N-Methyl-1-(6-methylpyridin-2-yl)propan-2-amine

Size: px
Start display at page:

Download "N-Methyl-1-(6-methylpyridin-2-yl)propan-2-amine"

Transcription

1 H C N CH HN CH. GENERAL INFORMATION IUPAC Name: CAS#: Not Available Synonyms: Source: Appearance: Not Available DEA Reference Material Collection Pale yellow powder UV max (nm): Not Determined. CHEMICAL AND PHYSICAL DATA. CHEMICAL DATA Form Chemical Formula Molecular Weight Melting Point ( o C) Base C 0 H 6 N 64 Not Determined HCl C 0 H 6 N HCl 00 Not Determined Page of 5

2 . QUALITATIVE DATA. NUCLEAR MAGNETIC RESONANCE Sample Preparation: Dilute analyte to ~4 mg/ml in CD OD containing TMS for 0 ppm reference and dimethylfumarate as quantitative internal standard. Instrument: Parameters: 0 0 MHz NMR spectrometer Spectral width: at least containing - ppm through ppm Pulse angle: 90 o Delay between pulses: 45 seconds HNMR: HCl Lot RM-0-0, 0MHz 9 8 Dimethylfumarate Dimethylfumarate Methanol Methanol TMS Page of 5

3 . GAS CHROMATOGRAPHY/MASS SPECTROMETRY Sample Preparation: Dilute analyte ~ mg/ml base extracted into CHCl. Instrument: Column: Carrier Gas: Temperatures: Agilent gas chromatograph operated in split mode with MS detector HP-5 MS (or equivalent); 0m x 0.5 mm x 0.5 m Helium at ml/min Injector: 80 o C MSD transfer line: 80 o C MS Source: 0 o C MS Quad: 50 o C Oven program: ) 00 o C initial temperature for.0 min ) Ramp to 80 o C at o C/min ) Hold final temperature for 9.0 min Injection Parameters: Split Ratio = 5:, L injected MS Parameters: Mass scan range: amu Threshold: 00 Tune file: stune.u Acquisition mode: scan Retention Time: 4.66 min EI Mass Spectrum: HCl, Lot RM-0-0 Intensity [x 0 6 ] EI m/z Intensity [x 0 4 ] EI m/z Page of 5

4 Page 4 of 5. INFRARED SPECTROSCOPY (FTIR) Insrument: FTIR with diamond ATR attachment ( bounce) Scan Parameters: Number of scans: Number of background scans: Resolution: 4 cm - Sample gain: 8 Aperture: 50 FTIR ATR (Diamond, Bounce): HCl, Lot RM-0-0) Wavenumber (cm-) %Transmittance Wavenumber (cm-) %Transmittance

5 4. ADDITIONAL RESOURCES No Literature available as of 0/06 Page 5 of 5

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Application Note Meeting the Korean Method ES 04607.1 for Naphthalene ( 나프탈렌 ) with Static and Dynamic Headspace GC/MS Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Introduction The Korean

More information

KNOW TRAINING MORE WITH. PerkinElmer Training Catalogue 2013 Relevant, interactive, rich in content.

KNOW TRAINING MORE WITH. PerkinElmer Training Catalogue 2013 Relevant, interactive, rich in content. PerkinElmer Training Catalogue 2013 Relevant, interactive, rich in content. KNOW MORE WITH TRAINING PERKINELMER MISSION For people. For the environment. For the shared goal of a healthier future. Improving

More information

Synthesis and spectroscopic properties of β meso directly linked porphyrin corrole hybrid compounds

Synthesis and spectroscopic properties of β meso directly linked porphyrin corrole hybrid compounds Supporting Information for Synthesis and spectroscopic properties of β meso directly linked porphyrin corrole hybrid compounds Baris Temelli * and Hilal Kalkan Address: Hacettepe University, Department

More information

Determination of the N-Nitrosamine Content in Rubber Articles Using the Agilent 7000A Triple Quadrupole GC/MS System

Determination of the N-Nitrosamine Content in Rubber Articles Using the Agilent 7000A Triple Quadrupole GC/MS System Determination of the N-Nitrosamine Content in Rubber Articles Using the Agilent 7A Triple Quadrupole GC/MS System Application Note Consumer Products Authors Yun Zou and Chongtian Yu Agilent Technologies

More information

Introducing the Agilent 7000A Triple Quadrupole GC/MS

Introducing the Agilent 7000A Triple Quadrupole GC/MS Introducing the Agilent 7000A Triple Quadrupole GC/MS Paul Zavitsanos, Joe Weitzel, Monty Benefiel and Terry Sheehan May 2008 Page 1 What s a Triple Quad? Page 2 GC/MS Triple Quad (QQQ) Collision Gas (N2

More information

Damage-free failure/defect analysis in electronics and semiconductor industries using micro-atr FTIR imaging

Damage-free failure/defect analysis in electronics and semiconductor industries using micro-atr FTIR imaging Damage-free failure/defect analysis in electronics and semiconductor industries using micro-atr FTIR imaging Application note Electronics and Semiconductor Authors Dr. Mustafa Kansiz and Dr. Kevin Grant

More information

Collision cell: Fragmentation efficiency > 90 % Duty cycle: Variable between 1 and 50 % (user selectable trapping time/scan time ratio) Transmission:

Collision cell: Fragmentation efficiency > 90 % Duty cycle: Variable between 1 and 50 % (user selectable trapping time/scan time ratio) Transmission: Finnigan MAT 95 XP-API-GC-Trap tandem Mass spectrometer system Tandem Mass Spectrometer based on high resolution magnetic sector and Ion-Trap MS/MS system with API ( ESI, APCI ) and EI/CI ionization Mass

More information

Supporting Information

Supporting Information Highly diastereoselective cyclopropanation of -methylstyrene catalyzed by a C 2 -symmetrical chiral iron porphyrin complex Daniela Intrieri, Stéphane Le Gac, Alessandro Caselli, Eric Rose, Bernard Boitrel,

More information

TechnicalNOTE EVALUATION OF LCT PREMIER NEGATIVE ION MODE ESI EXACT MASS MEASUREMENT PERFORMANCE. Introduction. Objective

TechnicalNOTE EVALUATION OF LCT PREMIER NEGATIVE ION MODE ESI EXACT MASS MEASUREMENT PERFORMANCE. Introduction. Objective Objective The aim of the study presented is to illustrate the enhanced performance of the Waters Micromass LCT Premier oa-tof Spectrometer. The data presented shows mass accuracy achieved over a twelve-hour

More information

Welcome! Denver, CO Houston, TX February 2009

Welcome! Denver, CO Houston, TX February 2009 Welcome! Denver, CO Houston, TX February 2009 Schedule & Introductions Announcements - Safety - Evaluation forms - Upcoming events > CBMSS, Feb 10 th, Nat l Jewish Center, www.cbmss.org (also Apr 22 nd

More information

Analytical Cumulative Exam November 21, 2008

Analytical Cumulative Exam November 21, 2008 Analytical Cumulative Eam November 21, 2008 Questions on this cume are drawn from the attached article, entitled Measurement of Sulfur Isotope Compositions by Tunable Laser Spectroscopy of SO 2 Where specific

More information

MT Gemcitabine, [5-3 H(N)]- Lot A NK

MT Gemcitabine, [5-3 H(N)]- Lot A NK MT-1572 Gemcitabine, [5-3 H(N)]- Lot 194-14-234-A-2514-NK A) All chromatograms were run using the HPLC method described on the Product Data Sheet. Concentrations and volumes: Standard solution concentration

More information

LC/MS/MS. Page Header. triple quadrupole mass spectrometer.

LC/MS/MS. Page Header. triple quadrupole mass spectrometer. LC/MS/MS VARIAN, INC. 320-MS Page Header triple quadrupole mass spectrometer www.varianinc.com VARIAN, INC. 320-MS Unsurpassed commitment to innovation Varian, Inc. is an innovator and leader in mass spectrometry

More information

Redox-Innocent Metal-Assisted Cleavage of S-S. Bond in a Disulfide-Containing Ligand.

Redox-Innocent Metal-Assisted Cleavage of S-S. Bond in a Disulfide-Containing Ligand. Supplementary Information for Redox-Innocent Metal-Assisted Cleavage of S-S Bond in a Disulfide-Containing Ligand. Charlène Esmieu, Maylis Orio, Laurent Le Pape, Colette Lebrun, Jacques Pécaut, Stéphane

More information

Integration and Optimization of Hardware and Software for a Differential Flow Modulated GCxGC

Integration and Optimization of Hardware and Software for a Differential Flow Modulated GCxGC Paper 170-2 Integration and Optimization of Hardware and Software for a Differential Flow Modulated GCxGC Roger Firor, James McCurry, Chunxiao Wang and Michael Feeney Agilent Technologies Wilmington, DE

More information

Synthesis of Novel Peptide Linkers: Simultaneous Cyclization and Labeling

Synthesis of Novel Peptide Linkers: Simultaneous Cyclization and Labeling UPPRTING INFRMATIN ynthesis of Novel Peptide Linkers: imultaneous Cyclization and Labeling Gajanan K. Dewkar, Pedro B. Carneiro, Matthew C. T. Hartman* Department of Chemistry and Massey Cancer Center,

More information

FT-IR.

FT-IR. FT-IR varian, inc. 610/620-IR ft-ir MICROSCOPY AND IMAGING SoLUTIONS www.varianinc.com VARIAN, INC. Setting the Standard Again When Only the Best Will Do The world leader in molecular spectroscopy innovation

More information

MT-540. D-Ribavirin, [ 3 H]- Lot A DG

MT-540. D-Ribavirin, [ 3 H]- Lot A DG MT-54 D-Ribavirin, [ 3 H]- Lot 165-145-225-A-2835-DG A) All chromatograms were run using the HPLC method described on the Product Data Sheet. Concentrations and volumes: Standard solution concentration

More information

FTIR microscopy and imaging for failure analysis in electronics manufacturing

FTIR microscopy and imaging for failure analysis in electronics manufacturing FTIR microscopy and imaging for failure analysis in electronics manufacturing Application Note Author Steven M. Barnett, Ellen V. Miseo, and Wayne Jalenak Agilent Technologies, Inc. Introduction The electronics

More information

SUPPLEMENTAL MATERIAL

SUPPLEMENTAL MATERIAL SUPPLEMENTAL MATERIAL 1 - Folios and areas of analysis Figure S1.1. Folio 4, areas of analysis for microxrf ( ), FORS ( ), micro-samples for Raman and FTIR ( ) and Raman in-situ ( ). Figure S1.2. Folio

More information

Profiling of Volatile Organic Compounds in Milk and Orange Juice Using Headspace Analysis

Profiling of Volatile Organic Compounds in Milk and Orange Juice Using Headspace Analysis Profiling of Volatile Organic Compounds in Milk and Orange Juice Using Headspace Analysis HT3 Application Note Introduction Volatile organic compounds, VOCs, present in liquid food products play a very

More information

High power and single frequency quantum. cascade lasers for gas sensing. Stéphane Blaser

High power and single frequency quantum. cascade lasers for gas sensing. Stéphane Blaser High power and single frequency quantum cascade lasers for gas sensing Stéphane Blaser Alpes Lasers: Yargo Bonetti Lubos Hvozdara Antoine Muller University of Neuchâtel: Marcella Giovannini Nicolas Hoyler

More information

Fingerprinting the oxidation state of U(IV) by

Fingerprinting the oxidation state of U(IV) by Fingerprinting the oxidation state of U(IV) by emission spectroscopy Emtithal Hashem, 1 Giulia Lorusso 2 Marco Evangelisti, 2 Thomas McCabe, 1 Carola Schulzke, 3 James A. Platts 4 and Robert J. Baker 1*

More information

Infrared Microscope AIM-9000 C103-E103B

Infrared Microscope AIM-9000 C103-E103B Infrared Microscope C103-E103B Wide-Field Camera Automatic Recognition Automatic Qualitative High Sensitivity Finally, a wide view on micro sample analysis Surface Micro SHIMADZU AIMs to provide analysis

More information

Infrared Microscope. Dedicated AIMsolution Software. Hisato Fukuda. 1. Introduction. 2. Automatic Contaminant Recognition Function

Infrared Microscope. Dedicated AIMsolution Software. Hisato Fukuda. 1. Introduction. 2. Automatic Contaminant Recognition Function C103-E120 Vol. 28 Infrared Microscope Dedicated AIMsolution Software ------- 02 Infrared Microscope Using Imaging Analysis ------- 05 EDXIR-Analysis EDX-FTIR Contaminant Finder/Material Inspector -------

More information

Beeswax adulteration issue: aspects of contamination and outcome

Beeswax adulteration issue: aspects of contamination and outcome Beeswax adulteration issue: aspects of contamination and outcome Lidija Svečnjak University of Zagreb Faculty of Agriculture Zagreb, Croatia Beeswax adulteration issue long-present and growing problem

More information

Chemistry 524--"Hour Exam"--Keiderling Mar. 19, pm SES

Chemistry 524--Hour Exam--Keiderling Mar. 19, pm SES Chemistry 524--"Hour Exam"--Keiderling Mar. 19, 2013 -- 2-4 pm -- 170 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils permitted. No open books allowed.

More information

New fraction collection features with the Agilent 220 micro plate sampler and micro plate sampling software rev. A.03. Technical Note.

New fraction collection features with the Agilent 220 micro plate sampler and micro plate sampling software rev. A.03. Technical Note. New fraction collection features with the Agilent 220 micro plate sampler and micro plate sampling software rev. A.03 Technical Note Introduction The Agilent 220 micro plate sampler (MPS) is an essential

More information

Examining Comprehensive Flow-Modulated Two-Dimensional Gas Chromatography

Examining Comprehensive Flow-Modulated Two-Dimensional Gas Chromatography Examining Comprehensive Flow-Modulated Two-Dimensional Gas Chromatography Roger L. Firor Senior Applications Chemist, Agilent Technologies roger_fi ror@agilent.com 1 Agilent Measurement Journal Gas chromatography

More information

Jian-Wei Liu, Jing Zheng, Jin-Long Wang, Jie Xu, Hui-Hui Li, Shu-Hong Yu*

Jian-Wei Liu, Jing Zheng, Jin-Long Wang, Jie Xu, Hui-Hui Li, Shu-Hong Yu* Supporting Information Ultrathin 18 O 49 Nanowire Assemblies for Electrochromic Devices Jian-ei Liu, Jing Zheng, Jin-Long ang, Jie Xu, Hui-Hui Li, Shu-Hong Yu* Experimental Section Synthesis and Assembly

More information

Micro Analytical Instruments - A System Approach. Jörg Müller Institut für Mikrosystemtechnik

Micro Analytical Instruments - A System Approach. Jörg Müller Institut für Mikrosystemtechnik Micro Analytical Instruments - A System Approach Jörg Müller Institut für Mikrosystemtechnik Technische h Universität i Hamburg-Harburg H b Outline Motivation Political Political Issues Design Principles

More information

Advanced Features of InfraTec Pyroelectric Detectors

Advanced Features of InfraTec Pyroelectric Detectors 1 Basics and Application of Variable Color Products The key element of InfraTec s variable color products is a silicon micro machined tunable narrow bandpass filter, which is fully integrated inside the

More information

Supporting Information

Supporting Information Supporting Information Eaton et al. 10.1073/pnas.1600789113 Additional Characterization and Simulation of CsPbX 3 Nanowires and Plates Atomic Force Microscopy Measurements. Atomic force microscopy (AFM)

More information

835 Vacuum Quality Monitor (VQM TM ) 2012 Brooks Automation, Inc. Proprietary Information

835 Vacuum Quality Monitor (VQM TM ) 2012 Brooks Automation, Inc. Proprietary Information 835 Vacuum Quality Monitor (VQM TM ) 2012 Brooks Automation, Inc. Proprietary Information The Revolutionary New Vacuum Quality Monitor World s Fastest, Accurate at low mass Single Gas Calibration Small

More information

The Extrel MAX Systems are UHV-compatible flange mounted Quadrupole Mass

The Extrel MAX Systems are UHV-compatible flange mounted Quadrupole Mass The Extrel MAX Systems are UHV-compatible flange mounted Quadrupole Mass Spectrometers. The Ionizer, Quadrupole Mass Filter, and Detector on a Mounting Flange are designed for inclusion in your experimental

More information

Material analysis by infrared mapping: A case study using a multilayer

Material analysis by infrared mapping: A case study using a multilayer Material analysis by infrared mapping: A case study using a multilayer paint sample Application Note Author Dr. Jonah Kirkwood, Dr. John Wilson and Dr. Mustafa Kansiz Agilent Technologies, Inc. Introduction

More information

Technical Specifications for the procurement of Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Technical Specifications for the procurement of Inductively Coupled Plasma Mass Spectrometer (ICP-MS) Technical Specifications for the procurement of Inductively Coupled Plasma Mass Spectrometer (ICP-MS) Preamble An Inductively Coupled Plasma Mass Spectrometer (Quadrupole ICP-MS) is proposed to be procured

More information

Agilent 6400 Series Triple Quadrupole LC/MS System

Agilent 6400 Series Triple Quadrupole LC/MS System Agilent 6400 Series Triple Quadrupole LC/MS System Concepts Guide The Big Picture Agilent Technologies Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by

More information

Markus Leuenberger1, Tesfaye Berhanu1, Peter Nyfeler1, David Kim-Hak2, John Hoffnagle2 and Minghua Sun2. Bern, Switzerland

Markus Leuenberger1, Tesfaye Berhanu1, Peter Nyfeler1, David Kim-Hak2, John Hoffnagle2 and Minghua Sun2. Bern, Switzerland GGMT-2017, Dübendorf, Switzerland August 30, 2017 Measurements of atmospheric oxygen using a newly built CRDS analyzer and comparison with a paramagnetic and an IRMS. Markus Leuenberger1, Tesfaye Berhanu1,

More information

Fast Methods for Small Molecules

Fast Methods for Small Molecules Fast Methods for Small Molecules Technical Overview Throughput is a key concern in many NMR laboratories, and using faster methods is one way to increase it. Traditionally, multidimensional NMR requires

More information

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application CONFIGURING Your Spectroscopy System For PEAK PERFORMANCE A guide to selecting the best Spectrometers, s, and s for your application Spectral Measurement System Spectral Measurement System Spectrograph

More information

Development and Applications of a Sample Compartment FTIR Microscope

Development and Applications of a Sample Compartment FTIR Microscope Application Note Development and Applications of a Sample Since the early to mid-1940 s, scientists using infrared spectroscopy have been trying to obtain spectral data from ever smaller samples. Starting

More information

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 4 of 4 (three out of four required, 20 points)

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 4 of 4 (three out of four required, 20 points) NAME Chem 203 Organic Spectroscopy Midterm Examination, Part II (60 points total) Problem 4 of 4 (three out of four required, 20 points) Saturday, November 15, 2014, 9 am -??? SUBMIT THREE OF THE FOUR

More information

Supporting Information

Supporting Information Supporting Information rganocatalytic Mitsunobu Reactions Tracy Yuen Sze But and Patrick H. Toy * Department of Chemistry, The University of Hong Kong, Pokfulam Road, Hong Kong, People s Republic of China

More information

New Developments in TDLAS NH3 Monitoring

New Developments in TDLAS NH3 Monitoring New Developments in TDLAS NH3 Monitoring Presented by John Pisano CEMTEK Environmental UCR (University of California at Riverside) Unisearch Associates Inc Outline What is a tunable diode laser (TDL) The

More information

Ultrasensitive LC MS/MS: Agilent 6470 and 6495 LC-QQQ

Ultrasensitive LC MS/MS: Agilent 6470 and 6495 LC-QQQ Ultrasensitive LC MS/MS: Agilent 6470 and 6495 LC-QQQ ROCK SOLID Performance for Trace-Level Quantitation Agilent Technologies 1 Brief History of Agilent 6400 QQQ LC/MS 6410 2006 6460 Agilent Jet Stream

More information

Terahertz Spectroscopic/ Imaging Analysis Systems

Terahertz Spectroscopic/ Imaging Analysis Systems Terahertz Spectroscopic/ Series Non-Destructive Analysis of Pharmaceuticals, Chemicals, Communication Materials, etc. Compact, High-Speed Terahertz Spectroscopic/ High-speed measurement functionality Compact,

More information

Quick Start MMRC Nicolet 6700 Revised 10/24/2018

Quick Start MMRC Nicolet 6700 Revised 10/24/2018 Do Not Remove Quick Start MMRC Nicolet 6700 Revised 10/24/2018 1. Fill out the logbook before you begin. 2. Before you start the Nicolet should be on with purge gas flowing. 3. Check that the purge gas

More information

High resolution cavity-enhanced absorption spectroscopy with a mode comb.

High resolution cavity-enhanced absorption spectroscopy with a mode comb. CRDS User meeting Cork University, sept-2006 High resolution cavity-enhanced absorption spectroscopy with a mode comb. T. Gherman, S. Kassi, J. C. Vial, N. Sadeghi, D. Romanini Laboratoire de Spectrométrie

More information

Doppler-Free Spetroscopy of Rubidium

Doppler-Free Spetroscopy of Rubidium Doppler-Free Spetroscopy of Rubidium Pranjal Vachaspati, Sabrina Pasterski MIT Department of Physics (Dated: April 17, 2013) We present a technique for spectroscopy of rubidium that eliminates doppler

More information

Automated Orthogonal Control System for Electrospray Ionization Mass Spectrometry

Automated Orthogonal Control System for Electrospray Ionization Mass Spectrometry utomated Orthogonal Control System for Electrospray Ionization Mass Spectrometry Gary. Valaskovic 1, James P. Murphy III 1, Mike S. Lee 2 1 New Objective Inc, Woburn, M 2 Milestone Development Services,

More information

Arrayed Acquisition in VNMR and on the Gemini 1

Arrayed Acquisition in VNMR and on the Gemini 1 Arrayed Acquisition in VNMR and on the Gemini 1 I. Arrays in VNMR Vnmr allows you to quickly define experiments in which a series of spectra can be obtained as a function of any NMR parameter. For example,

More information

*Corresponding author.

*Corresponding author. Supporting Information for: Ligand-Free, Quantum-Confined Cs 2 SnI 6 Perovskite Nanocrystals Dmitriy S. Dolzhnikov, Chen Wang, Yadong Xu, Mercouri G. Kanatzidis, and Emily A. Weiss * Department of Chemistry,

More information

Renishaw InVia Raman microscope

Renishaw InVia Raman microscope Laser Spectroscopy Labs Renishaw InVia Raman microscope Operation instructions 1. Turn On the power switch, system power switch is located towards the back of the system on the right hand side. Wait ~10

More information

830 Vacuum Quality Monitor

830 Vacuum Quality Monitor Granville-Phillips Vacuum Instrumentation 830 Vacuum Quality Monitor Featuring ART MS Technology February 2011 1 What is ART MS? ART MS Anharmonic Resonant Trap Mass Spectrometry or Auto Resonant Ion Trap

More information

NanoSpective, Inc Progress Drive Suite 137 Orlando, Florida

NanoSpective, Inc Progress Drive Suite 137 Orlando, Florida TEM Techniques Summary The TEM is an analytical instrument in which a thin membrane (typically < 100nm) is placed in the path of an energetic and highly coherent beam of electrons. Typical operating voltages

More information

Supplementary Figure 1. Scanning Electron Microscopy images of the pristine electrodes. (a) negative electrode and (b) positive electrode.

Supplementary Figure 1. Scanning Electron Microscopy images of the pristine electrodes. (a) negative electrode and (b) positive electrode. a b Supplementary Figure 1. Scanning Electron Microscopy images of the pristine electrodes. (a) negative electrode and (b) positive electrode. Images were performed using a FEI/Philips XL4 microscope with

More information

Supporting Information. 8. Real-time qpcr using a Ds-containing primer and fluorophor-dpxtps (Figures S1-S3).

Supporting Information. 8. Real-time qpcr using a Ds-containing primer and fluorophor-dpxtps (Figures S1-S3). Electronic Supplementary Material (ESI) for rganic & Biomolecular Chemistry Supporting Information 1. Chemical syntheses of Cy3- and Cy5-dPxTPs. 2. 1 MR spectrum of Cy3-dPxTP. 3. 31 P MR spectrum of Cy3-dPxTP.

More information

AccuTOF LC Training Course

AccuTOF LC Training Course 1 AccuTOF LC Training Course JEOL USA, Inc. 11 Dearborn Road Peabody, MA 01960 Area map Restaurants: Wendy s Bennigan s Bertucci s 2 AccuTOF LC Training Course Monday Basic principle and history of TOF

More information

Autoresonant Trap Mass Spectrometry

Autoresonant Trap Mass Spectrometry 8 th Harsh Environment Mass Spectrometry Wokshop Autoresonant Trap Mass Spectrometry (ART MS) Sept 21, 2011 Gerardo A. Brucker and G. Jeffery Rathbone Granville-Phillips Vacuum Instrumentation Brooks Automation,

More information

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 1 of 4 (three out of four required, 20 points)

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 1 of 4 (three out of four required, 20 points) NAME Chem 203 Organic Spectroscopy Midterm Examination, Part II (60 points total) Problem 1 of 4 (three out of four required, 20 points) Saturday, November 15, 2014, 9 am -??? SUBMIT THREE OF THE FOUR

More information

Identify extraneous chemicals that contributed towards the failure of actuating mechanism in inner vial

Identify extraneous chemicals that contributed towards the failure of actuating mechanism in inner vial Identify extraneous chemicals that contributed towards the failure of actuating mechanism in inner vial By Vishu Shah Consultek 460-D West Lambert Road Brea, CA 92821 1 Identify extraneous chemicals that

More information

Chem466 Lecture Notes. Spring, 2004

Chem466 Lecture Notes. Spring, 2004 Chem466 Lecture Notes Spring, 004 Overview of the course: Many of you will use instruments for chemical analyses in lab. settings. Some of you will go into careers (medicine, pharmacology, forensic science,

More information

Multiplying Interferometers

Multiplying Interferometers Multiplying Interferometers L1 * L2 T + iv R1 * R2 T - iv L1 * R2 Q + iu R1 * L2 Q - iu Since each antenna can output both L and R polarization, all 4 Stokes parameters are simultaneously measured without

More information

Practical Applications Of Infrared, Raman And Nuclear... Download Practical Applications Of Infrared Thermal...

Practical Applications Of Infrared, Raman And Nuclear... Download Practical Applications Of Infrared Thermal... Practical Applications Of Infrared Thermal Sensing And Imaging Equipment Third Edition Spie Tutorial Text Vol We have made it easy for you to find a PDF Ebooks without any digging. And by having access

More information

INDIAN INSTITUTE OF SCIENCE EDUCATION AND RESEARCH PUNE

INDIAN INSTITUTE OF SCIENCE EDUCATION AND RESEARCH PUNE INDIAN INSTITUTE OF SCIENCE EDUCATION AND RESEARCH PUNE CLARIFICATION ON TENDER NUMBER - IISER-PUR-0011-0012-15 ITEM DESCRIPTION- PROCUREMENT OF UV-VIS SPECTROPHOTOMETER & ATR-FTIR SPECTROSCOPY Refer our

More information

CRL MASS SPECTROMETRY FACILITY INSTRUMENT USER MANUAL MALDI MICRO. Operating Instructions. Basement Spec Lab:

CRL MASS SPECTROMETRY FACILITY INSTRUMENT USER MANUAL MALDI MICRO. Operating Instructions. Basement Spec Lab: CRL MASS SPECTROMETRY FACILITY INSTRUMENT USER MANUAL MALDI MICRO Operating Instructions Basement Spec Lab: 00.097 crs Page of 2 This is a guide to using the MALDI Micro for those who have received training.

More information

Measurement Method of High Absorbance (Low Transmittance) Samples by UH4150 INTRODUCTION

Measurement Method of High Absorbance (Low Transmittance) Samples by UH4150 INTRODUCTION INTRODUCTION With UH4150, a detector can be selected depending on the analysis purpose. When analyzing a solid sample which doesn t contain any diffuse components, by selecting the direct light detector,

More information

SUPPORTING INFORMATION

SUPPORTING INFORMATION Eur. J. Org. Chem. 2008 WILEY-VCH Verlag GmbH & Co. KGaA, 69451 Weinheim, 2008 ISSN 1434 193X SUPPORTING INFORMATION Title: Structural Elucidation with NMR Spectroscopy: Practical Strategies for Organic

More information

Microfocus Beam Condenser User Manual

Microfocus Beam Condenser User Manual Microfocus Beam Condenser User Manual 2I-02560 Issue 7 Microfocus Beam Condenser Microfocus Beam Condenser User Manual 2I-02560 Issue 7 3 User Manual Microfocus Beam Condenser P/N s GS02560 and GS02561

More information

Supporting Information

Supporting Information Supporting Information Development of Photostable Near-Infrared Cyanine Dyes Animesh Samanta, Marc Vendrell, Rajkumar Das and Young-Tae Chang. List of contents: 1. Synthetic procedures and characterization

More information

Supplementary Information

Supplementary Information Electronic Supplementary Material (ESI) for Analyst. This journal is The Royal Society of Chemistry 2018 Supplementary Information TANDEM TRAPPED ION MOBILITY SPECTROMETRY Fanny C. Liu, a, Mark E. Ridgeway,

More information

Analytical Chemistry in Forensic Science spectral imaging

Analytical Chemistry in Forensic Science spectral imaging Analytical Chemistry in Forensic Science spectral imaging W.T. Chan Department of Chemistry University of Hong Kong June 4, 2015 Plan of presentation Overview of chemical analysis in forensic science Principles

More information

A continuous-wave optical parametric oscillator for mid infrared photoacoustic trace gas detection

A continuous-wave optical parametric oscillator for mid infrared photoacoustic trace gas detection A continuous-wave optical parametric oscillator for mid infrared photoacoustic trace gas detection Frank Müller, Alexander Popp, Frank Kühnemann Institute of Applied Physics, University of Bonn, Wegelerstr.8,

More information

A Survey of Residual Solvents in Various Inks used in Consumer Product Packaging

A Survey of Residual Solvents in Various Inks used in Consumer Product Packaging A Survey of Residual Solvents in Various Inks used in Consumer Product Packaging Application Note Abstract Food manufacturers use various printing inks on their consumer product packaging. These inks are

More information

CHAPTER-V SUMMARY AND CONCLUSIONS

CHAPTER-V SUMMARY AND CONCLUSIONS CHAPTER-V SUMMARY AND CONCLUSIONS SUMMARY AND CONCLUSIONS The present work has been devoted to the differentiation and characterization of inkjet printed documents. All the four primary inks used in printers

More information

Your first NMR measurement

Your first NMR measurement Your first NMR measurement Introduction Select 10mM water in D2O as NMR sample. The NMR spectrum of such sample consists of only two signals: the water signal and the peak of the reference (TSP). Follow

More information

Agilent 4300 Handheld FTIR Spectrometer AT-SITE. IMMEDIATE RESULTS. TRUE NON-DESTRUCTIVE ANALYSIS.

Agilent 4300 Handheld FTIR Spectrometer AT-SITE. IMMEDIATE RESULTS. TRUE NON-DESTRUCTIVE ANALYSIS. Agilent 4300 Handheld FTIR Spectrometer AT-SITE. IMMEDIATE RESULTS. TRUE NON-DESTRUCTIVE ANALYSIS. BRING THE POWER OF FTIR SPECTROSCOPY OUT OF THE LAB AND TO THE SAMPLE From improving composite bonding

More information

Photograph: Norith Eckbo

Photograph: Norith Eckbo Photograph: Norith Eckbo Contaminants in Polar Regions PBDEs Particle deposition (high K OA ) HCHs High volatility: Single hoppers (high K AW ) Condensation DDT PCBs Evaporation Low volatility (low K AW

More information

Thermochromic Solid-State Emission of Dipyridyl Sulfoxide Cu(I) Complexes

Thermochromic Solid-State Emission of Dipyridyl Sulfoxide Cu(I) Complexes Supporting Information Thermochromic Solid-State Emission of Dipyridyl Sulfoxide Cu(I) Complexes Christopher M. Brown, Veronica Carta and Michael O. Wolf* Department of Chemistry, University of British

More information

Principios Básicos de RMN en sólidos destinado a usuarios. Gustavo Monti. Fa.M.A.F. Universidad Nacional de Córdoba Argentina

Principios Básicos de RMN en sólidos destinado a usuarios. Gustavo Monti. Fa.M.A.F. Universidad Nacional de Córdoba Argentina Principios Básicos de RMN en sólidos destinado a usuarios Gustavo Monti Fa.M.A.F. Universidad Nacional de Córdoba Argentina magnet 1 2 4 5 6 computer 3 Block diagrama of a traditional NMR spectrometer.

More information

LC/MS Purification System TRILUTION LC v3.0 User s Guide

LC/MS Purification System TRILUTION LC v3.0 User s Guide LC/MS Purification System TRILUTION LC v3.0 User s Guide Gilson, Inc. PO Box 620027 Middleton, WI 53562-0027, USA Tel: 608-836-1551 or 800-445-7661 Fax: 608-831-4451 Gilson S.A.S. 19, avenue des Entrepreneurs

More information

Electronic Supporting Information. General Experimental Details. Jack Li-Yang Chen and Margaret A. Brimble*

Electronic Supporting Information. General Experimental Details. Jack Li-Yang Chen and Margaret A. Brimble* Electronic Supporting Information Synthesis of the Bis-spiroacetal C 2 C 40 Moiety of the Antimitotic Agent Spirastrellolide B using a Bis-dithiane Deprotection / Spiroacetalisation Sequence Jack Li-Yang

More information

Observing Nightlights from Space with TEMPO James L. Carr 1,Xiong Liu 2, Brian D. Baker 3 and Kelly Chance 2

Observing Nightlights from Space with TEMPO James L. Carr 1,Xiong Liu 2, Brian D. Baker 3 and Kelly Chance 2 Observing Nightlights from Space with TEMPO James L. Carr 1,Xiong Liu 2, Brian D. Baker 3 and Kelly Chance 2 September 27, 2016 1 Carr Astronautics Corp., Greenbelt, MD, USA jcarr@carrastro.com 2 Harvard-Smithsonian

More information

System Parameter and Performance Comparison Between Agilent 7697A and Agilent G1888A Headspace Samplers for USP <467>

System Parameter and Performance Comparison Between Agilent 7697A and Agilent G1888A Headspace Samplers for USP <467> System Parameter and Performance Comparison Between Agilent 7697A and Agilent G888A Headspace Samplers for USP Technical Overview Introduction The Agilent 7697A Headspace Sampler introduced onboard

More information

The Best Way to Get the Right Answers... is in a Flash

The Best Way to Get the Right Answers... is in a Flash FlashQuant Workstation The Best Way to Get the Right Answers... is in a Flash FlashQuant Workstation: Integrating MALDI Speed with Triple Quad Selectivity 5.03E+02 0.011 1.40E+03 2.680 5.26E+02 0.011 1.30E+03

More information

Identifying Causes for Certain Types of Electrically Initiated Fires in Residential Circuits

Identifying Causes for Certain Types of Electrically Initiated Fires in Residential Circuits Identifying Causes for Certain Types of Electrically Initiated Fires in Residential Circuits John J. Shea Eaton Electrical 17 Industry Dr. Pittsburgh, PA 15275 johnjshea@eaton.com Some Sources of Potential

More information

Agilent 6400 Series Triple Quadrupole LC/MS System

Agilent 6400 Series Triple Quadrupole LC/MS System Agilent 6400 Series Triple Quadrupole LC/MS System Concepts Guide The Big Picture Agilent Technologies Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by

More information

Optimisation of LCMS Techniques for Complementary Medicines

Optimisation of LCMS Techniques for Complementary Medicines Optimisation of LCMS Techniques for Complementary Medicines Ashley Dowell Southern Cross Plant Science Southern Cross University Complementary Medicines Complementary medicines vitas erals herbal medicines

More information

Agilent 1260 Infinity Variable Wavelength Detector

Agilent 1260 Infinity Variable Wavelength Detector Agilent 1260 Infinity Variable Wavelength Detector Features, Technical Details, Specifications and Ordering Details Highest sensitivity for HPLC and RRLC The Agilent 1260 Infinity Variable Wavelength Detector

More information

An Introduction to Laser Diodes

An Introduction to Laser Diodes TRADEMARK OF INNOVATION An Introduction to Laser Diodes What's a Laser Diode? A laser diode is a semiconductor laser device that is very similar, in both form and operation, to a light-emitting diode (LED).

More information

Supporting Information

Supporting Information Supporting Information for Simple two-step synthesis of 2,4-disubstituted pyrroles and 3,5-disubstituted pyrrole-2-carbonitriles from enones Murat Kucukdisli 1, Dorota Ferenc 1, Marcel Heinz 2, Christine

More information

2-Micron high-repetition rate laser transmitter for coherent DIAL measurements of atmospheric CO2

2-Micron high-repetition rate laser transmitter for coherent DIAL measurements of atmospheric CO2 2-Micron high-repetition rate laser transmitter for coherent DIAL measurements of atmospheric CO2 Fabien Gibert, Dimitri Edouart, Claire Cénac, Florian Le Mounier, Pierre H. Flamant Laboratoire de Météorologie

More information

SWS SWS62221 Spectral Sensor. General Description. Block Diagram

SWS SWS62221 Spectral Sensor. General Description. Block Diagram SWS62221 Spectral Sensor General Description NeoSpectra sensors are the most compact and the lowest cost Fourier Transform InfraRed (FT-IR) spectrometers on the market today. They deliver the same functionality

More information

Installation and User Guide. FlexIR TM NIR Fiber Optic Accessory

Installation and User Guide. FlexIR TM NIR Fiber Optic Accessory Installation and User Guide FlexIR TM NIR Fiber Optic Accessory The information in this publication is provided for reference only. All information contained in this publication is believed to be correct

More information

Infrared wire grid polarizers: metrology, modeling, and laser damage threshold

Infrared wire grid polarizers: metrology, modeling, and laser damage threshold Infrared wire grid polarizers: metrology, modeling, and laser damage threshold Matthew George, Bin Wang, Jonathon Bergquist, Rumyana Petrova, Eric Gardner Moxtek Inc. Calcon 2013 Wire Grid Polarizer (WGP)

More information

Ion optics innovations for increased sensitivity in hybrid MS systems

Ion optics innovations for increased sensitivity in hybrid MS systems Ion optics innovations for increased sensitivity in hybrid MS systems Abstract Sensitivity, mass resolution, scan speed, spectral fidelity, mass accuracy, and mass range are just a few of the parameters

More information

OtO Photonics. Light Source Datasheet. Description. Balance Light Series LS-BA. Halogen Light Series LS-HA

OtO Photonics. Light Source Datasheet. Description. Balance Light Series LS-BA. Halogen Light Series LS-HA OtO Photonics Light Source Datasheet Description OtO Photonics provide many kinds of light sources to meet customer's need. Our light sources offer the stable and wide-band spectral output and proper for

More information

The FTNIR Myths... Misinformation or Truth

The FTNIR Myths... Misinformation or Truth The FTNIR Myths... Misinformation or Truth Recently we have heard from potential customers that they have been told that FTNIR instruments are inferior to dispersive or monochromator based NIR instruments.

More information

Bead Injection. MicroSI Chromatography on Renewable Column FLOWCELL MPV

Bead Injection. MicroSI Chromatography on Renewable Column FLOWCELL MPV Bead Injection BI exploits flow programming to meter, transport, capture, perfuse, monitor and discharge microspheres in order to separate and assay species that can be bound and released from immobilized

More information