Cathodoluminoscopic Spectroscopy at RPI

Size: px
Start display at page:

Download "Cathodoluminoscopic Spectroscopy at RPI"

Transcription

1 Cathodoluminoscopic Spectroscopy at RPI JRSC 3C26 Basic Operations Manual J.D. Price and J.M. Hanchar February 18, 2002

2 ii Introducing the Luminoscope... 1 Cathodoluminoscope setup 12 steps to shooting electrons... 3 Using WinSpec The art of spectrography... 4 Acquisition setup... 4 Establishing background correction... 5 Calibration check... 5 Viewing the spectrum... 5 Acquisition over a large range of wavelengths... 6 Using WinView Crude digital imaging... 6 When you re through... 7 Appendix 1 Calibration data... 8 Appendix 2 Default Settings (as of Aug. 7, 2001)... 9 Acquisition Menu... 9 EXPERIMENTAL SETUP... 9 Calibration Menu USAGE Spectrograph Menu MOVE DEFINE CALIBRATE Setup Menu... 11

3 Introducing the Luminoscope Luminescence is a general term for the emission of non-thermal radiation from a material, generally in the visible range, but including the infrared and ultraviolet spectrum. The emission is the result of incident energy imparted to the material, occurring when the material has certain patterns of electron energy levels which may be related to the type of structure, the presence of trace amounts of elements which serve as activators, or certain crystal defects. Luminescence includes phosphorescence and fluorescence; the former refers to samples which radiate energy for 10-8 seconds or more after removal of the excitation source, the latter to samples which cease to radiate 10-8 seconds or less after removal of the source. Cathodoluminescence refers to excitation using an electron source. A cathodoluminocope composed of a cathode ray tube, a vacuum chamber with a lead-glass porthole, and a compound microscope. A mechanical vacuum pump provides a reasonable vacuum to the chamber, and a power control unit controls the current and voltage applied to the filament within the tube. When current and voltage are applied to the tube, the filament produces electrons which are accelerated into the vacuum chamber. Small magnets control the direction of the electron path, which may be focused onto a sample within the chamber. In addition to the above components, the RPI luminoscope also has a Princeton Instruments spectrograph and charge coupled device Figure 1. The RPI Cathodoluminoscope spectroscopy system: the sample chamber (detail in figure 2) is attached to stage of an Olympus BH-2 polarizing compound microscope with trinocular head. Spectrograph and CCD are connected through the trinocular optical tube. Power source switches for the spectrograph and Hg lamp (calibration standard) are also indicated. (CCD, also called the detector). The spectrograph consists of two surfaces on a pivot; one is a mirrored grating surface with inclined facets, which spread-out incident light into a spectral range, in order to identify discrete wavelengths in the electromagnetic energy spectra, the other is a flat mirrored surface that reflects the image in the microscope. The CCD is able to digitize the spectrum or an image as distributed by the spectrograph. To drive all of this, the RPI luminoscope has a Pentium-based computer running Roper Scientific s WinSpec and WinView programs with the Windows 98 platform and a Princeton Instruments driver computer. Please note that an electron beam will generate energies above the visible electromagnetic spectrum, including x-rays. The lead-glass and brass fittings of the sample chamber are sufficient to prevent x-rays from exiting the chamber. It is therefore important that any damage to the sample holder, particularly the window be repaired prior to further use of the machine.

4 2 Figure 2. Detail of the sample chamber. Note that vacuum pump valve is shown in open position (parallel to the vacuum line) Figure 3. Front panel of the power control unit for the cathode ray tube, which includes a vacuum gauge for the sample chamber.

5 3 Cathodoluminoscope setup 12 steps to shooting electrons Set up of the cathodoluminoscope is uncomplicated provided that you are familiar with the basic pieces of the machine. If you have not done so already, please familiarize yourself figures 1-3 prior to starting the set up instructions. 1. Turn on vacuum pump, under the right end of the table. The switch is located on the power cord near the plug. Close the pump valve on right rear of sample chamber (the pump valve is closed when the handle is at a right angle to the flow direction) and open sample chamber valve. 2. Place the sample in the chamber. The sample holder will accept samples of 1 cm or less in height with dimensions of 7.5 x 4.5 cm or less. Whenever possible, make sure the sample is flat in the holder. Depending on the sample height, glass plates may need to be inserted between the sample and the holder to insure correct focusing distance. The holder may contain up to three petrographic thin sections, but two is recommended. 3. Close the chamber and open the close the sample chamber bleed valve and the pump valve. 4. Turn the control knob on the Luminoscope box to pumpdown. Note, the microscope lamp is plugged into the luminoscope box, and thus the control knob must be turned at least to pumpdown to turn on the microcope lamp. 5. Switch on the controller (black box with RS Princeton Instruments on face, located above the computer CPU case) and the spectrometer (small black box with red switch). 6. Allow the sample chamber to pump down to millitorrs (see vacuum gauge, figure 2). This usually takes ten minutes, which is adequate time for the electronic components to warm up. If the vacuum takes a long time to get to 100 millitorrs, then it is most likely that there is a leak in the o-ring on the Pb glass window. Try rotating the yellow ring around the glass to reseat the O- ring. If that fails to remedy the vacuum leak, replace the o-ring (spare o-rings may be ordered through Don Marshall ( ) for a modest cost. 7. As the vacuum pump evacuates the chamber, fully slide in the trinocular shutter, turn on the microscope lamp, and focus the view (see caution, right). 8. Switch the control knob to the regulated position. This regulates voltage and current by stabilizing the vacuum. 9. Switch on the electron tube by flipping the switch on the lower left of the Luminoscope face panel. 10. Slowly increase current to 0.7 ma. (Good current strong CL in many materials with least amount of damage). 11. Switch the meter knob to volts, and slowly turn the voltage knob until the meter reads 12 kv. Caution: the microscope was not specifically designed for use with the sample chamber; therefore exercise care when focusing or switching objective lenses. Do not let either objective lens touch the lead glass on the top of the sample holder. Such abuse will damage the objective lens and possibly crack the glass (an item worth more than $200). If you are unable to focus the sample, you need to open the sample chamber, and elevate the sample. 12. Adjust deflection magnets, on stage to center beam on the specimen under the objective. Loosen the magnet set screws to slide magnets in and out from lens area. While looking through the oculars of the microscope, move both magnets until you achieve the brightest intensity of luminescence in the sample.

6 4 Using WinSpec The art of spectrography Please make certain that you have carefully followed instructions concerning the setup of the luminoscope prior to making measurements. Before acquiring spectra or images, make certain that you have powered the controller (black box with RS Princeton Instruments on face) and the spectrometer (small black box with red switch). Note: the following provides the basic setup suitable for most users of the RPI luminoscope. More specific information on the instrumentation software is available in the WinSpec manual. Please cautiously undertake any deviations from the settings outlined below - some changes can greatly affect performance. The default settings are listed in Appendix 2 for reference. Acquisition setup Prior to software adjustments, you will want to close down the mechanical slit to 0.25 mm to increase the spectrographic resolution. 1. Start WinSpec32 (double click on desktop icon). Note: if you have recently (in the last minute) powered the spectrometer, you may encounter an error message indicating that the computer is not communicating with the spectrometer (ACTION SP-150 on COM1). This message pops up if the motor is initializing. Wait one minute and then click "Retry." 2. On the menu bar at the top of the window, find and click on Setup. On the resulting pull-down menu, select Detector Temperature and check that the specified temperature is -20 to -25 o C. Wait for the current temperature to reach the set temperature. Shortly thereafter the current temperature will lock. 3. Returning to the menu bar at the top of the window, click on Acquisition and select Experimental Setup from the resulting menu. By default, the Main tab will appear. Specify an Exposure Time (the time the shutter is to remain open). This will vary depending on the intensity of the light you are analyzing 0.1 seconds is good for really bright objects, such as the Hg lamp, 10 seconds is more appropriate for darker emitters, such as many geological materials. Set Number of Spectra to 1, and under the CCD Readout select Use Region of Interest and under accumulations, set Number to 1 4. Now click on the Data File Tab, and enter a name for the spectrum in text box at the top. Before entering the name, please click on the [...] box adjacent to the name box, and browse for or set up a folder to place your files (e.g. C:\Program Files\Princeton Instruments\WinSpec32\yourdata). In most cases you will want to not select Auto Increment File Name; enabling this feature will sequentially enumerate spectrum files. You may select whether you wish to Overwrite or Append files with the same name, and you may request confirmation to prevent overwriting files you wish to retain. In general, under Auto-save and prompts, select Don't auto-save or ask to save and leave the Use New Window for Each Run box unchecked. 5. Now click on the ROI Setup tab (note: these settings only apply if Use Region of Interest is selected on Main tab- step 3) and set Edit Pattern to 1. Select Spectroscopy Mode. Under Wavelength (the number of columns defined by x pixels, where 21 pixels = 1nm), set Start to 1, End to 576, and Group to 1 (This samples the full CCD - 1 to 576 pixels, a required setting for spectroscopy). Under Slit (software slit - the number of rows defined by y pixels), the Start may be any number from 1 to 384, and the Height may be any number between 1 and 384. Due to distortion of the spectra at the edges of the CCD, the best place to collect data is in the center. Consider setting Start to 192 and Height to 1; sampling the single centerline of pixels on the CCD. 6. Hit Store button to save your settings! 7. Now click on the Data Corrections Tab, and under Arithmetic make certain that Background is checked, and that a file name for the background appears in the following box (e.g. Back). To

7 5 begin with, leave Flatfield unchecked. Under CCD Blemishes, leave Remove unchecked and Off selected for Cosmic Ray Removal. 8. No parameters should be active under the ADC tab. 9. Click on Timing Tab and set Timing Mode to Free Run and Shutter Control to Normal Select Safe Mode (async). Set the Delay Time to Seconds and make certain Wait For TTL is unselected. Set Edge Trigger to +edge. 10. None of the items should be selected under the Processes Tab. 11. Click OK box at base of the window. Establishing background correction Now you are almost ready for acquisition. But before acquiring a spectrum, you must acquire a background. The background is the energy read by the CCD sees when the shutter is closed, and will be subtracted out from your observed spectra. 12. On the main menu bar, click on Acquisition, and select Acquire Background from the drop-down menu. A window will open and a background spectrum will appear after the prescribed acquisition time. The resulting spectrum should be a lot of noise with a low distribution of counts around the mean observation. What is focus (Acquistion menu or green button)? Focus is a command that repeats acquisition until stopped. It is designed to allow an operator to tune the focal length of the trinocular tube or the spectrograph-ccd tube without having to hit acquire each time. Although most users should not attempt to adjust these focal lengths, one may use focus in liu of acquire. Calibration check Prior to evaluation of an unknown sample, you will want to acquire a spectrum on a standard source to verify the calibration. In general, unless the spectrograph or CCD are moved, the machine remains calibrated. However, a quick check is always warranted. Please proceed with steps 13-23, which will also acquaint you with standard operation of the machine. 13. Find the mercury vapor wand lamp and power supply. Place the lamp under the low power objective lens on the microscope, and pull the trinocular shutter open (fully extended). If you have not done so already, adjust the mechanical slit opening to 0.25 µm. Warning, the lamp is a source of UV radiation, so refrain from staring at the lamp. Additionally, be careful not to pinch the wand lamp between the objective lens and the stage. 14. With the lamp in place, go to the spectrograph menu and select move. Specify an angle of 540, a position such that it is within 13 nm of a known peak. Click on Acquisition and select Acquire from the drop-down menu, or click the button with ACQ in a green circle on the custom tool bar. The shutter will click open and then close after the preset amount of time, and the data will display in the active window. If properly calibrated, you will see a significant peak at nm (see appendix 1). Viewing the spectrum If the spectrum is too small to observe of too large to fit in the window, you may rescale the data to fit. 15. Click on Display on the menu bar, and select Autoscale or Auto 5%-95% (or use one of the two buttons on the bottom far left of the spectrum) to fit the data to the window height. Autoscale will place the background on the x-axis and set the scale to the tallest peak in view. Auto 5%-95% will leave a margin on the top and bottom.

8 6 16. To determine the exact position and intensity of any part of the spectrum, click on the graph. Then click on View on the menu bar, and select Info or click on the info icon (a yellow "i" in a window) on the custom toolbar for the Info window. Acquisition over a large range of wavelengths You will note that the simple acquisition described above covers a 37 nm of the complete mercury spectrum. To conduct a thorough test of the calibration, you must do so using the Step and Glue procedure. This procedure takes a series of sequential 30 nm spectra over a specified range and produces a compiled spectrum. 17. Click on Acquisition and select Step and Glue... from the menu. 18. Under Glued Output Data File Name, specify a name for the final spectrum (e.g. HgLine). By clicking on the [...] box you may browse or create a folder for this file. 19. Under Gluing Parameters, state a starting value and an ending value. The RPI spectrometer uses a 1800 nm grooved plate with a 500 nm blaze; optimal settings for looking at the nm part of the spectrum. Keep that in mind when constructing a spectrum. Specify a minimum overlap of each spectrum; higher values are more time consuming. 20. Under Raw Data Files Setup, specify a File Names Template (this will be the prefix to all of the individual 30 nm spectra files). 21. Specify a File Increment Value (a value of 1 creates a nice sequence) that will be automatically placed on each 30 nm spectrum file. 22. Check the box prior to Reset After Each Run 23. When finished, click Run. The computer starts by moving to the center of the first 30nm spectrum, opens the shutter, counts for the prescribed time, closes the shutter and displays the spectrum. Then it moves to the center of the second 30 nm spectrum and repeats until the range in the gluing parameters is covered. When finished, a new window will open and the entire spectra will be displayed. If you are acquiring a spectrum on the Hg lamp for calibration verification, and the spectrum is not as presented in figure A1 (appendix 1), please consult Daniele Cherniak or Jon Price. Do not attempt to rectify the calibration without further instruction. Using WinView Crude digital imaging The CCD may also be used as a digital imaging device. However, there are several limitations to imaging on the RPI scope. 1. While the CCD is optimized to examine spectra with detailed resolution, it is a little too crude for producing high-resolution images. 2. The CCD is strictly a grayscale digitizer, so the lovely hues produced by CL activators are lost. 3. The pressure-sealed sample chamber compromises microscope objective height, and thus high-powered lenses are not an option. 4. While the turret is two sided, the mirrored side (1200 BLZ=MIRnm) is not an optimal mirror for imaging. While the best images are acquired using a separate camera, WinView may be used to capture crude pictures of Cl distribution, or (if the 1800 grating is employed) an optical resolution of diffraction peaks. Use of WinView is fairly straightforward, particularly as it looks A glitch after shifting back an forth from WinSpec to WinView, you may find that one or both of the programs will no longer restart. In that case, depress Ctrl-Alt-Delete simultaneously one time, find one of these programs on the list of running programs, and select end task. Repeat until all are removed from the list. Double click on the icon of the program you wish to start. and operates almost exactly like WinSpec. However, it is important to be mindful of the fact that it employs the settings file from WinSpec, and changes made here will affect WinSpec operation.

9 7 If you have already configured WinSpec as specified previously (such that you are able to obtain a calibration Hg spectra, then there are only a few things to change. If you have not configured Winspec, return to the previous section and do so before proceeding with this section. 1. If you wish to image what you see in the ocular lenses, you need to rotate the mirror side of the turret into place. In Winpec (reopen it if you have closed it) go to the Spectrograph Menu, and select MOVE..., and specify the 1200 BLZ=MIRnm. Set move to at 0.0. If you wish instead to image a particular peak in the spectra, got to the Spectrograph Menu, and select MOVE..., and keep the 1800 BLZ=500nm selected, and specify the peak angle in the box. 2. Open the mechanical slit to 3mm by rotation the vernier knob on the spectrograph. 3. Launch Winview (double click on desktop icon). 4. On the menu bar at the top of the window, click on Acquisition and select Experimental Setup from the resulting menu. As in Winspec, the Main tab page appears in front. Specify an Exposure Time. Again, this will vary depending on the intensity of the CL from the sample; 0.1 seconds is good for really bright objects, 10 seconds is more appropriate for many geological materials. Under the CCD Readout select Use Full chip. 5. Because you have changed the acquisition dimension of the chip from the single pixel side setting in the previous section (ROI) to the full chip, you must acquire a background. On the main menu bar, click on Acquisition, and select Acquire Background from the drop-down menu. A window will open and a background spectrum will appear after the prescribed acquisition time. The result will be a rather dark image. 6. Select Acquire on the Acquisition menu, and the image will appear in a separate window. Adjust shutter time (step 4) if necessary. When you re through 1. Switch off the electron tube with the toggle switch on the lower left side of the control panel. 2. Turn down the current and voltage knobs fully counter-clockwise 3. Turn the control knob to pump down. 4. Close the pump valve (rotate until it is 90 o to the line) 5. Open the chamber bleed valve to vent the sample chamber. 6. Switch off the pump. 7. Turn off the spectrometer and the Princeton Instruments control. 8. Shutdown the computer.

10 8 Appendix 1 Calibration data Wavelength (nm) Mercury * ** * * * ( ) * ( ) ( ) ( ) ( ) *indicates strong line ** indicates strongest line for the element Figure A1. CL spectrum for mercury.

11 9 Appendix 2 Default Settings (as of Aug. 7, 2001) The text in appendix 2 is formatted as follows: Bold - user entered parameters Underlined - software determined parameters Italized - comments by document author Acquisition Menu EXPERIMENTAL SETUP Here follows a typical setup for spectra acquisition (worked for Hg lamp) MAIN TAB Exposure time sec Number of Images: 1 CCD Readout Use region of Interest selected (or use full chip) Accumulations Number: 1 DATA FILE TAB Name: Spectrum01 C:\Program Files\Princeton Instruments\WinSpec32\Data Auto Increment File Name Enable unchecked Overwrite/Append Overwrite existing files Confirm before overwriting checked Auto-save and prompts Don't auto-save or ask to save Use new window for each run unchecked ROI SETUP TAB These settings only apply if "Use region of Interest" is selected on MAIN TAB Edit Pattern: 1 Number stored: 1 Imaging Mode unselected Spectroscopy Mode selected Wavelength Spectroscopy requires full width of CCD - 1 to 576 pixels Start: 1 End: 576 Group: 1 Slit Start: 192 Height: 1 Use Software Binning unselected DATA CORRECTIONS TAB Arithmetic Background checked Back Flatfield unchecked CCD Blemishes Remove unchecked

12 10 Cosmic Ray Removal Off selected Calibration Menu USAGE Calibration Mode Auto Spectro selected Calibration Units nm Save as Default Spectrograph Menu MOVE... Only initiate movement when the spectrometer motor is not already moving - listen for it Grating: 1800 BLZ=500nm (for spectrometry) or 1200 BLZ=MIRnm (for imaging) Move to: user's choice DEFINE... MAIN TAB Action SP150 on COM1 Use for Auto-spectro Calibration checked Excitation (doesn't apply unless using a laser source) Laser Wavelength: 0 nm Warn when crossing Laser Line: unselected GRATINGS TAB Grating Grooves Name BLZ=500nm Spectrometer side of turret 1800 Grating with 500 nm Blaze BLZ=MIRnm Mirror side of turret (imaging) Turret:1 Display warning on grating change unchecked (more of a hassel than its worth - just observe warnings listed under "Spectrograph, Move" Slew selected (Scan doesn't seem to work) CONNECT TAB COM1 Comm settings... button Baud: 9600 Data bits: 8 Par: None Stop bits: 1 Use Flow control: Unchecked CALIBRATE... warning: this is the spectrometer calibration - settings change spec angle and signal reception! Do not change these values unless you know what you are doing! Calibration based on Hg lamp, and standardized to a Dy-doped Zircon, J.D. Price, March 21, Spectrograph to calibrate ACTION SP150 on COM1

13 11 Grating to Calibrate 1800 BLZ=500nm Detector Pixel Width: 22 Magnification: 1 Grating movement slew selected (scan doesn't seem to work) Offset... button Selected grating: 1800 BLZ=500nm Reference Wavelength: (strong peak on the left side of the spectrometer range) Offset Value: Adjust... button Selected grating: 1800 BLZ=500nm Reference Wavelength: (intense peak on right of the calibration spectrum) Adjust Value: Dispersion... button Selected grating: 1800 BLZ=500nm Geometrics Focal Length (mm): Inclusion Angle: Detector Angle: Dispersion Calibration Lower Reference Wavelength: Higher Reference Wavelength: Next Target Wavelength: Setup Menu HARDWARE (refers to CCD controller - black box labled RS Princeton Instruments) CONTROLLER/CAMERA TAB ST133 version 3 EEV 576x384 [3ph] (setting on [6ph] will only utillize right half of screen) small shutter NTSC (shouldn't matter - this parameter refers to video display andwe don't have a video unit) Do not select user defined chip and timing! DISPLAY TAB Rotate Checked Reversed Checked Flip Checked image on CCD is rotated, reversed, flipped from optical image INTERFACE TAB Type: High-speed PCI the following are fixed parameters Interupt Level: IRQ 3 I/O address: 0x0d400 I/O address 2: d000 H I/O address 3: b800 H CLEANS/SKIPS TAB

14 Cleans Number of Cleans: 1 Number of Strips per Clean: 576 Vertical Skips: 2 Number of Blocks: 5 12

RENISHAW INVIA RAMAN SPECTROMETER

RENISHAW INVIA RAMAN SPECTROMETER STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples

More information

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop 1. Log in usage using the SMIF web site 2. Turn power

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

IBIL setup operation manual for SynerJY software version

IBIL setup operation manual for SynerJY software version IBIL setup operation manual for SynerJY software version 1.8.5.0 Manual version 1.0, 31/10/2008 Author: Carlos Marques Equipment Managers: Carlos Marques, +351219946084, cmarques@itn.pt Luís Alves, +351219946112,

More information

Renishaw InVia Raman microscope

Renishaw InVia Raman microscope Laser Spectroscopy Labs Renishaw InVia Raman microscope Operation instructions 1. Turn On the power switch, system power switch is located towards the back of the system on the right hand side. Wait ~10

More information

VS7550 VUV/UV Mini Spectrograph Operating Manual

VS7550 VUV/UV Mini Spectrograph Operating Manual Document RD 15 11 No: VS7550 VUV/UV Mini Spectrograph Operating Manual VS7550 Operating Manual 1 Table of Contents Table of Contents Overview Specifications Vacuum Interface Software and Drivers Packing

More information

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier). JEOL 200CX operating procedure Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Specimen loading 1.1. Unlock the TUMI system. 1.2. Load specimen(s) into the holder. If using the double tilt holder, ensure

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120)

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120) Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120) Please contact Dr. Amanda Henkes for training requests and assistance: 979-862-5959, amandahenkes@tamu.edu Hardware LN 2 FTIR FTIR camera 1

More information

WITec Alpha 300R Quick Operation Summary October 2018

WITec Alpha 300R Quick Operation Summary October 2018 WITec Alpha 300R Quick Operation Summary October 2018 This document is frequently updated if you feel information should be added, please indicate that to the facility manager (currently Philip Carubia,

More information

FEI Tecnai G 2 F20 Operating Procedures

FEI Tecnai G 2 F20 Operating Procedures FEI Tecnai G 2 F20 Operating Procedures 1. Startup (1) Sign-up in the microscope log-sheet. Please ensure you have written an account number for billing. (2) Log in to the computer: Login to your account

More information

Confocal Raman Microscopy (WITec Alpha 300R)

Confocal Raman Microscopy (WITec Alpha 300R) Confocal Raman Microscopy (WITec Alpha 300R) Please refer to Witec Alpha300R Confocal Raman Microscope User Manual for the details of the operating procedure. Sample preparation 1. Attach your sample on

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside December 21, 2017 (rev. 3.4) 1 Before you begin Complete

More information

Ph 3455 The Photoelectric Effect

Ph 3455 The Photoelectric Effect Ph 3455 The Photoelectric Effect Required background reading Tipler, Llewellyn, section 3-3 Prelab Questions 1. In this experiment you will be using a mercury lamp as the source of photons. At the yellow

More information

Things to check before start-up.

Things to check before start-up. Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your

More information

Instruction Manual for HyperScan Spectrometer

Instruction Manual for HyperScan Spectrometer August 2006 Version 1.1 Table of Contents Section Page 1 Hardware... 1 2 Mounting Procedure... 2 3 CCD Alignment... 6 4 Software... 7 5 Wiring Diagram... 19 1 HARDWARE While it is not necessary to have

More information

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual Scanning Electron Microscope FEI INSPECT F50 Step by step operation manual Scanning Electron Microscope, FEI Inspect F50 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check

More information

PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry

PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry Purpose PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry In this experiment, you will study the principles and applications of interferometry. Equipment and components PASCO

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside March 8, 2018 (rev. 3.5) 1 Before you begin Complete

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Follow the procedures below when you use the Hitachi 7100 TEM. Starting Session 1. Turn on the cold

More information

2. Raise HT to 200kVby following the procedure explained in 1.6.

2. Raise HT to 200kVby following the procedure explained in 1.6. JEOL 2100 MANUAL Quick check list 1. If needed, fill the reservoir with LN2 2. Raise HT to 200kVby following the procedure explained in 1.6. 3. Insert specimen holder into TEM (Insert holder in airlock,

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS Zeiss LSM 510 M eta Confocal M icroscope OPERATING INSTRUCTIONS Starting the System: 1. Turn the black knob on the laser box one-quarter turn from Off to On. You will hear the laser cooling mechanisms

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Sabeeh Irfan Ahmad, Physlab, 23 July 2016

Sabeeh Irfan Ahmad, Physlab, 23 July 2016 Horiba ihr550 Spectrometer QuickInstallation and Operation Guide Sabeeh Irfan Ahmad, Physlab, 23 July 2016 The Horiba ihr550 is an imaging spectrometer that can be used both as a spectrograph and as a

More information

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. http://staff.washington.edu/wpchan/if/cm100_inst.shtml Instructions for the Philips CM100 TEM and peripherals

More information

University of Wisconsin Chemistry 524 Spectroscopic Components *

University of Wisconsin Chemistry 524 Spectroscopic Components * University of Wisconsin Chemistry 524 Spectroscopic Components * In journal articles, presentations, and textbooks, chemical instruments are often represented as block diagrams. These block diagrams highlight

More information

RENISHAW RAMAN MICROSCOPE STANDARD OPERATING PROCEDURE

RENISHAW RAMAN MICROSCOPE STANDARD OPERATING PROCEDURE RENISHAW RAMAN MICROSCOPE STANDARD OPERATING PROCEDURE 09DJ00 CHRIS BUXEY (TECHNICIAN) X2715 C.BUXEY@SURREY.AC.UK V5.0 OCTOBER 2008 RENISHAW MICRORAMAN STANDARD OPERATING PROCEDURE 1.0 Foreword Chris Buxey:

More information

CATHODOLUMINESCENCE IMAGING UNIVERSITY OF ST ANDREWS

CATHODOLUMINESCENCE IMAGING UNIVERSITY OF ST ANDREWS CATHODOLUMINESCENCE IMAGING AT THE UNIVERSITY OF ST ANDREWS Instruction Guide by Adrian Finch 2 Noddy s Guide to Using the CL at St Andrews (SHORT version) 1. Switch on plugs at Wall (Microscope, Luminoscope

More information

Educational Spectrophotometer Accessory Kit and System OS-8537 and OS-8539

Educational Spectrophotometer Accessory Kit and System OS-8537 and OS-8539 GAIN 1 10 Instruction Manual with Experiment Guide and Teachers Notes 012-06575C *012-06575* Educational Spectrophotometer Accessory Kit and System OS-8537 and OS-8539 100 CI-6604A LIGHT SENSOR POLARIZER

More information

Bruker Dimension Icon AFM Quick User s Guide

Bruker Dimension Icon AFM Quick User s Guide Bruker Dimension Icon AFM Quick User s Guide March 3, 2015 GLA Contacts Jingjing Jiang (jjiang2@caltech.edu 626-616-6357) Xinghao Zhou (xzzhou@caltech.edu 626-375-0855) Bruker Tech Support (AFMSupport@bruker-nano.com

More information

Microscopy from Carl Zeiss

Microscopy from Carl Zeiss Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path

More information

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Operations Guide Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Nikon Eclipse 90i Operations Guide STARTUP Startup Powering Up Fluorescence

More information

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Warning Always adjust your specimen height before closing the chamber door to make sure your specimen will not hit the bottom of the lens;

More information

TRAINING MANUAL. Olympus FV1000

TRAINING MANUAL. Olympus FV1000 TRAINING MANUAL Olympus FV1000 September 2014 TABLE OF CONTENTS A. Start-Up Procedure... 1 B. Visual Observation under the Microscope... 1 C. Image Acquisition... 4 A brief Overview of the Settings...

More information

Simplified Instructions: Olympus Widefield Microscope S1230

Simplified Instructions: Olympus Widefield Microscope S1230 Contents General Microscope Operation Simple Image Capture Multi-Wavelength Capture Z-Series Timelapse Combining Capture Modes Synopsis of Other Functions Pages 2-23 24-40 41-47 48-56 57-59 60-68 69-83

More information

ThermaViz. Operating Manual. The Innovative Two-Wavelength Imaging Pyrometer

ThermaViz. Operating Manual. The Innovative Two-Wavelength Imaging Pyrometer ThermaViz The Innovative Two-Wavelength Imaging Pyrometer Operating Manual The integration of advanced optical diagnostics and intelligent materials processing for temperature measurement and process control.

More information

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling Zeiss LSM 510 Confocor III Training Notes Center for Cell Analysis & Modeling Confocor 3 Start Up Go to System Module Turn on Main Switch, System/ PC, and Components Switches Do you need the arc lamp?

More information

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE by Doug Bray Department of Biological Sciences University of Lethbridge Revised September, 2000 Note: The terms in bold in this document represent

More information

Practical work no. 3: Confocal Live Cell Microscopy

Practical work no. 3: Confocal Live Cell Microscopy Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside

More information

Ph 3455 The Franck-Hertz Experiment

Ph 3455 The Franck-Hertz Experiment Ph 3455 The Franck-Hertz Experiment Required background reading Tipler, Llewellyn, section 4-5 Prelab Questions 1. In this experiment, we will be using neon rather than mercury as described in the textbook.

More information

LEICA TCS SP5 AOBS TANDEM USER MANUAL

LEICA TCS SP5 AOBS TANDEM USER MANUAL LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE

More information

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the ECEN 4606 Lab 8 Spectroscopy SUMMARY: ROBLEM 1: Pedrotti 3 12-10. In this lab, you will design, build and test an optical spectrum analyzer and use it for both absorption and emission spectroscopy. The

More information

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5. Smith College August 2010 Operating Checklist for using the Laser Scanning Confocal Microscope Leica TCS SP5. CONTENT, page no. Startup, 1 Initial set-up, 1 Software, 2 Microscope Specimen observation

More information

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY References: http://virlab.virginia.edu/vl/home.htm (University of Virginia virtual lab. Click on the AFM link) An atomic force

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left

More information

Olympus Digital Microscope Camera (DP70) checklist

Olympus Digital Microscope Camera (DP70) checklist Smith College - July 2005 Olympus Digital Microscope Camera (DP70) checklist CONTENT, page no. Camera Information, 1 Startup, 1 Retrieve an Image, 2 Microscope Setup, 2 Capture, 3 Preview. 3 Color Balans,

More information

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000 The ideal K-12 science microscope solution User Guide for use with the Nova5000 NovaScope User Guide Information in this document is subject to change without notice. 2009 Fourier Systems Ltd. All rights

More information

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2:

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2: Image acquisition, managing and processing software TCapture Instruction Manual Key to the Instruction Manual TC is shortened name used for TCapture. Help Refer to [Help] >> [About TCapture] menu for software

More information

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 1. Utility Requirements a. System power is supplied by two 120 VAC/20 A circuits. When doing maintenance

More information

CCD User s Guide SBIG ST7E CCD camera and Macintosh ibook control computer with Meade flip mirror assembly mounted on LX200

CCD User s Guide SBIG ST7E CCD camera and Macintosh ibook control computer with Meade flip mirror assembly mounted on LX200 Massachusetts Institute of Technology Department of Earth, Atmospheric, and Planetary Sciences Handout 8 /week of 2002 March 18 12.409 Hands-On Astronomy, Spring 2002 CCD User s Guide SBIG ST7E CCD camera

More information

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Rotation By: Michael Case and Roy Grayzel, Acton Research Corporation Introduction The majority of modern spectrographs and scanning

More information

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400. Smith College August 2009 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check 1 Startup 1 Specimen Insertion 2 Filament Saturation 2 Beam Alignment

More information

Use of the HSW5 Spinning Disk Confocal Microscope Updated last May 25, 2010 OK

Use of the HSW5 Spinning Disk Confocal Microscope Updated last May 25, 2010 OK Use of the HSW5 Spinning Disk Confocal Microscope Updated last May 25, 2010 OK Getting Started: 2 Starting Micromanager and Loading a Configuration 3 The Main Micromanager GUI 3 Configuration Settings

More information

Basic Users Manual for Tecnai-F20 TEM

Basic Users Manual for Tecnai-F20 TEM Basic Users Manual for Tecnai-F20 TEM NB: This document contains my personal notes on the operating procedure of the Tecnai F20 and may be used as a rough guide for those new to the microscope. It may

More information

Operating the CCD Camera

Operating the CCD Camera Operating the CCD Camera 1995 Edition Incorporates ccd software for disk storage This eliminates problems with cc200 software 1 Setting Up Very little setup is required; the camera and its electronics

More information

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right Quick Guide to Operating FEI Titan Themis G2 200 (S)TEM: TEM mode Susheng Tan Nanoscale Fabrication and Characterization Facility, University of Pittsburgh Office: M104/B01 Benedum Hall, 412-383-5978,

More information

Goodman Cookbook. Goodman Spectrograph. Adapted by D. Sanmartim from L. Fraga's Guide. Sep SOAR Telescope

Goodman Cookbook. Goodman Spectrograph. Adapted by D. Sanmartim from L. Fraga's Guide. Sep SOAR Telescope Goodman Spectrograph 1 Goodman Spectrograph Documentation Goodman HTS Manual http://www.ctio.noao.edu/soar/content/goodman-hts-manual Goodman Overview http://www.ctio.noao.edu/soar/content/goodman-spectrograph-overview

More information

Before you start, make sure that you have a properly calibrated system to obtain high-quality images.

Before you start, make sure that you have a properly calibrated system to obtain high-quality images. CONTENT Step 1: Optimizing your Workspace for Acquisition... 1 Step 2: Tracing the Region of Interest... 2 Step 3: Camera (& Multichannel) Settings... 3 Step 4: Acquiring a Background Image (Brightfield)...

More information

User Operation of JEOL 1200 EX II

User Operation of JEOL 1200 EX II **Log onto Computer** Open item program Start Up Procedure User Operation of JEOL 1200 EX II 1. If scope is not running, locate an electron microscopy technician (EMT) to find out why not. 2. Turn up brightness

More information

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center LEO 912 TEM Short Manual Prepared/copyrighted by RH Berg Danforth Plant Science Center Specimen holder [1] Never touch the holder (outside of the O-ring, double-headed arrow) because finger oils will contaminate

More information

LSM 510 Training Notes

LSM 510 Training Notes LSM 510 Training Notes Turning on the system Turn on the arc lamp, found on the bench top left of the microscope. This supplies light for epifluorescence for viewing your samples through the microscope.

More information

Oriel MS260i TM 1/4 m Imaging Spectrograph

Oriel MS260i TM 1/4 m Imaging Spectrograph Oriel MS260i TM 1/4 m Imaging Spectrograph MS260i Spectrograph with 3 Track Fiber on input and InstaSpec CCD on output. The MS260i 1 4 m Imaging Spectrographs are economical, fully automated, multi-grating

More information

Leica Sp5 II Confocal User Guide

Leica Sp5 II Confocal User Guide Leica Sp5 II Confocal User Guide Turning on the Confocal System (instructions are posted in the room) 1. Turn on Laser Power Button 2. Turn Key to On position 3. Turn on Scanner Power Button 4. Turn on

More information

Instructions for the Experiment

Instructions for the Experiment Instructions for the Experiment Excitonic States in Atomically Thin Semiconductors 1. Introduction Alongside with electrical measurements, optical measurements are an indispensable tool for the study of

More information

Kit for building your own THz Time-Domain Spectrometer

Kit for building your own THz Time-Domain Spectrometer Kit for building your own THz Time-Domain Spectrometer 16/06/2016 1 Table of contents 0. Parts for the THz Kit... 3 1. Delay line... 4 2. Pulse generator and lock-in detector... 5 3. THz antennas... 6

More information

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only Version 1.0 Prepared by D. Turnbull February 21, 2007. Please submit any omissions to the Author Note: This SEM is a recent

More information

SlideBook 5. FRAP Imaging Module

SlideBook 5. FRAP Imaging Module SlideBook 5 FRAP Imaging Module for Windows XP User s Manual (Latest Revision 7.9.09) 1 Table of Contents TABLE OF CONTENTS...2 CONTACTING INTELLIGENT IMAGING INNOVATIONS, INC...3 MANUAL CONVENTIONS...4

More information

MS260i 1/4 M IMAGING SPECTROGRAPHS

MS260i 1/4 M IMAGING SPECTROGRAPHS MS260i 1/4 M IMAGING SPECTROGRAPHS ENTRANCE EXIT MS260i Spectrograph with 3 Track Fiber on input and InstaSpec IV CCD on output. Fig. 1 OPTICAL CONFIGURATION High resolution Up to three gratings, with

More information

Chlorophyll Fluorescence Imaging System

Chlorophyll Fluorescence Imaging System Quick Start Guide Chlorophyll Fluorescence Imaging System Quick Start Guide for Technologica FluorImager software for use with Technlogica CFImager hardware Copyright 2006 2015 TECHNOLOGICA LIMITED. All

More information

FLAMINGOS at the KPNO 2.1-m

FLAMINGOS at the KPNO 2.1-m FLAMINGOS at the KPNO 2.1-m Telescope Console Control Panels & GUIs used for Guiding Nick Raines & Richard Elston Version 0.1, 2003 October 21 FLAMINGOS at the 2.1-m: Guider Controls Page 1 of 10 Introduction

More information

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE University of Virginia ITC Academic Computing Health Sciences CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE Introduction The Olympus BH-2 microscope in ACHS s microscope lab has objectives from

More information

ScanArray Overview. Principle of Operation. Instrument Components

ScanArray Overview. Principle of Operation. Instrument Components ScanArray Overview The GSI Lumonics ScanArrayÒ Microarray Analysis System is a scanning laser confocal fluorescence microscope that is used to determine the fluorescence intensity of a two-dimensional

More information

VATTSpec Instructions Rev. 10/23/2015

VATTSpec Instructions Rev. 10/23/2015 VATTSpec Instructions Rev. 10/23/2015 Introduction VATTSpec is a medium resolution CCD range spectrograph with a skinny chip having excellent cosmetics. Its UA ITL chip, Serial Number 8228, has a gain

More information

LumaSpec 800S User Manual

LumaSpec 800S User Manual LumaSpec 800S User Manual Worldwide distribution VERSION 09112014 Prior Scientific, Ltd Cambridge, UK Prior Scientific, Inc Rockland, MA. USA Prior Scientific, GmbH Jena, Germany Prior Scientific KK Tokyo,

More information

STEM Spectrum Imaging Tutorial

STEM Spectrum Imaging Tutorial STEM Spectrum Imaging Tutorial Gatan, Inc. 5933 Coronado Lane, Pleasanton, CA 94588 Tel: (925) 463-0200 Fax: (925) 463-0204 April 2001 Contents 1 Introduction 1.1 What is Spectrum Imaging? 2 Hardware 3

More information

TriVista. Universal Raman Solution

TriVista. Universal Raman Solution TriVista Universal Raman Solution Why choose the Princeton Instruments/Acton TriVista? Overview Raman Spectroscopy systems can be derived from several dispersive components depending on the level of performance

More information

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack). FEI Tecnai F20 S/TEM: imaging in TEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/21/18 1. Filling the cold trap (if needed) 1.1. Prior to use, the cold trap needs

More information

FE-SEM SU-8020 Operating manual (Preliminary version)

FE-SEM SU-8020 Operating manual (Preliminary version) FE-SEM SU-8020 Operating manual (Preliminary version) 2016/04/11 Seimitsu Bunseki sitsu lab. Starting up 1.Turn on the Display switch. Windows OS is starting up 2. Select the user SU-8000. 3. Click the

More information

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017. Nikon AZ100 Laser Scanning Macro Confocal Microscope Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin May 2017 Contents 1 Introduction 2 2 Hardware - Startup 2 3 Software/Operation 4 3.1 Multidimensional

More information

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton

More information

Bruker Dimension Icon AFM Quick User s Guide

Bruker Dimension Icon AFM Quick User s Guide Bruker Dimension Icon AFM Quick User s Guide August 8 2014 GLA Contacts Jingjing Jiang (jjiang2@caltech.edu 626-616-6357) Xinghao Zhou (xzzhou@caltech.edu 626-375-0855) Bruker Tech Support (AFMSupport@bruker-nano.com

More information

User manual for Olympus SD-OSR spinning disk confocal microscope

User manual for Olympus SD-OSR spinning disk confocal microscope User manual for Olympus SD-OSR spinning disk confocal microscope Ved Prakash, PhD. Research imaging specialist Imaging & histology core University of Texas, Dallas ved.prakash@utdallas.edu Once you open

More information

Spectroscopy of Ruby Fluorescence Physics Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018

Spectroscopy of Ruby Fluorescence Physics Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018 1 Spectroscopy of Ruby Fluorescence Physics 3600 - Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018 I. INTRODUCTION The laser was invented in May 1960 by Theodor Maiman.

More information

This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface.

This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface. Tescan MIRA3 SEM: EDS using EDAX TEAM Nicholas G. Rudawski ngr@ufl.edu Cell: (805) 252-4916 Office: (352) 392-3077 Last updated: 12/04/17 This procedure assumes the user is already familiar with basic

More information

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION Revised November 15, 2017 INTRODUCTION The simplest and most commonly described examples of diffraction and interference from two-dimensional apertures

More information

Unit 8: Light and Optics

Unit 8: Light and Optics Objectives Unit 8: Light and Optics Explain why we see colors as combinations of three primary colors. Explain the dispersion of light by a prism. Understand how lenses and mirrors work. Explain thermal

More information

Supplementary Materials

Supplementary Materials Supplementary Materials In the supplementary materials of this paper we discuss some practical consideration for alignment of optical components to help unexperienced users to achieve a high performance

More information

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0 Instruction manual for T3DS software Release 4.0 Table of contents 0. Setup... 3 1. Start-up... 5 2. Input parameters and delay line control... 6 3. Slow scan measurement... 8 4. Fast scan measurement...

More information

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course) JEOL 7000F BASIC OPERATING INSTRUCTIONS-Ver.-2.0 Note: This is minimal operation checklist and does not replace the other reference manuals. Read the manual for Specimen Exchange (JEOL 7000 Specimen Exchange

More information

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer Brightfield Microscopy and Image Acquisition on Spotcam1 by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer Log onto the computer. Enter your username and password to log onto the server.

More information

FluorCam PAR- Absorptivity Module & NDVI Measurement

FluorCam PAR- Absorptivity Module & NDVI Measurement FluorCam PAR- Absorptivity Module & NDVI Measurement Instruction Manual Please read this manual before operating this product P PSI, spol. s r. o., Drásov 470, 664 24 Drásov, Czech Republic FAX: +420 511

More information

CNC Using the FlexiCam CNC and HMI Software. Guldbergsgade 29N, P0 E: T:

CNC Using the FlexiCam CNC and HMI Software. Guldbergsgade 29N, P0 E: T: CNC Using the FlexiCam CNC and HMI Software Guldbergsgade 29N, P0 E: makerlab@kea.dk T: +46 46 03 90 This grey box is the NC controller. Let s start by turning the red switch to the ON position, then press

More information

LSM 510 Meta Training Notes

LSM 510 Meta Training Notes LSM 510 Meta Training Notes Turning on the system Turn on X-Cite power supply. This supplies light for epifluorescence for viewing your samples through the microscope. Turn on the remote control switch.

More information

Olympus xcellence Software - basic user guide

Olympus xcellence Software - basic user guide Olympus xcellence Software - basic user guide This is a basic overview of setting up time lapse experiments using Olympus's xcellence software on BIU's IX81 inverted phase contrast system - the software

More information

Be aware that there is no universal notation for the various quantities.

Be aware that there is no universal notation for the various quantities. Fourier Optics v2.4 Ray tracing is limited in its ability to describe optics because it ignores the wave properties of light. Diffraction is needed to explain image spatial resolution and contrast and

More information

EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM

EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM GAIN 1 10 100 Instruction Manual and Experiment Guide for the PASCO scientific Model OS-8537 and OS-8539 012-06575A 3/98 EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM

More information

4.5.1 Mirroring Gain/Offset Registers GPIO CMV Snapshot Control... 14

4.5.1 Mirroring Gain/Offset Registers GPIO CMV Snapshot Control... 14 Thank you for choosing the MityCAM-C8000 from Critical Link. The MityCAM-C8000 MityViewer Quick Start Guide will guide you through the software installation process and the steps to acquire your first

More information

Point Calibration. July 3, 2012

Point Calibration. July 3, 2012 Point Calibration July 3, 2012 The purpose of the Point Calibration process is to generate a map of voltages (for galvos) or motor positions of the pointing device to the voltages or pixels of the reference

More information