Principle of X-Ray Systems

Size: px
Start display at page:

Download "Principle of X-Ray Systems"

Transcription

1 Principle of X-Ray Systems Hossein Ebrahimi Nasab PHYSICS OF X-RAYS Nature of X-rays Energy unit Interaction with matter INTERACTION WITH THE MATTER In vacuum: photon move along a straight line In materials, photon can be - transmitted (no interaction) - absorbed (they disappear, transferring their energy to the material), - scattered: they are deviated, with or without loss of energy X Ray Imaging INTERACTION WITH THE MATTER Absorption increases with the density of the material it is what we want to assess, and also what causes biological damage Transmission is what we can see photographically It is what we use to assess the amount of radiation absorbed Scatter is undesirable It interferes with the measurement of the transmitted radiation High density tissues (bones) result in low transmission Scattered radiation reduces the quality of the image History November 8, 1895 Wilhelm Conrad ROENTGEN, a German physicist, discovers accidentally unknown rays, able to pass through his hand and image the bones and his ring He called them. X-rays He was awarded the first Nobel Prize for physics in 1901 One year later, Elihu Thomson demonstrates use of X-rays for diagnosis of fractures and location of foreign objects and Eddy C. Jerman starts the first school for Radiographers PARTICLES ENERGY (Definition) When an electron is submitted to an electrical field, it is accelerated and gets a speed (an energy) proportional to the applied voltage. This energy is expressed in ev : Electron Volt When this electron reaches the positive electrode, it has a kinetic energy of 1000eV or 1 KeV KeV is a unit of energy used for all kinds of particles PRODUCING AN IMAGE The X-ray image displays the part of the X-ray beam not absorbed by the tissues. The less the absorption, the darker the film. The radiograph displays these differences in tissue density 1

2 Technology The heated filament produces a cloud of free electrons These electrons are attracted and accelerated towards the anode by the electrical field (kv) The flow of electrons is an electrical current (ma) The electrical power (100 kv x 100 ma = 10 kw) is converted into heat (99%) and X-ray power (1%) X-ray Spectrum: Variation with ma At a given energy, the rate of photon production is proportional to the ma 1 ma10 ma50 ma100 ma500 ma1000 ma the shape of the spectrum remains the same when ma is varied The Main Component HV Generator Power supply Exposure Exposure parameters control Image receptor Static (Radiography) Film Dynamic (Fluoroscopy) Image Intensifier + video camera X-ray tube Production of X-rays Collimator Beam size control Static & dynamic Digital detector+ Workstation X-ray Spectrum: Variation with Anode Material it changes the emission efficiency (continuous spectrum) FILTRATION Beam hardening (low energy filtration) Any material in the beam path absorbs more of the lower energy radiation (harmful for the patient) than the higher energy Regulation require a minimum of 2.5 mm Aluminum equivalent filtration THE X-RAY TUBE it determines the presence of characteristic radiations and their energy insert Tube casing (housing) X-ray Spectrum: Variation with kv kv changes the emission efficiency it changes the shape of the spectrum : maximum energy of photons in the spectrum (in kev) has the same value as tube HV (in kv) it has a significant effect on the relative amplitude of characteristic radiations Take Away 99% of the electrical power is converted into heat The X-ray beam is multi-energetic (spectrum) Continuous spectrum Specific peaks kv control mainly the maximum energy ma control the intensity of the beam Anode material defines the specific peaks Anode/filter combination enhances specific energies Tube casing (housing) The x-ray tube insert is mounted inside a protective housing. The housing is made of aluminum and lined with lead. The housing: Controls scatter and leakage radiation Isolates high voltage Cools the tube 2

3 X-RAY TUBE COMPONENT: Frame X-RAY TUBE COMPONENT: Anode The outer part of the insert (sometimes called the enclosure or frame) can be made of glass, steel, or copper. The tube is inserted inside a casing and immersed in oil for electrical insulation and cooling The casing also shields x-rays emitted in all directions The enclosure: Provides a structural base for the cathode and anode Provides high-voltage insulation Allows a vacuum to be maintained X-rays are produced by energy conversion when electrons strike the anode. The anode consists of 3 major components: A target, which receives a stream of electrons from the cathode A rotor of an induction motor, which holds the target and drives its rotation at a rate of approximately 10,000 rpm. This rotation dissipates the heat generated during the production of x-ray A bearing assembly, which supports the rotor and target assembly, and provides smooth rotation of the target Insert Compeonent Tube insert consist of the: Frame Cathod Anode X-RAY TUBE COMPONENT: Cathode(1) The cathode provides the electron stream for x-ray production. The cathode consists of: Filaments: which are coils of tungsten wire that, when heated, provide the electron stream for x-ray production Focusing cup: which surrounds the filament and concentrates, or focuses, the electron beam Rotore X-RAY TUBE COMPONENT: Cathode(2) Dual Focus Cup Some focusing cups have two filaments. X-ray Generation Filament and Cathode Anode Electrons e - beam X-Rays beam The large filament: used for higher power levels needed for dense anatomy. (High ma, high kv, or both) The small filament: used for lower dosage levels, which provides better resolution. The process by which the x-ray tube produces x-rays is very inefficient. Only one percent of the energy produced is converted to x-rays; the rest is dissipated as heat. 3

4 Focal Spot Characteristics Actual Focal Spot: The electron beam from the tube filament lands on the angled target surface and creates the focal spot. This area (rectangular area) called the actual focal spot. انتخاب ظرفيت حرارتي مناسب ظرفيت حرارتي 300KHU توان دفع حرارتي 40KHU/min ظرفيت حرارتي بدنه تيوب 1250KHU در صورت عدم دفع حرارت توسط تيوب در صورتي كه تابش ها تركيبي از نماهاي فوق باشد با متوسط 3000HU براي هر تابش مي توان 100 تابش داشت The actual focal spot is the area on the anode target where electrons strike, and where x-rays and heat are produced با توان دفع حرارتي 40KHU/min مي توان 4 عكس مهره كمري در دقيقه داشت Focal Spot Characteristics Effective Focal Spot: The area from which the x-ray beam appears to come. Viewed from the perspective of the central ray. Anode Heat Capacity Energy= Kv*mA*time Take Away An X-ray tube is characterized by: Focal spot size (single or dual) The smaller the focus, the better the spatial resolution Anode rotation speed Maximum heat storage capacity (khu or kj) The x-ray tube's effective focal spot is commonly called the focal spot. Focal Spot Characteristics Focal Spot Size and Shape: Length of the tube filament Diameter of the filament wire coil Width and length of the focusing cup slot Depth of the focusing cup Focal Spot Size and Target Angle: as the target angle increases, the effective focal spot size increases. The slope of the anode target allows a larger area to be heated while keeping the apparent area from which x-rays are produced as small as possible. مقدار انرژي حرارتي ايجاد شده در يك تابش اندام KV mas HU دست زانو قفسه سينه مهره كمري رخ شكم (چاق) مهره كمري جانبي The Generator The main functions of the x-ray generator are: Control kv applied to the x-ray tube, thereby controlling x-ray beam penetration Control the amount of ma in the tube, thereby controlling the quantity of x-ray beams Control exposure time and switching 4

5 Types Of Generators(1) Single phase generator Three phase generator High frequency generator Battery powered generator Capacitor discharge Types Of Generators(4) High frequency generator: Generator Control Console The control console variables include: kv ma Exposure time Types Of Generators(2) Single phase generator: Types Of Generators(5) Battery powered generator: Generator Parameters Importance of High KV & ma KV Penetration Good For Thick Patient ma Resolution High Contrast Image Power of Generator = Kilo Watt (KW) = KV x ma / 1000 mas = ma x Time (ma x sec) Types Of Generators(3) Three phase generator: Types Of Generators(6) Capacitor discharge: Tube Exposure Radiation Power Voltage Current Time Amount Unit kw kv ma sec mas Clinical Meaning Shorter Time No Penetration Contrast avoids motion blur Accumulated Dose Amount Correlation W = VxA mas=maxsec 5

6 Generator Classes Middle Power Generators Around 50kW For General Purpose Fluoroscopy is possible High Power Generators Around 80~100kW For long time exposure For high ma Fluoroscopy Low Power Generators Example : Less than 20kW Limited with Radiography General Radiology The collimator: Collimator(1) Controls the dimensions of the x-ray beam Aligns the x-ray beam, the patient, and the bucky prior to exposure Limits patient and technician exposure to radiation Reduces scatter radiation Cassette Tray The cassette tray is a mechanical component of the x-ray system. Purpose : Holds the cassette Centers the cassette System Overview Collimator(2) Two pairs of shutters control the beam dimensions. Patient Centering A light and mirror are built into the collimator to pre-position the x-ray beam. The simulation light allows the technologist to pre-center the beam on anatomical landmarks. Notice the cross hair on the patient's shoulder. Bucky A bucky provides the housing for three assemblies: Cassette tray Grid Automatic exposure control A bucky can be in a horizontal (table) or vertical (chest stand) orientation. 6

7 Automatic Exposure Controller (A.E.C) Automatic Exposure Control measures the dose of radiation that strikes the X-ray film behind the patient, and turns the X-ray system off when the predetermined dose for that screen-film combination has been reached. This assures that only the smallest required dose is administered. The resulting images all show a uniform blackening, and the danger is reduced that the X-ray examination might have to be repeated owing to an error in the image. In this way, automatic exposure timing also indirectly reduces the dose. Grid Design A radiographic grid is composed of strips of lead, separated by aluminum, a material that is relatively transparent to x- rays. The graphic shows how grid removes scatter radiation. Because the scatter radiation, represented by dotted lines, moves in various directions, it is largely absorbed by the lead strips. The transmitted x-rays pass untouched through the grid and reach the film. A E C In 1913, Gustav Bucky introduced the radiographic grid, a device placed between the patient and the cassette. The purpose of the grid is to: Reduce the amount of scatter radiation reaching the film Increase contrast in the x-ray image Scatter radiation is a noise factor. It causes a fogging which impairs radiographic quality, lessening image contrast. Scatter provides no useful information. Grid Grid Pattern(1) X-ray grids are available with either focused or parallel lead strips. Focused Grid: The lead strips of this grid are angled. A line drawn through each lead strip intersects at a point a specified distance from the grid, as shown here. Two important characteristics of the focused grid are: The distance from the grid to the point of intersection is called the focal distance of the Grid The density of the x-ray image is uniform from edge to edge, provided the grid is properly centered and leveled relative to the THE RADIOGRAPHIC FILM CASSETTE: The X-Ray Film GRID TYPES GRID SPECIFACTIONS Grid Pattern(2) Parallel Grid When the lead strips are not angled, they are all perpendicular to the face of the grid. This type of grid is termed parallel. Parallel grids have limited use because: The uniform density of an x-ray exposure is limited to a width of about 10 cm The advantage of a parallel grid is: significant only for an extremely long focal distance Latent Image Formation When the radiation interacts with the silver halide crystals in the film emulsion, the image on the film is produced. The image, which is not visible before processing, is called the latent image. An example of another type of latent image is fingerprints. If you touch an item, you leave your fingerprints even though you cannot see them on that item. When that item is treated, your fingerprints become visible. 7

8 Film Speed Film speed refers to the amount of radiation required to produce a radiograph of standard density (darkness). Film speed is determined by the following factors: The size of the silver halide crystals The thickness of the emulsion The presence of special radiosensitive dyes (Cont d) X-Ray Film Processing Processing is a series of steps that changes the latent image on the exposed film into a radiograph by producing a visible image on the film. Proper processing is just as important as exposure technique in producing diagnostic-quality radiographs. Radiographs that are nondiagnostic because of poor processing techniques must be retaken, exposing the patient to unnecessary radiation. Rinsing Rinsing of the films is necessary to remove the developer from the film so that the development process stops. Usually, agitating the film rack for 20 seconds is sufficient. This must be done under safelight conditions. Film Speed (Cont d) The film speed determines how much exposure time is required to produce the image on the film. A fast film requires less radiation; the film responds more quickly because the silver halide crystals in the emulsion are larger. The larger the crystals, the faster the film speed. This is the same principle as film speed on photographic film. F-speed film, the newest and fastest film on the market today, reduces radiation exposure to the patient by 20% to 60% compared with E-speed and D-speed film. The Five Steps in Processing Development Rinsing Fixation Washing Drying Fixing The acidic fixing solution removes the unexposed silver halide crystals from the film emulsion. The fixer also hardens the film emulsion during this process. For permanent fixation, the film is kept in the fixer for a minimum of 10 minutes. However, films may be removed from the fixing solution after 3 minutes for viewing. Films that are not properly fixed will fade and turn brown in a short time. Leaving films in the fixer for a long time (e.g., over a weekend) can remove the image from the film. Intensifying Screen An intensifying screen intensifies or increases the effect of the radiation and thus decreases the amount of exposure time needed. The intensifying screen is coated with a material called phosphor that gives off light when struck by x-radiation. The film inside the cassette is sandwiched between the intensifying screens and is affected by both the light produced by the phosphor and the x-radiation. Developing Developing is the first step in processing films. A chemical solution called the developer is used. The purpose of the developer is to chemically reduce the exposed silver halide crystals to black metallic silver. The developer solution also softens the film emulsion during this process. Washing After fixation, a water bath is used to wash the film. The washing step requires about 20 minutes to thoroughly remove all excess chemicals from the emulsion. 8

9 The final step in film processing is the drying of the films. Films may be air-dried at room temperature in a dust-free area or placed in a heated drying cabinet. Films must be completely dried before they can be handled for mounting and viewing. Drying Components of the Automatic Processor The processor housing covers all of the component parts. The film feed slot is for the unwrapped films to be inserted into the automatic processor. The roller film transporter is a system of rollers that rapidly moves the film through the compartments. The developer and fixer compartments holds the solutions.the film is transported directly from the developer into the fixer without a rinsing step. The water compartment holds circulating water. The drying chamber holds heated air and dries the wet film. INFLUENCE OF ENERGY (kv) Image Quality(2) The Automatic Processor Automatic film processing is a fast and simple method used to process x-ray films. Other than opening the film packet, all steps of film processing are handled by the automatic processor. Automatic film processing requires only 4 to 6 minutes for the development, fixing, washing, and drying of a film, whereas manual processing techniques require approximately 1 hour. (Cont d) Image Quality(3) INFLUENCE OF ENERGY (kv) The choice of kv depends on the patient anatomy and the object to be imaged Thick or dense part of the body: High kv, ex. Abdomen, pelvis: 80 to 100 kv Small or flat parts of the body: Low kv, ex. Breast, hand: 30 to 50 kv Then the necessary mas is elected to produce the clinically useful image The Automatic Processor (Cont d) The automatic processor maintains the correct temperature of the solutions and adjusts the processing time. Proper maintenance of the automatic processor reduces the chance of errors during film processing. Image Quality(1) The higher the energy, INFLUENCE OF ENERGY (kv) the higher the transmission ratio, the better the penetration Image Quality(4) INFLUENCE OF FOCAL SPOT SIZE: PENUMBRA Sharp projection from a fine focal spot High energy may be necessary in dense or thick areas (abdomen, pelvis) Very low energies are rarely of use and are harmful to the patient Large focal spot is responsible for blurr, leading to inability to distinguish between small, close objects LOSS OF SPATIAL RESOLUTION 9

10 Image Quality(5) The blurr effect, or Penumbra, can be decreased by: Increasing the Source to Object Distance Placing the image receptor as close as possible to the patient Take Away High energy (kv) = high penetration, low dose Low energy = high contrast Very low energy X-rays are undesirable There is an optimum energy to image each object The focal spot should be as small as possible. but thermal constraints Scattered radiation can be reduced by a grid, but it results in dose increase Mobile Radiology 10

LECTURE 1 The Radiographic Image

LECTURE 1 The Radiographic Image LECTURE 1 The Radiographic Image Prepared by:- KAMARUL AMIN ABDULLAH @ ABU BAKAR UiTM Faculty of Health Sciences Medical Imaging Department 11/23/2011 KAMARUL AMIN (C) 1 Lesson Objectives At the end of

More information

Radiology. Radiograph: Is the image of an object made with use of X- ray instead of light.

Radiology. Radiograph: Is the image of an object made with use of X- ray instead of light. Radiology د. اريج Lec. 3 X Ray Films Radiograph: Is the image of an object made with use of X- ray instead of light. Dental x- ray film: Is a recording media on which image of the object was made by exposing

More information

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage

More information

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,

More information

1-1. GENERAL 1-2. DISCOVERY OF X-RAYS

1-1. GENERAL 1-2. DISCOVERY OF X-RAYS 1-1. GENERAL Radiography is a highly technical field, indispensable to the modern dental practice, but presenting many potential hazards. The dental radiographic specialist must be thoroughly familiar

More information

X-ray Tube and Generator Basic principles and construction

X-ray Tube and Generator Basic principles and construction X-ray Tube and Generator Basic principles and construction Dr Slavik Tabakov - Production of X-rays and Patient Dose OBJECTIVES - X-ray tube construction - Anode - types, efficiency - Classical X-ray generator

More information

Beam-Restricting Devices

Beam-Restricting Devices Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the

More information

X-ray Tube and Generator Basic principles and construction

X-ray Tube and Generator Basic principles and construction X-ray Tube and Generator Basic principles and construction Dr Slavik Tabakov - Production of X-rays OBJECTIVES - X-ray tube construction - Anode - types, efficiency - X-ray tube working characteristics

More information

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. RA110 test 3 Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. An object 35 cm in width is radiographed at 100 cm SID and at a 50 cm SOD. What

More information

Film and processing quality assurance

Film and processing quality assurance Film and processing quality assurance Image Receptors Direct action non screen film Indirect action screen film Digital sensor Direct Action Non Screen Film Usually intra-oral film Non screen film reacts

More information

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS Course Number: RADG 112 Department: Radiography Course Title: Image Production & Eval. Semester: Spring Year: 1997 Objectives/ Unit One: Introduction

More information

PD233: Design of Biomedical Devices and Systems

PD233: Design of Biomedical Devices and Systems PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/

More information

- KiloVoltage. Technique 101: Getting Back to Basics

- KiloVoltage. Technique 101: Getting Back to Basics Why do I need to know technique? Technique 101: Getting Back to Basics Presented by: Thomas G. Sandridge, M.S., M.Ed., R.T.(R) Program Director Northwestern Memorial Hospital School of Radiography Chicago,

More information

Technical data CAMARGUE CS-VH50/300. VARIABLE Height Bucky Table With Ceiling Suspension

Technical data CAMARGUE CS-VH50/300. VARIABLE Height Bucky Table With Ceiling Suspension Technical data VARIABLE Height Bucky Table With Ceiling Suspension Model Variations CAMARGUE FH (Fixed Height) CAMARGUE FH Tomo CAMARGUE FH Ceiling suspension CAMARGUE VH (Variable Height) CAMARGUE VH

More information

X-RAY. Lecture No.4. Image Characteristics:

X-RAY. Lecture No.4. Image Characteristics: Lecture No.4 X-RAY أ.م.د. اسامة مراد ابراهيم Image Characteristics: *Radiographic density: It s the degree of blackness of the film. when a film is exposed by an x-ray beam (or by light in case of screenfilm

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

MXHF-1500RF is controlled by Digital key panel console that displays KV, ma and mas with APR menu programmed.

MXHF-1500RF is controlled by Digital key panel console that displays KV, ma and mas with APR menu programmed. R/F TV X-RAY SYSTEM DIAGNOSTIC RADIOGRAPHIC FLUOROSCOPIC TV SYSTEM MXHF-1500RF SYSTEM OUTLINE Product Data No. 041021-01 MXHF-1500RF is controlled by Digital key panel console that displays KV, ma and

More information

RADIOGRAPHIC EXPOSURE

RADIOGRAPHIC EXPOSURE RADIOGRAPHIC EXPOSURE Receptor Exposure Receptor Exposure the that interacts with the receptor. Computed Radiography ( ) requires a. Direct Digital Radiography (DR) requires a. Exposure Indicators Exposure

More information

KODAK Simulation Film / 7157

KODAK Simulation Film / 7157 TECHNICAL INFORMATION DATA SHEET Copyright, Eastman Kodak Company, 1998 1) Description KODAK Simulation Film / 7157 KODAK Simulation Film / 7157 is a high-speed, wide latitude, orthochromatic medical x-ray

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS Veterinary Science Preparatory Training for the Veterinary Assistant Floron C. Faries, Jr., DVM, MS Radiology Floron C. Faries, Jr., DVM, MS Objectives Determine the appropriate machine settings for making

More information

Ludlum Medical Physics

Ludlum Medical Physics Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to

More information

Seminar 8. Radiology S8 1

Seminar 8. Radiology S8 1 Seminar 8 Radiology Medical imaging. X-ray image formation. Energizing and controlling the X-ray tube. Image detectors. The acquisition of analog and digital images. Digital image processing. Selected

More information

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org INTRODUCTION

More information

Visibility of Detail

Visibility of Detail Visibility of Detail Radiographic Quality Quality radiographic images represents the, and information is for diagnosis. The of the anatomic structures and the accuracy of their ( ) determine the overall

More information

KODAK X-OMAT BT Film / 4530 / XBT

KODAK X-OMAT BT Film / 4530 / XBT TECHNICAL INFORMATION DATA SHEET Copyright, Eastman Kodak Company, 2003 KODAK X-OMAT BT Film / 4530 / XBT 1) Description KODAK X-OMAT BT Film / 4530 is a full-speed, blue-sensitive medical x-ray screen

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. RA202 image production class two Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. What removes excess chemistry from the film prior to it

More information

MILADY. Product Data. Page 1 of 8

MILADY. Product Data. Page 1 of 8 Page 1 of 8 The MILADY Mammographic Unit offers the best quality-to-price ratio to our customers worldwide. The unit advanced technology together with the application of industrial production standards,

More information

X-rays in medical diagnostics

X-rays in medical diagnostics X-rays in medical diagnostics S.Dolanski Babić 2017/18. History W.C.Röntgen (1845-1923) discovered a new type of radiation Nature, Jan. 23. 1896.; Science, Feb.14. 1896. X- rays: Induced the ionization

More information

KODAK INDUSTREX AA400 Film Versatility is the key.

KODAK INDUSTREX AA400 Film Versatility is the key. AIRCRAFT INSPECTION AERO-ENGINES ARCHEOLOGICAL ARTIFACTS ASSEMBLIES CASTINGS CONC ELECTRICAL COMPONENTS FORENSICS FORESTRY MUNITIONS BOMB DISPOSAL NUCLEAR APPLICAT PAINTINGS SCULPTURES PIPELINES TIRES

More information

Exposure System Selection

Exposure System Selection Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others

More information

Digital radiography. bucky table and wall stand as a dual detector or wireless system. Amadeo DR Systems

Digital radiography. bucky table and wall stand as a dual detector or wireless system. Amadeo DR Systems Amadeo DR Systems with dicom PACS DX-R Software X-ray Systems for the Future Digital radiography with Amadeo R-DR including bucky table and wall stand as a dual detector or wireless system High contrast

More information

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA RADT 2913 COMPREHENSIVE REVIEW 1 The CR cassette is backed by aluminum that: A. reflects x-rays B. absorbs x-rays C. captures the image D. transmits x-rays

More information

CR Basics and FAQ. Overview. Historical Perspective

CR Basics and FAQ. Overview. Historical Perspective Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

Maximizing clinical outcomes

Maximizing clinical outcomes Maximizing clinical outcomes Digital Tomosynthesis Dual Energy Subtraction Automated Long Length Imaging Improved image quality at a low dose Xray Xray Patented ISS capture technology promotes high sensitivity

More information

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS Course Number: RADG 212 Department: Radiography Course Title: Equip. Operation & Maint. Semester: Spring Year: 1997 Objectives/ Unit One: The X-ray

More information

KODAK Medical X-ray Film / General Purpose / Blue / 3528 / MXB

KODAK Medical X-ray Film / General Purpose / Blue / 3528 / MXB TECHNICAL INFORMATION DATA SHEET Copyright, Eastman Kodak Company, 1998 KODAK Medical X-ray Film / General Purpose / Blue / 3528 / MXB 1) Description KODAK Medical X-ray Film / General Purpose / Blue /

More information

DENTAL RADIOGRAPHY KAMARUL AMIN BIN ABU BAKAR

DENTAL RADIOGRAPHY KAMARUL AMIN BIN ABU BAKAR DENTAL RADIOGRAPHY KAMARUL AMIN BIN ABDULLAH @ ABU BAKAR Components of the Dental X-Ray Machine Dental x-ray machines may vary somewhat in size and appearance, but all machines will have three primary

More information

Collimation Assessment Using GAFCHROMIC XR-M2

Collimation Assessment Using GAFCHROMIC XR-M2 Collimation Assessment Using GAFCHROMIC XR-M2 I. Introduction A method of collimation assessment for GE Senographe full-field digital mammography (FFDM) systems is described that uses a self-developing

More information

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R Building 3, No.9, Jiuxianqiaodong Road, Chaoyang District, Beijing 100015, P.R. China E-mail:

More information

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013).

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013). CODE: RADT 151 INSTITUTE: Health Science TITLE: Radiographic Exposure DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of radiographic exposure selection and manipulation

More information

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging

More information

Rotating Anode X-ray Tube Assembly

Rotating Anode X-ray Tube Assembly RADII H18X H18Y Datasheet Rotating Anode X-ray Tube Assembly The tube unit is so constructed that an X-ray tube is sealed in a diagnostic type protective tube housing of electric shockproof, radiation

More information

Acquisition, Processing and Display

Acquisition, Processing and Display Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image

More information

Rotating Anode X-Ray Tube Housing Assembly. General Data

Rotating Anode X-Ray Tube Housing Assembly. General Data Display Devices & Components Company TECHNICAL DATA ROTANODE TM E7252X Rotating Anode X-Ray Tube Housing Assembly! High speed rotating anode X-ray tube housing assembly for high energy radiographic and

More information

Rotating Anode X-ray Tube

Rotating Anode X-ray Tube Datasheet RADII KL74-1.0/2.0-125 Rotating Anode X-ray Tube Rotating anode X-ray tube for the purpose of general diagnostic X-ray procedures. Specially processed Tungsten faced molybdenum target of 74 mm

More information

NPTEL NPTEL ONLINE COURSE. NPTEL Online Certification Course (NOC) NPTEL. Theory and Practice of Non Destructive Testing

NPTEL NPTEL ONLINE COURSE. NPTEL Online Certification Course (NOC) NPTEL. Theory and Practice of Non Destructive Testing NPTEL NPTEL ONLINE COURSE NPTEL Online Certification Course (NOC) NPTEL Theory and Practice of Non Destructive Testing Dr. Ranjit Bauri Dept. of Metallurgical & Materials Engineering IIT Madras, Chennai

More information

Photons interaction with matter

Photons interaction with matter ب س م هللا الر ح من الر حیم Photons interaction with matter Ionization Ionization is the process of removing an electron from an electrically neutral atom to produce an ion pair. An ion is an atom or subatomic

More information

10/3/2012. Study Harder

10/3/2012. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

MINIMUM REQUIREMENTS FOR A DARKROOM.

MINIMUM REQUIREMENTS FOR A DARKROOM. 1 THE DARKROOM Neill Serman. Aug, 2000 W + P. Ch. Pages 97 - It is important that the darkroom be designed to make film processing an efficient, precise and standardized procedure. Since processing procedures

More information

KODAK MIN-R S Film / 4906

KODAK MIN-R S Film / 4906 TECHNICAL INFORMATION DATA SHEET Copyright, Eastman Kodak Company, 2002 KODAK MIN-R S Film / 4906 1) Description KODAK MIN-R S Film / 4906 is a medium speed, dual coated, ortho-sensitive medical x-ray

More information

CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85

CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85 PD53-012 p Rotating Anode X-ray tube Assembly 0.3/0.8P32&0.6/1.2P32DK-85 GENERAL The Shimadzu 0.3/0.8P32DK-85 & 0.6/1.2P32DK-85, Rotating Anode X-ray tube assemblies are rated to 150kV and feature a 100mm

More information

Digital Radiography System

Digital Radiography System www.drgem.co.kr GXR-SD SERIES GXR-SD SERIES Digital Radiography System GXR-SD Series www.drgem.co.kr High Performance and Reliability The Optimal System for Digital Solution The GXR-SD Series is a diagnostic

More information

10/26/2015. Study Harder

10/26/2015. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 1 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT RADT 3463 COMPUTERIZED IMAGING Section I: Chapter

More information

Radiographic Testing (RT) [10]

Radiographic Testing (RT) [10] Radiographic Testing (RT) [10] Definition: An NDT method that utilizes x-rays or gamma radiation to detect discontinuities in materials, and to present their images on recording medium. 1> Electromagnetic

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

Corsix R 32 Accu PRODUCT DATA

Corsix R 32 Accu PRODUCT DATA PRODUCT DATA Product data Rev. 1 (April 2011) Page Description and Configurations 3 Technical Features 4 Radiological characteristics 4 Operating characteristics 5 Mechanical characteristics 6 Size and

More information

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory

More information

Contrast. Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subjective. Long Scale (Low Contrast)

Contrast. Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subjective. Long Scale (Low Contrast) Contrast Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subject Subjective Radiographic Long Scale (Low Contrast) Short Scale (High Contrast) Factors affecting

More information

P R E S E N T E D B Y. K A M A R U L A M I N A B D U L L A H Dip. MED. IMG., BSc. MED. IMG. (UiTM)

P R E S E N T E D B Y. K A M A R U L A M I N A B D U L L A H Dip. MED. IMG., BSc. MED. IMG. (UiTM) + - P R E S E N T E D B Y K A M A R U L A M I N A B D U L L A H Dip. MED. IMG., BSc. MED. IMG. (UiTM) 1 I N T R O D U C T I O N : An x-ray generator is a device that Supplies electrical power to x-ray

More information

Mammography: Physics of Imaging

Mammography: Physics of Imaging Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness

More information

SYLLABUS. 1. Identification of Subject:

SYLLABUS. 1. Identification of Subject: SYLLABUS Date/ Revision : 30 January 2017/1 Faculty : Life Sciences Approval : Dean, Faculty of Life Sciences SUBJECT : Biophysics 1. Identification of Subject: Name of Subject : Biophysics Code of Subject

More information

Electromagnetic Waves

Electromagnetic Waves Electromagnetic Waves What is an Electromagnetic Wave? An EM Wave is a disturbance that transfers energy through a field. A field is a area around an object where the object can apply a force on another

More information

FMT18 FLOOR MOUNTED SYSTEM

FMT18 FLOOR MOUNTED SYSTEM mas Time AEC 320 kvp 64 mas 320 ma 320 ma 320 DEN 0.0 mm Cu 17 in X 17 in 72.0 in FMT18 FLOOR MOUNTED SYSTEM with Synchronized Tracking System Overview Clinical Efficiency The FMT18 System was designed

More information

NDE SOLUTIONS RADIOGRAPHY COURSE OUTLINE

NDE SOLUTIONS RADIOGRAPHY COURSE OUTLINE NDE SOLUTIONS RADIOGRAPHY COURSE OUTLINE 80 Hour Course Length 1.0 NDT Qualification and Introduction (3 Hours) 1.1 NDT Introduction 1.2 NDT Qualification and Certification 1.2.1 NAS 410 1.2.2 SNT-TC-1A

More information

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This

More information

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893 Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -

More information

AeroDR X60. Motorized X-ray System

AeroDR X60. Motorized X-ray System AeroDR X60 Motorized X-ray System AeroDR X60 2 MOTORIZED X-RAY SYSTEM The AeroDR X60 is a high quality, cost effective, compact motorized floor mounted X-ray system, built around Konica Minolta s renowned

More information

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging The Physics of Radiation Oncology X-ray Imaging Charles E. Willis, Ph.D. DABR Associate Professor Department of Imaging Physics The University of Texas M.D. Anderson Cancer Center Houston, Texas Overview

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

DIGITAL RADIOGRAPHY ARTIFACTS

DIGITAL RADIOGRAPHY ARTIFACTS IMAGING LAB MPHY 487 DIGITAL RADIOGRAPHY ARTIFACTS Mohammad Esmael Alsulimane B.Sc, M.Sc Medical Physics Lecturer - Physics Department All Rights Reserved: Some information and figures in this presentation

More information

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED

More information

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated

More information

CONTRASTING VIEWS DIGITAL VS CONVENTIONAL RADIOGRAPHY

CONTRASTING VIEWS DIGITAL VS CONVENTIONAL RADIOGRAPHY Vet Times The website for the veterinary profession https://www.vettimes.co.uk CONTRASTING VIEWS DIGITAL VS CONVENTIONAL RADIOGRAPHY Author : PETRA AGTHE Categories : Vets Date : April 7, 2008 PETRA AGTHE

More information

RULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER USE OF X-RAY APPARATUS

RULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER USE OF X-RAY APPARATUS RULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER 0400-20-06 USE OF X-RAY APPARATUS TABLE OF CONTENTS 0400-20-06-.01 Purpose 0400-20-06-.06 Veterinary

More information

Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I

Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I Technical Data TD ROTANODE E7239X E7239FX E7239GX 197 Rotating Anode X-Ray Tube Housing Assembly Rotating anode X-ray tube housing assembly for the purpose of general diagnostic X-ray procedures. Specially

More information

Medical Imaging. Latest generation bucky room. system. radiology ahead

Medical Imaging. Latest generation bucky room. system. radiology ahead Medical Imaging Latest generation bucky room system radiology ahead system it grows efficiency, the productivity is on the up Moviplan ic System is the latest-generation X-ray equipment, combining all

More information

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer :

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer : Spectrophotometer An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer : Spectrophotometer components Excitation sources Deuterium Lamp Tungsten

More information

ROTANODE TM E7252X E7252FX E7252GX Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I Electrical :

ROTANODE TM E7252X E7252FX E7252GX Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I Electrical : Technical Data TD ROTANODE TM E7252X E7252FX E7252GX Rotating Anode X-Ray Tube Housing Assembly High speed rotating anode X-ray tube housing assembly for high energy radiographic and cine-fluoroscopic

More information

MULTIX PRO/TOP VERTIX PRO/TOP VERTIX Solitaire X-ray systems for radiographic rooms that make life easier

MULTIX PRO/TOP VERTIX PRO/TOP VERTIX Solitaire X-ray systems for radiographic rooms that make life easier MULTIX PRO/TOP VERTIX PRO/TOP VERTIX Solitaire X-ray systems for radiographic rooms that make life easier Over 1,000 systems sold! MULTIX PRO/TOP VERTIX PRO/TOP VERTIX Solitaire Faster and safer diagnosis

More information

BCA 9SK. 60 Hz on request Absorbed current on Stationary Vac and Vac in fluoro mode

BCA 9SK. 60 Hz on request Absorbed current on Stationary Vac and Vac in fluoro mode ELECTRICAL CHARACTERISTICS Voltage 230 Vac ±10% monophase standard 105 / 115 / 125 / 220 / 240 Vac ±10% monophase on request Frequency 50 Hz standard 60 Hz on request Absorbed current on Stationary 4.5

More information

Distributed by. Ecotron Anyview Series. The Intelligent C-Arm System

Distributed by. Ecotron Anyview Series. The Intelligent C-Arm System Distributed by Ecotron Anyview Series The Intelligent C-Arm System The Ecotron Anyview Series is a new standard for C-arm imaging Excellent radiographic imaging at an economical price Advanced X-ray Generator

More information

BCA 9RK. 60 Hz on request Absorbed current on Stationary Vac and Vac in fluoro mode

BCA 9RK. 60 Hz on request Absorbed current on Stationary Vac and Vac in fluoro mode ELECTRICAL CHARACTERISTICS RADIOLOGICAL DATA MONOBLOC Voltage 230 Vac ±10% monophase standard 105 / 115 / 125 / 220 / 240 Vac ±10% monophase on request Frequency 50 Hz standard 60 Hz on request Absorbed

More information

Introduction To NDT. BY: Omid HEIDARY

Introduction To NDT. BY: Omid HEIDARY Introduction To NDT BY: Omid HEIDARY NDT Methods Penetrant Testing Magnetic Particle Testing Eddy Current Testing Ultrasonic Testing Radiographic Testing Acoustic Emission Infrared Testing Visual Testing

More information

OTC18 OVERHEAD TUBE CRANE SYSTEM

OTC18 OVERHEAD TUBE CRANE SYSTEM OTC18 OVERHEAD TUBE CRANE SYSTEM System Overview Clinical Performance Versatile and intuitive, the OTC18M System delivers enhanced patient comfort and optimized workflow. Precisely designed to withstand

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

X-Rays and endoscopes

X-Rays and endoscopes X-Rays and endoscopes 1 What are X-rays? X-ray refers to electromagnetic radiation with a wavelength between 0.01nm - 10nm. increasing wavelength visible light ultraviolet x-ray increasing energy X-rays

More information

Conceptual Physics Fundamentals

Conceptual Physics Fundamentals Conceptual Physics Fundamentals Chapter 13: LIGHT WAVES This lecture will help you understand: Electromagnetic Spectrum Transparent and Opaque Materials Color Why the Sky is Blue, Sunsets are Red, and

More information

Examination of Pipe Welds by Image Plate Based Computed Radiography System

Examination of Pipe Welds by Image Plate Based Computed Radiography System Examination of Pipe Welds by Image Plate Based Computed Radiography System Sanjoy Das, M.S.Rana, Benny Sebastian, D. Mukherjee and K.K. Abdulla Atomic Fuels Division Bhabha Atomic Research Centre Mumbai

More information

NOTES ON PRINTING 08/06

NOTES ON PRINTING 08/06 NOTES ON PRINTING 08/06 1-Photographic materials are sensitive to most light sources. Never open your box of paper unless you are working under an appropriate yellow or red safelight. (Color photographic

More information

Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I Type B

Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I Type B Technical Data TD ROTANODE E7239X E7239FX E7239GX 0197 Rotating Anode X-Ray Tube Housing Assembly Rotating anode X-ray tube housing assembly for the purpose of general diagnostic X-ray procedures. Specially

More information

Anatomix RP X-Ray System

Anatomix RP X-Ray System T E C H N I C A L I N F O S H E E T Anatomix RP X-Ray System I. SYSTEM DESCRIPTION Anatomix RP liquid concentrates are formulated for use in all automatic processing machines for all standard time cycles,

More information

INTRODUCTION TO FLEXIBLE BRONCHOSCOPY. Fluoroscopy Synopsis HENRI G COLT MD SECOND EDITION THE BRONCHOSCOPY EDUCATION PROJECT SERIES

INTRODUCTION TO FLEXIBLE BRONCHOSCOPY. Fluoroscopy Synopsis HENRI G COLT MD SECOND EDITION THE BRONCHOSCOPY EDUCATION PROJECT SERIES SECOND EDITION INTRODUCTION TO FLEXIBLE BRONCHOSCOPY Fluoroscopy Synopsis HENRI G COLT MD With contributions from Dr. S. Murgu THE BRONCHOSCOPY EDUCATION PROJECT SERIES FLUOROSCOPY SYNOPSIS The purpose

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information