TN378: Openlab Module - FRET. Topic. Discussion

Size: px
Start display at page:

Download "TN378: Openlab Module - FRET. Topic. Discussion"

Transcription

1 TN378: Openlab Module - FRET Topic This technical note describes the use of the Openlab FRET module in Openlab and higher. Users of Openlab Server systems will require Openlab Server or higher on the server machine and Openlab or higher on the client machines to take advantage of this module. Discussion FRET (Fluorescence Resonance Energy Transfer) is the non-radiative transfer of energy from an excited fluorophore (donor) to another fluorophore (acceptor). FRET only occurs when the molecules concerned are in very close proximity (1-10nm). This makes it a valuable technique for studies where distances and interactions between molecules are of interest. The Openlab FRET module The Openlab FRET module allows accurate FRET measurements and generation of net FRET (or corrected FRET) and normalised FRET images for publication. The analysis currently offered by the module is the 3-image FRET technique, as described by Gordon et al. (1998) and Xia and Liu (2001). These references should be used to find further information on the technique and guidance on interpretation. Data acquisition Three images are collected for each data point, a Donor image, an Acceptor image, and a FRET image. One typical donor/acceptor pair is CFP & YFP. Thus: Donor image = Excite with CFP filter, CFP dichroic, emit with CFP filter Acceptor image = Excite with YFP filter, YFP dichroic, emit with YFP filter FRET image = Excite with CFP filter, CFP dichroic, emit with YFP filter The Openlab Automator should be used to acquire these images using the appropriate filters. The order of image acquisition is important in analysis, see below. The result is generated by the FRET module by combining the three input images on a pixel-by-pixel basis according to the chosen expression. The output is a floating - point value at each pixel. Bleed-through Reference to the expressions cited indicates that there are two constants required. To generate FRET images, the user requires two values (a and b) which represent the bleed-through or cross-talk of each dye into the FRET filter set. These values are calculated once and then used for all subsequent experiments. Page 1 of 5

2 A, called Donor bleed-through in the interface, is calculated by producing two images from a control sample labelled with the donor only. Each is background subtracted before calculation Donor bleed-through = Intensity of cell ROI in FRET Image / Intensity of cell ROI in Donor Image. B, called Acceptor bleed-through in the interface, is calculated by producing images from a control sample labelled with acceptor only. Each is background subtracted before calculation Acceptor bleed-through = Intensity of cell ROI in FRET Image / Intensity of cell ROI in Acceptor Image. This process is not carried out by the FRET dialog but all the necessary steps could be incorporated into an automation. Bleed-through values are normalised percentages between 0.0 and 1.0 Bleed-through must be recalculated if any change is made to the system which could affect transmission characteristics or sensitivity to wavelengths of light. Background Subtraction and Registration All images must be background-subtracted and in proper register before FRET analysis. The Openlab Core offers background subtraction using an image of background for each channel and the Registration module is available for correcting image translation. The FRET dialog Select one or more sets of input images in the Layers manager. A set of input images must meet the following criteria: A set consists of three images The set must be in order in the Layers manager. Accepted orders are Donor, Fret, Acceptor, Donor, Fret, Acceptor, or for minimum filter movement, Donor, Fret, Acceptor, Acceptor, Fret, Donor, Donor, Fret Images must be grayscale between 8 and 16 bit. All images in the set must be the same depth. All images must be General Image Layers. Page 2 of 5

3 Choose FRET from the Special menu Select the method of FRET calculation to be used. 2. Select the order in which the input images are presented in the layers manager. 3. Enter the Bleed through constants obtained from negative controls. 4. Enter values for thresholds in each image. This is to improve contrast between background and cells by setting background to Preview the output image. Zoom and pan the preview image. 6. Choose the set of input images from which the preview is generated. If there is more than one set of input images selected each will be listed in this pop-up. 7. Select the color look up table to be applied to the output image. A rainbow CLUT is the one most commonly used. 8. Select the distribution of the look up table. 9. Chose to generate an intensity modulated image in addition to the FRET output image by selecting a modulator image. 10. Choose the output range, the FRET values which will be displayed using the look up table. Click Auto to prompt the module to scan all selected images and find the maximum range of FRET values and set these as the output range. 11. Click OK to apply FRET calculation to all selected sets of input images and create output images. The FRET dialog will retain previously entered settings when closed Page 3 of 5

4 FRET Automator Task The FRET module will add a new task to the automator allowing batch processing of selected layers or generation of FRET images on line, as each set of input images is acquired. In the same way as for the FRET dialog, values for all parameters can be entered into the dialog or brought from variables. Alternatively the automator task can bring the last settings used in the main FRET dialog. This allows the user to set up the dialog as required, preview the results and automatically include them in an automation. Measurements from FRET result images The resultant images generated by the FRET dialog or automator task are density calibrated. Use the calibration bar stamp on the Paint Tools palette to show the net FRET values represented by the look up table. Use Openlab Measurements and Advanced Measurements to obtain values from regions of interest. In the measurements table the columns Min(Cal), Max(Cal), Mean (Cal) and Mode(Cal) will show the result of the chosen FRET calculation for that region Additional guidelines How do I obtain input threshold values? Use an intensity value obtained from outside an area of cells in each type of image. Input thresholds are necessary even though input images should be background subtracted. Determine input thresholds for each type of image using the HSI colorspy Page 4 of 5

5 What is an intensity modulated image? The IM image uses pseudocolor and intensity to display a FRET output image. Color is used to represent FRET values, intensity represents the intensity of one of the input images. This technique helps reflect spatial information since the intensity of the original images is reflected in the resultant image. To produce the IM image the intensity of each pixel of the FRET output image is scaled based on the range of intensities in the input images. The source image used for this is called the modulator image. The modulator image can be one of the source images, the brightest of the source images or the average of the source images. The range of colors in the IM image is determined by the look up table applied to the FRET output image. The range of intensities in the IM image is determined by the range of intensities in the modulator image. The IM image is a millions of colors image. It is for visualization. Troubleshooting Why is the FRET option under the special menu greyed out? Images have not been selected in the layers manager. A set of input images consists of 3 layers so 3 or more must be selected. The selected layers are not all General Image layers, or they are not 8-16 bit grayscale or they are not all the same depth. View the properties of the layers and change the selection as appropriate. References Gerald W. Gordon, Gail Berry, Xiao Huan Liang, Beth Levine, and Brian Herman Quantitative Fluorescence Resonance Energy Transfer Measurements Using Fluorescence Microscopy. Biophysical Journal : Zongping Xia and Yeuchueng Liu. Reliable and Global Measurement of Fluorescence Energy Transfer Using Fluorescence Microscopes. Biophysical Journal : Page 5 of 5

Advanced Live Cell Imaging

Advanced Live Cell Imaging FRET Analysis in Laser Scanning Microscopy What is FRET? FRET (fluorescence resonance energy transfer) is the non-radiative transfer of photon energy from an excited fluorophore (the donor) to another

More information

Shreyash Tandon M.S. III Year

Shreyash Tandon M.S. III Year Shreyash Tandon M.S. III Year 20091015 Confocal microscopy is a powerful tool for generating high-resolution images and 3-D reconstructions of a specimen by using point illumination and a spatial pinhole

More information

ZEISS LSM510META confocal manual

ZEISS LSM510META confocal manual ZEISS LSM510META confocal manual Switching on the system 1) Switch on the Remote Control button located on the table to the right of the microscope. This is the main switch for the whole system including

More information

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005 Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev Why use confocal microscopy? Principles of the laser scanning confocal microscope. Image resolution. Manipulating the

More information

April 2009 No.04 WIDEFIELD APPLICATION LETTER. resolution. FRET Sensitized Emission Wizard Widefield

April 2009 No.04 WIDEFIELD APPLICATION LETTER. resolution. FRET Sensitized Emission Wizard Widefield April 2009 No.04 WIDEFIELD APPLICATION LETTER resolution FRET Sensitized Emission Wizard Widefield FRET SE with the Leica Advanced Widefield Systems AF7000, AF6500 and AF6000 FRET Sensitized Emission (FRET

More information

Leica SP8 Resonant Confocal. Quick-Start Guide

Leica SP8 Resonant Confocal. Quick-Start Guide Leica SP8 Resonant Confocal Quick-Start Guide Contents Start-up Preparing for Imaging Part 1 On the scope Part 2 Software interface Part 3 Heat & CO2 incubation Part 4 Other hardware options Shut-down

More information

Calculate Ratiometric FRET-Images with the FRET-Image-Script

Calculate Ratiometric FRET-Images with the FRET-Image-Script Tutorial Calculate Ratiometric FRET-Images with the FRET-Image-Script Summary This tutorial shows step-by-step, how the "FRET Image" script of SymPhoTime 64 can be used to calculate pixel-by-pixel the

More information

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION Olympus VS120-L100 Slide Scanner Standard Operating Procedure Startup 1) Red power bar switch (behind monitor) 2) Computer 3) Login: UserVS120 account (no password) 4) Double click: WAIT FOR INITIALIZATION

More information

Confocal Microscopy. Kristin Jensen

Confocal Microscopy. Kristin Jensen Confocal Microscopy Kristin Jensen 17.11.05 References Cell Biological Applications of Confocal Microscopy, Brian Matsumoto, chapter 1 Studying protein dynamics in living cells,, Jennifer Lippincott-Schwartz

More information

Multifluorescence The Crosstalk Problem and Its Solution

Multifluorescence The Crosstalk Problem and Its Solution Multifluorescence The Crosstalk Problem and Its Solution If a specimen is labeled with more than one fluorochrome, each image channel should only show the emission signal of one of them. If, in a specimen

More information

IncuCyte ZOOM Scratch Wound Processing Overview

IncuCyte ZOOM Scratch Wound Processing Overview IncuCyte ZOOM Scratch Wound Processing Overview The IncuCyte ZOOM Scratch Wound assay utilizes the WoundMaker-IncuCyte ZOOM-ImageLock Plate system to analyze both 2D-migration and 3D-invasion in label-free,

More information

Cellular Bioengineering Boot Camp. Image Analysis

Cellular Bioengineering Boot Camp. Image Analysis Cellular Bioengineering Boot Camp Image Analysis Overview of the Lab Exercises Microscopy and Cellular Imaging The purpose of this laboratory exercise is to develop an understanding of the measurements

More information

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the

More information

In essence this means, that a certain proportion of a signal in one channel is actually derived from another dye spilling over into the channel.

In essence this means, that a certain proportion of a signal in one channel is actually derived from another dye spilling over into the channel. NNKQ táçéñáéäç=jìäíáåü~ååéä=råãáñáåö= _~ÅâÖêçìåÇ= Widefield Multichannel Unmixing is a new function for the removal of crosstalk between fluorescent dyes in multichannel images with up to eight fluorescence

More information

ScanArray Overview. Principle of Operation. Instrument Components

ScanArray Overview. Principle of Operation. Instrument Components ScanArray Overview The GSI Lumonics ScanArrayÒ Microarray Analysis System is a scanning laser confocal fluorescence microscope that is used to determine the fluorescence intensity of a two-dimensional

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and

More information

IncuCyte ZOOM Fluorescent Processing Overview

IncuCyte ZOOM Fluorescent Processing Overview IncuCyte ZOOM Fluorescent Processing Overview The IncuCyte ZOOM offers users the ability to acquire HD phase as well as dual wavelength fluorescent images of living cells producing multiplexed data that

More information

Last updated: May 2014 Y.DeGraaf

Last updated: May 2014 Y.DeGraaf FLINDERS MICROSCOPY BIOMEDICAL SERVICES AVAILABLE MICROSCOPES AND SPECIFICATIONS & INFORMATION REGARDING TRAINING FOR NEW USERS Last updated: May 2014 Y.DeGraaf If you have new staff or students (Honours/Masters

More information

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL MIF ZEISS LSM510 CONFOCAL USER PROTOCOL START-UP Turn on the Mercury Bulb Power Supply (if needed). Power-on the Control Box. Turn on the computer. Open the LSM 510 software. Choose Scan New Images and

More information

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling Zeiss LSM 510 Confocor III Training Notes Center for Cell Analysis & Modeling Confocor 3 Start Up Go to System Module Turn on Main Switch, System/ PC, and Components Switches Do you need the arc lamp?

More information

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide Operation 58. UltraVIEW VoX Getting Started Guide Overview This chapter outlines the basic methods used to operate the UltraVIEW VoX system. About other software Volocity places great demands on the computer

More information

Before you start, make sure that you have a properly calibrated system to obtain high-quality images.

Before you start, make sure that you have a properly calibrated system to obtain high-quality images. CONTENT Step 1: Optimizing your Workspace for Acquisition... 1 Step 2: Tracing the Region of Interest... 2 Step 3: Camera (& Multichannel) Settings... 3 Step 4: Acquiring a Background Image (Brightfield)...

More information

Locating Molecules Using GSD Technology Project Folders: Organization of Experiment Files...1

Locating Molecules Using GSD Technology Project Folders: Organization of Experiment Files...1 .....................................1 1 Project Folders: Organization of Experiment Files.................................1 2 Steps........................................................................2

More information

Dr. Bob on Colocalization or MSL Experiments In Learning Colocalization Using Image J

Dr. Bob on Colocalization or MSL Experiments In Learning Colocalization Using Image J Dr. Bob on Colocalization or MSL Experiments In Learning Colocalization Using Image J Confocal microscopy is used to test whether two fluorescently labeled molecules are associated with one another. If

More information

Zeiss 880 Training Notes Zen 2.3

Zeiss 880 Training Notes Zen 2.3 Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start

More information

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope ZEN 2009 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn Chameleon TiS laser key from Standby

More information

LSM 800 Confocal Microscope Standard Operation Protocol

LSM 800 Confocal Microscope Standard Operation Protocol LSM 800 Confocal Microscope Standard Operation Protocol Turning on the system 1. Switch on the Main switch (labeled 1 and 2 ) mounted on the wall. 2. Turn the Laser Key (labeled 3 ) 90 clockwise for power

More information

Leica SPEII confocal microscope. Short Manual

Leica SPEII confocal microscope. Short Manual Leica SPEII confocal microscope Short Manual Switching ON sequence: 1. Turn on the Workstation under the bench (top, far right). 2. Turn on the Supply Unit - Laser box (big green switch first and then

More information

Things to check before start-up.

Things to check before start-up. Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your

More information

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Training Guide for Leica SP8 Confocal/Multiphoton Microscope Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

Swept-Field User Guide

Swept-Field User Guide Swept-Field User Guide Note: for more details see the Prairie user manual at http://www.prairietechnologies.com/resources/software/prairieview.html Please report any problems to Julie Last (jalast@wisc.edu)

More information

Pattern Matching. Open an Image

Pattern Matching. Open an Image Pattern Matching Pattern Matching Analysis by decomposing a recorded image into different user-defined patterns. Display of the calculated data using an RGB false color model. The calculated amplitudes

More information

MetaMorph Imaging Handbook Update 6/4/13

MetaMorph Imaging Handbook Update 6/4/13 MetaMorph Imaging Handbook Update 6/4/13 Startup FIRST turn on mercury lamp (Fluorescence) Computer and monitor Qimaging Camera (on top) Uniblitz Shutters-2 Halogen Lamp (Transmitted Light) Computer Login

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

Image Analysis for Fluorescence

Image Analysis for Fluorescence Image Analysis for Fluorescence Terminology Table Image Analysis Macro Colocalization Intensity Dye AFI The extraction of meaningful information from digital images by means of digital image processing

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front

More information

(Quantitative Imaging for) Colocalisation Analysis

(Quantitative Imaging for) Colocalisation Analysis (Quantitative Imaging for) Colocalisation Analysis or Why Colour Merge / Overlay Images are EVIL! Special course for DIGS-BB PhD program What is an Image anyway..? An image is a representation of reality

More information

TEMScripts Auto Image Measurement (Particle) Manual. TEMScripts LLC. Last updated: 12/5/2016

TEMScripts Auto Image Measurement (Particle) Manual. TEMScripts LLC. Last updated: 12/5/2016 TEMScripts Auto Image Measurement (Particle) Manual TEMScripts LLC. Last updated: 12/5/2016 TS Auto Image Measurement- Particle Installation Close Digital Micrograph Copy following files to \\Gatan\DigitalMicrograph\PlugIns

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power

More information

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope ZEN 2012 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 If you require fluorescence imaging,

More information

FLIM on a wide field fluorescence microscope

FLIM on a wide field fluorescence microscope FLIM on a wide field fluorescence microscope L.K. van Geest, K.W.J. Stoop Lambert Instruments, Turfweg 4, 933TH Leutingewolde, The Netherlands. Phone: +3 50 50846, Fax: +3 50 500034, e-mail: lkvgeest@lambert-instruments.com

More information

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON N-SIM guide NIKON IMAGING CENTRE @ KING S COLLEGE LONDON Starting-up / Shut-down The NSIM hardware is calibrated after system warm-up occurs. It is recommended that you turn-on the system for at least

More information

Illumination Correction tutorial

Illumination Correction tutorial Illumination Correction tutorial I. Introduction The Correct Illumination Calculate and Correct Illumination Apply modules are intended to compensate for the non uniformities in illumination often present

More information

LSM 510 Meta Training Notes

LSM 510 Meta Training Notes LSM 510 Meta Training Notes Turning on the system Turn on X-Cite power supply. This supplies light for epifluorescence for viewing your samples through the microscope. Turn on the remote control switch.

More information

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope Training Guide for Carl Zeiss LSM 510 META Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON Components and System/PC switches

More information

Practical work no. 3: Confocal Live Cell Microscopy

Practical work no. 3: Confocal Live Cell Microscopy Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside

More information

(12) United States Patent (10) Patent No.: US 6,388,807 B1. Knebel et al. (45) Date of Patent: May 14, 2002

(12) United States Patent (10) Patent No.: US 6,388,807 B1. Knebel et al. (45) Date of Patent: May 14, 2002 USOO6388807B1 (12) United States Patent (10) Patent No.: Knebel et al. () Date of Patent: May 14, 2002 (54) CONFOCAL LASER SCANNING (56) References Cited MICROSCOPE U.S. PATENT DOCUMENTS (75) Inventors:

More information

Nasmyth Ultraview Vox User Protocol

Nasmyth Ultraview Vox User Protocol Nasmyth Ultraview Vox User Protocol Switch on all wall sockets labelled Nasmyth, switch camera on (power supply located on table behind monitor), switch on laser switch in laser rack, switch computer on

More information

User manual for Olympus SD-OSR spinning disk confocal microscope

User manual for Olympus SD-OSR spinning disk confocal microscope User manual for Olympus SD-OSR spinning disk confocal microscope Ved Prakash, PhD. Research imaging specialist Imaging & histology core University of Texas, Dallas ved.prakash@utdallas.edu Once you open

More information

3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit.

3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit. 1 Set up the confocal light path for imaging a green dye (e.g. Alexa488-EGFP). For example, under the Configuration Control window the light path could be set up as shown here using the 488 nm LASER (found

More information

Using the Nikon TE2000 Inverted Microscope

Using the Nikon TE2000 Inverted Microscope Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core Using the Nikon TE2000 Inverted Microscope Fluorescence image acquisition using Scanalytic s IPLab software and the B&W

More information

SlideBook 5. FRAP Imaging Module

SlideBook 5. FRAP Imaging Module SlideBook 5 FRAP Imaging Module for Windows XP User s Manual (Latest Revision 7.9.09) 1 Table of Contents TABLE OF CONTENTS...2 CONTACTING INTELLIGENT IMAGING INNOVATIONS, INC...3 MANUAL CONVENTIONS...4

More information

Advanced Technology and Manufacturing Institute. Zygo ZeScope

Advanced Technology and Manufacturing Institute. Zygo ZeScope Advanced Technology and Manufacturing Institute Zygo ZeScope Created by Andrew Miller ATAMI Oregon State University Revision Date Description Curator 0 8/31/2018 New Document Andrew Miller Zygo ZeScope

More information

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy, KTH Applied Physics Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy, 2009-06-05, 8-13, FB51 Allowed aids: Compendium Imaging Physics (handed out) Compendium Light Microscopy

More information

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017. Nikon AZ100 Laser Scanning Macro Confocal Microscope Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin May 2017 Contents 1 Introduction 2 2 Hardware - Startup 2 3 Software/Operation 4 3.1 Multidimensional

More information

Pixel shift in fluorescence microscopy

Pixel shift in fluorescence microscopy Pixel shift in fluorescence microscopy 1. Introduction Multicolor imaging in fluorescence microscopy is typically performed by sequentially acquiring images of different colors. An overlay of these images

More information

1 Co Localization and Working flow with the lsm700

1 Co Localization and Working flow with the lsm700 1 Co Localization and Working flow with the lsm700 Samples -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ BrDU with alexa 488. -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ no BrDU (but with

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Bi Imaging. Multicolor Imaging: The Important Question of Co-Localization. Anna Smallcombe Bio-Rad Laboratories, Hemel Hempstead, UK

Bi Imaging. Multicolor Imaging: The Important Question of Co-Localization. Anna Smallcombe Bio-Rad Laboratories, Hemel Hempstead, UK Multicolor Imaging: The Important Question of Co-Localization Anna Smallcombe Bio-Rad Laboratories, Hemel Hempstead, UK The use of specific fluorescent probes, combined with confocal or multiphoton microscopy

More information

Leica Sp5 II Confocal User Guide

Leica Sp5 II Confocal User Guide Leica Sp5 II Confocal User Guide Turning on the Confocal System (instructions are posted in the room) 1. Turn on Laser Power Button 2. Turn Key to On position 3. Turn on Scanner Power Button 4. Turn on

More information

Widefield 1. Switching on

Widefield 1. Switching on Widefield 1 Switching on 1. Ignite DG5 lamp - must be switched on first (if previous user has switched off, wait 30 min before igniting) 2. Wait 5s and then turn on the main DG5 controller switch. 3. DG5

More information

Supplemental Reference Guide

Supplemental Reference Guide Supplemental Reference Guide QuantiGene ViewRNA mirna ISH Cell Assay P/N 19167 Rev.A 120623 For research use only. Not for use in diagnostic procedures. Trademarks Affymetrix and, and QuantiGene are trademarks

More information

We attempted to separate the two dyes by acquiring images using a single excitation wavelength and just two emission wavelengths.

We attempted to separate the two dyes by acquiring images using a single excitation wavelength and just two emission wavelengths. TN437: Spectral Separation of monochrome images using Volocity 4.0 Introduction Spectral Separation is a technique that allows the user to separate images containing data from more than one fluorochrome

More information

Fundamentals of Light Microscopy II: Fluorescence, Deconvolution, Confocal, Multiphoton, Spectral microscopy. Integrated Microscopy Course

Fundamentals of Light Microscopy II: Fluorescence, Deconvolution, Confocal, Multiphoton, Spectral microscopy. Integrated Microscopy Course Fundamentals of Light Microscopy II: Fluorescence, Deconvolution, Confocal, Multiphoton, Spectral microscopy Integrated Microscopy Course Review Lecture 1: Microscopy Basics Light train Kohler illumination*

More information

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 Leica DMI AF6000LX Table of contents START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 ACQUIRE MODULE 6 SETTING THE LIGHTPATH 6

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s LSM 5 MP, LSM 510 and LSM 510 META M i c r o s c o p y f r o m C a r l Z e i s s Quick Guide Laser Scanning Microscopes LSM Software ZEN 2007 August 2007 We make it visible. Contents Page Contents... 1

More information

LSM 510 Training Notes

LSM 510 Training Notes LSM 510 Training Notes Turning on the system Turn on the arc lamp, found on the bench top left of the microscope. This supplies light for epifluorescence for viewing your samples through the microscope.

More information

Operating Instructions for Zeiss LSM 510

Operating Instructions for Zeiss LSM 510 Operating Instructions for Zeiss LSM 510 Location: GNL 6.312q (BSL3) Questions? Contact: Maxim Ivannikov, maivanni@utmb.edu 1 Attend A Complementary Training Before Using The Microscope All future users

More information

LEICA TCS SP5 AOBS TANDEM USER MANUAL

LEICA TCS SP5 AOBS TANDEM USER MANUAL LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE

More information

WHITE PAPER FAST PROTEIN INTERACTION BINDING CURVES WITH INO S F-HS CONFOCAL MICROSCOPE

WHITE PAPER FAST PROTEIN INTERACTION BINDING CURVES WITH INO S F-HS CONFOCAL MICROSCOPE WHITE PAPER FAST PROTEIN INTERACTION BINDING CURVES WITH INO S F-HS CONFOCAL MICROSCOPE Christian Tardif, Jean-Pierre Bouchard Pascal Gallant, Sebastien Roy, Ozzy Mermut September 2017 Introduction Protein-protein

More information

Quick Start Guide. Leica SP5 X

Quick Start Guide. Leica SP5 X Quick Start Guide Leica SP5 X Please note: Some of the information in this guide was taken from Leica Microsystems Leica TCS SP5 LAS AF Guide for New Users. This work is licensed under the Creative Commons

More information

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer Brightfield Microscopy and Image Acquisition on Spotcam1 by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer Log onto the computer. Enter your username and password to log onto the server.

More information

rainstorm User Guide STORM/PALM Image Processing Software

rainstorm User Guide STORM/PALM Image Processing Software rainstorm User Guide STORM/PALM Image Processing Software Eric Rees, Clemens Kaminski, Miklos Erdelyi, Dan Metcalf, Alex Knight Laser Analytics Group, University of Cambridge & Biotechnology Group, National

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Leica SP8 TCS Users Manual Follow the procedure for start up and log on as posted in the lab. Please log on with your account only and do not share your password with anyone. We track and confirm usage

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS Zeiss LSM 510 M eta Confocal M icroscope OPERATING INSTRUCTIONS Starting the System: 1. Turn the black knob on the laser box one-quarter turn from Off to On. You will hear the laser cooling mechanisms

More information

Microscopy from Carl Zeiss

Microscopy from Carl Zeiss Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path

More information

Volocity Tutorial Creating a Measurement Protocol in Volocity Software

Volocity Tutorial Creating a Measurement Protocol in Volocity Software TUTORIAL NOTE Cellular Imaging and Analysis Volocity Tutorial Creating a Measurement Protocol in Volocity Software This tutorial will illustrate the process of creating a Measurement Protocol in Volocity

More information

Spatial intensity distribution analysis Matlab user guide

Spatial intensity distribution analysis Matlab user guide Spatial intensity distribution analysis Matlab user guide August 2011 Guide on how to use the SpIDA graphical user interface. This little tutorial provides a step by step tutorial explaining how to get

More information

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report Introduction and Background Two-photon microscopy is a type of fluorescence microscopy using two-photon excitation. It

More information

Internal Medicine Imaging Core Emory University Department of Medicine

Internal Medicine Imaging Core Emory University Department of Medicine Internal Medicine Imaging Core Emory University Department of Medicine 1 OPERATION OF THE ZEISS LSM 510 META YOU MUST SIGN UP TO USE THE MICROSCOPE OR COMPUTER EVERY TIME NO EXCEPTIONS Before attempting

More information

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: Confocal imaging on the Leica TCS SP8 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: 4) Do not touch the microscope while the software is initializing. Choose your options: Turn

More information

TRAINING MANUAL. Olympus FV1000

TRAINING MANUAL. Olympus FV1000 TRAINING MANUAL Olympus FV1000 September 2014 TABLE OF CONTENTS A. Start-Up Procedure... 1 B. Visual Observation under the Microscope... 1 C. Image Acquisition... 4 A brief Overview of the Settings...

More information

Opterra II Multipoint Scanning Confocal Microscope. Innovation with Integrity

Opterra II Multipoint Scanning Confocal Microscope. Innovation with Integrity Opterra II Multipoint Scanning Confocal Microscope Enabling 4D Live-Cell Fluorescence Imaging through Speed, Sensitivity, Viability and Simplicity Innovation with Integrity Fluorescence Microscopy The

More information

Hyperspectral User Manual. [Type the author name]

Hyperspectral User Manual. [Type the author name] Hyperspectral User Manual [Type the author name] IMPORTANT: PLEASE READ CAREFULLY Limited Warranty CytoViva warrants for a period of one (1) year from the date of purchase from CytoViva, Inc. or an authorized

More information

Chapter 16: Batch Scanning

Chapter 16: Batch Scanning Chapter 16: Batch Scanning The Batch Scanning module allows users to scan and upload various versions of voter registration forms as a batch into the system. Once they are scanned, documents are available

More information

Guide to Confocal 5. Starting session

Guide to Confocal 5. Starting session Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope

More information

Microscope Confocal LSM510 META

Microscope Confocal LSM510 META Microscope Confocal LSM510 META Welcome to the Zeiss LSM 510 Meta Confocal tutorial. Before using the LSM 510 META, Log off any other computer that is open with your personal login. You will need to put

More information

RENISHAW INVIA RAMAN SPECTROMETER

RENISHAW INVIA RAMAN SPECTROMETER STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples

More information

personal DELTAVISION (pdv)

personal DELTAVISION (pdv) GUIDELINES AND HINTS Version 1.3 (March 2015) personal DELTAVISION (pdv) Epifluorescence microscope from Applied Precision Inc.: The microscope can be found in room 1.320. For details see the architectural

More information

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL START THE SYSTEM... 2 START ZEN SOFTWARE... 3 SET THE TEMPERATURE AND THE CO2 CONTROLLERS... OBSERVATION AT OCULARS... 5 STATIF PRESENTATION... 6 ACQUIRE ONE

More information

Quick Guide. NucleoCounter NC-3000

Quick Guide. NucleoCounter NC-3000 Quick Guide NucleoCounter NC-3000 Table of contents Setting up the FlexiCyte Protocol 2 Editing Image Capture and Analysis Parameters 3 Optimizing Exposure Time 4 Compensation for Spectral Overlap 6 Creating

More information

Camera Test Protocol. Introduction TABLE OF CONTENTS. Camera Test Protocol Technical Note Technical Note

Camera Test Protocol. Introduction TABLE OF CONTENTS. Camera Test Protocol Technical Note Technical Note Technical Note CMOS, EMCCD AND CCD CAMERAS FOR LIFE SCIENCES Camera Test Protocol Introduction The detector is one of the most important components of any microscope system. Accurate detector readings

More information

In-Vivo IMAGING SYSTEMS. A complete line of high resolution optical & X-ray systems for pre-clinical imaging

In-Vivo IMAGING SYSTEMS. A complete line of high resolution optical & X-ray systems for pre-clinical imaging In-Vivo IMAGING SYSTEMS A complete line of high resolution optical & X-ray systems for pre-clinical imaging In-Vivo Imaging Systems Carestream is a strong, successful, multi-billion dollar, international

More information

Benchtop System Quick Start

Benchtop System Quick Start Benchtop System Quick Start Release 5.2 Resonon Inc. Dec 11, 2018 CONTENTS 1 System Overview 1 2 Basic Data Acquisition 3 2.1 Data Modes................................................ 3 2.2 Start The

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left

More information

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each lens with a 1 Airy unit pinhole setting. Many water lenses

More information

Leica_Dye_Finder :53 Uhr Seite 6 Dye Finder LAS AF

Leica_Dye_Finder :53 Uhr Seite 6 Dye Finder LAS AF Dye Finder LAS AF Dye Finder Multicolor live cell fluorescence microscopy is limited by the availability of spectrally separable fluorescent dyes. Fluorescent dyes (or spectral GFP variants) with incongruent

More information

Image and Data Acquisition

Image and Data Acquisition Image and Data Acquisition LCP Image Acquisition Procedures This section provides guidelines for scanning images that will be added to the LCP image archive. By scanning the image, we obtain a digital

More information