics in Medicine & Biology Negar Omidvari 1, Jorge Cabello 1, Geoffrey Topping

More information

Development of an innovative LSO-SiPM detector module for high-performance Positron Emission Tomography

Development of an innovative LSO-SiPM detector module for high-performance Positron Emission Tomography Development of an innovative LSO-SiPM detector module for high-performance Positron Emission Tomography Maria Leonor Trigo Franco Frazão leonorfrazao@ist.utl.pt Instituto Superior Técnico, Lisboa, Portugal

More information

Development of a 256-channel Time-of-flight Electronics System For Neutron Beam Profiling

Development of a 256-channel Time-of-flight Electronics System For Neutron Beam Profiling JOURNAL OF L A TEX CLASS FILES, VOL. 14, NO. 8, AUGUST 2015 1 Development of a 256-channel Time-of-flight Electronics System For Neutron Beam Profiling Haolei Chen, Changqing Feng, Jiadong Hu, Laifu Luo,

More information

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A Nuclear Instruments and Methods in Physics Research A 614 (2010) 308 312 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima

More information

PET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002

PET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002 PET Detectors William W. Moses Lawrence Berkeley National Laboratory March 26, 2002 Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes

More information

Celesteion Time-of-Flight Technology

Celesteion Time-of-Flight Technology Celesteion Time-of-Flight Technology Bing Bai, PhD Clinical Sciences Manager, PET/CT Canon Medical Systems USA Introduction Improving the care for every patient while providing a high standard care to

More information

Design of a Novel Front-End Readout ASIC for PET Imaging System *

Design of a Novel Front-End Readout ASIC for PET Imaging System * Journal of Signal and Information Processing, 2013, 4, 129-133 http://dx.doi.org/10.4236/jsip.2013.42018 Published Online May 2013 (http://www.scirp.org/journal/jsip) 129 Design of a Novel Front-End Readout

More information

Design and development of compact readout electronics with silicon photomultiplier array for a compact imaging detector *

Design and development of compact readout electronics with silicon photomultiplier array for a compact imaging detector * CPC(HEP & NP), 2012, 36(10): 973 978 Chinese Physics C Vol. 36, No. 10, Oct., 2012 Design and development of compact readout electronics with silicon photomultiplier array for a compact imaging detector

More information

Investigation of a Transmission-Line Readout for Building PET Detector Modules

Investigation of a Transmission-Line Readout for Building PET Detector Modules Investigation of a Transmission-Line Readout for Building PET Detector Modules Contents 1. Introduction 2. Simulation Setup 3. Results 4.Summary and Plans Heejong Kim, Univ. of Chicago Pico-Second Workshop

More information

Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner

Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner Kisung Lee, Member, IEEE, Paul E. Kinahan, Senior Member, Robert S. Miyaoka, Member, IEEE, Jeffrey A. Fessler,

More information

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud University of Groningen Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you

More information

Conceptual Study of Brain Dedicated PET Improving Sensitivity

Conceptual Study of Brain Dedicated PET Improving Sensitivity Original Article PROGRESS in MEDICAL PHYSICS 27(4), Dec. 2016 https://doi.org/10.14316/pmp.2016.27.4.236 pissn 2508-4445, eissn 2508-4453 Conceptual Study of Brain Dedicated PET Improving Sensitivity Han-Back

More information

2594 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER /$ IEEE

2594 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER /$ IEEE 2594 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER 2009 Investigation of Depth of Interaction Encoding for a Pixelated LSO Array With a Single Multi-Channel PMT Yongfeng Yang, Member, IEEE,

More information

Timing Resolution Performance Comparison for Fast and Standard Outputs of SensL SiPM

Timing Resolution Performance Comparison for Fast and Standard Outputs of SensL SiPM Timing Resolution Performance Comparison for Fast and Standard Outputs of SensL SiPM Sergei Dolinsky, Geng Fu, and Adrian Ivan Abstract A new silicon photomultiplier (SiPM) with a unique fast output signal

More information

EndoTOFPET-US: an endoscopic Positron Emission Tomography detector for a novel multimodal medical imaging tool

EndoTOFPET-US: an endoscopic Positron Emission Tomography detector for a novel multimodal medical imaging tool : an endoscopic Positron Emission Tomography detector for a novel multimodal medical imaging tool DESY and University of Hamburg E-mail: daniele.cortinovis@desy.de The EndoTOFPET-US collaboration aims

More information

CAEN. Electronic Instrumentation. CAEN Silicon Photomultiplier Kit

CAEN. Electronic Instrumentation. CAEN Silicon Photomultiplier Kit CAEN Tools for Discovery Electronic Instrumentation CAEN Silicon Photomultiplier Kit CAEN realized a modular development kit dedicated to Silicon Photomultipliers, representing the state-of-the art in

More information

PoS(PhotoDet 2012)016

PoS(PhotoDet 2012)016 SiPM Photodetectors for Highest Time Resolution in PET, E. Auffray, B. Frisch, T. Meyer, P. Jarron, P. Lecoq European Organization for Nuclear Research (CERN), 1211 Geneva 23, Switzerland E-mail: stefan.gundacker@cern.ch

More information

APD Quantum Efficiency

APD Quantum Efficiency APD Quantum Efficiency Development of a 64-channel APD Detector Module with Individual Pixel Readout for Submillimeter Spatial Resolution in PET Philippe Bérard a, Mélanie Bergeron a, Catherine M. Pepin

More information

Performance Evaluation of SiPM Detectors for PET Imaging in the Presence of Magnetic Fields

Performance Evaluation of SiPM Detectors for PET Imaging in the Presence of Magnetic Fields 2008 IEEE Nuclear Science Symposium Conference Record M02-4 Performance Evaluation of SiPM Detectors for PET Imaging in the Presence of Magnetic Fields Samuel España, Student Member, IEEE, Gustavo Tapias,

More information

Simulation and evaluation of a cost-effective high-performance brain PET scanner.

Simulation and evaluation of a cost-effective high-performance brain PET scanner. Research Article http://www.alliedacademies.org/biomedical-imaging-and-bioengineering/ Simulation and evaluation of a cost-effective high-performance brain PET scanner. Musa S Musa *, Dilber U Ozsahin,

More information

764 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 3, JUNE 2004

764 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 3, JUNE 2004 764 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 3, JUNE 2004 Study of Low Noise Multichannel Readout Electronics for High Sensitivity PET Systems Based on Avalanche Photodiode Arrays Frezghi Habte,

More information

Design and development of compact readout electronics with silicon photomultiplier array for a compact imaging detector

Design and development of compact readout electronics with silicon photomultiplier array for a compact imaging detector University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2012 Design and development of compact readout

More information

Performance measurements of a depth-encoding PET detector module based on positionsensitive

Performance measurements of a depth-encoding PET detector module based on positionsensitive Home Search Collections Journals About Contact us My IOPscience Performance measurements of a depth-encoding PET detector module based on positionsensitive avalanche photodiode read-out This article has

More information

First Applications of the YAPPET Small Animal Scanner

First Applications of the YAPPET Small Animal Scanner First Applications of the YAPPET Small Animal Scanner Guido Zavattini Università di Ferrara CALOR2 Congress, Annecy - FRANCE YAP-PET scanner Scintillator: YAP:Ce Size: matrix of 2x2 match like crystals

More information

LaBr 3 :Ce, the latest crystal for nuclear medicine

LaBr 3 :Ce, the latest crystal for nuclear medicine 10th Topical Seminar on Innovative Particle and Radiation Detectors 1-5 October 2006 Siena, Italy LaBr 3 :Ce, the latest crystal for nuclear medicine Roberto Pani On behalf of SCINTIRAD Collaboration INFN

More information

Traditional analog QDC chain and Digital Pulse Processing [1]

Traditional analog QDC chain and Digital Pulse Processing [1] Giuliano Mini Viareggio April 22, 2010 Introduction The aim of this paper is to compare the energy resolution of two gamma ray spectroscopy setups based on two different acquisition chains; the first chain

More information

Development of a simplified readout for a compact gamma camera based on 2 2 H8500 multi-anode PSPMT array

Development of a simplified readout for a compact gamma camera based on 2 2 H8500 multi-anode PSPMT array University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2010 Development of a simplified readout for a

More information

Simulations of the J-PET detector response with the GATE package

Simulations of the J-PET detector response with the GATE package Simulations of the J-PET detector response with the GATE package Author: pawel.kowalski@ncbj.gov.pl 22nd to 24th September 2014 II Symposium on Positron Emission Tomography Outline 1. Introduction 2. Simulation

More information

Arrays of digital Silicon Photomultipliers Intrinsic performance and Application to Scintillator Readout

Arrays of digital Silicon Photomultipliers Intrinsic performance and Application to Scintillator Readout Arrays of digital Silicon Photomultipliers Intrinsic performance and Application to Scintillator Readout Carsten Degenhardt, Ben Zwaans, Thomas Frach, Rik de Gruyter Philips Digital Photon Counting NSS-MIC

More information

Development of LYSO detector modules for a charge-particle EDM polarimeter

Development of LYSO detector modules for a charge-particle EDM polarimeter Mitglied der Helmholtz-Gemeinschaft Development of LYSO detector modules for a charge-particle EDM polarimeter on behalf of the JEDI collaboration Dito Shergelashvili, PhD student @ SMART EDM_Lab, TSU,

More information

A Real-time Photoacoustic Imaging System with High Density Integrated Circuit

A Real-time Photoacoustic Imaging System with High Density Integrated Circuit 2011 3 rd International Conference on Signal Processing Systems (ICSPS 2011) IPCSIT vol. 48 (2012) (2012) IACSIT Press, Singapore DOI: 10.7763/IPCSIT.2012.V48.12 A Real-time Photoacoustic Imaging System

More information

Investigation of low noise, low cost readout electronics for high sensitivity PET systems based on Avalanche Photodiode arrays

Investigation of low noise, low cost readout electronics for high sensitivity PET systems based on Avalanche Photodiode arrays Investigation of low noise, low cost readout electronics for high sensitivity PET systems based on Avalanche Photodiode arrays Frezghi Habte, Member, IEEE and Craig S.Levin, Member, IEEE Abstract A compact,

More information

4 Time walk correction for TOF-PET detectors based on a monolithic scintillation crystal coupled to a photosensor array

4 Time walk correction for TOF-PET detectors based on a monolithic scintillation crystal coupled to a photosensor array 4 Time walk correction for TOF-PET detectors based on a monolithic scintillation crystal coupled to a photosensor array This chapter has been published as: R. Vinke, H. Löhner, D. Schaart, H. van Dam,

More information

Highlights of Poster Session I: SiPMs

Highlights of Poster Session I: SiPMs Highlights of Poster Session I: SiPMs Yuri Musienko* FNAL(USA)/INR(Moscow) NDIP 2011, Lyon, 5.07.2011 Y. Musienko (Iouri.Musienko@cern.ch) 1 Poster Session I 21 contributions on SiPM characterization and

More information

LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option

LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option Y. Bercier, Member, IEEE, M. Casey, Member, IEEE, J. Young, Member, IEEE, T. Wheelock, Member, IEEE, T. Gremillion Abstract-- Factors which

More information

Test and Simulation of Plastic Scintillator Strips readout by Silicon Photomultipliers

Test and Simulation of Plastic Scintillator Strips readout by Silicon Photomultipliers Test and Simulation of Plastic Scintillator Strips readout by Silicon Photomultipliers Tosi Nicolò, Balbi G., Boldini M., Cafaro V., Dallavalle G.M., D Antone I., Fabbri F., Giordano V., Lax I., Montanari

More information

PACS codes: Qx, Nc, Kv, v Keywords: Digital data acquisition, segmented HPGe detectors, clock and trigger distribution

PACS codes: Qx, Nc, Kv, v Keywords: Digital data acquisition, segmented HPGe detectors, clock and trigger distribution Clock and Trigger Synchronization between Several Chassis of Digital Data Acquisition Modules W. Hennig, H. Tan, M. Walby, P. Grudberg, A. Fallu-Labruyere, W.K. Warburton, XIA LLC, 31057 Genstar Road,

More information

Effects of Dark Counts on Digital Silicon Photomultipliers Performance

Effects of Dark Counts on Digital Silicon Photomultipliers Performance Effects of Dark Counts on Digital Silicon Photomultipliers Performance Radosław Marcinkowski, Samuel España, Roel Van Holen, Stefaan Vandenberghe Abstract Digital Silicon Photomultipliers (dsipm) are novel

More information

The digital Silicon Photomultiplier A novel Sensor for the Detection of Scintillation Light

The digital Silicon Photomultiplier A novel Sensor for the Detection of Scintillation Light The digital Silicon Photomultiplier A novel Sensor for the Detection of Scintillation Light Carsten Degenhardt, Gordian Prescher, Thomas Frach, Andreas Thon, Rik de Gruyter, Anja Schmitz, Rob Ballizany

More information

A Fast Waveform-Digitizing ASICbased DAQ for a Position & Time Sensing Large-Area Photo-Detector System

A Fast Waveform-Digitizing ASICbased DAQ for a Position & Time Sensing Large-Area Photo-Detector System A Fast Waveform-Digitizing ASICbased DAQ for a Position & Time Sensing Large-Area Photo-Detector System Eric Oberla on behalf of the LAPPD collaboration PHOTODET 2012 12-June-2012 Outline LAPPD overview:

More information

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE Takeyuki Hashimoto 1), Morio Onoe 2), Hiroshi Nakamura 3), Tamon Inouye 4), Hiromichi Jumonji 5), Iwao Takahashi 6); 1)Yokohama Soei

More information

arxiv: v1 [physics.ins-det] 12 Jul 2017

arxiv: v1 [physics.ins-det] 12 Jul 2017 Prepared for submission to JINST Multichannel FPGA based MVT system for high precision time (20 ps RMS) and charge measurement. arxiv:1707.03565v1 [physics.ins-det] 12 Jul 2017 M. Pałka a 1, P. Strzempek

More information

Total Absorption Dual Readout Calorimetry R&D

Total Absorption Dual Readout Calorimetry R&D Available online at www.sciencedirect.com Physics Procedia 37 (2012 ) 309 316 TIPP 2011 - Technology and Instrumentation for Particle Physics 2011 Total Absorption Dual Readout Calorimetry R&D B. Bilki

More information

PoS(PhotoDet 2012)022

PoS(PhotoDet 2012)022 SensL New Fast Timing Silicon Photomultiplier Kevin O`Neill 1 SensL Technologies Limited 6800 Airport Business Park, Cork, Ireland E-mail: koneill@sensl.com Nikolai Pavlov SensL Technologies Limited 6800

More information

Data Acquisition System for the Angra Project

Data Acquisition System for the Angra Project Angra Neutrino Project AngraNote 012-2009 (Draft) Data Acquisition System for the Angra Project H. P. Lima Jr, A. F. Barbosa, R. G. Gama Centro Brasileiro de Pesquisas Físicas - CBPF L. F. G. Gonzalez

More information

Silicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany.

Silicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany. KETEK GmbH Hofer Str. 3 81737 Munich Germany www.ketek.net info@ketek.net phone +49 89 673 467 70 fax +49 89 673 467 77 Silicon Photomultiplier Evaluation Kit Quick Start Guide Eval Kit Table of Contents

More information

Digital trigger system for the RED-100 detector based on the unit in VME standard

Digital trigger system for the RED-100 detector based on the unit in VME standard Journal of Physics: Conference Series PAPER OPEN ACCESS Digital trigger system for the RED-100 detector based on the unit in VME standard To cite this article: D Yu Akimov et al 2016 J. Phys.: Conf. Ser.

More information

A comparative study of the time performance between NINO and FlexToT ASICs

A comparative study of the time performance between NINO and FlexToT ASICs Journal of Instrumentation OPEN ACCESS A comparative study of the time performance between NINO and FlexToT ASICs To cite this article: I. Sarasola et al View the article online for updates and enhancements.

More information

A NOVEL FPGA-BASED DIGITAL APPROACH TO NEUTRON/ -RAY PULSE ACQUISITION AND DISCRIMINATION IN SCINTILLATORS

A NOVEL FPGA-BASED DIGITAL APPROACH TO NEUTRON/ -RAY PULSE ACQUISITION AND DISCRIMINATION IN SCINTILLATORS 10th ICALEPCS Int. Conf. on Accelerator & Large Expt. Physics Control Systems. Geneva, 10-14 Oct 2005, PO2.041-4 (2005) A NOVEL FPGA-BASED DIGITAL APPROACH TO NEUTRON/ -RAY PULSE ACQUISITION AND DISCRIMINATION

More information

CHAPTER 8 GENERIC PERFORMANCE MEASURES

CHAPTER 8 GENERIC PERFORMANCE MEASURES GENERIC PERFORMANCE MEASURES M.E. DAUBE-WITHERSPOON Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, United States of America 8.1. INTRINSIC AND EXTRINSIC MEASURES 8.1.1.

More information

A novel method based solely on FPGA units enabling measurement of time and charge of analog signals in Positron Emission Tomography

A novel method based solely on FPGA units enabling measurement of time and charge of analog signals in Positron Emission Tomography A novel method based solely on FPGA units enabling measurement of time and charge of analog signals in Positron Emission Tomography M. Pałka 1, T. Bednarski 1, P. Białas 1, E. Czerwiński 1, Ł. Kapłon 1,2,

More information

This article has been accepted for inclusion in a future issue of this journal. Content is final as presented, with the exception of pagination.

This article has been accepted for inclusion in a future issue of this journal. Content is final as presented, with the exception of pagination. IEEE TRANSACTIONS ON NUCLEAR SCIENCE 1 Waveform-Sampling Electronics for a Whole-Body Time-of-Flight PET Scanner W. J. Ashmanskas, B. C. LeGeyt, F. M. Newcomer, Member, IEEE, J. V. Panetta, W. A. Ryan,

More information

A high resolution FPGA based time-to-digital converter

A high resolution FPGA based time-to-digital converter A high resolution FPGA based time-to-digital converter Wei Wang, Yongmeng Dong, Jie Li, Hao Zhou, Pingbo Xiong, Zhenglin Yang School of Chongqing University of Posts and Telecommunications, Chongqing 465

More information

Initial evaluation of the Indiana small animal PET scanner

Initial evaluation of the Indiana small animal PET scanner Initial evaluation of the Indiana small animal PET scanner Ned C. Rouze, Member, IEEE, Victor C. Soon, John W. Young, Member, IEEE, Stefan Siegel, Member, IEEE, and Gary D. Hutchins, Member, IEEE Abstract

More information

Characterization of a 64 Channel PET Detector Using Photodiodes for Crystal Identification *

Characterization of a 64 Channel PET Detector Using Photodiodes for Crystal Identification * Characterization of a 64 Channel PET Detector Using Photodiodes for Crystal Identification * J. S. Huber, Member, IEEE, W.W. Moses, Senior Member, IEEE, S.E. Derenzo, Senior Member, IEEE, M.H. Ho, M.S.

More information

Positron Emission Tomography

Positron Emission Tomography Positron Emission Tomography UBC Physics & Astronomy / PHYS 409 1 Introduction Positron emission tomography (PET) is a non-invasive way to produce the functional 1 image of a patient. It works by injecting

More information

Development of front-end readout electronics for silicon strip. detectors

Development of front-end readout electronics for silicon strip. detectors Development of front-end readout electronics for silicon strip detectors QIAN Yi( 千奕 ) 1 SU Hong ( 苏弘 ) 1 KONG Jie( 孔洁 ) 1,2 DONG Cheng-Fu( 董成富 ) 1 MA Xiao-Li( 马晓莉 ) 1 LI Xiao-Gang ( 李小刚 ) 1 1 Institute

More information

Thomas Frach, Member, IEEE, Walter Ruetten, Member, IEEE, Klaus Fiedler, Gunnar Maehlum, Member, IEEE, Torsten Solf, and Andreas Thon

Thomas Frach, Member, IEEE, Walter Ruetten, Member, IEEE, Klaus Fiedler, Gunnar Maehlum, Member, IEEE, Torsten Solf, and Andreas Thon Assessment of Photodiodes as a Light Detector for PET Scanners Thomas Frach, Member, IEEE, Walter Ruetten, Member, IEEE, Klaus Fiedler, Gunnar Maehlum, Member, IEEE, Torsten Solf, and Andreas Thon Abstract

More information

Digital coincidence acquisition applied to portable β liquid scintillation counting device

Digital coincidence acquisition applied to portable β liquid scintillation counting device Nuclear Science and Techniques 24 (2013) 030401 Digital coincidence acquisition applied to portable β liquid scintillation counting device REN Zhongguo 1,2 HU Bitao 1 ZHAO Zhiping 2 LI Dongcang 1,* 1 School

More information

Time of Flight Measurement System using Time to Digital Converter (TDC7200)

Time of Flight Measurement System using Time to Digital Converter (TDC7200) Time of Flight Measurement System using Time to Digital Converter (TDC7200) Mehul J. Gosavi 1, Rushikesh L. Paropkari 1, Namrata S. Gaikwad 1, S. R Dugad 2, C. S. Garde 1, P.G. Gawande 1, R. A. Shukla

More information

Monte Carlo Simulation Study of a Dual-Plate PET Camera Dedicated to Breast Cancer Imaging

Monte Carlo Simulation Study of a Dual-Plate PET Camera Dedicated to Breast Cancer Imaging IEEE Nuclear Science Symposium Conference Record M-9 Monte Carlo Simulation Study of a Dual-Plate PET Camera Dedicated to Breast Cancer Imaging Jin Zhang, Member, IEEE, Peter D. Olcott, Member, IEEE, Angela

More information

60 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 1, FEBRUARY /$ IEEE

60 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 1, FEBRUARY /$ IEEE 60 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 1, FEBRUARY 2007 Prototype Parallel Readout System for Position Sensitive PMT Based Gamma Ray Imaging Systems Frezghi Habte, Member, IEEE, Peter D.

More information

Data Compression and Analysis Methods for High- Throughput Radiation Detector Systems

Data Compression and Analysis Methods for High- Throughput Radiation Detector Systems 1 Data Compression and Analysis Methods for High- Throughput Radiation Detector Systems John Mattingly Associate Professor, Nuclear Engineering North Carolina State University 2 Introduction The capabilities

More information

Performance evaluation of a new highsensitivity time-of-flight clinical PET/CT system

Performance evaluation of a new highsensitivity time-of-flight clinical PET/CT system Huo et al. EJNMMI Physics (2018) 5:29 https://doi.org/10.1186/s40658-018-0229-4 EJNMMI Physics ORIGINAL RESEARCH Open Access Performance evaluation of a new highsensitivity time-of-flight clinical PET/CT

More information

A Continuous Crystal Detector for TOF PET

A Continuous Crystal Detector for TOF PET 1 A Continuous Crystal Detector for TOF PET T. Szczęśniak, Member, IEEE, M. Moszyński, Fellow, IEEE, Ł. Świderski, Member, IEEE, A. Nassalski, Member, IEEE, A. Syntfeld-KaŜuch, Member, IEEE, P. Ojala,

More information

Hardware Trigger Processor for the MDT System

Hardware Trigger Processor for the MDT System University of Massachusetts Amherst E-mail: tcpaiva@cern.ch We are developing a low-latency hardware trigger processor for the Monitored Drift Tube system for the Muon Spectrometer of the ATLAS Experiment.

More information

DISCRETE crystal detector modules have traditionally been

DISCRETE crystal detector modules have traditionally been IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 5, OCTOBER 2006 2513 Performance Comparisons of Continuous Miniature Crystal Element (cmice) Detectors Tao Ling, Student Member, IEEE, Kisung Lee, and

More information

Multi-Channel Time Digitizing Systems

Multi-Channel Time Digitizing Systems 454 IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, VOL. 13, NO. 2, JUNE 2003 Multi-Channel Time Digitizing Systems Alex Kirichenko, Saad Sarwana, Deep Gupta, Irwin Rochwarger, and Oleg Mukhanov Abstract

More information

INDEX. Firmware for DPP (Digital Pulse Processing) DPP-PSD Digital Pulse Processing for Pulse Shape Discrimination

INDEX. Firmware for DPP (Digital Pulse Processing) DPP-PSD Digital Pulse Processing for Pulse Shape Discrimination Firmware for DPP (Digital Pulse Processing) Thanks to the powerful FPGAs available nowadays, it is possible to implement Digital Pulse Processing (DPP) algorithms directly on the acquisition boards and

More information

High granularity scintillating fiber trackers based on Silicon Photomultiplier

High granularity scintillating fiber trackers based on Silicon Photomultiplier High granularity scintillating fiber trackers based on Silicon Photomultiplier A. Papa Paul Scherrer Institut, Villigen, Switzerland E-mail: angela.papa@psi.ch Istituto Nazionale di Fisica Nucleare Sez.

More information

A low dead time vernier delay line TDC implemented in an actel flash-based FPGA

A low dead time vernier delay line TDC implemented in an actel flash-based FPGA Nuclear Science and Techniques 24 (2013) 040403 A low dead time vernier delay line TDC implemented in an actel flash-based FPGA QIN Xi 1,2 FENG Changqing 1,2,* ZHANG Deliang 1,2 ZHAO Lei 1,2 LIU Shubin

More information

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard.

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard. Chiara Secco PET Performance measurements of the new LSO-Based Whole Body PET/CT Scanner biograph 16 HI-REZ using the NEMA NU 2-2001 Standard. INTRODUCTION Since its introduction, CT has become a fundamental

More information

Combined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8

Combined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8 Combined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8 UNIVERSITY OF CAMBRIDGE 1.2 Rob C. Hawkes 1, Tim D. Fryer 1, Alun J. Lucas 1,2, Stefan B. Siegel 3, Richard

More information

FPGA-Based Pulse Pile-up Correction

FPGA-Based Pulse Pile-up Correction FPGA-Based Pulse Pile-up Correction M.D. Haselman 1, J. Pasko 1, S. Hauck 1, Senior Member IEEE, T.K. Lewellen 2, Fellow IEEE, R.S. Miyaoka 2, Member IEEE, 1 University of Washington Department of Electrical

More information

Tutors Dominik Dannheim, Thibault Frisson (CERN, Geneva, Switzerland)

Tutors Dominik Dannheim, Thibault Frisson (CERN, Geneva, Switzerland) Danube School on Instrumentation in Elementary Particle & Nuclear Physics University of Novi Sad, Serbia, September 8 th 13 th, 2014 Lab Experiment: Characterization of Silicon Photomultipliers Dominik

More information

The PennPET Explorer Scanner for Total Body Applications

The PennPET Explorer Scanner for Total Body Applications The PennPET Explorer Scanner for Total Body Applications JS Karp, MJ Geagan, G Muehllehner, ME Werner, T McDermott, JP Schmall, V Viswanath, University of Pennsylvania, Philadelphia, PA AE Perkins, C-H

More information

FPGA BASED DATA AQUISITION SYSTEMS FOR PHYSICS EXPERIMENTS

FPGA BASED DATA AQUISITION SYSTEMS FOR PHYSICS EXPERIMENTS INTERNATIONAL PHD PROJECTS IN APPLIED NUCLEAR PHYSICS AND INNOVATIVE TECHNOLOGIES This project is supported by the Foundation for Polish Science MPD program, co-financed by the European Union within the

More information

Study of a Prototype VP-PET Imaging System Based on highly. Pixelated CdZnTe Detectors

Study of a Prototype VP-PET Imaging System Based on highly. Pixelated CdZnTe Detectors Study of a Prototype VP-PET Imaging System Based on highly Pixelated CdZnTe Detectors Zheng-Qian Ye 1, Ying-Guo Li 1, Tian-Quan Wang 1, Ya-Ming Fan 1, Yong-Zhi Yin 1,*, Xi-Meng Chen 1 Affiliations: 1 School

More information

Hardware Trigger Processor for the MDT System

Hardware Trigger Processor for the MDT System University of Massachusetts Amherst E-mail: tcpaiva@cern.ch We are developing a low-latency hardware trigger processor for the Monitored Drift Tube system in the Muon spectrometer. The processor will fit

More information

Real Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor

Real Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor Real Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor Paul A. B. Scoullar a, Chris C. McLean a and Rob J. Evans b a Southern Innovation, Melbourne, Australia b Department of Electrical

More information

Cross-Strip Multiplexed Electro-Optical Coupled Scintillation Detector for Integrated PET/MRI

Cross-Strip Multiplexed Electro-Optical Coupled Scintillation Detector for Integrated PET/MRI IEEE TRANSACTIONS ON NUCLEAR SCIENCE 1 Cross-Strip Multiplexed Electro-Optical Coupled Scintillation Detector for Integrated PET/MRI Peter D. Olcott, Member, IEEE, GaryGlover, Member, IEEE, and CraigS.Levin,

More information

Contents. Why waveform? Waveform digitizer : Domino Ring Sampler CEX Beam test autumn 04. Summary

Contents. Why waveform? Waveform digitizer : Domino Ring Sampler CEX Beam test autumn 04. Summary Contents Why waveform? Waveform digitizer : Domino Ring Sampler CEX Beam test data @PSI autumn 04 Templates and time resolution Pulse Shape Discrimination Pile-up rejection Summary 2 In the MEG experiment

More information

An ASIC dedicated to the RPCs front-end. of the dimuon arm trigger in the ALICE experiment.

An ASIC dedicated to the RPCs front-end. of the dimuon arm trigger in the ALICE experiment. An ASIC dedicated to the RPCs front-end of the dimuon arm trigger in the ALICE experiment. L. Royer, G. Bohner, J. Lecoq for the ALICE collaboration Laboratoire de Physique Corpusculaire de Clermont-Ferrand

More information

Reconstruction Filtering in Industrial gamma-ray CT Application

Reconstruction Filtering in Industrial gamma-ray CT Application Reconstruction Filtering in Industrial gamma-ray CT Application Lakshminarayana Yenumula *, Rajesh V Acharya, Umesh Kumar, and Ashutosh Dash Industrial Tomography and Instrumentation Section, Isotope Production

More information

LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging

LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging 8th International Workshop on Radiation Imaging Detectors Pisa 2-6 July 2006 LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging Roberto Pani On behalf of SCINTIRAD Collaboration

More information

FPGA-Based Data Acquisition System for a Positron Emission Tomography (PET) Scanner

FPGA-Based Data Acquisition System for a Positron Emission Tomography (PET) Scanner FPGA-Based Data Acquisition System for a Positron Emission Tomography (PET) Scanner Michael Haselman 1, Robert Miyaoka 2, Thomas K. Lewellen 2, Scott Hauck 1 1 Department of Electrical Engineering, 2 Department

More information

Real-Time Digital Signal Processors with radiation detectors produced by TechnoAP

Real-Time Digital Signal Processors with radiation detectors produced by TechnoAP Real-Time Digital Signal Processors with radiation detectors produced by TechnoAP Lunch time Exhibitor presentation 2976-15 Mawatari, Hitachinaka-city, Ibaraki 312-0012, Japan Phone: +81-29-350-8011, FAX:

More information