1 Page 1 of 6 Description The measuring system is designed for flow measurement. Gas flow through the sensing section passes two PT 100 RTD transducers; one is used as a temperature sensing device whilst the other is used as a heater. The temperature sensor monitors the actual process value whilst the heater is maintained at a constant differential temperature above this by varying the power consumed by the sensor. The greater the mass flow, the greater the cooling effect and power required to maintain the differential temperature. The measured heater power is therefore a measure of the gas flow rate and is evaluated as a measuring signal using the transmitter electronics. The measuring systems consists of a transmitter and a sensor which can be mounted together (compact version) or separate (remote version). The electronic of the transmitter (type T-mass 6*) mounted inside the enclosure Proline G02. Type designation: T-MASS 6 * F ** - * * * * * * * * * * * * T-MASS 6 * I -* * * * * * * * * * * * * * Signal output A, B, C, D, E, H, J, K, L, M, N, P, Q, V, W, 0, 2, 3, 4, 5, 6, 7, 8, 9 = non-intrinsically safe outputs F = Ex ia PROFIBUS PA G = Ex ia Foundation Fieldbus R = Ex ia HART current output, active S = Ex ia HART current output, frequency output, active T = Ex ia HART current output, frequency output, passive U = Ex ia HART current output, passive Software Power supply Cable entry Housing A = Compact version 1 = Compact version, Ta -40 C G = Remote version 7 = Remote version, Ta -40 C Approvals B = Ex d [ia] IIC, Ex td A21 D = Ex de [ia] IIC, Ex td A21 N = Ex d [ia] IIC, Ex td A21 U = Ex d [ia] IIC, Ex td A21 3 = Ex d [ia] IIC, Ex td A21 5 = Ex de [ia] IIC, Ex td A21 Certificate Calibration Surface finishing Seal Process connection Material of tube / sensor Nominal diameter (type of sensor = F) or Insertion tube length (type of sensor = I) Type of electronic
2 Page 2 of 6 Parameters 1 Power supply (terminals no. 1 (L/+) and 2 (N/-) Nominal Voltage pure AC version AC 85 to 260 V DC/AC version DC 16 to 62 V or AC 20 to 55 V max. voltage Um AC 260 V 2 Signal circuits 2.1 Non-intrinsically safe signal circuits Type Terminal No. Safety parameters ***********p *************p with p = A, B, C, D, E, H, J, K, L, M, N, P, Q, V, W, 0, 2, 3, 4, 5, 6, 7, 8 or U m = 260 V, I max = 0.5A The service connector is limited for use with Endress+Hauser interfaces only, Um = 15 VDC. 2.2 Intrinsically safe signal circuits Type ***********F *************G Terminal No. Safety parameters Ex ia IIC Ii = 600 ma Pi = 8.5 W Ci 5 nf Li 10 µh
3 Page 3 of 6 Type ***********R *************R ***********S *************S ***********T *************T ***********U *************U Terminal No. Safety parameters Ex ia IIC Co = 0.16 µf Lo = 4.1 mh Co = 0.16 µf Lo = 4.1 mh Ex ia IIB Uo = 21.8 V Io = 90 ma Po = 491 mw Co = 1.16 µf Lo = 15 mh Ii = 10 ma Pi = 300 mw Ii = 500 ma Pi = 600 mw Uo = 21.8 V Io = 90 ma Po = 491 mw Co = 1.16 µf Lo = 15 mh Ii = 10 ma Pi = 300 mw Ii = 500 ma Pi = 600 mw Ii = 100 ma Pi = 1.25 W Ii = 100 ma Pi = 1.25 W
4 Page 4 of Intrinsically safe sensor circuits Terminals / Plugs for remote version 41, 42, 43, 44 = Sensor circuit For the connection of the remote sensor a cable with a max. length of 100 m and the following max values has to be used: Cable Inductance 0.9 µh per meter Cable capacitance 1 nf per meter 2.4 Medium and ambient temperatures The relationship between maximum ambient, medium temperatures and temperature class are as follows (for compact versions and sensor remote versions): max. medium temperature [ C] Temp. Class T4 T3 T2 T1 Ta max = 55 C T-mass 6*F**-************ Ta max = 60 C T-mass 6*F**-************ T-mass 6*I-************** Max. ambient temperature transmitter remote version type T-mass 6*F**-*******(G,7)**** and type T-mass 6*I-*********(G,7)**** 60 C Min. ambient temperature T-mass 6*F**-*******A**** T-mass 6*F**-*******G**** T-mass 6*I-*********A**** T-mass 6*I-*********G**** T-mass 6*F**-*******1**** T-mass 6*F**-*******7**** T-mass 6*I-*********1**** T-mass 6*I-*********7**** -20 C -40 C
5 Page 5 of 6 Marking Name of the manufacturer or its trademark Serial number Certificate number Ambient temperature range Warning: For the transmitter a delay time of 10 minutes after switching off the power before opening the enclosure has to be regarded. Compact versions Type/Order code Marking Gas Marking Dust ****** 2)1) **** Ex d[ia] IIC T1-T4 ******** 2)1) **** ****** 3)1) **** Ex de[ia] IIC T1-T4 ******** 3)1) **** ********3 1) **** Ga/Gb Ex d[ia] IIC T1-T4 4) ********5 1) **** Ga/Gb Ex de[ia] IIC T1-T4 4) 1) 2) 3) 4) At this place A or 1 will be inserted At this place B, N, U or 3 will be inserted At this place D or 5 will be inserted Comment: Sensor mounted in the wall to Zone 0 For types ***********F, *************F, ***********G and *************G additional marking with: Fisco Field device Remote Versions Transmitter Type/Order code Marking Gas Marking Dust ****** 2)1) **** Ex d[ia] IIC T1-T4 ******** 2)1) **** ****** 3)1) **** ******** 3)1) **** Ex de[ia] IIC T1-T4 1) 2) 3) At this place G or 7 will be inserted At this place B, N, U or 3 will be inserted At this place D or 5 will be inserted
6 Page 6 of 6 Sensor Type/Order code Marking Gas Marking Dust ****** 2)1) **** Ex ia IIC T1-T4 ******** 2)1) **** ****** 3)1) **** ******** 3)1) **** Ga/Gb Ex ia IIC T1-T4 4) 1) 2) 3) 4) At this place G or 7 will be inserted At this place B, N, U or D will be inserted At this place 3 or 5 will be inserted Comment: Sensor mounted in the wall to Zone 0
Rev ECN Description Approval Date AA Release to Approvals RCS 10/9/12
EB-20023139 Revision: AA Number of Pages: 13 Comments: Originator: RCS 10/05/12 Approved: RCS 10/05/12 Rev ECN Description Approval Date AA 1049085 Release to Approvals RCS 10/9/12 Micro Motion, Inc. Page
More informationRev ECN Description Approval Date AA Release to Approvals RCS 10/9/12
EB-20023137 Revision: AA Number of Pages: 13 Comments: Originator: RCS 10/05/12 Approved: RCS 10/05/12 Rev ECN Description Approval Date AA 1049085 Release to Approvals RCS 10/9/12 Micro Motion, Inc. Page
More informationTÜV NORD CERT. TÜV NORD CERT GmbH & Co. KG Am TÜV 1 D Hannover. Annexe to IECEx TUN , issue No. 0 Page 1 of 5.
GmbH & Co. KG Page 1 of 5 Electrical data: Base devices for 6DR50**-*****-**** Motherboard L250 Power supply / control current 4-20 ma... series connection (terminals 6+ and 7/8) 2-wire circuit with Hart
More informationAnnex 1 to Certificate of Conformity IECEx KEM X Issue No. 8 Rota Yokogawa Rotamass 3 Series Coriolis Mass Flowmeter
Description RotaMass 3 Series Coriolis Mass Flow Meter; - Integral Transmitter, Model RCCx; - Remote Converter, Model RCCF31; - Remote Converter, Model RCCR31; - Remote Detector Model RCCS3xxxx The Coriolis
More informationMicro Motion Model 5700 Transmitter IECEx Zone 2/22 Installation Instructions EPL Gc and EPL Dc
IECEx Installation Instructions EB-20025744 Rev. AC December 2015 Micro Motion Model 5700 Transmitter IECEx Zone 2/22 Installation Instructions EPL Gc and EPL Dc 5700 IECEx Installation Instructions EB-20025744
More informationAttachment to Certificate IECEx PTB X
Applicant: Electrical Apparatus: KROHNE Ltd. Measuring converter, type MFC400F Description of equipment The measuring converter, type MFC400F is used as part of a mass flow measuring system to determine
More informationProline Promass 40. Safety Instructions. NEPSI Zone 1, Zone 21
Safety Instructions Proline Promass 40 NEPSI Zone 1, Zone 21 This document is an integral part of the following Operating Instructions: BA00061D, Proline Promass 40 HART Contents General warnings.......................................................................
More informationMicro Motion Model 5700 Transmitter ATEX Zone 2/22 Installation Instructions II 3 G or II 3 (2) G/EPL Gc & II 3 D/EPL Dc
ATEX Installation Instructions EB-20025742 Rev. AC December 2015 Micro Motion Model 5700 Transmitter ATEX Zone 2/22 Installation Instructions II 3 G or II 3 (2) G/EPL Gc & II 3 D/EPL Dc 5700 ATEX Installation
More informationCertificate of Conformity
Certificate Holder: ABB Automation Products GmbH Schillerstrasse 72 D-32425 Minden GERMANY Electrical Apparatus: Type of Protection: 2600T Pressure Transmitter Models 265../267../269.. Communication protocol:
More informationIECEx Hazardous Area Approvals Fisher FIELDVUE DVC6200 Series Digital Valve Controllers
Instruction Manual Supplement DVC6200 Digital Valve Controllers IECEx Hazardous Area Approvals Fisher FIELDVUE DVC6200 Series Digital Valve Controllers SIS Hazardous Area Approvals and Special Instructions
More informationATEX Installation Drawings and Instructions
Installation Instructions P/N 20004414, Rev. C September 2008 ATEX Installation Drawings and Instructions For ATEX-approved transmitter installations Note: For hazardous installations in Europe, refer
More informationProline Promass 83, 84
Safety Instructions Proline Promass 83, 84 NEPSI Zone 1, Zone 21 This document is an integral part of the following Operating Instructions: BA00059D, Proline Promass 83 BA00109D, Proline Promass 84 BA00063D,
More information2-WIRE UNIVERSAL TEMPERATURE TRANSMITTER
SAFE INSTALLATION MANUAL (FM APPROVAL) -WIRE UNIVERSAL TEMPERATURE TRANSMITTER (HART communication, intrinsically safe/explosion-proof) MODEL B6U/B6U-B BEFORE USE... SAFETY PRECAUTIONS This manual describes
More informationDPharp Differential Pressure and Pressure Transmitters
User s Manual DPharp Differential Pressure and Pressure s Manual Change No. 12-021-2 For the products of EJX and EJA-E series with any of the following option codes, please refer to this manual change
More informationIntrinsic Safety Control Drawing
s Intrinsic Safety Control Drawing A5E00455287 system in total "No revision to drawing without prior Approval by FM Approvals!" PLC PC Non hazardous area PROFIBUS DP RS 485-IS Coupler PROFIBUS RS 485-IS
More informationCERTIFICATE OF COMPLIANCE
FM Approvals 1151 Boston Providence Turnpike P.O. Box 9102 Norwood, MA 0202 USA T: 781 72 400 F: 781-72-975 www.fmapprovals.com CERTIFICATE OF COMPLIANCE HAZARDOUS (CLASSIFIED) LOCATION ELECTRICAL EQUIPMENT
More informationSTT 3000 Series STT170 SMART TEMPERATURE TRANSMITTER Models STT171, STT173, STT17H, STT17F, STT17C
STT 3000 Series STT170 SMART TEMPERATURE TRANSMITTER Models STT171, STT173, STT17H, STT17F, STT17C 34-TT-03-07 3/06 PRODUCT SPECIFICATION SHEET OVERVIEW The Honeywell STT170 series of programmable temperature
More informationSTT 3000 Series STT170 SMART TEMPERATURE TRANSMITTER Models STT171, STT173, STT17H, STT17F, STT17C
STT 3000 Series STT170 SMART TEMPERATURE TRANSMITTER Models STT171, STT173, STT17H, STT17F, STT17C 6/07 PRODUCT SPECIFICATION SHEET OVERVIEW The Honeywell STT170 series of programmable temperature transmitters
More informationCSA Hazardous Area Approvals Fisher FIELDVUE DVC6200 Series Digital Valve Controllers
Instruction Manual Supplement DVC6200 Digital Valve Controllers CSA Hazardous Area Approvals Fisher FIELDVUE DVC6200 Series Digital Valve Controllers SIS Hazardous Area Approvals and Special Instructions
More informationDescription. Annexe to: IECEx CML X Issue 1. Isolating Amplifier D461
October 13, 2017 Annexe to: IECEx CML 15.0063X Issue 1 Applicant: Apparatus: Braun GmbH Isolating Amplifier D461 Description The Isolating Amplifier D461 is an intrinsic safety associated apparatus
More informationIECEx Certificate of Conformity Annexe
for Certificate No.: IECEx TSA 13.0023X Issue No.: 0 Description: The Measuring Gauge of type series SMALL * / / / / / / / / / / * / * is an intrinsically safe supplied apparatus providing different electronic
More informationATEX Installation Instructions for Micro Motion CMF400 Sensors with Booster Amplifier
Installation Instructions P/N MMI-20010152, Rev. A June 2007 for Micro Motion CMF400 Sensors with Booster Amplifier For ATEX-approved sensor installations Note: For hazardous installations in Europe, refer
More informationCSA Hazardous Area Approvals Fisher FIELDVUE DVC6005 Series Remote Mount Digital Valve Controllers
Instruction Manual Supplement DVC6005 Digital Valve Controllers CSA Hazardous Area Approvals Fisher FIELDVUE DVC6005 Series Remote Mount Digital Valve Controllers Hazardous Area Approvals and Special Instructions
More informationEx instruction manual. Rotamass Coriolis mass flow meter. IECEx. IM 01U10X02-00EN-R_002, 2nd edition,
Ex instruction manual Rotamass Coriolis mass flow meter IM 01U10X02-00EN-R_002, 2nd edition, 2016-09-22 Table of contents Table of contents 1 Introduction... 4 1.1 Scope of application... 4 1.2 Applicable
More informationSafety Instructions Proline Promass 300
XA01507D/06/EN/01.16 71327925 Products Solutions Services Safety Instructions Proline Promass 300 Class I, Zone 2 AEx/Ex na ic IIC Class I, Zone 2, AEx/Ex na nc ic IIC Class I Division 2 C US Document:
More informationR88000 T (including ATEX)
INSTRUCTION MANUAL R88000 T88000 (including ATEX) Factory set transmitters are ready to use. Connect according the labels. Otherwise follow the calibration instructions. Input signals ṮC RTD CONTENTS:
More informationFM Hazardous Area Approvals Fisher FIELDVUE DVC6200 Series Digital Valve Controllers
Instruction Manual Supplement DVC6200 Digital Valve Controllers FM Hazardous Area Approvals Fisher FIELDVUE DVC6200 Series Digital Valve Controllers SIS Hazardous Area Approvals and Special Instructions
More informationInstallation Instructions P/N MMI , Rev. AA July ATEX Installation Instructions for Micro Motion Model 3500 Transmitters
Installation Instructions P/N MMI-20011707, Rev. AA July 2009 ATEX Installation Instructions for Micro Motion Model 3500 Transmitters Note: For hazardous installations in Europe, refer to standard EN 60079-14
More informationCertificate of Compliance
Certificate of Compliance Certificate: 70163240 (166720) Issued to: WIKA Alexander Wiegand SE & Co. KG Alexander-Wiegand-Str 30 Klingenberg, Bayern 63911 GERMANY Attention: Harald Gollas The products listed
More informationSafety Instructions Proline Promag 500
XA01524D/06/EN/01.16 71327947 Products Solutions Services Safety Instructions Proline Promag 500 Cl.I, II, III Div.1 for XP (Ex d Flameproofed version) C US Document: XA01524D Safety instructions for electrical
More informationAnalog Input Module HART Ex i / I.S. Inputs, Channels Type 9461/
> 4 channels for 2-wire HART transmitters and 4 channels for 4-wire HART transmitters > Intrinsically safe inputs Ex ia IIC > Galvanic separation between inputs and system > Open-circuit and short-circuit
More informationCertificate of Conformity
Applicant: Moore Industries International Inc. 16650 Schoenborn Street North Hills CA 91343-6196 U.S.A Electrical Apparatus: Fully Encapsulated PC-Programmable Temperature Transmitters (TRX: nonisolated;
More informationTemperature Input Module for Zone 1 Series 9482/32
www.stahl.de > 8 channels for temperature sensors > Intrinsically safe inputs Ex ia > For Pt-, Ni- and Cu-resistance temperature detectors according to DIN, IEC and GOST in 2-, 3- and 4-wire circuits >
More informationUser's Manual. ROTAMASS Total Insight Coriolis Mass Flow and Density Meter Explosion Proof Type Manual IECEx IM 01U10X02-00EN-R
User's Manual ROTAMASS Total Insight Coriolis Mass Flow and Density Meter Explosion Proof Type Manual IECEx IM 01U10X02-00EN-R IM 01U10X02-00EN-R, 3rd edition, 2017-11-16 Table of contents Table of contents
More informationIP26 Series Current to Pneumatic Converter
Product Specifications 06.2016 PSS EVE0403 A-(en) IP26 Series Current to Pneumatic Converter The IP26 (current to pressure) transducer converts a 4 20mA electrical current signal to a proportional linear
More informationSTT 3000 Smart Temperature Transmitter Series STT170 Specifications
STT 3000 Smart Temperature Transmitter Series STT170 Specifications Model STT171, STT173, STT17H, STT17F, STT17C 34-TT-03-07 November 2010 Overview STT17H Features The Honeywell STT170 series of programmable
More informationEx instruction manual. Rotamass Coriolis mass flow meter ATEX. IM 01U10X01-00EN-R_002, 2nd edition,
Ex instruction manual Rotamass Coriolis mass flow meter ATEX IM 01U10X01-00EN-R_002, 2nd edition, 2016-09-22 Table of contents Table of contents 1 Language variants of ATEX documentation... 4 2 Introduction...
More informationEmerson (Appleton Group ATX)
These assemblies are available in AC-powered and DC-powered versions. They have space to land 48 or 96 CHARM I/O depending on the model chosen. The field I/O can be specified as either IS or NON-IS I/O.
More informationOPTIMASS / 300 / 010 Guidelines for the use of Coriolis meters in hazardous areas
OPTIMASS / 300 / 010 Guidelines for the use of Coriolis meters in hazardous areas 1000 Series Twin Straight Tube Coriolis Mass Flowmeter 2000 Series Twin Straight Tube Coriolis Mass Flowmeter 3000 Series
More informationExpert 1400 / Expert 3400 Submersible Hydrostatic Level Transmitters
Expert 1400 / Expert 3400 Submersible Hydrostatic Level Transmitters ll 2G EEx ia llc T6 As our products are continuously improved, we reserve the right to make any change in the stated specifications
More informationUniversal Intelligent 2-wire In-head Transmitters
Universal Intelligent 2-wire In-head Transmitters Actual size IPAQ-H is a universal and intelligent 2-wire In-head transmitter for temperature and other measurement applications. IPAQ-HX is the Intrinsic
More informationSafety Instructions Nivotester FTC625, FTC325
XA01351F-A/00/EN/01.14 71273077 Products Solutions Services Safety Instructions FTC625, FTC325 [Ex ia Ga] IIC/IIB TÜV 13.0903 X Segurança TÜV INMETRO OCP 0004 Document: XA01351F-A Safety instructions for
More informationProduct manual wire programmable RTD transmitter
Product manual 5332 2-wire programmable RTD transmitter PERFORMANCE MADE SMARTER TEMPERATURE I.S. INTERFACES COMMUNICATION INTERFACES MULTIFUNCTIONAL ISOLATION DISPLAY No. 5332V100-UK From serial no: 181903001
More informationMODEL: B6U-B B6-UNIT. [2] LCD DISPLAY 0: Without 1: With
Field-mounted Two-wire Signal Conditioners B6-UNIT 8: TIIS flameproof (CE not available) Confirm selectable combinations of approval and wiring conduit types in the table. 2-WIRE UNIVERSAL TEMPERATURE
More information3 Ex i Isolators. Temperature Transmitter with Output 0/ ma, without Limit Value Contact (Field Circuit Ex i) Type 9182/.0-5.
Ex i Isolators Temperature Transmitter with Output 0/4... 20 ma, without Limit Value Contact Field Circuit Ex i) One unit for nearly all temperature sensors indivdually configurable Intrinsically safe
More informationTwo-channel safety barriers Series 9002
Two-channel safety s Series 00 > Wide product range for all standard applications of the automation > Flexible and space-saving available in single and double channel versions > Time-saving installation
More informationCERTIFICATE OF COMPLIANCE
FM Approvals 1151 Boston Providence Turnpike P.O. Box 9102 Norwood, MA 02062 USA T: 781 762 4300 F: 781-762-9375 www.fmapprovals.com CERTIFICATE OF COMPLIANCE HAZARDOUS (CLASSIFIED) LOCATION ELECTRICAL
More informationIsolators A3/1. Vibration Transducer Supply Unit Series
> For vibration, acceleration and speed sensors in 2- and 3-wire design > Space-saving dual-channel version > Signal frequencies up to 50 khz > Easy setting by means of front-side rotary switch > Galvanic
More informationThie. Automation Products Rev. Chang Date Name Supersedes Dwg. : Part Class:
Page: 1 of 3 - - 2003 Date Name Name 27.03.03 Thiem. Title FM/CSA-Control-Document Appr. / 5 08.07.11 Thie. Std. 4 07.10.09 Lasa. 3 26.06.06 Thie. Drwg.-No. (Part-No.) 2 22.05.06 Thie. 901265 1 27.03.
More informationIntrinsically Safe Compact Controller CTR 210i. Intrinsically Safe Bargraph Indicator. BGI 210i
Intrinsically Safe Compact Controller CTR 210i Intrinsically Safe Bargraph Indicator BGI 210i DMT 02 ATEX E 148 2. Supplement also grahics display Revision 4 IBS BatchControl GmbH Im Sträßchen 2 4 Tel.:
More informationTemperature Transmitters TMT 136, TMT 137
Technical Information TI 066R/09/en Part No. 51000855 Temperature Transmitters TMT 136, TMT 137 For Pt100 Resistance Thermometers 2-wire head transmitter Standard- and instrinsically safe versions Head
More informationAnalog Universal Module HART for Zone 2 Series 9468/33
> 8 channels can be adjusted individually as analog inputs or outputs > Intrinsically safe inputs/outputs Ex ia > For 0/4 20 ma + HART signals > Line fault monitoring per channel > Diagnostics based on
More informationMCR-FL-TS-LP-I-EX SIL IEC Programmable Intrinsically Safe Loop-Powered Temperature Measuring Transducer With HART Protocol
SIL IEC 6 508 Programmable Intrinsically Safe Loop-Powered Temperature Measuring Transducer With HART Protocol Data Sheet 0068_en_03 PHOENIX CONTACT 0-07-09 Description Temperature FL ϑ Universal PC programmable
More informationA9000Px Power/Signal Adapter Installation Drawing for Emerson Transmitters
THIS DRAWING IS PROPERTY OF AND PROPRIETARY TO EMERSON. THIS DRAWING AND THE INFORMATION CONTAINED HEREIN SHALL NOT BE REPRODUCED NOR DISCLOSED TO ANYONE EXCEPT THOSE REQUIRING ITS USE FOR EMERSON AUTHORIZED
More informationD1061S INSTRUCTION MANUAL. RS422/RS485 Fieldbus Isolating Repeater Din-Rail Model D1061S. D RS422 / RS485 Fieldbus Isolating Repeater
DS INSTRUCTI MANUAL RS/RS Fieldbus Isolating Repeater Din-Rail Model DS D - RS / RS Fieldbus Isolating Repeater ISM00- Characteristics General Description: The single channel DIN Rail RS / RS Fieldbus
More informationGeneral Specifications
General Specifications FN310 Field Wireless Multi-Protocol Module GENERAL This General Specification (GS) describes the specifications for Field Wireless Multi-Protocol Module. Combined with Field Wireless
More informationTF 02 / TF 02-Ex. Head mounted temperature transmitters, FOUNDATION Fieldbus (H1), Pt 100 (RTD), thermocouples, 1 or 2 independent channels
TF 0 / TF 0-Ex Head mounted temperature transmitters, FOUNDATION Fieldbus (H), Pt 00 (RTD), thermocouples, or independent channels 0/-8.5 EN Input Resistance thermometer (,, wire circuit) Thermocouples
More informationOperator Manual. Transmitters Series KAT-... Analog transmitters in 2-wire technique
Operator Manual Transmitters Series KAT-... Analog transmitters in 2-wire technique Important Safety Notes! In the case of non-observance of the installation and safety notes or improper use of the transmitter
More information3 Ex i Isolators. Temperature Transmitter without Output 0/ ma, with Limit Value Contact (Field Circuit Ex i) Type 9182/.
Ex i Isolators Temperature Transmitter without Output 0/4... 20 ma, with Limit Value Contact Field Circuit Ex i) One unit for nearly all temperature sensors indivdually configurable Intrinsically safe
More informationGeneral Specifications
General Specifications EJX Series FOUNDATION TM Fieldbus Communication FOUNDATION fieldbus is the digital communication line for the field instruments, whose signal is internationally standardized by Fieldbus
More informationTwo-channel safety barriers Series 9002
Two-channel safety s Series 00 > Wide product range for all standard applications of the automation > Flexible and space-saving available in single and double channel versions > Time-saving installation
More informationSpecifications. Specifications. Transfer Characteristics Accuracy at 20 C (68 F) Temperature Drift
Appendix A Specifications 1797-IE8 and -IE8NF Input Modules Specifications Number of Inputs IS Input Type IS Module Type Resolution Transfer Characteristics Accuracy at 20 C (68 F) Temperature Drift Functional
More informationPROline promag 53 Division 1
XA 040D/06/en/05.01 50099923 FM+SGML 6.0 PROline promag 53 Division 1 Ex Ex documentation for the BA 053D and BA 054D operating instructions according to FACTORY MUTUAL standards documentation for the
More informationD1035S INSTRUCTION MANUAL. Frequency-Pulse Isolating Repeater DIN Rail Model D1035S. D Frequency-Pulse Isolating Repeater
D1035S INSTRUCTION MANUAL FrequencyPulse Isolating Repeater DIN Rail Model D1035S D1035 FrequencyPulse Isolating Repeater ISM007910 Characteristics General Description: The single channel DIN Rail FrequencyPulse
More informationRHE12. Hazardous Area Coriolis Mass Flow Transmitter. Features. Applications. Benefits
RHE12 Hazardous Area Coriolis Mass Flow Transmitter Features Field mounting Compact, pressure safe housing ATEX and CSA approvals for installation in hazardous areas 24 VDC power supply Configurable analog
More informationMODEL: B3HU. Space-saving Two-wire Signal Conditioners B3-UNIT
Space-saving Two-wire Signal Conditioners B3-UNIT 2-WIRE UNIVERSAL TEMPERATURE TRANSMITTER (HART communication, intrinsically safe) Functions & Features Universal input: mv, V, T/C, RTD, resistance, potentiometer
More informationGlass-fibre-reinforced polycarbonate. Dead zone Adaptive or fixed from % 20 ms. 60 ms. 60 ms. Approx. 1.3 kg. Approx. 0.9 kg. Approx. 5.
General data Mounting On linear actuators On quarter turn actuators ARCA-integrated or to VDI/VDE 3847-1 or IEC 534-6 (NAMUR) Range of stroke: 3... 130 mm Integrated to VDI/VDE 3847-2 or VDI/VDE 3845 Angle
More informationInstallation Instructions P/N MMI , Rev. AA July ATEX Installation Instructions for Micro Motion Model RFT9739 Transmitters
Installation Instructions P/N MMI-20011709, Rev. AA July 2009 ATEX Installation Instructions for Micro Motion Model RFT9739 Transmitters Note: For hazardous installations in Europe, refer to standard EN
More informationGeneral Specifications
General Specifications FN510 Field Wireless Multi-Function Module GENERAL This General Specification (GS) describes the specifications for Field Wireless Multi-Function Module. Combined with Field Wireless
More informationGlass-fibre-reinforced polycarbonate. Dead zone Adaptive or fixed from % 1.5 s adjustable. 20 ms. 60 ms. 60 ms. Approx. 1.
General data Mounting On linear actuators On quarter turn actuators ARCA-integrated or to VDI/VDE 3847-1 or IEC 534-6 (NAMUR) Range of stroke: 3... 130 mm Integrated to VDI/VDE 3847-2 or VDI/VDE 3845 Angle
More information09746E00. ATEX / IECEx NEC 505 NEC 506 NEC 500 Class I Class I Class II Class III Zone Zone Division Ex i
Temperature Transmitter Series 9182 www.stahl.de > One unit for nearly all temperature sensors indivdually configurable > Intrinsically safe input [Ex ia] IIC > Signal duplication possible > Galvanic isolation
More informationIsolators A3/1. Transmitter Supply Unit Field Circuit Ex i Series
Transmitter Supply Unit Series 9160 > Intrinsically safe input [Ex ia] IIC > Galvanic isolation between input, output and power supply > Open-circuit and short-circuit monitoring and messaging for input
More informationMiniature resistance thermometer Explosion-protected version Model TR34, thread-mounted
Electrical temperature measurement Miniature resistance thermometer Explosion-protected version Model TR34, thread-mounted WIKA data sheet TE 60.34 Applications Machine building, plant and vessel construction
More informationUniversal Intelligent 2-wire DIN Rail Transmitters
Universal Intelligent 2-wire DIN Rail Transmitters IPAQ-L is a universal and intelligent 2-wire DIN rail transmitter for temperature and other measurement applications. IPAQ-LX is the Intrinsic Safe version
More informationIsolators A3/1. Transmitter Supply Unit Field Circuit Ex i Series 9160
Series 9160 > Intrinsically safe input [Ex ia] IIC > Galvanic isolation between input, output and power supply > For use up to SIL 2, special version up to SIL 3 (IEC 61508) > High accuracy www.stahl.de
More informationIsolators A3/1. Transmitter Supply Unit Field Circuit Ex i Series 9160 Rev. F
Series 9160 Rev. F > Intrinsically safe input [Ex ia] IIC > Galvanic isolation between input, output and power supply > For use up to SIL 2, special version up to SIL 3 (IEC 61508) > High accuracy www.stahl.de
More informationCoriolis Massflowmeter
Coriolis Massflowmeter for liquids and gas measuring monitoring analysing Measuring range: 0-60 kg/h 0-60 000 kg/h water Accuracy: ±0,15 of reading ±zero-point stability pmax: PN40 tmax: -40 +180 C Connection:
More informationDigital Input Module 24 V Ex n Inputs, 16 Channels Series 9471
> 16 channels for active 0 / 24 V signals > Zone 2 / Division 2 version for connection of circuits acc. to Ex nl, Ex na, Nonincendive and Non-Ex > Galvanic separation between inputs and system > Two channels
More informationCoriolis Massflowmeter
Coriolis Massflowmeter for liquids and gas measuring monitoring analysing Measuring range: 0-0.8 kg / h 0-65 000 kg / h water Accuracy: ± 0.1 of reading ±zero-point stability (liquids) pmax: PN40 tmax:
More informationHead mounted Temperature Transmitter TF02/TF02-Ex
Data Sheet Head mounted Temperature Transmitter TF0/TF0-Ex FOUNDATION Fieldbus H, Pt 00 (RTD), thermocouples, or independent channels Input Resistance thermometer (-, -, -wire circuit) Thermocouples Resistance
More informationSingle-Channel Safety Barriers Series 9001
Single-Channel Safety Barriers Series 9001 > Broad product range for all standard applications in the world of automation > Flexible and space saving single and dual channel versions available > Time saving
More informationHART Compatible Intelligent 2-wire In-head Transmitters
HART Compatible Intelligent 2-wire In-head Transmitters Actual size MESO-H is a Smart and universal 2-wire Inhead transmitter for temperature and other measurement applications. MESO-HX is the Intrinsic
More informationDigital Input Module NAMUR Ex i / I.S. Inputs, 16 Channels for Zone 1 / Div. 1 Series 9470/22
www.stahl.de > 16 channels for contacts and NAMUR proximity switches (EN 60947-5-6) > Intrinsically safe inputs Ex ia IIC > Galvanic separation between inputs and system > Open-circuit and short-circuit
More information... for hydraulic and general applications in hazardous areas
ATEX Approved Inductive... for hydraulic and general applications in hazardous areas Inductive Namur sensors Explosion protected Ignition class "intrinsically safe" when used together a suitable certified
More informationTrendmaster * Dynamic Scanning Module
149744 Trendmaster * Dynamic Scanning Module Product Datasheet Bently Nevada * Asset Condition Monitoring Description The 149744 Trendmaster* Dynamic Scanning Module (DSM) is a compact rack-based data
More informationResistance thermometer Model TR30, compact design
Electrical temperature measurement Resistance thermometer Model TR30, compact design WIKA data sheet TE 60.30 Applications Machine building, plant and vessel construction Propulsion technology, hydraulics
More informationFieldIT Temperature Transmitter, rail mounted TH102/TH102-Ex
Data Sheet 10/-8.54 EN Level 0: Information FieldIT Temperature Transmitter, rail mounted TH102/TH102-Ex HART programmable, Pt 100 (RTD), thermocouples, electrical isolation Resistance thermometer (2,
More informationSET Installation and Operating Instructions. Level switch for two sensors
Labkotec Oy Myllyhaantie 6 FI-33960 PIRKKALA FINLAND Tel: + 358 29 006 260 Fax: + 358 29 006 1260 20.2.2013 Internet: www.labkotec.fi 1/14 SET-2000 Level switch for two sensors Copyright 2013 Labkotec
More informationSafety Instructions Proline Promass 100
XA00160D/06/EN/03.14 71261250 Products Solutions Services Safety Instructions Proline Promass 100 Modbus RS485 Cl.I Div.1, Zone 1 for IS (Ex i Intrinsically safe version) C US Document: XA00160D Safety
More informationProcess transmitter Model UPT-20, with pressure port Model UPT-21, with flush diaphragm
Electronic pressure measurement Process transmitter Model UPT-20, with pressure port Model UPT-21, with flush diaphragm WIKA data sheet PE 86.05 Applications Process technology Machine building and plant
More informationDigital temperature transmitter for resistance sensors Model T15.H, head mounting version Model T15.R, rail mounting version
Temperature Digital temperature transmitter for resistance sensors Model T15.H, head mounting version Model T15.R, rail mounting version WIKA data sheet TE 15.01 for further approvals see page 10 Applications
More informationIsolators A3/1.
> Compact limit value switch with 2 configurable limiting values and 4... 20 ma output > Suitable for 2-, 3-wire transmitters, 2-wire HART transmitters and ma sources > Intrinsically safe input [Ex ia]
More informationRHE11. Hazardous Area Multifunction Coriolis Flow Transmitter. Features. Applications. Benefits
RHE11 Hazardous Area Multifunction Coriolis Flow Transmitter Features Wall or Pipe Bracket Mount Built in safety barriers allow operation with RHM sensor in hazardous area Selectable Metric and English
More informationTMU. Coriolis Mass Flow Meter. for liquids and gas
Coriolis Mass Flow Meter for liquids and gas measuring monitoring analysing TMU S5 OO Measuring range: 0-60 kg/h 0-2200 t/h water OO Accuracy: ± 0.1% of reading ±zero point stability (from liquids up to
More informationProcessMaster, HygienicMaster FEX300, FEX500
SM/FEX300/FEX500/FM/CSA-EN ProcessMaster, HygienicMaster FEX300, FEX500 FM / cfm Div. 1, Div. 2 Safety information for electrical equipment for potentially explosive areas in accordance with Factory Mutual
More informationProduct Manual wire HART transmitter
Product Manual 6335 2wire HART transmitter PERFORMANCE MADE SMARTER TEMPERATURE I.S. INTERFACES COMMUNICATION INTERFACES MULTIFUNCTIONAL ISOLATION DISPLAY No. 6335V111UK From serial no.: 150951186 6 Product
More informationD1061S INSTRUCTION MANUAL. RS422/RS485 Fieldbus Isolating Repeater Din-Rail Model D1061S. D RS422 / RS485 Fieldbus Isolating Repeater
DS INSTRUCTI MANUAL RS/RS Fieldbus Isolating Repeater Din-Rail Model DS D - RS / RS Fieldbus Isolating Repeater ISM00- Characteristics General Description: The single channel DIN Rail RS / RS Fieldbus
More informationSEM210 SERIES. Programmable In-head Universal Temperature Transmitter INDEX SECTION CONTENTS PAGE NO.
INDEX SECTION CONTENTS PAGE NO. SEM210 SERIES 1.0 DESCRIPTION 1 2.0 SPECIFICATION 15 3.0 INSTALLATION 67 Programmable Inhead Universal Temperature Transmitter Status Instruments Ltd, Green Lane Business
More informationLoop Powered Indicator
Instruction Manual Model PD686 Loop Powered Indicator M397 January 3, 2019 JOWA USA, Inc. 59 Porter Road Littleton, MA 01460-1431 USA Tel: 978-486-9800 Fax: 978-486-0170 PD686 I.S. and N.I. NEMA 4X, IP67
More informationUser manual. Load cell with one built in amplifier KOSD-FA KIMD-FA KEND-FA Load cell with two built in amplifiers KOSD-FAD KIMD-FAD KEND-FAD
User manual Load cell with one built in amplifier KOSD-FA KIMD-FA KEND-FA Load cell with two built in amplifiers KOSD-FAD KIMD-FAD KEND-FAD Contents Precautions Intended use General 1 Specification 3
More information