MRI MRI REGISTRY REVIEW PHYSICAL PRINCIPLES OF IMAGE FORMATION ARTIFACTS SUPERCONDUCTIVE MAGNET ANAIBI MOLINA(R) (RT) (MR) (CT) T2 DEPHASING

Size: px
Start display at page:

Download "MRI MRI REGISTRY REVIEW PHYSICAL PRINCIPLES OF IMAGE FORMATION ARTIFACTS SUPERCONDUCTIVE MAGNET ANAIBI MOLINA(R) (RT) (MR) (CT) T2 DEPHASING"

Transcription

1 MRI ANAIBI MOLINA(R) (RT) (MR) (CT) T2 DEPHASING SUPERCONDUCTIVE MAGNET FREE INDUCTION DECAY ARTIFACTS MRI REGISTRY REVIEW PHYSICAL PRINCIPLES OF IMAGE FORMATION

2 Mri Registry Review Physical Principles of Image Formation Developmental Editor: Jimmy Vargas Graphics Project Editor: Deborah DeNicola Marielis Abreu Art/Design Coordinators: Jimmy Vargas Graphics Koko Designs Taylor Graphics Printer Services: Top Drawer Media Solutions Copyright 2016 MRI Registry Review 2016, Advanced Imaging Education. All rights reserved. No part of this information may be reproduced or used in any form in written, electronic, graphic and or mechanical including photocopying, taping, recording and or Web distribution without the written permission from the author Anaibi Molina. Disclaimer: The information of this book does not warrant or guarantee passing the MRI Examination. The student should gather the information necessary that he or she needs and review the material as necessary in order to pass the MRI registry examination. The information of this MRI Registry Review should not be considered as a medical advice. Each facility and medical entity has its own regulations and protocols. Please follow those regulations and protocols. Advanced Imaging Education disclaims any liability for the acts of the technologist, individuals, and or Physicians who receives any information from this educational material. 2

3 MRI REGISTRY REVIEW ANAIBI MOLINA(R) (RT) (MR) (CT) ALL RIGHT RESERVED. No part of this study guide is to be reproduced by any means of copying, recording and/or transmitted electronically without the permission from the author Anaibi Molina. 3

4 Physical Principles of Image Formation CONTENT CATEGORY Physical Principles of Image Formation ( PART III ) SECTION 1 ELECTROMAGNETISM SECTION 2 RADIOFREQUENCY SYSTEM SECTION 3 SECTION 4 SECTION 5 SECTION 6 GRADIENT SYSTEM NUCLEAR MAGNETISM TISSUE CHARACTERISTICS SPATIAL LOCALIZATION SECTION 7 SECTION 8 ARTIFACTS QUALITY CONTROL The ARRT Content Specifications for the Magnetic Resonance Imaging Examination are reprinted by permission of the ARRT. The ARRT Content Specifications for the Magnetic Resonance Imaging Examination are copyrighted by the ARRT. Notice: Some changes from the author have been made from the ARRT Content Specifications for the Magnetic Resonance Imaging Examination. 4

5 Contents Section 1 ELECTROMAGNETISM 6 A. FARADAY'S LAW... 6 B. TYPES OF MAGNETS... 6 C. MAGNETIC FIELD STRENGTH... 9 Magnetic fringe fields Section 2 RADIOFREQUENCY SYSTEM D. COIL CONFIGURATION General positioning E. TRANSMIT AND RECEIVE COILS General coil characteristics E.1 TRANSMIT AND RECEIVE BANDWIDTH F. PULSE PROFILE G. PHASE ARRAY Section 3 GRADIENT SYSTEM H. COIL CONFIGURATION Gradient strength or amplitude I. SLEW RATE J. RISE TIME K. DUTY CYCLE Section 4 NUCLEAR MAGNETISM 23 L. LARMOR EQUATION M. PRECESSION FREQUENCY N. GYROMAGNETIC RATIO O. RESONANCE P. RF PULSE Q. EQUILIBRIUM MAGNETIZATION R. ENERGY STATE TRANSITIONS...25 S.PHASE CONHERENCE...26 T. FREE INDUCTION DECAY Section 6 SPATIAL LOCALIZATION 30 X. VECTORS X, Y AND Z COORDINATES Y. PHYSICAL GRADIENT Slice Select, Phase Encoding and Frequency Encoding Z. K-SPACE(RAW DATA) Section 7 ARTIFACTS 36 A. CAUSE AND APPEARANCE Section 8 QUALITY CONTROL 59 A. SLICE THICKNESS B. SPATIAL RESOLUTION High Contrast Spatial Resolution C. CONTRAST RESOLUTION Low contrast object detectability D. SIGNAL TO NOISE RATIO E. CENTER FREQUENCY...// F. TRANSMIT GAIN OR ATTENUATION G. GEOMETRIC ACCURACY Quality Assurance Section 9 EQUIPMENT HANDLING 66 AND INSPECTION SYSTEM INDICATOR LIGHTS... H. 66 SAFETY LIGHTS AND LIGHT BULBS... I. 66 J. RF ROOM INTEGRITY AND DOOR SEALS... K. EMERGENCY CART L. COILS AND CABLES Section 5 TISSUE CHARACTERISTICS 27 U. T1, T2, T2*, PROTON DENSITY Indications V. FLOW W. DIFFUSION AND PERFUSION Indications 5

6 ne Section ELECTROMAGNETISM Electromagnetism is one of the fundamentals forces in nature. It deals with the physical relations between electricity and magnetism. A. FARADAY'S LAW The law of electromagnetic induction states that when a conductor is placed in a magnetic field; an electrical voltage will be induced in the conductor. Also when an electrical voltage is traveling along a conductor, a magnetic field will be produced around it. The higher the strength of the magnetic field, the greater the current flowing in the conductor. B. TYPES OF MAGNETS Divided into: Superconductive, Permanent and Resistive Superconductive Magnet (Electromagnet) Most common types of magnet. High field systems which mainly used in magnetic fields of 1.5 Tesla or higher. 85% of clinical MRI scanners are 1.5 Tesla. SUPERCONDUCTIVE MAGNET 6

7 Characteristics of superconductive magnets Closed field system Horizontal magnetic field Solenoid configuration= A coil is wound into a tightly packed helix. The superconductive magnet is a cylinder that has 55 to 70cm in diameter in order to accommodate a human The magnetic field is produced by the flow of current induced in the coil. Ohm's law governs the resistance of current along a wire which is directly proportional to the potential difference across two points. Ohm's law formula V (voltage) = I (current) x R (resistance within the wire) Superconductivity is a property of some materials in which there is zero electrical resistance at very low temperatures. Niobium-titanium alloy is used to make the coil windings for superconducting magnets. The coil windings are surrounded by liquid helium that when cooled below 4 K (kelvin) creates virtually zero current resistance and the magnetic field is maintained. Cryogens= Super-cooled substances such as liquid helium (most commonly used) or liquid nitrogen that remove current resistance in the coil windings. This creates superconductivity. Cryostat=A steel cylindrical tank which maintains the low temperature of the liquid helium and or nitrogen and does not allow it to boil off. 7 CRYOSTAT

8 Permanent magnets Low to mid field strength between 0.3 to 0.5 Tesla but recently there are a number of high field systems of 1.0 Tesla. PERMANENT MAGNET Characteristics Open field systems Vertical magnetic field Heavy weight Do not required electricity to power up the magnet Sensitive to temperature Small fringe field Fewer safety considerations with respect to the fringe field (magnetic field surrounding the magnet) which reduces the chance of projectiles. The use of ferromagnetic materials in permanent magnets such as iron, cobalt or steel will retain its magnetism after being exposed to the magnetic field. Ceramic bricks consisting of magnetic blocks of slabs which possess ferromagnetic properties that can be magnetized can also be used to produce permanent magnets. An option to use for claustrophobic patients 8

9 Resistive Magnet(Electromagnet) Low field strength less than 0.2 TO 0.3Tesla RESISTIVE MAGNET Characteristics Open field systems In order to maintain the magnetic field; a continuous flow of current must always be available. Magnetic field can be turn off horizontal or vertical magnetic field Light in weight High operating cost to maintain it High power consumption in order to keep the current in order to maintain the magnetic field. C. MAGNETIC FIELD STRENGTH Magnetic field strength is measured in tesla (T) or gauss (G) The Tesla unit is used to measure high field strength (1 tesla =10,000 gauss). The Gauss unit= is used to measures low field strength such as the fringe field (magnetic fields outside the bore of the magnet). Classified as: Low field strength Less than 0.3 Tesla Mid field strength Between 0.3 to 1.0 Tesla High field strength 1.5 Tesla to 3 Tesla 9 Ultra high field strength 14 Tesla used for research, spectroscopy and high resolution studies

10 Question 1. If 1 Tesla is equal to 10,000 Gauss, How many Gauss are in 0.3T? P Formula 1 Tesla 10,000 Gauss 0.3Tesla x Problem Solving 0.3Tesla x 10,000 Gauss=3000/1 Tesla=3000 Gauss Answer 3,000 gauss 3,000 Magnetic fringe fields The magnetic field outside of the magnet bore. 5 GAUSS LINE FRINGE FIELDS ISO CENTER FRINGE FIELD The magnetic field outside the bore of the magnet has a 5 gauss limit which represents the safety margin level for exposure to the magnetic fringe fields in patients that have a pacemaker. The use of magnetic shielding is used to reduce the area of the fringe field. Homogeneity The uniformity of the main magnetic field which is measured in parts per million (ppm). 10

11 wo Section.. RADIOFREQUENCY SYSTEM D. COIL CONFIGURATION The radiofrequency coil is the first inner layer component within the gantry. Commonly referred to as the body coil is capable of transmitting and receiving radiofrequencies.. E. TRANSMIT AND RECEIVE COILS Transmit coils= capable of transmitting the signal of an RF pulse. The primary RF transmitter is the body coil which is also capable of receiving the RF pulse. Most of the transmit coils are also capable of receiving the signal as well. MAIN BODY COIL Some other examples include head, knee and foot coils. Receive coils= only capable of receiving the signal of an RF pulse. The received coils send the signal to the MRI magnet in order for the computer to process the signal. These coils are placed on or around the surface of the area of interest. 11 Example: Surface (local) coils, Linear and or Phase array coils which are also referred as multichannel coils.

12 Surface (local) coils are placed directly on the surface of the area of interest which is usually small in size. These coils provide an excellent SNR but are only sensitive to tissue depths of 50 to 75% the coil's diameter. Example: ENDORECTAL SURFACE (LOCAL) COIL TEMPOROMANDIBULAR JOINT SURFACE (LOCAL) COIL Linear coils consist of a single loop of wire and provide a good signal since a small area of noise is detected. Example: The Helmholtz coil configuration which uses multiple coils with one single receiver. It consists of two circular coils parallel to each other. SENSE FLEX MEDIUM HELMHOLTZ COIL THAT CAN BE USED FOR SHOULDER AND OR WRIST 12

13 Another Helmholtz coil configuration can be used to obtain signal for the anterior neck and cervical spine. Example Phase Array Coils uses multiple coils with multiple receivers whose signal is combined to create images with an increased in SNR and yet covers a large anatomical area. It is sometimes used in combination with parallel imaging techniques in order to decrease the scan time. Example: brain, vascular neck, abdomen, pelvis and spine coil 8 CHANNEL PHASE ARRAY BRAIN COIL 8 CHANNEL PHASE ARRAY VASCULAR HEAD AND NECK COIL 13

14 8 CHANNEL PHASE ARRAY ABDOMEN AND PELVIS COIL 8 CHANNEL PHASE ARRAY SPINE COIL Trans-receiver coils= capable of transmitting and receiving the signal of an RF pulse. Notes 14

15 Example: Knee and foot quadrature coils, volume coils such as head and the body coil (the main magnet). The quadrature coil configura on uses wires that are perpendicular to one another. QUADRATURE KNEE AND FOOT TRANSMIT-RECEIVE COIL Volume coils can be used to accommodate a large volume of tissue but this causes a decrease in SNR. Example 1: HEAD (BIRDCAGE) TRANSMIT-RECEIVE COIL 15

16 HEAD TRANSMITCOIL HEAD-RECEIVE COIL MAIN BODY TRANSMIT-RECEIVE COIL General coil characteristics Use the smallest coil possible that fits the area of interest to be imaged. The correct coil that fits the area of interest should be used in order to have an increased in SNR. The closer the coil is to the area of interest, the greatest the signal detection. In general; the smaller the radiofrequency coil; the better the signal to noise ratio. In order to get the appropriate signal of the area of interest being imaged, the coil must be positioned perpendicular to the Bo field. 16

17 Example: COIL POSITIONING WAS PERPENDICULAR TO THE B0 FIELD COIL POSITIONING WAS OBLIQUE TO THE B0 FIELD COIL POSITIONING WAS PARALLEL TO THE B0 FIELD COIL POSITIONING WAS PERPENDICULAR TO THE B0 FIELD E.1 TRANSMIT AND RECEIVE BANDWIDTH Transmit Bandwidth Transmit bandwidth is used to define slice thickness. Once a certain gradient slope is applied, the transmit bandwidth will provide a range of radiofrequencies (RFs), centering about the Larmor frequency, transmitted to excite the slice, which should match the difference in precessional frequency between two points. 17

18 A narrow transmit bandwidth and/or steep slice select gradient is used to achieve thin slices. A broad transmit bandwidth and/or shallow slice select gradient is used to achieve thick slices. The resonant frequency bandwidth of the hydrogen protons to be visualized is transmitted in order to get resonance. A broad transmit bandwidth will give rise to: A shallow slice select gradient Low amplitude slice select gradient Increase slice thickness Decrease resolution Receive Bandwidth The range of frequencies the receiver can sample accurately, which must be mapped across the FOV. This is determined by the upper and lower limits of frequencies on the either side of the center frequency of the echo. The receive bandwidth is related to the slope of the frequency encoding gradient. Based on the Nyquist theorem Receive bandwidth = 2 x the highest frequency (Nyquist Frequency) An increase in receiver bandwidth will give rise to: Notes Faster sampling time Low minimum TE decrease chemical shift artifacts Decrease SNR 18

19 Noise Signal Noise Signal Noise Noise Broad Bandwidth Narrow Bandwidth 32 KHz bandwidth 12 KHz bandwidth Decrease Signal Increase Noise Increase Signal Decrease Noise F. PULSE PROFILE A waveform representation of the RF pulses that are received by the MR system. The center of the waveform is sampled during readout. The outer edges of the RF pulse waveform are not sampled but are still present. A gap of 30% of the slice thickness is used in order to prevent the cross excitation artifact. G. PHASE ARRAY A coil that uses from 4 to 32 receivers in order to increase SNR and yet coverage a larger area of anatomy. Please refer to the phase array coil section on page for more information. 19

20 Three Section. GRADIENT SYSTEM GRADIENTS Small electromagnets that are superimposed over the main magnetic field which uses wires of current to alter the external magnetic field. Gradients are the main source of noise in an MR system. 20

21 The banging noise that is experience when performing an MRI sequence occurs when rapid pulses of electricity are passed through the gradient coils which cause alterations in the external magnetic field. Gradient magnetic fields are used to spatially encode data in the slice, phase and frequency direction. H. COIL CONFIGURATION The order of the coils in a magnet from the inner to the outer layer of the magnet bore is as follows: The radiofrequency coil, the gradient coils, the shim coils and then the magnet coils. MAGNET SHIM COILS GRADIENT COILS RF COILS 21

22 Question nd 1. What coil represents the 2 layer component going from the inner to outer part of the magnet bore? A. the shim coil B. the radiofrequency coil C. the main body coil D. the gradient coil Answer D. the gradient coil Gradient strength or amplitude Defines how strong or steep a particular gradient is. It is measured in MilliTesla/ per meter (mt/m) or gauss per centimeter (G/cm). Typical maximum gradient strengths values range from 10-40mT/m. I. SLEW RATE The time it takes for a gradient to achieve maximum amplitude. It describes the rate of change in the speed and strength of the gradient amplitude and is measured in T/m/s. Slew rates can range between Mt/m/s. It can be calculated by dividing the maximum gradient amplitude by the rise time. Slew rate is the best indicator of gradient performance. J. RISE TIME (SPEED) The time it takes for a gradient to reach maximum amplitude. It is measured in microseconds. The shorter the rise time; the faster the gradients and therefore the smaller the echo spacing. Shorter echo spacing will give the advantage of increase resolution and more slices allowed/tr. K. DUTY CYCLE The time that the gradient is able to work at maximum amplitude. How long a gradient can work at its maximum strength and speed. It is measured in percentage (%). 22

23 Four Section. NUCLEAR MAGNETISM L. LARMOR EQUATION A mathematical equation which determines the frequency that has to be used in order to create resonance. It provides the precessional frequency that is needed in order to excite the specific spins that are required in order to obtain images of a specific area. The larmor equation describes the relationship between the static magnetic field and the gyromagnetic ratio of the hydrogen protons in specific tissues. In order to obtain the precessional frequency; the Larmor equation has to be applied. Larmor Equation The strength of the static magnetic field is multiply by a constant called gyromagnetic ratio. Wo= Bo x Y M. PRECESSION FREQUENCY It describes the wobbling motion of protons that spin around their axis in the presence of an external magnetic field. In order to obtain an MR image of a specific area; an RF pulse at the precessional frequency of the hydrogen protons to be imaged must be applied in order to get resonance. The precessional frequency of the protons in a magnetic field can be provided by using the Larmor equation. The larmor equation: Wo= Bo x Y Wo represents the precessional frequency (in MHz) Bo represents the strength of the static magnetic field (in Testla) Y represents the gyromagnetic ratio. Mhz/T 23 The precessional frequency of hydrogen at 1.5 is Tesla is MHz.

24 QUESTION 2. If the precessional frequency of the hydrogen proton is approximately 25 MHz, the magnetic field experienced by the patient's protons is. PROBLEM SOLVING 63.86Mhz 25Mhz 1.5Tesla x 25 MHz x 1.5Tesla=37.5/63.86 MHz=0.6Tesla. ANSWER 0.6Tesla QUESTION 3. In a 3.0 Tesla magnet, what is the precessional frequency of the hydrogen protons? PROBLEM SOLVING The gyromagnetic ratio of hydrogen at 1.0 tesla is 42.6 MHz. We just have to apply the Larmor equation (precessional frequency): Multiply the gyromagnetic ratio of the particular proton by the static magnetic field in order to find the precessional frequency. 42.6MHz x 1.0 Tesla =128 MHZ 3.0 Tesla ANSWER 128 MHZ N. GYROMAGNETIC RATIO The rate that a spin will precess when exposed to a magnetic field. This is a constant that defines the precessional frequency of an element at 1 Tesla. This gyromagnetic ratio is specific for each atom. It represents the ratio of the magnetic moment vs. the angular momentum of a particle. Different MR active nuclei have its unique gyromagnetic ratio which is specific for each atom. The hydrogen atom gyromagnetic ratio is equal to 42.6 MHz/T. 24

25 O. RESONANCE Resonance occurs when a nucleus is exposed to an oscillation that has a frequency close to its own and absorbs the energy from that force. This occurs by an RF pulse being transmitted at the precessional frequency of that particular atom in which its spins will flip into the transverse plane creating phase coherence (a signal induction). In order for energy transfer to occur; the system must resonate at the correct resonant frequency. Depending on the field strength, the higher the field strength, the faster and higher the resonant frequency needed. P. RF PULSE A pulse which is transmitted at exactly the resonant frequency of the hydrogen protons needed in order to obtain images of a specific location. This will tip the hydrogen protons into the transverse plane in order to obtain signal. The greater the amplitude and the longer the pulse is applied, the greater the flip angle. The RF field is oriented perpendicular to the main magnetic field in order to obtain signal. The RF energy used in MRI is an electromagnetic radiation that is non-ionizing. Q. EQUILIBRIUM MAGNETIZATION Thermal equilibrium is a condition in which more spins are aligned parallel than antiparallel to the main magnetic field. During thermal equilibrium, the vector that represents the excess spins aligned in the longitudinal plane is called net magnetization vector. Net magnetization vector is the sum of the contributions of all the magnetic moments of the individual protons. The magnetic field associated with the protons is known as the magnetic moment. R. ENERGY STATE TRANSITIONS When the hydrogen protons are placed in a magnetic field, the majority of them will align parallel in a lower energy state. Some of them but fewer will also be aligned antiparallel in a higher energy state. 25

26 Only the protons in the parallel energy state will add up to the net magnetization vector and are used to provide the MR signal S. PHASE COHERENCE It occurs when an RF pulse is transmitted at the precessional frequency of hydrogens protons for a specific area to be imaged. These spins tipped into the transverse plane and starts precessing along the same rotational path in phase. T. FREE INDUCTION DECAY(FID) After the RF pulse is turn off; the signal from the hydrogen protons will decay immediately and become out of plane which causes longitudinal magnetization to increase and out of phase which causes transverse magnetization to decreases. This causes a reduction of voltage induced in the receiver coil. The FID is a combination of true T2 decay which is produced by out of phase hydrogen protons and T2* decay dephasing which is produced by magnetic field inhomogeneities and spin-spin relaxation. Relative signal amplitude T2* decay T2 decay Time FID 26

27 Five Section. TISSUE CHARACTERISTICS During excitation The net magnetization of hydrogen protons is tipped into the transverse plane in which low energy spins enter the high energy state and begins to precess in-phase. During Relaxation The net magnetization of hydrogen protons is returning into the longitudinal plane in which high energy spins return to the low energy state and begin to precess out of phase and out of plane (loss of phase coherence). T. T1, T2, T2* AND PROTON DENSITY T1 Relaxation The process by which the removal of the RF pulse causes spins to relax back into the longitudinal plane. Also known as T1 RECOVERY (spin-lattice relaxation) (longitudinal relaxation) which is the time it takes for 63% of the tissue magnetization to recover in the longitudinal plane. Depending on the type of tissue, for example; fat will relax more quickly due to its molecular tumbling rate. Fat has a slow molecular tumbling rate and its molecules relax faster. Water has a fast molecular tumbling rate and its molecules relax slower. Notes 27

28 T2 relaxation The process by which the interactions between the magnetic fields of adjacent spins causes dephasing. Also known T2 DECAY (spin-spin relaxation) (transverse relaxation) which is the time it takes for 63% of a tissue magnetization to be lost in the transverse plane. (37% decay). Depending on the type of tissue, for example; fat will relax quicker due to its molecular tumbling rate. Fat has a slow molecular tumbling rate and its molecules relax faster. Water has a fast molecular tumbling rate and its molecules relax slower. T2* susceptibility The dephasing of the precessing spins due to magnetic field in homogeneities. AXIAL SUCEPTIBILITY WEIGHTED VENOGRAM OF THE BRAIN Proton (spin) Density The number of mobile hydrogen protons/unit volume of tissue. Each tissue has a different amount of hydrogen protons. Example: Brain tissue has a high number of mobile hydrogen protons while cortical bone has a low number of mobile hydrogen protons. 28

29 AXIAL PROTON DENSITY OF THE BRAIN U. FLOW Please refer to flow dynamics in section D. Data acquisition and processing. W. DIFFUSION AND PERFUSION DIFFUSION A functional technique that assesses the movement of molecules in the extracellular space due to random thermal motion. PERFUSION The technique that to assesses the microcirculation of blood in the tissues.. 29

30 Six Section. SPATIAL LOCALIZATION X. VECTORS Any quantity that has both direction and current and is usually represented by an arrow. The magnetic moment MR active nuclei that have a charge and are spinning acquires a magnetic property and can align with an external magnetic field. The net magnetization vector Produced as a result of the excess hydrogen protons aligned in the parallel direction to the external magnetic field. The NMV increases when the magnetic field strength is increased and results in an improved SNR Notes 30

31 X, Y, and Z coordinate system It is used in order to spatially localize the signal in 3 spatial dimensions. XY plane Referred to as the transverse plane where we obtain the signal. Closed magnets have a vertical XY plane while open magnets have a horizontal XY plane. Y CLOSED MAGNET Z Bo X OPEN MAGNET Y Z X Bo Z axis Referred to as the longitudinal plane. It is oriented in the same direction as the external magnetic field. 31 Closed magnets have a horizontal Z axis while open magnets have a vertical Z axis.

32 Isocenter Referred to as the area where the X Y and Z axes intercept. The anatomy of interest to be scanned is placed as close as possible in the isocenter as this area provides the most magnetic field homogeneity. X= Transverse plane(signal detection) Y= B1 field (radiofrequency field) Z= Longitudinal plane-bo (external magnetic field) Notes 32

33 X. PHYSICAL GRADIENT The actual pieces of hardware that makes up the gradients. Composed of the X, Y and Z gradients. The gradient system is composed of 3 coordinates in order to spatially locate the anatomical area of interest. SLICE SELECT GRADIENT (Z axis) Location of the anatomical slice plane. PHASE ENCODING GRADIENT (Y axis) Location of the signal in the short axis of the anatomy. FREQUENCY ENCODING GRADIENT (X axis) Location of the signal in the long axis of the anatomy. Slice Select Gradient The slice select gradient is turn on during the application of the RF pulse. It locates the signal and location of the particular slice. The slope of the slice select gradient will determine the slice thickness and gap. The slice select gradient determines the anatomical plane that is acquired. It is divided into: X, Y and Z gradient 33

34 X GRADIENT ALTERS THE MAGNETIC FIELD STRENGTH AND PRECESSIONAL FREQUENCY OF THE HYDROGEN SPINS. SELECTS SAGITTAL SLICES Y GRADIENT ALTERS THE MAGNETIC FIELD STRENGTH AND PRECESSIONAL FREQUENCY OF THE HYDROGEN SPINS. SELECTS CORONAL SLICES Z GRADIENT ALTERS THE MAGNETIC FIELD STRENGTH AND PRECESSIONAL FREQUENCY OF HYDROGEN SPINS. SELECTS AXIAL SLICES EXAMPLE: Y gradient Z gradient bore table X gradient At least two gradients need to be used for oblique imaging. 34

35 Phase Encoding Gradient The phase encoding gradient is switched on after the RF pulse. This gradient varies with amplitude with each TR. It locates the signal along the short-axis of anatomy. This is because motion artifacts occur along this gradient. Example: The X gradient is used as the phase encoding gradient for the acquisition of an axial slice of the head because it uses the short axis of the anatomy. This results in a decrease in scan time. The Y gradient is used as the phase encoding direction for the acquisition of an axial slice of the pelvis because it uses the short axis of the anatomy. This results in a decrease in scan time. Note: The slope of the phase encoding gradient also determines which line of K-space is filled. Example: The application of a shallow phase encoding gradient (high signal amplitude) will fill the center lines of K-space first and this will store the contrast information which contains the most SNR. The application of a steep phase encoding gradient (low signal amplitude) will fill the outer lines of K-space last and this will store the spatial resolution information. Frequency encoding gradient (readout gradient) It is turn on during the readout of the echo. It is used to collect the echoes. It locates the signal along the long-axis of anatomy. Logical gradients The computer software assigns letters to each physical gradient. Y. K-SPACE (RAW DATA) The spatial frequency domain that serves as a temporary storage space for data collection during image acquisition. K-space is composed of phase and frequency encoding steps. 35

36 even Section. ARTIFACTS An anomaly seen on an image not normally present that can cause distortion and obscure the anatomy being visualized. It can cause a confusing artifact appearance that may mimic pathology which can be misdiagnosed. Technologists must learn how to recognized artifacts and known what can be done to reduce or eliminate them. A. Cause and Appearance of artifacts 1. ALIASING (BACKFOLDING, PHASE WRAP, WRAP-AROUND) Cause When the FOV acquired is smaller than the area of interest that needs to be visualized; the tissue outside the selected FOV is undersampled. Appearance The tissue outside the selected FOV is undersampled along the phase axis which causes the anatomy outside the FOV to be mapped within the FOV. BRAIN ALIASING Occurs in the phase encoding direction. 36

37 ABDOMEN ALIASING Solution Use a larger FOV that covers the entire area of interest in the phase direction. Apply pre-saturation pulses to undesirable anatomy. Use no phase wrap (Fold-over suppression, Phase oversampling or anti-fold over). Notes 37

38 . Gibbs(Truncation) Artifact Cause Under-sampling of data. Appearance The interfaces between high and low signal intensity are incorrectly represented. Occurs in the phase encoding direction. TRUNCATION ARTIFACT (Black or white parallel lines appear adjacent to high contrast interfaces) Notes 38

39 It can also appear as lines seen along the inside of the spinal cord in sagittal spine imaging which can mimic a syrinx. Solution Increase # of phase encoding steps Increase the NSA (number of acquisitions) Chemical Shift(Type 1) Cause Occurs due to the slight difference in resonant frequency between hydrogen molecules residing in water compared to the resonant frequency of hydrogen molecules residing in fat. The size of the chemical shift artifact can be obtained as follows: A total receiver bandwidth is 20 khz with 256 pixels in the frequency encoding axis; the bandwidth/pixel can be calculated as 20,000/256=78Hz. 39

40 Since the fat and water frequency difference at 1.5Tesla is 220Hz; the size of the chemical shift artifact can be calculated as: (220Hz) / (78Hz/pixel) =2.8 pixels. The degree of chemical shift between fat and water depends on magnetic field strength, receive bandwidth and pixel size. To calculate the water-fat frequency difference; you multiply the larmor frequency at the specific field strength you are using x 3.5ppm. Example: In a 1.5T MRI scanner that operates at 64 MHz, the frequency difference between fat and water is calculated as follows. 6 6 Δf = (64 MHz)(3.5 ppm) = (64 x 10 Hz)(3.5 x10 ) 220 Hz Magnetic field strength Frequency difference between fat and water 1 Tesla 147 Hz 1.5 Tesla 220 Hz 3 Tesla 440 Hz Factors that increase the Chemical Shift(Type 1) Artifact: Increase magnetic field strength Decrease receive bandwidth Increase Pixel size Increase voxel volume 40

41 Appearance Black or white band seen at the edge of the interface between fat and water. Occurs in the frequency encoding axis. Example 1: Example 2: Black lines seen at the vertebral end plates SAG T1 41

42 Solution Decrease the magnetic field strength Increase the Receive bandwidth for a smaller water/fat shift. Decrease pixel size Decrease voxel volume Increase matrix Using a fat suppression technique such as a fat spectral suppression RF pulse Use a STIR sequence Chemical Shift Misregistration (out of phase artifact) (phase cancellation artifact) (Type 2) Cause The inability of the gradient in gradient echo sequences to rephase the spins efficiently. When fat and water are in phase within the same voxel; the signals will add up. However when fat and water are out of phase within the same voxel; the signals will cancel each other allowing chemical misregistration to take place and this produces a signal void. Selecting a TE that is approximately ½ the in-phase TE can generate out of phase images where the signal from fat is suppressed. Appearance Black boundary artifact Can appear in the phase or frequency axis. SIGNAL VOID AX GRE AX GRE 42 SIGNAL VOID

43 Ring of dark signal seen around certain organs where water and fat resides within the same voxel. SIGNAL VOID AX OUT OF PHASE Solution The use of a spin echo sequence Applying TE intervals of 4.4ms in which fat and water and in-phase with each other. Using a fat suppression technique such as fat spectral suppression RF pulse Using a STIR sequence The in and out of phase process is a function of echo time (TE). When a voxel is imaged with a TE containing fat and water out of phase, the signal will cancel out and produce a signal void in that voxel. To correct for the out of phase black boundary artifact; the operator should select a TE where fat and water in phase.

44 The following formula should be in order to obtained in-phase or out of phase images: 1 x Field strength 1 x 3.355= 2.2msec 1.5 Tesla TE= 0ms in phase TE= 2.2ms out of phase TE=4.4ms in phase In phase images can be obtained at 1.5 Tesla with TE's of 4.4msec, 8.8msec, etc. Out of phase images can be obtained at 1.5T with TE's of 2.2, 6.6, 11.0 msec. WATER FAT The arrow represents the milliseconds after RF pulse is applied. WATER FAT AND WATER FAT BRIGHT SIGNAL DARK SIGNAL Dixon Technique BRIGHT SIGNAL When doing out of phase imaging, a TE of 2.2ms is used to allow chemical misregistration to occur. A dark band will be seen around structures where fat and water resides. 44

45 Magnetic Susceptibility Refers to the property of a substance to become magnetized. It is used to describe the strength of magnetic forces acting on substances that are exposed to the magnetic field. Classified as: 1. Diamagnetic These substances have paired orbital electrons and display no magnetic field however when exposed to an external field; these substances begin to slightly repel with the external magnetic field. These substances have a low negative magnetic susceptibility and show a small decrease in magnetic field strength. Example: Lead, plastic, wood, copper, bismuth, silver, oxygen. Notes 45

46 2. Paramagnetic These substances have unpaired orbital electrons and display a small positive magnetic moment. When exposed to an external magnetic field; these substances are slightly attracted to the external magnetic field and align themselves with it. These substances have a small increase in magnetic susceptibility and show a small increase in magnetic field strength. Example: Aluminum, magnesium, tungsten, platinum, gadolinium contrast agents. 3. Super-paramagnetic These substances have an intermediate positive magnetic moment that is greater than paramagnetic but less than ferromagnetic substances. Example: Iron oxide particles and iron oxide contrast agents (generally used as T2 or T2* contrast agents in the liver). 4. Ferromagnetic These substances have half-filled electron shells and display a large magnetic moment. When exposed to an external magnetic field; these substances begin to be attracted to the external magnetic field with increase strength. These substances retain its magnetism even after being removed from the external magnetic field. Example: Iron, cobalt, steel and nickel. Non-magnetic These substances display no magnetic moment. Example: Glass, plastic and wood. 46

47 Magnetic susceptibility artifacts Cause The presence of metallic objects or naturally occurring iron in patients that have a hemorrhage causes the spins in the area to be imaged not to be affected by the excitation pulse because they precess at a vastly different precessional frequency. Example: Metallic teeth implants, titanium orthopedic implants, Regions of air/bone/soft tissue interfaces, iron particles in mascara. Appearance Regions of signal void or increase signal. SIGNAL VOID IN THE AREA OF THE DENTAL IMPLANT SURROUNDED BY AN INCREASED SIGNAL AX T1 47

48 Susceptibility artifacts due to iron particles in mascara which causes an increased signal Regions of air/bone soft tissue interfaces causes local gradients distortions due to differences in magnetic field susceptibility which causes an increase in signal. Factors that increase the artifact: Increase in magnetic field strength An increase in TE which allow more time for dephasing The use of gradient echo sequences The use of narrow bandwidth techniques The use of thicker slices Solution The patient must be changed into a gown that has no metallic snaps and ask to remove all types of metals that he or she might have. Avoid using T2*gradient echo sequences as this increases magnetic suceptibity artifacts. : 48

49 A fast spin echo sequences is helpful as this uses multiple 180 degrees RF pulses which are more efficient at rephasing spins. Example: SIGNAL VOID SIGNAL VOID AX GRE AX T2 FSE The use of parallel imaging techniques. When a patient has a metallic hardware that could not be removed; try to use spin echo imaging sequences. Unfortunately some patients cannot remove certain types of metals for example: dental implants. These patients are to be instructed that any heating or discomfort they might experience; to immediately inform to the technologist. 49

50 Radiofrequency (Zipper Artifact) Causes The most common cause of RF artifacts is an extraneous noise source which reaches the receiver coil in which the door of the RF shielded scanner room is not fully closed or the seal is defective. RF emissions from anesthesia machines inside the MRI room. RF signals from radios or machine malfunctions. Improper RF transmitter adjustments Stimulated echoes from imperfect slice selection profiles Appearance Band of alternative white and black areas. Can occur in the phase or frequency direction. RF ARTIFACT RF ARTIFACT This Zipper artifact is oriented perpendicular to the frequency direction) 50

51 RF artifacts oriented in the frequency direction are caused by stimulated echoes from imperfect slice selection profiles or improper RF transmitter adjustments. Example: RF ARTIFACT RF ARTIFACT SAG T1 COR T1 Solution Eliminate the source of noise by calling the service engineer to check for any RF leaks. Replacing any broken RF door seals. Improve the spoiling of transverse magnetization. 51

52 (Motion and flow artifacts) Phase Mismapping and or Ghosting Can be divided into: Motion and Flow artifacts Motion Artifacts Cause Voluntary motion occurs from patient movement and or breathing. Involuntary motion occurs from the heart beating or due to abdominal peristalsis. Appearance Seen as replications of moving anatomy across the phase axis of an image. Occurs in the phase encoding direction BREATHING ARTIFACTS SEEN AS MULTIPLE REPLICATIONS Breathing motion artifacts seen in the abdomen 52

53 Solution Voluntary Motion Proper instructions and communication provided to the patient to remain still and or follow breathing instructions during the examination. Making the patient as comfortable as possible. Example: Stabilization of the patient by using the aid of positioning sponges, taping and or supporting pads. Using breath hold techniques The use of pre-saturation pulses Example: In the area of the neck to prevent swallowing artifacts. Claustrophobic patients might benefit from receiving some anxiety medication to be able to help them relax during the exam. Swapping the phase and frequency encoding axes will shift the direction of the motion but will not eliminate them. Solution Involuntary Motion Application of cardiac or peripheral gating when applicable. The use of ultrafast sequences such as (HASTE OR EPI) can be used to acquired rapid images between 2 to 5 seconds without the use of breath hold techniques when necessary Physiologic motion such as peristalsis of the abdomen can be reduced with an appropriate pharmacological agent such as 1mg of glucagon. Increasing the number of signals averaged (NSA, NEX) will reduce artifacts and increase signal-to-noise but at expense of increased imaging time. Flow artifacts (ghosting) Blood flow is measured as the total volume of blood that passes a certain point in the vascular system within a given period of time. Blood flow can be laminar, spiral, vortex or turbulent. 53 Cause Blood flow and pulsations

54 Appearance Seen as ghosting and or replications of a vessel Solution Use Gradient moment nulling techniques Reducing the TE The use of spatial pre-saturation pulses Swapping the phase and frequency encoding axes will shift the direction of flow artifacts but will not eliminate them. 7. Partial volume averaging Cause The signal intensities from adjacent nuclei within the same voxel average togetherand this prevents a true representation of the individual spins in the tissues. This is cause by the use of large voxels, a large FOV, A coarse matrix and thick slices. As the slice thickness increases, there is an increase of different types of tissues projecting in the slice. When a large FOV is acquired, there is an increase superimposition of different MRI signals. 54

55 Appearance A loss of spatial resolution Solution Use thinner slices Decrease voxel volume Decrease the FOV Use a large(fine) matrix 8. Crosstalk (CROSS-EXCITATION) An RF pulse of a particular slice excites an adjacent slice that was previously excited therefore intercepting with each other and these extra RF pulse create a saturation of signal. Appearance A black line or band is seen in the intercepting area CROSSTALK Two Intercepting RF pulses causes a signal void CROSSTALK ARTIFACT CREATING A SIGNAL VOID 55

56 Solution In 2D multi-slice acquisition, use a gap between slices. Most sequences use a minimum 10% gap. Inversion recovery sequences a 20% gap. Avoid using steep angles between slice groups to prevent cross excitation between slices. Careful placement of stacks so they don't cross each other. Use an Interleaved slice acquisition. Example: Acquiring the odd number of slices first followed by the even slices. 10. Moiré Artifact Cause A combination of wrap around artifact (from anatomy that extends outside the FOV) and Inhomogeneity of the main magnetic field from one side of the body to the other (due to the patient's body pressed against the bore of the magnet). Appearance Areas of white and black bands seen along the edge of the FOV Moiré artifact Solution Using spin echo pulse sequences which compensates for magnetic field inhomogeneities. 56

57 Keep the patient arms away from touching the surface of the bore of the magnet when performing gradient echo sequences. Use of anti-aliasing techniques. 11. Parallel Imaging Artifacts Cause Increasing the acceleration factor results in a reduction in the number of phase encoding lines acquired which results in a small phase FOV. Appearance Anatomy from outside the FOV is folded into the anatomy inside the FOV along the phase encoding direction resulting in a wraparound artifact. WRAP-AROUND ARTIFACT Parallel imaging artifact Solution Reducing the acceleration factor(r factor) Proper use of a calibration scan. The SENSE and GRAPPA technique works either by unwrapping the aliasing in the images or, equivalently, by filling in the missing lines of k-space. 57

58 Other Artifacts Magic angle artifact Cause When structures containing collagen lies at a 55 degree angle to the main magnetic field; this causes a faster T2 decay. Appearance Areas of high signal intensity near tendons where collagen is usually present. SAG PD SAG T2 FAT SAT Solution Changing the positioning angle of the anatomical structure Lengthening the TE in will result in a slower T2 decay. 58

59 ight Section QUALITY CONTROL 4. Quality Control The process that evaluates standards of quality and takes corrective action when output doesn't meet standards. It requires the following steps: 1. Acceptance testing. 2. The establishing of baseline performance. 3. Detection and diagnosis of changes in the equipment performance. 4. Correction verification. The ACR requires several quality control tests to be performed in order for facilities to obtain accreditation. The following are quality control tests that are performed: A. SLICE THICKNESS The accuracy with which a prescribed slice achieves the desired thickness. This test is performed annually using an MRI ACR phantom. When doing an AXIAL T1 weighted ACR series; the acceptable ranges are between 5.0mm mm.. 59

60 Slice thickness plays an important role in obtaining a good MR quality image.. Depending on the area of interest; a thin or thick slice may be chosen. Example: Areas such as the orbits and internal auditory canal require thin 1mm slices in order to visualize the small structures. In this way; lesions will not be missed and there will be an adequate resolution. When the slice thickness is reduced by a factor of 2, the NSA must be increased by a factor of 4 in order to maintain the same SNR. Notes 60

61 B. SPATIAL RESOLUTION The amount of small visual detail that can be seen in an image. It provides the ability to distinguish two points as separate and distinct. High contrast spatial resolution The test is used to determine the accuracy of the scanner to resolve small objects. It determines the degree to which individuals small bright spots spaced closely together in arrays are discernible. It should be performed weekly using the ACR MRI phantom. When doing an AXIAL T1 weighted ACR series; the acceptable resolution should be 1.0mm or better with four holes in at least 1 row being recognized as separated and distinct How many holes comprise each array row and Column? UL A Single Array Pair UR 61

62 C. CONTRAST RESOLUTION The subtle differences in signal intensities from tissues being imaged. Low contrast object detectability The ability to assess the extent to which objects of low contrast are detectable in an MR image. This test is to be performed weekly using an ACR MRI phantom. If the number of visible spokes is reduced to more than 3; the service engineer is to be notified. Two complete spokes 1 2 Notes 62

63 D. SIGNAL TO NOISE RATIO Determines the accuracy of the SNR. E. CENTER FREQUENCY This test ensures that the correct Larmor frequency is transmitted by the RF antennae. It is performed weekly using an ACR MRI accreditation phantom. Ranges for the center frequency should not deviate by more than 1.5ppm between weekly measurements. Center frequency determines: 1. The frequency of the RF pulse used for excitation 2. The frequency of the RF pulse used for signal detection Question 4. If the center frequency of hydrogen in a 1.5 Tesla scanner is MHz, what is the center frequency of hydrogen in a 3.0 Tesla scanner? PROBLEM SOLVING MHz 1.5Tesla 3.0Tesla MHz x 3.0Tesla= /1.5Tesla=128 MHz ANSWER 128 MHz 63

64 F. TRANSMIT GAIN OR ATTENUATION This test measures the amplitude and or dura on of the RF pulse (determines the flip angle) and should be performed weekly using an ACR MRI phantom. G. GEOMETRIC ACCURACY Determines the accuracy in which the image represents the true dimensions of the body part being image. Determine by the gradient system. This test should be performed weekly using an MRI ACR phantom. Acceptable ranges of geometric accuracy should be +-2mm of the true values. st 1 SAGITTAL LOC 148mm + / - 2mm SLICE 1 190mm + / - 2mm Air- bubble Circular ROI SLICE 5 190mm + / - 2mm 64

65 Quality Assurance The process that aims to prevent quality control problems through planned and systematic activities including documentation which help assure the quality of services. The technologist and the engineer are responsible to do the quality control testing in the MRI machine. Preventive maintenance is to be done by a service engineer Monthly which can include installing any software updates that might be available. Notes 65

66 ight Section EQUIPMENT HANDLING AND INSPECTION Equipment handling and inspection The purpose of this test is to ensure that MRI system equipment components are working properly and electrically stable. The test should be performed weekly. Several exams include: 1. System indicator lights 2. Safety lights and light bulbs 3. RF room integrity and door deals 4. Emergency cart 5. Coils and cables H. SYSTEM INDICATOR LIGHTS Indicator lights will provide an accurate centering for the examination of interest. I. SAFETY LIGHTS AND LIGHT BULBS Safety lights will provide alertness to the personnel and or anyone regarding the static magnetic field. The flickering of light bulbs can cause RF artifacts that can affect image quality which can cause RF artifacts 66

67 . J. RF ROOM INTEGRITY AND DOOR SEALS Passive shielding in which iron, copper or steel plates are incorporated into the walls, ceiling, and/or floor of the magnet room known as (Faraday's cage) is used to reduce the magnetic field and provide room integrity by preventing RF artifacts. Patient monitoring devices that are MR conditional can also be the cause of RF artifacts. Door seals It must be shielded as this can prevent RF leak artifacts from entering the room. Copper shielding Missing RF copper shielding K. EMERGENCY CART Only label and inspected MR safe and or appropriate MR conditional emergency carts following appropriate protocols are to be used in the MRI suite. Notes 67

68 . L. COILS AND CABLES The technologist should inspect coils and cables for any signs of tear and wear as this can cause the possibility of a patient burn and or fire hazard. Any signs of wear or tear must be reported to the supervisor and service engineer. A tag must be placed on defective equipment by the service engineer. Example: Notes 68

69 69 Notes

70 References Physical Principles Of Image Formation Carolyn Kauth, Roth (2006). Basic and Advanced Principles of MRI: MRI Review Program for Technologists. Boothwyn, Pennsylvania: Imaging Education Associates Elster D, Allen and Burdette H, Jonathan (2001). Questions and Answers in MAGNETIC RESONANCE IMAGING. Second Edition. St Louis, Missouri: Mosby: An affiliate of Elsevier Science. Schering AG Berlin/Bergkamen (1990). MRI made Easy, Germany by Nationales Druckhaus Berlin 70

MRI SYSTEM COMPONENTS Module One

MRI SYSTEM COMPONENTS Module One MRI SYSTEM COMPONENTS Module One 1 MAIN COMPONENTS Magnet Gradient Coils RF Coils Host Computer / Electronic Support System Operator Console and Display Systems 2 3 4 5 Magnet Components 6 The magnet The

More information

Pulse Sequence Design and Image Procedures

Pulse Sequence Design and Image Procedures Pulse Sequence Design and Image Procedures 1 Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant 2 A pulse sequence is a timing diagram designed with a series of RF pulses, gradients switching, and signal readout

More information

MR Basics: Module 6 Pulse Sequences

MR Basics: Module 6 Pulse Sequences Module 6 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

MR Advance Techniques. Flow Phenomena. Class II

MR Advance Techniques. Flow Phenomena. Class II MR Advance Techniques Flow Phenomena Class II Flow Phenomena In this class we will explore different phenomenona produced from nuclei that move during the acquisition of data. Flowing nuclei exhibit different

More information

Background (~EE369B)

Background (~EE369B) Background (~EE369B) Magnetic Resonance Imaging D. Nishimura Overview of NMR Hardware Image formation and k-space Excitation k-space Signals and contrast Signal-to-Noise Ratio (SNR) Pulse Sequences 13

More information

Magnetic Resonance Imaging Principles, Methods, and Techniques

Magnetic Resonance Imaging Principles, Methods, and Techniques Magnetic Resonance Imaging Principles, Methods, and Techniques Perry Sprawls Jr., Emory University Publisher: Medical Physics Publishing Corporation Publication Place: Madison, Wisconsin Publication Date:

More information

1 Introduction. 2 The basic principles of NMR

1 Introduction. 2 The basic principles of NMR 1 Introduction Since 1977 when the first clinical MRI scanner was patented nuclear magnetic resonance imaging is increasingly being used for medical diagnosis and in scientific research and application

More information

MR Basics: Module 8 Image Quality

MR Basics: Module 8 Image Quality Module 8 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

MRI Metal Artifact Reduction

MRI Metal Artifact Reduction MRI Metal Artifact Reduction PD Dr. med. Reto Sutter University Hospital Balgrist Zurich University of Zurich OUTLINE Is this Patient suitable for MR Imaging? Metal artifact reduction Is this Patient suitable

More information

Magnetic Resonance Imaging

Magnetic Resonance Imaging Magnetic Resonance Imaging Principles, Methods, and Techniques Perry Sprawls, Ph.D., FACR, FAAPM, FIOMP Distinguished Emeritus Professor Department of Radiology Emory University Atlanta, Georgia Medical

More information

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Magnetic Resonance Imaging Spatial

More information

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 ACR MRI accreditation program MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics,

More information

H 2 O and fat imaging

H 2 O and fat imaging H 2 O and fat imaging Xu Feng Outline Introduction benefit from the separation of water and fat imaging Chemical Shift definition of chemical shift origin of chemical shift equations of chemical shift

More information

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010 (N)MR Imaging Lab Course Script FMP PhD Autumn School Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder Date: November 3rd, 2010 1 Purpose: Understanding the basic principles of MR imaging

More information

Pulse Sequence Design Made Easier

Pulse Sequence Design Made Easier Pulse Sequence Design Made Easier Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant gurumri@gmail.com 1 2 Pulse Sequences generally have the following characteristics: An RF line characterizing RF Pulse applications

More information

MAGNETIC RESONANCE IMAGING

MAGNETIC RESONANCE IMAGING CSEE 4620 Homework 3 Fall 2018 MAGNETIC RESONANCE IMAGING 1. THE PRIMARY MAGNET Magnetic resonance imaging requires a very strong static magnetic field to align the nuclei. Modern MRI scanners require

More information

Weber State University Radiologic Technology 4603

Weber State University Radiologic Technology 4603 Weber State University Radiologic Technology 4603 MRI Physics and Instrumentation Instructor: Rex T. Christensen MHA R.T. (R) (MR) (CT) (ARRT) CIIP Contact Info: E-mail: rexchristensen@weber.edu Phone:

More information

2014 M.S. Cohen all rights reserved

2014 M.S. Cohen all rights reserved 2014 M.S. Cohen all rights reserved mscohen@g.ucla.edu IMAGE QUALITY / ARTIFACTS SYRINGOMYELIA Source http://gait.aidi.udel.edu/res695/homepage/pd_ortho/educate/clincase/syrsco.htm Surgery is usually recommended

More information

Image Quality/Artifacts Frequency (MHz)

Image Quality/Artifacts Frequency (MHz) The Larmor Relation 84 Image Quality/Artifacts (MHz) 42 ω = γ X B = 2πf 84 0.0 1.0 2.0 Magnetic Field (Tesla) 1 A 1D Image Magnetic Field Gradients Magnet Field Strength Field Strength / Gradient Coil

More information

HETERONUCLEAR IMAGING. Topics to be Discussed:

HETERONUCLEAR IMAGING. Topics to be Discussed: HETERONUCLEAR IMAGING BioE-594 Advanced MRI By:- Rajitha Mullapudi 04/06/2006 Topics to be Discussed: What is heteronuclear imaging. Comparing the hardware of MRI and heteronuclear imaging. Clinical applications

More information

MRI at a Glance. Catherine Westbrook. Blackwell Science

MRI at a Glance. Catherine Westbrook. Blackwell Science MRI at a Glance Catherine Westbrook Blackwell Science MRI at a Glance MRI at a Glance CATHERINE WESTBROOK MSC DCRR CTC Director of Training and Education Lodestone Patient Care Ltd Blackwell Science 2002

More information

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Advanced MSK MRI Protocols at 3.0T Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Outline Why High Field for MSK? SNR and Relaxation Times Technical Issues Example

More information

Index COPYRIGHTED MATERIAL. Note: Page number followed by italics are for figures and bold are for tables, respectively.

Index COPYRIGHTED MATERIAL. Note: Page number followed by italics are for figures and bold are for tables, respectively. Note: Page number followed by italics are for figures and bold are for tables, respectively. abdominal imaging aliasing along the phase axis of abdomen, 256 entry-slice phenomenon (ESP) in, 283, 283 5

More information

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Dr John Ridgway Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Physics for clinicians: Part I Journal of Cardiovascular Magnetic Resonance 2010, 12:71 http://jcmr-online.com/content/12/1/71

More information

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements MR Data for Treatment Planning: Issues, Protocol Optimization, and s (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements TG-117 Use of MRI Data in Treatment Planning

More information

High Field MRI: Technology, Applications, Safety, and Limitations

High Field MRI: Technology, Applications, Safety, and Limitations High Field MRI: Technology, Applications, Safety, and Limitations R. Jason Stafford, Ph.D. The University of Texas M. D. Anderson Cancer Center, Houston, TX Introduction The amount of available signal

More information

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review MRES 7005 - Fast Imaging Techniques Module 2 Artefacts and Imaging Optimisation for single shot methods Content: Introduction Phase error Phase bandwidth Chemical shift review Chemical shift in pixels

More information

2 Hardware for Magnetic Resonance Imaging

2 Hardware for Magnetic Resonance Imaging Hardware for Magnetic Resonance Imaging 13 2 Hardware for Magnetic Resonance Imaging Kenneth W. Fishbein, Joseph C. McGowan, and Richard G. Spencer CONTENTS 2.1 Introduction 13 2.2 Magnets 13 2.2.1 Permanent

More information

Chapter 2. The Physics of Magnetic Resonance Imaging

Chapter 2. The Physics of Magnetic Resonance Imaging Chapter 2. The Physics of Magnetic Resonance Imaging 2.1. Introduction The origins of the Nuclear Magnetic Resonance (NMR) signal and how it is manipulated to form images are the subjects of this chapter.

More information

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging JOURNAL OF MAGNETIC RESONANCE IMAGING 20:1046 1051 (2004) Technical Note The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging James W. Goldfarb, PhD* Purpose: To describe a known (but undocumented)

More information

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images 10. Phase Cycling and Pulsed Field Gradients 10.1 Introduction to Phase Cycling - Quadrature images The selection of coherence transfer pathways (CTP) by phase cycling or PFGs is the tool that allows the

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

MRI Systems and Coil Technology

MRI Systems and Coil Technology MRI for Technologists MRI Systems and Coil Technology PROGRAM INFORMATION MRI for Technologists is a training program designed to meet the needs of radiologic technologists entering or working in the field

More information

磁振影像學 MRI 磁振假影與磁振安全 磁振假影. 本週課程內容 Hardware-related Artifacts 盧家鋒助理教授 磁振假影 磁振安全

磁振影像學 MRI 磁振假影與磁振安全 磁振假影. 本週課程內容   Hardware-related Artifacts 盧家鋒助理教授 磁振假影 磁振安全 本週課程內容 http://www.ym.edu.tw/~cflu 磁振假影 磁振安全 磁振影像學 MRI 磁振假影與磁振安全 盧家鋒助理教授 國立陽明大學生物醫學影像暨放射科學系 alvin4016@ym.edu.tw MRI The Basics (3rd edition) Chapter 18: Artifacts in MRI MRI in Practice, (4th edition) Chapter

More information

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP MR Data for Treatment Planning and Stereotactic Procedures: Sources of Distortion, Protocol Optimization, and Assessment (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements

More information

A Conceptual Tour of Pulsed NMR*

A Conceptual Tour of Pulsed NMR* A Conceptual Tour of Pulsed NMR* Many nuclei, but not all, possess both a magnetic moment, µ, and an angular momentum, L. Such particles are said to have spin. When the angular momentum and magnetic moment

More information

Potential Risks of MRI in Device Patients

Potential Risks of MRI in Device Patients Outline Potential Risks of MRI in Device Patients Redha Boubertakh r.boubertakh@qmul.ac.uk MRI and cardiac implantable electronic devices (CIED) Components of an MRI scanner MRI implant and device safety

More information

In an unmagnetized piece of iron, the atoms are arranged in domains. In each domain the atoms are aligned, but the domains themselves are random.

In an unmagnetized piece of iron, the atoms are arranged in domains. In each domain the atoms are aligned, but the domains themselves are random. 4/7 Properties of the Magnetic Force 1. Perpendicular to the field and velocity. 2. If the velocity and field are parallel, the force is zero. 3. Roughly (field and vel perp), the force is the product

More information

Magnetic Resonance Imaging and Radio Frequency. Part 1. Produced on behalf of Mid Sussex Amateur Radio Society by M5BTB

Magnetic Resonance Imaging and Radio Frequency. Part 1. Produced on behalf of Mid Sussex Amateur Radio Society by M5BTB Magnetic Resonance Imaging and Radio Frequency Part 1 Produced on behalf of Mid Sussex Amateur Radio Society by M5BTB Why Now? During 2011 my physical health was deteriorating, and a brain tumour was diagnosed

More information

2015 Spin echoes and projection imaging

2015 Spin echoes and projection imaging 1. Spin Echoes 1.1 Find f0, transmit amplitudes, and shim settings In order to acquire spin echoes, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week,

More information

MRI Anatomy and Positioning Series Module 12: Fat Suppression Techniques

MRI Anatomy and Positioning Series Module 12: Fat Suppression Techniques MRI Anatomy and Positioning Series Module 12: Fat Suppression Techniques 1 Introduction... 3 RF FatSat... 4 HOAST... 4 FatSat... 5 Segment FS... 8 PhaseCycle... 9 Water Excitation... 10 STIR... 12 FatSep...

More information

RAD 229: MRI Signals and Sequences

RAD 229: MRI Signals and Sequences RAD 229: MRI Signals and Sequences Brian Hargreaves All notes are on the course website web.stanford.edu/class/rad229 Course Goals Develop Intuition Understand MRI signals Exposure to numerous MRI sequences

More information

1.5T HIGH FIELD SMALL ANIMAL MRI

1.5T HIGH FIELD SMALL ANIMAL MRI 1.5T HIGH FIELD SMALL ANIMAL MRI Designed Specifically for Veterinarians TECHNICAL GUIDE ADVANCING THE ART AND SCIENCE OF VETERINARY MRI The PetVet is the only high-field MRI system designed specifically

More information

Hardware. MRI System. MRI system Multicoil Microstrip. Part1

Hardware. MRI System. MRI system Multicoil Microstrip. Part1 Hardware MRI system Multicoil Microstrip MRI System Part1 1 The MRI system is made up of a variety of subsystems. the Operator Workspace Gradient Driver subsystem The Physiological Acquisition Controller

More information

PHY3902 PHY3904. Nuclear magnetic resonance Laboratory Protocol

PHY3902 PHY3904. Nuclear magnetic resonance Laboratory Protocol PHY3902 PHY3904 Nuclear magnetic resonance Laboratory Protocol PHY3902 PHY3904 Nuclear magnetic resonance Laboratory Protocol GETTING STARTED You might be tempted now to put a sample in the probe and try

More information

Multi-channel SQUID-based Ultra-Low Field Magnetic Resonance Imaging in Unshielded Environment

Multi-channel SQUID-based Ultra-Low Field Magnetic Resonance Imaging in Unshielded Environment Multi-channel SQUID-based Ultra-Low Field Magnetic Resonance Imaging in Unshielded Environment Andrei Matlashov, Per Magnelind, Shaun Newman, Henrik Sandin, Algis Urbaitis, Petr Volegov, Michelle Espy

More information

RF and Electronic Design Perspective on Ultra-High Field MRI systems

RF and Electronic Design Perspective on Ultra-High Field MRI systems RF and Electronic Design Perspective on Ultra-High Field MRI systems A DISSERTATION SUBMITTED TO THE FACULTY OF THE GRADUATE SCHOOL OF THE UNIVERSITY OF MINNESOTA BY SUNG-MIN SOHN IN PARTIAL FULFILLMENT

More information

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter

Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter MR-guided RT in Heidelberg Funded by the German Research Foundation (DFG) Heidelberg consortium received

More information

SPECIFICATIONS FOR A 4.7 TESLA/310MM BORE ACTIVELY SHIELDED MAGNET SYSTEM

SPECIFICATIONS FOR A 4.7 TESLA/310MM BORE ACTIVELY SHIELDED MAGNET SYSTEM SPECIFICATIONS FOR A 4.7 TESLA/310MM BORE ACTIVELY SHIELDED MAGNET SYSTEM Prepared by:- Magnex Scientific Limited The Magnet Technology Centre 6 Mead Road Oxford Industrial Park Yarnton, Oxford OX5 1QU,

More information

EE469B: Assignment 2 Solutions

EE469B: Assignment 2 Solutions EE469B Fall 26-7 RF Pulse Design for MRI EE469B: Assignment 2 s Due Thursday Oct 3 Introduction This assignment concerns the design of small-tip-angle 2D excitation pulses based on spiral k-space trajectories.

More information

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves Experiment 1 Goal: Examine the effect caused by changing flip angle on image contrast in a simple gradient echo sequence and derive T1-curves. Image

More information

BOLD fmri: signal source, data acquisition, and interpretation

BOLD fmri: signal source, data acquisition, and interpretation BOLD fmri: signal source, data acquisition, and interpretation Cheryl Olman 4 th year student, Department of Neuroscience and Center for Magnetic Resonance Research Discussion series Week 1: Biological

More information

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo Gradient Spoiling Average balanced SSFP magnetization Reduce sensitivity to off-resonance FFE, FISP, GRASS, GRE, FAST, Field Echo 1 Gradient-Spoiled Sequence (GRE, FFE, FISP, GRASS) RF TR G z G y G x Signal

More information

MAGNETOSCOP Measurement of magnetic field strengths in the range 0.1 nanotesla to 1 millitesla

MAGNETOSCOP Measurement of magnetic field strengths in the range 0.1 nanotesla to 1 millitesla MAGNETOSCOP Measurement of magnetic field strengths in the range 0.1 nanotesla to 1 millitesla Extremely high sensitivity of 0.1 nanotesla with field and gradient probe Measurement of material permeabilities

More information

NMR Basics. Lecture 2

NMR Basics. Lecture 2 NMR Basics Lecture 2 Continuous wave (CW) vs. FT NMR There are two ways of tuning a piano: - key by key and recording each sound (or frequency). - or, kind of brutal, is to hit with a sledgehammer and

More information

Pulsed NMR Experiment Guide Kenneth Jackson Physics 173, Spring 2014 Professor Tsai

Pulsed NMR Experiment Guide Kenneth Jackson Physics 173, Spring 2014 Professor Tsai Pulsed NMR Experiment Guide Kenneth Jackson Physics 173, Spring 2014 Professor Tsai 1. Introduction NMR or nuclear magnetic resonance occurs when nuclei are placed in a magnetic field. It is a physical

More information

BACKGROUND. ** 78% of all MRI scanners have Image Quality problems. *** *** 25% of all Multi-Channel RF coils have at least one bad channel.

BACKGROUND. ** 78% of all MRI scanners have Image Quality problems. *** *** 25% of all Multi-Channel RF coils have at least one bad channel. Range of Results from over 534 ACR-mandated Annual MRI Performance Evaluations on over 204 Magnets from 8 Vendors Spanning a 10-year Period Moriel NessAiver, Ph.D. - Simply Physics - Baltimore, MD moriel@simplyphysics.com

More information

MRI Systems and Coil Technology

MRI Systems and Coil Technology MRI for Technologists MRI Systems and Coil Technology PROGRAM INFORMATION MRI for Technologists is a training program designed to meet the needs of radiologic technologists entering or working in the field

More information

The promise of high-field MRI. High Field MRI Technology, Applications, Safety, and Limitations. High-field Scanners

The promise of high-field MRI. High Field MRI Technology, Applications, Safety, and Limitations. High-field Scanners High Field MRI Technology, Applications, Safety, and Limitations R. Jason Stafford, Ph.D. Department of Imaging Physics The University of Texas M. D. Anderson Cancer Center Houston, TX The promise of high-field

More information

PULSED/CW NUCLEAR MAGNETIC RESONANCE

PULSED/CW NUCLEAR MAGNETIC RESONANCE PULSED/CW NUCLEAR MAGNETIC RESONANCE The Second Generation of TeachSpin s Classic Explore NMR for both Hydrogen (at 21 MHz) and Fluorine Nuclei Magnetic Field Stabilized to 1 part in 2 million Homogenize

More information

Echo-Planar Imaging for a 9.4 Tesla Vertical-Bore Superconducting Magnet Using an Unshielded Gradient Coil

Echo-Planar Imaging for a 9.4 Tesla Vertical-Bore Superconducting Magnet Using an Unshielded Gradient Coil Magn Reson Med Sci, Vol. XX, No. X, pp. XXX XXX, 2015 2016 Japanese Society for Magnetic Resonance in Medicine TECHNICAL NOTE by J-STAGE doi:10.2463/mrms.tn.2015-0123 Echo-Planar Imaging for a 9.4 Tesla

More information

System/Imaging Imperfections

System/Imaging Imperfections System/Imaging Imperfections B0 variations: Shim, Susceptibility B1 variations: Transmit, Receive Gradient Imperfections: Non-linearities Delays and Eddy currents Concomitant terms 1 B0 Variations - Off-Resonance

More information

Fundamentals Behind the 10 Most Common Magnetic Resonance Imaging Artifacts with Correction Strategies and 10 High-Yield Points

Fundamentals Behind the 10 Most Common Magnetic Resonance Imaging Artifacts with Correction Strategies and 10 High-Yield Points Fundamentals Behind the 10 Most Common Magnetic Resonance Imaging Artifacts with Correction Strategies and 10 High-Yield Points Award: Magna Cum Laude Poster No.: C-1248 Congress: ECR 2011 Type: Educational

More information

Liver imaging beyond expectations with Ingenia

Liver imaging beyond expectations with Ingenia Publication for the Philips MRI Community Issue 47 2012/3 Liver imaging beyond expectations with Ingenia Contributed by John Penatzer, RT, MR clinical product specialist, Cleveland, OH, USA Publication

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

functional MRI: A primer

functional MRI: A primer Activation Leads to: functional MRI: A primer CBF Increased +ΔR CBV Increased +ΔR (C+) O Utilization Increased slightly? Venous [O ] Increased -ΔR* Glucose Utilization Increased? Lactate BOLD R=/T R=/T

More information

Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils

Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils Magn Reson Med Sci doi:10.2463/mrms.tn.2016-0049 Published Online: March 27, 2017 TECHNICAL NOTE Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils

More information

PULSED NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706

PULSED NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706 (revised, 2/12/07) PULSED NUCLEAR MAGNETIC RESONANCE Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706 Abstract A pulsed nuclear magnetic resonance technique (spin-echo)

More information

SPECIFICATION FOR A 7.0 TESLA/400MM ROOM TEMPERATURE BORE MAGNET SYSTEM

SPECIFICATION FOR A 7.0 TESLA/400MM ROOM TEMPERATURE BORE MAGNET SYSTEM SPECIFICATION FOR A 7.0 TESLA/400MM ROOM TEMPERATURE BORE MAGNET SYSTEM Prepared by:- Magnex Scientific Limited The Magnet Technology Centre 6 Mead Road Oxford Industrial Park Yarnton, Oxford OX5 1QU,

More information

TimTX TrueShape. The parallel transmit architecture of the future. Answers for life.

TimTX TrueShape.  The parallel transmit architecture of the future. Answers for life. www.siemens.com/trueshape TimTX TrueShape The parallel transmit architecture of the future. The product/feature (mentioned herein) is not commercially available. Due to regulatory reasons its future availability

More information

Principles of MRI EE225E / BIO265. Lecture 21. Instructor: Miki Lustig UC Berkeley, EECS. M. Lustig, EECS UC Berkeley

Principles of MRI EE225E / BIO265. Lecture 21. Instructor: Miki Lustig UC Berkeley, EECS. M. Lustig, EECS UC Berkeley Principles of MRI Lecture 21 EE225E / BIO265 Instructor: Miki Lustig UC Berkeley, EECS Question What is the difference between the images? Answer Both T1-weighted spin-echo gradient-echo Lower SNR Meniscus

More information

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Val M. Runge, MD Institute for Diagnostic and Interventional Radiology Clinics for Neuroradiology and Nuclear Medicine University Hospital Zurich

More information

Digital Imaging CT & MR

Digital Imaging CT & MR Digital Imaging CT & MR January 22, 2008 Digital Radiography, CT and MRI generate images in a digital format What is a Digital Image? A digital image is made up of picture elements, pixels row by column

More information

Lab 8 6.S02 Spring 2013 MRI Projection Imaging

Lab 8 6.S02 Spring 2013 MRI Projection Imaging 1. Spin Echos 1.1 Find f0, TX amplitudes, and shim settings In order to acquire spin echos, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week, but these

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

NEMA Standards Publication MS (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging

NEMA Standards Publication MS (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging NEMA Standards Publication MS 1-2008 (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging Published by: National Electrical Manufacturers Association 1300 North

More information

GE Healthcare. Discovery MR T. Simply powerful. Powerfully simple.

GE Healthcare. Discovery MR T. Simply powerful. Powerfully simple. GE Healthcare Discovery MR750 3.0T Simply powerful. Powerfully simple. Break free. The breast images you need in only two sequences. A complete liver study in a 15-minute time slot. Routine fmri with shorter

More information

NUMERICAL DESIGN OF RESONATOR COILS FOR HIGH FIELD MAGNETIC RESONANCE IMAGING. A Thesis

NUMERICAL DESIGN OF RESONATOR COILS FOR HIGH FIELD MAGNETIC RESONANCE IMAGING. A Thesis NUMERICAL DESIGN OF RESONATOR COILS FOR HIGH FIELD MAGNETIC RESONANCE IMAGING A Thesis Presented in Partial Fulfillment of the Requirements for the Degree Bachelor of Science in the Graduate School of

More information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information Ordering Information Please contact us if you have any questions or if you would like a quote or delivery schedule regarding the Catphan phantom. phone 800-525-1190, or 518-692-1190 fax 518-692-3329 mail

More information

SIEMENS MAGNETOM Skyra syngo MR D13

SIEMENS MAGNETOM Skyra syngo MR D13 Page 1 of 12 SIEMENS MAGNETOM Skyra syngo MR D13 \\USER\CIND\StudyProtocols\PTSA\*ep2d_M0Map_p2_TE15 TA:7.9 s PAT:2 Voxel size:2.5 2.5 3.0 mm Rel. SNR:1.00 :epfid Properties Routine Contrast Prio Recon

More information

MAKING TRANSIENT ANTENNA MEASUREMENTS

MAKING TRANSIENT ANTENNA MEASUREMENTS MAKING TRANSIENT ANTENNA MEASUREMENTS Roger Dygert, Steven R. Nichols MI Technologies, 1125 Satellite Boulevard, Suite 100 Suwanee, GA 30024-4629 ABSTRACT In addition to steady state performance, antennas

More information

Improving high-field MRI using parallel excitation

Improving high-field MRI using parallel excitation review Improving high-field MRI using parallel excitation MRI at high magnetic field strengths promises to deliver clearer images of the body s structure and function. However, high-field MRI currently

More information

Compact yet Sophisticated

Compact yet Sophisticated Compact yet Sophisticated Hitachi has brought Open MRI one step further in its evolution, to better assist medical professionals who work at the forefront of healthcare. AIRIS Light MSK offers radiologists

More information

Philips Site Yearly Performance Evaluation Philips Openview 16-Jan-08. Table of Contents

Philips Site Yearly Performance Evaluation Philips Openview 16-Jan-08. Table of Contents Philips Site Yearly Performance Evaluation Philips Openview 6-Jan-8 Table of Contents Summary and Signature Page 2 Specific Comments 3 Site Information 4 Equipment Information 4 Table Position Accuracy

More information

SIGNA Explorer Lift revives our MR

SIGNA Explorer Lift revives our MR Seiji Shiotani, MD, PhD Seirei Fuji Hospital in Fuji City, Shizuoka, Japan Masayoshi Sugimura Seirei Fuji Hospital in Fuji City, Shizuoka, Japan SIGN Explorer Lift revives our MR The clinical usefulness

More information

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT) Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in

More information

Electron Spin Resonance v2.0

Electron Spin Resonance v2.0 Electron Spin Resonance v2.0 Background. This experiment measures the dimensionless g-factor (g s ) of an unpaired electron using the technique of Electron Spin Resonance, also known as Electron Paramagnetic

More information

Evaluation of MWT Materials Accusorb MRI Shield. Dr. E. Kanal, Department of Radiology, UPMC. Wednesday, April 21, 2010

Evaluation of MWT Materials Accusorb MRI Shield. Dr. E. Kanal, Department of Radiology, UPMC. Wednesday, April 21, 2010 Evaluation of MWT Materials Accusorb MRI Shield Dr. E. Kanal, Department of Radiology, UPMC Wednesday, April 21, 2010 Background: In almost all x-ray and ultrasonographic imaging examinations, in order

More information

Analysis of magnetic and electromagnetic field emissions produced by a MRI device

Analysis of magnetic and electromagnetic field emissions produced by a MRI device Sept. 8-1, 21, Kosice, Slovakia Analysis of magnetic and electromagnetic field emissions produced by a MRI device D. Giordano, M. Borsero, G. Crotti, M. ucca INRIM Istituto Nazionale di Ricerca Metrologica,

More information

SPECIFICATIONS FOR AN MRBR 7.0 TESLA / 210MM ACTIVELY SHIELDED MAGNET SYSTEM

SPECIFICATIONS FOR AN MRBR 7.0 TESLA / 210MM ACTIVELY SHIELDED MAGNET SYSTEM SPECIFICATIONS FOR AN MRBR 7.0 TESLA / 210MM ACTIVELY SHIELDED MAGNET SYSTEM Prepared by:- Magnex Scientific Limited The Magnet Technology Centre 6 Mead Road Oxford Industrial Park Yarnton, Oxford OX5

More information

Introduction to MR Hardware. RF Coils C M L C T. = g * B 0. Rotating magnetization produces alternating magnetic field

Introduction to MR Hardware. RF Coils C M L C T. = g * B 0. Rotating magnetization produces alternating magnetic field Introduction to MR Hardware RF Coils Dominik v. Elverfeldt Sep 5 th 2012 Courtesy of Hans Weber, Freiburg C M R = 50 Transmission = B 0 Reception L C T R Oscillating with Lamor frequency. B 1 field perpendicular

More information

Combined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8

Combined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8 Combined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8 UNIVERSITY OF CAMBRIDGE 1.2 Rob C. Hawkes 1, Tim D. Fryer 1, Alun J. Lucas 1,2, Stefan B. Siegel 3, Richard

More information

3D-Printed Microstrip Resonators for 4.7T MRI. Saeed Javidmehr. A thesis submitted in partial fulfillment of the requirements for the degree of

3D-Printed Microstrip Resonators for 4.7T MRI. Saeed Javidmehr. A thesis submitted in partial fulfillment of the requirements for the degree of 3D-Printed Microstrip Resonators for 4.7T MRI by Saeed Javidmehr A thesis submitted in partial fulfillment of the requirements for the degree of Master of Science in Electromagnetics and Microwaves Department

More information

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare GE Healthcare Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare There is excitement across the industry regarding the clinical potential of a hybrid

More information

ACRIN 6686 / RTOG 0825

ACRIN 6686 / RTOG 0825 ACRIN 6686 (RTOG 0825) Advanced MRI Imaging Manual ACRIN 6686 / RTOG 0825 A phase III double blind placebo controlled trial of conventional chemoradiation and adjuvant temozolomide plus bevacizumab vs

More information

TECHNICAL SPECIFICATIONS. FOR AN MRBR 7.0 TESLA / 160mm ACTIVELY SHIELDED ROOM TEMPERATURE BORE MAGNET SYSTEM

TECHNICAL SPECIFICATIONS. FOR AN MRBR 7.0 TESLA / 160mm ACTIVELY SHIELDED ROOM TEMPERATURE BORE MAGNET SYSTEM TECHNICAL SPECIFICATIONS FOR AN MRBR 7.0 TESLA / 160mm ACTIVELY SHIELDED ROOM TEMPERATURE BORE MAGNET SYSTEM Prepared by:- Magnex Scientific Limited The Magnet Technology Centre 6 Mead Road Oxford Industrial

More information

Radio Frequency Field

Radio Frequency Field Radio Frequency Field Radio Frequency Coils and RF Power Distribution RF Coil Maps Distribution of RF Power GE 750W RF maps courtesy of Tobias Gilk Siemens Prisma (courtesy Siemens) Radio Frequency Field

More information

System Options. Magnetic Property Measurement System. AC Susceptibility. AC Susceptibility Specifications

System Options. Magnetic Property Measurement System. AC Susceptibility. AC Susceptibility Specifications System Options AC Susceptibility Magnetic Property Measurement System Many materials display dissipative mechanisms when exposed to an oscillating magnetic field, and their susceptibility is described

More information

The Pulsed Resistive Low-Field MR Scanner

The Pulsed Resistive Low-Field MR Scanner 39 Chapter 3 The Pulsed Resistive Low-Field MR Scanner 3.1 Background In the remaining part of this work we are going to describe hyperpolarized gas relaxation, diffusion and MR imaging experiments. These

More information

Passive Tracking Exploiting Local Signal Conservation: The White Marker Phenomenon

Passive Tracking Exploiting Local Signal Conservation: The White Marker Phenomenon Passive Tracking Exploiting Local Signal Conservation: The White Marker Phenomenon Jan-Henry Seppenwoolde,* Max A. Viergever, and Chris J.G. Bakker Magnetic Resonance in Medicine 50:784 790 (2003) This

More information