Improved Separation and Quantification of Dioxin and Furan Congeners Using an Application-Specific Capillary Gas Chromatographic Column and GC-HRMS

Size: px
Start display at page:

Download "Improved Separation and Quantification of Dioxin and Furan Congeners Using an Application-Specific Capillary Gas Chromatographic Column and GC-HRMS"

Transcription

1 Improved Separation and Quantification of Dioxin and Furan Congeners Using an Application-Specific Capillary Gas Chromatographic Column and GC-HRMS Frank Dorman, Shawn Reese, Gary Stidsen Restek Corporation, Bellefonte, Pennsylvania Jack Cochran LECO Corporation, Las Vegas, Nevada Eric Reiner, Karen MacPherson, Terry Kolic Ministry of Environment, Etobicoke, Ontario

2 Analysis of Dioxins and Furans

3 History GC Analytical Methods All isomers/congeners can be separated from each other with a minimum of two columns (Ryan, J., 1991) and GC HRMS Typically used: 6-M GC column, 5 phenyl liquid phase, separates many congeners, but not with complete resolution HOWEVER.. Confirmation still required, generally with high-cyano liquid phases ex. for 2,3,7,8-TCDF Method 1613 suggests a 225-phase

4 Dioxin and Furan Analysis Dual column method Usually 5 diphenyl column and a high-cyano column (eg Rtx-225) Cyano columns have poorer lifetimes and lower maximum operating temperatures 5 diphenyl phases do not have the selectivity to accurately quantitate most samples USEPA requires 2,3,7,8-tcdf to be confirmed on a X-225 Desirable to have both columns in the same oven, and to improve the separation of the 5, or employ a single column which achieves the necessary separation

5 PCDD and PCDF Target List # of Chlorine #Dioxins #Furans tetra 22(1) 38(1) penta 14(1) 28(2) hexa 1(3) 16(4) hepta 2(1) 4(2) octa 1(1) 1(1) ( ) numbers are 2,3,7,8-substituted congeners

6 Tetra-dioxin resolution check standard on 5 diphenyl / 95 dimethyl stationary phase Jul21_ / : Voltage SIR 2 Channels EI e M X.25 mm i.d. X.25 um d.f. Time

7 How Resolution Affects Quantitation Quantification of 2 ng/ul Reference Standard (Results Based on Multi-level/Multi-component Calibration): Component A Component B Component A & B Rep Rep Rep Component A Rep Rep Rep Component B Rep Rep Rep

8 Achieving Analyte Separation Resolution R = 1/4 L / h x ( k / k+1 ) x ( α-1 1 / α ) Capacity Factor k = t R - t / t Selectivity α = k 2 / k 1 Thermodynamics: G G = H - T S G G = - RT ln K D

9 Modeling - Energies of Interaction F F H 3 C O Si Si O CH 3 F F G G Cl Cl Cl O O O O Cl Cl Cl Cl Cl F Si O F Separation ( α)( = G

10 3-Space Selectivity Model for 3 Compounds Surface=F H S

11 Example Result: Demeton-O on PDMS: Observed G = -1.14E4 J/mol Calculated G = -1.13E4 J/mol

12 Tetra-dioxin resolution check standard on 5 diphenyl / 95 dimethyl stationary phase Jul21_ / : Voltage SIR 2 Channels EI e M X.25 mm i.d. X.25 um d.f. Time

13 Rtx-Dioxin2 Capillary GC Column TCDD Resolution Mix 1238/

14 ay6_cs5 x4 TCDD Dioxin Window Voltage SIR 14 Channels EI e ay6_cs PCDD Voltage SIR 14 Channels EI e ay6_cs HxCDD Voltage SIR 14 Channels EI e ay6_cs HpCDD Voltage SIR 14 Channels EI e ay6_cs5 OCDD Voltage SIR 14 Channels EI e Time

15 ay6_cs TCDF Furan Window Voltage SIR 14 Channels EI e ay6_cs PCDF Voltage SIR 14 Channels EI e ay6_cs HxCDF Voltage SIR 14 Channels EI e ay6_cs HpCDF Voltage SIR 14 Channels EI e ay6_cs5 OCDF Voltage SIR 14 Channels EI e Time

16 GC program for RTX-Dioxin2 Gas Flow - constant 6m x.25mm x.25µm 1.5mL/min Initial Temp. 13 C - hold 1min 1st Temp. Ramp 2nd Temp. Ramp 3rd Temp. Ramp 4 C/min to 2 C, hold 3. C/min to 235 C, hold 5 C /min to 3 C, hold ~5-1min. Total Run Time ~43 min

17 TCDF Resolution on RTX-Dioxin2 TD TF WD may12_res : Voltage SIR 2 Channels EI e6 2347/ may12_res18 1 2: Voltage SIR 2 Channels EI e5 TCDFs 13C Time

18 TCDF ion channel for flyash extract c may12_res15 1 Flyash Extract : Voltage SIR 2 Channels EI e may12_res : Voltage SIR 2 Channels EI e6 13C12-2,3,7,8-tcdf Time

19 TCDF ion channel for sewage sludge c /1 jul4_dx3_s3 Sm (Mn, 2x1) 1 Sewage sludge : Voltage SIR 2 Channels EI e jul4_dx3_s3 Sm (Mn, 2x1) : Voltage SIR 2 Channels EI e6 13C tcdf 13C-2,3,6,7-tcdf Time

20 Table I - 2,3,7,8-TCDF (pg/g) Certified 5 phenyl X-225 Rtx-DIOXIN2 Value WMS ± 16 EC-2(DX-1) 88 n/a (±44) Sediment NIST n/a Flyash Flyash Biota Biota

21 TCDF ion channel for Biota sample c may9_res13 1 2: Voltage SIR 2 Channels EI e4 DPE ion channel may9_res13 1 tcdf 2: Voltage SIR 2 Channels EI e4 may9_res13 1 2: Voltage SIR 2 Channels EI e6 13 C-2378-tcdf Time

22 Potential resolution between DPEs and Furans c may9_res13 1 DPE ion channel 2: Voltage SIR 2 Channels EI e4 may9_res13 1 2: Voltage SIR 2 Channels EI e4 tcdf may9_res13 1 2: Voltage SIR 2 Channels EI e6 13 C-2378-tcdf Time

23 ay6_res TCDD : Voltage SIR 2 Channels EI e ay6_res : Voltage SIR 2 Channels EI e 2378-TCDD Time

24 5 Phenyl Column Resolution Mix Rtx-DIOXIN / tcdd 1237/38-tcdd C-1234-tcdd 13C-2378-tcdd Time

25 Table II - 2,3,7,8-TCDD (pg/g) Certified 5 phenyl X-225 Rtx-Dioxin2 Value WMS ±5.6 EC-2 (DX-1) 24 n/a (± 53) Sediment Flyash Flyash-2 <

26 ay6_res PCDF 3: Voltage SIR 2 Channels EI e ay6_res PCDF PCDF 3: Voltage SIR 2 Channels EI e Time

27 Table III - 1,2,3,7,8-PCDF (pg/g) Certified 5 phenyl Rtx-Dioxin2 Value WMS ± 5 EC-2(DX-1) ± 14 Sediment Flyash Flyash

28 Table IV- 2,3,4,7,8-PCDF (pg/g) Certified 5phenyl Rtx-Dioxin2 Value WMS ± 6.1 EC-2(DX-1) ± 32 Sediment Flyash Flyash

29 ay6_res PCDD 3: Voltage SIR 2 Channels EI e ay6_res : Voltage SIR 2 Channels EI e PCDD Time

30 HxCDF ion channel in Flyash sample c may12_res15 1 4: Voltage SIR 16 Channels EI e6 may12_res HxCDF HxCDF 13 C-HxCDFs 4: Voltage SIR 16 Channels EI e HxCDF HxCDF Time

31 Table V - 1,2,3,4,7,8-HxCDF (pg/g) Certified 5 phenyl Rtx-Dioxin2 Value WMS ± 24 EC-2 (DX-1) (± 276) Sediment Flyash Flyash

32 HxCDD ion channel for sediment EC-2 1uL may14_res3 1 4: Voltage SIR 16 Channels EI e5 may14_res HxCDD 13 C-HxCDDs 4: Voltage SIR 16 Channels EI e HxCDD HxCDD Time

33 Table VI - 1,2,3,7,8,9-HxCDD (pg/g) Certified 5 phenyl Rtx-Dioxin2 Value WMS ± 8 EC-2 (DX-1) (± 24) Sediment Flyash Flyash

34 ay6_res HpCDF 5: Voltage SIR 11 Channels EI e ay6_res HpCDF HpCDF : Voltage SIR 11 Channels EI e Time

35 ay6_res HpCDD : Voltage SIR 11 Channels EI e ay6_res : Voltage SIR 11 Channels EI e HpCDD HpCDD Time

36 ay6_res9 OCDD/OCDF : Voltage SIR 9 Channels EI e OCDF ay6_res : Voltage SIR 9 Channels EI e OCDD ay6_res : Voltage SIR 9 Channels EI e Time

37 Rtx-Dioxin2 Capillary Column Proprietary backbone stabilized polysiloxane designed for toxic dioxin and furan analysis by GC-HRMS 34 maximum operating temperature in standard fused silica tubing Can be used as secondary confirmation column for Rtx-Dioxin, or other primary columns like 5 diphenyl phases Can function as single column for stand-alone dioxin and furan analysis and has been accredited for use as such Clive Robinson ALS Environmental Queensland Australia Chromatographically separates the chlorinated diphenylethers from the furans especially important in biota matrices Columns available: Cat # M X.25 mm i.d. X.25 um d.f. Cat # M X.18 mm i.d. X.18 um d.f.

A new gold standard for Dioxin analysis A combined approach ; automated efficient clean-up together with a new Triple Quadrupole GC-MS/MS

A new gold standard for Dioxin analysis A combined approach ; automated efficient clean-up together with a new Triple Quadrupole GC-MS/MS A new gold standard for Dioxin analysis A combined approach ; automated efficient clean-up together with a new Triple Quadrupole GC-MS/MS Haruhiko Miyagawa, GCMS Product Manager Analytical &Measuring Instruments

More information

Persistant Organic Pollutants

Persistant Organic Pollutants Persistant Organic Pollutants Fast screening and precise confirmation with GC-MSMS and GC-HRMS. Hans-Joachim Huebschmann GC-MS Technology Manager Dirk Krumwiede Product & Application Specialist for Advanced

More information

RSR011 SEVERN TRENT LABORATORIES, INC. Run Date: 9/22/06 SAMPLE CONFIRMATION REPORT Page : 1 Time : 8:06:14 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - The following

More information

Prep TM Automated Extraction and Clean-up System for PCDDs, PCDFs and PCBs

Prep TM Automated Extraction and Clean-up System for PCDDs, PCDFs and PCBs Power-Prep Prep TM Automated Extraction and Clean-up System for PCDDs, PCDFs and PCBs Jef Focant and Hamid Shirkhan Dioxin Lab. BELGIUM FMS Inc. WWW.FMS-INC.Com USA Dioxins Analysis! Extraction (Soxhlet,

More information

Minnesota Department of Health Environmental Laboratory Accreditation Program

Minnesota Department of Health Environmental Laboratory Accreditation Program Minnesota Department of Health Environmental Laboratory Accreditation Program Issues accreditation to State Laboratory ID: 048-999-468 EPA Lab Code: 02693 ALS Group USA Corp dba ALS Environmental Houston

More information

Results from UNEP 2 nd Interlaboratory

Results from UNEP 2 nd Interlaboratory Results from UNEP 2 nd Interlaboratory Overview on the intercalibration data of dl-pops Van Bavel B 1, Van der Veen I 2, Nilsson H 1, De Boer J 2, Fiedler H 3 1 Örebro University, School of Science and

More information

Approaches to highest sensitivity analysis of selected Dioxin / Furan congeners:

Approaches to highest sensitivity analysis of selected Dioxin / Furan congeners: Approaches to highest sensitivity analysis of selected Dioxin / Furan congeners: a.) Cryogenic Peak Zone Focusing (cyro - PZF) b.) Fast GC Dirk Krumwiede, Jeff Focant, Heinz Mehlmann, Daniela Cavagnino,

More information

Thermo Scientific DFS

Thermo Scientific DFS Thermo Scientific DFS High Resolution Sector Field MS Instrument performance standards for Dioxin / Furan analysis Dirk Krumwiede, Product Specialist Trace MS Overview: presentation main topics 1.) NEW

More information

Date Issued: July 01, 2018 Expiration Date: June 30, 2019

Date Issued: July 01, 2018 Expiration Date: June 30, 2019 State of Florida Department of Health, Bureau of Public Health Laboratories This is to certify that VISTA ANALYTICAL LABORATORY INC. 1104 WINDFIELD WAY EL DORADO HILLS, CA 95762 has complied with Florida

More information

N-Methyl-1-(6-methylpyridin-2-yl)propan-2-amine

N-Methyl-1-(6-methylpyridin-2-yl)propan-2-amine H C N CH HN CH. GENERAL INFORMATION IUPAC Name: CAS#: Not Available Synonyms: Source: Appearance: Not Available DEA Reference Material Collection Pale yellow powder UV max (nm): Not Determined. CHEMICAL

More information

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Application Note Meeting the Korean Method ES 04607.1 for Naphthalene ( 나프탈렌 ) with Static and Dynamic Headspace GC/MS Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Introduction The Korean

More information

Photograph: Norith Eckbo

Photograph: Norith Eckbo Photograph: Norith Eckbo Contaminants in Polar Regions PBDEs Particle deposition (high K OA ) HCHs High volatility: Single hoppers (high K AW ) Condensation DDT PCBs Evaporation Low volatility (low K AW

More information

Introducing the Agilent 7000A Triple Quadrupole GC/MS

Introducing the Agilent 7000A Triple Quadrupole GC/MS Introducing the Agilent 7000A Triple Quadrupole GC/MS Paul Zavitsanos, Joe Weitzel, Monty Benefiel and Terry Sheehan May 2008 Page 1 What s a Triple Quad? Page 2 GC/MS Triple Quad (QQQ) Collision Gas (N2

More information

Profiling of Volatile Organic Compounds in Milk and Orange Juice Using Headspace Analysis

Profiling of Volatile Organic Compounds in Milk and Orange Juice Using Headspace Analysis Profiling of Volatile Organic Compounds in Milk and Orange Juice Using Headspace Analysis HT3 Application Note Introduction Volatile organic compounds, VOCs, present in liquid food products play a very

More information

EVALUATION OF PCDDs AND PCDFs IN WILD GAME TAKEN FROM THE FLOODPLAIN ALONG THE TITTABAWASSEE RIVER

EVALUATION OF PCDDs AND PCDFs IN WILD GAME TAKEN FROM THE FLOODPLAIN ALONG THE TITTABAWASSEE RIVER EVALUATION OF PCDDs AND PCDFs IN WILD GAME TAKEN FROM THE FLOODPLAIN ALONG THE TITTABAWASSEE RIVER Prepared by: ENTRIX, Inc. East Lansing, Michigan Prepared for: The Dow Chemical Company Midland, Michigan

More information

A Survey of Residual Solvents in Various Inks used in Consumer Product Packaging

A Survey of Residual Solvents in Various Inks used in Consumer Product Packaging A Survey of Residual Solvents in Various Inks used in Consumer Product Packaging Application Note Abstract Food manufacturers use various printing inks on their consumer product packaging. These inks are

More information

LC/MS/MS. Page Header. triple quadrupole mass spectrometer.

LC/MS/MS. Page Header. triple quadrupole mass spectrometer. LC/MS/MS VARIAN, INC. 320-MS Page Header triple quadrupole mass spectrometer www.varianinc.com VARIAN, INC. 320-MS Unsurpassed commitment to innovation Varian, Inc. is an innovator and leader in mass spectrometry

More information

Integration and Optimization of Hardware and Software for a Differential Flow Modulated GCxGC

Integration and Optimization of Hardware and Software for a Differential Flow Modulated GCxGC Paper 170-2 Integration and Optimization of Hardware and Software for a Differential Flow Modulated GCxGC Roger Firor, James McCurry, Chunxiao Wang and Michael Feeney Agilent Technologies Wilmington, DE

More information

GC Application Note. Quantification of nitrosamines in water by automated PAL SPME-Arrow and GC/MS.

GC Application Note. Quantification of nitrosamines in water by automated PAL SPME-Arrow and GC/MS. GC Application Note Quantification of nitrosamines in water by automated PAL SPME-Arrow and GC/MS www.palsystem.com Quantification of nitrosamines in water by automated PAL SPME-Arrow and GC/MS Peter Egli,

More information

Gas Chromatography and Troubleshooting

Gas Chromatography and Troubleshooting An Intensive 5 Day Training Course Gas Chromatography and Troubleshting for the Oil & Gas Industry 24-SEP-17 31 Dec - 04 Jan 2018, Dubai 02-06 Sep 2018, Dubai www.petroknowledge.com Gas Chromatography

More information

Determination of the N-Nitrosamine Content in Rubber Articles Using the Agilent 7000A Triple Quadrupole GC/MS System

Determination of the N-Nitrosamine Content in Rubber Articles Using the Agilent 7000A Triple Quadrupole GC/MS System Determination of the N-Nitrosamine Content in Rubber Articles Using the Agilent 7A Triple Quadrupole GC/MS System Application Note Consumer Products Authors Yun Zou and Chongtian Yu Agilent Technologies

More information

Ultrasensitive LC MS/MS: Agilent 6470 and 6495 LC-QQQ

Ultrasensitive LC MS/MS: Agilent 6470 and 6495 LC-QQQ Ultrasensitive LC MS/MS: Agilent 6470 and 6495 LC-QQQ ROCK SOLID Performance for Trace-Level Quantitation Agilent Technologies 1 Brief History of Agilent 6400 QQQ LC/MS 6410 2006 6460 Agilent Jet Stream

More information

Collision cell: Fragmentation efficiency > 90 % Duty cycle: Variable between 1 and 50 % (user selectable trapping time/scan time ratio) Transmission:

Collision cell: Fragmentation efficiency > 90 % Duty cycle: Variable between 1 and 50 % (user selectable trapping time/scan time ratio) Transmission: Finnigan MAT 95 XP-API-GC-Trap tandem Mass spectrometer system Tandem Mass Spectrometer based on high resolution magnetic sector and Ion-Trap MS/MS system with API ( ESI, APCI ) and EI/CI ionization Mass

More information

NPD - Nitrogen Phosphorus Detector

NPD - Nitrogen Phosphorus Detector NPD - Nitrogen Phosphorus Detector NPD Overview - The NPD detector is highly selective to Nitrogen and Phosphorus containing compounds. A precisely controlled Hydrogen and Air mixture surrounds the NPD

More information

Maximizing HPLC Speed and Resolution with the new 1100 DAD SL

Maximizing HPLC Speed and Resolution with the new 1100 DAD SL Welcome to our e-seminar: Maximizing HPLC Speed and Resolution with the new 1100 DAD SL Presenter: Dr. Stefan Schuette Outline Part 1 What do I need to do to maximize HPLC speed? Part 2 What is the performance

More information

Synthesis of Novel Peptide Linkers: Simultaneous Cyclization and Labeling

Synthesis of Novel Peptide Linkers: Simultaneous Cyclization and Labeling UPPRTING INFRMATIN ynthesis of Novel Peptide Linkers: imultaneous Cyclization and Labeling Gajanan K. Dewkar, Pedro B. Carneiro, Matthew C. T. Hartman* Department of Chemistry and Massey Cancer Center,

More information

Micro Analytical Instruments - A System Approach. Jörg Müller Institut für Mikrosystemtechnik

Micro Analytical Instruments - A System Approach. Jörg Müller Institut für Mikrosystemtechnik Micro Analytical Instruments - A System Approach Jörg Müller Institut für Mikrosystemtechnik Technische h Universität i Hamburg-Harburg H b Outline Motivation Political Political Issues Design Principles

More information

Project Planning Considerations For Emerging Contaminants. Ed Corl NAVSEA LQAO

Project Planning Considerations For Emerging Contaminants. Ed Corl NAVSEA LQAO Project Planning Considerations For Emerging Contaminants Ed Corl NAVSEA LQAO Report Documentation Page Form Approved OMB No. 0704-0188 Public reporting burden for the collection of information is estimated

More information

"What s inside your car?

What s inside your car? "What s inside your car? Car interior emissions testing using Thermal desorption GC/MS analysis Inge de Dobbeleer EU GCMS specialist ThermoFisher Lara Kelly Sales Manager Markes Thermal desorption The

More information

Supporting Information

Supporting Information Supporting Information rganocatalytic Mitsunobu Reactions Tracy Yuen Sze But and Patrick H. Toy * Department of Chemistry, The University of Hong Kong, Pokfulam Road, Hong Kong, People s Republic of China

More information

Examining Comprehensive Flow-Modulated Two-Dimensional Gas Chromatography

Examining Comprehensive Flow-Modulated Two-Dimensional Gas Chromatography Examining Comprehensive Flow-Modulated Two-Dimensional Gas Chromatography Roger L. Firor Senior Applications Chemist, Agilent Technologies roger_fi ror@agilent.com 1 Agilent Measurement Journal Gas chromatography

More information

PHOTOCOUPLER PS2505-1,-2,-4,PS2505L-1,-2,-4

PHOTOCOUPLER PS2505-1,-2,-4,PS2505L-1,-2,-4 PHOTOCOUPLER PS25-1,-2,-4,PS25L-1,-2,-4 HIGH ISOLATION VOLTAGE AC INPUT RESPONSE TYPE MULTI PHOTOCOUPLER SERIES NEPOC Series DESCRIPTION The PS25-1, -2, -4 and PS25L-1, -2, -4 are optically coupled isolators

More information

PHOTOCOUPLER PS2506-1,-2,-4,PS2506L-1,-2,-4

PHOTOCOUPLER PS2506-1,-2,-4,PS2506L-1,-2,-4 PHOTOCOUPLER PS2506-1,-2,-4,PS2506L-1,-2,-4 HIGH ISOLATION VOLTAGE AC INPUT, DARLINGTON TRANSISTOR TYPE MULTI PHOTOCOUPLER SERIES NEPOC Series DESCRIPTION The PS2506-1, -2, -4 and PS2506L-1, -2, -4 are

More information

New Certified Reference Materials for Chromatography

New Certified Reference Materials for Chromatography New Certified Reference Materials for Chromatography Dr. Frank Michel Frank.michel@sial.com sigma-aldrich.com/analytical 2012 Sigma-Aldrich Co. All rights reserved. Use of Certified Reference Materials

More information

Flash chromatography. MN Flash adsorbents a unique variety of phases. Separation efficiency and reproducibility

Flash chromatography. MN Flash adsorbents a unique variety of phases. Separation efficiency and reproducibility Flash chromatography MN Flash adsorbents a unique variety of phases Flash columns and cartridges from MACHEREY-NAGEL are available with all CHROMABOND SPE / Flash packings (more than 40 phases, e.g., C

More information

Rapid Resolution High Definition LC Columns

Rapid Resolution High Definition LC Columns Rapid Resolution High Definition LC Columns New Columns for Use with the 1290 Infinity LC Edward Kim Application Engineer August 11, 2010 Page 1 What are Rapid Resolution High Definition (RRHD) Columns?

More information

Site-specific Bioaccumulation of PCDD/PCDFs in Mothers and their Infants Living in Vicinity of Bien Hoa Airbase, Southern Vietnam

Site-specific Bioaccumulation of PCDD/PCDFs in Mothers and their Infants Living in Vicinity of Bien Hoa Airbase, Southern Vietnam 4 th International Forum on Sustainable Future in Asia 4 th NIES International Forum Site-specific Bioaccumulation of PCDD/PCDFs in Mothers and their Infants Living in Vicinity of Bien Hoa Airbase, Southern

More information

Volatile methyl siloxanes

Volatile methyl siloxanes Volatile methyl siloxanes Carrier in antiperspirants/deodorants Shampoo, conditioners, cosmetics VC exempt cleaning solvents Estimated global emissions to the atmosphere of >30000 tonnes/yr not including

More information

Capillary Flow Technology: GCxGC Flow Modulator GET A SECOND DIMENSION OF INFORMATION ON COMPLEX MIXTURES

Capillary Flow Technology: GCxGC Flow Modulator GET A SECOND DIMENSION OF INFORMATION ON COMPLEX MIXTURES Capillary Flow Technology: GCxGC Flow Modulator GET A SECOND DIMENSION OF INFORMATION ON COMPLEX MIXTURES Maximize information while collecting and analyzing GC data Comprehensive two-dimensional GC, or

More information

QUANTOF. High-resolution, accurate mass, quantitative time-of-flight MS technology

QUANTOF. High-resolution, accurate mass, quantitative time-of-flight MS technology QUANTOF High-resolution, accurate mass, quantitative time-of-flight MS technology Orthogonal-acceleration time-of-flight (oatof) mass spectrometers are invaluable tools for the detection and identification

More information

CONTAMINANTES ORGANICOS PERSISTENTES EN FAUNA SILVESTRE: PERSPECTIVA LATINOAMERICANA. Miguel A. Mora, Ph.D. Professor Texas A&M University

CONTAMINANTES ORGANICOS PERSISTENTES EN FAUNA SILVESTRE: PERSPECTIVA LATINOAMERICANA. Miguel A. Mora, Ph.D. Professor Texas A&M University CONTAMINANTES ORGANICOS PERSISTENTES EN FAUNA SILVESTRE: PERSPECTIVA LATINOAMERICANA Miguel A. Mora, Ph.D. Professor Texas A&M University Birds as Indicator Species Top of the food chain Photo: The Peregrine

More information

Dioxin Photoproducts of Triclosan and Its Chlorinated Derivatives in Sediment Cores. Chuck Sueper Pace Analytical Services, Inc.

Dioxin Photoproducts of Triclosan and Its Chlorinated Derivatives in Sediment Cores. Chuck Sueper Pace Analytical Services, Inc. Dioxin Photoproducts of Triclosan and Its Chlorinated Derivatives in Sediment Cores Chuck Sueper Pace Analytical Services, Inc. Background Anti-bacterial Soap 1980 1990 2000 2010 Background Anti-bacterial

More information

An Inclined Plane. Experiment OBJECTIVES MATERIALS

An Inclined Plane. Experiment OBJECTIVES MATERIALS Dual-Range Force Sensor An Inclined Plane Experiment 22 An inclined plane is a slanted surface used to raise objects. The sloping floor of a theater, a road over a mountain, and a ramp into a building

More information

A Review of Basic Concepts in Comprehensive Two-Dimensional Gas Chromatography

A Review of Basic Concepts in Comprehensive Two-Dimensional Gas Chromatography A Review of Basic Concepts in Comprehensive Two-Dimensional Gas Chromatography Ruby C.Y. Ong and Philip J. Marriott* Centre for Chromatography and Molecular Separations, Royal Melbourne Institute of Technology,

More information

The result of first SF6 inter-comparison experiment (SICE) from 2016 to 2017

The result of first SF6 inter-comparison experiment (SICE) from 2016 to 2017 The result of first SF6 inter-comparison experiment (SICE) from 2016 to 2017 Haeyoung LEE*, Sumin KIM, Hong-woo CHEO, Sang-Boom RYOO World Calibration Centre for SF6 National Institute of Meteorological

More information

Welcome! Denver, CO Houston, TX February 2009

Welcome! Denver, CO Houston, TX February 2009 Welcome! Denver, CO Houston, TX February 2009 Schedule & Introductions Announcements - Safety - Evaluation forms - Upcoming events > CBMSS, Feb 10 th, Nat l Jewish Center, www.cbmss.org (also Apr 22 nd

More information

High resolution ion mobility-mass spectrometry for separation and identification of isomeric lipids

High resolution ion mobility-mass spectrometry for separation and identification of isomeric lipids Electronic Supplementary Material (ESI) for Analyst. This journal is The Royal Society of Chemistry 2015 High resolution ion mobility-mass spectrometry for separation and identification of isomeric lipids

More information

Agilent 1260 Infinity Variable Wavelength Detector

Agilent 1260 Infinity Variable Wavelength Detector Agilent 1260 Infinity Variable Wavelength Detector Features, Technical Details, Specifications and Ordering Details Highest sensitivity for HPLC and RRLC The Agilent 1260 Infinity Variable Wavelength Detector

More information

HIGH ISOLATION VOLTAGE SINGLE TRANSISTOR TYPE MULTI PHOTOCOUPLER SERIES

HIGH ISOLATION VOLTAGE SINGLE TRANSISTOR TYPE MULTI PHOTOCOUPLER SERIES PHOTOCOUPLER PS2561A-1,PS2561AL-1,PS2561AL1-1,PS2561AL2-1 HIGH ISOLATION VOLTAGE SINGLE TRANSISTOR TYPE MULTI PHOTOCOUPLER SERIES NEPOC Series DESCRIPTION The PS2561A-1 is an optically coupled isolator

More information

HIGH ISOLATION VOLTAGE 4-PIN SOP PHOTOCOUPLER

HIGH ISOLATION VOLTAGE 4-PIN SOP PHOTOCOUPLER DESCRIPTION HIGH ISOLATION VOLTAGE 4-PIN SOP PHOTOCOUPLER PHOTOCOUPLER NEPOC Series The is an optically coupled isolator containing a GaAs light emitting diode and an NPN silicon phototransistor. This

More information

Agilent GC-MS Maintenance:

Agilent GC-MS Maintenance: Restek GC Accessories Agilent GC-MS Maintenance: Restek s Quick Reference Guide Replacement parts for Agilent ion sources now available. Optimize performance by upgrading ion source components and electron

More information

2014 Chromatography Supplies & Instruments Catalog

2014 Chromatography Supplies & Instruments Catalog 2014 Chromatography Supplies & Instruments Catalog http://www.scientifix.com.au ken@scientifix.com.au Fax: +61 (0)3 95437827 Page 1 Table of Contents Table Of Contents / How To Order.... 2 Who Is Visacon

More information

DMN3032LFDB. Features and Benefits. Product Summary. Description and Applications. Mechanical Data. Ordering Information (Note 4) Marking Information

DMN3032LFDB. Features and Benefits. Product Summary. Description and Applications. Mechanical Data. Ordering Information (Note 4) Marking Information YM DUAL N-CHANNEL ENHANCEMENT MODE MOSFET Product Summary BV DSS 3V R DS(ON) Max I D Max T A = +5 C 3mΩ @ V GS = V 6.A 4mΩ @ V GS = 4.5V 5.A Description and Applications This MOSFET is designed to minimize

More information

OPTIMASS 1000 Technical Datasheet

OPTIMASS 1000 Technical Datasheet OPTIMASS 1000 Technical Datasheet Sensor for mass flow First choice for universal applications Best price-performance ratio A wide range of options available with no restrictions The documentation is only

More information

The thin wall ESI tip appears unusually. advantageous for ensuring stable ESI. conditions and efficient ion transfers.

The thin wall ESI tip appears unusually. advantageous for ensuring stable ESI. conditions and efficient ion transfers. The thin wall ESI tip appears unusually advantageous for ensuring stable ESI conditions and efficient ion transfers. Valaskovic, Kelleher, McLafferty Science (1996), 273, 1199 perfect zero-dead-volume

More information

Column Chromatography System

Column Chromatography System Column Chromatography System ATTO Column Chromatography Equipment Purpose Purification of biosynthetic protein (Affinity chromato graphy, Ion exchange chromato graphy, etc.) Separation/purification of

More information

Precise Counting Scale GC. User Manual SNOWREX INTERNATIONAL CO., LTD. SRGC

Precise Counting Scale GC. User Manual SNOWREX INTERNATIONAL CO., LTD. SRGC Precise Counting Scale GC User Manual SNOWREX INTERNATIONAL CO., LTD. SRGC 20100415 Table of Contents Table of Contents...1 Specifications... 2 Basic specification... 2 Series specification(ec TYPE/OIML

More information

System Parameter and Performance Comparison Between Agilent 7697A and Agilent G1888A Headspace Samplers for USP <467>

System Parameter and Performance Comparison Between Agilent 7697A and Agilent G1888A Headspace Samplers for USP <467> System Parameter and Performance Comparison Between Agilent 7697A and Agilent G888A Headspace Samplers for USP Technical Overview Introduction The Agilent 7697A Headspace Sampler introduced onboard

More information

Automated Orthogonal Control System for Electrospray Ionization Mass Spectrometry

Automated Orthogonal Control System for Electrospray Ionization Mass Spectrometry utomated Orthogonal Control System for Electrospray Ionization Mass Spectrometry Gary. Valaskovic 1, James P. Murphy III 1, Mike S. Lee 2 1 New Objective Inc, Woburn, M 2 Milestone Development Services,

More information

PHOTOCOUPLER PS2801A-1,PS2801A-4

PHOTOCOUPLER PS2801A-1,PS2801A-4 PHOTOCOUPLER PS280-1,PS280-4 HIGH ISOLATION VOLTAGE SSOP PHOTOCOUPLER NEPOC Series DESCRIPTION The PS280-1 and PS280-4 are optically coupled isolators containing a GaAs light emitting diode and an NPN

More information

What is an image? Bernd Girod: EE368 Digital Image Processing Pixel Operations no. 1. A digital image can be written as a matrix

What is an image? Bernd Girod: EE368 Digital Image Processing Pixel Operations no. 1. A digital image can be written as a matrix What is an image? Definition: An image is a 2-dimensional light intensity function, f(x,y), where x and y are spatial coordinates, and f at (x,y) is related to the brightness of the image at that point.

More information

LONG TERM STATISTICS OF X-RAY SPECTROMETERS

LONG TERM STATISTICS OF X-RAY SPECTROMETERS 403 LONG TERM STATISTICS OF X-RAY SPECTROMETERS J. F. Dlouhy*, D. Mathieu Department of the Environment, Environmental Technology Center, River Road, Ottawa, Ontario, Canada Kl A OH3 K. N. Stoev Bulgarian

More information

Supporting Information

Supporting Information Supporting Information Admittance and Conductance Detection in the Capillary Scale. Weixiong Huang, Bikash Chouhan, Purnendu K. Dasgupta* Department of Chemistry and Biochemistry, The University of Texas

More information

Power supply. Calibration. Surface finishing. Seal. Process connection

Power supply. Calibration. Surface finishing. Seal. Process connection Page 1 of 6 Description The measuring system is designed for flow measurement. Gas flow through the sensing section passes two PT 100 RTD transducers; one is used as a temperature sensing device whilst

More information

Fabrication Techniques of Optical ICs

Fabrication Techniques of Optical ICs Fabrication Techniques of Optical ICs Processing Techniques Lift off Process Etching Process Patterning Techniques Photo Lithography Electron Beam Lithography Photo Resist ( Microposit MP1300) Electron

More information

FOR OPTICAL DAA, HIGH LINEAR 16-PIN SOP PHOTOCOUPLER

FOR OPTICAL DAA, HIGH LINEAR 16-PIN SOP PHOTOCOUPLER PHOTOCOUPLER PS8741 FOR OPTICAL DAA, HIGH LINEAR 16-PIN SOP PHOTOCOUPLER NEPOC Series DESCRIPTION The PS8741 is an optically coupled isolator containing a GaAs LED on the input side and two photodiodes

More information

FOR OPTICAL DAA, HIGH LINEAR 16-PIN SOP PHOTOCOUPLER

FOR OPTICAL DAA, HIGH LINEAR 16-PIN SOP PHOTOCOUPLER DESCRIPTION FOR OPTICAL DAA, HIGH LINEAR 16-PIN SOP PHOTOCOUPLER PHOTOCOUPLER PS8741 NEPOC Series The PS8741 is an optically coupled isolator containing a GaAs LED on the input side and two photodiodes

More information

3V OPERATION SILICON LDMOSFET RF POWER AMPLIFIER INTEGRATED CIRCUIT FOR 1.9 GHz PHS AND 2.4 GHz APPLICATIONS

3V OPERATION SILICON LDMOSFET RF POWER AMPLIFIER INTEGRATED CIRCUIT FOR 1.9 GHz PHS AND 2.4 GHz APPLICATIONS DESCRIPTION The µpd572tu is a silicon laterally diffused (LD) MOSFET IC designed for use as power amplifier 1.9 GHz PHS and 2.4 GHz applications. This IC consists of two stage amplifiers. The device is

More information

UCT SELECTRA C18 HPLC COLUMN INNOVATION THROUGH CHEMISTRY

UCT SELECTRA C18 HPLC COLUMN INNOVATION THROUGH CHEMISTRY UCT SELECTRA C18 HPLC COLUMN INNOVATION THROUGH CHEMISTRY SELECTRA C18 HPLC COLUMNS Benefits Good choice for easily converting current C18 methods Suitable for most traditional reverse phase analyses High

More information

The 5975C Series MSDs: Normalized Instrument Tuning. Technical Overview. Introduction

The 5975C Series MSDs: Normalized Instrument Tuning. Technical Overview. Introduction The 5975C Series MSDs: Normalized Instrument Tuning Technical Overview Jeffrey T. Kernan and Harry Prest Introduction The automated optimization of ion detection in a mass spectrometer, commonly referred

More information

2.0 Absolute Maximum Ratings 1 Parameter Min. Max. Unit.

2.0 Absolute Maximum Ratings 1 Parameter Min. Max. Unit. RFPO45 G.8263 1.0 Specification References a. Rakon part number (Doc ID) M6635LF (514241) b. 25.0MHz MERC 9x7 HOT CMOS 3.3V c. Version 1 (2014-04-01) d. Package 9.7 x 7.5 x 4.3 mm max. 2.0 Absolute Maximum

More information

SCIEX Presentation. Name of Speaker, Identification. QTRAP Functionalities. Alexandre Paccou Name of Speaker, Identification. SCIEX Presentation.

SCIEX Presentation. Name of Speaker, Identification. QTRAP Functionalities. Alexandre Paccou Name of Speaker, Identification. SCIEX Presentation. SCIEX Presentation Alexandre Paccou Name of Speaker, Identification Date SCIEX Presentation QTRAP Functionalities Hardware and Software Tipps und Tricks, Name of Speaker, Identification Date For Research

More information

Chem466 Lecture Notes. Spring, 2004

Chem466 Lecture Notes. Spring, 2004 Chem466 Lecture Notes Spring, 004 Overview of the course: Many of you will use instruments for chemical analyses in lab. settings. Some of you will go into careers (medicine, pharmacology, forensic science,

More information

File E Project 10ME Issued: August 26, 2010 Revised: October 27, 2012 REPORT

File E Project 10ME Issued: August 26, 2010 Revised: October 27, 2012 REPORT File E322469 Project 10ME03376 Issued: August 26, 2010 Revised: October 27, 2012 REPORT on COMPONENT - DRIVERS FOR LIGHT-EMITTING DIODES ARRAYS, MODULES AND CONTROLLERS Lutron Electronics Co Inc Coopersburg,

More information

LED Module. Data Sheet. Finger-64LED Module. Square type (Finger-SQ64B) SI-B 8V SI-B8U SI-B8T SI-B8R SI-B 8P

LED Module. Data Sheet. Finger-64LED Module. Square type (Finger-SQ64B) SI-B 8V SI-B8U SI-B8T SI-B8R SI-B 8P 3.0 1 / 10 Data Sheet Finger-64LED Module Model Name Type 700mA, 12V CCT Square type (Finger-SQ64B) Rectangular type (Finger-RT64B) Parts No. 3000K 3500K 4000K 5000K 6500K

More information

PHOTOCOUPLER PS2805-1,PS2805-4

PHOTOCOUPLER PS2805-1,PS2805-4 PHOTOCOUPLER PS2805-1,PS2805-4 HIGH ISOLATION VOLTAGE AC INPUT RESPONSE TYPE SSOP PHOTOCOUPLER NEPOC Series DESCRIPTION The PS2805-1 and PS2805-4 are optically coupled isolators containing GaAs light emitting

More information

HIGH CMR, 10 Mbps OPEN COLLECTOR OUTPUT TYPE 5-PIN SOP PHOTOCOUPLER

HIGH CMR, 10 Mbps OPEN COLLECTOR OUTPUT TYPE 5-PIN SOP PHOTOCOUPLER DESCRIPTION HIGH CMR, 1 Mbps OPEN COLLECTOR OUTPUT TYPE 5-PIN SOP PHOTOCOUPLER PHOTOCOUPLER PS9114 NEPOC Series The PS9114 is an optically coupled high-speed, isolator containing a GaAlAs LED on the input

More information

Ion Beam Lithography: faster writing strategies for features between 150nm and 1um

Ion Beam Lithography: faster writing strategies for features between 150nm and 1um Ion Beam Lithography: faster writing strategies for features between 150nm and 1um Brent P. Gila, Andes Trucco, David Hays Located in sunny Gainesville, FL (100 miles north of Disney World) https://nrf.aux.eng.ufl.edu/

More information

Optimisation of LCMS Techniques for Complementary Medicines

Optimisation of LCMS Techniques for Complementary Medicines Optimisation of LCMS Techniques for Complementary Medicines Ashley Dowell Southern Cross Plant Science Southern Cross University Complementary Medicines Complementary medicines vitas erals herbal medicines

More information

FAPAS Report Pesticide Residues in Green Tea. January February Page 1 of 32

FAPAS Report Pesticide Residues in Green Tea. January February Page 1 of 32 FAPAS Report 19162 Pesticide Residues in Green Tea N O T C O N TR O LL ED W H EN PR IN TE D January February 2014 Page 1 of 32 PARTICIPANT LABORATORY NUMBER Participants can log in to FAPAS SecureWeb at

More information

Haruo Saito. National Institute of Information and Communications Technology

Haruo Saito. National Institute of Information and Communications Technology Calibration system at NICT Haruo Saito National Institute of Information and Communications Technology Organization of NICT Content Calibration system Calibration system Carried in system and remote system

More information

File E Project 10ME Issued: August 26, 2010 Revised: October 27, 2012 REPORT

File E Project 10ME Issued: August 26, 2010 Revised: October 27, 2012 REPORT File E322469 Project 10ME03376 Issued: August 26, 2010 Revised: October 27, 2012 REPORT on COMPONENT - DRIVERS FOR LIGHT-EMITTING DIODES ARRAYS, MODULES AND CONTROLLERS Lutron Electronics Co Inc Coopersburg,

More information

4728 Gravois Ave. St. Louis, MO SIMAX

4728 Gravois Ave. St. Louis, MO SIMAX http://www.stemmerich.com 4728 Gravois Ave. St. Louis, MO 63116 314-832-7726 SALES 800-325-9528 FAX 314-832-7799 SIMAX DESCRIPTION: In 1837 The Kavalier Glassworks was established, and it has been making

More information

HIGH CMR, 10 Mbps TOTEM POLE OUTPUT TYPE 5-PIN SOP PHOTOCOUPLER

HIGH CMR, 10 Mbps TOTEM POLE OUTPUT TYPE 5-PIN SOP PHOTOCOUPLER PHOTOCOUPLER PS9115 HIGH CMR, 1 Mbps TOTEM POLE OUTPUT TYPE 5-PIN SOP PHOTOCOUPLER NEPOC Series DESCRIPTION The PS9115 is an optically coupled high-speed, totem pole output isolator containing a GaAlAs

More information

Detection and Quantitation of Brominated and Chlorinated Hydrocarbons by DART with Linear Ion Trap and Triple Quadrupole Technology

Detection and Quantitation of Brominated and Chlorinated Hydrocarbons by DART with Linear Ion Trap and Triple Quadrupole Technology Detection and Quantitation of Brominated and Chlorinated Hydrocarbons by DART with Linear Ion Trap and Triple Quadrupole Technology Mary L. Blackburn 1, Scott J. Harrison 2, and Maria C. Prieto Conaway

More information

JANUARY 30, 2012 TEST REPORT #211630B MIXED FLOWING GAS TESTING IPS1/IPT1 CONNECTOR SERIES SAMTEC, INC.

JANUARY 30, 2012 TEST REPORT #211630B MIXED FLOWING GAS TESTING IPS1/IPT1 CONNECTOR SERIES SAMTEC, INC. JANUARY 30, 2012 TEST REPORT #211630B MIXED FLOWING GAS TESTING IPS1/IPT1 CONNECTOR SERIES SAMTEC, INC. APPROVED BY: DOMINIC ARPINO PROJECT ENGINEERING MANAGER CONTECH RESEARCH, INC. ATTLEBORO, MA Contech

More information

Noise Calibration Systems and Accessories DATA SHEET / 4N-062

Noise Calibration Systems and Accessories DATA SHEET / 4N-062 Noise Calibration Systems and Accessories DATA SHEET / 4N-062 // MARCH 2018 Noise Calibration Systems and Components MT7149J99 WR10 75 110 GHz Noise Calibration System Introduction The Maury Noise Calibration

More information

Advancing EDS Analysis in the SEM Quantitative XRF. International Microscopy Congress, September 5 th, Outline

Advancing EDS Analysis in the SEM Quantitative XRF. International Microscopy Congress, September 5 th, Outline Advancing EDS Analysis in the SEM with in-situ Quantitative XRF Brian J. Cross (1) & Kenny C. Witherspoon (2) 1) CrossRoads Scientific, El Granada, CA 94018, USA 2) ixrf Systems, Inc., Houston, TX 77059,

More information

FiberLight Miniature UV-Vis Light Source for Mobile Spectroscopy and Online Process Control

FiberLight Miniature UV-Vis Light Source for Mobile Spectroscopy and Online Process Control FiberLight Miniature UV-Vis Light Source for Mobile Spectroscopy and Online Process Control FiberLight Miniature UV-Vis Light Source FiberLight DTM 6/10 FiberLight is a compact UV-Vis light source designed

More information

HIGH COLLECTOR TO EMITTER VOLTAGE SOP MULTI PHOTOCOUPLER

HIGH COLLECTOR TO EMITTER VOLTAGE SOP MULTI PHOTOCOUPLER DESCRIPTION PHOTOCOUPLER PS2732-1,PS2733-1 HIGH COLLECTOR TO EMITTER VOLTAGE SOP MULTI PHOTOCOUPLER NEPOC Series The PS2732-1 and PS2733-1 are optically coupled isolators containing a GaAs light emitting

More information

Useful GC/MS Sensitivity for Improved Environmental Analyses. Terry L. Sheehan, Ph.D. GC/MS Senior Product Manager Santa Clara, CA 7 February 2008

Useful GC/MS Sensitivity for Improved Environmental Analyses. Terry L. Sheehan, Ph.D. GC/MS Senior Product Manager Santa Clara, CA 7 February 2008 Useful GC/MS Sensitivity for Improved Environmental Analyses Terry L. Sheehan, Ph.D. GC/MS Senior Product Manager Santa Clara, CA 7 February 2008 Definitions Sensitivity Analytical definition: slope of

More information

DISCONTINUED LASER DIODE NX8311UD nm FOR LONG HAUL 2.5 Gb/s InGaAsP MQW-DFB LASER DIODE TOSA DESCRIPTION FEATURES

DISCONTINUED LASER DIODE NX8311UD nm FOR LONG HAUL 2.5 Gb/s InGaAsP MQW-DFB LASER DIODE TOSA DESCRIPTION FEATURES DESCRIPTION 1 310 nm FOR LONG HAUL.5 Gb/s InGaAsP MQW-DFB LASER DIODE TOSA The NX8311UD is a 1 310 nm Multiple Quantum Well (MQW) structured Distributed Feed-Back (DFB) laser diode TOSA (transmitter optical

More information

MSA XCell Sensors. Because every life has a purpose...

MSA XCell Sensors. Because every life has a purpose... MSA XCell Sensors Because every life has a purpose... XCell Sensors MSA XCell Sensors are a breakthrough in sensor design, enabling faster response and shorter span calibrations, saving you time and money.

More information

Features. Drain SOT23 D. Gate. Source. Part Number Case Packaging DMG3414UQ-7 SOT23 3,000/Tape & Reel DMG3414UQ-13 SOT23 10,000/Tape & Reel

Features. Drain SOT23 D. Gate. Source. Part Number Case Packaging DMG3414UQ-7 SOT23 3,000/Tape & Reel DMG3414UQ-13 SOT23 10,000/Tape & Reel N-CHANNEL ENHANCEMENT MODE MOSFET Product Summary Features V (BR)DSS 2V Description R DS(ON) max 25mΩ @ V GS = 4.5V I D max T A = +25 C 9A 29mΩ @ V GS = 2.5V 5.5A 37mΩ @ V GS = 1.8V 4.8A This MOSFET is

More information

A Residual Gas Analyzer for Dry Etching Process

A Residual Gas Analyzer for Dry Etching Process FFeature Article Article Makoto MATSUHAMA Concerning the dry process of the semiconductor device manufacturing, the monitoring of etching chamber conditions (pressure, temperature, gas concentration,...)

More information

Ground Fault Interrupter Earth Leakage Current Detector IK7104 TECHNICAL DATA

Ground Fault Interrupter Earth Leakage Current Detector IK7104 TECHNICAL DATA TECHNICAL DATA Ground Fault Interrupter Earth Leakage Current Detector IK7104 GENERAL DESCRIPTION The IL7104 is designed for use in earth leakage circuit interrupters for operation directly off the AC

More information

Supporting Information

Supporting Information Momiyama, Kanan, Liu page S1 Synthesis of Acyclic!,"-Unsaturated Ketones via Pd(II)-Catalyzed Intermolecular Reaction of Alkynamides and Alkenes Norie Momiyama, Matthew W. Kanan and David R. Liu* Department

More information

Finer Points of ICP-OES Setup and Operation

Finer Points of ICP-OES Setup and Operation Finer Points of ICP-OES Setup and Operation (Part 1) James Bartos Office of Indiana State Chemist Challenges with Fertilizers Broad conc ranges from low ppm to upper % level Want to test all nutrient elements

More information

IN COAXIAL PACKAGE FOR 2.5Gb/s APPLICATIONS

IN COAXIAL PACKAGE FOR 2.5Gb/s APPLICATIONS FEATURES NEC's φ50 µm InGaAs APD IN COAXIAL PACKAGE FOR 2.5Gb/s APPLICATIONS SMALL DARK CURRENT: ID = 7 na HIGH SENSITIVITY: S = 0.94 A/W at λ = 0 nm, M = S = 0.96 A/W at λ = 550 nm, M = HIGH SPEED RESPONSE:

More information

United Nations Stockholm Convention: PFOA Restrictions Sanjay Baliga, SEMI Global Headquarters

United Nations Stockholm Convention: PFOA Restrictions Sanjay Baliga, SEMI Global Headquarters United Nations Stockholm Convention: PFOA Restrictions Sanjay Baliga, SEMI Global Headquarters IHTESH Conference, Hsinchu, Taiwan (Friday, June 9, 2017) Presentation Agenda PFOA and Related Compounds Basics

More information