Development of dual MCP x-ray imager for 40 ~ 200 kev region

Size: px
Start display at page:

Download "Development of dual MCP x-ray imager for 40 ~ 200 kev region"

Transcription

1 Development of dual MCP x-ray imager for 40 ~ 200 kev region National ICF Diagnostics Working Group Meeting - October 6-8, 2015 N. Izumi, G. N. Hall, A. C. Carpenter, F. V. Allen, J. G. Cruz, B. Felker, J. Holder, J. D. Kilkenny, A. Lumbard, R. Montesanti, N. E. Palmer, K. Piston, G. Stone, M. Thao, R. Vern, R. Zacharias, O. L. Landen, R. Tommasini, D. Hargrove, D. K. Bradley and P. M. Bel LLNL-PRES This work was performed under the auspices of the U.S. Department of Energy by Lawrence Livermore National Laboratory under contract DE-AC52-07NA Lawrence Livermore National Security, LLC

2 Summary Using a novel dual MCP, we have demonstrated an imaging detector with 3.2x higher 59 kev (1) Compton radiography requires gated x-ray imager with DQE > 4% (2) Single MCP camera becomes noisy over 30 kev due to depth dependent electron gain of MCP (3) With using dual MCP, it is possible to suppress the noise due to depth dependent gain (4) With using with dual MCP, DQE = 4.5% was achieved (5) The 1 st dual MCP camera AXIS is in its production phase 2 2

3 MOTIVATION We need a gated imager for 40~200 kev x rays 3 3

4 Compton backlight Radiography based on Compton scattering can reveal the DT fuel assembly near bang time Tommasini, et al., PoP 18, 2011 Short pulse laser (ARC) 10 µm Au wire Imploded core Gated x-ray Detector Gating is essential to reduce background from hohlraum Expected contrast Photoelectric dominated High pass filter >40 kev Ideal contrast (OD ~1) for high density DT fuel near maximum compression time Flat cross section allows broadband radiography (40~ 200 kev) 10x-20x more signal than K-alpha High-pass filtration suppress the background from core emission Compton radiography requires gated imager which works well in 40~200 kev region 4 4

5 Goal of the project We need a DQE of 4% to achieve a SNR of 20 to measure fuel R with 5% rms accuracy SNR Single MCP Dual MCP 3% 5% 10% Obtainable accuracy of R measurement % Perfect Radiograph DQE (%) 300µm 5

6 Why do we need a new detector? Conventional MCP cameras becomes noisy in high energy regime 6 6

7 Standard MCP based camera becomes noisy for x ray energy over 30 kev Why does MCP image become noisy over 30keV? (1) Stochastic behavior of avalanche stream Multiplication has event-by-event variability (2) Avalanche starts from few electrons (3) Depth dependent MCP gain High energy x-rays penetrate deep into MCP NEXT slide 7

8 MCP gain depends on the depth of the detection event Depth effect is significant for high energy x rays MFP ( micron) X-ray MFP in MCP glass (4gm/cm3) Detection events on top surface experience high gain Deep events experience small gain x-ray energy (kev) 8 8

9 Due to this depth dependent gain, width of the pulse height response depends on x ray energy Narrow PHD for 5 kev photons Broad PHD for 59 kev photons 9 9

10 What is DQE? High DQE = low NF 10 1

11 Detective Quantum Efficiency (DQE) is a useful metric to evaluate SNR of noisy detectors (1) Quantum efficiency (QE) is a good metric for detectors which have narrow pulse height distribution (e.g. CCD, HPGe) QE detected_events (2) For detectors which have avalanche amplification incident noise, _ photons DQE is the quantity of interest ( e.g. Proportional counter, MCP) N N SNR QE N incident _ photons DQE SNR SNR SNR out in 2 DQE N out incident _ photons 11 1

12 Noise Factor (NF) is another useful metric which represents the relation of QE and DQE (1) Quantum Efficiency QE N N DQE where detected_events incident _ photons (2) Detective Quantum Efficiency QE NF NF Narrower pulse height response 2 cl 1 2 Cluster noise Standard deviation of the pulse height distribution Average of the pulse height distribution Lower NF Higher DQE 12 1

13 To achieve required SNR, we need a detector which has high DQE (small Noise Factor) Comparison of two detectors with the same QE Detected events per resolution element High DQE detector QE : 10% NF : 1 DQE : 10% Low DQE detector QE : 10% NF : 8 DQE : 1.25% Signal-to-Noise Ratio is determined by DQE 13 1

14 CONCEPT We can reduce noise using a dual MCP configuration 14

15 Proposed solution We are increasing DQE using a dual MCP configuration Dual MCP (AXIS) MCP Low gain MCP High gain phosphor FILM electron photon x X-ray The 1 st MCP (800 μm thick) is operated in low gain and works as thick photocathode The 2 nd MCP (460 μm thick) is operated in high gain and works as main amplifier Because of low gain of the 1 st MCP, the output is less sensitive to the depth of the detection high DQE 15 1

16 The dual MCP provides good performance regardless of gain setting Expected performance for 60 kev x-rays (Gain ~1000x) Dual MCP μm Single MCP 460 μm QE 13% 8 % DQE 4 ~ 4.5% Independent to total gain 1.4 % Depends on MCP gain Dynamic range Gate width Expected to be wide because PHD is narrower 0.5 ~ 1ns Limited by electron transit time High-end tail of the PHD saturates first (Events near entrance) 70 ~ 100 ps Limited by electron transit time 16 1

17 EXPERIMENT We measured DQE of MCPs 17 1

18 We measured QE and DQE of the dual MCP configuration camera using 59 kev RI source MCP Low gain MCP High gain Phosphor FOP electron PMT x Am 241 X-ray Charge sensitive amplifier Multi Channel Analyzer 18 1

19 Dual MCP The pulse height response changes as a function of bias voltage given to the MCP1 The maximum DQE = 4.5% was observed when MCP 1 bias = 625V When When Beyond 1 st the MCP 1 st this MCP turned point, is not DQE on, biased, x-rays goes detected all down the because events on are pulse height 1 st coming MCP response start from contributing the is expanding 2 MCP the (depth output dependent Both gain effect) QE and DQE go up 10 4 Pulse height response 1.0 Counts / pc QE and DQE 0.1 QE V2:600V DQE PMT output (pc) Total MCP gain 19

20 Dual MCP We optimized the bias voltage of 1 st MCP DQE = 4.5% was achieved successfully QE and DQE MCP1: 800 μm MCP2: 460 μm QE Requirement DQE 2 nd MCP only *Nominal MCP gain *Total MCP gain was changed by adjusting the bias voltage to MCP1 20 2

21 DQE of the dual MCP set us was measured at several energies using the same method % QE and DQE 10.00% 1.00% Monte Carlo L: 800 um Measured QE Measured DQE 0.10% X-ray energy 21

22 Current status of of AXIS 22

23 AXIS is DIM based x ray framing camera with two 40 mm x 40 mm frames 23 2

24 To use in harsh neutron background, KODAK TMAX 400 film is used for recording Dark Slide Film Pack Clamps 24 2

25 Assemble of MCP head is completed Phosphor HV supply Detector Transformers 9:1 Gain Transformers 25:1 Phosphor Slow Down Box Launch boards Bias Box Lemo 6 pin dc bias connector 50Ω Monitor Trace HV feed thrus Detector MCPs Grid Receiver boards HV contacts Gain MCPs 25 2

26 The AXIS head was assembled and currently being tested on HEX source at NSTec 26

27 The first AXIS image was obtained with using Manson source 27

28 Summary Using a novel dual MCP, we have demonstrated an imaging detector with 3.2x higher 59 kev (1) Compton radiography requires gated x-ray imager with DQE > 4% (2) Single MCP camera becomes noisy over 30 kev due to depth dependent electron gain of MCP (3) With using dual MCP, it is possible to suppress the noise due to depth dependent gain (4) With using with dual MCP, DQE = 4.5% was achieved (5) The 1 st dual MCP camera AXIS is in its production phase 28 2

High-resolution Penumbral Imaging on the NIF

High-resolution Penumbral Imaging on the NIF High-resolution Penumbral Imaging on the NIF October 6, 21 Benjamin Bachmann T. Hilsabeck (GA), J. Field, A. MacPhee, N. Masters, C. Reed (GA), T. Pardini, B. Spears, L. BenedeB, S. Nagel, N. Izumi, V.

More information

Scin%llators for high energy x- ray detec%on

Scin%llators for high energy x- ray detec%on Scin%llators for high energy x- ray detec%on NIF diagnos,cs workshop, Los Alamos Andrew MacPhee LLNL October 7th, 2015 LLNL-PRES-XXXXXX This work was performed under the auspices of the U.S. Department

More information

High collection efficiency MCPs for photon counting detectors

High collection efficiency MCPs for photon counting detectors High collection efficiency MCPs for photon counting detectors D. A. Orlov, * T. Ruardij, S. Duarte Pinto, R. Glazenborg and E. Kernen PHOTONIS Netherlands BV, Dwazziewegen 2, 9301 ZR Roden, The Netherlands

More information

Kirkpatrick-Baez Microscope for NIF

Kirkpatrick-Baez Microscope for NIF Kirkpatrick-Baez Microscope for NIF Diagnostic Workshop, Los Alamos 2015 L. A. Pickworth & the KBO team LLNL-PRES-XXXXXX This work was performed under the auspices of the U.S. Department of Energy by Lawrence

More information

Detectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014

Detectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014 Detectors for microscopy - CCDs, APDs and PMTs Antonia Göhler Nov 2014 Detectors/Sensors in general are devices that detect events or changes in quantities (intensities) and provide a corresponding output,

More information

NIF Neutron Bang Time Detector Development on OMEGA

NIF Neutron Bang Time Detector Development on OMEGA NIF Neutron Bang Time Detector Development on OMEGA 2400 2200 NBT2 scintillator bang time (ps) 2000 1800 1600 1400 1200 rms = 54 ps 1000 1000 1200 1400 1600 1800 2000 2200 2400 V. Yu. Glebov University

More information

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

Radiographic sensitivity improved by optimized high resolution X -ray detector design. DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Radiographic sensitivity improved by optimized high resolution X -ray detector

More information

Lecture 12 OPTICAL DETECTORS

Lecture 12 OPTICAL DETECTORS Lecture 12 OPTICL DETECTOS (eference: Optical Electronics in Modern Communications,. Yariv, Oxford, 1977, Ch. 11.) Photomultiplier Tube (PMT) Highly sensitive detector for light from near infrared ultraviolet

More information

Initial Results from the National Ignition Campaign on NIF

Initial Results from the National Ignition Campaign on NIF Initial Results from the National Ignition Campaign on NIF Presentation to 23 rd IAEA Fusion Energy Conference October 10-16, 2010 Daejeon, Republic of Korea John Lindl for the National Ignition Campaign

More information

Electron Beam Diagnosis Using K-edge Absorp8on of Laser-Compton Photons

Electron Beam Diagnosis Using K-edge Absorp8on of Laser-Compton Photons LLNL-PRES-740689 Electron Beam Diagnosis Using K-edge Absorp8on of Laser-Compton Photons Y. Hwang 1, D. J. Gibson 2, R. A. Marsh 2, T. Tajima 1, C. P. J. Barty 1 1 University of California, Irvine 2 Lawrence

More information

Characterization of 18mm Round and 50mm Square MCP-PMTs

Characterization of 18mm Round and 50mm Square MCP-PMTs Characterization of 18mm Round and 50mm Square MCP-PMTs Paul Hink, Robert Caracciolo, John Martin, Scott Moulzolf, Charlie Tomasetti, Joseph Wright BURLE INDUSTRIES, INC. 3 rd Beaune Conference New Developments

More information

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893 Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -

More information

Photomultiplier & Photodiode User Guide

Photomultiplier & Photodiode User Guide Photomultiplier & Photodiode User Guide This User Manual is intended to provide guidelines for the safe operation of Photek PMT Photomultiplier Tubes and Photodiodes. Please contact Sales or visit: www.photek.co.uk

More information

O.H.W. Siegmund, Experimental Astrophysics Group, Space Sciences Laboratory, 7 Gauss Way, University of California, Berkeley, CA 94720

O.H.W. Siegmund, Experimental Astrophysics Group, Space Sciences Laboratory, 7 Gauss Way, University of California, Berkeley, CA 94720 O.H.W. Siegmund, a Experimental Astrophysics Group, Space Sciences Laboratory, 7 Gauss Way, University of California, Berkeley, CA 94720 Microchannel Plate Development Efforts Microchannel Plates large

More information

Performance of Image Intensifiers in Radiographic Systems

Performance of Image Intensifiers in Radiographic Systems DOE/NV/11718--396 LA-UR-00-211 Performance of Image Intensifiers in Radiographic Systems Stuart A. Baker* a, Nicholas S. P. King b, Wilfred Lewis a, Stephen S. Lutz c, Dane V. Morgan a, Tim Schaefer a,

More information

CCD47-10 NIMO Back Illuminated Compact Pack High Performance CCD Sensor

CCD47-10 NIMO Back Illuminated Compact Pack High Performance CCD Sensor CCD47-10 NIMO Back Illuminated Compact Pack High Performance CCD Sensor FEATURES 1024 by 1024 Nominal (1056 by 1027 Usable Pixels) Image area 13.3 x 13.3mm Back Illuminated format for high quantum efficiency

More information

A Framed Monochromatic X-Ray Microscope for ICF

A Framed Monochromatic X-Ray Microscope for ICF A Framed Monochromatic X-Ray Microscope for ICF The Laser Fusion Experiments Groups from the Laboratory for Laser Energetics (LLE) and the Los Alamos National Laboratory (LANL) have jointly developed an

More information

Setting up digital imaging department!

Setting up digital imaging department! Outline Setting up digital imaging department! From screen/film to digital radiography PACS/Tele radiology Setting up digital department Digital Imaging Napapong Pongnapang, Ph.D. Department of Radiological

More information

Performance of the MCP-PMTs of the TOP counter in the first beam operation of the Belle II experiment

Performance of the MCP-PMTs of the TOP counter in the first beam operation of the Belle II experiment Performance of the MCP-PMTs of the TOP counter in the first beam operation of the Belle II experiment K. Matsuoka (KMI, Nagoya Univ.) on behalf of the Belle II TOP group 5th International Workshop on New

More information

Silicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany.

Silicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany. KETEK GmbH Hofer Str. 3 81737 Munich Germany www.ketek.net info@ketek.net phone +49 89 673 467 70 fax +49 89 673 467 77 Silicon Photomultiplier Evaluation Kit Quick Start Guide Eval Kit Table of Contents

More information

Photons and solid state detection

Photons and solid state detection Photons and solid state detection Photons represent discrete packets ( quanta ) of optical energy Energy is hc/! (h: Planck s constant, c: speed of light,! : wavelength) For solid state detection, photons

More information

Performance of Microchannel Plates Fabricated Using Atomic Layer Deposition

Performance of Microchannel Plates Fabricated Using Atomic Layer Deposition Performance of Microchannel Plates Fabricated Using Atomic Layer Deposition Andrey Elagin on behalf of the LAPPD collaboration Introduction Performance (timing) Conclusions Large Area Picosecond Photo

More information

Acquisition, Processing and Display

Acquisition, Processing and Display Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image

More information

Low Light Level CCD Performance and Issues

Low Light Level CCD Performance and Issues Low Light Level CCD Performance and Issues Nagaraja Bezawada UK Astronomy Technology Centre 04 July 2007 Overview of the Talk Introduction to L3CCD (EM CCD) ULTRASPEC Performance and Issues New L3 CCD

More information

Design and implementation of Dilation X-ray Imager for NIF "DIXI"

Design and implementation of Dilation X-ray Imager for NIF DIXI LLNL-CONF-644165 Design and implementation of Dilation X-ray Imager for NIF "DIXI" M. J. Ayers, S. R. Nagel, B. Felker, P. M. Bell, D. K. Bradley, K. Piston, J. Parker, Z. Lamb, J. D. Kilkenny, T. J. Hilsabeck,

More information

combustion diagnostics

combustion diagnostics 3. Instrumentation t ti for optical combustion diagnostics Equipment for combustion laser diagnostics 1) Laser/Laser system 2) Optics Lenses Polarizer Filters Mirrors Etc. 3) Detector CCD-camera Spectrometer

More information

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring

More information

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Unit thickness. Unit area. σ = NΔX = ΔI / I 0 Unit thickness I 0 ΔI I σ = ΔI I 0 NΔX = ΔI / I 0 NΔX Unit area Δx Average probability of reaction with atom for the incident photons at unit area with the thickness of Delta-X Atom number at unit area

More information

PMT tests at UMD. Vlasios Vasileiou Version st May 2006

PMT tests at UMD. Vlasios Vasileiou Version st May 2006 PMT tests at UMD Vlasios Vasileiou Version 1.0 1st May 2006 Abstract This memo describes the tests performed on three Milagro PMTs in UMD. Initially, pulse-height distributions of the PMT signals were

More information

CHAPTER 11 HPD (Hybrid Photo-Detector)

CHAPTER 11 HPD (Hybrid Photo-Detector) CHAPTER 11 HPD (Hybrid Photo-Detector) HPD (Hybrid Photo-Detector) is a completely new photomultiplier tube that incorporates a semiconductor element in an evacuated electron tube. In HPD operation, photoelectrons

More information

SPM Series Quick Start Experiment Guide Rev.1.0, May 2011

SPM Series Quick Start Experiment Guide Rev.1.0, May 2011 Experiment Guide Rev.1.0, May 2011 This document will assist a new user of SPM detectors to make observations and measurements that will verify that the detector is set-up and functioning correctly. The

More information

Introduction to X-ray Detectors for Synchrotron Radiation Applications

Introduction to X-ray Detectors for Synchrotron Radiation Applications Introduction to X-ray Detectors for Synchrotron Radiation Applications Pablo Fajardo Instrumentation Services and Development Division ESRF, Grenoble EIROforum School on Instrumentation (ESI 2011) Outline

More information

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage

More information

Advances in X-Ray Scintillator Technology Roger D. Durst Bruker AXS Inc.

Advances in X-Ray Scintillator Technology Roger D. Durst Bruker AXS Inc. Advances in X-Ray Scintillator Technology Roger D. Durst Inc. Acknowledgements T. Thorson, Y. Diawara, E. Westbrook, MBC J. Morse, ESRF C. Summers, Georgia Tech/PTCE B. Wagner, Georgia Tech/PTCE V. Valdna,

More information

Photon Count. for Brainies.

Photon Count. for Brainies. Page 1/12 Photon Count ounting for Brainies. 0. Preamble This document gives a general overview on InGaAs/InP, APD-based photon counting at telecom wavelengths. In common language, telecom wavelengths

More information

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26 Today s Outline - January 25, 2018 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today s Outline - January 25, 2018 HW #2 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today

More information

Advances in x-ray framing cameras at NIF to improve quantitative precision in x-ray imaging

Advances in x-ray framing cameras at NIF to improve quantitative precision in x-ray imaging Advances in x-ray framing cameras at NIF to improve quantitative precision in x-ray imaging L. R. Benedetti, 1 J. P. Holder, 1 M. Perkins, 1,2 C. G. Brown, 1 C. S. Anderson, 1 F. V. Allen, 1 R. B. Petre,

More information

MCP-PMT status. Samo Korpar. University of Maribor and Jožef Stefan Institute, Ljubljana Super KEKB - 3st Open Meeting, 7-9 July 2009

MCP-PMT status. Samo Korpar. University of Maribor and Jožef Stefan Institute, Ljubljana Super KEKB - 3st Open Meeting, 7-9 July 2009 , Ljubljana, 7-9 July 2009 Outline: MCP aging waveform readout (MPPC) summary (slide 1) Aging preliminary news from Photonis Old information: Current performance (no Al protection layer): 50% drop of efficiency

More information

PRACTICAL CONSIDERATIONS AND EFFECTS OF METALLIC SCREEN FLUORESCENCE AND BACKSCATTER CONTROL IN GAMMA COMPUTED RADIOGRAPHY

PRACTICAL CONSIDERATIONS AND EFFECTS OF METALLIC SCREEN FLUORESCENCE AND BACKSCATTER CONTROL IN GAMMA COMPUTED RADIOGRAPHY 19 th World Conference on Non-Destructive Testing 2016 PRACTICAL CONSIDERATIONS AND EFFECTS OF METALLIC SCREEN FLUORESCENCE AND BACKSCATTER CONTROL IN GAMMA COMPUTED RADIOGRAPHY Steven MANGO 1 1 Carestream

More information

arxiv:hep-ex/ v1 19 Apr 2002

arxiv:hep-ex/ v1 19 Apr 2002 STUDY OF THE AVALANCHE TO STREAMER TRANSITION IN GLASS RPC EXCITED BY UV LIGHT. arxiv:hep-ex/0204026v1 19 Apr 2002 Ammosov V., Gapienko V.,Kulemzin A., Semak A.,Sviridov Yu.,Zaets V. Institute for High

More information

XRF Instrumentation. Introduction to spectrometer

XRF Instrumentation. Introduction to spectrometer XRF Instrumentation Introduction to spectrometer AMPTEK, INC., Bedford, MA 01730 Ph: +1 781 275 2242 Fax: +1 781 275 3470 sales@amptek.com 1 Instrument Excitation source Sample X-ray tube or radioisotope

More information

Chemistry 985. Some constants: q e 1.602x10 19 Coul, ɛ x10 12 F/m h 6.626x10 34 J-s, c m/s, 1 atm = 760 Torr = 101,325 Pa

Chemistry 985. Some constants: q e 1.602x10 19 Coul, ɛ x10 12 F/m h 6.626x10 34 J-s, c m/s, 1 atm = 760 Torr = 101,325 Pa Chemistry 985 Fall, 2o17 Distributed: Mon., 17 Oct. 17, 8:30AM Exam # 1 OPEN BOOK Due: 17 Oct. 17, 10:00AM Some constants: q e 1.602x10 19 Coul, ɛ 0 8.854x10 12 F/m h 6.626x10 34 J-s, c 299 792 458 m/s,

More information

Abstract. Introduction

Abstract. Introduction Comparison of Electro-Optic Diagnostic Systems* K. G. Hagans and P. G. Sargis Lawrence Livermore National Laboratory Field Test Systems Division, Electronics Engineering Livermore, California 94550 Abstract

More information

Gamex CR 2.0 Program description and operating manual

Gamex CR 2.0 Program description and operating manual Gamex CR 2.0 Program description and operating manual Issue No. : 2.0 Date of Issue : Jan. 2013 Z.U.T. NDT SOFT http://www.ndtsoft.eu Copyright (c) 2013 by Z.U.T. NDT SOFT All Rights Reserved Disclaimer

More information

Performance of High Pixel Density Multi-anode Microchannel Plate Photomultiplier tubes

Performance of High Pixel Density Multi-anode Microchannel Plate Photomultiplier tubes Performance of High Pixel Density Multi-anode Microchannel Plate Photomultiplier tubes Thomas Conneely R&D Engineer, Photek LTD James Milnes, Jon Lapington, Steven Leach 1 page 1 Company overview Founded

More information

Advancement in development of photomultipliers dedicated to new scintillators studies.

Advancement in development of photomultipliers dedicated to new scintillators studies. Advancement in development of photomultipliers dedicated to new scintillators studies. Maciej Kapusta, Pascal Lavoutea, Florence Lherbet, Cyril Moussant, Paul Hink INTRODUCTION AND OUTLINE In the validation

More information

ABSTRACT. Supported by U.S. DoE grant No. DE-FG02-96ER54375

ABSTRACT. Supported by U.S. DoE grant No. DE-FG02-96ER54375 ABSTRACT A CCD imaging system is currently being developed for T e (,t) and bolometric measurements on the Pegasus Toroidal Experiment. Soft X-rays (E

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Advanced Optical Communications Prof. R. K. Shevgaonkar Department of Electrical Engineering Indian Institute of Technology, Bombay

Advanced Optical Communications Prof. R. K. Shevgaonkar Department of Electrical Engineering Indian Institute of Technology, Bombay Advanced Optical Communications Prof. R. K. Shevgaonkar Department of Electrical Engineering Indian Institute of Technology, Bombay Lecture No. # 27 EDFA In the last lecture, we talked about wavelength

More information

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices Moving from film to digital: A study of digital x-ray benefits, challenges and best practices H.U. Pöhler 1 and N. D Ademo 2 DÜRR NDT GmbH & Co. KG, Höpfigheimer Straße 22, Bietigheim-Bissingen, 74321,

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

CCD55-30 Inverted Mode Sensor High Performance CCD Sensor

CCD55-30 Inverted Mode Sensor High Performance CCD Sensor CCD55-3 Inverted Mode Sensor High Performance CCD Sensor FEATURES * 1252 (H) by 1152 (V) Pixel Format * 28 by 26 mm Active Area * Visible Light and X-Ray Sensitive * New Improved Very Low Noise Amplifier

More information

Application of CMOS sensors in radiation detection

Application of CMOS sensors in radiation detection Application of CMOS sensors in radiation detection S. Ashrafi Physics Faculty University of Tabriz 1 CMOS is a technology for making low power integrated circuits. CMOS Complementary Metal Oxide Semiconductor

More information

Solid-State Photomultiplier in CMOS Technology for Gamma-Ray Detection and Imaging Applications

Solid-State Photomultiplier in CMOS Technology for Gamma-Ray Detection and Imaging Applications Solid-State Photomultiplier in CMOS Technology for Gamma-Ray Detection and Imaging Applications Christopher Stapels, Member, IEEE, William G. Lawrence, James Christian, Member, IEEE, Michael R. Squillante,

More information

Development of New Large-Area Photosensors in the USA

Development of New Large-Area Photosensors in the USA Development of New Large-Area Photosensors in the USA @BURLE classical PMTs (separate talk) @UC Davis: (1) ReFerence Flat Panels for mass production (2) Light Amplifiers (flat and spherical) Daniel Ferenc

More information

J. A. Oertel, P-24 T. Archuleta, P-24 C. G. Peterson, P-23

J. A. Oertel, P-24 T. Archuleta, P-24 C. G. Peterson, P-23 . t LA-UR- 5 Title: Author@): Submitted to: Dual Microchannel Plate Module For A Gated Monochromatic XRay Imager. J. A. Oertel, P-24 T. Archuleta, P-24 C. G. Peterson, P-23 I 11th High Tempature Plasma

More information

CCD42-40 NIMO Back Illuminated High Performance CCD Sensor

CCD42-40 NIMO Back Illuminated High Performance CCD Sensor CCD42-40 NIMO Back Illuminated High Performance CCD Sensor FEATURES 2048 by 2048 pixel format 13.5 mm square pixels Image area 27.6 x 27.6 mm Back Illuminated format for high quantum efficiency Full-frame

More information

NM Module Section 2 6 th Edition Christian, Ch. 3

NM Module Section 2 6 th Edition Christian, Ch. 3 NM 4303 Module Section 2 6 th Edition Christian, Ch. 3 Gas Filled Chamber Voltage Gas filled chamber uses Hand held detectors cutie pie Geiger counter Dose calibrators Cutie pie Chamber voltage in Ionization

More information

Engineering Medical Optics BME136/251 Winter 2018

Engineering Medical Optics BME136/251 Winter 2018 Engineering Medical Optics BME136/251 Winter 2018 Monday/Wednesday 2:00-3:20 p.m. Beckman Laser Institute Library, MSTB 214 (lab) *1/17 UPDATE Wednesday, 1/17 Optics and Photonic Devices III: homework

More information

Dr. Jiří A. Mareš Institute of Physics Academy of Sciences of the Czech Republic Prague 6, Cukrovarnicka 10 Czech Republic

Dr. Jiří A. Mareš Institute of Physics Academy of Sciences of the Czech Republic Prague 6, Cukrovarnicka 10 Czech Republic Presentation of Activities Dr. Jiří A. Mareš Institute of Physics Academy of Sciences of the Czech Republic Prague 6, Cukrovarnicka 10 Czech Republic amares@fzu.cz Experimental set-up for scintillation

More information

RADIOGRAPHIC EXPOSURE

RADIOGRAPHIC EXPOSURE RADIOGRAPHIC EXPOSURE Receptor Exposure Receptor Exposure the that interacts with the receptor. Computed Radiography ( ) requires a. Direct Digital Radiography (DR) requires a. Exposure Indicators Exposure

More information

Superconducting single-photon detectors as photon-energy and polarization resolving devices. Roman Sobolewski

Superconducting single-photon detectors as photon-energy and polarization resolving devices. Roman Sobolewski Superconducting single-photon detectors as photon-energy and polarization resolving devices Roman Sobolewski Departments of Electrical and Computing Engineering Physics and Astronomy, Materials Science

More information

Lecture 18: Photodetectors

Lecture 18: Photodetectors Lecture 18: Photodetectors Contents 1 Introduction 1 2 Photodetector principle 2 3 Photoconductor 4 4 Photodiodes 6 4.1 Heterojunction photodiode.................... 8 4.2 Metal-semiconductor photodiode................

More information

Mammography: Physics of Imaging

Mammography: Physics of Imaging Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness

More information

Fully depleted, thick, monolithic CMOS pixels with high quantum efficiency

Fully depleted, thick, monolithic CMOS pixels with high quantum efficiency Fully depleted, thick, monolithic CMOS pixels with high quantum efficiency Andrew Clarke a*, Konstantin Stefanov a, Nicholas Johnston a and Andrew Holland a a Centre for Electronic Imaging, The Open University,

More information

MCPs and MCP based detectors. Raquel Ortega

MCPs and MCP based detectors. Raquel Ortega MCPs and MCP based detectors Raquel Ortega r.ortega@photonisusa.com Outline Introduction MCP characteristics: Gain Detection efficiency Time resolution Spatial resolution Detecting different particles

More information

Simulation of High Resistivity (CMOS) Pixels

Simulation of High Resistivity (CMOS) Pixels Simulation of High Resistivity (CMOS) Pixels Stefan Lauxtermann, Kadri Vural Sensor Creations Inc. AIDA-2020 CMOS Simulation Workshop May 13 th 2016 OUTLINE 1. Definition of High Resistivity Pixel Also

More information

Optical Receivers Theory and Operation

Optical Receivers Theory and Operation Optical Receivers Theory and Operation Photo Detectors Optical receivers convert optical signal (light) to electrical signal (current/voltage) Hence referred O/E Converter Photodetector is the fundamental

More information

High Performance. Image Intensifiers

High Performance. Image Intensifiers High Performance Image Intensifiers Image Intensifier Diodes PROXIFIER and MCP Image Intensifiers MCP-PROXIFIER Features Outstanding gain up to > 10 8 W/W High Quantum Efficiency up to 35 % Excellent Resolution

More information

Pixel hybrid photon detectors

Pixel hybrid photon detectors Pixel hybrid photon detectors for the LHCb-RICH system Ken Wyllie On behalf of the LHCb-RICH group CERN, Geneva, Switzerland 1 Outline of the talk Introduction The LHCb detector The RICH 2 counter Overall

More information

Working in Visible NHMFL

Working in Visible NHMFL Working in Visible Optics @ NHMFL NHMFL Summer School 05-19-2016 Stephen McGill Optical Energy Range Energy of Optical Spectroscopy Range SCM3 Optics Facility Energy Range of Optical Spectroscopy SCM3

More information

10/3/2012. Study Harder

10/3/2012. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

Extension of the MCP-PMT lifetime

Extension of the MCP-PMT lifetime RICH2016 Bled, Slovenia Sep. 6, 2016 Extension of the MCP-PMT lifetime K. Matsuoka (KMI, Nagoya Univ.) S. Hirose, T. Iijima, K. Inami, Y. Kato, K. Kobayashi, Y. Maeda, R. Omori, K. Suzuki (Nagoya Univ.)

More information

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES The current multiplication mechanism offered by dynodes makes photomultiplier tubes ideal for low-light-level measurement. As explained earlier, there

More information

Nano-structured superconducting single-photon detector

Nano-structured superconducting single-photon detector Nano-structured superconducting single-photon detector G. Gol'tsman *a, A. Korneev a,v. Izbenko a, K. Smirnov a, P. Kouminov a, B. Voronov a, A. Verevkin b, J. Zhang b, A. Pearlman b, W. Slysz b, and R.

More information

X-ray Detectors: What are the Needs?

X-ray Detectors: What are the Needs? X-ray Detectors: What are the Needs? Sol M. Gruner Physics Dept. & Cornell High Energy Synchrotron Source (CHESS) Ithaca, NY 14853 smg26@cornell.edu 1 simplified view of the Evolution of Imaging Synchrotron

More information

Timing and cross-talk properties of Burle multi-channel MCP PMTs

Timing and cross-talk properties of Burle multi-channel MCP PMTs Timing and cross-talk properties of Burle multi-channel MCP PMTs Peter Križan University of Ljubljana and J. Stefan Institute RICH07, October 15-20, 2007 Contents Motivation for fast single photon detection

More information

10/26/2015. Study Harder

10/26/2015. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

Infrared Single Shot Diagnostics for the Longitudinal. Profile of the Electron Bunches at FLASH. Disputation

Infrared Single Shot Diagnostics for the Longitudinal. Profile of the Electron Bunches at FLASH. Disputation Infrared Single Shot Diagnostics for the Longitudinal Profile of the Electron Bunches at FLASH Disputation Hossein Delsim-Hashemi Tuesday 22 July 2008 7/23/2008 2/ 35 Introduction m eb c 2 3 2 γ ω = +

More information

NON-AMPLIFIED HIGH SPEED PHOTODETECTOR USER S GUIDE

NON-AMPLIFIED HIGH SPEED PHOTODETECTOR USER S GUIDE NON-AMPLIFIED HIGH SPEED PHOTODETECTOR USER S GUIDE Thank you for purchasing your Non-amplified High Speed Photodetector. This user s guide will help answer any questions you may have regarding the safe

More information

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA RADT 2913 COMPREHENSIVE REVIEW 1 The CR cassette is backed by aluminum that: A. reflects x-rays B. absorbs x-rays C. captures the image D. transmits x-rays

More information

Bandpass Edge Dichroic Notch & More

Bandpass Edge Dichroic Notch & More Edmund Optics BROCHURE Filters COPYRIGHT 217 EDMUND OPTICS, INC. ALL RIGHTS RESERVED 1/17 Bandpass Edge Dichroic Notch & More Contact us for a Stock or Custom Quote Today! USA: +1-856-547-3488 EUROPE:

More information

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,

More information

Simulation studies of a novel, charge sharing, multi-anode MCP detector

Simulation studies of a novel, charge sharing, multi-anode MCP detector Simulation studies of a novel, charge sharing, multi-anode MCP detector Photek LTD E-mail: tom.conneely@photek.co.uk James Milnes Photek LTD E-mail: james.milnes@photek.co.uk Jon Lapington University of

More information

PRELIMINARY. Specifications are at array temperature of -30 C and package ambient temperature of 23 C All values are typical

PRELIMINARY. Specifications are at array temperature of -30 C and package ambient temperature of 23 C All values are typical DAPD NIR 5x5 Array+PCB 1550 Series: Discrete Amplification Photon Detector Array Including Pre-Amplifier Board The DAPDNIR 5x5 Array 1550 series takes advantage of the breakthrough Discrete Amplification

More information

Penumbral imaging with multi-penumbral-apertures and its heuristic reconstruction for nuclear reaction region diagnostics

Penumbral imaging with multi-penumbral-apertures and its heuristic reconstruction for nuclear reaction region diagnostics Journal of Physics: Conference Series Penumbral imaging with multi-penumbral-apertures and its heuristic reconstruction for nuclear reaction region diagnostics To cite this article: Tatsuki Ueda et al

More information

Examination of Pipe Welds by Image Plate Based Computed Radiography System

Examination of Pipe Welds by Image Plate Based Computed Radiography System Examination of Pipe Welds by Image Plate Based Computed Radiography System Sanjoy Das, M.S.Rana, Benny Sebastian, D. Mukherjee and K.K. Abdulla Atomic Fuels Division Bhabha Atomic Research Centre Mumbai

More information

Last class. This class. CCDs Fancy CCDs. Camera specs scmos

Last class. This class. CCDs Fancy CCDs. Camera specs scmos CCDs and scmos Last class CCDs Fancy CCDs This class Camera specs scmos Fancy CCD cameras: -Back thinned -> higher QE -Unexposed chip -> frame transfer -Electron multiplying -> higher SNR -Fancy ADC ->

More information

NON-AMPLIFIED PHOTODETECTOR USER S GUIDE

NON-AMPLIFIED PHOTODETECTOR USER S GUIDE NON-AMPLIFIED PHOTODETECTOR USER S GUIDE Thank you for purchasing your Non-amplified Photodetector. This user s guide will help answer any questions you may have regarding the safe use and optimal operation

More information

High Rep-Rate KrF Laser Development and Intense Pulse Interaction Experiments for IFE*

High Rep-Rate KrF Laser Development and Intense Pulse Interaction Experiments for IFE* High Rep-Rate KrF Laser Development and Intense Pulse Interaction Experiments for IFE* Y. Owadano, E. Takahashi, I. Okuda, I. Matsushima, Y. Matsumoto, S. Kato, E. Miura and H.Yashiro 1), K. Kuwahara 2)

More information

Advances in microchannel plate detectors for UV/visible Astronomy

Advances in microchannel plate detectors for UV/visible Astronomy Advances in microchannel plate detectors for UV/visible Astronomy Dr. O.H.W. Siegmund Space Sciences Laboratory, U.C. Berkeley Advances in:- Photocathodes (GaN, Diamond, GaAs) Microchannel plates (Silicon

More information

Computed Radiography of Resistance Temperature Sensor for Indian PHWR

Computed Radiography of Resistance Temperature Sensor for Indian PHWR National Seminar & Exhibition on Non-Destructive Evaluation, NDE 2014, Pune, December 4-6, 2014 (NDE-India 2014) Vol.20 No.6 (June 2015) - The e-journal of Nondestructive Testing - ISSN 1435-4934 www.ndt.net/?id=17831

More information

FIBER OPTICS. Prof. R.K. Shevgaonkar. Department of Electrical Engineering. Indian Institute of Technology, Bombay. Lecture: 20

FIBER OPTICS. Prof. R.K. Shevgaonkar. Department of Electrical Engineering. Indian Institute of Technology, Bombay. Lecture: 20 FIBER OPTICS Prof. R.K. Shevgaonkar Department of Electrical Engineering Indian Institute of Technology, Bombay Lecture: 20 Photo-Detectors and Detector Noise Fiber Optics, Prof. R.K. Shevgaonkar, Dept.

More information

Electronic Instrumentation for Radiation Detection Systems

Electronic Instrumentation for Radiation Detection Systems Electronic Instrumentation for Radiation Detection Systems January 23, 2018 Joshua W. Cates, Ph.D. and Craig S. Levin, Ph.D. Course Outline Lecture Overview Brief Review of Radiation Detectors Detector

More information

Broadband laser ranging for explosive experiments

Broadband laser ranging for explosive experiments DOE/NV/25946--2828 Broadband laser ranging for explosive experiments B. La Lone, B. Marshall, C. V. Bennett, a P. Younk, b K. Miller, E. Daykin National Security Technologies, LLC Special Technologies

More information

Measurements of MeV Photon Flashes in Petawatt Laser Experiments

Measurements of MeV Photon Flashes in Petawatt Laser Experiments UCRL-JC-131359 PREPRINT Measurements of MeV Photon Flashes in Petawatt Laser Experiments M. J. Moran, C. G. Brown, T. Cowan, S. Hatchett, A. Hunt, M. Key, D.M. Pennington, M. D. Perry, T. Phillips, C.

More information

Visibility of Detail

Visibility of Detail Visibility of Detail Radiographic Quality Quality radiographic images represents the, and information is for diagnosis. The of the anatomic structures and the accuracy of their ( ) determine the overall

More information

High Performance. Image Intensifiers

High Performance. Image Intensifiers High Performance Image Intensifiers Image Intensifier Diodes PROXIFIER and MCP Image Intensifiers MCP-PROXIFIER Features Outstanding gain up to > 10 8 W/W High Quantum Efficiency up to 35 % Excellent Resolution

More information

High Energy Digital Radiography & 3D-CT for Industrial Systems

High Energy Digital Radiography & 3D-CT for Industrial Systems DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France High Energy Digital Radiography & 3D-CT for Industrial Systems Non-Destructive

More information

Digital radiography (DR) post processing techniques for pediatric radiology

Digital radiography (DR) post processing techniques for pediatric radiology Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions

More information