Leica_Dye_Finder :53 Uhr Seite 6 Dye Finder LAS AF
|
|
- Clemence Gilbert
- 6 years ago
- Views:
Transcription
1 Dye Finder LAS AF
2 Dye Finder Multicolor live cell fluorescence microscopy is limited by the availability of spectrally separable fluorescent dyes. Fluorescent dyes (or spectral GFP variants) with incongruent excitation and emission maxima are commonly employed in combination with dichroic filters or beam splitters in order to separate their signals into different channels. In general, fluorescence spectra in liquid solutions tend to expand so that significant spectral overlap does occur, no matter how hard one tries to optimize a given set of fluorescent dyes and filters. The resulting multicolor images from such experiments show a mixture of colors with many color nuances in between the original selected channels. A lack of separation between the dyes often makes it difficult for users to identify which dye the different areas of a sample are derived from. This problem can be solved by the Leica Dye Finder. A Ch1 Ch2 Ovl B Ch1 Raw Baseline Median Separated Raw Baseline Median Separated Fig. 1: NIH 3T3 mouse fibroblast cells stably expressing H2A-YFP ( Ch2, red) were transiently transfected with Actin- GFP ( Ch1, green). GFP and YFP have strongly overlapping emission spectra with considerable spectral crosstalk ( Raw images). The crosstalk is most apparent in the overlay images as a yellow color ( Ovl ). Both Leica s Automatic Dye Separation tool (panel A) and Channel Dye Separation (panel B) were applied. Before dye separation the inbuilt baseline ( Baseline ) correction and a median filter ( Median ) with kernel size 3x3 were applied to reduce background and noise respectively. Both Automatic as well as Channel Dye Separation lead to efficient unmixing ( Separation ). Magnification 63x objective. Courtesy of Constantin Kappel, DKFZ, Heidelberg, Germany Ch2 Ovl
3 This software tool offers several different algorithms to separate the dyes after image acquisition and assigns the different color nuances back to the dye it originates from. As a result, the spatial distribution of Before Dye Separation After the dyes can easily be identified by the user. The Dye Finder tool can also help to separate signals from unbalanced fluorochromes. It allows an increase in acquisition speed, whilst still seeing well separated channels. Users of confocal systems can separate fluorochromes that are excited with one and the same laser line. The Dye Finder can be used on both the Leica AF6000 system series for widefield imaging and analysis and the Leica TCS SP5 confocal systems. Channel Automatic Dye Separation how it works The Leica Dye Finder offers three methods to separate fluorochromes with overlapping emission spectra from each other, i) Automatic Dye Separation, ii) Channel Dye Separation and iii) the Spectral Dye Separation. Automatic Dye Separation The Automatic Dye Separation tool employs an advanced clustering algorithm, resulting in robust unmixing of fluorescence channels (Fig. 1 A, Fig. 2). The user only needs to specify two parameters relating to separation stringency and image dynamics. The rest is done by the powerful Automatic Dye Separation function. The dyes are separated without any further user interaction. Fig. 2: Enlargement of areas from Fig. 1 in highlighted boxes. Channel Dye Separation the chosen regions of interest are shown for channel 1 in green and for channel 2 in red. Note the absence of actin fluorescence in the H2A channel, the low nuclear fluorescence in the actin channel (there is naturally cytosolic fluorescence above and below the nucleus) as well as a clearly red fluorescence signal in the nucleus of a non-transfected cell in the upper right corner. Magnification 63x objective. Courtesy of Constantin Kappel, DKFZ, Heidelberg, Germany Channel Dye Separation The Channel Dye Separation method uses reference regions to identify the distribution coefficients of the fluorochromes in the different channels. Ideally these regions are defined on separately acquired single dye images, but can also be areas within the specimen which clearly contain single dye regions. The user then defines all reference regions within the same sample (Fig. 2, red and green ROIs, Fig. 4). A color distribution box helps the user to easily identify the regions. Once these reference regions are defined for each channel, the Channel Dye Separation tool calculates an unmixing matrix and applies this matrix to the desired image set Mean Intensity x 10 3 all channels Fig 3: The color distribution box helps the user to identify single dye regions. Reference regions are characterized by a high degree of differentiation between the channels.
4 channel 1 channel 2 channel 3 channel 4 channel 5 channel 6 DAPI FITC/Alexa488 Cy3/TAMRA Texas Red Alexa633 Cy5 HeNe-633 HeNe-594 HeNe-543 Argon 488 Diodo-405 DAPI FITC/Alexa488 Cy3/TAMRA Texas Red Alexa633 Cy5 Spectral Dye Separation (available on TCS SP5 only) Provided the emission spectrum of each dye of a sample is known, the Spectral Dye Separation method can identify the spectral overlap to eliminate crosstalk between the fluorochromes (Fig. 5). The user only needs to assign the appropriate reference spectra to the dyes and apply the unmixing algorithm to the desired image set. Reference emission spectra can be measured with either a Lambdascan or extracted from literature and stored in the Spectra Database, a standard feature which is of the LAS AF core software application. Fig. 4: 5-Color FISH of a metaphase spread obtained from PHA stimulated human lymphocytes (XY); the image shows the signals in channels 1-6 (columns 1-6) upon excitation with corresponding laser lines before (lines 1-5) and after Channel Dye Separation (line 6). Color assignment after using the Channel Dye Separation Tool: grey, DAPI (entire DNA); green, Alexa488; yellow, TAMRA; red, Texas Red; magenta, Alexa633; blue, Cy5. Image courtesy of Dr. I. Solovei and Dr. M. Cremer, Bio Center, Department of Biology II, University of Munich (LMU).
5 Dye separation for fast cell dynamics and 4D live cell experiments For confocal systems, sequential scanning is the method of choice to avoid cross-talk of fluorochromes when imaging multi-labeled specimens. However, this method slows down acquisition speed and is therefore detrimental to life cell experiments. If acquisition speed is important, e.g. when dealing with fast cell dynamics or performing 4D live cell experiments, simultaneous acquisition of the signals is favourable. Any resulting cross-talk due to overlapping emission spectra of the fluorochromes, can easily be removed by adding a subsequent dye separation step with the Dye Finder. A A similar effect is seen when using multi bandpass filter cubes for widefield systems such as the AF6000 system series. The use of multi bandpass cubes avoids time consuming filter cube changes during the sequential image acquisition, by changing only the excitation wavelength and therefore increases acquisition speed. Consequently, the individual channel images in an overlay image cannot be clearly seen. The dye separation step can also serve as useful tool in this scenario. B Separation of fluorochromes excited with one laser line What if your sample is labeled with dye combinations like Alexa 488 and GFP or Alexa 633 and Cy5? These fluorochrome combinations all have one thing in common: highly overlapping emission spectra and an excitation with identical laser lines on confocal systems, i.e. sequential scanning is not possible! Previously impossible to separate, it is now an easy task to segregate such dye combinations using the Dye Finder. This tool offers much more flexibility in dye selection, offering better differentiation between dyes and convenient use of multi labeled samples. Separation of non-balanced fluorochromes Ideally, the labels of a specimen have the same emission intensity level. Unfortunately this is not always the case, especially when dealing with live cells labeled with several fluorescent proteins. If the emission intensities of the different fluorochromes are not well balanced and their emission spectra are highly overlapping, the weak signal can be dominated or even covered by the strong signal. If this misbalance cannot be compensated for during image acquisition via hardware parameters, the visualization and/or quantification of such weakly expressed signals is difficult. With the Dye Finder this problem is easily solved. Fig. 5: Drosophila melanogaster (Oocyte), Grey: Alexa 488, nuclei of terminal follicle cells; Green: Alexa 546, filamentous actin; Red: Alexa 568 adducin-like hts; Blue: TOTO-3, DNA A) overlay, simultaneous acquisition B) Color restoration after employing the Spectral Dye Separation Tool on a Lambda-Stack Courtesy of Dr. R. Pflanz, Max-Planck Institute for Biophysical Chemistry, Göttingen, Germany Author: Constantin Kappel, Div. Theoretical Bioinformatics, German Cancer Research Center (DKFZ), Heidelberg, Germany B in cooperation with Leica Microsystems
6 Copyright Leica Microsystems CMS GmbH Am Friedensplatz 3 D Mannheim, Germany Tel / Fax / LEICA and the Leica Logo are registered trademarks of Leica IR GmbH. Order no: September
Multifluorescence The Crosstalk Problem and Its Solution
Multifluorescence The Crosstalk Problem and Its Solution If a specimen is labeled with more than one fluorochrome, each image channel should only show the emission signal of one of them. If, in a specimen
More informationConfocal Application Letter No. 13. Sequential Scan for Leica TCS NT/SP systems
Confocal Application Letter No. 13 Sequential Scan for Leica TCS NT/SP systems Leica Microsystems Heidelberg GmbH Im Neuenheimer Feld 518 D-69120 Heidelberg Telephone +49 6221 4148 0 Fax +49 6221 414833
More informationConfocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:
Confocal imaging on the Leica TCS SP8 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: 4) Do not touch the microscope while the software is initializing. Choose your options: Turn
More information長庚大學共軛焦顯微鏡課程 長庚大學共軛焦顯微鏡課程. Spot light 長庚大學
長庚大學共軛焦顯微鏡課程 Spot light 長庚大學共軛焦顯微鏡課程 20071030 長庚大學 Basic principle of Laser Scanning Confocal Microscopy The application of LSM 510 META detector Multiphoton microscopy basic principle and introduction
More informationOperation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009
Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton
More informationLeica TCS SP8 Quick Start Guide
Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power
More informationLeica TCS SP8 Quick Start Guide
Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front
More informationZEISS LSM510META confocal manual
ZEISS LSM510META confocal manual Switching on the system 1) Switch on the Remote Control button located on the table to the right of the microscope. This is the main switch for the whole system including
More informationApril 2009 No.04 WIDEFIELD APPLICATION LETTER. resolution. FRET Sensitized Emission Wizard Widefield
April 2009 No.04 WIDEFIELD APPLICATION LETTER resolution FRET Sensitized Emission Wizard Widefield FRET SE with the Leica Advanced Widefield Systems AF7000, AF6500 and AF6000 FRET Sensitized Emission (FRET
More informationThings to check before start-up.
Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your
More informationMulti-channel imaging cytometry with a single detector
Multi-channel imaging cytometry with a single detector Sarah Locknar 1, John Barton 1, Mark Entwistle 2, Gary Carver 1 and Robert Johnson 1 1 Omega Optical, Brattleboro, VT 05301 2 Philadelphia Lightwave,
More informationLeica Sp5 II Confocal User Guide
Leica Sp5 II Confocal User Guide Turning on the Confocal System (instructions are posted in the room) 1. Turn on Laser Power Button 2. Turn Key to On position 3. Turn on Scanner Power Button 4. Turn on
More informationMicroscopy from Carl Zeiss
Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path
More informationLSM 510 META in Chang Gung University
Content LSM 510 META in Chang ung University LSM 510 META 路 理 The features and applications of LSM 510 META 01-09 Introduction of the hardware 10-12 Fluorescence observation in conventional microscope
More informationSensitive imaging of spectrally overlapping fluorochromes using the LSM 510 META
Invited Paper Sensitive imaging of spectrally overlapping fluorochromes using the LSM 510 META Mary E. Dickinson a*, Christopher W. Waters a, Gregory Bearman b, Ralf Wolleschensky c, Sebastian Tille d
More information1.The Problem LIGHT-LEVEL LEVEL IMAGING. light-level level Cameras. 3. Solutions. 2. Low-light LOW-LIGHT
LOW-LIGHT LIGHT-LEVEL LEVEL IMAGING 1.The Problem 2. Low-light light-level level Cameras 3. Solutions How Much Light? I. Illumination system: 75 W Xenon Arc (~1mW/nm in visible) 490/10 nm exciter filter
More informationWhy and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005
Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev Why use confocal microscopy? Principles of the laser scanning confocal microscope. Image resolution. Manipulating the
More informationQuick Start Guide. Leica SP5 X
Quick Start Guide Leica SP5 X Please note: Some of the information in this guide was taken from Leica Microsystems Leica TCS SP5 LAS AF Guide for New Users. This work is licensed under the Creative Commons
More informationTravel to New Dimensions- LSM 880. The Resolution of a Microscope is limited. The Resolution of a Microscope is limited. Image. Image. Object.
Travel to New Dimensions- LSM 880 LSM 880: The Power of Sensitivity Our Latest Member of the LSM 880 with GaAsP Detectors Sensitivity, and Ease of Use Innovative High-End Laser Scanning Microscopes from
More informationBi Imaging. Multicolor Imaging: The Important Question of Co-Localization. Anna Smallcombe Bio-Rad Laboratories, Hemel Hempstead, UK
Multicolor Imaging: The Important Question of Co-Localization Anna Smallcombe Bio-Rad Laboratories, Hemel Hempstead, UK The use of specific fluorescent probes, combined with confocal or multiphoton microscopy
More informationSupplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each
Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each lens with a 1 Airy unit pinhole setting. Many water lenses
More informationConfocal Laser Scanning Microscopy
Name of the Core Facility: Confocal Laser Scanning Microscopy CORE Forschungszentrum Immunologie Mainz Welcome to the CSLM Core Facility: The CLSM Core Facility enables working groups to incorporate high
More informationQuick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s
LSM 5 MP, LSM 510 and LSM 510 META M i c r o s c o p y f r o m C a r l Z e i s s Quick Guide Laser Scanning Microscopes LSM Software ZEN 2007 August 2007 We make it visible. Contents Page Contents... 1
More informationShreyash Tandon M.S. III Year
Shreyash Tandon M.S. III Year 20091015 Confocal microscopy is a powerful tool for generating high-resolution images and 3-D reconstructions of a specimen by using point illumination and a spatial pinhole
More informationConfocal Microscope. Confocal Microscope C2
Confocal Microscope Confocal Microscope C2 Confocal Microscope An essential microscopy laboratory instrument The C2 confocal microscope system comprises a new generation of Nikon confocal instruments designed
More informationOperating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.
Smith College August 2010 Operating Checklist for using the Laser Scanning Confocal Microscope Leica TCS SP5. CONTENT, page no. Startup, 1 Initial set-up, 1 Software, 2 Microscope Specimen observation
More informationOPERATING INSTRUCTIONS
Zeiss LSM 510 M eta Confocal M icroscope OPERATING INSTRUCTIONS Starting the System: 1. Turn the black knob on the laser box one-quarter turn from Off to On. You will hear the laser cooling mechanisms
More informationLeica TCS SP8. The New Milestone of Confocal Microscopy - New full range confocal platform - Major Instruments Co., Ltd.
Leica TCS SP8 The New Milestone of Confocal Microscopy - New full range confocal platform - Major Instruments Co., Ltd. Confocal Image -- Fluorescence Drosophila melanogaster, eye section Red: F-Actin,
More informationTraining Guide for Carl Zeiss LSM 510 META Confocal Microscope
Training Guide for Carl Zeiss LSM 510 META Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON Components and System/PC switches
More information1 Co Localization and Working flow with the lsm700
1 Co Localization and Working flow with the lsm700 Samples -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ BrDU with alexa 488. -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ no BrDU (but with
More informationDynamic Confocal Imaging of Living Brain. Advantages and risks of multiphoton microscopy in physiology
Dynamic Confocal Imaging of Living Brain Advantages and risks of multiphoton microscopy in physiology Confocal laser scanning microscopy In conventional optical microscopy focused and out-offocus light
More informationImaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal
Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal Todays Goal: Introduce some additional functionalities of the Leica SP8 confocal HyD vs. PMT detectors Dye Assistant Scanning By
More informationLeica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager
Leica TCS SL Confocal Training Neuroscience Imaging Facility The Ohio State University Rightmire Hall 614-292-1367 Staff Core Director Anthony Brown, Ph. D. 060 Rightmire Hall 614-292-1205 brown.2302@osu.edu
More informationZeiss 780 Training Notes
Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and
More informationLEICA TCS SP5 AOBS TANDEM USER MANUAL
LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE
More informationFundamentals of Light Microscopy II: Fluorescence, Deconvolution, Confocal, Multiphoton, Spectral microscopy. Integrated Microscopy Course
Fundamentals of Light Microscopy II: Fluorescence, Deconvolution, Confocal, Multiphoton, Spectral microscopy Integrated Microscopy Course Review Lecture 1: Microscopy Basics Light train Kohler illumination*
More informationWe attempted to separate the two dyes by acquiring images using a single excitation wavelength and just two emission wavelengths.
TN437: Spectral Separation of monochrome images using Volocity 4.0 Introduction Spectral Separation is a technique that allows the user to separate images containing data from more than one fluorochrome
More informationConfocal Microscope. Confocal Microscope C2
Confocal Microscope Confocal Microscope C2 Confocal Microscope An essential microscopy laboratory insturument The C2 confocal microscope system comprises a new generation of Nikon confocal instruments
More informationContents. Introduction
Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path and lasers... 12 Scanning
More informationBIO-RAD TECHNICAL NOTE 11. Written by: Anna Smallcombe and Duncan McMillan, Bio-Rad Microscopy Division, Hemel Hempstead, UK
BIO-RAD TECHNICAL NOTE 11 Co-localisation: how is it determined, and how is it analysed with the Bio-Rad LaserPix image analysis software? Written by: Anna Smallcombe and Duncan McMillan, Bio-Rad Microscopy
More informationConfocal Microscopy. (Increasing contrast and resolu6on using op6cal sec6oning) Lecture 7. November 2017
Confocal Microscopy (Increasing contrast and resolu6on using op6cal sec6oning) Lecture 7 November 2017 3 Flavours of Microscope Confocal Laser Scanning Problem: Out of Focus Light Spinning disc 2-Photon
More informationPractical work no. 3: Confocal Live Cell Microscopy
Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside
More informationCell Biology and Bioimaging Core
Cell Biology and Bioimaging Core Leica TCS SP5 Operating Instructions Starting up the instrument 1. First, log in the log book located on the confocal desk. Include your name, your lab s PI, an account
More informationFast Laser Raman Microscope RAMAN
Fast Laser Raman Microscope RAMAN - 11 www.nanophoton.jp Fast Raman Imaging A New Generation of Raman Microscope RAMAN-11 developed by Nanophoton was created by combining confocal laser microscope technology
More informationTitle: Leica SP5 Confocal User Manual
Title: Leica SP5 Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG07 or 05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk Website:
More informationLeica SPEII confocal microscope. Short Manual
Leica SPEII confocal microscope Short Manual Switching ON sequence: 1. Turn on the Workstation under the bench (top, far right). 2. Turn on the Supply Unit - Laser box (big green switch first and then
More informationZeiss 880 Training Notes Zen 2.3
Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start
More informationOpterra II Multipoint Scanning Confocal Microscope. Innovation with Integrity
Opterra II Multipoint Scanning Confocal Microscope Enabling 4D Live-Cell Fluorescence Imaging through Speed, Sensitivity, Viability and Simplicity Innovation with Integrity Fluorescence Microscopy The
More informationMicroscope Confocal Sp2 Upright.
Microscope Confocal Sp2 Upright. Welcome to the Leica Sp2 Confocal Upright tutorial. Before using the Sp2 Invert, You will need to put down your name on the reservation system = http://svintranet.epfl.ch/index.php?optio
More informationTraining Guide for Carl Zeiss LSM 880 with AiryScan FAST
Training Guide for Carl Zeiss LSM 880 with AiryScan FAST ZEN 2.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2018) Power ON Routine 1 2 Turn ON Main Switch from the remote control
More informationLSM 710 Confocal Microscope Standard Operation Protocol
LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.
More informationUser Guide to the IBIF Leica TCS SP8 MP Confocal Microscope
User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope This version: 7.24.14. Introduction The IBIF confocal microscope is made available on a fee-for-use-hour basis to all users who have been trained.
More informationIn essence this means, that a certain proportion of a signal in one channel is actually derived from another dye spilling over into the channel.
NNKQ táçéñáéäç=jìäíáåü~ååéä=råãáñáåö= _~ÅâÖêçìåÇ= Widefield Multichannel Unmixing is a new function for the removal of crosstalk between fluorescent dyes in multichannel images with up to eight fluorescence
More informationConfocal Microscopy. Kristin Jensen
Confocal Microscopy Kristin Jensen 17.11.05 References Cell Biological Applications of Confocal Microscopy, Brian Matsumoto, chapter 1 Studying protein dynamics in living cells,, Jennifer Lippincott-Schwartz
More informationNature Methods: doi: /nmeth Supplementary Figure 1. Schematic of 2P-ISIM AO optical setup.
Supplementary Figure 1 Schematic of 2P-ISIM AO optical setup. Excitation from a femtosecond laser is passed through intensity control and shuttering optics (1/2 λ wave plate, polarizing beam splitting
More informationTraining Guide for Leica SP8 Confocal/Multiphoton Microscope
Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence
More informationOpterra. Multipoint Scanning Confocal Microscope. Innovation with Integrity. Cell-Friendly, High-Speed, High-Resolution Imaging
Opterra Multipoint Scanning Confocal Microscope Cell-Friendly, High-Speed, High-Resolution Imaging Innovation with Integrity Fluorescence Microscopy Opterra Multipoint Scanning Confocal Microscope Superior
More information1. Editorial. N 9 June Content
N 9 June 2010 Content 1. Editorial 2. Timelapse: news and updates 3. n vivo rodent imaging setup available in Epalinges 4. 2010 Workshops 5. Spotlight on mage Stitching 1. Editorial We welcome new and
More informationLast updated: May 2014 Y.DeGraaf
FLINDERS MICROSCOPY BIOMEDICAL SERVICES AVAILABLE MICROSCOPES AND SPECIFICATIONS & INFORMATION REGARDING TRAINING FOR NEW USERS Last updated: May 2014 Y.DeGraaf If you have new staff or students (Honours/Masters
More informationFastest high definition Raman imaging. Fastest Laser Raman Microscope RAMAN
Fastest high definition Raman imaging Fastest Laser Raman Microscope RAMAN - 11 www.nanophoton.jp Observation A New Generation in Raman Observation RAMAN-11 developed by Nanophoton was newly created by
More informationQuick Guide. NucleoCounter NC-3000
Quick Guide NucleoCounter NC-3000 Table of contents Setting up the FlexiCyte Protocol 2 Editing Image Capture and Analysis Parameters 3 Optimizing Exposure Time 4 Compensation for Spectral Overlap 6 Creating
More informationFast Laser Raman Microscope RAMAN
Fast Laser Raman Microscope RAMAN - 11 www.nanophoton.jp Fast Raman Imaging A New Generation of Raman Microscope RAMAN-11 developed by Nanophoton was created by combining confocal laser microscope technology
More informationMicroscopy from Carl Zeiss LSM 700. Laser Scanning Microscope. High-End for All Uncompromised Quality and Operating Convenience
Microscopy from Carl Zeiss LSM 700 Laser Scanning Microscope High-End for All Uncompromised Quality and Operating Convenience 2 3 The New LSM 700 from Carl Zeiss From a specialists system to the high-end
More informationMicroscope Confocal LSM510 META
Microscope Confocal LSM510 META Welcome to the Zeiss LSM 510 Meta Confocal tutorial. Before using the LSM 510 META, Log off any other computer that is open with your personal login. You will need to put
More informationANSWER KEY Lab 2 (IGB): Bright Field and Fluorescence Optical Microscopy and Sectioning
Phys598BP Spring 2016 University of Illinois at Urbana-Champaign ANSWER KEY Lab 2 (IGB): Bright Field and Fluorescence Optical Microscopy and Sectioning Location: IGB Core Microscopy Facility Microscope:
More informationSpectral Imaging with the Opterra Multipoint Scanning Confocal
Spectral Imaging with the Opterra Multipoint Scanning Confocal Outline Opterra design overview Scan Modes Light Path Spectral Imaging with Opterra Drosophila larva heart. Opterra Design Overview Supravideo
More informationPixel shift in fluorescence microscopy
Pixel shift in fluorescence microscopy 1. Introduction Multicolor imaging in fluorescence microscopy is typically performed by sequentially acquiring images of different colors. An overlay of these images
More informationpicoemerald Tunable Two-Color ps Light Source Microscopy & Spectroscopy CARS SRS
picoemerald Tunable Two-Color ps Light Source Microscopy & Spectroscopy CARS SRS 1 picoemerald Two Colors in One Box Microscopy and Spectroscopy with a Tunable Two-Color Source CARS and SRS microscopy
More informationNikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON
N-SIM guide NIKON IMAGING CENTRE @ KING S COLLEGE LONDON Starting-up / Shut-down The NSIM hardware is calibrated after system warm-up occurs. It is recommended that you turn-on the system for at least
More informationPoint Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy
Bi177 Lecture 5 Adding the Third Dimension Wide-field Imaging Point Spread Function Deconvolution Confocal Laser Scanning Microscopy Confocal Aperture Optical aberrations Alternative Scanning Microscopy
More information(12) United States Patent (10) Patent No.: US 6,388,807 B1. Knebel et al. (45) Date of Patent: May 14, 2002
USOO6388807B1 (12) United States Patent (10) Patent No.: Knebel et al. () Date of Patent: May 14, 2002 (54) CONFOCAL LASER SCANNING (56) References Cited MICROSCOPE U.S. PATENT DOCUMENTS (75) Inventors:
More informationNikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.
Nikon AZ100 Laser Scanning Macro Confocal Microscope Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin May 2017 Contents 1 Introduction 2 2 Hardware - Startup 2 3 Software/Operation 4 3.1 Multidimensional
More informationLeica SP8 TCS Users Manual
Leica SP8 TCS Users Manual Follow the procedure for start up and log on as posted in the lab. Please log on with your account only and do not share your password with anyone. We track and confirm usage
More informationBoulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement
Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement CONFOCAL MICROSCOPY BioVis Uppsala, 2017 Jeremy Adler Matyas Molnar Dirk Pacholsky Widefield & Confocal Microscopy
More informationAdvanced Optical Microscopy lecture. 03. December 2012 Kai Wicker
Advanced Optical Microscopy lecture 03. December 2012 Kai Wicker Today: Optical transfer functions (OTF) and point spread functions (PSF) in incoherent imaging. 1. Quick revision: the incoherent wide-field
More informationLight Microscopy. Upon completion of this lecture, the student should be able to:
Light Light microscopy is based on the interaction of light and tissue components and can be used to study tissue features. Upon completion of this lecture, the student should be able to: 1- Explain the
More informationcontents TABLE OF The SECOM platform Applications - sections Applications - whole cells Features Integrated workflow Automated overlay
S E C O M TABLE OF contents The SECOM platform 4 Applications - sections 5 Applications - whole cells 8 Features 9 Integrated workflow 12 Automated overlay ODEMIS - integrated software Specifications 13
More informationLeica SP8 Resonant Confocal. Quick-Start Guide
Leica SP8 Resonant Confocal Quick-Start Guide Contents Start-up Preparing for Imaging Part 1 On the scope Part 2 Software interface Part 3 Heat & CO2 incubation Part 4 Other hardware options Shut-down
More informationZeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling
Zeiss LSM 510 Confocor III Training Notes Center for Cell Analysis & Modeling Confocor 3 Start Up Go to System Module Turn on Main Switch, System/ PC, and Components Switches Do you need the arc lamp?
More informationLSM 510 Meta Training Notes
LSM 510 Meta Training Notes Turning on the system Turn on X-Cite power supply. This supplies light for epifluorescence for viewing your samples through the microscope. Turn on the remote control switch.
More informationMULTIPHOTON MICROSCOPY. Matyas Molnar Dirk Pacholsky
MULTIPHOTON MICROSCOPY Matyas Molnar Dirk Pacholsky Information Information given here about 2 Photon microscopy were mainly taken from these sources: Background information on 2-Photon microscopy: http://micro.magnet.fsu.edu/primer/techniques/fluorescence/multiphoton/
More informationFinal Exam, 150 points PMB 185: Techniques in Light Microscopy
Final Exam, 150 points Name PMB 185: Techniques in Light Microscopy Point value is in parentheses at the end of each question. Note: GFP = green fluorescent protein ; CFP = cyan fluorescent protein ; YFP
More informationLSM 510 Training Notes
LSM 510 Training Notes Turning on the system Turn on the arc lamp, found on the bench top left of the microscope. This supplies light for epifluorescence for viewing your samples through the microscope.
More informationTRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO
TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010 Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped
More informationMULTIFOCAL MULTISPECTRAL DESCANNED DETECTION IN TPLSM
MULTIFOCAL MULTISPECTRAL DESCANNED DETECTION IN TPLSM T. Bergmann a1, M. Tiemann a, J. Martini a,b, K. Tönsing a and D. Anselmetti a a Bielefeld University, Experimental Biophysics & Applied Nanoscience,
More informationIn our previous lecture, we understood the vital parameters to be taken into consideration before data acquisition and scanning.
Interactomics: Protein Arrays & Label Free Biosensors Professor Sanjeeva Srivastava MOOC NPTEL Course Indian Institute of Technology Bombay Module 7 Lecture No 34 Software for Image scanning and data processing
More informationMetaXpress Software: Cell Scoring Module
MetaXpress Software: Cell Scoring Module Cell Scoring Module Overview The Cell Scoring module can be used to analyze cells imaged in 2 wavelengths W1 should be a stain for all nuclei (e.g. DAPI, Hoechst,
More informationFigure S1 Figure S1. Wild type IgG and FcRn colocalize in APPL+ TCs.
Figure S1 Figure S1. Wild type IgG and FcRn colocalize in APPL+ TCs. HMEC-1 cells were cotransfected with GFP-FcRn and β2m. Transfected cells were incubated with 200 μg/ml Alexa 647-wild type IgG (A) or
More informationSuper Resolution Microscope N-SIM E. Super Resolution Microscope
Super Resolution Microscope N-SIM E Super Resolution Microscope Explore Nano world with Nikon N-SIM E is a streamlined, affordable superresolution system that provides double the resolution of conventional
More informationTraining Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope
Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the
More informationSHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014
CIAN LSM1 or LSM2 short instructions, version 1.4, September 2014 page 1 of 6 SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 Before starting To work with LSM1
More information: fps. pco.edge. 1.4 electrons. 5.5 megapixel. pco. high speed. high resolution. low noise. high dynamic range. scientific CMOS camera
edge scientific CMOS camera low noise 1.4 electrons high resolution 5.5 megapixel high speed high dynamic range 22 000 :1 100 fps The new edge is a breakthrough in scientific imaging cameras, due to its
More informationBD LSRFortessa X-20. Special Order Product. Designed for limited space and boundless potential
BD LSRFortessa X-2 Special Order Product Designed for limited space and boundless potential Next generation high performance cell analyzer The BD LSRFortessa X-2 cell analyzer delivers high performance,
More informationSTART-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7
Leica DMI AF6000LX Table of contents START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 ACQUIRE MODULE 6 SETTING THE LIGHTPATH 6
More informationADVANCED METHODS FOR CONFOCAL MICROSCOPY II. Jean-Yves Chatton Sept. 2006
ADVANCED METHODS FOR CONFOCAL MICROSCOPY II Jean-Yves Chatton Sept. 2006 Workshop outline Confocal microscopy of living cells and tissues X-Z scanning Time series Bleach: FRAP, photoactivation Emission
More informationIN Cell Analyzer 6500HS
GE Healthcare IN Cell Analyzer 6500HS IN Cell Analyzer 6500HS is a fully-automated confocal cell imaging system from GE Healthcare. Building on the capabilities of earlier IN Cell Analyzer systems, it
More informationChemical Imaging. Whiskbroom Imaging. Staring Imaging. Pushbroom Imaging. Whiskbroom. Staring. Pushbroom
Chemical Imaging Whiskbroom Chemical Imaging (CI) combines different technologies like optical microscopy, digital imaging and molecular spectroscopy in combination with multivariate data analysis methods.
More informationBASICS OF CONFOCAL IMAGING (PART I)
BASICS OF CONFOCAL IMAGING (PART I) INTERNAL COURSE 2012 LIGHT MICROSCOPY Lateral resolution Transmission Fluorescence d min 1.22 NA obj NA cond 0 0 rairy 0.61 NAobj Ernst Abbe Lord Rayleigh Depth of field
More informationGuide to Confocal 5. Starting session
Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope
More informationConfocal Application Notes Vol. 5 July 2010
Tile Scan Prepared by Myriam Gastard, PhD Application and Technical Support Group, Leica Microsystems, Inc. In this issue of our Confocal Application Notes, proper set up of the Tile function enables you
More information